Rosemount 5081FG 2-Wire In Situ O2 Analyzer (550° to 1600°C) Manuals & Guides

Page 1

Reference Manual

00809-0100-4882, Rev AB
June 2019
Rosemount
TM
Two-Wire In-Situ Oxygen Analyzer (550 °C to 1600 °C)
Page 2
Page 3
Reference Manual
June 2019
Rosemount 5081FG
TOC-1
00809-0100-4882, Rev AB
SECTION i Introduction
SECTION 1 Description and Specifications
SECTION 2 Installation

Table of Contents

Preface .................................................................................................... iv
Definitions ................................................................................................ iv
Symbols ................................................................................................... iv
Technical Support Hotline: ................................................................. iv
Component Checklist ........................................................................... 1-1
System Overview .................................................................................. 1-3
Scope .............................................................................................. 1-3
System Description ......................................................................... 1-3
System Configuration ...................................................................... 1-4
System Features ............................................................................. 1-4
Handling the Analyzer ..................................................................... 1-6
System Considerations ................................................................... 1-6
Specifications ....................................................................................... 1-7
Pre-Installation ...................................................................................... 2-2
Inspect ............................................................................................. 2-2
Packing List ..................................................................................... 2-2
Mechanical Installation ......................................................................... 2-2
Locating Oxygen Probe ................................................................... 2-2
Installing Oxygen Probe .................................................................. 2-4
Locating Rosemount 5081 Transmitter ........................................... 2-7
Installing Rosemount 5081 Transmitter .......................................... 2-8
Electrical Installation ........................................................................... 2-11
General .......................................................................................... 2-11
Oxygen Probe Signal Connections ............................................... 2-12
Rosemount 5081 Transmitter 4-20 mA and Signal Connections . 2-13
Pneumatic Installation ........................................................................ 2-14
General .......................................................................................... 2-14
Reference Air Package ................................................................. 2-14
Instrument Air (Reference Air) ....................................................... 2-14
Calibration Gas .............................................................................. 2-15
SECTION 3 Startup and Operation
General ................................................................................................. 3-1
Power Up .............................................................................................. 3-1
Establishing Proper Calibration Gas Flow Rate ................................... 3-3
Operation .............................................................................................. 3-4
Overview ......................................................................................... 3-4
Display ............................................................................................. 3-4
Menu Tree ....................................................................................... 3-4
Navigation ....................................................................................... 3-6
Program Menu ...................................................................................... 3-7
Operator Adjustable Parameters ..................................................... 3-7
Code ................................................................................................ 3-8
Display Code ................................................................................... 3-9
Fault Val .......................................................................................... 3-9
Upper Range Val ........................................................................... 3-10
Page 4
Reference Manual
4882, Rev AB
June 2019
Rosemount 5081FG
TOC-2
00809-0100-
Cell T Hi ......................................................................................... 3-10
Reset Max Cell T ........................................................................... 3-11
Set O2 Filter Time .......................................................................... 3-11
Trim 4mA ....................................................................................... 3-12
Trim 20mA ..................................................................................... 3-13
Set Hi Bottle O Set Lo Bottle O Set O
Set Code ........................................................................................ 3-15
Diagnostics Menu ............................................................................... 3-16
Show Fault ..................................................................................... 3-16
T/C mV ........................................................................................... 3-17
O
Cell Impedance .............................................................................. 3-18
Current Slope ................................................................................. 3-18
Current Constant ........................................................................... 3-19
Previous Slope ............................................................................... 3-19
Previous Constant ......................................................................... 3-20
Max Cell T ...................................................................................... 3-20
SW Ver (SOFt) ............................................................................... 3-21
Unit Ser # (SEr) .............................................................................. 3-21
SW Build Number (bLdn) ............................................................... 3-21
SW Build Date (bd) ........................................................................ 3-21
Cal Check Menu .................................................................................. 3-21
In Manual ....................................................................................... 3-21
Accept High O Accept Low O
Slope .............................................................................................. 3-23
Constant ......................................................................................... 3-23
Tracking .............................................................................. 3-15
2
CELL mV .................................................................................. 3-17
2
.............................................................................. 3-14
2
............................................................................. 3-14
2
.............................................................................. 3-22
2
............................................................................... 3-22
2

SECTION 4 HART/AMS

SECTION 5 Troubleshooting
SECTION 6 Maintenance and Service
Overview ............................................................................................... 4-1
Field Communicator Signal Line Connections ...................................... 4-2
Field Communicator PC Connections ................................................... 4-4
Off-Line and On-Line Operations .......................................................... 4-4
HART/AMS Menu Tree ......................................................................... 4-4
Field Communicator Start Cal Check Method ...................................... 4-8
General ................................................................................................. 5-1
Probe Life .............................................................................................. 5-1
Fault Indications .................................................................................... 5-3
Identifying And Correcting Fault Indications ........................................
Calibration Passes, But Still Reads Incorrectly ..................................... 5-8
Probe Passes Calibration, O2 Still
Reads High ...................................................................................... 5-8
Overview ............................................................................................... 6-1
Electronics
Replacement ......................................................................................... 6-1
Display Board Replacement ............................................................ 6-1
Spare Board Stack Replacement .................................................... 6-3
Oxygen Probe Replacement ................................................................. 6-3
5-4
Page 5
Reference Manual
June 2019
Rosemount 5081FG
TOC-3
00809-0100-4882, Rev AB
SECTION 7 Replacement Parts
APPENDIX A Safety Data
Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
Safety Data Sheet for Ceramic Fiber Products . . . . . . . . . . . . . . . . A-3
APPENDIX B Return of Material
Returning Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
Page 6
Reference Manual
4882, Rev AB
June 2019
Rosemount 5081FG
TOC-4
00809-0100-
Page 7
Reference Manual
June 2019
Rosemount 5081FG
00809-0100-4882, Rev AB

ESSENTIAL INSTRUCTIONS

Oxygen Analyzers

READ THIS PAGE BEFORE PROCEEDING!

Emerson designs, manufactures and tests its products to meet many national and international standards. Because these inst ruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Emerson’s Rosemount products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidat i on.
Read all instructions prior to installing, operatin g, and servicing the product.
If you do not understand any of the instructions, contact your Emerson representative for clarification.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and maintenance of the product.
Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the proper electrical and
pressure sources.
To en sure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts specified by Emerson. Unauthorized parts and procedures can affect the product's performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protectiv e cov ers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
The information contained in this document is subje ct to change without notice.
If a Model 275/375 Universal HART® Communicator is used with this unit, the software within the Model 275/375 may require modification. If a software modification is required, please contact your local Emerson Service Group or National Response Center at 1-800-433-6076 or 1-888-433-6829.
Page 8
Reference Manual
4882, Rev AB
June 2019
Rosemount 5081FG
00809-0100-
WARNING
Physical access
Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of end users’ equipment. This could be intentional or unintentional and needs to be protected against.
Physical security is an important part of any security program and fundamental to protecting your system. Restrict physical access by unauthorized personnel to protect end users’ assets. This is true for all systems used within the facility.
Page 9
Reference Manual
June 2019
Rosemount 5081FG
00809-0100-4882, Rev AB

Section 1 Introduction

Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page i-iv

Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page i-iv
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page i-iv
PREFACE The purpose of this manual is to provide information concerning the
components, functions, installation and maintenance of the Rosemount 5081FG Two-Wire In-Situ Oxygen Analyzer (550° to 1600°C).
Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this module before operating it. Read this instruction manual c ompletely.
DEFINITIONS The following definitions apply to WARNINGS, CAUTIONS, and NOTES
found throughout this publication.
SYMBOLS
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in injury, death, or long-term health hazards of personnel.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.

NOTE

Highlights an essential operating procedure, co ndition, or statement.
:
EARTH (GROUND) TERMINAL
:
PROTECTIVE CONDUCT OR TERMINAL
:
RISK OF ELECTRICAL SHOCK
:
WARNING: REFER TO INSTRUCTION MANUAL

NOTE TO USERS

The number in the lower right corner of each illustration in this publication is a manual illustration number. It is not a part number, an d is not related to the illustration in any technical manner.
Page 10
Reference Manual
4882, Rev AB
June 2019
Rosemount 5081FG
00809-0100-

Technical Support Hotline:

For assistance with technical problems, please call the North American Response Center. The Response Center is staffed 24 hours a day, 7 days a week.
Phone: 1-800-654-RSMT (1-800-654-7768) Emerson may also be reached via the Internet
through email and the World Wide Web: Email: GAS.CSC@emerson.com
Page 11
Reference Manual
June 2019
Rosemount 5081FG
00809-0100-4882, Rev AB

Section 1 Description and Specifications

Component Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-1

System Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-3
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-7
COMPONENT CHECKLIST
A typical Rosemount Two-Wire In-Situ Oxygen Analyzer should contain the items shown in Figure 1-1. Record the part number, s erial number, and order number for each component of your system in the tabl e l ocated on the first page of this manual.
Also, use the product matrix in Table 1-1 at the end of thi s section to compare your order number against your unit. The first part of the matrix defines the model. The last part defines the various options and features of the Rosemount 5081FG Analyzer. Ensure the feature s and options specified by your order number are on or included with the unit.
Page 12
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
1-2
Figure 1-1. Typical System Package
00809-0100-
1.
Instruction Manual
2.
Rosemount 5081 Transmitter
3.
Oxygen Probe
4.
Adapter Plate with Mounting Hardware and Gasket (Optional)
5.
Infrared Remote Control (IRC)
6.
Reference Air Set (Optional)
7.
Field Communicator Package (Optional)
8.
Pipe Mounting Kit (Optional)
Page 13
Reference Manual
00809 May 2019
Rosemount 5081FG
1-3
-0100-4882, Rev AB
SYSTEM OVERVIEW Scope This Instruction Bulletin is designed to supply details needed to install, start
up, operate, and maintain the Rosemount Two-Wire In-Situ Oxygen Analyzer. The analyzer consists of an oxygen probe and Rosem ount 5081 transmitter. The signal conditioning electronics of the Rosemount 5081 transmitter outputs a 4-20 mA signal representing an O control (IRC) allows access to setup, calibration, and diagnostics. This same information, plus additional details, can be accessed with the HART field communicator or Asset Management Solutions (AM S) software.
value. An infrared remote
2
System Description The Rosemount Two-Wire In-Situ Oxygen Analyzer is designed to measure
the net concentration of oxygen in an industrial process; i.e., the oxygen remaining after all fuels have been oxidized. The oxygen probe is permanently positioned within an exhaust duct or stack and performs its task without the use of a sampling system. Rosemount 5081 transmitter is mounted remotely and conditions the oxygen pro be outputs.
The equipment measures oxygen percentage by reading the voltage developed across a heated electrochemical cel l, which consists of a small yttria stabilized, zirconia disc. Both sides of the disc are coated with porous metal electrodes. When operated at the proper tem perature, the millivolt output voltage of the cell is given by the following Nernst equation:
EMF = KT log10(P1/P2) + C
Where:
1.
P2 is the partial pressure of the oxygen in the measured gas on one side of the cell.
2.
P1 is the partial pressure of the oxygen in the reference air on the opposite side of the cell.
3.
T is the absolute temperature.
4.
C is the cell constant.
5.
K is an arithmetic constant.

NOTE

For best results, use clean, dry, instrument air (20. 95% oxygen) as the reference air.

NOTE

The probe uses a Type B thermocouple to measure th e cell temperature.
When the cell is at operating temperature and there are unequal oxygen concentrations across the cell, oxygen ions will t ravel from the high oxygen partial pressure side to the low oxygen partial pres sure side of the cell. The resulting logarithmic output voltage is approx im ately 50 mV per decade.
The output is proportional to the inverse logarithm of the oxygen concentration. Therefore, the output signal increases as the oxygen concentration of the sample gas decreases. This characteristic enables the Rosemount Two-Wire In-Situ Oxygen Analyzer to provide exceptional sensitivity at low oxygen concentrations.
Page 14
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
1-4
00809-0100-
The oxygen analyzer measures net oxygen concentration in the presence of all the products of combustion, including water vap or. Therefore, it may be considered an analysis on a “wet” basis. In compa rison with older methods, such as the portable apparatus, which provides an analysis on a “dry” gas basis, the “wet” analysis will, in general, indicate a lower percentage of oxygen. The difference will be proportional to the water content of the sampled gas stream.
System Configuration The equipment discussed in this manual consists of two major components:
the oxygen probe and the Rosemount 5081 transmit ter. Oxygen probes are available in three length options, providing In-Situ pene-
tration appropriate to the size of the stack or duct. The options on length are 20 in. (508 mm), 26 in. (660 mm), or 38 in. (965 mm).
Rosemount 5081 transmitter is a two-wire transmitter providing an isolated output, 4-20 mA, that is proportional to the measured oxygen concentration. A customer-supplied 24 VDC power source is required to simultaneously pro­vide power to the electronics and a 4-20 mA signal loop. The transmitt er accepts millivolt signals generated by the probe and produces the outputs to be used by other remotely connected devices. The ou t put is an isolated 4-20 mA linearized current.
System Features 1. The cell output voltage and sensitivity increase as the oxygen
concentration decreases.
2.
High process temperatures eliminate the need for external cell heating and increase cell accuracy.
3.
HART communication is standard. To use the HART capability, you must have either:
(a)
Field Communicator
(b)
Asset Management Solutions (AMS) software f or the PC
4.
Easy probe replacement due to the light-weight, com pact probe design.
5.
Remote location of the Rosemount 5081 transmitter removes the electronics from high temperature or corrosive environments.
6.
Power is supplied to the electronics through the 4-20 mA line for intrinsic safety (IS) purposes.
7.
Infrared remote control (IRC) allows interfacing without exposing the electronics.
Page 15
Reference Manual
00809 May 2019
Rosemount 5081FG
1-5
-0100-4882, Rev AB
8.
An operator can operate and diagnostically troubleshoot the Two-Wire In-Situ Oxygen Analyzer in one of two ways:
a.
Infrared Remote Control. The IRC allows access to faul t indication menus on the Rosemount 5081 transmitter LCD display. Calibration can be performed from the IRC keypad.
b.
Optional HART Interface (Figure 1-2). The Two-Wire In-Situ Oxygen Analyzer's 4-20 mA output line transmits an analog signal proportional to the oxygen level. The HART output is superimposed on the 4-20 mA output line. This information can be accessed through the following:
Field Communicator - The field communicator requires Device Description (DD) software specific to the Two-Wire In-Situ Oxygen Analyzer. The DD software will be supplied with many field communicators but can also be programmed int o existing units at most Fisher-Rosemount service offices. See Section 4: HART/AMS, for additional HART information.
Personal Computer (PC) - The use of a personal computer requires AMS software available from Fisher-Rosemount.
9.
Selected Distributed Control Systems - The use of distributed co nt rol systems requires input/output (I/O) hardwar e and A MS Security codes are provided to (by infrared remote control) prevent unintended changes to analyzers adjacent to the one being accessed.
10.
A calibration check procedure is provided to determine if the Rosemount Two-Wire In-Situ Oxygen Analyzer is correctly m easuring the net oxygen concentration in the industrial process.
Figure 1-2. Two-Wire In-Situ Oxygen Analyzer HART Connections and AMS Application
Page 16
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
1-6
00809-0100-
Handling the Analyzer The probe was specially packaged to prevent breakag e due to handling. Do
not remove the padding material from the probe unt il i m m ediately before installation.
It is important that printed circuit boards and integrated circuits are handled only when adequate antistatic precautions have been taken to prevent possible equipment damage.
The oxygen probe is designed for industrial applications. Treat each component of the system with care to avoid physical damage. Some probe components are made from ceramics, which are susceptible to shock when mishandled.
System Considerations Prior to installing your Rosemount Two-Wire In-Situ Oxygen Analyzer, make
sure you have all the components necessary to make the system installation. Ensure all the components are properly integrated to m ake t he system functional.
Figure 1-3. Typical System Installation
After verifying that you have all the components, select mounting locations and determine how each component will be placed in terms of available line voltage, ambient temperatures, environment al considerations, convenience, and serviceability. Figure 1-1 shows a typical sy st em package. A typical system installation is shown in Figure 1-3.
Page 17
Reference Manual
00809 May 2019
Rosemount 5081FG
1-7
-0100-4882, Rev AB

SPECIFICATIONS

A source of instrument air is optional at the oxygen probe for reference air use. Since the unit is equipped with an in place calibration feature, provisions can be made to permanently connect calibration gas b ottles to the oxygen probe.
If the calibration gas bottles will be permanently connected, a check valve is required next to the calibration fittings on the int egral electronics.
This check valve is to prevent breathing of the calibration gas line and subsequent flue gas condensation and corrosi on. The check valve is in addition to the stop valve in the calibration gas kit.

NOTE:

The electronics is rated Type 4X (IP65) and is capable of operat ion at temperatures up to 65°C (149°F).
Retain the packaging in which the Rosemount Tw o-Wire In-Situ Oxygen Analyzer arrived from the factory in case any components are to be shipped to another site. This packaging has been designe d to protect the product.
Transmitter
Net O2 Range 0 to 25% O System Accuracy ±1.5% of reading or 0.05% O2, whichever is greater System Response in Flue Gas Initial – less than 3 seconds,
Probe Lengths 20 in. (508 mm)
Process Temperature Limits: 550° to 1400°C (1022° to 2552°F)
Ambient Temperature Limits: -40° to 149°C (-40° to 300°F) Mounting and Mounting Position Vertical or horizontal Materials of Construction (Process Wetted Parts):
Inner Probe Outer Protection Tube
Probe Junction Box
T90 – less than 8 seconds
26 in. (660 mm) 38 in. (965 mm)
Operation to 1600°C (2912°F) with reduced cell life
Zirconia Alumina [1600°C (2912°F) limit] Inconel 600 [1000°C (1832°F) limit] Cast aluminum
2
Speed of Installation/Withdrawal 1 in. (25,4 mm) per minute Hazardous Area Certification See Section 8. Reference Air Requirement 100 m per minute (2,119 scfh) of clean, dry
Calibration Check Gas Fittings 1/4 in. tube fittings Cabling Two twisted pairs, shielded
Continued on Next Page
instrument air; 1/4 in. tube fittings
Page 18
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
1-8
Shipping Weight
10 lbs (4,5 kg)
Infrared Remote Control
Power Requirements
Three AAA atteries
Figure 1-4. Power Supply and Load Requirements
00809-0100-
Electronics
Electronics Enclosure TYPE 4X (IP65), weatherproof, and
corrosion-resistant Materials of Construction Low copper aluminum Ambient Temperature Limits -20° to 65°C (-4° to 149°F) Relative Humidity 95% with covers sealed Power Supply and Load Requirements See Figure 1-4 Inputs (from O
Output One 4-20 mA signal with superimposed digital HART
Hazardous Area Certification See Section 8. Power Transient Protection IEC 801-4
Hazardous Area Certification ATEX EEx ia IIC Class 1, Div 1, Group A, B, C, D
Probe) Two wires - O2 signal
2
Two wires - type B thermocouple
signal
Page 19
Reference Manual
00809 May 2019
Rosemount 5081FG
1-9
-0100-4882, Rev AB
Table 1-1. Product Matrix
5081FG High Temperature Oxygen Flue Gas Analyzer
Code Sensing Probe Type
1
20 in. (508 mm) probe, 1/4 in. tube fittings
2
26 in. (660 mm) probe, 1/4 in. tube fittings
3
38 in. (965 mm) probe, 1/4 in. tube fittings
Code Probe Outer Tube Material - Maximum Operating Temperature
1
Alumina - 2912°F (1600°C) maximum - 1.25 NPT mounting
2
Inconel Alloy - 1832°F (1000°C) maximum - 1.25 NPT mounting
Code Mounting Adapter- Stack Side
0
No adapter plate required uses 1.25 NPT ("0" must also be chosen under "Mounting Adapter" below)
1
New flanged installation - Square weld plate with studs (matches "Mounting Adapter" below)
2
Model 450 mounting ("4" must also be chosen under "Mounting Adapter" below)
3
Competitor's Mount ("5" must also be chosen under "Mounting Adapter" below)
Code Mounting Adapter - Probe Side
0
No adapter plate
1
ANSI 2 in. 150 lb flange to 1.25 NPT adapter (6 in. dia. flange, 4.75 in. BC with 4 x 0.75 in. dia. holes)
2
DIN to 1.25 NPT adapter (184 mm flange, 145 mm BC with 4 x 18 mm dia. holes)
3
JIS to 1.25 NPT adapter (155 mm flange, 130 mm BC with 4 x 13 mm dia. holes)
4
Model 450 to 1.25 NPT adapter
5
Competitor's mounting flange
Code Electronics & Housing TYPE 4X, IP65
H1 5081-G HART Electronics – ATEX/IECEx H2 5081-G HART Electronics - CSA H3 5081-G HART Electronics - FM
Code Housing Mounting
1
Surface or wall mounting
2
1/2 to 2 in. pipe mounting
Code Communications
1
No remote control
2
Infrared Remote Control (IRC) (LCD display through cover window
Code Calibration Accessories
1
No hardware
2
Calibration and reference air flowmeters and reference air pressure regulator
Code Armored Cable Length
00 No cable 11 20 ft (6 m) 12 40 ft (12 m) 13 60 ft (18 m) 14 80 ft (24 m) 15 100 ft (30 m) 16 150 ft (45 m) 17 200 ft (61 m) 18 300 ft (91 m) 19 400 ft (122 m) 20 500 ft (152 m)
5081FG 2 1 0 0 1 1 1 2 11 Example
Page 20
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
1-10
00809-0100-
Page 21
Reference Manual
00809 May 2019
Rosemount 5081FG
-0100-4882, Rev AB

Section 2 Installation

Pre-Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-2

Mechanical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-2
Electrical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-11
Pneumatic Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-14
Before installing this equipment, read the “Safety instructions for the wiring and installation of this apparatus” in Appendix A. Failure to follow safety instructions could result in serious injury or death.
Page 22
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
2-2
00809-0100-
PRE-INSTALLATION Inspect Carefully inspect the shipping container for any evi dence of damage. If the
container is damaged, notify the carrier immediately.
Packing List Confirm that all items shown on the packing list are present. Notify
Rosemount immediately if items are missing.
Before installing this equipment, read the “Safety instructions for the wiring and installation of this apparatus” in Appendix A. Failure to follow safety instructions could result in serious injury or death.

MECHANICAL INSTALLATION

Avoid installation locations near steam soot blowers.
Locating Oxygen Probe 1. The location of the oxygen probe in the stack or flue i s im portant for
maximum accuracy in the oxygen analyzing process. The pro be must be positioned so the gas it measures is representative of the process. Longer ducts may require several analyzers since the O to stratification. A point too near the wall of the duct, or the in-side radius
of a bend, may not provide a representative sample because of the very low flow conditions. The sensing point should be selected so the process gas temperature falls within a range of 55 0° to 1400°C (1022° to 2552°F). Figure 2-1 provides mechanical installation references.
2.
Check the flue or stack for holes and air leakage. The presence of this condition will substantially affect the accuracy of the oxygen reading. Therefore, either make the necessary repairs or install the probe upstream of any leakage.
3.
Ensure the area is clear of internal and external obstructions that will interfere with installation and maintenance access to the probe. Allow adequate clearance for probe removal (Figure 2-1).
can vary due
2
Page 23
Reference Manual
00809 May 2019
Rosemount 5081FG
2-3
-0100-4882, Rev AB
Figure 2-1. Probe Installation Details
Page 24
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
2-4

Installing Oxygen Probe

00809-0100-
The probe was specially packaged to prevent breakage due to handling. Do not remove the padding material from the probe until immediately before installation.
1.
Ensure all components are available to install the probe.

NOTE

Leave the probe inner protective cover in place until i nstal l ation. This is required to protect the ceramic cell during mov em ent.
2.
If using an optional adapter plate (Figure 2-2) or an optional mounting flange (Figure 2-3), weld or bolt the component onto the duct. The through hole in the stack or duct wall and refractory material must be 2 in. (50,8 mm) diameter, minimum.
3.
If the optional adapter plates are not used, a 2 in. NPT, schedule 40, pipe nipple (Figure 2-4) should be welded to the stack or duct wall. When a 2 in. NPT to 1.25 NPT adapter is threaded to the welded pipe nipple, the adapter provides the pipe threads needed f or the probe's process fitting.
4.
Where high particulate or slag is in the flue gas stream, it may be desirable to inset the probe in the refractory as shown in Figure 2-5. Use pipe couplings and nipples to adjust the probe insertion depth.
5.
Use high temperature material (alumina wool) to seal around the probe during insertion. This prevents hot gases from escaping or cold air from entering the stack or duct.
6.
Initially insert the probe to a depth of 3 in. (76,2 mm) or 1/2 the depth of the stack or duct refractory, whichever is greater.
After initial insertion, do not insert the probe at a rate exceeding 1 in. per minute (25.4 mm per minute) or damage to the probe may result due to thermal shock.
7.
After initial insertion, insert the probe at a rate of 1 i n. (25,4 mm) per minute until the probe is fully inserted.
8.
Install anti-seize compound on the pipe threads and scre w t he probe into the process flange or adapter.
9.
If insulation was removed to access the duct work for probe mounting, make sure the insulation is replaced. See Figure 2-4.
If the ducts will be washed down during outage, MAKE SURE to power down the probes and remove them from the wash area.
Page 25
Reference Manual
00809 May 2019
Rosemount 5081FG
2-5
-0100-4882, Rev AB
Figure 2-2. Optional Adapter Plate
Figure 2-3. Optional Probe Mounting Flange
Page 26
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
2-6
Figure 2-4. Horizontal Probe Installation
00809-0100-
Page 27
Reference Manual
00809 May 2019
Rosemount 5081FG
2-7
-0100-4882, Rev AB
Figure 2-5. Adjusting Probe Insertion Depth

Locating Rosemount 5081 Transmitter

1.
Ensure the Rosemount 5081 transmitter is easily acc essible for maintenance and service and for using the infrared remote control (if applicable).
Do not allow the temperature of the Rosemount 5081 transmitter exceed 65°C (149°F) or damage to the unit may result.
Page 28
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
2-8
2.

Installing Rosemount 5081 Transmitter

3.
1.
2.
Figure 2-6. Flat Surface Mounting Dimensions
00809-0100-
The ambient temperature of the transmitter housing must not exceed 65°C (149°F). Locate the electronics in an area where temperature extremes, vibration, and electromagnetic and radi o frequency interference are minimal.
Locate the Rosemount 5081 transmitter within 150 ft (45, 7 m) of the oxygen probe due to wiring and signal considerations.
Ensure all components are available to install the Rosemount 5081 transmitter.
Choose a method or location to mount the transmitter. a.
Flat Surface Mounting. The transmitter may be mounted on a flat surface using the threaded mounting holes located on the bottom of the transmitter housing. Refer to Figure 2-6 for inst al lation references.
b.
Pipe Mounting. An optional pipe mounting bracket i s available for this type of installation. Refer to Figure 2-7 for installation references.
Page 29
Reference Manual
00809 May 2019
Rosemount 5081FG
2-9
-0100-4882, Rev AB
Figure 2-7. Pipe Mounting Dimensions
Page 30
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
2-10
Figure 2-8. Display Positioning Assembly
00809-0100-
3.
For correct viewing orientation, the display may be changed 90 degrees, using the following procedure:
a.
Refer to Figure 2-8. Loosen the cover lock screw until the cover lock is disengaged from the knurled surface on the thread ed circuit end cap.
b.
Remove the circuit end cap.
c.
Remove the three screws retaining the display bo ard in place.
d.
Lift and rotate the display board 90 degrees either way.
e.
Reposition the display board on the standoffs. I nst al l and tighten all three screws.
f.
Install the circuit end cap and tighten the cover lo ck screw to secure the cover lock in place.
Page 31
Reference Manual
00809 May 2019
Rosemount 5081FG
2-11
-0100-4882, Rev AB

ELECTRICAL INSTALLATION

All wiring must conform to local and national codes.
For intrinsically safe applications, refer to the installation drawings on Emerson.com for the model 5081FG.
Disconnect and lock out power before connecting the unit to the power supply.
Install all protective equipment covers and safety ground leads after installation. Failure to install covers and ground leads could result in serious injury or death.
To meet the Safety Requirements of IEC 61010 (EC requirement), and ensure safe operation of this equipment, connection to the main electrical power supply must be made through a circuit breaker (min 10 A) which will disconnect all current-carrying conductors during a fault situation. This circuit breaker should also include a mechanically operated isolating switch. If not, then another external means of disconnecting the supply from the equipment should be located close by. Circuit breakers or switches must comply with a recognized standard such as IEC 947.
General The power supply and signal wiring should be shielded. Also, make sure the
signal wiring is grounded at the Rosemount 5081 transmitter end only. Do not ground the signal loop at more than one point. Twisted pai rs are recommended. Ground the transmitter housing to an earth ground to prevent unwanted electromagnetic interference ( EMI) or radio frequency interference (RFI).

NOTE

For optimum EMI/RFI immunity, shield the 4-20 mA current loop cable and enclose in an earth grounded metal conduit.

NOTE

Never run signal or sensor wiring in the same conduit, or open tray, with power cables. Keep signal or sensor wiring at least 12 i n. (0,3 m) away from other electrical equipment and 6.5 ft (2 m) from heavy elect rical equipment.
Page 32
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
2-12

Oxygen Probe Signal Connections

Figure 2-9. Oxygen Probe Terminal Block
00809-0100-
Moisture accumulation in the transmitter housing can affect its performance and may void its warranty.
It is necessary to prevent moisture from entering the Rosemount 5081 transmitter housing. The use of weather-tight cabl e glands is required. If conduit is used, plug and seal connections on the transmitter housing to prevent moisture accumulation in the terminal side of the housing.
1.
Two signals represent the O2 value and the cell temperature. The probe provides these values to the Rosemount 5081 transmitter for processing
and signal conditioning.
2.
Wiring connections for the probe are shown in Figure 2-9.
Page 33
Reference Manual
00809 May 2019
Rosemount 5081FG
2-13
-0100-4882, Rev AB

Rosemount 5081 Transmitter 4-20 mA and Signal Connections

Figure 2-10. Transmitter Terminal Block
1.
A 4-20 mA signal represents the O2 value. Superimposed on the 4-20 mA signal is HART information that is accessible through a
HART field communicator or AMS software.
2.
Two signals representing the O2 value and the cell temperature are supplied to the Rosemount 5081 transmitter from the oxygen probe.
3.
Wiring connections for the Rosemount 5081 tran smi t ter are shown in Figure 2-10.

NOTE

The ground arrangement shown in Figure 2-10 lim its the amount of noise introduced into the electronics.
4.
Connect wire shields to terminal 1. Connect earth ground as shown.
Page 34
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
2-14

PNEUMATIC INSTALLATION

00809-0100-
General Reference air is required for O
required during a calibration check. Refer to Figure 2-11 for the gas connections on the oxygen probe.
calculation, and calibration check gas is
2
Reference Air Package After the oxygen probe is installed, connect the reference air set. Install the
reference air set according to Figure 2-11.

Instrument Air (Reference Air)

Figure 2-11. Air Set, Plant Air Connections
Instrument air is required for reference. Refer to the reference air schematic, Figure 2-12. Use 10 psig (68,95 kPa gage) minimum, 225 psig (1551,38 kPa gage) at 0.2 scfh (100 m /min.); less than 40 parts-per-million total hydrocarbons. Regulator outlet pressure shoul d be set at 5 psi (35 kPa).
Figure 2-12. Reference Air Schematic
Page 35
Reference Manual
00809 May 2019
Rosemount 5081FG
2-15
-0100-4882, Rev AB
Calibration Gas Two calibration check gas concentrations are used with the Two-Wire In-Situ
Oxygen Analyzer: Low Gas - 0.4% O balance in nitrogen.
and High Gas - 8% O2, each with the
2
Do not use 100% nitrogen as a low gas (zero gas). It is suggested that gas for the low (zero)
Figure 2-13. Gas Connections at Oxygen Probe
be between 0.4% and 2.0% O than 40 parts per million. Failure to use proper gases will result in erroneous readings.
Do not use 100% nitrogen for the low (0%) check gas. See Figure 2-13 for the probe connections. Set both calibration check gases at the same flow rate: 5 scfh (2,5 /min).
. Do not use gases with hydrocarbon concentrations of more
2
Page 36
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
2-16
00809-0100-
Page 37
Reference Manual
00809 May 2019
Rosemount 5081FG
-0100-4882, Rev AB

Section 3 Startup and Operation

General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-1

Power Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-1
Establishing Proper Calibration Gas Flow Rate . . . . . . . page 3-3
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-4
Program Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-7
Diagnostics Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-16
Cal Check Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-21
GENERAL
Install all protective equipment covers and safety ground leads before equipment startup. Failure to install covers and ground leads could result in serious injury or death.
Verify Mechanical Installation
Ensure the Two-Wire In-Situ Oxygen Analy zer is installed correctly. See Mechanical Installation in Section 2: Installation for mechanical installation information.

Verify Terminal Block Wiring

Ensure the wiring of both the oxygen probe terminal block and Rosemount 5081 transmitter terminal block is correct. Refer to Electrical Installation in Section 2: Installation for electrical installation and wiring information.
POWER UP General
The Two-Wire In-Situ Oxygen Analyzer displays the current oxygen reading on the LCD face of the Rosemount 5081 transmitter. The O
cell temperature, and 4-20 mA output current are displayed as shown in Figure 3-1. This and other information may also be accessed using HART/AMS.

Startup Display

When the probe is first inserted into the stack, some time is required until the minimum operating temperature [550°C (1022°F)] is reached. Some time is also required for the electronics to reach an operating state. Therefore, when the unit is first powered up, a faulted operation display as shown in Figure 3-2 may be displayed by the transmitter until the probe operating temperature is reached and the electronics are working properly (approximately 5 minutes).
concentration,
2
Page 38
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
3-2
Figure 3-1. Normal Operation Display
00809-0100-

Operating Display

After the probe has reached operating temperature s, the Rosemount 5081 transmitter display should look similar to Figure 3-1. The display will now track the O
concentration, cell temperature, and 4-20 mA output current.
2
Figure 3-2. Faulted Operation Display
Page 39
Reference Manual
00809 May 2019
Rosemount 5081FG
3-3
-0100-4882, Rev AB
ESTABLISHING PROPER CALIBRATION GAS FLOW RATE
The calibration gas flow must be enough to ensure that no combustion flue gases mix with the calibration check gases and only clean, good calibration check gas surrounds the cell without expending excess gas (Figure 3-3). Monitor the O calibration check gas flow rate as follows:

NOTE

Only set the calibration check gas flow rate at startup. It is not necessary to perform this procedure for each calibration check.
1.
Adjust the calibration check gas flow to 5 scfh (2,5 /min. ) t o ensure the cell is surrounded by calibration check gas. Due t o the cooling effect of the gas, the cell temperature will decrease slightly, causing the O
concentration to drop. Once the electronics compensates for this effect, the O
2
2.
Next, slowly reduce the calibration check gas fl ow until the O concentration changes, which indicates that the cali bration check and flue gases are mixing. Increase the flow rate until t hi s effect is eliminated.
Figure 3-3. Proper Calibration Check Gas Flow Rate
concentration using an IRC or a field communicator. Set the
2
concentration will stabilize.
2
2
Page 40
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
3-4
00809-0100-
OPERATION Overview This section explains the operator controls and displays of the Two-Wire
In-Situ Oxygen Analyzer. The use of the Infrared Remote Control (IRC) and the Rosemount 5081 transmitter Liquid Crystal Display (LCD) are described in detail.
HART/AMS operation is not covered here. Refer to Section 4: HART/AMS.
Display The LCD on the circuit end of the Rosemount 5081 transmitter displays O
concentration, cell temperature, and 4-20 mA output current during normal operation (Figure 3-4). The LCD will also display fault conditions when they occur. To interact with the transmitter, use the IRC and navigate through a series of menus displayed on the LCD.
Menu Tree The screens that can be displayed are shown in the menu tree of Figure 3-5.
These screens are displayed on the LCD and are acce ss ed using the IRC keypad.
2
Figure 3-4. Normal Operation Display
\
Page 41
Reference Manual
00809 May 2019
Rosemount 5081FG
3-5
-0100-4882, Rev AB
Figure 3-5. Transmitter Menu Tree
Page 42
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
3-6
00809-0100-
Navigation The IRC in Figure 3-6 is used to interact with the Rosemount 5081 transmitter
and navigate through the screens on the LCD.
1.
Hold the IRC within 6 ft (1,8 m) of the Rosemount 5081 transmitter and within 15 degrees from the centerline of the transmit t er LCD. The amount of ambient light may also affect IRC performance:

NOTE

The LCD may react slowly to IRC commands. Allow suf ficient time between key presses to avoid undesired or repeated commands from accumulating in the command queue.
2.
Use the keys on the IRC to navigate through the menu screens. Refer to Figure 3-6. General usage is as follows:
a.
RESET. Returns to the PROCESS DISPLAY screen at the top of the menu tree. Any non-entered number in the exited state will be ignored, and the previous data will be used.
b.
HOLD. Not used.
c.
Left/Right Arrow. Moves left and right among editable digits on the display.
d.
Up/Down Arrow. Increases or decreases the value of the cu rr ently selected digit on the display.
e.
CAL. Accesses the CALCHECK MENU branch of the menu tree. Only works from the PROCESS DISPLAY screen.
Figure 3-6. Infrared Remote Control (IRC)
Page 43
Reference Manual
00809 May 2019
Rosemount 5081FG
3-7
CODE
Use to enter the access code for this analyzer; select 555
000 - 999 (excluding 000 and 555)
000 (no code)
to display designated analyzer code
FAULT VAL
Use to designate a 4-20mA value that when displayed
3.8mA - 24mA
3.6mA
will indicate an analyzer-faulted condition
UPPER RANGE VAL
Use to set upper O2% limit equivalent to 20mA output
2.0% - 25.0%
25% CELL T HI
Use to set upper cell temperature for no-fault condition
650°C - 1600°C
1600°C
SET O2 FILTER TIME
Use to adjust analyzer response time to changing O2%
0 - 300 seconds
0 seconds
SET HI BOTTLE O2
Use to define actual O2% of high calibration check gas
8%
SET LO BOTTLE O2
Use to define actual O2% of low calibration check gas
2%
SET CODE
Use to set security code for this analyzer
000 - 999 (excluding 000 and 555)
000 (no code)
-0100-4882, Rev AB
f.
PROG. Accesses the PROGRAM MENU branch of the menu tree. Only works from the PROCESS DISPLAY screen.
g.
DIAG. Accesses the DIAGNOSTICS MENU branch of the menu tree. Only works from the PROCESS DISPLAY screen.
h.
ENTER. Initiates the editing process and causes t he most significant digit of the edited item to start flashing. Also processes the entry so the previous value updates to the new value ent ered using the arrow keys. Failure to press ENTER before exiti ng a screen will cancel the input value and revert to the previous value.
i.
NEXT. Accesses the next user screen a s shown in the menu tree. Any non-entered number in the exited state will be i gnored, and the previous data will be used.
j.
EXIT. Exits from sub-branches of the menu tree where an exit option is explicitly shown. Otherwise, returns to the PROCESS DISPLAY screen at the top of the menu tree. Any non-entered num ber in the exited state will be ignored, and the previous data will be used.
PROGRAM MENU The PROGRAM MENU branch of the menu tree allows you to program and
edit some process parameters, faults, out put s, and security codes. To access this branch of the menu tree, press the PROG key on the IRC when in the PROCESS DISPLAY screen (Normal or Faulted). If secu rity is enabled, you must enter the analyzer code to gain further acces s t o the screens in this branch. Each screen in this branch is accessed sequentially using the NEXT

Operator Adjustable Parameters

Table 3-1. Adjusatble Parameters
Parameter Function Range Factory Default
key. Refer to Figure 3-5 during the following menu and screen descriptions.

NOTE

To edit a screen value, press ENTER to access the data field. Use the left and right arrow keys to move among the digits in the data field. Not e that the editable position will be flashing. To change the value of a digit, use the up and down arrow keys to increase or decrease the val ue. When finished editing, press ENTER to accept the value. To go to the next screen in the menu, press NEXT.
Figure 3-1 lists the range and default value of operat or-adjustable variables used by the Rosemount 5081 transmitter. Each of these variables may be adjusted using the PROGRAM MENU screens.
(adjust the slope of the analyzer)
Page 44
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
3-8
00809-0100-
Code Refer to Figure 3-7. After pressing the PROG key, this screen will display if
security is enabled (see Set Code). Use this screen to identify a specif ic analyzer in a process to prevent accessing an adjacent analyzer when using the IRC.
Press ENTER to begin editing. At this point, you can either specify the analyzer by its access code or view its code if it i s unk nown.
1.
To gain f urt her access to the screens in the PROGRAM M ENU branch, enter the correct three-digit analyzer access code using the arrow keys and press ENTER. If security is disabled, this screen does not appear and the system displays the FAULT VAL screen.
2.
If the analyzer access code is un-known, enter 555 and press ENTER to access the DISPLAY CODE screen. In t hat screen you will be able to view the analyzer access code.
Figure 3-7. Code
Page 45
Reference Manual
00809 May 2019
Rosemount 5081FG
3-9
-0100-4882, Rev AB
Display Code Refer to Figure 3-8. This screen is accessible from the CODE screen by
entering 555 and pressing ENTER. The DISPLAY CODE screen identifies the analyzer access code so you can return to the CODE screen and enter the code as described in Code. To return to the CODE screen, press NEXT.
Fault Val Refer to Figure 3-9. Use this screen to set the value t hat the 4-20 mA output
will drive to and display during a fault condition. Press ENTER t o begin editing. Use the arrow keys to enter a fault value. The f aul t value can be between 3.8 and 24 mA. Then, press ENTER to accept the value. Pressing NEXT displays the UPPER RANGE VAL screen. Refer to Section 5: Troubleshooting, for the actual fault conditions.
Figure 3-8. Display Code
Figure 3-9. Fault Val
Page 46
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
3-10
00809-0100-
Upper Range Val Refer to Figure 3-10. Use this screen to set the value of the upper range limit.
This value is the maximum limit of the O used to scale the 4-20 mA output. Press ENTER to begin editing. Use the
arrow keys to select and change the value. The upper range value can be between 0 and 25%. Then, press ENTER to accept the value. Pressing NEXT displays the CELL T HI screen.
concentration measurement and is
2
Cell T Hi Refer to Figure 3-11. Use this screen to set the value of the upper cell
temperature fault condition. This value is the m aximum allowed cell temperature before a fault condition is indicated. P ress ENTER to begin editing. Use the arrow keys to select and change the value. The value must be between 550°C and 1600°C. Press ENTER to accept the value. Pressing NEXT displays the RESET MAX CELL T screen.
Figure 3-10. Upper Range Val
Figure 3-11. Cell T Hi
Page 47
Reference Manual
00809 May 2019
Rosemount 5081FG
3-13
-0100-4882, Rev AB
Reset Max Cell T Refer to Figure 3-12. The transmitter tracks the maximum cell temperature
obtained. Use this screen to reset the maximum cell temperature attained value to the current cell temperature. Press ENTER to begin editing. Use the arrow keys to select and change the value (Y/N). Then, press ENTER to accept the value. Pressing NEXT displays the SET O
FILTER TIME screen.
2
Set O2 Filter Time Refer to Figure 3-13. In some applications it is beneficial to dampen the raw
signal coming from the cell. Use this screen to enter the am ount of time it
O
2
will take the O editing. Use the arrow keys to select and change the sc reen value to the O filter value (in seconds). Enter a value between 0 and 300 seconds and press ENTER to accept the value. Press NEXT to access the TRIM 4 mA? screen.
Figure 3-12. Reset Max Cell T
to reach 90% of the new reading. Press ENTER to begin
2
2
Figure 3-13. Set O2 Filter Time
Page 48
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
3-12
00809-0100-
Trim 4mA Refer to Figure 3-14. Use this screen to trim the 4 mA value of the 4-20 mA
output.

NOTE

Before trimming the 4 mA value you must break the l oop to add the ammeter. Power down the unit, connect the ammeter in serie s wit h Rosemount 5081 transmitter terminals 15(-) and 16(+), power up t he unit, and return to the TRIM 4 mA screen.
Press ENTER to begin editing. Use the arrow keys to select and change the screen value to the value displayed on the installed ammeter. Press ENTER to accept the value. After the value is entered, the unit cali brates itself to ensure it outputs 4 mA. Both the display and the amm eter will display 4 mA. Pressing EXIT returns to the initial TRIM 4 mA screen, and pressing NEXT displays the TRIM 20 mA? screen.
Figure 3-14. Trim 4mA
Page 49
Reference Manual
00809 May 2019
Rosemount 5081FG
3-13
-0100-4882, Rev AB
Trim 20mA Refer to Figure 3-15. Use this screen to trim the 20 mA value of the 4-20 mA
output.

NOTE

Before trimming the 20 mA value you must break the l oop to add the ammeter. Power down the unit, connect the amm eter in series with Rosemount 5081 transmitter terminals 15(-) and 16(+), power up the unit, and return to the TRIM 20 mA screen.
Figure 3-15. Trim 20mA
Press ENTER to begin editing. Use the arrow keys to select and change the screen value to the value displayed on the installed a m m eter. Press ENTER to accept the value. After the value is entered, the unit cali brates itself to ensure it outputs 20 mA. Both the display and the ammeter will display 20 mA. Pressing EXIT returns to the initial TRIM 20 mA sc reen, and pressing NEXT displays the SET HI BOTTLE O
screen.
2
Page 50
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
3-14
00809-0100-
Set Hi Bottle O
Set Lo Bottle O
2
2
Figure 3-16. Set Hi Bottle O
Refer to Figure 3-16. Use this screen to identify, wit hi n the electronics, the percentage of O begin editing. Use the arrow keys to select and change the screen value to
the O the value.
Refer to Figure 3-17. Use this screen to identify, wit hi n the electronics, the percentage of O begin editing. Use the arrow keys to select and change the screen value to
the O the value. Press NEXT to display the SET O
2
percentage of the high calibration check gas. Pres s E NTER to accept
2
percentage of the low calibration check gas. Press E NTER to accept
2
used as the high calibration check gas. Press E NTER to
2
used as the low calibration check gas. Press ENTER to
2
TRACKING screen.
2
Figure 3-17. Set Lo Bottle O
2
Page 51
Reference Manual
00809 May 2019
Rosemount 5081FG
3-15
-0100-4882, Rev AB
Set O2 Tracking Refer to Figure 3-18. Use this screen to permit the 4-20 mA line to track the
value during a calibration check. Press ENTER to begin editing. Use the
O
2
arrow keys to select Y or N. Entering Y (yes) will allow the 4-20 mA line to track the O
value steady during the calibration check. Press ENTER to accept the
O
2
value. Press NEXT to display the SET CODE screen.
value during the calibration check. Entering N (n o) will hold the
2
Set Code Refer to Figure 3-19. Use this screen to set the security code for the
Rosemount 5081 transmitter. Press ENTER to begin editing. Use the arrow keys to select and change the value. Select any value between 000 and 999, excluding 000 and 555. Code 000 indicates that no c ode is set. Code 555 accesses the DISPLAY CODE screen. Press ENTER to accept the value. Pressing NEXT returns to the FAULT VAL screen at the beginning of the
Figure 3-18. Set O2 Tracking
PROGRAM MENU.
Page 52
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
3-16
Figure 3-19. Set Code
00809-0100-
DIAGNOSTICS MENU The DIAGNOSTICS MENU branch of the menu tree allows you to examine
outputs, current faults, and unit information. None of the items in the DIAGNOSTICS MENU are editable. This branch of the menu tree may be accessed by pressing DIAG on the IRC when in the PRO CE SS DISPLAY screen (Normal or Faulted). Each screen in thi s branch is accessed sequentially by pressing NEXT. Refer to the menu in F i gure 3-5 when reviewing the following menu and screens.
Show Fault Refer to Figure 3-20. After pressing DIAG, this screen displays. Pressing
ENTER accesses a screen displaying the current faul t (if any). If more than one fault exists, and you are in the FAULT screen, press NEXT to go to the next fault. Information on the fault screens can be fou nd i n Section 5: Troubleshooting. Press EXIT to return from this fault sub-menu and press NEXT to access the T/C mV screen.
Page 53
Reference Manual
00809 May 2019
Rosemount 5081FG
3-17
-0100-4882, Rev AB
Figure 3-20. Show Fault
Page 54
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
3-18
00809-0100-
T/C mV Refer to Figure 3-21. Use this screen to examine t he cell thermocouple mV
output. Three decimal places are displayed. Pressing NEXT accesses the O CELL mV screen.
O2 CELL mV Refer to Figure 3-22. Use this screen to examine the O
Pressing NEXT accesses the CELL IMPEDANCE screen.
CELL mV output.
2
Figure 3-21. T/C mV
Figure 3-22. O
CELL mV
2
2
Page 55
Reference Manual
00809 May 2019
Rosemount 5081FG
3-19
-0100-4882, Rev AB
Cell Impedance Refer to Figure 3-23. Use this screen to examine the O
status. GOOD indicates the cell is operating normally. WARN indicates the cell has degraded but is still operational. HI indi cates that the cell has degraded but is still operational; however, fail ure will occur soon. Pressing NEXT accesses the CURRENT SLOPE screen.

NOTE

Temperature influences cell impedance. Wait until the cell is at operating temperature before checking cell impedance. I f checked before the cell reaches 550°C (1022°F), this screen displays a fail indication.
cell impedance
2
Current Slope Refer to Figure 3-24. Use this screen to examine the slope calculated from
the most recent calibration check. The slope is the amount of cell voltage generated for a given O
over the life of the probe. Tracking the slope will indic ate if the probe is degrading. Press NEXT to access the CURRENT CO NS TANT screen.
Figure 3-23. Cell Impedance
value. For each calibration check, record the slope
2
Figure 3-24. Current Slope
Page 56
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
3-20
00809-0100-
Current Constant Refer to Figure 3-25. Use this screen to examine the cell zero constant
calculated from the most recent calibration check. T he constant represents the voltage generated by the cell when no difference exists between the amount of O access the PREVIOUS SLOPE screen.
on the reference and process sides of the cell. Press NEXT to
2
Previous Slope Refer to Figure 3-26. Use this screen to examine the slope value stored from
the second to last calibration check. The slope is the amount of cell voltage generated for a given O
over the life of the probe. Tracking the slope will indicate if the probe is degrading. Press NEXT to access the PREVIOUS CO NST ANT screen.
Figure 3-25. Current Constant
value. For each calibration check, record the slope
2
Figure 3-26. Previous Slope
Page 57
Reference Manual
00809 May 2019
Rosemount 5081FG
3-21
-0100-4882, Rev AB
Previous Constant Refer to Figure 3-27. Use this screen to examine the cell zero constant stored
from the second to last calibration check. The constant represents the voltage generated by the cell when no difference exists between the amount of O
the reference and process sides of the cell. Press NEXT to access the MAX CELL T screen.
Max Cell T Refer to Figure 3-28. Use this screen to examine the maximum temperature
attained by the O2 cell. This value can be reset under the PROGRAM MENU. Pressing NEXT accesses the SW VER screen.
Figure 3-27. Previous Constant
Figure 3-28. Max Cell T
on
2
Page 58
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
3-22
00809-0100-
SW Ver (SOFt) Use this screen to see the software version number f or the Rosemount 5081
transmitter. Pressing NEXT accesses the U NI T SER # screen.
Unit Ser # (SEr) Use this screen to see the unit serial number for the Rosemo unt 5081
transmitter. Pressing NEXT accesses the SW BUILD NUMBER screen.
SW Build Number (bLdn) Use this screen to see the software build number for the Rosemount 5081
transmitter. Pressing NEXT accesses the SW BUILD DATE screen.
SW Build Date (bd) Use this screen to see the software build date for the Rosemount 5081
transmitter. Pressing NEXT returns to the beginni ng of the DIAGNOSTICS MENU branch (the SHOW FAULT screen).
CAL CHECK MENU The CALCHECK MENU branch of the menu tree (Figure 3-5) allows you to
perform a calibration check of the analyzer. Bef ore performing a calibration check, ensure the high calibration check gas and low calibration check gas O
percentages are entered into the electronics via the PROGRAM MENU. To set these values, refer to Set Hi Bottle O2 and Set Lo Bottle O2.
Once these values are set, access the CAL-CHECK MENU branch by pressing CAL on the IRC when in the PROCESS DISPLAY screen (Normal or Faulted). Each screen in this branch identifies a pro cess step in the calibration check procedure. The first screen in t he sequence is the IN MANUAL? screen.
2

In Manual?

Figure 3-29. In Manual?
Failure to remove the analyzer from automatic control loops prior to performing this procedure may result in a dangerous operating condition.
Refer to Figure 3-29. If the O control loops the loop must be placed in manual to begin a calibration check.
Once the analyzer is removed from any automatic control loops press ENTER to edit the screen. Use the arrow keys to select Y (yes); press ENTER to start a calibration check and to display the ACCEPT HIGH O
output value is used in any automatic process
2
screen.
2
Page 59
Reference Manual
00809 May 2019
Rosemount 5081FG
3-23
-0100-4882, Rev AB
Accept High O
Accept Low O
2
2
Figure 3-30. Accept High O
Refer to Figure 3-30. After pressing ENTER to begin the calibration check, the high calibration check gas starts to flow. Aft er waiting approximately three minutes for the displayed O
value to settle, press NEXT to accept the high
2
calibration check gas reading and apply the low calibration check gas. The next screen to display is the ACCEPT LOW O
screen.
2
Refer to Figure 3-31. Once the low calibration check gas is applied, wait approximately three minutes for the displayed O
value to settle. Once the
2
value settles, press NEXT to accept the reading and to display the SLOPE screen.
2
Figure 3-31. Accept Low O
2
Page 60
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
3-24
00809-0100-
Slope Refer to Figure 3-32. Use this screen to examine the slope calculated from
current calibration check. The slope is the amount of cell voltage generated for a given O life of the probe. Tracking the slope will indicate if t he probe is degrading. Press NEXT to access the CONSTANT screen.
Constant Refer to Figure 3-33. Use this screen to examine the cell zero constant
calculated from the current calibration check. The constant represents the voltage generated by the cell when no difference ex i sts between the amount
on the reference and process sides of the cell. Note this value for
of O
2
comparison against future calibration checks. P ress RESET or EXIT to return to the PROCESS DISPLAY screen.
Figure 3-32. Slope
value. After each calibration check, record the slope over the
2
Figure 3-33. Constant
Page 61
Reference Manual
00809 May 2019
Rosemount 5081FG
-0100-4882, Rev AB

Section 4 HART/AMS

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4-1

Field Communicator Signal Line Connections . . . . . . . . page 4-2
Field Communicator PC Connections . . . . . . . . . . . . . . . page 4-4
Off-Line and On-Line Operations . . . . . . . . . . . . . . . . . . . page 4-4
HART/AMS Menu Tree . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4-4
Field Communicator Start Cal Check Method . . . . . . . . . page 4-8
OVERVIEW The HART field communicator is a hand-held communications interface
device. It provides a common communications link to all microproces­sor-based instruments that are HART compatible. The field communicator has a liquid crystal display (LCD) and keypad. A pocket-sized manual, included with the field communicator, details the specific functions of the keypad keys.
To interface with the Rosemount 5081FG Analyzer, the field communicator requires a termination point along the 4-20 mA current loop and a minimum load resistance of 250 ohms between the field com m uni cator and the power supply.
The field communicator accomplishes its task using a frequency shift keying (FSK) technique. With the use of FSK, high-frequency digital communication signals are superimposed on the analyzer's 4-20 mA curre nt loop. The field communicator does not disturb the 4-20 mA signal, since no net energy is added to the loop.
The field communicator may be interfaced with a personal computer (PC), providing that special software has been installed. To connect the field com­municator to a PC, an interface adapter is required. Refer to the proper field communicator documentation regarding the P C interface option.
Page 62
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
4-2
00809-0100-

FIELD COMMUNICATOR SIGNAL LINE CONNECTIONS

The field communicator can connect to the analy zer 's analog output signal line at any wiring termination in the 4-20 mA current loop. There are two meth­ods of connecting the field communicator to the signal line. For applications in which the signal line has a load resistance of 250 ohms or more, refer to method 1. For applications in which the signal line load resistance is less than 250 ohms, refer to method 2.

Method 1, For Load Resistance 250 Ohms

Refer to Figure 4-1 and the following steps to connect the field communicator to a signal line < 250 ohms or more of load resistance.
Explosions can result in death or serious injury. Do not make connections to the field communicator's serial port, 4-20 mV signal line, or NiCad recharger jack in an explosive atmosphere.
Using the supplied lead set, connect the field comm unicator in parallel with to the Rosemount 5081FG Analyzer. Use any wiring termination points in the analog output 4-20 mA signal line.
Figure 4-1. Signal Line Connections, 250 Ohms Load Resistance
Page 63
Reference Manual
00809 May 2019
Rosemount 5081FG
4-3
-0100-4882, Rev AB

Method 2, For Load Resistance < 250 ohms

Refer to Figure 4-2 and the following steps to connect the field communicator to a signal line with < 250 ohms load resistance.
Explosions can result in death or serious injury. Do not make connections to the field communicator's serial port, 4-20 mA signal line, or NiCad recharger jack in an explosive atmosphere.
1.
At a convenient point break the analog output 4-20 mA signal line and install the optional 250 ohm load resistor.
2.
Plug the load resistor into the rear panel of the field communicator.
Figure 4-2. Signal Line Connections, < 250 Ohms Load Resistance
Page 64
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
4-4
00809-0100-

FIELD COMMUNICATOR PC CONNECTIONS

OFF-LINE AND ON-LINE OPERATIONS

There is an option to interface the field communicato r with a personal com­puter. Refer to the applicable field communicator documentation regarding the PC interface option.
The field communicator can be operated both of f-line and on-line. Off-line operations are those in which the communicator is not connected to
the Rosemount 5081FG Analyzer. Off-line operat ions can include interfacing the field communicator with a PC (refer to applicable HA RT docum ent ation regarding HART/PC applications.
In the on-line mode, the communicator is connected to the 4-20 mA analog output signal line. The communicator is conne ct ed i n parallel to the Rosemount 5081FG Analyzer or in parallel to the 250 ohm load resistor.

NOTE

If the field communicator is turned on while connected to the 4-20 mA analog output signal line, an undefined status indicatio n appears while the communicator warms up. Wait until the warm-up pe riod ends to continue.
The opening menu displayed on the LCD is different for on-line and off-line operations. When powering up a disconnected (off -line) communicator, the LCD will display the Main Menu. When powering up a connected (on-line) communicator, the LCD will display the On-lin e Menu. Refer to the field com­municator manual for detailed menu information.
HART/AMS MENU TREE This section consists of a menu tree for the field communicator. This menu is
specific to Two-Wire In -S i t u Oxygen Analyzer applications.
Page 65
Reference Manual
00809 May 2019
Rosemount 5081FG
4-5
-0100-4882, Rev AB
Figure 4-3. HART/AMS Menu Tree (Sheet 1 of 3)
Page 66
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
4-6
Figure 4-3. HART/AMS Menu Tree (Sheet 2 of 3)
00809-0100-
Page 67
Reference Manual
00809 May 2019
Rosemount 5081FG
4-7
-0100-4882, Rev AB
Figure 4-3. HART/AMS Menu Tree (Sheet 3 of 3)
Page 68
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
4-8
00809-0100-

FIELD COMMUNICATOR START CAL CHECK METHOD

To perform a calibration check on the Two-Wire In-Situ Oxygen Analyzer with the field communicator, use the following proced ure. If necessary, use the menu tree in Figure 4-3 (sheet 1 of 3) for reference.

NOTE

To select a menu item use either the up and down arrow keys to scroll to the menu item. Press the right arrow key or use the number keypad to select the menu item number. To return to a preceding menu, press t he l eft arrow key.

NOTE

Pressing ABORT at any time during this process will purge the calibration check gases and end the calibration check procedure.
1.
From the DEVICE SETUP SCREEN, select DIAG/SERVICE.
2.
From the DIAG/SERVICE screen, select O2 CALCHECK.
3.
Before starting the calibration check procedure, first set up the high calibration check gas, low calibration check gas, and tracking using HART/AMS.
4.
From the O2 CALCHECK screen, select menu item 1, START CALCHECK, to access the calibration check procedure.
Failure to remove the analyzer from automatic control loops prior to performing this procedure may result in a dangerous operating condition.
5.
In the first screen, a “Loop should be removed from automatic control" warning appears. Remove the analyzer from any automatic control loops to avoid undesirable equipment performance a nd press OK.
6.
The next screen prompts you to apply the high calibration check gas. This message will only display for approximately t hree seconds. Press OK.
7.
At this point, calibration check gas will flow for approximately three min­utes until the gas reading is taken. Once the gas is measured, the mes­sage, "Hi gas reading taken," displays for three seconds.
8.
Next, the screen prompts you to apply the low calibration check gas. Press OK.
9.
The low calibration check gas will flow for approximat el y three minutes until the reading is taken. Once the gas is measured, the message, "Low gas reading taken,” displays for three seconds.
10.
Next, the screen prompts you to disconnect the calibration check gases. Press OK. Once the gases are disconnected the system will purge the gases for approximately three minutes.
11.
When the “Loop may be returned to automatic control” note appears, return the analyzer to the automatic control loops previously removed and press OK.
Page 69
Reference Manual
00809 May 2019
Rosemount 5081FG
-0100-4882, Rev AB

Section 5 Troubleshooting

General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-1

Probe Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-1
Fault Indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-3
Identifying And Correcting Fault Indications . . . . . . . . . . page 5-4
Calibration Passes, But Still Reads Incorrectly . . . . . . . . page 5-8
GENERAL This troubleshooting section describes how to i dentify and isolate faults that
may develop in the Two-Wire In-Situ Oxygen Analyzer.
PROBE LIFE The zirconium oxide technology for measuring oxygen is very stable and
should provide accurate service for several year s. Life of the probe is negatively impacted by:
Continued operation at elevated temperatures abov e 1300°C (2372°F).
Operation in processes that contain high level s of sulfur, SO2, or other acidic compounds.
Operating conditions with simultaneously high levels of SO
are particularly damaging.
O
2
and low levels of
2
The health and accuracy of a given cell is closely related to the resistance, or impedance, of the cell. Figure 5-1 illustrates that the amount of output from a cell for a given O
value (represented as slope) will remain very stable to the
2
point where cell impedance increases to approximately 100 ohms. Frequently conduct calibration checks to look for t he following conditions:
Continued degradation of cell slope.
Sluggish response. (Note how long it takes the cell to res pond to the application of calibration check gases.) See Figure 5-2.
The slope will be valid only for the process temperatur e at which the calibration check gases are flowed, so no adjustments to the electronics are made as a result of a calibration check.
Note that cells exposed to temperatures above 1300°C (2372°F) may lose the ability to measure accurately and respond quickly when returned to the lowest end of the operating temperature range [550°C (1022°F)].
Page 70
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
5-2
Figure 5-1. Slope vs. Impedance
00809-0100-
Figure 5-2. Speed of Response
Page 71
Reference Manual
00809 May 2019
Rosemount 5081FG
5-3
-0100-4882, Rev AB
FAULT INDICATIONS The fault conditions for the Two-Wire In-Situ Oxygen Analyzer will be indi-
cated by the faulted operation display as shown in Figure 5-3. This screen displays when a fault that invalidates the O
error is corrected, the screen will return to a normal operation display unless another error exists.
Figure 5-3. Faulted Operation Display
reading is present. When the
2
Page 72
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
5-4
00809-0100-

IDENTIFYING AND CORRECTING FAULT INDICATIONS

Figure 5-4. Transmitter Terminal Block
A fault in the operation of the Two-Wire In-Situ Oxy gen Analyzer is indicated by the faulted operation display. If no faults exist the display will indicate NONE. Information on the current fault is found under the DIAGNOSTICS MENU as detailed in Section 3: Startup and Operatio n.
The following paragraphs describe the faults, possi ble causes, and corrective actions. Refer to Figure 5-4 as needed for test points and wiring information.

NOTE

Allow adequate time for the oxygen probe to reach its operating temperature [approximately 500°C (932°F)] before invest igating a fault. The SHOW FAULTS screen of the DIAGNOSTICS menu will indicate a fault until the unit reaches operating temperature.

NOTE

The probe uses a Type B thermocouple to measure the cell temperature. A Type B thermocouple output table may be useful for troubleshooting.
Page 73
Reference Manual
00809 May 2019
Rosemount 5081FG
5-5
-0100-4882, Rev AB
Figure 5-5. Fault 1, Open Thermocouple
Fault 1, Open Thermocouple
The thermocouple connection is open. The fault displays as shown in Figure 5-5.
1.
Refer to Figure 5-4 and check the thermocouple wiring connections at terminals 8 and 10. Ensure the wires are properly connected.
2.
Remove power. Disconnect the thermocouple wires (gray and red) from terminals 10 and 8. Measure the continuity across the gray and red thermocouple leads. The measurement should read approximately 1-2 ohms. Larger values indicate the thermocouple is open.
3.
If the thermocouple is open, replace the oxygen probe per Oxygen Probe Replacement in Section 6: Maintenance and Service.
Figure 5-6. Fault 2, Reversed Thermocouple
Fault 2, Reversed Thermocouple Active
The thermocouple connections are reversed. The fault displays as shown in Figure 5-6.
1.
Allow adequate time for the oxygen probe to reach operating temperatures. Probe temperatures below approximately 500°C (932°F) may result in this fault.
2.
Refer to Figure 5-4. Check the gray (to terminal 10) and red (to terminal 8) wires for the proper placement.
3.
Using a multimeter, measure between terminals 8(-) and 10(+). If the reading is negative, the thermocouple wiring is reversed. Rewire as necessary.
4.
If the wiring is correct and the probe is at operating temperature, then the transmitter electronics are bad. Replace the faulty analog or CPU board per Electronics Replacement in Section 6: Maintenance and Service.
Page 74
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
5-6
Figure 5-7. Fault 3, Shorted Thermocouple
00809-0100-
Fault 3, Shorted Thermocouple
The thermocouple connections are shorted. The fault displays as shown in Figure 5-7.
1.
Allow adequate time for the oxygen probe to reach operating temperatures. Probe temperatures below approximately 500°C (932°F) may result in this fault.
2.
Refer to Figure 5-4. Using a multimeter, measure between terminals 8(-) and 10(+).
3.
If the multimeter reading, in voltage mode, is between -0.5 and +0.5 mV, the thermocouple is shorted.
4.
If the thermocouple is shorted, replace the oxygen probe per Oxygen Probe Replacement in Section 6: Maintenance and Service.
5.
If the thermocouple is not shorted, then replace the faulty analog or CPU board per Electronics Replacement in Section 6: Maintenance and Service.
Figure 5-8. Fault 4, High Probe Temperature
Fault 4, High Probe Temperature
The probe's temperature has exceeded the maximum cell temperature setpoint. The fault displays as shown in Figure 5-8.
1.
If the probe temperature exceeds the maximum cell temperature setpoint, the 4-20 mA signal output will become invalid and go to the default value.
2.
Verify that the upper cell temperature setpoint is configured as desired under the PROGRAM MENU in Section 3: Startup and Operation.
Page 75
Reference Manual
00809 May 2019
Rosemount 5081FG
5-7
-0100-4882, Rev AB
Figure 5-9. Fault 5, O2 Cell Open
Fault 5, O2 Cell Open
The O displays as shown in Figure 5-9.
Check the cell output voltage at the probe terminals -- not at the electronics.
cell connection is open. The fault
2
1.
Allow adequate time for the oxygen probe to reach operating temperatures. Probe temperatures below approximately 500°C (932°F) may result in this fault.
2.
Refer to Figure 5-4 and check the O2 cell wiring connections at terminals 7 and 8. Ensure the wires are properly connected.

NOTE

3.
Apply low calibration check gas (0.4% O2). Measure the cell output from the O wires at the probe terminal block. The cell output should be 100 ±20 mV. If no voltage can be measured the cell is open.
4.
If the O2 cell is open, replace the oxygen probe per Oxygen Probe Replacement in Section 6: Maintenance and Service.
2
cell
Figure 5-10. Fault 6, Cell Impedance Too High
Fault 6, Cell Impedance Too High
The O The fault displays as shown in Figure 5-10.
cell impedance has exceeded 100 ohms.
2
1.
This fault is usually indicated in conjunction with Fault 5, Cell Open. Correcting Fault 5 should correct Fault 6.
2.
Fault 6 appears independently, the cell has degraded beyond specification.
3.
If the O2 cell has become too old, replace the oxygen probe per Oxygen Probe
Replacement in Section 6: Maintenance and Service.
Page 76
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
5-8
Figure 5-11. Fault 7, Reversed O2 Cell
00809-0100-
Fault 7, Reversed O
The O displays as shown in Figure 5-11.
cell connections are reversed. The fault
2
1.
Refer to Figure 5-4. Check the black (to terminal 7) and white (to terminal 8) wires for the proper placement. Rewire if necessary.
2.
Apply the low calibration check gas (0.4% O
3.
Using a multimeter measure between terminals 7(+) and 8(-). If the cell output reading is negative, the O reversed.
4.
If the wiring is correct, check if the multimeter reading is the same as the reading shown on the O diagnostics screen (Section 3: Startup and
Operation).
5.
If the reading is different the transmitter electronics are faulty. Replace the faulty analog or CPU board per Electronics Replacement in Section 6: Maintenance and Service.
).
2
Cell
2
cell wiring is
2
CELL mV
2

CALIBRATION PASSES, BUT STILL READS INCORRECTLY

Probe Passes Calibration, O2 Still Reads High
There are fault conditions where no alarm indicatio n i s pre sent and the probe passes calibration, but the O
reading may still be incorrect:
2
External Reference Air Leak - There may be a leak t hat is permitting ambient air to mix with the process gases. Since many combustion processes are slightly negative in pressure, ambient air can be dra wn into the cell area, biasing the O
1. Make sure that the calibration gas line is capped tightly between calibrations.
Bad Reference Side Cell Electrode - A bad reference side cell electrode can cause an elevated O
“Calibration Recommended” alarm and increasing cell impedance readings. A high cell impedance can be calibrated out, but if the i m pedance continues to increase rapidly, the sensing cell must be replaced.
reading upward.
2
reading. This fault is usually indicated by a frequent
2
Page 77
Reference Manual
00809 May 2019
Rosemount 5081FG
-0100-4882, Rev AB
Section 6 Maintenance and Service

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-1

Electronics Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-1
Oxygen Probe Replacement . . . . . . . . . . . . . . . . . . . . . . . page 6-3
OVERVIEW This section provides the procedures to maintain and service the Rosemount
5081FG Two-Wire In-Situ Oxygen Analyzer.
Install all protective equipment covers and safety ground leads after equipment repair or service. Failure to install covers and ground leads could result in serious injury or death.
Disconnect and lock out power before working on any electrical components.
ELECTRONICS REPLACEMENT
Display Board Replacement
Before replacing any electronic components v erify that the power to the Rosemount 5081 transmitter is removed. Refer t o Section 7: Replacement Parts for replacement part numbers.
Use the following procedure to replace the displ ay board (12, Figure 6-1).
1.
Loosen screw (15) until cover lock (16) disengages f rom the knurled surface of circuit end cap (14). Remove circuit end cap.
2.
Remove three screws (13) retaining the electroni cs in place.
3.
Lift display board (12) to disconnect from CPU board (11). The display board is plugged into the CPU board; use care when removing the display board.
4.
Align and plug the replacement display board (12) in t he desired orientation. Carefully plug the display board into the mating connector on CPU board (11). Ensure the display board is full y seated.
5.
Install and tighten all three screws (13).
6.
Install circuit end cap (14). Tighten cover lock screw (15) until cover lock (16) engages knurled surface of circuit end cap (14).
Page 78
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
6-2
Figure 6-1. Two-Wire In-Situ Oxygen Analyzer - Exploded View
00809-0100-
Page 79
Reference Manual
00809 May 2019
Rosemount 5081FG
6-3
-0100-4882, Rev AB

Spare Board Stack Replacement

The spare board stack is composed of the analog board (10, Figure 6-1) and the CPU board (11). Use the following procedure to re place either of these boards.
1.
Loosen cover lock screw (15) until cover lock (16) di sengages from the knurled surface of circuit end cap (14). Remove cir cuit end cap.
2.
Remove three screws (13).
3.
Lift display board (12) to disconnect from CPU board (11). The display board is plugged into the CPU board; use care when removing the display board.
4.
Remove terminal end cap (5).
5.
Remove two screws (4) and lockwashers (3). Lift terminal block (6) until the analog board (10) is unplugged from the terminal board.
6.
Lift failed board stack from housing (1) by the standoffs. Reinstall terminal block (6), lockwashers (3), and screws (4).

NOTE

Rosemount 5081 analyzers shipped after August 2008 incorporate a new display board. If display board replacement is nee ded for an earlier analyzer unit, both the display and spare board stack must be replaced as matching parts. The earlier model analyzer uses a ribbon cable to connect the display board (12) to the CPU board (10).
7.
Install new spare board stack into housing (1). Carefully seat the analog board onto housing pins. Press firmly on the CPU board standoffs to ensure good contact.
8.
Align and plug the replacement display board (12) int o the mating connector on the CPU board (11). Ensure the displ ay board is fully seated.
9.
Install and tighten all three screws.
10.
Install terminal end cap (5) and circuit end cap (14). Tighten cover lock screw (15) until cover lock (16) engages knurled surface of circuit end cap (14).

OXYGEN PROBE REPLACEMENT

The oxygen probe is designed with ceramic materials to provide maximum life at elevated temperatures and is not rebuildable. The condition of the sensing cell can be determined periodically by two methods:
Note the cell impedance at the electronics. When the im pedance displays a warning indication (WARN), increase the frequency of impedance readings. A cell with a sustained high impedance indication (HI) indicates a probe that is beyond its useful life.
Conduct a calibration check. Follow the prompts pr ovided by the electronics through the process of flowing t wo calibration check gases of known values. Record the generated slope an d constant values.
Page 80
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
6-4
00809-0100-
Use heat resistant gloves and clothing when removing the probe. The probe can be as hot as 1600
°C (2912°F). This can cause severe burns.
Do not install or remove probes from a process where pressures are more than a few inches of H personal injury.
O positive pressure. Hot gases may escape from the stack and cause severe
2
Do not insert or withdraw a probe into or out of a hot process faster than 1 in. (25,4 mm) per minute or instrument damage from thermal shock may occur.
Also, ash, slag, or other materials can build up on the probe body in some applications. If this buildup is causing difficulty when withdrawing the probe, DO NOT FORCE. Rotate the probe back and forth to attempt to loosen the material on the probe body. Or, wait until the process cools down and access the buildup from inside the furnace.
Probe replacement may be conducted online as long as the process in which the probe is mounted is operating at a negative, or slightly positive, pressure. Refer to Section 5: Troubleshooting for more information.
Refer to Table 7-1 for replacement probe part numbers. Before replacing the probe, verify that the reference air and calibration check gas lines are turned off and disconnected from the probe.
1.
Remove the end cap of the probe to expose the terminal block.
2.
Refer to Figure 6-2. Disconnect the four wires (two oxygen signal wires and two thermocouple wires) from the terminal block.
3.
Disconnect the reference air and the calibration check gas lines.
4.
Unscrew the probe from the stack and remove.
5.
Using a replacement probe, refer to Mechanical Installation in Section 2: Installation for mechanical installation instructions.
6.
Refer to Electrical Installation in Section 2: Installation for electrical installation instructions.
7.
Refer to Pneumatic Installation in Section 2: Installation for reference air and calibration check gas installation instructions.
Page 81
Reference Manual
00809 May 2019
Rosemount 5081FG
6-5
-0100-4882, Rev AB
Figure 6-2. Oxygen Probe Terminal Block
Page 82
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
6-6
00809-0100-
Page 83
Reference Manual
00809 May 2019
Rosemount 5081FG
Figure and
1-1, 2
6A00093G01
Rosemount 5081 Transmitter ATEX
1-1, 2
6A00093G02
Rosemount 5081 Transmitter CSA
1-1, 2
6A00093G03
Rosemount 5081 Transmitter FM
1-1, 3
5R10092G01
20" Replacement Oxygen Probe, with Alumina Outer
1-1, 3
5R10092G02
26" Replacement Oxygen Probe, with Alumina Outer
1-1, 3
5R10092G03
38" Replacement Oxygen Probe, with Alumina Outer
1-1, 3
5R10092G09
20" Replacement Oxygen Probe, with Inconel 600 Outer
1-1, 3
5R10092G10
26" Replacement Oxygen Probe, with Inconel 600 Outer
1-1, 3
5R10092G11
38" Replacement Oxygen Probe, with Inconel 600 Outer
6-1, 6
1A99777H04
Terminal Block
6-1, 10, 11,
1A99777H06
Spare Board Stack, HART Compatible, with Display
6-1, 12
1A99777H05
Display Board Assembly
6-1, 14
23593-01
Circuit End Cap (with Glass)
6-1, 5
1A99777H01
Terminal End Cap (witout Glass)
-0100-4882, Rev AB

Section 7 Replacement Parts

Table 7-1. Replacement Parts List
Index Number Part Number Description
Protection Tube
Protection Tube
Protection Tube
Protection Tube
Protection Tube
Protection Tube
12 and 13
Board Assembly

NOTE

Rosemount 5081 analyzers shipped after Augu st 2008 incorporate a new display board. If display board replacement is nee ded for an earlier analyzer unit, both the display and spare board stack must be r epla ced. The earlie r model analyzer uses a ribbon cable to connect the di splay board (12) to the CPU board (11).
Page 84
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
00809-0100-
Section 8 Rosemount 5081-G Product
Certifications

European Directive information

A copy of the EC Declaration of Conformity can be f ound at t he end of the Quick Start Guide. The most recent revision of the EC Declaration of Conformity can be found at Emerson.com/Rosemount

Ordinary location certification

As standard, the transmitter has been examined and t est ed to determine that the design meets the basic electrical, mechanical, and fire protection requirements by a nationally recognized test laboratory (NRTL) as accredited by the Federal Occupational Safety and Health Admi ni st ration (OSHA).
.

Installing equipment in North America

The US National Electrical Code (NEC) and the Canadian Electrical Code (CEC) permit the use of Division marked equipment in Zones and Zone marked equipment in Divi sions. The markings must be suitable for the area classification, gas, and temperature class. T hi s information is clearly defined in the respective codes.
USA
FM
Certificate: FM17US0021X Standards: FM Class 3600:2011, FM Class 3610:2015, FM Class 3611:2016
Markings:
FM Class 3615:2006, FM Class 3810:2005, ANSI/TYPE 250:1991
Intrinsically Safe for use in Class I, II and III, Division 1, Groups A, B, C, D, E, F, and G; T4 Ta = -20 °C to 70 °C; Per Control Drawing Numbers 1400227; 1400228 Nonincendive for Class I, Division 2, Groups A, B, C, D; T4 Ta = -20 °C to 70 °C; Per Control Drawing Numbers 1400227; 1400228 Dust-Ignitionproof for use in Class II and Class III, Division 1, Groups E, F, G; T6 Ta = -20 °C to 70 °C; Per Control Drawing Number 1400678 Explosionproof for use in Class I, Div 1, Groups B, C, and D; T6 Ta = -20 °C to 70 °C; Per Control Drawing Number 1400678 Type 4X

Canada

CSA
Certificate: 1132747 Standards: C22.2 No. 0-M1987, C22.2 No. 25-1966, C22.2 No. 30-M1986
Markings:
C22.2 No. 94-M91, C22.2 No 142-M1987, C22.2 No. 157-92, C22.2 No. 213-M1987
Intrinsically Safe for Class I Groups A, B, C, D; Class II Groups E, F, G; Class III; T4 Tamb = 70 °C, per Installation Drawing 1400229, 1400230 Non-Incendive for Class I, Div. 2 for Groups A, B, C, D; Class II, Div. 2, Groups F and G; Class III; T4 Tamb = 70 °C, per Installation Drawing 1400229, 1400230
Page 85
Reference Manual
00809 May 2019
Rosemount 5081FG
-0100-4882, Rev AB
Explosion-proof for Class I, Groups B, C, D; Class II, Groups E, F, G, Class III, T6 Tamb = 70 °C Type 4X

Europe

ATEX
Certificate: Baseefa04ATEX0052X Standards: EN 60079-0:2012+A11:2013
Markings: II 1 G

Special Conditions for Safe Use (X):

1. The equipment enclosure may contain light metals. The equipment must be installed in such a manner as to minimize the risk of impact or friction with other metal surfaces

EN 60079-11:2012
Ex ia IIC T4 Ga (-20°≤ Ta ≤ +65°C)

International

IECEx
Certificate: IECEx BAS 09.0159X Standards: IEC 60079-0:2011
Markings: Ex ia IIC T4 Ga
IEC 60079-11:2011 (-20°≤ Ta ≤ +65°C)

Special Conditions for Safe Use (X):

1. The Model 5081 enclosure may be made of aluminum alloy and given a protective polyurethane paint finish; however, care should be taken to protect it from impact or abrasion if located in a zone 0 environment.
Page 86
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
00809-0100-

Appendix A Safety Data

Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page A-1
Safety Data Sheet for Ceramic Fiber Products . . . . . . . . . . . . . page A-2
Page 87
Reference Manual
00809 May 2019
Rosemount 5081FG
A-1
-0100-4882, Rev AB
SAFETY INSTRUCTIONS IMPORTANT
SAFETY INSTRUCTIONS FOR THE WIRING
AND INSTALLATION OF THIS APPARATUS
The following safety instructions apply specifically to all EU member states. They should be strictly adhered to in order to assure compliance with the Low Voltage Directive. Non-EU states should also comply with the following unless superseded by local or Nati onal Standards.
1.
Adequate earth connections should be made to all earthing points, internal and external, where provided.
2.
After installation or troubleshooting, all safety covers and safety grounds must be replaced. The integrity of all earth terminals must be maintained at all times.
3.
Mains supply cords should comply with the requirements of IEC227 or IEC245.
4.
All wiring shall be suitable for use in an ambient tempe rat ure of greater than 75°C.
5.
All cable glands used should be of such internal dimensions as to provide adequate cable anchorage.
6.
To en sure safe operation of this equipment, connecti on to the mains supply should only be made through a circuit breaker which will disconnect all circuits carrying conductors during a fault situation. The circuit breaker may also include a mechanically operated isolating switch. If not, then another means of disconnecting the equipment from the supply must be provided and clearly marked as such. Circuit breakers or switches must comply with a recognized standard such as IEC947. All wiring must conform with any local standards.
7.
Where equipment or covers are marked with the symbol to the right, hazardous voltages are likely to be present beneath. These covers should only be removed when power is removed from the equipment - and then only by trained service personnel.
8.
Where equipment or covers are marked with the symbol to the right, there is a danger from hot surfaces beneath. These covers should only be removed by trained service personnel when power is removed from the equipment. Certain surfaces may remain hot to the touch.
9.
Where equipment or covers are marked with the symbol to the right, refer to the Operator Manual for instructions.
10.
All graphical symbols used in this product are from one or more of the following standards: EN61010-1, IEC417, and ISO3864.
11.
Where equipment or labels are marked "Do Not Open While Energized" or similar, there is a danger of ignition in areas whe re an explosive atmosphere is present. This equipment should only be opened when power is removed and adequate time as specified on t he label or in the instruction manual has been allowed for the equip m ent to cool down ­and then only by trained service personnel.
Page 88
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
A-2
00809-0100-
SAFETY DATA SHEET FOR CERAMIC FIBER PRODUCTS
JULY 1, 1996 SECTION I. IDENTIFICATION
PRODUCT NAME
Ceramic Fiber Heaters, Molded Insulation Modul es and Ceramic Fiber Radiant Heater Panels.

CHEMICAL FAMILY

Vitreous Aluminosilicate Fibers with Silicon Diox i de.

CHEMICAL NAME

N.A.

CHEMICAL FORMULA

N.A.

MANUFACTURER'S NAME AND ADDRESS

Watlow Columbia 2101 Pennsylvania Drive Columbia, MO 65202 573-814-1300, ext. 5170 573-474-9402

HEALTH HAZARD SUMMARY WARNING

Possible cancer hazard based on tests with laboratory animals.
May be irritating to skin, eyes and respiratory tract.
May be harmful if inhaled.
Cristobalite (crystalline silica) formed at high temperatures (above 1800ºF) can cause severe respiratory disease.
SECTION II. PHYSICAL DATA
APPEARANCE AND ODOR
Cream to white colored fiber shapes. With or without opt ional whit e to gray granular surface coating and/or optional black surface coating.
SPECIFIC WEIGHT: 12-25 LB./CUBIC FOOT BOILING POINT: N.A. VOLATILES (% BY WT.): N.A. WATER SOLUBILITY: N.A.
Page 89
Reference Manual
00809 May 2019
Rosemount 5081FG
A-3
-0100-4882, Rev AB
SECTION III. HAZARDOUS INGREDIENTS MATERIAL, QUANTITY, AND THRESHOLD/EXPOSURE LIMIT VALUES
Aluminosilicate (vitreous) 99+ % 1 fiber/cc TWA CAS. No. 142844-00-0610 fibers/cc CL Zirconium Silicate0-10% 5 mg/cubic meter (TLV) Black Surface Coating**0 - 1% 5 mg/cubic meter (TLV) Armorphous Silica/Silicon Dioxide0-10% 20 mppcf (6 mg/cubic meter) PEL (OSHA 1978) 3 gm cubic meter (Respirable dust): 10 mg/cubic meter, Intended TLV (ACGIH 1984-85) **Composition is a trade secret.
SECTION IV. FIRE AND EXPLOSION DATA
FLASH POINT: None FLAMMABILITY LIMITS: N.A. EXTINGUISHING MEDIA
Use extinguishing agent suitable for type of surrounding fire.

UNUSUAL FIRE AND EXPLOSION HAZARDS / SPECIAL FIRE FIGHTING PROCEDURES

N.A.
SECTION V. HEALTH HAZARD DATA
THRESHOLD LIMIT VALUE
(See Section III)

EFFECTS OF OVER EXPOSURE

EYE - Avoid contact with eyes. Slightly to moderately irritating. Abrasive action may cause damage to outer surface of eye.
INHALATION - May cause respiratory tract irritation. Repeated or prolonged breathing of particles of respirable size m ay cause inflammation of the lung leading to chest pain, difficult breathing, coughing and possible fibrotic change in the lung (Pneumoconiosis). Pre-existing medical conditions may be aggravated b y exposure: specifically, bronchial hyper-reactivity and chronic bronchial or lung disease.
INGESTION - May cause gastrointestinal disturbances. Symptoms may include irritation and nausea, vomiting and diarrhea.
SKIN - Slightly to moderate irritating. May cause irritation and inflammation due to mechanical reaction to sharp, broken ends of fibers.
Page 90
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
A-4
00809-0100-

EXPOSURE TO USED CERAMIC F IBER PRODUCT

Product which has been in service at elevated temperat ures (greater than 1800ºF/982ºC) may undergo partial conversion to cristobalite, a form of crystalline silica which can cause severe respiratory disease (Pneumoconiosis). The amount of cristobalite present will depend on the temperature and length of time in service. (See Secti on IX for permissible exposure levels).

SPECIAL TOXIC EFFECTS

The existing toxicology and epidemiology data bases for RCF's are still preliminary. Information will be updated as studies are completed and reviewed. The following is a review of the results t o date:

EPIDEMIOLOGY

At this time there are no known published reports demonstrating negative health outcomes of workers exposed t o ref ractory ceramic fiber (RCF). Epidemiologic investigations of RCF production workers are ongoing.
1.
There is no evidence of any fibrotic lung disease (interstitial fibrosis) whatsoever on x-ray.
2.
There is no evidence of any lung disease among those employees exposed to RCF that had never smoked.
3.
A statistical "trend" was observed in the exposed population between the duration of exposure to RCF and a decrease in some measures of pulmonary function. These observations are clinically insignificant. In other words, if these observations were made on a n i ndi vidual employee, the results would be interpreted as being within the normal range.
4.
Pleural plaques (thickening along the chest wall) have been observed in a small number of employees who had a long duration of employment. There are several occupational and non-occupational causes for pleural plaque. It should be noted that plaques are not "pre-cancer" nor are they associated with any measurable effect on lung function.

TOXICOLOGY

A number of studies on the health effects of inhalation exposure of rats and hamsters are available. Rats were exposed to RCF in a series of life-time nose-only inhalation studies. The animals were exposed to 30, 16, 9, and 3 mg/m3, which corresponds with approxi m at el y 200, 150, 75, and 25 fibers/cc.
Animals exposed to 30 and 16 mg/m3 were observed to have developed a pleural and parenchymal fibroses; ani mals exposed to 9 mg/m3 had developed a mild parenchymal fibrosis; ani m als exposed to the lowest dose were found to have the response typically observed any time a material is inhaled into the deep lung. While a stat ist i cally significant increase in lung tumors was observ ed following exposure to the highest dose, there was no excess lung cancers at the other doses. Two rats exposed to 30 mg/m3 and one rat exposed t o 9 m g/m3 developed masotheliomas.
Page 91
Reference Manual
00809 May 2019
Rosemount 5081FG
A-5
-0100-4882, Rev AB
The International Agency for Research on Cancer (IARC) reviewed the carcinogenicity data on man-made vitreous fibers (including ceramic fiber, glasswool, rockwool, and slagwool) in 1987. IA RC classified ceramic fiber, fibrous glasswool and mineral wool (rockwool and slagwool) as possible human carcinogens (Group 2B).

EMERGENCY FIRST AID PROCEDURES

EYE CONTACT - Flush eyes immediately with large amounts of water for approximately 15 minutes. Eye lids should be held away from the eyeball to insure thorough rinsing. Do not rub ey es. Get medical attention if irritation persists.
INHALATION - Remove person from source of exposure and move to fresh air. Some people may be sensitive to fiber induced irritation of the respiratory tract. If symptoms such as short ness of breath, coughing, wheezing or chest pain develop, seek medical attention. If person experiences continued breathing difficulties, administer oxygen until medical assistance can be rendered.
INGESTION - Do not induce vomiting. Get medical attention if irritation persists.
SKIN CONTACT - Do not rub or scratch exposed skin. Wash area of contact thoroughly with soap and water. Using a skin cream or lotion after washing may be helpful. Get medical attention if irritation persists.
SECTION VI. REACTIVITY DATA
STABILITY/CONDITIONS TO AVOID
Stable under normal conditions of use.

HAZARDOUS POLYMERIZATION/CONDITIONS TO AVOID

N.A.

INCOMPATIBILITY/MATERIALS TO AVOID

Incompatible with hydrofluoric acid and concentrated alkali.

HAZARDOUS DECOMPOSITIO N PRODUCTS

N.A.

SECTION VII. SPILL OR LEAK PROCEDURES

STEPS TO BE TAKEN IF MATERIAL IS RELEASED OR SPILLED
Where possible, use vacuum suction with HEPA filters to clean up spilled material. Use dust suppressant where sweeping if necessary. Avoid clean up procedure which may result in water pollut ion. (Observe Special Protection Information Section VII I.)

WASTE DISPOSAL METHODS

The transportation, treatment, and disposal of this waste material must be conducted in compliance with all applicable Federal, State, and Local regulations.
Page 92
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
A-6
00809-0100-
SECTION VIII. SPECIAL PROTECTION INFORMATION
RESPIRATORY PROTECTION
Use NIOSH or MSHA approved equipment when ai rborne exposure limits may be exceeded. NIOSH/MSHA approved breathing equipment may be required for non-routine and emergency use. (S ee S ect i on IX for suitable equipment).
Pending the results of long term health effects studies, engineering control of airborne fibers to the lowest levels attainable is advised.

VENTILATION

Ventilation should be used whenever possible to control or reduce airborne concentrations of fiber and dust. Carbon monoxide, carbon dioxide, oxides of nitrogen, reactive hydrocarbons and a small amount of formaldehyde may accompany binder burn off duri ng first heat. Use adequate ventilation or other precautions to elim i nat e vapors resulting from binder burn off. Exposure to burn off fumes ma y cause respiratory tract irritation, bronchial hyper-reactivity and asthmatic response.

SKIN PROTECTION

Wear gloves, hats and full body clothing to prevent skin contact. Use separate lockers for work clothes to prevent fiber t ransfer to street clothes. Wash work clothes separately from othe r clothing and rinse washing machine thoroughly after use.

EYE PROTECTION

Wear safety glasses or chemical worker's goggles to prevent eye contact. Do not wear contact lenses when working with this substance. Have eye baths readily available where eye contact can occur.

SECTION IX. SPECIAL PRECAUTIONS

PRECAUTIONS TO BE TAKEN IN HANDLING AND STORING
General cleanliness should be followed. The Toxicology data indicate that ceramic fiber should be handled with
caution. The handling practices described in this MSDS must be strictly followed. In particular, when handling refractory ceramic fiber in any application, special caution should be taken to avoi d unnecessary cutting and tearing of the material to minimize gene rat ion of airborne dust.
It is recommended that full body clothing be worn to reduce the potential for skin irritation. Washable or disposable clothing may be used. Do not take unwashed work clothing home. Work clothes sh oul d be washed separately from other clothing. Rinse washing machine thoroughly after use. If clothing is to be laundered by someone else, inform launderer of proper procedure. Work clothes and street clothes s houl d be kept separate to prevent contamination.
Product which has been in service at elevated temperat ures (greater than 1800°F/982°C) may undergo partial conversion to cristobalite, a form of crystalline silica. This reaction occurs at the furnace lining hot face. As a consequence, this material becomes more friable; special caution must be taken to minimize generation of ai r-borne dust. The amount of cristobalite present will depend on t he temperature and length in service.
Page 93
Reference Manual
00809 May 2019
Rosemount 5081FG
A-7
Concentration Personal Protective Equipment
-0100-4882, Rev AB
IARC has recently reviewed the animal, human, and o ther relevant experimental data on silica in order to critically evaluate and classify the cancer causing potential. Based on its review, IARC classified crystalline silica as a group 2A carcinogen (probable human carcinogen).
The OSHA permissible exposure limit (PEL for cristobalite is 0.05 mg/m3 (respirable dust). The ACGIH threshold l i m it value (TLV) for cristobalite is 0.05 mg/m3 (respirable dust) (ACGIH 1 991-92). Use NIOSH or MSHA approved equipment when airborne exposure limits may be exceeded. The minimum respiratory protect ion recommended for given airborne fiber or cristobalite concentrations are:

CONCENTRATION

3
0-1 fiber/cc or 0-0.05 mg/m cristobalite (the OSHA PEL)
Up to 5 fibers/cc or up to 10 times the OSHA PEL for cristobalite
Up to 25 fibers/cc or 50 times the OSHA PEL for cristobalite (2.5
3
)
mg/m
Greater than 25 fibers/cc or 50 times the OSHA PEL for cristobalite (2.5 mg/m
3
)
Optional disposable dust respirator (e.g. 3M 9970 or equivalent).
Half face, air purifying respirator equipped with high efficiency particulate air (HEPA) filter cartridges (e.g. 3M 6000 series with 2040 filter or equivalent).
Full face, air purifying respirator with high efficiency particulate air (HEPA) filter cartridges (e.g. 3M 7800S with 7255 filters or equivalent) or powered air purifying respirator (PARR) equipped with HEPA filter cartridges (e.g. 3M W3265S with W3267 filters or equivalent).
Full face, positive pressure supplied air respirator (e.g. 3M 7800S with W9435 hose & W3196 low pressure regulator kit connected to clean air supply or equivalent).
If airborne fiber or cristobalite concentrations are not known, as minimum protection, use NIOSH/MSHA appr oved half face, air purifying respirator with HEPA filter cartridges.
Insulation surface should be lightly sprayed with water before removal to suppress airborne dust. As water evaporates during removal, additional water should be sprayed on surfaces as needed. Only enough water should be sprayed to suppress dust so that water does not run onto the floor of the work area. To aid the wetting process, a surfactant can be used.
After RCF removal is completed, dust suppres sing cleaning methods, such as wet sweeping or vacuuming, should be used t o clean t he work area. If dry vacuuming is used, the vacuum must be equipped with HEPA filter. Air blowing or dry sweeping should not be used. Dust suppressing components can be used to clean up light dust.
Product packaging may contain product residue. D o not reuse except to reship or return Ceramic Fiber products to the fact ory .
Page 94
Reference Manual
4882, Rev AB
May 2019
Rosemount 5081FG
A-8
00809-0100-
Page 95
Reference Manual
00809 May 2019
Rosemount 5081FG
-0100-4882, Rev AB

Appendix B Return of Material

RETURNING MATERIAL SERVICE SUPPORT - To expedite t he return process outside of the United
States, contact the nearest Emerson representat i ve. Within the United States, call the Emerson Instrument and Valves Response Center using the 1-800-654-RSMT (7768) toll-free number. This center, available 24 hours a day, will assist you with any needed information or materials. The center will ask for product model and serial numbers, and will prov i de a Return Material Authorization (RMA) number. The center will also as k for the process material to which the product was last exposed. Emerson Instrument and Valves Response Center representatives will explain the additional information and procedures necessary to return goods exposed t o hazardous substances.
Page 96
Reference Manual
4882, Rev AB
June 2019
Rosemount 5081FG
B-2
00809-0100-
Page 97
Reference Manual
00809 June 2019
Rosemount 5081FG
-0100-4882, Rev AB

EC Declaration of Conformity

Page 98
Reference Manual
4882, Rev AB
June 2019
Rosemount 5081FG
B-2
00809-0100-
Page 99
Reference Manual
00809 June 2019
Rosemount 5081FG
部件名称
有害物质 / Hazardous Substances
六价铬
(Cr +6)
多溴联苯
(PBB)
多溴联苯醚
(PBDE)
Assembly
Assembly
-0100-4882, Rev AB
表格
Table 1: List of Model Parts with China RoHS Concentration above MCVs
1:
含有
China RoHS
管控物质超过最大浓度限值的部件型号列
Part Name
Lead (Pb)
Mercury
(Hg)
Cadmium
(Cd)
Hexavalent
Chromium
Polybrominated
biphenyls
Polybrominated
diphenyl ethers
电子组件
Electronics
X O O O O O
壳体组件
Housing
本表格系依据
This table is proposed in accordance with the provision of SJ/T11364
意为该部件的所有均质材料中该有害物质的含量均低于
O: O: Indicate that said hazardous subs tance in all of the homogeneous mat erials for this part is below the limit requirement of GB/T
26572.
: 意为在该部件所使用的所有均质材料里,至少有一类均质材料中该有害物质的含量高于
X X: Indicate that said hazardous s ubstance contained in at least one of the homogeneous materials used for this part is above the limit requirement of GB/T 26572.
SJ/T11364
X O O X O O
的规定而制作。
GB/T 26572
所规定的限量要求。
GB/T 26572
所规定的限量要求。
Page 100
Loading...