Rosemount BINOS 100 Series Analyzers including OXYNOS 100 and HYDROS 100-Software Version 5.1x-3rd Ed. Manuals & Guides

Rosemount Analytical
Microprocessor - Controlled Gas Analyzers
Operation Manual
Series 100
BINOS® 100
BINOS® 100 F
BINOS® 100 M
BINOS® 100 2M
OXYNOS® 100
HYDROS® 100
Software Version 5.1x
3. Edition 01/2003
ETC00781(3) Series 100 e 01/2003
Catalog - No.: ETC00781
Rosemount Analytical
This Maintenance & Operation Manual includes information about the operation of the instrument as well as additional indications and notes regarding maintenance, troubleshooting and repair.
Use qualified personnel only for troubleshooting, component replacement and internal adjustments.
Fisher-Rosemount MFG GmbH & Co. OHG assumes no liability for any omissions or errors in this
manual. Any liability for direct or consequential damages - which might occur in connection with the delivery or the use of this manual - is definitely excluded to the extend permitted by applicable law.
This instrument has left the factory in accordance to all applicable safety regulations. To maintain this operating condition, the user must strictly follow the instructions and consider the
warnings in this manual or provided on the instrument.
Misprints and alterations reserved
©
2001 by FISHER-ROSEMOUNT MFG GmbH & Co. OHG (PAD/ETC)
1. Edition: 10/2001 3. Edition: 01/2003
2. Edition: 11/2002
Read this operation manual carefully before attempting to operate the analyzer ! For expedient handling of reports of defects, please include the model and serial number which are printed on the instrument identity plate.
Look for the error check list too (see chapter 32. of this manual)
Fisher - Rosemount MFG GmbH & Co. OHG
European Technology Center
Process Analytic Division
Industriestrasse 1
D - 63594 Hasselroth • Germany
Phone + 49 (6055) 884-0
Telefax + 49 (6055) 884-209
Internet: http://www.processanalytic.com
ETC00781(3) Series 100 e 01/2003
SERIAL INTERFACE (OPTION)
NUMERICAL REPRESENTATION / BLOCK PARITY CHECK
Rosemount Analytical
12.4.6 Numerical Representations
Telegrams may contain integers or real numbers. The formats for these numbers are subject to the following restrictions.
Integers: - maximum value = 216 - 1
- positive numbers only accepted
- no decimal points allowed
Real: - maximum of 6 digits accepted
- no alphabetic characters (e.g. 2.2E-6) allowed
- analyzer output is 6 - digit real numbers
12.4.7 Block Parity Check
The master control computer may insert a message length parity byte into telegrams. These invariably consist of two characters.
The message - length parity byte is the cumulatively EXCLUSIVE - OR correlation of all previously transmitted characters of the telegram line. Representation is in hexadecimal format. For example, if the decimal value should be decimal 13, this will be represented by the two characters “OD”, i.e., 030H and 044H.
The verification procedure may be enabled or disabled at the analyzer (see chapter 12.3.4).
12 - 10
ETC00781(3) Series 100 e 01/2003
Rosemount Analytical
12.5 Instruction Syntax
Code definitions:
RP: receive parameters analyzer is accepting values SP: send parameters analyzer is sending values RI: receive instructions k: channel numbers 0 to 2 m: range number w: value
SERIAL INTERFACE (OPTION)
INSTRUCTION SYNTAX
<ID>: analyzer ID - no. for RS 485 mode of operation; follows start character LPB: message - length parity byte <CR>: terminate character
Receipt of any instruction codes not listed in the following section will be acknowledged by transmittal of status code 106. Future expansions will make use of code numbers not currently in use.
ETC00781(3) Series 100 e 01/2003
12 - 11
SERIAL INTERFACE (OPTION)
INSTRUCTION LIST
12.5.1 Instruction Listing
Instruction syntax: Instruction description:
$ID;001;k;LPB<CR> RI stand-by status (all valves closed)
$ID;002;k;LPB<CR> RI sample gas valve open
$ID;003;k;LPB<CR> RI zero gas valve open
$ID;005;m;k;LPB<CR> RI span gas valve open
$ID;006;LPB<CR> RI on - line status
$ID;007;LPB<CR> RI off - line status
$ID;008;LPB<CR>
*)
$ID;009;w;LPB<CR>
SP pump status
*)
RI pump status
*)
*)
w=0: pump Off, w=1: pump On
Rosemount Analytical
$ID;011;m;k;LPB<CR> SP at full scale range
$ID;013;k;LPB<CR> SP t
$ID;014;w;k;LPB<CR> RP t
(response time)
90
(response time)
90
$ID;017;k;LPB<CR> SP preflushing period (zero gas)
$ID;018;w;k;LPB<CR> RP preflushing period (zero gas)
$ID;019;k;LPB<CR> SP preflushing period (span gas)
$ID;020;w;k;LPB<CR> RP preflushing period (span gas)
$ID;023;k;LPB<CR> SP concentration
$ID;028;m;k;LPB<CR> SP span gas concentration
$ID;029;m;w;k;LPB<CR> RP span gas concentration
$ID;030;LPB<CR> SP status messages
$ID;031;t;LPB<CR> SP serial number (t=0, max. 10 characters)
channel identification (t=1: ch.1, t=2: ch. 2)
$ID;603;k;LPB<CR> SP gas component
$ID;604;k;LPB<CR> RI automatic zeroing
$ID;605;k;LPB<CR> RI automatic spanning
$ID;606;0;LPB<CR> RI automatic zeroing & spanning
$ID;626;LPB<CR> SP E-Prom / Software version
$ID;627;LPB<CR> SP failure message (possible error batt. is
clearing by read out)
$ID;645;0;LPB<CR> SP pressure value
$ID;646;w;LPB<CR> SP solenoid valve status (w=1: samplegas valve,
w=2: zerogas valve, w=4: spangas 1, w=8: spangas 2)
*) BINOS® 100 2M/F from program version 5.0 is required
12 - 12
ETC00781(3) Series 100 e 01/2003
Rosemount Analytical
SERIAL INTERFACE (OPTION)
RESPONSE TELEGRAMS
12.5.2 Response Telegrams
Response telegrams follow with the same syntax as the appropiate (SP-) commands (see 12.5).
12.5.2.1Status Messages (Instruction 030)
The response telegram for instruction
“$ID;030;LPB<CR> SP Status messages”
shows as follows:
$ID;030;a;b;c;LPB<CR>
This means:
a: OK-Status 0 = Relay without power 1 = Relay active
b: Value of variable “calibration” 0 = Relay without power >0 = Relay active
b Meaning
0 No Calibration 1 Zeroing channel 1 2 Zeroing channel 2 3 Zeroing channel 1 + 2 4 Spanning channel 1 5 Spanning channel 2 6 Spanning channel 1 + 2 7 Spanning channel 1 first, then channel 2 8 reserved 9 reserved 10 Waiting for flushing time and t90 response time
c: Gas sampling pump
*)
BINOS® 100 2M/F from program version 5.0 is required
ETC00781(3) Series 100 e 01/2003
*)
0 = Relay without power 1 = Relay active
(pump OFF) (pump ON)
12 - 13
SERIAL INTERFACE (OPTION)
RESPONSE TELEGRAMS
Rosemount Analytical
12.5.2.2 Pump Status (Instruction 008)
*)
The response telegram for instruction
“$ID;008;LPB<CR> SP Pump status”
shows as follows:
$ID;008;w;LPB<CR>
with w = pump status 0: Pump Off 1: Pump On
12 - 14
*) BINOS® 100 2M/F from program version 5.0 is required
ETC00781(3) Series 100 e 01/2003