All possible care has been taken in the preparation of this publication, but Emerson and its agents and distributors accept no
liability for any inaccuracies that may be found. This manual reflects the state of the product at the issue date below, but further
enhancements while in service may mean that the manual does not reflect your particular system.
Emerson reserves the right to make changes without notice both to this publication and the products which it describes.
Document number 00809-0100-4244 Rev AA, January 2020.
No part of this publication may be reproduced, stored in a retrieval system, or transmitted in any form or by any means electronic,
mechanical, photocopying, recording, or otherwise without the express prior written permission of the copyright holder.
If you require additional technical assistance, request help from cascade.support@emerson.com or Emerson distribution partners.
Send general inquiries to cascade.support@emerson.com.
All trademarks used within this document are the property of their respective owners.
Important instructions
Important
Users must read, understand and comply with the following information before proceeding.
All users, installers, operators, and maintainers must be familiar with operating the analyzer. To install, start up, operate, maintain
and service the analyzer in a safe manner, it is MANDATORY to read all additional instruction manuals shipped with the analyzer.
The following instruction manual(s) are available and/or referenced within this manual:
• Rosemount CT4400 Quick Start Guide
All instructions must be saved for future use. Contact your local service center or sales office if you are missing documents.
User information
Important
All users must read this page before proceeding!
Rosemount designs, manufactures, and tests its products to meet many national and international standards. The Rosemount
CT4400 is a sophisticated technical product, and to ensure it continues to operate as designed and within normal specifications, it
MUST be installed, used, and maintained correctly. The following instructions MUST be adhered to and integrated into your safety
program when installing, using, and maintaining Rosemount products.
• Failure to follow the proper instructions may cause:
— Loss of life.
— Personal injury
— Damage to property
— Damage to the analyzer
— Warranty invalidation
• Read all instructions prior to installing, operating, and servicing the analyzer.
• If you do not understand any of the instructions, contact your Rosemount Customer Care Representative for clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the analyzer.
• Inform and educate your personnel in the proper installation, operation, and maintenance of the analyzer.
• Install the analyzer as specified in the Installation Instructions of the manual and in accordance with all applicable local and
national codes.
• Connect the analyzer to the proper electrical and pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the analyzer.
• When replacement parts are required, ensure that only qualified people install replacement parts specified by Rosemount.
• Unauthorized parts and procedures can affect the analyzer’s performance, placing the safe operation of your process at risk,
and VOID YOUR WARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
2
Page 3
• To prevent electrical shock and personal injury, all equipment doors must be closed and protective covers in place, except
when maintenance is being performed by qualified personnel.
• The information contained in this document is subject to change without notice.
General safety notice residual risk
Installation, operation, and maintenance of the analyzer must be in accordance with these instructions.
When operated as intended and all applicable safety instructions are observed, an element of risk will remain to the emission of
gases hazardous to health when all gas connections have been made correctly.
To avoid exposure to the dangers, care must be taken when installing, operating, maintaining, and servicing the analyzer.
Authorized personnel
NOTICE
In-depth specialist knowledge is an absolute requirement for working with and on the analyzer. Personnel installing, operating,
servicing, and maintaining the analyzer must be instructed, trained and qualified with the operating company and the
manufacturer. It is the operating company's responsibility to:
• Train staff
• Observe safety regulations
• Follow the safety instructions and procedures in the product manual
Operators must:
• Be trained
• Read and understand all relevant sections of the product manual before commencing work
• Know the safety mechanisms and regulations
WARNING
To avoid explosions, loss of life, personal injury, and damage to this equipment and on-site property, do not install, operate,
maintain, or service this analyzer before reading and understanding this instruction manual and receiving appropriate training.
2012/19/EUWaste Electrical and Electronic Equipment (WEEE) Directive
USA 21 CFR 1040.1Laser products - CFR - Code of Federal Regulations Title 21 part
NEC® 505National Electrical Code (issued by ANSI: American National
BS EN 61326-1: 2013
EN 61326-1: 2013
IEC/EN 60825-1: 2014Safety of laser products part 1, equipment classification and
BS EN 61010-1 2010 IEC 61010-1 2010Safety requirements for electrical equipment for measurements,
N/AElectrical equipment for measurement, control, and laboratory use.
Description
1040.1
Standards Institute and NFPA 70: National Fire Protection
Association)
Electrical equipment for measurement, control, and laboratory use
- EMC requirements. General requirements.
requirements.
control, and laboratory use. General requirements.
EMC requirements. General requirements.
3
Page 4
Associated publications
Quick Start Guide
Compliance approvals
This product complies with USA 21 CFR 1040.10. It is also designed and manufactured
under an approved quality management system to ISO 9001:2015.
Emerson and the Rosemount CT4400 Gas Analyzer have satisfied the requirements
for applying the CE marking to the Rosemount CT4400 Gas Analyzer.
This equipment meets all requirements of the EMC and Low Voltage directives.
Waste disposal
Do not dispose of measuring tools into household waste.
Only for EC countries:
In accordance with European Directive 2012/19/EU for Waste Electrical and Electronic
Equipment and its implementation into national right, measuring tools that are no
longer usable must be collected separately and disposed of in an environmentally
correct manner.
Safety and information notices
DANGER
WILL CAUSE DEATH
Failure to follow this warning will result in death or serious injury to personnel.
WARNING
DANGER TO PERSONNEL
Failure to follow this warning may result in serious injury to personnel.
WARNING
SAFETY COMPLIANCE
Repairs or alterations are not permitted on any flame-proof paths, features, or joints.
Failure to follow this warning may compromise the product’s safety protection method and void the product certification.
NOTICE
Important messages will appear in this format.
Safety information
All authorized users, including installation, operation, and maintenance personnel, MUST observe the following safety
precautions and warnings.
DANGER
ELECTRIC SHOCK
The analyzer operates using mains voltage that is dangerous to life. Make sure that the power ON/OFF switch at the rear of the
panel is set to OFF and the power cable plug is removed BEFORE removing the top cover.
THE ANALYZER MUST BE EARTHED USING A GROUNDED THREE CORE CABLE suitable for the rating defined in the General
Characteristics (Table 1-3).
Failure to observe this precaution will cause death, personal injury, and/or damage to persons and/or property.
4
Page 5
DANGER
FAILURE TO LOCK-OUT THE GAS HANDLING SYSTEM
Always lock out the gas handling system when shutting down the analyzer. Unauthorized operation of the gas handling system
when maintenance is being performed on the analyzer or its associated pipes/hoses may result in highly flammable gas being
released, causing fire or explosion.
Death or personal injury may result if this is not observed.
DANGER
FAILURE TO VENT SAMPLE GAS.
The sample gas in the system must be vented to prevent fire or explosion during maintenance and to prevent damage to the
analyzer during startup.
The sample gas in the pipes leading to the analyzer must be purged to prevent hazards to personnel during maintenance. Purging
the sample gas must be done in accordance with the safe working procedures for the site.
Allow the analyzer and system for returning the sample gas to run for five minutes to allow any sample gas in the system to be
returned to the exhaust.
WARNING
LASER OPTICAL RADIATION EXPOSURE HAZARD
The Rosemount CT4400 is a Class 1 laser product. The lasers are fully enclosed so that no hazardous radiation is accessible under
normal operating conditions.
Opening the analyzer and attempting to perform adjustments or procedures other than those specified in this manual may
result in hazardous optical radiation exposure to personnel.
WARNING
HAZARDOUS SUBSTANCES
The analyzer may contain hazardous substances. Always handle the analyzer assemblies and components with extreme caution.
Wear personal protective equipment (PPE) when handling the equipment.
Gas handling components within the analyzer contain particulate matter residue from the sample gases. Over the life of the
analyzer, the concentration of particulate matter will become enriched within the gas handling components. When performing
repairs and maintenance on the analyzer:
— Handle used gas handling components with extreme caution.
— Avoid direct skin contact with used gas handling components.
— Do not smoke, drink, or eat in the work area.
— Wear goggles or eye shields.
— Wear a suitable face mask to protect against inhalation of particulate matter.
— Do not wet fingers, eyes, or any exposed skin.
— Pack used gas handling components for disposal in sealed packaging and label them Contaminated.
Dispose of contaminated items as hazardous material in accordance with applicable local, national, or international health and
safety regulations and pollution regulations.
5
Page 6
WARNING
HAZARDOUS GAS
The product stream that the analyzer is examining may be hazardous even at low concentrations. Therefore, take special care to
ensure that the sample gas return port either returns the sample gas to the product stream or discharges the sample gas to a
location that will not cause a hazard.
WARNING
HIGH PRESSURE GAS AND AIR
The calibration gas supply operates at a pressure that can cause injury, e.g., damage to eyes and skin or punctures from debris
blown by the high pressure gas or compressed air.
Always lock off or tag out the calibration gas supply when shutting down the analyzer.
WARNING
EXPLOSION HAZARD
Always lock-out/tag-out the gas handling system when shutting down the analyzer. Unauthorized operation of the gas handling
system when maintenance is being performed on the analyzer or its associated pipes/hoses may result in highly flammable gas
being released, causing fire or explosion.
Failure to observe this precaution will cause death, personal injury, and/or damage to persons.
WARNING
EXPLOSION HAZARD
The sample gas in the system must be vented to prevent fire or explosion during maintenance or during startup.
The sample gas in the pipes leading to the analyzer must be purged to prevent hazards to personnel during maintenance. Purging
the sample gas must be done in accordance with the safe working procedures for the site.
Allow the analyzer and the venting system to run for a minimum of five minutes to allow any sample gas in the system to be
returned to the exhaust.
WARNING
HEAVY ITEM
Handle the analyzer with caution during unpacking, installing, maintaining, and transporting to prevent crushing of hands, feet, or
other body parts.
The analyzer weighs 33 lb (15 kg) for the half rack system and 66 lb (30 kg) for the full rack system. Rosemount recommends that
the analyzer is moved by a minimum of two people when lifting and transporting the equipment.
Failure to properly handle the analyzer may cause injury to personnel, wear suitable protective gloves and protective footwear.
WARNING
PHYSICAL ACCESS
Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of end users’ equipment. This could
be intentional or unintentional and needs to be protected against.
Physical security is an important part of any security program and fundamental to protecting your system. Restrict physical access
by unauthorized personnel to protect end users’ assets. This is true for all systems used within the facility.
6
Page 7
CAUTION
EQUIPMENT DAMAGE
Always follow the start-up procedure. Damage to the analyzer may result from a failure to follow this procedure.
CAUTION
EQUIPMENT DAMAGE
Always follow the shutdown procedure. Damage to the analyzer may result from a failure to follow this procedure.
CAUTION
UNSERVICEABLE EQUIPMENT
If the pressure and temperature measurements are out of tolerance, refer to Troubleshooting and diagnostics for guidance.
CAUTION
EMC
This is a Class A product. In a domestic environment, this product may cause radio interference, in which case the user may be
required to take adequate measures.
CAUTION
EQUIPMENT DAMAGE
Ensure that the local power voltage where the unit is to be installed corresponds to the unit's nominal voltage as given on the
name plate label.
CAUTION
EQUIPMENT DAMAGE
Do not power up or try to operate the analyzer unless it is physically secure and all electrical and pneumatic connections to the
analyzer are in place.
Before commencing the start-up process, it is important to ensure that electrical power, sample gas handling facilities, and any
calibration gases that are required are available to the analyzer.
Safety and system labels and annotation
Safety Labels
The labels and annotation applied to the analyzer are specified in the table.
7
Page 8
Label typeExampleLocation
Identification label (including
the Serial No. (S/N) Model,
Config No. and Date of
Manufacture (DoM).
Fitted on the rear panel
Identification ratings label for
the Half Rack System.
Identification ratings label
for the Full Rack System.
Earth identification label.Backplate
Fitted on the rear panel of the
unit
Fitted on the rear panel of the
unit
8
Page 9
Label typeExampleLocation
CAUTION Laser Radiation
Rear panel
safety statement
AC power supply voltage labelFitted on inside rear panel
next to power input socket.
Abbreviations
The following symbols and abbreviations are used in this manual.
The Rosemount CT4400 Continuous Gas Analyzer, referred to hereafter as Rosemount
CT4400 is a multi-component Quantum Cascade™ Laser (QCL)/Tuned Diode Laser (TDL)
analyzer designed for applications where processing gas streams is required.
Optimized for cold/dry applications running at ambient pressure with low maintenance,
low cost of ownership, and easy integration, it can hold up to five laser modules and
measure up to six components simultaneously with enhanced dynamic range from sub
parts per million (sub ppm) to percentage levels in a process gas stream.
The analyzer is designed to be mounted in a standard 19 in. (482.6 mm) rack width as
either a half rack or full rack system depending on the gas analysis configuration required.
WARNING
HEAVY ITEM
Handle the analyzer with caution during unpacking, installing, maintaining, and
transporting to prevent crushing of hands, feet, or other body parts.
The analyzer weighs 33 lb (15 kg) for the half rack system and 66 lb (30 kg) for the full rack
system. Rosemount recommends that the analyzer is moved by a minimum of two people
when lifting and transporting the equipment.
Failure to properly handle the analyzer may cause injury to personnel, wear suitable
protective gloves and protective footwear.
1.2Equipment purpose and role
The analyzer is a multi-component QCL/TDL gas analysis system that can be configured to
measure concentrations of multiple small molecules in a gas sample that is provided to
the analyzer through a customer installed gas sample line.
The types of molecules that are measured depend on the system configuration.
Reference Manual13
Page 14
PlanReference Manual
January 202000809-0100-4244
Figure 1-1: Rosemount CT4400 Half Rack System
Figure 1-2: Rosemount CT4400 Full Rack System
The analyzer can be configured to detect and measure up to six different gases
simultaneously.
14Rosemount CT4400
Page 15
Reference Manual
00809-0100-4244January 2020
Plan
1.3System overview
A complete Rosemount CT4400 system consists of the analyzer, and the associated
interconnecting wiring and gas piping.
Measurement data from the analyzer can be displayed in the process control center. You
must provide a stable gas handling system and interconnecting wiring and gas piping;
Rosemount will supply the analyzer.
In Figure 1-3, the items supplied by Rosemount are colored orange; the items supplied by
the Customer are colored blue. Table 1-1 lists the main items of the system.
Figure 1-3: Complete Rosemount CT4400 Gas Analysis System
A. Gas handling system
B. Sample supply line
C. Sample return (exhaust) line
D. Rosemount CT4400 Gas Analyzer
E. Electrical power
F. Measurement data
G. Control center
The analyzer contains an optical system with a laser, components that provide an optical
path, sample inlet/outlet ports that can be connected to a customer provided gas
handling system and control/analysis electronics. The complete system operates from
either a 110 Vac 60Hz or a 230 Vac 50Hz supply.
Gas concentrations are measured using mid-infrared optical absorption spectroscopy. The
light source is provided by a quantum cascade laser or tuned diode laser, which produces
wavelength sweeps that cover the absorption lines of the gases. The light from the laser is
routed through an optical path to the analysis cell, which provides measurement of low
concentrations of the sample gas. An external sample handling system conditions the
sample gas and draws it through the analysis cell. The laser light exiting the multi-pass
analysis cell is directed to a receiver in the analyzer. The variation in the intensity of the
light in the vicinity of the absorption lines is measured, and the concentration is
determined using a spectral fitting routine.
There is no sample conditioning provided within the analyzer; this is the responsibility of
the Customer. The sample gas must be conditioned as detailed in Sample gas conditioning
requirements before entering the analyzer. Detailed characteristics of the analyzer are
given in Detailed system specifications.
Reference Manual15
Page 16
Plan
Reference Manual
January 202000809-0100-4244
Table 1-1: Main Items of the Rosemount CT4400 System
ItemName or descriptionSupplied byPart numberQuantityNotes
1Rosemount CT4400RosemountCT44001N/A
2Rosemount CT4400
software package,
version 5.7.8.0
minimum
3Gas handling systemCustomerCustomer
4Exhaust line hose (for
sample gas)
5Reference gas
cylinders (instrument
gas) for calibration
purposes
6Pressure regulatorCustomerCustomer
7Pneumatic T-pieceCustomerCustomer
8Excess flow lineCustomerCustomer
9110/230 Vac
power cable
10Cable from analyzer to
control center
RosemountN/A1Software is embedded on
choice
CustomerCustomer
choice
CustomerCustomer
choice
choice
choice
choice
RosemountCustomer
choice
CustomerCustomer
choice
the PC board.
Version described in the
manual
1Selected to provide steady
state conditions
1N/A
Dependent
upon number
of gases
being
measured
1 per gas
cylinder
2Required for calibration
1Required for calibration
13 core cable MUST be
1N/A
N/A
Required for calibration
earthed
11482.6 mm (19 in.) rack
system
CustomerCustomer
choice
1Holds the Rosemount
CT4400
1.4Customer information
This manual contains all the important information that must be followed to ensure the
correct operation and safety of personnel when operating the analyzer.
For information regarding installation, consult Install and the Quick Start Guide.
Emerson is committed to continuously improving its products and documentation. Every
effort will be made to include in the documentation any modifications by the
manufacturer. However, this document reflects the supplied analyzer at the revision date
on the front cover.
Should you require further information, or should particular problems arise that are not
covered in this manual, you can request additional help from Cascade Technical Support
(cascade.support@emerson.com) or Emerson distribution partners. Further contact
details for Emerson can be found on the back page of this manual.
16Rosemount CT4400
Page 17
Reference ManualPlan
00809-0100-4244January 2020
1.5Safety precautions and conditions for safe use
WARNING
Before installing or performing any maintenance on the analyzer, read and understand the
safety information given in the preliminary information of this manual.
The analyzer described in this manual has been quality control tested and left the
manufacturer in pristine condition. To achieve the correct and safe operation of this
product, it must be transported, installed, operated, and maintained as described by the
manufacturer.
All lasers used within the analyzer are Class 1. The lasers are fully enclosed so that no
hazardous radiation is accessible under normal operating conditions. Opening the
analyzer and attempting to perform adjustments or procedures other than those specified
in this manual may result in hazardous optical radiation exposure to personnel.
The analyzer has warning labels at appropriate positions in accordance with USA 21 CFR
1040.10.
1.6Qualified personnel
This manual provides installation, operation, and maintenance personnel with the level of
knowledge required to safely start, operate, and switch off the analyzer.
Only technically qualified personnel in the field of analysis and control who are familiar
with this manual and have been specially trained on the analyzer should install, operate,
switch off, and service the analyzer. Only qualified and trained persons have the required
specific knowledge to correctly interpret the general safety information, warnings, and
procedures given in this manual and apply them to this particular application. Emerson or
its distribution partners can provide this training on request.
Knowledge of the safety information within this manual and its technically correct
implementation are prerequisites for danger-free installation, operation, and maintenance
of the analyzer.
1.7Software version
The analyzer includes software that is used to control the operation of the analyzer. This
manual describes the software version as: 5.7.8.0 minimum.
Reference Manual17
Page 18
PlanReference Manual
January 202000809-0100-4244
1.8Detailed system specifications
Table 1-2 gives the physical characteristics of the analyzer. Schematic diagrams of the
analyzers are shown in Figure 1-4 and Figure 1-5 with mounting points in Figure 2-1 and
Figure 2-2. General and environmental characteristics for the analyzer are provided in
Table 1-3 and Table 1-4.
Table 1-2: Physical Characteristics
Rosemount™ CT4400
External dimensions
(Half Rack)
Weight (Half Rack)15
External dimensions
(Full Rack)
Weight (Full Rack)30
Value
623 x 215 x 172
24.5 x 8.5 x 6.8
33
658 x 483 x 172
25.9 x 19 x 6.8
66
UnitsComment
mm
in.
kg
lb
mm
in.
Kg
lb
Length x width x height
Nominal dimensions
Approximate weight
Length x width x height
Nominal dimensions
Approximate weight
Table 1-3: General Characteristics
Rosemount CT4400ValueUnitsComment
Supply voltage110/230 ± 10%VacN/A
Supply Frequency60/50 ± 5%HzN/A
Peak power consumption50WMax. consumption
Frame and structure materialN/AN/AAnodized and powder coated aluminum
Overvoltage categoryIIN/ANumeral defining a transient
Pollution degree2N/ANumeral indicating the level
Note
The Rosemount CT4400 Continuous Gas Analyzer is suitable for indoor use only.
1.8.1
Sample gas conditioning requirements
The sample gas temperature in Table 1-4 indicates the temperature range that the
analyzer can operate within.
ValueUnitsComment
overvoltage condition.
Overvoltage Category ll is
for equipment intended to
be supplied from the
building wiring.
of pollution that may be
present in the environment.
Pollution degree 2: only
non-conductive pollution
occurs except that
occasionally a temporary
conductivity caused by
condensation is expected.
1.8.2
The sample gas pressure in Table 1-4 shows the pressure range that the analyzer can
operate within. The sample gas pressure
ensure steady state conditions are being analyzed at all times. Sudden fluctuations in
pressure will affect the analysis being performed.
In the event of a power failure any residual sample gas should be purged from the analyzer
to ensure any corrosive gas elements are not left to contaminate/degrade the internal
components. Failure to do so could result in irreparable damage to the analyzer.
Sample filtration may be required to ensure the gas sample particulate size and density
does not exceed the requirements in Table 1-4.
MUST
not fluctuate more than -10 percent to
Optical description
The laser module is located in the core of the analyzer and the laser light beam passes
through the sample cell. The sample cell contains a set of mirrors to create a path through
the sample gas that is between 6.56 ft. (2 m) and 16.4 ft. (5 m) through multiple
reflections of the laser beam in the cell. The laser beam exits the cell at the opposite end
from where they entered and are directed using an optical block to a receiver.
By measuring and analyzing the light detected by the receiver unit, it is possible to
accurately determine the concentration of the target molecules within the gas sample.
20Rosemount CT4400
Page 21
Reference ManualPlan
00809-0100-4244January 2020
1.9Unpacking the analyzer
WARNING
HEAVY ITEM
Handle the analyzer with caution during unpacking, installing, maintaining, and
transporting to prevent crushing of hands, feet, or other body parts.
The analyzer weighs 33 lb (15 kg) for the half rack system and 66 lb (30 kg) for the full rack
system. Rosemount recommends that the analyzer is moved by a minimum of two people
when lifting and transporting the equipment.
Failure to properly handle the analyzer may cause injury to personnel, wear suitable
protective gloves and protective footwear.
Procedure
1. Visually inspect the packaging to ensure there has been no damage in transit or
storage.
2. Open the packaging and visually inspect the exterior of the analyzer for signs of
damage, corrosion, gas leaks, or signs of previous overheating if returning from
storage.
3. Report anything found to the maintenance organization.
4. Carefully lift the analyzer from the shipping container (handles are provided on the
full rack system to aid this process).
5. Place the analyzer on a flat, stable surface that can accommodate the weight (up to
66 lb [30 kg]). This step requires two people to lift and transport the analyzer.
Reference Manual21
Page 22
Plan
Reference Manual
January 202000809-0100-4244
Figure 1-6: Half Rack Front and Rear Views
A. Front panel
B. Display screen
C. User interface
D. Rear panel
E. Ethernet connection
F. Sample gas outlet
G. Sample gas inlet
H. Analog output (4-20 mA)
I. Mains power inlet
Figure 1-7: Full Rack Front and Rear Views
A. Front panel
B. Display screen
C. User interface
D. Rear panel
E. Ethernet connection
F. Sample gas outlet
G. Sample gas inlet
H. Analog output (4-20 mA)
I. Mains power inlet
22Rosemount CT4400
Page 23
Reference Manual
00809-0100-4244January 2020
Install
2Install
2.1Site selection and rack mounting
WARNING
HEAVY ITEM
Handle the analyzer with caution during unpacking, installing, maintaining, and
transporting to prevent crushing of hands, feet, or other body parts.
The analyzer weighs 33 lb (15 kg) for the half rack system and 66 lb (30 kg) for the full rack
system. Rosemount recommends that the analyzer is moved by a minimum of two people
when lifting and transporting the equipment.
Failure to properly handle the analyzer may cause injury to personnel, wear suitable
protective gloves and protective footwear.
Procedure
1. The analyzer MUST be installed in an enclosed environment that is protected from
the elements, humidity, and temperature variations for optimum performance.
2. Site selection for analyzer installation MUST account for the handles on the front
face of the full rack system and rear entry for the cable and tube connections (see
Figure 2-1 and Figure 2-2).
3. Screw in rubber feet have been fitted to the base of the analyzer – if required these
can be removed to aid assembly into the rack system being provided.
4. The full rack system has additional slotted holes on the front face for securing the
analyzer to the rack system (see Figure 2-2).
5. The minimum tubing and cabling allowance defined in Figure 2-1 and Figure 2-2 at
the rear of the analyzer MUST be accounted for when installing the rack system.
Figure 2-1: Half Rack Installation Requirements
A. Four off M4 mounting holes for customer supplied mounting ears (if required)
B. Three off M4 mounting holes for customer supplied rail (if required)
Reference Manual23
Page 24
InstallReference Manual
January 202000809-0100-4244
Figure 2-2: Full Rack Installation Requirements
A. Four off ⌀5.5mm mounting slots
B. Four off M4 mounting holes for customer supplied mounting ears (if required)
C. Three off M4 mounting holes for customer supplied rail (if required)
6. If rigid tubing is being connected to the analyzer further clearance will be required
to allow full access to the connections for installation and maintenance.
7. With rigid tubing fitted the analyzer will not be capable of being slid out of the rack
on the slides.
8. Install the analyzer in a standard 19 in. (482.6 mm) rack system.
Note
Supply of the rack, telescopic slide rails, fitting and securing the analyzer into the
rack is the responsibility of the Customer. (Typical requirement - 2 off telescopic
slides, 24 in. (610 mm) long, 121.3 lb (55 kg) maximum load 2 in. (50.8 mm) over
travel 4 in. (9.6 mm) slide thickness Lock-out Front disconnect.).
9. Installation of telescopic slides on the analyzer will be in accordance with the
manufacturer’s procedure the locations for the slide fixing points are provided in
Figure 2-1 and Figure 2-2.
24Rosemount CT4400
Page 25
Reference ManualInstall
00809-0100-4244January 2020
Figure 2-3: Analyzer With Telescopic Slides
10. Due to the weight, care MUST be taken when installing the analyzer in the rack.
2.2Gas inputs and outputs
The analyzer has provision for gas input and gas output, both of which are located on the
rear panel of the analyzer (Figure 2-4).
WARNING
HAZARDOUS GAS
The product stream that the analyzer is examining may be hazardous even at low
concentrations. Therefore, take special care to ensure that the sample gas return port
either returns the sample gas to the product stream or discharges the sample gas to a
location that will not cause a hazard.
Procedure
1. The sample supply line provided by the Customer must provide gas at a constant
temperature to the sample gas input port on the analyzer to prevent condensation
forming at any point in the sample supply line within the analyzer, which could
reduce the accuracy of the gas analysis.
2. The gas sample enters the analyzer through the sample gas input port
(C) Figure 2-4.
3. Once the gas sample passes through the measurement cell and it is expelled from
the analyzer through the sample gas return port (B) Figure 2-4.
4. The Customer is responsible for ensuring that all expelled gases are disposed of in
accordance with all local and national regulations.
5. Connections for the full rack system where a single gas sample inlet requiring dual
cell analysis or two gas samples requiring separate analysis are shown in Figure 2-4.
Reference Manual25
Page 26
InstallReference Manual
January 202000809-0100-4244
Figure 2-4: Gas Inlet and Outlet Connections
A. Rear panel
B. Sample gas outlet
C. Sample gas inlet
2.3Connecting the electrical/electronic inputs and
outputs
This section discusses making the electrical connections to the analyzer.
DANGER
ELECTRIC SHOCK
The analyzer operates using mains voltage that is dangerous to life. Make sure that the
power ON/OFF switch at the rear of the panel is set to OFF and the power cable plug is
removed.
The analyzer MUST be earthed using a grounded three core cable suitable for the rating
defined in the General Characteristics (Table 1-3).
Death or personal injury may result if this is not observed.
26Rosemount CT4400
Page 27
Reference ManualInstall
00809-0100-4244January 2020
Figure 2-5: Electrical/Electronic Inputs and Outputs
A. Rear panel
B. Ethernet connection
C. D sub connector type - 25 way
D. Mains power inlet
Procedure
1. Make electrical/electronic signal connections to the analyzer through connectors
located on the analyzer's rear panel (see Figure 2-5).
2. Use the wiring diagram to make the electrical connections as shown in Table 2-1,
Table 2-2, and Engineering drawings.
3. Power to the analyzer is connected through the Mains Power Inlet (Figure 2-5, D)
using the cable provided.
4. The Ethernet connector (Figure 2-5, B) provides an Ethernet output from the
analyzer that may be used for downloading data for failure diagnosis purposes.
5. The results of the gas analysis are output from the analyzer through the Analog
Output connection (Figure 2-5, C) and sent to your process control center.
Table 2-1: Analog Outputs Half Rack System, 25 Way Socket
PinsFunctionMeasurement (point
of display)
1 & 14Analog channel 0, set in config file (4-20 mA, SELV)Sample gas
measurement 1
2 & 15Analog channel 1, set in config file (4-20 mA , SELV)Sample gas
measurement 2
3 & 16Analog channel 2, set in config file (4-20 mA, SELV)Sample gas
measurement 3
4 & 17Analog channel 3, set in config file (4-20 mA, SELV)Sample gas
Damage to the analyzer may result from a failure to follow this procedure.
Always follow the start-up procedure.
The analyzer normally operates continuously. It should only be necessary to start up the
analyzer under the following circumstances:
• When the analyzer is first switched on following installation.
• Following repair or maintenance.
• When the analyzer has been switched off as part of a plant shutdown or maintenance.
3.2Preparation for use
The analyzer must be installed and fully commissioned prior to startup.
DANGER
ELECTRIC SHOCK
The analyzer operates using mains voltage that is dangerous to life. Make sure that the
power ON/OFF switch at the rear of the panel is set to OFF and the power cable plug is
removed.
The analyzer MUST be earthed using a grounded three core cable suitable for the rating
defined in the General Characteristics (Table 1-3).
Death or personal injury may result if this is not observed.
CAUTION
EQUIPMENT DAMAGE
Do not power up or try to operate the analyzer unless it is physically secure and all
electrical and pneumatic connections to the analyzer are in place.
Before commencing the start-up process, it is important to ensure that the electrical
power, sample gas handling and any calibration gases that are required are available.
Reference Manual29
Page 30
Start-up procedureReference Manual
January 202000809-0100-4244
3.3Start-up procedure
NOTICE
The gases shown in the screenshots and the measurements may be different from those
shown in your particular analyzer. They indicate the functionality of the software, which is
the same regardless of the gases being measured.
CAUTION
To stop the start-up procedure at any time, set the main circuit breaker to OFF.
To start the analyzer, perform the following steps:
Procedure
1. Visually inspect the exterior of the analyzer for signs of damage, corrosion, gas
leaks, or overheating. Report anything found to the maintenance organization.
2. Ensure that the analyzer has been correctly installed.
See Install.
3. Ensure that the On/Off switch at the rear of the analyzer is set to I (ON).
Figure 3-1: Rear Panel ON/OFF Switch
A. Rear panel
B. On/Off switch
4. The start-up sequence commences automatically under software control.
5. After a few minutes, the Gas Sensor Main screen (Figure 3-2) appears on the display
controller. If it does not, report the fault to maintenance.
30Rosemount CT4400
Page 31
Reference ManualStart-up procedure
00809-0100-4244January 2020
Figure 3-2: Gas Sensor Main Screen
6. Start/open the return system installed for the sample gas handling for the analyzer.
7. Start the gas handling system that conditions the sample gas before it is fed into
the analyzer.
8. Allow the analyzer sufficient time to stabilize in accordance with Table 1-3 before
steady state measurements are taken.
Reference Manual31
Page 32
Start-up procedureReference Manual
January 202000809-0100-4244
32Rosemount CT4400
Page 33
Reference ManualOperating the analyzer
00809-0100-4244January 2020
4Operating the analyzer
4.1Introduction
This chapter describes the controls, displays, and indicators on the Rosemount CT4400
and how to use the display controller located on the front panel of the analyzer.
NOTICE
The gases shown in the screenshots and the measurements may be different from those
shown in your particular analyzer. They indicate the functionality of the software, which is
the same regardless of the gases or gas concentrations being measured.
4.2Normal operation
The analyzer is designed for long term continuous operation, and therefore its normal
state is to be switched on and performing gas measurements. The analyzer is usually only
switched off for maintenance. The shutdown procedure used to switch off the analyzer is
described in Shutdown procedure.
Provided that the Start-up procedure has been followed, the analyzer does not require any
human intervention during normal operation other than occasional calibration checks as
described in Gas calibration.
During normal operation, either the Gas Sensor Main screen (Figure 4-1) or the Pressure andTemperature screen (Figure 4-2) is shown on the display controller. To toggle between
these two screens, press
Figure 4-1: Gas Sensor Main Screen
.
Reference Manual33
Page 34
Operating the analyzerReference Manual
January 202000809-0100-4244
Figure 4-2: Pressure and Temperature Screen
CAUTION
UNSERVICEABLE EQUIPMENT
If the pressure and temperature measurements are out of tolerance, refer to
Troubleshooting and diagnostics for guidance.
On both the Gas Sensor Main screen (Figure 4-1) and the Pressure and Temperature screen
(Figure 4-2) has no function when the analyzer is operating correctly. If, however, the
software detects a fault and an error message is displayed, press to get further
information on the error.
On both the Gas Sensor Main screen and the Pressure and Temperature screen, if the
analyzer makes more measurements than can fit on the display, use and to scroll
up and down the list.
34Rosemount CT4400
Page 35
Reference ManualOperating the analyzer
00809-0100-4244January 2020
4.3Display controller
Operation of the analyzer is controlled primarily through the display controller (Figure 4-3,
A).
Figure 4-3: Front Panel Display Controller
A. LCD display
B. Navigation buttons
The LCD display (A) can be used to display:
• Gas concentration measurements obtained.
• Operating temperature and pressure.
• Help screens.
• Step-by-step calibration.
• Diagnostics.
The navigation buttons (B) are configured to perform different functions according to
which software screen is shown on the LCD display.
Generally used to select an alternative function. Also allows you to toggle
between gas and physical measurements from the Home screen.
Referred to as ENTER.
The analyzer employs Intelligent Device Management, which enables self-monitoring and
diagnostics. This ensures that operators are made aware of malfunctions so they can take
appropriate action. Table 4-2 defines the symbols that may be displayed.
Table 4-2: Diagnostic symbols
System running
Maintenance required: still valid output signal
Out of specification: signal out of the specified range
Indicates the analyzer is performing a calibration or validation or that the
software has been deliberately stopped.
Failure: non-valid output signal
4.4Gas sensor main screen
When the analyzer is switched on, at the end of the startup procedure, the Gas Sensor
Main screen (Figure 4-4) appears.
The Gas Sensor Main screen is the screen that is normally displayed.
NOTICE
The gas concentrations shown in the following screenshots may be different from those
shown in your particular analyzer. The screenshots indicate the functionality of the
software, which is the same regardless of the gases or gas concentrations being measured.
Figure 4-4: Gas Sensor Main Screen
36Rosemount CT4400
Page 37
Reference ManualOperating the analyzer
00809-0100-4244January 2020
The Gas Sensor Main screen displays the gas concentration measurements obtained by
the analyzer. In the example shown in Figure 4-4, the gases Nitric Oxide (NO), Nitrous
Oxide (N2O), Carbon Monoxide (CO) and Ammonia (NH3) are being measured and, for
each gas, the concentration detected is in parts per million (ppm).
At the end of the start-up procedure, the gas measurements initially appear as 0.00 ppm
until the first readings are taken. After a few seconds, the initial gas concentrations are
displayed.
The Gas Sensor Main screen also shows the status of the analyzer. In the example shown
in Figure 4-4, the analyzer is Running and OK (e.g., no faults have been identified).
If a fault is identified, is displayed; if maintenance is required, is displayed.
is a link between the Gas Sensor Main screen (Figure 4-4) and the Pressure and
Temperature screen (Figure 4-5). Press
is a link to the Home Screen. Press to go to a Help screen (described in Help system).
On the Gas Sensor Main screen, has no function when the analyzer is operating correctly.
If, however, the software detects a fault, an error message is displayed. Press to get
further information on the error.
to toggle between the two screens.
4.5Pressure and temperature screen
The Pressure and Temperature screen (Figure 4-5) shows pressure and temperature
measurements taken inside the analyzer. The measurements shown in the manual are
indicative and may vary depending on the configuration of your individual system.
Figure 4-5: Pressure and Temperature Screen
The Cell Te reading is the temperature, in °C, of the analysis cell.
The Transd reading is the temperature, in °C, of the Transducer.
The Pressu reading is the pressure, in Torr, inside the analysis cell.
The Gas Te reading is the temperature, in °C, of the gas within the analysis cell.
Reference Manual37
Page 38
Operating the analyzerReference Manual
January 202000809-0100-4244
NOTICE
A Torr is a non-SI unit of pressure defined as 1/760 of standard atmospheric pressure and is
equal to the fluid pressure of 1 mm of mercury.
4.6Help system
The analyzer software includes a context-sensitive help system. Help messages are
displayed at either the top or bottom of the screen as appropriate.
When the
example of a help screen.
Figure 4-6: Example of a Help Screen
button is pressed it takes you to the home page. Figure 4-6 shows an
4.7Main menu
To access the Main menu (Figure 4-7), press on either the Gas Sensor Main screen
(Figure 4-4) or the Pressure and Temperature screen (Figure 4-5). The Main menu options
are:
DIAGNOSTICS:
Displays various parameters used in the internal calculations and
compares desired and actual parameters, for example, the analysis cell
pressure and temperature. The diagnostics routines and screens are
used to perform fault diagnosis.
FAULTS:
GAS SERVICE:
DATA SERVICE:
SYSTEM:
38Rosemount CT4400
Navigates to a screen that lists any faults affecting the analyzer. This
option is used as part of the failure diagnosis procedures.
Allows you to check the sensor readings against a known gas source
(verification) or, if necessary, to calibrate the analyzer against that
known gas source (calibration).
Navigates to a screen that allows data to be downloaded from the
analyzer. The downloaded data is used to diagnose faults.
Navigates to a screen that allows you to shut down the analyzer or
access system information, such as software versions or IP addresses.
Page 39
Reference ManualOperating the analyzer
00809-0100-4244January 2020
The main motherboard IP address can also be changed from this menu.
Some of these options are not available on older analyzers.
Figure 4-7: Main Menu
4.8BACK button
On most of the software screens, is configured as a BACK button. Press to return to
the previous screen.
Reference Manual39
Page 40
Operating the analyzerReference Manual
January 202000809-0100-4244
40Rosemount CT4400
Page 41
Reference ManualVerifying gas concentrations
00809-0100-4244January 2020
5Verifying gas concentrations
5.1Verification
Verification flows the known gas concentration through the analyzer and gives you a
display of the measurement, the cylinder value, and the difference between the two. You
can use verification to confirm that the analyzer is within tolerance. If it is out of tolerance,
perform a calibration (see Calibration).
5.1.1Zero verification
Zero verification confirms that when no sample gas is flowing through the analyzer, the
gas concentrations measured by the analyzer are zero. To zero verify, compare the
analyzer measurements to a known sample gas using the following procedure:
Procedure
1. Use nitrogen gas of instrument gas purity as the zero verification gas.
2. Ensure that a pressure regulator is connected to the nitrogen gas bottle.
3. Connect a hose from the nitrogen gas bottle through a T-piece to the sample supply
port on the rear of the analyzer.
4. Connect an excess flow line to the unused port on the T-piece and route the excess
flow line to a suitable extractor.
5. To ensure an accurate zero measurement, the analyzer must have been switched
on for a minimum of 30 minutes prior to verifying the reading.
6. On the display controller of the analyzer select the Main Menu screen.
7. Scroll
Figure 5-1: Gas Service
or and select GAS SERVICE. Press .
Reference Manual41
Page 42
Verifying gas concentrationsReference Manual
January 202000809-0100-4244
8. Scroll or and select PER GAS. Press .
Figure 5-2: Per Gas
9. Scroll or and select from the list the gas to be tested. Press .
Figure 5-3: Select Gas Screen
10. Scroll or and select RUN ZERO. Press .
Figure 5-4: For NH3
42Rosemount CT4400
Page 43
Reference ManualVerifying gas concentrations
00809-0100-4244January 2020
11. Allow the concentration readings to stabilize.
Figure 5-5: Result
This screen gives a reading of the concentration of the selected gas that is present
as an impurity in the nitrogen calibration gas.
In the example the gas being measured is NH3, and the analyzer has detected a
concentration of 0.866 ppm.
If the reading is within tolerance, no further action is required.
12. Press to end the zero verification process.
13. Press the appropriate button.
Option
Description
To return to the Gas Service Screen.
To select the next gas from the SelectGas screen.
To return to the Main Menu.
To return to the Gas Sensor Mainscreen (Figure 4-4).
Reference Manual43
Page 44
Verifying gas concentrationsReference Manual
January 202000809-0100-4244
Figure 5-6: Next Gas Service
If the reading is outside tolerance, follow the Zero calibration.
5.1.2
Span verification
To verify the span gas concentrations measured by the analyzer when reference gas is
flowing:
Procedure
1. Use a certified reference gas cylinder as the source of the span verification gas.
2. Ensure that a pressure regulator is connected to the reference gas bottle.
3. Connect a hose from the reference gas bottle, through a T-piece, to the sample
supply port on the rear panel of the analyzer.
4. Connect an excess flow line to the unused port on the T-piece and route the excess
flow line to a suitable extractor.
5. On the display controller, browse to the Main menu.
6. Press or , select GAS SERVICE. Press .
Figure 5-7: Main Menu
44Rosemount CT4400
Page 45
Reference ManualVerifying gas concentrations
00809-0100-4244January 2020
7. Scroll or and select PER GAS. Press .
Figure 5-8: Gas Service
8. Scroll or and select from the list the gas to be tested. Press .
Figure 5-9: Select Gas Selection
Reference Manual45
Page 46
Verifying gas concentrationsReference Manual
January 202000809-0100-4244
9. Scroll or and select RUN SPAN. Press .
Figure 5-10: For NH3
Note
Allow the concentration readings to stabilize.
10. Use or to highlight each digit in turn; then use or to increase or
decrease the value until the concentration displayed matches the cylinder you are
using. Once the concentration has been set, Press to accept.
Figure 5-11: Verify Input
11. The VERIFY SPAN screen will be visible while the analyzer makes the verification
measurement.
Allow the concentration readings to stabilize. After the purge time set has elapsed,
the RESULT screen will then be shown.
46Rosemount CT4400
Page 47
Reference ManualVerifying gas concentrations
00809-0100-4244January 2020
Figure 5-12: Verify Span
12. This RESULT screen gives a reading of the concentration for the selected gas that is
present as an impurity in the calibration gas. In the example the gas being
measured is NH3, and the analyzer has detected a concentration of 0.510 ppm.
Figure 5-13: Result Screen
If the reading is within tolerance, no further action is required.
If the reading is outside tolerance, follow the zero calibration procedure in Zero
calibration
13. Press to end the span verification process
Reference Manual47
Page 48
Verifying gas concentrationsReference Manual
January 202000809-0100-4244
14. Press the appropriate button
OptionDescription
To return to the Gas Service Screen.
To select the next gas from the SelectGas screen.
To return to the Main Menu.
To return to the Gas Sensor Main screen
(Figure 4-4).
Figure 5-14: Next Gas Service
48Rosemount CT4400
Page 49
Reference ManualGas calibration
00809-0100-4244January 2020
6Gas calibration
6.1Required tools
The gas concentrations measured by the analyzer can be validated against a known
sample gas or calibrated to match it by using the following gas calibration procedure.
To calibrate the analyzer, you need the following items:
• Nitrogen gas of instrument gas purity for use as a zero calibration gas
• Suitable span calibration gases for each gas measured
• Gas bottle pressure regulators
• Interconnecting hoses to connect the gas bottles to the analyzer
• A T-piece and excess flow line
NOTICE
In the case of gases, such as H2O, for which it is not normally possible to obtain calibrated
gas cylinders, it will usually be measured by the same laser as some other gas. Validating
the other gases measured by the analyzer (particularly any which are measured by the
same laser as H2O) can demonstrate that the system is functioning correctly, meaning
there is no need to calibrate the water measurement directly.
If calibration must be carried out (e.g., for legal requirements), it can be calibrated by
using a water vapor generator to supply a known concentration of water vapor.
6.2Gas service menu definitions
The GAS SERVICE menu screen can be accessed from the Main Menu screen.
This gives the operator access to a number of software routines and screens that are used
for calibration and maintenance of the analyzer.
Where more options are available than can be displayed on the screen use the
keys to scroll up or down onto the next page. The screen will continually wrap around until
the option required is highlighted.
Pressing the Forward key will move to the next screen where available as shown in the
GAS Service/Per Gas example below.
Pressing the screen Refresh key at any time will refresh the current display screen if
required by the operator.
or
PER GAS
CALIBR ALL
VERIFY ALL
Reference Manual49
allows for gas calibration or verification on a gas by gas basis.
not an option available for the Rosemount CT4400 analyzer.
not an option available for the Rosemount CT4400 analyzer.
Page 50
Gas calibrationReference Manual
January 202000809-0100-4244
CALIBR PARAMS
VERIFY PARAMS
STREAM SWITCH
Pressing the Back key will return to the previous screen. Pressing the Home key at
any time will return to the Main Menu screen.
6.3Calibration
Calibration flows the known cylinder gas through the analyzer and then adjusts the
readout until the measurement matches the cylinder. The analyzer then applies this
adjustment to all measurements until the next calibration. There are two circumstances in
which you may want to calibrate:
1. After you verify gas concentrations (see Verifying gas concentrations) and find that
the analyzer readings are outside of tolerance.
2. At regular intervals, such as once a month.
6.3.1
Zero calibration
The zero calibration procedure adjusts for any zero offset in the measurement. Zero
calibration is done by calibrating the analyzer measurements against a known sample gas
using the following procedure.
allows configuration of the purge time for Gas Calibration.
allows configuration of the purge time for Gas Verification.
not an option available for the Rosemount CT4400 analyzer.
Prerequisites
Run the analyzer at a stable temperature for at least thirty minutes prior to commencing
this procedure.
Procedure
1. Use nitrogen gas of "instrument gas" purity as the zero calibration gas.
2. Ensure that a pressure regulator is connected to the nitrogen gas bottle.
3. Connect a hose from the nitrogen gas bottle through a T-piece to the sample supply
port on the rear of the analyzer.
4. Connect an excess flow line to the unused port on the T-piece and route the excess
flow line to a suitable extractor.
5. On the display controller of the analyzer, browse to the Main menu as described in
Gas service menu definitions.
50Rosemount CT4400
Page 51
Reference ManualGas calibration
00809-0100-4244January 2020
6. Using and , select GAS SERVICE. Press .
Figure 6-1: Main Menu
7. Using and , select PER GAS. Press .
Figure 6-2: Per Gas
Reference Manual51
Page 52
Gas calibrationReference Manual
January 202000809-0100-4244
8. Scroll or and select from the list the gas to be tested. Press .
Figure 6-3: Select Gas Screen
Manually adjust the ZERO OFFSET
Procedure
1. Use
Figure 6-4: Zero Offset
or , select ZERO OFFSET. Press .
52Rosemount CT4400
Page 53
Reference ManualGas calibration
00809-0100-4244January 2020
2. Use or or as required to set required value, Press to confrim.
Figure 6-5: Gas Compensation - Zero Offset
ZERO CALIBRATE
Procedure
1. Press to start the ZERO CALIBRATION process.
Figure 6-6: Run Zero Calibration
Reference Manual53
Page 54
Gas calibrationReference Manual
January 202000809-0100-4244
2. Allow the concentration to stabilize.
Figure 6-7: Calibrate Zero for NH3
This screen gives a reading of the concentration of the selected gas that is present
as an impurity in the nitrogen calibration gas, as measured by the analyzer.
In the example shown the gas being measured is NH3 (ammonia), and the analyzer
has detected a concentration of 0.510 ppm (vol).
3. To abort the calibration process, press ▲ or ▼, otherwise the analyzer will perform
the ZERO CALIBRATION, either after waiting for the purge time, or when ► or is
pressed, whichever is sooner. The result of the calibration will then be displayed, as
below.
Figure 6-8: Result
4. Press the appropriate button
Option
Description
To return to the Gas Service Screen.
54Rosemount CT4400
Page 55
Reference ManualGas calibration
00809-0100-4244January 2020
OptionDescription
To select the next gas from the Select
Gas screen.
To return to the Main Menu.
To return to the Gas Sensor Main screen
(Figure 4-4).
Figure 6-9: Next Gas Service
6.3.2Span calibration
The span gas concentrations measured by the analyzer when reference gas is flowing can
be verified and, if necessary, calibrated against the known reference gas by using the
following procedure.
Prerequisites
Run the analyzer at a stable temperature for at least thirty minutes prior to commencing
this procedure.
Procedure
1. Use a certified reference gas cylinder as the source of the span calibration gas.
2. Ensure that a suitably rated pressure regulator is connected to the reference gas
bottle.
3. Connect a hose from the reference gas bottle, through a T-piece, to the sample
supply port on the rear panel of the analyzer.
4. Connect an excess flow line to the unused port on the T-piece and route the excess
flow line to a suitable extractor.
5. On the display controller of the analyzer select the Main Menu screen.
Reference Manual55
Page 56
Gas calibrationReference Manual
January 202000809-0100-4244
6. Use or and select GAS SERVICE. Press .
Figure 6-10: Select Gas Service
7. Use or and select PER GAS. Press .
Figure 6-11: Per Gas
56Rosemount CT4400
Page 57
Reference ManualGas calibration
00809-0100-4244January 2020
8. Use or and select from the list the gas to be tested. Press .
Figure 6-12: Select Gas
9. To manually adjust the
.
Figure 6-13: Span Factor
SPAN FACTOR, use
and to select SPAN FACTOR. Press
Reference Manual57
Page 58
Gas calibrationReference Manual
January 202000809-0100-4244
10. Use or or as required to set the required value. Press to confirm.
Figure 6-14: Gas Compensation Span Factor
Span calibrate
Procedure
1. Use to start the SPAN CALIBRATION process.
Figure 6-15: Run Span
58Rosemount CT4400
Page 59
Reference ManualGas calibration
00809-0100-4244January 2020
2. Use or or as required to set the calibration value for the test cylinder.
Press to confirm.
Figure 6-16: Verify Input
Figure 6-17: Calibrate Span Finish
Reference Manual59
Page 60
Gas calibrationReference Manual
January 202000809-0100-4244
3. To abort the calibration process, press or , otherwise the analyzer will
perform the SPAN CALIBRATION, either after waiting for the purge time, or when
or is pressed, whichever is sooner. The result of the calibration will then be
displayed, as below.
Figure 6-18: Span Calibration Result
4. Press the appropriate button.
Option
Description
To return to the Gas Service Screen.
To select the Next Gas from the SelectGas screen.
To return to the Main Menu.
To return to the Gas Sensor Main screen
(Figure 4-4).
60Rosemount CT4400
Page 61
Reference ManualGas calibration
00809-0100-4244January 2020
Figure 6-19: Next Gas Service Finish
Change the gas service calibration settings
The gas purge time can be adjusted as follows:
Procedure
1. Use
Figure 6-20: Select Gas Service
or and select GAS SERVICE. Press .
Reference Manual61
Page 62
Gas calibrationReference Manual
January 202000809-0100-4244
2. Use or and select CALIBR PARAMS. Press .
Figure 6-21: Gas Service
3. To adjust the length of time for the calibration gas to flush the cell use or to
select PURGE TIME. Press .
Note
AUTO PERIOD is not available on the Rosemount CT4400.
Figure 6-22: Calibrate Parameters
4. Use the or to select the number or duration to be changed. (seconds/
minutes/hours/days).
5. Use the or to change the number or duration value.
62Rosemount CT4400
Page 63
Reference ManualGas calibration
00809-0100-4244January 2020
6. Press to accept the change or HOME Key to cancel.
Figure 6-23: Purge Time
6.4Reference gas - suggested concentration
ranges
Calibration gas ranges are recommended to be between 80 percent and 100 percent of
the full range of the analyzer. Please contact your Rosemount Customer Care
Representative if you require further assistance at (cascade.support@emerson.com).
Reference Manual63
Page 64
Gas calibrationReference Manual
January 202000809-0100-4244
64Rosemount CT4400
Page 65
Reference ManualTroubleshooting and diagnostics
00809-0100-4244January 2020
7Troubleshooting and diagnostics
7.1Troubleshooting, repairs, and failure
diagnostics
The analyzer is specifically designed to run unattended for long periods, to automatically
resolve system issues, and to recover from power failures and return to a normal working
state without intervention.
This troubleshooting guide is intended to assist maintenance personnel when the analyzer
has not appeared to be working normally for a period of more than five minutes. If the
procedures in this chapter fail to return the analyzer to normal operations or do not
identify a fault, notify your Rosemount Customer Care Representative
(cascade.support@emerson.com) at which point the analyzer may require to be returned
for detailed examination and fault finding at a service center.
Observe and obey all safety precautions when performing preventative maintenance on
the analyzer.
Important
The troubleshooting and failure diagnosis procedures described in this chapter assume
that any host equipment provided by you is fully functional. Always make sure that the
host equipment is fully serviceable before performing failure diagnosis on the analyzer.
Failure diagnosis, repair, and maintenance must only be performed by:
• Maintenance engineers who have the necessary skills, and who have been authorized
to perform maintenance on the analyzer.
• Emerson customer care engineers.
In all the cases described in this chapter, maintenance personnel must perform the repairs
by directly replacing the faulty item with a known serviceable spare part supplied by
Emerson. All other items must be repaired or replaced by the manufacturer.
CAUTION
EQUIPMENT DAMAGE
In the event of a power failure any residual sample gas should be purged from the analyzer
to ensure any corrosive gas elements are not left to contaminate/degrade the internal
components. Failure to do so could result in irreparable damage to the analyzer.
Reference Manual65
Page 66
Troubleshooting and diagnosticsReference Manual
January 202000809-0100-4244
DANGER
ELECTRIC SHOCK
The analyzer operates using mains voltage that is dangerous to life. Make sure that the
power ON/OFF switch at the rear of the panel is set to OFF and the power cable plug is
removed.
The analyzer MUST be earthed using a grounded three core cable suitable for the rating
defined in the General Characteristics (Table 1-3).
Death or personal injury may result if this is not observed.
WARNING
HEAVY ITEM
Handle the analyzer with caution during unpacking, installing, maintaining, and
transporting to prevent crushing of hands, feet, or other body parts.
The analyzer weighs 33 lb (15 kg) for the half rack system and 66 lb (30 kg) for the full rack
system. Rosemount recommends that the analyzer is moved by a minimum of two people
when lifting and transporting the equipment.
Failure to properly handle the analyzer may cause injury to personnel, wear suitable
protective gloves and protective footwear.
WARNING
FIRE AND EXPLOSION
Always lock-out/tag-out the gas handling system when shutting down the analyzer.
Unauthorized operation of the gas handling system when maintenance is being
performed on the analyzer or its associated pipes/hoses may result in highly flammable
gas being released, causing fire or explosion.
Failure to observe this precaution will cause death, personal injury, and/or damage to
persons.
66Rosemount CT4400
Page 67
Reference ManualTroubleshooting and diagnostics
00809-0100-4244January 2020
WARNING
HAZARDOUS SUBSTANCES
The analyzer may contain hazardous substances. Always handle the analyzer assemblies
and components with extreme caution.
Gas handling components within the analyzer will contain particulate matter residue from
the sample gases. Over the life of the analyzer, the concentration of the particulate matter
will become enriched within the gas handling components.
When performing repairs or maintenance on the analyzer:
— Handle used gas handling components with extreme caution.
— Avoid direct skin contact with used gas handling components.
— Do not smoke, drink, or eat in the work area.
— Wear goggles or eye shields.
— Wear a suitable face mask to protect against inhalation of particulate matter.
— Do not wet fingers, eyes, or any exposed skin.
— Pack used gas handling components for disposal in sealed packaging and label them
Contaminated.
— Dispose of contaminated items as hazardous material according to the applicable local,
national, or international health and safety regulations and pollution regulations.
7.2Using the Built-In-Self Test (BIST) fault
diagnostics
The analyzer has a Built-In-Self-Test (BIST) function that can be used to perform failure
diagnosis of some functions other than a complete failure of the equipment. The BIST is
accessed and controlled through the display controller mounted on the front panel of the
analyzer. The BIST runs in the background when the analyzer is operating and continuously
monitors the analyzer for faults.
If the BIST detects a fault the STATUS display in the bottom left-hand corner of the MainScreen (Faults screen selection) will change from OK to a flashing .
Analysis of the fault detected can be performed using the indicated Fault menu screen or
the Diagnostics menu screen.
7.2.1
Fault menu
The BIST function can be used to perform failure diagnosis of some functions in situations
where there is a fault other than a complete failure of the equipment. The BIST is accessed
and controlled through the FAULTS screen accessed from the Main Menu screen mounted
on the front panel. This gives the operator access to any assess any Faults that the analyzer
has detected in the systems being monitored.
Reference Manual67
Page 68
Troubleshooting and diagnosticsReference Manual
January 202000809-0100-4244
Faults screen selection
Procedure
1. Where more Faults are available than can be displayed on the screen use the or
keys to scroll up or down onto the next page. The screen will continually wrap
around until the Fault required is highlighted.
2. Pressing the Forward key will move to the next screen where available as shown
in the Diagnostics/Faults example below. Pressing the screen Refresh key
at any
time will refresh the current display screen if required by the operator. Pressing the
Back key will return to the previous screen. Pressing the Home key at any
time will return to the Main Menu screen.
3. You can see the STATUS information from either the Gas Sensor Main screen (Figure
4-4) or the Pressure and Temperature screen (Figure 4-5).
• When the analyzer is running correctly, the STATUS displays Running and OK.
• If a fault is detected the STATUS display in the bottom left-hand corner of the
screen changes from OK to a flashing
.
4. Pressing the Forward key will move to the Main menu screen.
Figure 7-1: Faults Screen Selection
68Rosemount CT4400
Page 69
Reference ManualTroubleshooting and diagnostics
00809-0100-4244January 2020
5. Scroll or and select Faults. Press .
Figure 7-2: Main Menu - Faults Screen
6. Scroll and select the component to obtain details on the next screen. Press
to obtain details about the fault
Figure 7-3: Faults Selection Screen
Reference Manual69
Page 70
Troubleshooting and diagnosticsReference Manual
January 202000809-0100-4244
7. The example screen shows the Fit6/Path2 status is XShift Error. The XShift
value is not within the defined tolerance. When rectified, the status will change to
OK.
Figure 7-4: Fault Status Details Screen
7.2.2
8. Examine the data shown on the corresponding Faults screen to determine if a
component has failed or if it is operating outside of its correct parameters.
9. Contact your Emerson Customer Care Representative
(cascade.support@emerson.com) if you need further assistance with a
component's Fault Status.
Diagnostics menu screen
The Diagnostics screen lists those main components of the analyzer where problems can
be diagnosed using the BIST system. The Diagnostics screen also enables you to check on
the status and, where appropriate, the values of any of the listed components.
The Diagnostics menu screen can be accessed from the Main Menu screen.
This gives the operator access to a number of options when checking the overall health of
the analyzer.
Procedure
1. Where more options are available than can be displayed on the screen use the
or keys to scroll up or down onto the next page. The screen will continually wrap
around until the option required is highlighted.
2. Pressing the Forward key will move to the next screen where available as shown
in the Diagnostics/Config example below. Pressing the screen Refresh key at any
time will refresh the current display screen if required by the operator.
CONFIG
SENSORgives the status of the analyzer SENSOR(s) engine.
70Rosemount CT4400
gives the status of the analyzer CONFIGURATION file.
Page 71
Reference ManualTroubleshooting and diagnostics
00809-0100-4244January 2020
LASER XXXprovides the operator with additional information on each
LASER that the analyzer has been set up to monitor.
PATH XXXprovides the operator with information on the laser PATH for
each gas being monitored by the analyzer.
FIT XXX/PATH
XXX
provides the operator with information on the FIT/PATH for
each gas being monitored confirming if the gas concentration
measured is within specification.
BEKA RS232NOT AN OPERATOR REQUIRED OPTION.
CELL HEATERprovides the operator with information on the CELL HEATER.
XSTREAMNOT AN OPERATOR REQUIRED OPTION.
3. Pressing the Back key will return to the previous screen. Pressing the Home key
at any time will return to the Main Menu screen.
4. Scroll or and select DIAGNOSTICS. Press .
Figure 7-5: Main Menu
Example 1: Use or to view all the options available. Select the Item required
to obtain further details on the next screen. Press .
Reference Manual71
Page 72
Troubleshooting and diagnosticsReference Manual
January 202000809-0100-4244
Figure 7-6: Diagnostics Screen - CONFIG Selection
The example screen below shows the CONFIG option for the analyzer has been
selected and the current Status is OK.
Figure 7-7: Config Status
Example 2: Use or to view all the options available. Select the Item required
to obtain further details on the next screen. Press .
The example screen above shows the Laser 1 option for the analyzer has been
selected and the current Status is OK.
Figure 7-9: Laser 1 Status
Note
Additional operating information has been provided for this option.
7.3Configuring the 4-20mA outputs
The DATA SERVICE menu screen can be accessed from the Main Menu screen.
This gives the operator access to set the 4-20 mA output ranges for the gas being
analyzed.
Reference Manual73
Page 74
Troubleshooting and diagnosticsReference Manual
January 202000809-0100-4244
Procedure
1. Where more options are available than can be displayed on the screen use the
or keys to scroll up or down onto the next page. The screen will continually wrap
around until the option required is highlighted.
2. Pressing the Forward key will move to the next screen where available as shown
in the Data Service/Analog Ranges example below.
3. Pressing the screen Refresh key at any time will refresh the current display
screen if required by the operator.
ANALOG RANGES allows the operator to set the upper and lower limits for the
4-20 mA measurement output.
USB DATA DUMP NOT an option available for the Rosemount CT4400 analyzer.
TFTP TRANSFERNOT an option available on the Rosemount CT4400 analyzer.
4. Pressing the Back key will return to the previous screen.
5. Pressing the Home key at any time will return to the Main Menu screen.
6. Scroll or and select DATA SERVICE. Press .
Figure 7-10: Main Menu - Data Service
7. Scroll or and select ANALOG RANGES. Press .
74Rosemount CT4400
Page 75
Reference ManualTroubleshooting and diagnostics
00809-0100-4244January 2020
Figure 7-11: Data Service - Analog Ranges
8. Scroll or and select from the list of gases being tested. Press .
Figure 7-12: Analog Range Gas Selection Screen
9. Select the MIN option. Press .
Reference Manual75
Page 76
Troubleshooting and diagnosticsReference Manual
January 202000809-0100-4244
Figure 7-13: MIN Option Selected
10. Use the or to select the number to be changed.
11. Use the
or to change the number value.
Figure 7-14: Minimum Limit
12. Press to accept the change Home key to cancel.
13. Select the MAX option. Press .
76Rosemount CT4400
Page 77
Reference ManualTroubleshooting and diagnostics
00809-0100-4244January 2020
Figure 7-15: MAX Option Selected
14. Use the or to select the number to be changed.
15. Use the or to change the number value.
Figure 7-16: Maximum Limit
16. Press to accept the change Home key to cancel.
Reference Manual77
Page 78
Troubleshooting and diagnosticsReference Manual
January 202000809-0100-4244
7.4Visual examination
A visual examination of the analyzer is recommended as the next step in failure diagnosis if
the BIST fails to identify the problem.
DANGER
ELECTRIC SHOCK
The analyzer operates using mains voltage that is dangerous to life. Make sure that the
power ON/OFF switch at the rear of the panel is set to OFF and the power cable plug is
removed.
The analyzer MUST be earthed using a grounded three core cable suitable for the rating
defined in the General Characteristics (Table 1-3).
Death or personal injury may result if this is not observed.
Procedure
1. Power down the analyzer and allow it to cool.
2. Visually examine the exterior of the analyzer for signs of damage.
3. Perform a visual inspection of the electrical and gas connections on the rear panel of
the analyzer.
4. If any loose connections are found tighten as required.
7.5Failure diagnostics
If the BIST and the visual examination fail to identify the fault, perform the failure
diagnostics and recommended actions.
7.5.1
Gas reading abnormal
Potential cause
Gas reading abnormal
Recommended action
1. Check measurement validity flag.
2. Reading invalid.
3. refer to Measurement invalid flag.
Potential cause
Gas reading abnormal
Recommended action
1. Check measurement validity flag.
2. Reading valid.
78Rosemount CT4400
Page 79
Reference ManualTroubleshooting and diagnostics
00809-0100-4244January 2020
3. Perform a new Zero and Span calibration to ensure calibration factors are correct.
4. Readings return to normal.
5. Sensor required calibration.
Potential cause
Gas reading abnormal
Recommended action
1. Check measurement validity flag.
2. Reading valid.
3. Perform a new Zero and Span calibration to ensure calibration factors are correct.
4. Abnormal readings persist.
5. Cause Unknown.
6. Contact (cascade.support@emerson.com).
Related information
Measurement invalid flag.
7.5.2
Calibration/validation drift
Potential cause
Sensor reading invalid.
Recommended action
1. Check measurement validity flag.
2. Reading is invalid.
3. refer to Measurement invalid flag.
Potential cause
Sensor required calibration.
Recommended action
1. Check measurement validity flag.
2. Reading is valid.
3. Perform a new Zero and Span calibration to ensure calibration factors are correct.
4. Readings return to normal.
5. Sensor required calibration.
Potential cause
Sensor does not require calibration.
Recommended action
1. Check measurement validity flag.
2. Reading is valid.
Reference Manual79
Page 80
Troubleshooting and diagnosticsReference Manual
January 202000809-0100-4244
3. Perform a new Zero and Span calibration to ensure calibration factors are correct.
4. Readings do not return to normal.
5. Check calibration cylinder.
6. Calibration cylinder near empty.
7. Replace calibration cylinder.
8. Readings return to normal.
9. Cylinder was empty.
Potential cause
Calibration cylinder empty.
Recommended action
1. Check measurement validity flag.
2. Reading is valid.
3. Perform a new Zero and Span calibration to ensure calibration factors are correct.
4. Readings do not return to normal.
5. Check calibration cylinder.
6. Calibration cylinder near empty.
7. Replace calibration cylinder.
8. Readings do not return to normal.
9. Check the calibration pipework.
10. Pipework damaged.
11. Replace pipework.
12. Readings return to normal.
13. Pipework was damaged.
Potential cause
Pipework damaged empty.
Recommended action
1. Check measurement validity flag.
2. Reading is valid.
3. Perform a new Zero and Span calibration to ensure calibration factors are correct.
4. Readings do not return to normal.
5. Check calibration cylinder.
6. Calibration cylinder near empty.
7. Replace calibration cylinder.
8. Readings do not return to normal.
9. Check the calibration pipework.
10. Pipework damaged.
11. Replace pipework.
12. Readings do not return to normal.
80Rosemount CT4400
Page 81
Reference ManualTroubleshooting and diagnostics
00809-0100-4244January 2020
13. Cause Unknown.
14. Contact (cascade.support@emerson.com).
Related information
Measurement invalid flag
7.5.3Measurement invalid flag
Potential cause
Pressure or temperature is out of range.
Recommended actions
1. Check display controller to assess which measurements are invalid.
2. All gas readouts are invalid.
3. Use the display controller to check pressure and cell temperature.
4. Pressure and temperature are both within range.
5. Hardware failure.
6. Contact (cascade.support@emerson.com).
Potential cause
Pressure or temperature is out of range.
Recommended action
1. Check display controller to assess which measurements are invalid.
2. All gas readouts are invalid.
3. Use the display controller to check pressure and cell temperature.
4. Pressure is out of range.
5. Refer to Analysis cell pressure out of range.
Potential cause
Pressure or temperature is out of range.
Recommended action
1. One gas readout is invalid.
2. Refer to Single gas reading invalid
Related information
Single gas reading invalid
Analysis cell pressure out of range
Reference Manual81
Page 82
Troubleshooting and diagnosticsReference Manual
January 202000809-0100-4244
7.5.4Single gas reading invalid
Potential cause
Gas reading invalid
Recommended actions
1. Check which gas.
2. Check fault readings.
3. Fault reading active.
4. Use Fault to access further diagnostic information
5. Contact (cascade.support@emerson.com).
Potential cause
Gas reading invalid
Recommended action
1. Check which gas.
2. Check fault readings.
3. Fault reading not displayed.
4. Contact (cascade.support@emerson.com).
7.5.5
7.5.6
Analysis cell pressure out of range
Potential cause
Pressure was out of range.
Recommended action
1. Adjust inlet valve to change pressure.
2. Pressure returns to normal.
3. No further action required.
Potential cause
Pressure was out of range.
Recommended action
1. Adjust inlet valve to change pressure.
2. If pressure is unstable or impossible to adjust.
3. Contact (cascade.support@emerson.com).
Sensor frozen or unresponsive
82Rosemount CT4400
Page 83
Reference ManualTroubleshooting and diagnostics
00809-0100-4244January 2020
Potential cause
Computer freeze-up or connection failure.
Recommended actions
1. Attempt a reboot.
2. System reboots successfully.
3. Contact (cascade.support@emerson.com).
Potential cause
PC or HMI failure.
Recommended actions
1. Attempt a reboot.
2. System does not reboot.
3. Contact Cascade (cascade.support@emerson.com).
7.5.7
Other error not covered in other sections
Potential cause
Unknown
Recommended actions
Contact Cascade (cascade.support@emerson.com).
7.6Replacing the power socket fuses
The analyzer contains two external fuses located in the power socket attached to the rear
panel.
DANGER
ELECTRIC SHOCK
The analyzer operates using mains voltage that is dangerous to life. Make sure that the
power ON/OFF switch at the rear of the panel is set to OFF and the power cable plug is
removed.
The analyzer MUST be earthed using a grounded three core cable suitable for the rating
defined in the General Characteristics (Table 1-3).
Death or personal injury may result if this is not observed.
Procedure
1. Turn the ON/OFF switch on the back panel to OFF.
2. Disconnect the external power supply from the power socket.
3. Allow the analyzer to cool.
Reference Manual83
Page 84
Troubleshooting and diagnosticsReference Manual
January 202000809-0100-4244
WARNING
ELECTRIC SHOCK
Fuses MUST only be replaced with the fuses of the same type and rating as those
being removed (6.3A 250 VAC ceramic cartridge fuse, 5x20 mm, Speed F. IEC
60127 compliant). Failure to do so may result in personnel injury and or equipment
damage.
4. Carefully lever open the fuse cover (B) to gain access to the fuses (C).
Figure 7-17: Power Socket Fuses
A. Power socket
B. Fuse cover/fuse holder
C. Fuse
5. Remove the fuse (C) from the holder (B).
6. Fit a new fuse (C), of the same type and rating, into the fuse holder (B) (6.3A 250
VAC ceramic cartridge fuse, 5x20 mm, Speed F. IEC 60127 compliant).
7. Repeat for the second fuse.
8. Fully push the fuse holder (B) back into position in the power socket (A).
9. Reconnect the external power supply to the power socket on the back panel.
10. Start up the analyzer by turning the ON/OFF switch on the back panel to the ON
position.
11. The analyzer will automatically begin the startup process.
84Rosemount CT4400
Page 85
Reference ManualTroubleshooting and diagnostics
00809-0100-4244January 2020
7.7Cleaning the analysis cell mirrors
Cleaning the analysis cell mirrors will only be required if the sample gas is not conditioned
in accordance with gas requirements defined in the General characteristics Table 1-3 and
the Environmental characteristics Table 1-4.
For example, it is possible that the mirrors inside the cell may become contaminated if
contaminants are carried into the cell in water droplets in the gas being sampled if it is not
conditioned correctly the mirrors would then require to be cleaned to remove any
contamination from them.
The mirror surface is highly reflective and susceptible to damage if cleaned incorrectly or
excessively. Damage to the reflective coating on the mirrors WILL affect the analyzer
performance. Mirrors found to be damaged due to excessive cleaning or from corrosive
material residue will invalidate the warranty for the analyzer.
If the cell mirrors must be cleaned the analysis cell must only be opened in a CLEAN/DRY/DUST FREE area to prevent further contamination.
Observe all safety precautions before starting this procedure.
DANGER
ELECTRIC SHOCK
The analyzer operates using mains voltage that is dangerous to life. Make sure that the
power ON/OFF switch at the rear of the panel is set to OFF and the power cable plug is
removed.
The analyzer MUST be earthed using a grounded three core cable suitable for the rating
defined in the General Characteristics (Table 1-3).
Death or personal injury may result if this is not observed.
DANGER
FAILURE TO LOCK-OUT THE GAS HANDLING SYSTEM
Always lock out the gas handling system when shutting down the analyzer. Unauthorized
operation of the gas handling system when maintenance is being performed on the
analyzer or its associated pipes/hoses may result in highly flammable gas being released,
causing fire or explosion.
Death or personal injury may result if this is not observed.
Reference Manual85
Page 86
Troubleshooting and diagnosticsReference Manual
January 202000809-0100-4244
DANGER
FAILURE TO VENT SAMPLE GAS
The sample gas in the system must be vented to prevent fire or explosion during
maintenance and to prevent damage to the analyzer during startup.
The sample gas in the pipes leading to the analyzer must be purged to prevent hazards to
personnel during maintenance. Purging the sample gas must be done in accordance with
the safe working procedures for the site.
Allow the analyzer and system for returning the sample gas to run for five minutes to allow
any sample gas in the system to be returned to the exhaust.
WARNING
HEAVY ITEM
Handle the analyzer with caution during unpacking, installing, maintaining, and
transporting to prevent crushing of hands, feet, or other body parts.
The analyzer weighs 33 lb (15 kg) for the half rack system and 66 lb (30 kg) for the full rack
system. Rosemount recommends that the analyzer is moved by a minimum of two people
when lifting and transporting the equipment.
Failure to properly handle the analyzer may cause injury to personnel, wear suitable
protective gloves and protective footwear.
Procedure
1. Turn the ON/OFF switch on the back panel to OFF.
2. Disconnect the external power supply from the power socket.
3. VERIFY the sample gas handling system has been shut down and locked-out/
tagged-out.
4. The sample gas supply in the system MUST be vented to prevent fire or explosion
during maintenance.
5. Allow the analyzer to cool.
6. Disconnect all external cables. (Tag if necessary).
7. Disconnect, cap and label the tubing at the back of the analyzer connecting to the
sample handling system.
8. Remove the analyzer from the rack and place on a flat, stable surface that can
accommodate the weight of 33 lb (15kg) for the half rack and 66 lb (30 kg) for the
full rack system. This step requires two people to lift and transport the analyzer.
86Rosemount CT4400
Page 87
Reference ManualTroubleshooting and diagnostics
00809-0100-4244January 2020
Figure 7-18: Removing the Top Cover
9. The area selected MUST be clean and dust free.
10. Remove and retain the 12 off (24 off) M3 button head screws and plain washers
used to secure the top cover(s).
11. Lift the top cover(s) clear of the chassis assembly.
12. Examine the cover for signs of physical damage, inside and out.
13. Locate the two cam lock screws securing the hinged PCB tray, rotate a ¼ turn to
release.
Figure 7-19: Releasing the PCB Tray
Reference Manual87
Page 88
Troubleshooting and diagnosticsReference Manual
January 202000809-0100-4244
14. Carefully swing open the hinged PCB tray to gain access to the analysis cell. The
second analysis cell fitted in a Full rack system was fully accessible once the outer
cover was removed.
Figure 7-20: Accessing the Analysis Cell
15. Using an Allen key, remove and retain the six socket head cap screws and washers
(D) used to secure the cell cover (C) on the analysis cell (B). Care MUST be taken to
ensure the O-Ring (A) does not get damaged.
16. It may be necessary to use a small flat blade screwdriver to lever the cell cover off.
With the cover removed, the mirrors are now exposed.
88Rosemount CT4400
Page 89
Reference ManualTroubleshooting and diagnostics
00809-0100-4244January 2020
Figure 7-21: Opening the Analysis Cell
A. Analysis cell
B. O-ring
C. Cell cover
D. Fasteners
17. Clean the mirrors by applying a small quantity of reagent grade methanol to a
clean, dry, lens tissue and then gently wiping the mirrors with the lens tissue. Care
MUST be taken not to apply pressure to the lens during this operation to prevent
scratching the reflective surface.
Scratches to the mirror surface WILL affect the ability for the analyzer to operate
and WILL NOT be covered by the warranty for the analyzer. (If reagent grade
methanol is unavailable, then reagent grade IPA may be used instead.)
Reference Manual89
Page 90
Troubleshooting and diagnosticsReference Manual
January 202000809-0100-4244
Figure 7-22: Cleaning the Lens
NOTICE
The flow of gas through the cell means that any water droplets introduced into the
system will travel through the cell to the front mirror; therefore, it is the front mirror
that is more likely to be contaminated. The mirrors are coated with enhanced nickel
and should be clean and reflective. Cleaning the mirrors with methanol may not
visibly improve the mirror surface but should improve reflectivity in the midinfrared band used by the lasers.
18. Refit the cell cover (C) onto the analysis cell (B) and secure with the six socket head
cap screws and washers (D) removed previously.
19. Close the hinged PCB tray and secure in position using the two cam lock screws.
20. Refit the top cover and reinstall the analyzer in the rack.
90Rosemount CT4400
Page 91
Reference ManualShutdown procedure
00809-0100-4244January 2020
8Shutdown procedure
This chapter describes the procedure to be followed when shutting down the analyzer.
DANGER
ELECTRIC SHOCK
The analyzer operates using mains voltage that is dangerous to life. Make sure that the
power ON/OFF switch at the rear of the panel is set to OFF and the power cable plug is
removed.
The analyzer MUST be earthed using a grounded three core cable suitable for the rating
defined in the General Characteristics (Table 1-3).
Death or personal injury may result if this is not observed.
DANGER
FAILURE TO LOCK-OUT THE GAS HANDLING SYSTEM
Always lock out the gas handling system when shutting down the analyzer. Unauthorized
operation of the gas handling system when maintenance is being performed on the
analyzer or its associated pipes/hoses may result in highly flammable gas being released,
causing fire or explosion.
Death or personal injury may result if this is not observed.
DANGER
FAILURE TO VENT SAMPLE GAS
The sample gas in the system must be vented to prevent fire or explosion during
maintenance and to prevent damage to the analyzer during startup.
The sample gas in the pipes leading to the analyzer must be purged to prevent hazards to
personnel during maintenance. Purging the sample gas must be done in accordance with
the safe working procedures for the site.
Allow the analyzer and system for returning the sample gas to run for five minutes to allow
any sample gas in the system to be returned to the exhaust.
Reference Manual91
Page 92
Shutdown procedureReference Manual
January 202000809-0100-4244
CAUTION
EQUIPMENT DAMAGE
Always follow this shutdown procedure. Damage to the analyzer may result from a failure
to follow this procedure.
The analyzer normally operates continuously. It should only be necessary to shut down the
analyzer in the following circumstances:
— In order to perform repairs or maintenance on the analyzer.
— When the analyzer has to be switched off as part of a plant shutdown or plant maintenance.
Use the display controller to perform the shutdown procedure.
Operation of the analyzer is controlled through the display controller located on the front
panel of the analyzer.
NOTICE
The gas concentrations shown in the following screenshots may be different from those
shown in your particular analyzer. The screenshots indicate the functionality of the
software, which is the same regardless of the gases or gas concentrations being measured.
Procedure
1. Shut down the gas handling system that conditions the sample gas and feeds it to
the analyzer.
operation during maintenance, which could cause an escape of gas.
Lock-out the gas handling system to prevent
its unauthorized
WARNING
HIGH PRESSURE GAS AND AIR
The gas supply operates at a pressure that can cause injury, e.g., damage to eyes
and skin punctures from debris blown by high pressure gas or compressed air.
2. Purge any sample gas in the pipe/tube from the gas handling system to the analyzer
using factory air or a regulated nitrogen supply.
3. Allow the analyzer to run for five minutes with the purge gas connected, so that any
sample gas in the analyzer is vented to the exhaust.
4. Use the display controller on the front of the analyzer to perform the shutdown
process.
5. On the display controller, check that the gas concentration reads 0 ppm or 0%
before stopping the purge gas.
6. Turn off the purge gas supply to the analyzer. Lock-out the purge to prevent its
unauthorized operation.
7. The Shutdown procedure can be started as follows.
The SYSTEM menu screen can be accessed from the Main Menu screen.
92Rosemount CT4400
Page 93
Reference ManualShutdown procedure
00809-0100-4244January 2020
This gives the operator access to the Shutdown option for the analyzer which will
close the software. The other options are factory menus that the operator WILL NOT
need to access but are available for a Rosemount service technician during any
planned future maintenance.
8. Where more options are available than can be displayed on the screen use the
or keys to scroll up or down onto the next page. The screen will continually wrap
around until the option required is highlighted.
9. Pressing the Forward key will move to the next screen where available as shown
in the Data Service/Shutdown example below. Pressing the screen Refresh key at
any time will refresh the current display screen if required by the operator.
SAVE CONFIGNOT AN OPERATOR OPTION - saves the current
configuration to the SD card.
SHUTDOWNallows the OPERATOR to shutdown the software in a
controlled manner.
SHOW VERSIONSdisplays the software version installed on the analyzer.
NETWORK INFONOT AN OPERATOR OPTION – displays the network
information.
CHANGE IP ADDRESS NOT AN OPERATOR OPTION – displays one of the IP
address.
10. Pressing the Back key
will return to the previous screen.
11. Pressing the Home key at any time will return to the Main Menu screen.
12. Press or and select SYSTEM. Press .
Figure 8-1: Main Menu
Reference Manual93
Page 94
Shutdown procedureReference Manual
January 202000809-0100-4244
13. Press or and select SHUTDOWN. Press .
Figure 8-2: System
14. Press to accept. Press to reject.
Figure 8-3: Shutdown
94Rosemount CT4400
Page 95
Reference ManualShutdown procedure
00809-0100-4244January 2020
15. The screen WILL NOT go blank when the Shutdown has completed. Wait for the
heartbeat to stop flashing (Typically 10 -15 seconds) before turning off the power.
Figure 8-4: Shutdown
Reference Manual95
Page 96
Shutdown procedureReference Manual
January 202000809-0100-4244
96Rosemount CT4400
Page 97
Reference ManualPreventative maintenance
00809-0100-4244January 2020
9Preventative maintenance
As a minimum the analyzer MUST be visually inspected for signs of damage or corrosion at
the connections weekly to confirm the continued safe operation of the analyzer.
As a minimum the Zero and Span calibration for the analyzer MUST be run monthly to
confirm the continued accuracy of the analysis being performed. Any anomalies found
should be reported to the maintenance department and where in doubt the analyzer
should be switched off until rectified. Variation in customer sites may require these
activities to be performed more or less often than indicated.
Reference Manual97
Page 98
Preventative maintenanceReference Manual
January 202000809-0100-4244
98Rosemount CT4400
Page 99
Reference ManualRecommended spare parts
00809-0100-4244January 2020
10Recommended spare parts
It is not recommended for maintenance on the analyzer other than as defined in
Preventative maintenance to be conducted by unauthorized personnel.
Should the need arise to have the analyzer examined you MUST contact customer support
(cascade.support@emerson.com) or Emerson distribution partners for technical
assistance.
General inquiries should be sent to (cascade.support@emerson.com).
The only authorized spare parts that are available are listed in Table 10-1 below.
Table 10-1: Spare Parts List
Part NumbersDescription
P-6000-013326 mm Ferrule Nut (for sample gas connection)
P-6000-012726 mm back & front ferrules (for sample gas connection)
P-6000-00657AC power cable UK version
P-6000-00951AC power cable US version
P-6000-00952AC power cable Euro version
P-6000-01441AC power cable Chinese version
P-6000-010456.3 A 250 VAC ceramic cartridge fuse, 5x20 mm, speed F. IEC 60127
compliant
Reference Manual99
Page 100
Recommended spare partsReference Manual
January 202000809-0100-4244
100Rosemount CT4400
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.