1.0 About this guide
This guide provides instructions for installing the Rosemount 5708 3D Solids
Scanner in small vessel connections using the manhole mounting tool kit. It
does not provide instructions for mounting the scanner head, electrical
installation, or configuration.
Refer to the Rosemount 5708 Quick Start Guide
Reference Manual for more instructions. The manual and this guide are also
available electronically on Emerson.com/Rosemount.
Authorized personnel
All operations described in this document must be carried out by authorized, trained personnel only.
For safety and warranty reasons, any internal work on the devices must be carried out by
manufacturer-authorized personnel only.
Warnings about misuse
Inappropriate or incorrect use of the device may result in hazards and application-specific malfunctioning,
such as vessels overfill or damage to system components through incorrect mounting or adjustments.
If the device is used in a manner not specified in this document, the protection provided by the device will
be impaired.
General safety instructions
Consider local and national electrical codes and all common safety regulations and accident prevention
rules during installation.
Substitution of components may impair intrinsic safety.
For preventing ignition of flammable or combustible atmospheres, read, understand and adhere to the
manufacturer’s live maintenance procedures.
and Rosemount 5708
Contents
Install the antenna . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4
Rotate the antenna toward vessel center . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 9
2

June 2017
A
B
C
1x 3x 1x
4.10 (104)
Ø 1.73 (44)
1.1 Manhole mounting tool kit
Figure 1. Manhole Mounting Tool Kit (P/N 05708-3023-0001)
A. Installation adapter
B. Set screw
C. 2 mm hex key
Dimensions are in inches (millimeters).
Quick Start Guide
1.2 Required equipment
Standard tools (e.g. screwdriver, pliers)
60 mm wrench
Wrench to secure the mounting plate bolts
2 mm hex key
Two ropes of appropriate type and length (see page 6)
3

Quick Start Guide
2.0 Install the antenna
Step 1: Verify the O-rings are in place
June 2017
Step 2: Remove the nut from the neck tube
4

June 2017
Quick Start Guide
Step 3: Mount the installation adapter on the antenna
Step 4: Place the mounting plate and nut on the scanner’s
mounting hole
5

Quick Start Guide
Step 5: Pull the rope through to the manhole
Secure one end of the rope and pull the other end through the nut, mounting
plate, and mounting hole to the manhole.
Step 6: Tie the rope to the installation adapter
Note
Consider securing the antenna by tying an additional rope to the adapter from the
manhole opening to prevent the scanner from hitting the side walls.
June 2017
Step 7: Carefully lower the antenna into the manhole
6

June 2017
Quick Start Guide
Step 8: Pull the rope until the neck of the antenna is outside
the mounting hole
Step 9: Hand-tighten the nut
7

Quick Start Guide
Step 10: Remove the installation adapter
June 2017
H2 mm
Step 11: Tighten the mounting plate bolts
8

June 2017
Quick Start Guide
3.0 Rotate the antenna toward vessel center
Step 1: Slightly loosen the nut
Step 2: Rotate the antenna
The notch on top of the thread must be directed toward the vessel center.
9

Quick Start Guide
Step 3: Tighten the nut
Proceed with installation of scanner head. See Rosemount 5708 Quick Start
Guide and Rosemount 5708 Reference Manual for further installation
information.
June 2017
10

Global Headquarters
Emerson Automation Solutions
6021 Innovation Blvd.
Shakopee, MN 55379, USA
+1 800 999 9307 or +1 952 906 8888
+1 952 949 7001
RFQ.RMD-RCC@Emerson.com
North America Regional Office
Emerson Automation Solutions
8200 Market Blvd.
Chanhassen, MN 55317, USA
+1 800 999 9307 or +1 952 906 8888
+1 952 949 7001
RMT-NA.RCCRFQ@Emerson.com
Latin America Regional Office
Emerson Automation Solutions
1300 Concord Terrace, Suite 400
Sunrise, FL 33323, USA
+1 954 846 5030
+1 954 846 5121
RFQ.RMD-RCC@Emerson.com
Europe Regional Office
Emerson Automation Solutions
Neuhofstrasse 19a P.O. Box 1046
CH 6340 Baar
Switzerland
+41 (0) 41 768 6111
+41 (0) 41 768 6300
RFQ.RMD-RCC@Emerson.com
Asia Pacific Regional Office
Emerson Automation Solutions
1 Pandan Crescent
Singapore 128461
+65 6777 8211
+65 6777 0947
Enquiries@AP.Emerson.com
Middle East and Africa Regional Office
Emerson Automation Solutions
Emerson FZE P.O. Box 17033
Jebel Ali Free Zone - South 2
Dubai, United Arab Emi rates
+971 4 8118100
+971 4 8865465
RFQ.RMTMEA@Emerson.com
*00825-0200-4570*
Quick Start Guide
00825-0200-4570, Rev AA
Linkedin.com/company/Emerson-Automation-Solutions
Twitter.com/Rosemount_News
Facebook.com/Rosemount
Youtube.com/us er/RosemountMeasur ement
Google.com/+RosemountMeasurement
Standard Terms and Conditions of Sale can be found on the Ter ms
and Conditions of Sale page.
The Emerson logo is a trademark and service mark of Emerson
Electric Co.
Rosemount and Rosemount logotype are trademarks of Emerson.
All other marks are the property of their respective owners.
© 2017 Emerson. All rights reserved.
June 2017