Rosemount CT5100 Manuals & Guides

Rosemount™ CT5100
Continuous Gas Analyzer
Reference Manual
00809-0100-4511, Rev AA
May 2020
Preface
All possible care has been taken in the preparation of this publication, but Emerson and its agents and distributors accept no liability for any inaccuracies that may be found. This manual reflects the state of the product at the issue date below, but further enhancements while in service may mean that the manual does not reflect your particular system.
Emerson reserves the right to make changes without notice both to this publication and the products which it describes.
Document number: 00809-0100-4511
Rev AA, May 2020
©
Emerson 2020. All rights reserved.
No part of this publication may be reproduced, stored in a retrieval system, or transmitted in any form or by any means electronic, mechanical, photocopying, recording, or otherwise without the express prior written permission of the copyright holder.
If you require additional technical assistance, request help from cascade.support@emerson.com or Emerson distribution partners.
Send general inquiries to cascade.support@emerson.com.
All trademarks used within this document are the property of their respective owners.
Preliminary information
This section details important user information for the Rosemount CT5100 Continuous Gas Analyzer.
Important
Users must read, understand, and comply with the following information before proceeding.
All users, installers, operators, and maintainers must be familiar with operating the analyzer. To install, start up, operate, maintain, and service the analyzer in a safe manner, it is MANDATORY to read all additional documents shipped with the analyzer. The following information is also available and/or referenced in the Rosemount CT5100 Quick Start Guide: 00825-0100-4511.
Save all instructions for future use. Contact your local service center or sales office when missing documents.
Authorized personnel
In-depth specialist knowledge is an absolute requirement for working with and on the analyzer. Personnel installing, operating, servicing, and maintaining the analyzer must be instructed, trained, qualified, and authorized personnel of the operating company for hazardous areas and the manufacturer. It is the responsibility of the operating company to:
Train staff
Observe safety regulations
Follow the safety instructions and procedures in the product manual
Operators must:
Have been trained
Have read and understand all relevant sections of the product manual before commencing work
Know the safety mechanisms and regulations
WARNING
To avoid explosions, loss of life, personal injury, and damage to this equipment and on-site property, do not install, operate, maintain, or service this analyzer before reading and understanding this document and receiving appropriate training.
2
Important instructions
Important
All users must read this page before proceeding!
Emerson designs, manufactures and tests its products to meet many national and international standards. The Rosemount CT5100 is a sophisticated technical product, and to ensure it continues to operate as designed and within normal specifications it MUST be installed, used, and maintained correctly. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Emerson products.
Failure to follow the proper instructions may cause:Loss of life
Personal injury
Damage to property
Damage to this analyzer
Warranty invalidation
Read all instructions prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Emerson representative for clarification.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and maintenance of the product.
Install your equipment as specified in the installation instructions of the appropriate Reference Manual and in accordance with
applicable local and national codes.
Connect all products to the proper electrical and pressure sources.
To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts specified by Emerson.
Unauthorized parts and procedures can affect the product’s performance; place the safe operation of your process at risk, and
VOID YOUR WARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
3
Regulations and standards
Regulations/standards Description
2014/35/EU The Low Voltage Directive
2014/30/EU The Electromagnetic Compatibility Directive
2012/19/EU Waste Electrical and Electronic Equipment (WEEE) Directive
USA 21 CFR 1040.1 Laser products
NEC 505 National Electrical Code (issued by ANSI: American National Standards
BS EN 60825-1:2007 Safety of laser products. Equipment classification and requirements
BS EN 61010-1 2010 IEC 61010-1 2010
BS EN 61326-1: 2013 Electrical equipment for measurement, control, and laboratory use.
Associated publications
Rosemount CT5100 Quick Start Guide
Compliance approvals
Important
The Rosemount CT5100 analyzer is designed for use in Non Hazardous areas ONLY.
This product complies with USA 21 CFR 1040.10. This product is designed and manufactured under an approved quality management
system to ISO 9001: 2015.
Institute and NFPA 70: National Fire Protection Association)
(identical to IEC 608250-1 2007)
Safety requirements for electrical equipment for measurements, control, and laboratory use. General requirements
EMC requirements. General requirements
Waste disposal
Emerson and the Rosemount CT5100 have satisfied the requirements for applying the CE marking to the Rosemount CT5100 Gas Analyzer.
This equipment meets all requirements of the EMC and Low Voltage directives.
Do not dispose of measuring tools into household waste. Only for EC countries: In accordance with European Directive 2012/19/EU for Waste Electrical and
Electronic Equipment and its implementation into national right, measuring tools that are no longer usable must be collected separately and disposed of in an environmentally correct manner.
4
Safety and information notices
DANGER
WILL CAUSE DEATH
Failure to follow this warning will result in death or serious injury to personnel.
WARNING
DANGER TO PERSONNEL
Failure to follow this warning may result in death or serious injury to personnel.
CAUTION
MAY CAUSE DAMAGE TO EQUIPMENT
Failure to follow this warning may result in damage to the equipment.
NOTICE
Important or tip messages will appear in this format.
Safety precautions
Important
The precautions in this manual MUST NOT be changed amended or removed. All authorized users, installation, operation and maintenance personnel, must observe the following safety precautions and warnings. The Rosemount CT5100 analyzer is designed for use in Non Hazardous areas ONLY.
WARNING
ELECTRIC SHOCK
The analyzer operates using mains voltage, which may cause death or serious injury to personnel. Death, personal injury, and/or damage to persons and/or property may result if this is not observed.
Confirm that the circuit breakers are set to OFF and locked out and tagged out before removing the top cover or opening the front cover. The analyzer must be earthed.
Only trained, qualified personnel may install and connect power and signal cables. The installation/connection must be in accordance with all legislative requirements and applicable standards.
Only qualified personnel, familiar with potential risks, should install the analyzer.
DANGER
EXPLOSION HAZARD
Failure to observe this precaution will cause death, personal injury, and/or damage to persons.
The sample gas in the system must be vented to prevent fire or explosion during maintenance and to prevent damage to the analyzer during prolonged shutdowns.
The sample gas in the pipes leading to the analyzer must be purged for a minimum of 2 minutes and 30 seconds at a minimum flow rate of 280 L/m at 1.5 ± 0.5 bar to prevent hazards to personnel during maintenance.
Purge the sample gas in accordance with the safe working procedures for this site.
5
WARNING
FLAMMABLE SUBSTANCES
Some parts of the analyzer may reach temperatures of 374 °F (190 °C) and may present an ignition source. The interior of a analyzer is always hot unless it has been switched off and allowed to cool down. A fire may result if this precaution is not observed.
Exercise care when using oil, paint, cleaning rags, or other flammable substances near the analyzer.
WARNING
TRANSPORTATION HAZARD
The analyzer weighs 117 lb. (53 kg) and should always be lifted and moved using suitable lifting/moving equipment.
Handle the analyzer with caution during unpacking, installation, maintenance, and transport to prevent crushing of hands, feet, or other body parts.
Wear suitable protective gloves and protective footwear. When preparing the analyzer for transport by air, road, or rail, safeguard the analyzer against movement or break-away during transport by securely strapping it in place.
Use safety approved lifting equipment. Ensure that the equipment is tested, meets the lifting ratings for the weight of the equipment, and is in good operational condition.
WARNING
FIRE AND EXPLOSION
Failure to observe this warning could cause an explosion or potentially hazardous situation, which if not avoided, may cause death, personal injury, and/or damage to persons and/or property.
Do not open the analyzer's electrical compartment unless the atmosphere in the area is known to be below the ignitable concentration of combustible gases or materials, or unless all equipment within the protected enclosure is de-energized .
The protective gas supply valve must be kept open unless the atmosphere in the area is known to be below the ignitable concentration of combustible gases or materials or unless all equipment within the protected enclosure is de-energized.
Always lock out the gas handling system when shutting down the analyzer.
After opening the enclosure, do not restore power until the enclosure has been purged for a minimum of 2 minutes and 30 seconds at a minimum flow rate of 280 L/m at 1.5 ± 0.5 bar.
DO NOT operate the analyzer with doors or covers open.
Refer to local regulations as this may require a competent hot work supervisor to issue a hot work permit.
When the analyzer is out of order or if the pressurization unit shuts off due to a failure, all inputs and outputs connected to external equipment MUST be shut off.
This will ensure that no hazardous voltages are present within the analyzer enclosure when not pressurized.
Only properly trained personnel who understand the contents of all applicable manuals and related instructions should start up the analyzer.
Do not keep operating the analyzer if the enclosure shows permanent deformations after performing the overpressure test.
Use only replacement parts and components authorized by Emerson.
All replacement parts and components must be certified and approved for use in hazardous areas.
The analyzer contains a battery for data backup purposes.
Under normal operating conditions, there is no need to replace the battery during the analyzer life time. Battery replacement MUST only be conducted by Rosemount Customer Care personnel. It is NOT a customer serviceable item.
6
WARNING
FIRE, BURN, AND OPTICAL RADIATION EXPOSURE HAZARD
Electrical shock, thermal burns, or loss of vision may occur. Failure to observe this warning could cause an explosion or potentially hazardous situation, which if not avoided, may cause death, personal injury, and/or damage to persons and/or property.
Operators and service personnel do not have access to the laser/electrics or upper cell compartments for general maintenance or service.
WARNING
BURNS
Personal injury and/or damage to property may result if these precautions are not observed. These precautions are particularly important when working at heights.
Some parts of the analyzer may be heated to 374 °F (190°C). To prevent burns, do not touch any of the hot parts. All parts of a analyzer are always hot unless it has been switched off and allowed to cool down.
Before fitting, removing, or performing any maintenance on the analyzer, ensure that it has been switched off and allowed to cool for at least two hours. Before performing any maintenance on, or in the vicinity of, the analysis cell, allow the analyzer to cool for at least 12 hours as the analysis cell is insulated against heat loss.
When handling the analyzer, always wear suitable protective gloves.
If you receive a burn, seek medical treatment immediately.
WARNING
HAZARDOUS SUBSTANCES
Failure to observe this warning could cause a potentially hazardous situation, which if not avoided, may cause death, personal injury, and/or damage to persons and/or property.
The analyzer may contain hazardous substances. Always handle the analyzer assemblies and components with extreme caution. Wear personal protective equipment (PPE) when handling the equipment.
Gas handling components within the analyzer contain particulate matter residue from the sample gases. Over the life of the the analyzer, the concentration of particulate matter will become enriched within the gas handling components. When performing repairs and maintenance on the the analyzer:
Handle used gas handling components with extreme caution.
Avoid direct skin contact with used gas handling components.
Do not smoke, drink, or eat in the work area.
Wear goggles or eye shields.
Wear a suitable face mask to protect against inhalation of particulate matter.
Do not wet fingers, eyes, or any exposed skin.
Pack used gas handling components for disposal in sealed packaging and label them Contaminated.
Dispose of contaminated items as hazardous material in accordance with applicable local, national, or international health
and safety regulations and pollution regulations.
Take special care to ensure that the sample gas return port either returns the sample gas to the product stream or discharges the sample gas to a location that will not cause a hazard.
7
WARNING
OPTICAL RADIATION EXPOSURE HAZARD
There are three types of laser that may be included in the Rosemount CT5100: Quantum Cascade Lasers (QCLs), Interband Cascade Lasers (ICLs), and diode lasers. The lasers within the analyzer are Class 1. The characteristics of the lasers contained within the analyzer are given in the table below.
The emitted laser light is invisible (mid-infrared), and the combined laser powers are sufficiently low at the first accessible aperture that the unprotected eye will not be damaged. This class is eye safe under all operating conditions.
It is, however, possible to cause damage to the eye through not following correct procedures. Do not look at the laser with any kind of magnifier or optical measuring device.
Parameter QCL ICL Diode Comment
Operation mode Pulsed Pulsed Pulsed N/A
Lasers per system 1 - 6 1- 6 1- 6 Maximum of 6 lasers per system
Wavelength 4 - 10 µm 2 - 5 µm Approximately 760 nm N/A
Power < 5 mW < 5 mW < 5 mW Combined power of QCL at first accessible
Pulse duration < 1 µs < 1 µs < 5 µs N/A
Pulse repetition frequency
Duty cycle < 5 % < 5 % < 25 % N/A
The combined power of the QCL, ICL, and diode lasers at the first accessible aperture is < 9.62 mW.
The analyzer has warning labels in appropriate positions according to USA 21 CFR 1040.10. The location of laser safety labels on the analyzer is specified in Safety and system labels and annotation.
The use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous radiation exposure.
< 100 kHz < 100 kHz < 100 kHz N/A
aperture: < 9.62 mW
WARNING
HIGH PRESSURE GAS AND AIR
The calibration gas supply and compressed air supply operate at a pressure that can cause injury, e.g., damage to eyes and skin punctures from debris blown by the high pressure gas or compressed air.
Always lock off or tag out the calibration gas supply and compressed air supply when shutting down the analyzer.
The maximum gas pressure valve must not exceed 100 psig (690 kPa).
WARNING
EXPLOSIONS HAZARD DUE TO ELECTROSTATIC DISCHARGE
In the event of a sudden discharge from electrostatically charged devices or individuals, there is a risk of an explosion. Failure to observe this warning and or follow safety instructions could cause an explosion or potentially hazardous situation, which if not avoided, could result in death or serious injury.
Take suitable measures to ensure that no electrostatic discharge can build up in the explosions risk area.
Clean the device surface by gently wiping it with a damp or antistatic cloth only.
8
WARNING
HAZARD BY WRONG INPUT VOLTAGE
Applying a rated voltage other than specified on the analyzer´s nameplate label may cause an explosion, injury, or damage to the installation. Failure to observe this warning could cause an explosion or potentially hazardous situation, which if not avoided, may cause death, personal injury, and/or damage to persons and/or property.
This type of analyzer is always setup for a specific rated input voltage; see nameplate label.
Ensure the voltage at site of installation meets the rated analyzer input voltage.
WARNING
LOOSE ITEMS
Failure to observe this warning could cause a potentially hazardous situation, which if not avoided, could result in death or serious injury.
Do not place any loose items on top of the system or inside the compartments when doors / covers are open.
Confirm that all loose items, tools, and equipment are removed from compartments before closing doors and covers.
WARNING
MAINTENANCE/MODIFICATIONS
Failure to observe this warning could cause a potentially hazardous situation, which if not avoided, could result in death or serious injury.
On completion of any maintenance and or modifications verify:
All tools and equipment are removed.
No contamination (water/dust) is in the compartments.
Analyzer is wiped clean.
Vents are clear and not obstructed.
Verify that system is in a safe state for operation.
WARNING
Physical access
Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of end users’ equipment. This could be intentional or unintentional and needs to be protected against.
Physical security is an important part of any security program and fundamental to protecting your system. Restrict physical access by unauthorized personnel to protect end users’ assets. This is true for all systems used within the facility.
CAUTION
EQUIPMENT DAMAGE
Failure to perform pre-system start-up checks may cause damage to equipment.
Do not power up or try to operate the analyzer unless it is physically secure and all electrical and pneumatic connections to the analyzer are in place.
Before starting up the analyzer, ensure that electrical power, sample gas handling facilities, and any calibration gases that are required are available to the analyzer.
Always follow the Start-up procedure.
Always follow the Shutdown procedure.
9
CAUTION
UNSERVICEABLE EQUIPMENT
If the pressure and temperature screen does not display measurements similar to those shown in Figure 8-1 and Figure 8-2, refer to Troubleshooting and diagnostics.
CAUTION
EMC
This is a Class A product. In a domestic environment, this product may cause radio interference, in which case the user may be required to take adequate measures.
As a general principle, if any optical component other than the cell assembly, the laser modules, and the detectors is unserviceable, the analyzer must be repaired by Emerson. This is because the repair, replacement, and alignment of the optical components requires the use of special optical test/calibration equipment and procedures.
Some faults can only be repaired by Emerson. Where an item is unserviceable, and no replacement procedure is given in this manual, then the fault must be repaired by Emerson.
10
Reference Manual Contents
00809-0100-4511 May 2020

Contents

Chapter 1 Introduction.............................................................................................................15
1.1 Description................................................................................................................................ 15
1.2 Customer information............................................................................................................... 15
1.3 Safety precautions and conditions for safe use...........................................................................15
1.4 Qualified personnel....................................................................................................................16
1.5 Software version........................................................................................................................16
1.6 Glossary and abbreviations........................................................................................................ 17
Chapter 2 Theory of operation................................................................................................. 21
2.1 Overview................................................................................................................................... 21
2.2 Laser measurement principle.....................................................................................................21
2.3 Gas concentration measurements............................................................................................. 21
Chapter 3 Description.............................................................................................................. 23
3.1 Equipment purpose and role......................................................................................................23
3.2 System overview........................................................................................................................23
3.3 Gas inputs and outputs.............................................................................................................. 25
3.4 Connecting the electrical/electronic inputs and outputs............................................................27
3.5 Optical description.................................................................................................................... 31
Chapter 4 Specifications...........................................................................................................33
4.1 Gas detection............................................................................................................................ 33
4.2 Detailed system specifications...................................................................................................33
4.3 Safety and system labels and annotation................................................................................... 39
Chapter 5 Install.......................................................................................................................41
5.1 Site selection............................................................................................................................. 41
5.2 Unpacking................................................................................................................................. 43
5.3 Mounting the analyzer...............................................................................................................44
5.4 Connecting the electrical/electronic inputs and outputs............................................................49
Chapter 6 Controls and display controller.................................................................................57
6.1 Front panel controls and indicators............................................................................................57
6.2 Display controller.......................................................................................................................58
6.3 Gas Sensor Main screen............................................................................................................. 58
6.4 Pressure and Temperature screen..............................................................................................60
6.5 Help system...............................................................................................................................61
6.6 Main menu................................................................................................................................ 62
6.7 BACK button..............................................................................................................................62
Chapter 7 Start-up procedure...................................................................................................63
7.1 Introduction.............................................................................................................................. 63
Rosemount CT5100 11
Contents Reference Manual
May 2020 00809-0100-4511
7.2 Preparation for use.................................................................................................................... 63
7.3 Start-up procedure.................................................................................................................... 64
7.4 Front panel controls and indicators............................................................................................67
Chapter 8 Operating the analyzer.............................................................................................69
8.1 Introduction.............................................................................................................................. 69
8.2 Normal operation...................................................................................................................... 69
Chapter 9 Shutdown procedure............................................................................................... 71
9.1 Safety precautions .................................................................................................................... 71
9.2 Shutdown procedure................................................................................................................. 72
Chapter 10 Gas calibration......................................................................................................... 77
10.1 Required tools......................................................................................................................... 77
10.2 Main menu calibration routines............................................................................................... 78
Chapter 11 Troubleshooting and diagnostics............................................................................. 99
11.1 Use the Built-in-Self-Test (BIST) fault diagnostics..................................................................... 99
11.2 Use the Faults menu.............................................................................................................. 103
11.3 Failure diagnosis principles.................................................................................................... 104
11.4 Repairable faults.................................................................................................................... 105
11.5 Tools and test equipment...................................................................................................... 106
11.6 General troubleshooting and diagnostics information........................................................... 106
11.7 Navigate the Diagnostics menu............................................................................................. 106
11.8 Visually examine the analyzer................................................................................................ 109
11.9 Functional failure diagnostics................................................................................................ 109
Chapter 12 Repair and replacement......................................................................................... 115
12.1 Repair policy.......................................................................................................................... 115
12.2 Open the electrical compartment door..................................................................................115
12.3 Close the electrical compartment door..................................................................................117
12.4 Replace the fuses...................................................................................................................117
12.5 Replace the TEC board........................................................................................................... 121
12.6 Replace the peripheral board................................................................................................. 123
12.7 Replace the motherboard......................................................................................................125
12.8 Replace the Ethernet in/out relay modules............................................................................ 127
12.9 Lower (main) DIN rail electronic components........................................................................ 128
12.10 Replace a DC power supply..................................................................................................129
12.11 Replace the 10 Amp solid state relay....................................................................................131
12.12 Replace the Ethernet connector...........................................................................................131
12.13 Replace the laser module.....................................................................................................132
12.14 Replace the detector module...............................................................................................134
12.15 Replacing the O2 detector................................................................................................... 135
12.16 Replace the main terminal board......................................................................................... 138
12.17 Replace the mains power filter.............................................................................................139
12 Emerson.com/Rosemount
Reference Manual Contents
00809-0100-4511 May 2020
12.18 Remove the top cover..........................................................................................................140
12.19 Replace the pressure sensor.................................................................................................141
12.20 Replace the PT100 cell temperature sensor......................................................................... 142
12.21 Cleaning the cell mirrors...................................................................................................... 145
12.22 Spare parts list..................................................................................................................... 150
Chapter 13 Preventative maintenance..................................................................................... 153
13.1 Cleaning the analyzer............................................................................................................ 153
13.2 Maintaining the analyzer........................................................................................................154
13.3 Scheduled maintenance........................................................................................................ 154
Appendix A Engineering drawings............................................................................................ 155
Rosemount CT5100 13
Contents Reference Manual
May 2020 00809-0100-4511
14 Emerson.com/Rosemount
Reference Manual
00809-0100-4511 May 2020
Introduction

1 Introduction

1.1 Description

The Rosemount CT5100 Continuous Gas Analyzer, referred to hereafter as the Rosemount CT5100, is an electronic sensor that uses laser spectroscopy to perform analysis of process gas streams.
The function of the analyzer is to detect and measure up to 10 different types of gas at concentrations ranging from parts per million (ppm) to percentage levels in the process gas stream.
Important
The Rosemount CT5100 is designed for use in Non Hazardous areas ONLY.
Note
This manual is intended for the personnel who install, operate, and maintain the equipment.

1.2 Customer information

This manual contains all the important information that must be followed to ensure the correct operation and safety of personnel when operating the analyzer.
For information regarding installation, consult Install and the Quick Start Guide.
Emerson is committed to continuously improving its products and documentation. Every effort will be made to include in the documentation any modifications by the manufacturer. However, this document reflects the supplied analyzer at the revision date on the front cover.
Should you require further information, or should particular problems arise that are not covered in this manual, you can request additional help from Cascade Technical Support (cascade.support@emerson.com) or Emerson distribution partners. Further contact details for Emerson can be found on the back page of this manual.

1.3 Safety precautions and conditions for safe use

WARNING
SAFE USE PRECAUTIONS
Before installing or performing any maintenance on the analyzer, read and understand the safety information given in the preliminary information of this manual.
The analyzer described in this guide has been quality control tested and left the manufacturer in pristine condition. To achieve the correct and safe operation of this product, it must be transported, installed, operated, and maintained as described by the manufacturer.
Rosemount CT5100 15
Introduction
May 2020 00809-0100-4511
All lasers used within the analyzer are Class 1. The emitted laser light is invisible (mid­infrared) and the pulse duration so short that the unprotected eye will not be damaged.
The nature of the laser beam path and beam width further ensures that it should be impossible to cause any eye damage. The analyzer has warning labels at appropriate positions in accordance with USA 21 CFR 1040.10.
General safety notice/residual risk
Installation, operation, and maintenance of the analyzer must be in accordance with these instructions.
When operated as intended and all applicable safety instructions are observed, an element of risk will remain, including, but not limited to, the following:
The emission of gases hazardous to health may be possible when all gas connections
have been correctly made.
To avoid exposure to the dangers of residual risks, take particular care when installing,
operating, maintaining, and servicing the analyzer.
Reference Manual

1.4 Qualified personnel

In-depth specialist knowledge is an absolute requirement for working with and on the analyzer. Personnel installing, operating, servicing, and maintaining the analyzer must be instructed, trained, qualified, and authorized personnel of the operating company for hazardous areas and the manufacturer.
It is the operating company's responsibility to:
Train staff
Observe safety regulations
Follow the safety instructions and procedures in the product manual
Operators must:
Be trained
Read and understand all relevant sections of the product manual before commencing
work
Know the safety mechanisms and regulations
WARNING
To avoid explosions, loss of life, personal injury, and damage to this equipment and on-site property, do not install, operate, maintain, or service this analyzer before reading and understanding this reference manual and receiving appropriate training.

1.5 Software version

The analyzer includes software that is used to control the operation of the analyzer. This manual describes the software version as: 5.7.13.
16 Emerson.com/Rosemount
Reference Manual Introduction
00809-0100-4511 May 2020

1.6 Glossary and abbreviations

Abbreviation Description
©
% Percent
< Less than
° Degree
AC Alternating current
ATEX Explosive atmospheres
Barg Pressure, in units of bars, above or below atmospheric pressure
BS British Standard
C Celsius
CDA Compressed dry air
CE European Conformity
Copyright
CFR Code of Federal Regulations
CGA Continuous Gas Analyzer
CH
4
CO
2
DC Direct current
Deg Degree (temperature)
e.g. For example
EC European Community
EMC Electromagnetic compatibility
EU European Union
Hrs Hours
Hz Hertz
H2O Water
ICL Interband Cascade Laser
IEC International Electro-technical Commission
in. Inches
IP Ingress protection
Methane
Carbon dioxide
IPxx Ingress protection (xx are numbers that define the protection level)
IS Intrinsically safe
ISO International Organization for Standardization
k Thousand
kg Kilogram
Rosemount CT5100 17
Introduction Reference Manual
May 2020 00809-0100-4511
Abbreviation Description
kHz Kilo hertz
L Liter
lb. Pound
LCD Liquid crystal display
LED Light emitting diode
L/min Liters per minute
m Meter
3
m
mA Milliamp
Max Maximum
mBar milli-Bar
mbps Megabits per second
Cubic meter
mg Milligram
3
mg/m
Milligram/cubic meter
Mid IR Mid Infrared
min Minute
mm Millimeter
N
2
NEC
®
Nitrogen
National Electrical Code
NFPA National Fire Protection Association
nm Nanometer
NH
3
Ammonia
NO Nitric oxide
NO
2
Nitrogen dioxide
N2O Nitrous oxide
No. Number
O
2
Oxygen
PC Personal computer
PM Preventative maintenance
ppm Parts per million
psi Pounds per square inch
QCL Quantum Cascade Laser
TDL Tunable Diode Laser
Torr Unit of pressure defined as exactly 1/760 of a standard atmosphere
18 Emerson.com/Rosemount
Reference Manual Introduction
00809-0100-4511 May 2020
Abbreviation Description
UKAS United Kingdom Accreditation Service
USA United States of America
USB Universal serial bus
V Volt
VA Volt-ampere
Vac Volt alternating current
Vdc Volt direct current
W Watt
WEEE Waste electrical and electronic equipment
µm Micro-meter
Rosemount CT5100 19
Introduction Reference Manual
May 2020 00809-0100-4511
20 Emerson.com/Rosemount
Reference Manual Theory of operation
00809-0100-4511 May 2020

2 Theory of operation

2.1 Overview

The Rosemount CT5100 is a gas sensor system that can be configured to measure the concentrations of multiple small molecules carried in the gas sample. The types of molecules that are measured depend on the system configuration.
The analyzer can be configured to detect and measure up to 10 gases, with ranges varying from volume to percent (%) volume levels. A detailed description of the system is given in
Detailed system specifications .

2.2 Laser measurement principle

The analyzer can use up to six lasers to detect and measure gases. Each laser can measure between one and three gases. Inside the laser, which is about the size of a pin head, electrons cascade down a series of quantum wells, producing a photon at each step.
This cascade of electrons can produce between 20 and 100 photons per electron, giving the layers a higher output power than traditional semi-conductor lasers.
The lasing wavelength of a laser is determined by adjusting the physical thickness of the semiconductor layers, giving access to high power lasers covering the mid-infrared spectral region. The lasers have no need for cryogenic cooling and have excellent spectral quality in chirped mode and good tuneability.

2.3 Gas concentration measurements

In the analyzer, gas concentrations are measured using mid-infrared optical absorption spectroscopy. The laser light sources are operated to produce wavelength sweeps that cover the absorption lines of the gases to be measured.
Sample gas, which may contain impurity gases that are to be detected and measured, is conditioned and drawn into the analyzer. Inside the analyzer, the sample gas is fed into an analysis cell, where the beams from the Quantum Cascade Lasers (QCLs) are passed through the gas. The analysis cell contains a set of mirrors that bounce the light back and forth many times, which lengthens the path of the lasers through the gas.
On exiting the analysis cell, the light is detected by a receiver unit. The variation in the intensity of light in the vicinity of absorption lines for the gases being detected is measured, and the concentration is determined using a comprehensive spectral fitting routine.
Rosemount CT5100 21
Theory of operation Reference Manual
May 2020 00809-0100-4511
22 Emerson.com/Rosemount
Reference Manual
00809-0100-4511 May 2020
Description

3 Description

3.1 Equipment purpose and role

The types of molecules that are measured depend on the system configuration.
Figure 3-1: Rosemount CT5100 Continuous Gas Analyzer
The analyzer can be configured to detect and measure up to 10 different gases, depending on the combination of laser modules fitted.

3.2 System overview

A complete Rosemount CT5100 system consists of a gas handling system, the analyzer, and the associated interconnecting wiring and gas piping.
The customer MUST provide the gas handling system and interconnecting wiring and gas piping.
The Rosemount CT5100 is supplied by Emerson. The gas handling system may be provided by either the customer or Emerson, depending upon the specific installation. The
Rosemount CT5100 23
Description
Reference Manual
May 2020 00809-0100-4511
circuit breaker used to control the application of electrical power to the analyzer, the interconnecting wires, and gas piping are provided by the customer.
In Figure 3-2, the items supplied by Emerson are colored blue; the items supplied by the customer are colored purple. The green gas handling system may be provided by Emerson or the customer.
Table 3-1 lists the main items of the system.
Figure 3-2: Rosemount CT5100 Installation
A. Gas handling system
B. Sample supply line C. Sample return (exhaust) line D. Rosemount CT5100
E. Electrical power
F. Two pole main isolator complete with RCD G. Control center H. Measurement data
The analyzer contains an optical system with multiple lasers and a series of optical components that provide an optical path, a heated multi-pass analysis cell, and sample and outlet ports that can be connected to a gas handling system and control and analysis electronics. The number of lasers installed depends upon customer requirements. The complete system operates at either 110 or 240 Vac 50/60 Hz supply.
The analyzer uses mid-infrared optical absorption spectroscopy to measure gas concentrations. The light sources are Quantum Cascade Lasers (QCLs), which are operated to produce wavelength sweeps that cover the absorption lines of the gases. The light from each laser is routed through an optical path to the analysis cell, which provides measurement of low concentrations of the subject gases. An external sample handling system conditions the sample gas and draws it through the analysis cell. The light exits the multi-pass analysis cell and is directed to a receiver in the analyzer. The variation in the intensity of light in the vicinity of the absorption lines is measured, and the concentration is determined using a comprehensive spectral fitting routine.
There is no sample conditioning provided within the analyzer; the sampled gas must be brought within the parameters shown in Detailed system specifications before entering the analyzer. Detailed characteristics of the analyzer are also given in in Table 4-1.
24 Emerson.com/Rosemount
Reference Manual Description
00809-0100-4511 May 2020
Table 3-1: Main Items of the Rosemount CT5100 Installation
Item Name or description Supplied by Part number Quantity Notes
1 Rosemount CT5100 Emerson Rosemount CT5100 1 N/A
2 Rosemount CT5100 software
package, version 5.x.x.
3 Gas handling system Customer
4 Heated gas sample line hose Customer Customer choice 1 N/A
5 Exhaust line hose (for sample
gas)
6 Reference gas cylinders
(instrument gas) for calibration purposes
7 Pressure regulator Customer Customer choice 1 per gas
8 Pneumatic T-piece Customer Customer choice 1 Required for
9 Excess flow line Customer Customer choice 1 Required for
10 Power cables to Rosemount
CT5100
11 Cables from Rosemount
CT5100 to control center
Emerson N/A. Software is
embedded in the system.
Customer choice or (optionally by Emerson)
Customer Customer choice 1 N/A
Customer Customer choice Dependent upon
Customer Customer choice 1 N/A
Customer Customer choice 1 N/A
Emerson
1 Version
described in manual
1 N/A
N/A number of gases being measured
Required for cylinder
calibration
calibration
calibration
12 Main circuit breaker
complete with RCD
Customer Customer choice N/A N/A

3.3 Gas inputs and outputs

The analyzer has one gas input and one gas output. See Figure 3-3.
Rosemount CT5100 25
Description
Reference Manual
May 2020 00809-0100-4511
Figure 3-3: Gas Inlet and Outlet Connectors
A. Sample gas return port
B. Sample gas input port C. Top cover of Rosemount CT5100 (cell compartment) D. Laser/electrical compartment
Procedure
1. The gas sample that is to be measured for impurities enters the analyzer through
the sample gas input port located on top of the analyzer (see Figure 3-3.).
2. Once the gas sample has been examined for impurities, it is expelled from the
analyzer through the sample gas return port.
3. The sample gas supply line must be heated all the way to the sample gas input port
on the analyzer to prevent condensation forming in the line.
26 Emerson.com/Rosemount
Reference Manual Description
00809-0100-4511 May 2020
WARNING
HAZARDOUS GASES
The product stream that the analyzer is examining may be hazardous even at low concentrations.
Therefore, take special care to ensure that the sample gas return port either returns the sample gas to the product stream or discharges the sample gas to a location that will not cause a hazard.

3.4 Connecting the electrical/electronic inputs and outputs

CAUTION
If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
Rosemount CT5100 27
Description Reference Manual
May 2020 00809-0100-4511
Figure 3-4: Electrical/Electronic Connectors
A. Power entry point
B. Analog/digital entry point
C. Analog/digital entry point
Electrical/electronic signal connections to the analyzer are made through three electrical entry points located on the underside of the analyzer, as shown in Figure 3-4. Use the wiring diagram to make the electrical connections as shown in Engineering drawings and
Figure 3-5.
28 Emerson.com/Rosemount
Reference Manual Description
00809-0100-4511 May 2020
Figure 3-5: Signal Cable Outputs
A. To mains input fuses
B. Mains input filter C. To enclosure earth stud D. To digital modules
E. To digital modules
F. To analog modules G. To status relays H. Digital output terminals
I. Analog or digital output terminals J. Analog output terminals
K. Status outpput terminals
L. Mains input terminals
Table 3-2: Rosemount CT5100 System Wiring User Connections
Terminal Function
1 Sensor system supply (L)
2 Sensor system supply (N)
3 Earth
11 Digital output 1
12 Digital output 2
13 Digital output 3
14 Digital output 4
15 Digital output 5
16 Digital output 6
17 Digital output 7
18 Digital output 8
19 Digital output 9
Rosemount CT5100 29
Description Reference Manual
May 2020 00809-0100-4511
Table 3-2: Rosemount CT5100 System Wiring User Connections (continued)
Terminal Function
20 Digital output 10
21 Digital output 11
22 Digital output 12
23 Analog output 1
24 Analog output 2
25 Analog output 3
26 Analog output 4
27 Analog output 5
28 Analog output 6
29 Analog output 7
30 Analog output 8
31 Status output 1 (Check function)
32 Status output 2 (Maintenance required)
33 Status output 3 (Out of specification)
34 Status output 4 (Failed)
01 Analog or digital input/output
02 Analog or digital input/output
03 Analog or digital input/output
04 Analog or digital input/output
05 Analog or digital input/output
06 Analog or digital input/output
Electrical power is applied to the analyzer through the power entry point, Figure 3-4.
The power supply is 110 or 230 Vac, 50/60 Hz ± 10%. AC to DC. Power converters inside the analyzer automatically adjust in response to the input voltage level and ensure that the correct DC voltage is available inside the unit. The analyzer is electrically protected by an internal 5 A, 250 VA fast acting fuse on the instrumentation electrical supply line and an internal 2 A, 250 VA fuse on the purge electrical supply line.
WARNING
Failure to follow this warning may result in personal injury and/or damage to persons and/or property.
Make sure that the mains supply cable used is of a suitable rating for the unit power requirements and is of a three core earthed construction.
30 Emerson.com/Rosemount
Reference Manual Description
00809-0100-4511 May 2020
The digital outputs conduit (B and C) provides an Ethernet output from the instrument that may be used for downloading data for failure diagnosis purposes or for downloading data to the process control center.
The results of the gas analysis are output from the instrument through the 4-20 mA analog or Modbus® outputs and sent to the process control center.
The optional digital outputs provide fault indications to your process control center. Each digital output is connected to a normally closed relay, located inside the analyzer, which will open in response to the detection of a specific fault. The possible causes of a fault indication are:
1. The sample gas concentration is outside of specification (i.e., the sample gas
concentration has exceeded the measurement range of the instrument).
2. The analyzer is out of specification or has developed a fault.
NOTICE
The installation of the gas analyzer shall be in accordance with all local and national standards.

3.5 Optical description

The laser modules are located in the core of the analyzer. Each laser module produces a separate light beam, and these beams are combined linearly as the modules are aligned in the system. The combined beams are closely coupled, parallel, and coaxial about a virtual line. The laser light beams pass through a baseplate onto an optical steering assembly, which directs the laser beam through the sample cell.
The sample cell contains a set of mirrors to create a path through the sample gas that is between 0.66 ft. (0.2 m), 6.56 ft. (2 m), 16.4 ft. (5 m), and 19.2 ft. (15 m) through multiple reflections along the length of the cell. The laser beams exit the cell at the opposite end from where they entered and are directed using a second optical block to a receiver.
By measuring and analyzing the light detected by the receiver unit, it is possible to accurately determine the concentrations of the target molecules within the gas sample cell.
Rosemount CT5100 31
Description Reference Manual
May 2020 00809-0100-4511
32 Emerson.com/Rosemount
Reference Manual Specifications
00809-0100-4511 May 2020

4 Specifications

4.1 Gas detection

The analyzer is highly configurable in the gases that can be detected and their range of concentrations.

4.2 Detailed system specifications

Table 4-1 gives the physical characteristics of the analyzer. Schematic diagrams of the
sensor and mounting points are shown in Figure 4-1 and Figure 4-2. Table 4-2 gives the general characteristics of the analyzer.
Table 4-1: Physical Characteristics
Rosemount CT5100 Value Comment
External dimensions 22.64 x 11.73 x 28.11 in.
575 x 298 x 714 mm
Weight 117 lb.
53 kg
Length x width x height Nominal dimensions
Approximate weight
Table 4-2: General Characteristics
Rosemount CT5100 Value Units Comment
Instrument supply voltage 110 to 240 Vac 50/60 HZ ±10%
Peak power consumption 500 W Max consumption per gas
analyzer
Continuous steady-state power consumption
Electrical compartment enclosure N/A N/A 304 stainless steel
Optical compartment enclosure N/A N/A Polyester TGIC free powder
Wetted materials N/A N/A AISI 316 (EN 1.4401 grade)
300 W Once the gas analyzer has
stabilized and the analysis cell has reached the temperature set point
coated 304 stainless steel
stainless steel tubing and fittings including thermowell and pressure diaphragm, PFA coated aluminum cell body, PTFE seals, protected gold coated mirrors, CaF2 windows, and FKM (typically Viton™) or FFKM (typically Kalrez™) O-rings
Measurement technique N/A N/A Mid infrared (IR) absorption
spectroscopy
Mid IR source N/A N/A Quantum Cascade Laser
Rosemount CT5100 33
Specifications Reference Manual
May 2020 00809-0100-4511
Table 4-2: General Characteristics (continued)
Rosemount CT5100 Value Units Comment
Near IR source Interband Cascade Laser Diode
Laser
Laser classification Class 1 BS EN 60825-1: 2007 safety of
laser products. Equipment classification and requirements (identical to IEC 60825-1 2007)
Inlet gas port connector ¼
6
Outlet (exhaust) gas port connector ¼
6
Measurement result signals 4 to 20 mA 4 or 8 channel outputs, specify
Communication 10/100 Mbps Ethernet
Warm-up time 90 minutes N/A
in. mm
in. mm
Swagelok® type, factory­configured, specify on order
Swagelok type, factory­configured, specify on order
on order
34 Emerson.com/Rosemount
Reference Manual Specifications
00809-0100-4511 May 2020
Figure 4-1: Rosemount CT5100 Dimensions - Front View
Dimensions are in inches (mm).
A. Lifting eyelet
B. Ventilation
C. User interface
Rosemount CT5100 35
Specifications Reference Manual
May 2020 00809-0100-4511
Figure 4-2: Rosemount CT5100 Dimensions - Top View
Dimensions are in inches (mm).
A. Lifting eyelet
B. Sample return
C. Sample inlet
36 Emerson.com/Rosemount
Reference Manual Specifications
00809-0100-4511 May 2020
Figure 4-3: Rosemount CT5100 Dimensions - Bottom View
Dimensions are in inches (mm).
A. Cable glands
B. Earth point
Table 4-3: Environmental Characteristics
Environmental characteristic Value Units Comment
Operating temperature range -4 to 131
-20 to 55
Sample gas temperature range 131 to 374
50 to 190
Sample gas moisture content 30 (dependent on gas
type. Customer must confirm.)
Sample gas particulate density 5 mg/m
Sample gas particulate size 10 µm Maximum
IP code IP66 (electrical
compartment enclosure)
IP20 (optical compartment enclosure)
°F °C
°F °C
% Maximum
3
N/A IP to IEC 60529
Ambient temperature
Factory set, specify on order
Maximum
Rosemount CT5100 37
Specifications Reference Manual
May 2020 00809-0100-4511
Table 4-3: Environmental Characteristics (continued)
Environmental characteristic Value Units Comment
Sensor humidity range 10 to 95 % Relative humidity (non-
condensing) at 113 °F (45 °C)
38 Emerson.com/Rosemount
Reference Manual Specifications
00809-0100-4511 May 2020

4.3 Safety and system labels and annotation

The labels and annotation applied to the analyzer are specified in the table below.
Label type Example Location
Identification label (including serial number and model number)
Fuse identification label 1. Back plate
Ratings label Enclosure side panel
Emerson.com/RosemountGasAnalysis Front panel
2. Top right inside of door
Laser radiation CAUTION label Baseplate
Laser module identification label On each laser module housing
Intrinsically safe label 1. HMI
2. Intrinsically safe sensor barrier
Rosemount CT5100 39
Specifications Reference Manual
May 2020 00809-0100-4511
Label type Example Location
Terminal label Top left inside of door
Earth identification label Back plate
Manufacturer's label On analysis cell heater block
Electrical safety label On inside of electrical compartment
door
AC Power Supply Danger label 1. On outside of electrical
compartment door
2. On manifold block of air overpressure system
40 Emerson.com/Rosemount
Reference Manual Install
00809-0100-4511 May 2020

5 Install

5.1 Site selection

The Rosemount CT5100 has a T3 temperature classification which specifies the maximum surface temperature of the analyzer. Ensure that no combustible gas concentrations will be present, whether on a continual or occasional basis, which have an ignition temperature below the T-rating of the analyzer.
WARNING
FIRE AND EXPLOSION
Death, personal injury, and/or damage to persons and/or property may result if this is not observed.
The analyzer's electrical compartment must not be opened unless the atmosphere in the area is known to be below the ignitable concentration of combustible gases or materials, or unless all equipment within the protected enclosure is de-energized.
WARNING
ELECTRIC SHOCK
The analyzer operates using mains voltage, which may cause death or serious injury to personnel. Failure to observe this precaution will cause death, personal injury, and/or damage to persons and/or property.
Ensure that the circuit breakers are set to Off and locked out and tagged out off before removing the top cover or opening the front cover.
The analyzer is intended to be installed in a suitable Division 2 shelter to protect it from the elements.
Provide sufficient space around the analyzer to allow the maintenance and servicing of the unit.
Rosemount CT5100 41
Install Reference Manual
May 2020 00809-0100-4511
Figure 5-1: Clearance with Door Open
Dimensions are in inches (mm).
A. Door open
B. Door closed
C. Door opening arc
42 Emerson.com/Rosemount
Reference Manual Install
00809-0100-4511 May 2020

5.2 Unpacking

This procedure requires a minimum of two people to safely remove the equipment from the shipping container.
WARNING
HEAVY INSTRUMENT - LIFTING HAZARD
Handle the analyzer with caution during unpacking, installation, maintenance, and transport to prevent crushing of hands, feet, or other body parts.
The analyzer weighs 117 lb. (53 kg) and should always be lifted and moved using suitable lifting/moving equipment. Emerson recommends that a minimum of two people using suitable tools for transportation and lifting are employed.
Wear suitable protective gloves and protective footwear.
CAUTION
EQUIPMENT DAMAGE
Failure to observe this caution may cause damage to the equipment.
When preparing the analyzer for transport by air, road, or rail, safeguard the analyzer against movement or break-away during transport by securely strapping it in place.
WARNING
EXPLOSION HAZARD
Installing and wiring the analyzer must comply with all relevant national legislative requirements and regulations.
Consider all safety instructions within this manual and all associated analyzer instruction manuals.
WARNING
EXPLOSION HAZARD
Installing the analyzer requires opening the enclosure and working at the open unit. This is permitted only when both the analyzer and connected external circuitry are de-energized.
Depending on the local regulation, this may require a competent hot work supervisor to issue a hot work permit.
Rosemount CT5100 43
Install
May 2020 00809-0100-4511
Reference Manual
WARNING
HEAVY ITEM
Failure to propery handle the analyzer may cause injury to personnel.
Ensure the wall the analyzer is mounted on is solid, stable, and of suitable material to hold the weight of the analyzer.
Handle the analyzer with caution during unpacking, installing, maintaining, and transporting to prevent crushing of hands, feet, or other body parts.
The analyzer weighs 117 lb. (53 kg). Emerson recommends that a minimum of two people move and lift the analyzer. Wear suitable protective gloves and protective footwear.
CAUTION
SHOCK AND VIBRATION
Damage to the analyzer may result from a failure to follow this caution.
The analyzer contains sensitive electronic equipment. It MUST NOT be subjected to any shock and or vibration.
Procedure
1. On receipt of goods, look for any visible damage to the analyzer and verify that all items noted to be shipped were received. Record on the goods receipt note any damage or missing items, noting both the item(s) and quantity missing.
2. Visually inspect the exterior of the analyzer for signs of damage, corrosion, gas leaks, or signs of previously overheating.
3. Report anything found to the maintenance organization.
4. Attach suitably rated and tested lifting slings to the safety engineered lifting eye bolts mounted on top of the analyzer.
5. One person should carefully guide the equipment from the horizontal to vertical position while the other person lifts the equipment.
6. Use safety approved and tested lifting equipment to remove the analyzer from the shipping container and place it on a solid, level surface.
7. Ensure that the analyzer is stored in its protective plastic cover until installation.

5.3 Mounting the analyzer

This procedure requires two people to safely move and mount the Rosemount CT5100.
Procedure
1. Ensure that there is free space around the analyzer to allow ventilation of the upper part of the analyzer.
44 Emerson.com/Rosemount
Reference Manual Install
00809-0100-4511 May 2020
WARNING
HEAVY ITEM
Failure to propery handle the analyzer may cause injury to personnel.
Ensure the wall the analyzer is mounted on is solid, stable, and of suitable material to hold the weight of the analyzer.
Handle the analyzer with caution during unpacking, installing, maintaining, and transporting to prevent crushing of hands, feet, or other body parts.
The analyzer weighs 117 lb. (53 kg).
Emerson recommends that a minimum of two people move and lift the analyzer.
Wear suitable protective gloves and protective footwear.
Rosemount CT5100 45
Install
Reference Manual
May 2020 00809-0100-4511
2. Attach suitably rated and testing lifting slings to the safety engineered lifting eye bolts mounted on top of the analyzer.
Figure 5-2: Front View Dimensions
A. Lifting eye bolt B. Sample gas input port C. Sample gas return port
D. Lifting eye bolt
E. 0.413-in. (10.5 mm) diameter mounting bolts
3. One person should carefully guide the equipment while the other person operates the lifting equipment.
46 Emerson.com/Rosemount
Reference Manual Install
00809-0100-4511 May 2020
4. Use safety approved and tested lifting equipment to lift the analyzer from the stable platform.
5. Mount the analyzer using four M8 (-in.) fasteners to attach the wall mount brackets.
The bolts must be positioned in such a way to allow maximum use of all the thread length.
The installer must ensure that the fasteners used are suitable for the load and surface that the analyzer is mounted on.
In case you need to thread lock the fittings for extra security, only do this with compounds compatible with the zone classification of the installation location.
The four wall fixing points must be 0.413-in. (10.5 mm) diameter mounting holes. Ensure that the wall fixing points are capable of supporting a load of 242 lb (110 kg)
each; this includes a x 2 factor of safety. Figure 5-3 shows the locations of the mounting points on the analyzer. All mounting points are 0.413-in. (10.5 mm) diameter holes.
Confirm the bolts are secure. Do not overtighten the fasteners. The analyzer must be mounted using the four off factory fitted and predrilled holes
on the brace bars. Refer to Figure 5-2.
6. Remove the lifting eyes and retain them for future use. Threads must be protected with a suitable grease and plastic grommets. After mounting, do not place any additional load on the analyzer. Do not place or leave loose items on flat surfaces.
Rosemount CT5100 47
Install Reference Manual
May 2020 00809-0100-4511
Figure 5-3: Mounting Details
Dimensions are in inches (mm).
A. Mounting points
48 Emerson.com/Rosemount
Reference Manual
00809-0100-4511 May 2020
Figure 5-4: Clearance with Door Open
Install
A. Door open B. Door closed C. Door opening arc

5.4 Connecting the electrical/electronic inputs and outputs

5.4.1 AC power
Power is connected to the analyzer instrumentation through the power entry point (A) fitted to the base of the analyzer. Refer to Figure 5-5.
Rosemount CT5100 49
Install Reference Manual
May 2020 00809-0100-4511
Figure 5-5: Power Gland
A. Power and signal/exit points
B. M6 earth stud
The customer supplied circuit breaker, complete with RCD, must be in accordance with local and national standards.
50 Emerson.com/Rosemount
Reference Manual Install
00809-0100-4511 May 2020
Figure 5-6: Power Entry Point Connections
A. To mains input fuses
B. Mains input filter
C. To enclosure earth stud
Table 5-1: Mains Input Terminals User Connections
Terminal Function
1 Sensor system supply (L)
2 Sensor system supply (N)
3 Earth (E)
Electrical protection for the instrumentation circuitry of the analyzer is provided by fuses F1 and F2 located inside the analyzer. Refer to Figure 5-7.
The customer supplied power cable for the analyzer instrumentation will be connected to terminals 1 - 3.
Rosemount CT5100 51
Install Reference Manual
May 2020 00809-0100-4511
Table 5-2: Electrical Power Requirements
5.4.2
Electrical supply Power
consumption
Instrumentation supply voltage
500 W 100 to 230 Vac, 50/60 Hz ±10% 3.15 A
Voltage Fuse
internal fuses F1 and F2
For the electrical power wiring use 16 AWG stranded, three conductor copper or tin-plated copper power wire, rated for at least 250 Vac, of the required length. Cables must be terminated in the power entry points in accordance with local and national electrical codes. The full electrical wiring diagram is provided in Engineering drawings.
Fuses
Figure 5-7 shows the location of the fuses.
Figure 5-7: Fuses
A. Fuse F4 (24 Vdc supply)
B. Fuse F3 (12 Vdc supply) C. Fuse F1 (mains supply – live) D. Fuse F2 (mains supply – neutral)
E. Industrial power supply
Table 5-3: Fuse Requirements
Fuse Function Rating Schurter part number
1 Live line 110/240 Vac
(following mains filter)
2 Neutral line 110/240
Vac (following mains filter)
52 Emerson.com/Rosemount
3.15 A, 240 V, fast acting ceramic
3.15 A, 240 V, fast acting ceramic
0001.1009
0001.1009
Reference Manual Install
00809-0100-4511 May 2020
Table 5-3: Fuse Requirements (continued)
Fuse Function Rating Schurter part number
3 Analyzer 12 Vdc supply
rail
4 Analyzer 24 Vdc supply
rail
3.15 A, 240 V, fast acting ceramic
3.15 A, 240 V, fast acting ceramic
0001.1009
0001.1009
5.4.3 Connecting the sample supply and return line
Sample gas supply and sample return connections are ¼-in. (6 mm) ”Hamlet” compression tube fittings. To avoid the risk of gas leaks, confirm that these connections are made correctly and tightly. Both the sample gas supply pipe and the sample return pipe should be thermal insulated.
The maximum sample gas supply pressure is 2 BarG.
5.4.4
Connecting the signal cables
The signal cables are connected to the system through conduit outlets (B and C as shown in Figure 3-4). All signal cables are to be minimum 20 AWG tri-rated switchgear cable. Customer supplied conduit and cables are to be terminated in the conduit outlets in accordance with local and national electrical codes.
Rosemount CT5100 53
Install Reference Manual
May 2020 00809-0100-4511
Figure 5-8: Signal Cable Outputs
A. To mains input fuses
B. Mains input filter C. To enclosure earth stud D. To digital modules
E. To digital modules
F. To analog modules G. To status relays H. Digital output terminals
I. Analog or digital output terminals J. Analog output terminals
K. Status output terminals
L. Mains input terminals
Table 5-4: System Wiring
Terminal Function
1 Sensor system supply (L)
2 Sensor system supply (N)
3 Earth
11 Digital output 1
12 Digital output 2
13 Digital output 3
14 Digital output 4
15 Digital output 5
16 Digital output 6
17 Digital output 7
18 Digital output 8
54 Emerson.com/Rosemount
Reference Manual
Install
00809-0100-4511 May 2020
Table 5-4: System Wiring (continued)
Terminal Function
19 Digital output 9
20 Digital output 10
21 Digital output 11
22 Digital output 12
23 Analog output 1
24 Analog output 2
25 Analog output 3
26 Analog output 4
27 Analog output 5
28 Analog output 6
29 Analog output 7
30 Analog output 8
31 Status output 1 (Check function)
32 Status output 2 (Maintenance required)
33 Status output 3 (Out of specification)
34 Status output 4 (Failed)
01 Analog or digital input/output
02 Analog or digital input/output
03 Analog or digital input/output
04 Analog or digital input/output
05 Analog or digital input/output
06 Analog or digital input/output
5.4.5 Power input cables and circuit breaker
The power input cable circuit breaker can be tested as follows:
Set the main power circuit breaker to ON.
Check that the display controller lights up. The analyzer will then begin to power-up.
5.4.6
Temperature sensor and cell heater
The sample cell is controlled to operate at a pre-set temperature when it leaves the factory.
With the analyzer on, the cell will reach the pre-set operating temperature in approximately 90 minutes.
Rosemount CT5100 55
Install Reference Manual
May 2020 00809-0100-4511
5.4.7 Pressure sensor
The pressure sensor monitors the pressure in the analysis cell.
To test that it is functioning, check that a pressure reading is displayed under pressure on the Display Controller as described in Pressure and Temperature screen.
The reading will be approximately 760 Torr at atmospheric pressure. If desired, cap off the gas inlet and use an external pump to evacuate the cell. Verify that the pressure drops as expected.
NOTICE
A Torr is a non-SI unit of pressure, defined as 1/760 of standard atmospheric pressure, and is equal to the fluid pressure of 1 mm of mercury.
5.4.8
5.4.9
5.4.10
Analog output cable
In order to generate a 4-20 mA output, the analyzer must be left for 90 minutes to warm up, and the analysis cell must be at the correct pressure.
The 4-20 mA outputs will operate when the analysis cell temperature and pressure are within the required test range. It is not essential to flow sample gas through the system; nitrogen or atmospheric air will be adequate for this test.
With the analyzer at operating temperature and pressure, ensure that a current between 4 mA and 20 mA is generated on each 4-20 mA output. This can either be measured as a current with a multimeter, or as a gas concentration through the control station.
Seal glands
Where poured seal glands are a local industry requirement for cable termination, they should be made in accordance with the manufacturer’s instructions once acceptance tests have been made on the system to ensure that the analyzer does not need to be removed.
Commissioning
Once the sensor is fully installed as described above, it should be commissioned in accordance with the commissioning plan agreed between Emerson and the customer.
56 Emerson.com/Rosemount
Reference Manual Controls and display controller
00809-0100-4511 May 2020

6 Controls and display controller

6.1 Front panel controls and indicators

The Rosemount CT5100 is configured from the control display located on the front panel. Refer to Figure 6-1.
Figure 6-1: Front Panel
A. Display controller
NOTICE
ON/OFF circuit breaker
There is no ON/OFF switch on the analyzer. The external circuit breaker controls the application of electrical power to the analyzer.
The external circuit breaker is a simple two-pole ON/OFF circuit breaker that must to set to ON to permit the safe operation of the analyzer.
The display controller controls operation of the analyzer.Refer to Figure 6-2.
Rosemount CT5100 57
Controls and display controller Reference Manual
May 2020 00809-0100-4511

6.2 Display controller

The six buttons on the display controller control operation of the Rosemount CT5100:
Figure 6-2.
Figure 6-2: Display Controller Buttons
A. LCD display
B. Configurable button C. Scroll up button D. Scroll down button
The LCD displaycan be used to display:
Gas concentration measurements obtained
Operating temperature and pressure
Help screens
Step-by-step calibration
Diagnostics
Use the two scroll buttonsto scroll through the information on the LCD display. Use the right-hand scroll button to scroll up and the left-hand scroll button to scroll down.
The other four buttons are configured to perform different functions according to the software screen that is shown on the LCD display.

6.3 Gas Sensor Main screen

When the analyzer is switched on at the end of the start-up procedure, the Gas Sensor Main screen (Figure 6-3) appears.
The Gas Sensor Main screen is the screen that is normally displayed.
58 Emerson.com/Rosemount
Reference Manual Controls and display controller
00809-0100-4511 May 2020
NOTICE
The gas concentrations shown in the following screenshots may be different from those shown in your particular analyzer. The screenshots indicate the functionality of the software, which is the same regardless of the gases or gas concentrations being measured.
Figure 6-3: Gas Sensor Main Screen
The Gas Sensor Main screen displays the gas concentration measurements obtained by the analyzer. In the example shown in Figure 6-3 the gases nitric oxide (NO), nitrogen dioxide (NO2), oxygen (O2), carbon monoxide (CO), and carbon dioxide (CO2) are being measured and, for each gas, the concentration detected is in parts per million (ppm) or percentage as applicable.
The NOx reading is the total nitrogen oxide reading, a combination of nitric oxide (NO) and nitrogen dioxide (NO2).
At the end of the start-up procedure, the gas measurements initially appear as 0.00 ppm until the first readings are taken. After a few seconds, the initial gas concentrations are displayed.
The Gas Sensor Main screen also shows the status of the analyzer. In the example shown in
Figure 6-3, the analyzer is Running and OK (e.g., no faults have been identified). This area
of the display shows any errors detected by the software.
On the software screens, highlighted items are links to other screens in the software. To access a screen, press the button next to the highlighted item.
Rosemount CT5100 59
Controls and display controller Reference Manual
May 2020 00809-0100-4511
Figure 6-4: Gas Sensor Main Screen Buttons
A. MENU button
B. MENU text C. PAGE text D. PAGE button
E. HELP button
F. HELP text
G. STATUS button
PAGE
HELP
MENU
On the Gas Sensor Main screen, the STATUS button has no function when the analyzer is operating correctly. If, however, the software detects a fault, an error message is displayed. Press STATUS to get further information on the error.
A link between the Gas Sensor Main screen and the Pressure and Temperature screen (described in Pressure and Temperature screen). Press PAGE to toggle between these two screens.
A link to the Help system. Press HELP to go to the Help screen (described in Help
system).
A link to the Main menu of the software. Press MENU to go to the Main menu screen (described in Main menu).

6.4 Pressure and Temperature screen

The Pressure and Temperature screen (Figure 6-5) shows pressure and temperature measurements taken inside the analyzer.
60 Emerson.com/Rosemount
Reference Manual Controls and display controller
00809-0100-4511 May 2020
Figure 6-5: Pressure and Temperature Screen
The Cell reading is the temperature, in °C, for the analysis cell.
The Gas reading is the temperature, in °C, of the gas within the analysis cell.
The Pres reading is the pressure, in Torr, inside the analysis cell.
NOTICE
A Torr is a non-SI unit of pressure defined as 1/760 of standard atmospheric pressure and is equal to the fluid pressure of 1 mm of mercury.

6.5 Help system

The analyzer software includes a context-sensitive help system. Press HELP, which is available on most of the software screens, to open the help system.
The help system contains a number of different Help screens, each conveying a different message. As the help system is context-sensitive, the Help screen that appears is the one that is most appropriate to the software function engaged when HELP was pressed. Figure
6-6 shows an example of a Help screen.
Figure 6-6: Example of a Help Screen
Rosemount CT5100 61
Controls and display controller Reference Manual
May 2020 00809-0100-4511

6.6 Main menu

To access the Main menu (Figure 6-7), press MENU on either side of the Gas Sensor Main screen (Figure 6-3) or the Pressure and Temperature screen (Figure 6-5). The Main menu (Figure 6-7) is used for calibration, diagnostics, fault finding, downloading data, and shutting down the analyzer.
Figure 6-7: Main Menu

6.7 BACK button

On most of the software screens, the top lefthand button (Figure 6-8) is configured as a BACK button. Press BACK to return to the previous screen.
Figure 6-8: BACK Button
A. BACK button
62 Emerson.com/Rosemount
Reference Manual Start-up procedure
00809-0100-4511 May 2020

7 Start-up procedure

7.1 Introduction

CAUTION
EQUIPMENT DAMAGE
Damage to the analyzer may result from a failure to follow this procedure.
Always follow the start-up procedure.
The analyzer normally operates continuously. It should only be necessary to start up the analyzer under the following circumstances:
When the analyzer is first switched on following installation
Following repair or maintenance
When the analyzer has been switched off as part of a plant shutdown or maintenance

7.2 Preparation for use

Install and fully commission the Rosemount CT5100 before starting it up.
WARNING
BURN HAZARD
Personal injury and/or damage to property may result if these safety precautions are not observed.
Some parts of the analyzer may be heated to 374 °F (190 °C). To prevent burns, do not touch any of the hot parts. All components of a analyzer are hot unless it has been switched off and allowed to cool down.
Before fitting, removing, or performing any maintenance on the analyzer, make sure that it has been switched off and allowed to cool for at least two hours. Before performing any maintenance on or, in the vicinity of the analysis cell, allow the analyzer to cool for at least twelve hours, as the analysis cell is insulated against heat loss.
When handling the analyzer, always use suitable protective gloves.
These precautions are especially important when working at heights. If a burn is received, seek medical treatment immediately.
Rosemount CT5100 63
Start-up procedure Reference Manual
May 2020 00809-0100-4511
CAUTION
EQUIPMENT DAMAGE
Failure to perform pre-system start-up checks may cause damage to equipment.
Do not power up or try to operate the analyzer unless it is physically secure and all electrical and pneumatic connections to the analyzer are in place.
Before starting up, ensure that electrical power, sample gas handling facilities, and any required calibration gases are available to the analyzer.

7.3 Start-up procedure

NOTICE
The gases shown in the screenshots and the measurements thereof may be different from those shown in your Rosemount CT5100. They indicate the functionality of the software, which is the same regardless of the gases being measured.
NOTICE
To stop the start-up procedure at any time, set the main circuit breaker to OFF.
Procedure
1. Visually inspect the analyzer's exterior for signs of damage, corrosion, gas leaks, or
overheating. Report anything found to the maintenance organization.
2. Ensure that the analyzer has been installed correctly as described in Install.
3. Ensure that the top cover is fitted to the analyzer. If it is not, report it to the
maintenance organzation and do not proceed further until the top cover has been fitted.
4. Ensure that the door to the electrical compartment is closed and locked. If you
cannot close and lock the door, report it to the maintenance organization and do not proceed further until the door has been repaired.
5. Ensure that the gas handlying system is turned off.
6. Ensure that the external circuit breaker is set to OFF.
7. Visually examine the gas ports.
If necessary, refer to Figure 7-1 to ensure that the sample supply line and the sample return line are correctly attached.
64 Emerson.com/Rosemount
Reference Manual Start-up procedure
00809-0100-4511 May 2020
Figure 7-1: Gas Inlet and Outlet and Electrical Connections
A. Sample gas input port B. Sample gas return port C. Top cover (cell compartment)
D. Laser/electrical compartment
E. Analog/digital power entry point F. Power entry point
8. Ensure that the electrical connection has been made to the power entry point at the
base of the analyzer (Figure 7-1).
9. Start up the system that vents the sample gas exiting the analyzer.
10. Switch the circuit breaker to ON, applying electrical power to the analyzer.
The control PC that forms part of the analzyer is configured to automatically load the necessary gas sensor software and configuration files.
The software automatically starts the start-up sequence. After a few seconds, the
Gas Sensor Main Screen appears on the display controller.
Rosemount CT5100 65
Start-up procedure Reference Manual
May 2020 00809-0100-4511
Figure 7-2: Gas Sensor Main Screen
If the screen does not appear, report the fault to the maintenance organization.
11. Start up the system for venting the sample gas exiting the analyzer.
12. Start up the gas handling system that conditions the sample gas before it is fed into the analyzer.
At the end of the start-up procedure, that gas measurements initially appear as 0.00 ppm until the first readings are taken. After a few seconds, the initial gas concentrations are displayed.
The start-up procedure is now complete.
66 Emerson.com/Rosemount
Reference Manual Start-up procedure
00809-0100-4511 May 2020

7.4 Front panel controls and indicators

Configure the analyzer from the control display located on the front panel (Figure 7-3).
Figure 7-3: Front Panel
A. Display controller
NOTICE
ON/OFF circuit breaker
There are no ON/OFF switches on the analyzer. A customer-provided external circuit breaker controls the application of electrical power to the analyzer.
The circuit breaker is a simple two-pole ON/OFF circuit breaker that must be set to ON to permit the safe operation of the analyzer.
The display controller primarily controls operation of the analyzer (Figure 7-4).
Rosemount CT5100 67
Start-up procedure Reference Manual
May 2020 00809-0100-4511
Figure 7-4: Display Controller
WARNING
Failure to follow these instructions may cause warranty invalidation, property damage, and/or personal injury or death.
DO NOT, under any circumstances, press the top two buttons on the display controller simultaneously. This will override the system code supplied with the analyzer and display the default program screen for the controller.
This will lock the software controlling the analyzer requiring the analyzer to shut down externally and restarted risking a possible corruption of the software and loss of the analyzer operation during the shutdown and restart process.
It is essential that only trained, qualified personnel operate the controls on the analyzer.
Figure 7-5: Default Program Screen
68 Emerson.com/Rosemount
Reference Manual Operating the analyzer
00809-0100-4511 May 2020

8 Operating the analyzer

8.1 Introduction

This section describes the normal operation of the analyzer.
NOTICE
The gas concentrations shown in the following screenshots may be different from those shown in your particular analyzer. The screenshots indicate the functionality of the software, which is the same regardless of the gases or gas concentrations being measured.
WARNING
BURN HAZARD
Some parts of the analyzer may be heated to 374 °F (190 °C). All components of the analyzer are hot unless it has been switched off and allowed to cool down. Personal injury and/or damage to property may result if these safety precautions are not observed. These precautions are especially important when working at heights.
To prevent burns, do not touch any of the hot parts.
Before fitting, removing, or performing any maintenance on the analyzer, ensure that it has been switched off and allowed to cool for at least two hours. Before performing any maintenance on or in the vicinity of the analysis cell, allow the analyzer to cool for at least twelve hours, as the analysis cell is insulated against heat loss.
When handling the analyzer, always use suitable protective gloves.
If you receive a burn, seek medical treatment immediately.

8.2 Normal operation

The analyzer is designed for long term continuous operation, and therefore its normal state is to be switched on and performing gas measurements. The analyzer is usually only switched off for maintenance.
During normal operation, either the Gas Sensor Main screen (Figure 8-1) or the Pressure and Temperature screen (Figure 8-2) is shown on the display controller. To toggle between these two screens, press PAGE..
Rosemount CT5100 69
Operating the analyzer Reference Manual
May 2020 00809-0100-4511
Figure 8-1: Gas Sensor Main Screen
Figure 8-2: Pressure and Temperature Screen
70 Emerson.com/Rosemount
Reference Manual Shutdown procedure
00809-0100-4511 May 2020

9 Shutdown procedure

9.1 Safety precautions

CAUTION
EQUIPMENT DAMAGE
Damage to the analyzer may result from a failure to follow this procedure.
Always follow the shutdown procedure.
The analyzer normally operates continuously. It should only be necessary to shut down the analyzer in the following circumstances:
In order to perform repairs or maintenance on the analyzer
When the analyzer has to be switched off as part of a plant shutdown or plant
maintenance
When the analyzer is switched off for re-calibration
Use the display controller to perform the shutdown procedure. Refer to Display controller for the display controller navigation instructions.
NOTICE
The gas concentrations shown in the following screenshots may be different from those shown in your particular analyzer. The screenshots indicate the functionality of the software, which is the same regardless of the gases or gas concentrations being measured.
WARNING
BURN HAZARD
Some parts of the analyzer may be heated up to 374 °F (190 °C). All parts of the analyzer are hot unless it has been switched off and allowed to cool down. Personal injury and/or damage to property may result if these safety precautions are not observed. These precautions are particularly important when working at heights.
To prevent burns, do not touch any of the hot parts.
Before fitting, removing, or performing any maintenance on the analyzer, ensure that it has been switched off and allowed to cool for at least two hours. Before performing any maintenance on, or in the vicinity of, the analysis cell, allow the analyzer to cool for at least twelve hours, as the analysis cell is insulated against heat loss.
When handling the analyzer, always use suitable protective gloves.
If you receive a burn, seek medical treatment immediately.
Rosemount CT5100 71
Shutdown procedure Reference Manual
May 2020 00809-0100-4511

9.2 Shutdown procedure

WARNING
EXPLOSION HAZARD
Unauthorized operation of the gas handling system when maintenance is being performed on the analyzer or on its associated pipes/hoses may result in gas being released, causing fire or explosion. Failure to lock out the gas handling system may cause death.
Always lock out the gas handling system when shutting down the analyzer.
Procedure
1. Shut down the gas handling system that conditions the sample gas and feeds it to
the analyzer. Always lock-out the gas handling system to prevent its unauthorized operation during maintenance, which may cause an escape of gas.
WARNING
EXPLOSION HAZARD
Failure to vent sample gas may cause death.
Vent the sample gas in the system to prevent fire or explosion during maintenance and to prevent damage to the analyzer during shutdown.
Purge the sample gas in the pipes leading to the analyzer to prevent hazards to personnel during maintenance.
Purge the sample gas in accordance with safe working procedures for the site. Allow the analyzer and system for returning the sample gas to run for five
minutes to allow any sample gas in the analyzer to be returned to the exhaust.
2. Purge any sample gas in the pipe/hose from the gas handling system to the analyzer
using factory air or nitrogen supply.
WARNING
HIGH PRESSURE GAS AND AIR
The calibration gas supply and compressed air supply operate at a pressure that can cause injury (e.g. damage to eyes and skin punctures from debris blown by the high pressure gas or compressed air).
3. Turn off the gas used to purge the analyzer.
72 Emerson.com/Rosemount
Reference Manual Shutdown procedure
00809-0100-4511 May 2020
4. Press MENU on the display controller in either the Gas Sensor Main screen
(Figure 9-1) or the Pressure and Temperature screen (Figure 9-2).
Figure 9-1: Gas Sensor Main Screen
Figure 9-2: Pressure and Temperature Screen
The Main menu (Figure 9-3) opens.
Rosemount CT5100 73
Shutdown procedure Reference Manual
May 2020 00809-0100-4511
Figure 9-3: Main Menu
A. BACK button B. SELECT button C. Scroll up button
D. HELP button
E. Scroll down button
5. On the display controller, select System as shown in Figure 9-3.
6. Press SEL (select).
The System screen opens. Refer to Figure 9-4.
Figure 9-4: System Screen
A. BACK button B. SELECT button C. Scroll up button
D. HELP button
E. Scroll down button
74 Emerson.com/Rosemount
Reference Manual Shutdown procedure
00809-0100-4511 May 2020
7. Use scroll up (C) and scroll down (E) to select SHUTDOWN as shown in Figure 9-4.
Then press SEL (select). The Shutdown screen (Figure 9-5) opens.
Figure 9-5: Shutdown Screen
8. Press YES.
The analyzer shuts down.
9. Set the external secondary circuit breaker to Off. Lock-out and tag-out the
secondary circuit breaker.
10. Set the external main circuit breaker to Off. Lock-out and tag-out the main circuit breaker.
Rosemount CT5100 75
Shutdown procedure Reference Manual
May 2020 00809-0100-4511
76 Emerson.com/Rosemount
Reference Manual Gas calibration
00809-0100-4511 May 2020

10 Gas calibration

10.1 Required tools

The gas concentrations measured by the analyzer can be validated against a known sample gas or calibrated to match it by using the following gas calibration procedure.
To calibrate the analyzer, you need the following items:
Nitrogen gas of instrument gas purity for use as a zero calibration gas
Suitable span calibration gases for each gas measured
Gas bottle pressure regulators
Interconnecting hoses to connect the gas bottles to the analyzer
A T-piece and excess flow line
NOTICE
In the case of gases, such as H2O, for which it is not normally possible to obtain calibrated gas cylinders, it will usually be measured by the same laser as some other gas. Validating the other gases measured by the analyzer (particularly any which are measured by the same laser as H2O) can demonstrate that the system is functioning correctly, meaning there is no need to calibrate the water measurement directly.
If you need to calibrate (e.g., for legal requirements), you can use a water vapor generator to supply a known concentration of water vapor.
The validation procedure is essentially the same as the calibration procedure, except that it does not adjust the analyzer measurements to match the calibration standard.
WARNING
EQUIPMENT DAMAGE
Failure to follow may cause warranty invalidation, property damage, and/or personal injury or death.
DO NOT, under any circumstances, press the top two buttons on the display controller simultaneously. This will override the system code supplied with the analyzer and display the default program screen for the controller.
This will lock the software controlling the analyzer requiring the analyzer to shut down externally and restart, risking a possible corruption of the software and loss of the analyzer operation during the shutdown and restart process.
It is essential that only trained, qualified personnel operate the controls on the analyzer.
Rosemount CT5100 77
Gas calibration Reference Manual
May 2020 00809-0100-4511
Figure 10-1: Default Program Screen

10.2 Main menu calibration routines

Calibration is performed under the control of the analyzer software, using calibration routines built into the software.
Access calibration functions through the Main menu (Figure 10-4).
Procedure
1. To get the Main menu, press MENU on the Gas Sensor Main screen (Figure 10-2) or Pressure and Temperature screen (Figure 10-3).
NOTICE
Press BACK to cancel the procedure and return to the previous screen.
Figure 10-2: Gas Sensor Main Screen
78 Emerson.com/Rosemount
Reference Manual Gas calibration
00809-0100-4511 May 2020
Figure 10-3: Pressure and Temperature Screen
Figure 10-4: Main Menu
Use the Main menu (Figure 10-4) to access the software routines and screens necessary for this process.
2. Use the middle buttons (
Rosemount CT5100 79
and ) to scroll up or down to select GAS SERVICE.
Gas calibration Reference Manual
May 2020 00809-0100-4511
3. Press SEL.
Figure 10-5: Gas Service (PER GAS)
4. Use the middle buttons ( and ) to scroll up or down to select PER GAS.
5. Press SEL.
Figure 10-6: Select Gas (NO)
6. Use the middle buttons ( and ) to scroll up or down to select the desired gas (e.g., NO).
7. Press SEL.
80 Emerson.com/Rosemount
Reference Manual Gas calibration
00809-0100-4511 May 2020
8. To check or change the current offset value, select ZERO OFFSET and press CAL to view.
Figure 10-7: Action
Figure 10-8: Zero Offset Screen
9. Use the LEFT, RIGHT, INCR (increase), and DECR (decrease) buttons to enter the desired offset (-0.587313 in Figure 10-8).
10. Press OK to confirm.
The factory default offset is zero, and under normal circumstances, the offset should never be a positive number.
If it is shown as positive, press LEFT until the space to the left of the number is highlighted, then select INCR to change it to a minus (–) sign.
Rosemount CT5100 81
Gas calibration Reference Manual
May 2020 00809-0100-4511
11. To check or change the current span value, select SPAN FACTOR and press CAL to view.
Figure 10-9: Input Type
Figure 10-10: Span Factor Screen
12. Use the LEFT, RIGHT, INCR (increase), and DECR (decrease) buttons to enter the desired offset (1.06 ppm in Figure 10-10).
13. Press OK to confirm.
The factory default span factor is 1, and under normal circumstances, the offset should be between 0.8 and 1.2.
10.2.1
82 Emerson.com/Rosemount
Zero calibrate an individual gas
Procedure
1. To get to the Main menu, press MENU on the Gas Sensor Main screen (Figure 10-2) or the Pressure and Temperature screen (Figure 10-3).
Note
To cancel the procedure, press BACK to return to the previous screen.
Reference Manual Gas calibration
00809-0100-4511 May 2020
2. Use the middle buttons ( and ) to scroll up or down to select GAS SERVICE. Press SEL.
Figure 10-11: Main Menu (GAS SERVICE)
3. Use the middle buttons ( and ) to scroll up or down to select PER GAS. Press SEL.
Figure 10-12: Gas Service Menu (PER GAS)
Rosemount CT5100 83
Gas calibration Reference Manual
May 2020 00809-0100-4511
4. Use the middle buttons ( and ) to scroll up or down to select the desired gas (e.g., NO). Press SEL.
Figure 10-13: Select Gas (NO)
5. Use the middle buttons ( and ) to scroll up or down to select RUN ZERO.
Figure 10-14: Action Screen for NO (RUN ZERO)
84 Emerson.com/Rosemount
Reference Manual Gas calibration
00809-0100-4511 May 2020
6. Press CAL to calibrate, or VER to verify.
Figure 10-15 displays as the calibration progresses.
Figure 10-15: Calibration
Once complete, the program will automatically open the Result screen .
Figure 10-16: Result Screen
Rosemount CT5100 85
Gas calibration Reference Manual
May 2020 00809-0100-4511
7. Press OK.
The Verify next screen displays.
Figure 10-17: Verify Next Screen
8. Press DONE to finish.
To calibrate additional gases, select GAS and repeat Step 1 through Step 7.
10.2.2
To change types, select TYPE.
This completes the process, and the display will return to the Main menu.
Span calibrate an individual gas
Procedure
1. Gas Sensor MainTo get to the Main menu, press the MENU button on the screen (Figure 10-2) or the Pressure and Temperature screen (Figure 10-3).
Note
To cancel the procedure and return to the previous screen, press BACK.
86 Emerson.com/Rosemount
Reference Manual Gas calibration
00809-0100-4511 May 2020
2. Use the middle buttons ( and ) to scroll up or down to select GAS SERVICE. Press SEL.
Figure 10-18: Main Menu (GAS SERVICE)
3. Use the middle buttons ( and ) to scroll up or down to select PER GAS. Press the SEL button.
Figure 10-19: Gas Service (PER GAS)
Rosemount CT5100 87
Gas calibration Reference Manual
May 2020 00809-0100-4511
4. Use the middle buttons ( and ) to scroll up or down to select the desired gas (e.g. NO). Press SEL.
Figure 10-20: Select Gas (NO)
88 Emerson.com/Rosemount
Reference Manual Gas calibration
00809-0100-4511 May 2020
5. Use the middle buttons ( and ) to scroll up or down. Select RUN SPAN. Press CAL to calibrate, or VER to verify.
Figure 10-21: Action (RUN SPAN)
The Calibration screen show as the calibration progresses.
Figure 10-22: Calibration
Rosemount CT5100 89
Gas calibration Reference Manual
May 2020 00809-0100-4511
6. Using the LEFT, RIGHT, INCR (increase), and DECR (decrease) buttons, enter the cylinder concentration (e.g., 25 ppm).
Figure 10-23: Span Input
7. Press OK to confirm.
Once complete, the program will automatically open Span input For NO (Figure 10-23).
NOTICE
If an excessively high or low span factor is calculated (<0.7 or > 1.3 for example,) and if the calibration gas is a different concentration from that set above, it will be rejected and the previous span factor will be displayed and will continue to be used.
8. Press OK. The Result screen displays.
Figure 10-24: Result
90 Emerson.com/Rosemount
Reference Manual Gas calibration
00809-0100-4511 May 2020
9. Press DONE to finish.
Refer to Figure 10-25.
Figure 10-25: Calibration Complete
This completes the process, and the display will return to the Main menu.
10.2.3
Calibrate all gases (zero and span)
On systems with valve control, it is possible to run a calibration on all gases. The valves will sequentially switch through each of the span and zero gases.
This option is not practical to use on systems which require manual valve switching.
In order to set the cylinder concentration when using a new span gas cylinder, it is necessary to perform a single gas calibration as described in Span calibrate an individual
gas. This concentration will then be stored for future calibrations.
Procedure
1. To get to the Main menu, press MENU on the Gas Sensor Main screen (Figure 10-2) or the Pressure and Temperature screen (Figure 10-3).
NOTICE
To cancel the procedure and return to the previous screen, press BACK.
Rosemount CT5100 91
Gas calibration Reference Manual
May 2020 00809-0100-4511
2. Use the middle buttons ( and ) to scroll up or down to select GAS SERVICE. Press SEL.
Figure 10-26: Main Menu (GAS SERVICE)
92 Emerson.com/Rosemount
Reference Manual Gas calibration
00809-0100-4511 May 2020
3. Use the middle buttons ( and ) to scroll up or down to select CALIBR ALL. Press SEL.
Figure 10-27: Gas Service (CALIBR ALL)
The system will work through the zero and span calibration process for all readings. The Please wait screen will be displayed. This takes approximately one minute per gas depending on the purge time.
Figure 10-28: Calibration
The Finished screen displays when the zero and span calibration processes complete.
Rosemount CT5100 93
Gas calibration Reference Manual
May 2020 00809-0100-4511
Figure 10-29: Calibration (Finished)
4. Press BACK.
This completes the process, and the display returns to the Main menu.
10.2.4
Change the automatic calibration settings
Procedure
1. To get to the Main menu, press MENU on the Gas Sensor Main screen (Figure 10-2) or the Pressure and Temperature screen (Figure 10-3).
NOTICE
To cancel the procedure and return to the previous screen, press BACK.
2. Use the middle buttons ( and ) to scroll up or down to select GAS SERVICE. Press SEL.
Figure 10-30: Main Menu (GAS SERVICE)
94 Emerson.com/Rosemount
Reference Manual Gas calibration
00809-0100-4511 May 2020
3. Use the middle buttons ( and ) to scroll up or down to select CALIBR PARAMS or VERIFY PARAMS for validation settings). Press SEL.
Figure 10-31: Gas Service (CALIBR PARAMS)
This section applies to systems configured to perform automatic calibration at timed intervals only.
4. To adjust the frequency of automatic calibration, use the middle buttons ( and ) to scroll up or down
5. Select AUTO PERIOD.
6. Press SEL.
Figure 10-32: Calibr param (AUTO PERIOD)
Rosemount CT5100 95
Gas calibration Reference Manual
May 2020 00809-0100-4511
7. To change the time and the units (days, hours, minutes and seconds) between calibrations, use the LEFT, RIGHT, INCR (increase), and DECR (decrease) buttons to set the number and time interval.
Figure 10-33: (Calibration) Period
8. Press OK.
9. To adjust the length of time for the span gas to flush the cell, use the middle buttons ( and ) to scroll up or down and select PURGE TIME.
Figure 10-34: Calibr param (PURGE TIME)
10. Press SEL.
96 Emerson.com/Rosemount
Reference Manual Gas calibration
00809-0100-4511 May 2020
11. To change the time between opening the valve and performing the calibration, use the LEFT, RIGHT, INCR (increase), and DECR (decrease) buttons to set the number and time interval.
Figure 10-35: Purge Time
12. Press OK.
Rosemount CT5100 97
Gas calibration Reference Manual
May 2020 00809-0100-4511
98 Emerson.com/Rosemount
Reference Manual Troubleshooting and diagnostics
00809-0100-4511 May 2020

11 Troubleshooting and diagnostics

11.1 Use the Built-in-Self-Test (BIST) fault diagnostics

The analyzer has a Built-In-Self-Test (BIST) function that can be used to perform failure diagnosis of some functions in situations where there is a fault other than a complete failure of the equipment. To access and control the BIST, use the display controller mounted on the door of the electrical compartment.
The BIST runs in the background when the analyzer is operating and continuously monitors the analyzer for faults.
If the BIST detects a fault, the STATUS display in the bottom left-hand corner of the Gas Sensor Main screen (Figure 11-1) will change from OK to flashing the word FAILURE.
Figure 11-1: Gas Sensor Main Screen
A. FAILURE message
B. FAILURE button
Similarly, if the BIST detects a fault and the Pressure and Temperature screen (Figure 11-2) is being displayed, the STATUS display in the bottom left-hand corner of the screen will change from OK to flashing the word FAILURE.
Rosemount CT5100 99
Troubleshooting and diagnostics Reference Manual
May 2020 00809-0100-4511
Figure 11-2: Pressure and Temperature Screen
A. FAILURE message
B. FAILURE button
Procedure
1. On either the Gas Sensor Main screen (Figure 11-1) or the Pressure and Temperature screen (Figure 11-2), press FAILURE button to go to the Faults Menu screen (Figure
11-5). Controls and display controller provides details of how to use the controls on the
display controller and functions of the BACK and HELP buttons.
2. You can also access the FAULTS menu through the Main Menu (Figure 11-3) as follows:
100 Emerson.com/Rosemount
Loading...