The wetted sensor materials may not be compatible with process composition and operating
conditions.
Application compatibility is entirely the operator's responsibility.
CAUTION
Equipment damage
Do not exceed pressure and temperature specifications
Pressure: 65 psig (549 kPa abs) max.
Temperature: 32 to 122 °F (0 to 50 °C)
WARNING
Physical access
Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of end
users’ equipment. This could be intentional or unintentional and needs to be protected against.
Physical security is an important part of any security program and fundamental to protecting your
system. Restrict physical access by unauthorized personnel to protect end users’ assets. This is true for
all systems used within the facility.
Contents
First steps.....................................................................................................................................3
1. Inspect the shipping container. If it is damaged, contact the shipper
immediately for instructions.
2. If there is no apparent damage, unpack the container. Be sure all
items shown on the packing list are present. If items are missing,
notify Emerson immediately.
1.2Product description
Figure 1-1: Rosemount 499ACL-03 Sensor Parts
A. Membrane retainer
B. Membrane assembly
C. O-ring
D. Cathode
E. Electrolyte fill plug (wrap with pipe tape)
F. Pressure equalizing port
G. Sensor cable (integral cable shown)
1.3
Table 1-1: Sensor Specifications
Physical characteristicsSpecifications
PressureSensor must drain to open atmosphere. No back pressure.
Temperature (operating)32 to 122 °F (0 to 50 °C)
Process connection1-in. male national pipe thread (MNPT)
Wetted partsNoryl®, Viton®, silicone, wood, and Zitex® (PTFE)
Quick Start Guide3
Specifications
Page 4
Quick Start GuideNovember 2019
Table 1-1: Sensor Specifications (continued)
Physical characteristicsSpecifications
CathodeGold mesh
Linearity2% (typical)
Table 1-2: Other Specifications
TypePNWetted
Low flow cell
with bubble
sweeping
(1)
nozzle
(1) Flow cell must drain to open atmosphere. Do not install the sensor in a pressurized line.
Temperature and pressure specifications for the low flow cell exceed the temperature and
pressure specifications for the sensor.
24091-01Polycarbon-
materials
ate/
polyester,
316 stainless
steel, and
silicone
Process
connection
Compression
fitting for
¼-in. O.D.
tubing
Maximum
temperature
158 °F (70 °C) 90 psig
Maximum
pressure
(722 kPa abs)
4Emerson.com/Rosemount
Page 5
November 2019Quick Start Guide
2Install
Install sensor in the low flow cell (PN 24091-01) only. Keep the flow as
constant as possible between 1 and 4 gph (3.8 to 15 L/hr). The flow cell must
drain to open atmosphere.
Figure 2-1: Sensor Orientation
Install sensor within 45 degrees of vertical.
Quick Start Guide5
Page 6
Quick Start GuideNovember 2019
Figure 2-2: Low Flow Cell (PN 24091-00)
A. Inches
B. Millimeters
C. Outlet
D. Inlet
6Emerson.com/Rosemount
Page 7
November 2019Quick Start Guide
3Wire
NOTICE
For additional wiring information on this product, including sensor
combinations not shown here, please refer to the Liquid Transmitter Wiring
Diagrams.
Figure 3-1: Rosemount 499ACL-03 Sensor Wiring to Rosemount 1056
and 56 Transmitters
Table 3-1: Rosemount 499ACL-03 Sensor Wiring to Rosemount 1056 and 56
Transmitters
Terminal
number
1AWhiteResistance temperature
2BWhite/redRTD sense
3CRedRTD in
4DClearRTD shield
5N/AN/A+5 V out
6N/AN/A-4.5 V out
7DClearAnode shield
8EGrayAnode
9DClearCathode shield
Quick Start Guide7
LetterWire colorDescription
device (RTD) return
Page 8
Quick Start GuideNovember 2019
Table 3-1: Rosemount 499ACL-03 Sensor Wiring to Rosemount 1056 and 56
Transmitters (continued)
Terminal
number
10FOrangeCathode
LetterWire colorDescription
Figure 3-2: Rosemount 499ACL-03 Sensor Wiring to Rosemount 5081
Transmitter
Table 3-2: Rosemount 499ACL-03 Sensor Wiring to Rosemount 5081 Transmitter
Terminal
number
1N/AN/AN/A
2N/AN/A+0.8 V
3AWhiteRTD return
8Emerson.com/Rosemount
LetterWire colorDescription
Page 9
November 2019Quick Start Guide
Table 3-2: Rosemount 499ACL-03 Sensor Wiring to Rosemount 5081 Transmitter
(continued)
Terminal
number
4BWhite/redRTD sense
5CRedRTD in
6N/AN/AReference guard
7N/AN/AReference in
8DClearSolution ground
9N/AN/ApH guard
10N/AN/ApH in
11N/AN/A-5 V
12N/AN/A+5 V
13EGrayAnode
14FOrangeCathode
15N/AN/AHART®/FOUNDATION™ Fieldbus
16N/AN/AHART/FOUNDATION Fieldbus (+)
LetterWire colorDescription
(-)
Quick Start Guide9
Page 10
Quick Start GuideNovember 2019
Figure 3-3: Rosemount 499ACL-03 Sensor Wiring to Rosemount 1066
Transmitter
Note
Connect clear shield wires to solution ground terminal on TB 2. Use wire nut
and pigtail if necessary.
Table 3-3: Rosemount 499ACL-03 Wiring to Rosemount 1066
Transmitter
LetterColorTerminal description
AOrangeCathode
BGrayAnode
CWhiteReturn
DWhite/redSense
ERedRTD in
10Emerson.com/Rosemount
Page 11
November 2019Quick Start Guide
Figure 3-4: Rosemount 499ACL-03-01-54-VP Sensor Pin-out Diagram
(Top View of Connector End of Sensor)
Table 3-4: Pin-out Diagram
Terminal numberDescription
1Cathode
2N/A
3RTD sense
4Anode
5RTD return
6RTD in
When making a connection through a junction box (PN 23550-00), wire
point-to-point.
NOTICE
Use a wire nut and pigtail (included) when connecting several wires to the
same terminal.
Quick Start Guide11
Page 12
Quick Start GuideNovember 2019
4Calibrate
4.1Zero point calibration
Even in the absence of monochloramine, the sensor generates a small signal
called the zero current. Failing to correct for the zero current can introduce a
bias, particularly if the monochloramine concentration is small (<0.4 ppm).
Zero the sensor when it is first placed in service and every time the fill
solution is changed.
To zero the sensor:
Procedure
1. Pour a cup of deionized or bottled water.
2. Place the sensor in the water.
3. Wait until the sensor current has reached a stable low value (at least
two hours).
4. Follow the transmitter prompts for zeroing the sensor.
Note
Refer to the manual for the transmitter you are using (Rosemount
56, 1056, or 1066).
The zero current should be between -10 and +15 nA. For more information,
refer to the transmitter manual.
4.2
12Emerson.com/Rosemount
Full scale calibration
Because stable dilute monochloramine standards are not available, the
sensor must be calibrated against the results of a laboratory test run on a
grab sample of the process liquid.
Procedure
1. Place the sensor in the flow cell.
2. Start the sample and reagent flow.
3. Adjust the sample flow to between 1 and 4 gph.
4. Adjust the concentration so that it is near the upper end of the
operating range.
5. Wait for the readings to stabilize.
6. Follow the transmitter prompts to complete the calibration.
Note
Refer to the manual for the transmitter you are using (Rosemount
56, 1056, or 1066).
Page 13
November 2019Quick Start Guide
Be sure taking the sample does not alter flow to the sensor.
7. After calibration, go to the Diagnostics menu and check the
sensitivity.
The sensitivity should be between 250 and 450 nA/ppm. For more
information, refer to the transmitter manual.
Quick Start Guide13
Page 14
Quick Start GuideNovember 2019
5Maintenance
WARNING
Pressurized spray injury
Before removing the sensor, be absolutely certain that the process pressure
is reduced to 0 psig and the process temperature is lowered to a safe level!
5.1Cleaning the membrane
Keep the membrane clean and free from dirt and algae. Periodically inspect
the membrane. If it appears fouled, clean the membrane with water sprayed
from a wash bottle.
CAUTION
EQUIPMENT DAMAGE
Do not wipe the membrane with a tissue. Do not touch the membrane.
Doing so may damage the cathode, making the sensor unusable.
5.2Replacing the electrolyte solution and membrane
WARNING
Corrosive substance
Fill solution is corrosive.
Avoid contact with skin and eyes.
Consult Material Safety Data Sheet (MSDS) for safety information.
Procedure
1. Unscrew the membrane retainer.
2. Remove the membrane assembly and O-ring.
See Figure 1-1.
3. Hold the sensor over a container with the cathode pointing down.
4. Remove the fill plug.
5. Allow the electrolyte solution to drain out.
6. Remove the old pipe tape from the plug.
7. Wrap the plug with one or two turns of pipe tape..
8. Prepare a new membrane.
14Emerson.com/Rosemount
Page 15
November 2019Quick Start Guide
a) Hold the membrane assembly with the cup formed by the
membrane and membrane holder pointing up.
b) Fill the cup with electrolyte solution.
9. Hold the sensor at about a 45 degree angle with the cathode end
pointing up.
10. Add electrolyte solution through the fill hole until the liquid
overflows.
11. Tap the sensor near the threads to release trapped air bubbles.
12. Add more electrolyte solution if necessary.
13. Place the fill plug in the electrolyte port and begin screwing it in.
14. After several threads have engaged, rotate the sensor so that the
cathode is pointing up and continue tightening the fill plug.
Do not overtighten.
15. Place a new O-ring in the groove around the cathode post.
16. Cover the cathode with electrolyte solution; then place the
membrane assembly over the cathode.
17. Screw the membrane retainer in place.
18. Hold the sensor with the cathode end pointing down.
19. Give the sensor several sharp shakes to dislodge air bubbles trapped
behind the cathode.
The sensor may require several hours operating at the polarizing
voltage to equilibrate after the electolyte solution has been replaced.
Quick Start Guide15
Page 16
Quick Start GuideNovember 2019
6Accessories
Part #Description
23750-00Electrolyte fill plug with wooden osmotic pressure relief
9550094O-ring, Viton 2-014
33521-00Membrane retainer
23501-09Monochloramine membrane assembly: includes one
23502-09Monochloramine membrane kit: includes three
9210372Monochloramine sensor fill solution, 4 oz (120 mL)
The Emerson logo is a trademark and service
mark of Emerson Electric Co. Rosemount is a
mark of one of the Emerson family of companies.
All other marks are the property of their
respective owners.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.