Fisher-Rosemount GmbH & Co assumes no liability for any omissions or errors in this manual.
Any liability for direct or indirect damages, which might occur in connection with the deliv ery or the use of
this manual, is expressly e xcluded to the extend permitted by applicable law.
This instrument has left the works in good order according to safety regulations.
T o maintain this operating condition, the user must strictly follow the instructions and consider the warnings
in this manual or provided on the instrument.
T roubleshooting, component replacement and internal adjustments must be made by qualified
service personnel only.
According to the report No. “IBS/PFG-No. 41300392” about the approval of “DMT - Gesellschaft für
Forschung und Prüfung mbh, F achstelle für Sicherheit - Prüfstelle für Grubenbewetterung”, the stationary
®
gas analyzer BINOS
and of carbon dioxide between 0 and 80 % CO2. The system control with serial interfaces as described
CH
4
in this operation manual have not been subject to the DMT-approval. The DMT -in vestigation has no validity
for BINOS
Till completion of a separate operation manual of BINOS® 100 F, these supplements
to operation manual BINOS® 100 / OXYNOS® 100 are valid !
®
100 M (combined NDIR / Oxygen measurement).
100 is suitable for measuring the concentrations of methane between 0 and 80 %
According to the report No. 95CU054/B about the appro val of “TÜV Nord mbH” the gas analyzer BINOS
100(M) is suitable for CO measuring according to TI Air and 13. BlmSchV (large furnaces order) and for
measuring according to TI Air , 13th BlmSchV (large furnaces order) and 17th BlmSchV (incineration).
Read this operation manual carefully before attempting to operate the analyzer !
For expedient handling of reports of defects , please include the model and serial number which
can be read on the instrument identity plate.
Look for the error check list please too (see Item 29. of this manual)
Fisher - Rosemount GmbH & Co. (ETC)
®
Industriestrasse 1
D - 63594 Hasselroth • Germany
Phone + 49 (6055) 884-0
Telefax + 49 (6055) 884-209
Internet: http://www.processanalytic.com
90003786(1) BINOS® 100 F e [4.11] 23.06.98
Rosemount Analytical
SAFETY SUMMARY
Safety Summary
Outside and/or inside ML T or at operation manual resp. different symbols gives you a hint to special
sources of danger.
Source of danger !
See Operation Manual!
High V oltage !
GENERAL
Electrostatic Discharge (ESD) !
Explosives !
Hot components !
T oxic !
Risk to health !
MLT specific notes for the user !
In operation manual we will give partly additional informations to these symbols.
Strictly follow these instructions please !
90003786(1) BINOS® 100 F e [4.11] 23.06.98
S - 1
SAFETY SUMMARY
GENERAL
Rosemount Analytical
1.General
◆The following gener al safety precautions must be observed during all phases of operation,
service and repair of this instrument !
Failure to comply with these precautions or with specific w arnings elsewhere in this manual
violates safety standards of design, manufacture and intended use of this instrument !
Failure to comply with these precautions may lead to personal injury and damage to this
instrument !
◆Fisher-Rosemount GmbH & Co. does not take responsibility (liability) for the customer´s
failure to comply with these requirements !
◆Do not attempt internal service or adjustment unless other person, capable of rendering first
aid and resuscitation, is present !
◆Because of the danger of introducing additional hazards, do not perform any unauthorized
modification to the instrument !
Return the instrument to a Fisher-Rosemount Sales and Service office for service or repair
to ensure that safety features are maintained !
◆Instruments which appear damaged or defective should be made inoperativ e and secured
against unintended operation until they can be repaired by qualified service personnel.
S - 2
90003786(1) BINOS® 100 F e [4.11] 23.06.98
Rosemount Analytical
SAFETY SUMMARY
GENERAL
Operating personnel must not remove instrument covers !
Component replacement and internal adjustments must be made by qualified
service personnel only !
Read all operation manuals before attempting to operate with the instrument !
Be sure to observe the additional notes, safety precautions and warnings given
in the individual operation manuals !
Do not operate the instrument in the presence of flammable gases or explosive
atmosphere without supplementary protective measures !
At photometer or heated components there could be exist hot components !
Lift or carry housing with at least 2 persons because of the high weight of field
housing BINOS® 100 F (approx. 30 - 35 kg).
For easy transport use a suitable cart.
Guarantee, that the PG fittings together with pass through cables are hermetic
to be in agreement with protection class IP 65 (according to DIN standard
40050). The permissible outside diameters of the cables are 7 to 12 mm !
90003786(1) BINOS® 100 F e [4.11] 23.06.98
S - 3
SAFETY SUMMARY
GASES AND GAS CONDITIONING (SAMPLE HANDLING)
2.Gases and Gas Conditioning (Sample Handling)
Be sure to observe the safety regulations for the respective gases
(sample gas and test gases / span gases) and the gas bottles !
Inflammable or explosiv e gas mixtures must not be purged into the instrument
without supplementary protective measures !
To a void a danger to the operators by explosiv e, toxic or unhealthy gas
components, first purge the gas lines with ambient air or nitrogen (N2) before
cleaning or exchange parts of the gas paths.
Rosemount Analytical
S - 4
90003786(1) BINOS® 100 F e [4.11] 23.06.98
Rosemount Analytical
3.Supply Voltage
V erify whether the line voltage stated on the instrument ore power supply agrees
with that of your mains line!
Be sure to observe the safety precautions and warnings given by
manufacturer of power supply !
◆BINOS® 100 F is a Safety Class 1 instrument
SAFETY SUMMARY
SUPPLY VOLTAGE
Tha equipment is provided with a protective earth terminal.
T o prevent shoc k hazard, the instrument chassis and cabinet must be connected
to an electrical ground. The instrument must be connected to the AC pow er
supply mains through a three-conductor power cable, with the third wire firmly
connected to an electrical ground (safety ground) at the power outlet.
If the instrument is to be energized via an external power supply , that goes for the
power supply too .
Any interruption of the protective (grounding) conductor or disconnection of the
protective earth terminal will cause a potential shock hazard that could result in
personal injury. Deliberate disconnection is inadmissible / prohibited !
The BINOS® 100 F (field housing) has no switch with disconnect function.
The customer has to provide a switch or circuit break er into his installation. This
switch has to be installed near by analyz er, must be easily attainable f or operator
and has to be charaterized as disconnector for analyzer.
Cables to external dataprocessing have to be double-isolated against mains
voltage f or BINOS® 100 F !
If there is no guarantee, install lines internal that they hav e a distance of at least
5 mm to mains voltage lines.
This distance have to be guarantee permanent (e.g. via cable holder) !
90003786(1) BINOS® 100 F e [4.11] 23.06.98
S - 5
SAFETY SUMMARY
ANALYZER SPECIFIC NOTES FOR USER
4.Analyzer specific notes for the user
The installation site for the instrument has to be dry and remain above freezing
point at all times.
The instrument must be exposed neither to direct sunlight nor to strong sources
of heat. Be sure to observe the permissible ambient temperature !
For outdoor sites, we recommend to install the instrument in a protective cabinet.
At least, the instrument has to be protected against rain (e.g., shelter).
For Installation site please take care about safe way to reach the analyzer.
Rosemount Analytical
Do not interchange gas inlets and gas outlets !
All gases have to be supplied to the analyz er as conditioned gases !
When the instrument is used with corrosive gases, it is to be v erified that there
are no gas components which may damage the gas path components.
The exhaust gas lines have to be mounted in a declining, descending,
pressureless and frost-free and according to the valid emission legislation !
In case it is necessary to open the gas paths, close the analyzers
gas connections with PVC caps immediatly !
Use only from our factory optional delivered cables or equivalent shielded cab les
to be in agreement with the CE - conformity.
The customer has to guarantee, that the shield is be connected bothsided.
Shield and connectors housing have to be connected conductiv e.
Sub.-min.-D-plugs/soc kets have to be screw ed to the analyzer.
S - 6
90003786(1) BINOS® 100 F e [4.11] 23.06.98
Rosemount Analytical
5.Additional notes for service / maintenance
Operating personnel must not remove instrument covers !
Component replacement and internal adjustments must be made by qualified
service personnel only !
Always disconnect power, discharge circuits and remove external voltage
sources before troub leshooting, repair or replacement of any component !
Any work inside the instrument without switching off the power must be
performed by a specialist, who is familiar with the related danger, only !
SAFETY SUMMARY
ADDITIONAL NOTES FOR SERVICE / MAINTENANCE
In case of exchanging fuses the customer has to be certain that fuses of specified
type and rated current are used. It is prohibited to use repaired fuses or def ective
fuse holders or to short-circuit fuse carriers (fire hazard).
At photometer or heated parts there could be exist hot components !
90003786(1) BINOS® 100 F e [4.11] 23.06.98
S - 7
SAFETY SUMMARY
ELECTROSTATIC DISCHARGE
Rosemount Analytical
5.1Electrostatic Discharge
The electronic parts of the analyzer can be irreparably damaged if exposed to electrostatic
discharge (ESD).
The instrument is ESD protected when the covers have been secured and safety precautions
observed. When the housing is open, the internal components are not ESD protected anymore.
Although the electronic parts are reasonable safe to handle, you should be a ware of the following
considerations:
Best ESD example is when you w alked across a carpet and then touched an electrical grounded
metal doorknob. The tiny spark which has jumped is the result of electrostatic discharge (ESD).
You prevent ESD by doing the following:
Remove the charge from your body before opening the housing and maintain during work with
opened housing, that no electrostatic charge can be built up.
Ideally you are opening the housing and working at an ESD - protecting workstation.
Here you can wear a wrist trap.
However, if you do not have such a workstation, be sure to do the follo wing procedure exactly:
Discharge the electric charge from your body. Do this by touching a device that is grounded
electrically (any device that has a three - prong plug is grounded electrically when it is plugged into
a power receptacle).
This should be done several times during the operation with opened housing (especially after
leaving the service site because the movement on a low conducting floors or in the air might cause
additional ESDs).
S - 8
90003786(1) BINOS® 100 F e [4.11] 23.06.98
Rosemount Analytical
PREPARATION
INSTALLATION SITE
5.Preparation
Please check the packing and its contents immediately upon arrival.
If any item is damageg or lost you are kindly requested to notify the forwarder to undertake a
damage survey and report the loss or damage to us immediately.
5.1Installation Site
The analyzer must not operate in explosiv e atmosphere without supplementary
protective measures !
The installation site for the analyzer has to be dry and remain above freezing point
at all times. The analyzer must be e xposed neither to direct sunlight nor to strong
sources of heat.
Be sure to observe the permissible ambient temperatures (c.f. Item 20: T echnical
Data).
For Installation site please take care about safe way to reach the analyzer.
The analyzer has to be installed as near as possible to the sample point, in order to av oid low
response time caused by long sample gas lines.
In order to decrease the response time, a sample gas pump with a matching high pumping rate
may be used. Eventually, the ML T has to be operated in the bypass mode or by an o verflow valv e
to prevent too high flo w and too high pressure (Fig. 5-1).
Exhaust
Overpressure valve
Gas sampling pump
90003786(1) BINOS® 100 F e [4.11] 23.06.98
Analyzer
Filter
Fig. 5-1: BINOS® 100 F, Bypass installation
Flow meter
Exhaust
5 - 1
PREPARATION
GAS CONDITIONING (SAMPLE HANDLING)
Rosemount Analytical
5.2Gas Conditioning (Sample Handling)
The conditioning of the sample gas is of greatest importance for the successful operation of any
analyzer according to extractiv e method.
All gases have to be supplied to the analyz er as conditioned gases !
When the instrument is used with corrosive gases, it is to be v erified that there
are no gas components which may damage the gas path components.
The gas has to fullfil the following conditions:
It must be
❏free of condensable constituents
❏free of dust
❏free of aggressive constituents which are not compatible with the material of the gas
paths.
❏have temperatures and pressures which are within the specifications stated in “Technical
Data” of this manual.
Inflammable or explosive gas mixtures may not be introduced into the MLT
without supplementary protective measures !
When analysing vapours, the dewpoint of the sample gas has to be at least 10 °C below the
ambient temperature in order to avoid the precipitation of condensate in the gas paths .
Suitable gas conditionning hardware may be supplied or recommended for specific analytical
problems and operating conditions.
5.2.1Gas Flow
The gas flow rate should be within the range 0.2 l/min to maxi. 1.5 l/min !
A constant flow rate of about 1 l/min is recommended.
The gas flow rate for analyzer with paramagnetic oxygen sensor is
allowed to max. 1.0 l/min !
5 - 2
90003786(1) BINOS® 100 F e [4.11] 23.06.98
Rosemount Analytical
PREPARATION
GAS CONNECTIONS
5.3Gas Connections
The installed gas connections are specific to the different analyzers. The fittings are located on
the left bottom side (viewed from front) and are clearly marked:
IN = Gas inlet OUT = Gas outlet (exhaust)
For one-channel analyzer and dual-channel analyz ers tubed in series, only the 2 gas line fittings
for channel 1 are present. If the two channels are tube parallel, then all 4 fittings will be present.
Zero gas and span gas are introduced directly via the sample gas inlet. The test gas containers
have to be set up according to the current legislation.
Do not interchange gas inlets and gas outlets !
The exhaust gas lines have to be mounted in a declining, pressureless
and frost-free way and according to the valid emission legislation !
Be sure to observe the safety regulations for the respective gases
(sample gas and test gases / span gases) and the gas bottles !
Permissible gas pressure max. 1,500 hPa !
Wall mounting holder
Fig. 5-2: BINOS® 100 F, Gas connections (viewed from left side)
90003786(1) BINOS® 100 F e [4.11] 23.06.98
Front panel
Gas connections
5 - 3
PREPARATION
ADDITIONAL HINTS: WALL MOUNTING
Rosemount Analytical
5.4Additional Hints to Field Housing
5.4.1Wall Mounting
This housing of BINOS® 100 F is designed for wall mounting. For fastening points see Fig. 5-3
please.
operation
front panel
Fig. 5-3: Dimensional sketch / Drill drawing BINOS® 100 F [all dimensions in mm]
(optional electronics and photometer/sensors can be mounted into two separately housings)
Lift or carry housing with at least 2 persons because of the high weight of field
housing BINOS® 100 F (approx. 30 - 35 kg).
For easy transport use a suitable cart.
5 - 4
90003786(1) BINOS® 100 F e [4.11] 23.06.98
Rosemount Analytical
5.4.2 Electrical Connections
Be sure to observe the safety precautions and warnings !
Garantuee, that the PG fittings together with pass through cables are hermetic
to be in agreement with protection class IP 65 (according to DIN standard
40050). The permissible outside diameters of the cables are 7 to 12 mm !
a) Mains Supply
PREPARATION
ADDITIONAL HINTS: ELECTRICAL CONNECTIONS
The analyzer is specified for an operating voltage of 230 Vac or 120 Vac resp., 47-63 Hz..
Built-in power supply (manual switch between 230/120 Vac) is one power supply of type SL5.
❍ Opening of housing (front panel) (see Item 21.).
❍ Take mains line via PG fitting (Fig. 5-4) inside the housing.
Connect L and N to powerline filter (Fig. 5-6) via plug jacket (6,3x0,8 mm).
Connect PE via ring cable system to left ground conductor pin (Fig. 5-6).
Verify beforehand that the line voltage stated on the identification plate
(front door inside) agrees with that of your power supply line !
Verify that the position of input voltage switch of the power supply(s) agrees
with that of your power supply line (Fig. 24-2 and 5-5) !
The BINOS® 100 F (field housing) has no switch with disconnect function.
The customer has to provide a switch or circuit breaker into his installation. This
switch has to be installed near by analyzer, must be easily attainable for operator
and hase to be charaterized as disconnector for analyzer.
PG fittings
of data lines
PG fitting
of mains line
Wall mounting
holder
Fig. 5-4: BINOS® 100 F, PG fitting for lines (side view from left)
90003786(1) BINOS® 100 F e [4.11] 23.06.98
Front panel
Gas connections
5 - 5
PREPARATION
ADDITIONAL HINTS: ELECTRICAL CONNECTIONS
Rosemount Analytical
b)optional Data Lines
This are analog outputs, digital outputs, status output relays and serial interfaces.
❍Opening of housing (front panel) (see Item 21.).
❍Take lines via PG fittings (Fig. 5-4) inside the housing.
Connection is to be done to the respective terminal strips (Fig. 5-5 and 5-6).
Assignements of terminal strips is identically to assignment of connectors shown on page
26-1 of operation manual BINOS® / OXYNOS® 100 !
Cables to external dataprocessing have to be double-isolated against mains
voltage for analyz er !
If there is no garantuee, install lines internal that they hav e a distance of at least
5 mm to mains voltage lines.
This distance have to be garantuee permanent (e.g. via cable holder) !
Terminal strip X2:
analog outputs
Terminal strip Interface:
serial interface option
5 - 6
Power supply
Terminal strip X3:
digital outputs
Terminal strip X1:
output (status) relays option
Fig. 5-5: BINOS® 100 F, Data line connections
Inside view from front (without front door)
90003786(1) BINOS® 100 F e [4.11] 23.06.98
Rosemount Analytical
PREPARATION
ADDITIONAL HINTS: ELECTRICAL CONNECTIONS
Terminal strip X2:
analog outputs
Terminal strip X3:
digital outputs
Powerline filter
(connection of L and N of mains line)
output (status) relays option
Fig. 5-6: BINOS® 100 F, Connection data lines / mains line
90003786(1) BINOS® 100 F e [4.11] 23.06.98
Terminal strip X1:
(inside view, left side panel)
Ground conductor pin
(PE mains line)
5 - 7
PREPARATION
Rosemount Analytical
5 - 8
90003786(1) BINOS® 100 F e [4.11] 23.06.98
Rosemount Analytical
OPENING THE HOUSING
21.Opening the Housing
The housing must be opened for checking the electrical connections and for replacement or
cleaning of any of the components of the analyzer.
Be sure to observe Item 5. of the safety measures !
❍Disconnect all voltage supplies.
❍Open all 6 fasteners with a square key (Fig. 21-1)
❍Lift left side of front panel slightly and swing out front panel to the right side carefully.
Withdraw the photometer sliding carriage to the front carefully.
At photometer or heated parts there could be exist hot components !
Closing of the housing is performed in reverse order.
fastener for front panel
Operation front panel
Fig. 21-1: BINOS® 100 F (Field housing) (fastener for front panel)
90003786(1) BINOS® 100 F e [4.11] 23.06.98
fastener for front panel
21 - 1
OPENING THE HOUSING
Rosemount Analytical
21 - 2
90003786(1) BINOS® 100 F e [4.11] 23.06.98
Rosemount Analytical
TECHNICAL DATA
24. Technical Data
Certifications EN 50081-1, EN 50082-2, EN 61010-1
in preparation NAMUR, CSA NRTL/C, C-Tick, IS
Measuring ranges see order confirmation
Measuring components see order confirmation
24.1 Signal Outputs, Interfaces
analog [(optically isolated), 0 - 10 Vand 0 - 20 mA (R
≤ 500 Ω)
B
Offset and final concentration or
are free programmable] 2 - 10 Vand 4 - 20 mA (R
≤ 500 Ω),
B
adjustable via keyboard
Option: 0 (0,2) - 1 V and 0 (4) - 20 mA (R
digital, parallel (optically isolated) 2 threshold contacts per channel
Sample gas valve, Zero gas valve,
Span gas valve 1, Span gas valve 2
Dimensions65 x 125 x 103 mm (WxHxD)
InstallationMountable on DIN supporting rails TS35
90003786(1) BINOS® 100 F e [4.11] 23.06.98
Fig. 24-2: Sketch SL 5
24 - 3
TECHNICAL DATA
Rosemount Analytical
Needle valve (Option)
➤
12
% O
% O
2
1000 ppm
CO
para.Oxygen sensor
% O
chem.Oxygen sensor
paramagnetic
2
=
elektrochemical
2
=
3
4
1098765
Fig. A-1: BINOS
®
100 F, Operation Front Panel
1LED - Display (Channel 1)
2LED - Display (Channel 2)
3Function - LED for Options "Solenoid Valves / Gas Sampling Pump"
4Input setting control key DOWN
5Input setting control key UP
6ENTER key
7FUNCTION key
8Key for Option "Gas Sampling Pump"
9Flow Meter (Option)
10Fine Dust View Filter with Needle Valve (Option)
24 - 4
90003786(1) BINOS® 100 F e [4.11] 23.06.98