Temperature sensors are rarely inserted directly into an industrial process. They are installed into a thermowell to isolate them from the
potentially damaging process conditions of flow-induced stresses, high pressure, and corrosive chemical effects. Thermowells are
closed-end metal tubes or barstock installed into the process vessel or piping and become an integral pressure-tight part of the process
vessel or pipe. They permit the sensor to be quickly and easily removed from the process for calibration or replacement without requiring
a process shutdown and possible drainage of the pipe or vessel.
The Rosemount 114C Thermowell is made from solid barstock to ensure strength and integrity. The Rosemount 114C was designed to
accommodate a host of industry standard configurations, but has the flexibility to adapt to special configurations for different types of
applications.
Flexible design modifiers accommodate many process requirements
Different sizes of root (A), tip (B), and bore (d) diameters
Different tip (t) thickness as required
Numerous combinations for various industrial applications
Meet ASME PTC 19.3 TW standard with flexible design
A
B
t
d
Thermowell calculations ensure thermowell design
is compatible with process conditions
Thermowells inserted into any process are subjected to forces
from the fluid flow.
ASME released design standard PTC 19.3 TW that defines a
series of qualitative calculations to determine design suitability
to withstand process conditions.
Calculations are based on process conditions, process fluid,
installation method, thermowell geometry and material.
The thermowell calculation tool from Emerson is designed to
meet the ASME standard.
The free online software version can be used for preliminary
calculations. Click here to access the tool.
Formal calculations by our expert team of engineers are
available prior to order placement to ensure design is
acceptable.
Recommendations are also made if thermowell design is found
to be unacceptable for the application.
Order Thermowell Calculation option (R21) to receive a formal
report with product shipment.
Figure 1. Thermowell Calculation Report (R21)
Emerson.com/Rosemount
3
Page 4
Rosemount 114C
June 2019
Wide range of thermowell options and certificates for any application
Options for special testing requirements, such as External Hydrostatic Pressure Test (Q5) and Dye Penetration Test (Q73)
Options to ensure material traceability or compatibility, including Positive Material Identification or PMI (Q76), Material
®
Certification (Q8), Thermowell X-ray/Radiograph (Q81), and NACE
Options for special processing requirements such as Electropolishing (R20)
Approval (Q35)
Experience global consistency and local support from numerous worldwide
Emerson manufacturing sites
Selecting the proper thermowell for an application is an important activity as it impacts plant safety and measurement efficiency.
Thermowells are considered a wetted part; they physically become part of the pressure retaining system.
The four major factors to consider when selecting a thermowell for an application are described below:
Thermowell length
There is no standard formula to determine thermowell immersion length. However, there are a few common practices that the
process industry follows along with good engineering judgment. Ideally the thermowell tip should be located near the centerline in
turbulent flow conditions because this represents the most accurate process temperature.
To ensure optimal performance, a general guideline for immersion length into a pipe is as follows:
10⫻ the thermowell root diameter for air or gas
5⫻ the thermowell root diameter for liquids
Another guideline is at least one-third the way into the pipe for any measurement. The American Petroleum Institute (API) has a
specific recommendation of using an immersion length of the sensing element plus 50 mm (2-in.).
Mounting configuration
Consider how the thermowell is mounted on the pipe or tank. The process designer typically specifies what mating connection will
be used and the thermowell selected should match that connection. Temperature, pressure, and material are usually taken into
consideration to ensure the process connection will be adequate for the application. Welded, threaded, flanged, and Van Stone are
standard mounting configuration options.
Thermowell stem profile
Factors to be considered when selecting a stem style include the process pressure, required response speed of the measurement,
drag force of the fluid flow on the well, and the wake frequency. The stem or shank is the part of a thermowell inserted into the
process piping or vessel. Straight, stepped and tapered stem styles are available. Each profile has its advantages depending on the
need and situation.
Thermowell material
Rosemount Thermowells are supplied in most materials required for industrial applications. Standard materials are 316/316L
Stainless Steel, 304/304L Stainless Steel, and A105 Carbon Steel. For corrosive environments, special materials such as Alloy C-276
and Alloy 600 are also available. See the ordering table for a complete listing of standard materials. Contact your local Emerson
representative for additional material availability.
Emerson.com/Rosemount
5
Page 6
Rosemount 114C
Rosemount 114C Threaded Thermowells
June 2019
6
Emerson.com/Rosemount
Page 7
June 2019
Model
Instrument
connection threads
Head
length (H)
Thermowell
material
Stem
style
Process
connection
Mounting
style
Immersion
length (U)
Units
114CXXXXXXXXXXTXXXX
16151413121198765432110191817
(5) Units
English (E)
Metric (M)
(6-9) Immersion length (U)
25 to 1165 mm (M)
Example: 25 mm = 0025,
1165 mm =1165
1- to 42-in. (E)
Example: 2-in. = 0020,
42-in. =0420
Threaded thermowells are threaded into a process pipe or tank, allowing for easy installation and removal when necessary. While this
is a common mounting method, it has the lowest pressure rating of all mounting configuration options.
Figure 2. Standard Offering–Threaded
The common options shown in Figure 2 represent a partial offering; reference the Rosemount 114C Threaded Ordering Information
for a full list of available options.
Figure 3. Threaded Thermowell Components
A
A. Instrument connection
B. Process connection
Note
Wetted surface includes engaged threads and immersion length (U).
Emerson.com/Rosemount
H
Tapered threadsParallel threads
B
U
A
H. Head length
U. Immersion length
HU
B
7
Page 8
Rosemount 114C
June 2019
Use the form below to record your model code:
ModelUnits
Immersion
length (U)
Mounting
style
Process
connection
Stem
style
Thermowell
material
Head length (H)
Instrument
connection
Options
114CT
123456 through 91011 and 121314 and 1516 through 1819X X X X X
8
Emerson.com/Rosemount
Page 9
June 2019
U
Threaded ordering information
Figure 4. Model Number Ordering Example
ModelUnits
Immersion
length (U)
Mounting
style
Process
connection
Stem
style
Thermowell
material
Head length (H)
Rosemount 114C
Instrument
connection
Options
114C E 0060TA A1S C0 5 0A
12345 67891011 121314 1516 17 1819X X X X X
WR5,
Q76...
The numbers below the model number ordering example correlate to the character place numbers in the second column of the
ordering table.
Table 1. Rosemount 114C Threaded Ordering Information
The starred offerings (★) represent the most common options and should be selected for best delivery. The non-starred offerings are subject
to additional delivery lead time.
Place #s
1-4
★
Place #
★
ModelDetails
Barstock temperature
114C
thermowell
Dimension units
5
EEnglish units (in.)
Made with a standard bore diameter of 0.26-in. (6.6 mm) and tip wall
thickness of 0.25-in. (6.4 mm)
Ref.
page
N/A
Ref.
page
59
Specifies whether length units will be in inches (in.) or millimeters (mm)
★
MMetric units (mm)59
Tapered threadsParallel threads
Place #s
6-9
Immersion length (U)
U
Ref.
page
★
xxxx
Example of a 6.25-in. length where the second decimal is dropped off: 0062
xxxx mm, 25 to 2500 mm in 5 mm increments (when ordered with dimension units code M)
xxx.x-in., 1.00 to 100-in. in 1/4-in. increments (when ordered with dimension units code E)
★
xxxx
Example of a 50 mm length: 0050
Place #
10
★
Place #s
11-12
★
★
★
★
★
Emerson.com/Rosemount
Mounting style
TThreadedN/A
Process connection
1
AA
AB
AC1–11.5 NPTTaper ed threadsN/A
AD11/2–11.5 NPTTaper ed threadsN/A
AE
/2–14 NPTTaper ed threadsN/A
3
/4–14 NPTTaper ed threadsN/A
1
/2-in. BSPTTap ered t hreadsN/ A
59
59
Ref.
page
Ref.
page
9
Page 10
Rosemount 114C
June 2019
Table 1. Rosemount 114C Threaded Ordering Information
The starred offerings (★) represent the most common options and should be selected for best delivery. The non-starred offerings are subject
to additional delivery lead time.
★
★
★
★
★
3
AF
/4-in. BSPTTap ered t hreadsN/ A
AG1-in. BSPTTaper ed threadsN/A
DAM20 ⫻ 1.5pParallel threadsN/A
DBM24 ⫻ 1.5pParallel threadsN/A
DCM27 ⫻ 2pParallel threadsN/A
DDM33 ⫻ 2pParallel threadsN/A
★
★
★
1
DE
DF
/2-in. BSPF (G1/2)Parallel threadsN/A
3
/4-in. BSPF (G3/4)Parallel threadsN/A
DG1-in. BSPF (G1)Parallel threadsN/A
Place #
13
Stem styleDetailsImage
Minimum immersion length
★
1Straight
1-in. (25 mm) - Tapered threads
1.75-in. (45 mm) - Parallel threads
Minimum immersion length
★
2Tap e re d
1-in. (25 mm) - Tapered threads
1.75-in. (45 mm) - Parallel threads
Minimum immersion length
★
3Stepped
3-in. (75 mm) - Tapered threads
3.75-in. (95 mm) - Parallel threads
Place #s
14-15
★
Thermowell material Details
SC316/316L dual rated61
SD316/316L dual rated (NORSOK) Must order the Q8 Material Certificate to get NORSOK documentation62
★
SF304/304L dual rated61
★
CSCarbon steel (A-105)61
SG316Ti SST61
SL310 SST61
SM321 SST61
Ref.
page
60
60
60
Ref.
page
SN321H SST61
SR904L SST61
SP347 SST61
ABAlloy B361
ACAlloy C-27661
AGAlloy 2062
AHAlloy 40061
AKAlloy 60061
10
Emerson.com/Rosemount
Page 11
June 2019
H
Rosemount 114C
Table 1. Rosemount 114C Threaded Ordering Information
The starred offerings (★) represent the most common options and should be selected for best delivery. The non-starred offerings are subject
to additional delivery lead time.
AMAlloy 60161
ANAlloy 62561
APAlloy 80061
AQAlloy 800H/HT61
ARAlloy 82561
AUAlloy C-20261
ASAlloy F44 Mo661
MOMolybdenum61
Chrome-Moly Grade B-11/F-11
CA
Class II
Chrome-Moly Grade B-22/ F-22
CB
Class III
61
61
CCChrome-Moly Grade F-9161
NKNickel 20061
TTTitanium Grade 261
DSSuper duplex SST Grade F-5361
Super duplex SST Grade F-53 –
DT
NORSOK
Must order the Q8 Material Certificate to get NORSOK documentation62
DUDuplex 2205 Grade F5161
Duplex 2205 Grade F51 –
DV
NORSOK
Must order the Q8 Material Certificate to get NORSOK documentation62
Tapered threadsParallel threads
Place #
16-18
★
xxx
★
xxx
Head length (H)
xx.x-in., 1.75 to 11.25-in. in 1/4-in. increments (when ordered with dimension units code E)
Example of a 6.25-in. length where the second decimal is dropped off: 062 (default head length = 1.75-in.)
xxx mm, 40 to 225 mm in 5 mm increments (when ordered with dimension units code M)
Example of a 50 mm length: 050 (default head length = 45 mm)
H
Ref.
page
63
63
Emerson.com/Rosemount
11
Page 12
Rosemount 114C
Table 1. Rosemount 114C Threaded Ordering Information
The starred offerings (★) represent the most common options and should be selected for best delivery. The non-starred offerings are subject
to additional delivery lead time.
June 2019
Place #
19
★
★
Instrument connectionDetailsImage
1
A
B
C
DM18 ⫻ 1.5p64
EM20 ⫻ 1.5p64
FM24 ⫻ 1.5p64
GG 1/2-in. (BSPF)64
HG 3/4-in. (BSPF)64
JM27 ⫻ 2p64
KM14 ⫻ 1.5p64
/2–14 NPT
1
/2–14 NPSM64
3
/4–14 NPT64
Female threads
Options (include with selected model number)
Sensor/thermowell assemble to
options
★
★
Hand tight assembly of sensor
XT
and thermowell
Process-ready assembly of
XW
sensor and thermowell
Details
Ensures sensor is threaded into thermowell but only hand tightened65
Ensures sensor is threaded into thermowell and torqued for process-ready
installation
Ref.
page
64
Ref.
page
65
Extended product warranty
★
WR33-year limited warranty
★
WR55-year limited warranty65
This warranty option extends manufacturer’s warranty to three or five years
for manufacturer related defects
Thermowell calculation
★
R21Thermowell calculationSet of calculations to ensure thermowells are safe in certain process conditions65
NACE certification
★
Q35NACE certificationMeets MR0175/ISO 15156 and MR0103 requirements66
PMI testing
Q76PMI testingVerifies chemical composition of material66
Material certification
★
Q8Material certification
Certificate for material conformance and traceability in accordance with EN
10204 type 3.1
Ref.
page
65
Ref.
page
Ref.
page
Ref.
page
Ref.
page
67
12
Emerson.com/Rosemount
Page 13
June 2019
Table 1. Rosemount 114C Threaded Ordering Information
The starred offerings (★) represent the most common options and should be selected for best delivery. The non-starred offerings are subject
to additional delivery lead time.
Rosemount 114C
Material tests
M01Low temperature Charpy TestMeasures the low temperature ductility of the material67
M02Ultrasonic material testExamination of steel forgings for flaws and inclusions67
R14Finish < Ra 0.3 μm (12 μin)Improves surface roughness of thermowell68
Electropolish
R20ElectropolishImprove smoothness and surface quality68
Hydrostatic external pressure test
★
Q5Standard external pressure test
★
Q9Extended external pressure test Same as standard external pressure test but tested twice as long69
Verifies structural quality and checks for leaks at thermowell process
connection and stem
Hydrostatic internal pressure test
Ref.
page
Ref.
page
Ref.
page
Ref.
page
68
Ref.
page
★
Q85Standard internal pressure testVerify internal structural integrity of thermowell69
★
Q86Extended internal pressure test Same as standard internal pressure test but tested twice as long70
Canadian Registration Number
Q17Canadian Registration Number Canadian approvals for all provinces (Approved materials in reference section)70
Dye penetration test
★
Q73Dye penetration testChecks quality of material70
Wall thickness test
Q83Ultrasonic testChecks the bore concentricity of the thermowell71
Q84Radiography (X-ray) testChecks the bore concentricity of the thermowell72
Special cleaning
Q6Special cleaningOxygen enriched environment cleaning per ASTM G9371
Thermowell markings
Ref.
page
Ref.
page
Ref.
page
Ref.
page
Ref.
page
R40Test markings on thermowell
Emerson.com/Rosemount
External marking of the thermowell for specific tests (see reference page for
list of tests)
71
13
Page 14
Rosemount 114C
A
Table 1. Rosemount 114C Threaded Ordering Information
The starred offerings (★) represent the most common options and should be selected for best delivery. The non-starred offerings are subject
to additional delivery lead time.
June 2019
Plug and chain
R06Stainless steelProtects thermowell threads when sensor is not installed73
R23BrassProtects thermowell threads when sensor is not installed73
Spherical tip
R60Spherical tipChanges the flat tip to spherical72
Vent hole
R11Vent hole
Allows for the venting of a thermowell and for indication that thermowell
structural integrity has been compromised
Thermowells with wrench flats
R37Thermowells with wrench flats
Converts the two wrench flats to hex wrench flats; only applies to exotic
materials
Root diameter (A)
Ref.
page
Ref.
page
Ref.
page
73
Ref.
page
79
Ref.
page
x.xx-in., 0.36 to 3.15-in. in 0.01-in. increments (when ordered with dimension units code E)
xx.xx mm, 10 to 46 mm in 0.5-mm. increments (when ordered with dimension units code M)
Bxxx
Examples: Code B100 = 10.0 mm, Code B455 = 45.5 mm
B
Non-standard bore diameter (d)DetailsImage
D010.276-in./7.0 mm
d
D030.138-in./3.5 mm82
D040.386- in./9.8 mm82
D050.354-in./9.0 mm82
D060.433-in./11.0 mm82
Default = 0.26-in. (6.6 mm)
80
80
Ref.
page
81
81
Ref.
page
82
14
Emerson.com/Rosemount
Page 15
June 2019
t
A
B
C
D
C
Rosemount 114C
Table 1. Rosemount 114C Threaded Ordering Information
The starred offerings (★) represent the most common options and should be selected for best delivery. The non-starred offerings are subject
to additional delivery lead time.
Non-standard tip thickness (t)
Ref.
page
T010.197-in./5.0 mm
Default = 0.25-in. (6.4 mm)
T020.236-in./6.0 mm82
Threaded installation
Threaded thermowells are screwed into the process using a threaded fitting or directly into a tapped pipe if there is sufficient wall
thickness. Tapered threads will deform to each other to create a seal. Thread sealant and appropriate torque should be applied to
reduce risk of leaks.
Figure 5. Installation Components
82
A. Thermowell
B. Threaded fitting
C. Weld
D. Process
Emerson.com/Rosemount
15
Page 16
Rosemount 114C
.75 (19)
2.5 (63.5)
N
P
H
U
d
t
A
B
A
B
C
E
A
Threaded thermowell drawings
June 2019
Figure 6. Thread Mount Thermowell Drawings (Tapered Thread)
StraightTaperedStepped
A. Root diameter
B. Tip diameter
C. Total length (U + H)
E. Wrench allowance
F. Thread allowance
H. Head length
(1)
N. Instrument connection
P. Process connection
U. Immersion length
d. Bore diameter
t. Tip thickness
Table 2. Thread Mount Thermowells (Tapered Thread)
Hex size will be different depending on units selected (english and metric).
Wrench flats are used on exotic materials instead of hex flats. For hex flats on exotic materials, select option R37.
Additional root and tip diameters available.
1.Total length = U+ H.
Emerson.com/Rosemount
17
Page 18
Rosemount 114C
Rosemount 114C Flanged Thermowells
June 2019
18
Emerson.com/Rosemount
Page 19
June 2019
Model
Instrument
connection threads
Head
length (H)
Thermowell
material
Stem
style
Process
connection
Mounting
style
Immersion
length (U)
Units
114CXXXXXXXXXXXXXXX
16151413121198765432110191817
(5) Units
English (E)
Metric (M)
(6-9) Immersion length (U)
XXXX
XXXX
(10) Mounting style
(11-12) Process connections
AA
1-in. Class 150
AB
AC
1½-in. Class 150
2-in. Class 150
(13) Stem style
1
Straight stem
2
3
Tapered stem
Stepped stem
(14-15) Thermowell material
SC
316/316L SST
SF
CS
304/304L SST
Carbon steel
(16-18) Head length (H)
XXX
XXX
(19) Instrument connection
A
½-14 NPT
B
½-14 NPSM
Full penetration weld
F
P
Partial penetration
weld
Options
XX, XXX, XX
Common options
External pressure test
Q5
R21
Q35
Q8
NACE certification
Wake frequency
calculation
Material certification
Q73
Dye penetration test
AH
AJ
1-in. Class 300
1½-in. Class 300
AK
2-in. Class 300
0.5in to 42in (E)
Example: 0.5in=0005,
42in=0420
2.25in to 11.25in (E)
Example: 2.25in=022,
10in=100
G
Forged, no welds
25mm to 1165mm (M)
Example: 25mm=0025,
1165mm=1165
40mm to 225mm (M)
Example: 40mm=040,
225mm=225
DM18 x 1.5p
E
M20 x 1.5p
F
M24 x 1.5p
U
H
A
B
Rosemount 114C
Flanged thermowell overview
All Rosemount flanged thermowells are manufactured in accordance with ANSI B16.5. The flange to stem weld is in accordance to
ASME Section IX. There is also full traceability with material certifications available on request. Rosemount flanged thermowells are
available in two manufacturing configurations: full and partial penetration welds.
Full penetration weld (F)Partial penetration weld (P)Forged, no welds (G)
Stronger weld
joint per ASME
PTC 19.3 TW
Used for heavy
duty
applications
Emerson recommended option
Figure 8. Standard Offering–Flanged
Adequate for
most process
applications
Weld withstands
same pressure
and temperature rating as flange
Lower cost than full penetration weld
Highest fatigue
resistance per
ASME PTC 19.3
TW
Eliminates weld
qualifications and failures
Used in extreme process applications
The common options shown in Figure 8 represent a partial offering; reference the Rosemount 114C Flanged Ordering Information
for a full list of available options.
Figure 9. Flanged Thermowell Components
A. Instrument connection
B. Process connection
Note
Wetted surface includes flange face and immersion length (U).
Emerson.com/Rosemount
H. Head length
U. Immersion length
19
Page 20
Rosemount 114C
June 2019
Use the form below to record your model code:
ModelUnits
Immersion
length (U)
Mounting
style
Process
connection
Stem
style
Thermowell
material
Head length (H)
Instrument
connection
Options
114C
123456 through 91011 and 121314 and 1516 through 1819X X X X X
20
Emerson.com/Rosemount
Page 21
June 2019
U
Flanged ordering information
Figure 10. Model Number Ordering Example
ModelUnits
Immersion
length (U)
Mounting
style
Process
connection
Stem
style
Thermowell
material
Head length (H)
Rosemount 114C
Instrument
connection
Options
114C E 0150FA C1SC0 5 0A
12345 67891011 121314 1516 17 1819X X X X X
The numbers below the model number ordering example correlate to the character place numbers in the second column of the
ordering table.
Table 4. Rosemount 114C Flanged Ordering Information
The starred offerings (★) represent the most common options and should be selected for best delivery. The non-starred offerings are subject
to additional delivery lead time.
Place #s
1-4
★
Place #
★
★
ModelDetails
Made with a standard bore diameter of 0.26-in. (6.6 mm) and tip
114C Barstock temperature thermowell
Dimension units
5
EEnglish units (in.)
MMetric units (mm)59
wall thickness of 0.25-in. (6.4 mm).
Default ASME flange facing is raised face with spiral serrations.
Default EN 1092-1 flange facing is raised face Type B1.
Specifies whether length units will be in inches (in) or
millimeters (mm)
WR5,
Q76...
Ref.
page
N/A
Ref.
page
59
Place #
6-9
★
★
Place #
10
★
★
Immersion length (U)
xx.x-in., 1 to 100-in. in 1/4-in. increments (when ordered with dimension units code E)
xxxx
Example of a 6.25-in. length where the second decimal is dropped off: 0062
xxxx mm, 25 to 2500 mm in 5 mm increments (when ordered with dimension units code M)
xxxx
Example of a 25 mm length: 0025
Mounting style
PFlange, partial penetration weld
Weld refers to welding of the flange to thermowell stem
FFlange, full penetration weldN/A
GFlange, forgedSingle piece forging, no weldsN/A
Ref.
page
59
59
Ref.
page
N/A
Emerson.com/Rosemount
21
Page 22
Rosemount 114C
June 2019
Table 4. Rosemount 114C Flanged Ordering Information
The starred offerings (★) represent the most common options and should be selected for best delivery. The non-starred offerings are subject
to additional delivery lead time.
Place #s
11-12
Process connectionDetails
Ref.
page
Partial weld (P)Full penetration weld (F)Forged, no welds (G)
★
AA1-in. Class 1501-in. Class 1501-in. Class 150N/A
★
AB11/2-in. Class 15011/2-in. Class 15011/2-in. Class 150N/A
★
AC2-in. Class 1502-in. Class 1502-in. Class 150N/A
★
AD3-in. Class 1503-in. Class 1503-in. Class 150N/A
★
AE4-in. Class 1504-in. Class 1504-in. Class 150N/A
★
AF6-in. Class 1506-in. Class 1506-in. Class 150N/A
★
★
★
★
3
AG
/4-in. Class 300
AH1-in. Class 3001-in. Class 3001-in. Class 300N/A
AJ11/2-in. Class 30011/2-in. Class 30011/2-in. Class 300N/A
AK2-in. Class 3002-in. Class 3002-in. Class 300N/A
3
/4-in. Class 300
3
/4-in. Class 300N/A
AL1-in. Class 400/6001-in. Class 400/6001-in. Class 400/600N/A
AM11/2-in. Class 400/60011/2-in. Class 400/60011/2-in. Class 400/600N/A
AN2-in. Class 400/6002-in. Class 400/6002-in. Class 400/600N/A
APN/A1-in. Class 900/15001-in. Class 900/1500N/A
AQN/A11/2-in. Class 900/150011/2-in. Class 900/1500N/A
ARN/A2-in. Class 900/15002-in. Class 900/1500N/A
ATN /A11/2-in. Class 250011/2-in. Class 2500N/A
AUN/A2-in. Class 25002-in. Class 2500N/A
AV3-in. Class 3003-in. Class 3003-in. Class 300N/A
FADN 20/PN 2.5/6DN 20/PN 2.5/6N/AN/A
FEDN 20/PN 10/16/25/40DN 20/PN 10/16/25/40N/AN/A
FGDN 20/PN 63/100DN 20/PN 63/100N/A
N/A
GADN 25/PN 2.5/6DN 25/PN 2.5/6N/AN/A
GEDN 25/PN 10/16/25/40DN 25/PN 10/16/25/40N/AN/A
GGDN 25 PN 63/100DN 25/PN 63/100N/AN/A
JADN 40/PN 2.5/6DN 40/PN 2.5/6N/AN/A
JEDN 40/PN 10/16/25/40DN 40/PN 10/16/25/40N/A
N/A
JGDN 40/PN 63/100DN 40/PN 63/100N/AN/A
KADN 50/PN 2.5/6DN 50/PN 2.5/6N/AN/A
KCDN 50/PN 10/16DN 50/PN 10/16N/AN/A
KEDN 50/PN 25/40DN 50/PN 25/40N/AN/A
KFDN 50/PN 63DN 50/PN 63N/AN/A
22
Emerson.com/Rosemount
Page 23
June 2019
Table 4. Rosemount 114C Flanged Ordering Information
The starred offerings (★) represent the most common options and should be selected for best delivery. The non-starred offerings are subject
to additional delivery lead time.
Place #s
11-12
Process connectionDetails
Partial weld (P)Full penetration weld (F)Forged, no welds (G)
KGDN 50/PN 100DN 50/PN 100N/AN/A
LADN 65/PN 2.5/6DN 65/PN 2.5/6N/AN/A
LCDN 65/PN 10/16DN 65/PN 10/16N/AN/A
LEDN 65/PN 24/40DN 65/PN 24/40N/AN/A
LFDN 65/PN 63DN 65/PN 63N/AN/A
LGDN 65/PN 100DN 65/PN 100N/AN/A
MADN 80/PN 2.5/6DN 80/PN 2.5/6N/AN/A
MCDN 80/PN 10/16DN 80/PN 10/16N/AN/A
MEDN 80/PN 25/40DN 80/PN 25/40N/AN/A
MFDN 80/PN 63DN 80/PN 63N/AN/A
Rosemount 114C
Ref.
page
MGDN 80/PN 100DN 80/PN 100N/AN/A
NADN 100/PN 2.5/6DN 100/PN 2.5/6N/AN/A
NCDN 100/PN 10/16DN 100/PN 10/16N/AN/A
NEDN 100/PN 25/40DN 100/PN 25/40N/AN/A
NFDN 100/PN 63DN 100/PN 63N/AN/A
NGDN 100/PN 100DN 100/PN 100N/AN/A
Place #
13
★
★
Stem styleDetailsImage
1Straight
2Ta p er e d
Minimum immersion length
= 1-in. (25 mm)
Minimum immersion length
= 1-in. (25 mm)
Ref.
page
60
60
★
3Stepped
Emerson.com/Rosemount
Minimum immersion length
= 3-in. (75 mm)
60
23
Page 24
Rosemount 114C
Table 4. Rosemount 114C Flanged Ordering Information
The starred offerings (★) represent the most common options and should be selected for best delivery. The non-starred offerings are subject
to additional delivery lead time.
Place #s
14-15
★
★
★
Thermowell material Details
SC316/316L dual rated61
SD316/316L dual rated (NORSOK)
SF304/304L dual rated61
CSCarbon steel (A-105)61
SG316Ti SST61
SH316/316L SST with tantalum sheath
SJ316/316L SST with PFA coating61
SK304/304L SST with PTFE coating61
SL310 SST61
SM321 SST61
Must order the Q8 Material Certificate to get NORSOK
documentation
Only available as a straight stem profile with a 0.75 diameter thus
requires option A075 for English units
June 2019
Ref.
page
62
62
SN321H SST61
SR904L SST61
SP347 SST61
ABAlloy B361
ACAlloy C-27661
ADAlloy C-4 (with 304/304L SST flange) Not available with G (Fully forged) mounting style61
AEAlloy C-22 (with 304/304L SST flange) Not available with G (Fully forged) mounting style61
AFAlloy C-22 (with 316/316L SST flange) Not available with G (Fully forged) mounting style61
AGAlloy 2061
AHAlloy 40061
AJAlloy 400 (with 304/304L SST flange) Not available with G (Fully forged) mounting style61
AKAlloy 60061
ALAlloy 600 (with 304/304L SST flange) Not available with G (Fully forge) mounting style61
AMAlloy 60161
ANAlloy 62561
APAlloy 80061
AQAlloy 800H/HT61
ARAlloy 82561
AUAlloy C-2061
ASAlloy F44 Mo661
CAChrome-Moly Grade B-11/F-11 Class II61
CBChrome-Moly Grade B-22/ F-22 Class III61
24
Emerson.com/Rosemount
Page 25
June 2019
Table 4. Rosemount 114C Flanged Ordering Information
The starred offerings (★) represent the most common options and should be selected for best delivery. The non-starred offerings are subject
to additional delivery lead time.
CCChrome-Moly Grade F-9161
NKNickel 20061
TTTitanium Grade 261
DSSuper duplex SST Grade F-5361
DTSuper duplex – NORSOK
DUDuplex 2205 Grade F5161
DVDuplex 2205 – NORSOK
Place #s
16-18
★
★
Head length (H)
xx.x-in., 2.25 to 11.25-in. in 1/4-in. increments (when ordered with dimension units code E)
xxx
Example of a 6.25-in. length where the second decimal is dropped off: 062 (default head length = 2.25-in. for
flanges under Class 900)
xxx mm, 45 to 225 mm in 5 mm increments (when ordered with dimension units code M)
xxx
Example of a 50 mm length: 050 (default head length = 60 mm for flanges under Class 900)
Must order the Q8 Material Certificate to get NORSOK
documentation
Must order the Q8 Material Certificate to get NORSOK
documentation
Rosemount 114C
62
62
Ref.
page
63
63
Place #
19
★
★
Instrument connectionDetailsImage
1
A
B
C
DM18 ⫻ 1.5p64
EM20 ⫻ 1.5p64
FM24 ⫻ 1.5p64
GG 1/2-in. (BSPF)64
HG 3/4-in. (BSPF)64
JM27 ⫻ 2p64
KM14 ⫻ 1.5p64
/2–14 NPT
1
/2–14 NPSM64
3
/4–14 NPT64
Female threads
Options (include with selected model number)
Sensor/thermowell assemble to optionsDetails
★
★
Hand tight assembly of sensor and
XT
thermowell
Process-ready assembly of sensor and
XW
thermowell
Ensures sensor is threaded into thermowell but only hand tightened65
Ensures sensor is threaded into thermowell and torqued for
process-ready installation
Ref.
page
64
Ref.
page
65
Emerson.com/Rosemount
25
Page 26
Rosemount 114C
Table 4. Rosemount 114C Flanged Ordering Information
The starred offerings (★) represent the most common options and should be selected for best delivery. The non-starred offerings are subject
to additional delivery lead time.
Extended product warranty
★
WR33-year limited warranty
★
WR55-year limited warranty65
This warranty option extends manufacturer’s warranty to three or
five years for manufacturer related defects
June 2019
Ref.
page
65
Thermowell calculationDetails
Set of calculations to ensure thermowells are safe in certain process
conditions
★
R21Thermowell calculation
Note: If your chosen mounting style is flange with full penetration weld (F) then
you must choose dye penetration option (Q73) to meet the ASME PTC 19.3 TW
standard requirements.
NACE certification
★
Q35NACE certificationMeets MR0175/ISO 15156 and MR0103 requirements66
PMI testing
Q76PMI testingVerifies chemical composition of material66
Material certification
★
Q8Material certification
Certificate for material conformance and traceability in accordance
with EN 10204 type 3.1
Material tests
M01Low temperature Charpy Test
M02Ultrasonic material testExamination of steel forgings for flaws and inclusions67
Measures the low temperature ductility of the material. This option
drives a different weld documentation.
R14Finish < Ra 0.3 μm (12 μin)Improves surface roughness of thermowell68
Electropolish
R20ElectropolishImprove smoothness and surface quality68
Hydrostatic external pressure test
★
Q5Standard external pressure test
★
Q9Extended external pressure testSame as standard external pressure test but tested for twice as long69
Verifies structural quality and checks for leaks at thermowell
process connection and stem
Hydrostatic internal pressure test
★
Q85Standard internal pressure testVerifies internal structural integrity of thermowell69
★
Q86Extended internal pressure testSame as standard internal pressure test but tested for twice as long70
26
Emerson.com/Rosemount
Ref.
page
Ref.
page
Ref.
page
68
Ref.
page
Page 27
June 2019
Table 4. Rosemount 114C Flanged Ordering Information
The starred offerings (★) represent the most common options and should be selected for best delivery. The non-starred offerings are subject
to additional delivery lead time.
Canadian Registration Number
Q17Canadian Registration Number
Canadian approvals for all provinces (Approved materials in
reference section)
Dye penetration test
★
Q73Dye penetration testChecks quality of welds and material70
Rosemount 114C
Ref.
page
70
Ref.
page
Wall thickness test
Q83Ultrasonic testChecks the bore concentricity of the thermowell71
Special cleaning
Q6Special cleaningOxygen enriched environment cleaning per ASTM G9371
Thermowell markings
R40Test markings on thermowell
External marking of the thermowell for specific tests (see reference
page for list of tests)
Phased array test
Q80Phased array test
Verifies the quality of full penetration flange welds using ultrasonic
technology
X-ray/radiograph test
Q81X-ray/radiographVerifies quality of full penetration flange welds72
Flange welding documentation
Q66Procedure Qualification Records (PQR)
Welder Performance Qualifications
Q67
(WPQ)
Q68Welding Procedure Specifications (WPS)
A record of a test weld performed and tested rigorously to ensure
that the procedure will produce good weld.
A test certificate that shows whether a welder possesses the
necessary experience and knowledge to perform the specifications
of a particular weld procedure.
A formal written document that describes and provides direction to
a welder or welding operator for making sound and quality
production welds per code requirement.
Spherical tip
Ref.
page
Ref.
page
Ref.
page
71
Ref.
page
72
Ref.
page
Ref.
page
72
72
72
Ref.
page
R60Spherical tipChanges the flat tip to spherical72
Plug and chain
R06Stainless steel
Protects thermowell threads when sensor is not installsed
R23Brass73
Emerson.com/Rosemount
Ref.
page
73
27
Page 28
Rosemount 114C
B
Table 4. Rosemount 114C Flanged Ordering Information
The starred offerings (★) represent the most common options and should be selected for best delivery. The non-starred offerings are subject
to additional delivery lead time.
Vent hole
R11Vent holeAllows for the venting of a thermowell73
June 2019
Ref.
page
Flange face
R09Concentric serrationsConcentric serrations on flange face per ASME B16.574
R10FlatFlat flange face per ASME B16.5 or EN 1092-1 facing Type A74
R15Raised face, Type B2Raise face per EN 1092-1 facing Type B275
R16RTJRing type joint flange face per ASME B16.576
R18Groove, Type DGroove, Type D per EN 1092-176
R19Tongue, Type CTongue, Type C per EN 1092-177
R24Spigot, Type ESpigot Type E per EN 1092-178
R25Recess, Type FRecess Type F per EN 1092-178
A
Root diameter (A)
x.xx-in., 0.36 to 3.15-in. in 0.01-in. increments (when ordered with dimension units code E)
xx.x mm, 10 to 46 mm in 0.5 mm increments (when ordered with dimension units code M)
Bxxx
Examples: Code B100 = 10.0 mm, Code B455 = 45.5 mm
28
Ref.
page
81
81
Emerson.com/Rosemount
Page 29
June 2019
d
t
e
A
B
C
D
E
Rosemount 114C
Table 4. Rosemount 114C Flanged Ordering Information
The starred offerings (★) represent the most common options and should be selected for best delivery. The non-starred offerings are subject
to additional delivery lead time.
Non-standard bore diameter (d)DetailsImage
Ref.
page
D010.276-in./7.0 mm
D030.138-in./3.5 mm82
D040.386-in./9.8 mm82
Standard = 0.26-in. (6.6 mm)
D050.354-in./9.0 mm82
D060.433-in./11.0 mm82
Non-standard tip thickness (t)DetailsImage
T010.197-in./5.0 mm
Standard = 0.25-in. (6.4 mm)
T020.236-in./6.0 mm82
Fillet Radius (e)DetailsImage
E030.118-in./3 mm
Standard = 0.157-in.
E050.197-in./5 mm
(4 mm)
82
Ref.
page
82
Ref.
page
83
Flanged installation
Flanged thermowells are bolted to a mating flange which protrudes from the process. It is important to select appropriate gasket for
the process conditions, to provide a seal between the flange faces. The Rosemount 114C Thermowells come standard with a raised
face and spiral serrations designed per the ASME B16.5 standard. These should be installed with an inside bolt circle (IBC) gasket/ring
gasket, which extends to and is centered by the bolts. Other flange face options are available.
Figure 11. Installation Components
A. Thermowell
B. Bolt/washers
C. Ring gasket
D. Nozzle and mating flange
E. Process
Emerson.com/Rosemount
29
Page 30
Rosemount 114C
H
A
t
N
d
U
B
B
A
C
E
A
e
Flanged thermowell drawings
June 2019
Figure 12. Flange Mounted Thermowell Drawings
StraightTaperedStepped
(1)
A. Root diameter
B. Tip diameter
C. Lap flange
E. Total length (U + H)
H. Head length
N. Instrument connection
U. Immersion length
d. Bore diameter
e. Fillet radius
t. Tip thickness
1.Total length = U+H.
30
Emerson.com/Rosemount
Page 31
June 2019
Rosemount 114C
Table 5. Flange Mounted Thermowells
Process connection
Code P, flanged,
Code
partial penetration
weld
AA1-in. Class 1501-in. Class 1501-in. Class 15019 (.748)22.5 (.886)16 (.630)19 (.748)
AB11/2-in. Class 15011/2-in. Class 15011/2-in. Class 15021.5 (.846)26.5 (1.043)18 (.709)21.5 (.846)
AC2-in. Class 1502-in. Class 1502-in. Class 15021.5 (.846)26.5 (1.043)18 (.709)21.5 (.846)
AD3-in. Class 1503-in. Class 1503-in. Class 15021.5 (.846)26.5 (1.043)18 (.709)21.5 (.846)
AE4-in. Class 1504-in. Class 1504-in. Class 15021.5 (.846)26.5 (1.043)18 (.709)21.5 (.846)
AF6-in. Class 1506-in. Class 1506-in. Class 15021.5 (.846)26.5 (1.043)18 (.709)21.5 (.846)
AG
AH1-in. Class 3001-in. Class 3001-in. Class 30019 (.748)22.5 (.886)16 (.630)19 (.748)
AK2-in. Class 3002-in. Class 3002-in. Class 30021.5 (.846)26.5 (1.043)18 (.709)21.5 (.846)
AM1
AN2-in. Class 400/6002-in. Class 400/6002-in. Class 400/60021.5 (.846)26.5 (1.043)18 (.709)21.5 (.846)
APN/A1-in. Class 900/15001-in. Class 900/150019 (.748)22.5 (.886)16 (.630)19 (.748)
AQN/A1
ARN/A2-in. Class 900/15002-in. Class 900/150021.5 (.846)26.5 (1.043)18 (.709)21.5 (.846)
AUN/A2-in. Class 25002-in. Class 250021.5 (.846)26.5 (1.043)18 (.709)21.5 (.846)
AV3-in. Class 3003-in. Class 3003-in. Class 30021.5 (.846)26.5 (1.043)18 (.709)21.5 (.846)
42in=0420
25mm to 1165mm (M)
Example: 25mm=0025,
1165mm=1165
2.25in to 11.25in (E)
Example: 2.25in=022,
10in=100
40mm to 225mm (M)
Example: 40mm=040,
225mm=225
D
M18 x 1.5p
E
M20 x 1.5p
F
M24 x 1.5p
June 2019
Van Stone thermowell overview
Van Stone/lap Joint thermowells are mounted between the mating flange and lap joint flange. This unique design enables thermowell
designers to specify thermowell flange materials different than the thermowell stem material; flanges are easily replaceable. These
thermowells allow use of different thermowell materials for the flange contacting the process and overlaying flange which can save
material and manufacturing costs. They are a good choice for corrosive applications, because there are no welds so weld-joint
corrosion is eliminated. The Emerson standard for the Van Stone thermowell is a raised face style made of carbon steel. Other styles
and flange materials are also available.
The standard offering figure below shows the thermowell configurations that can typically be shipped in two weeks or less.
Figure 13. Standard Offering–Van Stone
The common options shown in Figure 13 represent a partial offering; reference the Rosemount 114C Van Stone Ordering Information
for a full list of available options.
Figure 14. Van Stone Thermowell Components
A
A. Instrument connection
B. Process connection
Note
Wetted surface includes flange face and immersion length (U).
34
H
B
U
H. Head length
U. Immersion length
Emerson.com/Rosemount
Page 35
June 2019
Rosemount 114C
Use the form below to record your model code:
ModelUnits
Immersion
length (U)
Mounting
style
Process
connection
Stem
style
Thermowell
material
Head length (H)
Instrument
connection
Options
114CV
123456 through 91011 and 121314 and 1516 through 1819X X X X X
Emerson.com/Rosemount
35
Page 36
Rosemount 114C
U
Van Stone ordering information
Figure 15. Model Number Ordering Example
ModelUnits
Immersion
length (U)
Mounting
style
Process
connection
Stem
style
Thermowell
material
Head length (H)
Instrument
connection
June 2019
Options
114C M 0150VA B1S C0 5 0A
12345 67891011 121314 1516 17 1819X X X X X
The numbers below the model number ordering example correlate to the character place numbers in the second column of the
ordering table.
Table 7. Rosemount 114C Van Stone Ordering Information
★The Standard offering represents the most common options. The starred options (★) should be selected for best delivery lead time.
The Expanded offering is subject to additional delivery lead time.
Place #s
1-4
★
114C Barstock temperature thermowell
Place #
★
EEnglish units (inches)
★
MMetric units (mm)59
ModelDetails
Dimension unitsDetails
5
Made with a standard bore diameter of 0.26-in. (6.6 mm) and tip
wall thickness of 0.25-in. (6.4 mm). Default cover flange material is
carbon steel.
Specifies whether length units will be in inches (in.) or
millimeters (mm)
WR5,
Q76...
Ref.
page
N/A
Ref.
page
59
Place #
6-9
★
xxxx
★
xxxx
Place #
10
★
VVan Stone, lap flangeDefault cover flange material is carbon steelN/A
Place #s
11-12
★
AA1-in. Class 150N/A
★
AB11/2-in. Class 150N/A
★
AC2-in. Class 150N/A
Immersion length (U)
xx.x-in., 1.0 to 100-in. in 1/4-in. increments (when ordered with dimension units code E)
Example of a 6.25-in. length where the second decimal is dropped off: 0062
xxxx mm, 25 to 2500 mm in 5 mm increments (when ordered with dimension units code M)
Example of a 25 mm length: 0025
Mounting styleDetails
Process connectionDetails
Ref.
page
59
59
Ref.
page
Ref.
page
36
Emerson.com/Rosemount
Page 37
June 2019
Table 7. Rosemount 114C Van Stone Ordering Information
★The Standard offering represents the most common options. The starred options (★) should be selected for best delivery lead time.
The Expanded offering is subject to additional delivery lead time.
★
AH1-in. Class 300N/A
★
AJ11/2-in. Class 300N/A
★
AK2-in. Class 300N/A
★
AL1-in. Class 400/600N/A
★
AM11/2-in. Class 400/600N/A
★
AN2-in. Class 400/600N/A
AP1-in. Class 900/1500N/A
AQ11/2-in. Class 900/1500N/A
AR2-in. Class 900/1500N/A
AS1-in. Class 2500N/A
AT11/2-in. Class 2500N/A
AU2-in. Class 2500N/A
Rosemount 114C
Place #
13
★
1Straight
★
2Ta pe r e d
★
3Stepped
Stem styleDetailsImage
Minimum immersion length
= 1-in. (25 mm)
Minimum immersion length
= 1-in. (25 mm)
Minimum immersion length
= 3-in. (75 mm)
Ref.
page
60
60
60
Place #s
14-15
★
SC316/316L dual rated61
SD316/316L dual rated (NORSOK)
★
SF304/304L dual rated61
Thermowell material Details
Must order the Q8 Material Certificate to get NORSOK
documentation
Emerson.com/Rosemount
Ref.
page
62
37
Page 38
Rosemount 114C
Table 7. Rosemount 114C Van Stone Ordering Information
★The Standard offering represents the most common options. The starred options (★) should be selected for best delivery lead time.
The Expanded offering is subject to additional delivery lead time.
★
CSCarbon steel (A-105)61
SG316Ti SST61
SH316/316L SST with tantalum sheath61
SJ316/316L SST with PFA coating61
SK304/304L SST with PTFE coating61
SL310 SST61
SM321 SST61
SN321H SST61
SR904L SST61
SP347 SST61
ABAlloy B361
ACAlloy C-27661
June 2019
AGAlloy 2061
AHAlloy 40061
AKAlloy 60061
AMAlloy 60161
ANAlloy 62561
APAlloy 80061
AQAlloy 800H/HT61
ARAlloy 82561
AUAlloy C-2061
MOMolybdenum61
CAChrome-Moly Grade B-11/F-11 Class II61
CBChrome-Moly Grade B-22/ F-22 Class III61
CCChrome-Moly Grade F-9161
NKNickel 20061
TTTitanium Grade 261
DSSuper duplex SST Grade F-5361
DTSuper duplex – NORSOK
DUDuplex 2205 Grade F5161
DVDuplex 2205 – NORSOK
Must order the Q8 Material Certificate to get NORSOK
documentation
Must order the Q8 Material Certificate to get NORSOK
documentation
62
62
38
Emerson.com/Rosemount
Page 39
June 2019
H
Table 7. Rosemount 114C Van Stone Ordering Information
★The Standard offering represents the most common options. The starred options (★) should be selected for best delivery lead time.
The Expanded offering is subject to additional delivery lead time.
Rosemount 114C
Place #s
16-18
★
xxx
★
xxx
Place #
19
★
A
★
B
C
DM18 ⫻ 1.5p64
EM20 ⫻ 1.5p64
FM24 ⫻ 1.5p64
Head length (H)
xx.x-in., 1.75to 11.25-in. in 1/4-in. increments (when ordered with Dimension units code E)
Example of a 6.25-in. length where the second decimal is dropped off: 062 (default head length = 2.25-in. for
flanges under Class 900)
xxx mm, 40 to 225 mm in 5 mm increments (when ordered with Dimension units code M)
Example of a 50 mm length: 050 (default head length = 60 mm for flanges under Class 900)
Instrument connectionDetailsImage
1
/2–14 NPT
1
/2–14 NPSM64
3
/4–14 NPT64
Female threads
Ref.
page
63
63
Ref.
page
64
GG 1/2-in. (BSPF)64
HG 3/4-in. (BSPF)64
JM27 ⫻ 2p64
KM14 ⫻ 1.5p64
Options (include with selected model number)
Sensor/thermowell assemble to optionsDetails
★
★
Extended product warrantyDetails
★
★
Hand tight assembly of sensor and
XT
thermowell
Process-ready assembly of sensor and
XW
thermowell
WR33-year limited warranty
WR55-year limited warranty65
Ref.
page
Ensures sensor is threaded into thermowell but only hand tightened65
Ensures sensor is threaded into thermowell and torqued for
process-ready installation
65
Ref.
page
This warranty option extends manufacturer’s warranty to three or
five years for manufacturer related defects
65
Emerson.com/Rosemount
39
Page 40
Rosemount 114C
Table 7. Rosemount 114C Van Stone Ordering Information
★The Standard offering represents the most common options. The starred options (★) should be selected for best delivery lead time.
The Expanded offering is subject to additional delivery lead time.
June 2019
Thermowell calculationDetails
★
R21Thermowell calculation
Set of calculations to ensure thermowells are safe in certain process
conditions
NACE certificationDetails
★
Q35NACE certificationMeets MR0175/ISO 15156 and MR0103 requirements66
Ref.
page
65
Ref.
page
PMI testingDetails
Q76PMI testingVerifies chemical composition of material66
Material certificationDetails
★
Q8Material certification
Certificate for material conformance and traceability in accordance
with EN 10204 type 3.1
Material testsDetails
M01Low temperature Charpy TestMeasures the low temperature ductility of the material67
M02Ultrasonic material testExamination of steel forgings for flaws and inclusions67
R14Finish < Ra 0.3 μm (12 μin)Improves surface roughness of thermowell68
ElectropolishDetails
R20ElectropolishImprove smoothness and surface quality68
Stepped instrument threads Image
Ref.
page
Ref.
page
67
Ref.
page
Ref.
page
Ref.
page
Ref.
page
R61Stepped instrument threads 68
Roughened stemDetails
R62Roughened stem
Roughens the last 4.7-in.
(120 mm) of the stem
40
Image
Emerson.com/Rosemount
Ref.
page
N/A
Page 41
June 2019
Table 7. Rosemount 114C Van Stone Ordering Information
★The Standard offering represents the most common options. The starred options (★) should be selected for best delivery lead time.
The Expanded offering is subject to additional delivery lead time.
Rosemount 114C
Hydrostatic external pressure testDetails
★
Q5Standard external pressure test
★
Q9Extended external pressure testSame as standard external pressure test but tested for twice as long69
Verifies structural quality and checks for leaks at thermowell process
connection and stem
Ref.
page
68
Hydrostatic internal pressure testDetails
★
Q85Standard internal pressure testVerifies internal structural integrity of thermowell69
★
Q86Extended internal pressure testSame as standard internal pressure test but tested for twice as long70
Canadian Registration NumberDetails
Q17Canadian Registration Number
Canadian approvals for all provinces (Approved materials in
reference section)
Dye penetration testDetails
★
Q73Dye penetration testChecks quality of welds and material70
Wall thickness testDetails
Q83Ultrasonic testChecks the bore concentricity of the thermowell71
Special cleaningDetails
Q6Special cleaningOxygen enriched environment cleaning per ASTM G9371
Thermowell markingsDetails
R40Test markings on thermowell
External marking of the thermowell for specific tests (see reference
page for list of tests)
Spherical tipDetails
Ref.
page
Ref.
page
70
Ref.
page
Ref.
page
Ref.
page
Ref.
page
71
Ref.
page
R60Spherical tipChanges the flat tip to spherical72
Plug and chainDetails
R06Stainless steel
Protects thermowell threads when sensor is not installed
R23Brass73
Vent holeDetails
R11Vent holeAllows for the venting of a thermowell73
Flange faceDetails
R09Concentric serrationsConcentric serrations on flange face per ASME B16.574
R16RTJRing type joint flange face per ASME B16.575
Emerson.com/Rosemount
Ref.
page
73
Ref.
page
Ref.
page
41
Page 42
Rosemount 114C
d
e
Table 7. Rosemount 114C Van Stone Ordering Information
★The Standard offering represents the most common options. The starred options (★) should be selected for best delivery lead time.
The Expanded offering is subject to additional delivery lead time.
June 2019
A
Root diameter (A)
x.xx-in., 0.36 to 3.15-in. in 0.01-in. increments (when ordered with dimension units code E)
xx.x mm, 10 to 46 mm in 0.5 mm. increments (when ordered with dimension units code M)
Bxxx
Examples: Code B100 = 10.0 mm, Code B455 = 45.5 mm
Non-standard bore diameter (d)DetailsImage
Ref.
page
80
80
Ref.
page
81
81
Ref.
page
D010.276-in./7.0 mm
D030.138-in./3.5 mm82
D040.386-in./9.8 mm82
D050.354-in./9.0 mm82
D060.433-in./11.0 mm82
Standard = 0.26-in.
(6.6 mm)
Non-standard tip thickness (t)Details Image
T010.197-in./5.0 mm
Standard = 0.25-in.
T020.236-in./6.0 mm
(6.4 mm)
t
Fillet Radius (e)Details Image
E030.118-in./3 mm
Standard = 0.157-in.
E050.197-in./5 mm
(4 mm)
82
Ref.
page
82
Ref.
page
83
42
Emerson.com/Rosemount
Page 43
June 2019
f
A
B
C
D
E
Rosemount 114C
Table 7. Rosemount 114C Van Stone Ordering Information
★The Standard offering represents the most common options. The starred options (★) should be selected for best delivery lead time.
The Expanded offering is subject to additional delivery lead time.
Van stone stub thicknessDetails Image
F010.591-in. (15 mm)
Ref.
page
Standard = 0.394-in.
F020.787-in. (20 mm)
(10 mm)
Lap flange material for Van Stone designDetails
C01No flangeProvides a Van Stone stem without a lap flange.84
C02316/316LSST flangeProvides a Van Stone stem with a 316/316LSST lap flange.84
C03Flange per stem material
Van Stone installation
Van Stone thermowells are installed using a lap joint flange
which slips over the stub end of the thermowell. The lap joint
flange has no flange face. Instead the flange is bolted over the
stub end which acts as the flange face and compresses the
gasket. The Rosemount 114C Thermowells come standard with
spiral serrations on the stub end designed per the ASME B16.5
standard. These should be installed with an inside bolt circle
(IBC) gasket/ring gasket, which extends to and is centered by the
bolts. Other flange face options are available.
Provides a Van Stone stem with a matching lap flange per stem
material. Coatings do not apply to lap flange.
Figure 16. Installation Components
83
Ref.
page
84
Emerson.com/Rosemount
A. Thermowell
B. Bolt/washers
C. Ring gasket
D. Nozzle and mating flange
E. Process
43
Page 44
Rosemount 114C
H
t
N
d
U
E
F
G
C
A
e
Van Stone thermowell drawings
June 2019
Figure 17. Van Stone/Lap Flanged Mounted Thermowell Drawings
StraightTaperedStepped
A
B
A. Root diameter
B. Tip diameter
C. Head diameter
E. ASME B16.5 lap flange
F. St ub th ic kn es s
H. Head length
N. Instrument connection (
U. Immersion length
1
/2-in. NPT)
(1)
A
B
d. Bore diameter
e. Fillet radius
t. Tip thickness
1.Total length = U + H.
44
Emerson.com/Rosemount
Page 45
June 2019
Rosemount 114C
Table 8. Van Stone/Lap Flanged Mounted Thermowells
25 to 1165 mm (M)
Example: 25 mm = 0025,
1165 mm = 1165
40 to 225 mm (M)
Example: 40 mm = 040,
225 mm = 225
AC1 1/4-in. Pipe
AD
1½-in. Pipe
Weld in (D)
DM18 x 1.5p
E
M20 x 1.5p
F
M24 x 1.5p
U
H
B
A
Rosemount 114C
Welded thermowell overview
Welded thermowells are permanently welded to process pipes or tanks. Welded thermowells have the highest pressure rating and
are generally used in applications with high velocity flow, high temperature, or extremely high pressure. They are necessary where a
leak-proof seal is required.
The standard offering figure below shows the thermowell configurations that can typically be shipped in two weeks or less.
Figure 18. Standard Offering–Welded
The common options shown in Figure 18 represent a partial offering; reference the Rosemount 114C Welded Ordering Information
for a full list of available options.
Figure 19. Welded Thermowell Components
A. Instrument connection
B. Process connection (dependent on weld point)
Note
Actual wetted surface varies; it is measured from the weld point to the thermowell tip.
Emerson.com/Rosemount
Socket weldWeld-in
H
A
U. Immersion length
H. Head length
B
U
47
Page 48
Rosemount 114C
June 2019
Use the form below to record your model code:
ModelUnits
Immersion
length (U)
Mounting
style
Process
connection
Stem
style
Thermowell
material
Head length (H)
Instrument
connection
Options
114C
123456 through 91011 and 121314 and 1516 through 1819X X X X X
48
Emerson.com/Rosemount
Page 49
June 2019
U
U
Welded ordering information
Figure 20. Model Number Ordering Example
ModelUnits
Immersion
length (U)
Mounting
style
Process
connection
Stem
style
Thermowell
material
Head length (H)
Rosemount 114C
Instrument
connection
Options
114C E 0060WA B1SC0 5 0A
12345 67891011 121314 1516 17 1819X X X X X
WR5,
Q76...
The numbers below the model number ordering example correlate to the character place numbers in the second column of the
ordering table.
Table 9. Rosemount 114C Welded Ordering Information
★The Standard offering represents the most common options. The starred options (★) should be selected for best delivery lead time.
The Expanded offering is subject to additional delivery lead time.
Place #s
1-4
★
114C Barstock temperature thermowell
Place #
★
EEnglish units (inches)
★
MMetric units (mm)59
Place #
6-9
ModelDetails
Dimension unitsDetails
5
Immersion length (U)
Made with a standard bore diameter of 0.26-in. (6.6 mm) and tip
wall thickness of 0.25-in. (6.4 mm)
Specifies whether length units will be in inches (in.) or
millimeters (mm)
Ref.
page
N/A
Ref.
page
59
Ref.
page
★
xxxx
Example of a 6.25-in. length where the second decimal is dropped off: 0062
xxxx mm, 25 to 2500 mm in 5 mm increments (when ordered with dimension units code M)
xxx-in., 1 to 100-in. in 1/4-in. increments (when ordered with dimension units code E)
★
xxxx
Example of a 50 mm length: 0050
Place #
10
★
WWelded–socket weldN/A
★
DWelded–weld-in (only available in tapered stem profile)N/A
Place #s
11-12
Mounting styleDetails
Process connectionsDetails
Welded–socket weld (W)Welded–weld-in (D) (only available in tapered stem profile)
★
★
★
★
3
AA
/4-in. pipe
AB1-in. pipe1-in. pipeN/A
AC11/4-in pipe11/4-in. pipeN/A
AD11/2-in. pipe 11/2-in. pipe N/A
AEN/A
3
/4-in. pipe N/A
Custom diameters (required for Root [Axxx] and tip [Bxxx]
modifications)
59
59
Ref.
page
Ref.
page
N/A
Emerson.com/Rosemount
49
Page 50
Rosemount 114C
Table 9. Rosemount 114C Welded Ordering Information
★The Standard offering represents the most common options. The starred options (★) should be selected for best delivery lead time.
The Expanded offering is subject to additional delivery lead time.
DAN/ADIN 43772-4-7 (18 h7/3.5 mm bore/M14)N/A
DBN/ADIN 43772-4-7 (24 h7/7.0 mm bore/M18)N/A
DCN/ADIN 43772-4-7 (26 h7/7.0 mm bore/G1/2 or M20)N/A
DDN/ADIN 43772-4-7 (26 h7/9.0 mm bore/G1/2 or M20)N/A
DEN/ADIN 43772-4-7 (32 h11/11.0 mm bore/G3/4 or M27)N/A
DHN/A
Place #
13
Stem styleDetailsImage
Custom diameters (required for root (Axxx) and tip ( Bxxx)
modifications
June 2019
N/A
Ref.
page
★
1Straight
★
2Ta p er e d
★
3Stepped
Place #s
14-15
★
SC316/316L dual rated61
SD316/316L dual rated (NORSOK)
★
SF304/304L dual rated61
★
CSCarbon steel (A-105)61
SG316Ti SST61
SL310 SST61
SM321 SST61
SN321H SST61
SR904L SST61
Thermowell material Details
Minimum immersion length =
1-in. (25 mm)
Minimum immersion length =
1-in. (25 mm)
Minimum immersion length =
3-in. (75 mm)
Must order the Q8 Material Certificate to get NORSOK
documentation
60
60
60
Ref.
page
62
SP347 SST61
ABAlloy B361
ACAlloy C-27661
AGAlloy 2061
AHAlloy 40061
AKAlloy 60061
AMAlloy 60161
ANAlloy 62561
APAlloy 80061
AQAlloy 800H/HT61
50
Emerson.com/Rosemount
Page 51
June 2019
H
Rosemount 114C
Table 9. Rosemount 114C Welded Ordering Information
★The Standard offering represents the most common options. The starred options (★) should be selected for best delivery lead time.
The Expanded offering is subject to additional delivery lead time.
ARAlloy 82561
AUAlloy C-2261
ASAlloy F44 Mo661
MOMolybdenum61
CAChrome-Moly Grade B-11/F-11 Class II61
CBChrome-Moly Grade B-22/ F-22 Class III61
CCChrome-Moly Grade F-9161
NKNickel 20061
TTTitanium Grade 261
DSSuper duplex SST Grade F-5361
DTSuper duplex – NORSOK
Must order the Q8 Material Certificate to get NORSOK
documentation
DUDuplex 2205 Grade F5161
DVDuplex 2205 – NORSOK
Place #s
16-18
Head length (H)
Must order the Q8 Material Certificate to get NORSOK
documentation
H
62
62
Ref.
page
★
xxx
Example of a 6.25-in. length where the second decimal is dropped off: 062 (default head length = 1.75-in.)
xxx mm, 40 to 225 mm in 5-mm increments (when ordered with dimension units code M)
xx.x-in., 1.75 to 11.25-in. in 1/4-in. increments (when ordered with dimension units code E)
★
xxx
Example of a 50 mm length: 050 (default head length = 45 mm)
Place #
19
★
A
★
B
C
Instrument connectionDetailsImage
1
/2–14 NPT
1
/2–14 NPSM64
3
/4–14 NPT64
DM18 ⫻ 1.5p64
EM20 ⫻ 1.5p64
Female threads
FM24 ⫻ 1.5p64
GG 1/2-in. (BSPF)64
HG 3/4-in. (BSPF)64
JM27 ⫻ 2p64
KM14 ⫻ 1.5p64
63
63
Ref.
page
64
Emerson.com/Rosemount
51
Page 52
Rosemount 114C
Table 9. Rosemount 114C Welded Ordering Information
★The Standard offering represents the most common options. The starred options (★) should be selected for best delivery lead time.
The Expanded offering is subject to additional delivery lead time.
June 2019
Options (include with selected model number)
Sensor/thermowell assemble to optionsDetails
★
Hand tight assembly of sensor and
XT
thermowell
Sensor is threaded into thermowell but only hand tightened65
Extended product warrantyDetails
★
WR33-year limited warranty
★
WR55-year limited warranty65
This warranty option extends manufacturer’s warranty to three or
five years for manufacturer related defects
Thermowell calculationDetails
★
R21Thermowell calculation
Set of calculations to ensure thermowells are safe in certain
process conditions
NACE certificationDetails
★
Q35NACE certificationMeets MR0175/ISO 15156 and MR0103 requirements66
PMI testingDetails
Q76PMI testingVerifies chemical composition of material66
Material certificationDetails
★
Q8Material certification
Certificate for material conformance and traceability in
accordance with EN 10204 type 3.1
Material testsDetails
Ref.
page
Ref.
page
65
Ref.
page
65
Ref.
page
Ref.
page
Ref.
page
67
Ref.
page
M01Low temperature Charpy TestMeasures the low temperature ductility of the material67
M02Ultrasonic material testExamination of steel forgings for flaws and inclusions67
R14Finish < Ra 0.3μm (12μin)Improves surface roughness of thermowell68
ElectropolishDetails
R20ElectropolishImproves smoothness and surface quality68
Hydrostatic internal pressure testDetails
★
Q85Standard internal pressure testVerifies internal structural integrity of thermowell69
★
Q86Externded internal pressure test
52
Same as standard internal pressure test but tested for twice as
long
Emerson.com/Rosemount
Ref.
page
Ref.
page
Ref.
page
70
Page 53
June 2019
A
B
Table 9. Rosemount 114C Welded Ordering Information
★The Standard offering represents the most common options. The starred options (★) should be selected for best delivery lead time.
The Expanded offering is subject to additional delivery lead time.
Rosemount 114C
Canadian Registration NumberDetails
Q17Canadian Registration Number
Canadian approvals for all provinces (Approved materials in
reference section)
Dye penetration testDetails
★
Q73Dye penetration testChecks quality of welds and material70
Ref.
page
70
Ref.
page
Wall thickness testDetails
Q83Ultrasonic testChecks the bore concentricity of the thermowell71
Special cleaningDetails
Q6Special cleaningOxygen enriched environment cleaning per ASTM G9371
Thermowell markingsDetails
R40Test markings on thermowell
External marking of the thermowell for specific tests (see
reference page for list of tests)
Spherical tipDetails
R60Spherical tipChanges the flat tip to spherical72
Plug and chainDetails
R06Stainless steel
R23Brass73
Protects thermowell threads when sensor is not installed
Vent holeDetails
Ref.
page
Ref.
page
Ref.
page
71
Ref.
page
Ref.
page
73
Ref.
page
R11Vent holeAllows for the venting of a thermowell73
Root diameter (A)
x.xx-in., 0.36 to 3.15-in. in 0.01-in. increments (when ordered with dimension units code E)
xx.xx mm, 10 to 46 mm in 0.5 mm. increments (when ordered with dimension units code M)
Bxxx
Examples: Code B100 = 10.0 mm, Code B455 = 45.5 mm
Emerson.com/Rosemount
Ref.
page
80
80
Ref.
page
81
81
53
Page 54
Rosemount 114C
A
B
C
D
E
F
June 2019
Table 9. Rosemount 114C Welded Ordering Information
★The Standard offering represents the most common options. The starred options (★) should be selected for best delivery lead time.
The Expanded offering is subject to additional delivery lead time.
Non-standard bore diameter (d)DetailsImage
Ref.
page
D010.276-in./7.0 mm
D030.138-in./3.5 mm82
D040.386-in./9.8 mm82
Standard = 0.26-in. (6.6 mm)
d
82
D050.354-in./9.0 mm82
D060.433-in./11.0 mm82
Non-standard tip thickness (t)DetailsImage
T010.197-in./5.0 mm
Ref.
page
82
Standard = 0.25-in. (6.4 mm)
T020.236-in./6.0 mm82
t
Socket weld installation
Socket weld thermowells are typically welded into a socket weld fitting. Welds should be designed according to the appropriate
standards. It is important to order a head length (H) that leaves enough space so the instrument threads will not be deformed by
welding at installation. The customer should also make sure the root diameter of the thermowell will fit through the inner diameter of
the weld fitting.
Note
When specified in a thermowell calculation, the unsupported length for a socket weld thermowell is from the point of weld (B shown
on Figure 21 on page 54) to the thermowell tip.
Figure 21. Installation Components
A. Thermowell
B. Weld
C. Socket weld fitting
1
D.
/16-in. gap
E. Weld
F. Process
54
Emerson.com/Rosemount
Page 55
June 2019
Rosemount 114C
Weld-in Type 4 thermowells according to DIN 43772
This section only defines the requirement necessary to provide a Type 4 thermowell according to the DIN 43772 Standard (for
ordering information on weld-in thermowells outside the DIN Standard, see Table 9 on page 49).
The illustration below shows the breakdown of a model according to the DIN Standard:
AECustomSpecified by design modifier “AXXX”Specified by design modifier “BXXX”
3
/4-in. pipe1.050 (26.67).748 (19)
1
/4-in. pipe1.660 (42.16)1.043 (26.5)
1
/2-in. pipe1.900 (48.26)1.250 (31.75)
1.Dimensions are in inches (millimeters).
1.Total length = U + H.
58
Emerson.com/Rosemount
Page 59
June 2019
U
Rosemount 114C
Ordering information detail
Dimension units
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
The Rosemount 114C Thermowell has the flexibility to be specified in either inches (E) or millimeters (M).
English units (inches)
If English is selected, all lengths will be in inches.
Metric
If metric is selected, all lengths will be in millimeters.
Immersion length (U)
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
The immersion length normally refers to the length of the thermowell stem beginning underneath the process connection to the tip
of the thermowell. This length is typically specified by the process designer but the general rule is at least one-third or one-half the
pipe diameter. Thermowells longer than 42-in. will be required to have an internal pressure test (Q85) performed to ensure the
internal cavity integrity has not been compromised. Parallel thread thermowells have a U length that actually includes the process
threads thus requiring an extra 1-in. (25 mm) for min. U length.
Table 16. Minimum Immersion Length by Profile Style
Profile
Straight1-in. (25 mm)2-in. (50 mm)
Tapered1-in. (25 mm)2-in. (50 mm)
Stepped3 -in. (75 mm)4-in. (100 mm)
Note
Long-length thermowells are those longer than 42-in. (1065 mm) and may be manufactured from two or three pieces of bar stock.
Minimum length
Emerson.com/Rosemount
Minimum length
(parallel threads)
59
Page 60
Rosemount 114C
June 2019
Stem style
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
Straight style thermowells (1)
Straight style thermowells have the same diameter along the entire immersion length. They
present the largest profile to the process medium and have the highest drag force compared
to other styles with the same root diameter. Because of the large tip diameter, there is more
mass to heat which slows the thermal response of the measurement assembly. The minimum
immersion length (U) allowed with this profile is 1-in. (25 mm) except for parallel threaded thermowells that have a minimum
immersion of 2-in. (50 mm).
Tapered style thermowells (2)
Tapered style thermowells have an outside diameter that decreases uniformly from root to
tip. For the same root diameter, this design represents a good compromise between straight
and stepped thermowells. It’s drag will be less than a straight style, but greater than a
stepped style. The response time will be faster than a straight style and slower than a stepped
style. The two general forms of a tapered stem are uniform (tapered from root to tip) and non-uniform (straight portion followed by
tapered portion). Because of its profile shape, it is a good compromise for strength between the two other styles. It is the common
choice for high velocity flow applications where the flow forces typically are too great to use a stepped well. The tapered design has
faster response than the straight style offering an optimal balance of strength and response time factors. The minimum immersion
length (U) allowed with this profile is 1-in. (25 mm) except for parallel threaded thermowells that have a minimum immersion of 2-in.
(50 mm). The maximum length allowed for the tapered profile is; Immersion length (U) + Head length (H) <= 4-in. (1067 mm)
Stepped style thermowells (3)
Stepped style thermowells have two straight sections with the smaller diameter straight
section at the tip. For the same root diameter as a straight profile thermowell, this design has
less profile exposure to the flowing process and exhibits less drag force and quicker response
time due to the smaller mass at the tip. In general, stepped thermowells will have thinner
walls. By the geometry of its design, the stepped well has a higher natural frequency than the other styles with the same root
diameter, and is less susceptible to vibration induced failure. Since this design has less material at the tip, its considered the best
thermowell for fast time response. The minimum immersion length (U) allowed with this profile is 3-in. (75 mm) except for parallel
threaded thermowells that have a minimum immersion of 4-in. (100 mm).
60
Emerson.com/Rosemount
Page 61
June 2019
Rosemount 114C
Thermowell material
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
The material of construction is typically the first consideration in choosing a thermowell for any given application. Three factors
affect the choice of material:
1. Chemical compatibility with the process media to which the thermowell will be exposed.
2. Temperature limits of the material
3. Compatibility with the process piping material to ensure solid, non-corroding welds and junctions.
It is important the thermowell conforms to the design specs of the pipe or vessel it will be inserted into to ensure structural and
material compatibility. The original process design most likely included temperature, pressure, and corrosive considerations as well
as cleaning procedures, agency approvals required, and conformance with codes or standards. Since an installed thermowell
essentially becomes part of the process, these original design considerations also apply to the thermowell and will drive the
thermowell material of construction and mounting type selection. International pressure vessel codes are explicit about the types of
materials and methods of construction allowed.
Table 17. Thermowell Materials
CodeThermowell materialFlange materialCode Thermowell materialFlange material
SC
316/316L SST dual rated (NORSOK)
(1)
SD
ASTM A479 NORSOK M-630 MDS D57
SF
SG
316/316L SST with Tantalum sheath
(2)
SH
SJ
SK
SL
SM
SN
SR
316/316L SST
UNS S31600/S31603
ASTM A479
UNS S31600/S31603
304/304L SST
UNS S30400/S30403
ASTM A479
316T i SST
UNS S31635
ASTM A479
UNS S31600/S31603
ASTM A479
316/316L SST w / PFA coating
UNS S31600/S31603
ASTM A479
304/304L SST with PTFE coating
UNS S30400/S30403
ASTM A479
310 SST
UNS S31008
ASTM A479
321 SST
UNS S32100
ASTM A479
321H SST
UNS S32109
ASTM A479
DIN 1.4
904L SST
UNS N08904
ASTM A479
DIN 1.4539
316/316L SST
UNS S31600/S31603
ASTM A182 or A240
(1)
316/316L SST dual rated (NORSOK)
UNS S31600/S31603
ASTM A182 NORSOK M-630 MDS D54
304/304L SST
UNS S30400/S30403
ASTM A182 or A240
316T i SST
UNS S31635
ASTM A182
316/316L SST with Tantalum sheath
UNS S31600/S31603
ASTM A182 or A240
Tantalum sheath UNS R05252AF
316/ 316L SST w / PFA coating
UNS S31600/S31603
ASTM A182 or A240
304/304L SST with PTFE coating
UNS S30400/S30403
ASTM A182 or A240
310 SST
UNS S31008
ASTM A182 or A240
321 SST
UNS S32100
ASTM A182 or A240
321H SST
UNS S32109
ASTM A182 or A240
DN 1.4878
904L SST
UNS N08904
ASTMor A240
DIN 1.4539
AM
SP
AB
AC
AD
AE
AG
AH
AJ
AK
AL
347 SST
UNS S34700
ASTM A479
DIN 1.4550
Alloy B3
UNS N10001
ASTM B335
Alloy C-276
UNS N10276
ASTM B574
Alloy C-4
UNS N06455
ASTM B574
Alloy C-22
UNS N06022
ASTM B574
Alloy C-22
UNS N06022
ASTM B574
Alloy 20
UNS N08020
ASTM B473
Alloy 400
UNS N04400
ASTM B164
Alloy 400
UNS N04400
ASTM B164
Alloy 600
UNS N06600
ASTM B166
Alloy 600
UNS N06600
ASTM B166
Alloy 601
UNS N06601
ASTM B166
DIN 2.4851
347 SST
UNS S34700
ASTM A182 or A240
DIN 1.4550
Alloy B3
UNS N10001
ASTM B333
Alloy C-276
UNS N10276
ASTM B462 or B575
304/304L SST
UNS S30400/S30403
ASTM A182 or A240
304/304L SST
UNS S30400/S30403
ASTM A182 or A240
316/316L SST
UNS S31600/S31603
ASTM A182 or A240
Alloy 20
UNS N08020
ASTM B462 or B463
Alloy 400
UNS N04400
ASTM B564 or B127
304/304L SST
UNS S30400/S30403
ASTM A182 or A240
Alloy 600
UNS N06600
ASTM B564 or B168
304/304L SST
UNS S30400/S30403
ASTM A182 or A240
Alloy 601
UNS N06601
ASTM B168 or B564
DIN 2.4851
Emerson.com/Rosemount
61
Page 62
Rosemount 114C
June 2019
Table 17. Thermowell Materials
CodeThermowell materialFlange materialCode Thermowell materialFlange material
AN
AP
AQ
AR
AU
AS
CS
TT
DS
Alloy 625
UNS N06625
ASTM B446
DIN 2.4856
Alloy 800
UNS N08800
ASTM B408
DIN 1.4876
Alloy 800H/HT
UNS N08811
ASTM B408
DIN 1.4959
Alloy 825
UNS N08825
ASTM B425
DIN 2.4858
Alloy C-22
UNS N06022
ASTM B574
DIN 2.4602
Alloy F44 Mo6
UNS S31254
ASTM A479
DIN 1.4547
Carbon steel
UNS K03504
ASTM A105
Titanium grade 2
UNS R50400
ASTM B348 GR 2
Super duplex
UNS 32750
ASTM A479 GR F53
Alloy 625
UNS N06625
ASTM B443 or B564
DIN 2.4856
Alloy 800
UNS N08800
ASTM B409 or B564
DIN 1.4876
Alloy 800H/HT
UNS N08811
ASTM B409 or B564
DIN 1.4959
Alloy 825
UNS N08825
ASTM B424 or B564
DIN 2.4858
Alloy C-22
UNS N06022
ASTM B567 or B575
DIN 2.4602
Alloy F44 Mo6
UNS S31254
ASTM A182 or A240
DIN 1.4547
Carbon steel
ASTM A105, A216 GR WCB, or A515 GR
UNS K03504
70
Titanium grade 2
UNS R50400
ASTM B381 GR 2
Super duplex
UNS 32750
ASTM A182 GR F53 or A240
Duplex 2205
UNS 31803
ASTM A479 GR F51
Chrome-Moly Grade F-91
UNS K90901
ASTM A182
Nickel 200
UNS N02200
ASTM B160
Molybdenum
UNS R03600
ASTM B387
Chrome-Moly Grade B-11
UNS K11797
ASTM A739 GR B-11
Chrome-Moly Grade B-22
UNS K21390
ASTM A739 GR B-22
Super duplex (NORSOK)
UNS 32750
AS TM A 47 9 GR F53 NO RSO K
M-630 MDS D57
Duplex 2205 (NORSOK)
UNS 31803
AS TM A 47 9 GR F51 NO RSO K
M-630 MDS D47
Molybdenum
UNS R03600
ASTM B387
DIN 1.5415 EN 10273
DT
DV
DU
CC
NK
MD
CA
CB
MO
(1)
(1)
Duplex 2205
UNS 31803
ASTM A182 GR F51or A240
Chrome-Moly Grade F-91
ASTM A182 GR F-9, A217 GR C12A,
ASTM A182 GR F-11 CL2 or A387
ASTM A182 GR F-22 CL3, A217 GR
ASTM A182 GR F53 NORSOK M-630
ASTM A182 GR F51 NORSOK M-630
UNS K90901
or A387 GR 91 CL2
Nickel 200
UNS N02200
ASTM B162 or B564
Molybdenum
UNS R03630
ASTM A204 or B386
Chrome-Moly Grade F-11
UNS K11572
GR11 CL2
Chrome-Moly Grade F-22
UNS K21590
WC9, or A387 GR22 CL2
Duplex 2205 (NORSOK)
UNS 31803
MDS D54
Duplex 2205 (NORSOK)
UNS 31803
MDS D44
Molybdenum
UNS R03630
ASTM A204 or B386
DIN 1.5415 EN 10222-2
1.Material supplier qualified per NORSOK M-650; material qualified per NORSOK M-630.
2.Sheath thickness = 0.01-in. (0.38 mm).
NORSOK
Rosemount 114 Thermowell ordered with NORSOK will have raw material from a NORSOK M-650 approved supplier, material
approver per NORSOK M-630 datasheet, and flange welding qualified to NORSOK M-601. The NORSOK material will also meet
requirements of NACE MR0175/ISO 15156.
Q8 should be ordered to receive the MTR. The MTR will come with an M-650 Qualification Test Record (QTR) cover sheet. Additional
testing required by the M-630 datasheet will be included in the MTR.
Some of the required testing for NORSOK Duplex in addition to ASTM requirements;
Micrographic examination at 400 to 500⫻ magnification
Ferrite content analysis according to ASTM E 562 or by image analysis according to ASTM E 1245. Ferrite content shall be within
35–55 percent.
Charpy V-notch testing according to ASTM A 370 at –46 °C. Minimum absorbed energy shall be 45 J average and 35 J single.
Corrosion testing according to ASTM G 48 Method A. No pitting at 20⫻ magnification; Weight loss shall be less than
2
.
4 g/m
Full details of the material requirements can be found in the ASTM and NORSOK M-630 standards.
62
Emerson.com/Rosemount
Page 63
June 2019
H
Rosemount 114C
Head length (H)
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
Head length is the distance from the bottom of the process connection to the top of the thermowell. Each style has a minimum head
length; the length specified must meet or exceed that minimum. It is shown below for all process connection styles.
Note
The industry standard minimum head length for flanged and Van Stone thermowells with connections under Class 900 (ASME B16.5)
or PN 100 (EN 1092-1) is 2.25-in. (60 mm).
Table 18. Recommended Minimum Head Length
Process connection
Threaded
Welded
1.Dimensions are in inches (millimeters).
Minimum head length (H)
1.75 (45)
Table 19. Recommended Minimum Head Length by Connection Class for ASME B16.5
Connection size
(1)
(1)
Connection class
Flanged150300400/600900/15002500
3
/4N/A1.75 (45)N/AN/AN/A
11.75 (45)2.00 (50)2.00 (50)2.50 (65)N/A
1
1
/21.75 (45)2.00 (50)2.00 (50)2.50 (65)3.00 (75)
21.75 (45)2.00 (50)2.00 (50)2.75 (70)3.25 (80)
32.00 (50)N/AN/AN/AN/A
42.00 (50)N/AN/AN/AN/A
62.00 (50)N/AN/AN/AN/A
Flanged with RTJ150300400/600900/15002500
3
/4N/A2.00 (50)N/AN/AN/A
11.75 (45)2.00 (50)2.00 (50)2.50 (65)N/A
1
1
/22.00 (50)2.00 (50)2.00 (50)2.50 (65)3.25 (80)
22.00 (50)2.00 (50)2.00 (50)2.75 (70)350 (85)
32.25 (60)N/AN/AN/AN/A
42.25 (60)N/AN/AN/AN/A
62.25 (60)N/AN/AN/AN/A
Emerson.com/Rosemount
63
Page 64
Rosemount 114C
June 2019
Table 19. Recommended Minimum Head Length by Connection Class for ASME B16.5
Connection size
Connection class
(1)
Van Stone150300400/600900/15002500
11.75 (45)1.75 (45)1.75 (45)2.00 (50)2.25 (60)
1
1
/21.75 (45)1.75 (45)1.75 (45)2.25 (60)2.75 (70)
21.75 (45)1.75 (45)2.00 (50)2.75 (70)3.25 (80)
Van Stone with RTJ150300400/600900/15002500
11.75 (45)1.75 (45)2.25 (60)2.25 (60)2.50 (65)
1
1
/21.75 (45)2.00 (50)2.00 (50)2.50 (65)3.00 (75)
21.75 (45)2.00 (50)2.25 (60)3.00 (75)3.50 (90)
1.Dimensions are in inches (millimeters).
Table 20. Recommended Minimum Head Length by Connection Class for EN 1092-1
Connection size
Connection class
(1)
FlangedPN 2.5/6PN 10/16PN 25/40PN 63PN 100
DN 20404550
DN 25404550
DN 40404550
DN 5040455560
DN 654045505560
DN 804045505560
DN 1004045505560
1.Dimensions are in millimeters.
Instrument connection
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
ThreadSpecificationInternal thread
1
/2–14 NPTSAE-AS 71082
1
/2–14 NPSMASME B1.20.1, 8 threads minimum
3
/4–14 NPTSAE-AS 71082
M18 ⫻ 1.5p
M20 ⫻ 1.5p
M24 ⫻ 1.5p
M27 ⫻ 2p
M14 ⫻ 1.5p
1
G
/2-in. (BSPF)ISO 228/1 (BS 2779)
G3/4-in. (BSPF)ISO 228/1 (BS 2779)
BS 3643
64
Emerson.com/Rosemount
Page 65
June 2019
Rosemount 114C
Sensor/thermowell assemble to options (XT, XW)
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
XT
This option is selected when a Rosemount 214C Sensor is ordered with the Rosemount 114C Thermowell. This ensures the sensor is
threaded into the thermowell, but only hand tightened.
XW
This option is selected when a Rosemount 214C Sensor is ordered with the Rosemount 114C Thermowell. This ensures the sensor is
threaded into the thermowell and torqued for a process-ready installation.
Extended product warranty (WR3, WR5)
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
The extended product warranty options are available in three or five year coverage plans. In the model string, order option codes
WR3 for a three year extended warranty or WR5 for a five year warranty. This coverage is an extension of the manufacturer's limited
warranty and states that the goods manufactured or services provided by seller will be free from defects in materials or workmanship
under normal use and care until the expiration of the applicable warranty period.
Thermowell calculation (R21)
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
The ASME PTC 19.3TW is internationally recognized as a mechanical design standard yielding reliable thermowell service in a wide
range of temperature measurement applications. It includes evaluation of stresses applied to a barstock thermowell as installed in a
process based on the design, material, mounting method, and process conditions. The documentation provided will detail the
process information, thermowell geometry, and comprehensive calculation analysis. It will also provide an acceptable or
unacceptable statement based on the analysis.
There are four quantitative criteria in ASME PTC 19.3 TW for a thermowell to be found acceptable for a particular set of process
conditions:
Frequency Limit: the resonant frequency of the thermowell must be sufficiently high so that destructive oscillations are not
excited by the fluid flow.
Dynamic Stress Limit: the maximum primary dynamic stress must not exceed the allowable fatigue stress limit. If the design
requires that the thermowell pass through the in-line resonance to get to the operating conditions, there is an additional fatigue
check at resonance.
Static Stress Limit: the maximum steady-state stress on the thermowell must not exceed the allowable stress, as determined by
the Von Mises criteria.
Hydrostatic Pressure Limit: the external pressure must not exceed the pressure ratings of the thermowell tip, shank, and flange
(or threads).
Emerson.com/Rosemount
65
Page 66
Rosemount 114C
In addition, the suitability of the thermowell material for the process environment must be considered. This means the designer must
evaluate how corrosion and erosion affects the thermowell as well as how exposure to the process conditions affects material
properties.
June 2019
For detailed information about this standard, refer to the Thermowell Calculations White Paper
should have a wake frequency calculation performed to ensure they are suitable for the process conditions in their application.
Emerson assumes that the customer has either done their own calculations or understands the risks of not having calculations done if
this option is not requested.
. Emerson advises that all thermowells
NACE certification (Q35)
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
This option certifies that thermowell materials used are compliant to NACE MR0175/ISO 15156 and NACE MR0103. The material
certification provided will list compliance to the referenced standard.
Material code
SC316/316L Dual Rated
SF304/304L Dual Rated
SL310 SST
SM321 SST
SN321H SST
SP347 SST
SR904L SST
ABAlloy B3
ACAlloy C-276
AGAlloy 20
AHAlloy 400
NACE certified material
Material code
AKAlloy 600
AMAlloy 601
ANAlloy 625
ARAlloy 800
AQAlloy 800H/HT
ARAlloy 825
ASAlloy F44 Mo6
AUAlloy C-22
CAChrome-Moly Grade B-11/F-11 Class II
CBChrome-Moly Grade B-22/ F-22 Class III
NACE certified material
PMI testing (Q76)
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
Positive Material Identification (PMI) is a test that verifies the thermowell material is as specified by the Rosemount 114C model code.
X-ray/radiograph fluorescence (XRF) is used to provide elemental analysis in a nondestructive manner. The certificate will provide
PMI results in comparison with the applicable material standards for each individual thermowell and state the reference standard.
Two points are provided on flanges. All other thermowell components (including welds) will have a single point. XRF will not detect
carbon in steels. PM I can be marke d on the thermowell by choosing option R40. Due to type of technology used carbon steel material
is exempt from this testing.
66
Emerson.com/Rosemount
Page 67
June 2019
Rosemount 114C
Material certification (Q8)
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
Material certificate and traceability in accordance with EN 10204 Type 3.1 Inspection Certificate. The certificate provided will
document the heat code, chemical analysis, and testing required by material standards.
Low temperature Charpy test (M01)
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
Test is performed in accordance with ASTM A370 and report will be include in the Material Traceability Report (Q8). This report must
be ordered if any documentation is required. Charpy test will be done to check toughness of the raw bar and flange material used for
the construction of the thermowell. The table below shows the material available with the option, test temperature, and acceptance
criteria.
Minimum: 55 J (41 ft-lb)
SJ – 316/316L with PFA coating
SK – 304/304L with PTFE coating
SM – 321 SST
When ordering this option with any weld documentation (Q66-Q68), note that there is a separate set of documents needed to meet
the test requirements. There are Q66-68 weld documents specifically for M01 tested material.
Ultrasonic material test (M02)
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
Ultrasonic examination will be done to check quality of the raw bar and flange material used for thermowell construction. The testing
shall be performed in accordance with procedures specified in ASTM A388 by a Level 2 inspector. Calibration and acceptance criteria
shall be per API 6A.
Emerson.com/Rosemount
67
Page 68
Rosemount 114C
μ
June 2019
Surface finish certification (Q16)
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
Thermowell surface finish is typically done to remove all the burrs and sharp edges which smooths the thermowell stem surface. The
Rosemount 114C comes with a standard surface finish of T32 in. CLA N6 (0.8 μm Ra) or better. This option provides a certificate
that documents the maximum surface finish reading for stem and flange (when applicable) and a pass/fail statement. Improved
surface finish options are also available for the Rosemount 114C (see options R14 and R20).
Surface finish <Ra 0.3 μm (12 μin) (R14)
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
Improves surface finish to be less than Ra 0.3 μm. An improved surface finish will increase corrosion resistance and make the
thermowell easier to clean. This is common in sanitary applications.
Electropolish (R20)
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
The electropolish process uses a combination of electrical current and chemicals to improve the surface finish. The surface will
appear shiny and polished. It can have an advantage over mechanical polishing because there is no cold work involved that can lead
to scratches, strains, metal debris, and embedded abrasives on the surface. An improved surface finish will increase corrosion
resistance and make the thermowell easier to clean. This is common in sanitary applications.
Stepped instrument threads (R61)
Back to Van Stone ordering table
Threads are recessed and begin 0.35-in. (9 mm) from the top of the face of the thermowell's instrument connection entry as shown
below:
Standard external hydrostatic pressure test (Q5)
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Thermowells are tested at room temperature for 10 minutes. Water is certified to have a chlorine content of less than 30 ppm. The
certificate will document the chlorine content, hydrostatic test pressure level, duration, and test results. The pressure rating (in psi)
for the different thermowell mounting styles is given below.
68
Emerson.com/Rosemount
Page 69
June 2019
Rosemount 114C
Flanged and Van Stone
Hydrostatic pressure test levels are in accordance with ASME B16.5. When the table below and the standard conflict, the standard
shall govern.
Thermowell material (psi)
Flange class (lb)
150300350425450450450
3007259001100112511251125
600145018002175222522502250
1500 (900)360045005400557556005625
2500600075009000927593009375
AH, AQ, TTSC, SD, SF, SG, SH, SJ, SK, SL, SM,
NK
SN, SP, AP, AM, AD, AE, AF, AJ, AL
CSSR
AG, AK, CA, AB, AC, CB,
CC, DU, DT, DV, AN, AR,
DIN flanged thermowells
Table 21. External Pressure Test–DIN
DIN flanged thermowells
Nominal pressure (bar)Test pressure (bar)
1640
40100
100250
Tes t to 2.5 ⫻ nominal pressure rating
Threaded thermowells
1500 psi
AU, AS, MO, DS
Extended external hydrostatic pressure test (Q9)
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Thermowells are tested at room temperature for 20 minutes. Water is certified to have a chlorine content of less than 30 ppm. The
certificate will document the chlorine content, hydrostatic test pressure level, duration, and test results. The pressure rating (in psi)
for the different thermowell mounting styles are the same as the standard external pressure test.
Standard internal hydrostatic pressure test (Q85)
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
This test is performed at room temperature for a minimum of 10 minutes to 3000 PSI. The water used here is certified to ensure a
chloride content of less than 30 PPM. The certificate provided will document the chloride content, hydrostatic test level, duration,
and results. Thermowells longer than 42-in. will be required to have an internal pressure test (Q85) performed to ensure the internal
cavity integrity has not been compromised.
Emerson.com/Rosemount
69
Page 70
Rosemount 114C
June 2019
Extended internal hydrostatic pressure test (Q86)
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
This test is performed at room temperature for a minimum of 20 minutes to 3000 PSI. The water used here is certified to ensure a
chloride content of less than 30 PPM. The certificate provided will document the chloride content, hydrostatic test level, duration,
and results. Thermowells longer than 42-in. will be required to have an standard internal pressure test (Q85) performed to ensure the
internal cavity integrity has not been compromised.
Canadian registration number (Q17)
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
Any pressure vessel, piping system, or fitting used in Canada is required by law to have a CRN (Canadian Registration Number). This
ensures all pressure vessels, piping systems, and fittings are built under appropriate quality control programs. This CRN is for all
Canadian provinces but the end destination province still needs to be known during the order process.
Dye or liquid penetration testings are performed by ASME Level II or III trained inspectors. These tests are all done in accordance to
ASME Section V, Article 6 with an acceptance criteria per ASME Section III, Div 1 NB-2546. The certificate will document the
inspectors name, dye penetration acceptance criteria, and test result.
70
Emerson.com/Rosemount
Page 71
June 2019
Rosemount 114C
Wall thickness test (Q83)
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
Ultrasonic examination performed to check stem wall thickness. Min and max wall thickness measurements shall be recorded
25 mm or 1-in. from the thermowell tip. Bore position should be 10 percent of minimum stem wall thickness at nominal dimensions
(see image below).
Special cleaning (Q6)
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
Special cleaning for oxygen/special service to be performed in accordance to ASTM G93. The procedure to be qualified using ASTM
G93 Type II quantitative tests. The documentation provided for this test will have a compliance statement to ASTM G93. All cleaned
thermowells will come in a sealed plastic bag to prevent contamination. Not available with carbon steel or any coated material.
Thermowell markings (R40)
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
This options provides the ability to have certain test markings on the thermowell. Below are the tests available for this option.
Q5 — Standard external pressure tests the values and units
Q76 — PMI will be marked on the head length portion of the thermowell and on the top of the flange if applicable
Q9 — Extended external pressure test values and units
Q85 — Standard internal pressure tests values and units
Q86 — Extended internal pressure tests values and units
Emerson.com/Rosemount
71
Page 72
Rosemount 114C
MAX = t+ 3.5
MIN = t
June 2019
Welding procedure qualification record (Q66)
Back to Flanged ordering table
A record of a test performed and tested to ensure that the procedure will produce a good weld. Documentation to be supplied in
accordance with QW-200.2 ASME Section IX.
The records for thermowells with option M01 (low temperature charpy) are different and are highlighted when choosing to ensure
the right documents are submitted to the customer.
Welder performance qualifications (Q67)
Back to Flanged ordering table
A test certificate that shows whether a welder possesses the necessary experience and knowledge to perform the specifications of a
particular weld procedure. Documentation to be supplied in accordance with QW-301.4 ASME Section IX.
Welding procedure specification (Q68)
Back to Flanged ordering table
A formal written document that describes and provides direction to a welder or welding operator for making sound and quality
production welds per code requirements. Documentation to be supplied in accordance with Article V, ASME Section IX.
Phased array ultrasonic test (Q80)
Back to Flanged ordering table
This test is only available on the full penetration weld flanged thermowells. Testing shall be performed in accordance with ASME
section V article 4. Inspection criteria shall be per ASME Section VIII Div 1 per UW3 and conducted by a Level 2 Inspector. The
certificate provided with this option will document the results and the inspectors acceptance criteria .
X-ray/radiograph test (Q81)
Back to Flanged ordering table
Weld joints to be examined for any internal imperfections. Available only with full penetration flanged thermowells. test shall be
performed according with ASME section V, article 24. Inspection criteria shall be per ASME section VIII, Div 1 per UW-51 by a level 2
inspector. Supplier to x-ray report for 10 years. Certificate shall include inspector, acceptance criteria and results. Marking made
during test shall be removed.
Spherical tip (R60)
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
Radius of spherical tip (B) is the same as the specified thermowell tip radius. Thermowell will still maintain specified “U” length.
72
Emerson.com/Rosemount
Page 73
June 2019
Rosemount 114C
Stainless steel plug and chain (R06)
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
The plug and chain are made from stainless steel. This plug is used to protect the thermowell threads when a sensor isn't installed. It
also keeps elements such as rain, dust, and dirt out of the thermowell.
Brass plug and chain (R23)
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
The plug and chain are made from brass. This plug is used to protect the thermowell threads when a sensor isn't installed. It also
keeps elements such as rain, dust, and dirt out of the thermowell.
Vent hole (R11)
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
The vent hole allows for the venting of a thermowell. Vent or weep holes are often used to prevent gas buildup in certain applications.
This option is useful in applications where gas build up is a concern. Process fluid leakage from the vent hole is an indicator of
thermowell failure.
Emerson.com/Rosemount
73
Page 74
Rosemount 114C
A
B
C
D
E
June 2019
Flange face – concentric serrations (R09)
Back to Flanged ordering table
Back to Van Stone ordering table
This option changes the flange face so it has concentric serrations covering the wetted portion of the flange raised face. It is installed
with an inside bolt circle (IBC) gasket/ring gasket, which extends to and is centered by the bolts. This flange face is designed per the
ASME B16.5 standard.
Figure 25. Installation Components
A. Thermowell
B. Bolt/washers
C. Ring gasket
D. Nozzle and mating flange
E. Process
Flange face – flat (R10)
Back to Flanged ordering table
This option changes the flange face so it has no raised section on the wetted portion of the flange face. The flat face is finished with
spiral serrations. This style is frequently used where the mating flange is made from a casting or fragile material. It can be installed
with ring gaskets or full face gaskets that extend past the bolt holes. This flange face is designed per the ASME B16.5 standard.
74
Emerson.com/Rosemount
Page 75
June 2019
A
B
C
D
E
A
B
C
D
E
Figure 26. Installation Components
A. Thermowell
B. Bolt/washers
C. Ring gasket
D. Nozzle and mating flange
E. Process
Raised face – Type B2 (R15)
Rosemount 114C
Back to Flanged ordering table
This option provides a smoother finish to the flange face compared to the standard Type B1 flange face.
Figure 27. Installation Components
A. Thermowell
B. Bolt/washers
C. Ring gasket
D. Nozzle and mating flange
E. Process
Emerson.com/Rosemount
75
Page 76
Rosemount 114C
A
B
C
D
E
June 2019
Flange face – RTJ (R16)
Back to Flanged ordering table
Back to Van Stone ordering table
This option changes the flange face so it has a ring type joint (RTJ). The RTJ flange face is common for high pressure applications using
Class 600 flanges or higher. Both mating flanges have grooves that can accept a RTJ gasket which is usually made of solid metal. This
flange face is designed per the ASME B16.5 standard.
Figure 28. Installation Components
A. Thermowell
B. Bolt/washers
C. Ring gasket
D. Nozzle and mating flange
E. Process
Flange face – groove, Type D (R18)
Back to Flanged ordering table
Back to Van Stone ordering table
Type C “tongue” will mount to Type D “groove”.
76
Emerson.com/Rosemount
Page 77
June 2019
Figure 29. Installation Components
A. Thermowell
B. Bolt/washers
C. Ring gasket
D. Nozzle and mating flange
E. Process
Flange face – tongue, Type C (R19)
Rosemount 114C
Back to Flanged ordering table
Back to Van Stone ordering table
Type C “tongue” will mount to Type D “groove”.
Figure 30. Installation Components
A. Thermowell
B. Bolt/washers
C. Ring gasket
D. Nozzle and mating flange
E. Process
Emerson.com/Rosemount
77
Page 78
Rosemount 114C
Flange face – spigot, Type E (R24)
Back to Flanged ordering table
Back to Van Stone ordering table
Type E “spigot” will mount to type F “recess”.
Figure 31. Installation Components
June 2019
A. Thermowell
B. Bolt/washers
C. Ring gasket
D. Nozzle and mating flange
E. Process
Flange face – recess, Type F (R25)
Back to Flanged ordering table
Back to Van Stone ordering table
Type E “spigot” will mount to type F “recess”.
78
Emerson.com/Rosemount
Page 79
June 2019
Figure 32. Installation Components
A. Thermowell
B. Bolt/washers
C. Ring gasket
D. Nozzle and mating flange
E. Process
Thermowells with wrench flats (R37)
Rosemount 114C
Back to Threaded ordering table
This option only applies to threaded thermowells made from exotic materials. By default, these thermowells are made with two
wrench flats; this option must be selected to get hex (6) wrench flats.
Larger root diameters will provide greater strength. Changing the root diameter is helpful when designing a thermowell to pass wake
frequency calculations.
Guidelines on specifying design modifiers based on the stem profile are as follows:
Straight – only root diameter (Axxx) should be specified
Tapered – both root (Axxx) and tip diameter (Bxxx) must be specified
Stepped – if root diameter (Axxx) only is specified, the tip will be standard 0.5-in diameter; if tip diameter (Bxxx) is ordered, root
diameter (Axxx) must also be specified
80
Emerson.com/Rosemount
Page 81
June 2019
B
Rosemount 114C
Tip diameter (B0XX)
Table 23. Sample Root Diameters
Code Dimension (E)Code Dimension (M)
A0400.4-in.A10010 mm
A0450.45-in.A11011 mm
A1001.00-in.A20520.5 mm
A3103.10-in.A79079 mm
A3153.15-in.A80080 mm
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
Smaller tip diameters will improve time response. Changing the tip diameter is helpful when designing a thermowell to pass wake
frequency calculations.
Guidelines on specifying design modifiers based on the stem profile are as follows:
Straight – only root diameter (Axxx) should be specified
Tapered – both root (Axxx) and tip diameter (Bxxx) must be specified
Stepped – if root diameter (Axxx) only is specified, the tip will be standard 0.5-in. diameter; if tip diameter (Bxxx) is ordered, root
diameter (Axxx) must also be specified
Table 24. Sample Tip Diameters
Code Dimension (E)Code Dimension (M)
B0400.4-in.B12012 mm
B0450.45-in.B13013 mm
B1001.00-in.B20520.5 mm
B1751.75-in.B45045 mm
B1801.80-in.B46046 mm
Emerson.com/Rosemount
81
Page 82
Rosemount 114C
d
June 2019
Bore diameter (d0X)
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
Bore diameter (d) can be selected to accommodate different temperature sensor sizes. Time response is improved when the sensor
and thermowells have a tighter fit.
Table 25. Available Bore Diameters
CodeDimensionCode Dimension (M)
D010.276-in./7.0 mmD050.354-in./9 mm
D030.138-in./3.5 mmD060.433-in./11 mm
D040.385-in./9.8 mm
Tip thickness (t0X)
Back to Threaded ordering table
Back to Flanged ordering table
Back to Van Stone ordering table
Back to Welded ordering table
Tip thickness (t) is specified as the minimum thickness and measured from the top of the gun drill web as shown in the figure below
t
Table 26. Available Tip Thicknesses
CodeDimension
T010.197-in./5.0 mm
T020.236-in./6.0 mm
82
Emerson.com/Rosemount
Page 83
June 2019
f
e
e
Van Stone stub thickness (F0X)
Back to Van Stone ordering table
Van Stone stub thickness is the thickness of the surface on which
the flange lies as shown in the figure below.
Table 27. Available Stub Thicknesses
CodeDimension
F010.591-in. (15 mm)
F020.787-in. (20 mm)
Fillet radius (E0X)
Rosemount 114C
Back to Flanged ordering table
Back to Van Stone ordering table
Lap Joint flanges are machined with a flat face and a fillet radius
to accommodate the stub end or pipe lap as shown in image
below.
Table 28. Available Fillet Radius
CodeDimension
E010.118-in. (3.0 mm)
E020.197-in. (5.0 mm)
Emerson.com/Rosemount
83
Page 84
Rosemount 114C
June 2019
Lap flange material for Van Stone design (C0X)
Back to Van Stone ordering table
This option is only available when the Van Stone (V) mounting configuration is selected. By default, a Van Stone thermowell comes
with a carbon steel A105 lap flange. These options give the choice of having the thermowell ordered without a flange, with a
316/316LSST flange, or with a flange of similar material as the thermowell stem. Below are some model string examples of the
standard offering and options for reference:
Example model: 114CE0030VAA2SC032A – carbon steel A105 lap flange with 316/316L SST thermowell stem provided (standard)
Example model: 114CE0030VAA2SC032AC01 – no lap flange, only thermowell stem provided
Example model: 114CE0030VAA2SC032AC02 – changes default carbon steel A105 lap cover flange to 316/316LSST flange
Example model: 114CE0030VAA2SC032AC03 – changes default cover flange to match thermowell stem material
Emerson Terms and Conditions of Sale are available upon request.
The Emerson logo is a trademark and service mark of Emerson Electric Co.
Rosemount is a mark of one of the Emerson family of companies.
All other marks are the property of their respective owners.