Rosemount BINOS 100 Series Analyzers including OXYNOS 100, HYDROS 100-1st Ed. Manuals & Guides

Page 1
Rosemount Analytical
Operation Manual
Series 100
Microprocessor - Controlled Gas Analyzers
BINOS® 100
BINOS® 100 F
BINOS® 100 M
BINOS® 100 2M
OXYNOS® 100 HYDROS® 100
1. Edition 06/2000
ETC00265(1) Series 100 e 23.06.2000
Catalog - No.: ETC00265
Page 2
Rosemount Analytical
This Maintenance & Operation Manual includes information about the operation of the instrument as well as additional indications and notes regarding maintenance, troubleshooting and repair.
Troubleshooting, component replacement and internal adjustments must be made by qualified service personnel only.
Fisher-Rosemount GmbH & Co assumes no liability for any omissions or errors in this manual. Any liability for direct or consequential damages - which might occur in connection with the delivery or the use of this manual - is definitely excluded to the extend permitted by applicable law.
This instrument has left the works in good order according to safety regulations.
To maintain this operating condition, the user must strictly follow the instructions and consider the
warnings in this manual or provided on the instrument.
Misprints and alterations reserved
©
2000 by FISHER-ROSEMOUNT GmbH & Co. (PAD/ETC)
1. Edition: 06/2000
Read this operation manual carefully before attempting to operate the analyzer ! For expedient handling of reports of defects, please include the model and serial number which can be read on the instrument identity plate. Look for the error check list please too (see chapter 32. of this manual)
Fisher - Rosemount GmbH & Co.
European Technology Center
Industriestrasse 1
D - 63594 Hasselroth • Germany
Phone + 49 (6055) 884-0
Telefax + 49 (6055) 884-209
Internet: http://www.processanalytic.com
ETC00265(1) Series 100 e 23.06.2000
Page 3
Rosemount Analytical
I. Intended Use Statement S - 1 II. Safety Symbols S - 1 III. General S - 2 IV. Gases and Gas Conditioning (Sample Handling) S - 4 V. Supply Voltage S - 4 VI. Analyzer specific notes for the user S - 6
CONTENTS
Table of Contents
SAFETY SUMMARY S - 1
VII. BINOS® 100 F specific notes for use in hazardous areas S - 7 VII.a Z purge for CSA-C/US Ex Zone 2 Non-Flammable AtmospheresS - 8 VIII. Additional notes for service / maintenance S - 9 VIII.a Electrostatic Discharge S - 10 IX. Operating Conditions according to DMT Approval S - 11
PREFACE P - 1
General Overview P - 1
a) Software Versions P - 1 b) Housing Versions P - 2
Area Classification P - 3 a) General Purpose P - 3 b) Hazardous Areas P - 3 Ex Zone 2 P - 3 Ex Zone 1 P - 3
ETC00265(1) Series 100 e xx.05.2000
I
Page 4
CONTENTS
DESCRIPTION
1. Technical Description 1 - 1
1.1 Front View 1 - 1
1.2 Rear Panel 1 - 5
1.3 Internal Construction 1 - 11
1.3.1 Internal Gas Paths 1 - 25 a) Gas Path Material 1 - 25 b) Gas Path Layout (internal tubing) 1 - 26
2. Measuring Principle 2 - 1
2.1 IR Measurement 2 - 1
Rosemount Analytical
2.1.1 Interference Filter Correlation (IFC Principle) 2 - 1
2.1.2 Opto-Pneumatic Measuring Principle 2 - 3
2.1.3 Technique 2 - 5
2.2 Oxygen Measurement 2 - 6
2.2.1 Paramagnetic Measurement 2 - 6
2.2.2 Elektrochemical Measurement 2 - 8
2.3 Thermal Conductivity Measurement 2 - 10
2.3.1 Sensor Design 2 - 10
2.3.2 Analysis Cell 2 - 10
2.3.3 Measurement Method 2 - 11
3. Photometer Assembly 3 - 1
3.1 Photometer with Pyroelectrical Detector (Solid-state detector) 3 - 1
3.2 Photometer with Gas Detector 3 - 4
4. (vacant)
II
ETC00265(1) Series 100 e xx.05.2000
Page 5
Rosemount Analytical
5. Preparation of Start-up 5 - 1
5.1 Installation Site 5 - 2
5.2 Gas Conditioning (Sample Handling) 5 - 3
5.2.1 Fine Dust Filter (Option BINOS® 100 2M/F) 5 - 4
5.2.2 Gas Sampling Pump (Option BINOS® 100 2M/F) 5 - 4
5.2.3 Pressure Sensor (Option) 5 - 4
5.2.4 Gas Flow 5 - 4
5.3 Gas Connections 5 - 5
5.3.1 Standard 5 - 5
5.3.2 Internal Solenoid Valves (Option BINOS® 100 2M/F) 5 - 8
CONTENTS
OPERATION
5.3.3 Purge gas connection of BINOS® 100 F for Ex zones 5 - 10 a) Continuous Purge (for CENELEC Ex Zone 1 Applications) 5 - 10 b) Z purge for US Ex Zone 2 Non-Flammable Atmospheres 5 - 11
5.4 Additional Hints to BINOS® 100 F (Field Housing) 5 - 12
5.4.1 Wall Mounting 5 - 13
5.4.2 Electrical Connections 5 - 14
6. Switching On 6 - 1
6.1 General 6 - 1
6.2 24 V DC Supply 6 - 2
6.3 230/120 V AC Supply 6 - 4
6.3.1 BINOS® 100 2M 6 - 4
6.3.2 BINOS® 100 F 6 - 5
7. Key Functions 7 - 1
7.1 FUNCTION 7 - 2
7.2 ENTER 7 - 3
7.3 INPUT - CONTROL 7 - 5
7.4 PUMP (BINOS® 100 2M/F only) 7 - 6
ETC00265(1) Series 100 e xx.05.2000
III
Page 6
CONTENTS
Rosemount Analytical
8. Entry of System Parameters 8 - 1
8.1 Pressure Correction 8 - 2
8.2 Cross Compensation (internal) 8 - 2
8.3 Cross Compensation Calibration (internal) 8 - 3
8.4 Hold 8 - 4
8.5 Automatic Calibration 8 - 4
8.6 Tolerance Check 8 - 5
8.7 Display Off 8 - 6
8.8 Analog Signal Outputs 8 - 7
8.9 Flushing Period 8 - 8
8.10 User Code 8 - 8
8.11 Response Time (t90)8 - 9
8.12 Offset (Begin of range) 8 - 10
8.13 End of Range Value 8 - 11
8.14 Reset 8 - 12
8.15 Program Version 8 - 13
8.16 Serial - No. 8 - 13
8.17 Pump 8 - 14
8.18 Pump Control 8 - 14
9. Calibration 9 - 1
9.1 Manual Calibration 9 - 2
9.1.1 Zeroing 9 - 2
9.1.2 Spanning 9 - 4
9.2 Time-Controlled Calibration Mode (Option) 9 - 7
9.2.1 Zeroing 9 - 7
9.2.2 Combined Zeroing and Spanning 9 - 9
9.3 Remote-Controlled Calibration Mode (Option) 9 - 10
10. Measurement / Switching Off 10 - 1
10.1 Measurement 10 - 1
10.2 Switching Off 10 - 2
11. Digital Outputs 11 - 1
11.1 Concentration Limits 11 - 2
11.2 Valve Control 11 - 4
11.3 Status Signals (Option non-voltage-carrying relay contacts) 11 - 4
IV
ETC00265(1) Series 100 e xx.05.2000
Page 7
Rosemount Analytical
12. Serial Interface (Option) 12 - 1
12.1 Retrofitting of Serial Interface / Status Signals 12 - 1
12.2 General 12 - 2
12.3 Start Up 12 - 4
12.3.1 RS 232 C 12 - 4
12.3.2 RS 485 12 - 5
12.3.3 Switching ON/OFF Interface Operation 12 - 6
12.3.4 Setting Interface Parameters 12 - 6
12.4 Telegram Syntax 12 - 8
12.4.1 Start Character ( “$” = Hex 24) 12 - 8
CONTENTS
OPTIONS
12.4.2 Terminate Character ( “CR” = Hex OD) 12 - 8
12.4.3 Instruction Code 12 - 8
12.4.4 Hyphen Character ( “;” = Hex 3B) 12 - 8
12.4.5 Status Telegram 12 - 9
12.4.6 Numerical Representations 12 - 10
12.4.7 Block Parity Check 12 - 10
12.5 Instruction Syntax 12 - 11
12.5.1 Instruction Listing 12 - 12
12.5.2 Response Telegrams 12 - 13
13. Digital Inputs/FOUNDATION Fieldbus (BINOS
®
100 2M/F only) 13 - 1
13.1 Digital Inputs 13 - 1
13.1.1 General 13 - 1
13.1.2 Start of Calibration 13 - 1
13.1.3 Valve Control 13 - 2
13.1.4 Pump Control 13 - 2
13.2 FOUNDATION™ Fieldbus 13 - 2
14. Cross Compensation/Setting of Respsonse time (TC only) 14 - 1
14.1 Cross Compensation 14 - 1
14.1.1 Previous Adjustment 14 - 2
14.1.2 Adjustment Procedure 14 - 3
15. (vacant)
ETC00265(1) Series 100 e xx.05.2000
V
Page 8
CONTENTS
TROUBLESHOOTING
16. Error List 16 - 1
17. Measuring Points of BKS and OXS 17 - 1
17.1 Measuring points of BKS 17 - 1
17.1.1 Supply Voltage + 6 V 17 - 1
17.1.2 Reference Voltage positive 17 - 1
17.1.3 Reference Voltage negative 17 - 2
17.1.4 Motor Drive (for IR channel only) 17 - 2
17.1.5 Temperature Sensor 17 - 3
17.1.6 Light Barrier Signal 17 - 4
Rosemount Analytical
17.1.7 Analog Preamplifiering 17 - 5
17.2 Measuring points of OXS (EO2 measurement) 17 - 6
17.2.1 Sensor Signal 17 - 6
18. Plug Pin Allocation of Printed Circuit Boards 18 - 1
18.1 Plug Pin Allocation of BKS 18 - 1
18.1.1 IR measurement without oxygen channel 18 - 2
18.1.2 Oxygen Measurement without IR channel 18 - 2
18.1.3 IR / Oxygen Measurement combined 18 - 3
18.1.4 TC Measurement without IR channel 18 - 3
18.1.5 IR / TC Measurement combined 18 - 4
18.1.6 Oxygen / TC Measurement combined 18 - 4
18.2 Plug Pin Allocation OXS (EO2 measurement only) 18 - 5
18.3 Plug Pin Allocation WAP 100 (TC measurement only) 18 - 6
19. Jumper Allocation of BKS 19 - 1
20. (vacant)
VI
ETC00265(1) Series 100 e xx.05.2000
Page 9
Rosemount Analytical
21. Fine Dust View Filter (Option) 21 - 1
22. Leak Testing 22 - 1
23. Housing 23 - 1
23.1 Cleaning of Housing Outside 23 - 1
23.2 Opening the Housing 23 - 2
23.2.1 1/4 19" Housing 23 - 2
23.2.2 BINOS® 100 2M 23 - 3
CONTENTS
MAINTENANCE
a) Housing Cover 23 - 3 b) Front Panel 23 - 4
23.2.3 BINOS® 100 F (Field Housing) 23 - 5
24. Replacement and Cleaning of Photometric Components 24 - 1
24.1 Removal of the Photometer Assembly 24 - 1
24.2 Light Source Replacement 24 - 2
24.3 Cleaning of Analysis Cells and Windows 24 - 3
24.3.1 Removal of Analysis Cells 24 - 3
24.3.2 Cleaning 24 - 4
24.3.3 Reinstalling of Analysis Cells 24 - 5
24.4 Chopper Replacement 24 - 6
24.5 Reinstalling of the Photometer Assembly 24 - 6
24.6 Physical Zeroing 24 - 7
24.6.1 Standard Photometer (not sealed version) 24 - 7
24.6.2 Sealed Photometer (Option) 24 - 8
ETC00265(1) Series 100 e xx.05.2000
VII
Page 10
CONTENTS
25. Check / Replacement of electrochemical Oxygen Sensor 25 - 1
25.1 Check of the Sensor 25 - 2
25.2 Replacement of the Sensor 25 - 3
25.2.1 Removal of the Sensor 25 - 3 a) Oxygen Measurement without IR channel 25 - 3 b) IR / Oxygen Measurement combined 25 - 5
25.2.2 Exchange of the Sensor 25 - 6
25.2.3 Reinstalling of the Sensor 25 - 6 a) Oxygen Measurement without IR channel 25 - 6 b) IR / Oxygen Measurement combined 25 - 6
25.2.4 Basic conditions for the Oxygen Sensor 25 - 7
Rosemount Analytical
26. (vacant)
27. TECHNICAL DATA 27 - 1
27.1 Options 27 - 1
27.2 Housing 27 - 2
27.3 Signal Inputs / Outputs, Interfaces 27 - 3
27.4 General Specifications 27 - 7
27.5 Voltage Supply 27 - 9
27.5.1 Electrical Safety 27 - 9
27.5.2 Power Supplys [UPS 01 T (Universal Power Supply) / SL10 / SL5] 27 - 10
VIII
ETC00265(1) Series 100 e xx.05.2000
Page 11
Rosemount Analytical
28. Replacing the EPROM 28 - 1
29. Pin Assignments 29 - 1
29.1 24 V DC Input 29 - 1
29.2 24 V DC Output 29 - 1
29.3 230/120 V AC Input 29 - 2
29.4 Analog Signal Outputs 29 - 2
29.5 Digital Signal Outputs 29 - 3
29.5.1 Standard Signal Outputs 29 - 3
29.5.2 Status Signals (Relay Outputs, Option) 29 - 3
CONTENTS
SUPPLEMENT
29.6 Serial Interface (Option) 29 - 4
29.7 Analog Signal Inputs (TC Option Cross Compensation) 29 - 4
29.8 Digital Signal Inputs (Option) 29 - 5
29.9 FOUNDATION™ Fieldbus (Option) 29 - 5
29.10 Wiring of Digital Inputs and Outputs to ensure electromagnetic immunity 29 - 6
29.10.1 Electrical connections (general) 29 - 6
29.10.2 Special Tips 29 - 8
30. Connection Cables 30 - 1
30.1 24 V DC 30 - 1
30.2 230/120 V AC Input (BINOS® 100 2M) 30 - 1
30.3 Sub D Sockets, 9 pin 30 - 2
30.4 Sub D Plugs, 9 pin 30 - 2
31. (vacant)
32. Failure Check List 32 - 1
INDEX I - 1
List of figures and tables I - 11
ETC00265(1) Series 100 e xx.05.2000
IX
Page 12
CONTENTS
Rosemount Analytical
X
ETC00265(1) Series 100 e xx.05.2000
Page 13
Rosemount Analytical
SAFETY SUMMARY
Safety Summary
I. Intended Use Statement
The series 100 instruments are intended for use an industrial measurement device only. It is not intended for use in medical, diagnostic, or life support applications, and no independent agency certifications or approvals are to be implied as covering such applications.
II. Safety Symbols
GENERAL
Outside and/or inside the 100 series instruments or at operation manual resp. different symbols give you a hint to special sources of danger.
Source of danger !
See Operation Manual!
Electrostatic Discharge (ESD) !
Explosives !
Hot components !
Toxic !
Risk to health !
Analyzer specific notes for the user !
For additional information to these safety symbols see operation manual. Strictly follow these instructions !
ETC00265(1) Series 100 e 23.06.2000
S - 1
Page 14
SAFETY SUMMARY
GENERAL
Rosemount Analytical
III. General
To avoid explosion, loss of life, personal injury and damages to this equipment and other property, all personnel authorized to install, operate and service this equipment should be thoroughly familiar with and strictly follow the instructions in this manual ! Save these instructions !
If this equipment is used in a manner not specified in these instructions, protective features may be impaired !
Correct and safe operation of analyzers calls for appropriate transportation and storage, expert installation and commissioning as well as correct operation and meticulous maintenance !
Fisher-Rosemount GmbH & Co. does not take responsibility (liability) for the customer´s failure to comply with these requirements !
Do not attempt internal service or adjustment unless other person, capable of rendering first aid and resuscitation, is present !
Because of the danger of introducing additional hazards, do not perform any unauthorized modification to the instrument ! Return the instrument to a Fisher-Rosemount Sales and Service office for service or repair to ensure that safety features are maintained !
Instruments which appear damaged or defective should be made inoperative and secured against unintended operation until they can be repaired by qualified service personnel.
S - 2
ETC00265(1) Series 100 e 23.06.2000
Page 15
Rosemount Analytical
SAFETY SUMMARY
GENERAL / GASES AND GAS CONDITIONING (SAMPLE HANDLING)
Do not open instrument when energized ! Component replacement and internal adjustments requires servicing by qualified personnel only !
Read this operation manual before attempting to operate with the instrument ! Be sure to observe the additional notes, safety precautions and warnings given in the operation manual !
Operate analyzer as table-top version or as rack-mountable version (built-in) only (except of BINOS® 100 F: designed for wall mounting only) !
Do not operate the instrument in the presence of flammable gases or explosive atmosphere without supplementary protective measures !
Hot components may exist at the photometer or in heated versions !
Lift or carry housing with at least 2 persons because of the high weight of field housing BINOS® 100 F (approx. 30 - 35 kg). For easy transport use a suitable cart.
Verify to BINOS® 100 F, that the PG fittings together with pass through cables are hermetic to be in agreement with protection class IP 65 (according to DIN stan­dard 40050). The permissible outside diameters of the cables are 7 to 12 mm !
ETC00265(1) Series 100 e 23.06.2000
S - 3
Page 16
SAFETY SUMMARY
GASES AND GAS CONDITIONING (SAMPLE HANDLING) / SUPPLY VOLTAGE
IV. Gases and Gas Conditioning (Sample Handling)
Be sure to observe the safety regulations for the respective gases (sample gas and test gases / span gases) and the gas bottles !
Inflammable or explosive gas mixtures must not be purged into the instrument without supplementary protective measures !
To avoid a danger to the operators by explosive, toxic or unhealthy gas components, first purge the gas lines with ambient air or nitrogen (N2) before cleaning or exchange parts of the gas paths.
Rosemount Analytical
V. Supply Voltage
The socket outlet shall be installed near the equipment and shall be easily accessible to disconnect the device from the socket outlet.
Verify whether the line voltage stated on the instrument ore power supply is in accordance with that of your mains line!
Be sure to observe the safety precautions and warnings given by manufacturer of power supply !
BINOS® 100(M), BINOS® 100 2M (external PS), HYDROS® 100 and OXYNOS® 100 are Safety Class III instruments.
Verify correct polarity for 24 V DC operation !
S - 4
Use only power supply VSE 2000, UPS 01 T, DP 157, SL5, SL10 (DP 157 and SL for rack mounting only) or equivalent power supplys to be in agreement with the CE conformity.
If using equivalent power supplys, they must have SELV output voltage !
For supply of more than one analyzer with one power supply, a fuse (T3.15A) is to be connect in series for each analyzer !
ETC00265(1) Series 100 e 23.06.2000
Page 17
Rosemount Analytical
BINOS® 100 2M (internal PS) and BINOS® 100 F are Safety Class 1 instruments
The analyzer is provided with a protective earth terminal. To prevent shock hazard, the instrument chassis and cabinet must be connected to an electrical ground. The instrument must be connected to the AC power supply mains through a three-conductor power cable, with the third wire firmly connected to an electrical ground (safety ground) at the power outlet. If the instrument is to be energized via an external power supply, that goes for the power supply too. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal injury. Deliberate disconnection is inadmissible / prohibited !
SAFETY SUMMARY
SUPPLY VOLTAGE
The analyzer BINOS® 100 F (field housing) has no switch with disconnect function. The customer has to provide a switch or circuit breaker into his installation. This switch has to be installed near by analyzer, must be easily attainable for operator and has to be characterized as disconnector for analyzer.
Cables to external data processing have to be double-insulated against mains voltage for analyzer BINOS® 100 F ! Use cables suitable for intrinsic safe applications only ! Install internal data lines that they have a distance to mains voltage lines of at least 5 mm. This distance has to be valid permanently (e.g. via cable holder) !
24 VDC supply to external components/analyzers with the internal power supply of BINOS® 100 2M requires a fuse to be connected in series to the consumer which limits the current consumption to max. 2 A !
Verify correct polarity for 24 V DC supply of external components !
ETC00265(1) Series 100 e 23.06.2000
S - 5
Page 18
SAFETY SUMMARY
ANALYZER SPECIFIC NOTES FOR USER
VI. Analyzer specific notes for the user
The installation site for the instrument has to be dry and remain above freezing point at all times.
The instrument must be exposed neither to direct sunlight nor to strong sources of heat. Be sure to observe the permissible ambient temperature ! For outdoor sites, we recommend to install the instrument in a protective cabinet. At least, the instrument has to be protected against rain (e.g., shelter).
Do not interchange gas inlets and gas outlets ! All gases have to be supplied to the analyzer as conditionned gases ! If corrosive gases are inserted into the instrument, it has is to be verified that there
Rosemount Analytical
are no gas components which may damage the gas path components.
Ensure that all gas connections are made as labeled and are leak free ! Improper gas connections could result in explosion and death ! The unit´s exhaust may contain hydrocarbons and other toxic gases such as carbon monoxide ! Carbon monoxide is highly toxic !
Permissible gas pressure max. 1,500 hPa !
The exhaust gas lines have to be mounted in a declining, descending, pressureless and frost-free and according to the valid emission legislation !
In case it is necessary to open the gas paths, close the analyzers gas connections with PVC caps immediatly !
Pressure of sample gas / test gases max. 1,500 hPa !
BINOS® 100 F lift points are labeled ! Labels showing down side for transport ! Do not use electronics of optional purge system” as handle !
S - 6
Use only optional delivered cables from our factory or equivalent shielded cables to be in agreement with the CE conformity. The customer has to prove that the shield is connected correctly (chapter 29.10). Shield and connectors housing have to be connected conductive. Sub. min. D plugs/sockets have to be screwed to the analyzer.
The analyzer (excepting BINOS® 100 F) is not in agreement with the CE conformity if optional terminal strip adapters are used In this case CE conformity must be declared by customer as “manufacturer of system”.
ETC00265(1) Series 100 e 23.06.2000
Page 19
Rosemount Analytical
BINOS® 100 F SPECIFIC NOTES FOR USE IN HAZARDOUS AREAS (EX ZONES)
VII. BINOS® 100 F specific notes for use in hazardous areas (Ex zones)
Be sure to observe the additional notes, safety precautions and warnings given in the individual manuals (simplified pressurization for CENELEC ex zone 2 /
EExp approved “purge system” for CENELEC ex zone 1) !
Wall mounting: Ex zone 1 and Ex zone 2 applications require additional space for safety related components (see Fig. 27-6, 27-7 and 27-8).
Pressure inside the housing must not exceed 5 hPa at normal operation or 10 hPa for a short time of less than 1/2 hour resp. !
SAFETY SUMMARY
Activation of pressurization bypass switch must be performed by a authorized person (for hazardous areas) only. This has to be in accordance with respective legislation only !
Be sure that there is no presence of flammable gases or explosive atmosphere before opening the housing !
Do not operate with doors open. Always disconnect power, discharge circuits and remove external voltage sources before troubleshooting, repair or replacement of any component ! After disconnecting power and and removal of external voltage wait at least 5 minutes before opening the housing !
Cleaning of BINOS® 100 F front panel for Ex Zones: Danger of electrostatic discharge ! Use damp cloth only for cleaning front panel !
ETC00265(1) Series 100 e 23.06.2000
S - 7
Page 20
SAFETY SUMMARY
Z PURGE FOR CSA-C/US EX ZONE 2 NON-FLAMMABLE ATMOSPHERES
VII.a Z purge for CSA-C/US Ex Zone 2 Non-Flammable Atmospheres
This enclosure shall not be opened unless the area is known to be free of flammable materials or unless all devices within have been de-energized !
Upon start-up or after loss of continuous dilution requiring switching off the electrical supply, purge for 22 minutes with flow rate approx. 23,5 scfh (13 l/min., see chapter 5.3.3) unless the internal atmosphere is known to be well below the lower explosive limit (LEL) !
This analyzer is not designed for analysis of flammable sample ! Introduction of flammable samples into this equipment could result in explosion,
Rosemount Analytical
causing severe personal injury, death or property damage !
Consult factory if flammable samples are to be measured !
Do not open while energized unless it is known that no explosive atmosphere is present !
S - 8
ETC00265(1) Series 100 e 23.06.2000
Page 21
Rosemount Analytical
VIII. Additional notes for service / maintenance
Do not open instrument when energized ! Component replacement and internal adjustments requires servicing by qualified personnel only !
Always disconnect power, discharge circuits and remove external voltage sources before troubleshooting, repair or replacement of any component !
Any work inside the instrument without switching off the power must be performed by a specialist who is familiar with the related danger, only !
SAFETY SUMMARY
ADDITIONAL NOTES FOR SERVICE / MAINTENANCE
To avoid a danger to the operators by explosive, toxic or unhealthy gas components, first purge the gas lines with ambient air or nitrogen (N2) before the gas paths are cleaned or parts are exchanged.
Do not open BINOS® 100 F for use in hazardous areas (Ex zones) while energized unless it is known that no explosive atmosphere is present !
Hot components may exist at the photometer or in heated versions !
In case of replacing fuses the customer has to be certain that fuses of specified type and rated current are used. It is prohibited to use repaired fuses or defective fuse holders or to short-circuit fuse carriers (fire hazard).
Cleaning of BINOS® 100 F front panel for Ex Zone 1: Danger of electrostatic discharge ! Use damp cloth only for cleaning front panel !
ETC00265(1) Series 100 e 23.06.2000
S - 9
Page 22
SAFETY SUMMARY
ELECTROSTATIC DISCHARGE
Rosemount Analytical
VIII.a Electrostatic Discharge
The electronic parts of the analyzer can be irreparably damaged if exposed to electrostatic discharge (ESD).
The instrument is ESD protected when the covers have been secured and safety precautions observed. When the housing is open, the internal components are not ESD protected anymore.
Although the electronic parts are reasonable safe to handle, you should be aware of the following considerations:
Best ESD example is when you walked across a carpet and then touched an electrical grounded metal doorknob. The tiny spark which has jumped is the result of electrostatic discharge (ESD).
You prevent ESD by doing the following:
Remove the charge from your body before opening the housing and maintain during work with opened housing, that no electrostatic charge can be built up.
Ideally you are opening the housing and working at an ESD protecting workstation. Here you can wear a wrist trap.
However, if you do not have such a workstation, be sure to do the following procedure exactly:
Discharge the electric charge from your body. Do this by touching a device that is grounded electrically (any device that has a three - prong plug is grounded electrically when it is plugged into a power receptacle). This should be done several times during the operation with opened housing (especially after leaving the service site because the movement on a low conducting floors or in the air might cause additional ESDs).
S - 10
ETC00265(1) Series 100 e 23.06.2000
Page 23
Rosemount Analytical
OPERATING CONDITIONS ACCORDING TO DMT APPROVAL
SAFETY SUMMARY
IX. Operating Conditions according to DMT Approval
(Chapter 6 of the supplement I to the DMT reports IBS/PFG-No. 41300392 NIII andIBS/PFG-No. 41300292 NIII” about the performance test of the stationary gas analyzers
BINOS® 100 (M/2M) and OXYNOS® 100.
According to the system version and measuring results included in this report, the stationary gas analyzer BINOS® 100 2M from Fisher-Rosemount GmbH & Co. is suitable for measuring the concentrations of methane between 0 and 80 % CH4, of carbon dioxide between 0 and 80 % CO2, of carbon monoxid between 0 - 200 ppm CO and 0 - 10 Vol.% CO and of oxygen between 0 - 10 Vol.-%, if the features and system version go conform with the details contained in the enclosed documents as stated in this report, if the analysis system is operated accordingly and if the following requirements are met:
When using the gas warning system, it must be ensured that the permissible variations will
not be exceeded, taking into account the systematic failures of the measuring signals (as indicated in this report) and the local operating conditions. Consider the Code of Pratice No. T032 of the Labor Association of the Chemical Industry "Usage of stationary gas warning systems for explosion protection".
Verify that the explosion protection requirements are met when using the gas warning
system.
Depending on the situation, it must be verified that the preset values are low enough to allow
the system to activate the necessary protection and emergency measures and, thus, to prevent any critical situations in a minimum period of time.
When at system installation, a release of one or both measuring components in the ambient
air might occur, its influence on the measuring result should be proved. A sealed cell or an external housing purging with sample-free air of measuring gases can be used, if required.
The operatability of the alarms and the displays of each system should be tested with clean
air and test gas after the initial operation, after each long-time interruption, and periodically. The tightness of gas pathes should also be tested. The tests must be documented by keeping accounts.
ETC00265(1) Series 100 e 23.06.2000
S - 11
Page 24
SAFETY SUMMARY
OPERATING CONDITIONS ACCORDING TO DMT APPROVAL
Rosemount Analytical
The intervals for the periodical tests must be settled by the person being responsible for the system´s security and in accordance with the Code of Pratice No. T023 of the Labor Association of the Chemical Industry "Maintenance of stationary gas warning systems for explosion protection".
Consider the superproportional dependency of the barometric pressure on the measured value for CO2.
The system control with serial interfaces described in this operation manual have not been subject to this investigation.
Sample gas condensation in analyzer (components) must be prevented by taking the necessary steps.
When the system is used with aggressive gases, it is to be verified that there are no gas components which might damage the gas path components.
Appropriate dust filters must precede the used systems.
The pressure and flow values recommended by the manufacturer should be observed. An
external monitoring of the sample gas flow through the analyzer should be provided.
The results of this investigation are based on the systems using software versions “3.03, “4.00”, “4.01” and “4.11”. A change of the software version used must be certified by the
Testing Association.
It should be ensured that the system parameters for the analog output have been correctly adjusted. End of range of low concentration should not be identical or lower than the begin of range. Disregarding these versions, the measurement range should be adjusted between 0 to 80 % CH4, 0 to 80 % CO2, 0 to 10 % CO or 0 to 10 % O2 resp. when the systems are used for explosion protection.
Read and follow the operation and maintenance manual supplied to and certified by PFG. It is important that the temperature is kept between 5 and 45 °C.
S - 12
ETC00265(1) Series 100 e 23.06.2000
Page 25
Rosemount Analytical
OPERATING CONDITIONS ACCORDING TO DMT APPROVAL
SAFETY SUMMARY
The analyzer housings must be provided with a permanent type plate indicating the name
of the manufacturer, model number, serial number, and the following reference and date of testing:
"IBS/PFG-Nr. 41300392" (for CH4, CO2 or CO)
"IBS/PFG-Nr. 41300292" (for O2)
Other designation requirements, such as these according to ElexV, are still valid. With this type plate, the manufacturer conformes that the features and technical data of the delivered system are identical with those described in this report. Any system which is not provided with such a type plate does not go conform with this report.
The chapter 6 of this report must be included in the operation and maintenance manual.
The manufacturer has to supply the customer with a copy of this report, if required.
A print of the report in an abridged version requires the agreement of PFG.
The results included in this report may not be altered in publications produced by the
manufacturer.
ETC00265(1) Series 100 e 23.06.2000
S - 13
Page 26
SAFETY SUMMARY
Rosemount Analytical
S - 14
ETC00265(1) Series 100 e 23.06.2000
Page 27
Rosemount Analytical
PREFACE
General Overview
The series 100 of analyzers offers multi-component, multi-method analysis. Different measure­ment methods can be combined in one analyzer. The following measuring methods of the individual measuring channels are possible:
IR = non-dispersive infrared measurement PO2= paramagnetic oxygen measurement EO2= electrochemical oxygen measurement TC = thermal conductivity measurement
PREFACE
All analyzers are designed to measure 1 or 2 gas components except of HYDROS® 100 and OXYNOS® 100 in case of PO2 measurement (both 1 channel only).
a) Software Versions
Different software versions and analyzer options are delivered:
BINOS® 100 (M), OXYNOS® 100, HYDROS® 100 (1/4 19" housing, external power supply):
Version 4.11 with optional RS 232/485 Interface (according to DMT Approval) Version 5.0 with optional RS 232/485 Interface
BINOS® 100 2M (1/2 19" housing, internal power supply):
Version 4.11 with optional RS 232/485 Interface (according to DMT Approval) Version 5.0 with optional RS 232/485 Interface
*)
BINOS® 100 2M (1/2 19" housing, external power supply):
Version 4.11 with optional RS 232/485 Interface (according to DMT Approval) Version 5.0 with optional RS 232/485 Interface*) and/or with optional 7 digital inputs Version 5.01 with optional 7 digital inputs and FOUNDATION™ Fieldbus
**)
BINOS® 100 F (field housing, internal power supply):
Version 4.11 with optional RS 232/485 Interface (according to DMT Approval) Version 5.0 with optional RS 232/485 Interface*) and/or with optional 7 digital inputs Version 5.01 with optional 7 digital inputs and optional FOUNDATION™ Fieldbus
*)
not in combination with FOUNDATION™ fieldbus
***)
not in combination with RS 232/485 interface
ETC00265(1) Series 100 e 23.06.2000
**)
P - 1
Page 28
PREFACE
Rosemount Analytical
b) Housing Versions
Different housing versions are delivered (for detailed informations see price list):
BINOS® 100 = 1/4 19" housing, ext. PS one or two IR channel BINOS® 100 M = 1/4 19" housing, ext. PS, one IR channel and one EO2 channel BINOS® 100 2M = 1/2 19" housing, one or two IR channel(s) or
internal or external PS, one IR channel and one EO2 channel or
one IR channel and one PO2 channel or one IR channel and one TC channel or one PO2 channel and one TC channel or one EO2 channel and one TC channel or one ot two PO2 channels one ot two EO2 channels one or two TC channels
with standard options: one internal sample gas pump
one internal solenoid valve block one integrated fine dust filter one integrated flow indicator
BINOS® 100 F = field housing, int. PS, measuring channels see BINOS® 100 2M
with standard options: see BINOS® 100 2M (dust filter for GP only) HYDROS® 100 = 1/4 19" housing, ext. PS, one TC channel OXYNOS® 100 = 1/4 19" housing, ext. PS, one or two EO2 channel or
one PO2 channel
P - 2
ETC00265(1) Series 100 e 23.06.2000
Page 29
Rosemount Analytical
Area Classification
a) General Purpose
All analyzer components are installed into a 1/4 19" housing (BINOS® 100 (M), OXYNOS® 100, HYDROS® 100) or a 1/2 19" enclosure (BINOS® 100 2M), 3 height units. These housings are go conform to DIN-standard protection class IP 20. The housings are available as rack-mountable or as table-top versions. The table-top housings are fitted with an additional carrying handle and additional rubber feets.
Additionally we can deliver a field housing version (BINOS® 100 F). All componets are installed
PREFACE
into a protection housing conforming to DIN-standard protection class IP 65 (approx. NEMA 4/4X). This enclosure is designed for wall mounting.
b) Hazardous Areas
For installation in hazardous areas the BINOS® 100 F is equipped with an impact tested front panel (according to CENELEC, EN 50014) with touch screen keypad. Optionally we can provide additional intrinsically safe I/O's and/or ex interface relays (couplers). Additionally we can built in a solvent resistant PO2 cell or an intrinsic safe PO2 or TC cell for potentially explosive atmosphere (consult factory).
Ex Zone 2
The BINOS® 100 F is additionally equipped with
- Simplified purge for CENELEC (EN 50016) Ex Zone 2 Applications
- Z purge for CSA-C/US Ex Zone 2 Non-Flammable Atmospheres
Ex Zone 1
The BINOS® 100 F is additionally equipped with
- Continuous Purge for CENELEC (EN 50016) Ex Zone 1 Applications
- Leakage Compensation for CENELEC (EN 50016) Ex Zone 1 Applications
ETC00265(1) Series 100 e 23.06.2000
P - 3
Page 30
PREFACE
Rosemount Analytical
P - 4
ETC00265(1) Series 100 e 23.06.2000
Page 31
Rosemount Analytical
TECHNICAL DESCRIPTION
1. Technical Description
The different analyzers are assembled in similar All analyzer components are installed into a 1/4 19" housing (BINOS® 100 (M), OXYNOS® 100, HYDROS® 100) or a 1/2 19" enclosure (BINOS® 100 2M), 3 height units tall. These housings are go conform to DIN-standard protection class IP 20. The housings are available as rack-mountable or as table-top versions. The table-top housings are fitted with an additional carrying handle and additional rubber feets. Additional we can deliver a field housing version (BINOS® 100 F). All componets are installed into a protection housing going conform to DIN-standard protection class IP 65 (approx. NEMA 4/4X).
FRONT VIEW
This enclosure is designed for wall mounting. For installation in hazardous areas the BINOS® 100 F is equipped with an impact tested front panel (according to CENELEC, EN 50014) with touch screen keypad. Optionally we can provide additional intrinsically safe I/O's and/or ex interface relays (couplers). For European ex zone 2 a simplified pressurization is installed and individual approval is provided. An EExp approved “purge system for European ex zone 1 (both according to CENELEC, EN 50016) or with Z Purge for measurement of non-flammable gases in hazardous areas (according to CSA-C/US for North American Ex zone 2) is another option.
1.1 Front View
The front panel includes the LED displays for both analysis channels and all of the analyzer operating controls. Additional the BINOS® 100 2M and BINOS® 100 F front panel is equipped with the function LED for the options “Solenoid Valves” and Gas Sampling Pump” and the key “PUMP” *) to switch on and switch off the gas sampling pump.
For general purpose applications with BINOS® 100 2M and BINOS® 100 F an optional fine dust filter with integrated needle valve and a flow meter can be built-in the front panel too. These options are not available if BINOS® 100 F is use in hazardous areas (ex zones).
ETC00265(1) Series 100 e 23.06.2000
1 - 1
Page 32
TECHNICAL DESCRIPTION
FRONT VIEW
1
Rosemount Analytical
11
12
250 ppm
CO
% O para. oxygen sensor
% O
2
2
FUNCTION ENTER INPUT - CONTROL
chem.
% O chem. oxygen sensor
7
paramagnetic
2
electrochemical
2
4
65
1 - 2
Fig. 1-1: BINOS® 100 (M), OXYNOS® 100, HYDROS® 100 Front view
1 LED display (channel 1) 2 LED display (channel 2) 4 Input setting control key DOWN 5 Input setting control key UP 6 Key ENTER 7 Key FUNCTION 11 Fastening screws for the carrying-strap bracket
or rack-mounting purposes
12 Housing cover fastening screw
ETC00265(1) Series 100 e 23.06.2000
Page 33
Rosemount Analytical
TECHNICAL DESCRIPTION
FRONT VIEW
Needle valve (option)
12
% O
2
1000 ppm
45678391011
Fig. 1-2: BINOS® 100 2M (standard version), front view
1 LED display (channel 1) 2 LED display (channel 2) 3 Function LED for options "Solenoid Valves / Gas Sampling Pump" 4 Input setting control key DOWN 5 Input setting control key UP 6 ENTER key 7 FUNCTION key 8 Key for option Gas Sampling Pump 9 Flow indicator (option) 10 Fine dust view filter with needle valve (option) 11 Fastening screws for the carrying strap bracket
or rack-mounting purposes
ETC00265(1) Series 100 e 23.06.2000
1 - 3
Page 34
TECHNICAL DESCRIPTION
FRONT VIEW
Wall mounting
holder
Fastener
front panel
Operation front panel
Rosemount Analytical
Fig. 1-2 for
OpenOpenOpen
OpenOpenOpe n
General Purpose
Fig. 1-4 for
hazardous areas
(ex zones)
Fig. 1-3: BINOS® 100 F, front view
3
1
SPAN SPAN ZERO
PUMP
2
SAMPLE
SWITCH CTRL
4
Front panel
1 - 4
FUNCTION ENTER INPU T-CONTROL
Fig. 1-4: BINOS® 100 F operation front panel for use in hazardous areas, front view
(impact tested with touch screen keypad)
ETC00265(1) Series 100 e 23.06.2000
Page 35
Rosemount Analytical
TECHNICAL DESCRIPTION
1.2 Rear Panel
The rear panels includes
the gas line fittings
the plug for the electrical supply input
the subminiature D mating socket for the analog signal outputs
the subminiature D plug for the digital outputs (concentration limits / valve control)
REAR PANEL
optionally the subminiature “D mating plug for analog signal inputs
(interference cross compensation, TC only)
optionally the subminiature D mating socket for the RS 232 C / RS 485 interface
optionally the subminiature D mating plug for the status signals (relay outputs)
Additional for BINOS® 100 2M/F
optionally the solenoid valve block
optionally the terminal strips for the 7 digital inputs
optionally the terminal strips for the FOUNDATION™ fieldbus
**)
**) ***)
*)
*)
not in combination with FOUNDATION™ fieldbus
**)
BINOS® 100 2M with external power supply only
***)
not in combination with RS 232/485 interface
ETC00265(1) Series 100 e 23.06.2000
1 - 5
Page 36
TECHNICAL DESCRIPTION
REAR PANEL
Rosemount Analytical
K1 K2 K1 K2
6
1
IN
OUT
CROSS COMP.
5
8
X1 OUTPUT
7
INTERFACE
24 V
max. W
X2 OUTPUT
3
12+ 1
X3 OUTPUT
2
MADE IN GERMANY
9
4
1 - 6
3
Fig. 1-5: BINOS® 100 (M), OXYNOS® 100 (EO2), HYDROS® 100, Rear view
1 Gas inlet line fittings
2 Analog signal output mating socket
3 24 VDC supply input terminal
4 Plug for Digital signal output
5 Gas outlet line fittings
6 Housing cover fastening screws
7 mating socket Serial Interface [RS 232 C / 485] (Option)
8 Plug for Output Relays (Option)
9 Plug for analog signal inputs
(interference cross compensation, HYDROS® 100 only)
ETC00265(1) Series 100 e 23.06.2000
Page 37
Rosemount Analytical
TECHNICAL DESCRIPTION
REAR PANEL
X2 OUTPUT
INTERFACE
X1 OUTPUT
6
7
IN
1
2
8
3
24 V
max. 40 W
3
12+1
OUT
5
4
X3 OUTPUT
MADE IN GERMANY
ETC00265(1) Series 100 e 23.06.2000
Fig. 1-6: OXYNOS® 100 (PO2), Rear view
1 Gas inlet line fittings 2 Analog signal output mating socket 3 24 VDC supply input terminal 4 Plug for Digital signal output 5 Gas outlet line fittings 6 Housing cover fastening screws 7 mating socket Serial Interface [RS 232 C / 485] (Option) 8 Plug for Output Relays (Option)
1 - 7
Page 38
TECHNICAL DESCRIPTION
REAR PANEL
12 34 56789
17
FLOW
Rosemount Analytical
10
11
Fig. 1-7: BINOS
15
®
100 2M, version A (shown with internal power supply), Rear view with all options
12131416
1 Gas inlet line fitting 22
nd
gas inlet line fitting (option) 3 Gas outlet line fitting 42
nd
gas outlet line fitting (option) 5 Plug for analog signal inputs
(interference cross compensation, TC only) 6 Solenoid valves: common gas outlet line fitting 7 Solenoid valves: Test gas inlet 1 8 Solenoid valves: Test gas inlet 2 9 24 VDC output (max. 2 A, see technical data)
10 Power supply [UPS 01 T (Universal Power Supply)] 11 Plug Power Supply (Mains line) 12 Solenoid valves: Zero gas inlet 13 Solenoid valves: Sample gas inlet 14 Plug Digital Outputs (threshold contacts) 15 Mating socket Serial Interface [RS 232 C / 485] (option) 16 Mating socket Analog Signal Outputs 17 Plug Output Relays (status signal option)
1 - 8
ETC00265(1) Series 100 e 23.06.2000
Page 39
Rosemount Analytical
TECHNICAL DESCRIPTION
REAR PANEL
1234 5
K1 K2 K1 K2
IN OUT
X1 OUTPUT
DURCHFLUSS
IN/OUT
X2 OUTPUT X3 OUTPUT
FLOW
MAX.1L/MIN
87
ANALOG IN
SPAN 1
SPAN 2
SAMPLE
ZERO
1
3
OUT
24V max.120W
2
9
DIGITAL IN
V+ V-
E1 E2
E3 E4 E5
E6 E7
FB+ FB­FB+ FB-
16 15 17 14 13 12
Fig. 1-8: BINOS
®
100 2M, version B (with external power supply), Rear view with all options
6
1 Gas inlet line fitting 22
nd
gas inlet line fitting (option) 3 Gas outlet line fitting 42
nd
gas outlet line fitting (option) 5 Plug for analog signal inputs
(interference cross compensation, TC only) 6 Solenoid valves: common gas outlet line fitting 7 Solenoid valves: Test gas inlet 1 8 Solenoid valves: Test gas inlet 2 9 Terminal strips for the 7 Digital Inputs (option)
10 Terminal strips for FOUNDATION™ Fieldbus (option) 11 24 V DC supply input terminal 12 Solenoid valves: Zero gas inlet 13 Solenoid valves: Sample gas inlet 14 Plug Digital Outputs (threshold contacts) 15 Mating socket Serial Interface 16 Mating socket Analog Signal Outputs 17 Plug Output Relays (status signal option)
*)
**)
[RS 232 C / 485] (option)
1011
ETC00265(1) Series 100 e 23.06.2000
*)
only possible if serial interface RS 232/485 is not request !
**)
only possible if FOUNDATION™ Fieldbus is not request !
1 - 9
Page 40
TECHNICAL DESCRIPTION
REAR PANEL
Rosemount Analytical
1234 5
OUTIN
X1 OUTPUT
IN/OUT
X4
!
DURCHFLUSS
FLOW
MAX.1L/MIN
X3 OUTPUTX2 OUTPUT
87
K2K1K2K1
ANALOG IN
SPAN 1
SPAN 2
SAMPLE
ZERO
OUT
24V 120W
1
8
7
!
SPAN 1
SPAN 2
OUT
2
3
SAMPLE
ZERO
16
Fig. 1-9: BINOS
15
14 13
17
®
100 2M (special version), Rear view with all options
6
12
11 12
1 Gas inlet line fitting channel 1 2 Gas inlet line fitting channel 2 3 Gas outlet line fitting channel 1 4 Gas outlet line fitting channel 2 5 Plug for analog signal inputs
(interference cross compensation, TC only) 6 Solenoid valves: common gas outlet line fitting 7 Solenoid valves: Test gas inlet 1 8 Solenoid valves: Test gas inlet 2 9 (open)
10 (open) 11 24 V DC supply input terminal 12 Solenoid valves: Zero gas inlet 13 Solenoid valves: Sample gas inlet 14 Plug Digital Outputs (threshold contacts) 15 Mating socket Serial Interface [RS 232 C / 485] (option) 16 Mating socket Analog Signal Outputs 17 Plug Output Relays (status signal option)
13
1 - 10
ETC00265(1) Series 100 e 23.06.2000
Page 41
Rosemount Analytical
1.3 Internal Construction
The analyzers includes the following components:
Depending on analyzer configuration
- one or two IR photometer benches
- one IR photometer and one EO2 sensor
- one IR photometer and one PO2 sensor
- one IR photometer and one TC sensor
- one PO2 sensor and one TC sensor
- one EO2 sensor and one TC sensor
TECHNICAL DESCRIPTION
INTERNAL CONSTRUCTION
- one or two PO2 sensors
- one ot two EO2 sensors
- one or two TC sensors
Optionally one pressure sensor (range of 800 to 1,100 hPa).
The concentration values computed by the analyzer will then be corrected to reflect the barometric pressure to eliminate faulty measurements due to changes in barometric pressure (see technical data).
Optionally one gas sampling pump (BINOS® 100 2M/F only, see chapter 7.4 and 8.17).
[pumping rate maxi. 2,5 l/min. (special solution with 2 pumps with parallel gas paths)].
BINOS® 100 2M (standard version) / BINOS® 100 F:
Integrated power supply (230/120 V AC).
Optionally solenoid valve unit
(BINOS® 100 2M/F only, special solution with 2 valve blocks with parallel gas paths).
For this case there are built-in 4 (8) solenoid valves (Sample Gas - Zero Gas ­Span Gas 1- Span Gas 2) at the analyzer. For manual or automatical adjustment the zero gas and the span gases will be fed to the solenoid valves controlled by the analyzer. If a solenoid valve is open there is illuminated a green LED (Fig. 1-2, Item 3) at the front panel.
ETC00265(1) Series 100 e 23.06.2000
1 - 11
Page 42
TECHNICAL DESCRIPTION
INTERNAL CONSTRUCTION
Cover metal plate
Rosemount Analytical
Gas line fittings
PCB BKS
(channel 2)
Front panel
IR photometer bench (depending on analyzer configuration)
(channel 1)
Pressure sensor (option)
1 - 12
Fig. 1-10: Inside View BINOS® 100
(1 IR channel analyzer, high measuring range with gas detector)
ETC00265(1) Series 100 e 23.06.2000
Page 43
Rosemount Analytical
TECHNICAL DESCRIPTION
INTERNAL CONSTRUCTION
Gas line fittings
Cover metal plate
PCB BKS
IR photometer bench
(depending on analyzer
configuration)
(channel 2)
electrochemical oxygen sensor with PCB OXS
(channel 1)
Pressure sensor (option)
ETC00265(1) Series 100 e 23.06.2000
Front panel
Fig. 1-11: Inside View BINOS® 100 M
(IR channel / electrochemical oxygen measurement)
1 - 13
Page 44
TECHNICAL DESCRIPTION
INTERNAL CONSTRUCTION
Gas line fittings
Security dust filter
Rosemount Analytical
O2 sensor
Inlet
Outlet
Heat exchanger
PCB BKS
to Sensor
from Sensor
Heat exchanger view "X" (180° rotate)
1 - 14
Front panel
Fig. 1-12: OXYNOS® 100, Inside View with paramagnetic sensor
ETC00265(1) Series 100 e 23.06.2000
Page 45
Rosemount Analytical
TECHNICAL DESCRIPTION
INTERNAL CONSTRUCTION
Gas line fittings
channel 2 channel 1
Security dust filter
Pressure sensor (Option)
Fig. 1-13: OXYNOS® 100, Inside view with electrochemical sensor
ETC00265(1) Series 100 e 23.06.2000
Front panel
1 - 15
Page 46
TECHNICAL DESCRIPTION
INTERNAL CONSTRUCTION
Rosemount Analytical
Gas line fittings
PCB BKS
PCB WAP 100
OUTINOUT
Out
Thermal conductivity sensor
IN
In
1 - 16
Front panel
Fig. 1-14: HYDROS® 100, Inside view
ETC00265(1) Series 100 e 23.06.2000
Page 47
Rosemount Analytical
Solenoid valves
(Option)
Power supply option
(UPS 01)
TECHNICAL DESCRIPTION
INTERNAL CONSTRUCTION
Gas line fittings
Circuit board BKS
(below physical bench)
IR photometer bench
(depending on analyzer
configuration)
Paramagnetic
oxygen sensor
(depending on analyzer
configuration)
(Channel 1)
(Channel 2)
Electrochemical
oxygen sensor with circuit board OXS (depending on analyzer
configuration)
(Channel 1)
Pressure sensor
(option)
Gas sampling pump
(option)
Fine dust filter
with integrated
needle valve for regulation of gas flow rate (option)
Fig. 1-15: Inside View dual-channel BINOS® 100 2M (version A)
ETC00265(1) Series 100 e 23.06.2000
Flow indicator
(option)
(IR channel / oxygen measurement, combined)
1 - 17
Page 48
TECHNICAL DESCRIPTION
INTERNAL CONSTRUCTION
Solenoid valves
(Option)
Power supply option
(UPS 01)
Rosemount Analytical
Gas line fittings
Circuit board BKS
(Channel 1)
Electrochemical
oxygen sensor
with circuit board OXS
Pressure sensor
(option)
Gas sampling pump
(option)
Fine dust filter
with integrated
needle valve for regulation of gas flow rate (option)
1 - 18
Flow indicator
(option)
Fig. 1-16: Inside View BINOS® 100 2M (version A)
(1 channel oxygen measurement, electrochemical)
ETC00265(1) Series 100 e 23.06.2000
Page 49
Rosemount Analytical
TECHNICAL DESCRIPTION
INTERNAL CONSTRUCTION
Gas line fittings
*
Solenoid valves
(Option)
Power supply option
(UPS 01)
PCB WAP 100
***
Circuit board BKS
(below physical bench)
IR photometer bench
(depending on analyzer
configuration)
Electrochemical
oxygen sensor with circuit board OXS (depending on analyzer
configuration)
Thermal conductivity
sensor
(depending on analyzer
configuration)
Fine dust filter
with integrated
needle valve for regulation of gas flow rate (option)
OUT
FILTERELEMENT 2um
42 707 676
FLOW
IN
B
C
Gas sampling pump
(option)
Flow indicator
(option)
Fig. 1-17: Inside View dual-channel BINOS® 100 2M (version A)
ETC00265(1) Series 100 e 23.06.2000
(IR / EO2 , TC / EO2 or IR / TC measurement, combined)
1 - 19
Page 50
TECHNICAL DESCRIPTION
INTERNAL CONSTRUCTION
Rosemount Analytical
Gas line fittings
***
Solenoid valves
(Option)
Power supply option
(UPS 01)
PCB WAP 100
*
Circuit board BKS
OUT
Paramagnetic
oxygens sensor
Fine dust filter
with integrated
needle valve for regulation of gas flow rate (option)
FILTERELEMENT 2um
42 707 676
FLOW
Thermal conductivity
sensor
IN
Gas sampling pump
(option)
Flow indicator
(option)
1 - 20
Fig. 1-18: Inside View dual-channel BINOS® 100 2M (version A)
(oxygen measurement (paramagnetic) / thermal conductivity measurement, combined)
ETC00265(1) Series 100 e 23.06.2000
Page 51
Rosemount Analytical
TECHNICAL DESCRIPTION
INTERNAL CONSTRUCTION
Gas line fittings
***
Solenoid valves
(Option)
Digital inputs and FOUNDATION
Fieldbus
(options)
PCB`s WAP 100 and
HEX 01
(intrinsically safe TC
measurement for
potentially explosive
atmosphere,
consult factory)
*
Circuit board BKS
(below physical bench)
IR photometer bench
(depending on analyzer
configuration)
Electrochemical
oxygen sensor with circuit board OXS (depending on analyzer
configuration)
OUT
(Channel 2)
(Channel 1)
Thermal conductivity
sensor
(depending on analyzer
configuration)
Fig. 1-19: Inside View dual-channel BINOS® 100 2M (version B)
ETC00265(1) Series 100 e 23.06.2000
(Channel 1)
IN
Gas sampling pump
(option)
(IR channel / oxygen measurement, combined)
1 - 21
Page 52
TECHNICAL DESCRIPTION
INTERNAL CONSTRUCTION
Rosemount Analytical
Gas line fittings
***
Solenoid valves
(Option)
Digital inputs and FOUNDATION
Fieldbus
(options)
PCB`s WAP 100 and
HEX 01
(intrinsically safe
measurement for
potentially explosive
atmosphere,
consult factory)
*
Circuit board BKS
OUT
Paramagnetic
oxygen sensor
1 - 22
Thermal conductivity
sensor
IN
Gas sampling pump
(option)
Fig. 1-20: Inside View dual-channel BINOS® 100 2M (version B)
(oxygen measurement (parmagnetic) - thermal conductivity measurement, combined)
ETC00265(1) Series 100 e 23.06.2000
Page 53
Rosemount Analytical
TECHNICAL DESCRIPTION
INTERNAL CONSTRUCTION
Gas line fittings
Solenoid valves
channel 2
(option)
Solenoid valves
channel 1
(option)
Gas sampling pump
channel 2 (option)
*
***
IR photometer benches
(depending on analyzer
configuration)
Fig. 1-21: Inside View dual-channel BINOS® 100 2M (special version)
ETC00265(1) Series 100 e 23.06.2000
Gas sampling pump
channel 1
(option)
(two gas sampling pumps and two solenoid valve blocks)
1 - 23
Page 54
TECHNICAL DESCRIPTION
INTERNAL CONSTRUCTION
Rosemount Analytical
Terminal strip X2:
analog outputs
X9
X11
1
2
1
1
F100.1
F100.2
X10
4321
Interface
Serial out
Terminal strip Interface:
serial interface option
Photometer / sensor assembly
is dependening on special
analyzer configuration
Terminal strip X1:
output (status) relays option
Fig. 1-22: Inside View BINOS® 100 F (field housing)
Power supplyTerminal strip X3:
digital outputs
1 - 24
ETC00265(1) Series 100 e 23.06.2000
Page 55
Rosemount Analytical
TECHNICAL DESCRIPTION
INTERNAL CONSTRUCTION
1.3.1 Internal Gas Paths
The materials used for the gas paths are selected according to the application. In marking such selection the diffusion rates of the individual gas components, their corrosivity, and the tempera­ture and pressure of the sampled gas must be taken into account.
a) Gas Path Material
The physical and chemical properties of the sampled gas and the operating conditions (temperature and pressure) of the analyzer determine the materials which are used for gas paths and gas fittings.
Safety Dust Filter
All analyzers are equipped with a PTFE safety dust filter. This filter is no substitute for the necessary dust filter to be provided with sample handling systems (described in chapter 5.2).
Fittings
As standard the analyzers are provided with PVDF fittings, 6/4 mm. The analyzers can be delivered with swagelok® fittings, stainless steel, 6/4 mm or 1/4" as option. Additional fittings are delivered as special options.
Tubing
As standard the analyzers are provided with Viton tubings or PTFE tubings (6/4 mm). Additional tubings (e.g. stainless steel) are delivered as special options.
ETC00265(1) Series 100 e 23.06.2000
1 - 25
Page 56
TECHNICAL DESCRIPTION
Rosemount Analytical
INTERNAL CONSTRUCTION
b) Gas Path Layout (internal tubing)
The principle various possible layouts of the internal gas lines are summarized in the following figures.
Analyzer
Gas outlet
(OUT K1)
Gas inlet
(IN K1)
Fig. 1-23: Tubing in series (BINOS® 100 (M), OXYNOS® 100, HYDROS® 100 analyzers)
Gas outlet
Analyzer
Gas inlet
(IN K1)
Gas inlet
(IN K2)
(OUT K1)
Gas outlet
(OUT K2)
1 - 26
Fig. 1-24 Tubing in parallel (BINOS® 100 (M), OXYNOS® 100, HYDROS® 100 analyzers)
ETC00265(1) Series 100 e 23.06.2000
Page 57
Rosemount Analytical
TECHNICAL DESCRIPTION
INTERNAL CONSTRUCTION
span gas 1
(SPAN 1 / V1)
span gas 2
(SPAN 2 / V2)
sample gas
(SAMPLE / V3)
zero gas
(ZERO / V4)
Gas outlet
(OUT)
Analyzer
solenoid valves
(Option)
sample gas pump
(Option)
Gas inlet
(IN K1)
Flow indicator,
Filter and throttle
(Options)
*)
not for BINOS® 100 F in hazardous areas
*)
Fig. 1-25: Tubing in series (BINOS® 100 2M analyzers)
(equipped with all options)
Gas outlet (OUT K1)
span gas 1
(SPAN 1 / V1)
not used
not used
span gas 2
(SPAN 2 / V2)
sample gas
sample gas pump
sample gas
zero gas
(Option)
zero gas
(ZERO / V4)
(ZERO / V4)
(SAMPLE / V3)
(SAMPLE / V3)
(OUT)
Gas outlet
(OUT)
Gas outlet
(IN K1)
Gas inlet
(IN K2)
Gas inlet
Gas outlet (OUT K1)
Flow indicator
Gas outlet (OUT K2)
Fig. 1-26: Tubing in parallel (special version of BINOS® 100 2M analyzer)
ETC00265(1) Series 100 e 23.06.2000
(equipped with all options)
1 - 27
Page 58
TECHNICAL DESCRIPTION
Rosemount Analytical
1 - 28
ETC00265(1) Series 100 e 23.06.2000
Page 59
Rosemount Analytical
MEASURING PRINCIPLE
IR MEASUREMENT
2. Measuring Principle
Depending on analyzer model different measuring methods will be used.
2.1 IR Measurement
The analyzers are non-dispersive infrared photometers (NDIR) using measurement of selective absorption in a column of gas. The measuring effect derived from absorption of infrared radiation is due to the gas being measured. The gas specific wavelengths of the absorption bands characterize the type of gas while the strength of the absorption gives a measure of the concentration of the component measured. Due to a rotation chopper wheel, the radiation intensities coming from measuring and reference side of the analysis cell produce periodically changing signals within the detector. The detector signal amplitude thus alternates between concentration dependent and concentra­tion independent values. The difference between the two is a reliable measure of the concentration of the absorbing gas component.
Dependent on measuring component and measuring concentration, two different measuring methods will be used.
2.1.1 Interference Filter Correlation (IFC Principle)
The undivided analysis cell is alternately illuminated with filtered light concentrated in one of two spectral separated wave length ranges. One of these two spectrally separated wave length bands is chosen to coincide with an absorption band of the sample gas, and the other is chosen such that none of the gas constituents expected to be encountered in practice absorbs anywhere within the band.
The spectral transmittance curves of the interference filters used in the 100 series analyzer and the spectral absorption of the gases CO and CO2 are shown in Fig. 2-1. It can be seen that the absorption bands of these gases each coincide with the passbands of one of the interference filters. The fourth interference filter, used for generating a reference signal, has its passband in a spectral region where none of these gases absorb. Most of the other gases of interest also do not absorb within the passband of this reference filter.
ETC00265(1) Series 100 e 23.06.2000
2 - 1
Page 60
MEASURING PRINCIPLE
IR MEASUREMENT
Rosemount Analytical
9075604530150
Transmittance [%]
HC
Transmittance [%]
CO
2
CO
2
Reference
4400 460042004000 48003000 3200 3400 3600 3800 5000 5200 5400 5600 5800 6000
Wave Length [nm]
CO
CO
Interference -
Absorption Band
Filter
18 36 54 72 900
Fig. 2-1: Absorption Bands of Sample Gases and Transmittance of the
Interference Filters used
The signal generation happens by a pyroelectrical (solid-state) detector.
The detector records the incoming IR radiation. This radiation intensity is reduced by the absorption of the gas at the corresponding wave lengths. By comparing the intensity at measuring and reference wave length an alternating voltage signal is developed. This signal results from cooling and heating of the pyroelectrical material of the detector.
2 - 2
ETC00265(1) Series 100 e 23.06.2000
Page 61
Rosemount Analytical
MEASURING PRINCIPLE
IR MEASUREMENT
2.1.2 Opto-Pneumatic Measuring Principle
A thermal radiator generates the infrared radiation passing through a chopper wheel. This radiation alternately passes through a filter cell and reaches the measuring and reference side of the analysis cell with equal intensity.
After passing another filter cell the radiation reaches the pneumatic detector.
The pneumatic detector compares and evaluates the radiation from the measuring and reference sides and converts them into voltage signals proportional to their intensity via a preamplifier.
The detector consists of a gas-filled absorption and a compensation chamber which are interconnected via a flow channel.
Absorption chamber
Flow channel with Microflow sensor
CaF2 Window
Gas intake connection
Compensation chamber
ETC00265(1) Series 100 e 23.06.2000
Fig. 2-2: Principle Design of Gas Detector
2 - 3
Page 62
MEASURING PRINCIPLE
IR MEASUREMENT
Rosemount Analytical
In principle the detector is filled with the infrared active gas to be measured and is only sensitive to this distinct gas with its characteristic absorption spectrum. The absorption chamber is sealed with a window which are transparent for infrared radiation [usually CaF2 (Calcium fluoride)].
When the IR radiation passes through the reference side of the analysis cell into the detector, no preabsorption occurs. Thus the gas inside the absorption chamber is heated, expands and some of it passes through the flow channel into the compensation chamber.
When the IR radiation passes through the open measurement side of the analysis cell into the detector, a part of it is absorbed depending on sample gas concentration. The gas in the absorption chamber then is heated less than in the case of radiation coming from reference side. Absorption chamber gas become colder, gas pressure in the absorption chamber is reduced and some gas of compensation chamber passes through the flow channel into the absorption chamber.
The flow channel geometry is designed in such a way that it hardly impedes the gas flow by restriction. Due to the radiation of chopper wheel, the different radiation intensities lead to periodically repeated flow pulses within the detector.
The microflow sensor evaluates this flow and converts it into electrical voltages. The electronics, which follow, evaluate the signals and convert them into the corresponding display format.
2 - 4
ETC00265(1) Series 100 e 23.06.2000
Page 63
Rosemount Analytical
123
123
MEASURING PRINCIPLE
IR MEASUREMENT
2.1.3 Technique
The broadband emission from two IR sources (in the case of dual channel analyzers) passes through the chopper blade, then, if IFC, through combinations of interference filters, if optopneumatic principle depending on application through an optical filter (reduction of influ­ences) and enters the analysis cells. The light transmitted through these cells is focused by filter cells onto the according detector. The preamplified detector output signal is sent to microprocessor circuitry, which converts the analytical signals to results expressed directly in physical concentra­tion units (Vol.-%, ppm, mg/Nm3 etc.).
Light source
Analysis cell measuring side
Analysis cell reference side
MOTOR
Duplex filter disc
Adapter cell (high measuring range)
Analysis cell (undivided)
Filter cell
Preamplifier
Filter cell
Gas detector
ETC00265(1) Series 100 e 23.06.2000
Pyroelectrical detector (solid-state detector)
Preamplifier
Chopper blade
Fig. 2-3: Principle Representation
2 - 5
Page 64
MEASURING PRINCIPLE
OXYGEN MEASUREMENT (PO2 PARAMAGNETIC PRINCIPLE)
Rosemount Analytical
2.2 Oxygen Measurement
Depending on analyzer model different measuring methods will be used.
The installed type of oxygen sensor is to identify at the channel code (see Fig. 1-1).
% O2 para. = paramagnetic Sensor % O2 chem. = electrochemical Sensor
2.2.1 Paramagnetic Measurement
The determination of O2 concentration is based on the paramagnetic principle (magneto­mechanic principle).
Two nitrogen filled (N2 is diamagnetic) quartz spheres are arranged in a "dumbbell" configuration and suspended free to rotate on a thin platinum ribbon in a cell. A small mirror that reflects a light beam coming from a light source to a photodetector, is mounted on this ribbon. A strong permanent magnet especially shaped to produce a strong highly inhomo­geneous magnetic field inside the analysis cell, is mounted outside the wall. When oxygen molecules enter the cell, their paramagnetism will cause them to be drawn towards the region of greatest magnetic field strength. The O2 molecules thus exert different forces which produce a torque acting on the sphere arrangement, and the suspended “dumbbell”, along with the mirror mounted on its suspension ribbon, will be angulary rotated away from the equilibrium position. The mirror then will deflect an incident light beam onto the photodetector which itself produces an electric voltage. The electric signal is amplified and fed back to a conducting coil at the “dumbbell”, forcing the suspended spheres back to the equilibrium position. The current required to generate the restoring torque to return the “dumbbell” to its equilibrium position is a direct measure of the O2 concentration in the gas mixture.
The complete analysis cell consists of analysis chamber, permanent magnet, processing electronics, and a temperature sensor. The sensor itself is thermostat controlled up to approx. 55 °C. For warming up the measuring gas is conducted via a heat-exchanger.
Optionally we have built-in a solvent resistant cell or an intrinsic safe cell for potentially explosive atmosphere.
2 - 6
ETC00265(1) Series 100 e 23.06.2000
Page 65
Rosemount Analytical
MEASURING PRINCIPLE
OXYGEN MEASUREMENT (PO2 PARAMAGNETIC PRINCIPLE)
ETC00265(1) Series 100 e 23.06.2000
Fig. 2-4: Principle Construction of paramagnetic Analysis Cell
1 Permanent magnet 2 Platinum wire 3 Mirror 4 Quartz spheres 5 Wire loop 6 Photodetector 7 Light source 8 Amplifier 9 Display
2 - 7
Page 66
MEASURING PRINCIPLE
Rosemount Analytical
OXYGEN MEASUREMENT (EO2 ELECTROCHEMICAL PRINCIPLE)
2.2.2 Elektrochemical Measurement
The determination of O2 concentrations is based on the principle of a galvanic cell. The principle structure of the oxygen sensor is shown in Fig. 2-5.
Lead wire (Anode)
Lead wire (Cathode)
Anode
O - ring (8)
Plastic disc (9)
Plastic top (10)
(1)
(Lead)
(Black)
Thermistor (5)
Acid electrolyte (3)
Sponge disc (7)
Cathode
(2)
Teflon membrane (4)
(Red)
Resistor (6)
(Gold film)
Fig. 2-5: Structure of electrochemical Oxygen Sensor
The oxygen senor incorporate a lead/gold oxygen cell with a lead anode (1) and a gold cathode (2), using a specific acid electrolyte. To avoide moisture losses at the gold electrode a sponge sheet is inserted on the purged side.
Oxygen molecules diffuse through a non-porous Teflon membrane (4) into the electrochemical cell and are reduced at the gold-cathode. Water results from this reaction. On the anode lead oxide is formed which is transferred into the electrolyte. The lead anode is regenerated continuously and the electrode potential therefore remains unchanged for a long time. The rate of diffusion and so the response time (t90) of the sensor is dependent on the thickness of the Teflon membrane.
2 - 8
ETC00265(1) Series 100 e 23.06.2000
Page 67
Rosemount Analytical
MEASURING PRINCIPLE
OXYGEN MEASUREMENT (EO2 ELECTROCHEMICAL PRINCIPLE)
(Red) (Black)
Thermistor (5)
(-)
Gold-
Cathode (2)
O2 + 4 H+ + 4 e- 2 H2O
Summary reaktion O
(11)
Electrolyte (3)
(ph 6)
+ 2 Pb 2 PbO
2
Resistor (6)
2 Pb + 2 H2O 2 PbO + 4 H+ + 4 e
Fig. 2-6: Reaction of galvanic cell
(+)
Lead-
Anode (1)
-
The electric current between the electrodes is proportional to the O2 concentration in the gas mixture to be measured. The signals are measured as terminal voltages of the resistor (6) and the thermistor (5) for temperature compensation.
The change in output voltages (mV) of the senor (11) represents the oxygen concentration.
Note !
Depending on measuring principle the electrochemical O oxygen. Admit cells continuously with sample gas of low grade oxygen concentration or with oxygen-free sample gas could result a reversible detuning of O2 sensitivity. The output signal will become instabil. For correct measurement the cells have to admit with a O We recommend to use the cells in intervall measurement (purge cells with conditioned ambient air at measurement breaks). If it is necessary to interrupt oxygen supply for several hours or days, the cell have to regenerate (admit cell for about one day with ambient air). Temporary flushing with nitrogen (N2) for less than 1 h (e.g. analyzer zeroing) will have no influence to measuring value.
cell needs a minimum internal consumption of
2
concentration of at least 2 Vol.-%.
2
ETC00265(1) Series 100 e 23.06.2000
2 - 9
Page 68
MEASURING PRINCIPLE
THERMAL CONDUCTIVITY (TC) MEASUREMENT
Rosemount Analytical
2.3 Thermal Conductivity Measurement
To measure gases like Hydrogen (H2), Argon (Ar) or Helium (He), the measurement method of thermal conductivity (TC) will be used.
2.3.1 Sensor Design
The sensor used consists of a system of annular chambers formed by inner and outer sections which may be rotates with respect to one other. The measurement cell itself, which is intercon­nected to an annular groove by transversely drilled passages, is situated within the inner section. The gas fittings are located on the external periphery of the outer section.
Fig. 2-7: Thermal conductivity sensor
2.3.2 Analysis Cell
Both the cell volume and the mass of its measurement resistor have been minimized on order to obtain short response time. A nickel resistor is sandwiched between two superimposed square ceramic plates which form the walls of the measurement cell. The cell volume is approximately 1 µl. A total of four such cells are integrated for form the sensor, two of these function as the measurement cells, and the other two functions as the reference cells. The latter may be either sealed off, or connected to a flow of a reference gas.
2 - 10
ETC00265(1) Series 100 e 23.06.2000
Page 69
Rosemount Analytical
THERMAL CONDUCTIVITY (TC) MEASUREMENT
MEASURING PRINCIPLE
2.3.3 Measurement Method
The entire measurement cell is thermostatted to a temperature of up to 75 °C. The four integral temperature sensors are electrically heated to a higher temperature. Their temperatures, and thus their electrical resistance, are dependent upon heat losses, which, in turn, result from heat transport in the surrounding gas to colder chamber walls. For otherwise stable conditions, this heat transport will be proportional to the thermal conductivity of the gas present between the sensor and the chamber wall. Interconnecting the four sensors into Wheatstone bridge circuit provides an electronic signal proportional to gas density. Electronic circuitry processes this signal to obtain standardized signal amplitudes, and transmits these to both an indicator instrument and to the signal output connector.
The annular chamber mentioned in section 2.31 (Fig. 2-7) is provided with two transverse passages, each of which is equipped with two temperature sensors. One of these transverse passages is subjected to a flow of the sample gas, while to other is subjected to a flow of reference gas (optional), or is sealed off (standard version). The gas flow will distribute itself between the transverse passages, or between the annular grooves on the periphery of the annular chamber, in a transverse passages with respect to the gas fittings.
This results in a variable bypass configuration.
If the transverse passages are aligned directly opposite the gas inlet and outlet fittings, there will result the shortest response times and an enhanced dependence of the analytical signal upon the sample-gas flow rate.
If the transverse passages are arranged aligned at 90° Angles to these gas fittings, the heat transport between sample gas and the sensor will be predominantly by diffusion (i.e. significantly slowed).
The dependence of the analytical signal upon sample gas flow rate will be minimized and the response time extended.
This arrangement has the advantage that any value between the two mentioned extremes may be set (see Fig. 2-8).
ETC00265(1) Series 100 e 23.06.2000
2 - 11
Page 70
MEASURING PRINCIPLE
THERMAL CONDUCTIVITY (TC) MEASUREMENT
T
T
Rosemount Analytical
Flow dependence
Strömungsabhängigkeit
Zeitkonstante
Timing constant
0° 90°45°
Optimaler
Bereich
∆α
Optimal
Range
Zelle
Cell
α
T
Fig. 2-8: Behavior of the time constant and the flowrate dependence of instrument responsibility
for various relative positions of the annular chamber sections.
The material in contact with the sample gas flow rate are Aluminium, Viton, stainless steel and ceramic. This provides for resistance to corrosion which might arise for some types of aggressive sample gas constitutions.
2 - 12
ETC00265(1) Series 100 e 23.06.2000
Page 71
Rosemount Analytical
PHOTOMETER ASSEMBLY
PYROELECTRICAL DETECTOR
3. Photometer Assembly
Depending on gas component and measuring range, different photometer assemblies will be realized in 100 series. Optional the photometer can be sealed to ambient air. In this case all parts are sealed with O- rings. The entire photometer assembly is mounted as a unit on the main circuit board (BKS) by means of a bracket. The main circuit board is inserted into guide rails in the analyzer housing, to which the front panel (membrane keypad) and the rear panel are assembled.
3.1 Photometer with Pyroelectrical Detector (Solid-state detector)
Fig. 3-1 shows the schematical photometer assembly for dual channel operation.
The base element for the photometer assembly is the chopper housing (03), upon which the light source (thermal radiator, 07), the analysis cell (cuvette, 09), and the signal detection unit [filter cell (14/15), pyroelectrical (solid-state) detector with integrated preamplifier (16)] are all mounted.
The chopper housing also incorporates the duplex filters (04/05) for the selection of spectral band­pass ranges from the broadband emission of the light sources.
Between the two halves of the chopper housing (03), which are sealed together with an O-ring, is the chopper blade, driven by a stepping motor. Both the chopper housing and the motor encapsulation are hermetically sealed with respect to the ambient in order to prevent entry of gases, such as atmospheric CO2, which could produce background absorptivity (preabsorption) leading to drift effects. An absorber material provides for constant removal of any traces of CO which may enter the interior of the chopper housing via diffusion.
2
The chopper housing additionally incorporates a photoelectric gate for providing a reference signal for the phase angle of the chopper blade, plus a temperature sensor (28) for monitoring continuously the photometer assembly temperature. This temperature information is used by the signal processing electronics for the compensation of thermal effects.
The analysis cells are merely aluminum tubes equipped with sample gas inlet and outlet fittings.
ETC00265(1) Series 100 e 23.06.2000
3 - 1
Page 72
PHOTOMETER ASSEMBLY
PYROELECTRICAL DETECTOR
Rosemount Analytical
This extremely simple and windowless design enables easy cleaning of the cells in the event of contamination. The only optical surfaces which also might become contaminated are the chopper windows and the windows of the filter cells; these are accessible upon removal of the cell body.
The filter cell (14/15) has a necked conical shape for optimal adaptation of the analysis cell beam cross - sectional profile to the active area of the detectors.
For high measurement ranges (up to 100 %), an adapter cell (10) is required. The use of a spacer ring (08) creates an analysis cell in the space between the exit window of the adapter cell and the entrance window of the filter cell.
3 - 2
ETC00265(1) Series 100 e 23.06.2000
Page 73
Rosemount Analytical
PHOTOMETER ASSEMBLY
PYROELECTRICAL DETECTOR
Legends:
03 Chopper Housing
04 / 05 Duplex Filter Disc
06 Zero - Adjustment Baffle (not for sealed photometer)
07 Light Source (thermal radiator)
08 Analysis Cell 1 - 7 mm (spacer ring)
09 Analysis Cell 50 - 200 mm
10 Adapter Cell
14/15 Filter Cell
16 Detector
17 Flange (light source)
18-21 O - Rings
22 Clamp (analysis cells 1-7 mm)
23 (24) Clamping Collar (analysis cells 1-7 mm)
25 Clamp (analysis cells 10-200 mm)
26 Light Source Mounting Screws
27 Mounting Screws for Analysis Cells/Adapter Cells
28 Temperature Sensor
ETC00265(1) Series 100 e 23.06.2000
Fig. 3-1: Photometer Assembly with Pyroelectrical Detector
3 - 3
Page 74
PHOTOMETER ASSEMBLY
GAS DETECTOR
Rosemount Analytical
3.2 Photometer with Gas Detector
Fig. 3-2 shows schematically the photometer assembly.
This assembly is similar to the assembly with pyroelectrical detector.
The analysis cells are separated into two halves by means of an internal wall along its axis and both ends are sealed with windows. This divided the analysis cell in measuring side and reference side. Sample gas is flowing through measuring side while the closed reference side contains inert gas (N2). To prevent measuring errors by preabsorption, two absorber, fitted to the gas connections of the reference side, absorb CO2 parts.
The filter cell has a single stage conical shape.
The gas detector is connected by a shielded cable to the separate preamplifier. For small measuring ranges the preamplifier is mounted at the analysis cell. For high measuring ranges the preamplifier is mounted at two holding clamps.
3 - 4
ETC00265(1) Series 100 e 23.06.2000
Page 75
Rosemount Analytical
PHOTOMETER ASSEMBLY
GAS DETECTOR
3 521
44
[example above: high measuring ranges (combined with electrochemical O2 measurement),
example below: small measuring ranges(without O2 measurement).
ETC00265(1) Series 100 e 23.06.2000
16 5 4 623
Fig. 3-2: Photometer Assembly 100 Series with Gas Detector
1 Analysis Cell 2 Filter Cell 3 Gas Detector 4 Holding Device 5 Preamplifier 6 Absorber
3 - 5
Page 76
PHOTOMETER ASSEMBLY
Rosemount Analytical
3 - 6
ETC00265(1) Series 100 e 23.06.2000
Page 77
Rosemount Analytical
PREPARATION
5. Preparation of Start-up
Please check the packing and its contents immediately upon arrival. If any item is damageg or lost you are kindly requested to notify the forwarder to undertake a damage survey and report the loss or damage to us immediately.
BINOS® 100 F: Unscrew transfer safety lock of photometer sliding carriage (Fig. 5-1) ! Lock the system during transport !
Fig. 5-1: BINOS® 100 F, Photometer safety lock
ETC00265(1) Series 100 e 23.06.2000
5 - 1
Page 78
PREPARATION
INSTALLATION SITE
5.1 Installation Site
The analyzers must not operate in explosive atmosphere without supplementary protective measures !
Free flow of air into and out of the analyzers (ventilation slits) must not be hindered by nearby objects or walls !
The installation site for the analyzers has to be dry and remain above freezing point at all times. The analyzers must be exposed neither to direct sunlight nor to strong sources of heat.
Rosemount Analytical
Be sure to observe the permissible ambient temperatures (c.f. chapter 27: Technical Data). For outdoor installation, we recommend to install the analyzers in a protective cabinet. At least, the analyzers has to be protected against rain (e.g., shelter).
The MLT has to be installed as near as possible to the sample point, in order to avoid low response time caused by long sample gas lines. In order to decrease the response time, a sample gas pump with a matching high pumping rate may be used. Eventually, the analyzers has to be operated in the bypass mode or by an overflow valve to prevent too high flow and too high pressure (Fig. 5-2).
Exhaust
5 - 2
Overpressure valve
Gas sampling pump
analyzer
Filter
Fig. 5-2: 100 series, Bypass installation
Flow meter
Exhaust
ETC00265(1) Series 100 e 23.06.2000
Page 79
Rosemount Analytical
GAS CONDITIONING (SAMPLE HANDLING)
PREPARATION
5.2 Gas Conditioning (Sample Handling)
The conditioning of the sample gas is of greatest importance for the successful operation of any analyzer according to extractive method.
All gases have to be supplied to the analyzers as conditionned gases ! If corrosive gases are inserted into the instrument, it has is to be verified that there are no gas components which may damage the gas path components.
The gas has to fullfil the following conditions:
It must
be free of condensable constituents be free of dust be free of aggressive constituents which are not compatible with the material of the gas
paths.
have temperatures and pressures which are within the specifications stated in “Technical
Data” of this manual.
Inflammable or explosive gas mixtures may not be introduced into the analyzers without supplementary protective measures !
When analysing vapours, the dewpoint of the sample gas has to be at least 10 °C below the ambient temperature in order to avoid the precipitation of condensate in the gas paths.
Suitable gas conditionning hardware may be supplied or recommended for specific analytical problems and operating conditions.
ETC00265(1) Series 100 e 23.06.2000
5 - 3
Page 80
PREPARATION
GAS CONDITIONING (SAMPLE HANDLING)
Rosemount Analytical
5.2.1 Fine Dust Filter (Option BINOS® 100 2M/F)
A fine dust filter having a pore size of 2 µm is inserted into the BINOS® 100 2M/F front panel as option (Fig. 1-2, Item 10, not for 2 channel analyzers with parallel gas paths and for BINOS® 100 F analyzers used in hazardous area (Ex zones)).
5.2.2 Gas Sampling Pump (Option BINOS® 100 2M/F)
Optional BINOS® 100 2M/F can be equipped with a gas sampling pump (pumping rate max. 2.5 l/min., see chapter 7.4). For special solutions with 2 parallel measuring channels and 2 gas sampling pumps consult factory.
Lifetime max. 5,000 operation hours !
5.2.3 Pressure Sensor (Option)
It is possible to integrate a pressure sensor with a range of 800 - 1,100 hPa. The concentration values computed by the analyzer will then be corrected to reflect the barometric pressure to eliminate faulty measurements due to changes in barometric pressure (see technical data, chapter 27.).
5.2.4 Gas Flow
The gas flow rate should be within the range 0.2 l/min to maxi. 1.5 l/min !
A constant flow rate of about 1 l/min is recommended.
The allowed gas flow rate for analyzers with paramagnetic oxygen sensor and for BINOS® 100 F analyzers used in hazardous area (Ex zones) is max. 1.0 l/min!
It is possible to integrate max. two flow indicators into BINOS® 100 2M/F located at operation front panel (Fig. 1-2 to 1-4). For standard BINOS® 100 2M/F flow control is done with a screw driver via an optional integrated throttle into the optional built-in dust filter (see chapter 5.2.1).
5 - 4
ETC00265(1) Series 100 e 23.06.2000
Page 81
Rosemount Analytical
PREPARATION
GAS CONNECTIONS
5.3 Gas Connections
The installed gas connections are depending on gas analyzer specification and model. All fittings are clearly marked. The fittings are located on the rear panel of the instrument or on the left bottom side.
The exhaust gas lines have to be mounted in a declining, pressureless and frost-free way and according to the valid emission legislation !
Do not interchange gas inlets and gas outlets !
Ensure that all gas connections are made as labeled and are leak free ! Improper gas connections could result in explosion and death ! The unit´s exhaust may contain hydrocarbons and other toxic gases such as carbon monoxide ! Carbon monoxide is highly toxic !
Permissible gas pressure max. 1,500 hPa !
5.3.1 Standard
Depending on analyzer version the following gas connections are installed:
in = Gas inlet out = Gas outlet
K 1 = measuring channel 1 K 2 = measuring channel 2
*)
Zero gas and span gas are introduced directly via the sample gas inlet. The test gas containers have to be set up according to the current legislation.
Be sure to observe the safety regulations for the respective gases (sample gas and test gases / span gases) and the gas bottles !
*)
for HYDROS® 100 with open reference side of sensor the gas fittings of channel 2 are used to
connect the reference gas.
ETC00265(1) Series 100 e 23.06.2000
5 - 5
Page 82
PREPARATION
GAS CONNECTIONS
Rosemount Analytical
K1 K2 K1 K2
Gas inlet
fittings
Gas outlet
IN
X1 OUTPUT
INTERFACE
24 V
max. W
OUT
CROSS COMP.
3
12+ 1
Fig. 5-3a: BINOS® 100 (M), OXYNOS® 100 (EO2), HYDROS® 100, gas connections
X2 OUTPUT
INTERFACE
X1 OUTPUT
IN
fittings
Gas inlet
fittings
5 - 6
24 V
max. 40 W
3
12+1
X3 OUTPUT
MADE IN GERMANY
Fig. 5-3b: OXYNOS® 100 (PO2), gas connections
OUT
Gas outlet
fittings
ETC00265(1) Series 100 e 23.06.2000
Page 83
Rosemount Analytical
PREPARATION
GAS CONNECTIONS
Gas inlet
fittings
Gas outlet
fittings
Fig. 5-3c: BINOS® 100 2M, standard gas connections
Wall mounting holder
ETC00265(1) Series 100 e 23.06.2000
Gas connections
Fig. 5-3d: BINOS® 100 F, standard gas connections
5 - 7
Page 84
PREPARATION
Rosemount Analytical
GAS CONNECTIONS
5.3.2 Internal Solenoid Valves (Option BINOS® 100 2M/F)
For operation with optional internal solenoid valves, the following indications have to be considered: All necessary gases have to be connected at the corresponding solenoid valve at an overpressure of 50 - maxi. 500 hPa. For 2 channel analyzer with parallel gas paths we can deliver a special solution with 2 valve blocks.
Be sure to observe the safety regulations for the respective gases (sample gas and test gases / span gases) and the gas bottles ! The test gas containers have to be set up according to the valid regulations.
If a solenoid valve is open there is illuminated a green LED (Fig. 1-2, Item 3) at the front panel.
Solenoid valve:
common gas outlet
(to standard gas inlet K1)
Solenoid valve:
span gas 1
Solenoid valve:
span gas 2
5 - 8
Solenoid valve:
sample gas
Solenoid valve:
zero gas
Fig. 5-4a: BINOS® 100 2M (version A), gas connections with solenoid valve option
ETC00265(1) Series 100 e 23.06.2000
Page 85
Rosemount Analytical
2
K1 K2 K1 K2
PREPARATION
Solenoid valve:
span gas 1
IN OUT
X1 OUTPUT
IN/OUT
X2 OUTPUT X3 OUTPUT
DURCHFLUSS MAX.1L/MIN
ANALOG IN
SAMPLE
ZERO
SPAN 1
SPAN 2
1
3
OUT
24V max.1
2
Fig. 5-4b: BINOS® 100 2M (version B), gas connections with solenoid valve option
Solenoid valve:
Span gas 1
channel 1
Solenoid valve:
not used
ANALOG IN
SPAN 1
!
SPAN 1
Solenoid valve:
span gas 2
Solenoid valve:
common gas outlet
(to standard gas inlet K1)
Solenoid valve:
sample gas
Solenoid valve:
zero gas
Solenoid valve:
not used
Solenoid valve:
Span gas 2
channel 2
Solenoid valve:
common gas outlet
channel 1
(to standard gas inlet K1)
Solenoid valve:
sample gas
channel 1
Solenoid valve:
zero gas
channel 1
Fig. 5-4c: BINOS® 100 2M (special version), gas connections with 2 solenoid valve blocks
ETC00265(1) Series 100 e 23.06.2000
SPAN 2
SAMPLE
ZERO
24V 120W
OUT
SPAN 2
Solenoid valve:
OUT
1
2
3
SAMPLE
Common gas outlet
channel 2
(to standard gas inlet K2)
Solenoid valve:
ZERO
sample gas
channel 2
Solenoid valve:
zero gas
channel 2
5 - 9
Page 86
PREPARATION
PURGE GAS CONNECTIONS OF BINOS® 100 F / CONTINUOUS PURGE
Rosemount Analytical
5.3.3 Purge gas connection of BINOS® 100 F for Ex zones
To connect purge gas to the BINOS® 100 F housing (Fig. 27-5 and 27-6) and for specifications of purge gas to simplified pressurization for European ex zone 2 or with an EExp approved “purge system for European ex zone 1 (both according to CENELEC, EN 50016) be sure to observe the additional notes, safety precautions and warnings given in the individual enclosed manuals !
Activation of pressurization bypass switch must be performed by a authorized person (for hazardous areas) only. This has to be in accordance with respective legislation only !
Purging can be accomplished with any inert gas including "instrument quality" air, nitrogen, or argon which contains no more than trace amounts of a combustible vapor. Compressed air is the most common and practical purging medium. "Instrument quality" air/compressed air as purge gas use for measurment of non-flammable sample gases only. Bottled nitrogen is uitilized most frequently in isolated places where no compressed air source is available. Saftey for human beings has to be taken into consideration in case of N2 purge in closed rooms, e.g. containers.
a) Continuous Purge (for CENELEC Ex Zone 1 Applications)
Connect purge gas (air) to the BINOS® 100 F inlet via Ø 10/8 mm tube fitting
(optionally 1/2" O.D.) connection (see Fig. 27-4) ! Outlet is done via Ø 14/12 mm tube fitting and without any backpressure.
Use purge gas (air) supply under pressure of 0.09 psig (6 mbar; 1,006 hPa abs.), resulting
in a flow rate of approx. 212 scfh (100 l/min.). This will provide an enclosure over-pressure of approx. 1.04 inch H2O (2.6 hPa).
(This is an example for H2 resulting in continuous dilution to 1/4 LEL with sample gas flow restriction to 1 l/min. max., measured in a dual housing version. Standard (single) housing or other gases require different gas streams to be documented separately)
Pressure inside the housing must not exceed 5 hPa at normal operation or 10 hPa for a short time of less than 1/2 hour resp. !
5 - 10
ETC00265(1) Series 100 e 23.06.2000
Page 87
Rosemount Analytical
PURGE GAS CONNECTIONS OF BINOS® 100 F / Z PURGE
PREPARATION
b) Z purge for CSA-C/US Ex Zone 2 Non-Flammable Atmospheres
This analyzer is not designed for analysis of flammable sample ! Introduction of flammable samples into this equipment could result in explosion,
causing severe personal injury, death or property damage !
Consult factory if flammable samples are to be measured !
Z Purge kit is designed for protection against the invasion of flammable gases into enclosure from outside atmosphere only. It does not provide protection against the release of inflammable gases contained in the sample gas via internal leakages.
Connect purge gas (air or N2) to the BINOS® 100 F inlet via 1/4" O.D. tube connection
(see Fig. 27-4) ! Outlet is done via 3/8" O.D. fitting with analyzer specific insert throttle.
Use purge gas supply under pressure of 58 psig (4 bar; 5,000 hPa abs.). . This will provide a flow rate approx. 55 scfh (26 l/min.). At this flow rate, five case volumes
of purge gas will pass through the instrument in 11 minutes.
After 11 minutes, reduce the pressure to 5.1 psig (350 mbar; 1,350 hPa abs.), resulting in a flow rate of approx. 18 scfh (8.5 l/min.). This will provide an enclosure over-pressure of approx. 0.24 inch H2O (0.6 hPa) for continuous purge.
Pressure inside the housing must not exceed 5 hPa at normal operation or 10 hPa for a short time of less than 1/2 hour resp. !
ETC00265(1) Series 100 e 23.06.2000
5 - 11
Page 88
PREPARATION
ADDITIONAL HINTS TO BINOS® 100 F (FIELD HOUSING)
Rosemount Analytical
5.4 Additional Hints to BINOS® 100 F (Field Housing)
In the field housing version BINOS® 100 F all componets are incorporated into a protection housing going conform to DIN-standard protection class IP 65 (approx. NEMA 4/4X). This housing is designed for wall mounting. For installation in hazardous areas the BINOS® 100 F is equipped with an impact tested front panel (according to CENELEC, EN 50014) with touch screen keypad. Optionally we can provide additional intrinsically safe I/O's and/or ex interface relays (couplers). For European ex zone 2 a simplified pressurization is installed and individual approval is provided. An EExp approved “purge system for European ex zone 1 (both according to CENELEC, EN 50016) or with Z Purge for measurement of non-flammable gases in hazardous areas (according to CSA-C/US for North American Ex zone 2) is another option.
Be sure to observe the additional notes, safety precautions and warnings given in the individual manuals (simplified pressurization for ex zone 2 / EExp approved “purge system” for ex zone 1) and in section “Safety Summary, subsection VII.!
5 - 12
ETC00265(1) Series 100 e 23.06.2000
Page 89
Rosemount Analytical
ADDITIONAL HINTS TO BINOS® 100 F (FIELD HOUSING)
PREPARATION
5.4.1 Wall Mounting
This housing is designed for wall mounting. For fastening points see Fig. 5-5 please.
Ex zone 1 and ex zone 2 applications require additional space for safety related components (see Fig. 27-4, 27-5 and 27-6).
Lift or carry housing with at least 2 persons because of the high weight of field housing BINOS® 100 F (approx. 30 - 35 kg for standard housing). For easy transport use a suitable cart.
Lift points are labeled ! Labels showing down side for transport ! Do not use electronics of optional purge system” as handle !
approx. 355
300
332
550
492
Glands
(connection cables)
Fig. 5-5: Dimensional sketch / Drill drawing BINOS® 100 F Standard version
ETC00265(1) Series 100 e 23.06.2000
[all dimensions in mm]
18
10
18
5 - 13
Page 90
PREPARATION
ADDITIONAL HINTS TO BINOS® 100 F (FIELD HOUSING)
5.4.2 Electrical Connections
Be sure to observe the safety precautions and warnings !
Verify, that the PG fittings together with pass through cables are hermetic to be in agreement with protection class IP 65 (according to DIN standard
40050). The permissible outside diameters of the cables are 7 to 12 mm !
a) Mains Supply
Rosemount Analytical
The analyzer is specified for an operating voltage of 230 V AC or 120 V AC resp., 47-63 Hz.. Built-in power supply (manual switch between 230/120 V AC) is either power supply of type SL5 or of type SL10.
Opening of housing (front panel) (see chapter 23.).
Take mains line via PG fitting (Fig. 5-6 or 5-5) inside the housing.
Connect L and N to powerline filter (Fig. 5-8) via plug jacket (6,3x0,8 mm). Connect PE via ring cable system to left ground conductor pin (Fig. 5-8). Alternatively the mains line is connected via terminal strips (Fig. 5-8).
Verify beforehand that the line voltage stated on the identification plate (front door inside) agrees with that of your power supply line ! Verify that the position of input voltage switch of the power supply(s) agrees with that of your power supply line (Fig. 27-7, 27-8 and 5-8) ! The BINOS® 100 F (field housing) has no switch with disconnect function. The customer has to provide a switch or circuit breaker into his installation. This switch has to be installed near by analyzer, must be easily attainable for operator
PG glands
of data lines
PG gland
of mains line
Wall mounting
holder
5 - 14
and hase to be charaterized as disconnector for analyzer.
Front panel
Gas connections
Fig. 5-6: BINOS® 100 F, PG fittings for lines (side view from left)
ETC00265(1) Series 100 e 23.06.2000
Page 91
Rosemount Analytical
ADDITIONAL HINTS TO BINOS® 100 F (FIELD HOUSING)
b) optional Data Lines
This are analog outputs, digital inputs/outputs and serial interfaces.
Opening of housing (front panel) (see Item 21).
Take lines via PG fittings (Fig. 5-6 and 5-5) inside the housing.
Connection is to be done to the respective terminal strips (Fig. 5-7 and 5-8). Assignement see chapter 2. please.
Cables to external dataprocessing have to be double-insulated against mains
PREPARATION
voltage for analyzer BINOS® 100 F ! Use cables suitable for intrinsic safe applications only ! Install internal data lines that they have a distance to mains voltage lines of at least 5 mm. This distance has to be valid permanently (e.g. via cable holder) !
Terminal strip X2:
analog outputs
X9
1
X11
1
2
1
F100.1
F100.2
X10
4321
Interface
Serial out
Terminal strip Interface:
serial interface option
Power supply
ETC00265(1) Series 100 e 23.06.2000
Terminal strip X3:
digital outputs
Terminal strip X1:
output (status) relays option
Fig. 5-7: BINOS® 100 F, Data line connections
Inside view from front (detail, without front door)
5 - 15
Page 92
PREPARATION
ADDITIONAL HINTS TO BINOS® 100 F (FIELD HOUSING)
Rosemount Analytical
Terminal strip X2:
analog outputs
Terminal strip X3:
digital outputs
X2 Analog out
X3 Digital out
X1 Status
Alternatively terminal strips
(connection of mains line)
PE
N L
N L
PE
Powerline filter
(connection of L and N of mains line)
5 - 16
Terminal strip X1:
output (status) relays option
Ground conductor pin
(PE mains line)
Fig. 5-8: BINOS® 100 F, Connection data lines / mains line
(inside view, left side panel)
ETC00265(1) Series 100 e 23.06.2000
Page 93
Rosemount Analytical
SWITCHING ON
6. Switching On
6.1 General
Be sure to observe the safety precautions and warnings !
Once the instrument has been correctly assembled and installed in accordance with the general instructions given in section 5., the equipment is ready for operation. The equipment is switched on by providing the required voltage.
GENERAL
The presence of the supply voltage will be indicated by the illumination of the LED displays. Upon connection of the supply voltage, the analyzer will perform a self diagnostic test routine. First the actual program version will be shown.
(analyzers without field bus) or (analyzers with field bus)
Finally either concentration values or error messages will be displayed If as a result of a battery fault the default values were charged, this will be shown by a flushing batt. This message will disappear after depressing any key.
Analyzer warming-up takes about 15 to 50 minutes, depending on the installed detectors !
Before starting an analysis, however, the following should be performed:
entry of the desired system parameters, calibration of the analyzer.
Note:
The "X’s" shown in the display indicate a number or combinations of numbers.
ETC00265(1) Series 100 e 23.06.2000
6 - 1
Page 94
SWITCHING ON
CROSS COMP
Rosemount Analytical
24 V DC SUPPLY
6.2 24 V DC Supply
The analyzers of series 100 (with except of BINOS® 100 F and BINOS® 100 2M with internal power supply) are specified for an operating voltage of 24 V DC. This voltage is be connected via a 3­pole XLR flange (male).
The DC supply voltage is to be provided by Option VSE 2000, UPS 01 T, DP 157, SL5, SL10 or equivalent power supply.
Connect power supply and analyzer (Fig. 6-1, Plug 24 V DC).
Verify correct polarity before operation (Fig. 29-1) !
Connect mains line and power supply.
Be sure to observe the safety precautions and warnings given by manufacturer of power supply !
X1 OUTPUT
INTERFACE
X2 OUTPUT
24 V
max. W
.
3
12+ 1
X3 OUTPUT
Plug 24 V DC
Input
6 - 2
MADE IN GERMANY
Fig. 6-1a: BINOS® 100 (M), OXYNOS® 100 (EO2), HYDROS® 100, Voltage supply
ETC00265(1) Series 100 e 23.06.2000
Page 95
Rosemount Analytical
SWITCHING ON
24 V DC SUPPLY
Plug 24 V DC
Input
X2 OUTPUT
INTERFACE
X1 OUTPUT
24 V
max. 40 W
2+
3
Fig. 6-1b: OXYNOS® 100 (PO2), Voltage supply
IN
OUT
K1 K2 K1 K2
DIGITAL IN
IN OUT
X1 OUTPUT
IN/OUT
X2 OUTPUT X3 OUTPUT
DURCHFLUSS MAX.1L/MIN
ANALOG IN
SAMPLE
ZERO
SPAN 1
SPAN 2
1
3
OUT
24V
max.120W
2
V+ V-
E1 E2
E3 E4
E5 E6 E7
FB+ FB­FB+
FB-
Plug 24 V DC
Fig. 6-1c: BINOS® 100 2M (version B) with external power supply, Voltage supply
Input
ETC00265(1) Series 100 e 23.06.2000
6 - 3
Page 96
SWITCHING ON
230/120 V AC SUPPLY
Rosemount Analytical
6.3 230/120 V AC Supply
6.3.1 BINOS® 100 2M
The analyzer BINOS® 100 2M is equipped optional with an internal “autoranging” power supply and is specified for an operating voltage of 230 V AC or 120 V AC resp., 47-63 Hz.
For supply of external components / analyzers there is built in a 24 VDC outlet (a 3-pole XLR flange, female, max. 2 A).
Connect internal power supply and external components (Fig. 6-2, 24 V DC out).
24 VDC supply to external components/analyzers with the internal power supply of BINOS® 100 2M requires a fuse to be connected in series to the consumer which limits the current consumption to max. 2 A !
Verify correct polarity for 24 V DC supply to external components before operation (Section 29.) !
Connect mains line and internal power supply (Fig. 6-2, V AC in).
Verify beforehand that the line voltage stated on the power supply is in accordance with that of your power supply line ! The socket outlet shall be installed near the equipment.
6 - 4
ETC00265(1) Series 100 e 23.06.2000
Page 97
Rosemount Analytical
SWITCHING ON
230/120 V AC SUPPLY
Socket 24 V DC out
Input
230/120 V AC
POWER
Fig. 6-2: BINOS® 100 2M (version A) shown with internal power supply, Voltage supply
6.3.2 BINOS® 100 F
The analyzer is specified for an operating voltage of 230 V AC or 120 V AC resp., 47-63 Hz.. Built-in power supply (manual switch between 230/120 VAC) is either power supply of type SL5 or of type SL10.
Once the instrument has been correctly assembled and installed in accordance with the general instructions given in section 5., 5.3.3 and 5.4, the equipment is ready for operation. The analyzer is switched on by providing the required voltage.
ETC00265(1) Series 100 e 23.06.2000
6 - 5
Page 98
SWITCHING ON
Rosemount Analytical
6 - 6
ETC00265(1) Series 100 e 23.06.2000
Page 99
Rosemount Analytical
KEY FUNCTIONS
7. Key Functions
The operation and programming of the analyzers is performed using the membrane - type keypad with its four keys as standard (see Fig. 1-1 and 1-2, Item 4 to 7). Additional the BINOS100® 2M/F analyzers can be equipped with a fithth key “PUMP” (see Fig. 1-
2. Item 8 and chapter 7.4).
BINOS® 100 F for use in hazardous areas is equipped with a touch screen keypad without fithth key for gas sampling pump.
Operator guidance prompts will appear on the 4 digit LED displays.
Battery buffering of the stored parameters prevents their loss in the absense of voltage supply.
ETC00265(1) Series 100 e 23.06.2000
7 - 1
Page 100
KEY FUNCTIONS
FUNCTION
Rosemount Analytical
7.1 FUNCTION
Pressing this key (Fig. 1-1 and 1-2, Item 3) addresses the individual analyzer functions in sequence. Merely addressing an analyzer function will not initiate an analyzer action or operation. The analyzer will continue to perform analysis until the ENTER key is pressed.
The following analyzer functions and their sequences (see also Fig. 7-1) are shown:
Zeroing channel 1
Zeroing channel 2
Spanning channel 1
Spanning channel 2
Interval Time for automatic Zeroing
Interval Time for automatic Spanning
Entry of concentration limits
Only in combination of internal solenoid valves or digital outputs and external solenoid valves , and
if Auto = 1
7 - 2
Entry of system parameters.
Entry of serial interface parameters
Only with Option RS 232 C/485 Serial Interface
ETC00265(1) Series 100 e 23.06.2000
Loading...