Rosemount 751 Operating Manual

Rosemount 751 Field Signal Indicator
Reference Manual
00809-0100-4378, Rev CE
February 2014
Reference Manual
NOTICE
00809-0100-4378, Rev CE
Rosemount 751 Field Signal Indicator
Title Page
February 2014
Read this manual before working with the product. For personal and system safety, and for optimum product performance, make sure you thoroughly understand the contents before installing, using, or maintaining this product.
Within the United States, Rosemount Inc. has two toll-free assistance numbers: Customer Central
Technical support, quoting, and order-related questions. 1-800-999-9307 (7:00 am to 7:00 pm CST) North American Response Center
Equipment service needs. 1-800-654-7768 (24 hours—includes Canada) Outside of the United States, contact your local Emerson Process Management
representative.
The products described in this document are NOT designed for nuclear-qualified applications. Using non-nuclear qualified products in applications that require nuclear-qualified hardware or products may cause inaccurate readings.
For information on Rosemount nuclear-qualified products, contact your local Emerson Process Management Sales Representative.
Reference Manual
00809-0100-4378, Rev CE
1Section 1: Introduction
2Section 2: Installation
Table of Contents
February 2014
1.1 LCD display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
1.2 Analog meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
1.3 Service support. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
1.4 Product recycling/disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
2.1 Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.2 Wiring diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
2.3 LCD display configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
2.3.1 Remove the cover. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
2.3.2 Position the decimal point and select the meter function . . . . . . . . . . . . .11
2.3.3 Store the information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
2.3.4 Set the display equivalent to a 4 mA signal . . . . . . . . . . . . . . . . . . . . . . . . . .12
2.3.5 Set the display equivalent to a 20 mA signal . . . . . . . . . . . . . . . . . . . . . . . . .12
2.3.6 Replace the cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
AAppendix A: Reference data
A.1 Housing specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
A.1.1 Physical specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
A.2 LCD display specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
A.2.1 Functional specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
A.2.2 Performance specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
A.2.3 Physical specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
A.3 Analog meter specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
A.3.1 Functional specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
A.3.2 Performance specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
A.3.3 Physical specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
A.4 Dimensional drawings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
A.5 Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
BAppendix B: Product Certifications
B.1 European directive information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
B.2 Ordinary location certification for FM approvals . . . . . . . . . . . . . . . . . . . . . . . . . . .21
B.2.1 North America . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
B.2.2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Europe22
Table of Contents
B.2.3 International . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
CAppendix C: Approval drawings
C.1 Declaration of conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34
Reference Manual
00809-0100-4378, Rev CE

Section 1 Introduction

LCD display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3
Analog meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4
Service support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5
Product recycling/disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5
The Rosemount 751 Field Signal Indicators provide a means of displaying important process variables. These devices operate with any two-wire transmitter that measures input variables such as pressure, flow, liquid level, or temperature. Rosemount indicators are ideal for installations where an integral meter would be difficult to view.
Rosemount 751 Indicators are designed for use in industrial environments where all-weather performance is necessary. These units are vibration- and corrosion-resistant, and explo­sion-proof or intrinsically safe. An LCD display or analog meter may be ordered to meet specific application requirements.
Section 1: Introduction
February 2014

1.1 LCD display

The LCD display requires an analog 4–20 mA dc output from a two-wire transmitter. It may be configured from a 4 mA point of –999 to 1000 and a 20 mA point of -999 to 9999. The sum of the 4 mA point and the span must not exceed 9999. The decimal point can be placed in any of three positions (X.X.X.X) or not used. Calibration adjustments are made using noninteractive zero and span buttons. The scaled meter may be labelled with the appropriate engineering units. A twenty-segment bar graph, on the bottom of the meter faceplate, represents the 4–20 mA signal directly.
The large 2
Figure 1-1. The 4 and 20 mA points may be changed by pressing the buttons on the meter
faceplate. The meter can be rotated in 90-degree increments within the enclosure for convenient viewing.
Figure 1-1. LCD display
1
/4-inch meter face has 1/2-inch-high characters for easy readability as shown in
LCD display
Introduction
Section 1: Introduction
PERCENT
0
20
40
60
80
100
February 2014

1.2 Analog meter

The analog meter requires an analog 4–20 mA dc, 10–50 mA dc, or 40–200 mV dc transmitter output from a two-wire transmitter. Several meter calibration options are available to suit the requirements of a particular application. Linear 0 to 100 percent meter scaling is adequate for the majority of applications. A logarithmic 0 to 100 percent scale is available for use with flow transmitters. As an option, the user can specify special meter scaling for direct readout in psi, gph, °F, °C, or other convenient engineering units.
Reference Manual
00809-0100-4378, Rev CE
The large 2
1
/4-inch diameter meter face has a two-inch long scale for easy readability as shown in
Figure 1-2. A meter-zero adjustment is located on the meter faceplate. The meter can be
rotated in 90° increments within the enclosure for convenient viewing.
Figure 1-2. Analog Meter
Analog Meter Face
Introduction
Reference Manual
00809-0100-4378, Rev CE

1.3 Service support

To expedite the return process outside of the United States, contact the nearest Emerson Process Management representative.
Within the United States, call the Emerson Process Management Instrument and Valves Response Center using the 1-800-654-RSMT (7768) toll-free number. This center, available 24 hours a day, will assist you with any needed information or materials.
The center will ask for product model and serial numbers, and will provide a Return Material Authorization (RMA) number. The center will also ask for the process material to which the product was last exposed.
Individuals who handle products exposed to a hazardous substance can avoid injury if they are informed of and understand the hazard. If the product being returned was exposed to a hazardous substance as defined by OSHA, a copy of the required Material Safety Data Sheet (MSDS) for each hazardous substance identified must be included with the returned goods.
Section 1: Introduction
February 2014
Emerson Process Management Instrument and Valves Response Center representatives will explain the additional information and procedures necessary to return goods exposed to hazardous substances.

1.4 Product recycling/disposal

Recycling of equipment and packaging should be taken into consideration and disposed of in accordance with local and national legislation/regulations.
Introduction
Reference Manual
A
B
C
D
E
F
G
H
I
J
K
L
M
N
00809-0100-4378, Rev CE

Section 2 Installation

Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 7
Wiring diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 9
LCD display configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 11

2.1 Assembly

The Rosemount 751 Field Signal Indicator is comprised of the components shown in Figure 2-1. The housing may contain an analog or LCD display. Both meters are independent of component parts and are completely interchangeable. Both meters plug into the terminal screws on the housing, as shown in Figure 2-1.
The meter subassembly contains the components shown in Figure 2-2.
Section 2: Installation
February 2014
Figure 2-1. Rosemount 751 Exploded View
A. Terminal Screws B. Housing O-Ring C. Field Wiring Terminals D. Loop Protection Diode E. Tapped Mounting Boss F. Optional Mounting Bracket G. Mounting Bolt with Washer
H. U-Bolt for 2-inch Pipe I. Housing J. Optional ¾ to ½-inch Conduit Reducing Bushing (if required) K. Meter L. Bushing M. Foam Spacer N. Housing Cover
7Installation
Section 2: Installation
A
B
C
D
K
J
I
H
G
F
E
February 2014
Figure 2-2. LCD display exploded view
Reference Manual
00809-0100-4378, Rev CE
A. Retaining Straps B. Mounting Screw into Housing
C. Washer for Retaining Strap D. Mounting Screws into Mounting Plate E. Terminal Screws (2) F. Mounting Plate G. Spacer Plate H. LCD Display I. Bushing J. Foam Spacer K. Configuration Buttons
Installation
Reference Manual
Power Supply
Power Supply
751
4–20 dc Input Signal for
Rosemount 2051
4–20 mA dc Input Signal for
Rosemount 3144P
+
+
Optional Ground
Load Resistor
T
2051
+
+
3144P
Optional Ground
Load Resistor
+
T
751
+
Power Supply
Power Supply
4–20 dc Input Signal for
Rosemount 3051S
4–20 mA dc Input Signal for
Rosemount 3051C
+
+
Optional Ground
Load Resistor
T
3051C
751
+
+
+
Optional Ground
Load Resistor
T
3051S
751
+
00809-0100-4378, Rev CE

2.2 Wiring diagrams

Use the following wiring diagrams to wire the Rosemount 751 Field Signal Indicator, in series or in parallel, with Rosemount transmitters. Use shielded cable for best results in electrically noisy environments.
It is recommended that the 751 indicator be wired in a series configuration when the 4-20 mA transmitter does not contain a test terminal. The 751 is designed so the analog or LCD display can be removed from the housing without impacting the integrity of the 4-20 mA loop. Removal of the entire 751 device from the series configuration will disrupt the loop.
Figure 2-3. Rosemount 751 Series Wiring Diagrams
Series Wiring Diagrams for Rosemount 3144P Temperature
transmitters and Rosemount 2051, 3051C, or 3051S Pressure
Section 2: Installation
February 2014
Tra ns mitt er s
Installation
Loading...
+ 29 hidden pages