Emerson designs, manufactures, and tests its products to meet many national and international standards. Because these
instruments are sophisticated technical products, you must properly install, use, and maintain them to ensure they continue to
operate within their normal specifications. The following instructions must be adhered to and integrated into your safety program
when installing, using, and maintaining Emerson products. Failure to follow the proper instructions may cause any one of the
following situations to occur: loss of life, personal injury, property damage, damage to this instrument, and warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the product.
• If this Reference Manual is not the correct one, call 1-800-854-8257 or 949-757-8500 to request the correct Reference
Manual. Save this Reference Manual for future reference.
• If you do not understand any of the instructions, contact your Emerson representative for clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper installation, operation, and maintenance of the product.
• Install equipment as specified in the installation instructions of the appropriate Reference Manual and per applicable local and
national codes. Connect all products to the proper electrical and pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
• When replacement parts are required, ensure that qualified people use replacement parts specified by Emerson.
Unauthorized parts and procedures can affect the product's performance, place the safe operation of your process at risk, and
VOID YOUR WARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed
by qualified people, to prevent electrical shock and personal injury.
WARNING
Physical access
Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of end users’ equipment. This could
be intentional or unintentional and needs to be protected against.
Physical security is an important part of any security program and fundamental to protecting your system. Restrict physical access
by unauthorized personnel to protect end users’ assets. This is true for all systems used within the facility.
Notice
ROSEMOUNT (“SELLER”) SHALL NOT BE LIABLE FOR TECHNICAL OR EDITORIAL ERRORS IN THIS MANUAL OR OMISSIONS FROM
THIS MANUAL. SELLER MAKES NO WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, WITH RESPECT TO THIS MANUAL AND, IN NO EVENT, SHALL
SELLER BE LIABLE FOR ANY SPECIAL OR CONSEQUENTIAL DAMAGES INCLUDING, BUT NOT LIMITED TO, LOSS OF PRODUCTION,
LOSS OF PROFITS, ETC.
PRODUCT NAMES USED HEREIN ARE FOR MANUFACTURER OR SUPPLIER IDENTIFICATION ONLY AND MAY BE TRADEMARKS/
REGISTERED TRADEMARKS OF THESE COMPANIES.
THE CONTENTS OF THIS PUBLICATION ARE PRESENTED FOR INFORMATIONAL PURPOSES ONLY, AND WHILE EVERY EFFORT HAS
BEEN MADE TO ENSURE THEIR ACCURACY, THEY ARE NOT TO BE CONSTRUED AS WARRANTIES OR GUARANTEES, EXPRESSED OR
IMPLIED, REGARDING THE PRODUCTS OR SERVICES DESCRIBED HEREIN OR THEIR USE OR APPLICABILITY. WE RESERVE THE RIGHT
TO MODIFY OR IMPROVE THE DESIGNS OR SPECIFICATIONS OF SUCH PRODUCTS AT ANY TIME.
SELLER DOES NOT ASSUME RESPONSIBILITY FOR THE SELECTION, USE, OR MAINTENANCE OF ANY PRODUCT. RESPONSIBILITY FOR
PROPER SELECTION, USE, AND MAINTENANCE OF ANY SELLER PRODUCT REMAINS SOLELY WITH THE PURCHASER AND END-USER.
Warranty
LIMITED WARRANTY: Subject to the limitations contained in Section 2 herein and except as otherwise expressly provided
1.
herein, Rosemount (“Seller”) warrants that the firmware will execute the programming instructions provided by Seller
and that the Goods manufactured or Services provided by Seller will be free from defects in materials or workmanship
under normal use and care until the expiration of the applicable warranty period. Goods are warranted for twelve (12)
months from the date of initial installation or eighteen (18) months from the date of shipment by Seller, whichever period
expires first. Consumables and Services are warranted for a period of 90 days from the date of shipment or completion of
the Services. Products purchased by Seller from a third party for resale to Buyer (“Resale Products”) shall carry only the
2
Page 3
warranty extended by the original manufacturer. Buyer agrees that Seller has no liability for Resale Products beyond
making a reasonable commercial effort to arrange for procurement and shipping of the Resale Products. If Buyer
discovers any warranty defects and notifies Seller thereof in writing during the applicable warranty period, Seller shall, at
its option, promptly correct any errors that are found by Seller in the firmware or Services, or repair or replace F.O.B. point
of manufacture that portion of the Goods or firmware found by Seller to be defective, or refund the purchase price of the
defective portion of the Goods/Services. All replacements or repairs necessitated by inadequate maintenance, normal
wear and usage, unsuitable power sources, unsuitable environmental conditions, accident, misuse, improper installation,
modification, repair, storage or handling, or any other cause not the fault of Seller are not covered by this limited
warranty, and shall be at Buyer's expense. Seller shall not be obligated to pay any costs or charges incurred by Buyer or any
other party except as may be agreed upon in writing in advance by an authorized Seller representative. All costs of
dismantling, reinstallation and freight, and the time and expenses of Seller's personnel for site travel and diagnosis under
this warranty clause shall be borne by Buyer unless accepted in writing by Seller. Goods repaired and parts replaced during
the warranty period shall be in warranty for the remainder of the original warranty period or ninety (90) days, whichever is
longer. This limited warranty is the only warranty made by Seller and can be amended only in a writing signed by an
authorized representative of Seller. Except as otherwise expressly provided in the Agreement, THERE ARE NO
REPRESENTATIONS OR WARRANTIES OF ANY KIND, EXPRESSED OR IMPLIED, AS TO MERCHANTABILITY, FITNESS FOR
PARTICULAR PURPOSE, OR ANY OTHER MATTER WITH RESPECT TO ANY OF THE GOODS OR SERVICES. It is understood
that corrosion or erosion of materials is not covered by our guarantee.
LIMITATION OF REMEDY AND LIABILITY: SELLER SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY DELAY IN
2.
PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF WARRANTY HEREUNDER SHALL BE LIMITED TO
REPAIR, CORRECTION, REPLACEMENT, OR REFUND OF PURCHASE PRICE UNDER THE LIMITED WARRANTY CLAUSE IN
SECTION 1 HEREIN. IN NO EVENT, REGARDLESS OF THE FORM OF THE CLAIM OR CAUSE OF ACTION (WHETHER BASED IN
CONTRACT, INFRINGEMENT, NEGLIGENCE, STRICT LIABILITY, OTHER TORT, OR OTHERWISE), SHALL SELLER'S LIABILITY TO
BUYER AND/OR ITS CUSTOMERS EXCEED THE PRICE TO BUYER OF THE SPECIFIC GOODS MANUFACTURED OR SERVICES
PROVIDED BY SELLER GIVING RISE TO THE CLAIM OR CAUSE OF ACTION. BUYER AGREES THAT IN NO EVENT SHALL
SELLER'S LIABILITY TO BUYER AND/OR ITS CUSTOMERS EXTEND TO INCLUDE INCIDENTAL, CONSEQUENTIAL OR PUNITIVE
DAMAGES. THE TERM “CONSEQUENTIAL DAMAGES” SHALL INCLUDE, BUT NOT BE LIMITED TO, LOSS OF ANTICIPATED
PROFITS, LOSS OF USE, LOSS OF REVENUE, AND COST OF CAPITAL.
1. Using the infrared remote control (IRC), press PROG, NEXT, NEXT, and ENTER in this
order.
2. Select the measurement type and unit of measurement.
3. Use the arrow keys to toggle between Celsius and Fahrenheit.
4. Press ENTER and then RESET.
5. Press PROG, NEXT, and ENTER in this order.
6. Use the arrow keys to toggle T AUTO between ON or OFF.
This determines whether the transmitter uses the process temperature (ON) or a
manual temperature (OFF).
7. Press ENTER.
8. If you select OFF, enter the manual temperature desired using the arrow keys.
9. Press ENTER.
Reference Manual 7
Page 8
Start-upReference Manual
October 201900809-0100-3581
8 Rosemount 5081
Page 9
Reference Manual Description and specifications
00809-0100-3581October 2019
2Description and specifications
2.1Features and applications
The Rosemount™ 5081 includes the following features:
General
The Rosemount 5081 explosion-proof transmitter is a loop powered device with a robust
design that serves several industrial, commercial, and municipal applications. It offers a
local operator interface (LOI) that can display values from a single measurement input.
This transmitter is compatible with a multitude of analytical sensors.
Analytical measurements
• pH/ORP
• Contacting conductivity
• Toroidal conductivity
• Dissolved oxygen
• Ozone
• Chlorine
Maintenance features
• Automatic two-point buffer calibration routine
• Automatic recognition of resistance temperature device (RTD)
• Sensor diagnostics
Diagnostics
Continuous monitoring of sensor performance along with warnings and fault messages to
alert the user of failures.
Enclosure
Explosion-proof and corrosion resistant
2.2General specifications
Table 2-1: General Specifications
HousingCast aluminum with epoxy coating. NEMA® 4X(IP65) and NEMA7B. Neoprene
O-ring seals.
Dimensions6.3 x 6.9 x 6.4 in. (160.5 x 175.3 x 161.3 mm) See the engineering drawings in
Engineering Drawings.
Conduit openings¾-in. female national pipe thread (FNPT)
Ambient temperature-4 to 149 °F (-20 to 65 °C)
Reference Manual 9
Page 10
Description and specificationsReference Manual
October 201900809-0100-3581
Table 2-1: General Specifications (continued)
Storage temperature-22 to 176 °F (-30 to 80 °C)
Relative humidity0 to 95% (non-condensing)
Weight / shipping weight10 lb./11 lb. (4.5 kg/5.0 kg)
DisplayFirst line: 7 segment LCD, 0.8 in. (20 mm) high. This line shows process
variables (pH/ORP, contacting conductivity, toroidal conductivity, etc.)
Second line: 7 segment LCD, 0.3 in. (7 mm) high. This line shows process
temperature, output current, warnings, faults, and messages during
calibration/programming.
Display board can be rotated 90 degrees clockwise or counterclockwise if
desired.
RFI/EMIMeets all industrial requirements of EN61326.
Diagnostics (may slightly vary based
on measurement type)
• Calibration error
• Low temperature error
• High temperature error
• Sensor failure
• Line failure
• CPU failure
• Calibration error
• Zero error
• Temperature slope error
• Sensor failure
• ROM failure
• Input warning
Once one of the above warnings/faults are diagnosed, the LOI will display a
message describing the failure detected.
Table 2-2: HART® Digital Communications
Power and load requirementsSupply voltage at the transmitter terminals should be at least 12 Vdc. Power
supply voltage should cover the voltage drop on the cable plus the external
load resistor required for HART communications (250 Ohms minimum).
Minimum power voltage is 12 Vdc. Maximum power voltage is 42.4 Vdc
(30 Vdc for intrinsically safe operation). Figure 4-1 shows the supply voltage
required to maintain 12 Vdc (upper line) and 30 Vdc (lower line) at the
transmitter terminals when the current is 22 mA.
Analog outputTwo-wire, 4-20 mA output with superimposed HART digital signal, scalable
over the operating range of the sensor.
Output accuracy±0.05 mA
10 Rosemount 5081
Page 11
Reference Manual Description and specifications
00809-0100-3581October 2019
Table 2-2: HART® Digital Communications (continued)
Variables assignable to• pH
• Temperature
• mV
• Glass impedance
• Reference impedance
• RTD resistance
• Oxidation reduction potential (ORP)
• Conductivity
• Resistivity
• Concentration
• Raw conductivity
• Chlorine
• Dissolved oxygen
• Dissolved ozone
Table 2-3: FOUNDATION™ Fieldbus Digital Communications
Power and load requirementsA power supply voltage of 9-32 Vdc at 22 mA is required.
A copy of the EU Declaration of Conformity can be found at the end of the Quick Start
Guide. The most recent revision of the EU Declaration of Conformity can be found at
Emerson.com/Rosemount.
2.3.2Ordinary Location Certification
As standard, the Power Module has been examined and tested to determine that the
design meets the basic electrical, mechanical, and fire protection requirements by a
nationally recognized test laboratory (NRTL) as accredited by the Federal Occupational
Safety and Health Administration (OSHA).
2.3.3
2.3.4
Installing equipment in North America
The US National Electrical Code® (NEC) and the Canadian Electrical Code (CEC) permit the
use of Division marked equipment in Zones and Zone marked equipment in Divisions. The
marking must be suitable for the area classification, gas, and temperature class. This
information is clearly defined in the respective codes.
USA
FM hazardous locations
Certificate
Standards
Markings
FM17US0021X
FM Class 3600:2011, FM Class 3610:2015, FM Class 3611:2016 FM Class
3615:2006, FM Class 3810:2005, ANSI/NEMA 250:1991
Intrinsically safe for use in Class I, II, and III, Division 1, Groups A, B,
C, D, E, F, and G; T4 Ta = -20 °C to 70 °C; per control drawing numbers
1400676; 1400677
Nonincendive for Class I, Division 2, Groups A, B, C, and D; T4 Ta = -20 °C
to 70 °C; per control drawing numbers 1400676; 1400677
Dust-ignitionproof for use in Class II and Class III, Division 1, Groups E, F,
and G; T6 Ta = -20 °C to 70 °C;per control drawing number 1400678
Explosionproof for use in Class I, Div 1, Groups B, C, and D; T6 Ta = -20 °C
to 70 °C; per control drawing number 1400678
Type 4X
Special Conditions for Safe Use (X):
1. The Rosemount 5081-T conductivity transmitters shall only be used with the
Rosemount 222, 225, 226, 228 (1 in. and 2 in. only) and 245 toroidal sensors.
Intrinsically safe for Class I Groups A, B, C, and D; Class II Groups E,
F, and G; Class III; T4 Tamb = 70 °C, per installation drawing 1400674
and 1400675
Non-incendive for Class I, Div. 2 for Groups A, B, C, and D; Class II, Div. 2,
Groups F and G; Class III; T4 Tamb = 70 °C, per installation drawing
1400674 and 1400675 (Rosemount 5081-A/P/C/T) and per 1700462
(Rosemount 5081-T)
Explosion-proof for Class I, Groups B, C, and D; Class II, Groups E, F, and
G, Class III, T6 Tamb = 70 °C
Type 4X
Certificate
Standards
Markings
Special Conditions for Safe Use (X):
1. The Rosemount 5081 enclosure may be made of aluminum alloy and given a
protective polyurethane paint finish; however, care should be taken to protect it
from impact or abrasion of located in a zone 0.
BAS02ATEX1284X
EN 60079-0:2012+A11:2013
EN 60079-11:2012
II 1 G
Ex ia IIC T4 Ga
(-20 °C ≤ Ta ≤ +65 °C)
Rosemount 5081-C liquid transmitter
ATEX
Certificate
Standards
Baseefa03ATEX0099X
EN 60079-0:2012+A11:2013
EN 60079-11:2012
Reference Manual 13
Page 14
Description and specificationsReference Manual
October 201900809-0100-3581
Markings
Special Conditions for Safe Use (X):
1. The equipment enclosure may contain light metals. The equipment must be
installed in such a manner as to minimize the risk of impact or friction with other
metal surfaces.
II 1 G
Ex ia IIC T4 Ga
(-20 °C ≤ Ta ≤ +65 °C)
Rosemount 5081-T liquid transmitter
ATEX
Certificate
Standards
Markings
Special Conditions for Safe Use (X):
Baseefa03ATEX0399X
EN 60079-0:2012+A11:2013
EN 60079-11:2012
II 1 G
Ex ia IIC T4 Ga
(-20 °C ≤ Ta ≤ +65°C)
2.3.7
1. The equipment may contain light metals. The equipment must be installed in such
a manner as to minimize the risk or impact or friction with other metal surfaces.
International
IECEx
Certificate
Standards
Markings
Special Conditions for Safe Use (X):
1. The Rosemount 5081 enclosure may be made of aluminum alloy and given a
protective polyurethane paint finish; however, care should be taken to protect it
from impact or abrasion if located in a zone 0 environment.
IECEx BAS 09.0159X
IEC 60079-0:2011
IEC 60079-11:2011
Ex ia IIC T4 Ga
(-20 °C ≤ Ta ≤ +65°C)
14 Rosemount 5081
Page 15
Reference Manual Description and specifications
00809-0100-3581October 2019
2.4Functional specifications
The sections below display the specifications for measuring different substances with the
Rosemount 5081.
2.4.1pH/ORP specifications
pH range0 to 14
ORP range-1400 to +1400 mV
Calibration/standardizationThe automatic buffer recognition uses stored buffer values and
their temperature curves for the most common buffer standards
available worldwide. The transmitter also performs a stabilization
check on the sensor in each buffer. To make a manual two-point
calibration, immerse the sensor in two different buffer solutions
and enter the pH values. The microprocessor automatically
calculates the slope which is used for self-diagnostics. The
transmitter displays an error message if the pH sensor is faulty.
The operator can read the slope on the display and/or manually
adjust it if desired. To complete an on-line, one-point
standardization process, enter the pH or ORP value of a grab
sample as measured by a lab reference.
Preamplifier locationUse a preamplifier to convert the high impedance pH electrode
signal to a low impedance signal for transmitter use. Use the
transmitter's integral preamplifier when the sensor to transmitter
distance is less than 15 ft. (4.5 m). Use a sensor with a built-in
preamplifier or a junction box if distance is longer than 15 ft. (4.5
m).
Automatic temperature
compensation
Accuracy±01 mv at 77 °F (25 °C) ± 0.01 pH
Repeatability±01 mv at 77 °F (25 °C) ±0.01 pH
Stability0.25% / year at 77 °F (25 °C)
External 3 or 4 wire Pt 100 resistance temperature device (RTD)
or Pt 1000 RTD, located in the sensor, compensates the pH
reading for temperature fluctuations. Compensation covers the
range 5 to 270 °F (-15 to 130 °C). The operator may also select
manual temperature compensation.
2.4.2Contacting conductivity specifications
Measured range0-20,000 µS/cm
CalibrationTo calibrate, immerse the sensor in a known solution and enter its value or
enter the cell constant for ultra-pure applications.
Automatic temperature compensation 3-wire Pt 100 or Pt 1000 resistance temperature device (RTD)
Conductivity: 32 to 392 °F (0 to 200 °C)
Resistivity: 32 to 212 °F (0 to 100 °C)
Low conductivity: 32 to 212 °F (0 to 100 °C)
Accuracy± 0.5% of reading and ±0.001 µS/cm
Reference Manual 15
Page 16
Description and specificationsReference Manual
October 201900809-0100-3581
Repeatability± 0.25% of reading
Stability0.25% of output range/month, non-cumulative
Ambient temperature coefficient± 0.05% of reading/°C
Temperature slope adjustment0 to 5%/°C
Other temperature compensation
algorithms
Compatible RTD100 Ohm or 1000 Ohm with automatic recognition
Ultra-pure water compensation
Cation conductivity
Raw (uncompensated) conductivity
2.4.3Toroidal conductivity specifications
Measured range50 to 2,000,000 µS/cm
CalibrationTo calibrate, immerse the sensor in a known solution and enter its value.
Automatic temperature compensation 3-wire Pt 100 resistance temperature device (RTD)
Conductivity: 32 to 392 °F (0 to 200 °C)
% concentration: 32 to 212 °F (0 to 100 °C)
Accuracy±1.0% of reading
Repeatability±0.25% of reading
Stability0.25% of output range/month, non-cumulative
Ambient temperature coefficient±0.2% of FS/°C
Temperature slope adjustment0 to 5% / °C
% concentration rangesSodium hydroxide: 0 to 15%
Hydrochloric acid: 0 to 16%
Sulfuric acid: 0 to 25% and 96 to 99.7%
Resolution0.01 ppm, 0.1 ppb for Rosemount 499ATrDO sensor
Temperature correction for
membrane permeability
CalibrationAir calibration (user must enter barometric pressure) or calibration against a
Automatic between 32 and 122 °F (0 and 50 °C). Can be disabled.
standard instrument
2.4.5Free chlorine specifications
Measurement range0-20 ppm (mg/L) as Cl
Resolution0.001 ppm (auto-ranges at 0.999 to 1.00 and 9.99 to 10.0)
Temperature correction for
membrane permeability
16 Rosemount 5081
Automatic between 32 and 122 °F (0 and 50 °C). Can be disabled.
2
Page 17
Reference Manual Description and specifications
00809-0100-3581October 2019
pH correctionAutomatic between pH 6.0 and 9.5. Manual pH correction is also available.
CalibrationCalibrate against grab sample with portable test kit.
2.4.6Total chlorine specifications
Measurement range0-20 ppm (mg/L) as Cl
Resolution0.001 ppm (auto-ranges at 0.999 to 1.00 and 9.99 to 10.0)
Temperature correction for
membrane permeability
CalibrationCalibrate against grab sample with portable test kit.
Automatic between 41 and 95 °F (5 and 35 °C). Can be disabled.
2
2.4.7Ozone specifications
Measurement range0-10 ppm (mg/L)
Resolution0.001 ppm (auto-ranges at 0.999 to 1.00 and 9.99 to 10.0)
Temperature correction for
membrane permeability
CalibrationAgainst grab sample analyzed using portable test kit.
Automatic between 41 and 95 °F (5 and 35 °C). Can be disabled.
2.4.8Percent oxygen in gas
Measurement range0 - 25% oxygen
Resolution0.1% - TBD
CalibrationAir calibration (automatic measurement of barometric pressure with internal
pressure sensor)
Sample pressure0 to 50 psig
Sample temperature32 to 110 °F (0 to 43 °C)
2.5Ordering information
The Rosemount 5081 two-wire transmitter is intended for the determination of pH/ORP,
conductivity (both contacting and toroidal), and for measurements using membranecovered amperometric sensors (oxygen, ozone, and chlorine). For free chlorine
measurements, which often require continuous pH correction, a second input for a pH
sensor is available. Use a hand-held infrared remote controller to locally configure and
calibrate the transmitter.
Rosemount 5081
OptionDescription
PpH/ORP
CContacting conductivity
Reference Manual 17
Smart two-wire microprocessor transmitter
Page 18
Description and specificationsReference Manual
October 201900809-0100-3581
TToroidal conductivity
AAmperometric (oxygen, ozone, and chlorine)
OptionDescription
HTAnalog 4-20 mA output with superimposed HART® digital signal
FFFOUNDATION™ Fieldbus digital output
FIFOUNDATION Fieldbus digital input with FISCO
OptionDescription
20Infrared remote controller included
21Infrared remote controller not included
OptionDescription
60No approval
67FM approved intrinsically safe, non-incendive (when used with
appropriate sensor and safety barrier), and explosion-proof
69CSA approved intrinsically safe, non-incendive (when used with
appropriate sensor and safety barrier), and explosion-proof
73ATEX/IECEx approved intriniscally safe (when used with
appropriate sensor and safety barrier)
Example5081-P-HT-20-67
18 Rosemount 5081
Page 19
Reference Manual
00809-0100-3581October 2019
Install
3Install
3.1Unpack and inspect
To unpack the instrument:
Procedure
1. Inspect the shipping container(s). If there is damage, contact the shipper
immediately for instructions.
2. If there is no apparent damage, unpack the container(s).
3. Ensure that all items shown on the packing list are present.
If items are missing, contact your local Customer Care representative
4. Save the shipping container and packaging.
They can be used to return the instrument to the factory in case of damage.
3.2Installation guidelines
1. The transmitter tolerates harsh environments. For best results, install the
transmitter in an area where temperature extremes, vibrations, and
electromagnetic and radio frequency interference are minimized or absent.
2. To prevent unintentional exposure of the transmitter circuitry to the plant
environment, keep the security lock in place over the circuit end cap. To remove the
circuit end cap, loosen the lock nut until the tab disengages from the cap end and
then unscrew the cover.
3. The transmitter has two ¾-in. conduit openings, one on each side of the housing.
Run sensor cable through the left side opening (as viewed from the wiring terminal
end of the transmitter) and run power wiring through the right side opening.
4. Use water tight cable glands to keep moisture out of the transmitter.
5. If using conduit, plug and seal the connections at the transmitter housing to
prevent moisture from getting inside the transmitter.
CAUTION
Equipment damage
Moisture accumulating in the transmitter housing can affect the performance of
the transmitter and may void the warranty.
6. If the transmitter is installed some distance from the sensor, a remote junction box
with preamplifier in the junction box or in the sensor may be necessary. Consult the
sensor reference manual for maximum cable lengths.
Reference Manual 19
Page 20
InstallReference Manual
October 201900809-0100-3581
3.3Orient the display board
The display board can be rotated 90 degrees, clockwise or counterclockwise, from the
original position. To reposition the display:
Procedure
1. Loosen the cover lock nut until the tab disengages from the circuit end cap.
Unscrew the cap.
2. Remove the three bolts holding the circuit board stack.
3. Lift and rotate the display board 90 degrees, clockwise or counterclockwise, into
the desired position.
4. Position the display board on the stand offs. Replace and tighten the bolts.
5. Replace the circuit end cap.
20 Rosemount 5081
Page 21
Reference Manual Install
00809-0100-3581October 2019
3.4Mount on a flat surface
Figure 3-1: Mounting Transmitter on a Flat Surface
A. Cover lock
B. Threaded cap (two places)
C. Terminal block (TB). Terminal end cap omitted for clarity this view.
D. ¼-in. - 20 threads (four places)
E. Surface plate (by others)
F. O-ring (two places)
G. Circuit end
H. Terminal end
I. Flat surface mounting pad hole pattern
Reference Manual 21
Page 22
Install
October 201900809-0100-3581
Reference Manual
3.5Mount on a pipe
Use pipe mounting kit (23820-00 or 23820-01).
Figure 3-2: Mounting Transmitter on a Pipe
A. Over look
B. Terminal block
C. Kit, 2-in. pipe/wall mounting bracket
Order PN2002577 as a separate item.
D. ½ - 20 threads
E. ¼ - 20 screw
Screws furnished with mounting kit only. Not furnished with transmitter.
F. 1½-in. to 2-in. pipe (customer furnished)
G. U-bolt
22 Rosemount 5081
Page 23
Reference Manual Install
00809-0100-3581October 2019
H. 5/16 washer
I. 6/18 - 18 nut
J. ¾ - 14 FNPT (two places)
K. 2-in. pipe supplied by customer
L. ¾ - 14 NPT (two places)
M. Threaded cap, two places
N. Circuit end
O. Terminal end
P. Terminal end cap omitted for clarity this view
Q. Four mounting holes
R. Bottom view
Reference Manual 23
Page 24
InstallReference Manual
October 201900809-0100-3581
24 Rosemount 5081
Page 25
Reference Manual
00809-0100-3581October 2019
Wire
4Wire
4.1Wiring overview
To find wiring diagrams for specific sensors, check the wiring sections of the reference
manuals for those particular sensors.
4.2Power supply/current loop
4.2.1Power supply overview
The tables below display the minimum and maximum voltages needed to operate the
transmitter.
Minimum supply voltage at the transmitter
terminals
Minimum power supply for load resister250 Ohms
Maximum power supply voltage42.0 Vdc
Maximum power supply voltage for intrinsically
safe installations
Figure 4-1: Power Supply Voltage for HART® or without HART Communication
Configurations
12.0 Vdc
30.0 Vdc
A. With HART communication
B. Without HART communication
Reference Manual 25
Page 26
WireReference Manual
October 201900809-0100-3581
Table 4-1: Values from Graph
Upper linePower supply voltage needed to provide 12 Vdc
at the transmitter terminals for a 22 mA current
Lower linePower supply voltage needed to provide 30 Vdc
for a 22 mA current
Maximum currentAbout 24 mA
Minimum load for digital communications250 Ohms
Minimum power supply voltage to supply the
12.0 Vdc lift off voltage at the transmitter
17.5 Vdc
4.2.2Wire to HART® or FOUNDATION™ Fieldbus communication
protocol
Procedure
1. Run the power/signal wiring through the opening nearest terminals 15 and 16.
2. Use shielded cable and ground the shield to the power supply.
3. To ground the transmitter, attach the shield to the grounding screw on the inside of
the transmitter case.
You can also use a third wire to connect the transmitter to earth ground.
Note
For optimum EMI/RFI immunity, shield the power supply and enclose it in an earth
ground metal conduit. Do not run power supply/signal wiring in the same conduit
or cable tray with AC power lines or with relay actuated signal cables. Keep power
supply/signal wiring at least 6 ft. (2 m) away from heavy electrical equipment. An
additional 0-1 mA current loop is available between TB-14 and TB-15. A 1 mA
current in this loop signifies a sensor fault. See Figure 4-2 for wiring instructions. See
Diagnostics and troubleshooting for more information about sensor faults.
26 Rosemount 5081
Page 27
Reference Manual Wire
00809-0100-3581October 2019
Figure 4-2: General Wiring Architecture
A. Filter
B. Terminators
C. Trunk
D. Spur
The power supply, filter, first terminator, and configuration device are typically
located in the control room.
Reference Manual 27
Page 28
WireReference Manual
October 201900809-0100-3581
28 Rosemount 5081
Page 29
Reference Manual Display and operate
00809-0100-3581October 2019
5Display and operate
5.1User interface and main display
The following are examples of the main (process) display screen (Figure 5-1) and the
program display screen (Figure 5-2).
Figure 5-1: Main Display Screen
A. Conductivity value
B. Temperature in °C or °F
Figure 5-2: Program Display Screen
A. Indicates HART® or FOUNDATION™ Fieldbus digital communiciations
B. Conductivity value
C. Units of display
D. Active menu: CALIBRATE, PROGRAM, or DIAGNOSE
E. Sub-menus, prompts, and diagnostic messages appear here.
F. Available commands for sub-menus, prompts, or diagnostic messages
G. Appears when transmitter is in hold
H. Appears when a disabling condition has occurred
Reference Manual 29
Page 30
Display and operateReference Manual
October 201900809-0100-3581
5.2Infrared remote control (IRC)
Use the IRC to read diagnostics messages, calibrate connected sensors, and program the
transmitter. Hold the IRC within 6 ft. (1.8 m) of the transmitter and less than 15 degrees
from the horizontal of the display window.
Figure 5-3: Infrared Remote Control (IRC) Functions
1. RESET
• End current operation and
return to the main display.
• Changes are not saved.
• Does not return the
transmitter to factory default
settings.
2. Editing (arrow) keys
• Change values of a flashing
display.
• Left and right arrows move
the cursor by one digit.
• Up and down arrows increase
or decrease the values and
navigate through the display
options.
3. CAL
• Access to Calibration menu.
4. PROG
• Access to Program menu.
5. DIAG
• Access to diagnostics.
6. HOLD
• Access to turn hold readings
on or off.
7. ENTER
• Advance to the next prompt.
• Store selected item.
• Store value in memory.
8. NEXT
• Advance to the next sub-
menu.
9. EXIT
• End current operation.
• Return to the first prompt in
the present sub-menu.
• Changes are not saved.
Guidelines for using IRC
• Do not use harsh chemicals or abrasive brushes when cleaning the remote control.
• If the green LED does not light when you press a key, the issue is probably a weak
battery. To restore operation, remove four screws to access and replace the two
batteries. Observe the two warning messages posted at the rear of the remote control.
• Requires two 1.5 V AAA batteries. If used in hazardous areas, replacement batteries
must be Energizer E92/EN92 or Duracell MN2400/PC2400.
• All functions for remote control PN 24479-00 are the same as those for the previous
remote control, PN 23572-00.
30 Rosemount 5081
Page 31
Reference Manual Display and operate
00809-0100-3581October 2019
5.3Menu system
There are three main menus: Calibrate, Program, and Diagnose. Calibrate and Program
menus have additional sub-menus as shown in the figures below.
Table 5-1: Program Menu
Displayed itemDefinition
OUtpUtCurrent output menu header
4MA4 mA current output (setpoint)
20MA20 mA current output (setpoint)
HoLdCurrent output on hold
FAULtFault condition current output setting
dPnCurrent output dampening time
tEStCurrent output test value
tAUtOAutomatic temperature compensation
tMAnManual temperature compensation
dISPLAYDisplay menu header
tYPMeasurement type
tEMP° C/° F toggle selection
OUtPUtCurrent (mA) or percent of full scale display
Figure 5-11: FOUNDATION Fieldbus Contacting Conductivity Menu Tree
Reference Manual 39
Page 40
5000 µS/cm
25.0C 12.00mA
CAL key
PROG key
DIAG key
HOLD key
Process Display
CALIBRATION
PROGRAM
DIAGNOSTICS
CALIbrAtE
SEnSOr 0
tEMP AdJ
CELL COnSt
tEMP SLOPE
tEMP
dISPLAY
SEtUP CuST
dEFAULT
AbS C
OFFSt
CELL COnSt
tSLOPE
CAL F
5081-T-FF
SoFt
HArd
FaULTs
Display and operateReference Manual
October 201900809-0100-3581
Figure 5-12: FOUNDATION Fieldbus Toroidal Conductivity Menu Tree
40 Rosemount 5081
Page 41
MAIN DISPLAY
CALIBRATEPROGRAMDIAGNOSE
sensor 0sensor Caltemp adjph Cal
(free chlorine only)
air cal
In process
press 760Grab spl
Cal 100
(O
2
only)
(All)
Temp 025.0
auto Cal
Man Cal
std pH
pH slope
Cal bf1
bf1 4.01
Cal bf2
bf2 10.00
Cal bf1
bf1 7.00
Cal bf2
bf2 7.00
std 07.00Slope 59.16
tempDEFAULT
Cal setup
Line freq
bar press
taUto On
tman 25.0
display
type --
Unit ppm
sensor stnd
Unit ppm
temp C
Output Cur
Code 000
if O
2
If Cl (free
and total)
If O
3
FACTORY NO
span Cal0 Cal
stabilize
limit 00.05
time 10SlOpe snGlSalnty 00.0
All
if chlorineif O
2
Line 60
PH ---
Pamp trans
Man 7.00
diaGnostiC
roffst 060
diaG OFF
lmptC On
GFI 1500
GFL 000
pH Cal
baUtO On
bUFFEr Std
StAbILISE
tIME 00
dELtA 0.02
Cl only
if On
if Off
Enter
Next
Unit nnHG
pman 760
type 02
sensor Cur
0 Current
bar press
5081-a-Ht
faUlts
MENU
SUB MENU
PROMPT
delta 0.05
sensitivty
Reference Manual Display and operate
00809-0100-3581October 2019
Figure 5-13: FOUNDATION Fieldbus Chorine, Dissolved Oxygen, and Ozone Menu Tree
Reference Manual 41
Page 42
Display and operateReference Manual
October 201900809-0100-3581
42 Rosemount 5081
Page 43
Reference Manual Programming basics
00809-0100-3581October 2019
6Programming basics
6.1Programming
The following can be programmed in the Rosemount™ 5081:
• 4-20 mA outputs
• Current generated by the transmitter during hold
• Current generated by the transmitter when fault is detected
• Automatic temperature correction (enable or disable)
• Type of measurement
• Measurement range
• Factory default setting
6.2Test 4-20 mA outputs and current generated
during hold and faults
Procedure
1. Press PROG on the remote controller.
The OUTPUT submenu appears.
2. Press ENTER.
The screen displays the 4mA prompt.
Reference Manual 43
Page 44
Programming basicsReference Manual
October 201900809-0100-3581
3. Use the arrow keys to change the setting. Press ENTER to save.
The screen displays the 20mA prompt.
4. Use the arrow keys to change the setting. Press ENTER to save.
The screen displays the HOLD prompt.
5. Use the arrow keys to change the setting to the output desired when the
transmitter is in hold mode.
The range is 3.80 to 22.0 mA. If you select 00.00, the transmitter will hold the
output value.
6. Press ENTER to save.
Note
The hold setting overrides the fault setting.
The screen displays the FAULT prompt.
7. Use the arrow keys to change the setting to the output desired when the
transmitter is in fault mode.
The range is 3.80 to 22.0 mA. If you select 00.00, the transmitter will hold the
output value.
8. Press ENTER to save.
The screen displays the DPN prompt.
9. Use the arrow keys to change the setting.
The range is 0 to 225.
44 Rosemount 5081
Page 45
Reference Manual Programming basics
00809-0100-3581October 2019
10. Press ENTER to save.
The screen displays the TEST prompt.
11. Use the arrow keys to enter the desired test current. Press ENTER to start the test.
Press EXIT to end the test.
12. Press RESET to return to the main display.
6.3Correct temperature
Procedure
1. Press PROG. Press NEXT until the TEMP submenu appears on the display.
2. Press ENTER.
The screen displays the T AUTO prompt.
3. Use the up or down arrow keys to enable (ON) or disable (OFF) the automatic
temperature correction feature. Press ENTER to save.
The screen displays the T MAN prompt.
4. Use the arrow keys to change to the desired temperature.
To enter negative numbers, press the left or the right arrow key until no digit is
flashing. then press the up or down arrow keys to display the negative sign. The
range is 23 to 266 ° F (-5.0 to 130 °C).
5. Press ENTER to save. Press RESET to return to the main display.
Reference Manual 45
Page 46
Programming basicsReference Manual
October 201900809-0100-3581
Note
If you disabled T AUTO in Step 3, then the transmitter will use the temperature
entered in this step in all subsequent measurements, no matter the actual process
temperature.
6.4Set up a custom curve (conductivity
measurements only)
Use custom curves to correlate conductivity to concentration of the measured liquid. The
Rosemount 5081 has programmable custom curves that can create a curve (second order)
from three to five user supplied points. If it only uses two points, then then the transmitter
creates a linear curve (straight lin). In order for the curve to be accurate, take the data
points from the liquid having the same reference temperature.
To obtain best results, select data points that are representative of the typical operating
range and have at least 5% difference in conductivity values. Observe the graph of
conductivity vs. concentration for the particular liquid to ensure that unsuitable points are
avoided. Unsuitable points include conductivity values with two concentrations associated
with them. In addition to unsuitable points, record any critical points - points that best
describe the curve. Following these general guidelines will provide optimal results.
Enter the first point (COND 1) at the normal operating condition. Then enter other points
above and below COND 1. Nonlinear conductivity curves require additional data points to
characterize these regions. Do not use the same data point more than once and only use
actual data (do not interpolate data points).
Note
The default values for the custom curve are three data points, reference temperature of
77 °F (25 °C), and a linear temperature slope of two percent/°C. This combination will yield
the best results in most applications. If the normal temperature is over 104 °F (40 °C) or
under 50 °F (10 °C), change the reference temperature to the normal process
temperature. If known, use the temperature slope at the reference temperature.
Procedure
1. From the main menu, select PROG and then press NEXT four times.
SETUP CUST appears.
2. Press ENTER.
T REF appears.
3. If needed, change the reference temperature from the factory default 77 °F (25 °C)
to a different reference temperature for the process. Press ENTER.
UNIT appears.
4. Press the up or down arrow to select the desired measurement units: µS
(microsiemens), mS (millisiemens), none (no units displayed), % (percent), or ppm
(parts per million); then press ENTER.
NUM PTS appears.
46 Rosemount 5081
Page 47
Reference Manual Programming basics
00809-0100-3581October 2019
5. Press up or down arrows to select the desired number of data points for a custom
conductivity curve. Select 2 to generate a linear relationship for conductivity and
concentration at the given temperature.
a) Enter the concentration for point 1 (displayed as µS 1). Press ENTER.
b) Enter the known conductivity for point 1 in µS/cm. Press ENTER.
c) Complete this process for additional known data points. Press ENTER.
CALC CUST appears briefly.
d) Press ENTER.
PROCESSING appears briefly; then APPLY CUST appears.
e) Press ENTER to register the custom curve into memory and return to the
SETUP CUST screen.
The custom curve will now be used to display and output all conductivity measurement
when the operator selects CUST as the measurement type in the Display menu.
6.5Restore to factory default settings
The operator can erase user-defined configurations to return the transmitter to the factory
default settings.
Procedure
1. Press PROG on the remote controller.
2. Press NEXT until DEFAULT appears in the display. Press ENTER.
3. Use the up and down arrows to toggle between NO and YES. Press ENTER to return
to the factory default settings.
6.6Set access (security) code
6.6.1Security overview
The access (security) code prevents program and calibration settings from accidental
changes. Emerson ships the transmitter with the access (security) code disabled.
6.6.2
Enter the access (security) code
Procedure
1. If calibration and program settings are protected with a security code, press PROG
or CAL on the infrared remote controller.
The ID screen appears.
2. Use the editing keys to enter the code. Press ENTER.
If the code is correct, the first sub-menu appears. If the security code is incorrect,
the process display reappears.
Reference Manual 47
Page 48
Programming basicsReference Manual
October 201900809-0100-3581
6.6.3Retrieve a lost security code
Procedure
1. If the you forget the security code, enter 555 at the ID prompt and press ENTER.
The transmitter displays the present code.
2. Press EXIT to return to the process display.
3. Press PROG or CAL.
The ID screen appears.
4. Use the editing keys to enter the security code. Then press ENTER.
The first sub-menu under the selected menu appears.
6.7Activate or deactivate HOLD
Activating HOLD keeps the transmitter output at the last value or sends the output to a
previously determined value. This is particularly useful when doing calibrations. During
calibrations, the sensor may be exposed to solutions that have concentrations outside the
normal range of the process. HOLD prevents false alarms and undesired operation as a
result of that reading. You can deactivate HOLD when the sensor is reinstalled in the
process stream and the readings have relatively stabilized.
Procedure
1. Press HOLD on the remote controller.
The HOLD prompt appears in the display.
2. Press 3 (up) or 5 (down) to toggle HOLD between ON and OFF.
3. Press ENTER to save.
48 Rosemount 5081
Page 49
Reference Manual Measurements
00809-0100-3581October 2019
7Measurements
7.1Calibrating pH sensors
7.1.1Calibration overview
The Calibration menu allows you to calibrate the pH and temperature response of the
sensor. To calibrate, use either the manual or the auto-calibration option. In both cases,
the transmitter does a two-point calibration for pH and one-point standardization against
a reference thermometer for temperature. In auto-calibration, the transmitter
automatically stores the temperature-corrected calibration data. In manual calibration,
you have to enter buffer values. The values are only registered when the readings are
stable (automatically determined by the transmitter). All calibration procedures are
guided by prompts on the display.
7.1.2
7.1.3
Calibration standards (buffer solutions)
• Calibrations are critical to make accurate pH measurements.
• Emerson recommends that the value of one of the pH buffer solutions is lower than the
pH of the process stream and the value of the other pH buffer solution is higher than
the pH of the process stream.
• For best results, make sure that the temperature of the solution is the same
temperature as that of the sensor. Allow the entire measurement cell, sensor, and
solution to reach relatively constant temperature.
• Using buffers at high temperatures can cause evaporation, which will change the
concentration of the buffer. The change in concentration causes a change in the pH,
thus resulting in an inaccurate calibration.
• Buffers have limited shelf lives. Do not use a buffer if the expiration date has passed.
Store buffers at controlled room temperatures.
• Do not reuse buffer solutions. Protect buffers from excessive exposure to air. Exposure
to air can cause changes in the pH of the buffer solution.
• Always rinse the sensor with DI water and remove excess water by dabbing with a clean
tissue before placing it in a buffer. Only dab the sensor and do not wipe it. Wiping the
sensor builds static charge which alters the reading.
Auto-calibrate
Prerequisites
Ensure that auto-calibration is turned ON and the appropriate buffers are used.
Procedure
1. Press CAL on the IRC to enter the CALIBRATE menu.
Reference Manual 49
Page 50
MeasurementsReference Manual
October 201900809-0100-3581
The CALIBRATE sub-menu appears.
2. At the CALIBRATE sub-menu, press ENTER.
The CAL bF1 prompt appears.
3. Rinse the sensor and place it in the first buffer. Be sure the glass bulb and the
temperature element are completely submerged. Keep the sensor at least 3 in.
below the liquid level. Do not rest the sensor on the glass bulb. Dislodge any
trapped bubbles by swirling the sensor body.
4. Press ENTER.
BF1 flashes until the measured pH meets the programmed stability limits. If the pH
reading is not relatively stable after 20 minutes, the transmitter automatically
leaves the CALIBRATE menu and returns to the main display. If this occurs, consult
Diagnostics and troubleshooting for assistance. Once the reading is stable, the
display moves forward by showing a flashing number with the nominal pH.
5. Change the value by using the up and down arrows until the correct pH of the buffer
solution appears. Press ENTER to save the first calibration point.
CAL BF2 appears.
6. Remove the sensor from the first buffer.
7. Rinse the sensor and place it in the second buffer solution. Again, make sure the
bulb is completely submerged and the bubbles are dislodged by swirling the sensor
body.
8. Press ENTER.
BF2 flashes until the reading is stable.
9. Repeat steps Step 4 and Step 5.
The calibration is now complete, but the transmitter remains in the CALIBRATE submenu for two minutes after you press ENTER.
10. Remove the sensor from the buffer and reinstall it into the process stream. If the
HOLD feature was used, be sure to turn off HOLD.
The sensor will not calibrate if the electrode slope (calculated by the transmitter
during calibration) is unacceptable. The transmitter displays a SLOPE ERR HI or
SLOPE ERR LO error message. Refer to Diagnostics and troubleshooting for
assistance.
Note
To display the electrode slope, press CAL on the IRC. The CALIBRATE sub-menu
appears. Press NEXT. The STD sub-menu appears. Press ENTER. The STD prompt
appears. Press ENTER again, and the slope appears on the display. For a good
sensor, the slope is generally 50 to 60 mV.
7.1.4
50 Rosemount 5081
Calibrate manually
Prerequisites
Ensure that auto-calibration is OFF and the appropriate buffers are used.
Procedure
1. Press CAL on the IRC to enter the CALIBRATE menu.
Page 51
Reference Manual Measurements
00809-0100-3581October 2019
The CALIBRATE sub-menu appears.
2. At the CALIBRATE sub-menu, press ENTER.
The CAL BF1 prompt appears.
3. Rinse the sensor with DI water and place it in the first buffer along with a calibrated
thermometer. Submerge the sensor tip at least 3 in. (76 mm) below the liquid level.
Do not rest the sensor on the glass bulb. Dislodge any trapped bubbles by swirling
the sensor body.
4. Once the pH reading and temperature are relatively stable, press ENTER. Use the
editing keys to change the flashing display to the appropriate pH of the buffer
solution. Press ENTER to save the value as buffer BF1.
The transmitter expects a reading to be entered within 20 minutes after the CAL
BF1 prompt. If you do not press ENTER, the transmitter exits the CALIBRATE menu
and returns to the process mode.
5. At the CALBF2 prompt, remove the sensor from the first buffer. Repeat steps Step 3
and Step 4.
The calibration is now complete, but the transmitter remains in the CALIBRATE submenu for two minutes after you press ENTER.
6. Remove the sensor from the buffer and reinstall it into the process stream. If the
HOLD feature was turned on, be sure to turn off HOLD.
The sensor will not calibrate if the electrode slope (calculated by the transmitter
during calibration) is unacceptable. The transmitter displays a SLOPE ERR HI or
SLOPE ERR LO error message. Refer to Diagnostics and troubleshooting for
assistance.
7.1.5
Note
To display the electrode slope, press CAL on the IRC. The CALIBRATE sub-menu
appears. Press NEXT. The STD sub-menu appears. Press ENTER. The STD prompt
appears. Press ENTER again, and the slope appears on the display. For a good
sensor, the slope is generally 50 to 60 mV.
Standardize for pH
Use standardization to match the transmitter values with those of another transmitter.
The transmitter converts the difference between the two pH values into an equivalent
voltage, called the reference offset.
Note
If a sensor that has been calibrated with buffers is standardized, then placing it back in the
buffer will show a different measured pH than that of the buffer due to the standardization
offset.
Procedure
1. Press CAL on the IRC to enter the CALIBRATE menu.
The CALIBRATE submenu appears.
2. At the CALIBRATE submenu, press NEXT.
The STD submenu appears.
3. Press ENTER to enter the STD prompt.
Reference Manual 51
Page 52
MeasurementsReference Manual
October 201900809-0100-3581
The process pH and temperature should be relatively stable.
4. Take a grab sample from the process stream as close to the pH sensor as possible.
This transmitter reading is called the pH_trans.
5. Measure the pH of the sample (pH_std) using the second pH meter.
6. Note the current process reading (pH_curr). Calculate the corrected reading from
the equation: ph_corr = ph_curr + (pH_std - pH_trans)
• pH_corr = corrected pH value
• pH_curr = current pH value
• pH_std = standard instrument's pH value
• pH_trans = measured sample's pH value
7. Use the editing keys to enter the pH_corr value calculated using the equation
above. Press ENTER to save the value.
If the corrected value is acceptable, then the display shows the slope (current
electrode slope).
8. If the slope is acceptable, press EXIT. If the slope is unacceptable, use the editing
keys to change it. Then press ENTER.
ORP is a function of temperature. The accuracy of a sensor/transmitter loop is about ±1 °C.
A new sensor does not need to be calibrated often. Only calibrate the loop when:
1. ±1 °C is not acceptable.
2. You suspect an error in temperature measurement
7.2.2
Calibrate temperature with an oxidation reduction
potential (ORP) sensor
Procedure
1. Place the ORP sensor and a reference thermometer in an insulated container of
water at ambient temperature. Keep the sensor submerged at least 3 in. (76 mm)
below the water level. Stir continuously and wait at least 20 minutes for the water,
sensor, and thermometer to reach a constant temperature.
2. Press CAL on the IRC to enter the CALIBRATE menu.
The STD submenu appears.
3. At the STD submenu, press NEXT.
The TEMP ADJ submenu appears.
52 Rosemount 5081
Page 53
Reference Manual Measurements
00809-0100-3581October 2019
4. Press ENTER to display the temperature editing prompt.
5. Compare the temperature displayed by the transmitter with the temperature
measured by the reference thermometer. If there is a discrepancy between the two
values, use the editing keys to change the measured value to that of the
thermometer.
Note
The reading cannot be changed more than 15 °C.
6. Press ENTER.
The transmitter saves the value, and the display returns to the TEMP ADJ submenu.
7. Press EXIT to leave the CALIBRATE menu.
8. Check for linearity by measuring the temperature of water: -10 to -15 °C and +10 to
+15 °C than the water used for calibration.
7.2.3
7.2.4
Standardize with an oxidation reduction potential (ORP)
sensor
ASTM d 1498-93 gives procedures for making iron (ii) - iron (iii) and quinhydrone ORP
standards. Emerson recommends the iron (ii) - iron (iii) standard. It is fairly easy to make
and has a shelf life of approximately one year; in contrast, quinhydrone standards contain
toxic quinhydrone and have only an eight hour shelf life. Iron (ii) - iron (iii) standard is
available from Rosemount™ as PN R508-16OZ. The ORP of the standard solution measured
against a silver-silver chloride reference electrode is 476 ± 20 mV at 77 °F (25 °C).
Procedure
1. Press CAL on the IRC to enter the CALIBRATE menu.
The STD submenu appears.
2. Rinse the sensor with DI water and place it in the ORP standard with a reference
thermometer. Ensure that the ORP sensor is submerged at least 3 in. (76 mm)
below the water level.
3. Once the temperature and ORP readings are stable, press ENTER.
4. Use the editing keys to change the flashing display to the desired ORP reading.
Press ENTER to save.
5. Press EXIT to return to the main display.
Match transmitter to oxidation reduction potential
(ORP) sensor's temperature element
Procedure
1. Press PROG on the IRC.
2. Press NEXT until the TEMP submenu appears in the display. Press ENTER.
3. Press EXIT to return to the main display.
Reference Manual 53
Page 54
MeasurementsReference Manual
October 201900809-0100-3581
7.3Calibrating contacting and toroidal
conductivity sensors
7.3.1Overview of calibrating contacting conductivity sensors
Contacting conductivity sensors do not require frequent calibrations. The sensors are
ready to be installed right out of the box.
7.3.2Calibrate contacting conductivity sensors
Procedure
1. Pres CAL on the IRC. Press ENTER.
The CALIBRATE submenu appears.
2. Place the sensor in a standard solution of known conductivity and allow the
measurement reading to become relatively stable.
3. Press ENTER to access the CAL segment with the flashing prompt.
4. Use the IRC editing keys to enter the conductivity value of the standard solution.
Press ENTER.
5. Press EXIT to return to the main display.
7.3.3
7.3.4
Zero the sensor
Procedure
1. Press CAL and then NEXT to enter the SENSOR 0 menu.
2. Press ENTER to access the SENSOR 0 submenu.
3. Hold the sensor in the air and press ENTER again to zero the sensor.
4. Press EXIT to return to the SENSOR 0 submenu.
Adjust temperature
Procedure
1. Press CAL and then NEXT until you see TEMP.
2. Press ENTER.
The TEMP submenu appears.
3. Place the sensor in any solution of known temperature. Allow the temperature of
the sensor to become relatively stable.
4. Use the editing keys of the IRC to change the values as needed.
5. Press ENTER to standardize the temperature reading and return to the TEMP ADJ
screen.
6. Press EXIT to return to the main display.
54 Rosemount 5081
Page 55
Reference Manual Measurements
00809-0100-3581October 2019
7.3.5Calibrate in-process
Procedure
1. Press CAL and then NEXT three times to access MTR CAL.
2. Use the IRC to go through the on-screen prompts.
3. Press EXIT to return to the main display.
7.3.6Calibrate temperature slope
Procedure
1. From the CAL menu, press NEXT until TEMP SLOPE displays.
2. Use the IRC arrow keys to enter the slope. Press ENTER to register the slope in the
memory of the transmitter. Press EXIT to return to the main screen.
3. If you don't know the temperature slope of the process, then refer to the below
guide:
7.3.7
• Acids: 1.0 percent to 1.6 percent per °C
• Bases: 1.8 percent to 2.2 percent per °C
• Salts: 2.2 percent to 3.0 percent per °C
• Water: 2.0 percent per °C
4. Press ENTER to proceed to the T SLOPE submenu with the flashing prompt. Use the
IRC keys to generate the desired slope value. Press ENTER.
5. Press EXIT to return to the main screen.
Calibrate output
Procedure
1. Wire an accurate milliammeter as shown in Figure 7-1.
Figure 7-1: Milliammeter Wiring
2. Press CAL on the IRC.
3. Press NEXT until the OUTPUT CAL submenu appears. Press ENTER.
Reference Manual 55
Page 56
MeasurementsReference Manual
October 201900809-0100-3581
4. Use the arrow keys to change the display to match the reading from the
milliammeter. Press ENTER.
5. Press RESET to return to the main display.
7.4Calibrating dissolved oxygen, ozone, free
chlorine, total chlorine, and ozone sensors
7.4.1Calibrating amperometric sensors
You can calibrate amperometric sensors in multiple ways: zeroing, air calibrations, dual
slope calibrations, and in process calibrations.
7.4.2
Zero the sensor
Procedure
1. Place the sensor in a solution of 5% sodium sulfite (Na2SO3) in water. Ensure that air
bubbles are not trapped against the membrane.
2. Go to the main display; press DIAG and then NEXT.
The SENSOR CUR prompt appears.
3. Press ENTER to view the sensor current. Make sure the sensor reaches its zero
current.
This may require several hours. Do not start the zero routine until the sensor has
been in zero solution for at least two hours.
4. Press CAL on the IRC.
The SENSOR 0 prompt appears.
5. Press ENTER.
The screen shows the appropriate value correlating to the zero value. The screen
shows 0.02. The reading must be below 0.02 ppm for the zero calibration to be
accepted.
6. To change the zero limit, use the editing keys and then press ENTER.
The TIME DELAY message appears and remains until the zero current is below the
concentration limit. If the current is already below the limit, TIME DELAY will not
appear.
7. To bypass the time delay, press ENTER.
When the procedure is complete, 0 DONE appears.
8. Press EXIT.
9. Press RESET to return to the main display.
56 Rosemount 5081
Page 57
Reference Manual Measurements
00809-0100-3581October 2019
7.4.3Calibrate in air (dissolved oxygen)
Procedure
1. Remove the sensor from the process liquid. Use a soft tissue and a wash bottle to
clean the membrane. Dry it by blotting.
The membrane must be completely dry during air calibration.
2. Pour some water into a beaker and suspend the sensor with the membrane
approximately ½ in. (approximately 1 cm) above the water surface.
Keep the sensor out of direct sunlight.
3. Monitor the dissolved oxygen and temperature reading. Once the readings are
relatively stable, begin the calibration.
It may take 10 - 15 minutes for the sensor reading in the air to stabilize.
4. Press CAL on the IRC.
5. Press NEXT.
The SENSOR CAL submenu appears.
6. Press ENTER.
The AIR CAL prompt appears.
7. Press ENTER.
8. Select the units. Then press NEXT.
9. Use the arrow keys to enter the barometric pressure. Press ENTER when done.
The TIME DELAY message appears and remains until the sensor reading is
relatively stable.
10. To bypass the time delay, press ENTER.
CAL DONE appears when the calibration is complete.
11. Press EXIT.
12. To return to the main display, press RESET.
7.4.4
Reference Manual 57
Calibrate in process (dissolved oxygen)
You can calibrate the sensor against a standard instrument.
Procedure
1. Press CAL on the IRC.
2. Press NEXT.
The SENSOR CAL submenu appears.
3. Press ENTER.
4. Press NEXT.
The AIR CAL prompt appears.
5. Press NEXT.
The INPROCESS prompt appears.
6. Press ENTER.
Page 58
MeasurementsReference Manual
October 201900809-0100-3581
The TIME DELAY message appears and will remain until the sensor is relatively
stable. To bypass the time delay, press ENTER. The GRAP SPL message appears.
7. Press ENTER.
8. Use the arrow keys to change the flashing display to the value indicated by the
standards instrument. Press ENTER.
9. Press RESET to return to the main display.
7.4.5Calibrate full scale (free chlorine, total chlorine, ozone)
Procedure
1. Place the sensor in the process liquid. Ensure that the pH sensor is calibrated or the
pH value is entered in case of manual pH correction, if applicable.
2. Adjust the chlorine concentration until it is near the upper end of the control range.
Wait until the reading is relatively stable before starting the calibration.
3. Press CAL on the IRC.
4. Press NEXT.
The SENSOR CAL submenu appears.
5. Press ENTER.
The TIME DELAY message appears and will remain until the sensor reading is
relatively stable. To bypass the time delay, press ENTER. The GRAB SPL prompt
appears.
6. Take a sample of the process liquid and determine the concentration of chlorine in
the sample. Press ENTER.
7. Press RESET to return to the main display.
7.4.6
58 Rosemount 5081
Calibrate dual slope (free chlorine, total chlorine)
Procedure
1. Zero the sensor.
2. Place the sensor in the process liquid. Ensure that the pH sensor is calibrated or the
pH valve is entered in case of manual pH correction, if applicable.
3. Press CAL on the IRC. Press NEXT.
The SENSOR CAL prompt appears.
4. Press ENTER.
The CAL PT1 prompt appears.
5. Adjust the chlorine concentration until it is near the upper end of the linear range of
the sensor. Press NEXT.
The TIME DELAY message appears and will remain until the sensor reading is
relatively stable. To bypass the time delay, press ENTER. The GRAB SPL prompt
appears.
6. Take a sample of the process liquid and determine the concentration of chlorine in
the sample.
7. Press ENTER.
Page 59
Reference Manual Measurements
00809-0100-3581October 2019
The PT1 prompt appears.
8. Use the arrow keys to change the flashing display to the concentration of chlorine
determined in the grab sample. Press ENTER to save.
The CAL PT2 prompt appears.
9. Adjust the concentration of chlorine until it is near the top end of the range. Press
ENTER.
The TIME DELAY message appears and will remain until the sensor reading is
relatively stable. To bypass the time delay, press ENTER. The GRAB SPL prompt
appears.
10. Take the sample of the process liquid and determine the concentration of chlorine
in the sample.
11. Press ENTER.
The PT2 prompt appears.
12. Use the arrow keys to change the flashing display to the concentration of chlorine
determined in the grab sample. Press ENTER to save.
13. Press RESET to return to the main display.
Reference Manual 59
Page 60
MeasurementsReference Manual
October 201900809-0100-3581
60 Rosemount 5081
Page 61
Reference Manual Diagnostics and troubleshooting
00809-0100-3581October 2019
8Diagnostics and troubleshooting
8.1Warning and fault messages
The Rosemount™ 5081 transmitter continuously monitors the measurement loop (sensor
and transmitter) for conditions that cause inaccurate measurements. When a problem
occurs, the transmitter displays either a warning or fault message. A warning alerts you
that a potentially system disabling condition exists. If you don't fix the problem, there is a
high probability that the measurement will be incorrect. A fault alerts you that a system
disabling condition exists. If a fault message is showing, regard all measurements as
incorrect.
When a warning condition exists:
1. The main display remains stable; it does not flash.
2. A warning message appears alternately with the temperature display. See Figure
8-1. See Troubleshooting when a diagnostic message is showing for an explanation
of the different warnings and suggested ways of correcting the problem.
Figure 8-1: Warning Annunciation
When a Fault exists:
1. The main display flashes.
2. The words FAULT and HOLD appear in the main display.
3. A fault message appears alternately with the temperature/output display. See
Figure 8-2. See Troubleshooting when a diagnostic message is showing for an
explanation of the different fault messages and suggested ways of correcting the
problem.
4. The output current will remain at the present value or go to the programmed fault
value.
5. If the transmitter is in Hold when the fault occurs, the output remains at the
programmed hold value. To alert you that a fault exists, the word FAULT appears in
the main display, and the display flashes. A fault or diagnostic message also
appears.
Reference Manual 61
Page 62
Diagnostics and troubleshootingReference Manual
October 201900809-0100-3581
6. If the transmitter is simulating an output current when the fault occurs, the
transmitter continues to generate the simulated current. To alert you that a fault
exists, the word FAULT appears in the display, and the display flashes.
Figure 8-2: Fault Annunciation
When a a fault occurs, the display appears as pictured above. To further alert you that
measurements are in error, the display flashes. Diagnostic messages appear in the
temperature/output area on the screen.
8.2pH/ORP diagnostics and troubleshooting
8.2.1Calibration errors
If an error occurs during calibration, an error message appears in the main display and the
transmitter does not update the calibration. The calibration errors are Std Err, SLOPE
Err LO, and SLOPE Err HI. See Troubleshooting when a diagnostic message is showing
for an explanation of the error messages and suggested ways of correcting the problem.
8.2.2
8.2.3
General troubleshooting
Procedure
1. Look for a diagnostic message on the display to help identify the problem.
Refer to Troubleshooting when a diagnostic message is showing for an explanation
of the message and a list of the possible problems that triggered it.
2. Refer to Troubleshooting when no diagnostic message is showing for common
measurement problems and the recommended actions to resolve them.
3. Follow the step-by-step troubleshooting approach, offered in Displaying diagnostic
variables, to diagnose and correct less common or more complex problems.
Troubleshooting when a diagnostic message is showing
The Rosemount 5081-P pH/ORP transmitter continuously monitors the measurement
loop (sensor and transmitter) for problems. If a problem is detected, the transmitter
62 Rosemount 5081
Page 63
Reference Manual Diagnostics and troubleshooting
00809-0100-3581October 2019
displays a fault or error message. The message appears in the temperature/output area of
the main display. The table lists each diagnostic message and the section to consult for
help.
MessageSection
GLASSFAILGLASSFAIL
GLASSWArnGLASSWArn
rEF FAILrEF FAIL
rEF WArnrEF WARn
CALIbrAtECALIbrAtE
tEMP HItEMP HI and tEMP LO
tEMP LOtEMP HI and tEMP LO
LInE FAILLinE FAIL
InPUt WArnInPUt WArn
SLOPE Err LOSLOPE Err LO
SLOPE Err HISLOPE Err HI
Std ErrStd Err
rOM FAILrOM FAIL or CPU FAIL
CPU FAILrOM FAIL or CPU FAIL
AdC WArnAdC WArn or CyCLE PWr
CyCLE PWrAdC WArn or CyCLE PWr
WrItE ErrWritE Err
FACt FAILFACt FAIL
GLASSFAIL
GLASSFAIL is an electrode fault message. It means the glass impedance is outside the
programmed Glass Fault High (GFH) or Glass Fault Low (GFL) limit. Glass Fault High
suggests the electrode is aging or the electrode is not immersed in the process liquid.
Glass Fault Low implies that the pH sensitive glass is cracked. GLASSFAIL also appears if
inappropriate limits have been entered into the transmitter.
If the measurement system was previously commissioned and operating correctly,
GLASSFAIL likely means a real problem exists. However, if the system is being started up
or if the advanced diagnostic feature is being used for the first time, GLASSFAIL could be
caused by a miswired sensor or by programmed limits that are not correct for the sensor.
Note
GLASSFAIL is a sensor diagnostic message. Sensor diagnostic messages are optional.
They can be turned off.
Recommended actions
1. Be sure the sensor is completely immersed in the process liquid.
Reference Manual 63
Page 64
Diagnostics and troubleshootingReference Manual
October 201900809-0100-3581
If the diagnostic message disappears, the sensor is in good condition.
If the diagnostic message remains, go to Step 2.
2. Measure the glass impedance. See Testing the transmitter by simulating the pH for
the procedure. Note the reading.
If the glass impedance is low (< 40 megohms)...
a. Check preamp location in program menus (PAMP = _____).
• If the location is incorrect, go to 2.b.
• If after selecting the correct location of Preamp in the Program menu, the
glass impedance is still low, go to Step b.
b. Calibrate the sensor.
• If the sensor calibrates properly...
1. The sensor is in good condition, but the Glass Fail Low (GFL) limit is
set too high.
2. Lower the GFL limit to about 10 megohms below the glass
impedance value (GIMP).
3. If the Glass Warning Low (GWL) message was also flashing, lower
the limit from its former value by the same amount the GFL was
lowered from its former value.
• If the sensor cannot be calibrated, the pH glass membrane is likely
cracked and the sensor must be replaced. The crack in the glass may not
be visible or may be difficult to see.
If the glass impedance is high (> 800 megohms)...
a. Check that the sensor is correctly wired to the transmitter. Pay particular
attention to the following:
1. For Rosemount PLUS (+) and TUpH sensors with integral preamplifiers,
the blue solution ground wire must be attached to TB-8 (SOL GND),
and the gray reference in wire must be attached to TB-7 (REF IN).
Note
TB-8 means terminal 8 on the terminal board.
2. If the sensor was wired with the blue solution ground wire unattached
and a jumper between the terminals TB-8 and TB-7, remove the
jumper and reattach the blue solution ground wire to TB-8. Keep the
gray reference in wire attached to TB-7.
3. For Rosemount PLUS (+) and TUpH sensors that do not have an
integral preamplifier, attach the blue solution ground wire to TB-8 or,
better, leave the blue wire unattached and jumper TB-7 to TB-8.
4. If the sensor does not have a blue solution ground wire, jumper
terminals TB-7 and TB-8.
• If the wiring was correct and the glass impedance is still too high,
go on to 2.b.
64 Rosemount 5081
Page 65
Reference Manual Diagnostics and troubleshooting
00809-0100-3581October 2019
• If correcting wiring errors causes the diagnostic message to
disappear, the sensor is in good condition.
• If after correcting wiring errors, the glass impedance is still high, go
on to Step b.
b. Inspect and clean the sensor. After cleaning the sensor, calibrate it. Be sure to
note the sensor slope.
• If cleaning the sensor lowers the impedance below 800 megohms:
1. The sensor is in good condition.
2. Return the calibrated sensor to service.
• If cleaning does not lower the glass impedance and the sensor can be
calibrated:
1. The sensor is probably in good condition; however, it may be
nearing the end of its life. The electrode slope is a good indicator of
remaining life.
SlopeCondition of sensor
54-60 mV/unit pHSensor is in good condition.
48-50 mV/unit pHSensor is nearing the end of its life. Once the
slope drops below 48 mV/unit pH, the sensor
can no longer be calibrated.
2. The Glass Fail High (GFH) limit is probably set too low for the
sensor. Set the GFH limit to about 150 megohms greater than the
measured glass impedance.
3. If the GLASSWArn message was also flashing, raise the GWH limit
from its former value by the same amount the GFH was raised from
its former value.
• If cleaning does not lower the glass impedance and the sensor cannot be
calibrated, the sensor has failed and should be replaced.
GLASSWArn
GLASSWArn is an electrode fault message. It means the glass impedance is outside the
programmed Gas Warning High (GWH) or Glass Warning Low (GWL) limit. Ideally, when
the measurement system exceeds the glass warning limits, you will have adequate time to
diagnose and correct problems before a failure occurs. High impedance implies the
electrode is aging or the sensor is not completely submerged in the process liquid. Low
impedance suggests the pH sensitive glass is cracked. The message also appears if
inappropriate limits have been entered into the transmitter.
If the measurement system was previously commissioned and operating correctly,
GLASSWArn likely means a real problem exists. However, if the system is being started up
or if the advanced diagnostic feature is being used for the first time, GLASSWArn could be
caused by a miswired sensor or by programmed limits that are not correct for the sensor.
Reference Manual 65
Page 66
Diagnostics and troubleshootingReference Manual
October 201900809-0100-3581
Note
GLASSWArn is a sensor diagnostic message. All sensor diagnostic messages are optional.
They can be turned off.
Troubleshooting GLASSWArn problems is exactly the same procedure as troubleshooting
GLASSFAIL problems. Refer to GLASSFAIL.
rEF FAIL
rEF FAIL is an electrode fault message. rEF FAIL means that the reference impedance
exceeds the programmed reference high fault (RFH) limit. A plugged or dry reference
junction is the usual cause of a high reference impedance. High reference impedance also
occurs if the sensor is not submerged in the process liquid or if inappropriate limits have
been entered into the transmitter.
If the operator previously commissioned the measurement system and it was operating
correctly, rEF FAIL likely means a real problem exists. However, if the operator is just
starting up the system or using the advanced diagnostic feature for the first time, rEF
FAIL could be caused by a miswired sensor or by programmed limits that are not correct
for the sensor.
Note
rEF FAIL is a sensor diagnostic message. All sensor diagnostic messages are optional.
They can be turned off.
Recommended actions
1. Be sure the sensor is completely immersed in the process liquid.
• If the diagnostic message disappears, the sensor is in good condition.
• If the diagnostic message remains, go to Step 2.
2. Check that the sensor is properly wired to the transmitter. Be sure the reference in
wire is attached to TB-7 and the solution ground wire is attached to TB-8.
Note
TB-8 means terminal 8 on the terminal board.
• If correcting wiring problems makes the diagnostic message disappear, the
sensor is in good condition.
• If the wiring is correct and the message still remains, go to Step 3.
3. Measure and make a note of the reference impedance (rIMP). See Testing the
transmitter by simulating the pH.
• If the reference impedance is low (< 70 kilohms)...
a. The reference electrode is in good condition. pH sensors manufactured
by Rosemount use low impedance silver/silver chloride reference
electrodes.
b. The reference failure high (RFH) limit is probably set too low. Change the
limit to a value about 50 kilohms greater than the measured reference
impedance. If rEF WARN was also displayed, change the reference
66 Rosemount 5081
Page 67
Reference Manual Diagnostics and troubleshooting
00809-0100-3581October 2019
warning high (RWH) limit to about 25 kilohms above the measured
reference impedance.
• If the reference impedance is high (> 70 kilohms)...
a. The sensor may be dirty, in which case cleaning it will lower the
impedance. If the sensor is rebuildable, the reference electrolyte may be
depleted. Finally, the sensor may be in good condition. The warning and
failure limits are simply set too high.
b. Inspect and clean the sensor. If the sensor is rebuildable, replace the
reference junction and replenish the electrolyte solution. Refer to the
sensor reference manual for details. Check the reference impedance
again.
— If cleaning the sensor reduces the impedance...
1. The sensor is in good condition. Calibrate the sensor and return
it to the process.
2. Change the reference failure high (RFH) limit to a value about
50 kilohms greater than the measured reference impedance. If
rEF WARN was also displayed, change the reference warning
high (RWH) limit to about 25 kilohms above the measured
reference impedance.
— If cleaning does not reduce the impedance and the sensor is not
rebuildable...
1. Try the reference junction rejuvenation procedure. The
rejuvenation procedure may not work. At best, it will get a little
more life out of a sensor with a plugged reference.
2. Whether or not the rejuvenation procedure worked, go to 3.c.
c. Recalibrate the sensor using the autocalibration procedure.
— If the sensor can be calibrated:
1. The sensor is in good condition. Return it to the process.
2. Change the reference failure high (RFH) limit to a value about
50 kilohms greater than the measured reference impedance. If
rEF WARN was also displayed, change the reference warning
high (RWH) limit to about 25 kilohms greater than the
measured reference impedance.
— If the sensor cannot be calibrated, the sensor has failed and must be
replaced.
rEF WARn
rEF WArn is an electrode fault message. It means the reference electrode impedance
exceeds the programmed reference warning high (RWH) limit. Ideally, when the
measurement system exceeds the warning limits, you will have adequate time to diagnose
and correct problems before a failure occurs. A high reference impedance implies that the
liquid junction is plugged or the reference electrolyte is depleted. The message also
appears if an inappropriate limit has been entered into the transmitter.
Reference Manual 67
Page 68
Diagnostics and troubleshootingReference Manual
October 201900809-0100-3581
If the measurement system was previously commissioned and operating correctly, rEF
WArn likely means a real problem exists. However, if the system is being started up or if the
advanced diagnostic feature is being used for the first time, rEF WArn could be caused by
a miswired sensor or by programmed limits that are not correct for the sensor.
Note
rEF WArn is a sensor diagnostic message. Sensor diagnostic messages are optional. They
can be turned off.
Troubleshooting rEF WArn problems is exactly the same as troubleshooting rEF FAIL
problems. Refer to rEF FAIL.
CALIbrAtE
CALIbrAte is a diagnostic intended for future use. If the CALIbrAte message is showing,
disable CALIbrAte.
tEMP HI and tEMP LO
tEMP HI and tEMP LO mean the transmitter has detected a problem with the
temperature measuring circuit. The problem may lie in the sensor, the cable, or the
transmitter. The determination of temperature is an integral part of the pH measurement.
Therefore, failure of the temperature measuring circuit is a system disabling condition.
However, in an emergency, automatic temperature compensation can be disabled and
the transmitter placed in manual temperature compensation. For manual temperature
compensation, choose a temperature equal to the average temperature of the process.
The resulting pH reading will be in error. The more variable the temperature and the
further the pH from 7, the greater the error.
Recommended actions
1. Check wiring, jumper settings, and software settings.
a. Check the wiring between the sensor and the transmitter. Pay particular
attention to TB=3 (RTD RTN), TB-4 (RTD SN), and TB-5 (RTD RTN ).
Note
TB-3 means terminal 3 on the terminal board.
b. Be sure the software settings match the type of resistance temperature
device (RTD) in the sensor.
• If the diagnostic message disappears, the sensor is in good condition.
• If the message persists, go to Step 2.
2. Check the sensor. Disconnect the RTD leads and measure the resistances shown in
Figure 8-3.
The measured resistance should agree with the value in Table 8-1 to within about
one percent. If the measured resistance is appreciably different (between one and
five percent) from the value shown, the operator can calibrate out the discrepancy.
68 Rosemount 5081
Page 69
Reference Manual Diagnostics and troubleshooting
00809-0100-3581October 2019
Figure 8-3: Three-Wire RTD
A. Compare resistance between these wires to the values in Table 8-1.
B. Resistance between these wires should be less than 2 ohms.
Consult Table 8-1 for resistance-temperature data. Lead resistance is about 0.05
ohm/ft at 77 °F (25 °C). Therefore, 15 feet of cable increases the resistance by about
1.5 ohm. The resistance between the RTD return and RTD sense leads should be less
than 2 ohms.
Table 8-1: RTD Resistance Values
TemperaturePt-100 resistancePt-1000 resistance
0 °C (32 °F)100.0 ohms1000 ohms
10 °C (50 °F)103.9 ohms1039 ohms
20 °C (68 °F)107.8 ohms1078 ohms
25 °C (77 °F)109.6 ohms1096 ohms
30 °C (86 °F)111.7 ohms1117 ohms
40 °C (104 °F)115.5 ohms1155 ohms
50 °C (122 °F)119.4 ohms1194 ohms
60 °C (140 °F)123.2 ohms1232 ohms
70 °C (158 °F)127.1 ohms1271 ohms
80 °C (176 °F)130.9 ohms1309 ohms
90 °C (194 °F)134.7 ohms1347 ohms
100 °C (212 °F)138.5 ohms1385 ohms
• If a connections is open or shorted and it should not be, the sensor has failed.
Replace the sensor.
Reference Manual 69
Page 70
Diagnostics and troubleshootingReference Manual
October 201900809-0100-3581
• If the measured resistances are acceptable, go to Step 3.
3. Check the transmitter. Disconnect the RTD sensor leads and wire the circuit shown
in Figure 8-4. Set the resistance to the value for 25 °C (77 °F) shown in Table 8-1. The
measured temperature should equal 25 °C (77 °F) within ±1 °C.
Figure 8-4: Temperature Simulation into the Rosemount 5081 pH/ORP
Transmitter
A. Terminal board
B. RTD return
C. RTD sense
D. RTD in
E. Jumper
F. Standard resistor(s) to stimulate RTD
• If the measured temperature is correct, the transmitter is working properly.
• If the measured temperature is incorrect, calibrate the transmitter against the
standard resistance equivalent to 77 °F (25 °C). Change the resistance and verify
that the temperature reading changes to the correct value.
— If the transmitter works properly after temperature calibration, the original
calibration was in error. Re-attach the RTD wires and check the temperature
performance of the sensor.
70 Rosemount 5081
Page 71
Reference Manual Diagnostics and troubleshooting
00809-0100-3581October 2019
— If the reading is still wrong, the transmitter electronics have failed. Replace
the sensor board stack.
LinE FAIL
LinE FAIL almost always means the transmitter is measuring an incorrect distance
between terminal TB-3 (resistance temperature device [RTD] return) and TB-4 (RTD
sense). These terminals are critical connections for the three-wire RTD measurement.
Figure 8-3 shows a three-wire RTD connection.
Recommended actions
1. Check for miswires and open connections at TB-3 and TB-4. Open connections can
be caused by loose connections, poor spade crimps, or broken wires. Be sure the
check junction boxes for proper pass through of all wires.
• If correcting a wiring problem makes the message disappear, the system is in
good condition.
• If the message is still showing, go to Step 2.
2. The RTD sense or the RTD return wire inside the sensor cable may be broken. Keep
the sensor wires attached and jumper TB-3 and TB-4.
• If the diagnostic message disappears, either the RTD return or RTD sense wire is
broken. To verify a broken wire, disconnect the leads and measure the resistance
between them. Installing the jumper completes the circuit, but bypasses the
three-wire function. The transmitter no longer corrects for changes in lead wire
resistance with temperature. replace the sensor as soon as possible.
• If the diagnostic message remains, go to Step 3.
3. The cable connecting the sensor to the transmitter may be too long. Test using a
sensor with a shorter cable.
• If shortening the cable eliminates the problem, move the transmitter closer to
the sensor. It may also be possible to increase diameter of the RTD wires.
Consult the factory for assistance.
• If the diagnostic message remains, go to Step 4.
4. Check the temperature of the transmitter. Simulate both temperature and pH. See
tEMP HI and tEMP LO (steps 2 and 3) for temperature simulation and Simulating
inputs - pH for pH simulation.
• If the transmitter fails either simulation, the electronic board stack should be
replaced.
• If the transmitter passes the simulations, the transmitter is not in good
condition, and the sensor should be replaced.
InPUt WArn
InPUt WArn means that the input value or the calculated pH is outside the measurement
range. The measured pH is less than -2 or greater than 16.
Reference Manual 71
Page 72
Diagnostics and troubleshootingReference Manual
October 201900809-0100-3581
Recommended actions
1. Check for miswires and open connections, particularly at TB-10. Open connections
can be caused by loose connections, poor spade crimps, or broken wires. Be sure to
check junction boxes for proper pass through of all wires.
• If correcting a wiring problem clears the message, the system is in good
condition.
• If the message is still showing, go to Step 2.
2. Check that the transmitter is working properly by simulating a pH input. See Testing
the transmitter by simulating the pH.
• If the transmitter does not respond to simulated inputs, replace the board stack.
• If the transmitter performs satisfactorily and the preamplifier is located in a
remote junction box or in a sensor-mounted junction box, go to Step 3.
• If the transmitter performs properly and the preamplifier is located in the
transmitter, the sensor has failed and should be replaced.
3. The problem may lie with the remote preamplifier or with the cable connecting the
preamplifier and junction box to the transmitter.
a) Be sure all wires between the junction box and transmitter are connected.
b) Use Rosemount cable.
Generic cable may not work.
• If the disagnostic message clears, the interconnecting cable was the problem.
• If the message remains, go to Step 4.
4. Confirm that the problem is with the remote preamplifier.
a) Wire the pH sensor directly to the transmitter.
b) Change the menu from PAMP=SnSr to trAnS for the test and return it to SnSr
afterwards.
• If the error message clears, the remote preamplifier is faulty. Replace the
preamplifier.
• If the error message remains, the sensor has failed. Replace the sensor.
SLOPE Err LO
SLOPE Err LO means that a two-point buffer calibration attempt has failed. The slope is
too low (<40 mV/pH) for a good measurement.
Recommended actions
1. Repeat the calibration.
a) Inaccurate buffers can cause a low slope. Repeat the calibration using fresh
buffers. Alkaline buffers, pH 10 or greater, are particularly susceptible to
changing value in air or with age. If you used a high pH buffer in the failed
calibration, try a lower pH buffer when repeating the calibration. For
example, use pH 4 and 7 buffer instead of pH 7 and 10 buffer.
72 Rosemount 5081
Page 73
Reference Manual Diagnostics and troubleshooting
00809-0100-3581October 2019
b) Allow adequate time for readings in buffer to become constant. If the sensor
was in a process substantially colder or hotter than the buffer, allow at least
20 minutes for readings in the buffer to stabilize. Alternatively, place the
sensor in a container of water at ambient temperatures for 20 minutes
before starting the calibration.
c) Be sure the correct buffer values are being entered during calibration.
• If the second calibration was successful, an error was made during the first
attempt.
• If the second calibration fails, go to Step 2.
2. Check wiring. Connections to TB-10, TB-7, and TB-8 are particularly important.
Recalibrate the sensor.
• If the wiring was the only problem, the sensor should calibrate.
• If the message persists, go to Step 3.
3. Inspect and clean the sensor. Recalibrate the sensor.
• If the sensor was dirty, it should calibrate after cleaning.
• If the message persists, go to Step 4.
4. Check for a faulty sensor.
• If a spare sensor is available, connect it to the transmitter. Use the auto
calibration procedure to calibrate the sensor.
— If the new sensor cannot be calibrated, the transmitter is faulty. Go to Step 5.
— If the new sensor can be calibrated, the old sensor has failed.
• If a spare sensor is not available, measure the glass impendance (GIMP). See
Testing the transmitter by simulating the pH.
— If the glass impedance is less than about 20 megohms, the glass has cracked
and the electrode must be replaced.
— If the glass impedance is greater than about 20 megohms, the sensor is
probably in good condition. Go to Step 5.
5. Check transmitter performance by simulating pH inputs. See Testing the
transmitter by simulating the pH.
• If the transmitter performs satisfactorily, go to Step 6.
• If the transmitter does not respond to stimulated inputs, replace the board
stack.
6. If the transmitter responds to simulated inputs, the problem must lie with the
sensor or the interconnecting wiring. Verify the interconnecting wiring point to
point. Fix or replace bad cable. If the cable is good, replace the pH sensor.
SLOPE Err HI
SLOPE Err HI means that a two-point buffer calibration attempt has failed. The slope is
too high (> 62 mV/pH) for a good measurement.
Reference Manual 73
Page 74
Diagnostics and troubleshootingReference Manual
October 201900809-0100-3581
Recommended actions
1. Repeat the calibration.
a) Inaccurate buffers can cause a high slope. Repeat the calibration using fresh
buffers. Alkaline buffers, pH 10 or greater, are particularly susceptible to
changing value in air or with age. If you used a high pH buffer in the failed
calibration, try a lower pH buffer when repeating the calibration. For
example, use pH4 and 7 buffer instead of pH 7 and 10 buffer.
b) Allow adequate time for readings in buffer to become constant. If the sensor
was in a process substantially colder or hotter than the buffer, allow at least
20 minutes for readings in the buffer to stabilize. Alternatively, place the
sensor in a container of water at ambient temperature for 20 minutes before
starting the calibration.
c) Be sure the correct buffer values are being entered during calibration. To
minimize errors caused by entering the wrong buffer values, use the
autocalibration procedure.
d) Verify that the temperature reading is accurate. Compare the sensor reading
against a thermometer known to be accurate. Recalibrate if necessary.
• If the second calibration was successful, an error was made during the first
attempt.
• If the second calibration fails, go to Step 2.
2. There is a remote possibility of a problem with the autocalibration program. Repeat
the calibration using the manual calibration procedure.
• If manual calibration was successful when autocalibration failed, the problem
might be with the sensor electronics. Call the factory for assistance.
• If manual calibration is not possible, go to Step 3.
3. Check the transmitter performance by simulating pH inputs. See Testing the
transmitter by simulating the pH.
• If the transmitter performs satisfactorily, go to Step 4.
• If the transmitter does not respond to stimulated inputs, replace the board
stack.
4. If the transmitter responds to simulated inputs, the problem must lie with the
sensor or the interconnecting wiring. Verify the interconnecting wiring point to
point. Fix or replace bad cable. If cable is good, replace the pH sensor.
Std Err
Std Err means the reference electrode voltage has changed drastically. Typical causes
are exposure to poisoning agents, sulfides or cyanides, or prolonged exposure to high
temperature.
Troubleshooting depends on the type of sensor.
74 Rosemount 5081
Page 75
Reference Manual Diagnostics and troubleshooting
00809-0100-3581October 2019
Recommended actions
1. If the sensor is rebuildable, replenish the electrolyte solution and replace the liquid
junction. Calibrate the sensor.
• If the sensor can be calibrated, the problem has been corrected.
• If the sensor cannot be calibrated, replace the sensor. If the sensor has separate
measuring and reference electrodes, replace only the reference electrode.
2. If the sensor is not rebuildable, try the reference electrode rejuvenation procedure.
• If the rejuvenated sensor can be calibrated, the problem has been corrected.
• If the sensor cannot be calibrated, replace the sensor.
rOM FAIL or CPU FAIL
rOM FAIL or CPU FAIL means the transmitter electronics have failed.
Recommended action
Replace the electronic board stack (PN 23992-02 [-HT] or PN 23992-03 [-HF]).
AdC WArn or CyCLE PWr
The Adc WArn or CyCLE PWr message appears momentarily when the transmitter has
recognized an internal calculation problem. The transmitter repeats the calculation, and
the message disappears once the calculation is successful. If the message is displayed
constantly, the transmitter electronics may be faulty.
Recommended actions
1. Check transmitter performance by simulating pH inputs. See Testing the
transmitter by simulating the pH.
• If the transmitter performs satisfactorily, go to Step 2.
• If the transmitter does not respond to simulated inputs, replace the board stack.
2. If the transmitter responds to simulated inputs, the problem must lie with the
sensor or the interconnecting wiring. Verify the interconnecting wiring point to
point. Fix or replace bad cable. If cable is good, replace the pH sensor.
WritE Err
WriTE Err means that jumper JP1 on the CPU board is not in place. If the sensor is not in
place, the transmitter cannot be programmed or calibrated.
Recommended actions
1. Check the position of jumper JP1 on the CPU board. If the jumper is hanging off one
of the pins, place it across both pins. If the jumper is missing entirely, use jumper
JP3 (50/60 Hz), which is not a critical jumper.
There are similar numbered jumpers on the analog board. The jumper to be
checked is on the CPU board, which is the center board in the stack.
2. Turn the power to the transmitter off and then back on.
Reference Manual 75
Page 76
Diagnostics and troubleshootingReference Manual
October 201900809-0100-3581
• Toggling the power should cause the message to disappear.
• If the message does not disappear, replace the electronic board stack.
FACt FAIL
FACt FAIL means the unit has not been factory calibrated. Call the factory. The
transmitter will probably need to be returned to the factory for calibration.
8.2.4Troubleshooting when no diagnostic message is
showing
If no diagnostic message is showing, locate the symptom(s) in the table below and refer to
the appropriate section for assistance.
SymptomSection
Id000 appears in display when trying to
program or calibrate transmitter
Error message flashing in displayTroubleshooting when a diagnostic message is
Transmitter does not respond to remote
controller
Calibration problems
SLOPE Err HI or SLOPE Err LO appears
after calibration attempt
bF1 or bF2 continuously flashes during auto
calibration
pH reading in buffer drifts during manual
calibration
Measurement problems
Sensor does not respond to known pH changesSensor does not respond to known pH changes
Buffer calibration is acceptable; process pH is
slightly different from expected value
Buffer calibration is acceptable; process pH is
grossly wrong and/or readings are noisy
Temperature reading is inaccurateTemperature reading is inaccurate
Id000 in display
showing
Transmitter not responding to IRC
SLOPE Err LO or SLOPE Err Hi appear after
calibration attempt
bF1 or bF2 continuously flashes during
autocalibration
pH reading in buffer drifts during manual
calibration
Buffer calibration is acceptable; process pH is
slightly different from expected value.
Buffer calibration is acceptable; process pH is
grossly different from expected value
Transmitter problems
No displayNo display
Display segments missing or display incorrectDisplay segments missing
Transmitter locked up, all display segments litTransmitter locks up
A security code has been programmed into the transmitter. The correct code must be
entered before the transmitter can be programmed or calibrated.
Transmitter not responding to IRC
Recommended actions
1. Be sure the transmitter is receiving the signal.
a) Clean the window in front of the IR detector.
The detector is a small rectangle just above the main display.
b) Hold the IRC at least 6 ft. (1.8 m) from the transmitter and not more than 15
degrees from the center.
c) Hold the IRC closer (within 2 ft. [.6 m]) in case the batteries are getting weak.
2. If Step 1 fails, check the IRC.
a) If a second transmitter is available, test the IRC on that transmitter. If a spare
transmitter is not available, continue with Step b.
b) The green LED, located just above and between the RESET and HOLD
buttons, should light when a key is pressed. A piece of black rubber film may
be covering the LED. Scrape the film away with your fingernail to expose the
LED.
The two clear LEDs on the front end of the IRC never light. They transmit the
invisible IR signal.
c) If the green LED does not light, the IRC is not working. Go to Step 3.
3. Take the IRC to a non-hazardous area and replace the two 1.5 Vdc AAA batteries.
• If the green LED lights, but the transmitter still does not respond, go to Step 4.
• If neither the LED lights nor the transmitter responds, replace the IRC.
4. Replace the transmitter display board.
SLOPE Err LO or SLOPE Err Hi appear after calibration attempt
Refer to SLOPE Err LO and SLOPE Err HI for assistance in solving calibration slope problems.
bF1 or bF2 continuously flashes during autocalibration
During autocalibration, bF1 or bF2 flashes until the pH reading of the sensor in the buffer
is stable.
Recommended actions
1. Check the stability limits.
If the stabilization range (prompt PH) is set too narrow or the stabilization time
(prompt tIME) is set too long, the transmitter will not accept buffer readings. A
good choice for PH is 0.02, and a good choice for tIME is 10 - 20 seconds.
Reference Manual 77
Page 78
Diagnostics and troubleshootingReference Manual
October 201900809-0100-3581
2. Allow adequate time for the temperature of the sensor to reach the temperature of
the buffer. If the sensor was in a process substantially hotter or colder than the
buffer, allow at least 20 minutes for readings in the buffer to stabilize. Alternatively,
place the sensor in a container of water at ambient temperature for 20 minutes
before starting the calibration.
3. Be sure to swirl the sensor after placing it in each new buffer solution.
4. Finally, check the sensor. Verify that wiring is correct. Also, the sensor may be dirty
or aged, or the reference junction may be depleted.
a) Check that the sensor is properly wired to the transmitter.
Pay particular attention to terminals TB-10 (mV in), TB-7 (reference), and
TB-8 (solution ground).
b) Clean the sensor.
c) If the sensor is not rebuildable, rejuvenate the reference junction.
d) If the sensor is rebuildable, replenish the reference electrolyte and replace
the liquid junction.
e) Replace the sensor.
A clean pH sensor should not drift in buffer.
pH reading in buffer drifts during manual calibration
Recommended actions
1. Allow adequate time for the temperature of the sensor to reach the temperature of
the buffer. If the sensor was in a process substantially hotter or colder than the
buffer, allow at least 20 minutes for readings in the buffer to stabilize. Alternatively,
place the sensor in a container of water at ambient temperature for 20 minutes
before starting the calibration.
2. Be sure to swirl the sensor after placing it in each new buffer solution.
3. Finally, check the sensor. Verify that wiring is correct. Also, the sensor may be dirty
or aged, or the reference junction may be depleted.
a) Check that the sensor is properly wired to the transmitter.
Pay particular attention to terminals TB-10 (mV in), TB-7 (reference), and
TB-8 (solution ground).
b) Clean the sensor.
c) If the sensor is not rebuildable, rejuvenate the reference junction.
d) If the sensor is rebuildable, replenish the reference electrolyte and replace
the liquid junction.
e) Replace the sensor.
A clean pH sensor should not drift in buffer.
78 Rosemount 5081
Page 79
Reference Manual Diagnostics and troubleshooting
00809-0100-3581October 2019
Sensor does not respond to known pH changes
Recommended actions
1. Verify that the change really happened. If you were checking pH response in
buffers, recheck performance with fresh buffers. If a process pH reading was not
what was expected, check the performance of the sensor in buffers. Also, use a
second pH meter to verify that the expected change in the process pH really
occurred.
2. Check the sensor. Verify that wiring is correct. Also, the sensor may be dirty or aged,
or the reference junction may be depleted. Check that the sensor is properly wired
to the transmitter.
See Wire. Pay particular attention to terminals TB-10 (mv in), TB-7 (reference), and
TB-8 (solution ground).
3. If a clean, properly wired sensor does not respond to pH changes, the glass bulb is
probably broken or cracked.
a) If a spare sensor is available, check the spare.
If the spare sensor responds to pH changes, the old sensor has failed.
If the spare sensor does not respond to pH changes, go to Step 4.
b) If a spare sensor is not available, check the glass impedance (GIMP) of the
existing sensor.
If the impedance is less than about 20 megohm, the pH sensor is cracked.
Replace the sensor.
If the impedance is greater than about 20 megohm, go to Step 4.
4. Check transmitter performance by simulating pH inputs.
See Testing the transmitter by simulating the pH.
If the transmitter responds to simulated inputs, the problem must lie with the
sensor or the interconnecting wiring. Verify the interconnecting wiring point to
point. Fix or replace bad cable. If cable is good, replace pH sensor.
If the transmitter does not respond to simulated inputs, replace the board stack.
Buffer calibration is acceptable; process pH is slightly
different from expected value.
Differences between pH readings made with an on-line instrument and a laboratory or
portable instrument are normal. The on-line instrument is subject to process variables (for
example, ground potentials, stray voltages, and orientation effects) that do not affect the
laboratory or portable instrument.
Buffer calibration is acceptable; process pH is grossly
different from expected value
The systems suggest a ground loop (measurement system connected to earth ground at
more than one point), a floating system (no earth ground), or noise being induced into the
transmitter by a sensor cabling.
Reference Manual 79
Page 80
Diagnostics and troubleshootingReference Manual
October 201900809-0100-3581
The problem arises from the process or installation. It is not a fault of the transmitter. The
problem should disappear once the sensor is taken out of the system.
Recommended actions
1. To confirm a ground loop:
a) Verify that the system works properly in buffers. Be sure there is no direct
electrical connection between the buffer containers and the process liquid or
piping.
b) Strip back the ends of a heavy gauge wire. Connect one end of the wire to the
process piping or place it in the process liquid. Place the other end of the wire
in the container of buffer with the sensor.
The wire makes an electrical connection between the process and sensor.
If similar symptoms develop after making the connection, a ground loop exists. If
no symptoms develop, a ground loop may or may not exist.
2. Check the grounding of the process.
The measurement system needs one path to ground: through the process liquid
and piping. Plastic piping, fiber glass tanks, and ungrounded or poorly grounded
vessels do not provide a path. A floating system can pick up stray voltages from
other electrical equipment.
a) Ground the piping or tank to a local earth ground.
Metal tees, grounding rings, or grounding rods may be required.
b) If problems persist, connect a wire from the ground connection at the dc
power supply to the transmitter case. Connect a second wire from the
transmitter case to the process.
These connections force the grounds to the same potential.
If the problem persists, simple grounding is not the problem. The sensor wiring is
probably carrying noise into the instrument. Go to Step 3.
3. Simplify the sensor wiring.
a) Disconnect all sensor wires at the transmitter except: TB-4 (resistance
temperature device [RTD] sense), TB-5 (RTD IN), TB-7 (Reference IN), and
TB-10 (pH/ORP IN). If you are using a remote preamplifier, disconnect the
wires at the input side of the junction box.
b) Tape back the ends of the disconnected wires, including all shield and drain
wires, to keep them from making accidental connections with other wires,
terminals, or the transmitter case.
c) Connect a jumper wire between TB-3 (RTD return) and TB-4 (RTD sense).
Connect a second jumper wire between TB-7 (Reference IN) and TB-8
(Solution ground).
d) Place the sensor back in the process liquid. If diagnostic measures such as
GLASSFAIL or REF WArn appear, turn off the sensor diagnostics.
If the symptoms disappear, interference was coming into the transmitter along one
of the sensor wires. The measurement system can be operated permanently with
simplified wiring.
80 Rosemount 5081
Page 81
Reference Manual Diagnostics and troubleshooting
00809-0100-3581October 2019
If symptoms still persist, go to Step 4.
4. Check for extra ground connections or reduced noise.
The electrode system is connected to earth ground through the process. If other
ground connections exist, there are multiple paths and ground loops are present.
Noise enters the measurement either by a direct connection, usually between the
cable and grounded metal, or by an indirect connection, usually EMI/RFI picked up
by the cable.
a) If the sensor cable is run inside conduit, there may be a short between the
cable and the conduit. Re-run the cable outside the conduit. If symptoms
disappear, then a short exists between the table and the conduit. Likely a
shield is exposed and is touching the conduit. Repair the cable and reinstall it
in the conduit.
b) To avoid induced noise in the sensor cable, run it as far away as possible from
the power cables, relays, and electric motors. Keep sensor wiring out of
crowded panels and cable trays.
c) Occasionally, noise can travel into the transmitter housing from the metal it
is mounted on. The noise is then radiated into the transmitter electronics. If
isolating the transmitter from its metal mounting eliminates the symptoms,
move the transmitter to a different location or mount it with isolating
materials.
If ground loop problems persist, consult the factory. It may require a visit from an
experienced service technician to solve plant-induced problems.
Temperature reading is inaccurate
Recommended actions
1. To troubleshoot temperature problems, refer to tEMP HI and tEMP LO.
2. Calibrate the temperature response of the sensor.
3. If necessary, automatic temperature compensation can be temporarily disabled
and the transmitter placed in manual temperature compensation. For manual
temperature, choose a temperature equal to the average temperature of the
process.
The resulting pH reading will be in error. The more variable the temperature and the
further from pH 7, the greater the error.
HART® communication problems
Recommended actions
1. If the handheld communicator software does not recognize the =transmitter, order
an upgrade from Rosemount at 800 999 9307.
2. Be sure the HART load and voltage requirements are met.
• HART communications requires a minimum 250 ohm load in the current loop.
• Install a 250 - 500 ohm resistor in series with the current loop. Check the actual
resistor value with an ohmmeter.
Reference Manual 81
Page 82
Diagnostics and troubleshootingReference Manual
October 201900809-0100-3581
• For HART communications, the power supply voltage must be at least 18 Vdc.
3. Be sure the HART communicator is properly connected.
• The communicator leads must be connected across the load.
• The communicator can be connected across the power terminals (TB-15 and
TB-16).
4. Verify that the handheld communicator is working correctly by testing it with
another HART Smart device.
If the communicator is working, the transmitter electronics may have failed. Call
Rosemount for assistance.
If the communicator seems to be malfunctioning, call Rosemount at 800 999 9307
for assistance.
No display
Recommended actions
1. Be sure power requirements are being met.
a. The positive voltage lead must be connected to TB-16.
b. Check dc voltage requirements and load restrictions.
2. Check for bad connections between the circuit boards. Be sure the ribbon cable
between the display and CPU boards is firmly seated in the socket on the CPU
board. Be sure the socket connection between the CPU and analog boards is firm.
Display segments missing
Recommended action
Replace the display board.
Transmitter locks up
Recommended actions
1. Turn the dc power off; then turn it back on.
2. If the problem persists, replace the electronic board stack (HART®: PN 23992-02;
FOUNDATION™ Fieldbus: PN 23992-03).
Transmitter periodically restarts itself
Recommended actions
1. The problem is usually related to improperly wired resistance temperature device
(RTD) input terminals.
a) The RTD return wire must be connected to TB-3. The RTD sense wire must be
connected to TB-4, and the RTD in wire must be connected to TB-5. If the pH
sensor does not have an RTD, connect a jumper wire across the terminals
TB-3 and TB-4 and a second jumper across TB-4 and TB-5.
82 Rosemount 5081
Page 83
Reference Manual Diagnostics and troubleshooting
00809-0100-3581October 2019
b) If you have jumpered the RTD connections as described in 1.a, turn off
automatic temperature compensation and operate the transmitter in
manual temperature mode.
2. If RTD wiring is correct and problems still persist:
a) Monitor the power supply. Be sure the power is not intermittent and the
correct voltage is present.
b) Try connecting the transmitter to a different power supply.
8.2.5Displaying diagnostic variables
This section describes how to display the diagnostic variables listed below:
Diagnostic measurements
1. Sensor voltage in mv (InPut)
2. Glass impedance in megohms (GIMP)
3. Reference temperature in kilohms (rIMP)
(1)
4. Temperature in °C
Diagnostic messages
1. Software version (VEr)
2. Display last three fault messages (ShoW FLt)
Note
Displays are read only.
Procedure
1. Press DIAG on the IRC to enter the Diagnostic menu.
Sensor voltage in mV (InPut) appears.
2. Press NEXT.
The temperature corrected glass impedance in megohms (GIMP) appears.
3. Press NEXT.
The reference impedance (rIMP) appears. For conventional low impedance silver/
silver chloride reference electrodes, the reference impedance has units of kilohms.
For the rare occasions when a high impedance reference is used, the units are
megohms.
4. Press NEXT.
The model number and software version (Ver) appears.
5. Press NEXT.
The temperature (tEMP) measured by the sensor appears.
6. Press NEXT.
The ShoW Fit submenu appears.
For high impedance reference electrodes, the reference impedance is in megohms.
(1)
Reference Manual 83
Page 84
Diagnostics and troubleshootingReference Manual
October 201900809-0100-3581
7. Press ENTER.
The most recent fault message appears in the display.
8. Press NEXT repeatedly to scroll through the stored messages.
The transmitter only remembers the three most recent messages. nonE appears if
there are no faults. Press EXIT to clear all the stored messages and return the
transmitter to the ShoW Fit display. If the transmitter loses power, all stored
warnings and fault messages are lost.
9. Press EXIT to return to the process display.
8.2.6Testing the transmitter by simulating the pH
Overview of simulating a pH input
This section describes how to simulate a pH input into the Rosemount 5081-P pH/ORP
transmitter. pH is directly proportionsl to voltage.
To simulate the pH measurement, connect a standard millivolt source to the transmitter.
If the transmitter is working properly, it will accurately measure the input voltage and
convert it to pH. Although the general procedure is the same, the wiring details depend on
the location of the preamplifier. Consult the table to find the correct procedure.
Preamplifier located inSection
TransmitterSimulate pH when the preamplifier is located in the transmitter
Remote junction boxSimulate pH when the preamplifier is located in a remote junction box or in a
sensor-mounted junction box
Sensor-mounted junction boxSimulate pH when the preamplifier is located in a remote junction box or in a
sensor-mounted junction box
SensorSimulate pH when preamplifier is in sensor
Simulate pH when the preamplifier is located in the
transmitter
Procedure
1. Program PAMP to transmitter.
2. Turn off sensor diagnostics.
3. Turn off automatic temperature compensation. Set manual temperature
compensation to 77 °F (25 °C).
4. Disconnect the sensor and wire transmitter as shown in Figure 8-5.
5. Attach a jumper between TB-7 (Reference IN) and TB-10 (pH IN).
6. Measure the voltage. Press DIAG on the infrared remote control (IRC).
The InPut voltage in millivolts appears in the temperature-output area. The main
display continues to show pH. The measured voltage should be 0 mV, and the pH
should be approximately 7. Because the calibration data in the transmitter may be
offsetting the input voltage, the displayed pH may not be exactly 7.0. If the actual
readings are close to expected, the transmitter is probably operating normally.
84 Rosemount 5081
Page 85
Reference Manual Diagnostics and troubleshooting
00809-0100-3581October 2019
7. If a standard millivolt source is available, remove the jumper between TB-7 and
TB-10 and connect the voltage source.
8. Calibrate the transmitter. Use 0.0 mV for pH 7 (bF1) and -177.4 mV for pH 10 (bF2).
If the transmitter is working, it should accept the calibration.
9. To check linearity, leave autocalibration and return to the main display. Set the
voltage source to the values in the table and verify that the pH reading matches the
expected value.
Voltage (mV)pH
295.82.00
118.35.00
-118.39.00
-295.812.00
Reference Manual 85
Page 86
Diagnostics and troubleshootingReference Manual
October 201900809-0100-3581
Example
Figure 8-5: pH simulation when the preamplifier is located in the transmitter
A. Transmitter
B. Earth ground
C. Shield
D. Resistance temperature device (RTD) return
E. -RTD sense
F. RTD in
G. Reference ground
H. Reference in
I. Solution ground
J. pH guard
K. pH in
L. +RTD sense
M. mV source
Simulate pH when the preamplifier is located in a remote
junction box or in a sensor-mounted junction box
Procedure
1. Program PAMP to sensor.
86 Rosemount 5081
Page 87
Reference Manual Diagnostics and troubleshooting
00809-0100-3581October 2019
2. Turn off sensor diagnostics.
3. Turn off automatic temperature compensation. Set manual temperature
compensation to 77 °F (25 °C).
4. Disconnect the sensor and wire the sensor side of the junction box as shown in
Figure 8-5. Leave the interconnecting cable between the junction box and
transmitter in place.
5. Attach a jumper between TB1-7 (Reference IN) and TB1-10 (pH IN)
6. From this point on, continue with steps 6 through 9 in Simulate pH when the
preamplifier is located in the transmitter.
For testing using a standard millivolt source. Be sure to remove the jumper between
TB1-7 and TB1-10 before connecting the standard millivolt source.
Reference Manual 87
Page 88
Diagnostics and troubleshootingReference Manual
October 201900809-0100-3581
Example
Figure 8-6: pH simulation when the preamplifier is located in a remote junction box
or in a sensor mounted junction box
A. Remote junction box: PN 23555-00 (includes preamplifier). Board PN 23557-00
B. mV source
C. Jumpers
D. No connection
E. Normally closed
F. Far side
G. Extension cable point to point wiring terminals 2 to 12
H. Earth ground
I. Shield
J. Resistance temperature device (RTD) return
K. -RTD sense
L. RTD in
M. Reference ground
N. Solution ground
O. pH guard
P. pH in
Q. +RTD sense
R. Transmitter
88 Rosemount 5081
Page 89
Reference Manual Diagnostics and troubleshooting
00809-0100-3581October 2019
Simulate pH when preamplifier is in sensor
The preamplifier in the sensor simply converts the high impedance signal into a low
impedance signal without amplifying it.
8.3Contacting and toroidal conductivity
diagnostics and troubleshooting
8.3.1Diagnose menu
The transmitter automatically monitors for fault conditions. The Diagnose menu displays
current variable settings and fault messages. The messages are defined in Table 8-2.
Troubleshoot contacting conductivity analyzers
Procedure
1. Look for a diagnostic fault message on the display to help pinpoint the problem.
Refer to Diagnostic messages for an explanation of the message recommended
actions to solve it.
2. Refer to Troubleshooting when no diagnostic message is showing for common loop
problems and the recommended actions to resolve them.
Display diagnostic values
Use the DIAG key on the IRC is used to access the Diagnose menu. The messages are
defined in Table 8-2.
Use the FAuLtS submenu to show the last three faults/warnings. The most recent is
displayed first. Use NEXT to scroll through the remaining faults. Press EXIT to clear all
fault/warnings and returns the FAuLtS segment. Disconnecting the transmitter removes all
fault messages from memory. The nonE message is displayed when no faults/warnings
have occurred.
Table 8-2: Diagnostic variables mnemonics
AbSAbsolute conductivity (µS/cm or mS/cm)
0 A irSensor zero in air
CELL ConStSensor cell constant (used in C mode)
tSLOPETemperature slope in %/° C
CAL FCalibration factor
SoFtSoftware version
HArdHardware version
FAuLtSShow fault messages
nonENo fault messages in memory
Reference Manual 89
Page 90
Diagnostics and troubleshootingReference Manual
October 201900809-0100-3581
8.3.2Fault conditions
The diagnostic program detects and diffentiates three classes of error conditions/
problems. System disabling problems are faults caused by failures in the loop or significant
variations in the process. System non-disabling problems are warnings and deal with input
or A to D conversion settings. The third class of detection problems are error messages
and occur when the calibration limits are exceeded.
Disabling faults
When a disabling fault occurs:
1. Both FAULT and HOLD annunciation fields become active (Figure 8-7).
2. The process variable flashes at the rate of 1 second on and 1 second off.
3. The appropriate fault message alternates with the normal Temperature/Current
output display (see Figure 8-7).
4. The output current loop is forced to run the non-zero fault value last entered or held
at last value if fault value = 0 if the transmitter is not in the TEST, HOLD, or Multidrop
operational modes.
5. A 0-1 mA output signal is available for external use when system disability
conditions are active. These conditions drive this output to 1 mA. Please contact
factory for specific application information.
Figure 8-7: Disabling Fault Annunciation
Non-disabling warnings
When a non-system-disabling condition occurs, a warning message is displayed.
The process variable does not flash. The appropriate message alternates with the
Temperature/Current output display (see Figure 8-8).
If more than one fault exists, the display will sequence through each diagnostic message.
This will continue until the cause of the fault has been corrected.
90 Rosemount 5081
Page 91
Reference Manual Diagnostics and troubleshooting
00809-0100-3581October 2019
Figure 8-8: Warning Annunciation
8.3.3Diagnostic messages
The transmitter's diagnostics constantly monitor the conductivity loop for possible
problems. If an operational problem is encountered, check the display for a fault or error
message. These are displayed in the Temperature/Current output segment of the display.
Note the message and refer to the following sections for a description of possible
problems that may have triggered the diagnostic message.
Faults
tEmP LO
Temperature is too low.
Recommended actions
1. Check wiring or sensor/process temperature.
2. Check resistance temperature device (RTD).
tEmP HI
Temperature is too high.
Recommended actions
1. Check wiring or sensor/process temperature.
2. Check resistance temperature device (RTD).
rtd FAIL
The resistance temperature device (RTD) sensor line fault limits have been exceeded into
the sensor.
Recommended action
Check wiring or check Program/Temp menu setting to verify the 100-3 or 100-4 sensor
type connected.
CPU FAIL
The CPU has failed during RAM or EEPROM verification.
Reference Manual 91
Page 92
Diagnostics and troubleshootingReference Manual
October 201900809-0100-3581
Recommended action
Recycle. If persistent, contact the factory.
FACt FAIL
FACt FAIL appears if the transmitter factory calibration message has been triggered.
A stray noise spike can cause this message to appear.
Recommended actions
1. If the pH reading seems acceptable, reset the calibration flag.
a) Press 2 on the infrared remote control (IRC) ten times to enter the factory
calibration menu.
The display does not change.
b) Press 3.
FActorYCAL appears in the display.
c) Press NEXT.
rEPAir appears in the display.
d) Press NEXT.
ConFIG appears in the display.
e) Press NEXT.
rESEt appears in the display.
f) Press NEXT.
rESEtCFG appears in the display.
g) Press ENTER.
rESET appears again.
h) Press NEXT.
FActorYCAL reappears.
i) Press ENTER.
FactOn appears in the display.
j) Press 3.
FactOFF appears.
k) Press ENTER to store the settings.
l) Press EXIT repeatedly until the main display reappears.
If the message does not clear or problems persist, the electronics have failed.
Replace the electronic board stack.
2. Contact factory.
rOm FAIL
The PROM failed the check-sum test.
92 Rosemount 5081
Page 93
Reference Manual Diagnostics and troubleshooting
00809-0100-3581October 2019
Recommended action
Contact factory.
CyCLE PoWr
A wrong value was detected during power-up.
Recommended action
Recycle the power.
SEnSor FAIL
Bad sensor means that the sensor current is a large negative number. SEnSor FAIL may
appear for a while when the sensor is first placed in service.
Recommended actions
1. Observe the sensor. Go to SEnSor Cur under the Diagnostic menu.
If the sensor is moving in the positive direction, there is probably nothing wrong.
The error message should soon disappear.
2. Verify that wiring is correct. Pay particular attention to the anode and cathode
connections.
3. Verify that the transmitter is configured for the correct measurment.
Configuring the measurement sets several things, including the polarizing voltage.
Applying the wrong polarizing voltage to the sensor can cause a negative current.
4. Replace the sensor membrane and electrolyte solution and clean the cathode if
necessary.
See the sensor reference manual for details.
5. Replace the sensor.
Warnings
InPUt WArn
InPUt WArn means that the input value or the calculated pH is outside the measurement
range. The measured pH is less than -2 or greater than 16.
Recommended actions
1. Check for miswires and open connections, particularly at TB-10. Open connections
can be caused by loose connections, poor spade crimps, or broken wires. Be sure to
check junction boxes for proper pass through of all wires.
• If correcting a wiring problem clears the message, the system is in good
condition.
• If the message is still showing, go to Step 2.
2. Check that the transmitter is working properly by simulating a pH input. See Testing
the transmitter by simulating the pH.
• If the transmitter does not respond to simulated inputs, replace the board stack.
Reference Manual 93
Page 94
Diagnostics and troubleshootingReference Manual
October 201900809-0100-3581
• If the transmitter performs satisfactorily and the preamplifier is located in a
remote junction box or in a sensor-mounted junction box, go to Step 3.
• If the transmitter performs properly and the preamplifier is located in the
transmitter, the sensor has failed and should be replaced.
3. The problem may lie with the remote preamplifier or with the cable connecting the
preamplifier and junction box to the transmitter.
a) Be sure all wires between the junction box and transmitter are connected.
b) Use Rosemount cable.
Generic cable may not work.
• If the disagnostic message clears, the interconnecting cable was the problem.
• If the message remains, go to Step 4.
4. Confirm that the problem is with the remote preamplifier.
a) Wire the pH sensor directly to the transmitter.
b) Change the menu from PAMP=SnSr to trAnS for the test and return it to SnSr
afterwards.
• If the error message clears, the remote preamplifier is faulty. Replace the
preamplifier.
• If the error message remains, the sensor has failed. Replace the sensor.
OvEr rAnGE and AMP FAIL
The current range setting has been exceeded or the measurment exceeds the display
limit.
This error message appears if the sensor current is too high. Normally, excessive sensor
current implies that the sensor is miswired or the sensor has failed.
Recommended actions
1. Verify the 4 and 20 mA settings in the Program/output menu.
2. Verify that the wiring is correct and connections are tight. Be sure to check
connections at the junction box if one is being used.
3. Replace the sensor membrane and electrolyte solution and clean the cathode if
necessary.
See the sensor reference manual for details.
4. Replace the sensor.
AdC Error
AdC Error means the analog to digital converter has failed.
Recommended actions
1. Verify that the sensor wiring is correct and connections are tight.
Be sure to check connections at the junction box if one is being used.
2. Disconnect sensor(s) and simulate temperature and sensor input.
94 Rosemount 5081
Page 95
Reference Manual Diagnostics and troubleshooting
00809-0100-3581October 2019
If the transmitter does not respond to simulated signals, the analog PCB has
probably failed.
3. Recycle the power.
4. Call the factory for assistance.
nEEd 0 CAL
The message nEEd 0 CAL means that the concentration of the analyte is too negative.
Recommended actions
1. Check the zero current (go to 0 CurrEnt under the Diagnose menu).
If the zero current is appreciably greater than the measurement current, the nEEd
0 CAL warning appears.
2. Verify that the zero current is close to the value given in the calibration section for
the analyte being determined.
3. Rezero the sensor.
Refer to the calibration and troubleshooting sections of the sensor reference
manual for more information.
Errors
CAl Err or OFFSEt Err
A calibration error has occurred between the standard and process.
Recommended action
Press RESET and repeat. Check calibration standards and unit configuration.
tSLOPE Err
The limit for T-2 in a two point calibration has been exceeded.
Recommended action
Press RESET and repeat the calibration/temp. slope menu setting.
-0- Err
Sensor zero limit has been exceeded.
Recommended actions
Press RESET and repeat the calibrate/sensor menu setting.
WritE Err
An attempt to write on the EEPROM has failed.
The jumper JP-1 on the CPU board has been removed.
SenSE OPEn
Most Rosemount sensors use a Pt100 or Pt1000 in a three-wire configuration. The in and
return leads connect the resistance temperature device (RTD) to the measuring circuit in
the transmitter. A third wire, called the sense line, is connected to the return lead. The
Reference Manual 95
Page 96
Diagnostics and troubleshootingReference Manual
October 201900809-0100-3581
sense line allows the transmitter to correct for the resistance of the in and return leads and
to correct for changes in lead wire resistance with changes in ambient temperature.
Recommended actions
1. Verify all wiring connections, including wiring in a junction box if one is being used.
2. Disconnect the RTD SENSE and RTD RETURN wires. Measure the resistance between
the leads.
It should be less than 5 Ω.
3. If the sense line is open, replace the sensor as soon as possible.
The transmitter can be operated with the sense line open. The measurement will be
less accurate, because the transmitter can no longer compensate for lead wire
resistance. However, if using the sensor at approximately constant ambient
temperatures, you can eliminate the lead wire resistance error by calibrating the
sensor to the measurement temperature. Errors caused by changes in ambient
temperature cannot be eliminated.
4. To make the error message disappear, connect the RTD SENSE and RETURN
terminals with a jumper.
8.3.4
Troubleshooting when no diagnostic message is
showing
The following sections identify some of the more common symptoms and suggest actions
to help resolve a problem. In general, wiring is the most common cause.
Wrong temperature reading
Recommended actions
1. Perform a temperature standardization.
2. Verify sensor's resistance temperature device (RTD).
Suspected temperature compensation problem
Recommended action
Check wiring.
Display segments missing
Recommended action
Replace the display board.
Transmitter locks up
The transmitter won't respond
Recommended actions
1. Replace PCB stack.
96 Rosemount 5081
Page 97
Reference Manual Diagnostics and troubleshooting
00809-0100-3581October 2019
2. Press RESET.
3. Check batteries in infrared remote control (IRC).
Erratic displays
Recommended action
Check sensors in process.
Transmitter won't respond to infrared remote control (IRC)
key presses
Recommended actions
1. Verify and clean ribbon cable connection on CPU board.
2. Check batteries in IRC.
Key press gives wrong selection
Recommended actions
1. Replace infrared remote control (IRC).
2. Check ribbon cable connection on CPU board.
Wrong or no current output
Recommended actions
1. Verify that output is not being overloaded.
2. Remove load.
3. Replace PCB stack.
No display or indicators
Recommended action
Replace PCB stack.
Excess input
Recommended action
Check sensor wiring.
Reverse input
Recommended action
Perform sensor zero.
Check sensor zero
The transmitter will not zero.
Reference Manual 97
Page 98
Diagnostics and troubleshootingReference Manual
October 201900809-0100-3581
Recommended action
Place the sensor in the air and access zero routine.
8.3.5Troubleshoot in the field
When it is apparent by grab sample analysis that the transmitter is giving inaccurate
readings, try the following:
Prerequisites
Make sure the sensor surfaces are totally wetted by the process and that air bubbles are
not trapped in the vicinity of electrodes. If you find air bubbles, change the installation
technique to eliminate air bubbles.
Procedure
1. Visually inspect the installation.
a) Check that the transmitter is mounted securely and that its internal parts are
properly connected.
b) Check all input and output wiring.
8.3.6
2. If Step 1 did not indicate the source of the problem, isolate the problem to either
the transmitter or the sensor.
3. To troubleshoot the sensor:
a) Disconnect the sensor from the transmitter.
b) Remove the sensor from the process.
c) Thoroughly dry the sensor electrodes.
Refer to the sensor reference manual for additional troubleshooting checks.
4. To troubleshoot the transmitter independently of the sensor, use an appropriate
resistor across the temperature input connectors and connect the conductivity
inputs to resistance decade box.
Troubleshoot conductivity measurement
If troubleshooting in the field does not resolve the error, try the following procedure to
troubleshoot a conductivity measurement problem in the process.
Note
Before starting this procedure, make sure that all wiring is correct.
Procedure
1. Remove the sensor from process and place the sensor in air. Zero the transmitter.
This step is for normal contacting only, not for low conductivity or resistivity.
If the transmitter zeros correctly, place the sensor in the process and standardize.
2. If the transmitter does not zero correctly:
a) Check diagnostic messages.
Refer to Diagnostic messages.
98 Rosemount 5081
Page 99
Reference Manual Diagnostics and troubleshooting
00809-0100-3581October 2019
b) Check wiring for short.
c) Consult service center.
3. If the sensor does not standardize:
a) Remove the sensor from the process and test it in a known conductivity
solution or against a certified conductivity instrument.
b) Check for ground loops and/or improper installation.
8.3.7Resistance temperature device (RTD) resistance values
Table 8-3 is a ready reference of RTD resistance values at various temperatures. Use these
values to test and evaluate the sensor.
Note
Resistance values are read across the RTD element and are based on the manufacturer's
stated values (±1%). Allow enough time for the RTD element in the sensor to stabilize to
the surrounding temperature.
Use the formula to determine the appropriate resistance value to use to simulate a
conductivity value.
8.4Chlorine, dissolved oxygen, and ozone
diagnostics
8.4.1Diagnostics overview
The Rosemount 5081 transmitter can display diagnostic information that is useful in
troubleshooting. The diagnostics available depend on the measurement being made. To
read diagnostic information, go to the main display and press DIAG on the infrared remote
controller. Press NEXT until the desired diagnostic message appears. Refer to the
appropriate section below for more information.
8.4.2
Diagnostic messages for dissolved oxygen
MessageSee section
TYPE O2TYPE O2
SEnSor CurSEnSor Cur
SEnSitvtYSEnSitviTY
0 CurrEnt0 CurrEnt.
bAr PreSSbAr PreSS.
5081-A-Ht5081-A-Ht.
FAULtSFAULtS
TYPE O2
Transmitter is measuring oxygen.
100 Rosemount 5081
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.