Rosemount 1085B-C Conductivity Microprocessor Analyzer Manuals & Guides

Model 1054B C
Conductivity Microprocessor Analyzer
Instruction Manual
PN 51-1054BC/rev.B April 2003
WARNING
ELECTRICAL SHOCK HAZARD
Relay contacts made to separate power sources must be disconnected before servicing.
Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA-70) and/or any other applicable national or local codes.
Unused cable conduit entries must be securely sealed by non-flammable closures to provide enclosure integrity in compliance with personal safety and environmental protection requirements.
For safety and proper performance this instru­ment must be connected to a properly grounded three-wire power source.
Proper relay use and configuration is the respon­sibility of the user.
Do not operate this instrument without front cover secured. Refer installation, operation and servicing to qualified personnel.
Be sure to disconnect all hazardous voltage before opening the enclosure.
The unused conduit openings need to be sealed with NEMA 4X or IP65 conduit plugs to maintain the ingress protection rating (IP65).
No external connection to the instrument of more than 43V peak allowed with the exception of power and relay terminals. Any violation will impair the safety protection provided.
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PRO-
CEEDING!
Rosemount Analytical designs, manufactures, and tests its products to meet many national and international stan­dards. Because these instruments are sophisticated tech­nical products, you must properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions must be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the product. If this Instruction Manual is not the correct manual, telephone 1-949-757-8500 and the requested manual will be provided. Save this Instruction Manual for future reference.
• If you do not understand any of the instructions, contact your Rosemount representative for clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper instal­lation, operation, and maintenance of the product.
• Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the proper electrical and pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
• When replacement parts are required, ensure that qual­ified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the product’s performance and place the safe operation of your process at risk. Look alike substitu­tions may result in fire, electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protec­tive covers are in place, except when maintenance is being performed by qualified persons, to prevent electri­cal shock and personal injury.
WARNING
This product is not intended for use in the light industrial, residential or commercial environment, per the instrument’s certification to EN50081-2.
Emerson Process Management
Rosemount Analytical Inc.
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250
http://www.RAuniloc.com
© Rosemount Analytical Inc. 2001
MODEL 1054B C TABLE OF CONTENTS
MODEL 1054B CONDUCTIVITY MICROPROCESSOR ANALYZER
TABLE OF CONTENTS
Section Title Page
1.0 DESCRIPTION AND SPECIFICATIONS........................................................... 1
1.1 Features and Applications ................................................................................. 1
1.2 Specifications .................................................................................................... 2
1.3 Ordering Information.......................................................................................... 4
2.0 INSTALLATION.................................................................................................. 5
2.1 General.............................................................................................................. 5
2.2 Unpacking and Inspection ................................................................................. 5
2.3 Mechanical Installations .................................................................................... 5
2.4 Electrical Wiring................................................................................................. 6
3.0 DESCRIPTION OF CONTROLS ....................................................................... 14
3.1 Keyboard Functions........................................................................................... 14
3.2 Item Selection and Value Adjustment Keys....................................................... 15
4.0 CONFIGURATION............................................................................................. 19
4.1 General ............................................................................................................. 19
4.2 Alarm 1 and 2 ................................................................................................... 21
4.3 Interval Timer .................................................................................................... 22
4.4 Temperature ..................................................................................................... 23
4.5 Current Output .................................................................................................. 23
4.6 Defaults ............................................................................................................. 24
4.7 Alarm Setpoint ................................................................................................... 25
4.8 Output Scale Expansion .................................................................................... 26
4.9 Simulate Output................................................................................................. 27
5.0 START-UP AND CALIBRATION........................................................................ 28
5.1 General.............................................................................................................. 28
5.2 Entering Cell Constant....................................................................................... 28
5.3 Temperature Calibration .................................................................................... 28
5.4 Initial Loop Calibration ....................................................................................... 28
5.5 Routine Standardization .................................................................................... 30
5.6 Sensor Maintenance.......................................................................................... 30
6.0 KEYBOARD SECURITY ................................................................................... 31
6.1 General.............................................................................................................. 31
6.2 Access Code ..................................................................................................... 31
7.0 THEORY OF OPERATION................................................................................ 31
8.0 DIAGNOSTICS AND TROUBLESHOOTING .................................................... 32
8.1 Diagnostics ........................................................................................................ 32
8.2 Troubleshooting................................................................................................. 33
8.3 Instrument Maintenance .................................................................................... 35
9.0 RETURN OF MATERIAL................................................................................... 37
i
MODEL 1054B C TABLE OF CONTENTS
LIST OF FIGURES
Figure No. Title Page
2-1 Panel Mounting Cutout...................................................................................... 7
2-2 Panel Mounting Tab Installation......................................................................... 8
2-3 Wall Mounting J-Box Installation........................................................................ 9
2-4 Wall Mounting J-Box Wiring............................................................................... 10
2-5 Pipe Mounting Installation ................................................................................. 11
2-6 Electrical Wiring................................................................................................. 12
2-7 Wall Mount Enclosure (Option -20).................................................................... 13
3-1 Function Select on Keypad................................................................................ 14
3-2 Accessing Editing Function ............................................................................... 14
3-3 Accessing Configuration Menus........................................................................ 14
3-4 LCD Display....................................................................................................... 15
3-5 Set Menu Items ................................................................................................. 18
4-1 Alarm 1 and 2 Configuration.............................................................................. 21
4-2 Interval Timer Configuration .............................................................................. 22
4-3 Timer Diagram for One Cycle ............................................................................ 22
4-4 Temperature Configuration................................................................................ 23
4-5 Current Output Configuration ............................................................................ 23
4-6 Default Configuration......................................................................................... 24
4-7 Alarm Setpoint................................................................................................... 25
4-8 Output Scale Expansion.................................................................................... 26
4-9 Simulate Current Output.................................................................................... 27
8-1 Simulate Conductivity Input............................................................................... 36
LIST OF TABLES
Table No. Title Page
1-1 Conductivity Range ........................................................................................... 3
1-2 Replacement Parts............................................................................................ 4
1-3 Accessories ....................................................................................................... 4
3-1 Key Description ................................................................................................. 16
3-2 Information Mnemonics ..................................................................................... 17
3-3 Set Function Mnemonics................................................................................... 17
4-1 Configuration Work Sheet ................................................................................. 20
4-2 Relay States for Various Conditions and Alarm/Default Configurations ............ 24
8-1 Fault Mnemonics ............................................................................................... 32
8-2 RTD Resistance Values..................................................................................... 32
8-3 Troubleshooting Guide ...................................................................................... 34
ii
1
MODEL 1054B C SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
1.1 FEATURES AND APPLICATIONS
The Model 1054B Microprocessor Analyzers, with the appropriate sensors, are designed to continuously meas­ure and control pH, ORP, conductivity, resistivity, ratio, percent concentration, dissolved oxygen, ozone or total free chlorine in industrial and municipal processes.
The Model 1054B Conductivity Analyzers are housed in a NEMA 4X (IP65) weatherproof, corrosion-resistant, flame retardant enclosure suitable for panel, pipe or wall mounting. All functions are accessed through the front panel membrane keyboard which features tactile feed­back. Measurement data may be read at any time. However, settings may be protected against accidental or unauthorized changes by a user selectable security code. The display indicates the measured value in engi­neering units as well as temperature, alarm status, hold output and fault conditions.
The 1054B transmits a user selected isolated current output which is continuously expandable over the meas­urement range for either direct or reverse action and can be displayed in milliamps or percent. Output dampening of 0-255 sec. is user selectable.
The output and relay default settings are user selectable for hold or fault mode operation. The hold output function allows manual control during routine sensor mainte­nance.
Continuous self diagnostics alert the operator to faults due to analyzer electronics, integral RTD failures, open wiring and process variable range problems. In the event of a fault condition or hold mode diagnosed by the ana­lyzer, the output will be set to a preset or last process value and the relays will be set to their default settings.
Dual alarms are a standard feature on the Model 1054B and are programmable for either high or low operation. Alarm 2 may be programmed as a fault alarm. Both alarms feature independent setpoints, adjustable hys­teresis and time delay action. The time delay is conven­ient when an alarm is used for corrective action, such as shutting down a demineralizer for regeneration. Time delay will ignore a temporary breakthrough and prevent shutting down a demineralizer unit prematurely. A dedi­cated interval timer with relay is also provided.
Automatic or manual temperature compensation is key­board selectable. The process temperature is accurately measured from an integral RTD in the sensor assembly and is read on the display. For greater accuracy, the tem­perature indication may be standardized to the process temperature. The temperature may be configured to read in °C or °F.
Calibration is easily accomplished by simply immersing the sensor in a known solution and entering the value. With a two point calibration, the Model 1054B will auto­matically calculate the temperature slope of the solution. Upon routine standardization a sensor cell factor value is calculated, and a trend of this value can be used to track sensor coating.
The Model 1054B Microprocessor Analyzer comes stan­dard with an LCD display. An LED display is available as an option.
• SELF DIAGNOSTICS with a user selectable fault alarm.
• KEYBOARD SECURITY is user selectable.
• NO BATTERY BACK-UP REQUIRED. Non-volatile EEPROM memory.
• DUAL ALARMS WITH PROGRAMMABLE LOGIC. A third relay is provided with timer func­tions.
• PROGRAMMABLE OUTPUT AND RELAY DEFAULTS for hold and fault modes.
• NEMA 4X (IP65) WEATHERPROOF CORROSION-RESISTANT ENCLOSURE.
2
Model 1054B C SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
1.2 SPECIFICATIONS -
Enclosure: Black, ABS, NEMA 4X, IP65,
CSA Enclosure 4 144 X 144 X 192 mm (5.7 X 5.7 X 7.6 in.)
Wall Mount Enclosure: NEMA 4X, Heavy duty
fiberglass, reinforced thermoplastic.
356.4 X 450.1 X 180.2 mm* (14 X 17.7 X 7.1 in.*)
Front Panel: Membrane keyboard with tactile feed-
back and user selectable security
Digital Display: LCD, black on grey
Optional, red LED Character Height: 18 mm (0.7 in.)
Electrical Classification:
FM Class I, Div. 2, Group A thru D
28 Vdc relays - 5.0 amps resistive only 150 mA - Groups A & B; 400 mA - Group C ;
540 mA - Group D; Ci = 0; Li = 0
CSA Class I, Div. 2, Group A thru D.
28 Vdc, 110 Vac & 230 Vac relays
5.0 Amps resistive only
Wall Mount Enclosure: General Purpose
Power: 100 - 127 VAC, 50/60 Hz ± 6%, 4.0 W
200 - 253 VAC, 50/60 Hz ± 6%, 4.0 W
Current Output: Isolated, 0-20 mA or 4-20 mA into
600 ohms maximum load at 115/230 Vac or 550 ohms maximum load at 100/200 Vac, Direct or Reverse Output Dampening: 0-255 seconds
Code -20 Wall Mount Enclosure does not meet CE requirements
*Includes latches and mounting feet
EMI/RFI: EN61326 LVD: EN61010-1 Ambient Temperature: -20 to 65°C (-4 to 149
°
F)
Ambient Humidity: LED max 95% RH
(LCD max 85% RH @ 50°C)
Alarms: Dual, field selectable High/Low, High/High,
Low/Low Alarm 2 configurable as a fault alarm Time Delay 0 to 254 seconds Dual Setpoints, continuously adjustable Hysteresis is adjustable up to 25% full scale for low side/High Alarm and high side/Low Alarm
Interval Timer: Interval: 10 min. to 2999 days
On Counts: 1 to 60 On Duration: 1 to 299.9 seconds Off Duration: 1 to 299.9 seconds Wait Duration: 1 to 299.9 seconds Controls dedicated relay
Relay Contacts: Epoxy Sealed Form A contacts,
SPST, Normally Open.
Resistive
Inductive
28 VDC 5.0 Amps 3.0 Amps 115 VAC 5.0 Amps 3.0 Amps 230 VAC 5.0 Amps 1.5 Amps
Weight/Shipping Weight: 1.1 kg/1.6 kg (2.5 lb/3.5 lb)
3
Model 1054B C SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
The Model 1054B Conductivity Analyzer measures over the range of 0-2 µS/cm to 0-1,000 mS/cm. Temperature slope may be adjusted anywhere between 0 and 5% to provide greater accuracy in chemical con­centration control. The temperature slope is factory set at 2% as a representative value, but each conductive solution has its own set of temperature vs. concentration curves. The Model 1054B C will automatically calcu­late the temperature slope for any given solution, or permit manual adjustment of the temperature slope if already known. On calibration the analyzer will also automatically correct for cell constant variations for better measurement accuracy.
ANALYZER SPECIFICATIONS @ 25°C Measurement Range: (See Table 1) Output Scale: Zero suppression: up to 90% full scale.
Span: from 10% to 100% full scale
Accuracy: ±0.5% of reading Repeatability: ±0.25% of reading Stability: ±0.25% month, non-cumulative Temperature Effect: 0.02% of reading/°C Temperature Compensation: -20 to 200°C
(-4 to 392°F) (automatic or manual)
Temperature Slope Adjustment: 0-5%/
°
C
RECOMMENDED SENSORS:
Model 140 Retractable Conductivity
Model 150 Insertion/Submersion Conductivity
Model 400 Screw-in Conductivity
Model 401-14 Screw-in Conductivity
Model 402 Retractable Conductivity
Model 403 Sanitary Flange Conductivity
Model 404 Low Flow Conductivity
FULL SCALE MICROSIEMENS/cm
TABLE 1-1. CONDUCTIVITY RANGE
Conductivity Sensor 150 140
Model Number 400 140 150 150
402/403 400/402/403
404
Cell Constant 0.1 0.2 0.5 1.0
Min. Range 2 4 100 200
Max. Range* 2,000 4,000 10,000 20,000
* Values shown are absolute conductivi-
ty. Maximum range will be reduced for compensated conductivity at elevated process temperatures.
4
MODEL 1054B C SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
1.3 ORDERING INFORMATION
The Model 1054B Microprocessor Analyzer: Housed in a corrosion resistant, weatherproof enclosure and oper-
ates on either 115 or 230 VAC, 50/60 Hz power. Standard features include digital display, isolated current output, dual alarms, and automatic and manual temperature compensation.
MODEL
1054B MICROPROCESSOR ANALYZER (3.5 lbs/1.5 kg)
Code Measurement
C Contacting Conductivity T Toroidal Conductivity
P/N DESCRIPTION
22966-00 PCB, LCD Digital Display
23025-01 Panel Mounting Kit
23739-00 PCB, Power Supply
23664-01 PCB, CPU, Conductivity
23245-01 PCB, LED Digital Display
23740-00 PCB, Motherboard
23695-04 Keyboard Overlay, LCD Version
23695-05 Keyboard Overlay, LED Version
33469-00 Enclosure, Body
33470-00 Enclosure, Rear Cover
32938-00 Gasket, Rear Cover
9100157 Fuse, 0.1A, 3AB, 250V, Slo-Blow
9100160 Fuse, .250A, 125V
9100189 Fuse, .750A, 125V
TABLE 1-2. Replacement Parts
CODE STANDARD ENCLOSURE OPTIONS
01 LCD Display 02 LED Display
CODE OPTIONS
20 Wall Mount Enclosure
1054BC 01 20 EXAMPLE
TABLE 1-3. Accessories
P/N DESCRIPTION
2001492 Tag, Stainless Steel, Specify
Marking
23053-00 Mounting Bracket, 2-inch Pipe
23054-01 Mounting Bracket, Wall, with
Junction Box
23268-01 Heater, 115 VAC, 50/60 Hz,
1054B (Code 20 Only)
23268-02 Heater, 230 VAC, 50/60 Hz,
1054B (Code 20 Only)
5
MODEL 1054B C SECTION 2.0
INSTALLATION
SECTION 2.0
INSTALLATION
2.1 GENERAL. Installation must be performed by a
trained technician. This analyzer's enclosure is suit­able for outdoor use. However, it should be located in an area where temperature extremes and vibrations are minimized or absent. Installation must be per­formed by a trained technician.
2.2 UNPACKING AND INSPECTION. Inspect the
analyzer for shipping damage. If damaged, notify the carrier immediately. Confirm that all items shown on the packing list are present. Notify Rosemount Analytical if items are missing.
2.3 MECHANICAL INSTALLATION. Select an instal-
lation site that is at least one foot from any high volt­age conduit, has easy access for operating personnel, and is not in direct sunlight. Mount the analyzer as fol­lows:
1. Remove the four screws that secure the rear
cover of the enclosure.
2. Remove the four screws holding the front panel
assembly of the enclosure and carefully pull the front panel and connected printed circuit boards straight out.
3. Follow the procedure for the appropriate mounting
configuration: Section 2.3.1 for panel mounting, Section 2.3.2 for wall mounting, Section 2.3.3 for pipe mounting.
2.3.1 Panel Mounting (Standard). The Model 1054B
is designed to fit into a DIN standard 137.9 mm x
137.9 mm (5.43 in. x 5.43 in.) panel cutout (Refer to
Figure 2-1 and Figure 2-2).
1. Prepare the analyzer as described in Section 2.3.
2. Install the mounting latches as described in Figure
2-2 (latches are shown oversize for clarity). If the latches are not installed exactly as shown, they will not work correctly. The screws provided are self-tapping. Tap the screw the full depth of the mounting latch (refer to side view) leaving a gap greater than the thickness of the cutout panel.
3. Align the latches as shown and insert the analyz­er enclosure through the front of the panel cutout. Tighten the screws for a firm fit. To avoid damag­ing the mounting latches, do not use excessive force.
4. Replace the front panel assembly. Circuit boards must align with the slots on the inside of the enclosure. Assure that the continuity wire is con­nected to the rear cover and the interface board’s closest mounting screw. Replace the door and four front panel screws.
2.3.2 Wall Mounting Plate with Junction Box
(P/N 23054-01). Refer to Figure 2-3 and Figure 2-4.
1. Prepare the analyzer as described in Section 2.3.
2. Mount the junction box and bracket to the analyz­er with the hardware provided. All wiring can be brought to the terminal strip prior to mounting the analyzer.
3. Place the metal stiffener on the inside of the ana­lyzer and mount the two 1/2-inch conduit fittings using two each weather seals as shown. Mount NEMA 4X conduit plug (included) into center con­duit hole.
4. Mount the analyzer to the junction box using the 1/2-inch conduit fittings.
5. Complete wiring from the analyzer to the junction box (Refer to Figure 2-4).
NOTE
Run sensor wiring out of the left opening (From front view) to J-Box. All others out right opening to J-Box.
2.3.3 Pipe Mounting (P/N 23053-00). The 2-inch pipe
mounting bracket includes a metal plate with a cutout for the analyzer (Refer to Section 2.3 for mounting the analyzer into the plate). Mounting details are shown in Figure 2-5.
2.3.4 Wall Mount Enclosure (option -20). See
Figure 2-7 for installation details.
6
MODEL 1054B C SECTION 2.0
INSTALLATION
2.4 ELECTRICAL WIRING. The Model 1054B has
three conduit openings in the bottom rear of the ana­lyzer housing which will accommodate 1/2-inch con­duit fittings. From the front view, the conduit opening on the left is for sensor wiring; the center is for signal output and the opening on the right is for timer, alarm, and AC connections. Sensor wiring should always be run in a separate conduit from power wiring. AC power wiring should be 14 gauge or greater.
NOTE
For maximum EMI/RFI protection the output cable should be shielded and enclosed in an earth grounded, rigid metal conduit. When wiring directly to the instrument connect the output cable’s outer shield to the transmitter’s earth ground via terminal 8 on TB3. When wiring to the wall mounting junc­tion box connect the output cable’s outer shield to the earth ground terminal on TB-A. The sensor cable should also be shield­ed. When wiring directly to the instru­ment connect the sensor cable’s outer shield to the transmitter’s earth ground via terminal 8 of TB-2. If the sensor cable’s outer shield is braided an appro­priate metal cable gland fitting may be used to connect to braid to earth ground via the instrument case. When wiring to the wall mounting junction box connect the sensor cable’s outer shield to the earth ground terminal on TB-A. The user must provide a means to dis­connect the main power supply in the form of circuit breaker or switch. The cir­cuit breaker or the switch must be locat­ed in close proximity to the instrument and identified as the disconnecting device for the instrument.
2.4.1 Power Input Wiring. The Model 1054B can be configured for either 115 VAC or 230 VAC power.
Connect AC power to TB1-8 and -9 (115 VAC) or TB1­7 and -8 (230 VAC) ground to the ground terminal at TB3-8 (refer to Figure 2-6).
CAUTION
The sensitivity and stability of the analyzer will be impaired if the input wiring is not grounded. DO NOT apply power to the analyzer until all electrical connections are verified and secure. The following precau­tions are a guide using UL 508 as a safe­guard for personnel and property.
1. AC connections and grounding must be in compli­ance with UL 508 and/or local electrical codes.
2. The metal stiffener is required to provide support and proper electrical continuity between conduit fittings.
3. This type 4/4X enclosure requires a conduit hub or equivalent that provides watertight connect, REF UL 508-26.10.
4. Watertight fittings/hubs that comply with the requirements of UL 514B are to be used.
5. Conduit hubs are to be connected to the conduit before the hub is connected to the enclosure, REF UL 508-26.10.
6. If the metal support plate is not used, plastic fit­tings must be used to prevent structural damage to the enclosure. Also, appropriate grounding lug and awg conductor must be used with the plastic fittings.
2.4.2 Output Wiring. The signal output and alarm
connections are made to terminals 1 through 6 of TB1 and TB3-1 and 2. (Refer to Figure 2-6).
7
MODEL 1054B C SECTION 2.0
INSTALLATION
FIGURE 2-1. Panel Mounting Cutout
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
DWG. NO. REV.
41054B01 B
8
MODEL 1054B C SECTION 2.0
INSTALLATION
FIGURE 2-2. Panel Mounting Tab Installation
DWG. NO. REV.
41054A26 A
9
MODEL 1054B C SECTION 2.0
INSTALLATION
FIGURE 2-3. Wall Mounting J-Box Installation
DWG. NO. REV.
41054A27 A
10
MODEL 1054B C SECTION 2.0
INSTALLATION
FIGURE 2-4. Wall Mounting J-Box Wiring
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
DWG. NO. REV.
41054B14 B
11
MODEL 1054B C SECTION 2.0
INSTALLATION
MILLIMETER
INCH
DWG. NO. REV.
41054B02 D
FIGURE 2-5. Pipe Mounting Installation
12
FIGURE 2-6. Electrical Wiring
MODEL 1054B C SECTION 2.0
INSTALLATION
DWG. NO. REV.
41054B04 D
13
MODEL 1054B C SECTION 2.0
INSTALLATION
FIGURE 2-7. Wall Mount Enclosure (Option -20)
DWG. NO. REV.
41054B43 A
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
14
Configuration is all accomplished through a series of menus located within the set mode menu. To access these set mode menus the ACCESS keypad is pressed TWICE in RAPID succession.
Once inside the Set mode menu, use the scroll key­pad to scroll through the menu list. When the menu desired is displayed, release the scroll keypad.
To enter the submenus press the SELECT keypad. If the submenu allows editing, the item will flash that can be edited. If not, use the scroll keypad to scroll through the next list of submenus. SELECT will enter this sub­menu and if it is editable, the field will flash.
To exit the menu and SAVE the new value, press the ENTER keypad.
To exit the menu without saving the edited value, press the PV keypad to jump out of the set menu program with out saving value. To change other parameters will require re-entering the set menu program.
Figure 3-4 explains the various fields surrounding the Primary process on the LC display.
Table 3-1 describes the functions accessible with the eight (8) keypads, the number of times to press the keypad to access, and its function when used with the select keypad and set menu.
Table 3-2 and Table 3-3 describe the meaning of the various mnemonics used on the display. They are cat­egorized by their use in either menus or as process information.
MODEL 1054B C SECTION 3.0
DESCRIPTION OF CONTROLS
SECTION 3.0
DESCRIPTION OF CONTROLS
3.1 KEYBOARD FUNCTIONS. All operations of the
Model 1054B microprocessor Analyzer are controlled by the eight (8) keypads on the front of the instrument. These keypads are used to do the following:
1. Display parameters other than the primary param­eter.
2. Edit setpoints for alarms, set up specific output current value for simulation, calibrate, tempera­ture, conductivity, etc.
3. Configure display for temperature units, for auto­matic temperature compensation, alarm usage, setting timer functions, security, and output range.
To view and not change parameters other than the pri­mary parameter requires only a simple keystroke rou­tine. As shown in Figure 3-1, a single keypress accesses the lower function printed on the keypad. Quick, double keypresses access the top function printed on the keypad.
Editing any of these parameters requires one more operation. After displaying the value associated with the parameter selected, press the SELECT keypad. As seen in Figure 3-2, this will display the numerical value, and the first digit will be flashing to indicate this value may be edited.
All changes to the operating program that set-up the instrument display are made through the set menu pro­gram. See Figure 3-5 at the end of this section.
FIGURE 3-1. Function Select on Keypad.
Single press of the keypad will access the present Conductivity reading. Read only.
1. Press Key.
2. AdJ shows briefly.
3. Numbers show with digit flashing.
1. Press Key twice.
2. Lo shows briefly.
3. Zero point is displayed.
SELECT
ZERO
ALARM
1
OUTPUT
PV
FIGURE 3-2. Accessing Editing Function.
1. Press twice in rapid succession.
2. See SEt on display. Confirms entry into set mode menu.
3. First menu item is displayed. Analyzer now ready to configure.
4. Use the SCROLL keypad to rotate through the available menus.
ACCESS
ENTER
SEt
Cin
ç
FIGURE 3-3. Accessing Configuration Menus
Quick double press will access the current output current value in mA or %. Read only.
ç
15
MODEL 1054B C SECTION 3.0
DESCRIPTION OF CONTROLS
FIGURE 3-4. LCD Display
RELAY 1 ACTIVATED
RELAY 2 ACTIVATED
CONDUCTIVITY: µS ­FLAG ON; mS ­FLAG BLINKING
% VALUE ­FLAG ON; mA ­FLAG BLINKING
HOLD STATUS FLAG ON; FAULT ­FLAG BLINKING
UPPER FUNC­TION PRESS TWICE QUICKLY
LOWER FUNC­TION PRESS ONCE
3.2 Item Selection and Value Adjustment Keys. The
three keys located on the lower right side of the keypad are used for menu navigation, value adjustment and entry, and item selection. These keys perform the fol­lowing functions:
A. SELECT/Shift (
ç) Key. This key is
used to select the displayed menu, or for shifting to the next digit in the Numeric Display.
B. SCROLL Key (é). This key is used to scroll through menu when selected, or scroll through digits on the active (flash­ing) Numeric Display, or move the deci­mal point and µS/mS display. Holding key
down auto scrolls display.
C. ACCESS/ENTER Key. This key is used to ACCESS the Set Mode (Section
4.1.2) and to ENTER the displayed value into memory (from Numeric Display).
ç
SELECT
é
ACCESS
ENTER
16
TABLE 3-1. Key Description
Displays - current output (mA or % full scale).
Set Function (w/SELECT) - Simulates cur­rent output.
Displays - low current setpoint.
Set Function (w/SELECT) - Sets low current point.
Displays - full scale output setpoint.
Set Function (w/SELECT) - Sets full scale output point.
Select sub menu (mnemonic display). Shift to next digit (numeric display).
Scroll through menu (mnemonic display). Scroll digits (numeric display). Scroll decimal position, µS/mS display. Holding key down auto scrolls the main set menu and each digit in the numeric display.
Press twice to access set-up menu. Enter displayed value into memory. Enter displayed menu item (flashing) into memory.
Two Point temperature slope calibra­tion.
Displays - Alarm 1 setpoint.
Set Function (w/SELECT) - Sets Alarm 1 setpoint.
Displays - conductivity.
Set Function (w/SELECT) - One point standardization of conductivity. (PV= Process Variable)
Initiates or removes analyzer from hold con­dition.
Displays - process temperature (°C or °F).
Set Function (w/SELECT) - One point standardization of temperature.
Displays - Alarm 2 setpoint.
Set Function (w/SELECT) - Sets Alarm 2 setpoint.
HOLD
TEMP
OUTPUT
PV
ZERO
ALARM 1
F.S.
ALARM 2
CAL
ç
SELECT
é
ACCESS
ENTER
Displays - temperature slope in percent.
Set Function (w/SELECT) - manually sets tem­perature slope.
SECOND FUNCTION (PRESS TWICE QUICKLY)MAIN FUNCTION (PRESS ONCE)
MODEL 1054B C SECTION 3.0
DESCRIPTION OF CONTROLS
17
MODEL 1054B C SECTION 3.0
DESCRIPTION OF CONTROLS
TABLE 3-2. Information Mnemonics
MNEMONIC DESCRIPTION
AdJ Adjustment to value reading bAd Incorrect entry Con Conductivity Display dOC Displays conductivity output (mA) HLd Analyzer in Hold Position
HI Displays high range value for current output
itr Interval timer activated
LO Displays low range value for current output LOC Access locked – enter security code Pct Displays conductivity output (percent) SEt Set mode SiP Simulates current output (percent) SiC Simulates current output (mA) SLP Displays temperature slope in percent SP1 Displays Alarm 1 setpoint SP2 Displays Alarm 2 setpoint Std Standardize conductivity tP1 Calibration Point 1 tP2 Calibration Point 2
AL1 Alarm 1 setup AL2 Alarm 2 setup Atc Automatic temp. comp. CEL Cell Constant
DC Temperature °C Cin Display Sensor input COd Security Code cnt Timer count CUr Config. current output Cur Config. fault output cur Default current setpoint dAY Days dFt Fault Configuration d-O Display output d-t Display temperature doc Display output in mA doF Delay off time don Delay on time dPn Dampen output
dtS LCD/LED Display test dur Timer duration
DF Temperature °F Fct Calibration Factor FLt Use alarm as fault alarm
Hi Relay action - high H-L Alarm logic
Hr Hours HYS Hysteresis int Interval period Int Timer setup
Lo Relay action - low non No action on fault oFF Relay open on fault OFF Alarm not used ont Timer on time
on Relay closed on fault
On Use alarm as process
alarm
OFt Timer off time OUt Current output
Pct
Display output in percent
rL1 Relay 1 fault setup rL2 Relay 2 fault setup SEC Seconds SHO Show fault history t-C Temperature config. tiL Timer - time remaining tOn Timer status UEr User version uin Minutes 420 4mA to 20mA output 020 0mA to 20mA output
TABLE 3-3. Set Function Mnemonics
18
MODEL 1054B C SECTION 3.0
DESCRIPTION OF CONTROLS
H-L
HYS
don
doF
On
OFF
Cin CEL Fct AL1
AL2
Int
t-C
OUt
UEr
dF
t
dtS
COd
SEt
Hi
Lo
tOn
int
cnt
ont
OFt
dur
ti L
oFF
on
SEC
uin
hr
dAY
*F
*C
doc
Pct
on
oFF
non
non
cur
rL1
rL2
Cur
SHO
on
oFF
On
FLt
OFF
420
020
d-t
Atc
dPn
CUr
d-O
FIGURE 3-5. Set Menu Items
19
MODEL 1054B C SECTION 4.0
CONFIGURATION
SECTION 4.0
CONFIGURATION
4.1 GENERAL. This section details all of the items avail-
able in the Set Mode to configure the analyzer to a spe­cific application.
4.1.1. Configuration Worksheet. The configuration worksheet on the following page should be filled out before proceeding with the analyzer’s configuration. This sheet gives a brief parameter description, the factory set­ting, and a space for user setting.
4.1.2 Set Mode Display Mnemonic SEt. Most of the analyzer's configuration is done while in the Set Mode. Please refer to Figure 3-5 for the layout of all menu items. All menu variables are written to the ana­lyzer's EEPROM (memory) when selected and remain there until changed. As these variables remain in memory even after the analyzer's power is removed, the analyzer configuration may be performed prior to installing it.
1. Power up the analyzer. Only power input wiring is required for analyzer configuration (Refer to Section 2.4.1). The analyzer's display will begin showing values and/or fault mnemonics. All fault mnemonics will be suppressed while the analyzer is in Set Mode (the fault flag will continue to blink).
2. Enter Set Mode. Pressing the ACCESS key twice in rapid succession will place the analyzer in Set Mode. The display will show SEt to confirm that it is in Set Mode. It will then display the first item in the set menu. The analyzer is now ready for user configuration.
NOTE:
If LOC displays, the Keyboard Security Code must be entered to access the Set Mode. (Refer to Section 6.0.)
3. Analyzer variables can be entered in any order. On initial configuration, however, it is recommend­ed that the variables be entered in the order shown on the worksheet. Refer to the configura­tion worksheet (Table 4-1). This will reduce the chance of accidentally omitting a needed variable.
20
MODEL 1054B C SECTION 4.0
CONFIGURATION
TABLE 4-1.
CONFIGURATION WORKSHEET
Use this worksheet to assist in the configuration of the analyzer. Date: _________________
RANGE FACTORY SET USER SET
A. Alarm 1 Setup (AL1)
1. Alarm Configuration (On/OFF) On _________
2. High or Low (H-L) (Hi/Lo) Lo _________
3. Hysteresis (HYS) 0-25 % of setpoint 0.00% _________
4. Delay Time On (don) 0-255 sec. 000 Seconds _________
5. Delay Time Off (doF) 0-255 sec. 000 Seconds _________
B. Alarm 2 Setup (AL2)
1. Alarm Configuration (On/FLt/OFF) On _________
2. High or Low (H-L) (Hi/Lo) Hi _________
3. Hysteresis (HYS) 0-25 % of setpoint 0.00% _________
4. Delay Time On (don) 0-255 sec 000 Seconds _________
5. Delay Time Off (doF) 0-255 sec 000 Seconds _________
C. Interval Timer (Int)
1. Active Status (tOn) (oFF/on) oFF _________
2. Interval Time (int) minimum 10 minutes 1 Day _________
3. Count (cnt) 1 to 60 5 _________
4. On Time (ont) 0 to 299.9 sec 1 Second _________
5. Off Time (OFt) 0 to 299.9 sec 1 Second _________
6. Duration (dur) 0 to 299.9 sec 2 Seconds _________
D. Temperature Setup (t-C)
1. Display Temperature (d-t) (°C/°F) °C _________
2. Automatic Temperature Compensation on _________ (Atc) (on/oFF)
a. Manual Temp. Value -20°C to 200°C _________
E. Current Output Setup (OUt)
1. mA Output (CUr) (020/420) 420 _________
2. Display Current Output (d-O) (Pct/doc) doc _________
3. Dampen Current Output (dPn) 0-255 sec. 0.0 Seconds _________
F. Default Setup (dFt)
1. Relay 1 Default (rL1) (non/oFF/on) non _________
2. Relay 2 Default (rL2) (non/oFF/on) non _________
3. Current Output Default (Cur) (non/cur) non _________
G. Keyboard Security Setup (COd)
1. Keyboard Security Required 001-999 _ _________
2. Keyboard Security Not Required 000 000 _________
Alarm Set Points
1. Alarm 1 (SP1) 0-1999 mS 0.00 mS _________
2. Alarm 2 (SP2) 0-1999 mS 1,000 mS _________
Current Output
1. Zero (0 or 4 mA) (Lo) 0-1,000 mS 0.00 mS _________
2. F.S. (20 mA) (HI) 0-1,000 mS 1,000 mS _________
21
MODEL 1054B C SECTION 4.0
CONFIGURATION
4.2. ALARM 1 AND 2. Display Mnemonic AL1 or AL2.
Used to set alarm relay logic. The alarms may be used to perform on-off process control. See note below.
A. ON. Display Mnemonic On. Select this item if Alarm 1 or 2 is to be used as a process alarm. See Steps D through G for further configuration.
B. OFF. Mnemonic OFF. Select this item if alarm 1 or 2 will not be used or to temporarily disable the alarm. Alarm 1 or 2 setpoint will display oFF if this item is selected. Omit Steps D through G.
C. Fault. Display Mnemonic FLt. (Alarm 2 only). Select to make Alarm 2 a fault alarm. Relay 2 will energize when the unit shows a fault condition. See Table 8-1 for a listing of the fault mnemonics and their descrip­tions. Alarm 2 setpoint will display flt if this item is selected. Omit Steps D through G.
D. Alarm Logic. Mnemonic H-L. Select this item for high or low alarm logic. High logic activates the alarm when the reading is greater than the set point value. Low logic activates the alarm when the reading is less than the set point value.
E. Relay Hysteresis. Display Mnemonic HYS. Sets the relay hysteresis (dead band) for deactivation after reading has passed the alarm set point. May be set from 0 to 25%. Use hysteresis when a specific con­ductivity should be reached before alarm deactivation.
F. Delay Time On. Display Mnemonic don. Sets time delay for relay activation after alarm set point is reached. May be set from 0 to 255 seconds.
G. Delay Time Off. Display Mnemonic doF. Sets time delay for relay deactivation after alarm set point is reached. May be set from 0 to 255 seconds. Alarm state restarts time from zero. Use when a fixed time should pass before relay deactivation occurs.
4.2.1 Alarm Configuration (AL1/AL2). Refer to Figure 4-1.
1. Enter Set Mode by pressing ACCESS key twice.
2. SCROLL (é ) until AL1 or AL2 appears on the dis­play.
3. SELECT to move to the next menu level. On, OFF or (AL2 only) FLt will display.
4. SCROLL (é ) to display desired item then SELECT.
5. If OFF is selected, display will show oFF to acknowledge. Press ENTER key to return to AL1 or AL2, concluding routine. Skip to Step 11.
If On is selected, display will show on to acknow­ledge, then display H-L. Proceed to Step 6.
If FLt is selected, display will show Flt to ack­nowledge. Press ENTER key to return to AL2.
6. SELECT H-L. Hi or Lo will display (flashing).
7. SCROLL (é ) to the desired item and ENTER it into memory. Display will return to H-L. If changes to relay activation logic are desired, proceed to Step 8, otherwise Step 12.
8. SCROLL (é ) to display HYS, don or doF then SELECT desired item. Numerical display will flash to indicate that a value is required.
9. Use SCROLL (é) and SHIFT (ç) to display the desired value.
10. ENTER value into memory. The analyzer will acknowledge and return to display of last item selected. Repeat Step 8 if further changes are desired, otherwise Step 12.
11. Repeat Step 3 for the other Alarm's settings as required.
12. To return to the first level of the Set Mode, Press the ACCESS key.
Figure 4-1. Alarm 1 and Alarm 2 Configuration
On
OFF
On
FLt
OFF
AL1
AL2
H-L
HYS
don
doF
Hi
Lo
SEt
22
MODEL 1054B C SECTION 4.0
CONFIGURATION
4. SCROLL (é) to display on or oFF and ENTER it into memory. If interval configuration is required, proceed to Step 5, otherwise Step 10.
5. SCROLL (é ) to display desired menu item. If in
t
is selected, go to Step 6, otherwise Step 10.
6. SCROLL (é ) to display desired interval period and SELECT it. Numerical Display will flash.
7. SCROLL ( é ) and SHIFT (ç ) to display the desired value and ENTER it into memory. Display will return to interval period menu.
8. Repeat Steps 6 and 7 as needed.
9. Press the ENTER key to return to the main timer menu.
10. SELECT the desired item. The Numerical Display will flash.
11. SCROLL (é ) and SHIFT (ç ) to display the desired value and ENTER it into memory.
12. Repeat Steps 5, 10, and 11 as required.
13. Press the ENTER key to return to Set Menu.
4.3 INTERVAL TIMER. Display Mnemonic Int. This item is used to set the interval timer's relay logic. The timer can be used for sensor maintenance, such as a wash cycle to clean the sensor in a bypass line. Choices are:
A. Interval Timer Enable/Disable. Display Mnemonic tOn. Select this item to begin interval cycle on or disable interval cycle oFF.
B. Interval Period. Display Mnemonic int. Select this item to set the time period between control cycles. SEC for seconds, uin for minutes, hr for hours, and dAY for days. May be set from a minimum of 10 minutes.
C. On Periods Per Cycle. Display Mnemonic cnt. Select this item to enter the number of on periods per cycle. May be set from 1 to 60 on periods.
D. Duration of On Periods. Display Mnemonic ont. Select this item to enter the relay activation time for each on period. May be set from 0.1 to 299.9 sec­onds.
E. Duration of OFF Periods. Display Mnemonic OFt. Select this item to enter the relay deactivation time between each on period during the control cycle. Valid when cntis 2 or greater. May be set from 0 to 299.9 seconds.
F. Sensor Recovery Time. Display Mnemonic dur. Select this option to enter the duration time after the last on period in a cycle. May be set from 0 to 299.0 seconds. The wait duration can be used for electrode recovery after a wash cycle.
G. Interval Time remaining. Display Mnemonic tiL. Select this item to display the time remaining to the next control cycle. If selected during the control cycle, display will show ---.
NOTE
The Model 1054B is placed on hold during the control cycle (from first on period through the wait duration). The analyzer will simulate a fault condition and briefly show itr every eight seconds. The dis­play will continue to show the measured value.
4.3.1 Interval Timer Configuration (Int).
Refer to Figure 4-2 and Figure 4-3.
1. Enter Set Mode by pressing ACCESS Key twice.
2. SCROLL (é) until Intappears on the display.
3. SELECT to move to the next menu level. tOn, will display.
SEt
OFF
On
tOn
int
cnt
ont
OFt
dur
ti L
SEC
uin
hr
dAY
Int
FIGURE 4-3.
Timer Diagram for One Cycle
RELAY
ACTIVATION
TIME
int
ont
dur
cnt = 1 OFt = 0
Figure 4-2. Interval Timer Configuration
23
MODEL 1054B C SECTION 4.0
CONFIGURATION
4.4 TEMPERATURE. Display Mnemonic t-C. Select this item
for temperature reading and compensation choices.
A. Temperature Display. Display Mnemonic d-t. Select this item to toggle between °F and °C temperature dis­play. The analyzer will show all temperatures in units selected until the selection is changed.
B. Automatic Temperature Compensation. Display Mnemonic A
t
c. The analyzer will use the temperature
input from the sensor for temperature correction when on is selected. When oFF is selected, the analyzer will use the value entered by the user for temperature correction. This manual temperature option is useful if the temperature sensor is faulty or not on line. Temperature specific faults will be disabled (refer to Section 8.0).
4.4.1 Temperature Configuration t-C. Refer to Figure 4-4.
1. Enter Set Mode by pressing ACCESS key twice
2. SCROLL (
é) until
t
-C appears on the display.
3. SELECT to move to the next menu level. d-
t
will
display.
4. SCROLL (
é) to display desired item then SELECT it.
5. If d-
t
is selected, display will show °C or °F.
If A
t
c is selected, display will show on or oFF.
6. SCROLL (
é) then ENTER desired item into memory.
7. If °C, °F or on are entered, display will return to the previous level (proceed to Step 9).
If oFF is selected, numerical display will flash indicat­ing that a process temperature value is required (proceed to Step 8).
8. Use SCROLL (
é ) and SHIFT (ç) to display the
desired value. ENTER value into memory.
9. Repeat Steps 4-8 as required for other item.
10. Press the ENTER key to return to Set Menu.
4.5 CURRENT OUTPUT. Display Mnemonic is OUt. This item is used to configure the output signal.
A. Output Dampening. Display Mnemonic dPn. Dampens the response of the signal output. This option is useful to minimize the effect of a noisy reading. The number entered is the sample time (in seconds) for an averaged output. Zero to 255 seconds may be entered.
B. mA Output Range. Display Mnemonic CUr. Selection of this item will allow choice of 0 to 20 mA or 4 to 20 mA output range.
C. Display Output. Display Mnemonic d-O. This item is used to select logic of output display. Selecting this item will allow the analyzer to display current output as mA (doc) or as a percent of full scale output range (Pc
t
).
4.5.1 Current Output Configuration OU
t
. Refer to
Figure 4-5.
1. Enter Set Mode by pressing the ACCESS key twice.
2. SCROLL (
é) until OU
t
appears on the display.
3. SELECT to move to the next menu level. dPn will display.
4. SCROLL (
é) then SELECT desired item.
5. If dPn is selected, numerical display will flash indicat­ing that a value is required (proceed to Step 6).
If CUr or d-O is selected, proceed to Step 7.
6. SCROLL (
é) then SHIFT (ç) to display the desired
value. ENTER into memory
7. SCROLL (
é) then ENTER desired item.
8. Repeat Steps 4-7 as required.
9. Press the ENTER key to return to the Set Menu.
t-C
SEt
Out
SEt
d-t
Atc
on
oFF
*F
*C
dPn
CUr
d-O
doc
Pct
420
020
Figure 4-5. Current Output Configuration
Figure 4-4. Temperature Configuration
24
MODEL 1054B C SECTION 4.0
CONFIGURATION
4.6 DEFAULTS. Display Mnemonic dFt. This item is
used to set the configuration of relays and output default conditions during fault or hold status. See Table 8-1 for a listing of the possible fault conditions which can be diagnosed by the analyzer. A hold sta­tus is initiated by pressing the HOLD key twice. (Press twice again to remove the hold.)
A. Relay 1 and 2. Display Mnemonic rL1 and rL2. The relays can be set to activate on, deactivate oFF, or hold present status non. See Table 4-2.
B. Current Output. Display Mnemonic Cur.The cur- rent output is held non or goes to a specified value cur during a fault condition. cur will probably be the most informative selection.
C. Fault History. Display Mnemonic SHO. Selecting this item will display the most recent detected faults. Press the SCROLL key once for each previous fault history. Pressing ACCESS will clear SHO history.
4.6.1 Default Configuration (dFt).
Refer to Figure 4-6.
1. Enter Set Mode by pressing the ACCESS key twice.
2. SCROLL (é
) until dFt appears on the display.
3. SELECT to move to the next menu level. rL1 will display.
4. SCROLL (é) then SELECT desired item.
5. Display will show next item selection. SCROLL (é) and ENTER desired item.
6. Repeat Steps 4 and 5 as required for other default settings rL2 and Cur. If cur is selected for Cur, press ENTER then use the SCROLL (é) and SHIFT (ç) keys to enter the desired current value in mA.
7. Press the ENTER key to return to Set Menu.
ANALYZER CONDITION
NORMAL HOLD FAULT
Set menu AL1/AL2 setting Set menu AL1/AL2 setting Set menu AL1/AL2 setting
On OFF FLt On OFF FLt On OFF FLt
(Alarm 2 (Alarm 2 (Alarm 2
only) only) only)
on Proc. det. + + +
oFF Proc. det. +
non Proc. det. Proc. det. Proc. det. +
Set Menu
default
(dFt)
setting
rL1/rL2
Proc. det.: Alarm state is determined by
the process value.
+:Relay will activate.
–:Relay will not activate.
Example: If you want the analyzer to activate relay 1 in
hold mode during calibration, set AL1 to On in Section 4.3, and set rL1 to on.
TABLE 4-2. Relay States for Various Conditions and Alarm/Default Configurations
SEt
dFt
rL1
rL2
Cur
SHO
on
oFF
non
non
cur
Figure 4-6. Default Configuration
25
MODEL 1054B C SECTION 4.0
CONFIGURATION
4.7 ALARM SETPOINT. The alarm setpoints should
be adjusted after completing the configuration proce­dure outlined in Sections 4.1 to 4.6 (Refer to Figure 4-7).
1. Press the PV key to ensure that the analyzer is not in Set Mode.
2. Press the ALARM 1 or ALARM 2 key. SP1 or SP2 will show briefly, followed by the Alarm 1 or Alarm 2 Setpoint.
NOTE:
If the alarm is set to OFF or FAULT (Alarm 2 only), the analyzer will display oFF or F1 t respectively (refer to Section
4.2, Alarm Configuration).
3. Press SELECT to adjust the value. The display will acknowledge briefly with AdJ followed by the Numeric Display with digit flashing.
4. SCROLL ( é ) and SHIFT (ç ) to display the desired value.
5. ENTER value into memory.
6. Repeat Steps 2 to 5 for the second setpoint.
NOTE
Selection of µS/mS and decimal posi­tions is achieved by pressing SHIFT (ç) until the µS/mS flag flashes, then SCROLL (é) until the desired combina- tion of decimal position and mS (quick flashing)/µS (slow flashing) flag are dis­played.
NOTE
Alarm logic may be changed from nor­mally open (N.O.) to normally closed (N.C.) by cutting circuits (W5, W7 & W9) on the power supply PCB and adding jumpers (W4, W6, & W8).
ZERO
ALARM
1
F.S.
ALARM
2
ACCESS
ç
SELECT
é
AdJ
SP1/2
ç
SELECT
ENTER
Press
Once
Press Once
Displays
Briefly
Displays
Briefly
Numeric
Display
Change to
desired
value
Press Once
Numeric
Display
of
Setpoint
FIGURE 4-7. Alarm Setpoint
26
MODEL 1054B C SECTION 4.0
CONFIGURATION
4.8 OUTPUT SCALE EXPANSION. This section should
be followed if it is desired to scale the current output range other than the factory setting of 0-20 millisiemen. The output zero and full scale value should be adjusted after completing the configuration procedure as outlined in Sections 4.1 to 4.6 (refer to Figure 4-8).
A. ZERO POINT (0 mA or 4 mA) (LO)
1. Press the PV key to ensure that the unit is not in Set Mode.
2. Press the ALARM 1 key twice. The display will show LO briefly then display the ZERO point.
3. Press SELECT to adjust the value. The display will acknowledge briefly with AdJ followed by the Numeric Display with digit flashing.
4. SCROLL ( é ) and SHIFT (ç ) to display the desired value.
5. ENTER value into memory. The display will show LO and display the entered value.
B. Full Scale (F.S.) Point (20 mA) (HI)
1. Press the PV key to ensure that the analyzer is not in Set Mode.
2. Press the ALARM 2 key twice. The display will show HI briefly then display the FULL SCALE point.
3. Press SELECT to adjust the value. The display will acknowledge briefly with AdJ followed by the Numeric Display with digit flashing.
4. SCROLL ( é ) and SHIFT (ç ) to display the desired value.
5. ENTER value into memory. The display will show HI and display the entered value.
NOTE
For a reverse output, enter the higher value for zero, and the lower value for the Full Scale.
NOTE
Selection of µS/mS and decimal positions is achieved by pressing SHIFT (ç) until the µS/mS flag flashes, then SCROLL (é) until the desired combination of deci­mal position and mS (quick flashing)/µS (slow flashing) flag are displayed.
ZERO
ALARM
1
F.S.
ALARM
2
ACCESS
ç
SELECT
é
AdJ
LO/HI
ç
SELECT
ENTER
Press Twice
Press Once
Displays
Briefly
Displays
Briefly
Numeric
Display
Change to
desired
value
Press Once
Numeric
Display
of
Setpoint
FIGURE 4-8. Output Scale Expansion
27
4.9 SIMULATE CURRENT OUTPUT. The output can
be simulated to check the operation of devices such as valves, pumps, or recorders. The output can be simulated in either current (mA) or percent of full scale, depending on how the output display d-O was configured in Section 4.5 (Refer to Figure 4-9).
A. Simulate Output in Percent SiP. The output can be simulated in percent if d-O in Section 4.5 was con­figured to display percent Pct.
1. Press the PV key once to insure that the analyzer is not in the Set Mode.
2. Press the OUTPUT key twice. The display will show Pct briefly, then display the output value in percent of full scale.
3. Press SELECT to simulate the output. The dis­play will briefly acknowledge with SiP followed by the Numeric Display with digit flashing.
4. SCROLL (
é ) and SHIFT ( ç ) to display the
desired value.
5. ENTER value into memory. The display will show Pct and display the entered value. Also, the dis­play will flash to acknowledge that the analyzer is placed on hold HLd. In hold mode the relays will be set as determined in Section 4.6.
6. To remove the analyzer from hold, press the HOLD key twice. The hold flag on the display will be removed and the display will stop flashing.
B. Simulate Output in Current SiC. The output can be simulated in mA units if d-O in Section 4.5 was con­figured to display current doc.
1. Press the PV key once to insure that the analyz­er is not in the Set Mode.
2. Press the OUTPUT key twice. The display will show dOC briefly, then display the output value in mA.
3. Press SELECT to simulate the output. the display will briefly acknowledge with Sic followed by the Numeric Display with digit flashing.
4. SCROLL (
é ) and SHIFT ( ç ) to display the
desired value.
5. ENTER value into memory. The display will show dOC and display the entered value. Also, the dis­play will flash to acknowledge that the analyzer is placed on hold HLd. In hold mode the relays will be set as determined in Section 4.6.
6. To remove the analyzer from hold, press the HOLD key twice. The hold flag on the display will be removed and the display will stop flashing.
MODEL 1054B C SECTION 4.0
CONFIGURATION
OUTPUT
COND
ACCESS
ç
SELECT
é
SiC/SiP
Pct/dOC
ç
SELECT
ENTER
Press Twice
Press
Once
Displays
Briefly
Displays
Briefly
Numeric
Display
Change to
desired
value
Press
Once
Numeric
Display
of Output
(Analyzer in
hold)
FIGURE 4-9. Simulate Current Output
28
MODEL 1054B C SECTION 5.0
START-UP AND CALIBRATION
SECTION 5.0
START-UP AND CALIBRATION
5.1 GENERAL. Calibration and operation of the Model
1054B should begin only after completion of the con­figuration of the analyzer. The sensor must be wired (including J-box and interconnecting cable) as it will be in operation.
NOTE READ THE ENTIRE CALIBRATION SECTION TO DETERMINE THE CALI­BRATION PLAN MOST SUITABLE FOR YOUR NEEDS.
5.2 Entering the Cell Constant. The first time the
analyzer is calibrated and any time there is a sensor change, the sensor cell constant must be entered into memory. Entering a cell constant into memory will reset the cell factor Fct to 1.0 and will initiate the ana­lyzer (the cell factor gives an indication of sensor scal­ing. Refer to Section 8.2.3).
1. Enter the Set Mode. Press the ACCESS key twice in rapid succession. The analyzer will dis­play SEt briefly then display Cin.
2. SCROLL (
é ) the menu until CEL is displayed,
then SELECT it. The Numerical display will flash to indicate that a value is desired.
3. Use SCROLL (é) and SHIFT (ç) to display the correct sensor cell constant and ENTER it into memory. This value can be found on the cable label (i.e., Sensor K= 1.00).
NOTE
Only adjust the cell constant when the con­ductivity sensor is replaced or serviced. Then always perform a restandardization. See Section 5.5.
5.3 Temperature Calibration. For accurate tempera-
ture correction, the temperature reading may need adjusting. The following steps must be performed with the sensor in the process or in a grab sample. For the most accurate results, standardization should be per­formed at or near operating temperature.
1. Observe the analyzer temperature reading by pressing the TEMP key. Allow the reading to sta­bilize to insure that the sensor has acclimated to the process temperature.
2. Compare the analyzer reading to a calibrated temperature reading device. If the reading requires adjusting, proceed to Step 3, otherwise, go to Section 5.4.
3. Press the TEMP key then the SELECT key to correct the temperature display. The analyzer will display AdJ briefly, then the Numeric Display will show with digit flashing.
4. SCROLL (é ) and SHIFT (ç) to display the cor­rect value and ENTER it into memory. Proceed to Section 5.4.
5.4 Initial Loop Calibration. Please read the entire
calibration section before proceeding to determine the best plan to follow.
A. Two Point Calibration - Standard Method. This is the recommended procedure for the initial calibration if the process's temperature slope is unknown. If any of the steps below are impossible or impractical, refer to the alternate Section 5.4 B.
1. Obtain a grab sample of the process to be meas­ured.
2. Determine the sample's conductivity using a cali­brated bench or portable analyzer. The analyzer must be able to reference the conductivity to 25°C, or the solution must be measured at 25°C. Note the reading. Insure that the analyzer is in hold. Press the HOLD key twice and observe the solid flag.
3. Immerse the analyzer's sensor into the process solution. The sensor body must be held away from the bottom and sides of the sample's con­tainer and the sensor cable must not be allowed to contact the solution. Shake the sensor to ensure that no air bubbles are present.
4. Adjust the sample's temperature to either the nor­mal high or normal low temperature of the process. To raise the sample's temperature, a hot plate with stirrer is recommended. To lower the process temperature, place the grab sample's container in an ice bath or let it slowly cool down.
29
MODEL 1054B C SECTION 5.0
START-UP AND CALIBRATION
A. Two Point Calibration. (continued)
5. Allow the sensor to acclimate to the solution. (The temperature reading should be stable.)
6. Press the CAL key. 15t displays briefly (if 2nd dis­plays, press CAL again), then the Numeric Adjustment window displays.
7. SCROLL (
é) and SHIFT (ç) to display the grab
sample's conductivity value at 25°C as noted in Step 2, then ENTER into memory.
8. Adjust the sample's temperature to the other nor­mal temperature extreme of the process. To raise the sample's temperature, a hot plate with stirrer is recommended. To lower the process temperature, place the grab sample container in an ice bath.
9. Allow the sensor to acclimate to the solution. (The temperature reading should be stable.)
10. Press the CAL key. 15t displays briefly (If 2nd dis­plays, press CAL again), then the Numeric Adjustment window displays.
11. SCROLL (é ) and SHIFT (ç) to display the grab sample's conductivity value 25°C as noted in Step 2, then ENTER into memory.
The analyzer will then calculate the true cell constant and the temperature slope then return to reading con­ductivity. The temperature slope of the process can now be read. Press the CAL key twice. The display will show SLP briefly then the calculated slope for the two calibration points. Place the sensor in the process, then remove the analyzer from hold by pressing the HOLD key twice again.
The slope may be calculated from the following formu­la:
% SLOPE/°C=
Conductivity T
max
(
Conductivity T
min
—1
)
T
X100
Where: Conductivity T
max
is the conductivity at the
maximum process temperature, Conductivity T
min
is the conductivity at the lower process temperature, and the ³T is the difference between the maximum and mini­mum process temperature.
EXAMPLE:
% SLOPE/°C=
45K
(
35K
—1
)
X100
60-50=10
=2.8%/°C
B. Single Point Calibration - Slope Known. This is the recommended procedure for the initial calibration if the temperature slope of the process is known.
If you do not know the exact temperature slope value, but wish to approximate it, refer to the following guide. However, the conductivity reading may have reduced accuracy compared to the value if the procedure in Section A is performed.
Acids: 1.0 to 1.6% per °C Bases: 1.8 to 2.2% per °C Salts: 2.2 to 3.0% per °C Water: 2.0% per °C
1. Press the CAL key twice. The analyzer will display SLP briefly, then show the temperature slope in memory.
2. SELECT to change the value. The analyzer will display AdJ briefly, then show the Numeric Display window.
3. SCROLL (
é) and SHIFT (ç) to display the proper
temperature slope for the process to be meas­ured, then ENTER into memory.
4. Obtain a grab sample of the process to be measured.
5. Determine the conductivity of the sample using a calibrated bench instrument or portable analyzer. The instrument must be able to reference the con­ductivity to 25°C or the solution must be measured at 25°C. Note the reading. Insure that the analyzer is in hold. Press the HOLD key twice and observe the solid flag.
6. Press the PV key once then press the SELECT key once. Std will display followed by the Numeric Display with digit flashing.
7. SCROLL (é ) and SHIFT (ç) to display the con­ductivity value you noted in Step 5, then ENTER it into memory.
8. Install the sensor in the process, then remove the analyzer from hold by pressing the HOLD key twice.
The analyzer will calculate the true cell constant after the initial calibration.
30
5.5 Routine Standardization. The sensor should be
standardized routinely if it is suspected that the process might degrade or coat the sensor. When a sensor cell constant is entered CEL is set to this value and the cell factor Fct is set to 1.000. The first stan­dardization recalculates the cell constant CEL. Subsequent standardizations will change the cell fac­tor Fct. Refer to Section 8.2.3 for a description of the cell factor.
To perform a standardization do the following:
1. Take a grab sample which is as close to the sen­sor as possible. Write down the value the analyz­er is reading at this time (C1).
2. Measure the conductivity of the grab sample using a calibrated bench analyzer referenced to 25°C/77°F or measured at 25°C. Write down this value (C2).
3. Before entering the reference value, note the value the analyzer is reading now (C3) and com­pare it to the value in Step 1. This accounts for the change while the grab sample is being measured.
4. Press the PV key once, then press SELECT. Std will display briefly followed by the Numeric display with flashing digit.
5. The corrected conductivity reference value may be determined by multiplying the value in Step 2 (C2) by the value noted in Step 3 (C3) and divid­ing the product by the analyzer value from Step 1 (C1):
C2x C
3
= C
RV
C
1
Enter this corrected reference value in the analyz­er using the SCROLL (é ) and SHIFT (ç) keys. Then press ENTER.
6. Note the cell factor value Fct. Press the ACCESS key twice quickly. SCROLL (é ) to Fct press SELECT and note this value. Keep track of this value to determine a sensor cleaning schedule.
5.6 Sensor Maintenance. Before performing mainte­nance or cleaning of the sensor, the Model 1054B C should be placed in hold. This will place the current output and relays in the states determined in Section 4.6. Before removing the sensor from the process, press the HOLD key twice. The HOLD flag will show to indicate the hold condi­tion.
Always reenter the cell constant (Section 5.2) and restandardize (Section 5.5) after cleaning or replace­ment of the sensor.
Replace the sensor back into the process and press the HOLD key twice again to remove the analyzer from hold. The hold flag will disappear.
MODEL 1054B C SECTION 5.0
START-UP AND CALIBRATION
31
MODEL 1054B C SECTION 6.0
KEYBOARD SECURITY
SECTION 6.0
KEYBOARD SECURITY
THEORY OF OPERATION. This section is a general
description of how the analyzer operates. This section is for those users who desire a greater understanding of the analyzer’s operation.
A square wave measurement circuit in the Model 1054B C Analyzer replaces the typical bridge circuit used in most conductivity analyzers, resulting in improved linearity, accuracy and a broad measurement range. The analyzer measures the absolute conductiv­ity of the measured process. The analyzer then cor­rects the conductivity to 25°C by accurately measuring the process temperature by means of a PT-100 RTD located in the conductivity sensor. The microprocessor also adjusts the amount of correction required for tem­perature compensation by means of a temperature slope adjustment.
The slope may be adjusted between 0-5%/°C either manually via the keyboard or automatically during bench or process calibration. This slope controls the amount of correction required in the temperature com­pensation circuit, and is specific to the process, giving you the most accurate conductivity reading possible.
The Model 1054B C analyzer can provide conductivity measurements as low as 1 uS/cm and as high as 1000 mS/cm full scale over a process temperature range of 0 to 200°C.
Rosemount Analytical also offers a booklet titled Conductance Data for Commonly Used Chemicals. This booklet includes measurement theory and conductance information for commonly used chemicals.
6.1 GENERAL. Display Mnemonic COd. Select this fea- ture to display the user defined security code. Any three digit number may be used for this code. 000 will disable the security feature. This item is used to pre­vent accidental changes to the calibration and configu­ration of the analyzer. When activated, the analyzer will allow all read functions to read normally. If an attempt is made to change a value, LOC will display followed by the Numeric Display ready for the code to be entered. A proper code will unlock the analyzer and the analyzer will return to the last function attempted. Any incorrect value will result in bAd briefly displaying. The analyzer will then return to numeric display and await the entry of the code. Once unlocked, the analyzer will allow access to all functions until the analyzer is either pow­ered down or no keystrokes are made for a period of two (2) minutes. If the code should be forgotten, press­ing and holding the ACCESS key for five (5) seconds will result in display of the code. Releasing the ACCESS key, then pressing ENTER will unlock the analyzer.
6.2 ACCESS CODE (COd).
1. Enter Set Mode by pressing ACCESS key twice.
2. SCROLL (
é) until COd appears on the display.
3. Press SELECT.
4. SCROLL (
é ) and SHIFT ( ç ) to display the
desired value, then ENTER it into memory.
NOTE
Entering 000 disables the keyboard securi­ty.
NOTE
Security feature will not activate until two (2) minutes without keyboard activity or power is removed from the analyzer then restored.
SECTION 7.0
THEORY OF OPERATION
32
MODEL 1054B C SECTION 8.0
DIAGNOSTICS AND TROUBLESHOOTING
SECTION 8.0
DIAGNOSTICS AND TROUBLESHOOTING
Display Description
EEP EEPROM write error (bad EEPROM chip).
CHS ROM failure (check sum error) (bad ROM chip).
Orn Overrange.
SEn Sensor line error or wire length error.
COP Computer not operating properly.
tcH High temperature compensation error.
tcL Low temperature compensation error.
Ein Input shorted.
rin Sensor miswired.
FAC Factory calibration required.
8.1 DIAGNOSTICS. The Model 1054B analyzer has a diagnostic feature which automatically searches for fault conditions that would cause an error in the meas­ured conductivity value. If such a condition occurs, the current output and relays will act as configured in default and the fault flag and display will flash. A fault code mnemonic will display at frequent intervals. If more than one fault condition exists, the display will sequence the faults at one second intervals. This will continue until the cause of the fault has been correct­ed. Display of fault mnemonics is suppressed when in Set Mode. Selecting the SHO item will display a history of the two most recent fault conditions unless SHO was cleared (Refer to Section 4.6).
NOTE
If the analyzer is in hold and a fault occurs, the mnemonic HLd will display during the fault sequence.
8.1.1 Fault Mnemonics. Table 8-1 (below) lists the fault mnemonics and describes the meaning of each.
8.1.2 Temperature Compensation. Table 8-2 (below)
is a ready reference of RTD resistance values at vari­ous temperatures. These are used for test and evalu­ation of the sensor.
NOTE
Ohmic values are read across the T.C. element and are based on the stated values (R
O
± .12%). Allow enough time for the T.C. element to stabilize to the surrounding temperature. Each 1°C change corresponds to a change of
0.385 ohms.
TABLE 8-1. Fault Mnemonics
TABLE 8-2. RTD Resistance Values
Temperature Resistance
0°C 100 ohms
10°C 103.90 ohms
20°C 107.70 ohms
25°C 109.62 ohms
30°C 111.67 ohms
40°C 115.54 ohms
50°C 119.40 ohms
60°C 123.24 ohms
70°C 127.07 ohms
80°C 130.89 ohms
90°C 134.70 ohms
100°C 138.50 ohms
110°C 142.29 ohms
120°C 146.06 ohms
130°C 149.82 ohms
140°C 153.58 ohms
150°C 157.31 ohms
160°C 161.04 ohms
170°C 164.76 ohms
180°C 168.46 ohms
190°C 172.16 ohms
200°C 175.84 ohms
33
MODEL 1054B C SECTION 8.0
DIAGNOSTICS AND TROUBLESHOOTING
8.2 TROUBLESHOOTING. The Model 1054B analyz-
er is designed with the state of the art microprocessor circuitry. This design incorporates programmed fea­tures that provide constant monitoring for fault condi­tions, and the reporting of these faults via Mnemonics on the instrument display screen. This aids in deter­mining where to start checking for the cause of failures, and in some instances, the ability to see changes that can be used to predict future degeneration of assem­blies before their complete failure.
8.2.1 Installation Failure. After completion of installa­tion the instrument should be checked for operation. Normally this would consist of Powering up the instru­ment and checking for:
1. A self diagnostic fault display. Refer to Table 8-1 for brief description of problem indicated by mnemonic. Table 8-3 provides a more compre­hensive problem explanation and actions that may help solve the problem.
2. A conductivity reading that is approximately cor­rect (depending upon sensor installation in either air or process). Refer to Section 8.2.3 for sensor checks.
3. Pressing several of the keypads to determine whether programming appears to be operational. Table 8-3 explains problems and actions that may be helpful in solving them.
4. Checking output for 4-20 mA output current.
8.2.2 After Operation. Troubleshooting this instrument
after previous operation should follow normal trou­bleshooting procedures. Check display. If power is O.K. the display mnemonic will direct you to the basic area of malfunction (Sensor, Printed Circuit Boards, calibra­tion, or temperature compensation).
Use Table 8-1 and Table 8-3 to determine area, possi­ble problem and actions to take to remedy fault.
Evaluate instrument electronics. This can be accomplished by simulating a known conductivity input and observation of instrument operation. To simulate sensor operation with known conductivity inputs, use the following procedures.
1. Disconnect the Sensor input leads from TB2-1, 3, 6, and 7.
2. Install decade box or resistor leads to TB2-1 and 3. (If decade box is not available, simulate desired conductivity input by either calculating using the for­mula given in Figure 8-1, or by using the Conductivity vs. resistance Table in Figure 8-1.)
3. Install a jumper between TB2-6 and 7. Check wiring with Figure 8-1.
4. Power up instrument and enter SEt menu.
5. Turn Atc to oFF.
6. Set manual temperature compensation to 25°C (See Section 4.4 and Figure 4-4).
7. Set Cell constant to 1.0 (See Section 5.1.1).
8. Evaluate analyzer response with previous responses.
Faulty display. If a faulty display is suspected, enter the SET menu and scroll through to the dtS option. This option will activate all display segments. See Figure 3-4.
Output Circuit Testing. To check for problems in the output circuit, bypass the sensor input and analyzer calculations by setting a known output current and checking item driven by output current and checking the operation of valves, pumps, recorders, etc. For directions on how to set output current, refer to Section 4.9.
8.2.3 Sensor Troubleshooting. In addition to the sen-
sor fault mnemonics, the analyzer can display informa­tion pertinent to determining if sensor has become coated, or if there is a conductivity versus temperature problem, or an application problem.
Sensor Coated. As the cell becomes coated, or affect­ed by the process, the cell factor will change. Tracking this change in cell factor will prevent use of a sensor that has lost its sensitivity because of contamination or damage.
CAUTION
Standardizing the instrument results in the cell factor being returned to 1.0.
This instrument tracks the change in calculated cell factor from the initial cell factor value of 1.0 every time the unit is standardized. The cell factor should be checked and tracked to set up a regular maintenance schedule and can be seen in the following manner:
1. Press ACCESS key twice.
2. SEt will be displayed briefly followed by Cin.
3. SCROLL (
é) to display Fct and press SELECT.
4. To return to normal operation, press PV.
34
MODEL 1054B C SECTION 8.0
DIAGNOSTICS AND TROUBLESHOOTING
TABLE 8-3. Troubleshooting Guide
SYMPTOM PROBLEM ACTION
Fct below 0.5 or above 2.0. 1. Old or coated sensor. 1. Clean or replace sensor. Actual range determined by user.
Analyzer value not the same 1. Grab sample incorrect. 1. Re-evaluate sample technique as grab sample of process. and equipment.
2. Unclear what is correct. 2. Bench test analyzer.
3. Analyzer out of calibration. 3. Recalibrate per Start-up and Calibration Section.
Fault code tcH/tcL/rin. 1. Miswire. 1. Check wiring between the
sensor and analyzer.
2. Open or shorted RTD. 2. Replace sensor.
Fault code Orn. 1. Process conductivity 1. Replace sensor with a sensor
too high for sensor in use. which has a higher cell constant
2. Process upset. (see Table 1-1).
2. Check for process control problem.
Fault code SEn. 1. Open wire between sensor 1. Repair wire/check connection.
and analyzer.
2. Cable length has been exceeded. 2. Locate analyzer within 250 ft.
Maximum cable length 250 ft. of sensor.
Fault code EEP. 1. Defective EEPROM. 1. Replace CPU PCB.
Fault code CHS. 1. Defective CPU. 1. Replace CPU PCB.
No alarm relay closure. 1. Defective power card. 1. Replace power PCB.
2. Defective CPU. 2. Replace CPU PCB.
No output current. 1. Defective power board. 1. Replace power PCB.
2. Miswired. 2. Check for short.
Low output current. 1. Circuit loading with excessive 1. Consult output loading limits
resistance on output. Model 1054B C specifications
(600 ohms max load).
Zero conductivity reading. 1. Sensor miswired. 1. Repair wire/connection.
2. Solids coating sensor. 2. Clean sensor.
3. Open wire in sensor. 3. Replace sensor or tube.
Fault code Ein. 1. Sensor miswired. 1. Repair wire connection. Very high conductivity reading. 2. Shorted sensor. 2. Replace sensor or tube.
35
MODEL 1054B C SECTION 8.0
DIAGNOSTICS AND TROUBLESHOOTING
Absolute Conductivity. As an aid in determining
whether a problem exists in the conductivity section of the sensor or analyzer, or the temperature compensat­ing circuits, the absolute conductivity (the uncorrected conductivity value, without temperature compensation) of the process can be displayed. To do so:
1. Press ACCESS key twice.
2. SEt will be displayed briefly followed by Cin.
3. SELECT Cin to read the absolute conductivity.
4. To return to normal operation, press PV.
Temperature Sensor accuracy. If the temperature sensor in the conductivity sensor is suspect, measur­ing the resistance across the T.C. element and com­paring the corresponding temperature reading can be used in the evaluation of the sensor. Allow enough time for the T.C. element to stabilize to the surrounding tem­perature. Each 1°C change corresponds to a change of 0.385 ohms.
8.2.4 Subassembly Replacement Considerations.
CPU Board Replacement. If a problem exists on the
CPU board, and replacement is required, specific pro­cedures included with the new board must be followed or the microprocessor will be improperly programmed. Should this occur, it will be necessary to return the ana­lyzer to the factory for reprogramming.
Power Board Replacement. If it becomes necessary to replace the power board, the CPU board will need to be recalibrated following the specific procedures that are included with the power board. Failure to follow these procedures exactly will cause the microproces­sor to be improperly programmed and require the return of the analyzer to the factory for reprogramming.
8.2.5 Instrument Electronic Check. This procedure will allow the operation of the analyzer to be evaluated by simulating a known conductivity input.
1. Disconnect the conductivity sensor input leads from TB2-1, 3, 6 and 7. Install decade box or resis­tor leads to TB2-1 and 3 and a jumper to TB2-6 and 7 (see Figure 8-1).
2. With instrument power on, enter the SEt menu and turn Atc to oFF. Set manual temperature compen­sation to 25°C (see Section 4.4 and Figure 4-3).
3. Set cell constant to 1.0 (see Section 5.1.1).
4. To simulate a desired conductivity input, an appro­priate resistance value may be calculated by Formula or selected from the conductivity (µmhos) vs resistance (ohms) table (see Figure 8-1).
5. Simulate conductivity input and evaluate the ana­lyzer response.
8.3 INSTRUMENT MAINTENANCE. To maintain the
appearance and extend the life of the enclosure, it should be cleaned on a regular basis using a mild soap and water solution followed by a clean water rinse.
36
FIGURE 8-1. Simulate Conductivity Input
MODEL 1054B C SECTION 8.0
DIAGNOSTICS AND TROUBLESHOOTING
Formula:
1
Conductivity =
____________
X 1,000,000
Resistance
Table:
Conductivity (µmhos) vs Resistance (ohms)
10 100,000
100 10,000
1,000 1,000 10,000 100 20,000 50
Decade Box or Resistor
Jumper
37
MODEL 1054B C SECTION 9.0
RETURN OF MATERIAL
SECTION 9.0
RETURN OF MATERIAL
9.1 GENERAL. To expedite the repair and return of
instruments, proper communication between the cus­tomer and the factory is important. A return material authorization (RMA) number is required. Call (949) 757-8500. The Return of Materials Request form is provided for you to copy and use in case the situation arises. The accuracy and completeness of this form will affect the processing time of your materials.
9.2 WARRANTY REPAIR. The following is the proce­dure for returning instruments still under warranty.
1. Contact the factory for authorization.
2. Complete a copy of the Return of Materials Request form as completely and accurately as possible.
3. To verify warranty, supply the factory sales order number or the original purchase order number. In the case of individual parts or sub-assemblies, the serial number on the mother unit must be sup­plied.
4. Carefully package the materials and enclose your Letter of Transmittal and the completed copy of the Return of Materials Request form. If possible, pack the materials in the same manner as it was received.
IMPORTANT
Please see second section of Return of Materials Request Form. Compliance to the OSHA requirements is mandatory for the safety of all personnel. MSDS forms and a certification that the instruments have been disinfected or detoxified are required.
5. Send the package prepaid to:
Rosemount Analytical Inc. Uniloc Division 2400 Barranca Parkway Irvine, CA 92606
Attn: Factory Repair
Mark the package: Returned for Repair
RMA# __________________
Model No. ______________
9.3 NON WARRANTY REPAIR.
1. Contact the factory for authorization.
2. Fill out a copy of the Return of Materials Request form as completely and accurately as possible.
3. Include a purchase order number and make sure to include the name and telephone number of the right individual to be contacted should additional information be needed.
4. Do Steps 4 and 5 of Section 9.2.
NOTE
Consult the factory for additional infor­mation regarding service or repair.
FROM: RETURN BILL TO:
_____________________________ _____________________________ _____________________________
_____________________________ _____________________________ _____________________________
_____________________________ _____________________________ _____________________________
CUSTOMER/USER MUST SUBMIT MATERIAL SAFETY SHEET (MSDS) OR COMPLETE STREAM COMPOSITION, AND/OR LETTER CERTIFYING THE MATERIALS HAVE BEEN DISINFECTED AND/OR DETOXIFIED WHEN RETURNING ANY PRODUCT, SAMPLE OR MATERIAL THAT HAVE BEEN EXPOSED TO OR USED IN AN ENVIRONMENT OR PROCESS THAT CONTAINS A HAZARDOUS MATERIAL ANY OF THE ABOVE THAT IS SUBMITTED TO ROSEMOUNT ANALYTICAL WITH- OUT THE MSDS WILL BE RETURNED TO SENDER C.O.D. FOR THE SAFETY AND HEALTH OF OUR EMPLOYEES. WE THANK YOU IN ADVANCE FOR COMPLIANCE TO THIS SUBJECT.
SENSOR OR CIRCUIT BOARD ONLY: (Please reference where from in MODEL / SER. NO. Column)
1. PART NO.__________________________1. MODEL _________________________________1. SER. NO. ________________
2. PART NO.__________________________2. MODEL _________________________________2. SER. NO. ________________
3. PART NO.__________________________3. MODEL _________________________________3. SER. NO. ________________
4. PART NO.__________________________4. MODEL _________________________________4. SER. NO. ________________
PLEASE CHECK ONE:
n REPAIR AND CALIBRATE n DEMO EQUIPMENT NO. __________________________
n EVALUATION n OTHER (EXPLAIN) _______________________________
n REPLACEMENT REQUIRED? n YES n NO _________________________________________________
DESCRIPTION OF MALFUNCTION:
______________________________________________________________________________________________________
______________________________________________________________________________________________________
______________________________________________________________________________________________________
WARRANTY REPAIR REQUESTED:
n YES-REFERENCE ORIGINAL ROSEMOUNT ANALYTICAL ORDER NO. ________________________________________
CUSTOMER PURCHASE ORDER NO. _________________________________________________
n NO-PROCEED WITH REPAIRS-INVOICE AGAINST P.O. NO. _________________________________________________
n NO-CONTACT WITH ESTIMATE OF REPAIR CHARGES: LETTER n __________________________________________
PHONE n ___________________________________________
NAME ____________________________________________________ PHONE _________________________________________
ADDRESS ___________________________________________________________________________________________________
______________________________________________________________ ZIP _________________________________________
RETURN AUTHORITY FOR CREDIT ADJUSTMENT [Please check appropriate box(s)]
n WRONG PART RECEIVED n REPLACEMENT RECEIVED
n DUPLICATE SHIPMENT REFERENCE ROSEMOUNT ANALYTICAL SALES ORDER NO. _________
n RETURN FOR CREDIT RETURN AUTHORIZED BY: ______________________________________
WARRANTY DEFECT____________________________________________________________________________________
_____________________________________________________________________________________________________
24-6047
RETURN OF MATERIALS REQUEST
•IMPORTANT!
This form must be completed to ensure expedient factory service.
R E P A
I
R
S T A T U S
R E A S O N
F O R
R E T U R N
C U
S
T O M
E R
N O T
I C E
T O
S E N D E R
Emerson Process Management
Rosemount Analytical Inc.
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250
http://www.RAuniloc.com
© Rosemount Analytical Inc. 2001
WARRANTY
Seller warrants that the firmware will execute the programming instructions provided by Seller, and that the Goods manufactured or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expira­tion of the applicable warranty period. Goods are warranted for twelve (12) months from the date of initial installation or eighteen (18) months from the date of shipment by Seller, whichever period expires first. Consumables, such as glass electrodes,
membranes, liquid junctions, electrolyte, o-rings, catalytic beads, etc., and Services are warranted for a period of 90 days from the date of shipment or provision.
Products purchased by Seller from a third party for resale to Buyer ("Resale Products") shall carry only the warranty extended by the original manufacturer. Buyer agrees that Seller has no liability for Resale Products beyond making a reasonable commercial effort to arrange for procurement and shipping of the Resale Products.
If Buyer discovers any warranty defects and notifies Seller thereof in writing during the applicable warranty period, Seller shall, at its option, promptly correct any errors that are found by Seller in the firmware or Services, or repair or replace F.O.B. point of man­ufacture that portion of the Goods or firmware found by Seller to be defective, or refund the purchase price of the defective por­tion of the Goods/Services.
All replacements or repairs necessitated by inadequate maintenance, normal wear and usage, unsuitable power sources, unsuit­able environmental conditions, accident, misuse, improper installation, modification, repair, storage or handling, or any other cause not the fault of Seller are not covered by this limited warranty, and shall be at Buyer's expense. Seller shall not be obligat­ed to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in writing in advance by an authorized Seller representative. All costs of dismantling, reinstallation and freight and the time and expenses of Seller's person­nel for site travel and diagnosis under this warranty clause shall be borne by Buyer unless accepted in writing by Seller.
Goods repaired and parts replaced during the warranty period shall be in warranty for the remainder of the original warranty peri­od or ninety (90) days, whichever is longer. This limited warranty is the only warranty made by Seller and can be amended only in a writing signed by an authorized representative of Seller. Except as otherwise expressly provided in the Agreement, THERE ARE NO REPRESENTATIONS OR WARRANTIES OF ANY KIND, EXPRESS OR IMPLIED, AS TO MERCHANTABILITY, FIT­NESS FOR PARTICULAR PURPOSE, OR ANY OTHER MATTER WITH RESPECTTO ANY OF THE GOODS OR SERVICES.
RETURN OF MATERIAL
Material returned for repair, whether in or out of warranty, should be shipped prepaid to:
Emerson Process Management
Liquid Division
2400 Barranca Parkway
Irvine, CA 92606
The shipping container should be marked:
Return for Repair Model
_______________________________
The returned material should be accompanied by a letter of transmittal which should include the following information (make a copy of the "Return of Materials Request" found on the last page of the Manual and provide the following thereon):
1. Location type of service, and length of time of service of the device.
2. Description of the faulty operation of the device and the circumstances of the failure.
3. Name and telephone number of the person to contact if there are questions about the returned material.
4. Statement as to whether warranty or non-warranty service is requested.
5. Complete shipping instructions for return of the material. Adherence to these procedures will expedite handling of the returned material and will prevent unnecessary additional charges
for inspection and testing to determine the problem with the device.
If the material is returned for out-of-warranty repairs, a purchase order for repairs should be enclosed.
Credit Cards for U.S. Purchases Only.
The right people, the right answers, right now.
ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
http://www.raihome.com
Emerson Process Management Liquid Division
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250
http://www.raihome.com
© Rosemount Analytical Inc. 2003
Loading...