Rosemount 4000 Percent Oxygen Sensor Manuals & Guides

• STABLE, RELIABLE amperometric Oxygen Sensor
• LONG LIFE, LOW MAINTENANCE rechargeable sensor
• ROBUST DESIGN for harsh applications
*
• RAPID CALIBRATION using ambient air
• COMPATIBLE WITH 5081-A, 1056 AND 1055 INSTRUMENTS
APPLICATIONS
The Model 4000 sensor is used to monitor percent oxygen concentrations in gaseous streams. The most common use is to monitor oxygen headspace concen­tration for nitrogen blanketing applications. The sen­sors are also used in other applications where oxygen levels are controlled and monitored.
FEATURES
The Model 4000 Series Percent Oxygen sensor is a membrane-covered amperometric sensor. The mem­brane consists of a gas permeable Teflon
®
1
mem­brane stretched tightly over a gold cathode. A silver anode and an electrolyte solution complete the inter­nal circuit. The sensor body is constructed of Ryton
®
2
, which can withstand exposure to hydrocar-
bons and other corrosive chemicals.
Gas permeates through the membrane, and the oxy­gen in the sample is reduced at the cathode. A volt­age is applied across the cathode and anode, generating an electrical current that is directly propor­tional to the oxygen concentration in the sample. Since the rate of oxygen diffusion through the mem­brane is temperature dependent, the sensor response must be corrected for permeability caused by temper­ature. A Pt 100 RTD in the sensor accurately meas­ures temperature, and the analyzer automatically performs the correction.
The sensor is easy to maintain. For calibration simply expose the sensor to ambient air and press the air calibration button. The analyzer measures the baro­metric pressure using an on-board pressure sensor
Model 4000
Percent Oxygen Sensor
Instruction Sheet
PN 51A-4000/rev.C January 2009
and calculates the equilibrium solubility of atmospher­ic oxygen at the prevailing temperature and pressure. Replacing the membrane requires no special tools or fixtures. To replenish the electrolyte solution, unscrew the fill plug, add the reagent, and replace the plug.
TYPES OF SENSORS:
Rechargeable with a fast response flow assembly —
allows minimum volume gas flow that permits mount­ing sensor in a flowing gas stream. Sample is supplied at slightly above atmospheric pressure, flows through the assembly, and discharges to atmospheric pres­sure. Internal volume is low to minimize sensor response time. Refer to Figure 1 for mounting instruc­tions.
Rechargeable In-line flow — In line pressure com­pensated flow assembly permits mounting the sensor in a variable pressure gas stream at pressures up to 50 psig. This may or may not include a gland on the sensor body. Refer to Figures 2, 3, and 4 for mounting instructions.
*
RYTON is resistant to 30% sulfuric acid, 85% phosphoric acid, 30% sodium hydroxide, gasoline, aliphatic alcohols, esters, ethers, and ketones, as well as to aromatic amines. It is not particularly suited for service in strong oxidizing agents, aliphatic amines, chlori­nated hydrocarbons, or aromatic nitrites, aldehydes, and nitro com­pounds.
1
Teflon is a registered trademark of E.I. du Pont de Nemours & Co.
2
Ryton is a registered trademark of Chevron Phillips Chemical Company LP.
SPECIFICATIONS
Range: 0-25% Oxygen Linearity: For constant sample temperature after correction for sensor zero offset: ±1% of full scale Repeatability: ± 0.1% of range Stability:
Zero drift ± 0.25% O2per week @ 25°C; Span drift ± 0.25% O2per week @ 25°C
Response Time: 90% in 20 seconds for a step change, using an equilibrated sensor at 25°C Sample Pressure: 0 to 50 PSIG Sample Temperature: 32 to 110°F Humidity: up to 95% non-condensing Wetted materials: Ryton
®
1
, Teflon
®
2
1
Ryton is a registered trademark of Chevron Phillips Chemical Company LP.
2
Teflon is a registered trademark of E.I. du Pont de Nemours & Co.
FIGURE 1. Rechargeable Sensor with Fast Response Flow Assembly (6A00184G01)
MODEL 4000 SPECIFICATIONS
2
FIGURE 2. Vertical Installation (Preferred)
FIGURE 3. Horizontal Installation
3
MODEL 4000 INSTALLATION
INSTRUMENT SET UP FOR % OXYGEN MEASUREMENT
For the 1055, access menu by selecting program, then measurement, configure Sensor 1. For manufacturer select Rosemount, for application select Other, for S1 Units select % sat.
For the 5081A, access the menu by using Program Key on the infrared remote. Select Display, Type=O2, Unit=%, Sensor=AdO.
For the 1056, access the menu by selecting program, then measurement, choose sensor 1 or 2 measurement = oxygen. In the “SN configure” screen, under “type” select “% 02 in gas” for “units” select “% oxygen in gas”.
FIGURE 6. Wiring to Model 5081 FIGURE 7. Wiring to Model 1056
FIGURE 4. Wiring to Model 1055-01-10-25
FIGURE 5. Wiring to Model 1055-01-11-25
MODEL 4000 WIRING
4
5
SYMPTOM PROBABLE CAUSE CORRECTIVE ACTION
Abnormally high oxygen readings 1. Hole in sensor membrane 1. Replace membrane (unable to calibrate) 2. Gold cathode loose 2. Replace sensor
3. Open RTD 3. Replace sensor
Abnormally low oxygen readings 1. High internal cell resistance 1. Replace sensor (unable to calibrate) 2. Membrane loose 2. Tighten cap / replace membrane
3. Contaminated electrolyte 3. Clean / recharge sensor
4. RTD shorted 4. Replace sensor
Sensor noisy (motion sensitive) 1. Membrane loose 1. Replace membrane
2. Low electrolyte level 2. Fill properly
3. Cathode contaminated 3. Replace sensor
Upscale reading with known Gold cathode loose Replace sensor oxygen-free sample
CALIBRATION
Air calibration is recommended. Using a certified span gas is an option, but since the concentration of oxygen in ambient air is close to 21% at sea level, this is the best solution. Refer to the instrument instruction manual for details on how to access the calibration menu.
Note: An “inprocess” calibration is required. Input 21% O2as the process concentration. For model 1056, air calibration can be performed.
TROUBLESHOOTING
The most frequent fault is a progressive development of insensitivity of the sensor. If sensor calibration is slug­gish, then the sensor should be recharged with new electrolyte and the membrane may need to be replaced. Listed below are some common troubleshooting techniques
MODEL 4000 MAINTENANCE
PROCEDURE TO RECHARGE THE SENSOR:
Refer to Figure 8 for an exploded view of the sensor.
1. Unscrew the knurled cap from the sensor body. Remove the membrane assembly. Empty all electrolyte from the sensor. Flush the sensor with distilled or deionized water to remove all particulate.
2. Place a piece of adhesive tape over the breather hole in the pressure compensation diaphragm port (not the slotted fill plug).
3. Examine the cathode for staining or uneven coloration, which indicates the cathode should be rejuvenated. Also inspect the grooves that surround the cathode for any deposited material, which will typically be white to gray in color. Most of these deposits are water soluble and can be removed via a water jet from a squeeze bottle. Any insoluble deposits may be removed with a toothpick, but care must be used to avoid deforming the grooves.
4. Disassemble the membrane assembly. Remove the retainer from the holder by placing your finger into the center hold of the holder and pressing your fingernail against the inner edge of the retainer. Remove and discard the old membrane.
5. Verify that the o-ring is properly positioned in the associated groove holder.
6. Holding a single membrane by the edges only, place it across the membrane holder and snap the retainer in place.
7. Using a sharp razor blade, carefully trim away excess membrane around the edges. Take care that the razor blade does not cut into the edges of the membrane assembly.
8. Set the sensor body on a flat surface with the cathode facing upward. Verify that the o-ring is properly posi­tioned around the cathode. Pour the electrolyte over the cathode / central post assembly so it runs down into the sensor electrolyte well. Fill the well to a level flush with the top of the sidewall. Put the membrane assembly directly onto the cathode so that the face of the holder fits against the o-ring in the end of sensor body. The membrane is now in place.
9. Carefully place the cap on the sensor body. Screw the cap on, finger-tight only. Then lay the sensor on its side with the side port up. Remove the side port screw, rubber pressure-compensating diagram, and washer. Add electrolyte, if necessary, to bring the level even with the shoulder. With the side port still facing up, tight­en the cap further until it is snug and the membrane is stretched taut across the cathode. Any excess elec­trolyte displaced from the well may now be removed by blotting with a tissue.
10. Insert the new rubber diaphragm into the side port, place the new washer over the diaphragm, and secure the side port screw. Do not overtighten the screw.
11. Inspect the sensor for possible leaks or damage to the membrane.
12. Remove the adhesive tape in step 2. The sensor is now ready for operation.
FIGURE 8. Rechargeable Sensor — Exploded View (6A00181G01)
6
MODEL 4000 MAINTENANCE
ACCESSORIES
PART # DESCRIPTION
6A00195G01 Cable, 10 ft
ORDERING INFORMATION
The Model 4000 sensor is housed in a molded Ryton
®
1
body for insertion or in line flow installations. A Pt 100 allows for accurate temperature compensation. The Model 4000 amperometric sensor provides real time measurement, excellent performance,a nd reliability in various detection and analysis applications.
PART # DESCRIPTION
6A00181G01 Percent Oxygen Sensor, Rechargeable, Ryton
®
1
6A00182G01 Percent Oxygen Sensor with Gland, Rechargeable, Ryton
®
1
6A00184G01 Percent Oxygen Sensor with Fast Response Flow Assembly, Rechargeable, Ryton
®
1
MODEL 4000 PERCENT OXYGEN SENSOR
SPARE PARTS
PART # DESCRIPTION
6A00194G01 Rebuild Kit
1
Ryton is a registered trademark of Chevron Phillips Chemical Company LP.
7
Credit Cards for U.S. Purchases Only.
The right people, the right answers, right now.
ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
http://www.raihome.com
Specifications subject to change without notice.
Emerson Process Management
Liquid Division
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250
http://www.raihome.com
© Rosemount Analytical Inc. 2009