Rosemount 400A Hydrocarbon Analyzer-Rev M (Cat # 194106 / Serial # beginning with 2000001) Manuals & Guides

Instruction Manual
748262-M May 2002
Model 400A
Hydrocarbon Analyzer
NOTE: This manual is applicable to Model 400A, Catalog Number 194106 with Serial Numbers beginning with 2000001.
http://www.processanalytic.com

ESSENTIAL INSTRUCTIONS

READ THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
Read all instructions prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Rosemount Analytical representative
for clarification.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and maintenance of
the product.
Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the
proper electrical and pressure sources.
To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the product’s performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
The information contained in this document is subject to change without notice.
Teflon is a Registered Trademark of E.I. duPont de Nemours and Co., Inc. SNOOP is a registered trademark of NUPRO Co.
Emerson Process Management
Rosemount Analytical Inc. Process Analytic Division
1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 e-mail: gas.csc@EmersonProcess.com
http://www.processanalytic.com
Model 400A
PREFACE...........................................................................................................................................P-1
Definitions ...........................................................................................................................................P-1
Safety Summary .................................................................................................................................P-2
General Precautions For Handling And Storing High Pressure Gas Cylinders .................................P-4
Documentation....................................................................................................................................P-5
1-0 DESCRIPTION AND SPECIFICATIONS..............................................................................1-1
1-1 Description.............................................................................................................................1-1
1-2 Analyzer Mounting Options ...................................................................................................1-1
1-3 Fuel Gas Options...................................................................................................................1-1
1-4 Output Options ......................................................................................................................1-2
a. Isolated Remote Range Change and Identification ........................................................1-3
b. Range Trim Option..........................................................................................................1-3
1-5 Sample Pump Option ............................................................................................................1-3
1-6 Gas Safety Features..............................................................................................................1-3
1-7 Specifications ........................................................................................................................1-4
Instruction Manual
748262-M
May 2002

TABLE OF CONTENTS

2-0 INSTALLATION ....................................................................................................................2-1
2-1 Unpacking..............................................................................................................................2-1
2-2 Location .................................................................................................................................2-1
2-3 Voltage Requirements...........................................................................................................2-1
2-4 Fuel and Air Requirements....................................................................................................2-2
a. Fuel Gas..........................................................................................................................2-2
b. Air....................................................................................................................................2-2
2-5 Sample Handling ...................................................................................................................2-2
2-6 Gas Connection.....................................................................................................................2-3
2-7 Leak Check............................................................................................................................2-4
2-8 Electrical Connections ...........................................................................................................2-4
a. Line Power Connection...................................................................................................2-5
b. Voltage Output Selection and Cable Connections for Recorder ....................................2-5
c. Voltage to Current Output Board (Optional) ...................................................................2-5
d. Auxiliary Contacts ...........................................................................................................2-5
e. Remote Range Control and Indication............................................................................2-5
f. Sample Pump Accessory................................................................................................2-7
3-0 OPERATION .........................................................................................................................3-1
3-1 Initial Startup and Calibration ................................................................................................3-1
a. Selection of Calibration Method and Associated Standard Gas(es)...............................3-3
3-2 Calibration Procedure............................................................................................................3-4
3-3 Range Switch ........................................................................................................................3-5
3-4 Range Trim Option ................................................................................................................3-6
3-5 Routine Operation .................................................................................................................3-8
3-6 Recommended Calibration Frequency..................................................................................3-8
3-7 Shutdown...............................................................................................................................3-8
3-8 Obtaining Maximum Sensitivity .............................................................................................3-8
Rosemount Analytical Inc. A Division of Emerson Process Management Contents i
Instruction Manual
748262-M May 2002
4-0 THEORY................................................................................................................................4-1
4-1 Principles of Operation ..........................................................................................................4-1
4-2 Burner ....................................................................................................................................4-1
a. Response to Different Hydrocarbons..............................................................................4-2
4-3 Analyzer Flow System ...........................................................................................................4-3
4-4 Preamplifier Board.................................................................................................................4-3
4-5 Main Electronics Board..........................................................................................................4-4
a. Post Amplifier ..................................................................................................................4-4
b. Digital Display .................................................................................................................4-4
c. Span ................................................................................................................................4-4
d. Range Trim (Option) .......................................................................................................4-4
e. Remote Range Control ...................................................................................................4-4
4-6 Heater/Fan Temperature Assembly ......................................................................................4-6
4-7 Ignition Circuit........................................................................................................................4-6
4-8 System Power Supplies.........................................................................................................4-6
4-9 Flame-Out Board ...................................................................................................................4-6
5-0 SERVICE AND TROUBLESHOOTING ................................................................................5-1
5-1 System Checkout ..................................................................................................................5-1
a. Amplifier Zero Adjustment...............................................................................................5-1
b. Zero Current Adjustment.................................................................................................5-1
c. Electrical Leakage Check ...............................................................................................5-1
d. Flame Ignition..................................................................................................................5-1
e. Noise Check....................................................................................................................5-1
f. Overall Sensitivity Check ................................................................................................5-1
g. Stability Check ................................................................................................................5-2
5-2 Servicing Flow System and Burner .......................................................................................5-2
a. Burner Disassembly and Cleaning..................................................................................5-2
b. Thermistor .......................................................................................................................5-3
c. Fuel and Air Restrictors ..................................................................................................5-3
d. Sample Capillary .............................................................................................................5-4
5-3 Troubleshooting.....................................................................................................................5-4
Model 400A
6-0 REPLACEMENT PARTS ......................................................................................................6-1
6-1 Circuit Board Replacement Policy .........................................................................................6-1
6-2 Recommended Replacement Parts List................................................................................6-1
6-3 Matrix .....................................................................................................................................6-2
7-0 RETURN OF MATERIAL ......................................................................................................7-1
7-1 Return Of Material .................................................................................................................7-1
7-2 Customer Service ..................................................................................................................7-1
7-3 Training..................................................................................................................................7-1
ii Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Model 400A
Figure 2-1. Range Input-Output Board..................................................................................... 2-6
Figure 2-2. Range Control and Output Connections................................................................ 2-7
Figure 3-1. Power Switch Location..............................................................................................3-2
Figure 3-2. Typical Curves of Downscale Response vs. Time for Various Hydrocarbons.......... 3-3
Figure 3-3. Example 1 Display .................................................................................................... 3-6
Figure 3-4. Example 2 Display .................................................................................................... 3-6
Figure 3-5. Example 3 Display .................................................................................................... 3-6
Figure 3-6. Range Trim Option Schematic.................................................................................. 3-8
Figure 3-7. Typical Curve of Analyzer Response vs. Pressure Setting on Sample Pressure .... 3-9
Figure 3-8. Typical Curves of Analyzer Response vs. Pressure Setting on Fuel Pressure
Figure 3-9. Typical Curves of Analyzer Response vs. Pressure Setting on Air Pressure
Figure 4-1. Flame Ionization Detection Theory ........................................................................... 4-1
Figure 5-1. Effect of Analyzer Temperature on Background Signal – Typical Curve.................. 5-2
Instruction Manual
748262-M
May 2002

LIST OF ILLUSTRATIONS

Regulator ............................................................................................................... 3-9
Regulator ............................................................................................................. 3-10

LIST OF TABLES

Table 1-1. Fuel Gas vs. Analyzer Characteristics................................................................... 1-3
Table 2-1. Range Control........................................................................................................ 2-6
Table 3-1. Internal Pressure Regulators Settings................................................................... 3-1
Table 3-2. Range Switch Settings........................................................................................... 3-5
Table 4-1. Approximate Effective Carbon Numbers ............................................................... 4-2
Table 4-2. Input Sensitivities...................................................................................................4-4
Table 4-3. Fullscale Sensitivity ............................................................................................... 4-4
Table 5-1. Troubleshooting ..................................................................................................... 5-5

LIST OF DRAWINGS

(LOCATED IN REAR OF MANUAL)
620424 Schematic Diagram, Preamplifier 620429 Schematic Diagram, Main Electronics 620434 Schematic Diagram, Isolated V/I 622883 Flow Diagram, Model 400A 623190 Burner Parts List 624003 Schematic Diagram, Temperature Control Board 654328 Installation Drawing, Model 400A 654905 Schematic Diagram, Relay Board 654910 Schematic Diagram, Relay Board 654913 Assembly Instructions, I/O Board 654998 Assembly Instructions, Relay Board 655354 Pictorial Wiring Diagram, Model 400A 657546 Schematic Diagram, Flame-Out Board
Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii
Instruction Manual
748262-M May 2002
Model 400A
iv Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model 400A

PREFACE

The purpose of this manual is to provide information concerning the components, functions, installation and maintenance of the 400A.
Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this module before operating it. Read this instruction manual completely.

DEFINITIONS

The following definitions apply to DANGERS, WARNINGS, CAUTIONS and NOTES found throughout this publication.
DANGER .
748262-M
May 2002
Highlights the presence of a hazard which will cause severe personal injury, death, or substantial property damage if the warning is ignored.
WARNING .
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in injury, death, or long-term health hazards of personnel.
CAUTION.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.
NOTE
Highlights an essential operating procedure, condition or statement.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-1
Instruction Manual
748262-M May 2002
Model 400A

SAFETY SUMMARY

To avoid explosion, loss of life, personal injury and damage to this equipment and on-site property, all personnel authorized to install, operate and service the Model 400A Hydrocarbon Analyzer should be thoroughly familiar with and strictly follow the instructions in this manual. Save these
instructions.

AUTHORIZED PERSONNEL

To avoid explosion, loss of life, personal injury and damage to this equipment and on-site property, all personnel authorized to install, operate and service the this equipment should be thoroughly familiar with and strictly follow the instructions in this manual. SAVE THESE INSTRUCTIONS.
DANGER.
ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel.
WARNING.
PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect safety of this product. Use only factory documented components for repair.
WARNING.
POSSIBLE EXPLOSION HAZARD
Ensure that all gas connections are made as labeled and are leak free. Improper gas connections could result in explosion or death.
P-2 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model 400A
WARNING .
POSSIBLE EXPLOSION HAZARD
Do not apply power to analyzer or ignite burner until all leak checks have been performed and until the environment of the analyzer has been determined to be non-hazardous. See Section 2-7, page 2-4 for leak check procedure.
This instrument uses a fuel containment hydrogen. The instrument is designed to protect against the formation of an explosive gas mixture within the enclosure. It must NOT be operated if the in­ternal ventilation fan is not functioning. Do NOT operate without factory installed fuel flow restric­tor in place.
Check the fuel supply and containment system for leaks, both inside and outside the analyzer, upon installation, before initial startup, during routine maintenance, or any time the integrity of the fuel containment system is broken, to assure that the system is leak-tight. Refer to Leak Check Procedure, Section 2-7, page 2-4.
The enclosure is protected by a continuous dilution air purge system, as recommended in IEC* Publication 79-2 (Third Edition - 1983). Do not operate unit unless the continuous dilution purge system is properly installed and functioning. Upon startup or loss of pressurization, purge for fif­teen minutes before restoring power unless the internal atmosphere is known to be non-explosive.
748262-M
May 2002
CAUTION .
HIGH PRESSURE GAS CYLINDERS
For safety and proper performance this instrument must be connected to a properly grounded three-wire source of power.
This analyzer requires periodic calibration with known zero and standard gases. See General Pre­cautions for Handling and Storing High Pressure Cylinders, page P-4 .
WARNING
TOPPLING HAZARD
This instrument’s internal pullout chassis is equipped with a safety stop latch located on the left side of the chassis.
When extracting the chassis, verify that the safety latch is in its proper (counter-clockwise) orientation.
If access to the rear of the chassis is required, the safety stop may be overridden by lifting the latch; how­ever, further extraction must be done very carefully to insure the chassis does not fall out of its enclosure.
If the instrument is located on top of a table or bench near the edge, and the chassis is extracted, it must be supported to prevent toppling.
Failure to observe these precautions could result in personal injury and/or damage to the product.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-3
Instruction Manual
748262-M May 2002
Model 400A
GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH
PRESSURE GAS CYLINDERS
Edited from selected paragraphs of the Compressed Gas Association's "Handbook of Compressed Gases" published in 1981
Compressed Gas Association 1235 Jefferson Davis Highway Arlington, Virginia 22202
Used by Permission
1. Never drop cylinders or permit them to strike each other violently.
2. Cylinders may be stored in the open, but in such cases, should be protected against extremes of weather and, to prevent rusting, from the dampness of the ground. Cylinders should be stored in the shade when located in areas where extreme temperatures are prevalent.
3. The valve protection cap should be left on each cylinder until it has been secured against a wall or bench, or placed in a cylinder stand, and is ready to be used.
4. Avoid dragging, rolling, or sliding cylinders, even for a short distance; they should be moved by using a suitable hand-truck.
5. Never tamper with safety devices in valves or cylinders.
6. Do not store full and empty cylinders together. Serious suckback can occur when an empty cylinder is attached to a pressurized system.
7. No part of cylinder should be subjected to a temperature higher than 125 never be permitted to come in contact with any part of a compressed gas cylinder.
8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding, precautions must be taken to prevent striking an arc against the cylinder.
°
F (52°C). A flame should
P-4 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model 400A

DOCUMENTATION

The following Model 400A instruction materials are available. Contact Customer Service Center or the local representative to order.
748262 Instruction Manual (this document)
748262-M
May 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-5
Instruction Manual
748262-M May 2002
Model 400A
P-6 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Model 400A
Instruction Manual
748262-M
May 2002
SECTION 1

DESCRIPTION AND SPECIFICATIONS

1-1 DESCRIPTION

The Model 400A Hydrocarbon Analyzer automatically and continuously measures the concentration of hydrocarbons in a gas stream. Typical applications include monitor­ing atmospheric air for low-level hydrocarbon contaminants and determining the hydrocar­bon content of exhaust emissions from inter­nal combustion engines.
The analyzer utilizes the flame ionization method of detection. The sensor is a burner in which a regulated flow of sample gas passes through a flame sustained by regulated flows of a fuel gas and air. Within the flame, the hy­drocarbon components of the sample stream undergo a complex ionization that produces electrons and positive ions. Polarized elec­trodes collect these ions, causing current to flow through an electronic measuring circuit. The ionization current is proportional to the rate at which carbon atoms enter the burner, and is therefore a measure of the concentra­tion of hydrocarbons in the original sample. The analyzer provides readout on a front-panel digital display and a selectable output for an accessory recorder.
descriptions of the principal standard and op­tional features.

1-2 ANALYZER MOUNTING OPTIONS

WARNING
INSTALLATION
For safety, the analyzer should be installed in a non-confined, ventilated space.
The standard analyzer is housed in a case designed for bench-top use, or if desired, the analyzer may be mounted in a cabinet or rack using RETMA spaced mounting holes. Outline dimensions are shown on DWG 654328.

1-3 FUEL GAS OPTIONS

For burner fuel gas, the standard analyzer re­quires 40% hydrogen/60% nitrogen or helium. Through installation of the optional 400A hy­drogen fuel kit (P/N 622576), the analyzer may be converted to use 100% hydrogen. This kit may be ordered as a factory installed option or supplied as an option for installation by the user.
To ensure stable, drift-free operation, particu­larly in high-sensitivity applications, an internal temperature controller maintains the analyzer interior at a constant 50°C. This feature mini­mizes temperature-dependent variations in (a) electronic current-measuring circuitry, and (b) adsorption/desorption equilibrium of back­ground hydrocarbons within the internal flow system.
To minimize system response time, an inter­nal sample-bypass feature provides high-velocity sample flow through the ana­lyzer.
The Model 400A may be equipped with vari­ous optional features in addition to, or instead of, the standard features of the basic instru­ment. The following paragraphs provide brief
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1
The preferred type of fuel depends on the particular application and the characteristics of the sample gas:
For measuring low-level hydrocarbons in am­bient air, or in other sample gas with relatively constant oxygen content, 100% hydrogen is preferable. It provides the highest obtainable sensitivity and the maximum stability. Zero drift caused by ambient temperature varia­tions of the fuel cylinder is somewhat lower for 100% hydrogen than for mixed fuel. (With ei­ther fuel, it is desirable to maintain cylinder temperature constant.)
For monitoring vehicular exhaust emissions, or other sample gas with varying oxygen content, mixed fuel is preferable; and a hy­drogen/helium mixture is more desirable than
Instruction Manual
748262-M May 2002
Model 400A
a hydrogen/nitrogen mixture. With this type of sample, the use of mixed fuel gas minimizes the error introduced by oxygen synergism. An effective way to reduce the effect of internal oxygen is to dilute it with an inert gas. This might be accomplished by a constant dilution of sample and calibration gases ahead of the burner but it is simpler and more accurate to provide that diluent in the form of premixed fuel. Both nitrogen and helium have been used as a diluent, but helium has proven to be most effective in improving the equality of re­sponse to the various species of hydrocar­bons.
As indicated earlier the flame output signal is optimum when the ratio of hydrogen flow to inert flow is about 40/60; therefore, this is the chosen composition for hydrogen/helium pre­mixed fuel.
The sample flow is kept low to maximize the dilution effect while still providing adequate sensitivity. The burner air flow is normally about four times the fuel flow, and changes have little effect on signal strength. For a given sample flow, the signal can be opti­mized by adjusting the fuel flow rate.
Typical flow rates with premixed fuel:
Fuel 100 cc/min
Sample 7 cc/min
Air 400 cc/min
It is worth noting that with a 40/60 premixed fuel, the above flows amount to 40 cc (8%) hydrogen, 67 cc (13%) inert plus sample and 400 cc (79%) air, which compare closely to the 30 cc (8%) hydrogen, 45 cc (12%) in­ert/sample and 300 cc (80%) air given earlier for straight hydrogen fuel.
Since the sample flow in the case of mixed fuel operation is only about one-sixth of that with straight hydrogen fuel, it is clear that higher sensitivity is obtained with straight hy­drogen fuel operation. However, in any appli­cation where the sample contains more than one species of hydrocarbon and/or a varying concentration of oxygen, the mixed fuel op­eration should be used.
The mixed fuel is recommended, not only for sample containing variable concentrations of oxygen, but also for two specific pure gas ap­plications. The first is the case of pure hydro­gen samples. The other is the case of pure oxygen samples. If straight oxygen samples are used with straight hydrogen fuel, the mix­ture entering the burner is essentially 40% H
/60% O2, which tends to produce an unsta-
2
ble signal. The mixed fuel works better. Note that the choice of fuel determines certain analyzer characteristics, as tabulated in Table 1-1.

1-4 OUTPUT OPTIONS

The standard analyzer provides (a) direct digital readout in percent of full scale on a front-panel display calibrated linearly from 0 to 100 %, (b) a selectable buffered output of 0 to
0.1 VDC, 1 VDC or 5 VDC suitable for a re­corder and (c) a 0 to 5 VDC un-buttered ac­cessory output.
An isolated output of 4 to 20 mA DC (max. load resistance 700 ohms) is obtainable through use of the optional current output board, P/N 620433, installed either during factory assembly or as a subsequent addition. When installed, this board uses the accessory 0 to 5 VDC output as an input signal and re­places this function at the output terminals.
1-2 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model 400A
Instruction Manual
748262-M
May 2002
ANALYZER CHARACTERISTICS 100% H2
Fullscale Sensitivity
Fuel Consumption 35 to 40 cc/min 75 to 80 cc/min
Operating Range for SAMPLE Pres­sure Regulator
Table 1-1. Fuel Gas vs. Analyzer Characteristics
a. Isolated Remote Range Change and
Identification
This option provides a 24 VDC operation of remote range and identification as well as providing terminals for flame out indi­cation.

b. Range Trim Option

This option provides individual potenti­ometers for each range to allow adjust­ment of individual differences in bottled gas.

1-5 SAMPLE PUMP OPTION

To provide the required sample flow, the sample gas must be under adequate pressure when applied to the analyzer inlet. Refer to Section 2-4, page 2-2. To permit analysis of gases at atmospheric or sub-atmospheric pressure, the analyzer may be shipped with a sample pump accessory, P/N 621062.
Adjustable from 1 ppm CH4 to 2% CH4
4 to 5 psig (27 to 34.5 kPa)
restrictor, ventilation holes in the enclosure and an internal circulation fan serve to dilute and dissipate the hydrogen fuel for a worst case leak to a safety factor below 25% of the LEL of hydrogen. The design basis for this system presumes 100% hydrogen fuel at 50 psig inlet pressure. 40% hydrogen fuel and lower inlet pressure serve to further reduce hydrogen concentration in the event of a leak. In reality, an open fitting leak would never oc­cur. As a leak developed the burner would eventually be starved of fuel and flame-out would occur at a leak equivalent of a loss of about 20 psig fuel pressure, thus actuating the fuel shut-off solenoid valve system.
If the sample is flammable, accessory kit P/N 624080, must be utilized. This kit provides a restrictor to limit sample flow and a solenoid valve to shut-off sample in the event of burner flame-out. The design basis for this kit pre­sumes a maximum sample flow rate of 470 cc/min and a sample with LEL not below that of hydrogen (4% v/v in air).
40% H2/60% N2 or 40%
H2/He
Adjustable from 4 ppm CH4 to 10% CH4
1.5 to 5 psig
(10.3 to 34.5 kPa)

1-6 GAS SAFETY FEATURES

The Model 400A is designed to provide a high degree of operational safety. In all analyzers, a front-panel LED indicates that the burner flame is lit. In addition, fuel gas is automati­cally shut off when a flame-out condition oc­curs.
All tubing ahead of the burner is rigid metallic tubing made up with ferrule/nut type compres­sion fittings. However, should there be an in­ternal fuel leak, an inlet fuel flow limiting
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-3
POSSIBLE EXPLOSION HAZARD
Protection against explosion depends upon special fuel flow limiting restrictor in fuel inlet fitting. Do not remove fuel inlet restrictor. Replace only with factory sup­plied fitting.
WARNING
Instruction Manual
748262-M May 2002

1-7 SPECIFICATIONS

Power Requirements..................... 115 VAC ±10%, 50/60 ±3 Hz, 250 W
Operating Temperature ................. 32°F to 110°F (0°C to 43°C)
Case Temperature ........................ Controlled at 122°F (50°C)
Ambient Humidity .......................... 95% relative humidity, but not in excess of 34°C wet bulb
Dimensions.................................... 8.75 in (22.2 cm) H
Weight ........................................... 22 lbs (10 kg)
Repeatability.................................. 1% of fullscale for successive identical samples
Response Time ............................. 90% of fullscale in 0.6 seconds with sample bypass flow at 3
Analyzer Fullscale Sensitivity
Standard ............................... Adjustable from 4 ppm CH4 to 1% CH4. (Adjustable from 100 ppm
Equipped with 100%
Hydrogen Fuel Assembly ...... Adjustable from 1 ppm CH4 to 0.25% CH4
Analyzer Fuel Gas Requirements
Standard Analyzer ................. 75 to 80 cc/min premixed fuel consisting of 40% hydrogen and 60%
Equipped with 100%
Hydrogen Fuel Assembly ...... 35 to 40 cc/min of clean, zero grade hydrogen (THC <0.5 ppm) at
Sample Gas Requirements
Non-Flammable Samples ...... 0.35 to 3.0 liters/minute at 5 to 10 psig (34 to 69 kPa)
Flammable Samples.............. 470 cc/minute maximum for safety1
Burner Air Requirements............... 350 to 400 cc/minute of zero grade (THC <1 ppm) air, supplied at
Sample Bypass Flow..................... 0.3 to 3.0 liters/minute
Stability .......................................... Electronic stability at maximum sensitivity is 1% of fullscale
Range ............................................ Eight ranges: 1, 2.5, 10, 25, 100, 250, 1000 and REMOTE.
Model 400A
temperature.
18.75 in. (47.6 cm) W
15.88 in. (39.7 cm) D Recommended panel cutout is 17.75 in. X 8 25 in. (45.1 cm x 21.0 cm). May be mounted in standard 19 inch rack.
liters/minute
CH4 to 10% CH4 using high-range capillary.)
nitrogen or helium (THC <0.5 ppm) supplied at 45 to 50 psig (309 to 344 kPa) at instrument
45 to 50 psig (309 to 344 kPa) at instrument
25 to 50 psig (172 to 344 kPa)
throughout ambient temperature range of 32°F to 110°F (0°C to 43°C). Built-in temperature controller minimizes effect of ambient temperature variations on internal flow and electronic systems.
In addition SPAN control provides continuously variable adjustment within a dynamic range of 4:1
1
Safety design basis presumes flammable sample having LEL not less than that of hydrogen (4% v/v in air).
1-4 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model 400A
Output............................................ 1) 0 to 5 VDC, 0 to 1 VDC, 0 to 0.1 VDC fully buffered - standard
Safety Features ............................. Flame-on indication and automatic flame-out fuel shutoff is
Contacts ........................................ Form A contact operates in parallel with flame-out fuel shut-off
Temperature Control ..................... Setpoint maintained at 122°F (50°C)
Data Display .................................. 3-1/2 digit LED, characters 0.52 inches high, range 0000 to 1999
Range Display ............................... 1 digit LED, character 0.52 inches high (1 to 7 normal ranges, 0 to
Remote Range Control.................. Standard, fully isolated range control and range ID is optional
Instruction Manual
748262-M
May 2002
(for 0 to 100.0%).
2) 4 to 20 mA isolated voltage to current - optional (maximum load resistance 700 ohms)
3) 0 to 5 VDC accessory output un-buffered - standard (for 0 to
100.0%) available when current option is not used.
standard.
All metal tubing with ferrule/nut compression fittings to minimize potential fuel leaks. Self-ventilated system maintains internal atmosphere below 25% of LEL for worst case internal leakage.
solenoid contact rating (24 VDC at 1 A) for sample shut-off by use of factory ordered kit (PN 624080) if sample is flammable (hydrogen).
remote control)
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-5
Instruction Manual
748262-M May 2002
Model 400A
1-6 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model 400A
Instruction Manual
748262-M
May 2002
SECTION 2

INSTALLATION

2-1 UNPACKING

Carefully examine the shipping carton and contents for signs of damage. Immediately notify the shipping carrier if the carton or its contents are damaged. Retain the carton and packing materials until the instrument is op­erational.

2-2 LOCATION

Install analyzer in a clean area, not subject to excessive vibration or extreme temperature variations. Preferably, the analyzer should be mounted near the sample stream, to minimize sample-transport time.
WARNING
INSTALLATION RESTRICTIONS
For safety, the analyzer should be installed in a non-confined, ventilated space. Do not block any of the vent holes at the top of each side panel of the instrument as they are part of the safety system.
A thermostatically controlled heating circuit holds internal temperature of the analyzer to the correct operating temperature for ambient temperatures in the range 32°F to 110°F (0°C to 43°C).

2-3 VOLTAGE REQUIREMENTS

WARNING
ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified per­sonnel.
For safety and proper performance this in­strument must be connected to a properly grounded three-wire source of power.
This instrument was shipped from the factory configured to operate on 115 VAC, 50/60 Hz electric power. Verify that the power source conforms to the requirements of the individual instrument, as noted on the name-rating plate.
NOTE:
230 VAC operation requires an accessory transformer mounted external to the in­strument. If external 230 VAC power is provided, do not change the switch setting on the Temperature Control Board, which is factory-selected for 115 VAC. A 230 VAC switch setting will cause the case tempera­ture controller to malfunction.
The cylinders of fuel, air, and calibration gas(es) should be located in an area of rela­tively constant ambient temperature.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1
Instruction Manual
748262-M May 2002
Model 400A

2-4 FUEL AND AIR REQUIREMENTS

WARNING
INSTALLATION RESTRICTIONS
Fuel, air and calibration gas cylinders are under pressure. Mishandling of gas cylin­ders could result in death, injury or prop­erty damage. Handle and store cylinders with extreme caution and in accordance with manufacturer's instructions. Refer to General Precautions for Handling and Storing Gas Cylinders, page P-4.
During normal operation, the analyzer uses fuel and air to maintain the burner flame. Cri­teria for selection of these gases are given in Sections 2-4a (page 2-2) and 2-4b (page 2-2). In addition, the analyzer requires suitable standard gas(es) for calibration. Refer to Sec­tion 3-1a (page 3-3).
Each gas used should be supplied from a tank or cylinder equipped with a clean, hydrocar­bon-free, two-stage regulator. In addition, a shutoff valve is desirable. Install the gas cylin­ders in an area of relatively constant ambient temperature.

a. Fuel Gas

NOTE:
Always assure the sample flow is pre­sent when using the 100% hydrogen fuel option. Absence of sample flow can result in burning of detector tip when using 100% hydrogen.

b. Air

Burner air should also be relatively free of hydrocarbons in order to assure a low background signal. Several grades of air are supplied by various gas vendors for this use. A maximum total hydrocarbon content of less than 1 ppm (THC < used as a zero standard).
An alternate source of pure air for burner and zero gas can be provided by a dia­phragm pump and heated palladium catalyst which effectively removes mod­erate amounts of both hydrocarbons and carbon monoxide from normal ambient air on a continuous basis.

2-5 SAMPLE HANDLING

CAUTION
BYPASS GAUGE PROTECTION
The standard analyzer is equipped to use only mixed fuel, i.e. 40% hydrogen/60% nitrogen or helium. Such blends are sup­plied by many gas vendors specifically for this use, with a guaranteed maximum to­tal hydrocarbon content of 0.5 ppm, measured as methane (THC < 0.5 ppm). This specification should be used when buying such mixtures.
When the analyzer is equipped with the optional hydrogen fuel kit, P/N 622576, 100% hydrogen fuel is to be used. This is also supplied by many gas vendors spe­cifically for this use, with the same guar­anteed total hydrocarbon content (THC <
0.5 ppm) which, again, should be speci­fied when buying the gas.
2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
When applying sample pressures greater than 5 psig, insure that the bypass regula­tor is fully open to protect the bypass gauge.
Operating range for the internal sample pres­sure regulator is 4 to 5 psig (28 to 35 kPa) for an analyzer using 100% hydrogen fuel, and
1.5 to 5 psig (10 to 35 kPa) for an analyzer using mixed fuel. With either fuel, sample and calibration gas(es) must be supplied to the sample inlet at a pressure slightly, but not ex­cessively higher than the desired setting on the internal sample pressure regulator. The criterion for correct supply pressure is that the gas flow discharged from the by-pass outlet must be between 0.5 and 3.0 liters/minute to operate within the control range of the sample pressure regulator, and preferably should be between 2 and 3 liters/minute to minimize system response time.
Model 400A
Instruction Manual
748262-M
May 2002
Note that use of excessive bypass flow will not only cause the sample pressure regulator to operate outside its control range, but will also result in rapid depletion of sample and standard gases.
If the analyzer is equipped with the accessory 400A sample pump, P/N 621062, the accept­able pressure range at the pump inlet is ap­proximately -1 to +2.5 psig (7 to 17 kPa). If the pump is used, it will automatically provide a sample bypass flow within the correct range. If the analyzer is not equipped with sample pump, adjustment of the bypass flow is ob­tainable by inserting an external flow control­ler or throttle valve into the external sample line, upstream from the sample inlet. Flow may be measured by connecting a flowmeter to the by-pass outlet.
WARNING
POSSIBLE EXPLOSION HAZARD
In the event that flammable sample is to be introduced into this analyzer, it must be equipped with accessory kit PN 624080, which restricts sample flow and provides automatic sample shutoff in the event of burner flameout. DO NOT OPERATE WITHOUT SAMPLE FLOW RESTRICTOR IN PLACE. The sample containment system should also be thoroughly leak checked. This kit is designed considering applica­tion on hydrogen sample (LEL=4% v/v). The instrument must not be used on a sample having a LEL less than 4% in air.
WARNING
POSSIBLE EXPLOSION HAZARD
Do not apply power to analyzer or ignite burner until all leak checks have been per­formed and until the environment of the analyzer has been determined to be non­hazardous. See Section 2-7 (page 2-4) for leak check procedure.
This analyzer has been designed for use in environments that do not contain combus­tible or explosive materials.
This analyzer uses a fuel containing hy­drogen. Leakage from the fuel contain­ment system can result in an explosion. The fuel supply and containment system, both inside and outside the analyzer, should be carefully checked for leaks upon installation, before initial startup, during routine maintenance or any time the integ­rity of the sample containment system is broken.
If hazardous sample is to be introduced into this analyzer, the leak check proce­dure should also be applied to the sample containment system, both inside and out­side the analyzer.
Proceed as follows:
1. Check analyzer to make sure that plugs and caps are removed from all inlet and outlet fittings.

2-6 GAS CONNECTION

For external gas lines, the use of all new tub­ing throughout is strongly recommended. Copper refrigeration tubing is preferred. Stainless steel tubing is less desirable, be­cause it contains hydrocarbon contaminants, necessitating thorough cleaning before in­stallation.
NOTE
In connecting gas supply lines and asso­ciated fittings, use Teflon tape only. Do not use pipe thread compound or other sub­stance with an organic base.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3
2. If a vent line is to be connected to exhaust outlet, use 1/2-inch ID tube slanted downward at least 10 degrees from hori­zontal.
NOTE:
Since water vapor is formed during oxida­tion of hydrogen, burner exhaust gas al­ways contains moisture, even if air and fuel entering the burner are completely dry. Unless exhaust line slants down, wa­ter may accumulate in line, causing back pressure and noisy readings. If exhaust line becomes blocked, water may back up in line and flood burner.
Instruction Manual
748262-M May 2002
Model 400A
3. If sample is toxic or noxious, or is to be reclaimed, connect by-pass outlet to suit­able disposal system. Do not use any de­vice causing back pressure on burner.
4. Clean external fuel, air, and sample lines and regulators. If necessary, heat lines with torch to drive out contaminants.
CAUTION
DISCONNECT ALL LINES
Do not perform this operation with the fuel, air and sample lines connected to the analyzer
5. Recommended method is to attach the tubing to either a nitrogen or helium cylin­der through a two-stage regulator, and adjust the regulator for a low flow of gas through the tubing. Use a propane or natural gas torch to heat the tubing to at least 300 slowly from the regulator end to the open end. This will remove contaminants from the inside walls of the tubing, and drive them out the open end.
6. Connect external fuel and air lines to fuel and air inlet fittings on analyzer. Connect external sample line to sample inlet on analyzer (or to inlet fitting on sample pump, if used).
7. Adjust regulators on fuel and air cylinders (or other gas supply sources) for appro­priate output pressure. Maximum permis­sible pressure at AIR and FUEL inlets of analyzer is 50 psig (345 kPa). The pres­sure at the AIR inlet must be at least 5 psig (35 kPa) higher than the desired set­ting on the air pressure gauge within the analyzer. Thus if the internal FUEL pres­sure regulators are to be set at a typical value of 25 psig (172 kPa), the pressure at the FUEL inlet must be set 15 to 20 pounds higher than the operating pres­sure.
8. Supply sample gas at appropriate pres­sure, as explained in Section 2-5 (page 2-
2). Sample bypass flow must be between
o
C, working the heat source
0.5 and 3.0 liters/minute for proper opera­tion. Preferably, it should be between 2.0 and 3.0 liters/minute, to minimize system response time. Flow may be measured by connecting a flowmeter to BY-PASS out­let.

2-7 LEAK CHECK

WARNING
POSSIBLE EXPLOSION HAZARD
Be particularly careful in checking for leaks in the fuel lines. Fuel gas leakage can cause an explosion.
Check all gas connections to ensure that they are leak free. Use of SNOOP (P/N 837801) or other suitable leak-test liquid is recom­mended. Do not use soap or other organic substances; they will contaminate the system, resulting in excessive noise and background current. To leak check the fuel containment system, it is necessary to have full operating pressure within the system. To accomplish this, hold the momentary IGNITE/PURGE switch in the up or PURGE position.

2-8 ELECTRICAL CONNECTIONS

WARNING
POSSIBLE EXPLOSION HAZARD
Before supplying electrical power to ana­lyzer, complete the gas connections and verify that fuel gas connections are leak free. Refer to Section 2-7 (page 2-4).
WARNING
ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified per­sonnel.
2-4 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 400A
Instruction Manual
748262-M
May 2002
a. Line Power Connection
The Model 400A is manufactured to oper­ate only on 115 VAC power. (Note that the power input box and power selection card display only 115 VAC operation.)
The power cable supplied is provided with a North American-style parallel blade grounded plug which must be inserted into a 3-wire grounded receptacle.
If 230 VAC operation is desired, an exter­nal isolated step-down transformer ac­cessory (not supplied) is required.
b. Voltage Output Selection and Cable
Connections for Recorder
The standard analyzer provides voltage output only. As shipped from the factory, the analyzer is set up for use with a 100 mV recorder (program header connecting pins E9 and E10).
A selection of three voltage output ranges, 0 to 0.1 VDC, 0 to 1.0 VDC or 0 to 5.0 VDC are available at terminals 3 and 4 of the lower six-position barrier strip labeled OUTPUT. Refer to Figure 2-1, page 2-6. To select 0 to 1.0 VDC range, connect pins E7 and E8 on the main electronics board assembly, P/N 620428, with the program header. For the 0 to 5.0 VDC range, connect pins E5 and E6, and for the 0 to 0.1 VDC range, connect pins E9 and E10, with the program header. This is a fully buffered signal and can be used with most types of voltage record­ers. A 10 VDC output displays 1999 on digital readout, indicating 99.9% over­range. A 5 VDC output displays 1000 on digital readout and indicates 100%, the normal instrument span.
c. Voltage to Current Output Board (Op-
tional)
The optional current output in the range of 4 to 20 mADC appears at terminals 5 and
6. Refer to Figure 2-1, page 2-6. This cur­rent may be transformed back to a volt­age using the appropriate resistor. This
fully isolated current board is an instru­ment option and mounts internally on the Model 400A front panel assembly. The maximum value of load resistor is 700 ohms. The 4 to 20 mA is valid over the range of 0 to 100%; the overrange capa­bility of 99.9% is not usable with this op­tion because maximum output of 20 mA corresponds to 5 VDC out.
If the voltage to current option is not util­ized, insert program header connecting pins E1/E2 and E3/E4 on the main elec­tronics board assembly to allow a voltage to appear at terminals 5 and 6 in place of the current.
d. Auxiliary Contacts
A Form A contact closure is available on pins 1 and 2 of the lower barrier strip at the rear of instrument. Refer to Figure 2-1, page 2-6. These contacts may be used with an existing alarm panel or an­nunciator system, providing the current and voltage limits are observed. The rat­ing for the contacts is 24 V at 1A DC. Contacts operate in parallel with the inter­nal fuel shutoff solenoid and may there­fore be used for external "flame out" indication.

e. Remote Range Control and Indication

The Model 400A allows remote control of range or alternatively remote control indi­cation; this is standard with each unit. As an option, a fully isolated interface may be installed. It performs the same function but assures electrical isolation between the analyzer and the control or indication module. The isolated range/indication op­tion requires an external 24 VDC power supply for operation.
Range Control (Non-Isolated)
Terminals 1 through 7 on the upper termi­nal block at the rear of the instrument are used for this function. Select RMT on the front panel RANGE switch. Connect the respective line, 1 through 7, to pin 3 (la­beled GND) on the lower barrier strip. The
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5
Instruction Manual
(+)
)
748262-M May 2002
Model 400A
front panel will indicate the range se­lected. Terminals 1 through 7 correspond to ranges 1 through 7. Table 2-1 below shows a simple range selection arrange­ment.
Range Indication (Non-Isolated)
When the RANGE switch is in any posi­tion, including RMT, the range selection of
1 2 3 4 5 6 7 8 9 10
1 2 3 4 5 6 7 REM 24V 24V
J8
24VDC MAX
1 AMP
CNT1 CNT2 SND OUT

RANGE CONTROL

RANGE RET
VI
(-
the instrument may be determined by sensing an output in terminals 1 through 8 on the upper terminal strip at the rear of the instrument. Any selected range, 1 through 7 or RMT (position 8), will be in­dicated as a low or ground signal with re­spect to pin 3 of the lower barrier strip labeled GND; all un-selected ranges will indicate a high or +5 VDC with respect to this same terminal.
1 2 3 4 5 6
OUTPUT
Figure 2-1. Range Input-Output Board
LOWER TB
UPPER TB
Terminal 3 1 2 34567
Range 1 X X
Range 2 X X
Range 3 X X
Range 4 X X
Range 5 X X
Range 6 X X
Range 7
XX
Table 2-1. Range Control
2-6 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 400A
Instruction Manual
748262-M
May 2002
Isolated Remote Range Control and Identification
The Remote Control Assembly is located on the rear panel of the instrument and is shown in Figure 2-2 below. Connections should be made to the terminals marked Range Control where a single line inter­ference is desired. By removing steering diodes on printed circuit board 654910, individual Range Control and Range I/O connections may be made. Connections are made to bare wire terminal which are easily removable to facilitate installation.
Wires should not be tinned prior to instal­lation to prevent premature loosening.
NOTE:
The terminal strips on the module use screws to secure bare wires. These wires should not be tinned or they may loosen prematurely.
Range changes are achieved by applying 24VDC at about 150 mW to the terminals marked "range control," using the com­mon terminal as return. Only one range may be selected at once. Range indica­tion is achieved by individual 24 VDC (1 A) relay contacts.
OUTPUT
V mA
+ - + - C NO
FLAME OUT 24VDC MAX
WARNING
ELECTRICAL HAZARD
Range indication relay contacts are limited to 24 VDC. Use of 115 VAC is unsafe.
Common control/sense connections may be made to the range control terminals.
Voltage and current output signals are also available on the range change mod­ule. Voltage output must be used with a load of greater than 10 kilohms, current output with less than 600 ohms. Voltage output interconnects should be shielded (use the V- output for shield connection) but current output shielding is less critical. If used, current output shielding should be grounded on the reception end (computer or chart recorder).
Remote Flame Out Indication
Terminals for flame out indication (TS12) are shown in Figure 2-2 below.

f. Sample Pump Accessory

If a sample pump is used, 115 VAC, 50/60 Hz power must be provided to the pump accessory independently. A power cord is provided with this option and mates with a standard 3-pin power con­nector at the rear of the housing.
TS12
CONTROL LINES 24 VDC, 150 Mw
REMOTE
24V ID
C 7 6 5 4 3 2 1
C 7 6 5 4 3 2 1
ALL CONTACTS 24 VDC MAX, 1 A MAX CURRENT OUTPUT 600 OHMS MAX
TS10 RANGE ID 24 VDC MAX
TS11 RANGE CONTROL 24 VDC
SINGLE LINE CONTROL/SENSE USE "RANGE CONTROL" CONNECTIONS
Figure 2-2. Range Control and Output Connections
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-7
Instruction Manual
748262-M May 2002
Model 400A
2-8 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 400A
Instruction Manual
748262-M
May 2002
SECTION 3

OPERATION

3-1 INITIAL STARTUP AND CALIBRATION

Section 3-1a (page 3-3) discusses calibration methods and the associated standard gases. Section 3-2 (page 3-4) explains the typical calibration procedure.
After installing analyzer per Section 2, pro­ceed as follows:
1. Set the RANGE multiplier switch (located inside the small door on the analyzer front panel) at 1000. Place POWER switch (lo­cated in the upper right-hand corner of the main electronics board on the back of the front panel) at ON. See Figure 3-1, page 3-2. The display should light up.
2. Set external regulators on air and fuel cylinders (or other gas supply sources) for suitable output pressure. Maximum per­missible pressure at AIR and FUEL inlets of analyzer is 50 psig (345 kPa). The pressure at the AIR inlet must be at least 5 psig (25 kPa) higher than the desired setting on the air pressure gauge within the analyzer. The internal FUEL pressure regulator is to be set at a typical value of 25 psig (172 kPa), the pressure at the FUEL inlet must be set at least 20 lbs. higher than the operation pressure.
3. If analyzer uses 100% hydrogen fuel, supply actual sample or other suitable gas to SAMPLE inlet port of analyzer (or to inlet fitting on sample pump, if provided). Note pressure and flow requirements ex­plained in Section 2-4, page 2-2.
CAUTION
POSSIBLE BURNER DAMAGE
If analyzer uses 100% hydrogen fuel, an adequate flow of sample or other gas must enter sample inlet at all times when flame is burning. Otherwise, burner will over­heat and damage burner tip.
If analyzer uses mixed fuel, sample flow may be initiated at this time, although it is not necessary.
4. Set internal pressure regulators at values appropriate to the fuel gas used. Refer to Table 3-1 below.
Internal
Pressure
Regulator
Air Fuel Sample
Table 3-1. Internal Pressure Regulators Settings
The PURGE/IGNITE switch has two posi­tions, IGNITE and PURGE.
5. With PURGE/IGNITE switch in PURGE position, wait about one minute for fuel gas to purge flow system. During purging period, rotate FUEL pressure regulator alternately clockwise and counterclock­wise several times, then return to setting specified in Step 4.
6. Briefly hold (2 to 4 sec.) PURGE/IGNITE switch in IGNITE position, then release. FLAME indicator should now be ON, indi­cating that flame is burning. If so, pro­ceed with following steps. If FLAME indicator does not stay on, the flame is not burning. Again actuate IGNITE switch If flame does not ignite after several at­tempts, refer to troubleshooting chart, Table 5-1 (page 5-5). If difficulty is experi­enced, allowing gas to flow for 20 or 30 seconds in the PURGE position prior to actuating the IGNITE switch may be helpful.
If ignition indication [FLAME ON] is ob­served without obtaining proper sensitiv­ity, refer to Table 5-1 (page 5-5).
100% H2 Fuel Mixed Fuel
psig/kPa psig//kPa
5/35 5/35
25/175 30/207
5/35 0/0
Fuel Gas
NOTE:
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-1
Instruction Manual
748262-M May 2002
Model 400A
NOTE:
When lighting the burner after extended shutdown, the instrument will require time to allow fuel to reach the burner. There­fore, extended operation of the switch may be required.
POWER
OFF ON
S2
Front Panel
Power Switch S2
Main Electronics Board
Figure 3-1. Power Switch Location
7. Increase setting on internal AIR pressure regulator to at least 15 psig (103 kPa). Recommended operating settings for in­ternal pressure regulators are AIR, 15 psig (103 kPa); FUEL, 25 psig (172 kPa). Verify that FLAME indicator is still on.
The analyzer is equipped with an auto­matic fuel shutoff solenoid. If flame goes out during subsequent operation, fuel gas flow will shut off automatically.
If analyzer has been in regular use, it is now ready for calibration, per Section 3-2 below (page 3-4), and then for normal op­eration. However, during initial startup, or startup following a prolonged shutdown, the following steps should first be per­formed.
8. Check for contamination in air and fuel systems:
a. Supply a clean, hydrocarbon-free gas,
such as pure nitrogen to the SAMPLE inlet. Adjust external flow controller or throttle valve so that flow discharged from BY-PASS outlet is between 0.5 and 3.0 liters/minute (preferably be-
tween 2.0 and 3.0 liters/minute). Set internal SAMPLE pressure regulator at
5.0 psig (34.5 kPa).
b. Set RANGE switch at 10, SPAN con-
trol at 1000 (maximum sensitivity), and ZERO control at 1000 (minimum zero suppression). Approximate fullscale sensitivity is now 10 ppm as methane for mixed fuels if the analyzer uses 100% hydrogen fuel, and 40 ppm methane if the analyzer uses mixed fuel.
c. Check display. Maximum acceptable
reading is 50% of fullscale. A higher reading indicates that the contamina­tion level is undesirably high. Exces­sive noise and baseline drift may result, depending on the desired oper­ating range. If the instrument is to be operated at high sensitivity, the source of the contamination must be deter­mined and the condition corrected. The most probable contamination sources are the fuel and air supplies, external regulators and connecting lines, and the internal flow system of the analyzer.
If the instrument is to be operated at a sensitivity low enough so that the noise and drift will not be observable on the display or recorder, removal of the source of contamination is unnec­essary.
9. With flame burning, allow system to stabi­lize for at least two hours, and preferably for a day. After initial startup, or after startup following a prolonged shutdown, the analyzer may display baseline drift for a considerable period of time, particularly on the more-sensitive ranges. Commonly, small amounts of hydrocarbons are pres­ent on the inner walls of the tubing in both the internal flow system and the external gas-supply system. Drift results from any factor influencing the equilibrium of these absorbed hydrocarbons. Typical causes are change of fuel cylinders or change in temperature or pressure. (Note that this type of drift occurs only when the flame is
3-2 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 400A
Instruction Manual
748262-M
May 2002
burning. If drift occurs when the flame is extinguished, the electronic circuitry is at fault.)
To minimize drift, use clean fuel and air, keep the analyzer clean, and locate the gas cylin­ders in an area of relatively constant ambient temperature.
a. Selection of Calibration Method and
Associated Standard Gas(es)
Preparatory to normal operation of the analyzer, it is necessary to select a suit­able calibration method and appropriate standard gas(es). Proper choice depends on the type of fuel gas, the intended op­erating range, and the desired accuracy. In all methods, the objective is to estab­lish both a downscale point and an up­scale point on the display or recorder chart. Different methods are described below.
Downscale Calibration P
O
The downscale calibration point is set with the ZERO control, by the appropriate one of two methods:
1. The generally preferred method is to adjust the ZERO control while a zero standard gas of low, accu­rately-known, hydrocarbon content is
int
entering the SAMPLE inlet port. This method is desirable with all analyzers, and is mandatory if the analyzer util­izes 100% hydrogen fuel. Typically, nitrogen (zero gas grade) is used as the zero gas. If desired, the burner air may be used as zero gas, provided that its hydrocarbon content is suffi­ciently low and accurately known. Al­though ideally the zero gas should be completely hydrocarbon-free, even the most carefully prepared bottled gas contains trace hydrocarbons. If the analyzer is to be used at high sensitivity, request that the supplier of the zero gas provide an exact deter­mination of its hydrocarbon content.
2. If the analyzer utilizes mixed fuel, an alternative method eliminates the re­quirement for a special zero gas. In­stead, the ZERO control is adjusted with no gas entering the SAMPLE in­let. This method is not possible with analyzer utilizing 100% hydrogen fuel, because the burner would overheat. Even with mixed fuel, the method is not recommended if the analyzer is to be used at sensitivity less than 100 ppm full scale.
3. Refer to Figure 3-2 below for typical curves of downscale response versus time for various hydrocarbons.
Sample CH4 (Methane)
Sample C3H8 (Propane)
Analyzer Response
Sample C6H
Time
14
(Hexane)
Figure 3-2. Typical Curves of Downscale Response vs. Time for Various Hydrocarbons
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-3
Instruction Manual
748262-M May 2002
Model 400A
Upscale Calibration Point
In all applications, the upscale calibration point is established by adjustment of the SPAN control, while a standard gas of ac­curately known hydrocarbon content is flowing into the SAMPLE inlet port. Since instrument response is linear, it is not necessary that the hydrocarbon content of the span gas fall within the desired oper­ating range. The instrument may be stan­dardized on one range, and then switched to another range without loss of accuracy. Commonly, a conveniently obtained stan­dard such as 100 ppm methane or 1000 ppm methane is used regardless of range.
A span gas consists of a specified con­centration of methane or other hydrocar­bon in a background gas such as nitrogen. Instrument response is affected by the composition of the background gas. Therefore, it is desirable that the span gas contain the same background gas as the actual sample. If so, the back­ground effect is automatically canceled out.
Standard Gas(es)
Upscale standard gas (and zero standard gas, if used) should be supplied from a tank or cylinder equipped with a clean, hydrocarbon-free, two-stage pressure regulator.
(34 kPa) or other desired value. (Recommended operating range is 4 to 5 psig [28.6 to 34 kPa] for analyzer using 100% hydrogen fuel, and 1.5 to 5 psig [10 to 34 kPa] for analyzer us­ing mixed fuel.)
CAUTION
POSSIBLE BURNER DAMAGE
If analyzer uses 100% hydrogen fuel, an adequate flow of sample or other gas must enter sample inlet at all times when flame is burning. Otherwise, burner will over­heat and damage burner tip.
b. Set RANGE switch at 10 and SPAN
potentiometer at 1000. The resultant approximate fullscale sensitivity is 10 ppm methane for an analyzer using 100% hydrogen fuel, and 40 ppm methane for an analyzer using mixed fuel.
c. Adjust ZERO control for reading of
zero (or appropriate near-zero value) on indicator or recorder. The resultant setting on the ZERO control is ap­proximately correct, and is sufficiently accurate for most applications. How­ever, if instrument is to be used for high sensitivity analysis, a recheck and possible slight readjustment will be made. Refer to Section 3-8, page 3-8.

3-2 CALIBRATION PROCEDURE

After completing startup procedure of Section 3-1a (page 3-3), calibrate analyzer as ex­plained below.
1. Set downscale calibration point as follows:
a. Supply zero gas to SAMPLE inlet. Ad-
just external flow controller or throttle valve so that flow discharge from BY-PASS outlet is between 0.5 and
3.0 liters/minute (preferably between
2.0 and 3.0 liters/minute). Set internal SAMPLE pressure regulator at 5 psig
3-4 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
2. Set upscale calibration point as follows:
a. Turn RANGE switch to setting appro-
priate to the particular span gas. Re­fer to Table 3-2, page 3-5.
b. Supply span gas to SAMPLE inlet.
Adjust external flow controller or throttle valve so that flow discharged from BY-PASS outlet is between 0.5 and 3.0 liters/minute (preferably be­tween 2.0 and 3.0 liters/minute). Ver­ify that reading on internal SAMPLE pressure gauge is 5 psig or other de-
Model 400A
Instruction Manual
748262-M
May 2002
sired value; if not, adjust SAMPLE pressure regulator as required.
c. Adjust SPAN control so that the dis-
play or recorder gives the desired in­dication. Lock SPAN control by pushing lever down.
Analyzer calibration is now sufficiently accurate for most applications. How­ever, if instrument is to be used for high sensitivity analysis, recheck zero

RANGE SWITCH SETTINGS

1 0 to 1 ppm CH4 0 to 4 ppm CH4
2.5 0 to 2.5 ppm CH4 0 to 10 ppm CH4
10 0 to 10 ppm CH4 0 to 40 ppm CH4
25 0 to 25 ppm CH4 0 to 100 ppm CH4
100 0 to 100 ppm CH4 0 to 400 ppm CH4
250 0 to 250 ppm CH4 0 to 1000 ppm CH4
1000 0 to 1000 ppm CH4 0 to 4000 ppm CH4
setting (refer to Section 3-8, page 3-
8). If recorder readout and display do not agree, correct display by adjusting R1 on front panel board.
d. Supply zero gas to SAMPLE inlet as in
Step 1. a. Set RANGE switch at 10. Note reading on indicator or recorder; if incorrect, adjust ZERO control as required. Lock ZERO control by pushing lever down. Analyzer is now ready for routine operation as ex­plained in Section 3-5 (page 3-8).
APPROXIMATE OPERATING RANGE SPAN CONTROL AT
1000
Note: For best results, calibrate with appropriate span gas every time the range is changed. For a clear understanding of the function of the Range Switch, see Section 3-3 below.
Table 3-2. Range Switch Settings

3-3 RANGE SWITCH

The operator can choose from seven range multipliers as represented by the settings on the Range Switch: 1, 2.5, 10, 25, 100, 250, and 1000.
Range 1 is the most sensitive, range 1000 the least sensitive. Range 1 is 10 times more sensitive than Range 10 and so forth.
The Range Switch settings (multipliers) and the display outputs do not represent hydro­carbon concentrations in percent or ppm. The LED display shows the percent of the fullscale
range which has been calibrated on the multi­plier selected. The display also shows which range multiplier has been selected on the Range Switch (1 = 1, 2 = 2.5, 3 = 10, 4 = 25, 5 = 100, etc.).
Range 1 has a maximum sensitivity of 1 ppm CH
4
(as 100% fullscale) if 100% H2 fuel is used, or 4 ppm CH Range 2.5 has a maximum sensitivity of 2.5 ppm CH mixed fuel.
4
with 100% H2, 10.0 ppm CH4 with
4
if mixed fuel is used.
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-5
Instruction Manual
748262-M May 2002
Example 1:
With the Range Switch set at 2.5, the operator can use, for example, 6 ppm CH calibrate the instrument for 20 ppm fullscale (30.0% x 20 ppm CH 3-3 below for display output and Range Switch setting.
4
= 6.0 ppm). See Figure
4
span gas to
Model 400A
Example 3:
Likewise, after the calibration in Example 1, Range 1 is automatically calibrated for 8 ppm as 100% fullscale. Therefore, when switching from Range 2.5 to 1 (which is 2.5 times more sensitive than 2.5), the display will show
4
75.0% fullscale for 6.0 ppm CH gas (because 75.0% x 8 ppm CH
4
CH
). See Figure 3-5 below.
calibration
4
= 6 ppm
30.0 2
PERCENT OF RANGE RANGE
RANGE MULTIPLIER
100
25
10
2.5
250
1000
1
RMT
Figure 3-3. Example 1 Display
Example 2:
After the calibration in Example 1, Range 10 is now automatically calibrated for 80 ppm
4
CH
as 100% fullscale. (Note that Range 10 is 4 times less sensitive than Range 2.5). Therefore, when the operator switches from Range 2.5 to Range 10, the display will show
7.5% fullscale for ppm CH since 7.5% x 80 ppm CH
4
calibration gas
4
= 6.0 ppm CH4.
See Figure 3-4 below.
75.0 1
10
2.5
Figure 3-4. Example 2 Display
7.5 3
10
2.5
Figure 3-5. Example 3 Display
NOTE:
The precision of the analyzer is ±1% full­scale of range. Analyzer should be cali­brated with a span gas that has a hydrocarbon concentration as close to the fullscale concentration as possible.

3-4 RANGE TRIM OPTION

The Range Trim option allows the user to ad­just each range separately to compensate for differences between calibration gas cylinders from range to range.
See Figure 3-6. The E1, E2 and E3 jumpers allow selection from "fixed" front span potenti­ometer use to "trim" individual potentiometers for each range. In the trim position, the ad­justment range of the individual pot is about 20% when the front panel span pot is set fully clockwise. If the front panel span pot is set fully counterclockwise, the individual pot range is about 5%.
3-6 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
748262-M May 2002
J11
3
SPAN
2
1
ZERO
AGND
DGND
TP9
6
5
4
16
9
+5V
-5V REF
R10 5K
R38 30K
R9 20K
C1 1uF
CR1
FIXED
E1
E2
R1
E3
TRIM
K1
2
10K
R2
10K
R3
10K
R4
10K
R5
10K
R6
10K
R7
10K
2
CR2 CR6
CR3
K1
6,7
12
K2
6,7
12
K3
6,7
12
K4
6,7
12
K5
6,7
12
K6
6,7
12
K7
6,7
12
CR4
K4K2
2
K3
2
CR5
Model 400A
J11
15
14
13
12
8
K5
2
K6
2
CR7
K7
2
7
10
11
11
11
11
11
11
11
11
Figure 3-6. Range Trim Option Schematic
3-8 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
748262-M May 2002

3-5 ROUTINE OPERATION

Model 400A
3-6 RECOMMENDED CALIBRATION FRE-
QUENCY
After calibrating instrument per Section 3-2 (page 3-4), proceed as follows:
Supply sample gas to SAMPLE inlet. Adjust external flow controller or throttle valve so that flow discharged from BY-PASS outlet is be­tween 0.5 and 3.0 liters/minute (preferably between 2.0 and 3.0 liters/minute). Note reading on BY-PASS pressure gauge. It should be the same as that used during ad­justment of the SPAN control; if not, adjust SAMPLE pressure regulator as required.
Turn RANGE switch to appropriate position. Indicator (and recorder, if used) will now automatically and continuously indicate the hydrocarbon content of the sample. Normally, readout is in terms of CH particular hydrocarbon present in the usual span gas. Note that readings obtained during operation depend on the type, as well as the concentration, of hydrocarbons in the sample.
If maximum accuracy and stability are de­sired, observe the operating requirements ex­plained in Section 3-8.
, since this is the
4
After initial startup, or startup following a pro­longed shutdown, the analyzer requires about one day for stabilization. For the first few days thereafter, calibrate daily. Subsequently, the frequency of calibration can be reduced as experience dictates, consistent with the accu­racy requirements of the particular application.

3-7 SHUTDOWN

WARNING
SHUTDOWN PROCEDURE
For safety in shutdown, always turn off fuel gas first, then the air and sample lines.

3-8 OBTAINING MAXIMUM SENSITIVITY

If maximum sensitivity is desired, it is neces­sary to use an optimum combination of set­tings on the SAMPLE, FUEL, and AIR pressure regulators. Settings must be deter­mined experimentally; however, the curves of Figure 3-7, Figure 3-8, and Figure 3-9 on the following pages may be used as a guide.
3-8 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 400A
1.0
Instruction Manual
748262-M
May 2002
1.0
0.8
RESPONSE
(100 ppm CH
4
fullcale)
0.6 SAMPLE: 100 ppm CH FUEL: 25 psig H AIR: 25 psig
0.4
0.2
0
0 2 3 4 5 6 7
SAMPLE PRESSURE
(psig)
4
in N
2
2
Figure 3-7. Typical Curve of Analyzer Response vs. Pressure Setting on Sample Pressure
0.8
AIR: 30 psig
AIR: 30 psig
RESPONSE
(100 ppm CH
4
fullcale)
0.6
AIR: 30 psig
0.4
0.2
0
5
34.4
10
68.8
103.2
SAMPLE PRESSURE
SAMPLE: 100 ppm CH4 in N2 @ 5 psig
15
20
137.6
psig
kPa
25
172.6
30
206.4
Figure 3-8. Typical Curves of Analyzer Response vs. Pressure Setting on Fuel Pressure Regulator
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-9
Instruction Manual
748262-M May 2002
Model 400A
1.0
15
103.2
psig kPa
2
2
2
20
137.6
25
172.630206.4
RESPONSE
(100 ppm CH4 fullcale)
0.8
0.6
0.4
0.2
FUEL: 30 psig H
FUEL: 25 ppm H
FUEL: 20 ppm H
SAMPLE: 100 ppm CH4 in N2 @ 5 psig
0
5
34.4
10
68.8
AIR PRESSURE
Figure 3-9. Typical Curves of Analyzer Response vs. Pressure Setting on Air Pressure Regulator
3-10 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 400A
Instruction Manual
748262-M
May 2002
SECTION 4

THEORY

4-1 PRINCIPLES OF OPERATION

The Model 400A Hydrocarbon Analyzer util­izes the flame ionization method of detection. The sensor is a burner in which a regulated flow of sample gas passes through a flame sustained by regulated flows of air and a fuel gas (hydrogen or a hydrogen/diluent mixture). Within the flame, the hydrocarbon compo­nents of the sample stream undergo a com­plex ionization that produces electrons and positive ions. Polarized electrodes collect these ions, causing current to flow through electronic measuring circuitry. Current flow is proportional to the rate at which carbon atoms enter the burner.

4-2 BURNER

Principle components of the burner are the manifold, burner jet, and the collector.
Streams of sample, fuel and air delivered by the analyzer flow system (Section 4-3) are routed through internal passages in the
manifold and into the interior of the burner. Here the sample and fuel pass through the burner jet and into the flame; the air stream flows around the periphery of the flame.
The burner jet and the collector function as electrodes. The jet is connected to the posi­tive terminal of the 90 VDC polarizing voltage. The collector is connected to the signal ampli­fier. The two polarized electrodes establish an electrostatic field in the vicinity of the flame. The field causes the charged particles formed during combustion to migrate. Elec­trons go to the burner jet; positive ions go to the collector. Thus a small ionization current flows between the two electrodes. Magnitude of the current depends on the concentration of carbon atoms in the sample. The burner cur­rent serves as the input signal to the elec­tronic measuring circuitry (Section 4-4).
Mounted on the burner are (1) igniter, driven by flame ignition circuit and (2) thermistor sensor for flame status indicator circuit.
Signal Conditioning
Negative Ion Collection Ring
+90V
Sample
Positive Carbon Ions
Air
Fuel
Figure 4-1. Flame Ionization Detection Theory
Rosemount Analytical Inc. A Division of Emerson Process Management Theory 4-1
Instruction Manual
748262-M May 2002
Model 400A

a. Response to Different Hydrocarbons

Both speed and magnitude of analyzer response are affected by the type of hy­drocarbon in the sample. Typical curves of response versus time for various hy­drocarbons are given in Figure 3-2, page 3-3.
Magnitude of the analyzer response to an atom of carbon depends on the chemical environment of this atom in the molecule. The characteristic response of a given type of atom may be expressed approxi- mately by a value designated the "effec­tive carbon number." The effective carbon number of a particular type of car­bon atom is defined as the ratio between the instrument response caused by an atom of this type and the instrument re­sponse caused by an aliphatic carbon atom. Table 4-1 (page 4-2) lists approxi-
mate effective carbon numbers for several types of carbon atoms. Although the in­strument does not respond directly to at­oms other than carbon, in some compounds certain other atoms do change instrument sensitivity to carbon. For this reason, values are listed for a few non-carbon atoms. Values in the table were determined experimentally, on a single analyzer. Because of slight varia­tions in characteristics of individual ana­lyzers, these values should be regarded as approximations only.
To determine the effective carbon number of a molecule of a given organic com­pound, algebraically add the individual values for the constituent atoms. Exam­ples of effective carbon numbers of mole­cules are: Butane (C
8H18
(C
1.4.
), 8; and ethyl alcohol (C2H5OH),
4H10
), 4; octane
TYPE OF ATOM OCCURRENCE EFFECTIVE CARBON NUMBER
Carbon In Aliphatic Compound +1.00 Carbon In Aromatic Compound +1.00 Carbon In Olefinic Compound +0.95 Carbon In Acetylenic Compound +1.30 Carbon In Carbonyl Radical 0.00 Carbon In Nitrile +0.30 Carbon In Ether -1.00 Carbon In Primary Alcohol -0.60 Carbon In Secondary Alcohol -0.75 Carbon In Tertiary Alcohol, Ester -0.25
Chlorine
Chlorine On Olefinic Carbon Atom +0.05
Nitrogen In Amine
Effective Carbon
numbers =
Instrument response caused by aliphatic carbon atom
As two or more chlorine atoms on single aliphatic carbon atom
Instrument response caused by atom of given type
Table 4-1. Approximate Effective Carbon Numbers
-0.12
Value similar to that for oxygen atom in corresponding alcohol
4-2 Theory Rosemount Analytical Inc. A Division of Emerson Process Management
Model 400A
Instruction Manual
748262-M
May 2002

4-3 ANALYZER FLOW SYSTEM

The internal flow system of the analyzer is shown in drawing 622883. Its basic function is to deliver regulated flows of sample, fuel, and air to the burner. In addition, the system routes the burner exhaust gas and sample bypass flow out of the analyzer through the corresponding outlet ports.
Suitable pressurized gases are supplied to the SAMPLE, FUEL, and AIR inlet ports. Each inlet fitting contains an internal filter.
Each of the three gas streams is routed to the burner via a flow control arrangement con­sisting of the following elements:
1. An adjustable pressure regulator. The AIR and FUEL pressure regulators pro­vide controlled pressure on the down­stream side. The SAMPLE pressure regulator is a back-pressure regulator that provides controlled pressure on the up­stream side, and discharges excess sam­ple through the BYPASS outlet. (Bypass feature provides high velocity sample flow through analyzer, to minimize system re­sponse time.)
2. A flow limiting element, selected to pass the appropriate gas flow when the pres­sure drop is adjusted to the correct value. The air and fuel streams utilize porous, sintered metallic restrictor elements mounted in fittings. The sample stream uses a calibrated capillary tube.
3. A gauge that indicates pressure at the in­let end of the corresponding restrictor or capillary. Ranges are SAMPLE pressure gauge, 0 to 5 psig (0 to 35 kPa); AIR and FUEL pressure gauges, 0 to 30 psig (0 TO 207 kPa).
further amplified by a post amplifier before being converted to a digital display suitable for direct data presentation. To cover the re­quired dynamic range, the amplifier is pro­vided with two gain ranges that differ by a factor of 100.
Output voltage from the preamp is a precise function of ionization current. In equation form:
e
out = -i
i
x R
f
Where i i = ionization current
R
= feed back resistance
f
The most sensitive gain range includes a trim adjustment so that inter-range correlation can be obtained over the entire signal span.
A buffer signal offering unity gain and noise filtration provide a low output impedance to drive the signal cable and post amplifier cir­cuits on the main circuit board. Selection of the low or high range feedback resistors is made by relay K1 on the preamplifier board. Refer to the preamplifier board schematic in the rear of this manual.
A variable offset current is injected into the summing node of the electrometer amplifier to compensate for background offset current. These currents influence the measurement procedure, and a variable voltage at the front panel allows the user to visually cancel these currents during the calibration procedure. Background current is due to unavoidable traces of carbonaceous material introduced into the burner flame by the fuel gas and air.

4-4 PREAMPLIFIER BOARD

The ionization current generated by the burner is measured by an electrometer pre­amplifier located adjacent to the burner as­sembly. This small current is amplified and transformed into a signal voltage that is then
Rosemount Analytical Inc. A Division of Emerson Process Management Theory 4-3
Instruction Manual
748262-M May 2002
Model 400A

4-5 MAIN ELECTRONICS BOARD

a. Post Amplifier

The post amplifier circuit comprises two amplifiers with various combinations of feedback resistances that are logically selected by the front panel RANGE switch or external inputs. The range selection process provides a total of seven possible gains corresponding to the input sensitivi­ties in Table 4-2, page 4-4.

b. Digital Display

The voltage originating from the signal amplifiers is made available to output terminals and directly to an analog to digital (A-D) converter for data display in digital form. The A-D converter is termed 3 1/4 digits, which implies a maximum reading of 1999. The one bit signifying overrange allows 100% excursion above the normal 99.9% range. Thus the user, depending upon span and zero offset conditions, may select the presentation to read directly in percent of fullscale and still have 100% overrange capability re­maining.

c. Span

To compensate for various calibration gases, provision is made to vary the gain with a variable span control. The variable span, accessed on the front panel through a calibrated potentiometer, allows gain variation of 400%.
Table 4-3 (page 4-4) gives the fullscale sensitivity of the current measuring cir­cuitry for the various settings on the RANGE switch and SPAN control.

d. Range Trim (Option)

The Range Trim will allow individual ad­justment on each range to compensate for different gas standards between ranges.

e. Remote Range Control

Internally all of the range terminals are pulled up to the system 5V power supply through 10K ohm resistors and are therefore suitable for connection to any TTL or equivalent logic circuit. See Sec­tion 2-8e, page 2-5.
RANGE
RANGE SWITCH SET-
TING
PREAMP GAIN POST AMP GAIN
11 HIGH 50
22.5 HIGH 20 310 HIGH 5 425 HIGH 2 5 100 LOW 50 6 250 LOW 20 7 1000 LOW 5 8RMT - REMOTE
Table 4-2. Input Sensitivities
RANGE SWITCH
FULLSCALE SENSITIVITY AMPS
SETTING MAX SPAN MIN SPAN
15 x 10 22 x 10 35 x 10 42 x 10 55 x 10 62 x 10
-12
-12
-11
-11
-10
-12
6.66 x 10
1.66 x 10
6.66 x 10
1.66 x 10
6.66 x 10
1.66 x 0
-12
-11
-11
-10
-10
-9
Table 4-3. Fullscale Sensitivity
4-4 Theory Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
748262-M May 2002
Model 400A

4-6 HEATER/FAN TEMPERATURE ASSEMBLY

The temperature controller is located in the heated compartment of the instrument. The RTD temperature sensor, in conjunction with a control circuit, maintains the internal tem­perature to meet performance specifications for the instrument. The temperature is con­trolled at 122°F (50°C). Refer to schematic drawing 624003 in the rear of this manual.
This temperature is maintained at a constant level to minimize temperature dependent variations in amplifier sensitivity and to pre­vent changes in absorption/desorption equilib­rium of trace hydrocarbons in the internal flow system. A blower fan runs continuously to circulate air and equalize the temperature throughout the analyzer.
The temperature control board is a multi­purpose board with a control resistor that is factory-selected for the model analyzer in which it is used. Schematic 624003 shows the correct jumper position for the Model 400A
NOTE:
230 VAC operation requires an accessory transformer mounted external to the in­strument. If external 230 VAC power is provided, do not change the switch setting on the temperature control board, which is factory-selected for 115 VAC.
A 230 VAC switch setting will cause the case temperature controller to malfunc­tion.

4-7 IGNITION CIRCUIT

The ignition circuit is used to light the burner flame during instrument startup. The principle circuit components are as follows:
IGNITE switch mounted on the preamp board and connected in series with the primary of the transformer

4-8 SYSTEM POWER SUPPLIES

Several power supplies are used within the Model 400A for the electrical functions. These are as follows:
±12 VDC
Used for most analog functions.
±5 VDC
These two voltages derived from a precision reference are used for zero offset and bias requirements.
+5 VDC
Used to power the digital circuit range decod­ing A-D display, etc.
+93 VDC
Used for the burner tip polarizing voltage.
+24 VDC
(unregulated) - Used for relays.
All of the power supplies are electronically regulated with the exception of the +24 VDC supply. Shorting of one supply to another may cause damage.

4-9 FLAME-OUT BOARD

Refer to schematic 657546 in the rear of this manual. Loss of the flame shuts down the fuel solenoid. If there is a flammable compo­nent in the sample, a fail safe solenoid acces­sory is available for wiring in parallel with the fuel solenoid.
Igniter element mounted in the burner cap
Step-down transformer (PN 621049)
4-6 Theory Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
748262-M May 2002
Model 400A
4-6 Theory Rosemount Analytical Inc. A Division of Emerson Process Management
Model 400A
Instruction Manual
748262-M
May 2002
SECTION 5

SERVICE AND TROUBLESHOOTING

WARNING
ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified per­sonnel.
For safety and proper performance this in­strument must be connected to a properly grounded three-wire source of power.
The power plug must be disconnected from the rear of the instrument before re­moving any of the boards and/or intercon­nect plugs.

5-1 SYSTEM CHECKOUT

If analyzer performance is unsatisfactory, make the following tests in the sequence given.

a. Amplifier Zero Adjustment

Place POWER switch at ON; RANGE switch at 1000; ZERO and SPAN potenti­ometers at 1000. With flame extin­guished, note reading on front panel indicator, or on potentiometric recorder if used. Reading should be zero; if not, ad­just trimming potentiometer R13 on am­plifier circuit board. Potentiometer R13 is adjusted by inserting a screwdriver through the lower hole in the amplifier shield.

b. Zero Current Adjustment

If a current recorder is used, verify that it reads zero. If not, adjust trimming potenti­ometer R1 on the current output board.

c. Electrical Leakage Check

Turn RANGE switch from position 1000 to position 1; indicator or recorder should still read 0 ±5%. If reading is outside this range, leakage is excessive. To deter­mine the source of the leakage, discon­nect the amplifier input cable from burner, and note response of indicator or re­corder. If abnormal reading persists, leak­age is in either the cable or electronics. If reading drops to zero, leakage is in the burner. Clean burner per Section 5-2a, page 5-2. Reconnect cable to burner.

d. Flame Ignition

Start up analyzer per Section Three, and attempt to ignite flame. If flame will not ig­nite refer to Table 5-1, page 5-5, Item 2.

e. Noise Check

With flame burning, observe indicator or recorder. If noise is greater than 2% of fullscale, refer to Table 5-1, page 5-5, Item 3.

f. Overall Sensitivity Check

With flame burning, supply a suitable span gas to SAMPLE inlet. Turn RANGE switch to a setting appropriate to the hy­drocarbon content of the particular span gas. Adjust SPAN control for reading of 100% on indicator or recorder. If the de­sired upscale reading is unobtainable by adjustment of the SPAN control, the fault may be in either the flow system (Table 5-1, page 5-5, Item 4) or in the electron­ics.
Rosemount Analytical Inc. A Division of Emerson Process Management Service and Troubleshooting 5-1
Instruction Manual
748262-M May 2002
BACKGROUND
SIGNAL
ppm CH
4
Model 400A
Note: Fuel 100% hydrocarbon. Fuel, air and zero gas cylinders at constant temperature.
0
-17.610-12.220-6.6
30
-1.1404.4
Figure 5-1. Effect of Analyzer Temperature on Background Signal – Typical Curve

g. Stability Check

Supply zero gas to SAMPLE inlet. Turn RANGE switch to Position 1. Observe in­dicator or recorder over several hours of operation. Drift greater than the stability specification may be due to malfunction­ing of the internal temperature control. For comparison, Figure 5-1 above shows the typical effect of ambient temperature variations on background reading for a standard analyzer, with internal tempera­ture controller functioning normally. In this example, temperature dependent varia­tions in background signal are very small. However, note that such high stability is obtainable only with exceedingly clean cylinder gases. Background signal, and temperature dependent variations in this signal, increase with level of trace hydro­carbon contaminants.
50 10
ANALYZER AMBIENT TEMPERATURE
60
15.57021.18025.69032.2

5-2 SERVICING FLOW SYSTEM AND BURNER

In preventive maintenance of the flow system, the most important precautions are (1) provi­sion for continuous removal of all combustion products, including water vapor, and (2) the use of great care to prevent the contamination of any component with hydrocarbons, even in trace amounts.
COMPONENT CONTAMINATION
Never touch burner tip, Teflon skirt, or combustion chamber with bare hands; al­ways use clean gloves or cloth. If this precaution is not observed, oil from skin will contaminate these components.

a. Burner Disassembly and Cleaning

100
37.8
CAUTION
110
43.3
120
48.9
F
°
C
°
If the internal temperature controller is functioning normally, apparent drift may be due to changes in ambient tempera­ture of the fuel and air cylinders. For fur­ther information, refer to Section 3-8 Obtaining Maximum Sensitivity (page 3-
8).
Disassemble the burner only if contami­nants must be removed. Combustion products or other contaminants which ac­cumulate inside the burner may form electrical leakage paths between the col­lector and the burner contact, resulting in noisy readings. If the instrument is to be operated at the highest sensitivity, traces
5-2 Service and Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management
Model 400A
Instruction Manual
748262-M
May 2002
of such contaminants can cause errone­ous readings. For best performance, it is necessary that the burner be kept free of any contamination.
WARNING
POSSIBLE EXPLOSION HAZARD
This analyzer uses a fuel containing hy­drogen. Leakage from the fuel contain­ment system can result in an explosion. The fuel supply and containment system, both inside and outside the analyzer, should be carefully checked for leaks upon installation, before initial startup, during routine maintenance and any time the in­tegrity of the containment system is bro­ken.
When burner requires cleaning, refer to Figure 4-1 (page 4-1) and DWG 623190 and proceed as follows:
1. Place POWER switch at OFF position and disconnect power cord.
7. Unscrew and remove burner tip as­sembly.
NOTE:
All items used for cleaning (tweezers, swabs, etc.) must be absolutely free of contamination.
8. Clean chimney assembly, combustion chamber, and burner tip assembly with alcohol, followed by a distilled water wash.
Using care not to touch internal parts, re­assemble burners as explained in the fol­lowing steps.
1. Holding burner tip assembly with clean tissue, screw finger-tight into manifold.
2. Push combustion chamber down onto manifold, taking care not to hit burner tip. Tighten clamp on combustion chamber.
2. Shut off fuel gas first, then air and sample gases.
3. Unscrew burner cap retainer ring and remove burner cap.
4. On combustion chamber, disconnect polarizing voltage cable and amplifier input cable.
5. Lift chimney from combustion cham­ber. Leave exhaust tubing connected to chimney unless old chimney is to be replaced.
6. Loosen clamp. Lifting straight up, re­move combustion chamber from manifold.
NOTE:
If old burner tip assembly is to be used again, do not touch it with bare hands or any materials likely to contaminate it with hydrocarbons, salt, etc.
3. Replace chimney on combustion chamber.
4. Replace burner cap.
5. Reconnect all leads.

b. Thermistor

The thermistor sensor for the FLAME OUT indicator circuit is mounted in the burner. See Figure 4-1 (page 4-1). Ther­mistor resistance should be approximately 100K ohms at 77 method requires that the comparator input signal be measured at the junction of R32 and R33. When the flame is burning nor­mally, the voltage at this point will be 0.1 VDC to 0.3 VDC, indicating that the ther­mistor circuit is functional and the flame temperature is correct.

c. Fuel and Air Restrictors

Fuel and air restrictors are porous, sin­tered metallic, restrictor elements
o
F (25oC). An alternate
Rosemount Analytical Inc. A Division of Emerson Process Management Service and Troubleshooting 5-3
Instruction Manual
748262-M May 2002
Model 400A
mounted within fittings. If a restrictor be­comes plugged, replace it. Do not attempt to clean restrictors with solvents. See DWG 622883.

d. Sample Capillary

The sample capillary is equipped with fit­tings, permitting convenient removal and replacement. If necessary, the capillary may be cleaned with acetone or methyl
ethyl ketone, followed by distilled water wash. Refer to warnings in Section 5-2a, page 5-2.

5-3 TROUBLESHOOTING

For troubleshooting the electronic system, refer to the appropriate schematic at the rear of this manual. For troubleshooting the burner system, refer to Table 5-1, page 5-5.
5-4 Service and Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management
Model 400A
SYMPTOM PROBABLE CAUSE POSSIBLE REMEDY
Indicator shows upscale reading when flame is out.
Electrical leakage in burner assembly. Clean burner per Section 5-2a, page 5-2.
Fuel gas emerging from burner jet di­luted with other gases because fuel system is insufficiently purged.
Purge FUEL pressure regulator by allowing gas to flow for several minutes, while turning regulator alternately clockwise and counterclockwise.
Instruction Manual
748262-M
May 2002
Burner will not ig­nite.
Indicator reading noisy
OSS OF SENSITIVITY
L
Sample pressure fluctuations
False flame (FLAME-ON) indi­cation of span gas produces little or no upscale deflection.
AIR and/or FUEL pressure regulator improperly adjusted.
No flow, or reduced flow, of fuel and/or air into burner combustion chamber.
Malfunction in igniter circuit.
Ignite leads improperly connected. Check igniter plugs for proper contact.
Igniter burned out.
Transformer circuit open. Check voltage. See drawings 620424 and 655354 IGNITE switch defective. Test switch for continuity. Replace if necessary. Contamination of flow system: fuel and air supplies, external pressure regula­tors, connecting tubing. Pressure regulator(s) and/or pressure gauge(s) clogged. Water or condensate in burner or ex­haust line.
Fuel and/or air filter clogged. Check filters; replace if necessary.
Plugged restrictor or capillary
Electrical elements of burner partially short circuited by combustion products Check valve in sample pump not func­tioning. Obstruction in bypass outlet. Examine bypass outlet, remove obstruction.
Flame may be lifting above burner tip (unusual condition).
Check readings on AIR and FUEL pressure gauges. Adjust AIR pressure regulator to increase or decrease air pressure slightly. Constriction in fuel and/or air passage in burner jet, restrictor, etc. Find cause of constriction and re­move. Remove cap from burner. Actuate IGNITE switch. Igniter should glow red. If not, check the following probable causes.
Replace igniter (glow plug) as shown in drawing
623190.
Replace fuel and/or air supply; clean or replace tubing and regulators per Sections 2-5, page 2-2.
Clean or replace regulators and gauges.
Clean burner and exhaust line
Verify that fuel and air restrictors and sample cap­illary are open. An abnormally low background signal, together with sensitivity loss, indicates plugged restrictor. Note that flame will not light unless fuel and air restrictors are open, but will not light even though sample capillary is completely closed. Disassembly and clean burner, as explained in Section 5-2a, page 5-2.
Examine check valve.
Connect a voltmeter from ground, as viewed with door open. If flame is properly lit, the voltage will be below 0.2 VDC. If voltage is high, readjust air­fuel ratio to obtain proper ignition. Re-light burner with richer fuel and air setting. FUEL: 30 psig AIR: 2 to 5 psig.
Table 5-1. Troubleshooting
Rosemount Analytical Inc. A Division of Emerson Process Management Service and Troubleshooting 5-5
Instruction Manual
748262-M May 2002
Model 400A
5-6 Service and Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management
Model 400A
WARNING
Instruction Manual
748262-M
May 2002
SECTION 6

REPLACEMENT PARTS

PARTS INTEGRITY
Tampering with or unauthorized substitution of components may adversely affect safety of this product. Use only factory-approved components for repair.

6-1 CIRCUIT BOARD REPLACEMENT POLICY

In most situations involving a malfunction of a circuit board, it is more practical to replace the board than to attempt isolation and replace­ment of the individual component. The cost of test and replacement will exceed the cost of a rebuilt assembly. As standard policy, rebuilt boards are available on an exchange basis.
Because of the exchange policy covering cir­cuit boards the following list does not include individual electronic components. If circum­stances necessitate replacement of an indi­vidual component which can be identified by inspection or from the schematic diagrams, obtain the replacement component from a lo­cal source of supply.
6-2 RECOMMENDED REPLACEMENT PARTS
LIST
655208 Board, Temperature Control 815187 Regulator, Sample 620423 Board , Preamp 888692 Regulator, Fuel/Air 655305 Heater in Fan Assembly 861984 Gauge, Fuel/Air 836482 Fan, Heater 617900 Glow Plug (Igniter) 621031 Cable, 16 Conductor 644055 Gauge, Sample 748262 Instruction Manual, 400A
(S/N 2000001 and up) 823484 Fuse, 2 AMP (115 VAC) 617902 Valve, Solenoid (Fuel Shutoff) 657545 Board, Flame-Out 620428 Board, Amplifier 621023 Board, Isolated Current Output, 4 to
20 mA 621017 Header 655178 Door, Clear Plastic 655203 Valve, Solenoid (Sample Shutoff)
(Option)
017154 Filter, Sample/Air
See drawing 623190 for listing of capillaries and restrictors, and for Burner Parts List.
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 6-1
Instruction Manual
748262-M May 2002

6-3 MATRIX

Each analyzer is configured per the customer sales order. Below is the CLD sales matrix which lists the various configurations available.
400A MODEL 400A HYDROCARBON ANALYZER
Code Seven Ranges with Remote Range Control, Range Multiplier: X1, X2.5, X10, X25, X100, X250, X1000
01 Seven ranges adjustable between 0-4 ppm through 0-1% 03 Seven ranges adjustable between 0-100 ppm through 0-10% 04 Seven ranges adjustable between 0-1 ppm through 0-2500 ppm (includes 100% H2 capillary) 99 Special
Code Output
01 Selectable, 0-100 mV, 0-1V, 0-5 VDC 02 Current, 4-20 mA Isolated 99 Special
Model 400A
To identify the configuration of an analyzer, locate the analyzer name-rating plate. The sales matrix identifier number appears on the analyzer name-rating plate.
Code Operation
01 117V, 50/60 Hz 99 Special
Code Sample Line
01 Standard (Brass/Copper) 02 Standard with Sample Shutoff Valve 03 Stainless Steel 04 Stainless Steel with Sample Shutoff Valve 99 Special
Code Case
01 Standard
02 Standard w/Tropicalization 99 Special
Code Range Control/ID Options
01 Features as selected above 02 Seven isolated remote range control, range trim adjustment, range ID 99 Special
Code Off Range Alarm
01 Features as selected above 02 Single contact 99 Special
400A 04 01 01 03 02 02 01 Example
6-2 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA 2000 TO2

RETURN OF MATERIAL

Instruction Manual
748262-M
May 2002
SECTION 7

7-1 RETURN OF MATERIAL

If factory repair of defective equipment is required, proceed as follows:
1. Secure a return authorization from a Rosemount Analytical Inc. Sales Office or Representative before returning the equipment. Equipment must be returned with complete identification in accordance with Rosemount instructions or it will not be accepted.
Rosemount CSC will provide the shipping address for your instrument.
In no event will Rosemount be responsible for equipment returned without proper authorization and identification.
2. Carefully pack the defective unit in a sturdy box with sufficient shock absorbing material to ensure no additional damage occurs during shipping.
3. In a cover letter, describe completely:
The symptoms that determined the
equipment is faulty.
The environment in which the
equipment was operating (housing, weather, vibration, dust, etc.).
Site from where the equipment was
removed.
Whether warranty or non-warranty
service is expected.
Complete shipping instructions for the
return of the equipment.
4. Enclose a cover letter and purchase order and ship the defective equipment according to instructions provided in the Rosemount Return Authorization, prepaid, to the address provided by Rosemount CSC.
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
If warranty service is expected, the defective unit will be carefully inspected and tested at the factory. If the failure was due to the conditions listed in the standard Rosemount warranty, the defective unit will be repaired or replaced at Rosemount’s option, and an operating unit will be returned to the customer in accordance with the shipping instructions furnished in the cover letter.
For equipment no longer under warranty, the equipment will be repaired at the factory and returned as directed by the purchase order and shipping instructions.

7-2 CUSTOMER SERVICE

For order administration, replacement Parts, application assistance, on-site or factory repair, service or maintenance contract information, contact:
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076

7-3 TRAINING

A comprehensive Factory Training Program of operator and service classes is available. For a copy of the Current Operator and Service Training Schedule contact the Technical Services Department at:
Rosemount Analytical Inc.
Customer Service Center
1-800-433-6076
Rosemount Analytical Inc. A Division of Emerson Process Management Return of Material 7-1
Instruction Manual
748262-M May 2002
Model 400A
7-2 Return of Material Rosemount Analytical Inc. A Division of Emerson Process Management

WARRANTY

Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in workmanship and material under normal use and service for a period of twelve (12) months from the date of shipment by Seller. Consumables, glass electrodes, membranes, liquid junctions, electrolyte, o-rings, etc., are warranted to be free from defects in workmanship and material under normal use and service for a period of ninety (90) days from date of shipment by Seller. Goods, part(s) and consumables proven by Seller to be defective in workmanship and/or material shall be replaced or repaired, free of charge, F.O.B. Seller's factory provided that the goods, part(s) or consumables are returned to Seller's designated factory, transportation charges prepaid, within the twelve (12) month period of warranty in the case of goods and part(s), and in the case of consumables, within the ninety (90) day period of warranty. This warranty shall be in effect for replacement or repaired goods, part(s) and the remaining portion of the ninety (90) day warranty in the case of consumables. A defect in goods, part(s) and consumables of the commercial unit shall not operate to condemn such commercial unit when such goods, part(s) and consumables are capable of being renewed, repaired or replaced.
The Seller shall not be liable to the Buyer, or to any other person, for the loss or damage directly or indirectly, arising from the use of the equipment or goods, from breach of any warranty, or from any other cause. All other warranties, expressed or implied are hereby excluded.
IN CONSIDERATION OF THE HEREIN STATED PURCHASE PRICE OF THE GOODS, SELLER GRANTS ONLY THE ABOVE STATED EXPRESS WARRANTY. NO OTHER WARRANTIES ARE GRANTED INCLUDING, BUT NOT LIMITED TO, EXPRESS AND IMPLIED WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Limitations of Remedy. SELLER SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY DELAY IN PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF WARRANTY SHALL BE LIMITED TO REPAIR OR REPLACEMENT UNDER THE STANDARD WARRANTY CLAUSE. IN NO CASE, REGARDLESS OF THE FORM OF THE CAUSE OF ACTION, SHALL SELLER'S LIABILITY EXCEED THE PRICE TO BUYER OF THE SPECIFIC GOODS MANUFACTURED BY SELLER GIVING RISE TO THE CAUSE OF ACTION. BUYER AGREES THAT IN NO EVENT SHALL SELLER'S LIABILITY EXTEND TO INCLUDE INCIDENTAL OR CONSEQUENTIAL DAMAGES. CONSEQUENTIAL DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO, LOSS OF ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVENUE, COST OF CAPITAL AND DAMAGE OR LOSS OF OTHER PROPERTY OR EQUIPMENT. IN NO EVENT SHALL SELLER BE OBLIGATED TO INDEMNIFY BUYER IN ANY MANNER NOR SHALL SELLER BE LIABLE FOR PROPERTY DAMAGE AND/OR THIRD PARTY CLAIMS COVERED BY UMBRELLA INSURANCE AND/OR INDEMNITY COVERAGE PROVIDED TO BUYER, ITS ASSIGNS, AND EACH SUCCESSOR INTEREST TO THE GOODS PROVIDED HEREUNDER.
Force Majeure. Seller shall not be liable for failure to perform due to labor strikes or acts beyond Seller's direct control.
Instruction Manual
748262-M May 2002
Model 400A
Emerson Process Management
Rosemount Analytical Inc. Process Analytic Division
1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 E gas.csc@emersonprocess.com
ASIA - PACIFIC Fisher-Rosemount Singapore Private Ltd.
1 Pandan Crescent Singapore 128461 Republic of Singapore T 65-777-8211 F 65-777-0947
http://www.processanalytic.com
© Rosemount Analytical Inc. 2001
Fisher-Rosemount GmbH & Co.
Industriestrasse 1 63594 Hasselroth Germany T 49-6055-884 0 F 49-6055-884209
EUROPE, MIDDLE EAST, AFRICA Fisher-Rosemount Ltd.
Heath Place Bognor Regis West Sussex PO22 9SH England T 44-1243-863121 F 44-1243-845354
LATIN AMERICA Fisher - Rosemount
Av. das Americas 3333 sala 1004 Rio de Janeiro, RJ Brazil 22631-003 T 55-21-2431-1882
Loading...