Rosemount BINOS 100 4P Analyzer-1st Ed. Manuals & Guides

Page 1
Rosemount Analytical
Operation Manual
BINOS® 100 4P
Microprocessor - Controlled
NDIR - Analyzer
90002927(1) BINOS® 100 4P e [1.00] 17.06.97
1. Edition 06/97
Catalog - No: 90 002 927
Managing The Process Better
Page 2
Rosemount Analytical
Fisher-Rosemount GmbH & Co assumes no liability for any omissions or errors in this man ual. Any liability for direct or indirect damages, which might occur in connection with the deliv ery or the use of this manual, is expressly excluded to the extend permitted by applicable law.
This instrument has left the works in good order according to safety regulations. T o maintain this operating condition, the user must strictly follow the instructions and consider the warnings in this manual or provided on the instrument.
Troubleshooting, component replacement and internal adjustments must be made by qualified service personnel only .
Misprints and alterations reserved
©
1997 by FISHER-ROSEMOUNT GmbH & Co. (PAD)
1. Edition: 06/97
Read this operation manual carefully before attempting to operate the analyz er ! For expedient handling of reports of defects , please include the model and serial number which can be read on the instrument identity plate. Look for the error check list, too (see Section 29 of this manual)
Fisher-Rosemount GmbH & Co.
Industriestrasse 1
D - 63594 Hasselroth
Phone (+49) 60 55 / 884 - 0
Telefax (+49) 60 55 / 884 - 209
90002927(1) BINOS® 100 4P e [1.00] 17.06.97
Page 3
Rosemount Analytical
SAFETY SUMMARY S - 1
General S - 1 Gases and Gas Conditioning (Sample Handling) S - 2 Supply Voltage S - 3 Connection Cables S - 3 Electrostatic Discharge S - 4

CONTENTS

CONTENTS

TECHNICAL DESCRIPTION

1. Setup 1 - 1
1.1 Front Panel 1 - 1
1.2 Rear Panel 1 - 1
1.3 Inside View 1 - 1
2. Photometer Assembly 2 - 1
2.1 Photometer 2 - 1
2.2 Detectors 2 - 2
3. Measuring Principle 3 - 1
3.1 Pyroelectrical Detector 3 - 1
3.2 Opto - Pneumatic Measuring Principle (Gas Detector) 3 - 3
3.3 Technique 3 - 5
4. Main Features 4 - 1
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
I
Page 4
CONTENTS

OPERATION

5. Preparation 5 - 1
5.1 Installation 5 - 1
5.2 Gas Conditioning (Sample Handling) 5 - 2
5.2.1 Gas Flow Rate 5 - 2
5.3 Gas Connections 5 - 3
6. Switching On / Switching Off 6 - 1
6.1 Switching On 6 - 1
6.2 Switching Off 6 - 3
Rosemount Analytical
7. Key Functions 7 - 1
7.1 FUNCTION 7 - 2
7.2 ENTER 7 - 3
7.3 INPUT - CONTROL 7 - 5
8. System Parameters 8 - 1
8.1 Enabling / Disabling System Parameter Function 8 - 1
8.2 Entry of System Parameters 8 - 2
8.2.1 Pressure Correction 8 - 3
8.2.2 Tolerance Check 8 - 3
8.2.3 Display Off 8 - 4
8.2.4 Analog Signal Outputs 8 - 4
8.2.5 Flushing Period 8 - 6
8.2.6 Response Time (t90) 8 - 6
8.2.7 Reset 8 - 8
8.2.8 Cross - Compensation 8 - 9
8.2.9 Cross - Compensation Calibration 8 - 9
8.2.10 Externally Located Switch 8 - 10
8.2.11 Program Version 8 - 11
8.2.12 Serial - No. 8 - 11
8.2.13 User Code 8 - 12
8.2.14 Gas - Cylinder Correction Factor 8 - 12
8.2.15 Ground Level Mode 8 - 14
II
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 5
Rosemount Analytical
9. Absolute Measurement 9 - 1
9.1 Setup of the Absolute Measurement Mode 9 - 1
9.2 Setting Analog Outputs 9 - 2
9.3 Calibration 9 - 4
9.3.1 Zeroing 9 - 5
9.3.2 Spanning 9 - 7
9.3.2a Enabling / Disabling Spanning Function 9 - 7
9.3.2b Performing of Spanning 9 - 8
9.4 Measurement 9 - 10
10. Differential Measurement 10 - 1
10.1 Ground Level Mode 10 - 1
CONTENTS
10.1.1 Automatic Determination 10 - 2
10.1.2 Determination of Ground Level by Selecting Parameters 10 - 3
10.2 Setup of the Differential Measurement Mode 10 - 4
10.2.1 Manual Ground Level Setting 10 - 5
10.3 Setting Analog Outputs 10 - 5
10.4 Calibration (Zeroing) 10 - 7
10.5 Measurement 10 - 9
11. Digital Outputs / Inputs 11 - 1
11.1 Digital Outputs 11 - 2
11.2 Digital Inputs 11 - 2
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
III
Page 6
CONTENTS
12. Serial Interface (Option) 12 - 1
12.1 Retrofitting of Serial Interface / Status Signals 12 - 1
12.2 General 12 - 2
12.3 Start Up 12 - 4
12.3.1 RS 232 C 12 - 5
12.3.2 RS 485 12 - 5
12.3.3 Switching ON/OFF Interface Operation 12 - 6
12.3.4 Setting Interface Parameters 12 - 6
12.4 Telegram Syntax 12 - 8
12.4.1 Start Character ( “$” = Hex 24) 12 - 8
12.4.2 Termination Character ( “CR” = Hex OD) 12 - 8
12.4.3 Instruction Code 12 - 8
Rosemount Analytical
12.4.4 Hyphen Character ( “;” = Hex 3B) 12 - 8
12.4.5 Status Telegram 12 - 9
12.4.6 Numerical Representations 12 - 10
12.4.7 Block Parity Check 12 - 10
12.5 Instruction Syntax 12 - 11
12.5.1 Instruction Listing 12 - 11
12.6 Datalogger 12 - 13
12.6.1 Manual Recording of Data in the Datalogger 12 - 13
12.6.2 Automatic Recording of Data in the Datalogger 12 - 14
12.6.3 Data Output 12 - 14
12.6.4 Deleting of Data from Datalogger 12 - 15
12.6.5 Display and Adjusting of Time 12 - 16
IV
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 7
Rosemount Analytical

TROUBLESHOOTING

13. Error List 13 - 1
14. Measuring Points of BKS 14 - 1
14.1 Supply Voltage + 6 V 14 - 1
14.2 Reference Voltage, Positive 14 - 1
14.3 Reference Voltage, Negative 14 - 2
14.4 Motor Drive 14 - 2
14.5 Temperature Sensor 14 - 3
14.6 Light Barrier Signal 14 - 3
14.7 Analog Preamplifying 14 - 4
CONTENTS
15. Plug Pin - Allocation of BKS 15 - 1
16. Jumper Allocation of BKS 16 - 1
17. (vacant)
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
V
Page 8
CONTENTS

MAINTENANCE 18 - 1

19. (vacant)
20. Leak Testing 20 - 1
21. Opening of the Housing 21 - 1
22. Replacement and Cleaning of Photometric Components 22 - 1
22.1 Removal of the Photometer Assembly 22 - 1
22.2 Light Source Replacement 22 - 2
22.3 Cleaning of Analysis Cells and Windows 22 - 3
Rosemount Analytical
22.3.1 Removal of Analysis Cells 22 - 3
22.3.2 Cleaning 22 - 4
22.3.3 Reinstalling Analysis Cells 22 - 5
22.4 Reinstalling the Photometer Assembly 22 - 6
22.5 Physical Zeroing 22 - 7
22.5.1 Standard Photometer (not sealed version) 22 - 7
22.5.2 Sealed Photometer (Option) 22 - 8
23. (vacant)
VI
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 9
Rosemount Analytical

TECHNICAL DATA 24 - 1

24.1 Housing / Environments 24 - 1
24.2 Signal Outputs / Inputs, Interfaces 24 - 2
24.3 Measurement Data / Gas Conditions 24 - 3
24.4 Voltage Supply 24 - 4

SUPPLEMENT

25. Replacing the EPROM 25 - 1
CONTENTS
26. Pin - Assignments 26 - 1
27. Connection Cable 27 - 1
28. (vacant)
29. Failure Check List 29 - 1
INDEX R - 1
List of Figures R - 7
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
VII
Page 10
CONTENTS
Rosemount Analytical
VIII
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 11
Rosemount Analytical

Safety Summary

1. General
SAFETY SUMMARY

GENERAL

In this manual we have used the following safety symbols
to draw your attention to strictly follow these instructions !
The following general saf ety precautions must be observed during all phases of operation, service and repair of this instrument ! Failure to comply with these precautions or with specific w arnings elsewhere in this manual violates safety standards of design, manuf acture and intended use of this instrument ! Failure to comply with these precautions may lead to personal injury and damage to this instrument !
Fisher-Rosemount GmbH & Co. assume no liability for the customer´s failure to comply with these requirements !
Do not attempt internal service or adjustment unless another person, capable of render­ing first aid and resuscitation, is present !
Because of the danger of introducing additional hazards, do not perform any unauthorized modification to the instrument ! Return the instrument to a Fisher-Rosemount Sales and Service office for service or repair to ensure that safety f eatures are maintained !
Operating personnel must not remove instrument covers ! Component replacement and internal adjustments must be made by qualified service personnel only !
Instruments which appear damaged or defective should be made inoperative and secured against unintended operation until they can be repaired by qualified service personnel.
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
S - 1
Page 12
SAFETY SUMMARY
GENERAL / GASES AND GAS CONDITIONING
Read this operation manual carefully before attempting to operate with the instrument !
Do not operate the instrument in the presence of flammable gases, explosive atmosphere or furnes without supplementary protective measures !
The installation site for the instrument has to be dry and remain above freezing point at all times. The instrument must be exposed neither to direct sunlight nor to strong sources of heat. Be sure to observe the permissible ambient temper ature ! For outdoor sites, we recommend to install the instrument in a protective cabinet. At least, the instrument has to be protected against rain (e.g., shelter).
Rosemount Analytical
Due to the high temperatures of photometer or heated components there is a danger of burns to the operators.

2. Gases and Gas Conditioning (Sample Handling)

Do not interchange gas inlets and gas outlets ! All gases have to be supplied to the system as conditioned gases ! When the instrument is used with corrosive gases, it is to be verified that there are no gas components which may damage the gas path components.
The exhaust gas lines hav e to be mounted in a dec lining, pressureless and frost-free way and according to the valid emission legislation !
Be sure to observe the safety regulations for the respective gases (sample gas and test gases / span gases) and the gas bottles !
S - 2
Inflammable or explosiv e gas mixtures must not be purged into the instrument without supplementary protective measures !
To avoid a danger to the operators by explosiv e, toxic or unhealthy gas components, first purge the gas lines with ambient air or nitrogen (N2) before cleaning or exchange parts of the gas paths.
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 13
Rosemount Analytical
SAFETY SUMMARY

SUPPLY VOLTAGE

3. Supply Voltage
Verify correct polarity for 24 V DC operation !
This product is a Safety Class 1 instrument (provided with a protective earth terminal).
To prevent shock hazard, the instrument chassis and cabinet must be connected to an electrical ground. The instrument must be connected to the AC power supply mains through a three-conductor power cable , with the third wire firmly connected to an electrical ground (safety ground) at the po wer outlet. If the instrument is to be energized via an external power supply, that goes for the po wer supply, too. Any interruption of the protective (grounding) conductor or disconnection of the protectiv e earth terminal will cause a potential shock hazard that could result in personal injury. Deliberate disconnection is inadmissible / prohibited !
Use only power supply VSE 2000, UPS 01 T or equiv alent po wer supplies to be in agreement with the CE conformity.
In case of exchanging fuses the customer has to be certain that fuses of specified type and rated current are used. It is prohibited to use repaired fuses or def ective fuse holders or to short-circuit fuse carriers (fire hazard).
Always disconnect power, discharge circuits and remove external voltage sources before troubleshooting, repair or replacement of any component !
Any work inside the instrument without switching off the power must be performed only by a specialist, who is familiar with the related danger !

4. Connection Cables

Use only from our factory optionally delivered cables or equiv alent shielded cables to be in agreement with the CE conformity. The customer has to guarantee that the shield is being connected bothsided.
By using optionally delivered terminal strip adapters the analyzer might not be in agreement with the CE conformity. In this case CE conformity is to be declared by customer as “manufacturer of system”.
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
S - 3
Page 14
SAFETY SUMMARY

ELECTROSTATIC DISCHARGE

Rosemount Analytical
5. Electrostatic Discharge
The electronic parts of the analyzer can be irreparably damaged if exposed to electrostatic discharge (ESD).
The instrument is ESD protected when the covers have been secured and safety precautions observed. When the housing is open, the internal components are not ESD protected any more.
Although the electronic parts are reasonably safe to handle, you should be aw are of the following considerations:
The best example for ESD is when y ou walk across a carpet then touch an electrically grounded metal doorknob: the tiny spark which jumps is a result of electrostatic discharge (ESD). You can prevent ESD b y removing the charge from your body before removing the analyzer´s housing and ensuring that no ESD can be built up while working with opened housing.
Ideally , the analyz er should be opened and serviced in an ESD protecting workstation: here y ou can wear a wrist trap. Should there be no such workstation available, be sure to carry out the following procedure:
Discharge the electric charge from your body. Do this by touching a device that is electrically grounded (any device that has a three - prong plug is electrically grounded when it is plugged into a power receptacle). This should be done several times during the operation with opened housing (especially after leaving the service site because the movement on low conducting floors or in the air might cause additional ESDs).
S - 4
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 15
Rosemount Analytical
FRONT PANEL / REAR PANEL
1. Setup
The analyzer is incorporated in a 1/4 19" rack-mounting housing, 3 height units.
The optional table-top housing is fitted additionally with a carrying strap and rubber feet.

1.1 Front Panel

The front panel (see Fig. A-1) includes the LED displays as well as all of the keys required for

SETUP

operating the analyzer.

1.2 Rear Panel

The rear panel (Fig. A-2) includes:
the gas line fittings
the plug for the electrical supply input
the subminiature “D” mating socket for the analog signal outputs
the subminiature “D” plug for the digital outputs (measurement mode / measuring range)
the 4-pin LEMOSA socket (connection to external switch of measurement mode / measuring
range)
optionally the subminiature “D” mating socket for the RS 232 C / 485 interface

1.3 Inside View

The inside view is shown in Fig. 1-1.
90002927(1) BINOS® 100 4P e [1.00] 17.06.97
1 - 1
Page 16
SETUP
INSIDE VIEW
Rosemount Analytical
Gas line fittings
Photometer with
gas detector
(CO2 channel)
Holding
device
Photometer with
pyroelectrical detector
(H2O channel)
1 - 2
Front panel
Fig. 1-1: Inside View BINOS® 100 4P
90002927(1) BINOS® 100 4P e [1.00] 17.06.97
Page 17
Rosemount Analytical

PHOTOMETER ASSEMBLY

2. Photometer Assembly
Depending on gas component and measuring range, different photometer assemblies may be realized in BINOS® 100 4P. Optionally the photometer can be sealed against ambient air . In this case all parts are sealed with O - rings. The entire photometer assembly is mounted as a unit on the main circuit board (BKS) by means of a bracket. The main circuit board is inserted into guide rails in the analyzer housing, to which the front panel (membrane keypad) and the rear panel are mounted.

2.1 Photometer

To enable both types of measurement - the standard absolute and the additional differential measurement (while taking the base concentration level into consideration) - a special photometer assembly has been developed (see Fig. 2-1).
The basic part of the photometer assembly is the chopper housing (01). The light source (thermal radiator, 02), the analysis cell (03), and the signal detection unit [including a filter cell (04) and a detector (05/07)] are all mounted on this chopper housing.
The chopper housing also incorporates the interference filters for the selection of spectral bandpass ranges from the broadband emission of the light sources.
Between the two halves of the chopper housing (03), which are sealed together with an O-ring, there is the chopper blade which is driven b y a stepping motor. Both the chopper housing and the motor encapsulation are hermetically sealed against ambient in order to prevent any gas from entering, such as atmospheric CO2, which could lead to background absorptivity (preabsorption) thus causing drift effects. An absorber pro vides f or constant remov al of any traces of CO2 which may enter the interior of the chopper housing via diffusion.
In addition the chopper housing incorporates a photoelectric gate for providing a reference signal for the phase angle of the chopper blade, and a temperature sensor (10) for monitoring continuously the temperature of the photometer assembly . The information thus obtained is used by the signal processing electronics to compensate thermal effects.
90002927(1) BINOS® 100 4P e [1.00] 17.06.97
2 - 1
Page 18
PHOTOMETER ASSEMBLY
Rosemount Analytical
Each analysis cell is divided in the middle by a partition wall in two halves: the measurement and the reference side. These sides are sealed at both ends by CaF2 windows. This allows the differential measurement to be carried out. The sample gas passes through the measurement side while the reference side is filled with either nitrogen or another ref erence gas .
The filter cell (04) consists of a cone which optimally adjusts the radiation cross section of the analysis cell to the detector surface.

2.2 Detectors

BINOS® 100 4P is equipped with two different detectors. The reason for this solution is the dependence on temperature of an H2O-filled gas detector (dew point undershoot). Therefore , for CO2 a gas detector is used and for H2O, a pyroelectrical (solid-state) detector.
The gas detector (05) and its preamplifier are interconnected by an isolated cable. The preamplifier (06) is mounted on the analysis cell (03).
The solid-state detector and its preamplifier are directly interconnected (07); both are mounted on the filter cell (04).
2 - 2
90002927(1) BINOS® 100 4P e [1.00] 17.06.97
Page 19
Rosemount Analytical
PHOTOMETER ASSEMBLY
05
04
06
07
04
Legends:
01 Chopper Housing 02 Light Source (thermal radiator) 03 Analysis Cell (200 mm) 04 Filter Cell 05 Gas Detector (CO2 measurement) 06 Preamplifier Gas Detector 07 Pyroelectrical Detector (H2O measurement) 08 Clamp (analysis cells 10-200 mm) 09 Mounting Screws for Analysis Cells 10 Temperature Sensor
03
09
01
02
90002927(1) BINOS® 100 4P e [1.00] 17.06.97
03
09
02
10

Fig. 2-1: Photometer Assembly BINOS® 100 4P

2 - 3
Page 20
PHOTOMETER ASSEMBLY
Rosemount Analytical
2 - 4
90002927(1) BINOS® 100 4P e [1.00] 17.06.97
Page 21
Rosemount Analytical

MEASURING PRINCIPLE

3. Measuring Principle
The analyzers are non - dispersive infrared photometers (NDIR). They detect the infrared light absorbed by the gas to be measured using the measurement of selective radiation in a column of gas.
The wavelength of the absorption bands is characteristic of the type of gas being measured, while the absorption strength gives a measure of the concentration of this gas component.
The detector receives signals succeeding in time. The difference between the absorption measured in the measurement and reference cells gives a measure of the difference in concen­tration of the gas component being measured.
To enable this measurement, a gas detector is used for CO2 and a pyroelectrical one for water vapour.

3.1 Pyroelectrical Detector

The spectral transmittance curves of the interference filter used in the BINOS® 100 4P analyzer and the spectral absorption of the gases CO2 and H2O are shown in Fig. 3-1. It can be seen that the absorption bands of these gases each coincide with the bandpass of one of the interference filters.
The signal is generated by a pyroelectrical detector. It records the incoming IR radiation; the intensity of this radiation is reduced by the absorption of the gas at the according wavelengths. By comparing the radiation of measurement side and reference side of the analysis cell, an alternating voltage signal is developed. This signal results from cooling and heating of the pyroelectrical material of the detector.
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
3 - 1
Page 22
MEASURING PRINCIPLE
Rosemount Analytical
100
Absorption Band
80 90
CO2 (330 ppm)
60 70
50
Transmittance [%]
40
010 20 30
4000 4500 5000 6000
5500
Wavelength [nm]
H2O (20 °C)
6500 7000 7500 8000
3 - 2
Fig. 3-1: Absorption Bands of Sample Gases and Transmittance of the
Interference Filters Used
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 23
Rosemount Analytical
MEASURING PRINCIPLE

3.2 Opto - Pneumatic Measuring Principle (Gas Detector)

A thermal radiator generates the infrared radiation which passes first through a chopper wheel. Then the radiation passes alternately through a filter cell and reaches the measurement and reference sides of the analysis cell with equal intensity.
After passing another filter cell, the radiation reaches the gas detector. The gas detector compares and evaluates the radiation coming from the measurement and reference sides and converts them into voltage signals proportional to their intensity via a preamplifier.
The detector consists of a gas-filled absorption and a compensation chamber, interconnected via a flow channel.
Absorption chamber
Flow channel with Microflow sensor
CaF2 Window
Gas intake connection
Compensation chamber
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Fig. 3-2: Principle Design of Gas Detector
3 - 3
Page 24
MEASURING PRINCIPLE
Rosemount Analytical
Standardly the detector is filled with the infrared active gas to be measured and is only sensitive to this distinct gas with its characteristic absorption spectrum. The absorption chamber is sealed with a window which is transparent for infrared radiation (usually CaF2, calcium fluoride).
When IR radiation passes through the reference side (purged with an unpressurized gas, which does not absorb within a specified spectral range; in most cases nitrogen [N2] is used) of the analysis cell into the detector, no absorption occurs. Thus the gas inside the absorption chamber is heated, it expands and some of it passes through the flow channel into the compensation chamber.
When the IR radiation passes through the measurement side (purged with sample gas) of the analysis cell into the detector, a part of it is absorbed depending on the gas concentration. Then the gas in the absorption chamber is heated less than in the case of the radiation coming from reference side. The absorption chamber cools down, the pressure is reduced, and a part of the gas passes through the flow channel into the measuring chamber.
The flow channel is designed so that it hardly impedes the gas flow by restriction. Due to the radiation of the chopper wheel, the different radiation intensities lead to periodically flow pulses within the detector.
The microflow sensor ev aluates and converts the pulses into electrical voltages. The electronics , which follow, evaluate the signals and convert them into the corresponding display format.
3 - 4
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 25
Rosemount Analytical
123
123
123
6
MEASURING PRINCIPLE

3.3 Technique

The broad emission coming from two IR sources (in the case of dual channel analyzers) passes a chopper blade, then passes through an optical filter and finally enters the analysis cell. The light transmitted through these cells is focused by filter cells onto the detector. The preamplified detector output signals are forwarded to the microprocessor electronics which converts the analytical signals to results expressed directly in physical concentration units such as Vol.-%, ppm, mg/Nm3 etc.
MOTOR
Light source Light source
Filter disc Filter disc
Analysis cell measuring side
Analysis cell reference side
Filter cell Filter cell
Gas detector
23456789012345678234567890123456789012345
CO
2
Preamplifier
H2O
Analysis cell measuring side Analysis cell reference side
Pyroelectrical detector (solid-state detector)
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Chopper blade
Fig. 3-3: Principle Representation
3 - 5
Page 26
MEASURING PRINCIPLE
Rosemount Analytical
3 - 6
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 27
Rosemount Analytical

MAIN FEATURES

4. Main Features
1/4 19" housing, 3 HU
Measurement mode and measuring range switch; absolute or difference mode
Ground level compensation
4 - digit LED - measuring value display and operators prompting via this displays for each
measuring channel
The response time (t90 time) can be adjusted separately for each measuring channel
Plausibility checks
Temperature compensations
Interference compensation f or reducing disturbing effects due to e xtraneous absorption of
secondary gas constituents
Analog signal outputs [0 (2) - 10 V {Option 0 (0,2) - 1 V} / 0 (4) - 20 mA], optically isolated
Digital interface for identification of measurement mode and measuring range
(max. 30 V DC / 30 mA, “Open Collector”, optically isolated)
Datalogger (option): data recording and output
RS 232 C/485 serial interface for data intercommunication with host computers (option)
Self - diagnostic procedures, plus maintenance and servicing support functions
Operator prompting for the avoidance of operator errors
90002927(1) BINOS® 100 4P e [1.00] 17.06.97
4 - 1
Page 28
MAIN FEATURES
Rosemount Analytical
4 - 2
90002927(1) BINOS® 100 4P e [1.00] 17.06.97
Page 29
Rosemount Analytical

PREPARATION

INSTALLATION SITE
5. Preparation
Please check the packing and its contents immediately upon arrival. If any item is damageg or lost you are kindly requested to notify the forwarder to undertake a damage survey and report the loss or damage to us immediately .

5.1 Installation

The analyzer must not operate in explosive atmosphere without supplementary protective measures !
The installation site for the analyzer has to be dry and remain above freezing point at all times. The analyzer must be exposed neither to direct sunlight nor to strong sources of heat. Be sure to observe the permissible ambient temperatures (cf. Section 24: Technical Data).
For outdoor installation, we recommend to install the analyzer in a protective cabinet. At least, the analyzer has to be protected against rain (e.g., shelter).
The analyzer has to be installed as near as possible to the sample point in order to avoid low response time caused by long sample gas lines. In order to decrease the response time, a sample gas pump with a matching high pumping rate may be used. Ev entually , the analyz er has to be operated in the bypass mode or by an o verflow v alve to prevent too high flow and too high pressure (Fig. 5-1).
Exhaust
Bypass valve
Gas sampling pump
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Analyzer
Flow meter
Filter
Fig. 5-1: BINOS® 100 4P, Bypass Installation
Exhaust
5 - 1
Page 30
PREPARATION

GAS CONDITIONING (SAMPLE HANDLING)

Rosemount Analytical
5.2 Gas Conditioning (Sample Handling)
The conditioning of the sample gas is of greatest importance for the successful operation of any analyzer according to extractiv e method.
Only conditioned gas has to be supplied to the analyzer !
The gas has to fulfil the following conditions.
It must be:
free of condensable constituents, free of dust, free of aggressive constituents which are not compatible with the material of the gas
paths, and
have temperatures and pressures which are within the specifications stated in Section 24,
"Technical Data" of this manual.
Inflammable or explosive gas mixtures may not be intr oduced into the analyzer without supplementary protective measures !
When analyzing vapors, the customer has to avoid the precipitation of condensate in the gas paths.
Suitable gas conditioning hardware may be supplied or recommended for specific analytical problems and operating conditions.
5.2.1 Gas Flow Rate
The gas flow rate should be within the range 0.2 l/min to max. 1.5 l/min !
A constant flow rate of about 1 l/min is recommended.
5 - 2
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 31
Rosemount Analytical
PREPARATION

GAS CONNECTIONS

5.3 Gas Connections
All the fittings for gas line connections are placed just on the rear panel of the analyzer and are clearly marked (Fig. 5-2 and Fig. A-2):
IN = gas inlet
OUT = gas outlet
M = measurement side
R = reference side
Do not interchange gas inlets and gas outlets !
The exhaust gas lines have to be mounted in a declining, pressureless and frost-free way and according to the valid emission legislation!
Zero gas, span gas and reference gas are introduced directly via the respectiv e gas inlet. The test gas containers hav e to be set up according to the current legislation. Be sure to observe the safety regulations f or the respective gases !
M1 M1 M2 M2
R1 R1 R2 R2
IN
OUT OUTIN
K1 K2
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
X1 OUTPUT
INTERFACE
ABS./DIFF.
Fig. 5-2: Gas Connections BINOS® 100 4P
5 - 3
Page 32
PREPARATION
Rosemount Analytical
5 - 4
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 33
Rosemount Analytical

SWITCHING ON / SWITCHING OFF

SWITCHING ON

6 . Switching On / Switching Off
6.1 Switching On
Once the analyzer has been correctly assembled and installed in accordance with the general instructions of Section 5. “Preparation”, the analyzer is ready for operation.
The analyzer is specified for an operating v oltage of 24 V DC (+ 20 % / - 50 %). Operation from 230 / 120 V AC requires the 24 V DC supply via VSE 2000 or equivalent power supply.
IN
X1 OUTPUT
INTERFACE
X2 OUTPUT
OUT OUTIN
K1 K2
24 VDC
MADE IN GERMANY
ABS./DIFF.
X3 OUTPUT
Plug
24 V DC
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Fig. 6-1: Supply Voltage BINOS® 100 4P
6 - 1
Page 34
SWITCHING ON / SWITCHING OFF
SWITCHING ON
Rosemount Analytical
Connect mains line and power supply.
Verify bef orehand that the line v oltage stated on the power suppl y agrees with that of your power supply line !
Connect power supply and analyzer (Fig. 6-1, Plug 24 V DC).
Verify correct polarity bef ore operation !
The presence of the supply voltage will be indicated by the illumination of the LED displays. Upon connection of the supply voltage, the analyzer will perform a self - diagnostic test routine. First the actual program version will be shown.
Finally either concentration values or error messages will be displayed.
If as a result of a battery fault the default values have been charged, this will be shown by a flushing
“batt.”! This message will disappear after depressing any ke y.
Analyzer warming-up takes about 15 to 50 minutes depending on the installed detectors !
Before starting an analysis, however, the following should be performed:
entry of the desired system parameters, calibrate the analyzer.
NOTE: The "X’s" sho wn in the displa y indicate a number or a combination of numbers.
6 - 2
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 35
Rosemount Analytical
SWITCHING ON / SWITCHING OFF

SWITCHING OFF

6.2 Switching Off
Before s witching off the analyzer, we recommend first flushing the gas lines for about 5 minutes with zeroing gas (N2) or adequate conditioned air. The full procedure for shutting down the analyzer is as follows:
Admit zeroing gas at the gas inlet fittings.
Set the gas flow rate to allowa b le rate.
After 5 minutes have elapsed:
Shut off the zeroing gas supply.
Switch off the analyzer by disconnecting the v oltage supply.
Close all gas line fittings immediately.
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
6 - 3
Page 36
SWITCHING ON / SWITCHING OFF
Rosemount Analytical
6 - 4
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 37
Rosemount Analytical

KEY FUNCTIONS

7. Key Functions
The operation and programming of the analyzer is perf ormed using the membrane - type keypad with its four keys (see Fig. A-1, Item 3 - 6).
Operator guidance prompts will appear on the 4 - digit LED - displays.
During power failure all parameters entered are secured by a battery-supplied buffer.
90002927(1) BINOS® 100 4P e [1.00] 17.06.97
7 - 1
Page 38
KEY FUNCTIONS

FUNCTION

Rosemount Analytical
7.1 FUNCTION
By pressing this key (Fig. A-1, Item 3) you can roll through all active functions of the analyz er. The selection of any of these functions does not lead to any action; the analyzer remains in the measurement mode. The following analyz er functions and their sequences (see also Fig. 7-1) are shown:
Zeroing channel 1
Zeroing channel 2
Recording of data in datalogger
Interval time for automatic recording of data in datalogger
Output of data from datalogger
Only with Option Serial Interface and if “P.Set = Yes”
Deleting of data in datalogger
Time setting and display
7 - 2
Selection of measurement mode and enabling of parameters
Spanning channel 1
Only if “tYPE = AbS.”
Spanning channel 2
90002927(1) BINOS® 100 4P e [1.00] 17.06.97
Page 39
Rosemount Analytical
KEY FUNCTIONS

ENTER

Entry of system parameters.
Entry of serial interface parameters
Only if “P.SET= YES”
With Option “Serial Interface” only
7.2 ENTER
The ENTER - key (Fig. A-1, Item 4) is used for the transf er of (keyed - in) numerical data to the corresponding operating parameters and for the initiation of certain operations, such as zeroing and spanning.
Several functions ma y be accessed only b y authorized persons. Theref ore these functions ha ve been secured by codes. When a code-secured functions is being selected for the first time, while rolling through a sequence of functions (starting at "0 -1"), and after pressing ENTER - key
will appear on the display.
Now the code must be entered. If the entered code is incorrect, the display remains unchanged, and the process of entering the code starts again with "0". When the code has been entered correctly, then you have access to the secured lev el.
This code has been set to the value “1” in our plant bef ore shipment.
90002927(1) BINOS® 100 4P e [1.00] 17.06.97
7 - 3
Page 40
7 - 4
ENTERFUNCTION
- - - -
- - - -
0..-1 0..-.1
0 -1
1
Fig. 7-1: BINOS
2
3
4
®
100 4P, Operating Function Matrix
5
Nominal
Calibrate
Value
0 -2 0..-2 0..-.2
Nominal
Calibrate
Value
d.in
tdin
d. Out
Zero Adjustment Channel 1 Zero Adjustment Channel 2
Data Input
Automatic Data Input
Interval-Time (min.); 0 = off
Data-Logger
Data Output
KEY FUNCTIONS
KEY FUNCTION OVERVIEW
6
7
8
9
90002927(1) BINOS
10
11
®
100 4P e [1.00] 17.06.97
PArA
12
d.Clr
t.SEt
SEL. tYPE Gnd1 Gnd2 P.SEt A.Out
Measuring
S -1 CodE S.-1 S..-1 S..-.1
CodES -2 S.-2 S..-2 S..-.2
CodESYS.-
CodESIP. On.-L
d.ClrCodE
Clear
MinHour
Set
Hours
Minutes
Reference Channel 1
Mode
Actual
Value
Actual Value
PrES toL. d.OFF L -0
Pressure
Clear Data
Set
Setting Time
Reference Channel 2
Nominal
Value
Nominal
Value
Air-
Tolerance
Id.No 232C
RS485-
On Line
Address
Parameter
display
Calibrate
Calibrate
Display
off
RS23 2C /
RS485
only with Option "Serial Interface" and with P.SEt =YES
Std./SCOP
anal. Output
0-1A
0 V abs.
Channel 1
1-1A 0-1d 1-1d 0-2A 1-2A 0-2d 1-2d
1 V abs.
0 V diff.
Channe l 1
Channel 1
1 V diff.
Chann el 1
Span Adjustment Channel 1 Span Adjustment Channel 2
t90.2 t90.3 t90.4 rES. C.On C.CAL ES. On P.-No S.-No S.-No CoDE FAC1 FAC2 4P.On
t90-Time
t90-Tim e Ch1 diff.
t90-Time Ch2 abs.
0 = 4800 1 = 2400 2 = 1200 3 = 600
Ch1 abs.
LPb.bAUd
longitud.-
parity
bAU d:
Live-Zero
Baudrate
tPur t90.1
Purge-
Delay
Echo
Echo
(RS232C)
0 V abs.
t90-Tim e
Ch2 diff.
1 V abs.
Chann e l 2
only with tYPE = AbS.
Reset of
Device
Chann e l 2
Parameter Serial Interface
0 V diff.
Channel 2
Cross-Corr.
on /o ff
1 V diff.
Chann e l 2
calibrate
Cross-Corr.
only with A.Out = SCOP
ext. Switch
Program-
active
Version
only with Option "Serial Interface"
Select Function
Serial-Nr.
Part 1
Serial-Nr.
Part 2
not
active
only with P.SEt = YES
System-Param eter
User-
Bottle-
Code
factor Ch1
Bottle-
factor Ch2
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
automatic
Reference
Rosemount Analytical
Page 41
Rosemount Analytical
KEY FUNCTIONS

INPUT - CONTROL

7.3 INPUT - CONTROL
By pressing this keys (Fig. A-1, Item 5 and 6) any parameter value entered may be changed. Depending on which key is pressed, the value is either enhanced or diminished by one digit.
UP previous v alue + 1 digit
DOWN previous v alue - 1 digit
Example: To setup the requested parameter value, the displayed characters can be adjusted by pressing each of the key, respectively. Now the correct value can be acknowledged by pressing ENTER on the secured level.
A suiting value range - which cannot be exceeded - belongs to each parameter. In addition all entries are subjected to a plausibility check as added protection against operator errors.
The counting speed can be increased by pressing the <UP> or <DOWN> k ey continuously . When rolling through the setup values from 0 to maximum or minimum value, the standard counting speed is activated to make the entering process easier.
Approx. 60 - 120 seconds after pressing any key, the analyzer automatically returns to the measuring mode.
90002927(1) BINOS® 100 4P e [1.00] 17.06.97
7 - 5
Page 42
KEY FUNCTIONS
Rosemount Analytical
7 - 6
90002927(1) BINOS® 100 4P e [1.00] 17.06.97
Page 43
Rosemount Analytical

ENABLING / DISABLING SYSTEM PARAMETER FUNCTION

ENTRY OF SYSTEM PARAMETERS

8. System Parameters

8.1 Enabling / Disabling System Parameter Function
It might not be necessary to have access to system parameter / spanning function. The appearance of this function may e ven disturb. Therefore, it is possible to skip this le v el completely.
Depress the key
until the text appears.
Depress the key
If the Code has not already been entered, the following display
will appear
Use the keys and to select the Code.
The display will show:
Depress the key
until the text appears.
Entry of “YES”: System parameter / spanning function is enabled Entry of “NO”: System parameter / spanning function is disabled
Effect the entry by using
and
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
8 - 1
Page 44

ENTRY OF SYSTEM PARAMETERS

ENTRY OF SYSTEM PARAMETERS
8.2 Entry of System Parameters
When the system parameters have to be entered or changed,
depress the key
until the display appears.
Rosemount Analytical
Depress the key
If the Code has not already been entered, the following display
will appear
Use the keys to select the Code.
and then depress
8 - 2
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 45
Rosemount Analytical
PRESSURE CORRECTION / TOLERANCE CHECK
ENTRY OF SYSTEM PARAMETERS
8.2.1 Pressure Correction
To eliminate faulty measurements due to changes in barometric pressure or sample gas pressure, the operator is offered the opportunity to enter the current pressure expressed in hPa (mbar) in a range of 800 to 1300 hPa. The concentration values computed by the analyzer will then be corrected to reflect the barometric pressure or sample gas pressure resp. entry.
The correction factor is calculated via following formula:
Correction factor =
1024 hPa
actual pressure
The entry is effected by using and
8.2.2 Tolerance Check
The tolerance function is for the activation and deactivation of the tolerance check procedure for various calibration gases.
If the tolerance check procedure has been activated, the microprocessor will verify during calibration procedures whether the used calibration gas shows a deviation of more than10 % from measuring range of zero (zero - level) or more than 10 % of the nominal concentration value entered resp. (span).
If this tolerance is exceeded, no calibration will be performed, and an error message will appear (see Section 13).
Entry of “YES”: Tolerance check is deactivated. Entry of “NO”: Tolerance check is activated.
Perform the entry by using and
To calculate the 10% value the measuring range of "AbS.2" is taken over from
"AbS.1" while being in absolute measuring range.
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
8 - 3
Page 46
ENTRY OF SYSTEM PARAMETERS
DISPLAY OFF / ANALOG SIGNAL OUTPUTS
Rosemount Analytical
8.2.3 Display Off
If "YES" is entered, the DISPLAY will be deactivated about 1 to 2 minutes after the last key depression. If an y key is depressed while the DISPLA Y is deactivated, all displa y elements will be reactivated without any further operation being initiated.
Entry of “YES”: Display will be deactivated Entry of “NO”: Display will not be deactivated
Entry is performed by using
followed by
8.2.4 Analog Signal Outputs
The analog signal outputs (optically isolated) are brought out to the 9-pin subminiature D­connector X2 on the analyzer rear panel.
Entry of “NO”: Output signal of 0 - 10 V (Option: 0 - 1 V) / 0 - 20 mA. Entry of “YES”: Output signal of 2 - 10 V (Option: 0.2 - 1 V) / 4 - 20 mA (life zero mode)
Use the keys
and for entry .
Note:
The begin of range concentration and the end of range concentration are freely programmable (see Sections 9.2 and 10.3). For type of voltage output (standard or option) look at order confirmation or identify plate.
8 - 4
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 47
Rosemount Analytical
ENTRY OF SYSTEM PARAMETERS
ANALOG SIGNAL OUTPUTS
M1 M1 M2 M2
R1 R1 R2 R2
Mating socket X 2
IN
X1 OUTPUT
INTERFACE
X2 OUTPUT
OUT OUTIN
K1 K2
24 VDC
MADE IN GERMANY
ABS./DIFF.
X3 OUTPUT
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Fig. 8-1: Mating Socket X 2 (Analog Signal Outputs)
5
1
6
9
1 2 0 (2) - 10 V DC [Option: 0 (0,2) - 1 V DC], channel 1 3 0 (4) - 20 mA, channel 1 (R 4 0 (2) - 10 V DC [Option: 0 (0,2) - 1 V DC], channel 2 5 0 (4) - 20 mA, channel 2 (R 6 7 8 9
⊥⊥
(V DC)
⊥⊥
⊥⊥
(mA)
⊥⊥
Fig. 8-2: Pin Assignments X 2 (Analog Signal Outputs)
500 )
B
500 )
B
8 - 5
Page 48
ENTRY OF SYSTEM PARAMETERS
FLUSHING PERIOD / RESPONSE TIME
Rosemount Analytical
8.2.5 Flushing Period
For calibration the gas paths must be supplied with sufficient calibration gas . The flushing period has to be fixed adequately; perform calibration only after a suitable flushing period (the calibration gas flow should be identical with sample gas flow).
This period may be selected in the range 0 - 99 sec. depending on calibration conditions.
Use the keys
and for entry.
8.2.6 Response Time (t90)
For some types of analysis an alteration of the analyzer damping factor , i.e. its electrical response time, t90, may be required. The operator is offered the option of selecting a response time optimal for each application.
The range of accepted entries is 2 - 15 sec.
Entry possibility for channel 1 (absolute measurement),
Use the key
and for the entry .
8 - 6
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 49
Rosemount Analytical
ENTRY OF SYSTEM PARAMETERS
RESPONSE TIME
Entry possibility for channel 1 (differential measurement).
Use the keys
and for the entry .
Entry possibility for channel 2 (absolute measurement).
Use the key
and for the entry .
Entry possibility for channel 2 (differential measurement).
Use the key
and for the entry .
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
8 - 7
Page 50
ENTRY OF SYSTEM PARAMETERS
RESET
Rosemount Analytical
8.2.7 Reset
The reset operation restores the settings of the analyzer to the parameters and calibration factors set in our factory at the time of its manufacture.
This is equivalent to switching off the electrical supply line and switching off the battery buffering of the RAM’s b y removing the battery jumper, J7.
All parameters and calibration factors entered by the user will be lost whenever a reset operation is performed.
The currently valid user identification code must be entered before a reset will be executed; this will prevent inadvertent resets.
Entry is performed by using
followed by
Whenever a reset operation is initiated, the analyz er operating program will be restarted, just as it is when the instrument is switched on at first (see Section 6).
Jumper J6, which activates the watchdog circuitry, must be inserted if the reset operation is to be executed correctly
8 - 8
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 51
Rosemount Analytical
CROSS-COMPENSATION / CROSS-COMPENSATION CALIBRATION
ENTRY OF SYSTEM PARAMETERS
8.2.8 Cross - Compensation
This control permits switching the electronic cross-compensation feature on and off. T h e cross-compensation feature is designed to minimize mutual interferences between the two gases (e.g., CO2 and CO) measured by the analyzer .
Entry of “NO”: Cross-compensation is disabled Entry of “YES”: Cross-compensation is enabled
Effect the entry by using
and
After changing this parameter a new spanning must be accomplished.
8.2.9 Cross - Compensation Calibration
Determination of cross-compensation correction factors is performed during the span adjustment. Pure test gases are required for this operation. Once cross-compensation corrections have been determined, span adjustments may be performed using test gas mixtures.
Entry of “NO”: Spanning without cross-compensation correction (test gas mixtures) Entry of “YES”: Spanning with cross-compensation correction (pure test gases)
Effect the entry by using
and
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
8 - 9
Page 52
ENTRY OF SYSTEM PARAMETERS
CROSS-COMPENSATION CALIBRATION / EXTERNALLY LOCATED SWITCH
Rosemount Analytical
To perform a calibration with cross-compensation correction, proceed as follows:
First perform a zeroing for both analysis channels (see Section 9.3.1). Then perform a spanning for both analysis channels as described in Section 9.3.2.
The spanning for the first of the analysis channels calibrated must then be repeated.
The entries described in Sections 8.2.8 and 8.2.9 must be “YES” for performanceof a calibration with cross-compensation correction ! Use only pure test gases !
When using test gas mixtures, “C.Cal” m ust be set to “NO” !
Spanning may be carried out only in the absolute measurement mode !
8.2.10 Externally Located Switch
Sometimes it is necessary to switch the measurement mode and/or measuring range of analog outputs by a separately located unit (see Section 11.2), i.e. not in the “SEL” menu. By entering “YES”, this function has been activated. Now it is no longer possible to change either measurement mode or measuring range in the “SEL ” menu; only the actual setup will be displayed.
Entry of “NO”: External switch (option) is disabled Entry of “YES”: External switch (option) is enabled
Effect the entry by using
and
8 - 10
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 53
Rosemount Analytical
ENTRY OF SYSTEM PARAMETERS
8.2.11 Program Version
The Program Version (No. of the installed software version) will be displayed.
Depress the key
8.2.12 Serial - No.
The Serial - No. will be displayed. (Please note this number for fur ther contact with our factory regarding maintenance, service etc.).
Depress the key
Continuation of Serial - No.
Depress the key
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
8 - 11
Page 54
ENTRY OF SYSTEM PARAMETERS
USER CODE / GAS - CYLINDER CORRECTION FACTOR
Rosemount Analytical
8.2.13 User Code
The value 1 has been set in our plant.
To prevent parameter alterations by unauthorized persons, the operator may specify another password (user code).
Use the keys
and for entry .
Please take care when completing the user code.
8.2.14 Gas - Cylinder Correction Factor
For differential measurement, it is sometime necessary to adjust the signal value to the set value. This is possible due to the gas - cylinder correction factor: using this factor , the sensitivity can be calibrated again during the differential measurement.
Example: FAC .1old: 1.000 Difference to set value: +38 Difference to actual value: +40 ppm
FAC1new: = = 0.950
Diff. to set value x FAC.1old 38 x 1.000
Diff. to actual value 40
Note:
Usually it is not necessary to adjust a gas - cylinder correction factor of 1.000. The difference between the set and the actual values may happen due to the test gas tolerances or during the analyzer calibration.
8 - 12
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 55
Rosemount Analytical
Cylinder correction factor f or channel 1
Entry possibility for
(differential measurement).
Use the keys
and for entry.
ENTRY OF SYSTEM PARAMETERS
GAS - CYLINDER CORRECTION FACTOR
Entry possibility for
cylinder correction factor for channel 2 (differential measurement).
Use the keys
and for entry.
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
8 - 13
Page 56
ENTRY OF SYSTEM PARAMETERS
GROUND LEVEL MODE
Rosemount Analytical
8.2.15 Ground Level Mode
With this parameter the type of ground level determination can be selected for differential measurement.
Entry of “YES”: The concentration present when switching over from absolute to
differential measurement is accepted by the analyzer as the valid ground level for differential measurement.
In addition the following gases must be present in the analyzer: Reference side: Zero gas
Entry of “NO”: The analyzer uses the concentration values set for parameter
When the system parameter “4P.On” is changed from “NO” to “YES” and the differential measurement mode is already active, then the analyzer continues to use the preset reference concentration v alues. To be able to change the ground lev el automatically, the analyzer must be switched over from absolute to differential measurement mode at first.
When "YES" has been selected for “4P.On”, then the parameters “Gnd1” and “Gnd2” (Section 10.2) show only the preset ground lev el. These parameters cannot be adjusted by hand any more.
Use the keys
Measurement side: Reference gas
“Gnd1” and/or “Gnd2” (can be found in the “SEL.” function line, see Item 10.2) as ground level for differential measurement.
and for entry.
Depress the key until
the display shows The analyzer is now is back to analysis
mode.
8 - 14
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 57
Rosemount Analytical

SETUP OF THE ABSOLUTE MEASUREMENT MODE

ABSOLUTE MEASUREMENT

9. Absolute Measurement
The analyzer can measure in two different modes: absolute and differential.
In absolute measurement mode the reference side of the analyzer is purged with an unpressurized gas, which does not absorb within a specified spectral range (flow: approx. 1 l/min)! In most cases nitrogen (N2) is used.
9.1 Setup of the Absolute Measurement Mode
To setup the absolute measurement mode, proceed as follows:
Depress the key
until the text appears.
Depress the key
If the Code has not already been entered, the following display
will appear
Use the keys to select the Code.
and then depress
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
9 - 1
Page 58
ABSOLUTE MEASUREMENT
SETUP ABSOLUTE MEASUREMENT MODE / SETTING ANALOG OUTPUTS
Rosemount Analytical
The display will show:
Use the keys and to select the variable “AbS. ”
Now the absolute measurement mode is active.
It is not possible to use different measurement modes for the two measuring channels. When "YES" has been chosen for the system parameter "ES.On" (i.e. external switch is active), then the switching over can be carried out only via this external switch. Only the actual setting is displayed on the monitor no w.
The display will show:
Depress the key until the display shows

9.2 Setting Analog Outputs

While setting the analog outputs, it can be chosen between standard values ("Std.") and a freely programmable magnifier (Scope, "SCOP").
Entry of “Std. ”: The factory range settings are taken ov er . The preset voltage and (standard) current ranges correspond to the following concentration range:
[0 ... End of range (1000 ppm or 3.0 % resp.)] When "Std." has been chosen, the "Scope" parameters will not be displayed any longer.
Entry of “SCOP”: The concentration range can be freely set. (scope) To select the variable and enter the desired range:
use the keys and
9 - 2
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 59
Rosemount Analytical
ABSOLUTE MEASUREMENT
ANALOG OUTPUT SETTING
When “SCOP” has been entered appears on the display.
Now the value for "channel 1 absolute" can be selected: it represents the begin of voltage/current range.
Entry is effected by using and
It appears
Now the value for "channel 1 absolute", which represents the end of voltage/current range, can be entered.
Example: “Life Zero” has been selected (system parameter “L-0” = “YES”). The analog output is being used. “0-1A” = 200 ppm; “1-1A” = 600 ppm
Concentration [ppm] 2 00 300 400 500 600 Current [mA] 4 8 12 16 20
If the same value has been selected for "0-1A" as for "1-1A", then 1 digit will be added automatically to the "1-1A" value.
The values for "channel 2 absolute"
must be entered, too.
When "YES" has been chosen for the system parameter "ES.On" (i.e. external switch is active), then the switching over from "Std." to "SCOP" can be carried out only via this external switch. Only the actual setting ma y be displayed on the monitor now.
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
9 - 3
Page 60
ABSOLUTE MEASUREMENT

CALIBRATION

Rosemount Analytical
9.3 Calibration
T o ensure correct measurement results, zeroing and spanning should be carried out once a week. Spanning can be performed only after zeroing.
For the calibration procedure the required test gases have to be fed to the analyzer through the respective gas inlets (cf. Section 5.3) with a no-back-pressure gas flow rate of about 1 l/min (the same as with sample gas) !
After switching on the analyzer, wait at least 15 to 50 minutes (depending on installed detectors) before admitting gas to the analyzer !
Zeroing
For zeroing, the measurement side of the analysis cell has to be purged with the same gas as its reference side.
Spanning
The span gas concentration should be in a range of 80 % - 110 % of full - scale range ! For lower span gas concentrations the measuring accuracy could be lower for sample gas concentrations, which are higher than the span gas concentration !
When using span gas mixtures the entry for “C.Cal” must be set to “NO” (see Section 8.2.9) !
9 - 4
The correct pressure must be entered before performing the calibration, if you want to have the possibility of pressure correction (see 8.2.1) !
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 61
Rosemount Analytical
ABSOLUTE MEASUREMENT
9.3.1 Zeroing
Zeroing will set the actually measured gas concentration to “zero”.
Depress the key
until the display shows (zeroing channel 1) or
ZEROING
(zeroing channel 2) resp .
Depress the key
The following display will appear
Use the keys to select the correct user code
and acknowledge by depressing
The display will show or resp.
The actual zero level will be displayed.
Wait until the entered flushing period and t90 time elapses.
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
9 - 5
Page 62
ABSOLUTE MEASUREMENT
ZEROING
Rosemount Analytical
Depress the key
The nominal value or will be displayed.
If the actual and nominal zero levels are identical, the next function can be selected then by using the FUNCTION - key (without zeroing).
If the two values differ , then
depress the key
The actual measuring value or will be displayed
T o start zeroing press again.
As soon as zeroing has finished,
the actual measuring value or resp. will be displayed.
To leave “calibration mode” press
9 - 6
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 63
Rosemount Analytical
ENABLING / DISABLING SPANNING FUNCTION
ABSOLUTE MEASUREMENT
9.3.2 Spanning
9.3.2a Enabling / Disabling Spanning Function
It might not be necessary to have access to system parameter / spanning function. The appearance of this function may ev en disturb. Theref ore , it is possib le to skip this level completely.
Depress the key
until the text appears.
Depress the key
If the Code has not already been entered, the following display
will appear
Use the keys and to select the Code.
The display will show:
Depress the key
until the text appears.
Entry of “YES”: System parameter / spanning function is enabled Entry of “NO”: System parameter / spanning function is disabled
Effect the entry by using
and
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
9 - 7
Page 64
ABSOLUTE MEASUREMENT
SPANNING
Rosemount Analytical
9.3.2b Performing of Spanning
V erification of the span calibration is essential f or accurate concentration measurement. Spanning can be performed only after zeroing. Spanning will set the actually measured gas concentration to the entered “span gas setpoint”.
Note: The span gas concentration should be in a range of 80 % - 110 % of full - scale range !
For lower span gas concentrations the measuring accuracy could be lower for sample gas concentrations, which are higher than the span gas concentration !
When using span gas mixtures the entry for “C.Cal” must be set to “NO” (see Section 8.2.9) ! The correct pressure must be entered before performing the calibration, if you want to have the possibility of pressure correction (see 8.2.1) !
Depress the key
until the display shows (spanning channel 1) or
(spanning channel 2) resp.
Depress the key
Enter the correct user code, if not already entered
The display will show or resp.
The actual concentration level will be displayed.
Wait until the entered flushing period and t
9 - 8
time elapses.
90
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 65
Rosemount Analytical
ABSOLUTE MEASUREMENT
Depress the key
The test gas setpoint or resp. will be displayed.
If necessary , enter the true test gas setpoint value (taken from the manuf acturer’s certification on the gas bottle) by
using the keys
SPANNING
and
The actual measuring value or resp. will be displayed
Leave calibration mode by pressing the FUNCTION - key (enter of nominal value without span calibration)
or press again to start spanning.
As soon as spanning has finished, the display indicates
the actual measuring value or resp.
To leave calibration mode press
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
9 - 9
Page 66
ABSOLUTE MEASUREMENT
MEASUREMENT
9.4 Measurement
Depress the key until
the display shows
The analyzer is now in the analysis mode.
Rosemount Analytical
In absolute measurement mode the reference side of the analyzer is purged with an unpressurized gas, which does not absorb within a specified spectral range. In most cases nitrogen (N2) is used for reference side. The measurement side is purged with sample gas (flow: approx. 1.0 l/min)!
Analyzer warming-up after switching on takes about 15 to 50 minutes, depending on the installed detectors !
Admit reference gas (nitrogen, N2) at the gas inlet fitting of reference side.
Admit sample gas at the gas inlet fitting of measuring side.
Set the gas flow rate to approx. 1 l/min.
Note !
No matter which program lev el is active , the program automatically returns to the measurement level 1 or 2 minutes after the last pressing of a key !
The right-hand decimal point of the display is pulsing to indicate absolute measurement mode.
It is possible to identify the measurement mode of the analyzer and measuring range of the analog outputs by digital outputs (see Section 11).
9 - 10
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 67
Rosemount Analytical

DIFFERENTIAL MEASUREMENT

GROUND LEVEL SETTINGS
10. Differential Measurement
The analyzer can measure in two different modes: absolute and differential.
In differential measurement mode the reference side of the analyzer is purged with an unpressur­ized reference gas, the measurement side is purged with sample gas (flow: approx. 1.0 l/min)!

10.1 Ground Level Mode

In this measurement mode, the reference concentration, which is used by the analyzer to measure the difference, is of great importance: it is the basis f or the differential measurement. The analyzer offers two possibilities to determinate the ground level.
Depress the key
until the display appears.
Depress the key
If the Code has not already been entered, the following display
will appear
Use the keys to select the Code
and then depress
The display will show:
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
10 - 1
Page 68
DIFFERENTIAL MEASUREMENT
GROUND LEVEL SETTINGS
Rosemount Analytical
Depress the key
until the text appears.
10.1.1 Automatic Determination
In this mode, the analyzer takes over the values present before the switching over from absolute to differential measurement.
Entry of “YES”: The concentration present when switching over from absolute to
differential measurement is accepted by the analyzer as the valid ground level for differential measurement.
In addition the following gases must be present in the analyzer: Reference side: Zero gas Measurement side: Reference gas
When the system parameter “4P.On” is changed from “NO” to “YES” and the differential measurement mode is already active, then the analyzer continues to use the preset reference concentration v alues. To be able to change the ground lev el automatically, the analyzer must be switched over from absolute to differential measurement mode at first.
When “YES” has been selected for “4P.On”, then the parameters “Gnd1” and “Gnd2” (section 10.2) show only the preset ground le vel: These parameters cannot be adjusted by hand any more.
Use the keys and to select the variable “Yes. ”
Depress the key until
the display shows The analyzer is now back to analysis
mode.
10 - 2
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 69
Rosemount Analytical
DIFFERENTIAL MEASUREMENT
GROUND LEVEL SETTINGS
10.1.2 Determination of Ground Level by Selecting Parameters
When this method is used, the ground level limits are preset values. This measurement mode does not require to be preset before switching over; it can be selected during a differential measurement as well.
Entry of “NO”: The analyzer uses the concentration values set for parameter
“Gnd1” and/or “Gnd2” (can be found in the “SEL.” function line, see Item 10.2) as ground level for differential measurement.
Use the keys and to select the v ariable “No”
Depress the key until
the display shows The analyzer is now back to analysis
mode.
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
10 - 3
Page 70
DIFFERENTIAL MEASUREMENT

SETUP OF THE DIFFERENTIAL MEASUREMENT MODE

10.2 Setup of the Differential Measurement Mode
To setup the differential measurement mode, proceed as follo ws:
Depress the key
until the text appears.
Depress the key
Rosemount Analytical
If the Code has not already been entered, the following display
will appear
Use the keys to select the Code
and then depress
The display will now show:
Use the keys and to select the variable “Diff. ”
Now the differential measurement mode is active. After switching over to differential measurement mode, the reference side must be purged with reference gas.
It is not possible to use different measurement modes for the two measuring channels. When "YES" has been chosen for the system parameter "ES.On" (i.e. external switch is active), then the switching over can be carried out only via this external switch. Only the actual setting is displayed on the monitor no w.
10 - 4
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 71
Rosemount Analytical
SETUP DIFFERENTIAL MEASUREMENT MODE / SETTING ANALOG OUTPUTS
DIFFERENTIAL MEASUREMENT
10.2.1 Manual Ground Level Setting
Use the keys and to set the variable to concentration of
the reference gas in channel 1.
Use the keys and to set the variable to concentration of
the reference gas in channel 2.
The display will show:
Depress the key

10.3 Setting Analog Outputs

While setting the analog outputs, it can be chosen between standard values ("Std.") and a freely programmable magnifier (Scope, "SCOP").
Entry of “Std. ”: The factory range settings are taken ov er . The preset voltage and (standard) current ranges correspond to the following concentration range:
Begin of Range End of Range Channel 1 - 50 ppm + 50 ppm Channel 2 - 1 % + 1 %
When "Std." has been chosen, the "Scope" parameters will not be displayed any longer.
Entry of “SCOP”: The concentration range can be set freely. (scope)
Use the keys and to select the variable.
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
10 - 5
Page 72
DIFFERENTIAL MEASUREMENT
SETTING ANALOG OUTPUT
Rosemount Analytical
When “SCOP” has been entered appears on the display.
Now the value for "channel 1 differential" can be selected which represents the begin of voltage/ current range.
Entry is effected by using and
It appears
Now the value for "channel 1 differential" can be entered which represents the end of voltage/ current range.
Example: “Life Zero” has been selected (system parameter “L-0” = “YES”). The analog output is being used. “0-1d” = -20 ppm; “1-1d” = +20 ppm
Concentration [ppm] -20 -10 0 10 20 Current [mA] 4 8 12 16 20
If the same value has been selected for "0-1d" as for "1-1d", then 1 digit will be added automatically to the "1-1d" value.
The values for "channel 2 differential"
must be entered, too.
When "YES" has been chosen for the system parameter "ES.On" (i.e. external switch is active), then the switching over from "Std." to "SCOP" can be carried out only via this external switch. Only the actual setting ma y be displayed on the monitor now.
10 - 6
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 73
Rosemount Analytical
DIFFERENTIAL MEASUREMENT

CALIBRATION (ZEROING)

10.4 Calibration (Zeroing)
To ensure correct measurement results, zeroing should be carried out once a week. Zeroing will set the actually measured gas concentration to “zero”.
For zeroing the measurement side of the analysis cell has to be purged with the same gas as its reference side through the respective gas inlets (cf. Section 5.3) with a no-back-pressure gas flow rate of about 1 l/min (the same as with sample gas) !
After switching on the analyzer, wait at least 15 to 50 minutes (depending on installed detectors) before admitting gas to the analyzer !
Depress the key
until the display shows (zeroing channel 1) or
(zeroing channel 2) resp .
Depress the key
The following display will appear
Use the keys to select the correct user code
and acknowledge by depressing
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
10 - 7
Page 74
DIFFERENTIAL MEASUREMENT
CALIBRATION (ZEROING)
Rosemount Analytical
The displays will show or resp.
The actual zero level will be displayed.
Wait until the entered flushing period and t90 time elapses.
Depress the key
The nominal value or will be displayed.
If the actual and nominal zero levels are identical, the next function can be selected then by using the FUNCTION - key (without zeroing).
If the two values differ , then
depress the key
The actual measuring value or will be displayed
To start zeroing depress again.
As soon as zeroing has finished,
the actual measuring value or resp. will be displayed.
To leave “calibration mode” depress
10 - 8
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 75
Rosemount Analytical
10.5 Measurement
Depress the key until
the display shows
The analyzer now is in the analysis mode.
DIFFERENTIAL MEASUREMENT

MEASUREMENT

In differential measurement mode the ref erence side of the analyzer is purged with an unpressur­ized reference gas, the measurement side is purged with sample gas (flow: approx. 1.0 l/min)!
Analyzer warming-up after switching on takes about 15 to 50 minutes, depending on the installed detectors !
Admit reference gas at the gas inlet fitting of reference side.
Admit sample gas at the gas inlet fitting of measuring side.
Set the gas flow rate to approx. 1 l/min.
Note !
No matter which program lev el is active, the prog ram automatically returns to the measurement lev el 1 or 2 minutes after the last pressing of a k ey!
The maximum absolute concentration of the gas to be measured is the upper limit of the measuring range for “AbS”. Above this limit linearization errors may occur. (Abs. Concentration of the gas measured = Ref. Concentr ation + Diff. Concentration)
The right-hand decimal point of the display is switched off during differential measurement mode .
It is possible to identify the measurement mode of the analyzer, measuring range of the analog outputs and the ground level mode b y digital outputs (see Section 11).
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
10 - 9
Page 76
DIFFERENTIAL MEASUREMENT
Rosemount Analytical
10 - 10
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 77
Rosemount Analytical
11. Digital Outputs / Inputs
All analyzer digital outputs and digital inputs appear on the rear panel (Fig. 11-1).
M1 M1 M2 M2
R1 R1 R2 R2

DIGITAL OUTPUTS / INPUTS

IN
X1 OUTPUT
INTERFACE
X2 OUTPUT
OUT OUTIN
K1 K2
24 VDC
MADE IN GERMANY
LEMOSA Socket
(Digital Inputs)
ABS./DIFF.
X3 OUTPUT
Plug X 3
(Digital Outputs)
Fig. 11-1: Rear Panel of BINOS® 100 4P (Digital Outputs / Digital Inputs)
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
11 - 1
Page 78
DIGITAL OUTPUTS / INPUTS
Rosemount Analytical

11.1 Digital Outputs

All digital outputs of the analyzer are brought out to plug X 3 on the rear panel (Fig. 11-1). The loading of the outputs (“Open Collector”) is max. 30 V DC / 30 mA.
1
5
9
6
1 Measuring mode (absolute / differential) 2 Analog output (“standard” / “scope”) 3 Ground level mode “4P-On” (on / off) 4 not used 5 6 not used 7 not used 8 not used 9 not used
⊥⊥
⊥⊥
Fig. 11-2: Pin - Assignments X 3 (Digital Outputs)

11.2 Digital Inputs

T o switch the measurement mode and/or measuring range of analog outputs via externally located switches (see Section 8.2.10) and not in the “SEL ” menu, the necessary digital inputs are brought out to the LEMOSA socket on the rear panel of the analyzer (Fig. 11-1). The loading of the customers contacts is min. 10 V DC / 1 mA.
To activate this function, system parameter “ES .ON” is set to “YES”. Now it is no longer possib le to change either measurement mode or measuring range in the “SEL ” menu; only the actual setup will be displayed.
white
brown
green
Fig. 11-3: Pin - Assignments LEMOSA Socket (Digital Inputs)
11 - 2
contact closed contact open
absolut differential measurement measurement
analog output analog output “standard” “scope”
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 79
Rosemount Analytical
123456
123456
12
12
12
12
12
12
123456
123456
12. Serial Interface (Option)

12.1 Retrofitting of Serial Interface / Status Signals

(status signals only: PCB BSI 10, Catalog - No.: 43 001 590, RS 232 - Interface: PCB BSI 10 with PCB SIF 232, Catalog - No.: CH 000 069 RS 485 - Interface: PCB BSI 10 with PCB SIF 485, Catalog - No.: CH 000 070,
see Section 12.3.2, too)
Be sure to observe the safety measures !

SERIAL INTERFACE (OPTION)

RETROFITTING
Opening the housing (see Section 21)
Connect circuit board to the threated bolts at the rear panel and mounting with the
washers and the screws .
Connect cable subject to code pin to BKS - pin connector J9.
Rear panel
Threated bolt
Rear panel
PCB BSI 10
J 9
J 9
Code pin
1
Fig. 12-1: Installation of PCB BSI 10
For retrofitting serial interface insert enclosed EPROM (see Section 25).
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
12 - 1
Page 80
SERIAL INTERFACE (OPTION)

GENERAL

Rosemount Analytical
12.2 General
The analyzer is equipped with a serial interface enabling communication with a host computer . The host computer can call up, prescribe, or alter parameters, as well as initiate analyz er operations, using standardized protocols. The optional BSI 10 plug in circuit board constitutes the hardw are interface. This ma y be configured as RS 232 C or RS 485 interface. The RS 485 interface permits networking several analyzers. Each analyzer may then be addressed using an assigned numerical ID - code.
Communications are always initiated b y the host computer, i.e., analyz er behaves passively until the host computer requests information from them or demands commencement of an action.
Communications use so - called “telegrams” being exchanged between the host computer and the analyzer(s). Syntax for these telegrams is established in protocols.
T elegrams alwa ys commence with the "$" start character, immediately followed b y a three - digit instruction code.
Subsequent elements of telegrams are segregated by the ";" hyphen c haracter.
The final element of all telegrams transmitted must be the “CR” termination character.
Upon receipt of the termination character, the analyzer attempts to e valuate the current contents of its input buffer as a valid telegram. If the syntax of the transmitted telegram is correct, the analyzer will transmit a response telegram to the host computer. This consists of the start character, an instruction code, requested data, a block - parity byte, and the termination character .
If the syntax of the transmitted telegram is not correct, the analyzer will transmit a status telegram containing an error message to the host computer. Each ter mination character reception thus initiates an analyzer response.
12 - 2
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 81
Rosemount Analytical
SERIAL INTERFACE (OPTION)
T o avoid detecting transmission errors, the host computer can insert a message - length parity byte immediately preceding the termination character for verification by the analyz er .
The analyzer invariably transmits message - length parity bytes immediately preceding termina­tion characters.
The time elapsed between the reception of star t characters and termination characters is not limited by the analyzer , i.e ., there are no “time - out” periods.
If the host computer transmits new characters before the analyzer has responded to the preceding telegram, the analyzer’ s input b uffer will reject them, i.e ., these characters will be ignored b y the
GENERAL
analyzer .
The transmission rate may be set between 600 and 4800 baud. An echo - mode may also be activated.
The analyzer software is configured so that telegrams may be sent to the host computer at time intervals of 150 ms and greater .
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
12 - 3
Page 82
SERIAL INTERFACE (OPTION)
Rosemount Analytical

START UP

12.3 Start Up
The analyzer has been set in our factory to RS 232 C or RS 485 interface via the plugged PCB SIF 232 or SIF 485 on the PCB BSI 10. The parameter 232c can also been set to “NO” or “YES” in the SIP (Serial Interface P arameters) line.
Interconnection to the interface is via the 9 - pin socket „Interface“ on the analyzer rear panel (Fig. 12-2).
Socket
“Interface”
M1 M1 M2 M2
R1 R1 R2 R2
IN
X1 OUTPUT
INTERFACE
X2 OUTPUT
OUT OUTIN
K1 K2
24 VDC
MADE IN GERMANY
ABS./DIFF.
X3 OUTPUT
12 - 4
Fig. 12-2: Socket “Interface” (Serial Interface)
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 83
Rosemount Analytical
12.3.1 RS 232 C
This interface requires a shielded cable having at least three internal conductors.
SERIAL INTERFACE (OPTION)
START UP
5
1
6
9
1 GND 2 RxD 3 TxD 4 not used 5 GND 6 not used 7 not used 8 not used 9 not used
Fig. 12-3: Pin - Assignments “RS 232 Interface”
12.3.2 RS 485
Configure 2- or 4- wire operation via solder bridge LB 1 of PCB SIF 485 before mounting the PCB. Connecting of [1 - 2] 2-wire operation is selected. Connecting of [2 - 3] 4-wire operation is active. Connect jumper P2 at both ends of interface connection (termination). For network operation with several analyzers via RS485 interface, termination has to be done at both ends of network connection only. For the other analyzers remo ve the jumper.
5
1
69
1 GND 2 RxD­3 RxD+ 4 TxD+ 5 TxD­6 not used 7 not used 8 not used 9 not used
Fig. 12-4: Pin - Assignments “RS 485 Interface”
In contrast to RS 232 C operation, simultaneous transmission and reception is not implemented in this standard. This would not result in damage to the electronics, but could lead to destruction of data. The analyzer beha ves passively in this mode of operation, i.e., it k eeps its transceiver set for reception whenever it is not transmitting. Since the time periods for transmission and reception are controlled by protocols, “data collisions” are excluded.
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
12 - 5
Page 84
SERIAL INTERFACE (OPTION)
START UP (INTERFACE - PARAMETER)
Rosemount Analytical
12.3.3 Switching ON/OFF Interface Operation
The analyzer may be set to either “on - line” or “off - line” status. This setting may be perf ormed either from the keypad or via telegram input.
Ke yboard setting:
SIP - parameter On.-L. = YES for on - line status SIP - parameter On.-L. = NO for off - line status
T elegram setting:
Instruction code 006: sets analyzer on - line status Instruction code 007: sets analyzer off - line status
If the analyzer is set to off - line status, it will accept only instruction code 006. All other instructions will be ignored and result in transmission of appropriate status telegrams.
12.3.4 Setting Interface Parameters
Agreement of interface parameters between analyzer and host computer is a fundamental requirement for communication without errors. The following analyzer parameters are concerned:
Baud rate: 600 / 1200 / 2400 / 4800 bits/s Data bits: 8 Stop bits: 2 Parity bit: none Echo mode: on / off (received characters will be retransmitted immediately) LPB-test: on / off (message - length parity check) ID No.: 0 to 99 (device ID No . in RS 485 mode)
12 - 6
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 85
Rosemount Analytical
All entries are made by using the keys
and
Depress the key
until appears,
SERIAL INTERFACE (OPTION)
START UP (INTERFACE - PARAMETER)
then depress the key
The unit is now ready for code entry, if such has not already been performed.
NO = off - line status YES = on - line status
Each device is assigned a device number for operation through the RS 485 interface (0 - 99).
Select interface type: NO = RS 485 YES = RS 232 C
Set baud rate: 0 = 4800 1 = 2400
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
2 = 1200 3 = 600
Echo-mode operation: NO = OFF YES = ON
Message - block parity check NO = OFF YES = ON
12 - 7
Page 86
SERIAL INTERFACE (OPTION)

TELEGRAM SYNTAX

Rosemount Analytical
12.4 Telegram Syntax
Telegrams are assembled as follows:
12.4.1 Start Character ( “$” = Hex 24)
If the start character is missing will be transmitted of an appropriate status telegram by the analyzer.
12.4.2 Termination Character ( “CR” = Hex OD)
If the termination character is missing, no decoding of the transmitted information will be performed, and the analyzer will not respond. No response message will be transmitted.
12.4.3 Instruction Code
Each instruction is assigned a unique three digit numerical instruction code. If a received instruction code should be other than three digits in length or contain non - numerical ASCII­characters, the analyzer will transmit an appropriate status telegram. Reception of unassigned instruction codes will also result in the transmission of a status telegram.
In the RS 232 C mode of operation, the instruction code immediately follows the start character; in the RS 485 mode of operation, the start character is followed by a two - digit de vice identification code, the separator character “;” and a three - digit instruction code, in this order .
12.4.4 Hyphen Character ( “;” = Hex 3B)
Individual elements of a telegram line are separated by this hyphen character. Missing hyphen characters can lead to misinterpretations of telegrams, and will result in transmission of an appropriate status telegram.
12 - 8
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 87
Rosemount Analytical
SERIAL INTERFACE (OPTION)
STATUS TELEGRAM
12.4.5 Status Telegram
If telegram syntax is faulty, or analyzer is unable to act upon an instruction received, then the analyzer will transmit a status telegram to the host computer .
These status telegrams are listed here for reference:
$ID;000;S100;LPB<CR> unrecognized instruction code $ID;000;S101;LPB<CR> LP - byte in error $ID;000;S102;LPB<CR> start character missing $ID;000;S103;LPB<CR> input buffer overflow $ID;xxx;S104;LPB<CR> analyzer off - line status $ID;xxx;S105;LPB<CR> text line too long $ID;xxx;S106;LPB<CR> undefined instruction $ID;xxx;S107;LPB<CR> invalid integer value $ID;xxx;S108;LPB<CR> numerical value outside defined range $ID;xxx;S109;LPB<CR> invalid failure/status code $ID;xxx;S110;LPB<CR> instruction can not be done here $ID;xxx;S111;LPB<CR> failure in transmitted character $ID;xxx;S112;LPB<CR> zeroing running $ID;xxx;S113;LPB<CR> spanning running $ID;xxx;S114;LPB<CR> invalid real number $ID;xxx;S115;LPB<CR> automatic calibration mode off $ID;xxx;S116;LPB<CR> parameter outside defined range $ID;xxx;S117;LPB<CR> preflushing period is running
xxx: instruction code I D: device ID No . in RS 485 mode LPB: message - length parity byte <CR>: termination character
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
12 - 9
Page 88
SERIAL INTERFACE (OPTION)
NUMERICAL REPRESENTATION / BLOCK PARITY CHECK
Rosemount Analytical
12.4.6 Numerical Representations
T elegrams may contain integers or real numbers. The formats for these numbers are subject to the following restrictions:
Integers: - Max. value = 216 - 1
- Positiv e numbers only accepted
- No decimal points allowed
Real: - Maximum of 6 digits accepted
- No alphabetic characters (e.g. 2.2E-6) allowed
- Analyzer output = 6 - digit real numbers
12.4.7 Block Parity Check
The master control computer may insert a message - length parity byte into telegrams. This invariably consists of two characters.
The message - length parity byte is the cumulatively EXCLUSIVE - OR correlation of all previously transmitted characters of the telegram line. Representation is in hexadecimal format. For example, if the decimal value should be decimal 13, this will be represented by the two characters “OD”, i.e., 030H and 044H.
The verification procedure may be enabled or disabled at the analyzer (see Section 12.3.4).
12 - 10
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 89
Rosemount Analytical
INSTRUCTION SYNTAX / INSTRUCTION LISTING

12.5 Instruction Syntax

Code definitions:
RP: Receive parameters analyzer is accepting values SP: Send parameters analyzer is sending values RI: Receive instructions k: Channel numbers 0 to 2 (0 = both channels) m: Measurement mode (0 = absolute / 1 = differential) w: Value (0 = N o o r scope / 1 = Yes or standard)
SERIAL INTERFACE (OPTION)
ff: Function line number pp : Parameter number <ID>: Analyzer ID - no . f or RS 485 mode of operation; f ollows start character LPB: Message - length parity byte <CR>: Termination character
Receipt of any instruction codes not listed in the following section will be acknowledged by transmission of status code 106. Future expansions will make use of code numbers not currently in use.
12.5.1 Instruction Listing
Instruction syntax: Instruction description:
$ID;001;LPB<CR> SP “4P.On” parameter
$ID;002;w;LPB<CR> RP “4P.On” parameter
$ID;003;k;LPB<CR> SP Ground level for differential measurement
$ID;004;w;k;LPB<CR> RP Ground level for differential measurement
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
answer “0”: Ground level of “GND” parameters is used (Section 10.2.1). answer “1”: Automatic ground level is used (Section 10.1.1).
w “0”: Use ground level of “GND” parameters (Section 10.2.1). w “1”: Use automatic ground level is used (Section 10.1.1).
answer: Differential measurement mode: entered ground level answer: Absolute measurement mode: “GND” parameters
Not for differential measurement mode, if “4P.On” = YES.
12 - 11
Page 90
SERIAL INTERFACE (OPTION)
INSTRUCTION LIST
Instruction syntax: Instruction description:
$ID;006;LPB<CR> RI on - line status $ID;007;LPB<CR> RI off - line status $ID;008;LPB<CR> SP datalogger data $ID;009;k;LPB<CR> SP measurement mode $ID;010;m;k;LPB<CR> RP measurement mode [k = 0 only]
Rosemount Analytical
$ID;011;m;k;LPB<CR> SP at full scale range $ID;013;k;LPB<CR> SP t $ID;014;w;k;LPB<CR> RP t
(response time) [k= 0 to 4]
90
(response time) [k= 1 to 4]
90
k = 0 all values k = 1 Ch. 1 abs. k = 2 Ch. 1 diff. k = 3 Ch. 2 abs. k = 4 Ch. 2 diff.
$ID;017;k;LPB<CR> SP preflushing period (zero gas) $ID;018;w;k;LPB<CR> RP preflushing period (zero gas) [k = 0 only] $ID;019;k;LPB<CR> SP preflushing period (span gas) $ID;020;w;k;LPB<CR> RP preflushing period (span gas) [k = 0 only] $ID;023;k;LPB<CR> SP concentration $ID;028;m;k;LPB<CR> SP span gas concentration $ID;029;m;w;k;LPB<CR> RP span gas concentration $ID;601;ff;pp;LPB<CR> SP analyzer parameter $ID;602;ff;pp;LPB<CR> RP analyzer parameter $ID;603;k;LPB<CR> SP gas component
12 - 12
$ID;604;k;LPB<CR> RI automatic zeroing $ID;605;k;LPB<CR> RI automatic spanning
$ID;627;LPB<CR> SP failure message (possible error batt. is
clearing by read out)
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 91
Rosemount Analytical
SERIAL INTERFACE (OPTION)
12.6 Datalogger
The analyzer has been equipped with a datalogger which instantly prints out all major data stored during the measurement: measuring values, measurement mode, error status, serial number, and time of up to 180 storage locations. An automatic function enables the automatic recording of measuring values at defined time intervals. Data output is accomplished e.g. by a printer connected via serial interface.
12.6.1 Manual Recording of Data in the Datalogger

DATALOGGER

If the measured value is displayed, then the actual data could be stored by just pressing
The measured value memory automatically switches over to the next storage location assigned. The initiated storage process is acknowledged by the following message, displayed for a short period of time:
(“XX” identifies the location for the next storage)
If all 180 storage locations are occupied, the error message “E.37” (OA error = Overflow Area error) appears on the monitor . In case that the new values must be stored, then press and the first lines will be overwritten.
Normally the message “E.37” disappears 1 or 2 minutes later.
The datalogger may be programmed via analyzer electronics, but this is possible only when “YES” has been selected for parameter “P.SEt” (see Section 8.1).
In this case the assigned storage location number is constantly display ed on the monitor .
In addition the storage location can now be freely assigned by pressing the keys
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
12 - 13
Page 92
SERIAL INTERFACE (OPTION)
DATALOGGER
Rosemount Analytical
12.6.2 Automatic Recording of Data in the Datalogger
In this menu the time intervals can be selected, when the automatic data storage have to be accomplished.
The range of accepted entries is 0 to 1440 minutes (= 24 hours) By selecting “0”, the automatic data storage function is disabled.
Use the keys and f or the entry.
Note:
If an OA error occurs during the automatic storage mode, no error message (“E.37”) will be displayed and the previously stored values will be automatically overwritten, first line at first.
12.6.3 Data Output
Data output is accomplished e.g. by a printer connected via serial interface. Make sure that the parameters (setting in “SIP. ”, see Section 12.3.4) of the serial interf aces are identical.
Baud rate: 600 / 1200 / 2400 / 4800 bits/s Data bits: 8 Stop bits: 2 Parity bit: none Echo mode: on / off (received characters will be retransmitted immediately)
To print out the stored data use the keys to select the variable “YES”.
and then depress
All data stored for the current operation/measurement will be printed out in A4 format and 65 lines per page. A line feed is acknowledged by printing of <CR> and <LF> signs.
12 - 14
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 93
Rosemount Analytical
SERIAL INTERFACE (OPTION)
12.6.4 Deleting of Data from Datalogger
The deleting of stored data is secured by code. To delete data proceed as follows:
Depress the key
until the display appears.
Depress the key
DATALOGGER
The display will appear
Use the keys to select the Code
and then depress
The display will appear.
Use the keys to select the variable “YES” and then depress
Now the stored data is deleted. New record begins with line 1.
Note:
It is not necessary to delete the stored data because they will be overwritten when an OA error occurs. Normally the error message “E.37” disappears 1 or 2 minutes later .
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
12 - 15
Page 94
SERIAL INTERFACE (OPTION)
DATALOGGER
Rosemount Analytical
12.6.5 Display and Adjusting of Time
In the time display , hours and minutes are separated b y a decimal point, which is pulsating at any second.
The analyzer is not equipped with a real-time unit. Theref ore the time is displayed only when the power supply is on. After turning off the power supply , the time must be adjusted ane w. The stored data remains stored in the battery-buffered RAM-memory .
To adjust the time display depress the key
until the display appears.
Depress the key
If the Code has not already been entered, the following display
will appear
Use the keys to select the Code and then depress
The display will appear.
Use the keys to set the hour and then depress
The display will appear.
Use the keys to set the minutes and then depress
Note:
By setting the minutes, the unit-internal second counter will be reset to “0”.
12 - 16
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 95
Rosemount Analytical
Error Code Possible Reasons Check / Correct
13. Error List
Some of the failures which ma y arise during measurement will be reported on the display in form of error codes.
When such a failure arises, the display's will show the concentration value
alternating with (E = ERROR).

ERROR LIST

Be sure to observe the safety measures for all workings at the anal yzer!
Error Code Possible Reasons Check / Correct
1. Displays are “switched OFF”.
No display.
2. Voltage supply absent.
1. Press any key.
Check parameter "dOFF" (see 8.2.3).
2. Check electrical supply (see Fig. A-2, Item 3).
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
3. Connection front panel /BKS absent.
3. Check connection BKB - BKS (X1) (see Section 15).
13 - 1
Page 96
ERROR LIST
Error Code Possible Reasons Check / Correct
Flushing.
Battery buffer faulty.
Rosemount Analytical
Check, if Jumper J 7 is plugged (see Section 16).
Channel 1.
Channel 2.
A/D - conversion
of end signal
absent.
The EPROM - default values were charged.
1. Jumper not or incorrect plugged.
2. Positive or negative reference voltage absent.
3. Light barrier signal absent.
Exchange battery, if battery voltage < 3.5 V (BKS - jumper J7 plugged).
The error is clearing after depressing any key or with serial interface instruction $627.
1. Channel 1: Check jumper J1 Channel 2: Check jumper J2 (see Section 16)
Switch analyzer off and then on again.
2. Check reference voltage (see Sections 14.2 and 14.3).
3. Check connection X9 / light barrier (see Section 15).
Temperature compensation
inoperative.
4. Chopper drive inoperative.
5. Supply voltage (internal 6 V DC) absent.
1. Start-up of A/D - conversion in temperature channel absent.
2. Supply voltage (internal 6 V DC) absent.
Check measuring point 14.6.
4. Check connection X2/chopper drive (see Section 15).
Check measuring point 14.4.
5. Check measuring point 14.1.
1. Switch analyzer off and then on again.
2. Check measuring point 14.1.
13 - 2
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 97
Rosemount Analytical
Error Code Possible Reasons Check / Correct
ERROR LIST
Channel 1
Channel 2.
Tolerance error.
Zero - gas value deviates
more than 10% of the
measuring range from zero.
Channel 1.
Channel 2.
1. Incorrect zero gas in use.
2. Photometer section contaminated.
3. Analyzer not calibrated.
1. Incorrect nominal value.
2. Incorrect span gas in use.
1. Check zero gas in use.
2. Check analysis cell and windows for contamination.
Cleaning of contaminated parts (see Section 22.3).
3. Switch off the tolerance check before starting an adjustment (see Section 8.2.2).
1. Enter the correct nominal value (certification of span gas bottle) (see Section 9.3.2).
2. Check span gas in use.
Use another or a new gas bottle.
Tolerance error.
Span - gas value deviates
more than 10% from
nominal value.
Channel 1.
Channel 2.
Measuring value more than
10% over full-scale range.
3. Photometer section contaminated.
4. Analyzer not calibrated.
1. Concentration of measuring gas too high.
Enter the correct nominal value
3. Check analysis cell and windows for contamination.
Cleaning of contaminated parts (see Section 22.3).
4. Switch off the tolerance check before starting an adjustment (see Section 8.2.2).
1. Check concentration of measuring gas.
Use another analyzer suitable for the concentration range involved.
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
13 - 3
Page 98
ERROR LIST
Error Code Possible Reasons Check / Correct
Rosemount Analytical
Datalogger
“Overflow Area”
EPROM check summary
defective.
Test for RAM - IC's
defective
Analog output absent.
Time - out for XON of serial interface.
All 180 storage locations are occupied.
1. EPROM faulty.
2. BKS faulty.
RAM - IC's / BKS faulty.
BKS faulty.
At drive of serial interface XON character is absent (Time out > 60 s).
The messages “E.37” disappears 1 or 2 minutes later.
1. Exchange EPROM (see Section 25.)
2. Exchange BKS.
Exchange BKS.
Exchange BKS.
Fluctuating or
erroneous display.
1. Leakage into gas circuit.
2. Ambient air contains gas constituent to be measured in excessive concentration.
3. Gas pressure subject to excessive fluctuations.
1. Perform a leakage check (see Section 20).
2. Replace absorber material for the light sources and chopper housing.
Use sealed photometer (option).
Flush out the analyzer.
3. Check the gas lines preceding and following the photometer.
Eliminate any restrictions found beyond the gas outlet fitting.
Reduce pumping rate or flow rate.
13 - 4
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Page 99
Rosemount Analytical
Error Code Possible Reasons Check / Correct
ERROR LIST
Fluctuating or
erroneous display.
4.Detector not connected.
5.Light source not connected or faulty.
6.Faulty analog preamplifying.
7.Contamination of the gas paths.
4.Check connections: BKS X5 / detector channel 1 BKS X6 / detector channel 2 (see Section 15).
5.Check connection: BKS X3(1/2)/light source channel1 BKS X3(4/5)/light source channel2 (see Section 15)
Light source is cold: For dual - IR - channel analyzer interchange the two light sources. Replace the suspect light source (see Section 23.2).
6.Check measuring point 14.7.
7.Check analysis cell and windows for contamination.
Cleaning of contaminated parts (see Section 22.3).
8.Barometric pressure effects.
9.Temperature below the dew point in the gas paths.
10.Faulty A/D converter.
Check gas paths and gas conditioning to contamination.
8.Enter the correct value for barometric pressure (see 8.1).
9.Check the temperature of the gas
paths and eliminate any reason of
condensation.
Maintain all temperatures at values at least 10 °C above the dew point of sample gas.
10.Exchange BKS.
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
13 - 5
Page 100
ERROR LIST
Error Code Possible Reasons Check / Correct
Response time too long
(t
time)
90
1. Incorrect response time ( t90 time).
2. Pumping rate inadequate.
3. Contamination of the gas paths.
Rosemount Analytical
1. Check the value for t90 time (see Section 8.2.6).
2. The feeder line between the sampling point and the analyzer is too long. Use a larger, external pump; consider adding a bypass line to the process stream for sampling purposes (see Section 5.1).
3. Check gas paths and gas conditioning to contamination.
Clean gas paths and exchange the filter elements.
13 - 6
90002927(1) BINOS® 100 4P e [1.00] 25.06.97
Loading...