Rosemount 5708 Operating Manual

Page 1
Rosemount™ 5708 3D Solids Scanner
Reference Manual
00809-0100-4570, Rev DA
November 2017
Page 2
Reference Manual
00809-0100-4570, Rev DA
Section Title
ii
Page 3
Reference Manual
00809-0100-4570, Rev DA

Contents

1Section 2: Contents
2Section 1: Introduction
3Section 2: Overview
Contents
November 2017
1.1 Product recycling/disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Storage and transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
2.1 Theory of operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.2 Applications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.2.1 Wide application range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.2.2 Advantages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.2.3 Models . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.3 Components of the Rosemount 5708 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
4Section 3: Mechanical Installation
3.1 Installation procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.2 Mounting considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.2.1 Location and positioning guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.2.2 Rosemount 5708 orientation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.2.3 Mounting location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.2.4 Site preparations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.2.5 Site and installation information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.3 Physical mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
5Section 4: Electrical Installation
4.1 Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.2 Cable selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.3 Hazardous areas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.4 Wiring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4.5 Different connection methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
4.5.1 4-20 mA connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
4.5.2 RS-485 communication. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
4.5.3 Communication using the Rosemount LinkPro. . . . . . . . . . . . . . . . . . . . . . 31
Contents
4.5.4 TCP/IP communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
i
Page 4
Contents
November 2017
Reference Manual
00809-0100-4570, Rev DA
6Section 5: Configuration - Rosemount 5708L
5.1 On-board configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
5.2 Setting the polling address. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
5.3 Configuring the Rosemount 5708L. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.3.1 Basic Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.3.2 Advanced Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.3.3 False Echo Mapping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
7Section 6: Configuration -
Rosemount 5708V and 5708S
6.1 On-board configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
6.2 Setting the polling address. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
8Section 7: Operation
7.1 Rosemount 3DVision/3DMultiVision overview . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
7.1.1 Server-client architecture . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
7.1.2 Multi-client. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
7.1.3 System requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
7.1.4 Network considerations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
7.2 Installing the Rosemount 3DVision software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
7.2.1 Required information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
7.2.2 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
7.2.3 Starting Rosemount 3DVision Server and Client. . . . . . . . . . . . . . . . . . . . . 51
7.3 First-time running . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
7.3.1 Creating a new project. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
7.3.2 Vessel configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
7.3.3 Advanced parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
7.3.4 Echo curve analysis. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
7.3.5 False echo mapping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
7.3.6 Vessel properties configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
7.4 Software navigation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
7.4.1 Project overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
7.4.2 Main screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
7.4.3 Main menu bar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
7.4.4 Project toolbar. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
7.4.5 Vessel selection behavior . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
7.4.6 Vessel pane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
ii
Contents
Page 5
Reference Manual
00809-0100-4570, Rev DA
Contents
November 2017
7.5 Advanced tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
7.5.1 Communication methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
7.5.2 Advanced parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
7.5.3 Device current simulation settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
7.5.4 Device output current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
7.5.5 Device output settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
7.5.6 Echo Curve Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
7.5.7 False Echo Mapping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
7.5.8 Device activations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
7.5.9 Model upgrade . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
7.5.10Reports Wizard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
7.5.11Event log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
7.5.12User management . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
7.5.13Material configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
7.5.143D visualization. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
7.5.15Multi-server connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
7.5.16Upgrading to Rosemount 3DMultiVision. . . . . . . . . . . . . . . . . . . . . . . . . . 132
7.5.17Center of Gravity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
7.5.18Level by Section Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134
9Section 8: Configuration -
Rosemount 5708 in a System
8.1 System overview and components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
8.2 Rosemount System Controller advantages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
8.3 System requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
8.4 Mounting the Rosemount System Controller. . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
8.5 Advanced commissioning steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142
10Section 9: Service and Troubleshooting
9.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
9.1.1 Preventive maintenance procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
9.1.2 Preventive maintenance frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
9.2 Log views . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
9.3 Vessel Data Collection (VDC). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
Contents
9.4 Removing the Rosemount 5708 head . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150
9.5 Changing the Rosemount 5708 antenna. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
9.6 Diagnostic messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152
9.7 Parameters and alerts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152
iii
Page 6
Contents
November 2017
Reference Manual
00809-0100-4570, Rev DA
9.7.1 Alert types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152
9.7.2 Selecting properties to be displayed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153
9.7.3 Setting alert threshold. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153
9.7.4 General tab. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 154
9.8 Service support. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155
9.8.1 Packaging and shipping of return products/parts . . . . . . . . . . . . . . . . . . 156
AAppendix A: Specifications and Reference Data
A.1 Product Certifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 157
A.2 Ordering Information, Specifications, and Drawings . . . . . . . . . . . . . . . . . . . . . 157
BAppendix B: Connection via Modbus RTU
B.1 Initial settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 159
B.2 Registers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 160
B.3 Communication examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 160
CAppendix C: Connection via RS-485
C.1 Physical layer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 161
C.2 Parameters and units. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 161
DAppendix D: Rosemount LinkPro Operating Instructions
D.1 About this document. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 165
D.1.1 Intended audience . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 165
D.1.2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 165
D.2 Rosemount LinkPro overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 166
D.2.1 Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 166
D.2.2 Area of application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 167
D.2.3 Technical concept. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 167
D.2.4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 167
D.2.5 GPRS modem. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 167
D.3 Technical specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 168
D.3.1 Modem specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 168
D.3.2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Rosemount LinkPro specifications169
D.4 Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170
D.4.1 Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170
D.4.2 Moisture . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170
D.4.3 Wall mounting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170
D.5 Rosemount LinkPro Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 172
iv
Contents
Page 7
Reference Manual
00809-0100-4570, Rev DA
EAppendix E: Accessories
Contents
November 2017
D.5.1 Connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 172
D.5.2 Wiring plan. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173
D.5.3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Modem connection174
D.6 Modem wizard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175
D.6.1 Getting started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175
D.6.2 Test performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175
D.6.3 Configure APN details . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177
D.6.4 Configure mode of operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177
D.7 Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178
E.0.1 Cable extender . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 181
E.0.2 Neck Extension . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182
E.0.3 Package components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183
E.1 Neck extension . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184
E.1.1 Required equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184
E.1.2 Assembling the neck extension . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184
E.2 Angle adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186
E.2.1 Parts of the adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186
E.2.2 Assembling the angle adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187
Contents
v
Page 8
Contents
November 2017
Reference Manual
00809-0100-4570, Rev DA
vi
Contents
Page 9
Reference Manual
NOTICE
00809-0100-4570, Rev DA
Rosemount™ 5708 3D Solids Scanner
Read this manual before working with the product. For personal and system safety, and for optimum product performance, make sure you thoroughly understand the contents before installing, using, or maintaining this product.
Within the United States, Emerson has two toll-free assistance numbers. Customer Central: 1-800-999-9307 (7:00 a.m. to 7:00 p.m. CST) Technical support, quoting, and order-related questions. North-American Response Center: Equipment service needs. 1-800-654-7768 (24 hours—includes Canada) For equipment service or support needs outside the United States, contact your local Emerson
representative.
Title Page
November 2017
Title Page
vi
Page 10
Title Page
November 2017
Reference Manual
00809-0100-4570, Rev DA
Failure to follow safe installation and service guidelines could result in death or serious injury.
Make sure the Rosemount 5708 is installed by qualified personnel and in accordance with applicable
code of practice.
Use the equipment only as specified in this manual. Failure to do so may impair the protection
provided by the equipment.
Inappropriate or incorrect use of the product may result in hazards and application-specific
malfunctioning, such as vessels overfill or damage to system components, through incorrect mounting or adjustments.
Do not perform any services other than those contained in this manual unless you are qualified.Any substitution of non-authorized parts or repair, other than exchanging the complete transmitter
head or antenna assembly, may jeopardize safety and is prohibited.
Unauthorized changes to the product are strictly prohibited as they may unintentionally and
unpredictably alter performance and jeopardize safety. Unauthorized changes that interfere with the integrity of the welds or mounting plates, such as making additional perforations, compromise product integrity and safety. Equipment ratings and certifications are no longer valid on any products that have been damaged or modified without the prior written permission of Emerson. Any continued use of product that has been damaged or modified without prior written authorization is at the customer's sole risk and expense.
Explosions could result in death or serious injury.
Verify that the operating environment of the transmitter is consistent with the appropriate hazardous
locations specifications.
To prevent ignition of flammable or combustible atmospheres, disconnect power before servicing.
Electrical shock can result in death or serious injury.
Avoid contact with the leads and terminals. High voltage that may be present on leads can cause
electrical shock.
Make sure the main power to the Rosemount 5708 is off and the lines to any other external power
source are disconnected or not powered while wiring the transmitter.
vii
The products described in this document are NOT designed for nuclear-qualified applications. Using non-nuclear qualified products in applications that require nuclear-qualified hardware or
products may cause inaccurate readings. For information on Rosemount nuclear-qualified products, contact your local Emerson Sales
Representative. This product is designed to meet FCC and R&TTE requirements for a non-intentional radiator. It does
not require any licensing whatsoever and has no tank restrictions associated with telecommunications issues.
This product complies with part 15 of the FCC rules. Operation is subject to the following two conditions: (1) This product may not cause harmful interference, and (2) this product must accept any interference received, including interference that may cause undesired operation.
Title Page
Page 11
Reference Manual
Procedures and instructions in this manual may require special precautions to ensure the safety of the personnel performing the operations. Information that raises potential safety issues is indicated by a warning symbol. Refer to the safety messages listed at the beginning of each section before performing an operation preceded by this symbol.
An indication of special instructions relevant to installations in hazardous locations.
00809-0100-4570, Rev DA

Section 1 Introduction

Introduction
November 2017
This reference manual provides detailed product related information, installation, setup and operation instructions for the Rosemount and carefully before unpacking and installing the products.
Indication of a potentially dangerous situation, which could result in serious injury to persons and/or damage to the Rosemount 5708.
5708. The manual is designed for trained personnel. Read it entirely

1.1 Product recycling/disposal

Recycling of equipment and packaging should be taken into consideration and disposed of in accordance with local and national legislation/regulations.

1.2 Storage and transport

The Rosemount 5708 is protected by special packaging during transport, and is guaranteed to handle normal loads during transport.
Introduction
1
Page 12
Introduction
November 2017
Reference Manual
00809-0100-4570, Rev DA
2
Introduction
Page 13
Reference Manual
00809-0100-4570, Rev DA

Section 2 Overview

Theory of operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3
Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3
Components of the Rosemount 5708 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4

2.1 Theory of operation

Rosemount™ 5708 delivers accurate volumetric measurement of bulk solids and powders – regardless of material type, product characteristics, storage silo type, size, bin or container, and harshness of the storage environment. The product incorporates unique dust-penetrating technology to achieve an unrivaled degree of process measurement and inventory control.
The Rosemount 5708 include an array of three antennas that generate low frequency acoustic signals and receive echo signals from the contents of the silo, bin, or other container type. Using these antennas, the unit measures not only the time/distance of each echoed signal but also its direction and generate a coordinate of the echo inside the vessel.
Rosemount 5708 Overview
November 2017
The built-in digital signal processor digitally samples and analyses the echoed signals and produces accurate measurements of the level and volume of the stored contents, and can generate a 3D visualization position and form of the material within the container for displaying on remote computer screens.

2.2 Applications

2.2.1 Wide application range
The Rosemount 5708 can measure practically any kind of solid material, stored in a variety of containers, including large open bins, bulk solid storage rooms, stockpiles and warehouses, mapping loads that randomly form over time inside silos, and many other challenging applications that were not possible until now. The Rosemount 5708 can measure ranges of up to 230 ft. (70 m) and generate 3D visualization of the material surface when installed according to the guidelines within this manual.
2.2.2 Advantages
Non-contacting measuring principle, easy to service and maintainMeasures minimum and maximum levelsSuitable for measuring all solid materials (including ones with low dielectric constants)Operates in dusty and moist environmentsProfiles of adhesions of materials to vessel wallsSelf-cleaning antenna3D mapping visualization tool that can be used for assistance with choosing a filling location for even
distribution of material (in vessels with multiple filling points)
Includes three transmitters and three receivers integrally assembled into a single unit
Rosemount 5708 Overview
3
Page 14
Rosemount 5708 Overview
November 2017
2.2.3 Models
The Rosemount 5708 consists of three sub-models: Rosemount 5708L, 5708V, and 5708S. All meet safety approvals outlined by ATEX, FM and others.
Rosemount 5708L
The model determines the average level of the stored contents and average distance from the Rosemount 5708 to the surface of the material. The Rosemount 5708L is suitable for narrow silos.
Rosemount 5708V
The Rosemount 5708V models yield highly accurate readings of level and volume. It is appropriate for vessels of up to 40 ft.ft. (12 m) in diameter and at least twice as high, open bins and stockpiles. It presents the minimum and maximum Level/Distance measurements along with the calculated average and volume.
Rosemount 5708S
The Rosemount 5708S models have the capability of generating a 3D visualization of the stored contents on a remote computer and to be mounted with additional devices in the same vessel. These features are highly useful for mapping build-up loads that form randomly over time and other irregularities and for monitoring inventory in large vessels.
Reference Manual
00809-0100-4570, Rev DA

2.3 Components of the Rosemount 5708

The Rosemount 5708 is divided into two parts:
AntennasHead
Figure 2-1. Components of the Rosemount 5708
A
B
A. Head B. Antennas
4
Rosemount 5708 Overview
Page 15
Reference Manual
00809-0100-4570, Rev DA
The antennas contain:
Three antennas, sealed with a customized tin foil2.04-in. (52 mm) thread to fasten the Rosemount 5708 to a mounting plate and a fastening nutTemperature sensorSealing rubber O-ringOrientation indications
The head contains:
The electronic board One gland and one blind glandGrounding screwFastening screw to the transducers
Rosemount 5708 Overview
November 2017
Rosemount 5708 Overview
5
Page 16
Rosemount 5708 Overview
November 2017
Reference Manual
00809-0100-4570, Rev DA
6
Rosemount 5708 Overview
Page 17
Reference Manual
00809-0100-4570, Rev DA

Section 3 Mechanical Installation

Installation procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 7
Mounting considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 7
Moisture and water condensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 8
Physical mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 17

3.1 Installation procedure

Make sure all required documentations are available: Application Evaluation Form (AEF) and Vessel
Drawings.
Compare locator suggestion or mounting location as suggested by Emerson
Rosemount
Perform manual measurements of the vessel dimensions, and compare results to the drawings. Check 24 Vdc input to all Rosemount 5708. Check RS-485 link between Rosemount 5708 and the other side. Make sure the resistance between D+
and D- is 60 ±5% Ω.
Check 4-20 mA connection to the local PLC. Perform Current Simulation from the Rosemount 5708
connected to the PLC.
Power up the Rosemount 5708 and check polling address.Set polling address. In daisy chain connection make sure that each Rosemount 5708 has a different
polling address and one of the Rosemount 5708 is set with polling address of 00.
In case of daisy chain connection, write the polling addresses on the vessel drawings for all Rosemount
5708.
Verify every Rosemount 5708 for proper functioning after power up. This can be verified by connecting
each device directly to a computer.
Verify sound signals transmission from all Rosemount 5708. Get material status (level/distance/volume) according to the process manager. Get all required information from the process manager in order to complete the Rosemount 5708
configuration.
5708 mounting positions.
Mechanical Installation
November 2017
with the actual

3.2 Mounting considerations

This chapter describes the necessary steps for proper installation of the Rosemount 5708 beginning with important pre-installation considerations such as environmental conditions, correct positioning and orientation, through the mounting and configuration process.
Mechanical Installation
7
Page 18
Mechanical Installation
November 2017
3.2.1 Location and positioning guidelines
Choosing the proper location to the unit should consider every aspect of the vessel and contained materials, including the silo or vessel dimensions, type of material and angle of repose, locations of filling and emptying points, maximum level of material, internal construction and moving parts and any other consideration which may possibly affect the performance.
Note
It is strongly recommended to install the unit according to the Locator PC software results provided by the factory for installation location and positioning. Contact Emerson Customer Support for assistance.
Moisture and water condensation
Use the recommended cable gland and tighten the cable connection. For additional protection against moisture, lead the connection cable downward in front of the cable entry. Rain and condensation water can thus drain off. This applies mainly to mounting in areas where moisture is expected (e.g. by cleaning processes), or on cooled or heated vessels.
Pressure
Reference Manual
00809-0100-4570, Rev DA
The process fitting must be sealed in the case of a low/high pressured vessel. Before usage, verify that the sealing material is resistant to the stored medium. The maximum allowed pressure is indicated on the type label of the sensor.
Installation location
Choosing the proper installation location for the unit is an important part of the installation process. The wrong mounting location may result in erroneous measurements or loss of performance.
The following factors must be taken into consideration while choosing the installation position:
Vessel dimensionsFilling and emptying point locationsInternal structure or support
Other restrictions related to the proximity of noisy devices (such as electrical motor) and any other element which may affect the proper operation of the Rosemount 5708.
Note
When mounting the Rosemount 5708, do not install it near the vessel wall. Installing it near the side wall will result in bad performance and will not be supported. The Rosemount 5708 cannot be mounted at a distance less than 24-in. (600 mm) from the wall or filling points.
Installing at the center of the vessel is not recommended, since the perfect symmetry from all sides toward the Rosemount 5708 may affect the resolution of the echo.
3.2.2 Rosemount 5708 orientation
Mounting direction
Mounting at a specific direction is important. The ridge on the antenna and the notch on the top of the thread (representing antenna no.1) should be directed toward the center of the vessel.
8
Mechanical Installation
Page 19
Reference Manual
00809-0100-4570, Rev DA
Figure 3-1. Mounting Direction 0° Indication toward Vessel Center
Mechanical Installation
November 2017
Figure 3-2. The 0° Indication
Mechanical Installation
Standpipe mounting
When mounting using a standpipe it should be assembled and positioned at a height that leaves at least
0.4-in. (10 mm) below the standpipe, for the antenna end protrudes, as shown in Figure 3-3.
9
Page 20
Mechanical Installation
0.4-in. (10 mm)
November 2017
Figure 3-3. Standpipe Mounting
Note
Any obstruction as well as rails, frames or support beams should never interfere with the acoustic beam transmitted and received, as shown in Figure 3-4.
Reference Manual
00809-0100-4570, Rev DA
Figure 3-4. Avoid Interference with the Acoustic Beam
Inflowing material
Do not mount the Rosemount 5708 in or above the filling stream, or close to the stream, or in the filling stream. The Rosemount 5708 should be located with a clear line of sight to the top of the material at high levels, where it will not be affected by the filling stream or the noise it creates.
10
Mechanical Installation
Page 21
Reference Manual
Minimum
24-in. (600 mm)
90°
00809-0100-4570, Rev DA
Figure 3-5. Inflowing Material
When choosing the installation location, consider the filling and emptying points and maintain at least 24-in. (600 mm) from the filling point.
Fitting
Mechanical Installation
November 2017
The fitting area should be prepared to ensure the antennas are positioned perpendicular to the ground, as shown in Figure 3-6.
Figure 3-6. Keep the Antennas in Vertical Position to the Ground
Mechanical Installation
11
Page 22
Mechanical Installation
90°
Minimum
24-in. (600 mm)
Minimum
24-in. (600 mm)
November 2017
3.2.3 Mounting location
Mount Rosemount 5708 perpendicular to the ground.
Keep necessary distance from side wall.
Reference Manual
00809-0100-4570, Rev DA
12
Keep necessary distance from filling points.
Mechanical Installation
Page 23
Reference Manual
Minimum 0.4-in. (10 mm)
for standpipe mounting
00809-0100-4570, Rev DA
Make sure there is no obstacle below the Rosemount 5708.
In the case of standpipe mounting, assemble and position the Rosemount 5708 at a height that leaves
at least 0.4-in. (10 mm) below the standpipe for the antenna end to protrude.
Mechanical Installation
November 2017
3.2.4 Site preparations
Prior to installation, complete and verify the site preparations described in this section. For optimal installation, ensure the Rosemount 5708 can be positioned and fitted according to the guidelines in
“Mounting location” on page 12.
Mechanical Installation
13
Page 24
Mechanical Installation
November 2017
Recommended tools for installation:
A set of small precision screwdrivers (for the terminal blocks)13 mm open wrench4 mm Hex key (preferably with a handle)Large adjustable wrenchWire cutter, pointed pliers, insulating tapeLaser measurement device or equivalentRS-485 to USB converter, including drivers120 Ω (RS-485) resistor 250 Ω (HARTPC or laptopDC voltmeter and Ohm meter
Complete the following steps before installing the Rosemount 5708.
Power
Make sure grounding is done properly. Connect one end of the cable shield to the power ground. It is
highly recommended to have the same potential grounding to all the Rosemount 5708.
Prepare a 24 Vdc power supply near the Rosemount 5708 mounting location.Make sure that you use proper cables for wiring. The Rosemount 5708 is a 4-wire device. The voltage
supply and data output (4-20 mA) are carried along two separate two-wire connection cables.
®
) resistor
Reference Manual
00809-0100-4570, Rev DA
Communications
For RS-485 communication, use shielded, twisted-pair cables with 120 Ω impedance. Make sure the
cables are approved for RS-485 communications.
Route communication cables in proper conduits. Use a proper cable type.For 4-20 mA communication, use shielded, twisted-pair, low resistance cables. Make sure the cables
are rated for analog signals.
For daisy chaining, a single 4-wire cable can be used, both for RS-485 and the
24 Vdc power supply.
3.2.5 Site and installation information
The information listed on these pages is necessary for configuration of the device.
Material characteristics
Material name:
Material density:
Max. temperature: °F °C Maximum pressure: Bar PSI
lbs/ft
3
tons/m3Angle of repose:
14
Mechanical Installation
Page 25
Reference Manual
B
A
Y
X
A
A
A
B
Y
X
A
A
B
A
B
Y
X
A
Y
X
A
00809-0100-4570, Rev DA
Vessel type and dimensions
Mechanical Installation
November 2017
Vessel details
(1)
ft m
Vessel type Cylindrical Rectangular
Flat
Cone
Top diameter (A): ______ Height (B): ______
Dome
Height (A): ______
Flat
Pyramid
Height (A): ______ X: ______ Y: ______
Top s hape
Cyl in der
Diameter (A): ______ Height (B): ______
Cube
Height (A): ______ X: ______ Y: ______
Center shape
Flat
Cone
Height (A): ______ Bottom diameter (B): ______
Pyramid
Height (A): ______ X: ______ Y: ______
Flat
Cone
Height (A): ______ Bottom diameter (B): ______
Pyramid
Height (A): ______ X: ______ Y: ______
Bottom shape
1. Required Field - Accurate Vessel dimensions are required to allow optimal Rosemount 5708 location determination.
Mechanical Installation
15
Page 26
Mechanical Installation
+Y
-Y
+X
-X
BA
November 2017
Vessel details
Internal structure like: Ladder, pipe, window, screw, door, rail, support beam or any other obstacle which is visible to the 5708 scanner. Drawings should be available.
Reference Manual
00809-0100-4570, Rev DA
Internal movement:
YES
NO If yes, please describe:
Rosemount 5708 and filling locations
X Y Offset from roof
1st Rosemount 5708 location:
2nd Rosemount 5708 location:
3rd Rosemount 5708 location:
Filling location:
Figure 3-7. Filling Port Locations
A. Rosemount 5708 (X1, Y1) B. Fill (X2,Y2)
When mounting more than one Rosemount 5708 on one silo the location of every device must be
specified.
When the application has more than a single filling point, all other filling points must be specified as
well.
16
Mechanical Installation
Page 27
Reference Manual
Full
Calibration
Empty Calibration
Total Vessel Height
Empty Calibration
Full
Calibration
Total Vessel Height
Reference Height
Reference Height
Dead Zone
00809-0100-4570, Rev DA
Full and empty calibration
Full and empty calibration levels are measured from the top of the vessel.The full and empty calibration levels represent the 20 mA (100%) and 4 mA (0%) of the volume
respectably.
The Rosemount 5708 has a 20-in. (0.5 m) dead zone starting from the top part of the antenna
assembly.
Figure 3-8. Full and Empty Calibration in Rectangular and Cylindrical Vessels
Mechanical Installation
November 2017
Full calibration: ft m
Empty calibration: ft m
Application process
Maximum filling rate: lbs/hour tons/hour
Maximum emptying rate: lbs/hour tons/hour
Total capacity when vessel is full: lbs tons

3.3 Physical mounting

Step 1: Check power and cables
1. Check the 24 Vdc with a voltmeter.
2. Check the resistance of the data communication lines.
3. Verify 60 Ω of resistance when connecting the 120 Ω resistors at both ends.
Mechanical Installation
17
Page 28
Mechanical Installation
Note
Make sure the hole diameter (D) in the center is 2.05-in. (52 mm).
D
November 2017
Step 2: Install the mounting plate
1. Verify the O-rings on the neck tube are in place.
2. Remove the nut from the neck tube.
Reference Manual
00809-0100-4570, Rev DA
3. Place the mounting plate over the neck tube.
18
Mechanical Installation
Page 29
Reference Manual
00809-0100-4570, Rev DA
4. Replace the nut and tighten it over the neck tube to the mounting plate.
5. Lower the antenna and mounting plate into the fitting location on the silo.
6. Bolt the mounting plate onto the flange of the silo.
Step 3: Rotate antenna toward silo center
Mechanical Installation
November 2017
1. Slightly loosen the nut that connects the antenna with the mounting plate.
2. Rotate the antenna. The notch on the top of the thread must be directed toward the center of the
silo.
Mechanical Installation
3. Tighten nut.
19
Page 30
Mechanical Installation
4x
A
November 2017
Step 4: Install head
1. Remove the rear panel.
A. Hex key (4 mm)
2. Remove the cable clamp.
Reference Manual
00809-0100-4570, Rev DA
20
3. Verify the presence of the O-ring on the tube neck.
Mechanical Installation
Page 31
Reference Manual
A
A. Antenna cable
00809-0100-4570, Rev DA
4. Gently insert the antenna cable through the Rosemount 5708 head.
Mechanical Installation
November 2017
5. Insert the Rosemount 5708 head onto the neck tube. a. Rotate the Rosemount 5708 head to the desired direction.The head may be installed in six
different positions. To avoid direction confusion, we can recommend to install the head in the same direction as the antennas.
b. Make sure to push the head all the way down until it fully contacts the top of the neck tube.
Mechanical Installation
21
Page 32
Mechanical Installation
A
B
A. Hex key (4 mm) B. Open wrench (13 mm)
November 2017
6. Tighten the front screw.
Reference Manual
00809-0100-4570, Rev DA
7. Re-mount the cable clamp.
8. Gently connect the antenna cable connector. Make sure the latch is clicked and locking the connector.
22
Mechanical Installation
Page 33
Reference Manual
00809-0100-4570, Rev DA

Section 4 Electrical Installation

Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 23
Cable selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 23
Hazardous areas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 23
Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 24
Different connection methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 29

4.1 Grounding

The Rosemount™ 5708 must be grounded electrostatically:
For internal grounding, use the power cable ground.For external grounding, use the earth potential equalization of the plant.
Electrical Installation
November 2017

4.2 Cable selection

Make sure to select the proper cable for each application.
For RS-485 communication, use shielded, twisted-pair cables with 120Ω impedance. Make sure the
cables are approved for RS-485 communications.
For 4-20 mA communication, use a low resistance cable and make sure the cables are rated for analog
signals.
For daisy chaining, a single 4-wire cable can be used, both for RS-485 and the
24 Vdc power supply. Make sure only to use a proper cable gaging (AWG).
Use 8-13 mm (20-24 AWG) diameter cables to ensure proper and effective sealing of the cable gland
entry opening.
Select a cable suitable for application (indoor or outdoor) and safety certified according to national
regulations.

4.3 Hazardous areas

In hazardous areas you should take note of the appropriate regulations, conformity and type of approval certificates of the sensors and power supply units. For further information on regulations, see Appendix A: Specifications and Reference Data. For more information on sensors
and power supply units, see “Wiring” on page 24.
Electrical Installation
23
Page 34
Electrical Installation
November 2017

4.4 Wiring

Reference Manual
00809-0100-4570, Rev DA
The units can be connected in different modes and configurations for different external systems such as PLC or DCS and communications on RS-485, Modbus to communication gateways such as Rosemount LinkPro for GMS or GPRS data relay and TCP/IP gateway. For in-depth details and explanations on wiring and communication, refer to “Different connection
methods” on page 29.
®
RTU, and also to RS-485 bus converting adapters
Communications
Note
If electromagnetic interference is expected, usage of a screened and twisted wired cable is recommended for the signal lines, which should be connected to the ground reference.
Always observe the following safety instructions:
Connections must be made only in the complete absence of line voltage.If over-voltage is expected, overvoltage arresters should be installed.Use only a safety-certified power supply with dual insulation between the primary and output for
powering the unit. The power supply output rating must be limited to 18-32 Vdc, 1A for a single Rosemount 5708, and not to be connected to a DC distribution network.
Power supply
For detailed information on power supply specifications, refer to Rosemount 5708 Product Data Sheet. 4-20 mA 4-wire: The power supply and signal current inputs must be carried over two separated pairs.
Connection procedure
For detailed information on the connection procedure, see “Different connection methods” on page 29.
Note
Use direct connection between the Rosemount 5708 and the plant (PLC/SCADA). For a remote connection, use the Rosemount LinkPro remote connection module or a local PC/laptop computer as detailed in the following drawing.
24
Electrical Installation
Page 35
Reference Manual
00809-0100-4570, Rev DA
Always observe the following safety instructions:
Connections must be made only in the complete absence of line voltage.If over-voltage is expected, overvoltage arresters should be installed.Use only a safety-certified power supply with dual insulation between the primary and output for
powering the unit. The power supply output rating must be limited to 18-32 Vdc, 1 A for a single Rosemount 5708, and not to be connected to a DC distribution network.
The 4-20mA lines should NOT be connected in multi-drop.
Using the Rosemount 3DVision to communicate with the units allow the user to choose one of several communication types: RS-485, GSM, GPRS or TCP/IP, for more information and details regarding the communications, refer to the “Communication methods” on page 91.
Wire the Rosemount 5708
Electrical Installation
November 2017
1. Verify the power supply is disconnected.
2. Loosen the compression nut of the cable gland entry.
3. Insert the cable through the cable gland entry.
Electrical Installation
25
Page 36
Electrical Installation
4-in. (100 mm)
0.4-in.
(10 mm)
November 2017
4. Remove approximately 4-in. (100 mm) of the cable mantle and strip approximately 0.4-in. (10 mm) off the edge of each conductor.
5. Loosen the terminal block screws located inside the Rosemount 5708 head.
Reference Manual
00809-0100-4570, Rev DA
Note
The two terminal blocks can be pulled out by unfastening the two side screws to allow for easier wiring into the terminals. When inserting the terminal blocks back into the head, it is advised to fasten the screws back into the electronics card in order to avoid a loose connection and or operational problems.
6. Connect the cable wires according to the wiring diagram. See “Different connection methods” on
page 29 for different connection methods.
Note
When connecting the last Rosemount 5708 in the chain, a 120 Ohm resistor must be connected as well. See Figure 4-6 on page 31. In any installation of the 5708 whether it is a single unit on a silo or multiple units, it is mandatory to use termination resistors at the ends of the daisy-chain connection. One resistor must be on the last scanner on the chain (connected to the +/- terminals of the RS485 and the other resistor on the other end of the line before converting the RS485 bus to a different communication protocol such as USB, TCP/IP and others.
26
Electrical Installation
Page 37
Reference Manual
­+
C
B
­+
A
+
-
00809-0100-4570, Rev DA
Figure 4-1. Wiring
Electrical Installation
November 2017
A. Feed in power supply (18-32 Vdc) B. 4-20 mA communication terminals C. RS-485/Modbus RTU communication terminals
7. Connect the external ground terminal.
The Rosemount 5708 must be grounded electrostatically.
For internal grounding, use the power cable ground. For external grounding, use the earth potential equalization of the plant.
Note
An internal cable grounding connection is also possible using the inner connection as shown below.
Electrical Installation
27
Page 38
Electrical Installation
A
B
November 2017
Figure 4-2. Grounds Screws
A. Internal ground screw B. External ground screw
Reference Manual
00809-0100-4570, Rev DA
8. Tighten the compression nut over the cable gland entry opening.
9. Attach the rear panel back to position and tighten the screws.
4x
28
Electrical Installation
Page 39
Reference Manual
3
4
D
A
B
C
00809-0100-4570, Rev DA

4.5 Different connection methods

4.5.1 4-20 mA connection
The Rosemount 5708 outputs the % of Volume as set in the configuration and between the Full and Empty calibration levels. The 4-20mA current output is available through ports 3 and 4 of the left green terminal block (as shown in Figure 4-3). Ports 3 and 4 are the negative and positive poles, respectively.
Figure 4-3. Port 3 and 4
Electrical Installation
November 2017
The 4-20mA line is connected directly from the Rosemount 5708 mounted on the vessel to the PLC/DCS/Display or any other device (as shown in Figure 4-4).
Figure 4-4. Rosemount 5708 Connected to PLC
A. Active B. 4-20 mA C. Passive D. PLC/Controller
Note
The maximum cable length of a 4-20 mA cable can be up to 2 miles (3 km).
This type of connection is active and not passive, hence the Rosemount 5708 is the active module and the PLC should be the passive module.
Electrical Installation
29
Page 40
Electrical Installation
November 2017
Figure 4-5. Wiring the 4-20mA to Emerson™ Wireless THUM™ Adapter
Reference Manual
00809-0100-4570, Rev DA
A
B
C D E F G
A. Rosemount 5708 housing ground shield B. THUM Adapter C. Green D. Yellow E. White F. Black G. Red
250 Ω
4.5.2 RS-485 communication
The unit includes an RS-485 communication port. This type of connection allows a PLC/DCS/laptop to communicate with the Rosemount 5708. It also allows a connection of multiple Rosemount 5708 in a system on the same RS-485 bus and communication with all devices using a single connection to the computer running the software.
It is required to use an appropriate cable rated for RS-485, the cable should be of twisted pair, have 120 Ω impedance and shielded. The total length of the cable should not reach 3280 ft (1000 m).
If multiple units of Rosemount 5708 is connected together, all devices must be connected in parallel mode in the RS-485 ports. Hence, all the '+' (positive) ports of the RS-485 should commonly be connected and all the '-' (negative) ports of the RS-485 should be connected commonly. The connection must be of Daisy-Chain type and have at each of the far ends of the chain 120 Ω resistor (such resistor is provided with the Rosemount 5708), as shown in Figure 4-6.
24DC
30
The communication cable must be grounded at one side. This ground needs to be the same grounding as the power supply ground.
Each Rosemount 5708 must be configured with a different polling address.
Electrical Installation
Page 41
Reference Manual
RS-485 (+)
120 Ω
RS-485 (-)
RS-485 (+)
120 Ω
RS-485 (-)
Rosemo unt
5708 2
Rosemo unt
5708 1
Rosemount
5708 3
Rosemo unt
5708 n
00809-0100-4570, Rev DA
Figure 4-6. Rosemount 5708 Connection
4.5.3 Communication using the Rosemount LinkPro
In all installations, the Rosemount 5708 can be connected to a GSM/GPRS modem in order to transfer the data over the cellular network. The Rosemount LinkPro should be connected on the RS-485 bus just like any Rosemount 5708, including cable and resistors as needed. For further details on wiring and establishing a connection, refer to the Appendix D: Rosemount LinkPro Operating Instructions.
Electrical Installation
November 2017
Note
In such communication mode, only one computer running the Rosemount 3DVision software can be connected to the Rosemount 5708.
4.5.4 TCP/IP communication
The RS-485 bus can be converted to TCP/IP communication. TCP/IP to RS-485 converter installation should be done as with any Rosemount 5708, including the resistors and daisy-chain considerations.
For more information on how to convert and transfer the RS-485 data over TCP/IP, see “TCP/IP” on
page 92.
Note
Consult the Emerson Technical Support team for assistance on the proper TCP/IP converter to use.
Electrical Installation
31
Page 42
Electrical Installation
November 2017
Reference Manual
00809-0100-4570, Rev DA
32
Electrical Installation
Page 43
Reference Manual
Navigates back within a function menu. Continuous 3 second press exits to the default screen.
Navigates upwards in the navigation list. Navigates right within a function.
Navigates downwards in the navigation list. Navigates left within a function.
Navigates to the right when within a function group. Stores a value once configured.
ESC
E
00809-0100-4570, Rev DA
Configuration - Rosemount 5708L

Section 5 Configuration - Rosemount 5708L

On-board configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 33
Setting the polling address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 34
Configuring the Rosemount 5708L . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 35

5.1 On-board configuration

The Rosemount™ 5708L can be completely configured via the LCD display. It is also possible to configure the Rosemount 5708L by using the Rosemount 3DVision software.
Figure 5-1. User Interface
Configuration - Rosemount 5708L
1. Connect power to the Rosemount 5708 and open the front cover.
The Rosemount 5708 initializes a self-test for about 30 seconds. The display remains blank during this time.
33
Page 44
Configuration - Rosemount 5708L
Rosemount 5708LNN Initialization
Initialization Please Wait...
m ft
<tag name>
3.45m Avg Dist.
█████████████▒▒▒▒▒
Menu
<tag name>
3.45m Avg Dist.
█████████████▒▒▒▒▒
Menu
Basic Settings Advanced Settings False Echo Mapping Polling Address
Polling Address
00
E
November 2017
Reference Manual
00809-0100-4570, Rev DA
2. After initialization, the version screen appears.
According to the factory default settings, after power initiation or restart, a screen prompts for configuration.
In the main menu, select Basic Settings.
3. Use the / keys to switch between the options. Press to select and continue with the settings or to exit to
E
the main screen.
4. When the startup process is complete, the following screen appears showing the current average distance measurement. The top line displays the Rosemount 5708 tag name. By default, this line is empty. Press to enter the main menu.
E
ESC

5.2 Setting the polling address

Setting the sensor address is mandatory when multiple sensors are connected over an RS-485­Multidrop (daisy chain). Set the addresses prior to use of the Rosemount 3DVision software.
1. From the main screen, select the key to enter the main menu.
2. In the main menu, use the / keys to scroll down to Polling Address.
Select to switch to the Polling Address configuration screen.
E
3. Use the key to switch between the two digits. Use the key to modify the value. The default polling address is 00. The polling address ranges from 00 to 63. Select to store the modified address and to exit to the main screen.
E
ESC
34
Configuration - Rosemount 5708L
Page 45
Reference Manual
<tag name>
3.45m Avg Dist.
█████████████▒▒▒▒▒
Menu
Basic Settings Advanced Settings False Echo Mapping Polling Address
m ft
Set Vessel Height
2
0.000 m
Cylindrical Vessel Rectangular Vessel
Set Vessel Diameter
1
0.000 m
E
Set Scanner Height
2
0.000 m
E
Configuration - Rosemount 5708L
00809-0100-4570, Rev DA

5.3 Configuring the Rosemount 5708L

5.3.1 Basic Settings
1. From the main screen, select the key to enter the main menu.
2. In the main menu, scroll down to Basic Settings, using the key.
Select to switch to the Basic Settings configuration screen.
E
3. Set the distance units, either meters (m) or feet (ft).
4. Set Vessel Height from the vessel bottom edge to the vessel top edge.
Use the key to switch between the two digits. Use the key to modify the value.
E
Configuration - Rosemount 5708L
5. Select the vessel shape, either Cylindrical or Rectangular.
a. Use the / keys to switch between the options. Select
to continue.
b. If Cylindrical is selected, set vessel diameter.
Use the key to switch between the two digits. Use the key to modify the value. Select to continue.
c. Set Scanner Height from the bottom edge of the vessel to the
mounting plate (point above the antenna). Use the key to switch between the two digits. Use the key to modify the value. Select to continue.
E
35
Page 46
Configuration - Rosemount 5708L
Scanner Distance
From Center 0
0.000 m
E
Set Vessel Width
0
10.000 m
E
Set Vessel Length
0
10.000 m
E
Set Scanner Height
2
0.000 m
E
Scanner X To Center
±000.00 m
E
Scanner Y To Center
±000.00 m
E
<tag name>
3.45m Avg Dist.
█████████████▒▒▒▒▒
Menu
November 2017
Reference Manual
00809-0100-4570, Rev DA
d. Set Scanner Distance from Center.
Use the key to switch between the two digits. Use the key to modify the value. Select to exit to the main screen.
6. If Rectangular is selected, set vessel width first (dimension on the x-axis).
a. Use the key to switch between the two digits. Use the
key to modify the value. Select to continue.
b. Set Vessel Length (dimension on the y-axis).
Use the key to switch between the two digits. Use the key to modify the value. Select to continue.
5.3.2 Advanced Settings
After configuring the Basic Settings, perform Advanced Settings configuration.
c. Set Scanner Height from the bottom edge of the vessel to the
mounting plate (point above the antenna). Use the key to switch between the two digits. Use the key to modify the value. Select to continue.
d. Set the distance from the x-axis, see Figure 3-7 on page 16.
Use the key to switch between the two digits. Use the key to modify the value. Select to continue.
e. Set the distance from the y-axis, see Figure 3-7 on page 16.
Use the key to switch between the two digits. Use the key to modify the value. Select to exit to the main screen.
36
1. From the main screen, select the key to enter the main menu.
E
Configuration - Rosemount 5708L
Page 47
Reference Manual
Basic Settings Advanced Settings False Echo Mapping Polling Address
Set Distance
To Full Calibration 0
0.500 m
Set Distance
To Empty Calibration 2
0.000 m
Set adapter Angle
0
0
Slow Process Regular Process Fast Process
Celsius Fahrenheit
Distance Level Volume Analog Output SNR
E
00809-0100-4570, Rev DA
Configuration - Rosemount 5708L
2. In the main menu, scroll down to Advanced Settings, using the key.
3. Select to switch to the Advanced Settings configuration
4. Set the distance from the mounting plate (top of the antenna) to
5. Set the distance from the mounting plate (top of the antenna) to
6. Select to keep the default setting.
E
screen.
the Full calibration point / 100% level point. Use the key to switch between the two digits. Use the key to modify the value. Select to continue.
the Empty calibration point / 0% level point. Use the key to switch between the two digits. Use the key to modify the value. Select to continue.
E
Note: if using an angle adapter, set the angle value. Use the key to switch between the two digits. Use the key to modify the value. Select to continue.
E
E
E
7. Set the process rate using the key. Select to continue.
Make sure to always use Regular Process.
For the Slow and Fast Process options, contact your local customer service.
8. Set the temperature units using the key. Select to continue.
9. Set the displayed parameter desired for the main screen using the
key. Select to exit to the main screen.
E
E
Configuration - Rosemount 5708L
37
Page 48
Configuration - Rosemount 5708L
<tag name>
3.45m Avg Dist.
█████████████▒▒▒▒▒
Menu
Basic Settings Advanced Settings False Echo Mapping Polling Address
Map False Echoes Reset Mapping
Distance To Map
False Echo
0
0.000 m
False Echo Area
E
Decline Mapping Approve Mapping
Decline Mapping Approve Mapping
E
November 2017
5.3.3 False Echo Mapping
When the configuration is finished, set False Echo Mapping.
Reference Manual
00809-0100-4570, Rev DA
1. From the main screen, select the key to enter the main menu.
2. In the main menu, scroll down to False Echo Mapping, using the
key.
Select to switch to the False Echo Mapping configuration
E
screen.
3. Select Map False Echoes for automatically mapping all the false echoes up to a certain distance. Or select Reset Mapping for deleting the stored mapped false echoes from the memory.
Use the key to go down the list. Select to continue.
a. If Map False Echoes is selected, set the distance from the top
of the antenna assembly to the end of the scanning point. Always make sure to map false echoes above material level. The recommended level is 3 ft. (1 m) above actual material level.
E
E
Use the key to switch between the two digits. Use the key to modify the value. Select to continue.
b. Approve or decline the false echo mapping operation.
Use the key to go down the list. Select to continue.
c. If Reset Mapping is selected, approve or decline the
operation. Use the key to go down the list. Select to continue.
E
38
Configuration - Rosemount 5708L
Page 49
Reference Manual
Navigates back within a function menu. Continuous 3 second press exits to the default screen.
Navigates upwards in the navigation list. Navigates right within a function.
Navigates downwards in the navigation list. Navigates left within a function.
Navigates to the right when within a function group. Stores a value once configured.
ESC
E
00809-0100-4570, Rev DA
Configuration - Rosemount 5708V and 5708S
Section 6 Configuration -
Rosemount 5708V and 5708S
On-board configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 39
Setting the polling address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 40

6.1 On-board configuration

For the Rosemount™ 5708V and 5708S only polling address configuration is done using the LCD display of the device.
The rest of the configuration is completed with the Rosemount 3DVision software. For further information, see Section 7: Operation.
November 2017
Figure 6-1. User Interface
1. Connect power to the Rosemount 5708 and open the front cover.
A self-test will start and run for about 30 seconds. The display remains blank during this time.
Configuration - Rosemount 5708V and 5708S
39
Page 50
Configuration - Rosemount 5708V and 5708S
Rosemount 5708SEV Initialization
Initialization Please Wait...
m ft
<tag name>
3.45m Avg Dist.
█████████████▒▒▒▒▒
Menu
<tag name>
3.45m Avg Dist.
█████████████▒▒▒▒▒
Menu
Basic Settings Advanced Settings False Echo Mapping Polling Address
Polling Address
0
0
E
November 2017
Reference Manual
00809-0100-4570, Rev DA
2. After initialization, the version screen appears.
According to the factory default settings, after power initiation or restart, a screen prompts for configuration.
In the main menu, select Basic Settings.
3. Use the / keys to switch between the options. Press to select and continue with the settings or to exit to
E
the main screen.
4. When the startup process is complete, the following screen appears showing the current average distance measurement. The top line displays the tag name. By default, this line is empty. Select to enter the main menu.
E
ESC

6.2 Setting the polling address

Setting the sensor address is mandatory when multiple sensors are connected over an RS-485­Multidrop (daisy chain). Set the addresses prior to use of the Rosemount 3DVision software.
1. From the main screen, select the key to enter the main menu.
2. In the main menu, use the / keys to scroll down to Polling Address.
Select to switch to the Polling Address configuration screen.
E
3. Use the key to switch between the two digits. Use the key to modify the value. The default polling address is 00. The polling address ranges from 00 to 63. Select to store the modified address and to exit to the main screen.
E
ESC
40
Configuration - Rosemount 5708V and 5708S
Page 51
Reference Manual
00809-0100-4570, Rev DA

Section 7 Operation

Rosemount 3DVision/3DMultiVision overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 41
Installing the Rosemount 3DVision software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 45
First-time running . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 54
Software navigation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 66
Advanced tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 91

7.1 Rosemount 3DVision/3DMultiVision overview

Rosemount™ 3DVision/3DMultiVision™ enables real-time assessment of bulk solids and powders, of any type or size, stored in storage vessels. The application periodically supplies information, providing the essential, real-time data required for critical decisions necessary for successful plant management.
Figure 7-1. Rosemount 3DMultiVision Output Example
Operation
November 2017
Operation
The Rosemount 3DVision works in conjunction with Rosemount 5708 providing measurement data to a Windows
monitored vessel and generates reports. The reports are used to increase operational efficiency, and provide the site managers with valuable real-time and historical information for effective decision making.
based computer running the programs. The program continuously collects data from the
41
Page 52
Operation
November 2017
Rosemount 3DMultiVision enables usage of a function on groups of vessels, even across multiple sites. Rosemount 3DMultiVision works in conjunction with the Rosemount 5708 to collect data from all monitored vessels in all sites and geographic regions, to combine and merge the data, and to generate and analyze various types of reports.
Basically, both 3DVision and 3DMultiVision have the same capabilities, the one major difference is that the 3DMultiVision allows the user to view measurements from all vessels in one screen while in the 3DVision, it is possible to view only one vessel at a time.
7.1.1 Server-client architecture
The Rosemount 3DVision/3DMultiVision software is comprised of two components: a server and a client.
The Rosemount 3DVision Ser ver runs on a designated computer and is connected directly or remotely to every Rosemount 5708 in a single, local site or in multiple remote sites. The data is stored on the server computer which generates the reports, collects and stores readings, and transfers the information to all connected Rosemount 3DVision clients. The server computer should be running and logging at all times.
The Rosemount 3DVision Client is a graphical and interactive program, allowing the user to receive online measurement data, view a 3D visualization of the material stored in the vessels, add or remove sites, vessels, and Rosemount 5708, and manage alerts and reports. The Rosemount 3DVision Client supports several authorization levels of different privileges such as full-management authorized system administrators, technicians, and view-only operators.
Reference Manual
00809-0100-4570, Rev DA
7.1.2 Multi-client
The Rosemount 3DVision software is designed to work in a multi-client environment. Authorized users, connected to the same LAN or via external connections (WAN), have access to both real-time and historical data for all devices connected to the server.
Figure 7-2. Rosemount 3DVision Computer Topology
A
B
Local IP Address
192.168.1.100 Port 22222
Local IP Address
A. Remote 3DVision Client B. Rosemount 3DVision Server Computer C. Switch D. Router E. Internet F. Local Client (LAN)
192.168.1.104
E
Local IP Address
192.168.1.105
F
INTERNET
E
C
D
External Static IP
212.235.113.24
Local IP Address
192.168.1.106
Address
42
Operation
Page 53
Reference Manual
+ -
PS OUT
Display
+ -
PS OUT
Display
++--
PS IN 4...20mA
+ -
RS-485 PS IN 4...20mA RS-485
++-- + -
++--
PS IN 4...20mA
+ -
RS-485
++--
PS IN 4...20mA
+ -
RS-485
A
CD
A
B
E
00809-0100-4570, Rev DA
Figure 7-3. Rosemount 3DVision Server Connection Example
Operation
November 2017
A.120 Ω resistor B. PLC / DCS / Display C. 24 Vdc power supply D. Rosemount 3DVision Server E. RS-485 to USB converter
The Rosemount 3DVision Client application can connect to multiple Servers simultaneously. This feature enables the user to view data from multiple plants (without relation to geographical location) in a single screen. In addition, each Rosemount 3DVision Server can connect and supply information to an unlimited number of Clients. These features enable high flexibility in inventory management and amount of personnel reviewing the information.
When more than one client is logged as technician, only one can perform changes to the vessel or the configuration. If, for example, one user opened the Configuration Wizard, the second user that opens the wizard will get the following error message:
Figure 7-4. Setting Information Done by Another User
Operation
43
Page 54
Operation
November 2017
7.1.3 System requirements
Refer to Rosemount 5708 Quick Start Guide for the Software Installation requirements for the server and client.
7.1.4 Network considerations
When installing the Rosemount 3DVision Clients and Servers in a secured network or when it is needed to access the Rosemount 3DVision Server from an external connection (WAN – Wide Area Network) it is required to open TCP Ports in Firewalls, Routers or other devices or applications that might block such communication.
The required TCP Ports are:
22220 – Control Port22221 – Used for data Connection22222 – Connection of Rosemount 3DVision Client to Server22223 – Logs Download Data Port
Additional TCP ports may be required when integrating the 3DVision to SCADA/DCS system using Modbus communication protocol. See Table 7-5 on page 72 Server options - SCADA Configuration, and
Figure 7-33 on page 77.
Reference Manual
00809-0100-4570, Rev DA
44
Operation
Page 55
Reference Manual
00809-0100-4570, Rev DA

7.2 Installing the Rosemount 3DVision software

After completing the recommended preparations listed in the prerequisite information section below, the Rosemount 3DVision program is ready to be installed.
7.2.1 Required information
Gather the following information related to each Rosemount 5708 connection prior to the software installation and Rosemount 5708 connections:
Polling address of each Rosemount 5708 [00 – 63]Method of connection to the Rosemount 3DVision Server computerVessel dimensions, filling points and Rosemount 5708 locationProduct information – Density (for mass calculation) and product name
For more information on various methods of connecting to the Rosemount 5708, refer to Section 4:
Different connection methods.
7.2.2 Installation
For initial configuration, it is recommended to install both Rosemount 3DVision Server and Rosemount 3DVision Client on the same computer. However, it is possible to install the Rosemount 3DVision Server and Client on separate computers and connect to them accordingly.
Operation
November 2017
If you are using the Rosemount System Controller unit, skip to “Starting Rosemount 3DVision Server and
Client” on page 51.
It is recommended to have an active Internet connection during the installation process, as updated versions of Microsoft .NET Framework and other components may exist.
Note
Do not install the server application on remote computers. It is not necessary and may result in slow operation.
To install the software:
1. Insert the USB flash drive into the USB port or download the installation file from the support page in
Emerson website and execute the “Installer.exe” file. The installation wizard starts automatically and
the Rosemount 3DVision setup wizard appears.
If the installation program does not start automatically, run “Installer.exe” from the flash drive.
Operation
45
Page 56
Operation
November 2017
Reference Manual
00809-0100-4570, Rev DA
2. Select Install Rosemount 3DVision.
Note
If the computer where Rosemount 3DVision is being installed does not have Microsoft .NET Framework
4.0 installed, the Rosemount 3DVision setup wizard automatically detects and prompts you to install
this component. The installation file of the .NET Framework is provided in the flash drive supplied with the software and Rosemount System Controller.
3. Select your preferred language and OK to proceed. The installer's welcome box appears.
4. Select Cancel to stop the installation process and close the window. OR Select Next to proceed.
46
Operation
Page 57
Reference Manual
00809-0100-4570, Rev DA
5. Select Full install and select Next to install Rosemount 3DVision with all available features. The setup wizard completes the software installation. OR Select Custom install for remote client installations, and select Next to change the installation folder and select applications to install.
Operation
November 2017
6. Select “I do not accept the ...” to cancel and exit the installation. OR Select “I accept the ...” and select Next to proceed with the installation.
Operation
47
Page 58
Operation
November 2017
Reference Manual
00809-0100-4570, Rev DA
7. Select for which users the software should be available to:
Install for all users to make the software available to all computer users.Install for current user to make the software available to the present user account only.
8. Select Next to confirm and proceed to the next step. OR Select Back to return to the previous step or select Cancel to stop the installation.
9. When the Custom Install option is selected:
a. Check the required options for the applications to be installed. You may choose to install Client
app files only, Server app files only, or both.
Note
When installing a remote client, install the Client app files only.
b. Select Next to install the selected applications.
OR Select Back to return to the previous step or Cancel to stop the installation.
48
Operation
Page 59
Reference Manual
00809-0100-4570, Rev DA
c. If Server app files check-box was previously checked, select the installation method for the
Operation
November 2017
Rosemount 3DVision Server. The Server can be installed either as an Application (Process) or as a
Service.
Note
When installing the Server as a Service, the server starts automatically when the OS starts and does not require the user to log in.
d. Select Next to choose the installation folder.
OR Select Back to return to the previous step or Cancel to stop the installation.
e. To change the default installation folder (if required), select Browse and specify the desired
location for the installation.
Operation
49
Page 60
Operation
November 2017
Reference Manual
00809-0100-4570, Rev DA
f. Select Next to proceed with the installation.
10. Select the Windows Start Menu folder where you want the program’s shortcuts to be added in the
Choose Start Menu Folder window. You can also enter a name to create a new folder.
50
11. Select Install to proceed to install the software. Select Back to return to the previous step or Cancel
to stop the installation.
Operation
Page 61
Reference Manual
00809-0100-4570, Rev DA
12. Select Finish to complete the installation and close the setup wizard.
Following a successful installation, the installation package creates a desktop shortcut in accordance with the applications installed (Rosemount 3DVision Server, Client, or both). A shortcut is also placed in the Startup folder in Program Files.
Operation
November 2017
7.2.3 Starting Rosemount 3DVision Server and Client
When executing the Rosemount 3DVision Client for the first time, the system asks for details required to establish a connection to the Rosemount 3DVision Server. Following a successful connection, a new window appears to enter a user name and password. The default user level is Administrator, with the default user name: admin and password: admin.
1. Double-click the “Rosemount 3DVision Client” icon to start the client application.
The application opens and the Rosemount 3DVision Server Connection window appears. On the first connection attempt, the Server Name and Server Address fields are empty and the relevant details must be entered. The connection information is kept for subsequent connections.
Operation
51
Page 62
Operation
November 2017
Reference Manual
00809-0100-4570, Rev DA
a. In the Server Name field, enter your own description for the system managed by that server.
To activate the Server Name field, select Edit... Then the Server Configuration window appears.
b. In the Server Address field, enter the IP address of the computer running the Rosemount 3DVision
Server:
If both client and server applications are installed and running on the same computer, enter
the computer local IP address, or type IP 127.0.0.1 or type “localhost”.
If the client application is running on a remote computer, enter the IP address of the
computer running the Rosemount 3DVision Server application. c. In the Server Port field, verify that the server port is set to 22222. d. Select OK. e. Select Stay Signed In to automatically sign in on the next execution of the Rosemount 3DVision
Client application.
f. Select Slow Connection to increase timers and timeouts of various commands in case of slow
communication between the Rosemount 3DVision Server and the Rosemount 5708.
g. Select the Auto Reconnection after option to allow the Rosemount 3DVision Client to
automatically reconnect (with a delay of 20 seconds) to the Rosemount 3DVision Server.
52
h. Select Connect. The application connects and the Login dialog box appears.
2. Enter User Name and Password. The default username and password are both "admin" (case sensitive).
Note
It is possible to sign out from the File menu under: File > Exit.
Operation
Page 63
Reference Manual
00809-0100-4570, Rev DA
3. Define the user names, passwords, and permission levels of the various user accounts prior to setting up projects in the system. At least one user account must have the administrator permission level. For further details, refer to “Users Management…” on page 74.
Note
To log in to the technician mode of the system, enter the technician mode user name and password. For further details, refer to “Switch User…” on page 74.
4. The server application may need some time to initialize communications with the Rosemount 5708 and clients. Ensure the server icon appears on the task bar as shown.
The server should remain running. Manually running the server is required only if the computer is restarted and the server does not start automatically.
When installing the server, it is added to the Microsoft Windows Start Up files automatically.
Operation
November 2017
Operation
53
Page 64
Operation
November 2017

7.3 First-time running

The following steps are required to set a single vessel:
1. Creating a new project (see page 54).
2. Vessel configuration (see page 58).
3. Advanced parameters (see page 62).
4. Echo curve analysis (see page 62).
5. False echo mapping (see page 62).
6. Vessel properties configuration (see page 62).
These steps are basic and mandatory for the completion of a vessel commissioning. For additional changes and modifications of the vessel and for the Rosemount 3DVision Software, refer
to the various software menus starting on page 66.
7.3.1 Creating a new project
Following a successful login, the software automatically opens the last project. If no project exists, create a new project by selecting the New Project icon in the center of the screen or by selecting File > New… or by importing a saved project by selecting File > Import Project to Server….
Reference Manual
00809-0100-4570, Rev DA
Figure 7-5. New and Open Project Buttons
54
Operation
Page 65
Reference Manual
00809-0100-4570, Rev DA
To create a new project:
1. Select New Project. The Wizard window appears.
2. In the Name field, enter a name for the new project.
3. In the # Sites field, enter the number of sites the project should manage (each site may contain multiple vessels).
Figure 7-6. Project Wizard - Site Name
Operation
November 2017
4. Select Next.
Note
If you are running the Rosemount 3DVision application you can specify one site and one vessel only. An upgrade to the Rosemount 3DMultiVision version is required for configuring more than one site. Contact your local Emerson
representative for purchasing a license key for upgrading to Rosemount
3DMultiVision.
5. Enter the site name (limited to 16 characters), number of vessels in the site, and number of Rosemount 5708 in each vessel. Sites, vessels, and Rosemount 5708 can be edited later.
6. Select the connection type between the computer and the Rosemount 5708 in the site.
RS-485, TCP/IP, GPRS and GPRS+SMS connections are available. The very basic connection method is RS-485. For further information on the connection methods, refer to “Communication methods” on
page 91.
Operation
55
Page 66
Operation
November 2017
Reference Manual
00809-0100-4570, Rev DA
7. Select the serial COM port for the server PC connection.
Figure 7-7. Project Wizard - Site Connection
8. Select Finish.
The new project is created and the vessels and sites appear on the main screen (Figure 7-8).
Note
When a project includes more than a single site, each site should be defined individually with the number of vessels, the number of Rosemount 5708 units in each vessel, connection method, and serial COM port. Selecting Next switches to the next site. Repeat site configurations for each site. In the following example, a single site is defined.
Note
It is possible to change at any time the communication method, switch user to technician, select the device from the device selection drop-list appear above the 3D visualization (see Figure 7-15 on
page 63), and go to the parameters tab, disconnect the Rosemount 5708 and set the new
communication method and parameters. Select Save to save the project.
The project window is built according to the project hierarchy of Sites > Vessels > Scanners.
56
Operation
Page 67
Reference Manual
00809-0100-4570, Rev DA
Figure 7-8. Project Window
Operation
November 2017
Multiple sites of the project may be defined. The defined sites appear and can be selected from the site selection bar.
When selecting a site, its relevant information and configuration options appear in the main work area organized in two tabs: Vessels and Logs. Refer to “Main screen” on page 66 for display area descriptions.
Multiple vessels may be defined for each site. The vessels are represented by icons appearing on the vessel selection bar. When a site is selected, the corresponding vessels are displayed in the vessel selection bar.
Selecting a vessel from the vessel selection bar displays various types of information and configuration options related to that vessel in the main work area. This information is organized in tabs: Overview, Logs, and Parameters. Right-clicking a vessel icon opens a menu with options related to that vessel.
Multiple devices may be defined under each vessel. A device can be selected and accessed from the device selection list (refer to “Main screen” on page 66).
Note
Selecting a Rosemount 5708 for configuration under a specific vessel can be done only by logging in as a user with Technician or Senior Technician privileges.
Refer to “Users Management…” on page 74 for further details regarding user levels and authorizations.
Operation
57
Page 68
Operation
November 2017
7.3.2 Vessel configuration
Following successful connection, it is required to configure the devices according to the physical installation and with information of the vessel’s dimensions and Rosemount 5708 mounting position.
The wizard is a 4-step procedure to configure the Rosemount 5708 to start working in the application installed, during the Configuration Wizard’s steps, there is a 3D visualization of the vessel representing the actions in each step.
To configure the vessel and devices
1. Select the vessel from the vessel selection bar (refer to “Main screen” on page 66).
2. On the Device menu, select Device Configuration Wizard…. The Configuration Wizards opens (Figure 7-9 on page 59).
3. Set the Measurement Units in use for Distance and for Tem pe ra tu re.
4. Select the proper shape for the vessel top from the Top Shape field. The options are: Flat, Cone, Dome, or Pyramid. Set the dimensions:
For a Cone shape, set the height and the top diameter of the cone.For a Dome shape, set the height of the dome.For a Pyramid shape, set the top section height, width and length of the vessel top at the X and Y
directions.
Reference Manual
00809-0100-4570, Rev DA
5. Select the proper shape for the vessel center from the Center Shape field. The options are: Cylinder or Cube or cylinder sector. Set the dimensions:
For a cylinder shape, set the height and the diameter.For a cubic shape, set the height, width and length of the vessel according to the X and Y axis,
respectively.
For a cylinder sector, set both bottom and top shapes to flat, set the internal and external
diameters, the sector angle (from 15 and up to 90 degrees) and set the sector height.
6. Select the proper shape for the vessel bottom section from the Bottom Shape field. The options are: Flat, Cone, Dome, Pyramid or Inverted Cone. Set the dimensions:
For a Cone shape, set the height and the bottom diameter of the bottom section.For a Dome shape, set the dome height.For a pyramid shape, set the height, width and length of the vessel bottom section at the X and Y
directions. If the vessel shape is asymmetrical (for both cone and pyramid shapes), it is possible to set the
bottom center as an offset from the entire shape center. The X Position and Y Position are the offset of the bottom section from the center.
For Inverted cone, set the height, bottom diameter and top diameter. The inverted cone can be configured only inside the center shape and only when it is set as a cylinder. Top diameter must be smaller than bottom diameter.
The 3D visualization in this step shows the complete vessel. When setting one of the shapes, the 3D visualization will highlight the shape in red to indicate the part being configured at that moment.
58
7. Select Next.
Operation
Page 69
Reference Manual
00809-0100-4570, Rev DA
Figure 7-9. Configuration Wizard Step 1 - Vessel Dimensions
Operation
November 2017
8. Set the Rosemount 5708 position on the silo top.
It is possible to set X and Y coordinates measured from the center of the vessel and set the offset of the Rosemount 5708 from the top of the vessel at the mounting position.
Z position is calculated automatically when the X and Y coordinates modified or there is a change to the offset from the top.
The angle is being calculated automatically when the X, Y coordinates changes, When Senior Technician user is logged in, there is an option to modify the angle manually by checking the Set device angle manually.
When entering this step, the 3D visualization is showing the silo from the top and indicating the Rosemount 5708 in the vessel and showing a red arrow indicating the direction of the device that should match the physical installation.
Operation
59
Page 70
Operation
November 2017
Reference Manual
00809-0100-4570, Rev DA
Figure 7-10. Configuration Wizard Step 2 - Rosemount 5708 Positioning
9. Select Next.
10. Add the filling points on the top of the vessel. Up to nine filling points may be defined.
Select Add and enter the X and Y coordinates in the Filling Points table and press Enter after
every number entered. Filling points are marked as an inverted orange cone at the set position.
Select Delete to remove the selected filling point.Select Clear to remove all filling points.
Figure 7-11. Configuration Wizard Step 3 - Device Position
60
11. Select Next.
Operation
Page 71
Reference Manual
00809-0100-4570, Rev DA
12. Configure the vessel full and empty calibration levels. Figure 7-12 shows the full and empty
calibration fields to set. It is possible to set the full and empty calibration as measured from the top (Distance) of the vessel or measured from the bottom (Level) of the vessel.
Set the distance from the top of the vessel to set 100% (20 mA) (see Figure 7-12 on page 61) in
Set the Empty Calibration by setting the level or the distance from the top of the vessel to set the
Select the Default button to set the full and empty calibration levels to default values.Select View Table to open a window showing a summary table of each Rosemount 5708 in the
13. Select Finish.
Figure 7-12. Configuration Wizard Step 4 - Full and Empty Calibration Levels
Operation
November 2017
the Full Calibration section. OR Set the maximal material level in the Level (Bottom) field. Dead band is 1.64 ft (0.5 m) from the top of the Rosemount 5708 antenna case.
0% (4mA) (see Figure 7-12). Default value is set to the bottom of the vessel.
vessel and the actual full and empty calibration levels configured, see Figure 7-13 on page 61.
Operation
Figure 7-13. Full and Empty Calibration Levels Summary Table
14. When the configuration wizard settings are completed, select the Finish button. The parameters will
be uploaded to all devices in the vessel.
61
Page 72
Operation
November 2017
7.3.3 Advanced parameters
It is recommended to adjust some of the advanced parameters. Open the advanced parameter window (select Device > Advanced Parameters… or press F3) in order to
set general parameters, such as Max Capacity, Filling and Emptying Rates, Material Slope and Damping Time. Other parameters may be configured and fine adjustments can be done at a later stage. Refer to
“Advanced parameters” on page 96 for detailed instructions and descriptions.
7.3.4 Echo curve analysis
When first configuring the vessel, it is recommended to perform Echo Curve Analysis, see chapter
“Performing Echo Curve Analysis” on page 109.
View the Echo Curve Analysis results in echo curve viewer. For additional information, see chapter
“Viewing Echo Curve Analysis” on page 110.
Using the Echo Curve Analysis it is possible to determine if any of the advanced parameters need additional changes.
7.3.5 False echo mapping
Reference Manual
00809-0100-4570, Rev DA
Using the Echo Curve Analysis result, it is possible to determine if false echo mapping actions are needed. For additional information see “False Echo Mapping” on page 112.
7.3.6 Vessel properties configuration
The example below shows how to configure a single site with a single vessel and two Rosemount 5708S.
Note
In order to configure a vessel with more than one Rosemount 5708, all devices must have the same model number (e.g. Rosemount 5708S). Contact your local Emerson representative to purchase a license key for upgrading options.
To configure vessel properties:
1. Select the vessel from the vessel selection bar (refer to “Main screen” on page 66).
2. Right-click the vessel and select Property… from the drop-down menu. The Edit Base Properties window opens (Figure 7-14).
3. Enter the vessel name and description. Select OK.
Figure 7-14. Edit Base Properties Window
62
4. Double-click the vessel in the vessel selection bar to open the vessel measurements and details in the main work area (Figure 7-15). The Overview tab shows how many devices are set for this vessel and their connection status.
Operation
Page 73
Reference Manual
00809-0100-4570, Rev DA
Note
If the selected vessel contains more than one device and the user has technician privileges, a device selection list appears (see Figure 7-15). Select the required device for configuration from the list.
Figure 7-15. Rosemount 3DMultiVision Vessel Overview Tab
Operation
November 2017
Operation
5. Select the Parameters tab to configure the Rosemount 5708 parameters (Figure 7-16 on page 64).
Note
Rosemount 5708 parameters can be modified only in technician mode. For further information on the connection methods, refer to “Communication methods” on page 91.
63
Page 74
Operation
November 2017
Reference Manual
00809-0100-4570, Rev DA
Figure 7-16. Rosemount 5708 Parameters Tab
6. Select the Devices tab to set connection type, serial COM port, and polling address.
Figure 7-17. Rosemount 3DMultiVision Vessel Devices Tab
7. Set Connection Type (the unsupported types are visible but not selectable).
Note
Connection type is enabled for editing only when the Rosemount 5708 is disconnected.
For further information on the connection methods, refer to “Communication methods” on page 91.
64
8. Set Serial COM Port.
9. Set Polling Addresses.
Operation
Page 75
Reference Manual
00809-0100-4570, Rev DA
10. Configure each Rosemount 5708 set for the vessel by selecting them from the device selection list
and entering their communication parameters.
11. Once all devices are configured, connect to all the devices in the vessel by right-clicking the vessel
name and selecting the Connect option from the drop-down menu.
12. When a connection is established, select the site from the site selection bar (as shown to the right).
The vessel profile screen appears with the results of the scanned measurements.
13. To view a 3-dimensional profile, right-click on a vessel in the main work area and select Show 3D
visualization from the drop-down menu. The profile window opens (Figure 7-18).
Figure 7-18. Product 3-Dimensional Profile
Operation
November 2017
Operation
Note
When connecting to the Rosemount 5708S, the 3D visualization of the product in the vessel is available by double-clicking the vessel.
For further information on the 3D visualization, refer to “3D visualization” on page 128.
65
Page 76
Operation
A. Main menu bar: Includes the main menus: File, Communication, Edit, Device, View, Tools, and Help.
B. Project toolbar: Includes various tools for project management and operations including project opening and saving,
connection to sites, and data loading.
C. Site selection bar: A project allows multi-site management, where each site may include multiple vessels. In the site
selection bar, a site can be selected for monitoring and operations.
D. Vessel selection bar: Lists all vessels included in the selected site. In the vessel selection bar, a vessel can be selected for
monitoring and various operations.
E. Control tabs: Includes tabs related to the selected item (site, vessel, or Rosemount 5708). Each selected tab is displayed in
the main work area.
F. Main work area: Provides graphical and textual information and properties, and different modification and control
options related to the selected item (site, vessel, or Rosemount 5708) and tab.
G. Status bar: Displays actual information related to different operations performed, such as downloading data and
connection.
A
B D
E
F
G
C
November 2017

7.4 Software navigation

This chapter provides an overview of the various toolbars, menus, and projects.
7.4.1 Project overview
The Rosemount 3DMultiVision software allows for defining different projects. Each project is organized in the following hierarchical structure:
Top Level: The projectSecond Level: A list of sitesThird Level: A list of vessels located in the siteFourth Level: A list of devices associated with each vessel
For example, a single project may include multiple sites (or multiple groups of vessels), where each site/group may contain multiple vessels, with a minimum of one Rosemount 5708 per vessel.
7.4.2 Main screen
Figure 7-19 displays the main screen areas followed by their descriptions.
Reference Manual
00809-0100-4570, Rev DA
Figure 7-19. Main Screen
66
Operation
Page 77
Reference Manual
00809-0100-4570, Rev DA
7.4.3 Main menu bar
The Rosemount 3DVision Client main menu bar accesses the main menu options.
Figure 7-20. Main Menu Bar
File menu
The File menu includes the following options:
Figure 7-21. File Menu
Operation
November 2017
Table 7-1. File Menu
Option Description
New... Starts a new, empty project.
Open... Opens an existing project.
Save Project Saves the current project.
Save Project As... Saves current project as a different name.
Export Project from Server… Exports a project in XML format.
Import Project to Server… Imports a project in XML format.
Delete Project Deletes a project from the server. A window opens with a list of all projects
Browse to Local Folder… Allows browsing the software installation folder.
Exit Closes the application.
Note
The Import and Export functions are for maintenance. They enable downloading projects to a local computer for modifications and Software upgrades, and then uploading the modified projects back to the server.
stored in the server and the desired project is selected.
Operation
67
Page 78
Operation
November 2017
Reference Manual
00809-0100-4570, Rev DA
Note
Projects can only be deleted if there is at least one project remaining. The current project cannot be deleted.
Communication menu
The available options in the Communication menu change in accordance with the selected item. If a vessel is selected, it corresponds to a vessel connection. If a Rosemount 5708 is selected, it displays Rosemount 5708 related commands, e.g. Connect & Disconnect.
The following options are available when a vessel is selected:
Figure 7-22. Communication Menu
Table 7-2. Communication Menu
Option Description
Connect All Connect to all the devices in the selected site.
Disconnect All Disconnect from all the devices in the selected site.
Load from Vessels Download all data from all vessels.
Connect Vessel Connect to all devices in the selected vessel.
Disconnect Vessel Disconnect all devices in the selected vessel.
Load from Vessel Download all data from all devices in the selected vessel.
68
Operation
Page 79
Reference Manual
00809-0100-4570, Rev DA
Edit menu
The Edit menu includes the following options:
Figure 7-23. Edit Menu
Table 7-3. Edit Menu
Option Description
Operation
November 2017
Add Add sites, vessels, or devices depending on the level selected in
Site On the site level it adds sites or vessels.
Vessel On the vessel level it adds devices.
Scanner On the device level it is disabled.
Delete Delete sites, vessels, or devices depending on the level selected
Site On the site level it deletes sites.
Vessel On the vessel level it deletes vessels.
Scanner On the device level it deletes devices.
Properties Set a name and description to a site or a vessel
Site On the site level define the name and description of the site.
Vessel On the vessel level define the name and description of the vessel.
Vessel graphic figures Additional option (only available when a vessel is selected): Add
Edit Project... An interface option to rearrange the vessels and sites per
Save Selected Vessel Configuration As... This operation saves the selected vessel configuration into an
the project pane.
in the project pane.
text labels on the 3D visualization (X, Y, and Z coordinates are referenced to the center axis located at the bottom center of the vessel).
requirements.
XML file for maintenance and support issues.
Operation
Load Vessel Configuration OnLine... This operation uploads a saved XML file with the vessel
OffLine Configuration When in Offline mode there are options to work on existing
Manual Load to Selected Vessel... This operation loads parameters manually when the software is
Cancel Manual Load from All Vessels This operations cancels manual load from all the vessels in the
Browse to Vessel AutoSaved File Folder...
configuration into the project that is running online.
project.
offline and not connected to devices.
project.
This operation opens a window within the folder with all the saved configuration XML files.
69
Page 80
Operation
November 2017
Reference Manual
00809-0100-4570, Rev DA
Figure 7-24. Edit Project Option
Figure 7-25. Vessel Graphic Figures
70
Figure 7-26. Example of Adding a Text Label to the 3D visualization
Operation
Page 81
Reference Manual
00809-0100-4570, Rev DA
Device menu
The device menu includes the following options:
Figure 7-27. Device Menu
Operation
November 2017
Table 7-4. Device Menu
Option Shortcut Description
Device Configuration Wizard... F4 Opens the configuration wizard for the selected vessel, see
Advanced Parameters... F3 For additional info, refer to “Advanced parameters” on
Device Current Simulation Settings…
Device Output Current… F8 For additional information, refer to “Device output current”
Device Output Settings… F7 For complete information about all functions under the
Echo Curve Analysis… Ctrl+Z For complete information about all functions under the
Echo Curve Analyze Viewer… Ctrl+V For complete information about all functions under the
Device False Echo Mapping… N/A For complete information about all functions under the
Devices Activations… N/A Allows different types of Resets, Update Firmware and to test
Update Model... N/A Allows a Rosemount 5708L or 5708V to be upgraded to a
F6 For additional info, refer to “Device current simulation
“Vessel configuration” on page 58.
page 96.
settings” on page 101.
on page 102.
Output Settings window, refer to “Device output settings” on
page 103.
Output Settings window, refer to “Echo Curve Analysis” on
page 109.
Output Settings window, refer to “Viewing Echo Curve
Analysis” on page 110.
Output Settings window, refer to “False Echo Mapping” on
page 112.
communication quality. For complete information about all functions under the Output Settings window, refer to “Device
activations” on page 114.
Rosemount 5708V or 5708S.
Operation
71
Page 82
Operation
November 2017
Reference Manual
00809-0100-4570, Rev DA
Note
Some of the options allow changes only if the user is a Technician or a Senior Technician. Refer to “Users
Management…” on page 74 for additional information. Viewer has no option to do any changes under
this menu, Administrator can do only some of the actions.
Tools menu
The Tools menu offers useful tools such as reporting and administration tools, material configuration, Server options, user management and Demo mode.
Figure 7-28. Tools Menu
72
Table 7-5. Tools Menu
Option Description
Reports Report files are created and stored in Excel worksheet format and can be
Reports Definition Wizard... (F9)
Export Present Project Reports from Server
Browse to selected project report folder
Show last generated report
opened, viewed, and manipulated using Excel.
Note
To open and view report files in Excel® format, Microsoft Excel must be installed on the computer.
Opens a wizard to set, edit and generate reports.
“Reports Wizard” on page 116.
Opens a date picker to select the From and To dates for the project reports. Select OK to retrieve the list of corresponding report files from the server (Figure 7-29). Select the required reports from the list. Click Select to browse to a folder to store the selected report.
Opens a file browse window in the project report files location on the local hard drive. Available only if the Client is running on the same computer as the server.
Opens the last generated report in Excel. Available only if the Client is running on the same computer as the server. See Figure 7-29.
Operation
Page 83
Reference Manual
00809-0100-4570, Rev DA
Option Description
Generate report for selected vessel (logs data)
Select logs directory from local folder…
Operation
November 2017
Browse to local folder to retrieve saved reports
Download from server
Hardware Inventory Ta bl e …
Measurement Summary Ta bl e …
Events Log Viewer... Displays a list of events tracking of changes in the project,
Material Configuration… Opens the Material Configuration window for defining new material types
Server Options... Summary: Opens a window with a summary of the server information. See
Select time and date of the required information. The generated file is a .pdf and will be saved at a selected location.
Displays a special report including details for all project devices. For more details, refer to “Hardware inventory table” on page 125.
Displays a special report of vessel storage measurements for all project vessels. For more details, refer to “Measurement summary table” on
page 125.
and color and for setting bulk density for mass calculation. This data is displayed in the vessel images. For further details, refer to “Material
configuration” on page 127.
Figure 7-30.
Configuration: Allows setting the Server's Local folder to save all files.
Active TCP Connections: Retrieves and displays a list of all active TCP
connections on Server computer and contains a list of all existing connections to the server. Each connection displayed includes the local port, remote port, TCP/IP address, and status. See Figure 7-31.
Active TCP Listeners: Retrieves and displays a list of all existing TCP ports listeners on the server computer. See Figure 7-32.
Scada Configuration: Allows SCADA systems to be integrated to the server. Opens a window for activating integration support and specifying up to four TCP listening ports, allowing up to 4 different SCADA systems to retrieve information from the server. The communication type is Modbus RTU encapsulated in TCP/IP packets or Modbus TCP. In addition, a folder path can be set to store a file containing the 3D visualization to be popped up from the SCADA system. See Figure 7-33.
Operation
License Key: Upgrade from the single-vessel Rosemount 3DVision software license to the Rosemount 3DMultiVision software license by entering a new license key. Also entering license codes for Center of Gravity and Output by Sections activations. For a detailed description, refer to “Upgrading to
Rosemount 3DMultiVision” on page 132.
Paste the License key you received into the field and select Apply. The Rosemount 3DVision Client title bar switches from Rosemount 3DVision to Rosemount 3DMultiVision, if the key is correct. See Figure 7-35.
Server Options... (Continued)
Connect to Server… Connects the client to a server (if not already connected), or disconnects the
Sign Out Cancels the 'Stay Signed In' option previously selected when the Rosemount
The Don’t remind check-box on Figure 7-34 is for not getting the notification on the top window screen part that there is a remaining X days until the free 30-days trial Rosemount 3DMultiVision license is over.
Logged Clients: Shows a list of clients currently connected to the Rosemount 3DVision Server. See Figure 7-35.
current connection and connects to a different server. For further details regarding server connection, refer to “First-time running” on page 54.
3DVision Client was executed.
73
Page 84
Operation
November 2017
Reference Manual
00809-0100-4570, Rev DA
Option Description
Switch User… Logs out the current user and logs in as a different user. This function opens a
window requesting the user name and password.
Note
Each user account has its own defined permission level. If you switch to a user account with a lower permission level, you will be more limited in the actions you can perform.
Users Management… Opens the Users dialog box displaying a list of all the user accounts defined
on the server and their permission levels. For further details regarding user levels and authorizations, refer to “User management” on page 126. This option is available to users with Administrators privileges. See Figure 7-36.
The Users dialog box enables adding, removing, and changing properties of users.
Users Authorization is Enabled: Un-checking this field will allow
opening the Client without the need to enter user name and password.
Add: Select to add a new user. A dialog box opens requesting a name,
password, and permission level for the new user.
Remove: Select the user account you want to delete, and then select
Remove.
Properties: Select a user from the list and select Properties to change
their properties. A dialog box opens enabling the user to change the name, password, and permission level for the selected user.
Close: Select to accept the changes and close the Users dialog box.
Client Options... Site - Vessel: See Figure 7-37.
Site view manual: Allows multiple vessels in a single display without the
need to scroll the window.
Vessel width limits: Defines the minimum and maximum width
dimensions of the vessel in site view. Measured in pixels. When the window is stretched or reduced, the size of the vessel will change between these values.
Vessel height limits: Defines the minimum and maximum heights of the
vessel in site view.
Group by name/material/site name: Defined in parallel to vessel limit
definition, to give full flexibility in different types of layouts and switching between layouts.
Reduced vessel display: When reducing the size of window, the vessel
size in site view will be smaller up to the limit of minimum size configured in this window. Values on the right side will disappear and only the first in the list will appear above the vessel drawing.
Enlarge text area width in vessel display: If checked, it will enlarge the
text above the vessel's image. By default, it is not checked and the text above the vessel's view will be small.
Enlarge Sites tool bar to maximum: When working with multiple sites,
if checked, it will squeeze all the sites in the menu bar as much as possible, else it will give the option to scroll left and right for sites display.
Show vessels chart lines according to site: In the vessel's logs tab, all
the logs of the vessels in the site (group) that the vessel is related to will be shown. For example, if we have three sites, and in one of the sites we have 7 vessels, selecting this option, all the seven related trends in one log chart will be displayed.
74
Operation
Page 85
Reference Manual
00809-0100-4570, Rev DA
Option Description
Operation
November 2017
Client Options... (Continued)
Refresh Display (F5) This option refreshes the screen display in the case of communication
Demo Activates the software in Demo mode. This option includes selecting a demo
Project: See Figure 7-38.
Auto save project on vessel change: If checked, it will automatically save
all changes made to the project.
Show save project warning on vessel change: if checked, for every
change done in the project, it will give a pop up window to alert the user that a change is planned, and the user must save in order to continue.
Logo Image File...: The user can add a (company) logo/image on the top
right corner of the Rosemount 3DVision main screen.
Change Language: See Figure 7-39. Enables changing the software's language.
Server Connection Mode: See Figure 7-40. Enables the user to choose between single server connection or multiple servers connection from a client's Rosemount 3DVision.
Translate Application: Allows the user to translate the Rosemount 3DVision to any language and send the translation file to the manufacturer to embed in future versions.
Sites/Vessels Filtering: See Figure 7-41. In a complex project with multiple sites and vessels, the software allows an option of viewing a partial set of selected vessels and sites only, per the preference of the user.
problems.
site, performing a simulated operation of the sites, various components, and creating logs. For further details, refer to “Demo mode” on page 126.
Figure 7-29. Report Files Selection
Operation
75
Page 86
Operation
November 2017
Reference Manual
00809-0100-4570, Rev DA
Figure 7-30. Server Options... - Summary
Figure 7-31. Server Options... - Active TCP Connections
76
Operation
Page 87
Reference Manual
00809-0100-4570, Rev DA
Figure 7-32. Server Options... - Active TCP Listeners
Figure 7-33. Server Options... - Scada Configuration
Operation
November 2017
Operation
Figure 7-34. Server Options... - License Key
77
Page 88
Operation
November 2017
Reference Manual
00809-0100-4570, Rev DA
Figure 7-35. Server Options - Logged In Clients
Figure 7-36. Users Management Dialog Box
78
Operation
Page 89
Reference Manual
00809-0100-4570, Rev DA
Figure 7-37. Client Configuration Window - Site - Vessel
Operation
November 2017
Operation
79
Page 90
Operation
November 2017
Reference Manual
00809-0100-4570, Rev DA
Figure 7-38. Client Configuration - Project
Figure 7-39. Client Configuration - Change Language
80
Operation
Page 91
Reference Manual
00809-0100-4570, Rev DA
Figure 7-40. Client Configuration - Server Connection Mode
Operation
November 2017
Figure 7-41. Client Options / Sites - Vessels Filtering
Operation
81
Page 92
Operation
November 2017
Reference Manual
00809-0100-4570, Rev DA
Help menu
The Help menu includes a link to the Rosemount 5708 Quick Start Guide and to this reference manual. The About option opens a window with version details, Emerson support number, and a link to the
Emerson website.
Figure 7-42. Help Menu
82
Operation
Page 93
Reference Manual
00809-0100-4570, Rev DA
7.4.4 Project toolbar
Figure 7-43. Project Toolbar
Function Description
Operation
November 2017
Level/Distance Data Toggles the measurement type displayed in the vessel by either Level or
Connect Vessel Connects the selected Vessel to the application.
Disconnect Vessel Disconnects the selected Vessel from the application.
Load from Vessel
Wizard This button is a shortcut for opening the Wizard.
Echo Curve A shortcut to open the Echo Curve Analysis window, same as clicking
VDC Stands for Vessel Data Collector - a tool to help users collect all relevant
Distance.
Level refers to the level of material measured from the bottom of the silo. Distance refers to the distance of the material from the Rosemount
5708.
The selection applies for measurements in all viewing panes, at all view levels, for the corresponding values (min/max/average).
Downloads all measurements from the selected vessel to the application. All data is automatically downloaded from all devices to the server.
Device > Echo Curve Analysis or by clicking Ctrl-Z
information from the vessel and to send it to the factory. Need to add another chapter showing the windows and options on the VDC.
Operation
83
Page 94
Operation
Sort By: Name view
Sort By Manual
November 2017
Reference Manual
00809-0100-4570, Rev DA
Sort By Organizes the vessels by group.
All: Arrange by the numbering order as defined in the project.Material: Displays the vessels ordered alphabetically by material.Name: Displays the vessels ordered alphabetically by name.All Same Size: Displays all vessels as the same size in one window
(without scrolling).
Manual 4x4, 5x5 and 6x4: Displays the vessels arranged in a matrix
with the selected dimensions.
Manual: Displays the vessels arranged in a matrix with dimensions as
defined in Too ls > Client Options > Site - Vessel > Site View Manual Layout.
When selecting a sorting option other than All, the site display changes and shows the vessels in groups arranged according to the sorting criteria, as shown in the following image:
The Manual sorting options display the vessels with different sizes in accordance with the selected option. Smaller sizes may be more convenient when managing a large site. For example, selecting Manual 4x4 displays the vessels as follows:
Connection status Indicates the client connection to the server. See “Aler t types” on
page 152.
84
Operation
Page 95
Reference Manual
00809-0100-4570, Rev DA
7.4.5 Vessel selection behavior
Vessels are displayed in blocks in the main work area. Each vessel is represented by a rectangle with the following information:
The vessel image containing the material stored using the selected material color at its current
level/distance.
Stored material name and device model.Empty and full calibration levels represented by orange and red lines respectively.Level measurements: Average level/distance, maximum level and minimum level/distance, in the
configured measurement units, and volume in percentage. Additional information can be added by right clicking the vessel. For more information, refer to “Parameters and alerts” on page 152.
A connection status indicator: Note that in the case of a connection failure, the indicator and the entire
background of the rectangle becomes red.
Figure 7-44. Main Work Area
Operation
November 2017
Operation
85
Page 96
Operation
A
B
November 2017
Reference Manual
00809-0100-4570, Rev DA
Double-clicking a vessel icon in the vessel selection bar replaces the display with a detailed vessel pane in the main work area, with the Overview tab selected.
Figure 7-45. Detailed Vessel Pane when Double-clicking a Vessel Icon
A. Double-click a vessel icon to view a detailed vessel pane in the main work area B. Detailed vessel pane
Right-clicking the vessel icon opens a drop-down menu with the following features:
Connect: Connects to the devices on the vessel.Disconnect: Disconnects communication between the server and the Rosemount 5708.Load from Vessel: Downloads all parameters from the devices connected to the software.Generate report (logs data) - Download from server... : Opens the Search Log window for selecting
a report date and time. Then it downloads the corresponding historical data from the server and opens a Save As dialog box to save the report file.
- The report file is created in PDF format and includes the date, time, project, and vessel details, and volume/mass and alert parameters for the selected date and time.
- The report also includes a 3D visualization of the corresponding vessel.
Generate report (logs data) - From local folder...: Opens a Browse for Folder dialog box to browse to
the folder where the related data is stored. Creates a report as described above, but per data stored locally.
Property...: Opens the Parameters Definition window allowing the user to select the visible material
properties in the representing rectangle and the alert types to be used for that vessel. For further details regarding alerts, see “3D visualization” on page 128.
86
Operation
Page 97
Reference Manual
A
B
00809-0100-4570, Rev DA
Figure 7-46. Vessel Selection
Operation
November 2017
A. Right-click the vessel icon to open a drop-down menu B. Click a vessel in the main work area to select the vessel
A single click selects the vessel. The selected rectangle becomes outlined with a yellow frame, and the corresponding vessel icon in the vessel selection bar is selected.
When right-clicking a selected vessel in site view, the drop-down menu opens with the following added options:
Show 3D Profile...: Opens a 3D visualization window.Connect: Connects communication between the selected vessel Rosemount 5708 and the server.Disconnect: Disconnects communication between the selected vessel Rosemount 5708 and the
server.
Load from Vessel: Loads data from the vessel Rosemount 5708 and updates the displayed status and
parameters.
Device Configuration Wizard...: Opens the Configuration Wizard for the selected vessel.Advanced Parameters…: Opens the Advanced Parameters window.Echo Curve Analysis…: Opens the Echo Curve Analysis window.False Echo Mapping…: Opens the False Echo Mapping window.Edit: Allows for adding a new device to the selected vessel or deleting the entire vessel.Generate report (logs data) – Download from server...: Opens the Search Log window for selecting
a report date and time. It then downloads the corresponding historical data from the server and opens a Save As dialog box to save the report file.
- The report file is created in PDF format and includes the date, time, project and vessel details, and volume/mass and alert parameters for the selected date and time.
Operation
- The report also includes a 3D visualization of the corresponding vessel.
87
Page 98
Operation
A
November 2017
Reference Manual
00809-0100-4570, Rev DA
Generate report (logs data) - From local folder...: Opens a Browse for Folder dialog box to browse to
the folder where the related data is stored. Creates a report as described above, but per data stored locally.
Material Selection...: Opens the Material Configuration window allowing the user to select a stored
material for the specific vessel.
Properties...: Opens the Parameters Definition window allowing the user to select the visible material
properties in the representing rectangle and the alert types to be used for that vessel. For further details regarding alerts, refer to “Parameters and alerts” on page 152.
Figure 7-47. Right-clicking a Selected Vessel in Site View
A. Right-click a selected vessel to open a drop-down menu
88
Operation
Page 99
Reference Manual
A
B
00809-0100-4570, Rev DA
7.4.6 Vessel pane
The Vessel pane provides vessel monitoring options and basic vessel operations (Figure 7-48). This pane opens in the main work area when a vessel is selected from the vessel selection bar.
Figure 7-48. Vessel Pane
Operation
November 2017
A. Vessel selected from the vessel selection bar B. Device drop-down list
Overview tab
Selecting the Overview tab displays two panes in the main work area. The left pane includes the vessel connection information, alerts, measurements, and indications related
to the selected vessel:
Vessel name, connection status of the vessel, and selected devicesAlerts, if they existDevice names, versions, and communication detailsAverage, Minimum and Maximum Level/DistanceVolume (%), Volume in the selected units and max scaleMass (if density is defined) and max scaleTem pe ra t ur eSNRRosemount 5708 Output CurrentA graphical representation of the vessel, including the vessel shape, maximum and minimum levels,
and connection status indicator.
Material button: Opens the Material Configuration window which allows for managing the materials in the vessel. The material can be chosen out of a pre-defined list of materials and their representing colors.
Operation
89
Page 100
Operation
November 2017
Reference Manual
00809-0100-4570, Rev DA
When a device is selected from the device list (when the user is with technician privileges), the following information is also displayed:
Scanner NameConnection TypeCOM Port Polling AddressS/N – electronic card serial numberHardware VersionFirmware VersionDevice model
Logs tab
The Logs tab provides a chart representation of historical measurements data and allows for different manipulations of chart inspection. For further details, refer to “Log views” on page 146.
Parameters tab
Selecting the Parameters tab, the following parameters are displayed in the main work area:
Vessel SCADA IDScannerPoll AddrS/NHardwareFirmwareDevice TypeAvg. LevelMax. LevelMin. LevelVolu meTem pe ra t ur eSNRCurrent Output (mA)
90
Devices Tab
Selecting the Devices tab displays and allows modifying the communication settings. In addition, it show the Rosemount 5708 Polling Address, or a list of device polling addresses for a vessel with multiple Rosemount 5708 in a system configuration vessel. Option modifications are not allowed when devices are connected. Click Disconnect next to the device or at the toolbar and modify communication as needed. Multi-scanner vessels may include Rosemount 5708 units with different polling addresses that are not in series.
For a detailed explanation regarding the various ways of connect to a Rosemount 5708,
“Communication methods” on page 91.
Operation
Loading...