Rosemount CCO 5500 Carbon Monoxide (CO) Analyzer-Rev 1.0 Manuals & Guides

Page 1
Rosemount™ CCO 5500
Carbon Monoxide (CO) Analyzer
Reference Manual
00809-0100-4550, Rev AA
June 2018
Page 2

Essential instructions

EmersonTM designs, manufactures, and tests its RosemountTM products to meet many national and international standards. Because these instruments are sophisticated technical products, you must properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions must be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount products. Failure to follow the proper instructions may cause any one of the following situations to occur: loss of life, personal injury, property damage, damage to this instrument, and warranty invalidation.
Read all instructions prior to installing, operating, and servicing this product. If this instruction manual is not the correct
manual, call 1-855-724-2638, and we will provide the requested manual. Save this manual for future reference.
If you do not understand any of the instructions, contact your Emerson representative for clarification.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and maintenance of the product.
Install your equipment as specified in the installation instructions of the appropriate manual and per applicable local and
national codes. Connect all products to the proper electrical and pressure sources.
To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount.
Unauthorized parts and procedures can affect the product's performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Look alike substitutions may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed
by qualified people, to prevent electrical shock and personal injury.
The information contained in this document is subject to change without notice.

Warranty

RosemountTM warrants that the equipment manufactured and sold by it will, upon shipment, be free of defects in workmanship or material. Should any failure to conform to this warranty become apparent during a period of one year after the date of shipment, Rosemount shall, upon prompt written notice from the purchaser, correct such nonconformity by repair or replacement, F.O.B. factory of the defective part or parts. Correction in the manner provided above shall constitute a fulfillment of all liabilities of Rosemount with respect to the quality of the equipment.
THE FOREGOING WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES OF QUALITY WHETHER WRITTEN, ORAL, OR IMPLIED (INCLUDING ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR PURPOSE).
The remedy(ies) provided above shall be purchase'rs sole remedy(ies) for any failure of Rosemount to comply with the warranty provisions, whether claims by the purchaser are based in contract or in tort (including negligence).
Rosemount does not warrant equipment against normal deterioration due to environment. Factors such as corrosive gases and solid particulates can be detrimental and can create the need for repair or replacement as part of normal wear and tear during the warranty period.
Equipment supplied by Rosemount but not manufactured by it will be subject to the same warranty as is extended to Rosemount by the original manufacturer.
At the time of installation, it is important that the required services are supplied to the system and that the electronic controller is set up at least to the point where it is controlling the sensor heater. This will ensure that, should there be a delay between installation and full commissioning, the sensor being supplied with ac power and reference air will not be subjected to component deterioration.

Preface

The purpose of this manual is to provide information concerning the components, functions, installation, and maintenance of the RosemountTM CCO 5500.
Some sections may describe equipment not used in your configuration. Become thoroughly familiar with the operation of this module before operating it. Read this reference manual completely.

Definitions

The following definitions apply to WARNINGS, CAUTIONS, and NOTICES found throughout this publication.
Page 3
WARNING!
Highlights an operation or maintenance procedure, condition, statement, etc. that if not strictly observed, could result in injury, death, or long-term health hazards of personnel.
CAUTION!
Highlights an operation or maintenance procedure, practice, condition, statement, etc. that if not strictly observed, could result in damage to or destruction of equipment or loss of effectiveness.
NOTICE
Highlights an essential operating procedure, condition, or statement.

Symbols

Earth (ground) terminal
:
Protective conduit or terminal
:
Risk of electrical shock
:
Warning: Refer to Instruction Manual
:
Page 4
Page 5

Contents

Contents
Chapter 1 Description and specifications ..........................................................................................1
1.1 Component checklist ..................................................................................................................... 1
1.2 Overview ........................................................................................................................................3
1.3 System description .........................................................................................................................3
1.4 Specifications .................................................................................................................................5
1.5 Rosemount™ CCO 5500 ordering information ................................................................................ 6
Chapter 2 Install ...............................................................................................................................7
2.1 Unpack the equipment ...................................................................................................................7
2.2 Safety considerations ..................................................................................................................... 7
2.3 Cable requirements ........................................................................................................................8
2.4 Selecting location ...........................................................................................................................8
2.4.1 Points to consider ............................................................................................................9
2.5 Mechanical installation ...................................................................................................................9
2.5.1 Mount flange assemblies ............................................................................................... 10
2.5.2 Isolating valves .............................................................................................................. 11
2.5.3 Purge air supply .............................................................................................................12
2.5.4 Air purge units ...............................................................................................................13
2.5.5 Source and receiver units ...............................................................................................14
2.5.6 Control unit ................................................................................................................... 15
2.5.7 Power supply unit .......................................................................................................... 16
2.6 Electrical data ...............................................................................................................................17
2.6.1 AC supplies ....................................................................................................................17
2.6.2 Outputs .........................................................................................................................17
2.6.3 Normalizing inputs ........................................................................................................ 17
2.6.4 Plant status input ...........................................................................................................17
2.7 Electrical connections .................................................................................................................. 18
2.7.1 Installation of cables ......................................................................................................19
2.7.2 Cable connections ......................................................................................................... 19
Chapter 3 Configuration and startup .............................................................................................. 25
3.1 Introduction .................................................................................................................................25
3.2 Safety considerations ................................................................................................................... 25
3.3 Power up the Rosemount™ CCO 5500 Analyzer ............................................................................ 25
3.4 Alignment .................................................................................................................................... 26
3.5 Detector levels ............................................................................................................................. 29
3.5.1 Receiver gain adjustment .............................................................................................. 30
3.5.2 Control unit gain adjustment .........................................................................................31
3.6 Source adjustments ..................................................................................................................... 33
3.6.1 Source intensity .............................................................................................................33
3.6.2 Chopper frequency ........................................................................................................34
3.7 Set up mode .................................................................................................................................35
3.7.1 Enter security code ........................................................................................................ 36
3.7.2 Set averages .................................................................................................................. 36
3.7.3 Configure O/P ................................................................................................................37
3.7.4 Parameters ....................................................................................................................40
3.7.5 Normalization ................................................................................................................43
Reference Manual i
Page 6
Contents
3.7.6 Reset averages ...............................................................................................................46
3.7.7 Calibrate ........................................................................................................................46
3.8 Current output calibration ............................................................................................................50
Chapter 4 Operation ...................................................................................................................... 51
4.1 Introduction .................................................................................................................................51
4.1.1 Measurement ................................................................................................................ 51
4.1.2 Calibration .....................................................................................................................52
4.2 Startup and operation .................................................................................................................. 52
4.3 Modes of operation ...................................................................................................................... 52
4.4 Keypad operation .........................................................................................................................53
4.5 Menu tree .................................................................................................................................... 53
4.6 Operating mode ...........................................................................................................................55
4.7 Parameters ...................................................................................................................................56
4.7.1 Identification ................................................................................................................. 56
4.7.2 Parameters .................................................................................................................... 56
4.7.3 Averages ........................................................................................................................57
4.7.4 Output ...........................................................................................................................57
4.7.5 Alarm .............................................................................................................................57
4.7.6 Plant status ....................................................................................................................57
4.8 Normalization .............................................................................................................................. 57
4.9 Diagnostic mode ..........................................................................................................................58
4.9.1 Detector outputs ........................................................................................................... 58
4.9.2 Modulation (chopper motor) frequency ........................................................................ 59
4.9.3 YVals and CO ppm ......................................................................................................... 59
4.9.4 Calibration data ............................................................................................................. 60
4.9.5 Fault condition .............................................................................................................. 60
4.10 Set up mode .................................................................................................................................61
4.10.1 Enter security code ........................................................................................................ 61
4.10.2 Set averages .................................................................................................................. 62
4.10.3 Configure O/P ................................................................................................................63
4.10.4 Parameters .................................................................................................................... 65
4.10.5 Normalization ................................................................................................................69
4.10.6 Reset averages ...............................................................................................................71
4.10.7 Calibrate ........................................................................................................................ 72
4.11 Check Cell mode .......................................................................................................................... 76
4.12 Shutdown procedure ................................................................................................................... 76
4.13 Routine checks .............................................................................................................................76
4.13.1 Notes for using a Rosemount™ check cell .......................................................................76
4.13.2 Alarms and emergency conditions .................................................................................79
4.13.3 Emergency shutdown procedure ................................................................................... 79
4.13.4 Isolation procedure ........................................................................................................79
Chapter 5 Maintenance .................................................................................................................. 81
5.1 Preventative maintenance ........................................................................................................... 81
5.1.1 Cleaning windows ..........................................................................................................81
5.2 Corrective maintenance ............................................................................................................... 82
5.2.1 Replace heater element ................................................................................................. 82
5.2.2 Replace chopper motor assembly .................................................................................. 84
5.2.3 Replace source unit gas cell ........................................................................................... 86
ii Rosemount CCO 5500
Page 7
Contents
5.2.4 Replace receiver unit gas cell ......................................................................................... 87
5.2.5 Electronics ..................................................................................................................... 88
5.3 Adjust span factor ........................................................................................................................ 88
5.3.1 Reset span factor ........................................................................................................... 89
Chapter 6 Troubleshooting ............................................................................................................ 91
6.1 Finding faults with the keypad ......................................................................................................91
6.1.1 Data valid LED out ..........................................................................................................91
6.2 Troubleshooting tables ................................................................................................................ 92
6.3 Component tests ......................................................................................................................... 96
6.3.1 Heater cartridge ............................................................................................................ 96
6.3.2 Chopper motor ..............................................................................................................96
6.4 LED indications .............................................................................................................................97
6.5 Test points ................................................................................................................................... 98
Chapter 7 Spare parts ...................................................................................................................101
Appendices and reference
Appendix A Theory of operation ......................................................................................................103
A.1 Infrared source unit .................................................................................................................... 103
A.2 Infrared receiver unit ..................................................................................................................104
A.3 Control unit ................................................................................................................................105
A.4 Power supply unit .......................................................................................................................105
A.5 Air purge .................................................................................................................................... 105
A.6 Isolating valves ...........................................................................................................................105
A.7 Principles and modes of operation ............................................................................................. 106
A.7.1 Calculation of gas concentration ................................................................................. 106
A.7.2 Error compensation ..................................................................................................... 106
A.7.3 Calculation sequence ...................................................................................................107
A.7.4 Normalization equations ............................................................................................. 107
A.7.5 Principles of cross-duct gas analyzers .......................................................................... 109
Appendix B Return equipment to the factory .................................................................................. 113
Reference Manual iii
Page 8
Contents
iv Rosemount CCO 5500
Page 9

Description and specifications

1 Description and specifications

1.1 Component checklist

A typical Rosemount™ CCO 5500 Carbon Monoxide Analyzer should contain the items shown in Figure 1-1. Record the part number, serial number, and order number for your system.
Reference Manual 1
Page 10
Description and specifications
Typical System PackageFigure 1-1:
A. Control unit B. Quick Start Guide C. Power supply D. Hardware E. Receiver F. Interconnect cable G. 33 ft, (10 m) cables H. Source I. Gaskets (4)
Also, use the product matrix in Section 1.5 to compare your order number against your unit. Ensure the features and options specified by your order number are on or included with the unit.
2 Rosemount CCO 5500
Page 11

1.2 Overview

Rapid advances in design of across the duct infrared gas analyzers have led to the general acceptance of this technique for monitoring gas levels in flue gases of power generation boilers and large industrial process steam boilers.
The Rosemount™ CCO 5500 Carbon Monoxide (CO) analyzer is designed to operate on duct widths of less than 26 ft. (8 m) at flue gas temperatures up to 572 °F (300 °C).
NOTICE
The instrument can achieve temperatures up to 1200 °F (650 °C), but degradation in instrument accuracy will occur.
The rugged construction makes installation extremely simple, and through the use use of microprocessor technology, the Rosemount CCO 5500 has many advanced features:
Serial data facility to allow communication between analyzers and a central data
logging station.
User-definable output in either mg/m3, mg/Nm3, or ppm.
Four rolling averages are held, selectable from 10 seconds to 30 days.
Integral, back-lit 32 character LCD provides diagnostic and measurement
information.
Plant status input to prevent emissions dilution during off periods.
Description and specifications
1.3

System description

The Rosemount
An infrared source unit to project a beam of infrared radiation across the duct.
A receiver to measure radiation.
A power supply unit to provide the necessary power rails.
A control unit to compute the gas concentration from the signals provided by the
receiver unit.
Each of these units is designed to be rugged and flexible. They are all fully sealed to IP65 standards and are suitable for outside mounting without the need for further weatherproof enclosures.
CCO 5500 Analyzer consists of four items (Figure 1-2):
Reference Manual 3
Page 12
Description and specifications
Typical System LayoutFigure 1-2:
A. Source B. Air purge C. Site mounting flange D. Receiver E. Pressure regulator F. Purge air G. Isolation valve (by customer if used) H. Cable 33 ft. (10 m) standard (by Rosemount) I. Control unit J. Analog outputs, normalizing inputs, and serial data port K. AC power in and relay contact outputs L. Power supply unit
NOTICE
The maximum cable length allowed between the power supply and the receiver is 82 ft (25 m). The maximum cable length allowed between the power supply and the transmitter is 33 ft (10 m).
4 Rosemount CCO 5500
Page 13

1.4 Specifications

System Measurement SpecificationsTable 1-1:
Span Selectable 0-100 ppm to 0-10,000 ppm within
Accuracy ±2% of measurement or ±5 ppm, whichever is
Path length 1.6 to 26.2 ft. (0.5 to 8 m)
Process temperature range 32 to 1202 °F (0 to 650 °C)
Display units ppm
Averaging Four averages selectable from 10 sec to 30 days
Environmental SpecificationsTable 1-2:
Description and specifications
the range of 200 to 6,000 ppm.m
greater
mg/m3 (measured)
mg/Nm3 (normalized)
Material of construction Cast aluminum, fully sealed to IP65
Ambient temperature limits -4 to 158 °F (-20 to 70 °C)
Installation SpecificationsTable 1-3:
Probe mounting 6.5 in. (165 mm) flange with 4.9 in. (125 mm)
bolt circle
Inputs Temperature: 4 - 20 mA
Pressure: 4 - 20 mA
Plant status dry contact relay
Outputs Analog 4 - 20 mA isolated, 500 Ω max.
Power requirements 85-132/170-264 Vac, 50/60 Hz, 50 VA
Air purge consumption l liter/sec at 1 bar (compressed air)
5 liter/sec (blower air)
Reference Manual 5
Page 14
Description and specifications
1.5 Rosemount™ CCO 5500 ordering information
Model Description
CCO 5500 Carbon Monoxide Analyzer - across-the-stack
Level 1 power requirements
01 110/220 Vac, 50/60 Hz
Level 2 control module display/keypad
01 English
Level 3 calibration options
00 None
01 Calibration check cell and holder
6 Rosemount CCO 5500
Page 15
2 Install
WARNING!
Before installing this equipment, read Essential Instructions. Failure to follow safety instructions could result in serious injury or death.
WARNING!
ELECTRICAL HAZARD Install all protective equipment covers and safety ground leads after installation. Failure to install covers and ground leads could result in serious injury or death.
WARNING!
ELECTRIC SHOCK Before making any electrical connections, make sure the AC power supply is first switched off. Failure to do so could cause personal injury or even death. Make sure that the voltage and frequency of the AC supply match the designations on the analyzer component tags.

Install

2.1 Unpack the equipment

CCO 5500 Carbon Monoxide (CO) Analyzer should contain the
2.2
A typical Rosemount following items.
Refer to Figure 1-1 for an illustration of each of these components. Record the part number, serial number, and order number for each major component of your system.
1. Source with 33 ft. (10 m) of cable and air purge
2. Receiver with 33 ft. (10 m) of cable and air purge
3. Interconnect cable 3 ft. (0.91 m)
4. Control unit
5. Power supply
6. Gaskets (four)
7. Selected screws and washers

Safety considerations

Power is supplied to the whole system via the power supply unit. During installation, do not connect the system to the facility power source until all units are in place and fully wired. If used, keep the isolating valves closed. You must turn off the compressed air
Reference Manual 7
Page 16
Install
supplied to the purges until the full installation is complete. If you do any servicing or rewiring, ensure that the power supply is isolated. During configuration, the system requires electrical power, compressed air, and open isolating valves.

2.3 Cable requirements

1. Power supply to signal processor - seven-core, overall screen, multi-stranded, 6/0.2 mm. 0.5 mm2.
NOTICE
Although shielded cable is specified for the interconnecting cable, it is not necessary to ground the cable.
2. Current loop output - any suitable two-conductor cable, maximum length depends on keeping output load within the 500 ohm maximum load requirement.
3. Contact outputs - any two-conductor cable capable of supplying the power to the warning device/relay, etc. 250 V, 10 A maximum.
4. A.C. power - any suitable three-conductor power cable capable of transmitting 50 VA.
5. Serial data link (if required) - twin twisted pair shielded cable.
6. Analog inputs - any suitable two-conductor cable; Rosemount™ instruments have an internal impedance of 240 ohms for these inputs.
2.4

Selecting location

Rosemount the weather. The instrument is fully sealed and requires no further enclosures or protection. The specific location of the instrument depends on the application and user requirements. Consider the following when choosing a site.
Refer to Figure 1-2 for a typical system arrangement.
1. The site must be accessible at both sides of the duct for servicing the source and
2. The site should be as free from extremes of temperature and vibration as possible.
3. Flue gas temperatures should not exceed 572 °F (300 °C) at the point of
4. There must be an uninterrupted sight path available between the source and
designed the equipment for mounting on boiler ducting or stacks open to
receiver.
Permissible ambient temperature range is -4 °F to 158 °F (-20 °C to 70 °C).
measurment.
NOTICE
The instrument can achieve temperatures up to 1200 °F (600 °C), but degradation in instrument accuracy will occur.
receiver.
8 Rosemount CCO 5500
Page 17
5. The maximum cable length allowed between the power supply and the source is 33 ft. (10 m).
6. The maximum cable length between the power supply and the receiver is 82 ft. (25 m).

2.4.1 Points to consider

Path length
1. Too long [> 26 ft (8 m)]: low energy available.
2. Too short [< 1.6 ft (0.5 m): optical problems
Flue gas temperature
1. Too low (< dewpoint): potential water droplets.
2. Too high [> 572 °F (> 300 °C)]: reduced sensitivity.
Ambient temperature
1. Too low [< -4 °F (< -20 °C)]: condensation on lenses.
2. Too high [> 158 °F (> 70 °C)]: potential instrument problems.
Install
2.5
Measurement range
1. Minimum range depends on acceptable measurement uncertainty which is 10 ppm­meters. For example, for the level of uncertainty to be below 2% of range, the minimum range would be 500 ppm-meters.
NOTICE
10 ppm CO = 12.5 mg/m
2. For increased sensitivity (reduced uncertainty of measurement), the path length must be maximized.
3. Maximum range is 6,000 ppm-meters.
3
NOTICE
To correct ppm-meters to effective ppm, divide the path length (in meters).

Mechanical installation

The transmitter and receiver units are mounted on a site mounting flange on opposite sides of the duct. To protect operators, Rosemount™ recommends using an isolating valve for ducts that operate at a higher than atmospheric pressure.
Use a stand-off pipe [nominal bore 3 in. (75 mm) - not supplied] between the duct and the site mounting flange. The pipe should be long enough to clear the equipment from any duct lagging; it also helps to insulate the equipment from any high duct temperatures.
Reference Manual 9
Page 18
Install

2.5.1 Mount flange assemblies

Complete the following steps to mount the flange assemblies for the analyzer.
1. Form two mounting holes on opposite sides of the stack according to the considerations in Section 2.4.
These holes should accept a slip fit with the stand-off pipe.
2. Weld the stand-off pipes to the site mounting flanges as shown in Figure 2-1.
Site Mounting Flange AssemblyFigure 2-1:
A. Duct wall B. Lagging C. Stand-off pipe (if used) 2.95 in. (75 mm) dia. nominal D. ØM8 hole 4.92 in. (125 mm) BC (4) places E. Site mounting flange F. Bracing fillets
3. With the stand-off pipes and site mounting flanges welded together, insert the mounting flange assemblies into their mounting holes.
4. Position the mounting flange assemblies so the four threaded mounting holes are located as shown in Figure 2-1.
10 Rosemount CCO 5500
Page 19
NOTICE
Rosemount™ suggests that you tack weld the stand-off pipe to the duct and check the alignment visually before making a complete weld.
5. Look through one of the mounting flange assemblies.
If the you can see the orifice across the stack clearly, the alignment is satisfactory. The alignment of these holes is not critical; an integral adjustable mount can compensate for up to 4 degrees of misalignment.
6. Weld the assemblies in place.
To avoid vibration and movement, you may need to fit spreader plates or bracing fillets on the mounting flange assembly as shown in Figure 2-1.

2.5.2 Isolating valves

To protect operators, Rosemount™ recommends that you use customer supplied isolating valves (Figure 2-2) for ducts that operate at higher than atmospheric pressure. Valve selection and installation is your responsibility.
Install
Isolating Valve and Air Purge ArrangmentFigure 2-2:
A. Isolating valve (customer supplied) B. Adjusting nuts C. Locking nuts D. Air purge port E. Pressure regulator assembly F. Rear flange G. Front flange H. Site mounting flange
Reference Manual 11
Page 20
Install
After the isolating valves are installed in the site mounting flanges, connect the purge air supply and install air purge units according to the following instructions.

2.5.3 Purge air supply

The purpose of the purge air is to keep the windows of the source and the receiver clean. Always connect the purge air supply to the air purge units before you install the air purge units on the process duct. Purge air may be supplied by one of the following three methods.
Negative pressure duct
If the duct operates at a negative pressure under all firing conditions, you may simply leave the air purge inlets open and allow the negative draft in the duct to draw in ambient air.
You must supply the air purge units for positive pressure ducts with compressed air or blower air to prevent contamination of the source and receiver units.
Compressed air
You may use compressed air to provide the air flow required. An air supply of 14.7 psi (1 bar) is required, and the consumption is 2.2 cfm (1 liter/second) per purge. Use a fine control flow regulator and filter.
Blower air
You may use a blower to provide the air to the air purge. Customers may specify their own blower. The blower should deliver 11 cfm (5 liters/second) per purge against the working pressure of the duct.
12 Rosemount CCO 5500
Page 21
Air Purge MountingFigure 2-3:
A. Site mounting flange or isolating valve flange B. Adusting nuts C. Locking nuts D. Air purge E. Rear flange F. Front flange
Install
2.5.4

Air purge units

Use the general procedure that follows to install the air purge units on the site mounting flanges or on the exposed flanges of the customer supplied isolating valves, if used.
CAUTION!
CONTAMINATION
Always connect and turn on the purge air supply to both air purge units before mounting the air purge units. Failure to flow purge air may allow the optical surfaces of the source and receiver units to become severely contaminated.
Procedure
1. Remove the four locking nuts holding the front flange to the rear flange (Figure 2-3).
2. Carefully work and pull the front flange from the air purge unit.
3. Align the four holes on the front flange with the four holes on the site mounting flange.
4. Fasten the front flange to the site mounting flange with the four countersunk screws and gaskets provided.
5. Connect and turn on a compressed air or blower air supply to the purge unit. Always flow purge air before installing an air purge unit on the duct.
Reference Manual 13
Page 22
Install
6. Install the air purge unit on the front flange as shown.
7. Install and tighten the four locking nuts removed in Step 1.

2.5.5 Source and receiver units

Use the following procedure to install the source and receiver units on the air purge units.
1. Insert a flexible gasket between the air purge unit and the source or receiver unit.
2. Dowel pins (Figure 2-4) ensure that the source and receiver units and the air purge units mount in a fixed rotary position. Align the dowel pin and dowel pin hole.
Air Purge and Source Unit Mounting FeaturesFigure 2-4:
A. Locking nut B. Window plate C. Source D. Alignment dowel pin hole E. Threaded screw hole F. Rear face of air purge
3. Attach the source or receiver to the rear face of the air purge and install the four screws provided (Figure 2-5).
14 Rosemount CCO 5500
Page 23
Air Purge and Source Units (Installed)Figure 2-5:
A. Site mounting flange B. Front flange C. Gasket D. Source or receiver E. Window plate F. Air purge G. Rear flange
Install

2.5.6 Control unit

Rosemount™ supplies adequate cableto locate the control unit up to 33 ft. (10 m) from the receiver. Do not exceed the 33 ft. (10 m) cable length.
1. Loosen the four captive cover screws and remove the cover.
2. Unplug the ribbon cable connector on the cover side.
3. Fasten the control unit to a firm vertical support. Install four mounting screws in the mounting holes provided.
Refer to Figure 2-6 for mounting dimensions.
Reference Manual 15
Page 24
Install
Mounting Dimensions for the Control and Power Supply UnitsFigure 2-6:
2.5.7
A. Cover seal - Note that the mounting holes are outside the extent of the seal. B. Cover C. Base D. 4 Holes for M6 mounting screws E. Cable gland entry blanking page F. Approx. 6 in. (150 mm) free space required below box for cables G. Assembled box 4.3 in. (110 mm) deep
NOTICE
The unit mounting holes are located outside the seal. You do not need to seal the mounting holes after installation or to remove the circuit boards from the unit prior to mounting.

Power supply unit

Rosemount supplies adequate cableto locate the power supply unit up to 33 ft. (10 m) from the source unit. You may use a maximum cable length of 82 ft. (25 m) to connect the power supply unit to the receiver. Do not exceed the 82 ft. (25 m) maximum cable length.
Dimensions and mounting hole locations are identical to the control unit and are shown in
Figure 2-6.
16 Rosemount CCO 5500
Page 25

2.6 Electrical data

2.6.1 AC supplies

You can power the Rosemount™ CCO 5500 Analyzer from either 85-135 Vac or 170-264 Vac at 50/60 Hz. A switch within the power supply unit selects the input voltage, and an internal 2 A fuse protects the instrument. The analyzer tolerates voltage fluctuations within these ranges without losing performance. The total power requirement for the analyzer is less than 50 VA.

2.6.2 Outputs

Three analyzer outputs are available:
1. Selectable, fully isolated 4-20 mA or 0-20 mA % CO concentration, 500 ohms maximum load.
2. Single pole, switching relays (rated 250 V, 10 A) for the following outputs:
Alarm trigger at a selectable gas threshold.
Data-valid indication active during power failure and any equipment fault
condition. See Chapter 6 for further details.
3. Four-wire serial data link for two-way communication between the control unit and a distributed control system or other process control system.
Install
2.6.3
2.6.4

Normalizing inputs

The analyzer can hold pressure, temperature, and oxygen values to normalize the calculated gas value to standard conditions. The instrument may read these values using the following methods:
1. Fixed value from the keypad.
2. 4-20 mA outputs from measurment transducers. You can set the ranges represented by these inputs from within the processor. These are analog process inputs to the control unit.
3. When the analyzer is part of an integrated system, the serial data line can carry the normalizing values.

Plant status input

The plant status input parameter is available to prevent the rolling average data from being diluted by measurements made while the plant is shut down. The parameter is governed by one of three choices:
1. Serial input (from an integrated system)
2. Logic input (terminals PS1 and PS2 in the control unit)
3. Multiple (five variables)
a. Temperature
Reference Manual 17
Page 26
Install
b. Oxygen
c. Pressure
d. Water vapor
e. Logic input
You can set these parameters in Mode 5. Chapter 4 describes each of these parameters.
During normal operation, the plant status registers as ON. However, if the plant status input is lost, the status changes to OFF, and the averaging data (seconds, minutes, hours, days) is not updated.
NOTICE
During normal operation, do not link terminals PS1 and PS2 together.

2.7 Electrical connections

All equipment wiring must conform to local and national codes. Read and observe the following instructions before making electrical connections.
WARNING!
ELECTRIC SHOCK Disconnect and lock out power before connecting the power supply to the analyzer.
WARNING!
ELECTRIC SHOCK Install all protective covers and safety ground leads after installation. Failure to install covers and ground leads could result in serious injury or death.
WARNING!
ELECTRIC SHOCK To meet the safety requirements of IEC 1010 (EC requirement) and ensure safe operation of this equipment, connect to the main electrical power supply through a circuit breaker (min. 10 A) which will disconnect all current-carrying conductors during a fault situation.This circuit breaker should also included a mechanically operated isolating switch. If not, then locate another external means of disconnecting the supply from the equipment close by. Circuit breakers or switches must comply with a recognized standard, such as IEC 947.
NOTICE
To maintain proper earth grounding, ensure a positive connection exists between the transmitter housing and earth. The connecting ground wire must be 14 AWG minimum.
18 Rosemount CCO 5500
Page 27
NOTICE
Line voltage, signal, and relay wiring must be rated for at least 221 °F (105 °C). Make sure that the voltage and frequency of the AC power supply match the required power specifications.
NOTICE
If metal conduit is used with the power supply unit and/or the source unit, bond the conduit reliably to protective earth. Grounding points inside the units are not bonded to PE and do not provide adequate grounding.

2.7.1 Installation of cables

Decide routing for all non-power cables (both those supplied by Rosemount™ and those sourced locally). Use common routing wherever possible and install leaving sufficient free­end length to make final connections.
Install power cables separately using different routes if possible to reduce the risk of cross­interference. Leave sufficient free-end length to make final connections.
Install
Rosemount supplied cables are provided with ferrite beads fitted to all cores to protect against interference. Do not modify the cables without consulting Rosemount.

2.7.2 Cable connections

Use the following procedure to make cable connections between the source, receiver, power supply unit, and control unit.
Figure 2-7 displays a system wiring diagram. Figure 2-8 displays the location of power
supply and control unit connectors, etc.
Reference Manual 19
Page 28
Install
System Wiring DiagramFigure 2-7:
20 Rosemount CCO 5500
Page 29
Install
Wiring Connector Locations (Power Supply Board)Figure 2-8:
A. 110/220 Vac power in B. Data-valid relay contacts C. Alarm relay contacts D. Interconnect cable contacts
Reference Manual 21
Page 30
Install
Wiring Connector Locations (Control Board)Figure 2-9:
A. Plant status in B. Analog out C. Receiver cable in D. Serial data input
Procedure
1. Install the receiver cable in the center rear cable port of the control unit enclosure. Provide adequate free wire length for making connections to the control board terminals 16 through 22. Tighten the cable gland nut.
2. Connect the receiver cable wires to the control board terminals 16 through 22 according to the wiring diagram, Figure 2-7.
Do not connect the receiver cable shield wire at the control unit.
22 Rosemount CCO 5500
Page 31
Install
3. Install the source cable in the right front cable port of the control unit enclosure. Provide adequate free wire length for making connections to the control board terminals 8 through 12. Tighten the cable gland nut.
4. Connect the source cable wires to the control board terminals 8 through 12 according to the wiring diagram, Figure 2-7.
Do not connect the source cable shield wire at the control unit.
5. Install one end of the power supply to control unit interconnect cable in the center front cable port of the control unit enclosure. Provide adequate free wire length for making connections to the control board terminals 1 through 7. Tighten the cable gland nut.
6. Connect the cable wires to the control board terminals 1 through 7 according to the wiring diagram, Figure 2-7.
Do not connect the cable shield wire at the control unit.
7. Install the opposite end of the power supply to control unit interconnect cable in one of the right hand cable ports of the power supply unit enclosure. Provide adequate free wire length for making connections to the power supply board terminals 1 through 7. Tighten the cable gland nut.
8. Connect the cable wires to the power supply board terminals 1 through 7 according to the wiring diagram, Figure 2-7.
Do not connect the cable shield wire at the power supply unit.
9. Install the 110/220 Vac power cable in one of the left hand cable ports of the power supply unit enclosure. Provide adequate free wire length for making connections to the power supply board terminals: L, N, and E. Tighten the cable gland nut.
10. Connect the cable wires to the power supply board terminals L, N, and E according to the wiring diagram, Figure 2-7.
Do not connect the power cable to the facility power source at this time.
11. Verify that the power switch is in the correct position.
The voltage position selected must match the voltage supplied to the Rosemount CCO 5500 Analyzer at your facility.
12. Connect two separate earth ground leads to the ground screws located on the left hand side of the power supply unit and control unit enclosures.
NOTICE
To maintain proper earth grounding, ensure a positive connection exists between the enclosures and the earth. The connecting ground wires must be 14 AWG minimum.
Reference Manual 23
Page 32
Install
24 Rosemount CCO 5500
Page 33
3 Configuration and startup

3.1 Introduction

You may need two hours or more to configure the instrument, and you need to complete the following tasks:
Power up
Alignment*
Gain adjustment*
Set operating parameters
Calibration*
Note
*Perform these operations when a clean stack condition exists.

Configuration and startup

3.2 Safety considerations

The power supply unit supplies power to the analyzer system. Before removing any equipment covers, lock out and tag out power to the supply unit.
WARNING!
ELECTRIC SHOCK Disconnect and lock out power before connecting power to the analyzer.
3.3
Power up the Rosemount™ CCO 5500 Analyzer
Use the following procedure to power up the analyzer.
1. Make sure that the voltage and frequency of the AC power supply match the required power specifications.
2. With the AC power supply locked out and tagged off, unscrew and remove the power supply unit cover.
3. Select the correct power supply voltage using the AC power selector switch shown in Figure 3-1.
Reference Manual 25
Page 34
Configuration and startup
Power SupplyFigure 3-1:
A. AC power selector switch B. Rail indication LED
3.4
4. Power up the Rosemount CCO 5500 Analyzer and verify that the power supply rail indication LED (Figure 3-1) lights up.
5. Install and fasten the power supply unit cover.
6. Check that the LCD display is functioning at the control unit.
While the source unit is warming up, the LCD display shows WAITING FOR REFERENCE. When the source unit reaches an adequate temperature for the reference to be detected,
the LCD display shows the message STABILIZING REF, along with the frequency and mark/space ratio. See Section 4.9 for further details.
The reference frequency takes some time to stabilize (about five minutes from cold startup). When the reference frequency is within tolerance for 10 consecutive measurement cycles, the instrument automatically changes to the OPERATING mode. This is Mode 1 and is indicated by a number 1 appearing in the top left corner of the LCD
display. The display shows a reading in ppm; this is not an accurate reading until all configuration and startup procedures are completed.
Before conducting the alignment procedure, allow 30 minutes for the source temperature to become stable.

Alignment

For the instrument to operate properly, the source and receiver units must be aligned. Rosemount™ has built in a degree of optical redundancy; normal duct movements do not affect the operation of the instrument. Read and understand this entire procedure before starting the alignment.
26 Rosemount CCO 5500
Page 35
Configuration and startup
1. See Figure 2-5. Unscrew the four screws that secure the receiver to the air purge. Remove and place the receiver in a safe location.
2. Go to the source unit location. To align the source, turn the adjusting nuts (Figure 3-2). Use opposing adjusting nuts to align the source unit in one plane, then the other.
Note
You can achieve receiver alignment by monitoring the output of the detector directly. Use a voltmeter set to AC volts (10 V max.) to measure across test points S0V and S2 for D3, and S0V and S1 for D1 on the receiver control board (Figure 3-3). This alignment method is useful when the receiver is not located near the control unit.
Alignment FeaturesFigure 3-2:
A. Adjusting nuts B. Locking nuts C. Air purge D. Rear flange E. Front flange
Reference Manual 27
Page 36
Configuration and startup
Receiver Test PointsFigure 3-3:
A. S1 white B. S2 blue C. S0V green D. Voltage indication LEDs
3. Adjust the alignment until the bright red disc of the source is located centrally in the field of view when viewed from the receiver air purge. When the source unit is aligned, tighten the locking nuts.
4. Install and tighten the receiver on its air purge using the four screws removed in
Step 1.
5. At the control unit keypad, press MODE four times to select SET UP mode (Mode 5). When 5 (SET UP) is displayed on the LCD, press ENTER to access the SET UP
mode.
Note
The analyzer uses a security code to prevent unauthorized alteration of settings. The default code set at the factory is 0000.
The keypad cursor flashes over the first digit of the security code.
28 Rosemount CCO 5500
Page 37
Configuration and startup
6. Use the arrow keys to enter the desired value for this digit. Press ENTER to select the displayed value; the cursor moves to the second digit. Select the value for the second digit and press ENTER. Continue this process for each digit of the security code.
When the fourth digit is correctly entered, the processor enters Mode 5.
7. When in Mode 5, select Calibrate using the keypad arrow keys; press ENTER to access the Calibrate menu.
8. Use the arrow keys to select SET DETECTORS; press ENTER.
The display shows the D1 and D2 detector levels.
CAUTION!
If the analyzer is not in SET UP mode, the gas cell at the source unit periodically interrupts the IR beam and make alignment difficult.
9. Adjust the receiver alignment, using the adjusting nuts, Figure 3-2. Adjust in one plane, then in the other.
As a rule, the D2 detector level is affected to a greater extent by adjustment in one particular plane. The D1 detector level is affected more by adjustment in the other plane.
10. Make sure the maximum possible values of both D1 and D2 are reached. After you achieve alignment, tighten the locking nuts.
Note
The alignment of the receiver unit is important. Make sure to obtain the maximum values of D1 and D2.
11. If the displayed detector level is below 5000, increase the gain at the control unit to between 12000 and 14000.. If the detector level is above 15000, reduce the gain to between 12000 and 14000.
Refer to Section 3.5 for details.
12. To fine tune the alignment, repeat adjustments of Step 9 at the source unit. Again, make sure the values of D1 and D2 are appropriate. Lock the source unit in place when the maximum values are achieved.
13. When you have properly completed the alignment, there is rarely any need for further alignment adjustments.
3.5
Reference Manual 29

Detector levels

The gain of the detector signals is set in two locations:
1. In the receiver, two potentiometers set the gain. Refer to Section 3.5.1.
Page 38
Configuration and startup
2. In the control unit trim, potentiometers adjust the level of the D1 and D2 signals before they enter the microprocessor. Refer to Section 3.5.1.
It is essential that you properly conduct the alignment procedure and obtain a maximum detector signal before attempting to optimize the detector levels.

3.5.1 Receiver gain adjustment

To give an optimum signal-to-noise ratio, the detector levels must be maximized. For the best signal-to-noise ratio, you must set the gain of the detector signals in the receiver to a maximum without saturating. Rosemount™ sets the gains at a path length of 6.5 f.t (2 m). If the path length is above 13 ft (4 m) or below 5 ft (1.5 m), you may need to adjust to optimize the detector levels.
1. Enter Mode 5 Calibrate Set Detectors and display the value of D2/D1.
2. Loosen the receiver cable gland so the receiver cable can slip when you remove the window plate.
3. Pull the window plate (Figure 3-4) from the receiver to access the detectors.
Receiver Trim PotsFigure 3-4:
A. Trim pot for end detector D1 B. Trim pot for side detector D2
4. Trim potentiometer(s) set the gain.
An AC voltmeter measures the gain levels.
5. Connect the voltmeter to the test points on the control unit (18/20 for D1 and 19/20 for D2). Increase the gain using the trim pot at the end detector D2 until the voltage is between 1 - 4 Vac.
30 Rosemount CCO 5500
Page 39
6. Repeat Step 5 for the side detector D1 measuring across the test points.
7. When the detector levels are satisfactory, replace the cover.
Note
If the duct is operating and a high opacity may be in the path, reduce the set voltages to 2 V rms maximum. This should prevent saturation should the opacity level drop off.

3.5.2 Control unit gain adjustment

After you have optimized the detector level(s) at the receiver, optimize the levels within the microprocessor. Make the adjustment with two trim potentiometers in the control unit.
1. Set the gain to a minimum by turning the D2 detector trim pot (Figure 3-5) fully clockwise.
D2 is a 20-turn potentiometer.
Configuration and startup
Reference Manual 31
Page 40
Configuration and startup
Gain Adjust PotentiometersFigure 3-5:
A. Trim pot B. Test point C. LED D. D2 trim pot E. D1 trim pot F. Plant G. Analog H. Receiver I. Serial data
2. Enter Mode 5 Set Detectors and display the values of D2 and D1. Turn the trim pot counterclockwise until the D2 level is between 12,000 and 15,000. Allow time between adjustments for the readings to settle.
Note
If the duct is operating and the opacity levels are high, reduce the D2 level to about 8,500. This should prevent saturation should the opacity level drop off.
32 Rosemount CCO 5500
Page 41
3. To ensure that the detector signal is not saturating, observe the saturation count signal displayed next to the detector levels. If a SAT # of more than 0 is displayed, turn the trim pot slightly to reduce the gain. Reduce the gain until a SAT # of 0 is displayed.
4. If saturation is indicated with the trim pot turned fully clockwise, reduce the gain in the receiver and repeat the procedure.
5. Repeat Step 1 through Step 3 for the D1 level using the D1 trim pot.
Note
Rosemount™ has designed the circuits so that wherever saturation occurs (receiver or control unit), the microprocessor always detects it. If the displayed detector levels cannot be set to within this band or saturation cannot be avoided, optimize the detector levels ar the receiver. Refer to Section 3.5.1.

3.6 Source adjustments

Two trim potentiometers within the source unit allow adjustments to be made to the intensity of the source and to the frequency of the chopper motor. Rosemount™ sets these at the factory, and they rarely need adjustment. Rosemount recommends that you consult the company before making any adjustments within the source unit.
Configuration and startup
3.6.1
CAUTION!
DECREASED SOURCE LIFE
Increasing source intensity may severely reduce the source life.

Source intensity

Complete the following steps to adjust the source intensity in the analyzer.
1. Loosen and slide the source unit cable gland so the cable can slip when the rear cover plate is removed.
2. Unscrew and remove the rear cover plate.
3. A trim pot (Figure 3-6) allows adjustment to the intensity of the source. Turn the trim pot clockwise to increase the source intensity.
Reference Manual 33
Page 42
Configuration and startup
Source Intensity Trim PotFigure 3-6:
A. Trim pot

3.6.2 Chopper frequency

Complete the following steps to adjust the analyzer's chopper frequency.
1. Unscrew and remove the source from the air purge.
2. Unscrew and pull the window plate from the front of the source unit.
See Figure 3-7. A trim pot (VR1) allows you to adjust the frequency of the chopper motor.
34 Rosemount CCO 5500
Page 43
Configuration and startup
Chopper Frequency Trim PotFigure 3-7:
3.7
A. Trim pot VR1
3. To increase the chopper frequency, turn the trim pot counterclockwise.

Set up mode

To prevent any unauthorized changes, you must enter a four digit security code to enter the set up mode. You must set operating parameters in the instrument for proper analyzer operation. All operating parametrs are set within the control unit using the SET UP mode. In the SET UP mode, parameters are held in non-volatile memory and retained in the event of a power loss.
Even if the measured data is not going to be normalized, you must set the normalizing parameters to ensure proper analyzer operation.
Note
When you select SET UP mode, the instrument suspends motor operations, and the Data Valid LED goes dim. If no key is pressed within five seconds after selecting SET UP mode, the instrument control reverts to OPERATING mode.
Chapter 4 lists all parameters in full. Basic details are given here for configuration purposes.
To aid configuration and to record any subsequent changes to the operating parameters, Rosemount recommends completing to provide a record of the instrument setup.
Reference Manual 35
Page 44
Configuration and startup
Procedure
1. Press MODE unil the number 5 is displayed in the top left corner.
After you correctly enter the security code, there are six submodes of operation from which the set up parameters may be changed. These six submodes are:
a. Set Averages: The four averaging stack times (seconds, minutes, hours, and
days) may be set as required.
b. Configure O/P - Analog output setup: Origin, units, span, rolling average, and
fault condition.
c. Parameters: The following are set from this mode: security code, identity
number, path length, alarm level, cal factor, and plant status.
d. Normalization: You may set up all normalization parameters from this mode.
e. Reset Average: Select this submode to reset the four averaging stacks.
f. Calibrate: Set the outputs of the detectors and the basic calibration for the
instrument.
2. Use the arrow keys to toggle between these six options and press ENTER when the desired option is displayed.

3.7.1 Enter security code

Complete the following steps to enter the security code.
1. Once the display is as shown here, press ENTER to gain access to SET UP mode.
The cursor flashes over the first digit of the security code number.
2. Select the required first digit with the arrow keys and press ENTER.
3. Repeat this procedure for the remaining three numbers.
If the code is correct after you press ENTER on the last digit, the sequence is continued. If it is not correct, the instrument returns to OPERATING mode. Refer to
Section 3.7.4 for further details.
Note
Rosemount™ sets the code number to 0000 at the factory; you should change it in SET UP mode.
3.7.2

Set averages

The instrument calculates four separate averages. These are defined in units of seconds, minutes, hours, and days. You can use any of the four averaging stacks to provide the instrument's analog output. You can set each averaging time within predefined limits.
36 Rosemount CCO 5500
Page 45
Configuration and startup
1. Press ENTER when this display is shown.
The display now shows one of the averages.
2. Use the arrow keys to select the average time that you need to change and press ENTER to change it.
You can now change the value using the arrow keys and confirm it by pressing ENTER.
3. Set the seconds averaging stack to the required value.
This is limited to within 10 to 60 seconds in 10-second intervals.
4. Set the minutes averaging stack to the required value.
3.7.3
This is limited to within 1 to 60 minutes in 1-minute intervals.
5. Set the hours averaging stack to the required value.
This is limited to within 1 to 24 hours in 1-hour intervals.
6. Set the days averaging stack to the required value.
This is limited to within 1 to 30 days in 1-day intervals.

Configure O/P

You can set up the analog loop output from this mode.
1. Press ENTER while this display is shown to access it.
Reference Manual 37
Page 46
Configuration and startup
2. Press the arrow keys to step through the available options.
3. Press ENTER to access and change the displayed parameters of each of the six available options listed below:
a. Output
b. Averages
c. Units
d. Output Span
e. Fault Condition
f. Set mA Output
Output
You can set an origin of 0 or 4 mA for the current loop output. The arrow keys toggle between these two options. Press ENTER to enter the new value.
Average
You may use any of the four averaging stacks (seconds, minutes, hours, and days) for the analog output. You can select them with the arrow keys and enter them with the ENTER key.
Units
The analog output can represent the gas concentration in units of mg/m3, mg/Nm3, or vpm. The arrow keys toggle between these three options. Press ENTER to enter the new value.
Output Span
Select the required span using the arrow keys for each digit. Press ENTER to enter the value of each digit. The units are displayed in vpm, mg/m3, or mg/Nm3, depending on what has been selected beforehand.
38 Rosemount CCO 5500
Page 47
Configuration and startup
Note
Once you select Output Span, the current value is displayed for one second. The first digit of the display then defaults to zero; thus you must re-enter the span value for the unit to function correctly.
Fault condition
If a fault condition occurs, you may set the current output of the instrument to one of the following options:
Procedure
1. Set the output at 0 mA - ZERO.
2. Adjust the output to the calculated gas concentration even though a fault condition exists - MEAS.
3. Hold the last selected gas concentration - HOLD.
4. Set the output to full scale (20 mA) - F.S.
Press the arrow keys to select one of these options; when the desired output is displayed, press ENTER to confirm.
Set mA output
Note
This is set at the factory and should not be altered without due consideration.
From this option, the current levels of the analog output are set up. Press ENTER to select it, and the instrument prompts you to set the current levels at 0 and 20 mA.
When this is displayed, set the current output to 0 mA as measured with a calibrated current meter across the analog ouput loop terminals; do not connect anything else to these terminals when setting up the output.
Use the two arrow keys to adjust the value; the UP arrow takes the current output up, and the DOWN arrow takes it down. Press ENTER when the correct output is displayed on the ammeter.
Reference Manual 39
Page 48
Configuration and startup
Note
Zero mA should be set up no matter what has been selected as the base of the current output. This is factory set.
In a similar manner to the above, set the current output level to 20 mA.

3.7.4 Parameters

Complete the following steps to set the parameters.
1. With this option displayed, press ENTER to access the list of six available options.
The arrow keys cycle through these options.
2. When the option you need to change is displayed, press ENTER.
3. When you have made all required changes, select EXIT and press ENTER.
The six available options are:
a. Security Number
b. Identity Number
c. Path Length
d. Alarm
e. Cal. Factor
f. Plant Status Input
Security Number
To prevent any unauthorized tampering with the setup information, it is important to change the security code from the factory setting.
Select each digit with ENTER and change it with the arrow keys.
Note
It is important to make note of this number; otherwise, it will not be possible to change the instrument set up.
40 Rosemount CCO 5500
Page 49
Configuration and startup
Identity Number
If you are using the system as part of an integral monitoring system and are using the serial input and outputs, the central processor requires a Device Identity to identify each instrument. This number must be unique for each equipment item and can be set from 1 to 30 as required.
Path Length
Note
Once you select path length, the instrument displays the current value for one second. The first value of the display then defaults to zero; therefore, you must re-enter the value to calculate the gas concentrations correctly.
The transmissivity of any gas depends both on the concentration and on the path length through which the radiation is transmitted. Similarly, the output of the Rosemount™ CCO 5500 Analyzer gas monitor also depends on the path length of the flue gas through which the radiation is transmitted.
Refer to Section 2.4.1.
The Rosemount CCO 5500 Analyzer is sensitive to the product of concentration and path length. In order to obtain a true value of concentration of gas, you must input the correct path length into the processor. The processor then uses the value to produce a final value of gas concentration.
Note
The path length entered must represent the length of the actual gas pass, not the flange to flange dimension between the source and receiver.
Alarm
A contact output is available to warn of a high gas concentration. You may trigger the contact output from any of the four averaging stacks. Select the source with the arrow key and enter it with ENTER.
Reference Manual 41
Page 50
Configuration and startup
Select the units for the alarm; these may be different than the units selected for the analog output.
After you select the source, the instrument requires a level that will trigger the output. Set the desired level with the arrow keys.
Cal Factor
Note
First, record the original Cal Factor before entering this mode as displayed in Mode 4 Calibration Data. The Cal Factor is lost when the menu option is entered.
During the calibration routine, the instrument calculates a Cal Factor which sets the basic calibration of the instrument. You may change this value from this mode.
Note
Since this value controls the calibration of the instrument, only change if necessary.
Plant Status Input
Use this to determine whether the plant is operating under correct conditions.
There is a choice of three controls for plant status: Logic Input, Serial Input, and Multiple. You can only use one to control plant status at any one time.
Logic Input
If the PS1 and PS2 terminals are linked in the control unit, the logic contact is made, and the plant status is OFF. You may link these terminals manually during a plant shut down, or you may wire them to a switch/contact outside the unit (e.g., a value that opens and closes the duct). Press ENTER to select this option when the Plant Status Serial Input option is displayed.
Serial input
42 Rosemount CCO 5500
Page 51
Configuration and startup
If you select this option, the criteria controlling plant status are transmitted via the serial data link. Press ENTER when the Plant Status I/P Serial Input option is displayed to select this option.
Multiple
Four options are available here. Press ENTER when the Plant Status I/P Multiple option is displayed, and the first option Temperature is displayed. Use the arrow keys to toggle YES or NO. NO means that the temperature threshold is not used to determine plant status. If you select YES, the display enters the display below. Configure the instrument for temperature threshold. Press ENTER when it is correctly configured, and the display moves to the next option Oxygen. After you set the last option, the Logic Input display returns to the PARAMETERS Plant Status I/P option. Use the DOWN arrow to scroll down to EXIT and press ENTER. The plant status is now full configured.
Plant status is only OFF if all options selected are registering plant status OFF. If any one of them is not fulfilling plant status OFF conditions, then the instrument registers plant status ON.
- Temperature
A value is set here for the temperature threshold. While the temperature (taken from the normalizing temperature) is above the threshold value, plant status is ON. If the temperature drops below the threshold, plant status is OFF, and only the seconds averaging stack updates.
- Oxygen
Oxygen is set and used in a similar manner to the temperature threshold. However, if the normalizing oxygen level rises above the threshold, plant status is OFF. For plant status ON, the oxygen level must be below the threshold.
- Water Vapor
This is set and used in a similar manner to the temperature threshold. If the normalizing water vapor level falls below the threshold, plant status is OFF. For plant status ON, the water vapor level must be above the threshold.
- Logic Input
Select YES or NO and press ENTER. For plant status to be ON, the logic input (PS1 and PS2) must be open circuit; for plant status to be OFF, the logic input must be closed circuit. After you configure this option, the menu exits to the Multiple option. Use the DOWN arrow to select EXIT and press ENTER.
3.7.5
Reference Manual 43

Normalization

Complete the following steps to set the Normalization parameters.
1. Press ENTER to access one of the four normalizing parameters listed below.
Page 52
Configuration and startup
2. Use the arrow keys to cycle through the four options and press ENTER once the desired option is selected.
a. Temperature
b. Oxygen
c. Pressure
d. Water Vapor
Once selected, each of the above normalizing parameters have the same suboptions: Set Standard Levels and Set Values. Under Set Values, are three more suboptions that determine how the instrument reads the data: Analog Input, Serial Input, and Keypad Input.
Set Standard Levels
Each normalizing parameter normalizes the measured gas concentration to standard conditions of temperature, oxygen, pressure, and water vapor. Set these levels within this option. Use the arrow keys to change each displayed normalizing standard value.
Temperature
Always use an analog input for temperature correction; this ensures that the flue gas temperature is being measured continuously and accurately. Connect the analog output of the temperatue transducer into the Rosemount™ analyzer and select the analog input option. This value is used to normalize the gas concentration measurement and to correct for the effects of temperature on the IR absorption spectrum.
If you use the Keypad Input, and the gas temperature is higher than 572 ° F (300 °C), the compensation algorithm becomes less precise, and instrument accuracy deteriorates accordingly. Rosemount does not recommend this.
Note
If normalization is not required, the instrument must hold the temperature of the gas in the duct using the Analog Input option.
Press ENTER with the Temperature option selected to access Set Standard Levels and Set Values. Use the arrow keys to toggle between these options and Exit.
Set Values
You can bring the normalizing data into the instrument in one of three ways:
Analog Input
44 Rosemount CCO 5500
Page 53
Configuration and startup
This uses the 4-20 mA inputs within the processor to receive the measured transducer data. The values at 4 mA and 20 mA will be requested should this option be selected.
Serial Input
If you use an input unit, all normalizing data can be transmitted via the serial data line.
You can enter a fixed value via the keypad. This is suitable where the value is stable
to about ±5%.
With an integrated system, set the lead analyzer's normalizing parameters to the 4-20 mA inputs. Then set all the other analyzers to serial, and the normalization parameters are transmitted down the serial data highway.
Oxygen
To correct the data to standard levels of oxygen, you must enter an estimate of the oxygen at the point of measurement. If the oxygen level is being continuously measured, connect the analog output of the oxygen analyzer into the Rosemount™ CCO 5500 Analyzer and select Analog Input. You must define this input as either WET or DRY depending on how the measurement is made. After you define the wet or dry, you need to define the Analog Input values; set the 4 mA and 20 mA values. If the oxgyen level is relatively constant through all firing conditions, you may use a fixed keypad input.
With an integrated system, you can take the oxygen data to the instrument via the serial data line.
Note
If normalization is not required, you must set the normalizing parameters for oxygen in the instrument.
Pressure
To correct the data to a standard pressure, normally 14.7 psi (101 kPa), you must determine the pressure at the point of measurement. If the flue pressure is relatively constant through all firing conditions, then you may use a fixed keypad input. If the pressure, is not constant, measure it and bring it into the instrument via the 4-20 mA analog input within the processor.
Reference Manual 45
Page 54
Configuration and startup
Note
If normalization is not required, set the normalizing parameters for pressure in the instrument to
14.7 psi (101 kPa)(Standard Level and Keypad Input).
With an integrated system, you can take the pressure data to the instrument via the serial data line.
Water Vapor
An across the duct monitor measures the gas concentration under wet conditions. Unlike a sampling system, the gas has not been preconditioned in any way before you make a measurement.
When the water vapor is at a relatively fixed level, set the standard level to DRY to normalize it to dry conditions. Use a fixed value in the keypad option representing the expected water vapor produced for the fuel type. If the measurement is not to be normalized for water vapor, set the standard level to WET.
With an integrated system, you can take the water vapor data to the instrument via the serial data line.

3.7.6 Reset averages

Note
Resetting averages causes the rolling average data to be cleared from memory.
You can reset the average values that are currently held in the four averaging stacks using this option; this ereases the current average that is held in all of the averaging stacks. Select this option by pressing ENTER and using the arrow keys. The instrument requests confirmation before the averages are reset.
Note
If you select this option, all the data in the averaging stacks is reset, and the data for as much as 30 days is lost.
3.7.7

Calibrate

From this option, you may display the two detector levels and conduct a basic calibration. While in this mode, the gas cell is not moved; this gives an immediate response for setting up the detector levels. A Cal Factor that is calculated during a calibration routine sets the basic calibration of the instrument. Press ENTER while this is displayed to see the following options:
1. Set Detectors
46 Rosemount CCO 5500
Page 55
Configuration and startup
2. Span Adjust
3. Calibrate
Set detectors
You can display both the D1 and D2 levels; you can also display saturation counts. To give an immediate response to any alterations that are required, the filters and gas cells are not moved during this operation.
Refer to Section 3.5 for a discussion of the detector level and saturation count.
Span adjust
Note
Rosemount™ initially sets the span factor at the factory; do not adjust it unless the instrument sensitivity is suspected. In any case, Rosemount recommends that you record the orginal value before making adjustments.
You can adjust instrument sensitivity if a known concentration of gas exists between the source and receiver units and instrument sensitivity is supsected. If a problem arises, consult Rosemount.
You may need to adjust the span factor if you have fitted new gas cells or filters.
Calibrate
Re-enter the Mode 5 CALIBRATE menu and proceed to the Calibrate option.
You can calculate the basic calibration of the instrument from this routine. It is preferable to conduct this operation with the plant shut down to ensure a zero gas concentration within the duct. If this is not possible, the instrument can calibrate to a known value of the gas concentration - the calibration target.
Reference Manual 47
Page 56
Configuration and startup
Set the calibration target either to a known value of the gas concentration or to zero. Then set the desired number of cycles over which the calibration factor is determined (Rosemount™ recommends a minimum of 30). Now run the calibration, and the display will show a countdown during its execution. When the calibration is complete, the new Cal Factor is displayed for about five seconds, and the instrument exits the calibration routine.
Note
You must run the calibration routine during commissioning; otherwise, the instrument will not be able to calculate the true level of gas within the duct.
Do not run the calibration routine unless reasonable conditions exist in the duct. If it is not the initial calibration, record the Cal Factor from the Parameters option before running the calibration.
To aid configuration and to record any subsequent changes to the operating parameters,
Table 3-1 lists all of the options available and can be used as a record of the operating
parameters.
Instrument SettingsTable 3-1:
Parameter Config
Averages
Seconds
Minutes
Hours
Days
Output
0 or 4 mA base
Units
Average
48 Rosemount CCO 5500
Page 57
Instrument Settings (continued)Table 3-1:
Parameter Config
Fault condition
Parameters
Path length
Alarm source
Alarm units
Alarm levels
Normalization
Temperature
Standard level °F (°C)
I/P °F (°C) @ 4 mA
I/P °F(°C) at 20 mA
Keypad input °C (not ideal)
Serial input
Configuration and startup
Oxygen
Standard level %
Wet or dry gas
I/P % @ 4 mA
I/P % @ 20 mA
Keypad input %
Serial input
Pressure
Standard level psi (kPa)
I/P psi (kPa) @ 4 mA
I/P psi (kPa) @ 20 mA
Keypad input
Serial input
Water vapor
Standard level (wet %/dry
Keypad input %
Measured value
After the instrument calibrates, it calculates a cal factor; this determines the calibration of the instrument. The instrument displays the cal factor for a few seconds after calibrating, and you can also interrogate the cal factor from Mode 4 (DIAGNOSTIC mode). Enter the cal factor into the table below as a record of instrument operation.
Reference Manual 49
Page 58
Configuration and startup
Note
Rosemount™ obtained factory settings under the following conditions:
2 m path length
Clean conditions
Calibration DataTable 3-2:
Detector outputs Factory Config
D2
D1
E1
E2
Calibration data
Cal factor
Span factor
Output calibration
Set zero
Set span

3.8 Current output calibration

Now set up the current output using a calibrated multimeter set to DC current, 20 mA max. Conduct this procedure as follows:
Note
Rosemount™ sets the current output at the factory; do not alter it.
Procedure
1. Connect the multimeter to the output terminals within the control unit terminals +mA and -mA.
2. Enter Mode 5 Configure Output Fault Condition Set Zero and adjust the level using the arrow keys until 0 mA is recorded. Record the value in brackets on the
display in Table 3-2.
3. Enter Mode 5 Configure Output Fault Condition Set Span and adjust the level using the arrow keys until 20 mA is recorded. Record the value in brackets on the
display in Table 3-2.
50 Rosemount CCO 5500
Page 59
4 Operation

4.1 Introduction

After you configure the Rosemount™ CCO 5500, it measures the gas levels between the source and receiver and produces an output proportional to the gas levels. An integral 32­character LCD display also shows the calculated levels.
The Rosemount CCO 5500 Analyzer allows you to interrogate the microprocessor to observe the system parameters and to change them if required.
The Rosemount CCO 5500 uses a menu-based program; you can gain access to it by the key panel mounted on the lid of the control unit (Figure 4-1).
Control Unit KeypadFigure 4-1:

Operation

A. 32-character liquid crystal display B. Keypad C. Data valid and alarm LEDs
4.1.1
Reference Manual 51

Measurement

Once the analyzer has completed configuration, it measures the absorption of IR radiation and calculates a parameter Y.
Page 60
Operation
Refer to Section A.7. The analyzer uses this value to produce a final concentration of gas that can be normalized to standard conditions and averaged over a time ranging from 10 seconds to 30 days.
The instrument computes four averages, any of which can be used to drive the analog output or displayed on the integral 32-character LCD.

4.1.2 Calibration

During the configuration procedure, the analyzer conducts a calibration that sets the system gains to produce a zero or known gas level. Once the analyzer has conducted the routine, precision filters, which do not change, fix the calibration of the instrument .

4.2 Startup and operation

Power up the system and wait for 30 minutes. This allows time for the infrared source to heat up. Once the receiver is detecting a signal, you will see a reading on the control unit display. This should be in normal OPERATING mode, Mode 1 (shown by a number 1 at the
top left corner of the LCD); the display also shows a reading in vpm, mg/m3, or Nmg/m3. If this appears, the system is functioning properly.

4.3 Modes of operation

The instrument has six modes of operation identified by a number in the top left corner of the display.
Mode 1: Operating Mode
Displays average gas concentration.
Mode 2: Parameters
Displays operating parameters.
Mode 3: Normalization
Displays normalization data.
Mode 4: Diagnostics
Investigates instrument operation. The instrument continually checks itself; if a complicaiton exists, the instrument automatically selects this mode and displays the fault on the display.
Mode 5: Set up Mode
Sets operating parameters. You must enter the opening parameters for the instrument to function correctly. You can only access this mode using a security code.
52 Rosemount CCO 5500
Page 61
Mode 6: Check Cell Mode
Used to verify the instrument's operation and calibration.
Note
The outputs of the instrument are unaffected by key operation in all modes except the Set up mode.

4.4 Keypad operation

You can access each mode sequentially by each push of the MODE key.
Figure 4-1 illustrates the display and keys of the control unit. After you select a mode, use
the arrow keys to select the various options within these modes. Use the ENTER key to input the displayed value and step the cursor to the next option where applicable.
1. MODE key: Press the MODE key to either take the instrument to the next mode of operation or back to the OPERATING mode if pressed from within a mode.
2. Arrow keys: Press the arrow keys to do one of two things, depending on the position in the program:
It increases or decreases the displayed value. If you hold down the key, it
scrolls quickly to the desired value.
It steps through the available options within a mode or submode.
3. ENTER key: Press the ENTER key to do one of two things, depending on the position in the program:
It inputs the displayed parameter value.
It selects the displayed mode or option from within a mode or submode.
Operation
Note
Allow time for the instrument to respond to a key instruction; otherwise, it may record a double key entry.
4.5
Reference Manual 53

Menu tree

Figure 4-2 shows the arrangement of the Rosemount™ CCO 5500 Analyzer menu tree.
Page 62
Operating mode
Parameters
Normalization
Diagnostics
Set up mode
vpm, mg/m
3
,
mg/Nm
3
, Measure, Normalized, Sec, Min, Hour, Day
Identification
Parameters
Averages
Output alarm Plant status Exit
Analyzer type EPROM program ID Identity number
Path length Span factor O/P fault
Sec, Min, Hour, Day
On, Off Plant St. I/P
Temperature threshold Oxygen threshold Water vapor threshold
Temperature Oxygen Pressure Water vapor Exit
Detector outputs Mod. Frequency Yval & CO ppm Calibration data Fault condition Exit
Security code
Set averages
Configure O/P
0000 to 9999
Seconds
Minutes
Hours
Days
Exit
10 to 60 s @ 10 s intervals
01 to 60 m @ 01 m intervals
01 to 24 h @ 01 h intervals
01 to 30 d @ 01 d intervals
Output
Average
Units
Set span
Fault condition
Set mA O/P
Exit
(0 or 4) to 20
Sec, Min, Hour, Day
vpm, mg/m
3
, mg/Nm
3
0 to 9999 span
Zero, Mea, Hold, F.S.
Set zero Set span
Operation
Rosemount CCO 5500 Analyzer Menu TreeFigure 4-2:
54 Rosemount CCO 5500
Page 63
Set up mode
Check cell mode
Parameters
Normalization
Reset averages
Calibrate
Security
Identity
Path length
Alarm
Cal. factor
Plant St. I/P
Exit
4 digit code
01 to 30
Gas path length in mm
Source
Units Alarm level (0 to 9999)
0 to 9999
Logic Serial Multiple Exit
Sec, Min, Hour, Day
Vpm, mg/m
3
, mg/Nm
3
Temperature
Oxygen
Water vapor
Det. level
Logic
Yes, No
Yes, No
Yes, No
Yes, No
Yes, No
Temperature
Oxygen
Pressure
Water vapor
Exit
Set std. levels Set values Exit
Analog input Serial input Keypad input Exit
Set std. levels Set values Exit
Set std. levels Set values Exit
Set std. levels Set values Exit
Analog input Serial input Keypad input Exit
Analog input Serial input Keypad input Exit
Serial input Keypad input Exit
Yes, No
Set detectors Span adjust Calibrate Exit
Yes, No
Operation

4.6 Operating mode

From this mode of operation, you may alter the averaging time of the displayed gas concentration to one of the other averaging stacks and observe the measured/normalized gas measurements. When in this mode, the display appears similar to that shown below. If the display is not similar to this, press MODE until number 1 appears at the top left corner
of the display.
Reference Manual 55
Page 64
Operation
To change the data displayed, press ENTER, and a flashing cursor appears at the beginning of the concentration units, i.e., vpm or mg/Nm3. The arrow keys now change the
highlighted parameter. Each push of ENTER selects another of the parameters in the following order:
1. Concentration units: vpm, mg/m3 (or mg/Nm3).
2. Measured or normalized display.
3. Averaging time: seconds, minutes, hours, or days.
Press ENTER when the cursor is flashing on the averaging time, and the cursor disappears from the display. You may press ENTER again if required to bring the cursor back onto the display.

4.7 Parameters

4.7.1
4.7.2
In this mode, you can examine the parameters set within the SET UP mode, but you cannot change them. Press MODE until the number 2 appears in the top left corner of the display;
then press ENTER. Use the arrow keys to scroll through the available options; press ENTER to display one of the selected options below.
1. Identification
2. Parameters
3. Averages
4. Output Alarm
5. Plant Status
Press ENTER again to exit from each option.
Refer to Section 4.10 for further details of the display information and how to change the held parameters.

Identification

You can display the analyzer type, identity number, and EPROM program ID from this option. Use the arrow keys to scroll between these options.

Parameters

You can display the following parameters from this option; select them using the arrow keys.
1. Path Length: The path length currently used to calculate the gas calculation.
56 Rosemount CCO 5500
Page 65
2. Span Factor: You can adjust the sensitivity of the instrument from the SET UP mode CALIBRATE option. Rosemount™ initially set the span factor at the factory using known gas concentrations.
3. O/P (Output) Fault: Should a fault condition occur, you can set the analog input from one of four options.

4.7.3 Averages

Select this option to display the times set for each of the four averaging stacks: seconds, minutes, hours, and days.

4.7.4 Output

You can display the base, span, and averaging of the analog output from this option.

4.7.5 Alarm

A changeover relay contact output is available to indicate a high gas concentration. You can examine the level at which the output is operated and the averaging stack from which the gas value is obtained from this display.
Operation

4.7.6 Plant status

When plant status is OFF, the minutes, hours, and days averaging stacks do not update.
Note
When the plant status is OFF, pollutant levels are zero. It is not normally permitted to use plant OFF zero levels to reduce the recorded mean entitled pollutant levels.
You can use this function to ensure data is only collected when the plant is fully operational. You can use three options to determine plant status ON or OFF: Logic Input,
Serial Input, and Multiple. Multiple has four options: Temperature threshold, Oxygen threshold, Water Vapor threshold, and Logic Input. You can view the plant status and its governing factor from
this display.
4.8

Normalization

Complete the following steps to display and/or edit the normalization parameters for the analyzer.
1. Press MODE until you see the number 3 in the top left corner of the display.
From this mode, you can display the normalization parameters currently used.
2. Press ENTER to enter the routine and use the arrow keys to select which of the normalizing parameters to display.
Reference Manual 57
Page 66
Operation
Listed below are the four normalizing parameters accessible through this mode:
a. Temperature
b. Oxygen
c. Pressure
d. Water Vapor
3. When you see the required normalizing parameter, press ENTER to display the normalization data. Press ENTER again to exit the parameter.
For each of the four normalizing parameters, the display appears similar to that shown below.
Normalization ExampleFigure 4-3:
A. Selected parameter B. Units C. Standard level required (wet or dry for water vapor) D. Parameter source
k = keypad input
a = analog input
s = serial data input
m = measured input
4.9
4.9.1
58 Rosemount CCO 5500

Diagnostic mode

You may examine the detector levels, chopper blade frequency, Y parameter, and the fault condition from this mode. Press MODE until the number 4 appears in the top left corner of
the display and press ENTER to access this mode. Once in Diagnostics, the five suboptions are as follows:
1. Detector outputs
2. Modulation Frequency
3. Yval and CO ppm
4. Calibration Data
5. Fault Conditions

Detector outputs

This mode displays detector levels from the detector. D1 is the reference level and should always be less than D2. The level of D2 should be between 10,000 and 20,000.
Page 67
E1 and E2 are the detector levels with the gas cell within the source unit in the sight path and will be roughly ½ of the D2 and D1 levels. The analyzer may also display smoothed detector values; these are noted as d1, d2, e1, and e2.
Sat. # indicates whether the detector signals are saturating within the micro-processor (this value should always be zero). If it displays a number other than zero, it indicates saturation, and you should adjust the detector gain. Refer to Section 3.5.
Phase is the time correction applied for the calculation of the detector levels. This will be between 0.1 and 5.9 milliseconds. The instrument calculates this value, and you may not adjust it.

4.9.2 Modulation (chopper motor) frequency

The chopper blade should chop the IR radiation at a frequency of about 37 Hz. The processor measures this frequency and displays it from this option.
Operation
4.9.3
As the chopper blades interrupt the IR beam, they split the radiation into two. Half the time the blade obscures the beam, and half the time the beam radiates across the duct. The value of Mark/space should be between 0.9 and 1.1 where:
Mark/space = Time IR beam obscured/Time IR beam clear.

YVals and CO ppm

A parameter Y determines the calculation of the gas concentration; refer to Section A.7.5.
As a check on the program operation, you may view this parameter and the resulting raw gas calculation here.
The term Yx is the second averaging stack's held value, and the term Y(60) is the 60­second raw value from which all of the other averaging stacks are calculated. These gas values represent the raw data before averaging for the corresponding Y values. You can also display the Z values by pressing the arrow keys. The Z values are the adjusted Y values used to compensate for cross sensitivities in the measurement range.
Reference Manual 59
Page 68
Operation

4.9.4 Calibration data

You can examine the calibration factors determined during the calibration routine (SCcal) and the value currently being used (SCwkg) from this display. If the two values are different, this indicates a change in instrument temperature between the time of calibration and the current temperature. Press one of the arrow keys to examine the temperature information.
Temperature has a small effect on the filter/gas cell characteristics compensated for by the instrument. Temperature measurement is made within the receiver.

4.9.5 Fault condition

To display the current fault condition, press ENTER while this is displayed.
The instrument automatically selects this display mode if a fault condition occurs. The instrument recognized the following fault conditions:
1. ALL CLEAR: No fault condition.
2. Det. Saturated: The detector level gain within either the receiver or the control unit is too high for the current duct conditions.
3. Low Det. Level: Detector levels are too low (< 3,000).
4. Mod. Freq.O.R.: Chopper motor frequency is out of range (< 30 Hz or > 45 Hz).
5. Reference Fail.: No reference signal from the source unit.
6. Cal. Fact. O.R.: After the calibration routine, the calculated Set Cal factor is out of range. Refer to Chapter 6.
Press the arrow key to observe the previous fault condition.
Note
If a fault condition exists, the instrument does not update the minutes, hours, and days averages. Refer to Section 6.1.1.
60 Rosemount CCO 5500
Page 69

4.10 Set up mode

You can change all operating parameters (averaging times, output settings, normalization parameters, path length, calibration, etc.) from this mode. To prevent any unauthorized changes, you must enter a four digit code before you can enter this mode.
Note
After you select this mode, the instrument suspends its operation and extinguishes the Data Valid LED. If you don't press a key within five seconds after selecting this mode, the Rosemount™ CCO 5500 Analyzer will revert to the normal OPERATING mode.
Rosemount recommends that you complete Table 4-1 to provide a record of the instrument setup in order to aid configuration and to record any subsequent changes to the operating parameters.
Press MODE until the number 5 is displayed in the top left corner. After you have correctly entered the security code, there are six submodes of operation from which you can change
the setup parameters. These six submodes are listed below:
1. Set Averages: You may set the four averaging stack times (seconds, minutes, hours, and days) as required.
2. Configure O/P: Analog output setup: origin, units, span, rolling average, and fault condition.
3. Parameters: You may set the following from this mode: security code, identity number, path length, alarm level, cal factor, and plant status.
4. Normalization: You may set up all normalization parameters from this mode.
5. Reset Average: Select this submode to reset the four averaging stacks.
6. Calibrate: You can set the outputs of the detectors and the basic calibration of the instrument.
Operation
Use the arrow keys to toggle between these six options and press ENTER when the desired option is displayed.
4.10.1
Reference Manual 61

Enter security code

Complete the following steps to enter the security code.
1. Once the display is as shown here, press ENTER to gain access to SET UP mode.
The cursor flashes over the first digit of the security code number.
2. Select the required first digit with the arrow keys and press ENTER.
3. Repeat this procedure for the remaining three numbers.
Page 70
Operation
If the code is correct after you press ENTER on the last digit, the sequence is continued. If it is not correct, the instrument returns to OPERATING mode. Refer to
Section 3.7.4 for further details.
Note
Rosemount™ sets the code number to 0000 at the factory; you should change it in SET UP mode.

4.10.2 Set averages

The instrument calculates four separate averages. These are defined in units of seconds, minutes, hours, and days. You can use any of the four averaging stacks to provide the instrument's analog output. You can set each averaging time within predefined limits.
1. Press ENTER when this display is shown.
The display now shows one of the averages.
2. Use the arrow keys to select the average time that you need to change and press ENTER to change it.
You can now change the value using the arrow keys and confirm it by pressing ENTER.
3. Set the seconds averaging stack to the required value.
This is limited to within 10 to 60 seconds in 10-second intervals.
4. Set the minutes averaging stack to the required value.
This is limited to within 1 to 60 minutes in 1-minute intervals.
5. Set the hours averaging stack to the required value.
This is limited to within 1 to 24 hours in 1-hour intervals.
6. Set the days averaging stack to the required value.
62 Rosemount CCO 5500
Page 71
This is limited to within 1 to 30 days in 1-day intervals.

4.10.3 Configure O/P

You can set up the analog loop output from this mode.
1. Press ENTER while this display is shown to access it.
2. Press the arrow keys to step through the available options.
3. Press ENTER to access and change the displayed parameters of each of the six available options listed below:
a. Output
b. Averages
c. Units
d. Output Span
e. Fault Condition
f. Set mA Output
Operation
Output
You can set an origin of 0 or 4 mA for the current loop output. The arrow keys toggle between these two options. Press ENTER to enter the new value.
Average
You may use any of the four averaging stacks (seconds, minutes, hours, and days) for the analog output. You can select them with the arrow keys and enter them with the ENTER key.
Reference Manual 63
Page 72
Operation
Units
The analog output can represent the gas concentration in units of mg/m3, mg/Nm3, or vpm. The arrow keys toggle between these three options. Press ENTER to enter the new value.
Output Span
Select the required span using the arrow keys for each digit. Press ENTER to enter the value of each digit. The units are displayed in vpm, mg/m3, or mg/Nm3, depending on what has been selected beforehand.
Note
Once you select Output Span, the current value is displayed for one second. The first digit of the display then defaults to zero; thus you must re-enter the span value for the unit to function correctly.
Fault condition
If a fault condition occurs, you may set the current output of the instrument to one of the following options:
Procedure
1. Set the output at 0 mA - ZERO.
2. Adjust the output to the calculated gas concentration even though a fault condition exists - MEAS.
3. Hold the last selected gas concentration - HOLD.
4. Set the output to full scale (20 mA) - F.S.
Press the arrow keys to select one of these options; when the desired output is displayed, press ENTER to confirm.
Set mA output
Note
This is set at the factory and should not be altered without due consideration.
64 Rosemount CCO 5500
Page 73
Operation
From this option, the current levels of the analog output are set up. Press ENTER to select it, and the instrument prompts you to set the current levels at 0 and 20 mA.
When this is displayed, set the current output to 0 mA as measured with a calibrated current meter across the analog ouput loop terminals; do not connect anything else to these terminals when setting up the output.
Use the two arrow keys to adjust the value; the UP arrow takes the current output up, and the DOWN arrow takes it down. Press ENTER when the correct output is displayed on the ammeter.
Note
Zero mA should be set up no matter what has been selected as the base of the current output. This is factory set.
In a similar manner to the above, set the current output level to 20 mA.

4.10.4 Parameters

Complete the following steps to set the parameters.
1. With this option displayed, press ENTER to access the list of six available options.
The arrow keys cycle through these options.
2. When the option you need to change is displayed, press ENTER.
3. When you have made all required changes, select EXIT and press ENTER.
The six available options are:
a. Security Number
b. Identity Number
c. Path Length
d. Alarm
e. Cal. Factor
f. Plant Status Input
Reference Manual 65
Page 74
Operation
Security Number
To prevent any unauthorized tampering with the setup information, it is important to change the security code from the factory setting.
Select each digit with ENTER and change it with the arrow keys.
Note
It is important to make note of this number; otherwise, it will not be possible to change the instrument set up.
Identity Number
If you are using the system as part of an integral monitoring system and are using the serial input and outputs, the central processor requires a Device Identity to identify each instrument. This number must be unique for each equipment item and can be set from 1 to 30 as required.
Path Length
Note
Once you select path length, the instrument displays the current value for one second. The first value of the display then defaults to zero; therefore, you must re-enter the value to calculate the gas concentrations correctly.
The transmissivity of any gas depends both on the concentration and on the path length through which the radiation is transmitted. Similarly, the output of the Rosemount™ CCO 5500 Analyzer gas monitor also depends on the path length of the flue gas through which the radiation is transmitted.
Refer to Section 2.4.1.
The Rosemount CCO 5500 Analyzer is sensitive to the product of concentration and path length. In order to obtain a true value of concentration of gas, you must input the correct path length into the processor. The processor then uses the value to produce a final value of gas concentration.
Note
The path length entered must represent the length of the actual gas pass, not the flange to flange dimension between the source and receiver.
66 Rosemount CCO 5500
Page 75
Operation
Alarm
A contact output is available to warn of a high gas concentration. You may trigger the contact output from any of the four averaging stacks. Select the source with the arrow key and enter it with ENTER.
Select the units for the alarm; these may be different than the units selected for the analog output.
After you select the source, the instrument requires a level that will trigger the output. Set the desired level with the arrow keys.
Cal Factor
Note
First, record the original Cal Factor before entering this mode as displayed in Mode 4 Calibration Data. The Cal Factor is lost when the menu option is entered.
During the calibration routine, the instrument calculates a Cal Factor which sets the basic calibration of the instrument. You may change this value from this mode.
Note
Since this value controls the calibration of the instrument, only change if necessary.
Plant Status Input
Use this to determine whether the plant is operating under correct conditions.
There is a choice of three controls for plant status: Logic Input, Serial Input, and Multiple. You can only use one to control plant status at any one time.
Reference Manual 67
Page 76
Operation
Logic Input
If the PS1 and PS2 terminals are linked in the control unit, the logic contact is made, and the plant status is OFF. You may link these terminals manually during a plant shut down, or you may wire them to a switch/contact outside the unit (e.g., a value that opens and closes the duct). Press ENTER to select this option when the Plant Status Serial Input option is displayed.
Serial input
If you select this option, the criteria controlling plant status are transmitted via the serial data link. Press ENTER when the Plant Status I/P Serial Input option is displayed to select this option.
Multiple
Four options are available here. Press ENTER when the Plant Status I/P Multiple option is displayed, and the first option Temperature is displayed. Use the arrow keys to toggle YES or NO. NO means that the temperature threshold is not used to determine plant status. If you select YES, the display enters the display below. Configure the instrument for temperature threshold. Press ENTER when it is correctly configured, and the display moves to the next option Oxygen. After you set the last option, the Logic Input display returns to the PARAMETERS Plant Status I/P option. Use the DOWN arrow to scroll down to EXIT and press ENTER. The plant status is now full configured.
Plant status is only OFF if all options selected are registering plant status OFF. If any one of them is not fulfilling plant status OFF conditions, then the instrument registers plant status ON.
- Temperature
A value is set here for the temperature threshold. While the temperature (taken from the normalizing temperature) is above the threshold value, plant status is ON. If the temperature drops below the threshold, plant status is OFF, and only the seconds averaging stack updates.
- Oxygen
Oxygen is set and used in a similar manner to the temperature threshold. However, if the normalizing oxygen level rises above the threshold, plant status is OFF. For plant status ON, the oxygen level must be below the threshold.
- Water Vapor
68 Rosemount CCO 5500
Page 77
This is set and used in a similar manner to the temperature threshold. If the normalizing water vapor level falls below the threshold, plant status is OFF. For plant status ON, the water vapor level must be above the threshold.
- Logic Input
Select YES or NO and press ENTER. For plant status to be ON, the logic input (PS1 and PS2) must be open circuit; for plant status to be OFF, the logic input must be closed circuit. After you configure this option, the menu exits to the Multiple option. Use the DOWN arrow to select EXIT and press ENTER.

4.10.5 Normalization

Complete the following steps to set the Normalization parameters.
1. Press ENTER to access one of the four normalizing parameters listed below.
Operation
2. Use the arrow keys to cycle through the four options and press ENTER once the desired option is selected.
a. Temperature
b. Oxygen
c. Pressure
d. Water Vapor
Once selected, each of the above normalizing parameters have the same suboptions: Set Standard Levels and Set Values. Under Set Values, are three more suboptions that determine how the instrument reads the data: Analog Input, Serial Input, and Keypad Input.
Set Standard Levels
Each normalizing parameter normalizes the measured gas concentration to standard conditions of temperature, oxygen, pressure, and water vapor. Set these levels within this option. Use the arrow keys to change each displayed normalizing standard value.
Temperature
Always use an analog input for temperature correction; this ensures that the flue gas temperature is being measured continuously and accurately. Connect the analog output of the temperatue transducer into the Rosemount™ analyzer and select the analog input option. This value is used to normalize the gas concentration measurement and to correct for the effects of temperature on the IR absorption spectrum.
Reference Manual 69
Page 78
Operation
If you use the Keypad Input, and the gas temperature is higher than 572 ° F (300 °C), the compensation algorithm becomes less precise, and instrument accuracy deteriorates accordingly. Rosemount does not recommend this.
Note
If normalization is not required, the instrument must hold the temperature of the gas in the duct using the Analog Input option.
Press ENTER with the Temperature option selected to access Set Standard Levels and Set Values. Use the arrow keys to toggle between these options and Exit.
Set Values
You can bring the normalizing data into the instrument in one of three ways:
Analog Input
This uses the 4-20 mA inputs within the processor to receive the measured transducer data. The values at 4 mA and 20 mA will be requested should this option be selected.
Serial Input
If you use an input unit, all normalizing data can be transmitted via the serial data line.
You can enter a fixed value via the keypad. This is suitable where the value is stable
to about ±5%.
With an integrated system, set the lead analyzer's normalizing parameters to the 4-20 mA inputs. Then set all the other analyzers to serial, and the normalization parameters are transmitted down the serial data highway.
Oxygen
To correct the data to standard levels of oxygen, you must enter an estimate of the oxygen at the point of measurement. If the oxygen level is being continuously measured, connect the analog output of the oxygen analyzer into the Rosemount™ CCO 5500 Analyzer and select Analog Input. You must define this input as either WET or DRY depending on how the
70 Rosemount CCO 5500
Page 79
Operation
measurement is made. After you define the wet or dry, you need to define the Analog Input values; set the 4 mA and 20 mA values. If the oxgyen level is relatively constant through all firing conditions, you may use a fixed keypad input.
With an integrated system, you can take the oxygen data to the instrument via the serial data line.
Note
If normalization is not required, you must set the normalizing parameters for oxygen in the instrument.
Pressure
To correct the data to a standard pressure, normally 14.7 psi (101 kPa), you must determine the pressure at the point of measurement. If the flue pressure is relatively constant through all firing conditions, then you may use a fixed keypad input. If the pressure, is not constant, measure it and bring it into the instrument via the 4-20 mA analog input within the processor.
4.10.6
Note
If normalization is not required, set the normalizing parameters for pressure in the instrument to
14.7 psi (101 kPa)(Standard Level and Keypad Input).
With an integrated system, you can take the pressure data to the instrument via the serial data line.
Water Vapor
An across the duct monitor measures the gas concentration under wet conditions. Unlike a sampling system, the gas has not been preconditioned in any way before you make a measurement.
When the water vapor is at a relatively fixed level, set the standard level to DRY to normalize it to dry conditions. Use a fixed value in the keypad option representing the expected water vapor produced for the fuel type. If the measurement is not to be normalized for water vapor, set the standard level to WET.
With an integrated system, you can take the water vapor data to the instrument via the serial data line.

Reset averages

Note
Resetting averages causes the rolling average data to be cleared from memory.
Reference Manual 71
Page 80
Operation
You can reset the average values that are currently held in the four averaging stacks using this option; this ereases the current average that is held in all of the averaging stacks. Select this option by pressing ENTER and using the arrow keys. The instrument requests confirmation before the averages are reset.
Note
If you select this option, all the data in the averaging stacks is reset, and the data for as much as 30 days is lost.

4.10.7 Calibrate

From this option, you may display the two detector levels and conduct a basic calibration. While in this mode, the gas cell is not moved; this gives an immediate response for setting up the detector levels. A Cal Factor that is calculated during a calibration routine sets the basic calibration of the instrument. Press ENTER while this is displayed to see the following options:
1. Set Detectors
2. Span Adjust
3. Calibrate
Set detectors
You can display both the D1 and D2 levels; you can also display saturation counts. To give an immediate response to any alterations that are required, the filters and gas cells are not moved during this operation.
Refer to Section 3.5 for a discussion of the detector level and saturation count.
Span adjust
Note
Rosemount™ initially sets the span factor at the factory; do not adjust it unless the instrument sensitivity is suspected. In any case, Rosemount recommends that you record the orginal value before making adjustments.
72 Rosemount CCO 5500
Page 81
Operation
You can adjust instrument sensitivity if a known concentration of gas exists between the source and receiver units and instrument sensitivity is supsected. If a problem arises, consult Rosemount.
You may need to adjust the span factor if you have fitted new gas cells or filters.
Calibrate
Re-enter the Mode 5 CALIBRATE menu and proceed to the Calibrate option.
You can calculate the basic calibration of the instrument from this routine. It is preferable to conduct this operation with the plant shut down to ensure a zero gas concentration within the duct. If this is not possible, the instrument can calibrate to a known value of the gas concentration - the calibration target.
Set the calibration target either to a known value of the gas concentration or to zero. Then set the desired number of cycles over which the calibration factor is determined (Rosemount™ recommends a minimum of 30). Now run the calibration, and the display will show a countdown during its execution. When the calibration is complete, the new Cal Factor is displayed for about five seconds, and the instrument exits the calibration routine.
Note
You must run the calibration routine during commissioning; otherwise, the instrument will not be able to calculate the true level of gas within the duct.
Reference Manual 73
Page 82
Operation
Do not run the calibration routine unless reasonable conditions exist in the duct. If it is not the initial calibration, record the Cal Factor from the Parameters option before running the calibration.
To aid configuration and to record any subsequent changes to the operating parameters,
#xd_5bc5da84c7ce5b06--2a6d55ac-163c9dfb12c--7ead/table_jlb_4y5_b2b lists all of the
options available and can be used as a record of the operating parameters.
Instrument SettingsTable 4-1:
Parameter Config
Averages
Seconds
Minutes
Hours
Days
Output
0 or 4 mA base
Units
Average
Fault condition
Parameters
Path length
Alarm source
Alarm units
Alarm levels
Normalization
Temperature
Standard level °F (°C)
I/P °F (°C) @ 4 mA
I/P °F(°C) at 20 mA
Keypad input °C (not ideal)
Serial input
Oxygen
Standard level %
Wet or dry gas
I/P % @ 4 mA
I/P % @ 20 mA
Keypad input %
Serial input
74 Rosemount CCO 5500
Page 83
Operation
Instrument Settings (continued)Table 4-1:
Parameter Config
Pressure
Standard level psi (kPa)
I/P psi (kPa) @ 4 mA
I/P psi (kPa) @ 20 mA
Keypad input
Serial input
Water vapor
Standard level (wet %/dry
Keypad input %
Measured value
After the calibration is conducted, a cal factor is calculated; this determines the calibration of the instrument. The cal factor is displayed for a few seconds after a calibration has been conducted, and you can aslo interrogate it from Mode 4 (DIAGNOSTIC mode). Enter the cal factor into the table below as a record of instrument operation.
Note
Rosemount™ obtained factory settings under the following conditions:
2 m path length
Clean conditions
Calibration DataTable 4-2:
Detector outputs Factory Config
D2
D1
E1
E2
Calibration data
Cal factor
Span factor
Output calibration
Set zero
Set span
Reference Manual 75
Page 84
Operation

4.11 Check Cell mode

Use this mode with a Rosemount™ check cell. It verifies calibration and operation of the analyzer. For details for this mode's use, refer to Section 4.13.1.
Procedure
1. Press ENTER on viewing this display if you need to use the CHECK CELL mode.
2. Use the arrow keys to toggle the display to YES and press ENTER again.
You have now selected the CHECK CELL mode. Note that the display will default back to Mode 1 if you don't select YES within five seconds. Do not insert a the check cell before accessing this option. Refer to Section 4.13.1.

4.12 Shutdown procedure

You do not need to shut down the system unless you are performing maintenance.
1. In this case, power down the power supply (and thus the control unit and analyzer heads).
2. Close the manually operated isolating valves on the analyzer heads.
3. Proceed with service.
4. When completed, open the valves and power up the system.
Postrequisites
You may need to calibrate the system depending on the type of maintenance you are doing. For details, refer to Chapter 6
4.13
4.13.1 Notes for using a Rosemount™ check cell

Routine checks

The check cell and holder are optional items available from Rosemount. Rosemount has designed the check cell to verify the reading of Rosemount cross duct analyzers. When placed within the measurement path, the check cell can generate a known increase in gas concentration.
76 Rosemount CCO 5500
Page 85
Operation
Prerequisites
Note
Place the check cell at the receiver side.
Measurement conditions
For absolute verification, you must check the instrument when no measurement gas is present. If a background concentration of measurement gas is present, an increase will still be generated, but the net effect will be complex.
Mode 6
The ideal time to perform a check cell test is with the plant off, auto zero condition on, and the analyzer well stabilized at zero. Do not insert the check cell in any other mode than Mode 6.
The analyzer does not update rolling averages while in Mode 6.
Procedure
1. Enter Mode 6 on the control unit before inserting the check cell into the analyzer. Press ENTER when Mode 6 is displayed.
2. Use an arrow key to toggle from NO to YES and press ENTER to access the check cell function.
If you don't select this within five seconds, the instrument returns to the normal OPERATING mode.
3. When you see this option, insert the check cell observing the procedure outlined below.
4. Remove the two screws retaining the cover on the check cell holder (Figure 4-4).
Reference Manual 77
Page 86
Operation
Check Cell HolderFigure 4-4:
A. Check cell cover B. Receiver C. Check cell holder
5. Insert the check cell into the check cell holder and replace and tighten the screws.
You can insert the cell in either direction. Refer to Figure 4-5.
Check CellFigure 4-5:
A. Gas cell window
Introducing the check cell may cause an initial major disturbance to the instrument operation.
6. Wait for the instrument reading to settle (five to ten minutes) and record the gas measurement with the cell in position.
7. Remove the check cell and wait for the analyzer to return to zero (another five to ten minutes).
8. Replace the cover on the check cell holder and place MODE on the control unit.
The instrument now returns to OPERATING mode.
78 Rosemount CCO 5500
Page 87
You can do this test with the plant on line, but any pollutant gas present (it will probably be residing at a different temperature) will interfere with the check cell value.
Note
Do not insert the check cell in any other mode as this will influence the recorded rolling averages.
Rosemount fixes the calibration of the Rosemount CCO 5500 Analyzer at the point of manufacture. If gross errors exist, this could suggest an instrument malfunction. If you observe minor errors, please check the procedure and if necessary, return the gas cell for recertification.

4.13.2 Alarms and emergency conditions

The alarm thresholds for the system are set as described in the normal OPERATING mode. When an alarm condition occurs, the red LED on the control unit lights up. This goes out when the alarm condition has cleared. When the condition has cleared, the software records the fact that an alarm condition occurred. The 4 to 20 mA output from the analyzer will also alter according to the pollution levels detected.
Operation

4.13.3 Emergency shutdown procedure

This is the same as normal shutdown.
1. Remove power from the power supply.
The whole system closes down.
2. Close the isolating valves if required and stop the flow of air to the air purges.
4.13.4

Isolation procedure

Complete the following steps to isolate the analyzer.
1. Shut down power to the power supply.
2. Shut off compressed air to the air purges and close the isolating valves.
Reference Manual 79
Page 88
Operation
80 Rosemount CCO 5500
Page 89
5 Maintenance

5.1 Preventative maintenance

Rosemount designs this equipment to keep maintenance to an absolute minimum.

5.1.1 Cleaning windows

It is important that you keep the optical windows of both the source and receiver reasonably clean. Keep any mounting tubes free from build-up of dust and fly ash. Clean the optical window (Figure 5-1) every six months and more frequently for dirty processes.
Optical WindowFigure 5-1:

Maintenance

A. Air purge B. Source C. Window
WARNING!
HIGH TEMPERATURE AND DANGEROUS VAPORS Take great care when removing the Rosemount™ CCO 5500 Analyzer from a positive pressure stack or duct. The source may be very hot, and there may be dangerous vapors present. Observe all required safety practices.
Procedure
1. Unbolt and remove the source and receiver from their air purges.
2. Wipe the windows with a soft dry cloth.
Reference Manual 81
Page 90
Maintenance

5.2 Corrective maintenance

The following maintenance procedures describe the necessary steps for removing and replacing failed elements of the Rosemount™ CCO 5500 Analyzer.
WARNING!
ELECTRIC SHOCK Install all protective equipment covers and safety ground leads after equipment repair or service. Failure to install covers and ground leads could result in serious injury or death.

5.2.1 Replace heater element

The heater element has a finite life and at some stage you will have to replace it. Rosemount™ designed the unit to give a minimum of two years continuous operation; you can replace the heater element on site when necessary. Replace a failed heater element according to the following procedure.
Procedure
1. Shut down and lock out power to the Rosemount CCO 5500 Analyzer.
2. Unbolt and remove the source unit from the air purge.
3. Loosen the source unit cable gland (Figure 5-2) so the cable can slip when the rear cover plate is removed.
Source Rear Cover PlateFigure 5-2:
A. Rear cover plate B. Cable gland
4. Unsccrew and remove the rear cover plate.
5. Unscrew and remove the circuit board (Figure 5-3).
82 Rosemount CCO 5500
Page 91
Source Unit Circuit BoardFigure 5-3:
A. Circuit board mounting screw B. Circuit board
Maintenance
6. Remove the terminal nuts (Figure 5-4) and remove the heater wires.
Heater ElementFigure 5-4:
A. Terminal nuts B. Captive screw C. Heater
7. Unscrew the three captive screws.
8. Remove and discard the heater assembly.
9. Install the new heater assembly and tighten the three captive screws.
Reference Manual 83
Page 92
Maintenance
10. Install the two wires and terminal nuts removed in Step 6.
11. Install the circuit board and retaining screws removed in Step 5.
12. Install and fasten the rear cover plate.
13. Power up the Rosemount CCO 5500 Analyzer and allow fifteen minutes for the heater to reach operating temperature.
The analyzer starts to monitor the gas levels.

5.2.2 Replace chopper motor assembly

Complete the following steps to replace the chopper motor assembly in the analyzer.
1. Shut down and lock out power to the Rosemount™ CCO 5500 Analyzer.
2. Unbolt and remove the source unit from the air purge.
3. Remove the window plate (Figure 5-5) from the source unit.
Source Unit Circuit Board and Center Plate AssemblyFigure 5-5:
A. Black wire (M-) B. Red wire (M+) C. Window plate D. Circuit board and center plate assembly E. Brass screws
84 Rosemount CCO 5500
Page 93
Maintenance
4. Unscrew the three brass screws that secure the circuit board and center plate assembly.
5. Lift the circuit board and center plate assembly out of the source unit.
6. De-solder the red and black chopper motor power wires from the circuit board.
7. Unscrew and remove the three chopper motor screws (Figure 5-6).
Chopper Motor AssemblyFigure 5-6:
A. Chopper motor assembly B. Chopper motor retaining screws C. Center plate
8. Install the new chopper motor and retaining screws removed in Step 7.
9. Solder the chopper motor power wires to the circuit board. Solder the red wire to the M+ terminal and the black wire to the M- terminal.
10. Install the center plate and circuit board assembly. Secure the assembly with the three brass crews removed in Step 4.
11. Line up and install the window plate onto the source unit. Fully seat the window plate flange in the source unit.
12. Secure the source unit to the air purge using the four screws removed in Step 2. Tighten the screws evenly.
13. Power up the Rosemount CCO 5500 Analyzer and check the chopper motor frequency in Mode 4 (DIAGNOSTICS). Adjust using the trim potentiometer as described in Section 3.6.
Reference Manual 85
Page 94
Maintenance

5.2.3 Replace source unit gas cell

Replace the source unit gas cell according to the following procedure.
1. Shut down and lock out power to the Rosemount™ CCO 5500 Analyzer.
2. Unbolt and remove the source from the air purge.
3. Remove the window plate (Figure 5-5) from the source unit.
4. Unscrew the three brass screws that secure the circuit board and center plate assembly.
5. Lift the circuit board and center plate assembly out of the source unit.
6. Loosen the setscrew at the end of the gas cell assembly (Figure 5-7).
Source Gas CellFigure 5-7:
A. Gas cell B. Setscrew
7. Pry the gas cell from the stepper motor shaft.
8. Insert a new gas cell onto the stepper motor shaft and tighten the setscrew.
9. Install the center plate and circuit board assembly and secure with the three brass screws removed in Step 4.
10. Line up and install the window plate onto the source unit. Fully seat the window plate flange in the source unit.
86 Rosemount CCO 5500
Page 95
11. Secure the source unit to the air purge using the four screws removed in Step 2. Tighten the screws evenly.
12. Power up and recalibrate the Rosemount CCO 5500 Analyzer.
Refer to Calibrate.

5.2.4 Replace receiver unit gas cell

Replace the receiver unit gas cell according to the following procedure.
1. Shut down and lock out power to the Rosemount™ CCO 5500 Analyzer.
2. Unbolt and remove the receiver unit from its air purge.
3. Loosen the receiver unit cable gland so the cable can slip when the window plate is removed.
4. Unscrew and pull the window plate (Figure 5-8) from the front of the receiver unit.
Receiver Gas CellFigure 5-8:
Maintenance
A. End detector PCB B. Window plate C. M6 bolts D. M4 x 16 screws E. Side detector PCB F. Gas cell assembly
5. Use a marker pen to match-mark all components of the receiver stack.
6. Remove the four M4x16 screws that secure the side detector PCB to the receiver stack.
Reference Manual 87
Page 96
Maintenance
7. Remove the four M6 bolts that secure the detector stack components.
8. Observe the orientation of the gas cell assembly in the detector stack.
9. Slide the gas cell assembly out of the detector stack.
10. Replace the gas cell using the same orientation observed in Step 8.
(The notch in the gas cell assembly should be nearest to the large circuit board secured to the detector stack).
11. Place the end detector on top of the detector stack.
CAUTION!
EQUIPMENT DAMAGE
Do not overtighten the detector stack mounting bolts. Over-tightening the mounting bolts may fracture the gas cell.
12. Install and gently tighten the M6 mounting bolts removed in Step 7.
13. Position the side detector against the receiver stack. Install and tighten the M4x16 side detector mounting screws.
14. Line up and install the window plate onto the receiver unit. Fully seat the window plate flange in the receiver unit.
15. Secure the receiver unit to the air purge using the four screws removed in Step 2. Tighten the screws evenly.
16. Power up and recalibrate the Rosemount CCO 5500 Analyzer.
5.2.5
5.3
Refer to Calibrate.

Electronics

The electronics require no routine maintenance. They are all solid state and undergo a rigorous factory burn-in procedure. If there is any doubt about the electronics' performance, interrogate the control unit from the keypad to determine whether or not they are functioning properly.
Refer to Chapter 6.

Adjust span factor

The span factor does not require periodic adjustment. However, if either of the following repair actions are performed, you may need to adjust the span factor.
Gas cell changed either in the receiver or source.
Interference filter changed in the receiver.
Use the following procedure to adjust the span factor.
88 Rosemount CCO 5500
Page 97
Maintenance
Procedure
1. At the control unit keypad, select Mode 5 Configure O/P Set Span.
2. Set the span factor to 1000.
3. Calibrate the analyzer under zero conditions.
Refer to Calibrate.
4. Select Mode 5 Normalization. Set the temperature to 68 °F (20 °C).
5. Select Mode 5 Parameters Pathlength to enter the 3.28 ft. (1 m) path length.
Note
The check cell reading must be less than 999 ppmm. If greater than 999 ppmm, you must enter a larger path length to reduce the effective ppmm value of the check cell. Select a path
length such that ppmm/path length is less than 999.
6. Select Mode 5 Configure O/P Set Span. Enter the span factor indicated for the check cell in Mode 6. If this value exceeds 999 ppm, divide by the path length
entered above to determine the span.
7. Press ENTER and wait for the second reading to appear under the span factor option.
5.3.1
This should read the same as the value entered in Step 6.
8. Using the up and down arrow keys, adjust the span factor until the second reading agrees with the certified check value (divided by the path length if appropriate).
9. Record the span factor and press ENTER.
10. Confirm the span factor by selecting Mode 2 Parameters Span Factor and checking the value.

Reset span factor

An example of resetting the span factor is outlined below.
1. You insert a certified check cell of 2880 ppm into the check cell holder with the analyzer in Mode 6. You take a reading of 3240.
Due to the value of the check cell exceeding 999 ppm, you need to use a suitable path length to reduce the effective value of the check cell.
2. In this case, you choose a path length of 13.1 ft. (4 m); thus the effective value of the check cell is 720 ppm (2880/4).
However, in Mode 6 it reads higher (3240); this value is effectively 810 ppm at a 13.1 ft. (4 m) path length.
3. You key 810 in as the first value in the span factor option.
4. Then you adjust the span factor until the second value reads 720.
5. Press ENTER to set the span factor.
Reference Manual 89
Page 98
Maintenance
90 Rosemount CCO 5500
Page 99
6 Troubleshooting

6.1 Finding faults with the keypad

WARNING!
ELECTRIC SHOCK Install all protective equipment ground covers and safety ground leads after troubleshooting. Failure to install covers and ground leads could result in serious injury or death.
If a fault occurs, the control unit display switches from its current mode of operation to the DIAGNOSTICS mode and displays the current fault condition. Refer to Section 6.1.1 for further information on the fault conditions. If the fault clears, the display stays in DIAGNOSTICS mode and displays All Clear.
Enter DIAGNOSTICS mode from the instrument's keypad; you can do this at any time without interrupting or disturbing the analog outputs of the equipment. As an initial guide to equipment performance, typical values for instrument operation are given below.

Troubleshooting

Rosemount CCO 5500
6.1.1
D2 min
3,000 20,000 3,000 20,000 2,000 2,900
where:
Modulation frequency = 30 to 45 Hz
Mark/space ratio = 0.9 to 1.1
Saturation count = 50 maximum for all analyzers
If the values are outside of the above ranges, the Data valid LED extinguishes, and the fault condition is displayed.
D2 max D1 min D1 max Set Cal min Set Cal max

Data valid LED out

If one or more fault conditions occur, the data valid LED on the front panel extinguishes, the data valid relay operates, and the instrument automatically enters the DIAGNOSTICS mode to display the fault condition.
Fault ConditionsTable 6-1:
Fault condition Possible causes
Detector signal saturated (Sat # over 50) Incorrect detector gain adjustment.
Reference Manual 91
Page 100
Troubleshooting
Fault Conditions (continued)Table 6-1:
Fault condition Possible causes
Instrument condition during high opacity condi­tions which have now cleared.
Low detector levels (D1 < 3,000) High opacity in duct.
Dirty windows.
Incorrect detector gain adjustment.
Cal factor out of range Calibration conducted during unstable duct
conditions.
Poor alignment.
Incorrect detector gain adjustment.
Modulation frequency is below 30 Hz or above 45 Hz.
Reference signal failure Chopper motor fails.
Poor supply voltage.
Faulty chopper motor.
Incorrect setting within the source unit.
6.2
Note
If the instrument recognizes a fault condition, it does not update the minutes, hours, and day averages.

Troubleshooting tables

The troubleshooting tables provide fault diagnosis, possible causes, and the appropriate actions if you suspect an instrument fault. Note the symptoms and when the fault has occurred. Refer to the appropriate tables.
Note
Be sure to use the tables from the top down and pay particular attention when "proceed to next test" and "proceed to next possible cause" are indicated.
The troubleshooting tables indicate which of the four units failed:
Source unit
Control unit
Receiver unit
Power supply unit
You can then return the faulty unit to Rosemount™ for repair.
92 Rosemount CCO 5500
Loading...