MODEL 54e C SPECIFICATIONS
PHYSICAL SPECIFICATIONS - GENERAL
Enclosure: Epoxy-painted cast aluminum
NEMA 4X (IP65),144 X 144 X 132mm, DIN size
(5.7 X 5.7 X 5.2 in.)
Front Panel: Membrane keyboard with tactile feed
back and user selectable security. Light gray, blue
and white overlay. Dark gray enclosure, light gray
bezel.
Display: Three-line LED back-lit dot matrix LCD
(7.0 x 3.5 cm). The display contrast is compensated for ambient temperature.
Process Variable Character Height: 16mm (0.6 inch)
Electrical Classification:
Class I, Division 2, Groups A, B, C, & D.
T5 Ta=50°C. Dust ignition proof: Class II,
Division 1, Groups E, F, & G; Class III.
CSA-LR34186:
Max. relay contact rating: 28 Vdc; 110 Vac;
230 Vac; 6 amps resistive
FM: Max. relay contact rating: 28 Vdc resistive
150 mA - Groups A & B;
400 mA - Group C;
540 mA - Group D
Power:
Code -01: 100 - 127 VAC, 50/60 Hz ± 6%, 6.0 W
200 - 253 VAC, 50/60 Hz ± 6%, 6.0 W
Code -02: 20 - 30 VDC, 6.0 W
Current Outputs:
Output 1: Process, Raw conductivity, or Temp.
Output 2: Process, Raw conductivity, or Temp.
Each output is galvanically isolated, 0-20 mA or
4-20 mA into 500 ohms maximum load at 24 Vdc
or 115/230 Vac or 500 ohms maximum load at
100/200 Vac.Output 1 includes digital signal
4-20 mA superimposed HART (Code -09 only).
EMI/RFI :EN61326
LVD (Code -01 only): EN61010-1
Ambient Temperature:
0 to 50°C (32 to 122°F)
NOTE: The analyzer is operable from
-20 to 60°C
(-4 to 140°F) with some degradation in display
performance.
Relative Humidity: 95%, non-condensing
Alarms: Relay 1 - Process, Interval, or Time
Proportional Control (code -20)
Relay 2 - Process, Interval, or Time Proportional
Control (code -20)
Relay 3 - Process, Interval, or Time Proportional
Control (code -20)
Relay 4 - Fault alarm
Each relay has a dedicated LED on the front panel.
Relay Contacts: Relays 1-3: Epoxy sealed form A
contacts, SPST, normally open.
Relay 4: Epoxy sealed form C, SPDT.
Resistive Inductive
28 Vdc 5.0 Amps 3.0 Amps
115 Vac 5.0 Amps 3.0 Amps
230 Vac 5.0 Amps 1.5 Amps
Weight/Shipping Weight: 1.1 kg/1.6 kg (2.5 lb/3.5 lb)
INSTRUMENT SPECIFICATIONS @ 25°C
Measurement Range: –15 to 200°C (5 to 392°F)
Contacting: 0-20,000 µS/cm
Toroidal: 0-2 S/cm
Accuracy of Analyzer: (Analyzer connected to simulated sensor input)
Contacting Sensors: ±0.5% of reading, ±.005 µS/cm
Inductive Sensors: ±1% of reading, 200 µS/cm to 2 S/cm, ±5 µS/cm
Repeatability: ±0.25% of reading
Stability: ±0.25% of output range/month, noncumulative
Ambient Temperature Coefficient: ± 0.01% of reading/°C
Temperature Compensation: -15 to 200°C (5 to 392°F) (automatic or manual)
Temperature Correction: High purity water (dilute sodium chloride), cation conductivity (dilute hydrochloric
acid), linear temperature coefficient (0.0 to 5.00%/°C), or none. High purity water and cation conductivity
temperature correction apply between 0 and 100°C. Linear temperature coefficient can be applied between
-5 and 200°C (23 to 392°F).