Fisher-Rosemount GmbH & Co assumes no liability for any omissions or errors in this man ual.
Any liability for direct or indirect damages, which might occur in connection with the deliv ery or the use of
this manual, is expressly excluded to the extend permitted by applicable law.
This instrument has left the works in good order according to safety regulations.
T o maintain this operating condition, the user must strictly follow the instructions and consider the warnings
in this manual or provided on the instrument.
Troubleshooting, component replacement and internal adjustments must be made by qualified
service personnel only .
Read this operation manual carefully before attempting to operate the analyz er !
For expedient handling of reports of defects , please include the model and serial number which
can be read on the instrument identity plate.
Look for the error check list, too (see Section 29 of this manual)
In this manual we have used the following safety symbols
to draw your attention to strictly follow these instructions !
◆The following general saf ety precautions must be observed during all phases of operation,
service and repair of this instrument !
Failure to comply with these precautions or with specific w arnings elsewhere in this manual
violates safety standards of design, manuf acture and intended use of this instrument !
Failure to comply with these precautions may lead to personal injury and damage to this
instrument !
◆Fisher-Rosemount GmbH & Co. assume no liability for the customer´s failure to comply with
these requirements !
◆Do not attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present !
◆Because of the danger of introducing additional hazards, do not perform any unauthorized
modification to the instrument !
Return the instrument to a Fisher-Rosemount Sales and Service office for service or repair
to ensure that safety f eatures are maintained !
◆Operating personnel must not remove instrument covers !
Component replacement and internal adjustments must be made by qualified service
personnel only !
◆Instruments which appear damaged or defective should be made inoperative and secured
against unintended operation until they can be repaired by qualified service personnel.
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
S - 1
Page 12
SAFETY SUMMARY
GENERAL / GASES AND GAS CONDITIONING
Read this operation manual carefully before attempting to operate with the
instrument !
Do not operate the instrument in the presence of flammable gases, explosive
atmosphere or furnes without supplementary protective measures !
The installation site for the instrument has to be dry and remain above freezingpoint at all times.
The instrument must be exposed neither to direct sunlight nor to strong sources
of heat. Be sure to observe the permissible ambient temper ature !
For outdoor sites, we recommend to install the instrument in a protective cabinet.
At least, the instrument has to be protected against rain (e.g., shelter).
Rosemount Analytical
Due to the high temperatures of photometer or heated components there is a
danger of burns to the operators.
2.Gases and Gas Conditioning (Sample Handling)
Do not interchange gas inlets and gas outlets !
All gases have to be supplied to the system as conditioned gases !
When the instrument is used with corrosive gases, it is to be verified that there
are no gas components which may damage the gas path components.
The exhaust gas lines hav e to be mounted in a dec lining, pressureless
and frost-free way and according to the valid emission legislation !
Be sure to observe the safety regulations for the respective gases
(sample gas and test gases / span gases) and the gas bottles !
S - 2
Inflammable or explosiv e gas mixtures must not be purged into the instrument
without supplementary protective measures !
To avoid a danger to the operators by explosiv e, toxic or unhealthy gas
components, first purge the gas lines with ambient air or nitrogen (N2) before
cleaning or exchange parts of the gas paths.
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 13
Rosemount Analytical
SAFETY SUMMARY
SUPPLY VOLTAGE
3.Supply Voltage
◆Verify correct polarity for 24 V DC operation !
◆This product is a Safety Class 1 instrument (provided with a protective earth terminal).
To prevent shock hazard, the instrument chassis and cabinet must be connected to an
electrical ground. The instrument must be connected to the AC power supply mains through
a three-conductor power cable , with the third wire firmly connected to an electrical ground
(safety ground) at the po wer outlet. If the instrument is to be energized via an external power
supply, that goes for the po wer supply, too.
Any interruption of the protective (grounding) conductor or disconnection of the protectiv e
earth terminal will cause a potential shock hazard that could result in personal injury.
Deliberate disconnection is inadmissible / prohibited !
◆Use only power supply VSE 2000, UPS 01 T or equiv alent po wer supplies to be in
agreement with the CE conformity.
◆In case of exchanging fuses the customer has to be certain that fuses of specified type and
rated current are used. It is prohibited to use repaired fuses or def ective fuse holders or to
short-circuit fuse carriers (fire hazard).
◆Always disconnect power, discharge circuits and remove external voltage sources before
troubleshooting, repair or replacement of any component !
Any work inside the instrument without switching off the power must be
performed only by a specialist, who is familiar with the related danger !
4.Connection Cables
◆Use only from our factory optionally delivered cables or equiv alent shielded cables to be in
agreement with the CE conformity.
The customer has to guarantee that the shield is being connected bothsided.
◆By using optionally delivered terminal strip adapters the analyzer might not be in agreement
with the CE conformity. In this case CE conformity is to be declared by customer as
“manufacturer of system”.
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
S - 3
Page 14
SAFETY SUMMARY
ELECTROSTATIC DISCHARGE
Rosemount Analytical
5.Electrostatic Discharge
The electronic parts of the analyzer can be irreparably damaged if exposed to electrostatic
discharge (ESD).
The instrument is ESD protected when the covers have been secured and safety precautions
observed. When the housing is open, the internal components are not ESD protected any more.
Although the electronic parts are reasonably safe to handle, you should be aw are of the following
considerations:
The best example for ESD is when y ou walk across a carpet then touch an electrically grounded
metal doorknob: the tiny spark which jumps is a result of electrostatic discharge (ESD). You can
prevent ESD b y removing the charge from your body before removing the analyzer´s housing and
ensuring that no ESD can be built up while working with opened housing.
Ideally , the analyz er should be opened and serviced in an ESD protecting workstation: here y ou
can wear a wrist trap. Should there be no such workstation available, be sure to carry out the
following procedure:
Discharge the electric charge from your body. Do this by touching a device that is electrically
grounded (any device that has a three - prong plug is electrically grounded when it is plugged into
a power receptacle).
This should be done several times during the operation with opened housing (especially after
leaving the service site because the movement on low conducting floors or in the air might cause
additional ESDs).
S - 4
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 15
Rosemount Analytical
FRONT PANEL / REAR PANEL
1.Setup
The analyzer is incorporated in a 1/4 19" rack-mounting housing, 3 height units.
The optional table-top housing is fitted additionally with a carrying strap and rubber feet.
1.1Front Panel
The front panel (see Fig. A-1) includes the LED displays as well as all of the keys required for
SETUP
operating the analyzer.
1.2Rear Panel
The rear panel (Fig. A-2) includes:
❏the gas line fittings
❏the plug for the electrical supply input
❏the subminiature “D” mating socket for the analog signal outputs
❏the subminiature “D” plug for the digital outputs (measurement mode / measuring range)
❏the 4-pin LEMOSA socket (connection to external switch of measurement mode / measuring
range)
❏optionally the subminiature “D” mating socket for the RS 232 C / 485 interface
1.3Inside View
The inside view is shown in Fig. 1-1.
90002927(1) BINOS® 100 4P e [1.00] 17.06.97
1 - 1
Page 16
SETUP
INSIDE VIEW
Rosemount Analytical
Gas line fittings
Photometer with
gas detector
(CO2 channel)
Holding
device
Photometer with
pyroelectrical detector
(H2O channel)
1 - 2
Front panel
Fig. 1-1: Inside View BINOS® 100 4P
90002927(1) BINOS® 100 4P e [1.00] 17.06.97
Page 17
Rosemount Analytical
PHOTOMETER ASSEMBLY
2.Photometer Assembly
Depending on gas component and measuring range, different photometer assemblies may be
realized in BINOS® 100 4P. Optionally the photometer can be sealed against ambient air . In this
case all parts are sealed with O - rings.
The entire photometer assembly is mounted as a unit on the main circuit board (BKS) by means
of a bracket. The main circuit board is inserted into guide rails in the analyzer housing, to which
the front panel (membrane keypad) and the rear panel are mounted.
2.1Photometer
To enable both types of measurement - the standard absolute and the additional differential
measurement (while taking the base concentration level into consideration) - a special photometer
assembly has been developed (see Fig. 2-1).
The basic part of the photometer assembly is the chopper housing (01). The light source (thermal
radiator, 02), the analysis cell (03), and the signal detection unit [including a filter cell (04) and a
detector (05/07)] are all mounted on this chopper housing.
The chopper housing also incorporates the interference filters for the selection of spectral
bandpass ranges from the broadband emission of the light sources.
Between the two halves of the chopper housing (03), which are sealed together with an O-ring,
there is the chopper blade which is driven b y a stepping motor. Both the chopper housing and the
motor encapsulation are hermetically sealed against ambient in order to prevent any gas from
entering, such as atmospheric CO2, which could lead to background absorptivity (preabsorption)
thus causing drift effects. An absorber pro vides f or constant remov al of any traces of CO2 which
may enter the interior of the chopper housing via diffusion.
In addition the chopper housing incorporates a photoelectric gate for providing a reference signal
for the phase angle of the chopper blade, and a temperature sensor (10) for monitoring
continuously the temperature of the photometer assembly . The information thus obtained is used
by the signal processing electronics to compensate thermal effects.
90002927(1) BINOS® 100 4P e [1.00] 17.06.97
2 - 1
Page 18
PHOTOMETER ASSEMBLY
Rosemount Analytical
Each analysis cell is divided in the middle by a partition wall in two halves: the measurement and
the reference side. These sides are sealed at both ends by CaF2 windows. This allows the
differential measurement to be carried out. The sample gas passes through the measurement side
while the reference side is filled with either nitrogen or another ref erence gas .
The filter cell (04) consists of a cone which optimally adjusts the radiation cross section of the
analysis cell to the detector surface.
2.2Detectors
BINOS® 100 4P is equipped with two different detectors. The reason for this solution is the
dependence on temperature of an H2O-filled gas detector (dew point undershoot). Therefore , for
CO2 a gas detector is used and for H2O, a pyroelectrical (solid-state) detector.
The gas detector (05) and its preamplifier are interconnected by an isolated cable. The preamplifier
(06) is mounted on the analysis cell (03).
The solid-state detector and its preamplifier are directly interconnected (07); both are mounted
on the filter cell (04).
The analyzers are non - dispersive infrared photometers (NDIR).
They detect the infrared light absorbed by the gas to be measured using the measurement of
selective radiation in a column of gas.
The wavelength of the absorption bands is characteristic of the type of gas being measured, while
the absorption strength gives a measure of the concentration of this gas component.
The detector receives signals succeeding in time. The difference between the absorption
measured in the measurement and reference cells gives a measure of the difference in concentration of the gas component being measured.
To enable this measurement, a gas detector is used for CO2 and a pyroelectrical one for
water vapour.
3.1Pyroelectrical Detector
The spectral transmittance curves of the interference filter used in the BINOS® 100 4P analyzer
and the spectral absorption of the gases CO2 and H2O are shown in Fig. 3-1. It can be seen that
the absorption bands of these gases each coincide with the bandpass of one of the interference
filters.
The signal is generated by a pyroelectrical detector. It records the incoming IR radiation; the
intensity of this radiation is reduced by the absorption of the gas at the according wavelengths. By
comparing the radiation of measurement side and reference side of the analysis cell, an alternating
voltage signal is developed. This signal results from cooling and heating of the pyroelectrical
material of the detector.
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
3 - 1
Page 22
MEASURING PRINCIPLE
Rosemount Analytical
100
Absorption Band
8090
CO2 (330 ppm)
6070
50
Transmittance [%]
40
010 20 30
4000450050006000
5500
Wavelength [nm]
H2O (20 °C)
6500700075008000
3 - 2
Fig. 3-1: Absorption Bands of Sample Gases and Transmittance of the
A thermal radiator generates the infrared radiation which passes first through a chopper wheel.
Then the radiation passes alternately through a filter cell and reaches the measurement and
reference sides of the analysis cell with equal intensity.
After passing another filter cell, the radiation reaches the gas detector.
The gas detector compares and evaluates the radiation coming from the measurement and
reference sides and converts them into voltage signals proportional to their intensity via a
preamplifier.
The detector consists of a gas-filled absorption and a compensation chamber, interconnected via
a flow channel.
Absorption chamber
Flow channel with
Microflow sensor
CaF2 Window
Gas intake connection
Compensation chamber
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Fig. 3-2: Principle Design of Gas Detector
3 - 3
Page 24
MEASURING PRINCIPLE
Rosemount Analytical
Standardly the detector is filled with the infrared active gas to be measured and is only sensitive
to this distinct gas with its characteristic absorption spectrum. The absorption chamber is sealed
with a window which is transparent for infrared radiation (usually CaF2, calcium fluoride).
When IR radiation passes through the reference side (purged with an unpressurized gas, which
does not absorb within a specified spectral range; in most cases nitrogen [N2] is used) of the
analysis cell into the detector, no absorption occurs. Thus the gas inside the absorption chamber
is heated, it expands and some of it passes through the flow channel into the compensation
chamber.
When the IR radiation passes through the measurement side (purged with sample gas) of the
analysis cell into the detector, a part of it is absorbed depending on the gas concentration. Then
the gas in the absorption chamber is heated less than in the case of the radiation coming from
reference side. The absorption chamber cools down, the pressure is reduced, and a part of the gas
passes through the flow channel into the measuring chamber.
The flow channel is designed so that it hardly impedes the gas flow by restriction. Due to the
radiation of the chopper wheel, the different radiation intensities lead to periodically flow pulses
within the detector.
The microflow sensor ev aluates and converts the pulses into electrical voltages. The electronics ,
which follow, evaluate the signals and convert them into the corresponding display format.
3 - 4
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 25
Rosemount Analytical
123
123
123
6
MEASURING PRINCIPLE
3.3Technique
The broad emission coming from two IR sources (in the case of dual channel analyzers)
passes a chopper blade, then passes through an optical filter and finally enters the analysis
cell. The light transmitted through these cells is focused by filter cells onto the detector. The
preamplified detector output signals are forwarded to the microprocessor electronics which
converts the analytical signals to results expressed directly in physical concentration units
such as Vol.-%, ppm, mg/Nm3 etc.
MOTOR
Light sourceLight source
Filter discFilter disc
Analysis cell measuring side
Analysis cell reference side
Filter cellFilter cell
Gas detector
23456789012345678234567890123456789012345
CO
2
Preamplifier
H2O
Analysis cell measuring side
Analysis cell reference side
Pyroelectrical detector
(solid-state detector)
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Chopper blade
Fig. 3-3: Principle Representation
3 - 5
Page 26
MEASURING PRINCIPLE
Rosemount Analytical
3 - 6
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 27
Rosemount Analytical
MAIN FEATURES
4.Main Features
◆1/4 19" housing, 3 HU
◆Measurement mode and measuring range switch; absolute or difference mode
◆Ground level compensation
◆4 - digit LED - measuring value display and operators prompting via this displays for each
measuring channel
◆The response time (t90 time) can be adjusted separately for each measuring channel
◆Plausibility checks
◆Temperature compensations
◆Interference compensation f or reducing disturbing effects due to e xtraneous absorption of
◆Digital interface for identification of measurement mode and measuring range
(max. 30 V DC / 30 mA, “Open Collector”, optically isolated)
◆Datalogger (option): data recording and output
◆RS 232 C/485 serial interface for data intercommunication with host computers (option)
◆Self - diagnostic procedures, plus maintenance and servicing support functions
◆Operator prompting for the avoidance of operator errors
90002927(1) BINOS® 100 4P e [1.00] 17.06.97
4 - 1
Page 28
MAIN FEATURES
Rosemount Analytical
4 - 2
90002927(1) BINOS® 100 4P e [1.00] 17.06.97
Page 29
Rosemount Analytical
PREPARATION
INSTALLATION SITE
5.Preparation
Please check the packing and its contents immediately upon arrival.
If any item is damageg or lost you are kindly requested to notify the forwarder to undertake a
damage survey and report the loss or damage to us immediately .
5.1Installation
The analyzer must not operate in explosive atmosphere without supplementary protective
measures !
The installation site for the analyzer has to be dry and remain above freezing point at all times.
The analyzer must be exposed neither to direct sunlight nor to strong sources of heat.
Be sure to observe the permissible ambient temperatures (cf. Section 24: Technical Data).
For outdoor installation, we recommend to install the analyzer in a protective cabinet. At least, the
analyzer has to be protected against rain (e.g., shelter).
The analyzer has to be installed as near as possible to the sample point in order to avoid low
response time caused by long sample gas lines.
In order to decrease the response time, a sample gas pump with a matching high pumping rate may
be used. Ev entually , the analyz er has to be operated in the bypass mode or by an o verflow v alve
to prevent too high flow and too high pressure (Fig. 5-1).
Exhaust
Bypass valve
Gas sampling pump
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Analyzer
Flow meter
Filter
Fig. 5-1: BINOS® 100 4P, Bypass Installation
Exhaust
5 - 1
Page 30
PREPARATION
GAS CONDITIONING (SAMPLE HANDLING)
Rosemount Analytical
5.2Gas Conditioning (Sample Handling)
The conditioning of the sample gas is of greatest importance for the successful operation of any
analyzer according to extractiv e method.
Only conditioned gas has to be supplied to the analyzer !
The gas has to fulfil the following conditions.
It must be:
❏free of condensable constituents,
❏free of dust,
❏free of aggressive constituents which are not compatible with the material of the gas
paths, and
❏have temperatures and pressures which are within the specifications stated in Section 24,
"Technical Data" of this manual.
Inflammable or explosive gas mixtures may not be intr oduced into the analyzer
without supplementary protective measures !
When analyzing vapors, the customer has to avoid the precipitation of condensate in the gas
paths.
Suitable gas conditioning hardware may be supplied or recommended for specific analytical
problems and operating conditions.
5.2.1Gas Flow Rate
The gas flow rate should be within the range 0.2 l/min to max. 1.5 l/min !
A constant flow rate of about 1 l/min is recommended.
5 - 2
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 31
Rosemount Analytical
PREPARATION
GAS CONNECTIONS
5.3Gas Connections
All the fittings for gas line connections are placed just on the rear panel of the analyzer and are
clearly marked (Fig. 5-2 and Fig. A-2):
IN = gas inlet
OUT = gas outlet
M = measurement side
R = reference side
Do not interchange gas inlets and gas outlets !
The exhaust gas lines have to be mounted in a declining, pressureless and frost-free way
and according to the valid emission legislation!
Zero gas, span gas and reference gas are introduced directly via the respectiv e gas inlet. The test
gas containers hav e to be set up according to the current legislation.
Be sure to observe the safety regulations f or the respective gases !
M1M1M2M2
R1R1R2R2
IN
OUTOUTIN
K1K2
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
X1 OUTPUT
INTERFACE
ABS./DIFF.
Fig. 5-2: Gas Connections BINOS® 100 4P
5 - 3
Page 32
PREPARATION
Rosemount Analytical
5 - 4
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 33
Rosemount Analytical
SWITCHING ON / SWITCHING OFF
SWITCHING ON
6 .Switching On / Switching Off
6.1Switching On
Once the analyzer has been correctly assembled and installed in accordance with the general
instructions of Section 5. “Preparation”, the analyzer is ready for operation.
The analyzer is specified for an operating v oltage of 24 V DC (+ 20 % / - 50 %).
Operation from 230 / 120 V AC requires the 24 V DC supply via VSE 2000 or equivalent power
supply.
IN
X1 OUTPUT
INTERFACE
X2 OUTPUT
OUTOUTIN
K1K2
24 VDC
MADE IN GERMANY
ABS./DIFF.
X3 OUTPUT
Plug
24 V DC
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Fig. 6-1: Supply Voltage BINOS® 100 4P
6 - 1
Page 34
SWITCHING ON / SWITCHING OFF
SWITCHING ON
Rosemount Analytical
❍Connect mains line and power supply.
Verify bef orehand that the line v oltage stated on the power suppl y agrees
with that of your power supply line !
❍Connect power supply and analyzer (Fig. 6-1, Plug 24 V DC).
Verify correct polarity bef ore operation !
The presence of the supply voltage will be indicated by the illumination of the LED displays.
Upon connection of the supply voltage, the analyzer will perform a self - diagnostic test routine.
First the actual program version will be shown.
Finally either concentration values or error messages will be displayed.
If as a result of a battery fault the default values have been charged, this will be shown by a flushing
“batt.”! This message will disappear after depressing any ke y.
Analyzer warming-up takes about 15 to 50 minutes depending on the
installed detectors !
Before starting an analysis, however, the following should be performed:
❏entry of the desired system parameters,
❏calibrate the analyzer.
NOTE:
The "X’s" sho wn in the displa y indicate a number or a combination of numbers.
6 - 2
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 35
Rosemount Analytical
SWITCHING ON / SWITCHING OFF
SWITCHING OFF
6.2Switching Off
Before s witching off the analyzer, we recommend first flushing the gas lines for about 5 minutes
with zeroing gas (N2) or adequate conditioned air. The full procedure for shutting down the analyzer
is as follows:
❍Admit zeroing gas at the gas inlet fittings.
❍Set the gas flow rate to allowa b le rate.
After 5 minutes have elapsed:
❍Shut off the zeroing gas supply.
❍Switch off the analyzer by disconnecting the v oltage supply.
❍Close all gas line fittings immediately.
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
6 - 3
Page 36
SWITCHING ON / SWITCHING OFF
Rosemount Analytical
6 - 4
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 37
Rosemount Analytical
KEY FUNCTIONS
7.Key Functions
The operation and programming of the analyzer is perf ormed using the membrane - type keypad
with its four keys (see Fig. A-1, Item 3 - 6).
Operator guidance prompts will appear on the 4 - digit LED - displays.
During power failure all parameters entered are secured by a battery-supplied buffer.
90002927(1) BINOS® 100 4P e [1.00] 17.06.97
7 - 1
Page 38
KEY FUNCTIONS
FUNCTION
Rosemount Analytical
7.1FUNCTION
By pressing this key (Fig. A-1, Item 3) you can roll through all active functions of the analyz er. The
selection of any of these functions does not lead to any action; the analyzer remains in the
measurement mode. The following analyz er functions and their sequences (see also Fig. 7-1) are
shown:
Zeroing channel 1
Zeroing channel 2
Recording of data in datalogger
Interval time for automatic recording of data in datalogger
Output of data from datalogger
Only with Option Serial Interface
and if “P.Set = Yes”
Deleting of data in datalogger
Time setting and display
7 - 2
Selection of measurement mode and enabling of parameters
Spanning channel 1
Only if “tYPE = AbS.”
Spanning channel 2
90002927(1) BINOS® 100 4P e [1.00] 17.06.97
Page 39
Rosemount Analytical
KEY FUNCTIONS
ENTER
Entry of system parameters.
Entry of serial interface parameters
Only if “P.SET= YES”
With Option “Serial Interface” only
7.2ENTER
The ENTER - key (Fig. A-1, Item 4) is used for the transf er of (keyed - in) numerical data to the
corresponding operating parameters and for the initiation of certain operations, such as zeroing
and spanning.
Several functions ma y be accessed only b y authorized persons. Theref ore these functions ha ve
been secured by codes. When a code-secured functions is being selected for the first time, while
rolling through a sequence of functions (starting at "0 -1"), and after pressing ENTER - key
will appear on the display.
Now the code must be entered. If the entered code is incorrect, the display remains unchanged,
and the process of entering the code starts again with "0".
When the code has been entered correctly, then you have access to the secured lev el.
This code has been set to the value “1” in our plant bef ore shipment.
90002927(1) BINOS® 100 4P e [1.00] 17.06.97
7 - 3
Page 40
7 - 4
ENTERFUNCTION
- - - -
- - - -
0..-1 0..-.1
0 -1
1
Fig. 7-1: BINOS
2
3
4
®
100 4P, Operating Function Matrix
5
Nominal
Calibrate
Value
0 -2 0..-2 0..-.2
Nominal
Calibrate
Value
d.in
tdin
d. Out
Zero Adjustment Channel 1
Zero Adjustment Channel 2
Data Input
Automatic Data Input
Interval-Time (min.); 0 = off
Data-Logger
Data Output
KEY FUNCTIONS
KEY FUNCTION OVERVIEW
6
7
8
9
90002927(1) BINOS
10
11
®
100 4P e [1.00] 17.06.97
PArA
12
d.Clr
t.SEt
SEL. tYPEGnd1 Gnd2 P.SEt A.Out
Measuring
S -1 CodES.-1S..-1 S..-.1
CodES -2S.-2 S..-2 S..-.2
CodESYS.-
CodESIP.On.-L
d.ClrCodE
Clear
MinHour
Set
Hours
Minutes
Reference
Channel 1
Mode
Actual
Value
Actual
Value
PrEStoL. d.OFF L -0
Pressure
Clear Data
Set
Setting Time
Reference
Channel 2
Nominal
Value
Nominal
Value
Air-
Tolerance
Id.No 232C
RS485-
On Line
Address
Parameter
display
Calibrate
Calibrate
Display
off
RS23 2C /
RS485
only with Option
"Serial Interface"
and with P.SEt =YES
By pressing this keys (Fig. A-1, Item 5 and 6) any parameter value entered may be changed.
Depending on which key is pressed, the value is either enhanced or diminished by one digit.
UPprevious v alue + 1 digit
DOWNprevious v alue - 1 digit
Example:
To setup the requested parameter value, the displayed characters can be adjusted by
pressing each of the key, respectively. Now the correct value can be acknowledged by
pressing ENTER on the secured level.
A suiting value range - which cannot be exceeded - belongs to each parameter. In addition
all entries are subjected to a plausibility check as added protection against operator errors.
The counting speed can be increased by pressing the <UP> or <DOWN> k ey continuously . When
rolling through the setup values from 0 to maximum or minimum value, the standard counting
speed is activated to make the entering process easier.
Approx. 60 - 120 seconds after pressing any key, the analyzer automatically
returns to the measuring mode.
90002927(1) BINOS® 100 4P e [1.00] 17.06.97
7 - 5
Page 42
KEY FUNCTIONS
Rosemount Analytical
7 - 6
90002927(1) BINOS® 100 4P e [1.00] 17.06.97
Page 43
Rosemount Analytical
ENABLING / DISABLING SYSTEM PARAMETER FUNCTION
ENTRY OF SYSTEM PARAMETERS
8.System Parameters
8.1Enabling / Disabling System Parameter Function
It might not be necessary to have access to system parameter / spanning function. The appearance
of this function may e ven disturb. Therefore, it is possible to skip this le v el completely.
Depress the key
until the textappears.
Depress the key
If the Code has not already been entered, the following display
will appear
Use the keysandto select the Code.
The display will show:
Depress the key
until the textappears.
Entry of “YES”:System parameter / spanning function is enabled
Entry of “NO”:System parameter / spanning function is disabled
Effect the entry by using
and
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
8 - 1
Page 44
ENTRY OF SYSTEM PARAMETERS
ENTRY OF SYSTEM PARAMETERS
8.2Entry of System Parameters
When the system parameters have to be entered or changed,
depress the key
until the displayappears.
Rosemount Analytical
Depress the key
If the Code has not already been entered, the following display
will appear
Use the keysto select the Code.
and then depress
8 - 2
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 45
Rosemount Analytical
PRESSURE CORRECTION / TOLERANCE CHECK
ENTRY OF SYSTEM PARAMETERS
8.2.1 Pressure Correction
To eliminate faulty measurements due to changes in barometric pressure or sample gas pressure,
the operator is offered the opportunity to enter the current pressure expressed in hPa (mbar) in a
range of 800 to 1300 hPa. The concentration values computed by the analyzer will then be
corrected to reflect the barometric pressure or sample gas pressure resp. entry.
The correction factor is calculated via following formula:
Correction factor =
1024 hPa
actual pressure
The entry is effected by using and
8.2.2 Tolerance Check
The tolerance function is for the activation and deactivation of the tolerance check procedure for
various calibration gases.
If the tolerance check procedure has been activated, the microprocessor will verify during
calibration procedures whether the used calibration gas shows a deviation of more than10 %
from measuring range of zero (zero - level) or more than 10 % of the nominal concentration value
entered resp. (span).
If this tolerance is exceeded, no calibration will be performed, and an error message will
appear (see Section 13).
Entry of “YES”: Tolerance check is deactivated.
Entry of “NO”: Tolerance check is activated.
Perform the entry by using and
To calculate the 10% value the measuring range of "AbS.2" is taken over from
"AbS.1" while being in absolute measuring range.
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
8 - 3
Page 46
ENTRY OF SYSTEM PARAMETERS
DISPLAY OFF / ANALOG SIGNAL OUTPUTS
Rosemount Analytical
8.2.3Display Off
If "YES" is entered, the DISPLAY will be deactivated about 1 to 2 minutes after the last key
depression. If an y key is depressed while the DISPLA Y is deactivated, all displa y elements will be
reactivated without any further operation being initiated.
Entry of “YES”:Display will be deactivated
Entry of “NO”:Display will not be deactivated
Entry is performed by using
followed by
8.2.4Analog Signal Outputs
The analog signal outputs (optically isolated) are brought out to the 9-pin subminiature Dconnector X2 on the analyzer rear panel.
Entry of “NO”:Output signal of 0 - 10 V (Option: 0 - 1 V) / 0 - 20 mA.
Entry of “YES”:Output signal of 2 - 10 V (Option: 0.2 - 1 V) / 4 - 20 mA (life zero mode)
Use the keys
andfor entry .
Note:
The begin of range concentration and the end of range concentration are freely programmable
(see Sections 9.2 and 10.3).
For type of voltage output (standard or option) look at order confirmation or identify plate.
8 - 4
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 47
Rosemount Analytical
ENTRY OF SYSTEM PARAMETERS
ANALOG SIGNAL OUTPUTS
M1M1M2M2
R1R1R2R2
Mating socket X 2
IN
X1 OUTPUT
INTERFACE
X2 OUTPUT
OUTOUTIN
K1K2
24 VDC
MADE IN GERMANY
ABS./DIFF.
X3 OUTPUT
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Fig. 8-1: Mating Socket X 2 (Analog Signal Outputs)
5
1
6
9
1
20 (2) - 10 V DC [Option: 0 (0,2) - 1 V DC], channel 1
30 (4) - 20 mA, channel 1 (R
40 (2) - 10 V DC [Option: 0 (0,2) - 1 V DC], channel 2
50 (4) - 20 mA, channel 2 (R
6
7
8
9
⊥⊥
⊥ (V DC)
⊥⊥
⊥⊥
⊥ (mA)
⊥⊥
Fig. 8-2: Pin Assignments X 2 (Analog Signal Outputs)
≤ 500 Ω)
B
≤ 500 Ω)
B
8 - 5
Page 48
ENTRY OF SYSTEM PARAMETERS
FLUSHING PERIOD / RESPONSE TIME
Rosemount Analytical
8.2.5Flushing Period
For calibration the gas paths must be supplied with sufficient calibration gas . The flushing period
has to be fixed adequately; perform calibration only after a suitable flushing period (the calibration
gas flow should be identical with sample gas flow).
This period may be selected in the range 0 - 99 sec. depending on calibration conditions.
Use the keys
andfor entry.
8.2.6Response Time (t90)
For some types of analysis an alteration of the analyzer damping factor , i.e. its electrical response
time, t90, may be required. The operator is offered the option of selecting a response time optimal
for each application.
The reset operation restores the settings of the analyzer to the parameters and calibration factors
set in our factory at the time of its manufacture.
This is equivalent to switching off the electrical supply line and switching off the battery buffering
of the RAM’s b y removing the battery jumper, J7.
All parameters and calibration factors entered by the user will be lost whenever a reset
operation is performed.
The currently valid user identification code must be entered before a reset will be executed; this
will prevent inadvertent resets.
Entry is performed by using
followed by
Whenever a reset operation is initiated, the analyz er operating program will be restarted, just as
it is when the instrument is switched on at first (see Section 6).
Jumper J6, which activates the watchdog circuitry, must be inserted if the
reset operation is to be executed correctly
This control permits switching the electronic cross-compensation feature on and off.T h e
cross-compensation feature is designed to minimize mutual interferences between the two gases
(e.g., CO2 and CO) measured by the analyzer .
Entry of “NO”:Cross-compensation is disabled
Entry of “YES”:Cross-compensation is enabled
Effect the entry by using
and
After changing this parameter a new spanning must be accomplished.
8.2.9Cross - Compensation Calibration
Determination of cross-compensation correction factors is performed during the span adjustment.
Pure test gases are required for this operation. Once cross-compensation corrections have been
determined, span adjustments may be performed using test gas mixtures.
Entry of “NO”:Spanning without cross-compensation correction (test gas mixtures)
Entry of “YES”:Spanning with cross-compensation correction (pure test gases)
Effect the entry by using
and
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
8 - 9
Page 52
ENTRY OF SYSTEM PARAMETERS
CROSS-COMPENSATION CALIBRATION / EXTERNALLY LOCATED SWITCH
Rosemount Analytical
To perform a calibration with cross-compensation correction, proceed as follows:
First perform a zeroing for both analysis channels (see Section 9.3.1).
Then perform a spanning for both analysis channels as described in Section 9.3.2.
The spanning for the first of the analysis channels calibrated must then be repeated.
The entries described in Sections 8.2.8 and 8.2.9 must be “YES” for
performanceof a calibration with cross-compensation correction !
Use only pure test gases !
When using test gas mixtures, “C.Cal” m ust be set to “NO” !
Spanning may be carried out only in the absolute measurement mode !
8.2.10Externally Located Switch
Sometimes it is necessary to switch the measurement mode and/or measuring range of analog
outputs by a separately located unit (see Section 11.2), i.e. not in the “SEL” menu.
By entering “YES”, this function has been activated. Now it is no longer possible to change either
measurement mode or measuring range in the “SEL ” menu; only the actual setup will be displayed.
Entry of “NO”:External switch (option) is disabled
Entry of “YES”:External switch (option) is enabled
Effect the entry by using
and
8 - 10
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 53
Rosemount Analytical
ENTRY OF SYSTEM PARAMETERS
8.2.11Program Version
The Program Version (No. of the installed software version) will be displayed.
Depress the key
8.2.12Serial - No.
The Serial - No. will be displayed. (Please note this number for fur ther contact with our factory
regarding maintenance, service etc.).
Depress the key
Continuation of Serial - No.
Depress the key
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
8 - 11
Page 54
ENTRY OF SYSTEM PARAMETERS
USER CODE / GAS - CYLINDER CORRECTION FACTOR
Rosemount Analytical
8.2.13User Code
The value 1 has been set in our plant.
To prevent parameter alterations by unauthorized persons, the operator may specify another
password (user code).
Use the keys
andfor entry .
Please take care when completing the user code.
8.2.14Gas - Cylinder Correction Factor
For differential measurement, it is sometime necessary to adjust the signal value to the set value.
This is possible due to the gas - cylinder correction factor: using this factor , the sensitivity can be
calibrated again during the differential measurement.
Example:
FAC .1old: 1.000
Difference to set value:+38
Difference to actual value: +40 ppm
FAC1new:== 0.950
Diff. to set value x FAC.1old38 x 1.000
Diff. to actual value40
Note:
Usually it is not necessary to adjust a gas - cylinder correction factor of 1.000. The difference
between the set and the actual values may happen due to the test gas tolerances or during the
analyzer calibration.
8 - 12
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 55
Rosemount Analytical
Cylinder correction factor f or channel 1
Entry possibility for
(differential measurement).
Use the keys
andfor entry.
ENTRY OF SYSTEM PARAMETERS
GAS - CYLINDER CORRECTION FACTOR
Entry possibility for
cylinder correction factor for channel 2
(differential measurement).
Use the keys
andfor entry.
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
8 - 13
Page 56
ENTRY OF SYSTEM PARAMETERS
GROUND LEVEL MODE
Rosemount Analytical
8.2.15Ground Level Mode
With this parameter the type of ground level determination can be selected for differential
measurement.
Entry of “YES”:The concentration present when switching over from absolute to
differential measurement is accepted by the analyzer as the valid
ground level for differential measurement.
In addition the following gases must be present in the analyzer:
Reference side:Zero gas
Entry of “NO”:The analyzer uses the concentration values set for parameter
When the system parameter “4P.On” is changed from “NO” to “YES” and the differential
measurement mode is already active, then the analyzer continues to use the preset
reference concentration v alues. To be able to change the ground lev el automatically, the
analyzer must be switched over from absolute to differential measurement mode at first.
When "YES" has been selected for “4P.On”, then the parameters “Gnd1” and “Gnd2”
(Section 10.2) show only the preset ground lev el. These parameters cannot be adjusted
by hand any more.
Use the keys
Measurement side:Reference gas
“Gnd1” and/or “Gnd2” (can be found in the “SEL.” function line,
see Item 10.2) as ground level for differential measurement.
andfor entry.
Depress the keyuntil
the display showsThe analyzer is now is back to analysis
mode.
8 - 14
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 57
Rosemount Analytical
SETUP OF THE ABSOLUTE MEASUREMENT MODE
ABSOLUTE MEASUREMENT
9.Absolute Measurement
The analyzer can measure in two different modes: absolute and differential.
In absolute measurement mode the reference side of the analyzer is purged with an unpressurized
gas, which does not absorb within a specified spectral range (flow: approx. 1 l/min)! In most cases
nitrogen (N2) is used.
9.1Setup of the Absolute Measurement Mode
To setup the absolute measurement mode, proceed as follows:
Depress the key
until the textappears.
Depress the key
If the Code has not already been entered, the following display
will appear
Use the keysto select the Code.
and then depress
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
9 - 1
Page 58
ABSOLUTE MEASUREMENT
SETUP ABSOLUTE MEASUREMENT MODE / SETTING ANALOG OUTPUTS
Rosemount Analytical
The display will show:
Use the keysandto select the variable “AbS. ”
Now the absolute measurement mode is active.
It is not possible to use different measurement modes for the two measuring channels.
When "YES" has been chosen for the system parameter "ES.On" (i.e. external switch
is active), then the switching over can be carried out only via this external switch. Only
the actual setting is displayed on the monitor no w.
The display will show:
Depress the keyuntil the display shows
9.2Setting Analog Outputs
While setting the analog outputs, it can be chosen between standard values ("Std.") and a freely
programmable magnifier (Scope, "SCOP").
Entry of “Std. ”:The factory range settings are taken ov er . The preset voltage and
(standard)current ranges correspond to the following concentration range:
[0 ... End of range (1000 ppm or 3.0 % resp.)]
When "Std." has been chosen, the "Scope" parameters will not be
displayed any longer.
Entry of “SCOP”:The concentration range can be freely set.
(scope)To select the variable and enter the desired range:
use the keysand
9 - 2
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 59
Rosemount Analytical
ABSOLUTE MEASUREMENT
ANALOG OUTPUT SETTING
When “SCOP” has been enteredappears on the display.
Now the value for "channel 1 absolute" can be selected: it represents the begin of voltage/current
range.
Entry is effected by usingand
It appears
Now the value for "channel 1 absolute", which represents the end of voltage/current range, can be
entered.
Example:
“Life Zero” has been selected (system parameter “L-0” = “YES”).
The analog output is being used. “0-1A” = 200 ppm; “1-1A” = 600 ppm
Concentration [ppm]2 00300400500600
Current [mA]48121620
If the same value has been selected for "0-1A" as for "1-1A", then 1 digit will be added
automatically to the "1-1A" value.
The values for "channel 2 absolute"
must be entered, too.
When "YES" has been chosen for the system parameter "ES.On" (i.e. external switch
is active), then the switching over from "Std." to "SCOP" can be carried out only via
this external switch. Only the actual setting ma y be displayed on the monitor now.
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
9 - 3
Page 60
ABSOLUTE MEASUREMENT
CALIBRATION
Rosemount Analytical
9.3Calibration
T o ensure correct measurement results, zeroing and spanning should be carried out once a week.
Spanning can be performed only after zeroing.
For the calibration procedure the required test gases have to be fed to the analyzer through the
respective gas inlets (cf. Section 5.3) with a no-back-pressure gas flow rate of about 1 l/min (the
same as with sample gas) !
After switching on the analyzer, wait at least 15 to 50 minutes
(depending on installed detectors) before admitting gas to the analyzer !
Zeroing
For zeroing, the measurement side of the analysis cell has to be purged with the same gas
as its reference side.
Spanning
The span gas concentration should be in a range of 80 % - 110 % of full - scale range !
For lower span gas concentrations the measuring accuracy could be lower for sample
gas concentrations, which are higher than the span gas concentration !
When using span gas mixtures the entry for “C.Cal” must be set to “NO”
(see Section 8.2.9) !
9 - 4
The correct pressure must be entered before performing the calibration,
if you want to have the possibility of pressure correction (see 8.2.1) !
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 61
Rosemount Analytical
ABSOLUTE MEASUREMENT
9.3.1Zeroing
Zeroing will set the actually measured gas concentration to “zero”.
Depress the key
until the display shows(zeroing channel 1) or
ZEROING
(zeroing channel 2) resp .
Depress the key
The following display will appear
Use the keysto select the correct user code
and acknowledge by depressing
The display will show orresp.
The actual zero level will be displayed.
Wait until the entered flushing period and t90 time elapses.
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
9 - 5
Page 62
ABSOLUTE MEASUREMENT
ZEROING
Rosemount Analytical
Depress the key
The nominal value orwill be displayed.
If the actual and nominal zero levels are identical, the next function can be selected then by using
the FUNCTION - key (without zeroing).
If the two values differ , then
depress the key
The actual measuring value orwill be displayed
T o start zeroing pressagain.
As soon as zeroing has finished,
the actual measuring value orresp. will be displayed.
To leave “calibration mode” press
9 - 6
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 63
Rosemount Analytical
ENABLING / DISABLING SPANNING FUNCTION
ABSOLUTE MEASUREMENT
9.3.2Spanning
9.3.2aEnabling / Disabling Spanning Function
It might not be necessary to have access to system parameter / spanning function. The appearance
of this function may ev en disturb. Theref ore , it is possib le to skip this level completely.
Depress the key
until the textappears.
Depress the key
If the Code has not already been entered, the following display
will appear
Use the keysandto select the Code.
The display will show:
Depress the key
until the textappears.
Entry of “YES”:System parameter / spanning function is enabled
Entry of “NO”:System parameter / spanning function is disabled
Effect the entry by using
and
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
9 - 7
Page 64
ABSOLUTE MEASUREMENT
SPANNING
Rosemount Analytical
9.3.2b Performing of Spanning
V erification of the span calibration is essential f or accurate concentration measurement.
Spanning can be performed only after zeroing.
Spanning will set the actually measured gas concentration to the entered “span gas setpoint”.
Note:The span gas concentration should be in a range of 80 % - 110 % of full - scale range !
For lower span gas concentrations the measuring accuracy could be lower for sample gas
concentrations, which are higher than the span gas concentration !
When using span gas mixtures the entry for “C.Cal” must be set to “NO”
(see Section 8.2.9) !
The correct pressure must be entered before performing the calibration,
if you want to have the possibility of pressure correction (see 8.2.1) !
Depress the key
until the display shows(spanning channel 1) or
(spanning channel 2) resp.
Depress the key
Enter the correct user code, if not already entered
The display will show orresp.
The actual concentration level will be displayed.
Wait until the entered flushing period and t
9 - 8
time elapses.
90
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 65
Rosemount Analytical
ABSOLUTE MEASUREMENT
Depress the key
The test gas setpoint orresp. will be displayed.
If necessary , enter the true test gas setpoint value (taken from the manuf acturer’s certification on
the gas bottle) by
using the keys
SPANNING
and
The actual measuring value orresp. will be displayed
Leave calibration mode by pressing the FUNCTION - key (enter of nominal value without span
calibration)
or pressagain to start spanning.
As soon as spanning has finished, the display indicates
the actual measuring value orresp.
To leave calibration mode press
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
9 - 9
Page 66
ABSOLUTE MEASUREMENT
MEASUREMENT
9.4Measurement
Depress the keyuntil
the display shows
The analyzer is now in the analysis mode.
Rosemount Analytical
In absolute measurement mode the reference side of the analyzer is purged with an unpressurized
gas, which does not absorb within a specified spectral range. In most cases nitrogen (N2) is used
for reference side. The measurement side is purged with sample gas (flow: approx. 1.0 l/min)!
Analyzer warming-up after switching on takes about 15 to 50 minutes,
depending on the installed detectors !
❍Admit reference gas (nitrogen, N2) at the gas inlet fitting of reference side.
❍Admit sample gas at the gas inlet fitting of measuring side.
❍Set the gas flow rate to approx. 1 l/min.
Note !
No matter which program lev el is active , the program automatically returns to the measurement
level 1 or 2 minutes after the last pressing of a key !
The right-hand decimal point of the display is pulsing to indicate absolute measurement mode.
It is possible to identify the measurement mode of the analyzer and measuring range of the analog
outputs by digital outputs (see Section 11).
9 - 10
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 67
Rosemount Analytical
DIFFERENTIAL MEASUREMENT
GROUND LEVEL SETTINGS
10.Differential Measurement
The analyzer can measure in two different modes: absolute and differential.
In differential measurement mode the reference side of the analyzer is purged with an unpressurized reference gas, the measurement side is purged with sample gas (flow: approx. 1.0 l/min)!
10.1Ground Level Mode
In this measurement mode, the reference concentration, which is used by the analyzer to measure
the difference, is of great importance: it is the basis f or the differential measurement. The analyzer
offers two possibilities to determinate the ground level.
Depress the key
until the displayappears.
Depress the key
If the Code has not already been entered, the following display
will appear
Use the keysto select the Code
and then depress
The display will show:
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
10 - 1
Page 68
DIFFERENTIAL MEASUREMENT
GROUND LEVEL SETTINGS
Rosemount Analytical
Depress the key
until the textappears.
10.1.1Automatic Determination
In this mode, the analyzer takes over the values present before the switching over from absolute
to differential measurement.
Entry of “YES”:The concentration present when switching over from absolute to
differential measurement is accepted by the analyzer as the valid
ground level for differential measurement.
In addition the following gases must be present in the analyzer:
Reference side:Zero gas
Measurement side:Reference gas
When the system parameter “4P.On” is changed from “NO” to “YES” and the differential
measurement mode is already active, then the analyzer continues to use the preset
reference concentration v alues. To be able to change the ground lev el automatically, the
analyzer must be switched over from absolute to differential measurement mode at first.
When “YES” has been selected for “4P.On”, then the parameters “Gnd1” and “Gnd2”
(section 10.2) show only the preset ground le vel: These parameters cannot be adjusted
by hand any more.
Use the keysandto select the variable “Yes. ”
Depress the keyuntil
the display showsThe analyzer is now back to analysis
mode.
10 - 2
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 69
Rosemount Analytical
DIFFERENTIAL MEASUREMENT
GROUND LEVEL SETTINGS
10.1.2Determination of Ground Level by Selecting Parameters
When this method is used, the ground level limits are preset values. This measurement mode does
not require to be preset before switching over; it can be selected during a differential measurement
as well.
Entry of “NO”:The analyzer uses the concentration values set for parameter
“Gnd1” and/or “Gnd2” (can be found in the “SEL.” function line,
see Item 10.2) as ground level for differential measurement.
Use the keysandto select the v ariable “No”
Depress the keyuntil
the display showsThe analyzer is now back to analysis
mode.
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
10 - 3
Page 70
DIFFERENTIAL MEASUREMENT
SETUP OF THE DIFFERENTIAL MEASUREMENT MODE
10.2Setup of the Differential Measurement Mode
To setup the differential measurement mode, proceed as follo ws:
Depress the key
until the textappears.
Depress the key
Rosemount Analytical
If the Code has not already been entered, the following display
will appear
Use the keysto select the Code
and then depress
The display will now show:
Use the keysandto select the variable “Diff. ”
Now the differential measurement mode is active. After switching over to differential measurement
mode, the reference side must be purged with reference gas.
It is not possible to use different measurement modes for the two measuring channels.
When "YES" has been chosen for the system parameter "ES.On" (i.e. external switch
is active), then the switching over can be carried out only via this external switch. Only
the actual setting is displayed on the monitor no w.
10 - 4
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 71
Rosemount Analytical
SETUP DIFFERENTIAL MEASUREMENT MODE / SETTING ANALOG OUTPUTS
DIFFERENTIAL MEASUREMENT
10.2.1Manual Ground Level Setting
Use the keysandto set the variable to concentration of
the reference gas in channel 1.
Use the keysandto set the variable to concentration of
the reference gas in channel 2.
The display will show:
Depress the key
10.3Setting Analog Outputs
While setting the analog outputs, it can be chosen between standard values ("Std.") and a freely
programmable magnifier (Scope, "SCOP").
Entry of “Std. ”:The factory range settings are taken ov er . The preset voltage and
(standard)current ranges correspond to the following concentration range:
Begin of RangeEnd of Range
Channel 1- 50 ppm+ 50 ppm
Channel 2- 1 %+ 1 %
When "Std." has been chosen, the "Scope" parameters will not be
displayed any longer.
Entry of “SCOP”:The concentration range can be set freely.
(scope)
Use the keysandto select the variable.
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
10 - 5
Page 72
DIFFERENTIAL MEASUREMENT
SETTING ANALOG OUTPUT
Rosemount Analytical
When “SCOP” has been enteredappears on the display.
Now the value for "channel 1 differential" can be selected which represents the begin of voltage/
current range.
Entry is effected by usingand
It appears
Now the value for "channel 1 differential" can be entered which represents the end of voltage/
current range.
Example:
“Life Zero” has been selected (system parameter “L-0” = “YES”).
The analog output is being used. “0-1d” = -20 ppm; “1-1d” = +20 ppm
Concentration [ppm]-20-1001020
Current [mA]48121620
If the same value has been selected for "0-1d" as for "1-1d", then 1 digit will be added
automatically to the "1-1d" value.
The values for "channel 2 differential"
must be entered, too.
When "YES" has been chosen for the system parameter "ES.On" (i.e. external switch
is active), then the switching over from "Std." to "SCOP" can be carried out only via
this external switch. Only the actual setting ma y be displayed on the monitor now.
10 - 6
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 73
Rosemount Analytical
DIFFERENTIAL MEASUREMENT
CALIBRATION (ZEROING)
10.4Calibration (Zeroing)
To ensure correct measurement results, zeroing should be carried out once a week.
Zeroing will set the actually measured gas concentration to “zero”.
For zeroing the measurement side of the analysis cell has to be purged with the same gas as its
reference side through the respective gas inlets (cf. Section 5.3) with a no-back-pressure gas flow
rate of about 1 l/min (the same as with sample gas) !
After switching on the analyzer, wait at least 15 to 50 minutes
(depending on installed detectors) before admitting gas to the analyzer !
Depress the key
until the display shows(zeroing channel 1) or
(zeroing channel 2) resp .
Depress the key
The following display will appear
Use the keysto select the correct user code
and acknowledge by depressing
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
10 - 7
Page 74
DIFFERENTIAL MEASUREMENT
CALIBRATION (ZEROING)
Rosemount Analytical
The displays will showorresp.
The actual zero level will be displayed.
Wait until the entered flushing period and t90 time elapses.
Depress the key
The nominal valueorwill be displayed.
If the actual and nominal zero levels are identical, the next function can be selected then by using
the FUNCTION - key (without zeroing).
If the two values differ , then
depress the key
The actual measuring valueorwill be displayed
To start zeroing depressagain.
As soon as zeroing has finished,
the actual measuring valueorresp. will be displayed.
To leave “calibration mode” depress
10 - 8
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 75
Rosemount Analytical
10.5Measurement
Depress the keyuntil
the display shows
The analyzer now is in the analysis mode.
DIFFERENTIAL MEASUREMENT
MEASUREMENT
In differential measurement mode the ref erence side of the analyzer is purged with an unpressurized reference gas, the measurement side is purged with sample gas (flow: approx. 1.0 l/min)!
Analyzer warming-up after switching on takes about 15 to 50 minutes,
depending on the installed detectors !
❍Admit reference gas at the gas inlet fitting of reference side.
❍Admit sample gas at the gas inlet fitting of measuring side.
❍Set the gas flow rate to approx. 1 l/min.
Note !
No matter which program lev el is active, the prog ram automatically returns to the measurement
lev el 1 or 2 minutes after the last pressing of a k ey!
The maximum absolute concentration of the gas to be measured is the upper limit of
the measuring range for “AbS”. Above this limit linearization errors may occur.
(Abs. Concentration of the gas measured = Ref. Concentr ation + Diff. Concentration)
The right-hand decimal point of the display is switched off during differential measurement mode .
It is possible to identify the measurement mode of the analyzer, measuring range of the analog
outputs and the ground level mode b y digital outputs (see Section 11).
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
10 - 9
Page 76
DIFFERENTIAL MEASUREMENT
Rosemount Analytical
10 - 10
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 77
Rosemount Analytical
11.Digital Outputs / Inputs
All analyzer digital outputs and digital inputs appear on the rear panel (Fig. 11-1).
M1M1M2M2
R1R1R2R2
DIGITAL OUTPUTS / INPUTS
IN
X1 OUTPUT
INTERFACE
X2 OUTPUT
OUTOUTIN
K1K2
24 VDC
MADE IN GERMANY
LEMOSA Socket
(Digital Inputs)
ABS./DIFF.
X3 OUTPUT
Plug X 3
(Digital Outputs)
Fig. 11-1: Rear Panel of BINOS® 100 4P (Digital Outputs / Digital Inputs)
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
11 - 1
Page 78
DIGITAL OUTPUTS / INPUTS
Rosemount Analytical
11.1Digital Outputs
All digital outputs of the analyzer are brought out to plug X 3 on the rear panel (Fig. 11-1). The
loading of the outputs (“Open Collector”) is max. 30 V DC / 30 mA.
1
5
9
6
1Measuring mode (absolute / differential)
2Analog output (“standard” / “scope”)
3Ground level mode “4P-On” (on / off)
4not used
5
6not used
7not used
8not used
9not used
⊥⊥
⊥
⊥⊥
Fig. 11-2: Pin - Assignments X 3 (Digital Outputs)
11.2Digital Inputs
T o switch the measurement mode and/or measuring range of analog outputs via externally located
switches (see Section 8.2.10) and not in the “SEL ” menu, the necessary digital inputs are brought
out to the LEMOSA socket on the rear panel of the analyzer (Fig. 11-1).
The loading of the customers contacts is min. 10 V DC / 1 mA.
To activate this function, system parameter “ES .ON” is set to “YES”. Now it is no longer possib le
to change either measurement mode or measuring range in the “SEL ” menu; only the actual setup
will be displayed.
❍Connect circuit board to the threated bolts at the rear panel and mounting with the
washers and the screws .
❍Connect cable subject to code pin to BKS - pin connector J9.
Rear panel
Threated bolt
Rear panel
PCB BSI 10
J 9
J 9
Code pin
1
Fig. 12-1: Installation of PCB BSI 10
❍For retrofitting serial interface insert enclosed EPROM (see Section 25).
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
12 - 1
Page 80
SERIAL INTERFACE (OPTION)
GENERAL
Rosemount Analytical
12.2General
The analyzer is equipped with a serial interface enabling communication with a host computer . The
host computer can call up, prescribe, or alter parameters, as well as initiate analyz er operations,
using standardized protocols. The optional BSI 10 plug in circuit board constitutes the hardw are
interface. This ma y be configured as RS 232 C or RS 485 interface. The RS 485 interface permits
networking several analyzers. Each analyzer may then be addressed using an assigned
numerical ID - code.
Communications are always initiated b y the host computer, i.e., analyz er behaves passively until
the host computer requests information from them or demands commencement of an action.
Communications use so - called “telegrams” being exchanged between the host computer and
the analyzer(s). Syntax for these telegrams is established in protocols.
T elegrams alwa ys commence with the "$" start character, immediately followed b y a three - digit
instruction code.
Subsequent elements of telegrams are segregated by the ";" hyphen c haracter.
The final element of all telegrams transmitted must be the “CR” termination character.
Upon receipt of the termination character, the analyzer attempts to e valuate the current contents
of its input buffer as a valid telegram. If the syntax of the transmitted telegram is correct, the analyzer
will transmit a response telegram to the host computer. This consists of the start character, an
instruction code, requested data, a block - parity byte, and the termination character .
If the syntax of the transmitted telegram is not correct, the analyzer will transmit a status telegram
containing an error message to the host computer. Each ter mination character reception thus
initiates an analyzer response.
12 - 2
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 81
Rosemount Analytical
SERIAL INTERFACE (OPTION)
T o avoid detecting transmission errors, the host computer can insert a message - length parity byte
immediately preceding the termination character for verification by the analyz er .
The time elapsed between the reception of star t characters and termination characters is not
limited by the analyzer , i.e ., there are no “time - out” periods.
If the host computer transmits new characters before the analyzer has responded to the preceding
telegram, the analyzer’ s input b uffer will reject them, i.e ., these characters will be ignored b y the
GENERAL
analyzer .
The transmission rate may be set between 600 and 4800 baud.
An echo - mode may also be activated.
The analyzer software is configured so that telegrams may be sent to the host computer at time
intervals of 150 ms and greater .
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
12 - 3
Page 82
SERIAL INTERFACE (OPTION)
Rosemount Analytical
START UP
12.3Start Up
The analyzer has been set in our factory to RS 232 C or RS 485 interface via the plugged PCB
SIF 232 or SIF 485 on the PCB BSI 10.
The parameter 232c can also been set to “NO” or “YES” in the SIP (Serial Interface P arameters)
line.
Interconnection to the interface is via the 9 - pin socket „Interface“ on the analyzer rear panel
(Fig. 12-2).
Socket
“Interface”
M1M1M2M2
R1R1R2R2
IN
X1 OUTPUT
INTERFACE
X2 OUTPUT
OUTOUTIN
K1K2
24 VDC
MADE IN GERMANY
ABS./DIFF.
X3 OUTPUT
12 - 4
Fig. 12-2: Socket “Interface” (Serial Interface)
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 83
Rosemount Analytical
12.3.1RS 232 C
This interface requires a shielded cable having at least three internal conductors.
SERIAL INTERFACE (OPTION)
START UP
5
1
6
9
1GND
2RxD
3TxD
4not used
5GND
6not used
7not used
8not used
9not used
Fig. 12-3: Pin - Assignments “RS 232 Interface”
12.3.2RS 485
Configure 2- or 4- wire operation via solder bridge LB 1 of PCB SIF 485 before mounting the PCB.
Connecting of [1 - 2] 2-wire operation is selected. Connecting of [2 - 3] 4-wire operation is active.
Connect jumper P2 at both ends of interface connection (termination). For network operation with
several analyzers via RS485 interface, termination has to be done at both ends of network
connection only. For the other analyzers remo ve the jumper.
5
1
69
1GND
2RxD3RxD+
4TxD+
5TxD6not used
7not used
8not used
9not used
Fig. 12-4: Pin - Assignments “RS 485 Interface”
In contrast to RS 232 C operation, simultaneous transmission and reception is not implemented
in this standard. This would not result in damage to the electronics, but could lead to destruction
of data. The analyzer beha ves passively in this mode of operation, i.e., it k eeps its transceiver set
for reception whenever it is not transmitting. Since the time periods for transmission and reception
are controlled by protocols, “data collisions” are excluded.
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
12 - 5
Page 84
SERIAL INTERFACE (OPTION)
START UP (INTERFACE - PARAMETER)
Rosemount Analytical
12.3.3Switching ON/OFF Interface Operation
The analyzer may be set to either “on - line” or “off - line” status. This setting may be perf ormed
either from the keypad or via telegram input.
Ke yboard setting:
SIP - parameter On.-L. = YESfor on - line status
SIP - parameter On.-L. = NOfor off - line status
T elegram setting:
Instruction code 006: sets analyzer on - line status
Instruction code 007: sets analyzer off - line status
If the analyzer is set to off - line status, it will accept only instruction code 006. All other instructions
will be ignored and result in transmission of appropriate status telegrams.
12.3.4Setting Interface Parameters
Agreement of interface parameters between analyzer and host computer is a fundamental
requirement for communication without errors.
The following analyzer parameters are concerned:
❏Baud rate:600 / 1200 / 2400 / 4800 bits/s
❏Data bits:8
❏Stop bits:2
❏Parity bit:none
❏Echo mode:on / off (received characters will be retransmitted immediately)
❏LPB-test:on / off (message - length parity check)
❏ID No.:0 to 99 (device ID No . in RS 485 mode)
12 - 6
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 85
Rosemount Analytical
All entries are made by using the keys
and
Depress the key
untilappears,
SERIAL INTERFACE (OPTION)
START UP (INTERFACE - PARAMETER)
then depress the key
The unit is now ready for code entry, if such has not already been performed.
NO= off - line status
YES= on - line status
Each device is assigned a device number for operation
through the RS 485 interface (0 - 99).
Select interface type:
NO = RS 485YES = RS 232 C
Set baud rate:
0 = 48001 = 2400
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
2 = 12003 = 600
Echo-mode operation:
NO = OFF YES = ON
Message - block parity check
NO = OFF YES = ON
12 - 7
Page 86
SERIAL INTERFACE (OPTION)
TELEGRAM SYNTAX
Rosemount Analytical
12.4Telegram Syntax
Telegrams are assembled as follows:
12.4.1Start Character ( “$” = Hex 24)
If the start character is missing will be transmitted of an appropriate status telegram by the analyzer.
12.4.2Termination Character ( “CR” = Hex OD)
If the termination character is missing, no decoding of the transmitted information will be performed,
and the analyzer will not respond. No response message will be transmitted.
12.4.3Instruction Code
Each instruction is assigned a unique three digit numerical instruction code. If a received
instruction code should be other than three digits in length or contain non - numerical ASCIIcharacters, the analyzer will transmit an appropriate status telegram. Reception of unassigned
instruction codes will also result in the transmission of a status telegram.
In the RS 232 C mode of operation, the instruction code immediately follows the start character;
in the RS 485 mode of operation, the start character is followed by a two - digit de vice identification
code, the separator character “;” and a three - digit instruction code, in this order .
12.4.4Hyphen Character ( “;” = Hex 3B)
Individual elements of a telegram line are separated by this hyphen character. Missing hyphen
characters can lead to misinterpretations of telegrams, and will result in transmission of an
appropriate status telegram.
12 - 8
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 87
Rosemount Analytical
SERIAL INTERFACE (OPTION)
STATUS TELEGRAM
12.4.5Status Telegram
If telegram syntax is faulty, or analyzer is unable to act upon an instruction received, then the
analyzer will transmit a status telegram to the host computer .
These status telegrams are listed here for reference:
$ID;000;S100;LPB<CR>unrecognized instruction code
$ID;000;S101;LPB<CR>LP - byte in error
$ID;000;S102;LPB<CR>start character missing
$ID;000;S103;LPB<CR>input buffer overflow
$ID;xxx;S104;LPB<CR>analyzer off - line status
$ID;xxx;S105;LPB<CR>text line too long
$ID;xxx;S106;LPB<CR>undefined instruction
$ID;xxx;S107;LPB<CR>invalid integer value
$ID;xxx;S108;LPB<CR>numerical value outside defined range
$ID;xxx;S109;LPB<CR>invalid failure/status code
$ID;xxx;S110;LPB<CR>instruction can not be done here
$ID;xxx;S111;LPB<CR>failure in transmitted character
$ID;xxx;S112;LPB<CR>zeroing running
$ID;xxx;S113;LPB<CR>spanning running
$ID;xxx;S114;LPB<CR>invalid real number
$ID;xxx;S115;LPB<CR>automatic calibration mode off
$ID;xxx;S116;LPB<CR>parameter outside defined range
$ID;xxx;S117;LPB<CR>preflushing period is running
xxx:instruction code
I D:device ID No . in RS 485 mode
LPB:message - length parity byte
<CR>:termination character
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
12 - 9
Page 88
SERIAL INTERFACE (OPTION)
NUMERICAL REPRESENTATION / BLOCK PARITY CHECK
Rosemount Analytical
12.4.6Numerical Representations
T elegrams may contain integers or real numbers. The formats for these numbers are subject to the
following restrictions:
Integers:- Max. value = 216 - 1
- Positiv e numbers only accepted
- No decimal points allowed
Real:- Maximum of 6 digits accepted
- No alphabetic characters (e.g. 2.2E-6) allowed
- Analyzer output = 6 - digit real numbers
12.4.7Block Parity Check
The master control computer may insert a message - length parity byte into telegrams. This
invariably consists of two characters.
The message - length parity byte is the cumulatively EXCLUSIVE - OR correlation of all previously
transmitted characters of the telegram line. Representation is in hexadecimal format. For example,
if the decimal value should be decimal 13, this will be represented by the two characters “OD”, i.e.,
030H and 044H.
The verification procedure may be enabled or disabled at the analyzer (see Section 12.3.4).
12 - 10
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 89
Rosemount Analytical
INSTRUCTION SYNTAX / INSTRUCTION LISTING
12.5Instruction Syntax
Code definitions:
RP:Receive parametersanalyzer is accepting values
SP:Send parametersanalyzer is sending values
RI:Receive instructions
k:Channel numbers 0 to 2(0 = both channels)
m:Measurement mode(0 = absolute / 1 = differential)
w:Value(0 = N o o r scope / 1 = Yes or standard)
SERIAL INTERFACE (OPTION)
ff:Function line number
pp :Parameter number
<ID>:Analyzer ID - no . f or RS 485 mode of operation; f ollows start character
LPB:Message - length parity byte
<CR>: Termination character
Receipt of any instruction codes not listed in the following section will be acknowledged by
transmission of status code 106. Future expansions will make use of code numbers not currently
in use.
12.5.1Instruction Listing
Instruction syntax:Instruction description:
$ID;001;LPB<CR>SP “4P.On” parameter
$ID;002;w;LPB<CR>RP “4P.On” parameter
$ID;003;k;LPB<CR>SP Ground level for differential measurement
$ID;004;w;k;LPB<CR>RP Ground level for differential measurement
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
answer “0”: Ground level of “GND” parameters is used (Section 10.2.1).
answer “1”: Automatic ground level is used (Section 10.1.1).
w “0”: Use ground level of “GND” parameters (Section 10.2.1).
w “1”: Use automatic ground level is used (Section 10.1.1).
Not for differential measurement mode, if “4P.On” = YES.
12 - 11
Page 90
SERIAL INTERFACE (OPTION)
INSTRUCTION LIST
Instruction syntax:Instruction description:
$ID;006;LPB<CR>RI on - line status
$ID;007;LPB<CR>RI off - line status
$ID;008;LPB<CR>SP datalogger data
$ID;009;k;LPB<CR>SP measurement mode
$ID;010;m;k;LPB<CR>RP measurement mode [k = 0 only]
Rosemount Analytical
$ID;011;m;k;LPB<CR>SP at full scale range
$ID;013;k;LPB<CR>SP t
$ID;014;w;k;LPB<CR>RP t
(response time) [k= 0 to 4]
90
(response time) [k= 1 to 4]
90
k = 0 all values
k = 1 Ch. 1 abs.
k = 2 Ch. 1 diff.
k = 3 Ch. 2 abs.
k = 4 Ch. 2 diff.
$ID;017;k;LPB<CR>SP preflushing period (zero gas)
$ID;018;w;k;LPB<CR>RP preflushing period (zero gas) [k = 0 only]
$ID;019;k;LPB<CR>SP preflushing period (span gas)
$ID;020;w;k;LPB<CR>RP preflushing period (span gas) [k = 0 only]
$ID;023;k;LPB<CR>SP concentration
$ID;028;m;k;LPB<CR>SP span gas concentration
$ID;029;m;w;k;LPB<CR>RP span gas concentration
$ID;601;ff;pp;LPB<CR>SP analyzer parameter
$ID;602;ff;pp;LPB<CR>RP analyzer parameter
$ID;603;k;LPB<CR>SP gas component
$ID;627;LPB<CR>SP failure message (possible error batt. is
clearing by read out)
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 91
Rosemount Analytical
SERIAL INTERFACE (OPTION)
12.6Datalogger
The analyzer has been equipped with a datalogger which instantly prints out all major data stored
during the measurement: measuring values, measurement mode, error status, serial number, and
time of up to 180 storage locations. An automatic function enables the automatic recording of
measuring values at defined time intervals. Data output is accomplished e.g. by a printer
connected via serial interface.
12.6.1Manual Recording of Data in the Datalogger
DATALOGGER
If the measured value is displayed, then the actual data could be stored by just pressing
The measured value memory automatically switches over to the next storage location assigned.
The initiated storage process is acknowledged by the following message, displayed for a short
period of time:
(“XX” identifies the location for the next storage)
If all 180 storage locations are occupied, the error message “E.37” (OA error =
Overflow Area error) appears on the monitor . In case that the new values must be
stored, then press and the first lines will be overwritten.
Normally the message “E.37” disappears 1 or 2 minutes later.
The datalogger may be programmed via analyzer electronics, but this is possible only when “YES”
has been selected for parameter “P.SEt” (see Section 8.1).
In this case the assigned storage location number is constantly display ed on the monitor .
In addition the storage location can now be freely assigned by pressing the keys
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
12 - 13
Page 92
SERIAL INTERFACE (OPTION)
DATALOGGER
Rosemount Analytical
12.6.2Automatic Recording of Data in the Datalogger
In this menu the time intervals can be selected, when the automatic data storage have to be
accomplished.
The range of accepted entries is 0 to 1440 minutes (= 24 hours)
By selecting “0”, the automatic data storage function is disabled.
Use the keysandf or the entry.
Note:
If an OA error occurs during the automatic storage mode, no error message (“E.37”) will be
displayed and the previously stored values will be automatically overwritten, first line at first.
12.6.3Data Output
Data output is accomplished e.g. by a printer connected via serial interface. Make sure that the
parameters (setting in “SIP. ”, see Section 12.3.4) of the serial interf aces are identical.
❏Baud rate:600 / 1200 / 2400 / 4800 bits/s
❏Data bits:8
❏Stop bits:2
❏Parity bit:none
❏Echo mode:on / off (received characters will be retransmitted immediately)
To print out the stored data use the keysto select the variable “YES”.
and then depress
All data stored for the current operation/measurement will be printed out in A4 format and 65 lines
per page. A line feed is acknowledged by printing of <CR> and <LF> signs.
12 - 14
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 93
Rosemount Analytical
SERIAL INTERFACE (OPTION)
12.6.4Deleting of Data from Datalogger
The deleting of stored data is secured by code. To delete data proceed as follows:
Depress the key
until the displayappears.
Depress the key
DATALOGGER
The displaywill appear
Use the keysto select the Code
and then depress
The displaywill appear.
Use the keysto select the variable “YES” and then depress
Now the stored data is deleted. New record begins with line 1.
Note:
It is not necessary to delete the stored data because they will be overwritten when an OA error
occurs. Normally the error message “E.37” disappears 1 or 2 minutes later .
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
12 - 15
Page 94
SERIAL INTERFACE (OPTION)
DATALOGGER
Rosemount Analytical
12.6.5Display and Adjusting of Time
In the time display , hours and minutes are separated b y a decimal point, which is pulsating at any
second.
The analyzer is not equipped with a real-time unit. Theref ore the time is displayed only when the
power supply is on. After turning off the power supply , the time must be adjusted ane w. The stored
data remains stored in the battery-buffered RAM-memory .
To adjust the time display depress the key
until the displayappears.
Depress the key
If the Code has not already been entered, the following display
will appear
Use the keysto select the Code and then depress
The displaywill appear.
Use the keysto set the hour and then depress
The displaywill appear.
Use the keysto set the minutes and then depress
Note:
By setting the minutes, the unit-internal second counter will be reset to “0”.
12 - 16
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 95
Rosemount Analytical
Error CodePossible ReasonsCheck / Correct
13.Error List
Some of the failures which ma y arise during measurement will be reported on the display in form
of error codes.
When such a failure arises, the display's will show the concentration value
alternating with(E = ERROR).
ERROR LIST
Be sure to observe the safety measures for all workings at the anal yzer!
Error CodePossible ReasonsCheck / Correct
1. Displays are “switched OFF”.
No display.
2. Voltage supply absent.
1. Press any key.
Check parameter "dOFF"
(see 8.2.3).
2. Check electrical supply
(see Fig. A-2, Item 3).
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
3. Connection front panel /BKS
absent.
3. Check connection
BKB - BKS (X1) (see Section 15).
13 - 1
Page 96
ERROR LIST
Error CodePossible ReasonsCheck / Correct
Flushing.
Battery buffer faulty.
Rosemount Analytical
Check, if Jumper J 7
is plugged (see Section 16).
Channel 1.
Channel 2.
A/D - conversion
of end signal
absent.
The EPROM - default
values were charged.
1. Jumper not or incorrect
plugged.
2. Positive or negative
reference voltage absent.
3. Light barrier signal absent.
Exchange battery,
if battery voltage < 3.5 V
(BKS - jumper J7 plugged).
The error is clearing after
depressing any key or with
serial interface instruction $627.
Light source is cold:
For dual - IR - channel analyzer
interchange the two light sources.
Replace the suspect light source
(see Section 23.2).
6.Check measuring point 14.7.
7.Check analysis cell and windows
for contamination.
Cleaning of contaminated parts
(see Section 22.3).
8.Barometric pressure
effects.
9.Temperature below the
dew point in the gas paths.
10.Faulty A/D converter.
Check gas paths and gas
conditioning to contamination.
8.Enter the correct value for
barometric pressure (see 8.1).
9.Check the temperature of the gas
paths and eliminate any reason of
condensation.
Maintain all temperatures at
values at least 10 °C above the
dew point of sample gas.
10.Exchange BKS.
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
13 - 5
Page 100
ERROR LIST
Error CodePossible ReasonsCheck / Correct
Response time too long
(t
time)
90
1. Incorrect response time
( t90 time).
2. Pumping rate inadequate.
3. Contamination of the gas
paths.
Rosemount Analytical
1. Check the value for t90 time
(see Section 8.2.6).
2. The feeder line between the
sampling point and the analyzer is
too long.
Use a larger, external pump;
consider adding a bypass line to the
process stream for sampling
purposes (see Section 5.1).
3. Check gas paths and gas
conditioning to contamination.
Clean gas paths and exchange the
filter elements.
13 - 6
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.