Rosemount 951C NOx Analyzer-Rev N Manuals & Guides

Page 1
Rosemount Analytical
ODEL
M
951C
NO
NSTRUCTION MANUAL
X ANALYZER
PN 748214
Page 2
N
OTICE
The information contained in this document is subject to change without notice.
Teflon® is a registered trademark of E.I. duPont de Nemours and Co., Inc. Alconox is a registered trademark of Alconox, Inc.
SNOOP® is a registered trademark of NUPRO Co.
Manual Part Number 748214-N September 1997 Printed in U.S.A.
Rosemount Analytical Inc.
4125 East La Palma Avenue
NAHEIM
A
ALIFORNIA
, C
92807-1802
Page 3
C
ONTENTS
P
REFACE
PURPOSE/SAFETY SUMMARY........................................................................P-1
SPECIFICATIONS - LO RANGE ........................................................................P-4
SPECIFICATIONS - HI RANGE .........................................................................P-5
CUSTOMER SERVICE, TECHNICAL ASSISTANCE AND FIELD SERVICE ....P-6
RETURNING PARTS TO THE FACTORY .........................................................P-6
TRAINING ......................................................................................................P-6
DOCUMENTATION............................................................................................P-6
COMPLIANCES..................................................................................................P-7
CONDENSED STARTUP AND CALIBRATION PROCEDURE..........................P-9
S
ECTION
1.1 OVERVIEW ...............................................................................................1-1
1.2 APPLICATIONS.........................................................................................1-2
S
ECTION
2.1 UNPACKING.............................................................................................2-1
2.2 LOCATION ...............................................................................................2-1
2.3 VOLTAGE REQUIREMENTS...................................................................2-1
2.4 ELECTRICAL CONNECTIONS ................................................................2-2
2.5 GAS REQUIREMENTS ............................................................................2-5
2.6 SAMPLE REQUIREMENTS .....................................................................2-6
2.7 GAS CONNECTIONS...............................................................................2-6
2.8 LEAK TEST ..............................................................................................2-7
1. I
2. I
2.4.1 Line Power Connections.............................................................2-2
2.4.2 Potentiometric Recorder Connections.........................................2-4
2.4.3 Current Recorder Connections....................................................2-4
NTRODUCTION
NSTALLATION
748214-N
September 1997
Rosemount Analytical
i
Page 4
M
ODEL
951C NOX A
NALYZER
S
ECTION
3.1 FRONT PANEL INDICATORS AND CONTROLS.....................................3-1
3.2 STARTUP PROCEDURE.........................................................................3-2
3.3 CALIBRATION .........................................................................................3-4
3.4 ROUTINE OPERATION...........................................................................3-5
3.5 CONVERTER TEMPERATURE ADJUSTMENT PROCEDURE..............3-5
3.6 MEASUREMENT OF CONVERTER EFFICIENCY..................................3-7
3.7 RECOMMENDED CALIBRATION FREQUENCY....................................3-8
S
ECTION
3. I
3.1.1 Display ......................................................................................3-1
3.1.2 Range Selection.........................................................................3-1
3.1.3 Sample Pressure Gauge............................................................3-2
3.1.4 Ozone Pressure..........................................................................3-2
3.1.5 Zero and Span Potentiometers...................................................3-2
3.1.6 Ozone Interlock ..........................................................................3-2
3.3.1 Zero Calibration..........................................................................3-4
3.3.2 Upscale Calibration ....................................................................3-4
4. T
NITIAL STARTUP AND OPERATION
HEORY
4.1 NITRIC OXIDE DETERMINATION BY CHEMILUMINESCENCE
METHOD ...................................................................................... 4-1
4.2 ANALYZER FLOW SYSTEM....................................................................4-1
4.2.1 Flow of Sample, Standard Gas or Zero Gas to
Reaction Chamber........................................................4-1
4.2.2 Ozone Generation......................................................................4-2
4.3 SIGNAL PROCESSING ELECTRONICS SYSTEM.................................4-2
4.4 ANALYZER THERMAL SYSTEM.............................................................4-3
S
ECTION
5.1 SYSTEM CHECKS AND ADJUSTMENTS................................................5-1
5.2 SERVICING FLOW SYSTEM ..................................................................5-3
5. R
5.1.1 Display Fullscale Span Adjustment............................................5-1
5.1.2 Overall Sensitivity.......................................................................5-2
5.1.3 Ozone Output.............................................................................5-2
5.1.4 Background Current ...................................................................5-3
5.1.5 Excessive Photomultiplier Dark Current.....................................5-3
5.2.1 Cleaning Sample Capillary .........................................................5-3
5.2.2 Ozone Restrictor Fitting..............................................................5-4
OUTINE SERVICING
5.3 PHOTOMULTIPLIER TUBE/REACTION CHAMBER ..............................5-5
ii
Rosemount Analytical
September 1997
748214-N
Page 5
C
ONTENTS
S
ECTION
5.4 OZONE GENERATION SYSTEM.............................................................5-8
5.5 CONVERTER ASSEMBLY.......................................................................5-9
5.6 SERVICING ELECTRONIC CIRCUITRY..................................................5-10
S
ECTION
6.1 CIRCUIT BOARD REPLACEMENT POLICY.............................................6-1
6.2 REPLACEMENT PARTS...........................................................................6-1
ENERAL PRECAUTIONS FOR HANDLING
G
ARRANTY
W
IELD SERVICE AND REPAIR FACILITIES
F
5. R
5.3.1 Photomultipler Tube/Reaction Chamber Removal......................5-5
5.3.2 Cleaning Reaction Chamber.......................................................5-5
5.3.3 Photomultiplier Tube and Housing..............................................5-7
5.3.4 Replacement of Photomultiplier Tube.........................................5-7
5.4.1 Lamp/Housing Removal..............................................................5-8
5.4.2 UV Lamp Replacement...............................................................5-9
5.4.3 Power Supply Removal...............................................................5-9
6. R
OUTINE SERVICING (CONTINUED
EPLACEMENT PARTS
TORING HIGH PRESSURE CYLINDERS
& S
)
F
IGURES
2-1 POWER SUPPLY BOARD VOLTAGE SELECT SWITCHES...................2-3
2-2 TEMPERATURE CONTROL BOARD .......................................................2-3
2-3 CABLE GLAND..........................................................................................2-4
2-4 REAR VIEW OF MODEL 951C (COVER REMOVED)...............................2-5
3-1 MODEL 951C CONTROLS, INDICATORS AND ADJUSTMENTS ...........3-1
3-2 SIGNAL BOARD.......................................................................................3-3
3-3 POWER SUPPLY BOARD.......................................................................3-4
4-1 ANALYZER SIGNAL CONDITIONING CIRCUIT......................................4-3
4-2 ANALYZER THERMAL SYSTEM.............................................................4-4
6-1 MAJOR ASSEMBLIES OF THE MODEL 951C........................................6-3
6-2 PHOTOMULTIPLIER HOUSING ASSEMBLY..........................................6-4
6-3 CONVERTER ASSEMBLY.......................................................................6-5
6-4 CASE HEATER TEMPERATURE CONTROL ASSEMBLY .....................6-6
748214-N
September 1997
Rosemount Analytical
iii
Page 6
M
T
951C NOX A
ODEL
ABLES
NALYZER
3-1 RESISTANCE OF CONVERTER TEMPERATURE SENSOR VS.
TEMPERATURE...........................................................................3-6
D
RAWINGS (LOCATED IN REAR OF MANUAL
654063 INSTALLATION DRAWING 654090 FLOW DIAGRAM, LOW RANGE 654093 FLOW DIAGRAM, HIGH RANGE
)
iv
Rosemount Analytical
September 1997
748214-N
Page 7
P
REFACE
P
URPOSE/SAFETY SUMMARY
To avoid explosi on, l oss of l i fe, per sonal i nj ury and damage to this equipment and on-site property, all personnel authorized to install, operate and service the Model 951C NOx Analyzer should be thoroughly familiar with and strictly follow the instructions in this manual.
If this equipment is used in a manner not specified in these instructions, protective systems may be impaired.
Save these instructions.
DANGER
personal injury, death, or substantial property damage if the warning is ignored.
WARNING
personal injury, death, or substantial property damage if the warning is ignored.
CAUTION minor
NOTE
important but not hazard-related.
is used to indicate the presence of a hazard which
is used to indicate the presence of a hazard which
is used to indicate the presence of a hazard which
personal injury or property damage if the warning is ignored.
is used to indicate installation, operation or maintenance information which is
will
cause
can
cause
will or can
severe
severe
cause
WARNING: ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel.
This instrument was shipped from factory set up to operate on 115 volt 50/60 Hz. For operation on 230 volt 50/60 Hz, refer to Section 2.3.
For safety and proper performance this instrument must be connected to a properly grounded three-wire source of power.
WARNING: INTERNAL ULTRAVIOLET LIGHT HAZARD
Ultraviolet light from the ozone generator can cause permanent eye damage. Do not look directly at the ultraviolet source in ozone generator. Use of ultraviolet filtering glasses is recommended.
748214-N
September 1997
Rosemount Analytical
P-1
Page 8
M
ODEL
951C NOX A
NALYZER
WARNING: TOXIC CHEMICAL HAZARD
This instrument generates ozone which is toxic by inhalation and is a strong irritant to throat and lungs. Ozone is also a strong oxidizing agent. Its presence is detected by a characteristic pungent odor.
The instrument exhaust contains both ozone and nitrogen dioxide, both toxic by inhalation, and may contain other constituents of the sample gas which may be toxic. Such gases include various oxides of nitrogen, unburned hydrocarbons, carbon monoxide and other products of combustion reactions. Carbon monoxide is highly toxic and can cause headache, nausea, loss of consciousness, and death.
Avoid inhalation of the ozone produced within the analyzer and avoid inhalation of the sample and exhaust products transported within the analyzer. Avoid inhalation of the combined exhaust products at the exhaust fitting.
Keep all tube fittings tight to avoid leaks. See Section 2.6 for Leak Test Procedure.
Connect rear exhaust outlet to outside vent by a 1/4 inch (6.3 mm) or larger stainless steel or Teflon* line. Check vent line and connections for leakage.
WARNING: PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect safety of this product. Use only factory documented components for repair.
WARNING: HIGH PRESSURE GAS CYLINDERS
This instrument requires periodic calibration with a known standard gas. See Paragraphs 2.5 and 3.3. See also General Precautions for Handling and Storing High Pressure Gas Cylinders, following Section Six.
P-2
Rosemount Analytical
September 1997
748214-N
Page 9
P
WARNING: TOXIC AND OXIDIZING GAS HAZARDS
The ozone generator lamp contains mercury. Lamp breakage could result in mercury exposure. Mercury is highly toxic if absorbed through skin or ingested, or if vapors are inhaled.
HANDLE LAMP ASSEMBLY WITH EXTREME CARE.
If lamp is broken, avoid skin contact and inhalation in the area of the lamp or the mercury spill.
Immediately clean up and dispose of the mercury spill and lamp residue as follows:
Wearing rubber gloves and goggles, collect all droplets of
mercury by means of a suction pump and aspirator bottle with long capillary tube. Alternatively, a commercially available mercury spill clean-up kit, such as J. T. Baker product No. 4439-01, is recommended.
REFACE
Carefully sweep any remaining mercury and lamp debris into
a dust pan. Carefully transfer all mercury, lamp residue and debris into a plastic bottle which can be tightly capped. Label and return to hazardous material reclamation center.
Do not place in trash, incinerate or flush down sewer.
Cover any fine droplets of mercury in non-accessible
crevices with calcium polysulfide and sulfur dust.
CAUTION: TOPPLING HAZARD
This instrument’s internal pullout chassis is equipped with a safety stop latch located on the left side of the chassis.
When extracting the chassis, verify that the safety latch is in its proper (counter-clockwise) orientation.
If access to the rear of the chassis is required, the safety stop may be overridden by lifting the latch; however, further extraction must be done very carefully to insure the chassis does not fall out of its enclosure.
If the instrument is located on top of a table or bench near the edge, and the chassis is extracted, it must be supported to prevent toppling.
Failure to observe these precautions could result in personal injury and/or damage to the product.
748214-N
September 1997
Rosemount Analytical
P-3
Page 10
M
ODEL
951C NOX A
NALYZER
S
PECIFICATIONS
R
ANGES
R
EPEATABILITY
Z
ERO/SPAN DRIFT
R
ESPONSE TIME
(ELECTRONIC
S
ENSITIVITY
D
ETECTOR OPERATING
P
RESSURE
T
OTAL SAMPLE FLOW RATE
S
AMPLE PRESSURE
O
ZONE GENERATOR GAS
A
MBIENT TEMPERATURE
R
ANGE
:
:
:
:
:
:
:
+ F
- LO R
)
LOW
:
:
ANGE
0 to 10, 0 to 25, 0 to 100, 0 to 250 ppm NOx within 0.1 ppm or ±1% of fullscale, whichever is greater less than ±0.1 ppm or ±1% of fullscale, whichever is
greater, in 24 hours at constant temperature less than ±0.2 ppm or ±2% of fullscale, whichever is
greater, over any 10°C interval from 4 to 40°C (for rate change of 10°C or less per hour)
90% of fullscale in less than 1 minute
less than 0.1 ppm or 1% of fullscale, whichever is greater atmospheric
1 Liter per minute at 20 psig
:
138 kPa (20 psig) U.S.P. breathing-grade air 4 to 40°C (40 to 104°F)
A
NALOG OUTPUT
P
OWER REQUIREMENTS
E
NCLOSURE
D
IMENSIONS
W
EIGHT
:
:
:
Potentiometric: Isolated Current:
0 to +5 VDC, 2000 ohm minimum load
Field-selectable 0 to 20 or 4 to 20 mA,
700 ohm maximum load
Display
: Digital, 4-1/2 digit LCD, readout in engineering
units, backlighted
:
115/230 VAC ±10%, 50/60 ±3 Hz, 570 W maximum General purpose for installation in weather-protected
areas
22.0 cm (8.7 in) H
48.3 cm (19 in.) W
48.3 cm (19 in.) D
22.2 kg (49 lbs) approximate
P-4
Rosemount Analytical
September 1997
748214-N
Page 11
P
REFACE
S
PECIFICATIONS
R
ANGES
R
EPEATABILITY
Z
ERO/SPAN DRIFT
R
ESPONSE TIME
(ELECTRONIC
S
ENSITIVITY
D
ETECTOR OPERATING
P
RESSURE
T
OTAL SAMPLE FLOW RATE
S
AMPLE PRESSURE
O
ZONE GENERATOR GAS
A
MBIENT TEMPERATURE
R
ANGE
:
:
:
:
:
:
:
+ F
- HI R
)
LOW
:
:
ANGE
0 to 100, 0 to 250, 0 to 1000, 0 to 2500 ppm NOx within 0.1 ppm or ±1% of fullscale, whichever is greater less than ±1.0 ppm or ±1% of fullscale, whichever is
greater, in 24 hours at constant temperature less than ±2.0 ppm or ±2% of fullscale, whichever is
greater, over any 10°C interval from 4 to 40°C (for rate change of 10°C or less per hour)
90% of fullscale in less than 1 minute
less than 0.1 ppm or 1% of fullscale, whichever is greater atmospheric
1 Liter per minute at 20 psig
:
138 kPa (20 psig) U.S.P. breathing-grade air 4 to 40°C (40 to 104°F)
A
NALOG OUTPUT
P
OWER REQUIREMENTS
E
NCLOSURE
D
IMENSIONS
W
EIGHT
:
:
:
:
Potentiometric: Isolated Current:
0 to +5 VDC, 2000 ohm minimum load
Field-selectable 0 to 20 or 4 to 20 mA,
700 ohm maximum load
Display:
Digital, 4-1/2 digit LCD, readout in engineering
units, backlighted
:
115/230 VAC ±10%, 50/60 ±3 Hz, 570 W maximum General purpose for installation in weather-protected
areas
22.0 cm (8.7 in) H
48.3 cm (19 in.) W
48.3 cm (19 in.) D
22.2 kg (49 lbs) approximate
748214-N
September 1997
Rosemount Analytical
P-5
Page 12
M
ODEL
951C NOX A
NALYZER
C
USTOMER SERVICE
For order administration, replacement Parts, application assistance, on-site or factory repair, service or maintenance contract information, contact:
R
ETURNING PARTS TO THE FACTORY
Before returning parts, contact the Customer Service Center and request a Returned Materials Authorization (RMA) number. Please have the following information when you call:
Order Number.
Prior authorization by the factory must be obtained before returned materials will be accepted. Unauthorized returns will be returned to the sender, freight collect.
When returning any product or component that has been exposed to a toxic, corrosive or other hazardous material or used in such a hazardous environment, the user must attach an appropriate Material Safety Data Sheet (M.S.D.S.) or a written certification that the material has been decontaminated, disinfected and/or detoxified.
Model Number, Serial Number, and Purchase Order Number or Sales
, T
ECHNICAL ASSIST ANCE AND FIELD SERVICE
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
Return to:
Rosemount Analytical Inc.
4125 East La Palma Avenue
Anaheim, California 92807-1802
T
RAINING
A comprehensive Factory Training Program of operator and service classes is available. For a copy of the the Technical Services Department at:
D
OCUMENTATION
The following Model 951C NOx Analyzer instruction materials are available. Contact Customer Service or the local representative to order.
748214 Instruction Manual (this document)
Current Operator and Service Training Schedule
Rosemount Analytical Inc.
Phone: 1-714-986-7600
FAX: 1-714-577-8006
contact
P-6
Rosemount Analytical
September 1997
748214-N
Page 13
C
9
6
OMPLIANCES
This product satisfies all obligations of all relevant standards of the EMC framework in Australia and New Zealand.
N
P
REFACE
748214-N
September 1997
Rosemount Analytical
P-7
Page 14
M
ODEL
N
OTES
951C NOX A
NALYZER
P-8
Rosemount Analytical
September 1997
748214-N
Page 15
C
ONDENSED STARTUP AND CALIBRATION
P
ROCEDURE
The following summarized instructions on startup and calibration are intended for operators already familiar with the analyzer.
For initial startup, refer to detailed instructions provided in Section 3.
1. Set slider switch on the Signal Board (Figure 3-2) to 250 ppm (see Figure 3-2).
2. Apply power to the analyzer. The analyzer will now require approximately one to two hours for temperature equilibrium before being ready for calibration.
3. Verify that the pressure regulator on the cylinder of zero gas (nitrogen or air) or sample gas is set for supply pressure of 10 to 17 psig.
4. Verify that the pressure regulator on the cylinder of air (ozonator supply) is set for supply pressure of 20 to 25 psig.
5. Establish correct pressure of sample gas:
a. Supply sample gas to rear-panel SAMPLE inlet at 10 to 17 psig (normally 15
psig).
b. Adjust SAMPLE Back Pressure Regulator so that SAMPLE Pressure Gauge
indicates the value appropriate to the desired operating range (normal operating pressure is 3 to 5 psig). See Figure 3-1.
6. Establish correct pressure of zero gas:
a. Supply zero gas to rear panel SAMPLE inlet and set to 15 psig. b. Note reading on SAMPLE Pressure Gauge. It should be the same as in Step 5b.
If not, adjust output pressure regulator on the zero gas cylinder as required.
7. Establish correct pressure of upscale standard gas:
a. Supply upscale standard gas to rear panel SAMPLE inlet. b. Note reading on SAMPLE Pressure Gauge. It should be the same as in Step 6b.
If not, adjust output regulator on cylinder of upscale standard gas as required.
748214-N
September 1997
Rosemount Analytical
P-9
Page 16
M
ODEL
951C NOX A
NALYZER
Note Supply pressure for sample, upscale standard gas and zero air must be
the same. If not, the readout will be in error.
8. Zero Calibration. a. Set PPM RANGE Switch for range to be used for sample analysis. Set SPAN
Control at normal operating setting, if known, or at about mid-range if normal
setting is not known. b. Supply zero gas to rear panel SAMPLE inlet. c. Adjust ZERO Control for reading of zero on meter or recorder.
9. Upscale Calibration. a. Set PPM RANGE Switch at setting appropriate to the particular span gas. b. Supply upscale standard gas of accurately known NOx content to rear panel
SAMPLE inlet.
c. Adjust SPAN Control so that reading on meter or recorder is equal to the know
parts-per-million concentration of NOx in the span gas.
Note It is the responsibility of the user to measure efficiency of the NO
-to-NO
2
converter during initial startup, and thereafter at intervals appropriate to the application, normally once a month.
P-10
Rosemount Analytical
September 1997
748214-N
Page 17
I
NTRODUCTION
1
1.1 OVERVIEW
The Model 951C NOx Analyzer is designed to measure NOx using one of two sets of ranges designated as Hi or Lo. The Hi Range set consists of spans with ranges of 0-100, 0-250, 0-1000, and 0-2500 ppm NOx. The Lo Range set consists of spans with ranges of 0-10, 0-25, 0-100, and 0-250 ppm NOx.
The NOx analyzer continuously analyzes a flowing gas sample for NOx [nitric oxide (NO) plus nitrogen dioxide (NO2)]. The sum of the concentrations is continuously reported as NOx.
The analyzer is based on the chemiluminescence method of NO detection. The sample is continuously passed through a heated bed of vitreous carbon, in which NO2 is reduced to NO. Any NO initially present in the sample passes through the converter unchanged, and any NO2 is converted to an approximately equivalent (95%) amount of NO.
The NO is quantitatively converted to NO2 by gas-phase oxidation with molecular ozone produced within the analyzer from air supplied by an external cylinder. During this reaction, approximately 10% of the NO2 molecules are elevated to an electronically excited state, followed by immediate decay to the non-excited state, accompanied by emission of photons. These photons are detected by a photomultiplier tube, which in turn generates a DC current proportional to the concentration of NOx in the sample stream. The current is then amplified and used to drive a front panel display and to provide potentiometric and isolated current outputs.
To minimize system response time, an internal sample-bypass feature provides high-velocity sample flow through the analyzer.
The display blanks when the analyzer is 10% or more over-range. Selecting a less sensitive (higher) range restores the display function.
The case heater assembly of the Model 951C maintains the internal temperature at approximately 50oC (122oF).
748214-N
September 1997
Rosemount Analytical
1-1
Page 18
M
ODEL
951C NOX A
NALYZER
1.2 APPLICATIONS
The Model 951C Analyzer has specific applications in the following areas:
Oxides of nitrogen (NOx) emissions from the combustion of fossil fuels in:
Vehicle engine exhaust Incinerators Boilers Gas appliances Turbine exhaust
Nitric acid plant emissions
Ammonia in pollution control equipment (with converter)
Nitric oxide emissions from decaying organic material (i.e., landfills)
1-2
Rosemount Analytical
September 1997
748214-N
Page 19
I
NSTALLATION
2
2.1 UNPACKING
Carefully examine the shipping carton and contents for signs of damage. Immediately notify the shipping carrier if the carton or its contents are damaged. Retain the carton and packing material until the instrument is operational.
2.2 LOCATION
See drawing 654063 for Outline and Mounting dimensions. Install analyzer in a clean area, free from moisture and excessive vibration, at a stable
temperature within 4 to 40°C. The analyzer should be mounted near the sample source to minimize
sample-transport time. A temperature control system maintains the internal temperature of analyzer at 50°C
(122°F) to ensure proper operation over an ambient temperature range of 4°C to 40°C (40°F to 110°F). Temperatures outside these limits necessitate use of special temperature-controlling equipment or environmental protection. Also, the ambient temperature should not change at a rate exceeding 10°C/hr.
The cylinders of air and span gas should be located in an area of constant ambient temperature (±10°C).
2.3 VOLTAGE REQUIREMENTS
WARNING: ELECTRICAL SHOCK HAZARD
For safety and proper performance this instrument must be connected to a properly grounded three-wire source of power.
This instrument was shipped from the factory set up to operate on 115 VAC, 50/60 Hz electric power. For operation on 230 VAC, 50/60 Hz, position voltage select switches S1, S2, S3 (located on the Power Supply Board, Figure 2-1) and S3 (located on the Temperature Control Board, Figure 2-2) must be in the 230 VAC position.
748214-N
September 1997
Rosemount Analytical
2-1
Page 20
M
951C NOX A
ODEL
NALYZER
Refer to Figure 2-4. Remove the 6.25 A fuse (P/N 902413) and replace with the 3.15 A fuse (P/N 898587) provided in the shipping kit.
2.4 ELECTRICAL CONNECTIONS
The power and output (recorder and current) cable glands are supplied loose in the shipping kit to allow cable installation to connectors or terminal strips.
Cable Gland Part No.
Power 899330 Recorder 899329
Remove rear cover to access terminals. Route each cable through the cable gland and connect to the appropriate connector or terminal strip, tighten the gland.
2.4.1 L
INE POWER CONNECTIONS
Refer to Figures 2-3, 2-4 and drawing 654063. If this instrument is located on a bench or table top or is installed in a protected rack, panel or cabinet, power may be connected via a 3-wire flexible power cord, minimum 18 AWG (max. O.D. 0.480", min. O./D. 0.270"), through the hole labeled POWER, utilizing connector gland (P/N
899330) provided. Route the power cable through the cable gland and connect the leads to TB1. Tighten
the cable gland adequately to prevent rotation or slippage of the power cable. Since the rear terminals do not slide out with the chassis, no excess power cable slack is necessary.
The following power cord and/or support feet (for bench top use) are available:
Power Cord 634061
North American power cord set (10 foot)
Enclosure Support Kit 634958
Enclosure support feet (4)
Pow er Cord/Encl osure Support Kit 654008
North American power cord set (10 foot)
Enclosure support feet (4)
If the instrument is permanently mounted in an open panel or rack, use electrical metal tubing or conduit.
2-2
Rosemount Analytical
September 1997
748214-N
Page 21
I
NSTALLATION
IGURE
F
J5
J20
1
1
S2
230V
230V
115V
115V
115V
Set switch window for voltage required.
2-1. P
OWER SUPPLY BOARD VOLTAGE SELECT SWITCHES
1
115V
S1
J3
230V
115V
S2
230V
S1
230V
115V
655340 POWER SUPPLY BD
230V
115V115V 115V
115V
S3
CS
C10
S3
115V
115V
IGURE
F
SENSOR
AR1
J18
R10 R11 R7 R8
C2
CR1
C
B
Q2
C1
E
+
R18R19
R4
R3
K
A
R13
R2R1
Q1
G
CR2
R17R16 R12
TEMP CONTROL BD
Set switch window for voltage required.
2-2. T
EMPERATURE CONTROL BOARD
S3
3 2 1
U2
U1
J17
POWER LINE J5
2
3
1
S3
230
115
115
POWER SUPPLY
J11
C4
R15
R6
C3
R9 R5
CR
1
E
B
J19
Q3
TEST
C
R14
1
1 2 1 2 3
T.I.F. HEATER
748214-N
September 1997
Rosemount Analytical
2-3
Page 22
M
951C NOX A
ODEL
NALYZER
2.4.2 P
OTENTIOMETRIC RECORDER CONNECTIONS
Refer to Figures 2-3, 2-4 and drawing 654063. Potentiometric recorder connec-tions are made on the rear panel. Route the potentiometric recorder cable through the cable gland in the hole labeled RECORDER OUTPUT and connect to VOLT OUTPUT terminals.
Potentiometric recorder cable specifications are as follows:
Distance from recorder to analyzer: 1000 feet (305 meters) maximum
Input impedance: Greater than 2000 ohms
Cable (user supplied): Two-conductor, shielded, min. 20 AWG
Voltage output: 0 to +5 VDC
2.4.3 C
URRENT RECORDER CONNECTIONS
Refer to Figures 2-3, 2-4 and drawing 654063. Current recorder connections are made on the rear panel. Route the current recorder cable through the cable gland in the hole labeled RECORDER OUTPUT and connect to CUR OUTPUT terminals
Current recorder interconnection cable specs are as follows:
Distance the recorder from analyzer: 3000 feet (915 meters).maximum
Load resistance: Less than 700 Ohms.
Cable (user supplied): Two-conductor, shielded, min. 20 AWG
As supplied by the factory, the current output produces a zero of 4 mA. The current output may be adjusted to produce a zero of 0 mA as follows:
1. Zero the instrument as in Section 3.4.
2. Adjust R23, the zero-adjust potentiometer on the Power Supply Board, to produce 0 mA current output.
INTERIOR EXTERIOR
Nut Gland Nut
IGURE
F
2-4
2-3. C
Rosemount Analytical
ABLE GLAND
Cable
September 1997
Case Wall
748214-N
Page 23
I
NSTALLATION
IGURE
F
20 PSI (138 kPa)
NOMINAL
2-4. R
SAMPLE
EXHAUSTAIR
IN
IN
10 PSI - 17 PSI
(70 kPa - 120 kPa)
EAR VIEW OF MODEL
951C (
RECORDER
OUTPUT
FUSE
CUR OUTPUT
Current Output Connections
COVER REMOVED
L1/HOT
L2/NEUT
GND
VOLT OUTPUT
+ - G + -
)
POWER
2.5 GAS REQUIREMENTS
The instrument requires two gases normally supplied from cylinders. They are:
Fuse
AC Power Connections
Recorder Connections
IR
A
(U.S.P. B
REATHING GRADE
)
This is used as both (a) an oxygen source for generation of the ozone required for the chemiluminescence reaction, and (b) a standard gas for zero calibration (nitrogen can also be used). Gas for each purpose must be supplied from a separate cylinder due to different pressure requirements at ozonator and zero inlets.
PAN GAS
S
This is a standard gas of accurately known composition, used to set an upscale calibration point. The usual span gas is NO or NO2 in a background of nitrogen.
WARNING: HIGH PRESSURE GAS CYLINDERS
This instrument requires periodic calibration with a known standard gas. See Paragraphs 2.5 and 3.3. See also General Precautions for Handling and Storing High Pressure Gas Cylinders, following Section Six.
748214-N
September 1997
Rosemount Analytical
2-5
Page 24
M
951C NOX A
ODEL
NALYZER
Note For maximum calibration accuracy, the concentration of NO in the span
gas should be similar to that in the sample gas. Also, the span gas should be supplied to the rear panel SAMPLE inlet at the same pressure as the sample gas. To ensure constant pressure, a pressure regulator may be utilized immediately upstream from the SAMPLE inlet.
Each gas used should be supplied from a tank or cylinder equipped with a clean, non-corrosive type, two-stage regulator. In addition, a shut-off valve is desirable. Install the gas cylinders in an area of relatively constant ambient temperature.
2.6 SAMPLE REQUIREMENTS
The sample must be clean and dry before entering the analyzer. In general, before admission to the analyzer, the sample should be filtered to eliminate particles larger than two microns and have a dew point below 90°F (32°C). The factory can provide technical assistance if desired.
Proper supply pressure for sample, zero and span gases for the Model 951C is 20 psig (138 kPa).
2.7 GAS CONNECTIONS
WARNING: TOXIC AND OXIDIZING GAS HAZARDS
This instrument generates ozone which is toxic by inhalation and is a strong irritant to throat and lungs. Ozone is also a strong oxidizing agent. Its presence is detected by a characteristic pungent odor.
The instrument exhaust contains both ozone and nitrogen dioxide, both toxic by inhalation, and may contain other constituents of the sample gas which may be toxic. Such gases include various oxides of nitrogen, unburned hydrocarbons, carbon monoxide and other products of combustion reactions. Carbon monoxide is highly toxic and can cause headache, nausea, loss of consciousness, and death.
Avoid inhalation of the ozone produced within the analyzer and avoid inhalation of the sample and exhaust products transported within the analyzer. Avoid inhalation of the combined exhaust products at the exhaust fitting.
Keep all tube fittings tight to avoid leaks. See Section 2.8 for Leak Test Procedure.
Connect rear exhaust outlet to outside vent by a 1/4 inch (6.3 mm) or larger stainless steel or Teflon line. Check vent line and connections for leakage.
2-6
Rosemount Analytical
September 1997
748214-N
Page 25
1. Remove plugs and caps from all inlet and outlet fittings. (See Figure 2-4.)
2. Connect EXHAUST outlet to external vent via tubing with O.D. of 1/4-inch (6.3 mm) or larger. Use only stainless steel or Teflon tubing.
3. Connect external lines from ozonator air and sample sources to corresponding rear panel inlet ports. For sample line, stainless steel tubing is recommended.
4. Adjust regulator on ozonator air cylinder for output pressure of 20 to 25 psig (138 to 172 kPa). At least 20 psig should be present at rear of analyzer.
5. Supply sample gas to rear panel SAMPLE inlet at appropriate pressure: 20 psig (138 kPa). The nominal input pressure is 20 psig (138 kPa).
2.8 LEAK TEST
The following test is designed for sample pressure up to 5 psig (35 kPa).
1. Supply air or inert gas such as nitrogen at 5 psig (35 kPa) to analyzer sample and air input fittings.
I
NSTALLATION
2. Seal off analyzer exhaust fitting with a tube cap.
3. Use a suitable test liquid such as SNOOP (P/N 837801) to detect leaks. Cover all fittings, seals, or possible leak sources.
4. Check for bubbling or foaming which indicates leakage, and repair as required. Any leakage must be corrected before introduction of sample and/or application of electrical power.
748214-N
September 1997
Rosemount Analytical
2-7
Page 26
M
ODEL
N
OTES
951C NOX A
NALYZER
2-8
Rosemount Analytical
September 1997
748214-N
Page 27
I
(S1)
(
)
NITIAL STARTUP AND OPERATION
3.1 FRONT PANEL INDICATORS AND CONTROLS
3
3.1.1 D
ISPLAY
The display is a 4-digit liquid crystal device which always displays NOx concentration in parts-per-million. See Figure 3-1.
3.1.2 R
ANGE SELECTION
The Model 951C has eight customer selectable ranges, four LO ranges (10 ppm, 25 ppm, 100 ppm and 250 ppm) and four HI ranges (100 ppm, 250 ppm 1000 ppm and 2500 ppm). The range is selected by positioning the RANGE Switch (S1) and the three jumpers on the Signal Board to the d esired range controlling the recorde r output. Refer to Figure 3-2.
The display blanks for values 10% in excess of the range maximum. Moving the switch to the left selects a higher fullscale value and restores the display.
POWER SUPPLY BOARD (See Figure 3-3)
CASE HEATER TEMPERATURE CONTROL ASSEMBLY (See Figure 6-4) TEMPERATURE CONTROL BOARD (See Figure 2-2)
Voltage Select (S3)
Voltage Select (S3)
Voltage Select (S1)
Voltage Select (S2)
SAMPLE PRESSURE GAUGE
SAMPLE PRESSURE REGULATOR (Adjustment Knob)
SIGNAL BOARD (See Figure 3-2)
Display (Signal Board DS1)
Adj. (R8)
Gain (R24)
Signal (R20)
Cal (R18)
Range Select Switch
Convertor Heater (R9)
Converter
Temp Check
(S4)
PMT High Voltage
R30
TP2
Current Output Zero (R23)
Current Output Span (R20)
Zero Control (Signal Board R100)
IGURE
F
748214-N
3-1. M
ODEL
Span Control (Signal Board R101)
951C C
Ozone Indicator Lamp (Signal Board DS2)
ONTROLS
NDICATORS AND ADJUSTMENTS
, I
September 1997
Rosemount Analytical
3-1
Page 28
M
951C NOX A
ODEL
NALYZER
3.1.3 S
AMPLE PRESSURE GAUGE
The internal SAMPLE pressure (nominally 4 psig, 28 kPa) is adjusted by rotation of the Sample Pressure Regulator. See Figure 3-1.
3.1.4 O
ZONE PRESSURE
The OZONE pressure is determined by the pressure regulator of the air supply cylinder. A nominal pressure of 20 to 25 psig (138 to 172 kPa) is recommended. Proper operation is indicated when the front panel OZONE indicator lamp is lit.
Note If ozone lamp does not light, increase pressure slightly by adjusting
pressure regulator control on the air cylinder.
3.1.5 Z
ERO AND SPAN POTENTIOMETERS
See Figures 3-1 and 3-2. Screwdriver access holes through the front panel allow adjustments of the ZERO and SPAN potentiometers (R100 and R101 on Signal Board.
3.1.6 O
ZONE INTERLOCK
The ozone-producing UV lamp will not ignite or stay lit unless adequate air pressure is present at the AIR inlet (see Figure 2-4). Nominal set point pressure is 20 to 25 psig.
3.2 STARTUP PROCEDURE
The following are detailed instructions on startup and calibration.
1. Supply electrical power to the analyzer. The analyzer will require approximately two hours for temperature equilibration before calibration.
2. On Signal Board, Figure 3-2, set PPM RANGE Switch (S1) to 250 ppm.
3. Establish correct pressure for air by the following: a. Adjust OZONE Pressure Regulator so that OZONE Pressure Gauge indicates
20 to 25 psig (138 to 172 kPa).
b. To establish correct pressure of zero gas, supply zero gas to rear panel SAMPLE
inlet. Note reading on internal SAMPLE Pressure Gauge. It should be the same as the nominal 4 psig (28 kPa) SAMPLE pressure indicated on the internal SAMPLE pressure gauge. This should remain constant when the analyzer input SAMPLE is switched from calibration gas standard to a zero gas standard. This may be assured by setting the delivery from the SAMPLE and the zero gas cylinder of span gas cylinder to the same value of delivery pressure, nominally 20 psig (138 kPa). If not, adjust output pressure regulator on zero g as cylinder as required.
3-2
Rosemount Analytical
September 1997
748214-N
Page 29
I
654050 SIGNAL CONTROL BD
NITIAL STARTUP AND OPERATION
Lo Hi
E6 E7
Lo Hi
E1 E2
E4
E3
E1 E2
E4
E3
E5
E5
E6 E7
C
DS2
SIGNAL IN
E6 E7
R28
R29
C14
R34
RP3
1
K3
C15
R37
AR4
R38
RP2
AR5
AR6
R39 R40 R23
J1
C12
C16
CR23
SPAN
10 100 25 250
TP6 TP 5 TP 4 TP3 TP2 TP1
E1 E2
E3
DP SELECT
DS1
U7
+
C13
R35
R17
E
B
C9
R36
Q3
C
R101
CW S CCW
E4
RANGE
E5
CR1
R14 R12
C5
S1
10 100 25 250
RP1
1
U1
CR13
CR12
CR14
CR11
CR2
C8
U2
R22
1 (10) 10 2 (25) 25 3 (100) 100 4 (250) 250
S1
CR7
CR6
CR10
CR8
R19
CR9
C6
C10
R13
+
ADJ. GAIN SIG. CAL.
R41
R27
R25 R24 R20 R18
CR15
C11
+
CR3
CR5
C7
CR19
CR4
R21
R1R2R3R4R5R6R7R8R9
R15
R30 AR1
R31
R16
U3
ZERO
R100
Range
(S1 pos)
T1
CR17
CR16
CR22
CR20
CR21
VR1
CW S CCW
R42
CR18
R32 R33
Lo
+
C17
B
Q1 CE
AR2
R10
R11
B
C
Q2
ppm
Fullscale
C3
E
(S1 pos)
1 (10) 100 2 (25) 250 3 (100) 1000 4 (250) 2500
C2 C1
AR3
C4
J3
K1
K2
CS
Range
Hi
ppm
Fullscale
R101 - SPAN Potentiometer R100 - ZERO Potentiometer
IGURE
F
3-2. S
IGNAL BOARD
4. Establish correct pressure of sample gas by the following: a. Supply sample gas to rear panel SAMPLE inlet. b. Adjust SAMPLE Backpressure Regulator so internal SAMPLE Pressure Gauge
indicates the value appropriate to the desired operating range.
Note Inability to obtain a flow of one liter per minute at the EXHAUST outlet
usually indicates insufficient sample supply pressure at the SAMPLE inlet. Use a 2400 cc flowmeter (i.e., Brooks P/N 1350) at the EXHAUST outlet to measure flow.
5. Establish correct flow of upscale standard gas by the following:
a. Supply upscale standard gas to rear panel SAMPLE inlet.
b. Note reading on internal SAMPLE Pressure Gauge. It should be the same as in
Step 3b.
Note Supply pressures for sample and upscale standard gases must be the
same. Otherwise, readout will be in error.
The analyzer is now ready for calibration.
748214-N
September 1997
Rosemount Analytical
3-3
Page 30
M
3
Q
Q
A
R16
Q2
Q1
R14
R18
R12
Q3
C9
CR5
9
R78
U23
U
C16
C18
35
66
Q
3
CR16
U22
36R3
3
CR9
C
CR23
R69
J
33
A
J
J
J16
S1
S2
S3
951C NOX A
ODEL
NALYZER
TP13
CR4
CONVERTOR
TP13
CR4
CONVERTOR
Q15
E
B
C
1
2
7
TP1
TP1
CONV TEMP
CHECK
C8
R77
CR2
R27
CONV TEMP
CHECK
C8
TP2 TP5 TP6 TP7 TP8
R2
S4
R4
R6 R13
C
B
E
CR6
R15
R7
Q14
C
E
B
B
U21
R62
R63
5
+
C20
R28
C
Q4
C17
TP2 TP5 TP6 TP7 TP8
S4
K
G
R3
A
B
C
E
AR1
1
R71
R68
R64
R34
B
E
R
CONV
+
R67
HTR
C7
R6
R7
R11
J19
1
R72
RP1
U20
CR24
R76
R75
R74
R
R
R7
R
R37
2
AR2
C19
R
CURRENT
SPAN
R22 R24 R25 R21 R19
CR15
R20
R26
VR2
U7
VR1
VR3
TP14 TP15
VR6
U8
1 2 3 4
C4
+
R23
R20
CURRENT
OUTPUT
SPAN
ZERO
TP3 TP4
C11
+
J11
1
CR7
+
C13
I
+
G O
VR4
C14
CR8
+
C12
CR14
O G
I
+
C15
C3
+
+
CR1
+
+
C2
+
+
VR5
I G O
CR18
C29
+
VR7
C1
G I O
+
CR2
+
C6 C5
K1
VR8
O G I
C25
+
C27
+
CR19
+
C26
R61
R56
R60
R59
J6
1
CR3
R1
12
E
2
C23
C21
R48
B
R46
C
R44
C
R43 R41
Q8
O
VR9
I
J18
C28
1
+
C
E
B
Q13
CR20
+
R58
R57
J20
1
1
TP9 TP10 TP11 TP12
R54
C
E
+
R47
+
R49
E
R42
Q9
C
R40
B
G
E
10
R53
J5
1
CR13
1
Q24
B
E
R4
U9
R68
R82
R81 R80 R45
R51 CR12
R55
J15
J3
S2
230V
115V
115V
C
J17
1
C22
J4
1
1
S1
230V
115V
115V
655340 POWER SUPPLY BD
J9
1
2
1
J12
230V
115V
1
1
J8
1
1
U2
R17 R18
C10
S3
115V
CS
R9
R30
CONV
PMT
HTR
HV
THERMO COOLER
C
CR10
R29
TP14 TP15
CR1
R2
J14
OUTPUT
ZERO
J13
Q7
B
E
R3
RMT
HV
THERMO COOLER
J19
R38
R39
C
Q6
B
E
R11
1
CR17
115V
230V
115V
115V
230V
115V
115V
IGURE
F
3-3. P
230V
115V
OWER SUPPLY BOARD
3.3 CALIBRATION
3.3.1 Z
3.3.2 U
ERO CALIBRATION
1. On the Signal Board, Figure 3-2, set PPM RANGE Switch for the same range that will be used during sample analysis. Set SPAN Control at about mid-range.
2. Supply zero gas to rear panel SAMPLE inlet.
3. After a stable reading is reached, adjust the zero by inserting a screwdriver in the ZERO slot on the front of the analyzer and turning until zero reading is obtained.
PSCALE CALIBRATION
1. On the Signal Board, Figure 3-2, set PPM RANGE Switch to the position appropriate to the particular span gas.
2. Supply upscale standard gas of accurately known NOx content to rear panel SAMPLE inlet.
3-4
Rosemount Analytical
September 1997
748214-N
Page 31
I
NITIAL STARTUP AND OPERATION
3. Adjust SPAN Control so that reading on display or recorder is equal to the known parts-per-million concentration of NOx in the span gas. If the correct reading is not initially attainable by adjustment of the SPAN Control, make the electronic adjustment in Step 4.
4. If necessary, increase sensitivity by raising photomultiplier voltage. This will interact with zero. Repeat Zero Calibration and Upscale Calibration (through step 3).
3.4 ROUTINE OPERATION
After calibrating analyzer per Section 3.3, supply sample to SAMPLE inlet. Set PPM RANGE Switch in appropriate position. The instrument will now continuously analyze the sample stream.
The Model 951C is designed for continuous operation. Normally, it is never turned off except for servicing or for a prolonged shutdown.
Note During periods of shutdown, turn off the ozone lamp by shutting off the
input air source.
3.5 CONVERTER TEMPERATURE ADJUSTMENT PROCEDURE
Once the appropriate high voltage and electronic gain have been selected such that the named calibration gas value is indicated by the Model 951C, the instrument is ready for adjustment of the converter temperature.
The vitreous carbon converter used in this analyzer has a low surface area which gradually increases during high temperature operation of the converter material.
Initially, the temperature of the peak of the converter efficiency starts at a relatively high value because significant heat must be supplied to make the converter active enough to reduce the input nitrogen dioxide to nitric oxide at the required 95% level. During the operation of the analyzer, the temperature of the peak will fall as the surface area of the converter is increased and less external energy is required to cause adequate conversion.
In extreme cases, where converter re-profiling has not been conducted, the converter is so active that it not only reduces nitrogen dioxide to nitric oxide, but it reduces the nitric oxide produced to nitrogen, which is not detected by the chemiluminescence reaction. The remedy in this case is to adjust the converter temperature to a lower value to improve the converter efficiency.
It is important that the converter temperature be periodically profiled to assure that it is running at its peak efficiency. An interval of one week is recommended. The nominal range of operational temperatures for the converter is 275°C to 400°C (527°F to 750°F). The operating temperature of the converter may be conveniently checked by momentarily depressing switch S4 on the Power Supply Board while monitoring the
748214-N
September 1997
Rosemount Analytical
3-5
Page 32
M
951C NOX A
ODEL
NALYZER
resistance across terminals TP1 and TP2. Table 3-1 allows for conversion of the observed resistance to the operating temperature for the converter.
Follow this procedure to optimize the operating temperature of the converter:
1. Power instrument and allow it to stabilize at operating temperature (one to two hours). Measure the operating temperature of the converter by the technique described above. Note the value for future reference.
2. Admit a calibration gas of known (NO2) concentration into the analyzer and note the concentration value determined when the full response has been achieved.
3. Refer to Figure 3-3. Turn the converter temperature adjust potentiometer R9, on the Power Supply Board one turn
counterclockwise
from the setting established at the factory, and allow fifteen minutes for operation at the new lower temperature setpoint. Recheck the response and note the value for later use.
4. Increase the temperature of the converter by rotating the converter temperature adjust potentiometer, R9, one quarter turn
clockwise
, wait fifteen minutes for thermal equilibrium and then re-measure the NO2 calibration gas value. Note its value. Repeat this procedure of one quarter turn adjustments of the potentiometer, waiting for thermal stability and determination of the calibration gas value until either a 95% value is obtained or the final one quarter turn adjustment gives an efficiency increase of less than one percent.
5. Decrease the temperature of converter operation by rotating the converter temperature adjust potentiometer one eighth of a turn counterclockwise. This places the converter at a temperature suitable for low ammonia interference and efficient NO2 conversion. Re-measure the indicated converter temperature and compare it to the initially recorded value.
ABLE
T
3-1. R
TEMPERATURE
(°C)
0 400
25 438 100 552 200 704 250 780 300 856 350 932 400 1008 450 1084
ESIST ANCE OF CONVERTER TEMPERA TURE SENSOR VS
RESISTANCE
(Ohms)
EMPERA TURE
. T
3-6
Rosemount Analytical
September 1997
748214-N
Page 33
I
NITIAL STARTUP AND OPERATION
Note Converter temperature is not a direct measure of converter efficiency.
Temperature measurement is for reference purposes only.
3.6 MEASUREMENT OF CONVERTER EFFICIENCY
It is the responsibility of the user to measure efficiency of the NO2-to-NO converter during initial startup, and thereafter at intervals appropriate to the application (normally once a month).
The above procedure optimizes the operating temperature of the converter. It also serves as an efficiency check if the concentration of NO2 in the calibration gas is documented accurate relative to National Institute of Standards and Technology (NIST) Reference Materials. If the concentration of the nitrogen dioxide calibration gas is not known accurately, this procedure still serves to adequately provide the correct converter operating temperature.
If the only available known standard is the nitric oxide calibration standard, the following procedure may be performed. This procedure checks converter efficiency through the utilization of gas-phase oxidation of nitric oxide into nitrogen dioxide over a range of nitrogen dioxide concentrations. This technique is abstracted and adapted from 40 CFR, Pt. 60, App. A, Method 20, Paragraph 5.6.
1. Select the appropriate instrument range.
2. Admit a nitric oxide in nitrogen NIST traceable calibration gas of a value between 45% and 55% of the instrument range selected to a clean, evacuated, leak tight Tedlar bag. Dilute this gas approximately 1:1 with a 20.9% oxygen, purified air.
3. Immediately attach the bag outlet to the input of the pump supplying pressurized gas to the analyzer. It is important to use a sample delivery pump which does not consume nitrogen dioxide as it delivers sample to the analyzer. Losses of nitrogen dioxide in the pump will be reported as converter inefficiency.
4. Operate the analyzer and continue to sample the diluted nitric oxide sample for a period of at least thirty minutes. If the nitrogen dioxide to nitric oxide conversion is at the 100% level, the instrument response will be stable at the highest value noted.
5. If the response at the end of the thirty minute period decreases more than 2.0 percent of the highest peak value observed, the system is not acceptable and corrections must be made before repeating the check. If it is determined that observed subnormal conversion efficiencies are real, and not due to errors introduced by nitrogen dioxide consumption in the sample pump or other parts of the sample handling system, verify that the converter is peaked at the optimum temperature before replacing with a new converter.
748214-N
September 1997
Rosemount Analytical
3-7
Page 34
M
951C NOX A
ODEL
NALYZER
3.7 RECOMMENDED CALIBRATION FREQUENCY
After initial startup or startup following a shutdown, the analyzer requires about two hours for stabilization before it is ready for calibration. Maximum permissible interval between calibrations depends on the analytical accuracy required, and therefore cannot be specified. It is recommended that initially the instrument be calibrated at least once every 8 hours. This practice should continue until experience indicates that some other interval is more appropriate.
3-8
Rosemount Analytical
September 1997
748214-N
Page 35
T
HEORY
4
4.1 NITRIC OXIDE DETERMINATION BY CHEMILUMINESCENCE METHOD
The chemiluminescence method for detection of nitric oxide (NO) is based on its reaction with ozone (O3) to produce nitrogen dioxide (NO2) and oxygen (O2). Some of the NO2 molecules thus produced are initially in an electronically excited state (NO2*). These revert immediately to the ground state, with emission of photons (essentially red light).
The reactions involved are:
NO + O3 → NO2* + O NO2* → NO2 + Red Light
As NO and O3 mix in the reaction chamber, the intensity of the emitted red light is proportional to the concentration of NO.
(Any NO2 initially present in the sample is reduced to NO by a heated bed of vitreous carbon through which the sample is passed before being routed to the reaction chamber.)
The intensity of the emitted red light is measured by a photomultiplier tube (PMT), which produces a current of approximately 3 X 10-9 amperes per part-pe r-million of NO in the reaction chamber.
4.2 ANALYZER FLOW SYSTEM
The analyzer flow system is shown in drawing 654090. Its basic function is to deliver regulated flows of sample, calibration gas, or zero gas and ozonized air to the reaction chamber. The discharge from the reaction chamber flows from the analyzer via the EXHAUST outlet.
2
4.2.1 F
748214-N
LOW OF SAMPLE
Suitably pressurized sample, standard gas or zero gas is supplied to the rear panel SAMPLE inlet.
TANDA RD GAS OR ZERO GAS TO REACTION CHAMBER
, S
September 1997
Rosemount Analytical
4-1
Page 36
M
951C NOX A
ODEL
NALYZER
The flow rate of the selected gas into the reaction chamber is controlled by a back pressure regulator inside the analyzer. It provides an adjustable, controlled pressure on the upstream side, where gas is supplied to the calibrated, flow-limiting sample capillary. The regulator is adjusted for appropriate reading on the internal SAMPLE Pressure Gauge. For operation at NO and NO2 levels below 250 ppm, correct setting on the SAMPLE Pressure Gauge is 4 psig (28 kPa). This results in a flow of approximately 60 to 80 cc/min to the reaction chamber.
Excess sample is discharged with the effluent from the reaction chamber via the EXHAUST outlet. Bypass flow is set by the restrictor at 1 L/min (nominal) to ensure proper functioning of the SAMPLE Pressure Regulator and rapid system response. Excessive changes, on the order of 5 psig (35 kPa), in the pressure of the sample or standard gas will affect the bypass flow rate and can affect accuracy.
4.2.2 O
ZONE GENERATION
Suitably pressurized air from an external cylinder is supplied to the rear panel AIR inlet. The proper pressure setting is 20 to 25 psig (138 to 172 kPa). W ithin the ozone generator, a portion of the oxygen in the air is converted to ozone by exposure to an ultraviolet lamp. The reaction is:
UV
3O2 → 2O
3
From the generator, the ozonized air flows into the reaction chamber for use in the chemiluminescence reaction.
4.3 SIGNAL PROCESSING ELECTRONICS SYSTEM
A block diagram of the signal-processing electronics is shown in Figure 4-1. Basic functions of these electronics are acceptance of PMT output and conversion of it to potentiometric and isolated current outputs, and providing a visual display of the concentration of the NOx in the sample stream. All functions except the high-voltage source and the voltage-to-current converter are contained on the Signal Control PC Board, 654050. The two exceptions are located on the Power Supply Board, 654059.
The PMT drives a high input impedance amplifier which produces a voltage between 0 and approximately 5 volts. The front panel Zero Control injects a small current into the PMT amplifier to null any current from the PMT which is not related to the concentration of NOx in the sample stream.
The PMT amplifier drives a programmable gain amplifier (PGA). The gain of the PGA is controlled by the Range Switch.
The PGA drives the Span Amplifier. The gain of this amplifier is controlled by the front panel Span Control. The output of the Span Amplifier is a voltage which is properly scaled to represent the concentration of NOx in the sample stream.
The Span Amplifier drives the front panel Display and associated electronics, and the isolated current output. It also provides the potentiometric output.
4-2
Rosemount Analytical
September 1997
748214-N
Page 37
4.4 ANALYZER THERMAL SYSTEM
The Analyzer Thermal System is shown in Figure 4-2. Its basic function is to provide a stable thermal environment for the PMT.
The temperature of the PMT must be held within a half-degree band at approximately 18°C if it is to produce a useful signal for low concentrations of NOx. This is accomplished by means of a solid-state cooler which houses the PMT. The heat which is radiated from the cooler is carried away by the Cooler Fan.
The solid-state cooler must work against a relatively constant load in order to maintain the temperature of the PMT. This load is produced by a case heater and exhaust fan which control the temperature inside the case within a one-degree band (approximately 50°C for ambient temperatures from 4°C to 40°C).
The electronics which support the Analyzer Thermal System and the NO2-to-NO Converter are contained on the Power Supply Board.
T
HEORY
IGURE
F
Photomultiplier
Tube
High Voltage
Supply
4-1. A
Signal/Control Board
PMT
Amplifier
Zero Cont rol
Programmable
Gain
Amplifier
Range Switch Span Control
Span
Amplifier
Power Supply Board
NALYZER SIGNAL CONDITIONING CIRCUIT
Display
Voltage-to-Current
Converter
Potentiometric Output
Isolated Current Output
748214-N
September 1997
Rosemount Analytical
4-3
Page 38
M
y
951C NOX A
ODEL
NALYZER
IGURE
F
4-2. A
INLET VENT HOLES
SOLID-STATE COOLER
CASE HEATER
Fan Heater
NALYZER THERMAL SYSTEM
Top View of Anal
PMT
Cooling Fins
Cooler Fan
FRONT PANEL
zer
EXHAUST FAN
4-4
Rosemount Analytical
September 1997
748214-N
Page 39
R
OUTINE SERVICING
5
WARNING: ELECTRICAL SHOCK HAZARD
Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel.
WARNING: INTERNAL ULTRAVIOLET LIGHT HAZARD
Ultraviolet light from the ozone generator can cause permanent eye damage. Do not look directly at the ultraviolet source in ozone generator. Use of ultraviolet filtering glasses is recommended.
Note The photomultiplier tube must not be exposed to ambient light. If the
photomultiplier tube is exposed to light while the power is on, either through a loose fitting on the reaction chamber or any other leak, it will be destroyed. If exposed to ambient light with the power off, the tube will be noisy for some period of time. Unless appropriate precautions are observed, light can strike the tube upon removal of fittings from the reaction chamber.
5.1 SYSTEM CHECKS AND ADJUSTMENTS
The following procedures may be used to determine the cause of unsatisfactory instrument performance, or to make adjustments following replacement of components. If a recorder is available, use it for convenience and maximum accuracy in the various tests.
5.1.1 D
ISPLAY FULLSCALE SPAN ADJUSTMENT
If a recorder is used, and has been properly zeroed, it should agree with the display reading. If not, obtain agreement by adjustment of R20 on the Signal/Control Board (see Figures 3-1, 3-2). If agreement cannot be reached, check the recorder. If the recorder is functioning properly, replace the amplifier board.
748214-N
September 1997
Rosemount Analytical
5-1
Page 40
M
951C NOX A
ODEL
NALYZER
5.1.2 O
VERALL SENSITIVITY
Principal factors that determine overall sensitivity of the analyzer are the following: (a) sample flow rate to the reaction chamber, (b) sensitivity of the photomultiplier tube (PMT), and (c) PMT high voltage. If specified fullscale readings are unobtainable by adjustment of the SPAN Control, sensitivity is subnormal. The cause of reduced sensitivity may be in either the flow system (See Section 5.2) or the electronic circuitry (See Section 5.6).
If either the High Voltage Board or the Phototube/Reaction Chamber Assembly has been replaced, a readjustment of the high voltage will probably be required to obtain the correct overall sensitivity. Adjust R30 on the Power Supply Board (see Figures 3-1, 3-3) clockwise to increase (negative) the photomultiplier high voltage and sensitivity, or counterclockwise to decrease (negative) the voltage and sensitivity. The adjustment range is about -650 V to -2100 V for the regulated DC voltage applied to the photomultiplier tube. Nominal setting is -1100 volts. However, the voltage should be adjusted as required for overall system sensitivity.
5.1.3 O
ZONE OUTPUT
To check for adequate output from the ozone lamp, a convenient technique is to calibrate the analyzer on a high level NO standard such as 250 ppm NO at the nominal 4.0 psi internal sample pressure setpoint, and note the reading. The sample pressure setpoint is then sequentially set to pressures of 3.0, 2.0, and 1.0 psi after a stable span gas reading has obtained at the higher pressure setpoint. The span gas value will change as the pressure is changed. The difference in span gas value between two successive sample pressure levels should be approximately the same for the 4.0 to 3.0, 3.0 to 2.0, and 2.0 to 1.0 pressure steps.
If the size of the span gas value difference increases as the sample pressure is lowered, the analyzer output is limited by the amount of ozone production from the lamp and the two additional checks should be made. First, verify that the sample flow (not including bypass) does not exceed the nominal 60 to 80 cc/min, at 4.0 psi internal sample pressure. Second, substitute another lamp to see if the ozone output is increased.
If no other ozone lamp is available, the analyzer sample input pressure may be reduced to the pressure where the ozone limitation is not present. If the lamp output is low and the sample pressure is reduced to restore operation to the condition where ozone limitation is not occurring, some degradation in analyzer response time characteristics may occur.
WARNING: TOXIC GAS HAZARD
Use extreme caution in troubleshooting the ozone generator. Ozone is toxic.
5-2
Rosemount Analytical
September 1997
748214-N
Page 41
R
OUTINE SERVICING
5.1.4 B
ACKGROUND CURRENT
With zero air supplied to rear panel SAMPLE inlet, excessive background current is evidenced by the inability to obtain zero display reading with adjustment of the ZERO Control. If this cannot be accomplished, the cause must be found and corrected. The fault may be in either the electronic circuitry or the sample flow system.
First, establish proper performance of the electronic circuitry. Turn on analyzer power. Verify that ZERO Control and amplifier are functioning properly. Then, check for excessive photomultiplier dark current and/or contamination of the reaction chamber or sample flow system as follows:
5.1.5 E
XCESSIVE PHOTOMULTIPLIER DARK CURRENT
To check, shut off all flow to the ozone generator. Turn off ozone generator. Supply cylinder air to rear panel SAMPLE inlet. Note response on display or recorder. If back­ground is still excessive, possible causes are:
leakage of ambient light to photomultiplier tube
defective photomultiplier tube
electrical leakage in socket assembly
ONTAMINATION OF REACTION CHAMBER OR SAMPLE FLOW SYSTEM
C
.
See Section 5.4.1.
5.2 SERVICING FLOW SYSTEM
To facilitate servicing and testing, the Model 951C has front drawer access. Drawing 654090 shows flow system details, including fittings, thread specifications
and connecting tubing.
5.2.1 C
LEANING SAMPLE CAPILLARY
If clogging of sample ca pillary is suspected, measure flo w rate as described below.
1. Turn off instrument power and shut off all gases.
2. Refer to Figures 6-1 and 6-3. Cover and shade the fittings on the reaction chamber with a dark cloth or other light-shielding material. Remove the fitting associated with the sample capillary and place a cap over the open fitting to prevent entry of stray light.
748214-N
September 1997
Rosemount Analytical
5-3
Page 42
M
951C NOX A
ODEL
NALYZER
Note: If the opened fitting is inadvertently exposed to ambient light, the
instrument will temporarily give a highly noisy background reading. If so, this condition may be corrected by leaving the instrument on, with high voltage on, for several hours. If high voltage is on during exposure, the photomultiplier tube will be destroyed.
3. With instrument power off, supply suitable test gas (dry nitrogen or air) to rear-panel SAMPLE inlet.
4. Connect a flowmeter to open end of sample capillary. Adjust internal SAMPLE Pressure Regulator to normal operating setting of 4 psig (28 kPa). Verify that flowmeter indicates appropriate flow of 60 to 80 cc/min.
5. If flow is correct, restore analyzer to normal operation.
6. If flow is low, the capillary requires cleaning or replacement (Proceed with the step 5 below).
7. Clean capillary with denatured alcohol, and purge with dry nitrogen or air f or one minute. Reconnect capillary.
8. With the photomultiplier still covered, slowly insert the free end of the capillary into the corresponding fitting on the reaction chamber. Push the capillary in until it touches bottom against the internal fitting. Then tighten fitting 1/4 turn past finger tight.
Note: Do not overtighten capillary internal fitting, as overtightened fittings may
restrict the sample flow.
5.2.2 O
ZONE RESTRICTOR FITTING
With instrument power off, supply suitable test gas (dry nitrogen or air) to rear panel AIR inlet. Cover photomultiplier housing with a dark cloth. At the fittings on the reaction chamber, disconnect the ozone tube and place a cap over the open fitting to prevent entry of ambient light. Connect a flowmeter to open end of ozone tube. Adjust the OZONE Pressure Regulator so that the OZONE Pressure Gauge indicates normal operating pressure of 20 to 25 psig (138 to 172 kPa). Verify that test flowmeter indicates an appropriate flow of 500 to 600 cc/min for 20 psig.
Subnormal flow indicates clogging in the flow path that supplies air to the ozone generator. This path contains a Restrictor (P/N 655519), consisting of a metal fitting with internal fritted (metal membrane) restrictor to reduce pressure. The fitting is upstream from the inlet port of the ozone generator. If the internal restrictor becomes plugged, the assembly (P/N 655519) must be replaced as it cannot normally be cleaned satisfactorily.
5-4
Rosemount Analytical
September 1997
748214-N
Page 43
R
OUTINE SERVICING
5.3 PHOTOMULTIPLIER TUBE/REACTION CHAMBER
This assembly consists of the photomultiplier tube and socket, the thermoelectric cooler, and the reaction chamber. Refer to Figure 6-1 for location and details of mounting. Refer to Figure 6-3 for information on the assembly.
The assembly must be removed from the analyzer in order to clean the reaction chamber or to replace the photomultiplier tube.
5.3.1 P
HOTOMULTIPLIER TUBE/REACTION CHAMBER REMOVAL
To remove the photomultiplier tube/reaction chamber assembly from the analyzer, do the follow:
1. Disconnect power from the analyzer.
2. Release pressure from SAMPLE and AIR supplies.
3. Unplug the electrical cable from the Power Supply PC Board.
4. Disconnect the high-voltage cable and the signal cable from the left side of the assembly. Note the two mounting screws just below the connectors.
5. Uncouple the sample and ozone capillaries and the exhaust line from the right side of the assembly. Note the two mounting screws just below the fittings.
6. Loosen the screws described in steps 4 and 5 above.
7. Lift the assembly from the analyzer.
8. Replace the assembly by reversing the order of steps 1 through 7.
5.3.2 C
LEANING REACTION CHAMBER
Note: Photomultiplier tube will be permanently damaged if exposed to ambient
light while powered with high voltage. Photomultiplier tube will develop temporary electronic noise if exposed to ambient light with high voltage OFF. A temporary noisy condition may be corrected by leaving instrument on, with high voltage on, for several hours. The required recovery time depends on intensity and duration of the previous exposure. Noise level on the most sensitive range usually drops to normal within 24 hours.
If sample gas is properly filtered, the reaction chamber should not require frequent cleaning. In event of carryover or contamination, however, the chamber should be disassembled to permit cleaning the quartz window and the optical filter. The following procedure is recommended.
748214-N
September 1997
Rosemount Analytical
5-5
Page 44
M
951C NOX A
ODEL
NALYZER
1. Cover and shade the Reaction Chamber/Photomultiplier Assembly with a dark cloth or other light-shielding material.
Note: Always wear surgical rubber gloves when handling the reaction chamber to
prevent contamination from handling.
2. Note the orientation of the fittings. Slowly rotate and withdraw the reaction chamber from the housing. Ensure that no light strikes the photomultiplier tube.
3. Unscrew plastic end cap, thus freeing the quartz window and the red plastic optical filter. Note the sequence in which these are assembled.
4. Clean the reaction chamber by the appropriate one of the following two methods, standard or alternate. The standard method is applicable in most cases. The alternate method is applicable when the instrument has shown high residual fluorescence. That condition is indicated by high residual currents on a zero gas and high differentials between zero gas readings obtained with the ozone lamp on and off.
TANDA RD CLEANING PROCEDURE
S
Using a stiff plastic bristle brush, such as a toothbrush, scrub the Teflon surface and gas ports of the reaction chamber with clean distilled water and Alconox* detergent (P/N 634929). Alconox detergent is included in the shipping kit provided with the Model 951C NOx Analyzer, and is available from Sargent-Welch Scientific Company under its catalog number S-195650-A.
Using Alconox and clean, soft facial tissue (NOT industrial wipes), carefully clean the quartz window. Vigorously flush reaction chamber and quartz window with clean distilled water. Blow out all possible water from internal passages of reaction chamber. Dry reaction chamber and quartz window in a warm oven at 125°F to 150°F (52°C to 66°C) for 30 to 45 minutes or purge-dry the parts with dry cylinder air or nitrogen to eliminate all moisture.
WARNING: ACID HAZARD
Hydrochloric acid (HC1) is a strong acid. It is irritating to the skin, mucous membranes, eyes and respiratory tract. Direct contact causes severe chemical burns.
Avoid Contact with eyes and skin and avoid breathing fumes. Use in hood or well ventilated place. Wear goggles, rubber gloves and protective clothing.
5-6
Rosemount Analytical
September 1997
748214-N
Page 45
R
OUTINE SERVICING
LTERNATE CLEANING PROCEDURE
A
OR HIGH RESIDUAL FLUORESCENCE
- F
Holding the reaction chamber by the tube fittings, and using appropriate caution, immerse the white Teflon part of the chamber in 50% concentrated Reagent Grade hydrochloric acid. After five minutes, rinse thoroughly with de-ionized water, then air dry as in the standard cleaning method above.
Place parts in position and press on end-cap so that mating threads engage properly, without cross threading. Turn mating parts in one continuous motion until the parts mesh. Do not over-torque.
With reaction chamber now assembled, replace and reconnect it in reversed removal sequence. Orient as noted in step 2.
5.3.3 P
HOTOMULTIPLIER TUBE AND HOUSING
The photomultiplier tube operates at high DC voltages (nominal setting is -1100 volts) and generates small currents that are highly amplified by the signal-conditioning circuitry. It is therefore important that ambient humidity and condensed water vapor be excluded from the interior of the photomultiplier housing. Ambient humidity can result in electrical leakage, observed as abnormally high dark current. Water vapor or condensed moisture in contact with the photomultiplier tube may result in an abnormally high noise level during instrument readout on zero air or upscale standard gas.
The Photomultiplier Tube/Reaction Chamber Assembly incorporates several features for exclusion of humidity and moisture. The photomultiplier socket assembly is potted with high impedance silicone rubber compound and is sealed from external influences with epoxy and rubber gasket material. The socket assembly and the reaction chamber are sealed with O-rings into opposite ends of the tubular photomultiplier housing. The socket end of the housing may be sealed with either one or two O-rings, depending on the length of the phototube.
5.3.4 R
EPLACEMENT OF PHOTOMULTIPLIER TUBE
The photomultiplier tube assembly must be removed from the housing in order to replace the tube. To remove, do the following:
1. Note the orientation of the connectors.
2. Slowly rotate and withdraw the socket assembly from the housing. Note the orientation and placement of the metal shield and the black plastic insulating cover.
3. Carefully unplug the photomultiplier tube from the socket.
4. Plug a new tube into the socket.
5. Orient the metal shield and black plastic insulator as noted in step 2.
6. Carefully rotate and insert the tube, shield and cover into the housing. Orient as noted in step 1.
748214-N
September 1997
Rosemount Analytical
5-7
Page 46
M
951C NOX A
ODEL
NALYZER
5.4 OZONE GENERATION SYSTEM
WARNING: TOXIC CHEMICAL HAZARD
The ozone generator lamp contains mercury. Lamp breakage could result in mercury exposure. Mercury is highly toxic if absorbed through skin or ingested, or if vapors are inhaled.
Handle lamp assembly with extreme care. If lamp is broken, avoid skin contact and inhalation in the area of the lamp
or the mercury spill. Immediately clean up and dispose of the mercury spill and lamp residue as
follows: Wearing rubber gloves and goggles, collect all droplets of mercury by
means of a suction pump and aspirator bottle with long capillary tube. Alternatively, a commercially available mercury spill clean-up kit, such as J. T. Baker product No. 4439-01, is recommended.
Carefully sweep any remaining mercury and lamp debris into a dust pan. Carefully transfer all mercury, lamp residue and debris into a plastic bottle which can be tightly capped. Label and return to hazardous material reclamation center.
Do not place in trash, incinerate or flush down sewer. Cover any fine droplets of mercury in non-accessible crevices with calcium
polysulfide and sulfur dust.
This system consists of the ultraviolet lamp, lamp housing, and power supply. Refer to Figure 6-1 for location and details of mounting.
5.4.1 L
AMP/HOUSING REMOVAL
To remove the lamp and housing, do the follow:
1. Disconnect power from the instrument.
2. Release pressure from SAMPLE and AIR supplies.
3. Disconnect the air supply tubing from the front of the housing.
4. Disconnect the ozone tube leading to the reaction chamber.
5. Disconnect the power cable from the Power Supply.
6. Uncouple the two Velcro straps which secure the housing to power supply.
7. Lift the housing from the analyzer.
5-8
Rosemount Analytical
September 1997
748214-N
Page 47
R
OUTINE SERVICING
5.4.2 UV L
AMP REPLACEMENT
To replace the lamp, do the following:
1. Unscrew and remove end cap.
2. Unscrew aluminum outer lamp housing tube from lamp base, using care not to hit or touch lamp assembly.
Note: Do not touch lamp. Fingerprints may cause a decrease in lamp output.
1. Replace O-ring in lamp base with new O-ring supplied in kit.
2. Insert replacement lamp assembly using care not to hit or touch lamp housing.
3. Insert new O-ring into new end cap. Screw end cap onto end of lamp housing.
4. Replace the lamp and housing by reversing the steps in this section.
5.4.3 P
OWER SUPPLY REMOVAL
To remove the Power Supply, do the following: Refer to Figure 6-1.
1. Remove the lamp and housing as in Section 5.4.2.
2. Disconnect the power lead from the Power Supply Board.
3. Remove the two screws which secure the Power Supply to the bottom plate of the analyzer.
4. Lift the Power Supply from the analyzer.
5. Replace the Power Supply by reversing the order of the steps in this section.
5.5 CONVERTER ASSEMBLY
To check the heater blanket, verify the continuity of the heater coil. To check the temperature sensor, refer to Section 3.4 and measure its resistance
when instrument power is off (should be about 440 ohms) and when instrument power is on (should range from 800 to 1,000 ohms). See Table 3-1.
To remove the glass converter tube (see Figure 6-4):
1. Carefully disconnect the blue silicon connectors from the ends of the inlet and outlet tubes.
748214-N
September 1997
Rosemount Analytical
5-9
Page 48
M
951C NOX A
ODEL
NALYZER
The inlet tube is partially filled with glass wool and has a larger inside diameter than the outlet tube. Further, the outlet tube and the sample capillary (P/N
615989) connect to the same stainless steel tee.
2. Release the assembly and disconnect the heater and sensor connectors from the temperature control board.
3. Remove the lacing from the heater blanket, and remove the converter tube. Note the position of the temperature sensor and its leads as the aluminum foil is unwrapped.
4. Replace the defective part and reassemble. The temperature sensor should contact the converter tube with the top of the sensor at the midpoint of the converter. Route sensor leads axially to the outer end.
5. Condition the converter as described in Sections 3.4 and 3.5.
5.6 SERVICING ELECTRONIC CIRCUITRY
For troubleshooting the electronic system, refer to Section 4 and the appropriate pictorial diagrams at the back of the manual. The electronic system utilizes printed circuit boards with solid-state components. After a malfunction is traced to a particular board, the recommended procedure is to return it to the factory for repair.
5-10
Rosemount Analytical
September 1997
748214-N
Page 49
R
EPLACEMENT PARTS
6
The following parts are recommended for routine maintenance and troubleshooting of the Model 951C NOx Analyzer. If the troubleshooting procedures do not resolve the problem, contact your local Rosemount Analytical service office. A list of Rosemount Analytical Service Centers is located in the back of this manual. Figures 6-1 through 6­5 show locations of components and assemblies.
WARNING: PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect safety of this product. Use only factory documented components for repair.
6.1 CIRCUIT BOARD REPLACEMENT POLICY
In most situations involving a malfunction of a circuit board, it is more practical to replace the board than to attempt isolation and replacement of the individual component. The cost of test and replacement will exceed the cost of a rebuilt assembly from the factory.
The following list does not include individual electronic components. If circumstances necessitate replacement of an individual component which can be identified by inspection or from the schematic diagrams, obtain the replacement component from a local source of supply.
6.2 REPLACEMENT PARTS
OMMON PARTS
C
Refer to Figure 6-1.
655519 Air Restrictor Fitting 657091 Capacitor Assembly 655166 Capillary, Bypass 655589 Capillary, Sample Hi 623719 Capillary, Sample Lo 654068 Temperature Control Assembly 654070 Converter Assembly
748214-N
September 1997
Rosemount Analytical
6-1
Page 50
M
951C NOX A
ODEL
NALYZER
655303 Exhaust Fan 654052 Fan Assembly 898587 Fuse 3.15 A 902413 Fuse 6.25 A 654390 I/O Assembly 652173 Ozone Generator 658156 Ozone Generator UV Lamp Replacement Kit 655129 Ozone Generator Power Supply 654062 Photomultiplier Assembly 655332 Power Supply Assembly 654085 Pressure Switch 623936 Sample Flow Restrictor 644055 Sample Pressure Gauge 815187 Sample Regulator * 622917 Sensor, Temperature 654050 Signal Board 654878 Transformer/Inductor Assembly
HOTOMULTIPLIER ASSEMBLY
P
Refer to Figure 6-2.
654943 Housing 649541 Insulating Washer 636318 Magnetic Shield 630916 Magnetic Shield 001522 O-Ring 008423 O-Ring, Photomultiplier 655168 Photomultiplier Tube 654381 Reaction Chamber 654086 Socket Assembly 639722 Thermal Shield
ONVERTER ASSEMBLY
C
654070
Refer to Figure 6-3.
632784 Connector, Blue Silicone 657127 Heater 632782 Temperature Sensor 632795 Tube
654062
EMPERATURE CONTROL ASSEMBLY
T
Refer to Figure 6-4.
6-2
Rosemount Analytical
September 1997
654068
748214-N
Page 51
622733 Fan 622732 Heater 655335 Temperature Control Board 900492 Thermal Fuse
Power Supply Assembly 655332
Case Heater Temperature Control Assembly 654068 (See Figure 6-4)
Photomultiplier Assembly 654062 (See Figure 6-2)
I/O Assembly 654390
Transformer/Inductor Assembly 654878
R
EPLACEMENT PARTS
Sample Flow Restrictor 623936
Exhaust Fan 655303
Fan Assembly 654052
Pressure Switch 654085
IGURE
F
748214-N
6-1. M
Air Restrictor 655519 (on Ozone Generator inlet)
Ozone Generator Power Supply
Sample Regulator 815187
Sample Pressure Gauge 644055
655129
AJOR ASSEMBLIES OF THE MODEL
Ozone Generator 658156
UV Lamp Replacement Kit 658156 Sensor 622917
951C
September 1997
Converter Assembly 654070 (See Figure 6-3)
Rosemount Analytical
6-3
Page 52
M
951C NOX A
ODEL
Magnetic Shield 636318
NALYZER
Socket Assembly 654086
O-Rings 001522
Thermocooler Housing 654943
PVC Tube
Photomultiplier Tube 655168
Insulating Washer 649541
Thermal Shield 639722
O-Ring 008423
Reaction Chamber 654381
Magnetic Shield 630916
Photomultiplier Tube
Magnetic Shield 630916
Photomultiplier Tube and Magnetic Shield (PN 630916) to be Flush
Note: Silicone lubricant to be applied to o-rings.
IGURE
F
6-4
6-2. P
Rosemount Analytical
HOTOMULTIPLIER HOUSING ASSEMBLY
September 1997
.3
PVC Tube
748214-N
Page 53
Connectors, Glass Tube
Temperature Sensor 632782
R
EPLACEMENT PARTS
IGURE
F
6-3. C
Converter Tube - Packed 632795
Heater 657127
ONVERTER ASSEMBLY
748214-N
September 1997
Rosemount Analytical
6-5
Page 54
M
951C NOX A
ODEL
NALYZER
Temperature Control Board 655335
Thermal Fuse 900492
IGURE
F
6-4. C
Heater Cable Orientation
Fan 622733
Heater 633732
ASE HEATER TEMPERATURE CONTROL ASSEMBLY
6-6
Rosemount Analytical
September 1997
748214-N
Page 55
ENERAL PRECAUTIONS FOR HANDLING AND
G
TORING HIGH PRESSURE GAS CYLINDERS
S
Edited from selected paragraphs of the Compressed Gas Association's "Handbook of Compressed Gases" published in 1981 Compressed Gas Association 1235 Jefferson Davis Highway Arlington, Virginia 22202 Used by Permission
1. Never drop cylinders or permit them to strike each other violently.
2. Cylinders may be stored in the open, but in such cases, should be protected against extremes of weather and, to prevent rusting, from the dampness of the ground. Cylinders should be stored in the shade when located in areas where extreme temperatures are prevalent.
3. The valve protection cap should be left on each cylinder until it has been secured against a wall or bench, or placed in a cylinder stand, and is ready to be used.
4. Avoid dragging, rolling, or sliding cylinders, even for a short distance; they should be moved by using a suitable hand-truck.
5. Never tamper with safety devices in valves or cylinders.
6. Do not store full and empty cylinders together. Serious suckback can occur when an empty cylinder is attached to a pressurized system.
7. No part of cylinder should be subjected to a temperature higher than 125°F (52°C). A flame should never be permitted to come in contact with any part of a compressed gas cylinder.
8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding, precautions must be taken to prevent striking an arc against the cylinder.
4125 E
AST LA PALMA AVENUE
Rosemount Analytical Inc.
• A J
ULY
, C
NAHEIM
ALIFORNIA
1997 • 748525-C • P
92807-1802 • 714-986-7600 • FAX 714-577-8006
RINTED IN
USA
Page 56
(blank)
Page 57
ARRANTY
W
Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in workmanship and material under normal use and service for a period of twelve (12) months from the date of shipment by Seller. Consumables, glass electrodes, membranes, liquid junctions, electrolyte, o-rings, etc., are warranted to be free from defects in workmanship and material under normal use and service for a period of ninety (90) days from date of shipment by Seller. Goods, part(s) and consumables proven by Seller to be defective in workmanship and/or material shall be replaced or repaired, free of charge, F.O.B. Seller's factory provided that the goods, part(s) or consumables are returned to Seller's designated factory, transportation charges prepaid, within the twelve (12) month period of warranty in the case of goods and part(s), and in the case of consumables, within the ninety (90) day period of warranty. This warranty shall be in effect for replacement or repaired goods, part(s) and the remaining portion of the ninety (90) day warranty in the case of consumables. A defect in goods, part(s) and consumables of the commercial unit shall not operate to condemn such commercial unit when such goods, part(s) and consumables are capable of being renewed, repaired or replaced.
The Seller shall not be liable to the Buyer, or to any other person, for the loss or damage directly or indirectly, arising from the use of the equipment or goods, from breach of any warranty, or from any other cause. All other warranties, expressed or implied are hereby excluded.
IN CONSIDERATION OF THE HEREIN STATED PURCHASE PRICE OF THE GOODS, SELLER GRANTS ONLY THE ABOVE STATED EXPRESS WARRANTY. NO OTHER W ARRANTIES ARE GRANTED INCLUDING, BUT NOT LIMITED TO, EXPRESS AND IMPLIED WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Limitations of Remedy. SELLER SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY DELAY IN PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF WARRANTY SHALL BE LIMITED TO REPAIR OR REPLACEMENT UNDER THE STANDARD W ARRANTY CLAUSE. IN NO CASE, REGARDLESS OF THE FORM OF THE CAUSE OF ACTION, SHALL SELLER'S LIABILITY EXCEED THE PRICE TO BUYER OF THE SPECIFIC GOODS MANUFACTURED BY SELLER GIVING RISE TO THE CAUSE OF ACTION. BUYER AGREES THAT IN NO EVENT SHALL SELLER'S LIABILITY EXTEND TO INCLUDE INCIDENTAL OR CONSEQUENTIAL DAMAGES. CONSEQUENTIAL DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO, LOSS OF ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVENUE, COST OF CAPITAL AND DAMAGE OR LOSS OF OTHER PROPERTY OR EQUIPMENT. IN NO EVENT SHALL SELLER BE OBLIGATED TO INDEMNIFY BUYER IN ANY MANNER NOR SHALL SELLER BE LIABLE FOR PROPERTY DAMAGE AND/OR THIRD PARTY CLAIMS COVERED BY UMBRELLA INSURANCE AND/OR INDEMNITY COVERAGE PROVIDED TO BUYER, ITS ASSIGNS, AND EACH SUCCESSOR INTEREST TO THE GOODS PROVIDED HEREUNDER.
Force Majeure. Seller shall not be liable for failure to perform due to labor strikes or acts beyond Seller's direct control.
Rosemount Analytical
4125 E
AST LA PALMA AVENUE
Rosemount Analytical Inc.
• A
F
EBRUARY 1997 • 7485189-C • PRINTED IN USA
NAHEIM
, C
ALIFORNIA
92807-1802 • 714-986-7600 • FAX 714-577-8006
Page 58
(blank)
Page 59
IELD SERVICE AND REPAIR FACILITIES
F
Field service and repair facilities are located worldwide.
U.S.A.
To obtain field service on-site or assistance with a service problem, contact (24 hours, 7 days a week):
National Response Center
1-800-654-7768
INTERNATIONAL
Contact your local Rosemount Sales and Service office for service support.
FACTORY
For order administration, replacement Parts, application assistance, on-site or factory repair, service or maintenance contract information, contact:
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
RETURNING PARTS TO THE FACTORY
Before returning parts, contact the Customer Service Center and request a Returned Materials Authorization (RMA) number. Please have the following information when you call:
Model Number, Serial Number, and Purchase Order Number or Sales Order Number.
Prior authorization by the factory must be obtained before returned materials will be accepted. Unauthorized returns will be returned to the sende r, f re ight collect.
When return ing any product or compon ent that has been expo sed to a toxic, co rrosive or other hazardous material or used in such a hazardous environment, the user must attach an appropriate Material Safety Data Sheet (M.S.D.S.) or a written certification that the material has been decontaminated, disinfected and/or detoxified.
Return to:
Rosemount Analytical Inc.
4125 East La Palma Avenue
Anaheim, California 92807-1802
4125 E
AST LA PALMA AVENUE
Rosemount Analytical Inc.
• A J
, C
NAHEIM
ULY 1997 • 748190-G • PRINTED IN USA
ALIFORNIA
92807-1802 • 714-986-7600 • FAX 714-577-8006
Page 60
(blank)
Loading...