Rosemount 401 Endurance Conductivity Sensor Abridged Manuals & Guides

Page 1
Model 401
ENDURANCE™Conductivity Sensor
Instruction Sheet
PN 51A-401/rev.J January 2011
INSTALLATION
If the sensor is installed in a sidestream with the sample draining to open atmosphere, bubbles may accumulate on the electrodes. Trapped bubbles will cause errors. Normally, as bubbles accu­mulate the conductivity reading drifts down. To control bubble formation, apply a small amount of back pressure to the sensor.
q
q
FIGURE 1. Sensor Orientation FIGURE 2. Sensor Installation
SENSOR SPECIFICATIONS
SPECIFICATIONS
MODEL 401
Wetted Materials graphite, Kynar, epoxy, EPDM
Temperature Range 32-212°F (0-100°C)
Maximum Pressure 200 psig (1481 kPa abs)
For additional information, please visit our website
at www.emersonprocess.com/raihome/liquid/.
The wetted sensor materials may not be compatible with process com position and operating conditions. Application compat ibility is entirely the
responsi-
bility of the user.
CAUTION
SENSOR/PROCESS APPLICATION COMPATIBILITY
WARNING
Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is lowered to a safe level!
WARNING
trapped air
30q
trapped sludge
30
flow
flow
flow
recommended
not
recommended
flow
Page 2
MODEL 401 WIRING
2
WIRING DIAGRAMS
WIRING
WIRE COLOR AND CONNECTIONS IN SENSOR
COLOR FUNCTION
Gray Connects to outer electrode
Clear Coaxial shield for gray wire
Orange Connects to inner electrode
Clear Coaxial shield for orange wire
Red
White with red stripe
White
Clear Shield for all RTD lead wires
RTD
RTD in
RTD sense
RTD return
FIGURE 3. Model 54eC Wiring
FIGURE 4. Model 56 and 1056 Wiring
FIGURE 5. Model Xmt-C-10 Wiring (Panel)
FIGURE 6. Model Xmt-C-11 Wiring (Pipe or Wall)
Page 3
3
MODEL 401 WIRING
RCV B
RCV A
RSHLD
DRVB
RTN
SENSE
RTD IN
SHLD
CLEAR
WHITE/RED
RED
DRVA
DSHLD
WHITE
CLEAR
GRAY
ORANGE
CLEAR
TB1TB2
WIRING THROUGH A JUNCTION BOX
If wiring connections are made through a remote junction box (PN 23550-00), wire point-to-point. Use cable 23747-00 (factory-terminated) or 9200275 (no terminations).
FIGURE 8. Model 5081C Wiring
FIGURE 7. Model 1066 Wiring
CLEANING THE SENSOR
Use a warm detergent solution and a soft brush or pipe cleaner to remove oil and scale. Isopropyl alcohol (rubbing alco­hol) can also be used to remove oily films. Avoid using strong mineral acids to clean conductivity sensors.
CALIBRATION
ENDURANCE conductivity sensors are calibrated at the factory and do not need calibration when first placed in service. Simply enter the cell constant printed on the label into the analyzer.
After a period of service, the sensor may require calibration. The sensor can be calibrated against a solution having known conductivity or against a referee meter and sensor. If using a standard solution, choose one having conductivity of about 5000 uS/cm.
For more information about calibrating contacting conductivity sensors, refer to application sheet ADS 43-024, available on the Rosemount Analytical website.
Page 4
Credit Cards for U.S. Purchases Only.
The right people, the right answers, right now.
ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
http://www.raihome.com
Specifications subject to change without notice.
Emerson Process Management
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250
http://www.raihome.com
© Rosemount Analytical Inc. 2011
MODEL 401 TROUBLESHOOTING
TROUBLESHOOTING
PROBLEM PROBABLE CAUSE SOLUTION
Off-scale reading Wiring is wrong. Verify wiring.
Temperature element is open or shorted. Check temperature element for open or
short circuits. See Figure 9.
Sensor is not in process stream. Be sure sensor is completely submerged in
process stream.
Sensor has failed. Call the factory.
Noisy reading Sensor is improperly installed in process Be sure sensor is completely submerged
stream. in process stream.
Reading seems wrong (lower Bubbles trapped in sensor. Be sure sensor is properly oriented in or higher than expected) pipe or flow cell. See Figures 1 & 2
Apply back pressure to flow cell.
Wrong temperature correction algorithm. Check that temperature correction is
appropriate for the sample. See analyzer manual for more information.
Wrong cell constant Verify that the correct cell constant has been
entered inthe analyzer and that the cell constant is appropriate for the conductivity of the sample. See analyzer manual.
Sluggish response Electrodes are fouled Clean electrodes
Sensor is installed in dead area in piping Move sensor to a location more
representative of the process liquid.
4
FIGURE 9. Checking Temperature Element
Disconnect leads and measure resistances shown. The measured resistance at room temperature should be close to the value in the table.
q
temperature
C
0 100.0 1000 10 103.9 1039 20 107.8 1078 30 111.7 1117 40 115.5 1155 50 119.4 1194
RTD
red
red/white
white
R
8
resistance in :
Pt 100 Pt 1000
Loading...