Rosemount BINOS 100 Series Analyzers including OXYNOS 100, HYDROS 100 and CAT 100 Addendum to ETC00781 Regarding BKS 20 Hardware Upgrade-1st Ed. Manuals & Guides

Instruction Sheet
HAS1xE-IS-BKS October 2004
OXYNOS® 100, HYDROS® 100, CAT 100
Instruction Manual Addendum to Manual ETC00781
Series 100 Gas Analyzers
BINOS® 100, BINOS® 100 M
BINOS® 100 2M, BINOS® 100 F
Regarding BKS 20 Hardware Upgrade
www.EmersonProcess.com
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
Emerson Process Management (Rosemount Analytical) designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using and maintaining Emerson Process Management (Rosemount Analytical) products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
Read all instructions prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Emerson Process
Management (Rosemount Analytical) representative for clarification.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and maintenance of the product.
Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes.
Connect all products to the proper electrical and pressure sources.
T o ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts specified by Emerson Process Management (Rosemount Analytical). Unauthorized parts and procedures can affect the product’s performance, place the safe operation of your process at risk, and VOID YOUR W ARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
The information contained in this document is subject to change without notice. 1st edition: 2004-10
Emerson Process Management Manufacturing GmbH & Co. OHG
Industriestrasse 1 D-63594 Hasselroth Germany T +49 (0) 6055 884-0 F +49 (0) 6055 884-209 Internet: www.EmersonProcess.com
Instruction Sheet
HAS1xE-IS-BKS 10/2004
BKS 20 Upgrade
INTRODUCTION
The BINOS® 100 series analyzers are subject to a consistent improvement as well as all EMERSON PROCESS MANAGEMENT products. Within this improvement the main board BKS 10 was replaced by the successor BKS 20. Due to the short term development of this new printed circuit board the BINOS® 100 analyzer series instruction manual ETC00781 could not be adapted in time. This means that all illustrations and references within this manual still refer to the old board. While during operation this is not a disadvantage, it may cause problems when servicing and maintaining the analyzer . For this reasons EMERSON PROCESS MANAGEMENT has created this instruction sheet.
This Instruction sheet is intended to be used in conjunction with the BINOS® 100 analyzer series instruction manual ETC00781 delivered together with the analyzer.
It deals with service aspects and replaces respectively completes the corresponding chapters of the original instruction manual ETC00781.
We recommend to replace the corresponding chapters inside the manual by the chapters of this sheet. For this the pages in this sheet are numbered corresponding to the manual.
A new section is the attachment containing block diagrams for the different analyzer variations.
Contents
Chapter 16............... List of Failures.......................................................... Pages 16-1 to 16-8
Chapter 17............... Measuring Points ..................................................... Pages 17-1 to 17-6
Chapter 18............... Allocation of Connectors........................................... Pages 18-1 to 18-6
Chapter 19............... BKS Jumper Configuration....................................... Pages 19-1 to 19-2
Attachment............... Block Diagrams ...........................................................Pages A-1 to A-6
EMERSON Process Management Manufacturing GmbH & Co. OHG I-1
BKS 20 Upgrade
Instruction Sheet
HAS1xE-IS-BKS
10/2004
EMERSON Process Management Manufacturing GmbH & Co. OHGI-2
Instruction Sheet
HAS1xE-IS-BKS 10/2004
Error Code Possible Reasons Check / Correct
16. List of Failures
BKS 20 Upgrade
Some of the failures possibly coming up during operation are reported on the displays by the use of error codes.
In such a case the display's show the current concentration value alternating with
(E = error , XX = code).
The following table lists all error codes, some related reasons and measures to repair .
Error Code Possible Reasons Check / Correct
1. Displays are “switched OFF”
If there is an "error message", an optional digital status signal "Failure" may be set (see chapter 10.3)!
Be sure to observe the safety measures while working at the analyzer!
1. Press any key.
Check parameter dOFF (see chapter 8.7).
2. Voltage supply missing
Display dark
BINOS® 100 F
3. Connection front panel /BKS interrupted
EMERSON Process Management Manufacturing GmbH & Co. OHG 16-1
2. Check connection mains to P/S
Check external power supply
Check electrical supply (see Fig. 1-5 and 1-6, Item 3 Fig. 1-7 to 1-9, Item 10).
Check internal fuse on board BKS (s. page 18-1)
Check internal fuses F1 and F2
3. Check connection BKB to BKS (X1) (see chapter 18.1).
BKS 20 Upgrade
Error Code Possible Reasons Check / Correct
Instruction Sheet
HAS1xE-IS-BKS
10/2004
Flashing
Channel 1
Channel 2
Battery buffer failure
The EPROM default values have been loaded
If only one code is shown:
1. A/D converter of related channel is not driven correctly
If both codes are shown:
2. Positive or negative reference voltage missing
Check, if Jumper J 7 is set (see chapter 19.).
Replace battery, if battery voltage < 3,5 V (BKS - Jumper J7 set).
The error code is cleared by pressing any key or by serial interface instruction $627.
1. Switch analyzer off and then on again.
2. Check reference voltage (see chapter 17.1.2/17.1.3).
A/D Conversion Interrupt
missing
Temperature compensation
incorrect
3. Light barrier signal missing
4. IR channel: Chopper drive inoperative
5. Supply voltage (internal 6 V DC) missing
1. Failed to start A/D­conversion for temperature channel
2. Supply voltage (internal 6 V DC) absent.
3. Check connection X9 / light barrier (see chapter 18.)
Solder bridge LB 18 not configured properly, check measuring point (17.1.5)
4. Check connection X2 / chopper drive (see chapter 18)
Check measuring point (17.1.4)
5. Check measuring point (17.1.1)
1. Switch analyzer off and then on again.
2. Check measuring point 17.1.1
EMERSON Process Management Manufacturing GmbH & Co. OHG16-2
Instruction Sheet
HAS1xE-IS-BKS 10/2004
Error Code Possible Reasons Check / Correct
BKS 20 Upgrade
Channel 1
Channel 2
Tolerance error
Channel 1
Channel 2
Tolerance error
Zero gas value differs from
zero more than 10% of
measuring range .
Channel 1
Channel 2
1. Span gas in use exceeds measuring range
1. Wrong zero gas in use.
2. IR channel: Photometer section contaminated.
3. Analyzer not calibrated.
1. Wrong nominal value.
2. Wrong span gas in use.
1. Check span gas in use and replace by gas with correct nominal value
1. Check zero gas in use.
2. Check analysis cell and windows for contamination.
Clean contaminated parts (see chapter 24.3).
3. Switch off the tolerance check before starting a calibration (see chapter 8.6).
1. Enter the correct nominal value (refer to span gas bottle certificate) (see chapter 9.1.2).
2. Check span gas in use. Use another or a new gas bottle. Enter the correct nominal value
Tolerance error
Span gas value differs from
nominal value more than
10% .
Channel 1
Channel 2
Measuring value more than
10% over full scale range.
EMERSON Process Management Manufacturing GmbH & Co. OHG 16-3
3. IR channel: Photometer section contaminated.
4. Analyzer not calibrated.
1. Concentration of sample gas too high.
3. Check analysis cell and windows for contamination.
Clean contaminated parts (see chapter 24.3).
4. Switch off the tolerance check before starting a calibration (see chapter 8.6).
1. Check concentration of sample gas
2. Use an analyzer suitable for the existing concentration range
BKS 20 Upgrade
Error Code Possible Reasons Check / Correct
Instruction Sheet
HAS1xE-IS-BKS
10/2004
Channel 1
Channel 2
Failure while entering limits
Channel 1
Channel 2
1. Absorber in NVE: lifetime cycle ended
2. Parameter “AbS.” not set to “0”
XON (serial interface) timed out.
wrong limit value entered due to shift of decimal point
Wrong nominal value entered due to shift of decimal point
1. Replace absorber. Enter menu “SYS.-PArA” and set parameter “AbS” to “2” (new life time cycle)
2. Enter menu “SYS.-PArA” and set parameter “AbS.” to “0”
XON - character is missing while driving the serial interface (time out > 60 s).
Correct limit value in line “LI.”
1. Correct nominal value by performing a span gas calibration
2. Correct nominal value via serial interface by sending “$029”
Wrong nominal value entered
for span gas
Channel 1
Channel 2
Wrong offset value entered
Wrong offset value entered due to shift of decimal point
EMERSON Process Management Manufacturing GmbH & Co. OHG16-4
1. Correct offset value “OFS.x”
Instruction Sheet
HAS1xE-IS-BKS 10/2004
Error Code Possible Reasons Check / Correct
BKS 20 Upgrade
Pressure measurement
failure
EPROM checksum
failure
Test for SRAM - IC's
failed
1. Measuring range failure.
2. Connection faulty.
3. Pressure sensor faulty.
1. EPROM faulty.
2. BKS faulty.
SRAM IC's / BKS faulty.
1. Pressure not within the sensor measuring range
2. Check connection P1 (at BAF 01) / pressure sensor (see chapter 18.).
3. Replace pressure sensor.
1. Replace EPROM (see chapter 28.).
2. Replace BKS.
Replace BKS.
Analog output signal
missing
Fluctuating or
erroneous display
BKS faulty.
1. Leakage in gas path
2. Ambient air contains gas component to be measured in a concentration influencing measurement
Replace BKS.
1. Perform a leakage test (see chapter 22)
2. Replace absorber material for IR sources and chopper housing
Use sealed photometer (Option)
Purge analyzer with inert gas
EMERSON Process Management Manufacturing GmbH & Co. OHG 16-5
BKS 20 Upgrade
Error Code Possible Reasons Check / Correct
Instruction Sheet
HAS1xE-IS-BKS
10/2004
Fluctuating or
erroneous display
(cont’d)
3. Gas pressure subject to considerable variation
4. IR detector / Oxygen senor not connected.
5. Electrochemical oxygen sensor is weared
6. IR channel: IR source not connected or faulty.
3. Check the gas lines preceding and beyond the sensor cell.
Eliminate any restrictions beyond the gas outlet fitting.
Reduce pump rate or flow rate.
4. Check connections: BKS X5 detector channel 1 (Oxygen sensor or IR detector) BKS X6 detector channel 2 (IR detector) (see 18.)
5. Replace sensor (see 25.)
6. Checkconnection: BKS X3(1/2) - IR source channel 1 BKS X4(1/2) - IR source channel 2 (see 18.)
7. Faulty analog preamplifying
8. Contamination of the gas paths.
If IR source housing is cold: For dual IR channel analyzer interchange the two IR sources. Replace the suspect IR source (see 24.2).
7. Check measuring point 17.1.7 or
17.2.1 resp.
8. Check analysis cell and windows for contamination.
Clean contaminated parts (see 24.3).
Check gas paths for contamination.
EMERSON Process Management Manufacturing GmbH & Co. OHG16-6
Instruction Sheet
HAS1xE-IS-BKS 10/2004
Error Code Possible Reasons Check / Correct
BKS 20 Upgrade
Fluctuating or
erroneous display
(cont’d)
Response time too long
(t
time)
90
9. Influence by barometric pressure
10. Temperature in the gas paths below dew point.
11. A/D converter failure
1. Incorrect response time ( t
time) settings.
90
2. Pump rate too low
9. Enter the correct value for baro­metric pressure (see chapter 8.1). Pressure sensor failure (E.37).
10. Check the temperature of the gas paths and remove any cold spots
Keep all temperatures at values at least 10 °C above the dew point of the gas.
11. Replace BKS.
1. Check the value for t
90
time
(see chapter 8.11).
2. The feeder line between the sampling point and the analyzer is too long. Use a larger, external pump; consider adding a bypass line to the process stream for sampling purposes (see chapter 5.1).
3. Contamination of the gas paths.
3. Check gas paths and gas conditioning for contamination.
Clean gas paths and replace the filter elements.
EMERSON Process Management Manufacturing GmbH & Co. OHG 16-7
BKS 20 Upgrade
Error Code Possible Reasons Check / Correct
Instruction Sheet
HAS1xE-IS-BKS
10/2004
No gas flow
1. Optional needle valve not opened
2. Sample gas pump (Option) is not switched on
3. Membrane of pump is defective
4. Sample gas pump is defective
5. Solenoid valves (Option) defective
1. Open needle valve (Option, see Fig. 1-2)
2. Press “PUMP” key
3. Replace membrane of the pump.
4. Replace sample gas pump
5. External valves: Check connection between valves and analyzer`s digitaloutputs
Valves controlled by optional serial interface / digital inputs: Any valve activated?
6. Contamination of gas paths
Check the valve seat of the solenoid valves and replace if necessary.
Replace solenoid valves
6. Check gas paths including filters for contamination.
Clean gas lines and replace the filter elements.
EMERSON Process Management Manufacturing GmbH & Co. OHG16-8
Instruction Sheet
HAS1xE-IS-BKS 10/2004
17 . Measuring Points of BKS and OXS
Be sure to observe the safety measures !
17.1 Measuring points of BKS
(Connector’s pins 1 on the board are identified by a mark or a square solder pad)
All measuring points are measured against ground (X 1 1 or X 28) !
17.1.1 Supply Voltage +6 V
BKS 20 Upgrade
Measuring point: X14 Measuring device: DVM Signal: + 6 V DC (+10 mV / -200 mV)
(adjust with potentiometer R90, if necessary)
Failure: No signal Possible reasons: a) Supply voltage is missing
b) Supply voltage < 9 V
or of reverse polarity c) BKS faulty (replace) d) Fuse F2 opened
17.1.2 Positive Reference Voltage
Measuring point: X5, pin 6 Measuring device: DVM Signal: + 5,535 V DC (± 60 mV) Possible reasons: a) Supply voltage (+6V) is faulty
b) Reference voltage +3 V not
correct (measure at C179,
pin 1(left pin in fig. 17-1a))
EMERSON Process Management Manufacturing GmbH & Co. OHG
Front panel side
Fig. 17-1a: PCB “BKS”, measuring points
17-1
BKS 20 Upgrade
17.1.3 Negative Reference Voltage
Measuring point: X5, pin 4 Measuring device: DVM Signal: inverted positive
reference voltage
The difference between negative reference voltage and positive reference voltage must not
Instruction Sheet
HAS1xE-IS-BKS
10/2004
be more than 10 mV (U
ref. pos.
+ U
ref. neg.
!
If the difference is higher , replace BKS.
17.1.4 Temperature Sensor
Measuring point: X8 Measuring device: DVM Signal: approx. 0 ± 500 mV DC
(at ambient temperature)
Possible reasons
- IR measurement or paramagnetic oxygen measurement a) Temperature sensor not connected
(see chapter 18.1). b) Temperature sensor faulty (replace sensor). c) Broken temperature sensor cable
≤ ± 10 mV)
(replace sensor). d) BKS faulty (replace BKS).
- Electrochemical oxygen measurement: a) Temperature sensor not connected
(see chapter 18.2). b) PCB OXS faulty (replace OXS). c) PCB BKS faulty (replace BKS).
17-2
Front panel side
Fig. 17-1b: PCB “BKS”, measuring points
EMERSON Process Management Manufacturing GmbH & Co. OHG
Instruction Sheet
HAS1xE-IS-BKS 10/2004
17.1.5 Light Barrier Signal
Measuring point: X9, pin 2 Measuring device: Oscilloscope Signal: square impulse, level see below
Frequency = 24 Hz (± 0,1 Hz)
Failure: No signal or incorrect measuring values
Possible reasons (IR measurement):
Expected signal level: U = 6 VSS (± 0,3 V)
BKS 20 Upgrade
a) Chopper not connected (see chapter 18.1). b) Chopper inoperative (switch analyzer off and then on again). c) Light barrier not connected (see chapter 18.1). d) Light barrier defect (replace chopper) e) Broken cable of light barrier or faulty light barrier (replace
chopper).
f) BKS faulty (replace BKS).
Possible reasons (oxygen measurement without IR channel):
Expected signal level: LOW < 0.45 V; HIGH > 2.4 V (TTL logic signal)
a) Solder bridge LB 18 not configured properly (pins 1-2 connected) b) µP not operating (switch analyzer off and on) c) BKS faulty (replace BKS).
EMERSON Process Management Manufacturing GmbH & Co. OHG
17-3
BKS 20 Upgrade
17.1.6 Analog Preamplifiers
a) Paramagnetic Oxygen measurement
Measuring point: X 25 channel 1 Measuring device: DVM Signal: At zero gas purge: 0 V dc (± 50 mV)
At ambient air (approx. 21 Vol. - % O2):
100 % O2 measuring cell: approx. 840 mV 25 % O2 measuring cell: approx. 3.36 V
Instruction Sheet
HAS1xE-IS-BKS
10/2004
Failure: No signal or incorrect measuring values. Possible reasons: a) oxygen sensor not connected (see chapter 18.).
a) IR measurement
Measuring point: X 25 channel 1 (not for analyzer with oxygen measurement)
Measuring device: DVM Signal: At zero gas purge: 0 V dc (± 100 mV)
Failure: No signal or incorrect measuring values.
(for identification of cell see nameplate label on cell)
b) oxygen sensor faulty (replace sensor). c) BKS faulty (replace BKS).
X 27 channel 2
There should be a minimum difference of 600 mV (measuring ranges < 1000 ppm difference 500 mV) between
zeropoint voltage and sensitivity voltage.
17-4
Possible reasons: a) Detector not connected (see chapter 18.).
b) Detector faulty (replace detector). c) BKS faulty (replace BKS).
EMERSON Process Management Manufacturing GmbH & Co. OHG
Instruction Sheet
HAS1xE-IS-BKS 10/2004
BKS 20 Upgrade
17.2 Measuring points of OXS (electrochemical oxygen measurement)
17.2.1 Sensor Signal
Measuring point: Tp 1 (Signal)
Tp 2 (
⊥⊥
⊥)
⊥⊥
Measuring device: DVM
Signal: At ambient air (approx. 21 Vol. - % O2): approx. 3.36 V (new cell)
Failure: No signal or faulty voltage
Possible reasons: a) Oxygen sensor not connected to PCB "OXS"
b) PCB “OXS” not connected / faulty c) Oxygen sensor faulty or weared (replace) d) BKS faulty (replace)
Note !
If the measured value is less than 2.8 V at gas flow with ambient air , the sensor is weared. Replace the sensor.
Tp 2
Tp 1
Fig. 17-2: PCB “OXS”, assembled, horizontal projection
EMERSON Process Management Manufacturing GmbH & Co. OHG
17-5
BKS 20 Upgrade
Instruction Sheet
HAS1xE-IS-BKS
10/2004
17-6
EMERSON Process Management Manufacturing GmbH & Co. OHG
Instruction Sheet
HAS1xE-IS-BKS 10/2004
BKS 20 Upgrade
18. Allocation of Connectors on Printed Circuit Boards
Be sure to observe the safety measures !
18.1 BKS
X19 (+) X20 (GND) 24 V DC power supply input
F2 Fuse for 24 V DC input
type: Wickmann 372 rating: T 4 A / 250 V
P 1 or P2 24 V DC distribution for
optional internal load (e.g. heater for paramagn. O2 sensor)
X 1 Front panel BXF (P8)
X 16 Digital Outputs, parallel
X 18 Analog Outputs
J 9 Option BSI 10:
Status signals and serial interface
Front panel
Fig. 18-1: PCB “BKS”, location of connectors
EMERSON Process Management Manufacturing GmbH & Co. OHG 18-1
BKS 20 Upgrade
18.1.1 IR measurement without Oxygen Channel
X 2 Chopper
X 3 (pins 1 & 2) IR source channel 1
X 4 (pins 1 & 2) IR source channel 2
X 5 IR detector (channel 1)
X 6 IR detector (channel 2)
Instruction Sheet
HAS1xE-IS-BKS
10/2004
X 7 Temperature sensor (chopper)
X 9 Light barrier (chopper)
18.1.2 Oxygen Measurement without IR channel
X 5 Oxygen sensor (channel 1)
[ electrochemical measurement: PCB OXS (cable P1, 5 pin connector)]
X 6 Oxygen sensor (channel 2)
[electrochemical measurement: PCB OXS (cable P1, 5 pin connector)]
X 7 Temperature sensor
[electrochemical measurement: PCB OXS (cable P1, 2 pin connector)]
EMERSON Process Management Manufacturing GmbH & Co. OHG18-2
Instruction Sheet
HAS1xE-IS-BKS 10/2004
18.1.3 IR / Oxygen Measurement Combination
X 2 Chopper
X 4 (pins 1 & 2) IR source channel 2
X 5 Oxygen sensor
X 6 IR detector (channel 2)
X 7 Temperature sensor (chopper)
BKS 20 Upgrade
X 9 Light barrier (chopper)
18.1.4 TC Measurement without IR channel
In preparation !
EMERSON Process Management Manufacturing GmbH & Co. OHG 18-3
BKS 20 Upgrade
18.1.5 IR / TC Measurement combined
X 2 Chopper
X 4 (pins 1 & 2) IR source channel 2
X 5 TC sensor (from PCB WAP 100, cable P15, 5 pin connector))
X 6 IR detector (channel 2)
Instruction Sheet
HAS1xE-IS-BKS
10/2004
X 7 Temperature sensor (chopper)
X 9 Light barrier (chopper)
18.1.6 Oxygen / TC Measurement combined
In preparation !
EMERSON Process Management Manufacturing GmbH & Co. OHG18-4
Instruction Sheet
HAS1xE-IS-BKS 10/2004
18.2 OXS (electrochemical oxygen measurement)
Strip P2 Oxygen sensor
Cable P1, 5 pin connector PCB BKS, X 5 (sensor signal)
Cable P1, 2 pin connector PCB BKS, X 7 (temperature sensor)
(not used in combination with IR measurement)
View A: PCB OXS assembled, horizontal projection
BKS 20 Upgrade
PCB “OXS”
Connection
View A
↓↓
↓↓
Oxygen sensor (Pin-base“P 2”)
Fig. 18-2: PCB “OXS”
Cable “P 1”
Sensor cell
EMERSON Process Management Manufacturing GmbH & Co. OHG 18-5
Instruction Sheet
BKS 20 Upgrade
18.3 WAP 100 (TC measurement)
P2 24 V DC supply from PCB BKS (P1 or P2)
T erminals P 18 to sensor
Cable P15, 5 - pin connector PCB BKS, X 5 (sensor signal)
P 10 PCB BSE 01 (analog input for cross compensation)
HAS1xE-IS-BKS
10/2004
P 16 sensor heater
P 10
orange
P 18
1
P 16
P 15
blue
green
1
yellow
Fig. 18-3: PCB “WAP 100”,
(principle drawing)
1
P 2
EMERSON Process Management Manufacturing GmbH & Co. OHG18-6
Instruction Sheet
HAS1xE-IS-BKS 10/2004
19. BKS Jumper Configuration
There is only one jumper on board BKS 20: J7, which activates battery buffering for the SRAM data. Opening J7 deletes all data changed by the operator when power supply is disconnected. The analyzer configuration in this case is reset to default values (stored into EPROM at factory startup).
BKS 20 Upgrade
J7 Buffer Battery
(normally closed)
Front panel
Fig. 19-1: PCB “BKS”
EMERSON Process Management Manufacturing GmbH & Co. OHG 19-1
BKS 20 Upgrade
Instruction Sheet
HAS1xE-IS-BKS
10/2004
EMERSON Process Management Manufacturing GmbH & Co. OHG19-2
Instruction Sheet
HAS1xE-IS-BKS 10/2004
BKS 20 Upgrade
A. ATTACHMENT
A.1 BINOS® 100 (M) / HYDROS® 100 / OXYNOS® 100 Block Diagram
X19 (+)
X20 (GND)
EMERSON Process Management Manufacturing GmbH & Co. OHG A-1
BKS 20 Upgrade
A.2 BINOS® 100 2M Block Diagram
Instruction Sheet
HAS1xE-IS-BKS
10/2004
X19 (+)
X20 (GND)
EMERSON Process Management Manufacturing GmbH & Co. OHGA-2
Instruction Sheet
HAS1xE-IS-BKS 10/2004
BKS 20 Upgrade
X19 (+)
X20 (GND)
EMERSON Process Management Manufacturing GmbH & Co. OHG A-3
BKS 20 Upgrade
A.3 BINOS® 100 2M for CAT Block Diagram
Instruction Sheet
HAS1xE-IS-BKS
10/2004
X19 (+)
X20 (GND)
EMERSON Process Management Manufacturing GmbH & Co. OHGA-4
Instruction Sheet
HAS1xE-IS-BKS 10/2004
A.4 CAT 100 Block Diagram
BKS 20 Upgrade
EMERSON Process Management Manufacturing GmbH & Co. OHG A-5
BKS 20 Upgrade
A.5 BINOS® 100 F Block Diagram
X19 (+)
X20 (GND)
Instruction Sheet
HAS1xE-IS-BKS
10/2004
EMERSON Process Management Manufacturing GmbH & Co. OHGA-6
BKS 20 Upgrade
EUROPE Emerson Process Management Manufacturing GmbH & Co. OHG
Industriestrasse 1 63594 Hasselroth Germany T +49 (6055) 884-0 F +49 (6055) 884-209 Internet: www.emersonprocess.de
EUROPE, MIDDLE EAST , AFRICA Emerson Process Management Shared Services Limited
Heath Place Bognor Regis West Sussex PO22 9SH England T +44-1243-863121 F +44-1243-845354 Internet: www.emersonprocess.co.uk
Instruction Sheet
HAS1xE-IS-BKS
10/2004
NORTH AMERICA Rosemount Analytical Inc. Process Analytic Division
1201 N. Main St. Orrville, OH 44667-0901 T +1 (330) 682-9010 F +1 (330) 684-4434 Internet: www.emersonprocess.com
LATIN AMERICA Emerson Process Management Ltda
Avenida Hollingsworth, 325 Iporanga-Sorocabe­SP 18087-000 Brazil T:+55 (152) 38-3788 F:+55 (152) 38-3300 Internet: www.emersonprocess.com.br
ASIA - PACIFIC Emerson Process Management Asia Pacific Pte Ltd
1 Pandan Crescent Singapore 128461 Tel +65 6777 8211 Fax +65 6777 0947 Internet: www.ap.emersonprocess.com
© Emerson Process Management Manufacturing GmbH & Co. OHG 2004
Loading...