Rosemount Flame Photometric Detector Module for Models 500 and 700 Gas Chromatographs Rev A Manuals & Guides

Page 1
FLAME PHOTOMETRIC DETECTOR
MODULE
FOR MODEL 500 & 700
GAS CHROMATOGRAPHS
HARDWARE REFERENCE MANUAL
ROSEMOUNT ANALYTICAL
TULLIBODY, SCOTLAND
Part Number : 09902 0066
Revision A MARCH 2010
Page 2
NOTICE
ROSEMOUNT ANALYTICAL. ("RAI") SHALL NOT BE LIABLE FOR TECHNICAL OR EDITORIAL ERRORS IN THIS MANUAL OR OMISSIONS FROM THIS MANUAL. RAI MAKES NO WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE WITH RESPECT TO THIS MANUAL AND, IN NO EVENT, SHALL RAI BE LIABLE FOR ANY SPECIAL OR CONSEQUENTIAL DAMAGES INCLUDING, BUT NOT LIMITED TO, LOSS OF PRODUCTION, LOSS OF PROFITS, ETC.
PRODUCT NAMES USED HEREIN ARE FOR MANUFACTURER OR SUPPLIER IDENTIFICATION ONLY AND MAY BE TRADEMARKS/REGISTERED TRADEMARKS OF THESE COMPANIES.
THE CONTENTS OF THIS PUBLICATION ARE PRESENTED FOR INFORMATIONAL PURPOSES ONLY, AND WHILE EVERY EFFORT HAS BEEN MADE TO ENSURE THEIR ACCURACY, THEY ARE NOT TO BE CONSTRUED AS WARRANTIES OR GUARANTEES, EXPRESSED OR IMPLIED, REGARDING THE PRODUCTS OR SERVICES DESCRIBED HEREIN OR THEIR USE OR APPLICABILITY. WE RESERVE THE RIGHT TO MODIFY OR IMPROVE THE DESIGNS OR SPECIFICATIONS OF SUCH PRODUCTS AT ANY TIME.
RAI DOES NOT ASSUME RESPONSIBILITY FOR THE SELECTION, USE OR MAINTENANCE OF ANY PRODUCT. RESPONSIBILITY FOR PROPER SELECTION, USE AND MAINTENANCE OF ANY RAI PRODUCT REMAINS SOLELY WITH THE PURCHASER AND END-USER.
EMERSON PROCESS MANAGEMENT, ROSEMOUNT ANALYTICAL, THE ROSEMOUNT ANALYTICAL LOGO, AND THE EMERSON PROCESS MANAGEMENT LOGO ARE REGISTERED TRADEMARKS OF EMERSON ELECTRIC INC.
COPYRIGHT ® 2010
ROSEMOUNT ANALYTICAL, TULLIBODY, SCOTLAND
All rights reserved. No part of this work may be reproduced or copied in any form or by any means -graphic, electronic or mechanical - without first receiving the permission of Rosemount Analytical, Tullibody, Scotland.
PREFACE
Page 3
1. LIMITED WARRANTY: Subject to the limitations contained in Section 2 herein and except
as otherwise expressly provided herein, Rosemount Analytical. ("RAI") warrants that the firmware will execute the programming instructions provided by RAI, and that the Goods manufactured or Services provided by RAI will be free from defects in materials or workmanship under normal use and care until the expiration of the applicable warranty period. Goods are warranted for twelve (12) months from the date of initial installation or eighteen (18) months from the date of shipment by RAI, whichever period expires first. Consumables and Services are warranted for a period of 90 days from the date of shipment or completion of the Services. Products purchased by RAI from a third party for resale to Buyer ("Resale Products") shall carry only the warranty extended by the original manufacturer. Buyer agrees that RAI has no liability for Resale Products beyond making a reasonable commercial effort to arrange for procurement and shipping of the Resale Products. If Buyer discovers any warranty defects and notifies RAI thereof in writing during the applicable warranty period, RAI shall, at its option, promptly correct any errors that are found by RAI in the firmware or Services, or repair or replace F.O.B. point of manufacture that portion of the Goods or firmware found by RAI to be defective, or refund the purchase price of the defective portion of the Goods/Services. All replacements or repairs necessitated by inadequate maintenance, normal wear and usage, unsuitable power sources, unsuitable environmental conditions, accident, misuse, improper installation, modification, repair, storage or handling, or any other cause not the fault of RAI are not covered by this limited warranty, and shall be at Buyer's expense. RAI shall not be obligated to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in writing in advance by an authorized RAI representative. All costs of dismantling, reinstallation and freight and the time and expenses of RAI personnel for site travel and diagnosis under this warranty clause shall be borne by Buyer unless accepted in writing by RAI. Goods repaired and parts replaced during the warranty period shall be in warranty for the remainder of the original warranty period or ninety (90) days, whichever is longer. This limited warranty is the only warranty made by RAI and can be amended only in writing signed by an authorized representative of RAI. Except as otherwise expressly provided in the Agreement, THERE ARE NO REPRESENTATIONS OR WARRANTIES OF ANY KIND, EXPRESSED OR IMPLIED, AS TO MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE, OR ANY OTHER MATTER WITH RESPECT TO ANY OF THE GOODS OR SERVICES. It is understood that corrosion or erosion of materials is not covered by our guarantee.
2. LIMITATION OF REMEDY AND LIABILITY: RAI SHALL NOT BE LIABLE FOR
DAMAGES CAUSED BY DELAY IN PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF WARRANTY HEREUNDER SHALL BE LIMITED TO REPAIR, CORRECTION, REPLACEMENT OR REFUND OF PURCHASE PRICE UNDER THE LIMITED WARRANTY CLAUSE IN SECTION 1 HEREIN. IN NO EVENT, REGARDLESS OF THE FORM OF THE CLAIM OR CAUSE OF ACTION (WHETHER BASED IN CONTRACT, INFRINGEMENT, NEGLIGENCE, STRICT LIABILITY, OTHER TORT OR OTHERWISE), SHALL RAIS LIABILITY TO BUYER AND/OR ITS CUSTOMERS EXCEED THE PRICE TO BUYER OF THE SPECIFIC GOODS MANUFACTURED OR SERVICES PROVIDED BY RAI GIVING RISE TO THE CLAIM OR CAUSE OF ACTION. BUYER AGREES THAT IN NO EVENT SHALL RAIS LIABILITY TO BUYER AND/OR ITS CUSTOMERS EXTEND TO INCLUDE INCIDENTAL, CONSEQUENTIAL OR PUNITIVE DAMAGES. THE TERM "CONSEQUENTIAL DAMAGES" SHALL INCLUDE, BUT NOT BE LIMITED TO, LOSS OF ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVENUE AND COST OF CAPITAL.
PREFACE
Page 4
TABLE OF CONTENTS
SECTION DESCRIPTION
1.0 DESCRIPTION
1.1 PURPOSE OF THIS MANUAL
1.2 INTRODUCTION
For details of software function & system requirements, please refer to the MON software manual or GC hardware manual.
1.3 THEORY OF OPERATION
1.3.1 THE ANALYSER DETECTOR
1.4 GLOSSARY
2.0 EQUIPMENT DESCRIPTION
2.1 GENERAL INFORMATION
2.2 CONNECTION TO GC
2.3 ENVIRONMENTAL CONSIDERATIONS
2.4 UTILITIES
2.5 MODEL 500 FPD
2.6 MODEL 700 FPD
2.7 MODEL 700 FPD FRONT ENTRY
2.8 VENTING
3.0 OPERATION
4.0 MAINTENANCE
5.0 FAULT FINDING
APPENDIX A CERTIFICATION APPENDIX B DRAWINGS APPENDIX C DETECTOR MANUAL APPENDIX D PID CONTROLLER MANUAL APPENDIX E SPARE PARTS LIST APPENDIX F TEST PROCEDURES
Page 5
SECTION 1 : DESCRIPTION
1.1 The Purpose of This Manual
The Rosemount Analytical FPD Module Hardware Reference Manual (P/N 09902 0066) is intended as a user's guide to accompany the Rosemount Analytical FPD Module for use with Rosemount Analytical Model 500 & Model 700 Gas Chromatographs.
NOTE: For software operation instructions, see the Rosemount Analytical
MON 2000 Software for Gas Chromatograph User Manual (P/N 3­9000-522).
For operating instructions for the Model 500 or Model 700 Gas Chromatograph, see the appropriate GC User Manual.
This manual provides the following information:
Section 1 Description
A general description of the FPD Module and its components. A brief description of the GC Systems software, user interfaces, and capabilities. Introduction to GC theory of operation and terminology.
Section 2 Equipment Description
Guidelines for sampling system and gas connections. Descriptions of Analyzer subsystems and components. Descriptions of GC Controller subsystems and components.
Section 3 Operation
Instructions for operating the GC System by means of its built-in keypad and liquid crystal display (LCD), if provided.
Section 4 Maintenance
Instructions for regular maintenance and care of the GC System hardware. Instructions for troubleshooting, repair, and service of the GC System hardware.
Section 5 Fault Finding List of boards, valves, and other components suggested as spare parts.
Appendices
Appendices with additional, helpful reference materials and drawings.
Page 6
1.2 Introduction
The Rosemount Analytical FPD Module is a Flame Photometric Detector that is factory engineered to be used in conjunction with either a Model 500 or a Model 700 Gas Chromatograph. The FPD Module can be used as a single detector, to measure low levels of sulphur compounds in natural gas, or as a second detector in conjunction with a Thermal Conductivity Detector (TCD) mounted in the GC where the application requires full analysis of the natural gas plus sulphur compounds. The FPD Module typically consists of three major components, flame cell, the photomultiplier tube, and the electrometer circuit board :
The Flame Cell : Located in the lower enclosure, the flame cell has connections for fuel gas (Hydrogen), HC Free air, sample injection (process gas plus Nitrogen carrier) and an exhaust pipe. It is fitted with an RTD to monitor the temperature when running, and an ignitor to light the fuel gas.
The Photomultiplier Tube : Located in the lower enclosure, the photomultiplier tube is the sensor that measures the light that is emitted from the flame cell during operation using an optical filter to allow only sulphur wavelength light to be seen by the photomultiplier detector. It has one high voltage lead and one signal lead that takes the signal from the detector to the electrometer. The leads are co-axial type cables.
The Electrometer Board : Located in the upper enclosure, the electrometer board amplifies and processes the signal data from the detector, and sends it to the CPU board on the GC (2350A controller on Model 500). It also provides the ignition circuit, controls the relight function, generates the flame out alarm, and operates the Hydrogen shut off valve.
Page 7
1.3 Theory of Operation NOTE: See also section 1.4, the "Glossary" section of this manual, for definitions
of some of the terminology used in the following explanations.
1.3.1 The Analyser Detector
The detection system in a Rosemount Analytical FPD Module is a Flame Photometric Detector. This uses the reactions of Sulphur components in a Hydrogen/Air flame as a source of analytical detection. The source of the FPDs signal is derived from the light produced by an excited molecule created in the flames combustion, that is, a
photochemical process called chemiluminescence.
FIGURE 1.1 : FPD DETECTOR
The analysis begins when a fixed volume of sample is injected into the column by operation of the sample valve. The sample is moved through the column by the continuous flow of carrier gas. As successive components elute from the GC column system, they are burned in the flame cell. An optical filter is fitted between the flame cell and the photomultiplier tube (PMT). This allows only the wavelength of the emission band for Sulphurs, 394 nm, to pass through to the PMT.
A thermocouple is fitted to the flame cell to ensure that the flame is present. If the flame is not detected, the Electrometer shuts off the Hydrogen to the flame cell. It then supplies a voltage to the igniter, waits 5 seconds and opens the Hydrogen shut off valve. It then repeats this process for a maximum of ten times until the flame is lit. If it does not succeed, then the Hydrogen is shut off, an alarm is set on the GC controller and the unit awaits attention from the operator.
Page 8
FIGURE 1.2 : ELUTION OF COMPONENTS
1. Carrier gas only at the detector
2. First component begins to elute from the columns and is sensed by the detector.
3. Peak concentration of first component.
4. The second component begins to elute from the columns and is sensed by the detector.
5. Peak concentration of the second component.
The signal is sent from PMT to the Electrometer to be amplified. The Electrometer also provides the PMT with the high voltage it requires to operate and the auto relight circuits.
The signal is then sent to the preamplifier board for further amplification. In addition the preamplifier converts the voltage signal to a 4 to 20 milliamp (mA) current loop for transmission to the GC controller. The signal is proportional to the concentration of a component detected in the gas sample. The preamplifier provides four different gain channels as well as compensation for baseline drift. The signals are sent to the GC Controller for computation, recording on a printer, or viewing on a PC monitor or LCD.
In the quiescent condition (prior to injecting a sample), the detector is exposed to pure carrier gas. In this condition, the output from the detector is electrically nulled. The detector output is set to approximately 1 mV DC. This is measured on the red and black terminals on the preamplifier board, and adjusted using the potentiometer (R38) on the electrometer PCB.
Page 9
1.4 GLOSSARY
Auto Zero: Automatic zeroing of the preamplifier. May be entered into the Controller to
take place at any time during the analysis when either the component is not eluting or the baseline is steady (not normally used).
Chromatogram: A permanent record of the detector output. A chromatograph is
obtained from a PC interfaced with the detector output through the GC Controller. A typical chromatogram displays all component peaks, and gain changes. It may be viewed in colour as it is processed on a PC VGA display. Tick marks recorded on the chromatogram by the GC Controller indicate where timed events take place.
Component: Any one of several different gases that may appear in a sample mixture.
For example, sample gas usually contains the following components: ethyl mercaptan, t­butyl mercaptan, methyl ethyl sulphide, diethyl sulphide, hydrogen sulphide and carbonyl sulphide.
Response Factor: Correction factor for each component as determined by the
calibration. It is defined by the equation:
or =
ARFn = Area response factor for component “n” in area per mole percent (%)
HRFn = Height response factor for component “n”
Arean = Area associated with component “n” in calibration gas
Htn = Height associated with component “n” in mole % in calibration gas
Caln = Amount of component “n” in mole % in calibration gas
Retention time : The time in seconds that elapses between the start of analysis ( 0 seconds) and the sensing of the maximum concentration of each component by the analyser detector.
Page 10
SECTION 2 : DESCRIPTION OF EQUIPMENT
2.1 General Information
There are three different versions of the FPD module : Model 500 FPD ; Model 700 FPD ; Model 700 Front Entry FPD. All versions are covered by the same ATEX certificate Sira06ATEX1174 (see Appendix A). The differences between the versions will be detailed in later sections of this manual.
FIG 2.1 : 500 FPD FIG 2.2 : 700 FPD FIG 2.3 : 700 FRONT ENTRY
2.2 Gas connections.
For all FPD analysers used to measure low range sulphur components, consideration should be given to the use of sulphur inert or equivalent tubing for all calibration gas & process gas connections. All internal process pipework, columns etc are silcosteel by design. If 316 or other stainless steel piping is used, the sulphur components will adhere to the internal surface of the pipe, and will continue to do so until such times are the entire internal surface is coated or “conditioned”. This will result in lower than expected levels of sulphur components reaching the detector for measurement. If the sulphur concentration in the line decreases, sulphur components will detach from the wall of the tubing, resulting in an artificially high reading at the detector. Conditioning may take one week or longer, depending on the levels of sulphur components and the length of the pipe runs.
2.3 Environmental Considerations
All Rosemount Analytical FPD detectors are sensitive to changes in temperature and pressure. It is recommended that FPD analysers are located in shelters that have stable temperature & pressure. The use of positive pressurisation for shelters is not recommended.
Page 11
2.4 Utilities
GAS
SUPPLY PRESSURE
TYPICAL FLOWRATE
HYDROGEN
5 BAR
120 CC/MIN
HC FREE AIR
5 BAR
200 CC/MIN
NITROGEN
8 BAR
15 CC/MIN
SAMPLE GAS
3 BAR
100 CC/MIN
GAS
CYLINDER SIZE
RECOMMENDED QTY
HYDROGEN
50 LITRE / 200 BAR
2
HC FREE AIR
50 LITRE / 200 BAR
2
NITROGEN
50 LITRE / 200 BAR
1
Rosemount Analytical FPD analysers require the following utility gases :
Hydrogen 99.995% purity
Hydrocarbon Free Air
Nitrogen 99.995% purity (carrier gas)
Helium 99.995% purity (optional 2nd carrier – application specific)
Calibration gas – application specific
All utility & process gas connections are Swagelok 1/8” double ferrule compression fittings. Metric conversion kits can be supplied on request.
Supply Voltage : Either 230 Vac or 110Vac.
TYPICAL PRESSURE & FLOW RATE INFORMATION
These are typical values supplied for information only. Actual values are application specific.
A water container is necessary for use with the FPD Module to collect the condensed water from the FPD vent. It is not recommended to pipe the vent away unless a continuous downward slope on the pipe and no back pressure or obstruction by water can be guaranteed.
Page 12
2.5 Model 500 FPD
The Model 500 FPD module consists of three Exd GUB enclosures mounted on a frame, plus an Exd solenoid which acts as a Hydrogen shut-off valve. These enclosures contain the following :
1. Electrometer assembly in GUB 5 enclosure.
2. Flame cell and photometric detector tube in GUB 5 enclosure.
3. Transformer (either 230/110Vac or 110/110Vac) in GUB 4 enclosure.
4. Hydrogen shut-off valve.
FIG 2.4 : ELECTROMETER ASSEMBLY FIG 2.5 : FLAME CELL & FP TUBE
FIG 2.6 : H2 SHUT OFF VALVE FIG 2.7 : TRANSFORMER
The FPD module needs to located as close as possible to a Model 500 GC to minimise the length of sample tubing between the two parts, and therefore to keep the cycle time as short as possible.
Page 13
FIG 2.8 : Model 500 FPD
The tubing required to operate the FPD flame cell is 1/16” OD 0.010” ID. All tubing enters the GUB enclosure containing the flame cell via a specially designed tubing gland. All internal fittings are Swagelok double ferrule type compression fittings.
FIG 2.9 : TUBING GLAND
Page 14
2.6 Model 700 FPD
The Model 700 FPD module consists of four Exd GUB enclosures mounted on a frame, plus an Exd solenoid valve which acts as a Hydrogen shut-off valve. The Model 700 FPD requires an additional enclosure to house temperature control equipment that is available internally on a Model 500 GC, but not on a Model 700 GC. These enclosures contain the following :
1. Electrometer assembly in GUB 5 enclosure. (see FIG 2.4)
2. Flame cell and photometric detector tube in GUB 5 enclosure. (see FIG 2.5)
3. PID Temperature controller & relay.
4. Transformer (either 230/110Vac or 110/110Vac) in GUB 4 enclosure. (see FIG 2.7)
5. Hydrogen shut-off valve. (see FIG 2.6)
FIG 2.10 : PID TEMPERATURE CONTROLLER & RELAY
The FPD module needs to be located as close as possible to a Model 700 GC to minimise the length of sample tubing between the two parts, and therefore to keep the cycle time as short as possible. (see FIG 2.2)
The tubing required to operate the FPD flame cell is 1/16” OD 0.010” ID. All tubing enters the GUB enclosure containing the flame cell via a specially designed tubing gland. All internal fittings are Swagelok double ferrule type compression fittings.
Page 15
2.7 Model 700 FPD Front Entry
The Model 700 FPD Front Entry is comprised of the same components as the standard 700 FPD, but an additional frame has been added to allow all the enclosures to be mounted on the front of the unit. This allows the unit to be located close to a wall, as no rear access is required for installation or maintenance. (see FIG 2.3)
2.8 Venting.
All Rosemount Analytical FPD modules have a vent from the flame cell that exits the GUB enclosure via a proprietary Exd breather/drain/flame arrestor assembly. The exhaust from the flame cell emits water vapour as a result of burning Hydrogen as fuel. This vapour condenses in the exhaust tubing outside the GUB enclosure, and can be seen as drips of water.
The FPD exhaust must be allowed to vent to atmosphere. It must not be subjected to any back pressure as this will have a detrimental effect on the detector, and may cause the flame to extinguish.
For information on Model 500 GC & 700 GC venting, please consult the appropriate hardware manual.
Page 16
SECTION 3.0 : OPERATION
The Rosemount Analytical FPD module is operated as a separate detector. It is controlled by, and reports to the GC controller. The flow rates for the utility gases and the carrier gas
are factory set, and are specific to each individual detector. These should only be adjusted
by fully trained and authorised personnel.
The FPD module is identified as detector # 1 on MON software. When used in conjunction with a TCD detector, the FPD is detector # 1 and the TCD is detector # 2.
For GC controller and MON software operation, please consult the appropriate manual.
For FPD operation, please refer to Appendix C in this manual.
Page 17
SECTION 4 : MAINTENANCE
4.1 The Rosemount Analytical FPD module is a complex piece of equipment, and needs to be regularly maintained, preferably as part of a planned maintenance regime.
There are only two important operations that need to be conducted routinely :
1. Replace the Flame Cell and Photometric Tube O-rings (18 – 24 months).
2. Lubricate the stem of the Hydrogen shut-off valve (12 months).
For both of these operations, the GC needs to be shut down, and the appropriate permits and permissions gained before commencing.
Maintenance operations should only be carried out by trained & authorised personnel. Failure to maintain the FPD module correctly may cause loss of functionality, and can
result in catastrophic failure.
See APPENDIX B for the correct drawings, and APPENDIX D for the recommended spare parts list.
Page 18
SECTION 5.0 : FAULT FINDING
Fault Symptoms
Possible Cure
When monitoring the baseline in MON, there are no upsets present when the auto re-light circuit fires.
If no voltage, remove coax connector.
If voltage is present check signal coax.
Check High Voltage is present on coax.
Approx. -600VDC
If voltage now present on board, check coax cable.
Check BNC coax connectors are tight.
If there is no voltage or the signal cable is ok, replace electrometer.
If upsets are being seen but there are no peaks when gas is injected.
Vary the H2/Air ratio
Check the 12v GND wiring to the electrometer board. The two GND terminals on connector 2 are not linked on board. If there are three black wires ensure that pins 1 & 4 are connected to the power supply. The other wire is for the flame cell GND.
Check the tubing going into the bottom of the flame cell. Loosen fitting and pull tubing downwards while watching CGM.
If peaks appear then the tubing will need to be cut.
Check to see if there is flow, from the metering valve next to the heater block.
Check the sample is getting to the flame cell
Try replacing the columns one at a time.
Check you are getting carrier through port 1 with valve 2 on and through port 5 with valve 2 off. If not check the vents on the Alcon valve for back pressure.
Fault finding on Rosemount Analytical FPD modules should only be carried out by competent trained personnel.
This section of the manual is not intended to be a definitive list of every fault that can occur on a FPD module. It only details the most common faults.
Page 19
Air & H2 flows are set correctly and the unit fails to stay lit.
Using a digital thermometer connected to the thermocouple wires coming from the bottom of the flame-cell, check that the temperature is 160 ˚C.
Use the OVERRIDE function on the electrometer when ignitor signal can be seen on a chromatogram and observe to see if it stays lit. If the flame stays on, the problem is with the temperature sensor circuit.
Check flame out thermocouple wires.
Ensure no insulation is trapped under screw on terminal strip.
Try pulling the sample tube out when it is attempting to light in case the tube is affecting the fuel mixture.
Ensure the ignitor is connected.
Replace the flame cell and try again.
Ensure that the signal wires are connected to the correct place remember that the White signal wire should be connected to the TC+ of the CON5.
Unit give good size sample peaks, then after a while the peaks are not present but the relight still gives good peaks.
There might be „soot on the sample tube going to the flame-cell. Pull tube down slightly whilst watching the CGM to see if that cures the fault.
Flame cell temperature cannot be controlled.
Check the flame cell thermistor.
The resistance is approx 100KΩ at ambient. Resistance goes down as temperature goes up.
Flame cell temperature is erratic.
Check that the thermistor has not been pushed right through the flame cell.
In later models, the flame cell will be „blanked at end of holes to ensure that this cannot happen. Flame cell temperature is erratic.
Page 20
Check there is enough heat-sink compound fitted around sensors.
Unable to balance the bridge
Check the BNC connectors for the signal in and the high voltage. Ensure that they are tight
Cut off the flame and check the response from the detector on a live CGM.
Try changing the filter.
Restrictor metering valve seems to be restricting the output flow completely.
Apply snoop to the two fittings at the bottom of the metering valve.
Change the metering valve
Peaks are very small or appear to be back to front
Check N2 flow into union at flame-cell.
This should be no less than 15cc/min.
Noisy baseline and/or very big dips on the baseline
Check the air supply, should be no lower that 500psi in the cylinder
Page 21
APPENDIX A
CERTIFICATION
1. ATEX Certificate Sira06ATEX1174
2. EC Declaration of Conformity
Above certificates were correct at the time this manual was issued. Please contact factory for updates.
Page 22
Page 23
Page 24
Page 25
Page 26
Page 27
Page 28
Page 29
APPENDIX B
DRAWINGS
Model 500 FPD Drawings
DUK 7233/013/1 GENERAL ARRANGEMENT : MODEL 500 FPD MODULE
DUK 7233/002/1 GENERAL ARRANGEMENT : MODEL 500 FPD ANALYSER
DUK 7233/039/1 GENERAL ARRANGEMENT : MODEL 500 FPD DUAL ANALYSER
DUK 7233/028/1 POWER WIRING DIAGRAM : 500 FPD C/W AUX STREAM SWITCHING
DUK 7233/029/1 WIRING DIAGRAM : FPD RELIGHT FAILURE ALARM
DUK 7233/030/1 POWER WIRING DIARAM : MODEL 500 FPD
DUK 7233/033/1 INTERCONNECTION DIAGRAM : 500 FPD / 2350A CONTROLLER
DUK 7233/034/1 WIRING DIAGRAM : 500 FPD / 2350A CONTROLLER
DUK 7233/048/1 WIRING DIARGAM : 500 FPD / 2350A – 6 x 6 PORT VALVES
DUK 7233/056/1 POWER DISTRIBUTION : 500 FPD DUAL ANALYSER C/W 2 x TRACE
HEAT
DUK 7233/062/1 WIRING DIAGRAM : 500 FPD / 2350A – 6 x 6 PORT – AUX STREAM
SWITCHING
DUK 7233/031/1 FLOW DIAGRAM : 500 FPD : TBM/THT
DUK 7233/032/1 FLOW DIAGRAM : 500 FPD : SULPHUR
DUK 7233/035/1 FLOW DIAGRAM : 500 FPD : C6+/SULPHUR
DUK 7233/045/1 FLOW DIAGRAM : 500 FPD : C6+/SULPHUR
DUK 7233/046/1 FLOW DIAGRAM : 500 FPD : DUAL RANGE DMS
DUK 7233/047/1 FLOW DIAGRAM : 500 FPD : SULPHUR – 6 x 6 PORT – 20 MINUTE
CYCLE
DUK 7233/049/1 FLOW DIAGRAM : 500 FPD : DUAL RANGE DMS
DUK 7233/050/1 FLOW DIAGRAM : 500 FPD : C6+/SULPHUR
DUK 7233/051/1 FLOW DIAGRAM : 500 FPD : C6+/MERCAPTANS
DUK 7233/052/1 FLOW DIAGRAM : 500 FPD : C6+/SULPHUR
DUK 7233/053/1 FLOW DIAGRAM : 500 FPD : C6+/SULPHUR – HIGH C3 & C4 – VALCO 6
PORT INJECTION VALVE
DUK 7233/055/1 FLOW DIAGRAM : 500 FPD : SULPHUR – 6 x 6 PORT - 30 MINUTE CYCLE
DUK 7233/057/1 FLOW DIAGRAM : 500 FPD : TOTAL SULPHUR – 5 x 6 PORT
DUK 7233/063/1 FLOW DIAGRAM : 500 FPD : MM/H2S/COS – 5 x 6 PORT
Page 30
Model 700 FPD Drawings
DUK 7204/074/1 GENERAL ARRANGEMENT : 700 FPD ANALYSER
DUK 7204/100/1 GENERAL ARRANGEMENT : 700 FPD MODULE
DUK 7204/156/1 GENERAL ARRANGEMENT : 700 FPD MODULE FRONT ENTRY
DUK 7204/102/1 INTERNAL CABLE WIRING : 700 FPD ANALYSER
DUK 7204/103/1 POWER WIRING DIAGRAM : 700 FPD ANALYSER
DUK 7204/129/1 FLOW DIAGRAM : DET 1 BFM, DCV. DET 2 BFV. He/N2 CARRIERS
DUK 7204/134/1 FLOW DIAGRAM : DET 1 BFM, DCV. EXTERNAL STR SWITCH
DUK 7204/136/1 FLOW DIAGRAM : DET 1 BFM, SWV
DUK 7204/137/1 FLOW DIAGRAM : DET 1 BFM, DCV. INTERNAL STR SWITCH
DUK 7204/138/1 FLOW DIAGRAM : DET 1 BFM. DET 2 BFV, HCV
DUK 7204/159/1 FLOW DIAGRAM : DET 1 BFM, SWV, BFV
DUK 7204/170/1 FLOW DIAGRAM : DET 1 BFM, DCV. DET 2 BFM
Page 31
Page 32
Page 33
Page 34
Page 35
Page 36
Page 37
Page 38
Page 39
Page 40
Page 41
Page 42
Page 43
Page 44
Page 45
Page 46
Page 47
Page 48
Page 49
Page 50
Page 51
Page 52
Page 53
Page 54
Page 55
Page 56
Page 57
Page 58
Page 59
Page 60
Page 61
Page 62
Page 63
Page 64
Page 65
Page 66
Page 67
Page 68
Page 69
Page 70
Page 71
Page 72
Page 73
Page 74
Page 75
Page 76
Page 77
APPENDIX C
DETECTOR MANUAL
Page 78
Page 79
Page 80
Page 81
Page 82
Page 83
Page 84
Page 85
Page 86
Page 87
Page 88
Page 89
Page 90
Page 91
Page 92
Page 93
Page 94
Page 95
Page 96
Page 97
Page 98
Page 99
APPENDIX D
PID CONTROLLER MANUAL
Page 100
Loading...