Rosemount 3150 Series Nuclear Pressure
Transmitters including the Rosemount
3152, 3153 and 3154
Reference Manual
00809-0100-4835 Rev BD
July 2016
Rosemount 3150 Series
NOTICE
Rosemount Nuclear Instruments, Inc. satisfies all obligations
Rosemount 3150 Series Nuclear Pressur e Transmitters
including the Rosemount 3152, 3153 and 3154
Read this manual before working with the product. For personal and system
safety, and for optimum performance, make sure you thoroughly understand the
contents of this manual before installing, using or maintaining this product.
For Assistance:
Within the United States, contact Rosemount Nuclear Instruments, Inc. at
1-952-949-5210.
Outside the United States, contact the nearest Rosemount representative.
Customer Feedback:
Your feedback is important to us, please send comments or suggestions to:
Chan.RNII-CustomerFeedback@emersonprocess.com
coming from legislation to harmonize product requirements in
the European Union.
i
Rosemount 3150 Series
Reference Manual
00809-0100-4835 Rev BD
July 2016
Rosemount Nucl ear Instruments, I nc. Warranty and
Limitations of Remedy
The warranty and limitations of remedy applicable to this Rosemount equipment are as stated on the reverse
side of the current Rosemount quotation and customer acknowledgment forms.
RETURN OF MATERIAL
Authorization for return is required from Rosemount Nuclear Instruments, Inc. prior to shipment. Contact
Rosemount Nuclear Instruments, Inc. (1-952-949-5210) for details on obtaining Return Material Authorization
(RMA). Rosemount Nuclear Instruments will not accept any returned materi al without a Return Material Authorization. Material returned without authorization is subject to return to customer.
Material returned for repair, whether in or out of warranty, should be shipped prepaid to:
Rosemount Nuclear Instruments, Inc.
8200 Market Boulevard
Chanhassen, MN 55317
USA
IMPORTANT
Rosemount 3152, 3153 and 3154 Series Pressure Transmitters are designed for Nuclear Class 1E usage, and
have been tested to the standards shown below:
• IEEE Std 323™-1974, -1983 and -2003
• IEEE Std 344™-1975, -1987 and -2004
These transmitters are manufactured under aquality system that meets the requirements of 10CFR50
Appendix B, 10CFR Part 21, ISO 9001, NQA-1, KTA 1401, KTA 3507, CSA N285.0, CSA Z299 and the
applicable portions of IAEA-50-C-Q. During qualification testing, interfaces were defined between the
transmitter and its environment that are essential to meeting requirements of the qualification standards listed
above. Specifically, to ensure compliance with 10CFR Part 21, the transmitter must comply with the
requirements herein and in the applicable Rosemount qualification report(s) throughout its installation,
operation and maintenance. It is incumbent upon the user to ensure that the Rosemount Nuclear Instruments,
Inc.’s component traceability program is continued throughout the life of the transmitter.
In order to maintain the qualified status of the transmitter, the essential environmental interfaces must not be
compromised. Performance of any operations on the transmitter other than those specifically authorized in
this manual has the potential for compromising an essential environmental interface. Where the manual
uses the terms requirement, mandatory, must or required, the instructions so referenced must be
carefully followed. Rosemount Nuclear Instruments, Inc. expressly disclaims all responsibility and liability for
transmitters for which the foregoing has not been complied with by the user.
• RCC-E-2002
• KTA 3505-2005
ii
Reference Manual
00809-0100-4835 Rev BD
July 2016
Rosemount 3150 Series
Cover,
throughout
Cover,
throughout
Throughout
Throughout
Updated references and terminology for consistency throughout document.
Minimum time constant for damping adjustment set to maximum position for
3153 and 3154 transmitters (output code T) changed from 0.8 sec to 1.2 sec.
Relocated step 2 in the Preliminary Electric al Housing Reassembly Section to
the Electronics Housing Cover Installation Section and revised for clarity.
Updated Electronics Assembly Installation procedure to include additional
instruction for installing the connector plug.
Added Figure 5-7 – Connector Plug Detail and updated subsequent figure and
page numbers accordingly.
NOTE
The above Revision Status list summarizes the changes made. Please refer to both manuals for complete
comparison details.
Revision Status
Changes from March 2015 (Rev BC) to July 2016 (Rev BD)
Page
(Rev BC)
3-11 3-11
5-11 & 5-13 5-11 & 5-14
5-12 5-12
5-12 to 5-18 5-12 to 5-19
Page
(Rev BD)
Changes
Document revision change from March 2015 to July 2016, Rev BC to Rev BD
This manual is designed to assist in installing, operating and
NOTE
Refer to the applicable Rosemount Qualification/Test Reports, Product
performance specifications and dimensional drawings for each model.
Unique Transmitter Serial Number
is stamped here
Transmitter Model Number is
stamped here
Factory Calibrated Span
is stamped here
Transmitter Maximum Power
Maximum Working
Pressure is stamped here
SECTION 1: INTRODUCTION
maintaining the Rosemount 3150 Series Pressure Transmitters.
Instructions for the 3152, 3 153 and 3154 models are included in this
manual. Where differences in instructions between the models occur
they shall be noted within those instructions. The manual is organized
into the following sections:
Data Sheets and/or Specification Drawings for details on testing,
Figure 1-1 shows the standard transmitter nameplate and where
transmitter information is stamped onto the nameplate. Nameplate
material is stainless steel.
Figure 1-1 – Standard Transmitter
Nameplate
Section 2: Installation
Provides general, mechanical, and electrical installation
considerations.
Section 3: Calibration
Provides transmitter calibration procedures.
Section 4: Operation
Provides a description of how the transmitter operates.
Section 5: Maintenance and Troubleshooting
Provides basic hardware troubleshooting considerations including
disassembly and reassembly procedures and post assembly tests.
Section 6: Transmitter Spare Pa rts
Provides order information for transmitter spare parts.
o Process Connections
o Impulse Piping
o Mounting Configurations
o Conduit
o Electrical Housing
• Electrical Considerations
• Installation Procedures
o Mechanical
precautions to ensure the safety of the personnel performing the
operation. Refer to the following safety messages before performing
Explosions can result in death or injury.
• Do not remove the transmitter covers in explosive
• Verify that the operating atmosphere of the transmitter is
environments when the circuit is live.
consistent with the appropriate qualification parameters.
Rosemount 3150 Series
Reference Manual
00809-0100-4835 Rev BD
July 2016
WARNING
GENERAL
CONSIDERATIONS
Measurement accuracy depends upon proper installation of the
MECHANICAL
CONSIDERATIONS
This section contains information you should consider when preparing
WARNING
NOTE
For steam service, do not blow down impulse piping through the
lines with water before resuming measurement.
Electrical shock can result in death or serious injury.
• Avoid contact with the leads and terminals.
Process leaks could result in death or serious injury.
•Install and tighten all four flange bolts before applying
pressure.
•Do not attempt to loosen or remove flange bolts while the
transmitter is in service.
Replacement equipment or spare parts not approved by
Rosemount Nuclear Instruments, Inc. for use could reduce the
pressure retaining capabilities of the transmitter and may render
the instrument dangerous or adversely impact i ts qualified
status.
•Use only components supplied with the Rosemount 3152,
3153 or 3154 transmitter or designated by Rosemount
Nuclear Instruments, Inc. as spare parts for the 3152,
3153 or 3154.
Improper assembly of mounting bracket to traditional process
flange can damage sensor module.
•For safe assembly of bracket to transmitter traditional
process flange, bolts must break back plane of flange
web (i.e. bolt hole), but must not contact module housing.
Use only the RNII approved bolts supplied with the
bracket.
transmitter and its associated impulse piping and valves. Mount the
transmitter close to the process and use a minimum of piping to
achieve best accuracy. For flow measurement, proper installation of
the primary element is also critical to accuracy. Also, consider the
need for easy access, personnel safety, practical field calibration and a
suitable transmitter environment. Transmitter installation should
minimize the effects of temperature gradients and fluctuations, and
avoid vibration and shock during normal operation.
to mount the transmitter. Read this section carefully before
proceeding to the mechanical installation procedure. Proper
installation is mandatory to assure seismic qualification.
Do not attempt to loosen or remove flange bolts while the transmitter
is in service.
transmitter. Flush the lines with the transmitter isolated and refill the
2-2
Reference Manual
00809-0100-4835 Rev BD
July 2016
Rosemount 3150 Series
NOTE
When the transmitter is mounted on its side, position the traditional
liquid service.
Mount the Rosemount 3150 Series transmitter to a rigid support (i.e.
position.
NOTE
The transmitter is calibrated in an upright position at the factory.
Calibration.
process flanges to ensure proper venting or draining. Keep drain/vent
connections oriented on the bottom for gas service and on the top for
one with a fundamental mechanical resonant frequency of 40 Hz or
greater). Two mounting options are qualified for the transmitter: panel
mount or 2-inch pipe mount. A stainless steel panel bracket is
provided with the 3154. For the 3152 and 3153 series, the user has
the option of specifying either the stamped carbon steel panel bracket
or the stainless steel panel bracket.
Refer to Figure 2-5 for qualified mounting configurations for both the
panel and pipe mount options.
Orientation with respect to gravity is not critical to qualification. For
maximum accuracy, zero the transmitter after installation to cancel any
zero shift that may occur due to liquid head effect caused by mounting
Mounting the transmitter in another position may cause the zero po int to
shift by an amount equivalent to the internal liquid head within the
sensor module induced by the varied mounting position. For maximum
accuracy, zero the transmitter to cancel this effect per Section 3:
2-3
Rosemount 3150 Series
Reference Manual
00809-0100-4835 Rev BD
July 2016
Process Connections and
Interfaces
Mount the process flanges with sufficient clearance for process
thickness for the pressure involved.
connections. For safety reasons, place the drain/vent valves so the
process fluid is directed away from possible human contact when the
vents are used. Also consider that access to the vent/drain valve(s)
and process connection(s) may be required for plant specific
operations (i.e. calibration, draining, etc.)
Process tubing must be installed to prevent any added mechanical
stress on the transmitter under seismic conditions. Use stress-relief
loops in the process tubing or separately support the process tubing
close to the transmitter.
Typical connections on the transmitter flanges are ¼ - 18 NPT or 3/8
inch Swagelok ™. Use your plant-approved, qualified thread sealant
when making threaded connections. The end-user is responsible for
the qualification of the threaded seal interface on all ¼ - 18 NPT
interfaces.
Transmitters with options including 3/8 inch Swagelok™ are shipped
with front ferrule, rear ferrule and nut. Place these fittings on the
tubing with the orientation and relative position shown in Figure 2-1.
Use process tubing with 3/8 inch outside diameter and of suitable
2-4
Reference Manual
00809-0100-4835 Rev BD
July 2016
Rosemount 3150 Series
Figure 2-1 – Swagelok™
Compression Fitting Detail
Impulse Piping
The Swagelok™ tube fittings come completely assembled and are
ready for immediate use.
because dirt or foreign materials may get into the fitting and cause
le
tubing rests firmly on the shoulder of the fitting and the nut is finger
tight. Tighten the nut one
To reconnect, insert the tubing with p
until the front ferrule sits in the fitting. Tighten the nut by hand, then
rotate one
position. Then snug it slightly with a wrench.
For more detailed i
Swagelok™ tube fittings, refer to:
If drain/vent valves are opened to
the value in Table 5
when closing.
The piping between the process and the transmitter must accurately
transfer the pressure to obtain accurate measurements. There are fi
possible sources of error: pressure transfer (such as obstruction), leaks,
friction loss (particularly if purging is used), trapped gas in a liquid line
or liquid in a gas line and density variations between the legs.
The best location for the transmit
depends on the process itself. Use the following guidelines to
determine transmitter location and placement of impulse piping:
•Make sure both impulse legs are the same temperature
Dimensions are nominal in inches (mm)
aks. Insert the tubing into the Swagelok™ tube fitting, make sure the
-quarter turn more or to the original one-and-one-quarter tight
• Keep impulse piping as short as possible
• For liquid service, slope the impulse piping at least 1 inch per
foot (8 cm per m) upward from the transmitter toward the
process connection
•For gas service, slope the impulse piping at least 1 inch per foot
(8 cm per m) downward from the transmitter toward the process
connection
•Avoid high points in liquid lines and low points in gas lines
Do not disassemble them before use
-and-one-quarter turns. Do not over-tighten.
re-swaged ferrules into the fitting
nformation regarding the specifications and use of
Fittings Catalog MS-01-140
“Gaugeable Tube Fittings and Adapter Fittings”
www.swagelok.com
-2 in Section 5 Maintenance and Troubleshooting
bleed process lines, torque stems to
ter in relation to the process pipe
ve
2-5
Rosemount 3150 Series
Reference Manual
July 2016
Mounting Configurations
•Use impulse piping of large enough diameter to avoid friction
l out of direct contact with
Refer to Figure 2
configurations:
Liquid Flow Measurement
Gas Flow Measurement
Stea
negligible.
NOTE
The mounting configurations described above and depicted in Figure 22 are based on general
Where applicable, specific plant approved installation practices should
be used.
NOTE
In steam or other elevated temperature services, it is important that
temperatures at the process flanges not exceed 25
vacuum service, these limits are reduced to 220oF (104oC).
• Vent all gas from liquid piping legs
• Vent all liquid from gas piping legs
• When using a sealing fluid, fill both piping legs to the same level
• When purging, make the purge connection close to the process
• Keep corrosive or hot process materia
• Prevent sediment deposits in the impulse piping
• Keep the liquid balanced on both legs of the impulse piping
• Avoid conditions that might allow process fluid to freeze within
• Make sure the impulse piping is of adequate strength to be
• Place taps to the side of the line to prevent sediment deposits
• Mount the transmitter beside or below the taps so gases vent
00809-0100-4835 Rev BD
effects and blockage
taps and purge through equal lengths of the same size pipe –
avoid purging through the transmitter
the transmitter
the process flange
compatible with anticipated pressure.
-2 for examples of the following mounting
on the process isolators.
into the process lines.
• Place taps in the top or side of the line.
• Mount the transmitter beside or above the taps to drain liquid
into the process line.
m Flow Measurement
• Place taps to the side of the line.
• Mount the transmitter below the taps to ensure that impulse
piping will remain filled with condensate.
•Fill impulse lines with water to prevent steam from contacting
the transmitter directly and to ensure accurate measurement
start-up. Condensate chambers are not typically necessary
since the volumetric displacement of the transmitter is
industry “best practice” recommendations.
0oF (121oC). In
2-6
Reference Manual
00809-0100-4835 Rev BD
July 2016
Rosemount 3150 Series
Conduit
Electrical Housing
The conduit connections to the transmitter are threaded. Options
f the unused opening with a compatible thread type
IMPORTANT
For all 3152 and 3153 transmitters, install the conduit plug (provided
with
values outlined in
Table 5-2.
Rosemount Nuclear Instruments, Inc.
Figure 2-2 Transmitter Installation
Examples (liquid, gas or steam)
Please note that transmitters depicted in Figure 2-2 are intended for reference only.
available are ½ -14 ANPT, M20, PG 13.5 and G1/2. Two openings are
available on the 3152 and 3153 transmitter housings for convenient
installation. Close of
stainless steel pipe plug. Use your plant-approved, qualified thread
sealant on the conduit connection threads.
the transmitter) in the unused conduit opening per the torque
The 3154 has one conduit connection.
Use a qualified conduit seal at the conduit entry to prevent moisture
from accumulating in the terminal side of the housing during accident
conditions. Certain option codes provide a qualified connector with the
connector factory assembled to the transmitter. To prevent the conduit
from adding mechanical stress to the transmitter during seismic
disturbances, use flexible conduit or support the conduit near the
transmitter. Install the conduit seal in accordance with the
manufacturer’s instructions or use the procedure in this section.
The standard transmitter orientation is shown in dimensional drawings
found in this manual (see Figure 2-6). The electronics housing cannot be rotated in the field. For more information, please contact
Section 5: Maintenance and Troubleshooting,
2-7
Rosemount 3150 Series
Reference Manual
00809-0100-4835 Rev BD
July 2016
ELECTRICAL
CONSIDERATIONS
This section contains information you should consider when preparing
receivers (controller, indicator, computer).
The power supply must supply at least 12 volts to the transmitter
to make electrical connections to the transmitter. Read this section
carefully before proceeding to the electrical installation procedure.
Rosemount 3150 Series transmitters provide a 4-20 mA signal when
connected to a suitable dc power source. Figure 2-3 illustrates a typical
signal loop consisting of a transmitter, power supply, and various
Supply
terminals at 20 mA signal, or the maximum output current required for
proper system operation. Any power supply ripple appears in the
output load. The power supply versus load limitation relationship is
shown in Figure 2-4. See qualification reports for additional details.
The loop load is the sum of the resistance of the signal leads and the
load resistance of the receivers.
2-8
Reference Manual
00809-0100-4835 Rev BD
July 2016
Rosemount 3150 Series
POWER SUPPLY (VDC)
2500
1500
2000
500
1000
12
20
25
30
35
40
45
50
48
13.5
15
38
55
IEEE
REGION
DESIGN
REGION
LOAD (OHMS)
53
2150
1725
60
43
0
2150
POWER SUPPLY (VDC)
2500
1500
500
12
1000
20
25
30
35
40
45
50
15
55
KTA
DESIGN
REGION
LOAD (OHMS)
53
60
43
0
1900
2000
Figure 2-4 – Transmitter Supply
Voltage vs. Load
Figure 2-4a – IEEE Qualified and Design Regions
(applicable to 3152N, 3153N and 3154N models only)
QUALIFIED
Figure 2-4b – KTA Qualified and Design Regions
(applicable to 3152K and 3154K models only)
QUALIFIED
REGION
2-9
Rosemount 3150 Series
Reference Manual
July 2016
Signal wiring need not be shielded, but twis ted pairs yi eld the best
in the signal
), and
45
43
POWER SUPPLY (VDC)
2500
2000
1500
500
12
1000
20
25
30
35
40
50
15 55
RCC
DESIGN
REGION
LOAD (OHMS)
53
2150
2050
60
0
38
750
Figure 2-4c – RCC-E Qualified and Design Regions
(applicable to 3153K and 3154K models only)
00809-0100-4835 Rev BD
QUALIFIED
REGION
results. Shielded cable should be used for best results in electrically
noisy environments. Do not run signal wiring in conduit or open trays
with AC power wiring, or near heavy electrical equipment. Signal wiring
may be ungrounded (floating) or grounded at any one point
loop.
For installations with EMC performance requirements, consult the
Rosemount Nuclear Instruments, Inc. EMC test reports for additional
details regarding recommended practices for electrical wiring per
various national and international codes and regulations.
The transmitter case may be grounded or ungrounded. Grounding
should be completed in accordance with national and local electrical
codes. Transmitter case can be grounded using either the internal or
external ground connection.
• Internal Ground Connection: The Internal Ground
• External Ground Assembly: The External Ground location is
Connection screw is inside the terminal side of the electronics
housing. The screw is identified by a ground symbol (
is standard on all 3150 Series transmitters.
indicated by the ground symbol ( ) on the module. An
External Ground Assem bl y kit can be ordered as an option on
the 3150 Series transmitter. This kit can also be ordered as a
spare part. Please contact RNII for ordering information.
2-10
Reference Manual
00809-0100-4835 Rev BD
July 2016
Rosemount 3150 Series
The capacitance sensing element uses alternating current to generate a
rough the
INSTALLATION
PROCEDURES
Mechanical
Installation consists of mounting the transmitter and conduit and making
WARNING
Improper assembly of mounting bracket to transmitter traditional
process
supplied with the bracket.
.
see
Troubleshooting.
capacitance signal. This alternating current is developed in an oscillator
circuit with a nominal frequency of 110 kHz +/- 11 kHz. This 110 kHz
signal is capacitively-coupled to the transmitter case ground th
sensing element. Because of this coupling, a voltage may be imposed
across the load, depending on choice of grounding.
This impressed voltage, which is seen as high frequency noise, has no
effect on most instruments. Computers with short sampling times in a
circuit where the negative transmitter terminal is grounded detect a
significant noise signal. Filter this signal out by using a large capacitor
(1 uf) or a 110 kHz LC filter across the load. Signal loops at any other
point are negligibly affected by this noise and do not need filtering.
electrical and process connections. The procedures for each operation
follow.
– Transmitter
flange can damage sensor module.
•For safe assembly of bracket to traditional flange, bolts must
break back plane of flange web (i.e. bolt hole), but must not
contact module housing. Use only the RNII approved bolts
1. Attach the mounting bracket to the mounting location as follows:
Panel Mount
Mount the bracket to a panel or other flat surface (for
illustration see Figur e 2-5). Please note that the bolts
required for this step are customer supplied hardware.
Based on qualification tests performed by Rosemount, the
bolts listed in Table 2-1 are recommended for the bracketto-customer interface. Torque each bolt to value shown in
Table 5-2 in Section 5 Maintenance and
Troubleshooting
Pipe Mount
Assemble the bracket kit to a 2-inch pipe (for illustration
Figure 2-5). Torque each bolt to value shown in Table 5-2
in Section 5 Maintenance and Troubleshooting.
2. Attach the transmitter to the mounting bracket (for illustration
see Figure 2-5). Use the four 7/16-20 x ¾ inch bolts with
washers supplied with the transmitter. Torque each bolt to
value shown in Table 5-2 in Section 5 Maintenance and
2-11
Rosemount 3150 Series
Reference Manual
00809-0100-4835 Rev BD
July 2016
Table 2-1 – Recommended bolts for
bracket-to-customer interface
(1) The Bracket Code can be found in the 13
(2) Bracket Codes 1 and 5 are available on the 3152 and 3153-series only
(3) Bracket Codes 2 and 7 are available on the 3152, 3153 and 3154-series
(4) Bracket Codes 3 and 8 are available on the 3152, 3153 and 3154-series; this bracket code includes the listed bracket along
with the pipe mount hardware (Rosemount P/N 01154-0044-0001)
BRACKET
CODE
(1)
0
(2)
1
(3)
2
(4)
3
(2)
5
(3)
7
(4)
8
BRACKET TYPE
No Bracket
Supplied
Carbon Steel
Panel Bracket
SST Panel
Bracket
SST 2” Pipe
Mount Bracket
Carbon Steel
Panel Bracket
SST Panel
Bracket
SST 2” Pipe
Mount Bracket
th
position of the 3152, 3153 and 3154 model strings
RECOMMENDED BOLT
FOR BRACKET TO
CUSTOMER INTERFACE
N/A
5/16-18 UNC 2A Grade 2
3/8-24 UNF 2A Grade 2
3/8-24 UNF 2A Grade 2
5/16-18 UNC 2A Grade 2
3/8-24 UNF 2A Grade 2
3/8-24 UNF 2A Grade 2
2-12
Reference Manual
00809-0100-4835 Rev BD
July 2016
Rosemount 3150 Series
NOTE: All dimensions are nominal in
inches (millimeters)
Figure 2-5 – Typical Transmitter Mounting Bracket Configuration, Traditional Flange
(1) (2)
(1) Transmitter and bracket orientation with respect to gravity will not impact qualification.
(2) Transmitters can alternatively be mounted inside bracket (as shown below) or with process connection posit i oned adj acent to bracket
(not shown).
(3) A pipe-mount kit with three (3) U-bolts (not shown) is also available. The use of a third U-bolt is necessary to meet KTA seismic
requirements in installations where the KTA Airplane Crash (APC) value of 8g is applicable,. Please consult the applicable RNII Product
Data Sheet (PDS) for ordering information.
2-13
Rosemount 3150 Series
Reference Manual
July 2016
NOTE: All dimensions are nominal in inches (millimeters)
NOTE: All dimensions are nominal in inches (millimeters)
Figure 2-6 – Transmitter Dimensional
Drawings
00809-0100-4835 Rev BD
3152, 3153 Traditional Flange
3154 Traditional Flange
2-14
Loading...
+ 56 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.