Rosemount 498CL pH-Independent Free Chlorine Sensor Quick Start Guide

Page 1
Rosemount™ 498CL
pH-Independent Free Chlorine Sensor
Quick Start Guide
LIQ-QSG-498CL, Rev M
July 2017
Page 2
Page 3

Contents

Contents
1.1 Unpacking and inspection .............................................................................................................. 1
1.2 Specifications .................................................................................................................................1
4.1 Cleaning the membrane .................................................................................................................7
4.2 Replacing the electrolyte slurry and membrane ............................................................................. 7
4.3 Storage .......................................................................................................................................... 9
4.4 Rejuvenating a sensor following improper storage ......................................................................... 9
Quick Start Guide i
Page 4
Contents
ii Rosemount 498CL
Page 5
1 Plan

1.1 Unpacking and inspection

1. Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions.
2. Save the box.
3. If there is no apparent damage, unpack the container. Be sure all items shown on the packing list are present. If items are missing, notify Rosemount immediately.

1.2 Specifications

Sensor specificationsTable 1-1:
Physical characteristics Specifications
Linear range 0 to 20 ppm (mg/L) as Cl2. For higher ranges, consult the fac-
tory.
Pressure Sample must drain to open atmosphere. No back pressure
allowed.
Temperature (operating) 0 to 50 °C (32 to 122 °F)
Wetted materials PVC, polyethersulfone, polyester, Viton
stainless steel.
Cathode Gold mesh
Accuracy Accuracy depends on the accuracy of the chemical test used
to calibrate the sensor.
Linearity between 0 and 20 ppm 1% (per IEC 60746)
Linearity between 0 and 2 ppm ±0.05 ppm following calibration at about 2 ppm
Sensitivity to pH Between 6.5 and 10 sensor signal changes <4% per unit
change in pH. Below pH 6.5, the change is < 1% per unit change in pH.
Sample conductivity > 10 µS/cm
Interferences Monochloramine, dichloramine, and permanganate
Response time < 2 minutes to 90% of final value following step change at
1.2 gal/hr (75 mL/min) and 25 °C (77 °F)
Electrolyte life 3 months (approx.)
Cable length (standard integral cable) 25 ft (7.6 m)
Cable length (maximum) 300 ft (91 m)
Sample flow 1.2 gal/hr (75 mL/min). Changing flow from 0.5 to 2.5 gph
(30 to 160 mL/min) increases the sensor signal by about 30%. Increasing flow increases pH dependence.
(1)
, silicone, and 316

Plan

Quick Start Guide 1
Page 6
Plan
Sensor specifications (continued)Table 1-1:
Physical characteristics Specifications
Process connections Sensor must be used in flow cell PN 24091-01
Weight/shipping weight
(2)
Sensor with integral cable: 2 lb/3 lb (1.0 kg/1.5 kg)
Sensor with VP connector: 1 lb/2 lb (0.5 kg/1.0 kg)
(1) Viton is a registered trademark of EI duPont de Nemours. (2) Weights and shipping weights are rounded up to the nearest whole pound or 0.5 kg.
Flow cell specificationsTable 1-2:
Part number
Wetted materials Polycarbonate, polyester, 316 stainless steel, and silicone
Maximm inlet pressure 65 psig (549 kPa abs)
Process connections 1/4 in. OD tubing compression fitting or 1/4 in. FNPT
(1) Sold separately
(1)
24091-01 (low flow cell with bubble sweeping nozzle)
CAUTION!
SENSOR/PROCESS APPLICATION COMPATIBILITY
The wetted sensor materials may not be compatible with process composition and operating conditions. Application compatibility is entirely the responsiblity of the user.
2 Rosemount 498CL
Page 7
2 Install
Complete the following steps to install the sensor in the low flow cell. The sensor leaves the factory with a shipping membrane in place.
CAUTION!
EQUIPMENT DAMAGE
Do not remove the shipping membrane until you are ready to put the sensor in service.
Procedure
1. Remove the red protective cap from the sensor.
2. Holding the membrane end pointing up (cable end pointing down), unscrew the retainer cap and remove the shipping membrane.
See Figure 2-1. It is not necessary to remove the O-ring. Save the shipping membrane. It should be reinstalled when the sensor is not in use.

Install

Rosemount 498CL SensorFigure 2-1:
3. Still holding the membrane end pointing up, install the chlorine membrane.
The chlorine membrane is in the plastic bag attached to the sensor.
4. Screw the retainer back in place.
5.
Place the sensor in the flow cell and start the flow.
CAUTION!
EQUIPMENT DAMAGE
Do not store or operate the sensor for more than about 45 minutes in a non-flowing sample.
The sensor must drain to open atmosphere without back pressure. Do not leave the sensor in the air.
Quick Start Guide 3
Page 8
Install
InstallationFigure 2-2:
The plastic adapter shipped with the flow cell is not used.
Keep sample flow constant and between 1 and 2 gph (60 to 120 mL/min).
4 Rosemount 498CL
Page 9
3 Wire
NOTICE
For additional wiring information on this product, including sensor combinations not shown here, please refer to the Liquid Transmitter Wiring Diagrams.

Wire

Figure 3-1:
Rosemount 498CL Sensor Wiring to Rosemount 1056 and 56 Transmitters
Quick Start Guide 5
Page 10
Wire
Figure 3-2:
Pin out Diagram for Rosemout 498CL-01-VP Sensor (Top View of Connector End of Sensor)
6 Rosemount 498CL
Page 11
4 Maintenance
The sensor requires routine maintenance which includes cleaning the membrane as well as periodically replacing the membrane and fill slurry. The fill slurry and membrane should be replaced at approximately three-month intervals. The frequency of cleaning can be determined only by experience.

4.1 Cleaning the membrane

Keep the membrane clean and free from dirt and algae. Clean the membrane with water sprayed from a wash bottle.
CAUTION!
EQUIPMENT DAMAGE

Maintenance

Do not use tissues to clean the membrane. Touching the membrane may bend the cathode and permanently damage the sensor.

4.2 Replacing the electrolyte slurry and membrane

Complete the following steps to replace the electrolyte slurry and membrane in the sensor.
WARNING!
HARMFUL SUBSTANCE Fill solution and solid may cause irritation. Avoid contact with skin and eyes. May be harmful if swallowed. Read and follow manual.
Procedure
1. Unscrew the membrane retainer and remove the membrane assembly and O-ring.
See Figure 2-1.
2. Remove the fill plug.
3. Empty all remaining fill slurry from the sensor. Rinse with deionized water until there is no significant amount of solid left in the sensor.
4. Place a few drops of water in the replacement membrane assembly and place it on the mesh cathode.
Quick Start Guide 7
Page 12
Maintenance
CAUTION!
EQUIPMENT DAMAGE
Do not touch the mesh cathode. Doing so may bend the mesh and permanently damage the sensor.
5. Screw the membrane retainer into place.
6. Obtain one bottle of saturated succinic acid (PN 9210381, 40 mL) and one bottle of succinic acid crystals (PN 9210379, 40 g) from the electrolyte kit. Remove the red cap from the fill spout on each bottle.
7. Using a razor blade or scissors, cut the fill spout on the bottle of dry succinic acid just below the line on the spout.
8. Hold the sensor with the membrane end pointed slightly upward. Insert the spout of the bottle of succinic acid solution into the fill port. Squeeze the bottle until half of the solution has been transferred to the electrolyte chamber.
9. Pour the solid succinic acid crystals into the fill port.
If the crystals accumulate in the fill hole, shake or tap the sensor to gently unblock the port.
10. Use the remainder of the succinic acid solution to rinse crystals adhering to the threads in the sensor. Keep adding solution until it overflows the fill port. Tap the sensor a few times to be sure no air bubbles are trapped in the sensor.
11. Screw the fill plug back into place until it is flush with the body.
12. Hold the sensor with the membrane end pointing down and give it a few shakes as though shaking down a fever thermometer.
Shaking helps clear bubbles that might have become trapped behind the mesh cathode.
The sensor may require several hours operating at the polarizing voltage to equilibrate after the electrolyte had been replaced.
NOTICE
Be sure to place the sensor in flowing, chlorinated water for equilibration.
8 Rosemount 498CL
Page 13

4.3 Storage

The sensor must be stored in a flowing sample. Minimum sample flow is 0.5 gph (30 mL/min). Do not let the membrane dry out. If the sensor cannot be stored in a flowing sample:
1. Turn off power to the transmitter.
2. Remove the sensor from the flow cell.
3. Replace the membrane with the shipping membrane provided with the sensor.
For a replacement shipping membrane, order PN 23501-00.
4. To protect the sensor from physical damage, store it in the dry flow cell.

4.4 Rejuvenating a sensor following improper storage

If the sensor is stored in a stagnant sample for more than a day or two, it can become contaminated with copper. Corrosion of the external copper electrode produces copper ions, which diffuse through the membrane into the sensor. If the sensor was powered up during storage, copper will plate out on the cathode. If the sensor was not powered up, the copper will start plating out as soon as the polarizing voltage is applied. Once copper has coated the cathode, the sensor zero current will become very high, and the sensor will be unusable until the cathode has been cleaned.
1. If the sensor was not powered up during storage, do not apply power. Complete the
following steps:
Maintenance
a. Empty the fill slurry and thoroughly rinse the sensor with deionized water.
b. Refill the sensor with fresh fill slurry.
c. Let the sensor run in flowing chlorinated water overnight.
d. Zero and calibrate the sensor.
See the transmitter instruction manual for details. The zero current should be less than about 100 nA.
2. If the sensor was powered up during storage, the cathode is probably coated with
metallic copper. Complete the following steps:
a. Disconnect the sensor from the transmitter.
b. Remove the membrane and clean out the fill slurry.
c. Immerse the mesh cathode in 10% nitric acid solution (10 mL of concentrated
nitric acid in 90 mL of water) for about 5 minutes.
d. Rinse thoroughly with deionized water.
e. Refill the sensor with fresh slurry and install a new membrane.
f. Let the sensor run overnight in flowing chlorinated water.
g. Zero and calibrate the sensor.
Quick Start Guide 9
Page 14
Maintenance
The zero current should be less than about 100 nA.
10 Rosemount 498CL
Page 15
5 Accessories
Rosemount 498CL pH-Independent Free Chlorine Sensor accessoriesTable 5-1:
Part number Description
24150-01 Variopol, 6.0 interconnecting cable, 10 ft (3 m)
24150-02 Variopol 6.0 interconnecting cable, 50 ft (15 m)
24091-01 Cell, low flow, 1/4 in. inlet and outlet, with bubble shedding nozzle
33970-00 Fill plug
33521-03 Membrane retainer cap
23501-10 pH-independent free chlorine membrane assembly, includes one
23502-10 pH-independent free chlorine membrane assembly, includes three
24146-00 pH-independent free chlorine sensor electrolyte kit, includes three

Accessories

membrane assembly and one O-ring
membrane assemblies and three O-rings
bottles of saturated succinic acid solution and three bottles of suc­cinic acid crystals
Quick Start Guide 11
Page 16
www.Emerson.com/RosemountLiquidAnalysis
Emerson Automation Solutions 8200 Market Blvd Chanhassen, MN 55317 Toll Free +1 800 999 9307 F +1 952 949 7001
liquid.csc@emerson.com www.Emerson.com/RosemountLiquidAnalysis
EUROPE
Emerson Automation Solutions Neuhofstrasse 19a P.O. Box 1046 CH-6340 Baar Switzerland T + 41 (0) 41 768 6111 F + 41 (0) 41 768 6300
liquid.csc@emerson.com www.Emerson.com/RosemountLiquidAnalysis
LIQ-QSG-498CL
Rev M
2017
MIDDLE EAST AND AFRICA
Emerson Automation Solutions Emerson FZE Jebel Ali Free Zone Dubai, United Arab Emirates, P.O. Box 17033 T +971 4 811 8100 F +971 4 886 5465
liquid.csc@emerson.com www.Emerson.com/RosemountLiquidAnalysis
ASIA-PACIFIC
Emerson Automation Solutions 1 Pandan Crescent Singapore 128461 Singapore T +65 777 8211 F +65 777 0947
liquid.csc@emerson.com www.Emerson.com/RosemountLiquidAnalysis
©
2017 Emerson. All rights reserved.
The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a mark of one of the Emerson family of companies. All other marks are the property of their respective owners. The contents of this publication are presented for information purposes only, and while effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, expressed or implied, regarding the products or services described herein or their use or applicability. All sales are governed by our terms and conditions, which are available on request. We reserve the right to modify or improve the designs or specifications of our products at any time without notice.
Loading...