Rosemount Hx338 pH Steam Sterilizable and Autoclavable pH Sensor Quick Start Guide

Rosemount™ Hx338
Steam Sterilizable and Autoclavable pH Sensor
Quick Start Guide
LIQ-QSG-Hx338, Rev M
July 2017

Safety Information

Do not exceed temperature and pressure limitations of 266 °F (130 °C) and 43 psig (600 kPa, 4 bar).
HIGH PRESSURE AND TEMPERATURE HAZARD
Before removing the senosr, reduce the process pressure to 0 psig and cool down the process temperature.
Failure to reduce the pressure and temperature may cause serious injury to personnel.
CORROSIVE SUBSTANCE
Internal electrolyte solution may cause skin or eye irritation.
TOXIC AND CORROSIVE SUBSTANCE
Hydrofluoric acid and the mixture of hydrochloric acid and thiourea is toxic and highly corrosive. Avoid skin contact and wear protective gloves. Use only in a well-ventilated area. Do not inhale fumes. In case of an accident, consult a doctor immediately.
CAUTION!
EQUIPMENT DAMAGE
The wetted sensor materials may not be compatible with process composition and operating conditions. Application compatibility is entirely your responsibility.

Contents

Contents
Chapter 1 Plan ..................................................................................................................................1
1.1 Unpacking ......................................................................................................................................1
1.2 Storage and handling ..................................................................................................................... 1
1.3 Specifications .................................................................................................................................1
Chapter 2 Install ...............................................................................................................................3
2.1 Installing the sensor ....................................................................................................................... 3
Chapter 3 Calibration and maintenance ...........................................................................................9
3.1 Two point buffer calibration ........................................................................................................... 9
3.2 Recommended pH sensor standardization ..................................................................................... 9
Chapter 4 Troubleshooting ............................................................................................................ 11
4.1 Cleaning or replacing the electrodes ............................................................................................ 11
4.2 Accessories .................................................................................................................................. 11
Rosemount Hx338 pH Sensor i
Contents
ii Quick Start Guide
1 Plan

1.1 Unpacking

Electrode preparation
1. Remove electrode from shipping container.
2. Remove the protective boot covering the electrode bulb.
3. Rinse away salt film with clean water and then shake the electrode so that the internal solution fills the bulb, thus removing any air trapped there.
Note
Do not allow lubricant to coat electrode bulb or reference junction. If it does, wipe it clean before installation.

Plan

1.2 Storage and handling

Proper storage and handling instructions of the electrodes must be followed to ensure correct measurements.
1. Store the electrodes in their original shipping containers until needed.
2. Do not store at temperatures below -5 °C (23 °F).
3. Store the electrodes with a protective cap containing KCI solution (PN 9210342).
4. For overnight storage, immerse the sensor in tap water or 4 pH buffer solution.
5. Use the startup and calibration procedure described in Section 3.1 to properly calibrate the sensors.
1.3 Specifications

SpecificationsTable 1-1:

Hx338 pH Sensor Specifications Materials and Units
Operating Temperature Range 32 to 266 °F (0 to 130 °C)
Maximum Pressure 43 psig (600 kPa abs, 4 bar)
pH Range 2 to 12 (0 to 14 short periods)
Wetted Materials Glass, EPDM (FDA approved), Ceramic (Diaphragm)
Process Connections PG 13.5 Threads
Cable Connector Single Pole S7 Connector Plug or VP6 Cable Connection
Minimum Conductivity of Sample 50 µS/cm
Rosemount Hx338 pH Sensor 1
Plan
2 Quick Start Guide
2 Install

2.1 Installing the sensor

Rosemount Hx338 dimensional drawingFigure 2-1:

Install

Rosemount Hx338 pH Sensor 3
Install
A. VP 6.0 plug B.
SW17 free spinning PG 13.5 thread
C. Washer D. O-ring
Procedure
1.
Mount the sensor in a vertical pipe run with flow from top to bottom.
If the sensor must be mounted in a horizontal pipe run, orient the sensor in the or 9 o'clock position.
2. Orient the sensor within 80 ° of vertical. See Figure 2-2.
Sensor OrientationFigure 2-2:
3. Wire the sensor to the transmitter. See wiring diagrams below.
4 Quick Start Guide
Install
Rosemount Hx338-01-72 wiring to Rosemount 1066 TransmitterFigure 2-3:
Rosemount Hx338-01 wiring to Rosemount 1066 TransmitterFigure 2-4:
Rosemount Hx338 pH Sensor 5
Install
Rosemount Hx338-01-72 wiring to Rosemount 5081 TransmitterFigure 2-5:
Rosemount Hx338-01 wiring to Rosemount 5081 TransmitterFigure 2-6:
6 Quick Start Guide
Install
Rosemount Hx338-01-72 wiring to Rosemount 1057 TransmitterFigure 2-7:
Rosemount Hx338-01 wiring to Rosemount 1057 TransmitterFigure 2-8:
Rosemount Hx338 pH Sensor 7
Install
Figure 2-9:
Figure 2-10:
Rosemount Hx338-01-72 wiring to Rosemount 1056 and Rosemount 56 Transmitters
Rosemount Hx338-01 wiring to Rosemount 1056 and Rosemount 56 Transmitters
8 Quick Start Guide

Calibration and maintenance

3 Calibration and maintenance

3.1 Two point buffer calibration

Prerequisites
Select two stable buffer solutions, preferably pH 4.0 and 7.0 (pH buffers other than pH 4.0 and pH 7.0 can be used as long as the pH values are at least two pH units apart).
Note
A pH 7.0 buffer solution reads an mV value of approximately zero, and pH buffers read approximately ±59.1 mV for each pH unit above or below pH 7.0. Check the pH buffer manufacturer specifications for millivolt values at various temperatures, as it may affect the actual value of the buffer solution mV/pH value.
Procedure
1. Immerse sensor in the first buffer solution. Allow sensor to equilibrate to the buffer temperature (to avoid errors due to temperature differences between the buffer solution and sensor temperature) and wait for reading to stabilize.
The value of the buffer can now be acknowledged by the transmitter.
2. Once the first buffer has been acknowledged by the transmitter, rinse the buffer solution off of the sensor with distilled or deionized water.
3. Repeat steps 1 and 2 using the second buffer solution.
The theoretical slope value, according to the Nernst equation for calculating pH, is approximately 59.17 mV/pH. Over time the sensor will age, both in the process and in storage, and this will result in reduced slope values. To ensure accurate readings, it is recommended that the electrode be replaced when the slope value falls below 47 to 49 mV/pH.

3.2 Recommended pH sensor standardization

For maximum accuracy, the sensor can be standardized on-line or with a process grab sample after a buffer calibration has been performed and the sensor has been conditioned to the process. Standardization accounts for the sensor junction potential and other interferences. Standardization will not change the sensor’s slope, but will simply adjust the transmitter’s reading to match that of the known process pH.
Rosemount Hx338 pH Sensor 9
Calibration and maintenance
10 Quick Start Guide
4 Troubleshooting

4.1 Cleaning or replacing the electrodes

Electrodes should respond rapidly. Sluggishness, offsets, and erratic readings are indicators that the electrodes may need cleaning or replacement.
1. Use a mild non-abrasive detergent to clean and remove oil deposits from the electrode.
2. Soak electrodes for 30 to 60 minutes in a 5% hydrochloric acid solution to remove scale deposits.
Temperature effect on life expectancy: If glass electrode life expectancy is 100% at 25 °C (77 °F), then it will be approximately 25% at 80 °C (176 °F), and approximately 5% at 120 °C (248 °F).

Troubleshooting

4.2 Accessories

Part number Description
24281-00 15 ft (4.6 m) cable with mating VP8 connector
24281-01 25 ft (7.6 m) cable with mating VP8 connector
24281-02 2.5 ft (0.8 m) cable with mating VP8 connector
24281-05 4 ft (1.3 m) cable with mating VP8 connector
24281-06 10 ft (3 m) cable with mating VP8 connector
24281-07 20 ft (6.1 m) cable with mating VP8 connector
24281-08 30 ft (9.1 m) cable with mating VP8 connector
23645-06 15 ft (4.6 m) cable with mating VP8 connector with BNC on transmitter end
23645-07 15 ft (4.6 m) cable with mating VP8 connector with bare wires on transmitter end
9160494 3.3 ft (1 m) cable with AS7 connector, bare wire on analyzer end
9160495 16.4 ft (5 m) cable with AS7 connector, bare wire on analyzer end
9160496 32.8 ft (10 m) cable with AS7 connector, bare wire on analyzer end
9160498 16.4 ft (5 m) cable with AS7 connector, BNC on analyzer end
9160478 Insertion assembly, 70 mm insertion, for 120 mm sensor, G 1-1/4 in.
9160483 15 degree weld-in socket, G 1-1/4 in. thread, 44 mm
9260484 Service kit for insertion mounting assembly
9210012 Buffer solution, pH 4.01, 16 oz
9210013 Buffer solution, pH 6.86, 16 oz
9210014 Buffer solution, pH 9.18, 16 oz
Rosemount Hx338 pH Sensor 11
Troubleshooting
12 Quick Start Guide
Troubleshooting
Rosemount Hx338 pH Sensor 13
Troubleshooting
14 Quick Start Guide
Troubleshooting
Rosemount Hx338 pH Sensor 15
www.Emerson.com/RosemountLiquidAnalysis
Emerson Automation Solutions 8200 Market Blvd Chanhassen, MN 55317 Toll Free +1 800 999 9307 F +1 952 949 7001
liquid.csc@emerson.com www.Emerson.com/RosemountLiquidAnalysis
EUROPE
Emerson Automation Solutions Neuhofstrasse 19a P.O. Box 1046 CH-6340 Baar Switzerland T + 41 (0) 41 768 6111 F + 41 (0) 41 768 6300
liquid.csc@emerson.com www.Emerson.com/RosemountLiquidAnalysis
LIQ-QSG-Hx338
Rev M
2017
MIDDLE EAST AND AFRICA
Emerson Automation Solutions Emerson FZE Jebel Ali Free Zone Dubai, United Arab Emirates, P.O. Box 17033 T +971 4 811 8100 F +971 4 886 5465
liquid.csc@emerson.com www.Emerson.com/RosemountLiquidAnalysis
ASIA-PACIFIC
Emerson Automation Solutions 1 Pandan Crescent Singapore 128461 Singapore T +65 777 8211 F +65 777 0947
liquid.csc@emerson.com www.Emerson.com/RosemountLiquidAnalysis
©
2017 Rosemount. All rights reserved.
The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a mark of one of the Emerson Process Management family of companies. All other marks are the property of their respective owners. The contents of this publication are presented for information purposes only, and, while effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, express or implied, regarding the products or services described herein or their use or applicability. All sales are governed by our terms and conditions, which are available on request. We reserve the right to modify or improve the designs or specifications of our products at any time without notice.
Loading...