Rosemount 399VP General Purpose pH/ORP Sensor Abridged Manuals & Guides

Page 1
Model 399VP
General Purpose pH/ORP Sensor
Instruction Sheet
PN 51A-399VP/rev.J January 2011
MODEL 399VP
Insertion/Submersion Sensor
with new VP8 connector
CAUTION
The solution used during calibration is an acid and should be handled with care. Follow the directions of the acid manufacturer. Wear the proper protective equipment. Do not let the solution come in contact with skin or clothing. If contact with skin is made, immediately rinse with clean water.
The wetted sensor materials may not be compatible with process com position and operating conditions. Application compat ibility is entirely the
responsi-
bility of the user.
CAUTION
SENSOR/PROCESS APPLICATION COMPATIBILITY
SPECIFICATIONS
Sensor Type: General purpose 399VP Measured Range: pH: A
CCUGLASS 0-14
ORP: -1500 to +1500 mV
Percent Linearity Over pH Range:
Operating Temperature: 0° to 85°C (32° to 185°F)
Au tomatic temperature compensation 0° to 85°C (32° to 185°F), Temperature compensation is not required for 399 ORP when used with Models 1060, or 1181 ORP
Maximum Pressure: 790 kPa [abs] (100 psig) at 65°C (see
Graph 1 below)
Flow Rate: up to 2 ft/sec Materials of Construction: Tefzel, glass, ceramic, & Viton Materials of ORP: Platinum Process Connections: 1 in. MNPT, 2 places VP8 Cable: use PN 24281-00 Weight/Shipping Weight: 0.45 kg/0.9 kg (1 lb/2 lb)
• High pH glass provides greater accuracy over 12.5 pH
• HF glass provides improved resistance to hydrofluoric acid up to 300 ppm
GPLR Hemi Bulb Flat Bulb
1-2 pH 94% 93%
2-12 pH 99% 98%
12-13 pH 97% 95%
13-14 pH 92%
For additional information, please visit our website at
www.emersonprocess.com/raihome/liquid/.
ATEX DIRECTIVE
Special Conditions for safe use
1. All pH/ORP sensors have a plastic enclosure which must only be cleaned with a damp cloth to avoid the danger due to a build up of an electro­static charge.
2. All pH/ORP sensor Models are intended to be in contact with the process fluid and may not meet the 500V r.m.s. a.c. test to earth. This must be taken into consideration at installation.
WARNING
Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is lowered to a safe level!
Page 2
2
STORAGE
1. It is recommended that electrodes be stored in their original shipping containers until needed.
2. Do not store at temperatures below -5°C (23°F).
3. Electrodes should be stored with a protective cap containing KCl solution (PN 9210342).
4. For overnight storage, immerse the sensor in tap water or 4 pH buffer solution.
5. A pH glass electrode has a limited shelf life of one year.
ELECTRODE PREPARATION
1. Remove electrode from shipping container.
2. Remove the protective boot covering the electrode bulb.
3. Rinse away salt film with clean water, then shake the electrode so that the internal solution fills the bulb, thus removing any air trapped there.
NOTE
Do not allow lubricant to coat electrode bulb or reference junction. If it does, wipe it clean before installation.
INSTALLATION
For sensor orientation, see Figure 1.
For sensor dimensions, see Figure 2.
For wiring, see Figure 3 - 40.
TWO POINT BUFFER CALIBRATION
Select two stable buffer solutions, preferably pH 4.0 and 10.0 (pH buffers other than pH 4.0 and pH 10.0 can be used as long as the pH values are at least two pH units apart).
NOTE
A pH 7 buffer solution reads a mV value of approx. zero, and pH buffers read approximately ± 59.1 mV for each pH unit above or below pH 7. Check the pH buffer manufacturer specifications for millivolt values at various temperatures since it may affect the actual value of the buffer solution mV/pH value.
1. Immerse sensor in the first buffer solution. Allow sensor to equilibrate to the buffer temperature (to avoid errors due to temperature differences between the buffer solution and sensor tempera­ture) and wait for reading to stabilize. Value of buffer can now be acknowledged by analyzer/ transmitter.
2. Once the first buffer has been acknowledged by the analyzer/transmitter, rinse the buffer solution off of the sensor with distilled or deionized water.
3. Repeat steps 1 and 2 using the second buffer solution.
4. The theoretical slope value, according to the Nernst equation for calculating pH, is approximately 59.17 mV/pH. Over time the sensor will age, both in the process and in storage, and will result in reduced slope values. To ensure accurate readings, it is recommended that the electrode be replaced when the slope value falls below 47 to 49 mV/pH.
RECOMMENDED pH SENSOR STANDARDIZATION
For maximum accuracy, the sensor can be standard­ized on-line or with a process grab sample after a buffer calibration has been performed and the sensor has been conditioned to the process. Standardization accounts for the sensor junction potential and other interferences. Standardization will not change the sensor’s slope but will simply adjust the analyzer’s reading to match that of the known process pH.
MAINTENANCE FOR pH ELECTRODE
Electrodes should respond rapidly. Sluggishness, offsets, and erratic readings are indicators that the electrodes may need cleaning or replacement.
1. To remove oil deposit, clean the electrode with a mild non-abrasive detergent.
2. To remove scale deposits, soak electrodes for 30 to 60 minutes in a 5% hydrochloric acid solution.
3. Temperature effect on life expectancy: If glass electrode life expectancy is 100% @ 25°C (77°F), then it will be approximately 25% @ 80°C (176°F), and approximately 5% @ 120°C (248°F).
MODEL 399VP INSTALLATION
Page 3
3
MODEL 399VP INSTALLATION
FIGURE 2. Sensor Dimensions
DWG. NO. REV.
40399VP01 A
INCH
MILLIMETER
INSTALLATION
FIGURE 1. Sensor Orientation
Page 4
MODEL 399VP INSTALLATION
4
ORP CALIBRATION
1. After making an electrical connection between the sensor and the instrument, obtain a standard solution of saturated quinhydrone (PN R508-8OZ). This can also be made quite simply by adding a few crystals of quinhydrone to either pH 4 or pH 7 buffer. Quinhydrone is only slightly soluble, there­fore only a few crystals will be required.
2. Immerse the sensor in the standard solution. Allow 1-2 minutes for the ORP sensor to stabilize.
3. Adjust the standardize control of the instrument to the solution value shown in the table below. The resulting potentials, measured with a clean plat­inum electrode and saturated KCl/AgCl reference electrode, should be within +/- 20 millivolts of the value shown in the table below. Solution temperature
must be noted to ensure accurate interpretation of results. The ORP value of saturated quinhydrone solution is not stable over long periods of time. Therefore, these standards should be made up fresh each time they are used.
4. Remove the sensor from the buffer, rinse, and install in the process.
MAINTENANCE FOR ORP ELECTRODE
Electrodes should respond rapidly. Sluggishness, off­sets, and erratic readings are indicators that the elec­trodes may need cleaning or replacement.
1. To remove oil deposit, clean the electrode with a
mild non-abrasive detergent.
2. To remove scale deposits, soak electrodes for 30 to
60 minutes in a 5% hydrochloric acid solution.
3. ORP (metallic) electrodes should be polished with
moistened baking soda.
pH 4 Solution pH 7 Solution
Temp °C 20 25 30 20 25 30
mV Potential 268 264 260 94 87 80
ORP of Saturated Quinhydrone Solution (millivolts)
CAUTION
Hydrofluoric acid and the mixture of Hydrochloric acid and thiourea is toxic and highly corrosive. Avoid skin contact, wear protective gloves. Use only in a well ventilated area. Do not inhale fumes. In case of an accident, consult a doctor immediately.
FIGURE 5. VP8 Cable, instrument end
FIGURE 4. VP8 Cable, sensor end
Page 5
MODEL 399VP WIRING
5
WIRING
FIGURE 7. Model 81 VP Wiring FIGURE 8. Model 1181 VP Wiring
FIGURE 9. Model 54 VP Wiring through Remote Junction Box
GREEN
VP6
(PN 24281-00) VP8
GREEN
VP6 cable
VP8 cable (PN 24281-00)
FIGURE 6. Model 1056. This is the standard wiring for VP8 cable assembly. VP8 cable works both with
VP8 and VP6 sensor connector. VP8 cable is required for SMART communication.
Page 6
6
MODEL 399VP WIRING
FIGURE 11. Model 2081 VP Wiring
FIGURE 13. Model 2081 VP Wiring through
Remote Junction Box
FIGURE 12. Model 1181, 1050/1060, and 1003/1023
VP6 Wiring through Remote Junction Box
FIGURE 10. Model 54 and 54e VP6 Wiring
GREEN
GREEN
Page 7
7
MODEL 399VP WIRING
FIGURE 16. Model 54/54e VP8 Wiring
FIGURE 14. Model 81 VP6 Wiring through
Remote Junction Box
FIGURE 15. Model 3081, 4081, and 5081 VP
Wiring through Remote Junction Box
VP6
VP6
399VP
(PN 24281-00) VP8
Page 8
MODEL 399VP WIRING
8
FIGURE 18. Model 3081, 4081, and 5081 Wiring. This
is standard VP8 wiring for 399VP. VP8 cable works
with both VP6 and VP8 sensor connector. VP8 cable
is required for SMART communication.
FIGURE 17. Model 1055-22-32 VP6 Wiring
FIGURE 19. Model 1054 VP Wiring
FIGURE 20. Model 1054A/B and 2054 VP6 Wiring
399VP
VP6
(PN 24281-00) VP8
Page 9
9
FIGURE 22. Model 1054A/B & 2054 VP6 Wiring
through a Remote Junction Box
FIGURE 21. Model 1054 VP6 Wiring through
a Remote Junction Box
MODEL 399VP WIRING
FIGURE 23. Model 1055 -01 -10 -22 Wiring.
This is standard VP8 wiring for 399VP. VP8 cable works with both VP6 and VP8 sensor connector
Note: Place RC jumper (P/N 23988-00) across “Reference In” to “Solution Ground” when using dual sensor inputs and the pH sensor doesn’t have a solution ground.
Page 10
10
MODEL 399VP WIRING
FIGURE 25. Model Xmt Wiring. This is standard VP8
wiring for 399VP. VP8 cable assembly works with both
VP8 and VP6 assembly
FIGURE 24. Model SCL-(P/Q) VP6 Wiring
FIGURE 26. Model 1055-22-32 integral cable and VP6 Wiring through Remote Junction Boxes
399VP
Page 11
FIGURE 27. Wiring Model 399VP-09, 400, 401, 402, 403, and 404 to Model 1055 (Pipe/Wall Mount).
DWG. NO. REV.
41055107 C
11
MODEL 399VP WIRING
FIGURE 29. Model 399VP-09 to Model 6081 VP6 Wiring
(no reference diagnostics)
FIGURE 28. Model 399VP-09 to Model 6081 VP8 Wiring
(no reference diagnostics)
Note: Do not connect the Blue wire to Solution Ground but Instead terminate with a wire nut and jumper Solution Ground to Reference In. For models Xmt and 1056 termi­nate Inner Drain with a wire nut.
Note: Do not connect the Blue wire to Solution Ground but Instead terminate with a wire nut and jumper Solution Ground to Reference In. For models Xmt and 1056 termi­nate Inner Drain with a wire nut.
Page 12
Credit Cards for U.S. Purchases Only.
The right people, the right answers, right now.
ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
http://www.raihome.com
Specifications subject to change without notice.
Emerson Process Management
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250
http://www.raihome.com
© Rosemount Analytical Inc. 2011
Sira MC070110/00
ACCESSORIES
PART NUMBER DESCRIPTION
23645-06 15 ft cable with mating VP6 connector, prepped with BNC on analyzer end 23645-07 15 ft cable with mating VP6 connector, prepped without BNC on analyzer end 24281-00 15 ft. cable with mating VP8 connector 24281-01 25 ft. cable with mating VP8 connector 24281-02 2.5 ft. cable with mating VP8 connector 24281-03 50 ft. cable with mating VP8 connector 24281-04 100 ft. cable with mating VP8 connector 24281-05 4 ft. cable with mating VP8 connector 24281-06 10 ft. cable with mating VP8 connector 24281-07 20 ft. cable with mating VP8 connector 24281-08 30 ft. cable with mating VP8 connector
915240-03 Process tee with NPT connection threads
22719-02 Junction box 12707-00 Jet cleaner 23242-02 Insertion adapter 24091-00 Low flow cell
8
Loading...