All possible care has been taken in the preparation of this publication, but Emerson and its agents and distributors accept no
liability for any inaccuracies that may be found. This manual reflects the state of the product at the issue date below, but further
enhancements while in service may mean that the manual does not reflect your particular system.
Emerson reserves the right to make changes without notice both to this publication and the products which it describes.
No part of this publication may be reproduced, stored in a retrieval system, or transmitted in any form or by any means electronic,
mechanical, photocopying, recording, or otherwise without the express prior written permission of the copyright holder.
If you require additional technical assistance, request help from cascade.support@emerson.com or Emerson distribution partners.
Send general inquiries to cascade.support@emerson.com.
All trademarks used within this document are the property of their respective owners.
Preliminary information
This section details important user information for the Rosemount CT5100 Continuous Gas Analyzer.
Important
Users must read, understand, and comply with the following information before proceeding.
All users, installers, operators, and maintainers must be familiar with operating the analyzer. To install, start up, operate, maintain,
and service the analyzer in a safe manner, it is MANDATORY to read all additional documents shipped with the analyzer. The
following information is also available and/or referenced in the Rosemount CT5100 Quick Start Guide: 00825-0100-4511.
Save all instructions for future use. Contact your local service center or sales office when missing documents.
Authorized personnel
In-depth specialist knowledge is an absolute requirement for working with and on the analyzer. Personnel installing, operating,
servicing, and maintaining the analyzer must be instructed, trained, qualified, and authorized personnel of the operating company
for hazardous areas and the manufacturer. It is the responsibility of the operating company to:
• Train staff
• Observe safety regulations
• Follow the safety instructions and procedures in the product manual
Operators must:
• Have been trained
• Have read and understand all relevant sections of the product manual before commencing work
• Know the safety mechanisms and regulations
WARNING
To avoid explosions, loss of life, personal injury, and damage to this equipment and on-site property, do not install, operate,
maintain, or service this analyzer before reading and understanding this document and receiving appropriate training.
2
Important instructions
Important
All users must read this page before proceeding!
Emerson designs, manufactures and tests its products to meet many national and international standards. The Rosemount
CT5100 is a sophisticated technical product, and to ensure it continues to operate as designed and within normal specifications it
MUST be installed, used, and maintained correctly. The following instructions MUST be adhered to and integrated into your safety
program when installing, using, and maintaining Emerson products.
• Failure to follow the proper instructions may cause:
— Loss of life
— Personal injury
— Damage to property
— Damage to this analyzer
— Warranty invalidation
• Read all instructions prior to installing, operating, and servicing the product.
• If you do not understand any of the instructions, contact your Emerson representative for clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper installation, operation, and maintenance of the product.
• Install your equipment as specified in the installation instructions of the appropriate Reference Manual and in accordance with
applicable local and national codes.
• Connect all products to the proper electrical and pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
• When replacement parts are required, ensure that qualified people use replacement parts specified by Emerson.
• Unauthorized parts and procedures can affect the product’s performance; place the safe operation of your process at risk, and
VOID YOUR WARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
2012/19/EUWaste Electrical and Electronic Equipment (WEEE) Directive
USA 21 CFR 1040.1Laser products
NEC 505National Electrical Code (issued by ANSI: American National Standards
BS EN 60825-1:2007Safety of laser products. Equipment classification and requirements
BS EN 61010-1 2010
IEC 61010-1 2010
BS EN 61326-1: 2013Electrical equipment for measurement, control, and laboratory use.
Associated publications
Rosemount CT5100 Quick Start Guide
Compliance approvals
Important
The Rosemount CT5100 analyzer is designed for use in Non Hazardous areas ONLY.
This product complies with USA 21 CFR 1040.10.
This product is designed and manufactured under an approved quality management
system to ISO 9001: 2015.
Institute and NFPA 70: National Fire Protection Association)
(identical to IEC 608250-1 2007)
Safety requirements for electrical equipment for measurements,
control, and laboratory use. General requirements
EMC requirements. General requirements
Waste disposal
Emerson and the Rosemount CT5100 have satisfied the requirements for applying
the CE marking to the Rosemount CT5100 Gas Analyzer.
This equipment meets all requirements of the EMC and Low Voltage directives.
Do not dispose of measuring tools into household waste.
Only for EC countries:
In accordance with European Directive 2012/19/EU for Waste Electrical and
Electronic Equipment and its implementation into national right, measuring tools
that are no longer usable must be collected separately and disposed of in an
environmentally correct manner.
4
Safety and information notices
DANGER
WILL CAUSE DEATH
Failure to follow this warning will result in death or serious injury to personnel.
WARNING
DANGER TO PERSONNEL
Failure to follow this warning may result in death or serious injury to personnel.
CAUTION
MAY CAUSE DAMAGE TO EQUIPMENT
Failure to follow this warning may result in damage to the equipment.
NOTICE
Important or tip messages will appear in this format.
Safety precautions
Important
The precautions in this manual MUST NOT be changed amended or removed. All authorized users, installation, operation and
maintenance personnel, must observe the following safety precautions and warnings.
The Rosemount CT5100 analyzer is designed for use in Non Hazardous areas ONLY.
WARNING
ELECTRIC SHOCK
The analyzer operates using mains voltage, which may cause death or serious injury to personnel. Death, personal injury, and/or
damage to persons and/or property may result if this is not observed.
Confirm that the circuit breakers are set to OFF and locked out and tagged out before removing the top cover or opening the front
cover. The analyzer must be earthed.
Only trained, qualified personnel may install and connect power and signal cables. The installation/connection must be in
accordance with all legislative requirements and applicable standards.
Only qualified personnel, familiar with potential risks, should install the analyzer.
DANGER
EXPLOSION HAZARD
Failure to observe this precaution will cause death, personal injury, and/or damage to persons.
The sample gas in the system must be vented to prevent fire or explosion during maintenance and to prevent damage to the
analyzer during prolonged shutdowns.
The sample gas in the pipes leading to the analyzer must be purged for a minimum of 2 minutes and 30 seconds at a minimum
flow rate of 280 L/m at 1.5 ± 0.5 bar to prevent hazards to personnel during maintenance.
Purge the sample gas in accordance with the safe working procedures for this site.
5
WARNING
FLAMMABLE SUBSTANCES
Some parts of the analyzer may reach temperatures of 374 °F (190 °C) and may present an ignition source. The interior of a
analyzer is always hot unless it has been switched off and allowed to cool down. A fire may result if this precaution is not observed.
Exercise care when using oil, paint, cleaning rags, or other flammable substances near the analyzer.
WARNING
TRANSPORTATION HAZARD
The analyzer weighs 117 lb. (53 kg) and should always be lifted and moved using suitable lifting/moving equipment.
Handle the analyzer with caution during unpacking, installation, maintenance, and transport to prevent crushing of hands, feet, or
other body parts.
Wear suitable protective gloves and protective footwear. When preparing the analyzer for transport by air, road, or rail, safeguard
the analyzer against movement or break-away during transport by securely strapping it in place.
Use safety approved lifting equipment. Ensure that the equipment is tested, meets the lifting ratings for the weight of the
equipment, and is in good operational condition.
WARNING
FIRE AND EXPLOSION
Failure to observe this warning could cause an explosion or potentially hazardous situation, which if not avoided, may cause death,
personal injury, and/or damage to persons and/or property.
Do not open the analyzer's electrical compartment unless the atmosphere in the area is known to be below the ignitable
concentration of combustible gases or materials, or unless all equipment within the protected enclosure is de-energized .
The protective gas supply valve must be kept open unless the atmosphere in the area is known to be below the ignitable
concentration of combustible gases or materials or unless all equipment within the protected enclosure is de-energized.
Always lock out the gas handling system when shutting down the analyzer.
After opening the enclosure, do not restore power until the enclosure has been purged for a minimum of 2 minutes and 30
seconds at a minimum flow rate of 280 L/m at 1.5 ± 0.5 bar.
DO NOT operate the analyzer with doors or covers open.
Refer to local regulations as this may require a competent hot work supervisor to issue a hot work permit.
When the analyzer is out of order or if the pressurization unit shuts off due to a failure, all inputs and outputs connected to
external equipment MUST be shut off.
This will ensure that no hazardous voltages are present within the analyzer enclosure when not pressurized.
Only properly trained personnel who understand the contents of all applicable manuals and related instructions should start up
the analyzer.
Do not keep operating the analyzer if the enclosure shows permanent deformations after performing the overpressure test.
Use only replacement parts and components authorized by Emerson.
All replacement parts and components must be certified and approved for use in hazardous areas.
The analyzer contains a battery for data backup purposes.
Under normal operating conditions, there is no need to replace the battery during the analyzer life time. Battery replacement
MUST only be conducted by Rosemount Customer Care personnel. It is NOT a customer serviceable item.
6
WARNING
FIRE, BURN, AND OPTICAL RADIATION EXPOSURE HAZARD
Electrical shock, thermal burns, or loss of vision may occur. Failure to observe this warning could cause an explosion or potentially
hazardous situation, which if not avoided, may cause death, personal injury, and/or damage to persons and/or property.
Operators and service personnel do not have access to the laser/electrics or upper cell compartments for general maintenance or
service.
WARNING
BURNS
Personal injury and/or damage to property may result if these precautions are not observed. These precautions are particularly
important when working at heights.
Some parts of the analyzer may be heated to 374 °F (190°C). To prevent burns, do not touch any of the hot parts. All parts of a
analyzer are always hot unless it has been switched off and allowed to cool down.
Before fitting, removing, or performing any maintenance on the analyzer, ensure that it has been switched off and allowed to cool
for at least two hours. Before performing any maintenance on, or in the vicinity of, the analysis cell, allow the analyzer to cool for
at least 12 hours as the analysis cell is insulated against heat loss.
When handling the analyzer, always wear suitable protective gloves.
If you receive a burn, seek medical treatment immediately.
WARNING
HAZARDOUS SUBSTANCES
Failure to observe this warning could cause a potentially hazardous situation, which if not avoided, may cause death, personal
injury, and/or damage to persons and/or property.
The analyzer may contain hazardous substances. Always handle the analyzer assemblies and components with extreme caution.
Wear personal protective equipment (PPE) when handling the equipment.
Gas handling components within the analyzer contain particulate matter residue from the sample gases. Over the life of the the
analyzer, the concentration of particulate matter will become enriched within the gas handling components. When performing
repairs and maintenance on the the analyzer:
— Handle used gas handling components with extreme caution.
— Avoid direct skin contact with used gas handling components.
— Do not smoke, drink, or eat in the work area.
— Wear goggles or eye shields.
— Wear a suitable face mask to protect against inhalation of particulate matter.
— Do not wet fingers, eyes, or any exposed skin.
— Pack used gas handling components for disposal in sealed packaging and label them Contaminated.
— Dispose of contaminated items as hazardous material in accordance with applicable local, national, or international health
and safety regulations and pollution regulations.
Take special care to ensure that the sample gas return port either returns the sample gas to the product stream or discharges the
sample gas to a location that will not cause a hazard.
7
WARNING
OPTICAL RADIATION EXPOSURE HAZARD
There are three types of laser that may be included in the Rosemount CT5100: Quantum Cascade Lasers (QCLs), Interband
Cascade Lasers (ICLs), and diode lasers. The lasers within the analyzer are Class 1. The characteristics of the lasers contained within
the analyzer are given in the table below.
The emitted laser light is invisible (mid-infrared), and the combined laser powers are sufficiently low at the first accessible aperture
that the unprotected eye will not be damaged. This class is eye safe under all operating conditions.
It is, however, possible to cause damage to the eye through not following correct procedures. Do not look at the laser with any
kind of magnifier or optical measuring device.
ParameterQCLICLDiodeComment
Operation modePulsedPulsedPulsedN/A
Lasers per system1 - 61- 61- 6Maximum of 6 lasers per system
Power< 5 mW< 5 mW< 5 mWCombined power of QCL at first accessible
Pulse duration< 1 µs< 1 µs< 5 µsN/A
Pulse repetition
frequency
Duty cycle< 5 %< 5 %< 25 %N/A
The combined power of the QCL, ICL, and diode lasers at the first accessible aperture is < 9.62 mW.
The analyzer has warning labels in appropriate positions according to USA 21 CFR 1040.10. The location of laser safety labels on
the analyzer is specified in Safety and system labels and annotation.
The use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous
radiation exposure.
< 100 kHz< 100 kHz< 100 kHzN/A
aperture: < 9.62 mW
WARNING
HIGH PRESSURE GAS AND AIR
The calibration gas supply and compressed air supply operate at a pressure that can cause injury, e.g., damage to eyes and skin
punctures from debris blown by the high pressure gas or compressed air.
Always lock off or tag out the calibration gas supply and compressed air supply when shutting down the analyzer.
The maximum gas pressure valve must not exceed 100 psig (690 kPa).
WARNING
EXPLOSIONS HAZARD DUE TO ELECTROSTATIC DISCHARGE
In the event of a sudden discharge from electrostatically charged devices or individuals, there is a risk of an explosion. Failure to
observe this warning and or follow safety instructions could cause an explosion or potentially hazardous situation, which if not
avoided, could result in death or serious injury.
Take suitable measures to ensure that no electrostatic discharge can build up in the explosions risk area.
Clean the device surface by gently wiping it with a damp or antistatic cloth only.
8
WARNING
HAZARD BY WRONG INPUT VOLTAGE
Applying a rated voltage other than specified on the analyzer´s nameplate label may cause an explosion, injury, or damage to the
installation. Failure to observe this warning could cause an explosion or potentially hazardous situation, which if not avoided, may
cause death, personal injury, and/or damage to persons and/or property.
This type of analyzer is always setup for a specific rated input voltage; see nameplate label.
Ensure the voltage at site of installation meets the rated analyzer input voltage.
WARNING
LOOSE ITEMS
Failure to observe this warning could cause a potentially hazardous situation, which if not avoided, could result in death or serious
injury.
Do not place any loose items on top of the system or inside the compartments when doors / covers are open.
Confirm that all loose items, tools, and equipment are removed from compartments before closing doors and covers.
WARNING
MAINTENANCE/MODIFICATIONS
Failure to observe this warning could cause a potentially hazardous situation, which if not avoided, could result in death or serious
injury.
On completion of any maintenance and or modifications verify:
— All tools and equipment are removed.
— No contamination (water/dust) is in the compartments.
— Analyzer is wiped clean.
— Vents are clear and not obstructed.
— Verify that system is in a safe state for operation.
WARNING
Physical access
Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of end users’ equipment. This could
be intentional or unintentional and needs to be protected against.
Physical security is an important part of any security program and fundamental to protecting your system. Restrict physical access
by unauthorized personnel to protect end users’ assets. This is true for all systems used within the facility.
CAUTION
EQUIPMENT DAMAGE
Failure to perform pre-system start-up checks may cause damage to equipment.
Do not power up or try to operate the analyzer unless it is physically secure and all electrical and pneumatic connections to the
analyzer are in place.
Before starting up the analyzer, ensure that electrical power, sample gas handling facilities, and any calibration gases that are
required are available to the analyzer.
Always follow the Start-up procedure.
Always follow the Shutdown procedure.
9
CAUTION
UNSERVICEABLE EQUIPMENT
If the pressure and temperature screen does not display measurements similar to those shown in Figure 8-1 and Figure 8-2, refer
to Troubleshooting and diagnostics.
CAUTION
EMC
This is a Class A product. In a domestic environment, this product may cause radio interference, in which case the user may be
required to take adequate measures.
As a general principle, if any optical component other than the cell assembly, the laser modules, and the detectors is
unserviceable, the analyzer must be repaired by Emerson. This is because the repair, replacement, and alignment of the optical
components requires the use of special optical test/calibration equipment and procedures.
Some faults can only be repaired by Emerson. Where an item is unserviceable, and no replacement procedure is given in this
manual, then the fault must be repaired by Emerson.
The Rosemount CT5100 Continuous Gas Analyzer, referred to hereafter as the Rosemount
CT5100, is an electronic sensor that uses laser spectroscopy to perform analysis of process
gas streams.
The function of the analyzer is to detect and measure up to 10 different types of gas at
concentrations ranging from parts per million (ppm) to percentage levels in the process
gas stream.
Important
The Rosemount CT5100 is designed for use in Non Hazardous areas ONLY.
Note
This manual is intended for the personnel who install, operate, and maintain the
equipment.
1.2Customer information
This manual contains all the important information that must be followed to ensure the
correct operation and safety of personnel when operating the analyzer.
For information regarding installation, consult Install and the Quick Start Guide.
Emerson is committed to continuously improving its products and documentation. Every
effort will be made to include in the documentation any modifications by the
manufacturer. However, this document reflects the supplied analyzer at the revision date
on the front cover.
Should you require further information, or should particular problems arise that are not
covered in this manual, you can request additional help from Cascade Technical Support
(cascade.support@emerson.com) or Emerson distribution partners. Further contact
details for Emerson can be found on the back page of this manual.
1.3Safety precautions and conditions for safe use
WARNING
SAFE USE PRECAUTIONS
Before installing or performing any maintenance on the analyzer, read and understand the
safety information given in the preliminary information of this manual.
The analyzer described in this guide has been quality control tested and left the
manufacturer in pristine condition. To achieve the correct and safe operation of this
product, it must be transported, installed, operated, and maintained as described by the
manufacturer.
Rosemount CT510015
Introduction
May 202000809-0100-4511
All lasers used within the analyzer are Class 1. The emitted laser light is invisible (midinfrared) and the pulse duration so short that the unprotected eye will not be damaged.
The nature of the laser beam path and beam width further ensures that it should be
impossible to cause any eye damage. The analyzer has warning labels at appropriate
positions in accordance with USA 21 CFR 1040.10.
General safety notice/residual risk
Installation, operation, and maintenance of the analyzer must be in accordance with these
instructions.
When operated as intended and all applicable safety instructions are observed, an element
of risk will remain, including, but not limited to, the following:
• The emission of gases hazardous to health may be possible when all gas connections
have been correctly made.
• To avoid exposure to the dangers of residual risks, take particular care when installing,
operating, maintaining, and servicing the analyzer.
Reference Manual
1.4Qualified personnel
In-depth specialist knowledge is an absolute requirement for working with and on the
analyzer. Personnel installing, operating, servicing, and maintaining the analyzer must be
instructed, trained, qualified, and authorized personnel of the operating company for
hazardous areas and the manufacturer.
It is the operating company's responsibility to:
• Train staff
• Observe safety regulations
• Follow the safety instructions and procedures in the product manual
Operators must:
• Be trained
• Read and understand all relevant sections of the product manual before commencing
work
• Know the safety mechanisms and regulations
WARNING
To avoid explosions, loss of life, personal injury, and damage to this equipment and on-site
property, do not install, operate, maintain, or service this analyzer before reading and
understanding this reference manual and receiving appropriate training.
1.5Software version
The analyzer includes software that is used to control the operation of the analyzer. This
manual describes the software version as: 5.7.13.
BargPressure, in units of bars, above or below atmospheric pressure
BSBritish Standard
CCelsius
CDACompressed dry air
CEEuropean Conformity
Copyright
CFRCode of Federal Regulations
CGAContinuous Gas Analyzer
CH
4
CO
2
DCDirect current
DegDegree (temperature)
e.g.For example
ECEuropean Community
EMCElectromagnetic compatibility
EUEuropean Union
HrsHours
HzHertz
H2OWater
ICLInterband Cascade Laser
IECInternational Electro-technical Commission
in.Inches
IPIngress protection
Methane
Carbon dioxide
IPxxIngress protection (xx are numbers that define the protection level)
ISIntrinsically safe
ISOInternational Organization for Standardization
kThousand
kgKilogram
Rosemount CT510017
IntroductionReference Manual
May 202000809-0100-4511
AbbreviationDescription
kHzKilo hertz
LLiter
lb.Pound
LCDLiquid crystal display
LEDLight emitting diode
L/minLiters per minute
mMeter
3
m
mAMilliamp
MaxMaximum
mBarmilli-Bar
mbpsMegabits per second
Cubic meter
mgMilligram
3
mg/m
Milligram/cubic meter
Mid IRMid Infrared
minMinute
mmMillimeter
N
2
NEC
®
Nitrogen
National Electrical Code
NFPANational Fire Protection Association
nmNanometer
NH
3
Ammonia
NONitric oxide
NO
2
Nitrogen dioxide
N2ONitrous oxide
No.Number
O
2
Oxygen
PCPersonal computer
PMPreventative maintenance
ppmParts per million
psiPounds per square inch
QCLQuantum Cascade Laser
TDLTunable Diode Laser
TorrUnit of pressure defined as exactly 1/760 of a standard atmosphere
18Emerson.com/Rosemount
Reference ManualIntroduction
00809-0100-4511May 2020
AbbreviationDescription
UKASUnited Kingdom Accreditation Service
USAUnited States of America
USBUniversal serial bus
VVolt
VAVolt-ampere
VacVolt alternating current
VdcVolt direct current
WWatt
WEEEWaste electrical and electronic equipment
µmMicro-meter
Rosemount CT510019
IntroductionReference Manual
May 202000809-0100-4511
20Emerson.com/Rosemount
Reference ManualTheory of operation
00809-0100-4511May 2020
2Theory of operation
2.1Overview
The Rosemount CT5100 is a gas sensor system that can be configured to measure the
concentrations of multiple small molecules carried in the gas sample. The types of
molecules that are measured depend on the system configuration.
The analyzer can be configured to detect and measure up to 10 gases, with ranges varying
from volume to percent (%) volume levels. A detailed description of the system is given in
Detailed system specifications .
2.2Laser measurement principle
The analyzer can use up to six lasers to detect and measure gases. Each laser can measure
between one and three gases. Inside the laser, which is about the size of a pin head,
electrons cascade down a series of quantum wells, producing a photon at each step.
This cascade of electrons can produce between 20 and 100 photons per electron, giving
the layers a higher output power than traditional semi-conductor lasers.
The lasing wavelength of a laser is determined by adjusting the physical thickness of the
semiconductor layers, giving access to high power lasers covering the mid-infrared
spectral region. The lasers have no need for cryogenic cooling and have excellent spectral
quality in chirped mode and good tuneability.
2.3Gas concentration measurements
In the analyzer, gas concentrations are measured using mid-infrared optical absorption
spectroscopy. The laser light sources are operated to produce wavelength sweeps that
cover the absorption lines of the gases to be measured.
Sample gas, which may contain impurity gases that are to be detected and measured, is
conditioned and drawn into the analyzer. Inside the analyzer, the sample gas is fed into an
analysis cell, where the beams from the Quantum Cascade Lasers (QCLs) are passed
through the gas. The analysis cell contains a set of mirrors that bounce the light back and
forth many times, which lengthens the path of the lasers through the gas.
On exiting the analysis cell, the light is detected by a receiver unit. The variation in the
intensity of light in the vicinity of absorption lines for the gases being detected is
measured, and the concentration is determined using a comprehensive spectral fitting
routine.
Rosemount CT510021
Theory of operationReference Manual
May 202000809-0100-4511
22Emerson.com/Rosemount
Reference Manual
00809-0100-4511May 2020
Description
3Description
3.1Equipment purpose and role
The types of molecules that are measured depend on the system configuration.
Figure 3-1: Rosemount CT5100 Continuous Gas Analyzer
The analyzer can be configured to detect and measure up to 10 different gases, depending
on the combination of laser modules fitted.
3.2System overview
A complete Rosemount CT5100 system consists of a gas handling system, the analyzer,
and the associated interconnecting wiring and gas piping.
The customer MUST provide the gas handling system and interconnecting wiring and gas
piping.
The Rosemount CT5100 is supplied by Emerson. The gas handling system may be
provided by either the customer or Emerson, depending upon the specific installation. The
Rosemount CT510023
Description
Reference Manual
May 202000809-0100-4511
circuit breaker used to control the application of electrical power to the analyzer, the
interconnecting wires, and gas piping are provided by the customer.
In Figure 3-2, the items supplied by Emerson are colored blue; the items supplied by the
customer are colored purple. The green gas handling system may be provided by Emerson
or the customer.
Table 3-1 lists the main items of the system.
Figure 3-2: Rosemount CT5100 Installation
A. Gas handling system
B. Sample supply line
C. Sample return (exhaust) line
D. Rosemount CT5100
E. Electrical power
F. Two pole main isolator complete with RCD
G. Control center
H. Measurement data
The analyzer contains an optical system with multiple lasers and a series of optical
components that provide an optical path, a heated multi-pass analysis cell, and sample
and outlet ports that can be connected to a gas handling system and control and analysis
electronics. The number of lasers installed depends upon customer requirements. The
complete system operates at either 110 or 240 Vac 50/60 Hz supply.
The analyzer uses mid-infrared optical absorption spectroscopy to measure gas
concentrations. The light sources are Quantum Cascade Lasers (QCLs), which are operated
to produce wavelength sweeps that cover the absorption lines of the gases. The light from
each laser is routed through an optical path to the analysis cell, which provides
measurement of low concentrations of the subject gases. An external sample handling
system conditions the sample gas and draws it through the analysis cell. The light exits the
multi-pass analysis cell and is directed to a receiver in the analyzer. The variation in the
intensity of light in the vicinity of the absorption lines is measured, and the concentration
is determined using a comprehensive spectral fitting routine.
There is no sample conditioning provided within the analyzer; the sampled gas must be
brought within the parameters shown in Detailed system specifications before entering
the analyzer. Detailed characteristics of the analyzer are also given in in Table 4-1.
24Emerson.com/Rosemount
Reference ManualDescription
00809-0100-4511May 2020
Table 3-1: Main Items of the Rosemount CT5100 Installation
ItemName or descriptionSupplied byPart numberQuantityNotes
1Rosemount CT5100EmersonRosemount CT51001N/A
2Rosemount CT5100 software
package, version 5.x.x.
3Gas handling systemCustomer
4Heated gas sample line hoseCustomerCustomer choice1N/A
5Exhaust line hose (for sample
gas)
6Reference gas cylinders
(instrument gas) for
calibration purposes
7Pressure regulatorCustomerCustomer choice1 per gas
8Pneumatic T-pieceCustomerCustomer choice1Required for
9Excess flow lineCustomerCustomer choice1Required for
10Power cables to Rosemount
CT5100
11Cables from Rosemount
CT5100 to control center
EmersonN/A. Software is
embedded in the
system.
Customer choice or
(optionally by
Emerson)
CustomerCustomer choice1N/A
CustomerCustomer choiceDependent upon
CustomerCustomer choice1N/A
CustomerCustomer choice1N/A
Emerson
1Version
described in
manual
1N/A
N/A
number of gases
being measured
Required for
cylinder
calibration
calibration
calibration
12Main circuit breaker
complete with RCD
CustomerCustomer choiceN/AN/A
3.3Gas inputs and outputs
The analyzer has one gas input and one gas output. See Figure 3-3.
Rosemount CT510025
Description
Reference Manual
May 202000809-0100-4511
Figure 3-3: Gas Inlet and Outlet Connectors
A. Sample gas return port
B. Sample gas input port
C. Top cover of Rosemount CT5100 (cell compartment)
D. Laser/electrical compartment
Procedure
1. The gas sample that is to be measured for impurities enters the analyzer through
the sample gas input port located on top of the analyzer (see Figure 3-3.).
2. Once the gas sample has been examined for impurities, it is expelled from the
analyzer through the sample gas return port.
3. The sample gas supply line must be heated all the way to the sample gas input port
on the analyzer to prevent condensation forming in the line.
26Emerson.com/Rosemount
Reference ManualDescription
00809-0100-4511May 2020
WARNING
HAZARDOUS GASES
The product stream that the analyzer is examining may be hazardous even at low
concentrations.
Therefore, take special care to ensure that the sample gas return port either returns
the sample gas to the product stream or discharges the sample gas to a location
that will not cause a hazard.
3.4Connecting the electrical/electronic inputs and
outputs
CAUTION
If the equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
Rosemount CT510027
DescriptionReference Manual
May 202000809-0100-4511
Figure 3-4: Electrical/Electronic Connectors
A. Power entry point
B. Analog/digital entry point
C. Analog/digital entry point
Electrical/electronic signal connections to the analyzer are made through three electrical
entry points located on the underside of the analyzer, as shown in Figure 3-4. Use the
wiring diagram to make the electrical connections as shown in Engineering drawings and
Figure 3-5.
28Emerson.com/Rosemount
Reference ManualDescription
00809-0100-4511May 2020
Figure 3-5: Signal Cable Outputs
A. To mains input fuses
B. Mains input filter
C. To enclosure earth stud
D. To digital modules
E. To digital modules
F. To analog modules
G. To status relays
H. Digital output terminals
I. Analog or digital output terminals
J. Analog output terminals
K. Status outpput terminals
L. Mains input terminals
Table 3-2: Rosemount CT5100 System Wiring User Connections
TerminalFunction
1Sensor system supply (L)
2Sensor system supply (N)
3Earth
11Digital output 1
12Digital output 2
13Digital output 3
14Digital output 4
15Digital output 5
16Digital output 6
17Digital output 7
18Digital output 8
19Digital output 9
Rosemount CT510029
DescriptionReference Manual
May 202000809-0100-4511
Table 3-2: Rosemount CT5100 System Wiring User Connections (continued)
TerminalFunction
20Digital output 10
21Digital output 11
22Digital output 12
23Analog output 1
24Analog output 2
25Analog output 3
26Analog output 4
27Analog output 5
28Analog output 6
29Analog output 7
30Analog output 8
31Status output 1 (Check function)
32Status output 2 (Maintenance required)
33Status output 3 (Out of specification)
34Status output 4 (Failed)
01Analog or digital input/output
02Analog or digital input/output
03Analog or digital input/output
04Analog or digital input/output
05Analog or digital input/output
06Analog or digital input/output
Electrical power is applied to the analyzer through the power entry point, Figure 3-4.
The power supply is 110 or 230 Vac, 50/60 Hz ± 10%. AC to DC. Power converters inside
the analyzer automatically adjust in response to the input voltage level and ensure that the
correct DC voltage is available inside the unit. The analyzer is electrically protected by an
internal 5 A, 250 VA fast acting fuse on the instrumentation electrical supply line and an
internal 2 A, 250 VA fuse on the purge electrical supply line.
WARNING
Failure to follow this warning may result in personal injury and/or damage to persons
and/or property.
Make sure that the mains supply cable used is of a suitable rating for the unit power
requirements and is of a three core earthed construction.
30Emerson.com/Rosemount
Reference ManualDescription
00809-0100-4511May 2020
The digital outputs conduit (B and C) provides an Ethernet output from the instrument
that may be used for downloading data for failure diagnosis purposes or for downloading
data to the process control center.
The results of the gas analysis are output from the instrument through the 4-20 mA analog
or Modbus® outputs and sent to the process control center.
The optional digital outputs provide fault indications to your process control center. Each
digital output is connected to a normally closed relay, located inside the analyzer, which
will open in response to the detection of a specific fault. The possible causes of a fault
indication are:
1. The sample gas concentration is outside of specification (i.e., the sample gas
concentration has exceeded the measurement range of the instrument).
2. The analyzer is out of specification or has developed a fault.
NOTICE
The installation of the gas analyzer shall be in accordance with all local and national
standards.
3.5Optical description
The laser modules are located in the core of the analyzer. Each laser module produces a
separate light beam, and these beams are combined linearly as the modules are aligned in
the system. The combined beams are closely coupled, parallel, and coaxial about a virtual
line. The laser light beams pass through a baseplate onto an optical steering assembly,
which directs the laser beam through the sample cell.
The sample cell contains a set of mirrors to create a path through the sample gas that is
between 0.66 ft. (0.2 m), 6.56 ft. (2 m), 16.4 ft. (5 m), and 19.2 ft. (15 m) through multiple
reflections along the length of the cell. The laser beams exit the cell at the opposite end
from where they entered and are directed using a second optical block to a receiver.
By measuring and analyzing the light detected by the receiver unit, it is possible to
accurately determine the concentrations of the target molecules within the gas sample
cell.
Rosemount CT510031
DescriptionReference Manual
May 202000809-0100-4511
32Emerson.com/Rosemount
Reference ManualSpecifications
00809-0100-4511May 2020
4Specifications
4.1Gas detection
The analyzer is highly configurable in the gases that can be detected and their range of
concentrations.
4.2Detailed system specifications
Table 4-1 gives the physical characteristics of the analyzer. Schematic diagrams of the
sensor and mounting points are shown in Figure 4-1 and Figure 4-2. Table 4-2 gives the
general characteristics of the analyzer.
Table 4-1: Physical Characteristics
Rosemount CT5100ValueComment
External dimensions22.64 x 11.73 x 28.11 in.
575 x 298 x 714 mm
Weight117 lb.
53 kg
Length x width x height
Nominal dimensions
Approximate weight
Table 4-2: General Characteristics
Rosemount CT5100ValueUnitsComment
Instrument supply voltage110 to 240Vac50/60 HZ ±10%
Sensor humidity range10 to 95%Relative humidity (non-
condensing) at 113 °F (45 °C)
38Emerson.com/Rosemount
Reference ManualSpecifications
00809-0100-4511May 2020
4.3Safety and system labels and annotation
The labels and annotation applied to the analyzer are specified in the table below.
Label typeExampleLocation
Identification label (including serial
number and model number)
Fuse identification label1. Back plate
Ratings labelEnclosure side panel
Emerson.com/RosemountGasAnalysis Front panel
2. Top right inside of door
Laser radiation CAUTION labelBaseplate
Laser module identification labelOn each laser module housing
Intrinsically safe label1. HMI
2. Intrinsically safe sensor barrier
Rosemount CT510039
SpecificationsReference Manual
May 202000809-0100-4511
Label typeExampleLocation
Terminal labelTop left inside of door
Earth identification labelBack plate
Manufacturer's labelOn analysis cell heater block
Electrical safety labelOn inside of electrical compartment
door
AC Power Supply Danger label1. On outside of electrical
compartment door
2. On manifold block of air
overpressure system
40Emerson.com/Rosemount
Reference ManualInstall
00809-0100-4511May 2020
5Install
5.1Site selection
The Rosemount CT5100 has a T3 temperature classification which specifies the maximum
surface temperature of the analyzer. Ensure that no combustible gas concentrations will
be present, whether on a continual or occasional basis, which have an ignition
temperature below the T-rating of the analyzer.
WARNING
FIRE AND EXPLOSION
Death, personal injury, and/or damage to persons and/or property may result if this is not
observed.
The analyzer's electrical compartment must not be opened unless the atmosphere in the
area is known to be below the ignitable concentration of combustible gases or materials,
or unless all equipment within the protected enclosure is de-energized.
WARNING
ELECTRIC SHOCK
The analyzer operates using mains voltage, which may cause death or serious injury to
personnel. Failure to observe this precaution will cause death, personal injury, and/or
damage to persons and/or property.
Ensure that the circuit breakers are set to Off and locked out and tagged out off before
removing the top cover or opening the front cover.
The analyzer is intended to be installed in a suitable Division 2 shelter to protect it from the
elements.
Provide sufficient space around the analyzer to allow the maintenance and servicing of the
unit.
Rosemount CT510041
InstallReference Manual
May 202000809-0100-4511
Figure 5-1: Clearance with Door Open
Dimensions are in inches (mm).
A. Door open
B. Door closed
C. Door opening arc
42Emerson.com/Rosemount
Reference ManualInstall
00809-0100-4511May 2020
5.2Unpacking
This procedure requires a minimum of two people to safely remove the equipment from
the shipping container.
WARNING
HEAVY INSTRUMENT - LIFTING HAZARD
Handle the analyzer with caution during unpacking, installation, maintenance, and
transport to prevent crushing of hands, feet, or other body parts.
The analyzer weighs 117 lb. (53 kg) and should always be lifted and moved using suitable
lifting/moving equipment. Emerson recommends that a minimum of two people using
suitable tools for transportation and lifting are employed.
Wear suitable protective gloves and protective footwear.
CAUTION
EQUIPMENT DAMAGE
Failure to observe this caution may cause damage to the equipment.
When preparing the analyzer for transport by air, road, or rail, safeguard the analyzer
against movement or break-away during transport by securely strapping it in place.
WARNING
EXPLOSION HAZARD
Installing and wiring the analyzer must comply with all relevant national legislative
requirements and regulations.
Consider all safety instructions within this manual and all associated analyzer instruction
manuals.
WARNING
EXPLOSION HAZARD
Installing the analyzer requires opening the enclosure and working at the open unit. This is
permitted only when both the analyzer and connected external circuitry are de-energized.
Depending on the local regulation, this may require a competent hot work supervisor to
issue a hot work permit.
Rosemount CT510043
Install
May 202000809-0100-4511
Reference Manual
WARNING
HEAVY ITEM
Failure to propery handle the analyzer may cause injury to personnel.
Ensure the wall the analyzer is mounted on is solid, stable, and of suitable material to
hold the weight of the analyzer.
Handle the analyzer with caution during unpacking, installing, maintaining, and
transporting to prevent crushing of hands, feet, or other body parts.
The analyzer weighs 117 lb. (53 kg).
Emerson recommends that a minimum of two people move and lift the analyzer.
Wear suitable protective gloves and protective footwear.
CAUTION
SHOCK AND VIBRATION
Damage to the analyzer may result from a failure to follow this caution.
The analyzer contains sensitive electronic equipment. It MUST NOT be subjected to any
shock and or vibration.
Procedure
1. On receipt of goods, look for any visible damage to the analyzer and verify that all
items noted to be shipped were received. Record on the goods receipt note any
damage or missing items, noting both the item(s) and quantity missing.
2. Visually inspect the exterior of the analyzer for signs of damage, corrosion, gas
leaks, or signs of previously overheating.
3. Report anything found to the maintenance organization.
4. Attach suitably rated and tested lifting slings to the safety engineered lifting eye
bolts mounted on top of the analyzer.
5. One person should carefully guide the equipment from the horizontal to vertical
position while the other person lifts the equipment.
6. Use safety approved and tested lifting equipment to remove the analyzer from the
shipping container and place it on a solid, level surface.
7. Ensure that the analyzer is stored in its protective plastic cover until installation.
5.3Mounting the analyzer
This procedure requires two people to safely move and mount the Rosemount CT5100.
Procedure
1. Ensure that there is free space around the analyzer to allow ventilation of the upper
part of the analyzer.
44Emerson.com/Rosemount
Reference ManualInstall
00809-0100-4511May 2020
WARNING
HEAVY ITEM
Failure to propery handle the analyzer may cause injury to personnel.
Ensure the wall the analyzer is mounted on is solid, stable, and of suitable material
to hold the weight of the analyzer.
Handle the analyzer with caution during unpacking, installing, maintaining, and
transporting to prevent crushing of hands, feet, or other body parts.
The analyzer weighs 117 lb. (53 kg).
Emerson recommends that a minimum of two people move and lift the analyzer.
Wear suitable protective gloves and protective footwear.
Rosemount CT510045
Install
Reference Manual
May 202000809-0100-4511
2. Attach suitably rated and testing lifting slings to the safety engineered lifting eye
bolts mounted on top of the analyzer.
Figure 5-2: Front View Dimensions
A. Lifting eye bolt
B. Sample gas input port
C. Sample gas return port
D. Lifting eye bolt
E. 0.413-in. (10.5 mm) diameter mounting bolts
3. One person should carefully guide the equipment while the other person operates
the lifting equipment.
46Emerson.com/Rosemount
Reference ManualInstall
00809-0100-4511May 2020
4. Use safety approved and tested lifting equipment to lift the analyzer from the stable
platform.
5. Mount the analyzer using four M8 (⅜-in.) fasteners to attach the wall mount
brackets.
The bolts must be positioned in such a way to allow maximum use of all the thread
length.
The installer must ensure that the fasteners used are suitable for the load and
surface that the analyzer is mounted on.
In case you need to thread lock the fittings for extra security, only do this with
compounds compatible with the zone classification of the installation location.
The four wall fixing points must be 0.413-in. (10.5 mm) diameter mounting holes.
Ensure that the wall fixing points are capable of supporting a load of 242 lb (110 kg)
each; this includes a x 2 factor of safety. Figure 5-3 shows the locations of the
mounting points on the analyzer. All mounting points are 0.413-in. (10.5 mm)
diameter holes.
Confirm the bolts are secure. Do not overtighten the fasteners.
The analyzer must be mounted using the four off factory fitted and predrilled holes
on the brace bars. Refer to Figure 5-2.
6. Remove the lifting eyes and retain them for future use.
Threads must be protected with a suitable grease and plastic grommets.
After mounting, do not place any additional load on the analyzer.
Do not place or leave loose items on flat surfaces.
Rosemount CT510047
InstallReference Manual
May 202000809-0100-4511
Figure 5-3: Mounting Details
Dimensions are in inches (mm).
A. Mounting points
48Emerson.com/Rosemount
Reference Manual
00809-0100-4511May 2020
Figure 5-4: Clearance with Door Open
Install
A. Door open
B. Door closed
C. Door opening arc
5.4Connecting the electrical/electronic inputs and
outputs
5.4.1AC power
Power is connected to the analyzer instrumentation through the power entry point (A)
fitted to the base of the analyzer. Refer to Figure 5-5.
Rosemount CT510049
InstallReference Manual
May 202000809-0100-4511
Figure 5-5: Power Gland
A. Power and signal/exit points
B. M6 earth stud
The customer supplied circuit breaker, complete with RCD, must be in accordance with
local and national standards.
50Emerson.com/Rosemount
Reference ManualInstall
00809-0100-4511May 2020
Figure 5-6: Power Entry Point Connections
A. To mains input fuses
B. Mains input filter
C. To enclosure earth stud
Table 5-1: Mains Input Terminals User Connections
TerminalFunction
1Sensor system supply (L)
2Sensor system supply (N)
3Earth (E)
Electrical protection for the instrumentation circuitry of the analyzer is provided by fuses
F1 and F2 located inside the analyzer. Refer to Figure 5-7.
The customer supplied power cable for the analyzer instrumentation will be connected to
terminals 1 - 3.
Rosemount CT510051
InstallReference Manual
May 202000809-0100-4511
Table 5-2: Electrical Power Requirements
5.4.2
Electrical supplyPower
consumption
Instrumentation supply
voltage
500 W100 to 230 Vac, 50/60 Hz ±10%3.15 A
VoltageFuse
internal
fuses
F1 and F2
For the electrical power wiring use 16 AWG stranded, three conductor copper or tin-plated
copper power wire, rated for at least 250 Vac, of the required length. Cables must be
terminated in the power entry points in accordance with local and national electrical
codes. The full electrical wiring diagram is provided in Engineering drawings.
Fuses
Figure 5-7 shows the location of the fuses.
Figure 5-7: Fuses
A. Fuse F4 (24 Vdc supply)
B. Fuse F3 (12 Vdc supply)
C. Fuse F1 (mains supply – live)
D. Fuse F2 (mains supply – neutral)
E. Industrial power supply
Table 5-3: Fuse Requirements
FuseFunctionRatingSchurter part number
1Live line 110/240 Vac
(following mains filter)
2Neutral line 110/240
Vac (following mains
filter)
52Emerson.com/Rosemount
3.15 A, 240 V, fast
acting ceramic
3.15 A, 240 V, fast
acting ceramic
0001.1009
0001.1009
Reference ManualInstall
00809-0100-4511May 2020
Table 5-3: Fuse Requirements (continued)
FuseFunctionRatingSchurter part number
3Analyzer 12 Vdc supply
rail
4Analyzer 24 Vdc supply
rail
3.15 A, 240 V, fast
acting ceramic
3.15 A, 240 V, fast
acting ceramic
0001.1009
0001.1009
5.4.3Connecting the sample supply and return line
Sample gas supply and sample return connections are ¼-in. (6 mm) ”Hamlet”
compression tube fittings. To avoid the risk of gas leaks, confirm that these connections
are made correctly and tightly. Both the sample gas supply pipe and the sample return
pipe should be thermal insulated.
The maximum sample gas supply pressure is 2 BarG.
5.4.4
Connecting the signal cables
The signal cables are connected to the system through conduit outlets (B and C as shown
in Figure 3-4). All signal cables are to be minimum 20 AWG tri-rated switchgear cable.
Customer supplied conduit and cables are to be terminated in the conduit outlets in
accordance with local and national electrical codes.
Rosemount CT510053
InstallReference Manual
May 202000809-0100-4511
Figure 5-8: Signal Cable Outputs
A. To mains input fuses
B. Mains input filter
C. To enclosure earth stud
D. To digital modules
E. To digital modules
F. To analog modules
G. To status relays
H. Digital output terminals
I. Analog or digital output terminals
J. Analog output terminals
K. Status output terminals
L. Mains input terminals
Table 5-4: System Wiring
TerminalFunction
1Sensor system supply (L)
2Sensor system supply (N)
3Earth
11Digital output 1
12Digital output 2
13Digital output 3
14Digital output 4
15Digital output 5
16Digital output 6
17Digital output 7
18Digital output 8
54Emerson.com/Rosemount
Reference Manual
Install
00809-0100-4511May 2020
Table 5-4: System Wiring (continued)
TerminalFunction
19Digital output 9
20Digital output 10
21Digital output 11
22Digital output 12
23Analog output 1
24Analog output 2
25Analog output 3
26Analog output 4
27Analog output 5
28Analog output 6
29Analog output 7
30Analog output 8
31Status output 1 (Check function)
32Status output 2 (Maintenance required)
33Status output 3 (Out of specification)
34Status output 4 (Failed)
01Analog or digital input/output
02Analog or digital input/output
03Analog or digital input/output
04Analog or digital input/output
05Analog or digital input/output
06Analog or digital input/output
5.4.5Power input cables and circuit breaker
The power input cable circuit breaker can be tested as follows:
• Set the main power circuit breaker to ON.
• Check that the display controller lights up. The analyzer will then begin to power-up.
5.4.6
Temperature sensor and cell heater
The sample cell is controlled to operate at a pre-set temperature when it leaves the
factory.
With the analyzer on, the cell will reach the pre-set operating temperature in
approximately 90 minutes.
Rosemount CT510055
InstallReference Manual
May 202000809-0100-4511
5.4.7Pressure sensor
The pressure sensor monitors the pressure in the analysis cell.
To test that it is functioning, check that a pressure reading is displayed under pressure on
the Display Controller as described in Pressure and Temperature screen.
The reading will be approximately 760 Torr at atmospheric pressure. If desired, cap off the
gas inlet and use an external pump to evacuate the cell. Verify that the pressure drops as
expected.
NOTICE
A Torr is a non-SI unit of pressure, defined as 1/760 of standard atmospheric pressure, and
is equal to the fluid pressure of 1 mm of mercury.
5.4.8
5.4.9
5.4.10
Analog output cable
In order to generate a 4-20 mA output, the analyzer must be left for 90 minutes to warm
up, and the analysis cell must be at the correct pressure.
The 4-20 mA outputs will operate when the analysis cell temperature and pressure are
within the required test range. It is not essential to flow sample gas through the system;
nitrogen or atmospheric air will be adequate for this test.
With the analyzer at operating temperature and pressure, ensure that a current between
4 mA and 20 mA is generated on each 4-20 mA output. This can either be measured as a
current with a multimeter, or as a gas concentration through the control station.
Seal glands
Where poured seal glands are a local industry requirement for cable termination, they
should be made in accordance with the manufacturer’s instructions once acceptance tests
have been made on the system to ensure that the analyzer does not need to be removed.
Commissioning
Once the sensor is fully installed as described above, it should be commissioned in
accordance with the commissioning plan agreed between Emerson and the customer.
56Emerson.com/Rosemount
Reference ManualControls and display controller
00809-0100-4511May 2020
6Controls and display controller
6.1Front panel controls and indicators
The Rosemount CT5100 is configured from the control display located on the front panel.
Refer to Figure 6-1.
Figure 6-1: Front Panel
A. Display controller
NOTICE
ON/OFF circuit breaker
There is no ON/OFF switch on the analyzer. The external circuit breaker controls the
application of electrical power to the analyzer.
The external circuit breaker is a simple two-pole ON/OFF circuit breaker that must to set to
ON to permit the safe operation of the analyzer.
The display controller controls operation of the analyzer.Refer to Figure 6-2.
Rosemount CT510057
Controls and display controllerReference Manual
May 202000809-0100-4511
6.2Display controller
The six buttons on the display controller control operation of the Rosemount CT5100:
Figure 6-2.
Figure 6-2: Display Controller Buttons
A. LCD display
B. Configurable button
C. Scroll up button
D. Scroll down button
The LCD displaycan be used to display:
• Gas concentration measurements obtained
• Operating temperature and pressure
• Help screens
• Step-by-step calibration
• Diagnostics
Use the two scroll buttonsto scroll through the information on the LCD display. Use the
right-hand scroll button to scroll up and the left-hand scroll button to scroll down.
The other four buttons are configured to perform different functions according to the
software screen that is shown on the LCD display.
6.3Gas Sensor Main screen
When the analyzer is switched on at the end of the start-up procedure, the Gas Sensor Main
screen (Figure 6-3) appears.
The Gas Sensor Main screen is the screen that is normally displayed.
58Emerson.com/Rosemount
Reference ManualControls and display controller
00809-0100-4511May 2020
NOTICE
The gas concentrations shown in the following screenshots may be different from those
shown in your particular analyzer. The screenshots indicate the functionality of the
software, which is the same regardless of the gases or gas concentrations being measured.
Figure 6-3: Gas Sensor Main Screen
The Gas Sensor Main screen displays the gas concentration measurements obtained by the
analyzer. In the example shown in Figure 6-3 the gases nitric oxide (NO), nitrogen dioxide
(NO2), oxygen (O2), carbon monoxide (CO), and carbon dioxide (CO2) are being measured
and, for each gas, the concentration detected is in parts per million (ppm) or percentage
as applicable.
The NOx reading is the total nitrogen oxide reading, a combination of nitric oxide (NO) and
nitrogen dioxide (NO2).
At the end of the start-up procedure, the gas measurements initially appear as 0.00 ppm
until the first readings are taken. After a few seconds, the initial gas concentrations are
displayed.
The Gas Sensor Main screen also shows the status of the analyzer. In the example shown in
Figure 6-3, the analyzer is Running and OK (e.g., no faults have been identified). This area
of the display shows any errors detected by the software.
On the software screens, highlighted items are links to other screens in the software. To
access a screen, press the button next to the highlighted item.
Rosemount CT510059
Controls and display controllerReference Manual
May 202000809-0100-4511
Figure 6-4: Gas Sensor Main Screen Buttons
A. MENU button
B. MENU text
C. PAGE text
D. PAGE button
E. HELP button
F. HELP text
G. STATUS button
PAGE
HELP
MENU
On the Gas Sensor Main screen, the STATUS button has no function when the analyzer is
operating correctly. If, however, the software detects a fault, an error message is
displayed. Press STATUS to get further information on the error.
A link between the Gas Sensor Main screen and the Pressure and Temperature
screen (described in Pressure and Temperature screen). Press PAGE to toggle
between these two screens.
A link to the Help system. Press HELP to go to the Help screen (described in Help
system).
A link to the Main menu of the software. Press MENU to go to the Main menu
screen (described in Main menu).
6.4Pressure and Temperature screen
The Pressure and Temperature screen (Figure 6-5) shows pressure and temperature
measurements taken inside the analyzer.
60Emerson.com/Rosemount
Reference ManualControls and display controller
00809-0100-4511May 2020
Figure 6-5: Pressure and Temperature Screen
The Cell reading is the temperature, in °C, for the analysis cell.
The Gas reading is the temperature, in °C, of the gas within the analysis cell.
The Pres reading is the pressure, in Torr, inside the analysis cell.
NOTICE
A Torr is a non-SI unit of pressure defined as 1/760 of standard atmospheric pressure and is
equal to the fluid pressure of 1 mm of mercury.
6.5Help system
The analyzer software includes a context-sensitive help system. Press HELP, which is
available on most of the software screens, to open the help system.
The help system contains a number of different Help screens, each conveying a different
message. As the help system is context-sensitive, the Help screen that appears is the one
that is most appropriate to the software function engaged when HELP was pressed. Figure
6-6 shows an example of a Help screen.
Figure 6-6: Example of a Help Screen
Rosemount CT510061
Controls and display controllerReference Manual
May 202000809-0100-4511
6.6Main menu
To access the Main menu (Figure 6-7), press MENU on either side of the Gas Sensor Main
screen (Figure 6-3) or the Pressure and Temperature screen (Figure 6-5). The Main menu
(Figure 6-7) is used for calibration, diagnostics, fault finding, downloading data, and
shutting down the analyzer.
Figure 6-7: Main Menu
6.7BACK button
On most of the software screens, the top lefthand button (Figure 6-8) is configured as a
BACK button. Press BACK to return to the previous screen.
Figure 6-8: BACK Button
A. BACK button
62Emerson.com/Rosemount
Reference ManualStart-up procedure
00809-0100-4511May 2020
7Start-up procedure
7.1Introduction
CAUTION
EQUIPMENT DAMAGE
Damage to the analyzer may result from a failure to follow this procedure.
Always follow the start-up procedure.
The analyzer normally operates continuously. It should only be necessary to start up the
analyzer under the following circumstances:
• When the analyzer is first switched on following installation
• Following repair or maintenance
• When the analyzer has been switched off as part of a plant shutdown or maintenance
7.2Preparation for use
Install and fully commission the Rosemount CT5100 before starting it up.
WARNING
BURN HAZARD
Personal injury and/or damage to property may result if these safety precautions are not
observed.
Some parts of the analyzer may be heated to 374 °F (190 °C). To prevent burns, do not
touch any of the hot parts. All components of a analyzer are hot unless it has been
switched off and allowed to cool down.
Before fitting, removing, or performing any maintenance on the analyzer, make sure that
it has been switched off and allowed to cool for at least two hours. Before performing any
maintenance on or, in the vicinity of the analysis cell, allow the analyzer to cool for at least
twelve hours, as the analysis cell is insulated against heat loss.
When handling the analyzer, always use suitable protective gloves.
These precautions are especially important when working at heights. If a burn is received,
seek medical treatment immediately.
Rosemount CT510063
Start-up procedureReference Manual
May 202000809-0100-4511
CAUTION
EQUIPMENT DAMAGE
Failure to perform pre-system start-up checks may cause damage to equipment.
Do not power up or try to operate the analyzer unless it is physically secure and all
electrical and pneumatic connections to the analyzer are in place.
Before starting up, ensure that electrical power, sample gas handling facilities, and any
required calibration gases are available to the analyzer.
7.3Start-up procedure
NOTICE
The gases shown in the screenshots and the measurements thereof may be different from
those shown in your Rosemount CT5100. They indicate the functionality of the software,
which is the same regardless of the gases being measured.
NOTICE
To stop the start-up procedure at any time, set the main circuit breaker to OFF.
Procedure
1. Visually inspect the analyzer's exterior for signs of damage, corrosion, gas leaks, or
overheating. Report anything found to the maintenance organization.
2. Ensure that the analyzer has been installed correctly as described in Install.
3. Ensure that the top cover is fitted to the analyzer. If it is not, report it to the
maintenance organzation and do not proceed further until the top cover has been
fitted.
4. Ensure that the door to the electrical compartment is closed and locked. If you
cannot close and lock the door, report it to the maintenance organization and do
not proceed further until the door has been repaired.
5. Ensure that the gas handlying system is turned off.
6. Ensure that the external circuit breaker is set to OFF.
7. Visually examine the gas ports.
If necessary, refer to Figure 7-1 to ensure that the sample supply line and the
sample return line are correctly attached.
64Emerson.com/Rosemount
Reference ManualStart-up procedure
00809-0100-4511May 2020
Figure 7-1: Gas Inlet and Outlet and Electrical Connections
A. Sample gas input port
B. Sample gas return port
C. Top cover (cell compartment)
D. Laser/electrical compartment
E. Analog/digital power entry point
F. Power entry point
8. Ensure that the electrical connection has been made to the power entry point at the
base of the analyzer (Figure 7-1).
9. Start up the system that vents the sample gas exiting the analyzer.
10. Switch the circuit breaker to ON, applying electrical power to the analyzer.
The control PC that forms part of the analzyer is configured to automatically load
the necessary gas sensor software and configuration files.
The software automatically starts the start-up sequence. After a few seconds, the
Gas Sensor Main Screen appears on the display controller.
Rosemount CT510065
Start-up procedureReference Manual
May 202000809-0100-4511
Figure 7-2: Gas Sensor Main Screen
If the screen does not appear, report the fault to the maintenance organization.
11. Start up the system for venting the sample gas exiting the analyzer.
12. Start up the gas handling system that conditions the sample gas before it is fed into
the analyzer.
At the end of the start-up procedure, that gas measurements initially appear as 0.00 ppm
until the first readings are taken. After a few seconds, the initial gas concentrations are
displayed.
The start-up procedure is now complete.
66Emerson.com/Rosemount
Reference ManualStart-up procedure
00809-0100-4511May 2020
7.4Front panel controls and indicators
Configure the analyzer from the control display located on the front panel (Figure 7-3).
Figure 7-3: Front Panel
A. Display controller
NOTICE
ON/OFF circuit breaker
There are no ON/OFF switches on the analyzer. A customer-provided external circuit
breaker controls the application of electrical power to the analyzer.
The circuit breaker is a simple two-pole ON/OFF circuit breaker that must be set to ON to
permit the safe operation of the analyzer.
The display controller primarily controls operation of the analyzer (Figure 7-4).
Rosemount CT510067
Start-up procedureReference Manual
May 202000809-0100-4511
Figure 7-4: Display Controller
WARNING
Failure to follow these instructions may cause warranty invalidation, property damage,
and/or personal injury or death.
DO NOT, under any circumstances, press the top two buttons on the display controller
simultaneously. This will override the system code supplied with the analyzer and display
the default program screen for the controller.
This will lock the software controlling the analyzer requiring the analyzer to shut down
externally and restarted risking a possible corruption of the software and loss of the
analyzer operation during the shutdown and restart process.
It is essential that only trained, qualified personnel operate the controls on the analyzer.
Figure 7-5: Default Program Screen
68Emerson.com/Rosemount
Reference ManualOperating the analyzer
00809-0100-4511May 2020
8Operating the analyzer
8.1Introduction
This section describes the normal operation of the analyzer.
NOTICE
The gas concentrations shown in the following screenshots may be different from those
shown in your particular analyzer. The screenshots indicate the functionality of the
software, which is the same regardless of the gases or gas concentrations being measured.
WARNING
BURN HAZARD
Some parts of the analyzer may be heated to 374 °F (190 °C). All components of the
analyzer are hot unless it has been switched off and allowed to cool down. Personal injury
and/or damage to property may result if these safety precautions are not observed. These
precautions are especially important when working at heights.
To prevent burns, do not touch any of the hot parts.
Before fitting, removing, or performing any maintenance on the analyzer, ensure that it
has been switched off and allowed to cool for at least two hours. Before performing any
maintenance on or in the vicinity of the analysis cell, allow the analyzer to cool for at least
twelve hours, as the analysis cell is insulated against heat loss.
When handling the analyzer, always use suitable protective gloves.
If you receive a burn, seek medical treatment immediately.
8.2Normal operation
The analyzer is designed for long term continuous operation, and therefore its normal
state is to be switched on and performing gas measurements. The analyzer is usually only
switched off for maintenance.
During normal operation, either the Gas Sensor Main screen (Figure 8-1) or the Pressure andTemperature screen (Figure 8-2) is shown on the display controller. To toggle between
these two screens, press PAGE..
Rosemount CT510069
Operating the analyzerReference Manual
May 202000809-0100-4511
Figure 8-1: Gas Sensor Main Screen
Figure 8-2: Pressure and Temperature Screen
70Emerson.com/Rosemount
Reference ManualShutdown procedure
00809-0100-4511May 2020
9Shutdown procedure
9.1Safety precautions
CAUTION
EQUIPMENT DAMAGE
Damage to the analyzer may result from a failure to follow this procedure.
Always follow the shutdown procedure.
The analyzer normally operates continuously. It should only be necessary to shut down the
analyzer in the following circumstances:
• In order to perform repairs or maintenance on the analyzer
• When the analyzer has to be switched off as part of a plant shutdown or plant
maintenance
• When the analyzer is switched off for re-calibration
Use the display controller to perform the shutdown procedure. Refer to Display controller
for the display controller navigation instructions.
NOTICE
The gas concentrations shown in the following screenshots may be different from those
shown in your particular analyzer. The screenshots indicate the functionality of the
software, which is the same regardless of the gases or gas concentrations being measured.
WARNING
BURN HAZARD
Some parts of the analyzer may be heated up to 374 °F (190 °C). All parts of the analyzer
are hot unless it has been switched off and allowed to cool down. Personal injury and/or
damage to property may result if these safety precautions are not observed. These
precautions are particularly important when working at heights.
To prevent burns, do not touch any of the hot parts.
Before fitting, removing, or performing any maintenance on the analyzer, ensure that it
has been switched off and allowed to cool for at least two hours. Before performing any
maintenance on, or in the vicinity of, the analysis cell, allow the analyzer to cool for at least
twelve hours, as the analysis cell is insulated against heat loss.
When handling the analyzer, always use suitable protective gloves.
If you receive a burn, seek medical treatment immediately.
Rosemount CT510071
Shutdown procedureReference Manual
May 202000809-0100-4511
9.2Shutdown procedure
WARNING
EXPLOSION HAZARD
Unauthorized operation of the gas handling system when maintenance is being
performed on the analyzer or on its associated pipes/hoses may result in gas being
released, causing fire or explosion. Failure to lock out the gas handling system may cause
death.
Always lock out the gas handling system when shutting down the analyzer.
Procedure
1. Shut down the gas handling system that conditions the sample gas and feeds it to
the analyzer. Always lock-out the gas handling system to prevent its unauthorized
operation during maintenance, which may cause an escape of gas.
WARNING
EXPLOSION HAZARD
Failure to vent sample gas may cause death.
Vent the sample gas in the system to prevent fire or explosion during
maintenance and to prevent damage to the analyzer during shutdown.
Purge the sample gas in the pipes leading to the analyzer to prevent hazards to
personnel during maintenance.
Purge the sample gas in accordance with safe working procedures for the site.
Allow the analyzer and system for returning the sample gas to run for five
minutes to allow any sample gas in the analyzer to be returned to the exhaust.
2. Purge any sample gas in the pipe/hose from the gas handling system to the analyzer
using factory air or nitrogen supply.
WARNING
HIGH PRESSURE GAS AND AIR
The calibration gas supply and compressed air supply operate at a pressure that can
cause injury (e.g. damage to eyes and skin punctures from debris blown by the high
pressure gas or compressed air).
3. Turn off the gas used to purge the analyzer.
72Emerson.com/Rosemount
Reference ManualShutdown procedure
00809-0100-4511May 2020
4. Press MENU on the display controller in either the Gas Sensor Main screen
(Figure 9-1) or the Pressure and Temperature screen (Figure 9-2).
Figure 9-1: Gas Sensor Main Screen
Figure 9-2: Pressure and Temperature Screen
The Main menu (Figure 9-3) opens.
Rosemount CT510073
Shutdown procedureReference Manual
May 202000809-0100-4511
Figure 9-3: Main Menu
A. BACK button
B. SELECT button
C. Scroll up button
D. HELP button
E. Scroll down button
5. On the display controller, select System as shown in Figure 9-3.
6. Press SEL (select).
The System screen opens. Refer to Figure 9-4.
Figure 9-4: System Screen
A. BACK button
B. SELECT button
C. Scroll up button
D. HELP button
E. Scroll down button
74Emerson.com/Rosemount
Reference ManualShutdown procedure
00809-0100-4511May 2020
7. Use scroll up (C) and scroll down (E) to select SHUTDOWN as shown in Figure 9-4.
Then press SEL (select).
The Shutdown screen (Figure 9-5) opens.
Figure 9-5: Shutdown Screen
8. Press YES.
The analyzer shuts down.
9. Set the external secondary circuit breaker to Off. Lock-out and tag-out the
secondary circuit breaker.
10. Set the external main circuit breaker to Off. Lock-out and tag-out the main circuit
breaker.
Rosemount CT510075
Shutdown procedureReference Manual
May 202000809-0100-4511
76Emerson.com/Rosemount
Reference ManualGas calibration
00809-0100-4511May 2020
10Gas calibration
10.1Required tools
The gas concentrations measured by the analyzer can be validated against a known
sample gas or calibrated to match it by using the following gas calibration procedure.
To calibrate the analyzer, you need the following items:
• Nitrogen gas of instrument gas purity for use as a zero calibration gas
• Suitable span calibration gases for each gas measured
• Gas bottle pressure regulators
• Interconnecting hoses to connect the gas bottles to the analyzer
• A T-piece and excess flow line
NOTICE
In the case of gases, such as H2O, for which it is not normally possible to obtain calibrated
gas cylinders, it will usually be measured by the same laser as some other gas. Validating
the other gases measured by the analyzer (particularly any which are measured by the
same laser as H2O) can demonstrate that the system is functioning correctly, meaning
there is no need to calibrate the water measurement directly.
If you need to calibrate (e.g., for legal requirements), you can use a water vapor generator
to supply a known concentration of water vapor.
The validation procedure is essentially the same as the calibration procedure, except that
it does not adjust the analyzer measurements to match the calibration standard.
WARNING
EQUIPMENT DAMAGE
Failure to follow may cause warranty invalidation, property damage, and/or personal injury
or death.
DO NOT, under any circumstances, press the top two buttons on the display controller
simultaneously. This will override the system code supplied with the analyzer and display
the default program screen for the controller.
This will lock the software controlling the analyzer requiring the analyzer to shut down
externally and restart, risking a possible corruption of the software and loss of the analyzer
operation during the shutdown and restart process.
It is essential that only trained, qualified personnel operate the controls on the analyzer.
Rosemount CT510077
Gas calibrationReference Manual
May 202000809-0100-4511
Figure 10-1: Default Program Screen
10.2Main menu calibration routines
Calibration is performed under the control of the analyzer software, using calibration
routines built into the software.
Access calibration functions through the Main menu (Figure 10-4).
Procedure
1. To get the Main menu, press MENU on the Gas Sensor Main screen (Figure 10-2) or
Pressure and Temperature screen (Figure 10-3).
NOTICE
Press BACK to cancel the procedure and return to the previous screen.
Figure 10-2: Gas Sensor Main Screen
78Emerson.com/Rosemount
Reference ManualGas calibration
00809-0100-4511May 2020
Figure 10-3: Pressure and Temperature Screen
Figure 10-4: Main Menu
Use the Main menu (Figure 10-4) to access the software routines and screens
necessary for this process.
2. Use the middle buttons (
Rosemount CT510079
and ) to scroll up or down to select GAS SERVICE.
Gas calibrationReference Manual
May 202000809-0100-4511
3. Press SEL.
Figure 10-5: Gas Service (PER GAS)
4. Use the middle buttons ( and ) to scroll up or down to select PER GAS.
5. Press SEL.
Figure 10-6: Select Gas (NO)
6. Use the middle buttons ( and ) to scroll up or down to select the desired gas (e.g.,
NO).
7. Press SEL.
80Emerson.com/Rosemount
Reference ManualGas calibration
00809-0100-4511May 2020
8. To check or change the current offset value, select ZERO OFFSET and press CAL to
view.
Figure 10-7: Action
Figure 10-8: Zero Offset Screen
9. Use the LEFT, RIGHT, INCR (increase), and DECR (decrease) buttons to enter the
desired offset (-0.587313 in Figure 10-8).
10. Press OK to confirm.
The factory default offset is zero, and under normal circumstances, the offset
should never be a positive number.
If it is shown as positive, press LEFT until the space to the left of the number is
highlighted, then select INCR to change it to a minus (–) sign.
Rosemount CT510081
Gas calibrationReference Manual
May 202000809-0100-4511
11. To check or change the current span value, select SPAN FACTOR and press CAL to
view.
Figure 10-9: Input Type
Figure 10-10: Span Factor Screen
12. Use the LEFT, RIGHT, INCR (increase), and DECR (decrease) buttons to enter the
desired offset (1.06 ppm in Figure 10-10).
13. Press OK to confirm.
The factory default span factor is 1, and under normal circumstances, the offset
should be between 0.8 and 1.2.
10.2.1
82Emerson.com/Rosemount
Zero calibrate an individual gas
Procedure
1. To get to the Main menu, press MENU on the Gas Sensor Main screen (Figure 10-2) or
the Pressure and Temperature screen (Figure 10-3).
Note
To cancel the procedure, press BACK to return to the previous screen.
Reference ManualGas calibration
00809-0100-4511May 2020
2. Use the middle buttons ( and ) to scroll up or down to select GAS SERVICE. Press
SEL.
Figure 10-11: Main Menu (GAS SERVICE)
3. Use the middle buttons ( and ) to scroll up or down to select PER GAS. Press SEL.
Figure 10-12: Gas Service Menu (PER GAS)
Rosemount CT510083
Gas calibrationReference Manual
May 202000809-0100-4511
4. Use the middle buttons ( and ) to scroll up or down to select the desired gas (e.g.,
NO). Press SEL.
Figure 10-13: Select Gas (NO)
5. Use the middle buttons ( and ) to scroll up or down to select RUN ZERO.
Figure 10-14: Action Screen for NO (RUN ZERO)
84Emerson.com/Rosemount
Reference ManualGas calibration
00809-0100-4511May 2020
6. Press CAL to calibrate, or VER to verify.
Figure 10-15 displays as the calibration progresses.
Figure 10-15: Calibration
Once complete, the program will automatically open the Result screen .
Figure 10-16: Result Screen
Rosemount CT510085
Gas calibrationReference Manual
May 202000809-0100-4511
7. Press OK.
The Verify next screen displays.
Figure 10-17: Verify Next Screen
8. Press DONE to finish.
• To calibrate additional gases, select GAS and repeat Step 1 through Step 7.
10.2.2
• To change types, select TYPE.
This completes the process, and the display will return to the Main menu.
Span calibrate an individual gas
Procedure
1. Gas Sensor MainTo get to the Main menu, press the MENU button on the screen
(Figure 10-2) or the Pressure and Temperature screen (Figure 10-3).
Note
To cancel the procedure and return to the previous screen, press BACK.
86Emerson.com/Rosemount
Reference ManualGas calibration
00809-0100-4511May 2020
2. Use the middle buttons ( and ) to scroll up or down to select GAS SERVICE. Press
SEL.
Figure 10-18: Main Menu (GAS SERVICE)
3. Use the middle buttons ( and ) to scroll up or down to select PER GAS. Press the
SEL button.
Figure 10-19: Gas Service (PER GAS)
Rosemount CT510087
Gas calibrationReference Manual
May 202000809-0100-4511
4. Use the middle buttons ( and ) to scroll up or down to select the desired gas (e.g.
NO). Press SEL.
Figure 10-20: Select Gas (NO)
88Emerson.com/Rosemount
Reference ManualGas calibration
00809-0100-4511May 2020
5. Use the middle buttons ( and ) to scroll up or down. Select RUN SPAN. Press CAL
to calibrate, or VER to verify.
Figure 10-21: Action (RUN SPAN)
The Calibration screen show as the calibration progresses.
Figure 10-22: Calibration
Rosemount CT510089
Gas calibrationReference Manual
May 202000809-0100-4511
6. Using the LEFT, RIGHT, INCR (increase), and DECR (decrease) buttons, enter the
cylinder concentration (e.g., 25 ppm).
Figure 10-23: Span Input
7. Press OK to confirm.
Once complete, the program will automatically open Span input For NO
(Figure 10-23).
NOTICE
If an excessively high or low span factor is calculated (<0.7 or > 1.3 for example,) and
if the calibration gas is a different concentration from that set above, it will be
rejected and the previous span factor will be displayed and will continue to be used.
8. Press OK.
The Result screen displays.
Figure 10-24: Result
90Emerson.com/Rosemount
Reference ManualGas calibration
00809-0100-4511May 2020
9. Press DONE to finish.
Refer to Figure 10-25.
Figure 10-25: Calibration Complete
This completes the process, and the display will return to the Main menu.
10.2.3
Calibrate all gases (zero and span)
On systems with valve control, it is possible to run a calibration on all gases. The valves will
sequentially switch through each of the span and zero gases.
This option is not practical to use on systems which require manual valve switching.
In order to set the cylinder concentration when using a new span gas cylinder, it is
necessary to perform a single gas calibration as described in Span calibrate an individual
gas. This concentration will then be stored for future calibrations.
Procedure
1. To get to the Main menu, press MENU on the Gas Sensor Main screen (Figure 10-2) or
the Pressure and Temperature screen (Figure 10-3).
NOTICE
To cancel the procedure and return to the previous screen, press BACK.
Rosemount CT510091
Gas calibrationReference Manual
May 202000809-0100-4511
2. Use the middle buttons ( and ) to scroll up or down to select GAS SERVICE. Press
SEL.
Figure 10-26: Main Menu (GAS SERVICE)
92Emerson.com/Rosemount
Reference ManualGas calibration
00809-0100-4511May 2020
3. Use the middle buttons ( and ) to scroll up or down to select CALIBR ALL. Press
SEL.
Figure 10-27: Gas Service (CALIBR ALL)
The system will work through the zero and span calibration process for all readings.
The Please wait screen will be displayed. This takes approximately one minute per
gas depending on the purge time.
Figure 10-28: Calibration
The Finished screen displays when the zero and span calibration processes
complete.
Rosemount CT510093
Gas calibrationReference Manual
May 202000809-0100-4511
Figure 10-29: Calibration (Finished)
4. Press BACK.
This completes the process, and the display returns to the Main menu.
10.2.4
Change the automatic calibration settings
Procedure
1. To get to the Main menu, press MENU on the Gas Sensor Main screen (Figure 10-2) or
the Pressure and Temperature screen (Figure 10-3).
NOTICE
To cancel the procedure and return to the previous screen, press BACK.
2. Use the middle buttons ( and ) to scroll up or down to select GAS SERVICE. Press
SEL.
Figure 10-30: Main Menu (GAS SERVICE)
94Emerson.com/Rosemount
Reference ManualGas calibration
00809-0100-4511May 2020
3. Use the middle buttons ( and ) to scroll up or down to select CALIBR PARAMS or
VERIFY PARAMS for validation settings). Press SEL.
Figure 10-31: Gas Service (CALIBR PARAMS)
This section applies to systems configured to perform automatic calibration at
timed intervals only.
4. To adjust the frequency of automatic calibration, use the middle buttons ( and )
to scroll up or down
5. Select AUTO PERIOD.
6. Press SEL.
Figure 10-32: Calibr param (AUTO PERIOD)
Rosemount CT510095
Gas calibrationReference Manual
May 202000809-0100-4511
7. To change the time and the units (days, hours, minutes and seconds) between
calibrations, use the LEFT, RIGHT, INCR (increase), and DECR (decrease) buttons to
set the number and time interval.
Figure 10-33: (Calibration) Period
8. Press OK.
9. To adjust the length of time for the span gas to flush the cell, use the middle
buttons ( and ) to scroll up or down and select PURGE TIME.
Figure 10-34: Calibr param (PURGE TIME)
10. Press SEL.
96Emerson.com/Rosemount
Reference ManualGas calibration
00809-0100-4511May 2020
11. To change the time between opening the valve and performing the calibration, use
the LEFT, RIGHT, INCR (increase), and DECR (decrease) buttons to set the number
and time interval.
Figure 10-35: Purge Time
12. Press OK.
Rosemount CT510097
Gas calibrationReference Manual
May 202000809-0100-4511
98Emerson.com/Rosemount
Reference ManualTroubleshooting and diagnostics
00809-0100-4511May 2020
11Troubleshooting and diagnostics
11.1Use the Built-in-Self-Test (BIST) fault
diagnostics
The analyzer has a Built-In-Self-Test (BIST) function that can be used to perform failure
diagnosis of some functions in situations where there is a fault other than a complete
failure of the equipment. To access and control the BIST, use the display controller
mounted on the door of the electrical compartment.
The BIST runs in the background when the analyzer is operating and continuously
monitors the analyzer for faults.
If the BIST detects a fault, the STATUS display in the bottom left-hand corner of the GasSensor Main screen (Figure 11-1) will change from OK to flashing the word FAILURE.
Figure 11-1: Gas Sensor Main Screen
A. FAILURE message
B. FAILURE button
Similarly, if the BIST detects a fault and the Pressure and Temperature screen (Figure 11-2) is
being displayed, the STATUS display in the bottom left-hand corner of the screen will
change from OK to flashing the word FAILURE.
Rosemount CT510099
Troubleshooting and diagnosticsReference Manual
May 202000809-0100-4511
Figure 11-2: Pressure and Temperature Screen
A. FAILURE message
B. FAILURE button
Procedure
1. On either the Gas Sensor Main screen (Figure 11-1) or the Pressure and Temperature
screen (Figure 11-2), press FAILURE button to go to the Faults Menu screen (Figure
11-5).
Controls and display controller provides details of how to use the controls on the
display controller and functions of the BACK and HELP buttons.
2. You can also access the FAULTS menu through the Main Menu (Figure 11-3) as
follows:
100Emerson.com/Rosemount
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.