Rosemount Analytical In Situ Oxygen Probes-Instruction Bulletin Addendum-Rev. 3.0 Manuals & Guides

Optional Isolation Valving System Used With Rosemount Analytical In Situ Oxygen Probes
Instruction Bulletin Addendum IB – ISO1000 Rev. 3- June, 2005
Customer Support Center-in US 800-433-6076 International 440-914-1261
Oxymitter 4000
Part no. ____________ Serial no. ____________ Order no. ____________
Visit our website at:
On-Line Ordering Available!
www.processanalytic.com
Applicability: Isolation Valving Option, Oxy-Iso 2000
Utilized with Rosemount Analytical’s Oxymitter 4000/5000 series of O
analyzers or with Rosemount Analytical’s CENELEC World
2
Class O
Analyzers. May be used with pressure-balancing option,
2
p/n 3D39811Gxx

General

Rosemount Analytical’s line of in situ oxygen analyzers utilize zirconium oxide (ZrO
) sensing technology which is sensitive to pressure variations in the
2
process. Rosemount Analytical’s Oxymitter and World Class probes place the sensing cell at the end of an in situ probe that inserts directly into the process gas stream. This arrangement provides fast response and is very resistant to pluggage from particulate material or the acids that frequently condense within normal flue gas. The ZrO and an output change of approximately 1% of reading range, or 1% O
sensing technology is sensitive to pressure changes in the process
2
(not 1 % of full scale
) for every 4 inches of H2O pressure or vacuum in the process
2
can be expected. See the instruction bulleton for the separate “pressure balancing system” for a more complete description of the pressure effects, and our accommodation to nullify these effects. Process pressures of more than 1 PSI also present the challenge of removing or inserting the probe for maintenance purposes while the process is on line. Hot, noxious combustion gases can make the removal and installation procedure hazardous, especially when the operation is conducted at elevation and on less than adequate catwalks. This isolation valving system facilitates the insertion and withdrawal of the probe for service while the process is under pressure. This valving system is recommended if the process cannot be shut down for probe maintenance.

Unpacking

The Isolation Valving system will be shipped separately from the O
probe and other
2
options, such as pressure balancing and automatic calibration gas sequencer. The isolation valving system is shipped as a compact assembly, but after unpacking, it’s best to disassemble the guide bars and packing box so that many lighter components can be installed as opposed to one heavy assembly.

Mechanical/Pneumatic Installation- Isolation Valving System

The process must be off-line before conducting the installation and the temperatures at the installation point must have operating temperatures of less than 1,300
0
F (7050 C). Process ductwork must be prepared by mounting a 3-inch, 300-pound nozzle with a length no longer than 12 inches. A 3-inch, 300-pound ANSI flange is required at the end of this nozzle. Ensure that the installation will permit the probe to insert at least 6 inches into the process ductwork. (See insertion/removal envelope drawing at the end of this manual). Place the provided gasket onto the flange and bolt the isolation gate valve to the nozzle
IB Addendum ISO1000
2
flange with the packing box assembly and O2 probe removed. O2 probe installation may be conducted with the process on-line once the Isolation Valve is installed and closed. Install the packing box housing weldment onto the isolation valve.
(Isolation valve would be to the right, but is removed in this photo sequence for clarity.)
IB Addendum ISO1000
3
Next install the removable guide rods, and then the sliding probe carrier assembly.
IB Addendum ISO1000
4
Probe carrier assembly.
Remove the flame arrestor from the end of the Oxymitter O2 probe, and place the probe into the cradle of the sliding carrier assembly with the probe flange behind the carrier assembly, and lock the retaining bar up. Note that the L­shaped piece at the 12 o’clock position will contain the probe from sliding back under pressure, but also permit the probe to be rotated back and forth to assist removal.
IB Addendum ISO1000
5
String the following components onto the probe body:
- adjustment plate for gasket compression ring
- 8-hole packing box retaining ring
- gasket compression ring
- packing box , without rope gaskets
IB Addendum ISO1000
6
Insert the three rings of packing material into the packing box, pushing each ring to the back of the box. Ensure that the point where the gasket ends meet is indexed from gasket to gasket. For instance, it would be good for the ends of gasket #1 to meet at a 12 o’clock orientation, gasket # 2 ends to meet at 4 o’clock, and gasket # 3 ends to meet at 8 o’clock.
IB Addendum ISO1000
7
Before inserting the probe into the packing box housing, ensure that the face gasket at the back of the housing is in place. Also shown is the split ring with knurled teeth. This is depicted for clarity, but must be removed for probe insertion, and reassembled around the probe body.
IB Addendum ISO1000
8
Reinstall the flame arrestor/ diffusion element assembly onto the end of the probe. Make sure to use Rosemount’s white mineral antisieze compound on the threads. Do not use any petroleum based antisieze. Remember to orient the V-deflector shield (if using a ceramic diffusion element) into the direction of the process flow. Slide the carrier assembly forward so that the probe inserts through the packing box weldment, and to the closed gate of the gate valve.
IB Addendum ISO1000
9
A split knurled ring is provided to scrape off cemented catalyst material that may be deposited onto the probe body during subsequent removal. Insert the split ring into the back of the packing box housing around the probe body. Ensure that the small hole in the top half of the split ring fits over the pin in the back of the weldment.
IB Addendum ISO1000
10
Loading...
+ 21 hidden pages