Rosemount 1057 Three-Input Intelligent Analyzer Manuals & Guides

Reference Manual
0809-0100-3157
0
Rev. AD
March 2020
Rosemount
1057
Essential Instructions
Read this page before proceeding
Your instrument purchase from Emerson is one of the finest available for your particular application. These instruments have been designed, and tested to meet many national and international standards. Experience indicates that its performance is directly relat­ed to the quality of the installation and knowledge of the user in operating and maintaining the instrument. To ensure their contin­ued operation to the design specifications, personnel should read this manual thoroughly before proceeding with installation, commissioning, operation, and maintenance of this instrument. If this equipment is used in a manner not specified by the manufac­turer, the protection provided by it against hazards september be impaired.
ailure to follow the proper instructions september cause any one of the following situations to occur: Loss of life; personal
F
injury; property damage; damage to this instrument; and warranty invalidation.
Ensure that you have received the correct model and options from your purchase order. Verify that this manual covers your
model and options. If not, call 1-800-854-8257 or 949-757-8500 to request correct manual.
For clarification of instructions, contact your Rosemount representative.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Use only qualified personnel to install, operate, update, program and maintain the product.
Educate your personnel in the proper installation, operation, and maintenance of the product.
Install equipment as specified in the Installation section of this manual. Follow appropriate local and national codes. Only
connect the product to electrical sources specified in this manual.
Use only factory documented components for repair. Tampering or unauthorized substitution of parts and procedures can
affect the performance and cause unsafe operation of your process.
All instrument enclosures must be closed and protective covers must be in place unless qualified personnel are performing
maintenance.
WARNING
RIS
K
O
F
E
L
ECTRI
CAL
S
HO
C
K
I
ns
tallation
M
ain pow
Do
not
Signal
N
on-
j
umper
U
nuse
wi
t
h p
con
d
Elect n
ation
Ope
r
Saf
ety an
Proper
Equipment pr
er
oper
w
ir
ing c
metallic
w
ir
es
d
ca
b
e
rson
a
u
i
t
p
l
u
g
r
ical inst
al o
r lo
ate
o
nly with
d perf
use
an
and
w
ir
ate
onnec
c
able s
.
l
e
co
l
sa
s t
o ma
allation
cal co
o
d con
otec
s
er
vic
ed to
or
ener
nd
ui
fe
t
y a
i
n
de
f
ro
rm
ance requ
figu
ted
thr
oughout
ing of
this
s
epar
ate pow
giz
e ins
tr
ted in this
train
t
e nd t
a
mu
s. n
t pan
box
reliefs
do not
ntr
i
e
s m
ust
e
nv
i
ronm
i
n
t
he
i
n
g
re
st
be
in
e
l f
astened and
ire that this instrument be connected and properly grounded through a three-wire power source.
r
atio
n
is
th
pr
oduc
er
s
ourc
ument w
mus
b
e
se
e
n
t
a
l
p
ss p
rot
accordan
e
r
espo
by
double ins
t
s
eptember
e
mus
ith
c
t be
rated
pr
ovide
cur
e
l
y
rot
e
ct
e
ct
i
on
ce with the National Electrical Code (ANSI/NFPA-70) and/or any other applicable
in place.
n
sibility of the user.
ex
t
be
as
e open!
at
leas
gr
ounding betw
se
a
l
e
d
i
on re
q
ra
t
i
n
g
ulation.
pos
e
dis
c
onnec
t
240 V
b
y
no
u
i
re
me
(
Typ
e
per
n-f
n
4X)
l
a
t
s. Un
s
onnel
to
ted before
for
Eur
opean
een
c
onduit
m
m
a
b
l
e
u
se
d
con
dangerous
s
er
vic
ing.
mains
c
onnec
cl
o
sur
e
s
d
u
i
t
op
voltages
operation.
tions
to
p
r
o
vi
e
n
i
n
g
!
U
d
e
s mu
e
s
.
e
ncl
st
gr
ounding
o
sur
b
e
se
e
i
nte
a
l
e
d
wi
ty
g
t
r
i h Typ
pe bus
ty i
n co
hings
m
p
e
4X or IP
l
i
WARNING
Physical access
Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of end users’ equip­ment. This could be intentional or unintentional and needs to be protected against.
Physical security is an important part of any security program and fundamental to protecting your system. Restrict physical access by unauthorized personnel to protect end users’ assets. This is true for all systems used within the facility.
a
and
nce
66
Essential Instructions i
Quick Start Guide
1. Refer to Section 2 for mechanical installation instructions.
2. Wire sensor(s) to the signal boards. See Section 3 for wiring instructions. Refer to the sensor instruction sheet for additional details. Make current output, alarm relay and power connections.
3. Once connections are secured and verified, close panel and apply power to the analyzer.
WARNING
R
IS
K
O
F
E
LE
CT
R
I
CA
L
S
H
O
C
K
El
ec
tr
i
c
al
i
n
s
tal
l
ati
on
mu
s
t be
i
n
ac
c
or
dan
c
e
w
i
th
th
e N
ati
on
al
El
ec
tr
i
c
al
Code (A
oth
er
appl
i
c
abl
e
n
ati
on
al
or
l
oc
al
c
odes
.
4. When the analyzer is powered up for the first time, Quick Start screens appear. Quick Start operating tips are as follows:
a. A backlit field shows the position of the cursor.
b. To move the cursor left or right, use the keys to the left or right of the ENTER key. To scroll up or down or to increase or decrease the value of a digit use the keys above and below the ENTER key . Use the left or right keys to move the decimal point.
c. Press ENTER to store a setting. Press EXIT to leave without storing changes. Pressing EXIT during Quick Start returns the display to the initial start-up screen (select language).
5. Complete the steps as shown in the Quick Start Guide flow diagram, Figure A .
6. After the last step, the main display appears. The outputs are assigned to default values.
7. To change output, and temperature-related settings, go to the main menu and choose Program. Follow the prompts. For a general guide to the Program menu, see the Quick Reference Guide, Figure B.
8. To return the analyzer to the default settings, choose Reset Analyzer under the Program menu.
N
S
I
/
N
F
P
A
-
7
0
)
an
d/
or
an
y
Quick Start Guide ii
Quick Start Guide
Figure A. Quick Start Guide
Quick Start Guideiii
Quick Reference Guide
Figure B. Model 1057 Menu Tree
Quick Reference Guide iv
Reference Manual
00809-0100-3157
Contents
Table of Contents
March 2020
Quick Start Guide ...................................................................................................................iii
Quick Reference Guide............................................................................................................v
About This Document............................................................................................................vi
Table of Contents..................................................................................................................vii
Section 1: Description and Specifications
1.1 Features and Applications...........................................................................................1
1.2 Specifications - General................................................................................................2
1.3 Contacting Conductivity (Codes -20, -30, -40).............................................................4
1.4 pH/ORP/ISE (Codes -22, -32, -42).................................................................................5
Section 2: Installation
2.1 Unpacking and Inspection............................................................................................7
2.2 Installation....................................................................................................................7
Section 3: Wiring
3.1 General...................................................................................................................... 11
3.2 Preparing Conduit Openings......................................................................................12
3.3 Preparing Sensor Cable..............................................................................................12
3.4 Power, Output and Sensor Connections ....................................................................12
Section 4: Display and operation
4.1 User Interface.............................................................................................................17
4.2 Instrument Keypad ....................................................................................................17
4.3 Main Display...............................................................................................................18
4.4 Menu System .............................................................................................................19
Section 5: Programming the Analyzer – Basics
5.1 General.......................................................................................................................21
5.2 Changing the Startup Settings...................................................................................21
5.3 Choosing Temperature Units and Automatic/Manual
Temperature Compensation......................................................................................22
5.4 Configuring and Ranging The Current Outputs .........................................................22
5.5 Setting a Security Code ..............................................................................................24
5.6 Security Access...........................................................................................................24
5.7 Using Hold..................................................................................................................25
5.8 Resetting the Factory Default Settings.......................................................................25
5.9 Programming Alarm Relays .......................................................................................26
Table of Contents v
Table of Contents
March 2020
Section 6: Programming Measurements
Section 7: Calibration
Section 8: Return of Material
Reference Manual
00809-0100-3157
6.1 Programming Measurements - Introduction...........................................................31
6.2 pH Measurement Programming................................................................................31
.3ORP Measurement Programming..............................................................................33
6
6.4 Contacting Conductivity ............................................................................................34
7.1 Unpacking and Inspection..........................................................................................41
7.2 pH Calibration ............................................................................................................41
7.3 ORP Calibration..........................................................................................................44
7.4 Contacting Conductivity Calibration..........................................................................45
7.5 Calibrating Temperature............................................................................................48
8.1 General...................................................................................................................... 53
8.2 Warranty Repair .........................................................................................................53
8.3 Non-Warranty Repair .................................................................................................53
Table of Contentsvi
Reference Manual
00809-0100-3157
Section 1: Description and Specifications
Section 1: Description and Specifications
March 2020
M
L
Easy to install – modular boards, removable connectors, easy to wire power, sensors, an
Intuitive menu screens with advanced diagnostics and help screens.
SMART - Enabled - compatible with SMART pH sensors.
Eight menu languages.
Four analog outputs.
UL and CSA approved.
ulti-parameter instrument – up to three inputs. Choose pH/ORP/ISE or contacting
conductivity/resistivity in any combination.
arge display – large easy-to-read process measurements.
outputs.
1.1 Features and Applications
The Model 1057 analyzer offers three sensor inputs and four current outputs thus reducing the cost per loop and saving panel space. The pH signal input board supports pH, ORP, and Ion-Selective Electrode measurements. The conductivity signal input board supports contacting conductivity, resistivity, total dissolved solids, salinity and percent concentration curves for special applications. The modular design allows signal input boards to be field replaced making configuration changes easy. Conveniently, live process values are always displayed during programming and calibration routines. Standard features include isolated inputs, eight embedded local languages, four 4-20mA current outputs, four alarm relays and removable connectors for power and current outputs.
Quick Start Programming: Exclusive quick start screens appear the first time the Model 1057 is powered. The instrument auto-recognizes each measurement board and prompts the user to configure each sensor loop in a few quick steps for immediate deployment.
Menus: Menu screens for calibrating and programming are simple and intuitive. Plain language prompts and help screens guide the user through these procedures.
4-Electrode Conductivity: For applications requiring wide range conductivity measurements, use Rosemount Model 410VP 4-electrode sensor. It is not affected by fouling and is supported by the same contacting conductivity signal board as traditional 2-electrode sensors.
Three Sensor Inputs: The Model 1057 accepts one, two or three isolated inputs. Inputs are isolated from other signal sources and earth ground.
Four Current Outputs: Four 0/4-20 mA current outputs are electrically isolated. Outputs are fully scalable and can be programmed to linear or logrithmic modes. Output dampening can be enabled with time constants from 0 to 999 seconds.
Enclosure: The instrument fits standard ½ DIN panel cutouts. The versatile enclosure design supports panel-mount, pipe-mount, and surface/wall-mount installations.
Smart pH Sensors: Avoid buffer calibrations in the field. Use Rosemount SMART pH sensors to automatically calibrate the measurement loop when connected to Model 1057. Choose from a complete range of SMART pH sensors suited to most applications.
Section 1: Description and Specifications 1
Section 1: Description and Specifications
March 2020
Security Access Codes: Two levels of security access are available. Program one access code for
outine calibration and hold of current outputs; program another access code for all menus and
r functions.
Diagnostics: The analyzer continuously monitors itself and the sensor(s) for problematic conditions.
he display flashes fault and/or warning when these conditions occur.
T
Diagnostics Faults Warnings
ensor 1
S Sensor 2 Sensor 3 Out 1: 12.05 mA Out 2: 12.05 mA Out 3: 12.05 mA Out 4: 12.05 mA 1057PPC03AN Instr SW VER: 3.12 AC Freq. Used: 60Hz
Information about each condition is quickly accessible by pressing DIAG on the keypad. User help screens are displayed for most fault and warning conditions to assist in troubleshooting.
Reference Manual
00809-0100-3157
Display: The high-contrast LCD provides live measurement readouts in large digits and shows up to six additional process variables or diagnostic parameters. The display is back-lit and the format can
be customized to meet user requirements.
Local Languages: Rosemount extends its worldwide reach by offering eight local languages – English, French, German, Italian, Spanish, Portuguese, Chinese and Russian. Every unit includes user programming menus; calibration routines; faults and warnings; and user help screens in all eight lan guages. The displayed language can be easily set and changed using the menus.
1.2 Specifications - General
Enclosure: Polycarbonate. Type: CSA 4X (IP65). Dimensions: Overall 155 x 155 x 131mm (6.10 x 6.10 x 5.15 in.). Cutout: 1/2 DIN 139mm x 139mm
(5.45 x 5.45 in.) Minimum depth for panel mount insatllations 101.6 mm (4.0 in).
Conduit Openings: Accepts 1/2” or PG13.5 conduit fitings Display: Monochromatic graphic liquid crystal display. 128 x 96 pixel display resolution.
Backlit. Active display area: 58 x 78mm (2.3 x 3.0 in.).
Ambient Temperature and Humidity: 0 to 55 °C (32 to 131 °F). RH 5 to 95 (non-condensing) Storage Temperature Effect: -20 to 60 °C (-4 to 140 °F) Power: Code 02: 20 to 30 Vdc. 15 W. Code 03: 84 to 265 Vac, 47.5 to 65.0 Hz, switching. 15 W.
Equipment protected by double insulation
Section 1: Description and Specifications2
U
L
U
S
C
L
I
S
T
E
D
Reference Manual
00809-0100-3157
Section 1: Description and Specifications
March 2020
Hazardous Location Approvals:
Options for CSA:02, 03, 20, 21, 22, 24, 25, 26, 30, 31, 32, 34, 35, 36, 38, 40, 41, 42, 44, 45, 46, 48, UL
Class I, Division 2, Groups A, B, C, & D
Class Il, Division 2, Groups E, F, & G
Class Ill T4 Tamb= 55 °C (applies to all classes) Type 4X, IP66 Non-Incendive Field Wiring (NIFW) may be used when installed per drawing
1400680. The ‘C’ and ‘US’ indicators adjacent to the CSA Mark signify that the product has been evaluated to the applicable CSA and ANSI/UL Standards, for use in Canada and the U.S. respectively.
Evaluated to CSA Standard 22.2 No. 0-10, 0.4-04, 25-1996, 94- M1991, 142-M1987, 213-M1987, 60529-2005/2015. ANSI/IEC 60529-2004/2011. ANSI/ISA 12.12.01:2007. UL No. 50, 11th Ed., 508,17th Ed.
Ordinary Locations (only with UL ordering option):
Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, a
temporary conductivity caused by condensation must be expected.
Altitude: for use up to 2000 meter (6562 ft.)
EMI/RFI effect
Meets all industrial requirements of EN61326.
Analog communication
No effect on the values being given if using a 4-20 mA analog signal with shielded, twisted pair wiring.
LVD: EN 61010-1
Alarm Relays: Four alarm relays for process measurement(s) or temperature. Any relay can be
configured as a fault alarm instead of a process alarm. Each relay can be configured independently and each can be programmed with interval timer settings.
Maximum Relay Current
Resistive
28 Vdc 5.0 A
115 Vac 5.0 A
230 Vac 5.0 A
Inputs: Up to three sensor inputs-electrically isolated.
Relays: Form C, SPDT, epoxy sealed
Inductive load: 1/8 HP motor (max.), 120/240 Vac
Inductive load: Four 4-20 mA or 0-20 mA isolated current outputs. Fully scalable. Max Load: 550
Ohms.
Current Output Accuracy: ±0.05 mA @25 °C
Section 1: Description and Specifications 3
Section 1: Description and Specifications
March 2020
Reference Manual
00809-0100-3157
Terminal Connections Rating:
Power connector (3-leads): 24-12 AWG wire size. Signal board terminal blocks: 26-16 AWG wire size. Current output connectors (4-leads): 24-16 AWG wire size. Alarm relay terminal blocks: 24-12 AWG wire siz
Weight/Shipping Weight: (rounded up to nearest lb or nearest 0.5 kg): 3 lbs/4 lbs (1.5 kg/2.0 kg)
1.3 Contacting Conductivity (Codes -20, -30 and -40)
Measures conductivity in the range 0 to 600,000 µS/cm (600mS/cm). Measurement choices are conductivity, resistivity, total dissolved solids, salinity, and % concentration. The % concentration selection includes the choice of five common solutions (0-12% NaOH,0-15% HCl, 0-20% NaCl, and 0-25% or 96-99.7% H2SO4). The conductivity concentration algorithms for these solutions are fully temperature compensated. Three temperature compensation options are available: manual slope (X%/°C), high purity water (dilute sodium chloride), and cation conductivity (dilute hydrochloric acid). Temperature compensation can be disabled, allowing the analyzer to display raw conductivity. For more information concerning the use and operation of the contacting conductivity sensors, refer to the product data sheets.
Note: When contacting conductivity sensors are used for sensor 1 and sensor 2, Model 1057 can derive an inferred pH value called pHCalc. pHCalc is calculated pH, not directly measured pH.
Note: Selected 4-electrode, high-range contacting conductivity sensors are compatible with Model 1056.
Input filter: time constant 1 - 999 seconds, default 2 seconds Response time: 3 seconds to 100% of final reading Salinity: Uses Practical Salinity Scale
Total Dissolved Solids: Calculated by multiplying conductivity at 25 °C by 0.65 Recommended Sensors For Conductivity
All Rosemount ENDURANCE Model 400 series conductivity sensors (Pt 1000 RTD) and Model 410 sensor.
Temperature Specifications:
Temperature range 0-200 °C
Temperature Accuracy, Pt-1000, 0-50 °C
Temperature Accuracy, Pt-1000, Temp. > 50 °C
± 0.1 °C
± 0.5 °C
Performance Specifications
Recommended Range – Contacting Conductivity
Cell 0.01µS/cm 0.1µS/cm 1.0µS/cm 10µS/cm 100µS/cm 1000µS/cm 10mS/cm 100mS/cm 1000mS/cm Constant
0.01
0.1
1.0
4-electrode
0.01µS/cm to 200µS/cm
0.1µS/cm to 2000µS/cm
1 µS/cm to 20mS/cm
200µS/cm to 6000µS/cm
2000µS/cm to 60mS/cm
20mS/cm to 600mS/cm
2 µS/cm to 300mS/cm
Cell Constant Linearity
±0.6% of reading in recommended range
+2 to -10% of reading outside high recommended range
±5% of reading outside low recommended range
±4% of reading in recommended range
Section 1: Description and Specifications4
Reference Manual
00809-0100-3157
Section 1: Description and Specifications
1.4 pH/ORP/ISE (Codes -22, -32, and -42)
For use with any standard pH or ORP sensor. Measurement choices are pH, ORP, Redox, ammonia, fluoride or custom ISE. The automatic buffer recognition feature uses stored buffer values and their temperature curves for the most common buffer standards available worldwide. The analyzer will recognize the value of the buffer being measured and perform a self stabilization check on the sensor before completing the calibration. Manual or automatic temperature compensation is menu selectable. Change in pH due to process temperature can be compensated using a programmable temperature coefficient. For more information concerning the use and operation of the pH or ORP sensors, refer to the product data sheets.
Model 1057 can also derive an inferred pH value called pHCalc (calculated pH). pHCalc can be derived and displayed when two contacting conductivity sensors are used as sensor 1 and sensor 2.
Performance Specifications - Analyzer (pH Input)
Measurement Range [pH]: 0 to 14 pH
Accuracy: ±0.01 pH
Diagnostics: Glass impedance, reference impedance
Temperature coefficient: ±0.002pH/ °C
Solution temperature correction: Pure water, dilute base and custom.
Buffer recognition: NIST, DIN 19266, JIS 8802, and BSI.
Input filter: Time constant 1 - 999 seconds, default 4 seconds.
Response time: 5 seconds to 100%
Sensor Compatibility: Model 1057 is also compatible with SMART pH sensors from Rosemount.
Choose from 14 SMART sensor models for a wide array of applications and process conditions.
March 2020
Performance Specifications - Analyzer (ORP Input)
Measurement Range [ORP]: -1500 to +1500 mV
Accuracy: ± 1 mV
Temperature coefficient: ±0.12mV / °C
Input filter: Time constant 1 - 999 seconds,
default 4 seconds.
Response time: 5 seconds to 100% of final reading
Recommended Sensors For ORP: All standard
ORP sensors.
Temperature Specifications:
Temperature range 0-150 °C
Temperature Accuracy, Pt-100, 0-50 °C ± 0.5 °C
Temperature Accuracy, Temp. > 50 °C ± 1 °C
Fourteen SMART sensor models include PERpH-X, TUpH and general purpose pH sensor families.
Section 1: Description and Specifications 5
Section 1: Description and Specifications
March 2020
Reference Manual
00809-0100-3157
Section 1: Description and Specifications6
Reference Manual
00809-0100-3157
Section 2: Installation
March 2020
Section 2: Installation
2.1 Unpacking and Inspection
nspect the shipping container. If it is damaged, contact the shipper immediately for instructions.
I Save the box. If there is no apparent damage, unpack the container. Be sure all items shown on the pack­ing list are present. If items are missing, notify Rosemount immediately.
2.2 Installation
2.2.1 General Information
1. Although the analyzer is suitable for outdoor use, do not install it in direct sunlight or in areas of extreme temperatures.
2. Install the analyzer in an area where vibration and electromagnetic and radio frequency interference are minimized or absent.
3. Keep the analyzer and sensor wiring at least one foot from high voltage conductors. Be sure
there is easy access to the analyzer.
4. The analyzer is suitable for panel mounting (Figure 2-1) or Wall and pipe mounting (Figure 2-2).
WARNING
RIS
K
O
F
E
Elec
tr
ic
al
an
other applic
L
ECTRI
ins
tallation
able national
CAL
S
mus
HO
t
be
C
K
in
ac
c
ordanc
e
w
ith
the
N
ational
Elec
tric
al
Code
(A
N
SI
/
N
F
PA
-
70)
and/
or
or
loc
al
c
odes
.
Section 2: Installation 7
Section 2: Installation
March 2020
Figure 2-1 Panel Mounting Dimensions
Reference Manual
00809-0100-3157
Front View
Bottom View
Side View
The front panel is hinged at the bottom. The panel swings down for easy access to the wiring locations. Panel mounting seal integrity (4/4X) for outdoor applications is the responsibility of the end user.
Section 2: Installation8
Reference Manual
00809-0100-3157
Figure 2-2 Pipe and Wall Mounting Dimensions
Section 2: Installation
March 2020
Wall / Surface Mount
Front View
Bottom View
Side View
Pipe Mount
Side View
The front panel is hinged at the bottom. The panel swings down for easy access to the wiring locations.
Section 2: Installation 9
Section 2: Installation
OR
OR
OR
U4
R32
C
47
2
1 J
2
6
5
1
J4
4
R
11
U9
U18
U15
U16
U13
U5
C49
C
46
R77
R
73
U3
R
3
R1
1
J3
2
U1
C
8
U2
R9
R4
R10
Z
6
C16
R
23
R
15
U10
C6
R
6
+
C
19
+
C
20
+
C
28
C
24
+
C
21
+
+
C
55
R75
C41
R58
R
57
C
43
C
1
3
U6
R
12
C52
R49
C
37
C
42
R67
C
38
R
70
R
71
C54
R
72
C
56
R
74
R68
C23
C35
R50
C25
C26
R
28
R
22
C10
R30
Z
3
Z
4
R
35
C15
Z
5
C12
C14
R46
U
11
R41
C
27
C36
R
51
R47
J
1
C
30
R38
R
63
C
51
C
29
R64
R44
U
12
U8
R20
R
31
R
36
Y1
C
32
C
3
3
C34
R
8
R14
R21
R19
R16
R54
C
44
R
55
R
56
C45
R2
R69
U
7
C7
Q
3
R29
Z
1
C
5
R17
Z2
C
3
C
4
R5
R
61
R65
R66
C18
R
27
C1
R
7
R13
U20
C
48
C50
R62
R25
C
2
U
19
C22
R76
R
53
R45
C17
R60
R33
U
22
U
14
U
26
D
1
U
2
1
C39
T
B1
-5SHLD
RTD RTN
1
+5
SENSE
REF
GND
RTD IN
pH
SMART pH/ORP
ASSY 24312- REV
SHLD
R40
U
1
2
1
Z4
R
6
4
3 7
R
6
6
R
C3
C2
Z3
C4
5
6
R
U2
1
RTD RTN
CONTACTING
CONDUCTIVITY
ASSY 24355- REV
SHLDSHLD
RTD IN
SEN SEN
4CT4CT
SEN
SHLD
C
3
5
R40
R
5
2
C
3
8
C17
R
5
4
R
4
9
8
5
R
C44
C20
Q3
R44
R41
C
1
3
2
4
R
R13
R
5
5
C
12
5
3
R
9
2
C
C24
R
3
7
R
6
0
C
34
R36
R
7
5
R70
C
2
5
74C
C40
R20
U
1
4
C42
C62
C5
R45
C41
C54
C
1
C
39
C45
R9
R6
R68
C49
R67
1
5
C
C
5
5
C57
R63
R
3
R
12
R
1
C63
R25
R18
R30
Z
1
C65
C64
R7
3
5
C
Z9
C23
C30
+
C52
U
1
7
R
48
U16
1
R
74
R19
C18
U19
U2
R69
R72
U1
C21
R
3
1
R
4
6
R14
R
4
R5
1
3
C
C28
+
R57
U26
R53
7
2
C
+
0
5
C
+
C14
Z8
R
2
R
6
2
R61
R8
U
1
0
R11
R
76
R15
C16
R28
C56
R59
C
22
0
1
R
C26
+
U22
U
2
3
C
6
U11
TB1
2
3
C
C
3
6
C59
C7
C48
Z2
U24
Y1
R38
U
2
5
U4
U15
R
71
R47
C43
U
2
8
9
2
U
C58
C46
J1
U
6
U
5
TB2
1
6
C
8
C9
C10
C11
C15
C66
C
6
7
L1
Z5
U9
C60
+
U27
U20
6
Z
Z7
C37
1
6
C
POWER SUPPLY
ALARM
WIRING (VAC)
(OPTIONAL)
ANALOG OUTPUT
SENSOR 1
ANY CSA APPROVED DEVICE
OR SIMPLE APPARATUS
UNCLASSIFIED AREA
1057
CLASS 1 DIVISION 2, GROUPS ABCD 0-50°C
CLASS II, III DIVISION 2 GROUPS EFG
6. SIMPLE APPARATUS IS DEFINED AS AN ELECTRICAL DEVICE THAT DOES NOT GENERATE MORE
1.3W. CONTACTING CONDUCTIVITY SENSORS AND pH, ORP, AMPEROMETRIC SENSORS WITHOUT
PREAMPS QUALIFY AS SIMPLE APPARATUS.
5
NON-INCENDIVE FIELD WIRING METHODS MAY BE USED FOR CONNECTING SENSORS TO THE 20/30 AND 22/32
OPTION BOARDS. SENSORS MUST BE CSA APPROVED AS NON-INCENDIVE FOR CLASS I, DIVISION 2,
Voc AND Isc LISTED IN TABLES 1A TO 1C
AND THE Ci AND Li OF THE SENSOR AND INTERCONNECTED WIRING MUST BE
Ca AND La LISTED IN TABLES
1A TO 1C
OR BE CLASSIFIED AS SIMPLE APPARATUS
.
4
INSTRUMENT ENCLOSURE. AFTER TERMINATION, WRAP N.I. FIELD WIRING WITHIN ENCLOSURE WITH MYLAR TAPE, TO
ENSURE ADEQUATE DOUBLE INSULATION REMAINS.
3
GROUND CONNECTION MAY BE MADE IN HAZARDOUS AREA.
2. SEAL REQUIRED AT EACH CONDUIT ENTRANCE, WHEN CONDUIT IS USED.
1. INSTALLATION MUST CONFORM TO THE CEC.
SENSOR 2 (OPTIONAL)
ANY CSA APPROVED DEVICE OR
SIMPLE APPARATUS
UNCLASSIFIED AREA
METAL CONDUIT
METAL CONDUIT
METAL CONDUIT
METAL CONDUIT
SENSOR CABLE
IS SHIELDED
5
METAL CONDUIT
3
4
4
SENSOR 1
ANY CSA APPROVED DEVICE OR
SIMPLE APPARATUS
WARNING
IF THE SENSOR TIP HAS EXPOSED ELECTRODES,
THEN IT MUST ONLY BE USED IN A NON-FLAMMABLE LIQUID PROCESS
TABLE 1A
ENTITY PARAMETERS FOR
OPTION 22/32 (pH / ORP / ISE SENSOR BOARD)
TABLE 1B
ENTITY PARAMETERS FOR OPTION 20/30
(CONTACTING CONDUCTIVITY SENSOR BOARD)
SENSOR 3 (OPTIONAL)
ANY CSA APPROVED DEVICE OR
SIMPLE APPARATUS
SENSOR 3 (OPTIONAL)
ANY CSA APPROVED DEVICE OR
SIMPLE APPARATUS
SENSOR 2 (OPTIONAL)
ANY CSA APPROVED DEVICE OR
SIMPLE APPARATUS
METAL CONDUIT
SENSOR CABLE
IS SHIELDED
5
5
SENSOR CABLE
IS SHIELDED
NON-INCENDIVE FIELD WIRING CONNECTIONS
FOR CLASS 1, DIVISION 2, GROUPS ABCD
OPTION 22/32 (pH/ORP SENSOR BOARD)
OPTION 20/30 (CONTACTING CONDUCTIVITY SENSOR BOARD)
OUTPUT
PARAMETERS
pH TB1
CONNECTOR
Voc, Vo
Isc, Io
115 mA
276.8 mW
Ca
La
6 mH
OUTPUT
PARAMETERS
CONDUCTIVITY
CONNECTORS
TB1, TB2
Voc, Vo
6.633 V
Isc, Io
30.45 mA
50.5 mW
Ca
La
85 mH
NOTES: UNLESS OTHERWISE SPECIFIED
SCALE: 1:1
WEIGHT:
SIZE
D
DWG NO
SHEET 1 OF 1
B
1400680
REV
THIS DOCUMENT IS CERTIFIED BY
CSA (REVISION A)
REVISIONS ARE NOT PERMITTED
WITHOUT CSA APPROVAL
March 2020
Figure 2-3 Non Incendive Field Wiring Installation (CSA)
Reference Manual
00809-0100-3157
Section 2: Installation10
Reference Manual
00809-0100-3157
Section 3: Wiring
3.1 General
The Model 1057 is easy to wire. It includes removable connectors and slide-out signal input boards.
3.1.1 Removable Connectors and Signal Input Boards
Model 1057 uses removable signal input boards and communication boards for ease of wiring and installation. Each of the signal input boards can be partially or completely removed from the enclosure for wiring. The Model 1057 has three slots for placement of up to three
signal input boards and one communication board.
Figure 3-1 Model 1057 Signal Input Boards
Section 3: Wiring
March 2020
Slot 1 – Left
Signal board 3
Note: If the UL option code has been ordered, a plastic insulator shield surrounds the entire power supply board (AC power supply only). The protective insulator shield does not appear in this photo.
3.1.2 Signal Input Boards
Slots 1, 2 and 3 are for signal input measurement boards. Wire the sensor leads to the measurement board following the lead locations marked on the board. After wiring the sensor leads to the signal board, carefully slide the wired board fully into the enclosure slot and take up the excess sensor cable through the cable gland. Tighten the cable gland nut to secure the cable and ensure a sealed enclosure. Note: that signal input board 3 is inserted into slot 1. Board 3 is inverted in the slot to allow board components to face to the right. Board 3 uses a long ribbon cable to connect to the main PCB. Boards 1 and 2 use a split ribbon cable to connect both signal boards to a common connector on the main board.
Slot 2 – Center
Signal board 2
Slot 3 – Right
Signal board 1
3.1.3 Alarm Relays
Four alarm relays are supplied with the switching power supply (84 to 265 Vac, 03 order code) and the 24 Vdc power supply (20-30 Vdc, 02 order code). All relays can be used for process measure ments or temperature. Any relay can be configured as a fault alarm instead of a process alarm. Each relay can be configured independently and each can be programmed as an interval timer, typically used to activate pumps or control valves. As process alarms, alarm logic (high or low acti vation or USP*) and deadband are user-programmable. Customer-defined failsafe operation is supported as a programmable menu function to allow all relays to be energized or not-energized as a default condition upon powering the analyzer. The USP alarm can be programmed to activate when the conductivity is within a user-selectable percentage of the limit. USP alarming is available only when a contacting conductivity measurement board is installed.
Section 3: Wiring 11
Section 3: Wiring
March 2020
3.2 Preparing Conduit Openings
There are six conduit openings in all configurations of Model 1057. (Note that four plugs are provid-
ed upon shipment.)
Note: Use watertight fittings and hubs that comply with your requirements. Connect the conduit
ub to the conduit before attaching the fitting to the analyzer.
h
3.3 Preparing Sensor Cable
The Model 1057 is intended for use with all Rosemount pH/ORP and contacting conductivity sensors. Refer to the sensor installation instructions for details on preparing sensor cables.
3.4 Power, Output and Sensor Connections
All field wiring must be rated for 75 °C or higher. Each instrument includes a printed label inside the enclosure stating this wiring requirement.
3.4.1 Power Wiring
Two Power Supplies are offered for Model 1057:
a. 24 Vdc (20 – 30V) Power Supply (-02 ordering code)
b. 84 – 265 Vac Switching Power Supply (-03 ordering code)
AC mains (115 or 230V) leads and 24 Vdc leads are wired to the Power Supply board which is mounted vertically on the left side of the main enclosure cavity. Each lead location is clearly marked on the Power Supply board. Wire the power leads to the Power Supply board using the lead markings on
the board.
Reference Manual
00809-0100-3157
Figure 3-2 24 Vdc Power Supply (-02 ordering code)
This power supply automatically detects DC power and accepts 20 Vdc to 30 Vdc inputs. Four programmable alarm relays are included.
Figure 3-3 Switching AC Power Supply (-03 ordering code)
This power supply automatically detects AC line conditions and switchesto the proper line volt­age and line frequency. Four programmable alarm relays are included.
12
Section 3: Wiring
Reference Manual
00809-0100-3157
3.4.2 Current Output Wiring
All instruments are shipped with four 4-20 mA current outputs. Wiring locations for the outputs are on the main board which is mounted on the hinged door of the instrument. Wire the output leads to the correct position on the Main board connectors using the lead markings (+/positive, -/negative)on the
oard. Male mating connectors are provided with each unit. Use a 3/32” wide standard blade
b screwdriver.
Note: Twisted pairs are required to minimize noise pickup in the 4-20 mA current outputs. For high EMI/RFI environments, shielded sensor wire is required and recommended in all other installations.
3.4.3 Alarm Relay Wiring
Four alarm relays are supplied with the switching power supply (84 to 265 Vac, -03 order code) and the 24 Vdc power supply (20-30 Vdc, -02 order code). Wire the relay leads on each of the independ­ent relays to the correct position on the power supply board using the printed lead markings (NO/Normally Open, NC/Normally Closed, or Com/Common) on the board.
Figure 3-4 24 Vdc Power Supply (-02 ordering code)
NO1 COM1
NC1 NO2
COM2
NC2
NO3
COM3
NC3
NO4
COM4
NC4
Section 3: Wiring
March 2020
RELAY 1
RELAY 2
RELAY 3
RELAY 4
WARNING
RIS
K
O
F E
L
ECTRI
CAL
S
HO
C
K
Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA-70) and/or an
othe
r
applicable n
ation
al or local codes.
3.4.4 Sensor Wiring to Signal Boards
Wire the correct sensor leads to the measurement board using the lead locations marked directly on the board. After wiring the sensor leads to the signal board, carefully slide the wired board fully into the enclosure slot and take up the excess sensor cable through the cable gland.
For best EMI/RFI protection use shielded output signal cable enclosed in an earth-grounded metal conduit. Connect the shield to earth ground. AC wiring should be 14 gauge or greater. Provide a switch or breaker to disconnect the analyzer from the main power supply. Install the switch or breaker near the analyzer and label it as the disconnecting device for the analyzer. Keep sensor and output signal wiring separate from power wiring. Do not run sensor and power wiring in the same conduit or close together in a cable tray.
Section 3: Wiring 13
Section 3: Wiring
March 2020
3.4.5 Signal Board Wiring
igure 3-5 Contacting Conductivity Signal Board and Sensor Cable Leads
F
Reference Manual
00809-0100-3157
Figure 3-6 pH/ORP/ISE Signal Board and Sensor Cable Leads
Section 3: Wiring14
Reference Manual
00809-0100-3157
Figure 3-7 Power Wiring for Model 1057 84-265 Vac Power Supply (-03 ordering code)
Section 3: Wiring
March 2020
Figure 3-8 Output Wiring for Model 1057 Main PCB
Section 3: Wiring 15
Section 3: Wiring
March 2020
igure 3-9 Power Wiring for Model 1057 24 Vdc Power Supply (-02 ordering code)
F
Reference Manual
00809-0100-3157
Section 3: Wiring16
Reference Manual
00809-0100-3157
Section 4: Display and Operation
4.1 User Interface
he Model 1057 has a large display which shows three live
T measurement readouts in large digits and up to six additional process variables or diagnostic parameters concurrently. The display is back-lit and the format can be customized to meet user requirements. The intuitive menu system allows access to Calibration, Hold (of current outputs), Programming, and Display functions by pressing the MENU button. In addition, a dedicated DIAGNOSTIC button is available to provide access to useful operational information on installed sensor(s) and any problematic conditions that might occur. The display flashes Fault and/or Warning when these conditions occur. Help screens are displayed for most fault and warning conditions to guide the user in troubleshooting.During calibration and programming, key presses cause different dis plays to appear. The displays are self-explanatory and guide
the user step-by-step through the procedure.
Section 4: Display and Operation
March 2020
4.2 Instrument Keypad
There are four function keys and four selection keys on the instrument keypad.
Function Keys
The MENU key is used to access menus for programming and calibrating the instrument. Four top­level menu items appear when pressing the MENU key:
Calibrate – Calibrate attached sensors and analog
outputs.
Hold– Suspend current outputs.
Program – Program outputs, measurement,
temperature, security and reset.
Display – Program display format, language, warnings,
and contrast.
Pressing MENU always causes the main menu screen to appear. Pressing MENUfollowed by EXITcauses the main display to appear.
Pressing the DIAG key displays active Faults and Warnings, and provides detailed instrument information and sensor diagnostics including: faults, warnings, sensor 1, 2 and 3 information, current outputs live values, model configuration string e.g. 1057PPC03AN, Instrument Software version, and AC frequency. Pressing ENTERon Sensor 1 or Sensor 2 provides useful diagnostics and information (as applicable): measurement, sensor type, raw signal value, cell constant, zero offset and temperature. Offset, selected measurement range, cable resistance, temperature sensor resistance, signal board software version.
The ENTERkey - Pressing ENTERstores numbers and settings and moves the display to the next screen.
The EXITkey - Pressing EXITreturns to the previous screen without storing changes.
Section 4: Display and Operation 17
Section 4: Display and Operation
March 2020
Selection Keys
Surrounding the ENTERkey, four selection keys – up, down, right and left, move the cursor to all areas of the screen while using the menus.
Selection keys are used to:
1. Select items on the menu screens
2. Scroll up and down the menu lists.
3. Enter or edit numeric values.
4. Move the cursor to the right or left
5. Select measurement units during operations
4.3 Main Display
The Model 1057 displays one, two or three primary measurement values, up to six secondary measurement values, a fault and warning banner, alarm relay flags.
4.3.1 Process Measurements
Three process variables are displayed if three signal boards are installed. One process variable and process temperature is displayed if one signal board is installed with one sensor. The upper display area shows the Sensor 1 process reading. The center display area shows the Sensor 2 process read ing. For dual conductivity, the display areas can be assigned to different process variables as follows:
4.3.2 Secondary Values
Up to six secondary values are shown in six display quadrants at the bottom of the screen. All four secondary value positions can be programmed by the user to any display parameter available. Possible secondary values include:
Process variables for display - examples
Measure 1 Measure 2 Measure 3  Reject  Pass Ratio Blank pH Calc
Reference Manual
Displayable Secondary Values
Slope 1, 2, 3 Output 1 mA
Ref Off 1, 2, 3 Output 2 mA
Gl Imp 1, 2, 3 Output 3 mA
Ref Imp 1, 2, 3
Raw 1, 2, 3
mV Input 1, 2, 3 Output 2 %
Temp 1, 2, 3 Output 3 %
Man Temp 1, 2, 3 Output 4 %
Measure 1, 2, 3 Blank
Output 4 mA
Output 1 %
00809-0100-3157
4.3.3 Fault and Warning Banner
If the analyzer detects a problem with itself or the sensor the word Fault or Warning will appear at the bottom of the display. A fault requires immediate attention. A warning indicates a problematic condition or an impending failure. For troubleshooting assitance, press Diag.
4.3.4 Formatting the Main Display
The main display screen can be programmed to show primary process variables, secondary process variables and diagnostics.
Section 4: Display and Operation18
Reference Manual
00809-0100-3157
4.3.4 Formatting the Main Display
The main display screen can be programmed to show primary process variables, secondary process
ariables and diagnostics.
v
. Press MENU.
1
. Scroll down to Display. Press ENTER.
2
3. Main Format is highlighted. Press ENTER.
4. The Sensor 1 process value is highlighted in reverse video. Press the selection keys to navigate down to the screen sections that you wish to program. Press ENTER.
5. Choose the desired display parameter or diagnostic for each of the four display sections in the lower screen.
6. Continue to navigate and program all desired screen sections. Press MENU and EXIT. The screen returns to the main display.
For single sensor configurations, the default display shows the live process measurement in the upper display area and temperature in the center display area. The user can select to disable the display of temperature in the center display area using the Main Format function. See guide you through programming the main display to select process parameters and diagnostics of your choice.
For dual sensor configurations, the default display shows Sensor 1 live process measurement in the display area one and Sensor 2 live process measurement temperature in the display area two. See Figure 4-1 to guide you through programming the main display to select process parameters and diagnostics of your choice.
Section 4: Display and Operation
March 2020
Figure 4-1 to
4.4 Menu System
Model 1057 uses a scroll and select menu system. Pressing the MENU key at any time opens the top-level menu including Calibrate, Hold, Program and Display functions.
To find a menu item, scroll with the up and down keys until the item is highlighted. Continue to scroll and select menu items until the desired function is chosen. To select the item, press ENTER. To return to a previous menu level or to enable the main live display, press the EXIT key repeatedly. To return immediately to the main display from any menu level, simply press MENU then EXIT.
The selection keys have the following functions:
The Down key (below ENTER) decrements numerical values, moves the decimal place one place to
The Leftkey (left of ENTER) moves the cursor to the left.
The Rightkey (right of ENTER) moves the cursor to the right.
To access desired menu functions, use the Quick Reference Figure B. During all menu displays (except main display format and Quick Start), the live process measurements and secondary measurement values are displayed in the top two lines of the upper display area. This conveniently allows display of the live values during important calibration and programming operations.
The Upkey (above ENTER) increments numerical values, moves the decimal place one place
to the right, or selects units of measurement.
the left, or selects units of measurement
Menu screens will time out after two minutes and return to the main live display.
Section 4: Display and Operation 19
Section 4: Display and Operation
March 2020
igure 4-1 Configuring the Main Display
F
Reference Manual
00809-0100-3157
Section 4: Display and Operation20
Reference Manual
00809-0100-3157
Section 5: Programming the Analyzer - Basics
Section 5: Programming the Analyzer - Basics
5.1 General
This section describes the following programming functions:
Changing the measurement type, measurement units and temperature units.
Choose temperature units and manual or automatic temperature compensation mode
Configure and assign values to the current outputs
Set a security code for two levels of security access
Accessing menu functions using a security code
Enabling and disabling Hold mode for current outputs
Choosing the frequency of the AC power (needed for optimum noise rejection)
Resetting all factory defaults, calibration data only, or current output settings only
March 2020
5.2 Changing the Startup Settings
5.2.1 Purpose
To change the measurement type, measurement units, or temperature units that were initially entered in Quick Start, choose the Reset Analyzer function (Section 5.8) or access the Program menus for sensor 1, 2 or 3 (Section 6). The following choices for specific measurement type, measurement units are available for each sensor measurement board.
Table 5-1 Measurments and measurement units
Signal board Available measurements Measurements units:
pH/ORP (-22, -32, -42)
Contacting conductivity
(-20, -30, -40)
Temperature (all) Temperature °C, °F
pH, ORP, Redox, Ammonia, Fluoride, Custom ISE
Conductivity, Resistivity, TDS, Salinity, NaOH (0-12%), HCl (0-15%), Low H2SO4, High H2SO4, NaCl (0-20%), Custom Curve
5.2.2 Procedure
Follow the Reset Analyzer procedure (Section 5.8) to reconfigure the analyzer to display new measurements or measurement units. To change the specific measurement or measurement units
for each signal board type, refer to the Program menu for the appropriate measurement (Section 6).
pH, mV (ORP) %, ppm, mg/L, ppb, µg/L, (ISE)
µS/cm, mS/cm, S/cm
% (concentration)
Section 5: Programming the Analyzer - Basics 21
Section 5: Programming the Analyzer - Basics
March 2020
Reference Manual
00809-0100-3157
5.3 Choosing Temperature Units and Automatic/Manual Temperature Compensation
5.3.1 Purpose
Most liquid analytical measurements (except ORP) require temperature compensation. The Model 1057 performs temperature compensation automatically by applying internal tempera-
ture correction algorithms. Temperature correction can also be turned off. If temperature correction is
off, the Model 1057 uses the temperature entered by the user in all temperature correction calculations.
S1: 1.234µS/cm 123.4°C S2: 12.34pH 123.4°C
3:123.4µS/cm 123.4°C
S
Temperature Units: °C S1 Temp Comp: Auto S2 Temp Comp: Auto
5.3.2 Procedure
Follow the menu screens in Figure 5-1 to select automatic or manual temp compensation, set the manual reference temperature, and to program temperature units as °C or °F.
Figure 5-1 Choosing Temperature Units and Manual Auto Temp Compensation
S3 Temp Comp: Auto
5.4 Configuring and Ranging The Current Outputs
5.4.1 Purpose
The Model 1057 accepts inputs from three sensors and has four analog current outputs. Ranging the outputs means assigning values to the low (0 or 4 mA) and high (20 mA) outputs. This section provides a guide for configuring and ranging the outputs. Always configure the outputs first.
5.4.2 Definitions
1. Current ouputs - The analyzer provides a continuous output current (4-20 mA or 0-20 mA) directly proportional to the process variable or temperature. The low and high current outputs can be set to any value.
2. Assigning outputs - Assign a measurement to outputs 1, 2, 3, or 4.
3. Dampen - Output dampening smooths out noisy readings. It also increases the response time of the output. Output dampening does not affect the response time of the display.
4. Mode - The current output can be made directly proportional to the displayed value (linear mode)
or directly proportional to the common logarithm of the displayed value (log mode).
Section 5: Programming the Analyzer - Basics22
Reference Manual
00809-0100-3157
Section 5: Programming the Analyzer - Basics
5.4.3 Procedure - Configure Outputs
Under the Program/Outputs menu, the adjacent screen appears to allow configuration of the outputs. Follow the menu screens in Figure 5-2 to configure the outputs.
5.4.4 Procedure - Assigning Measurements the Low and
High Current Outputs
The adjacent screen appears when entering the Assign function under Program/Output/Configure. These screens allow you to assign a measurement, process value, or temperature input to each output. Follow the menu screens in Figure 5-2 to assign measurements to the outputs.
5.4.5 Procedure - Ranging the Current Outputs
The adjacent screen appears under Program/Output/Range. Enter a value for 4mA and 20mA (or 0 mA and 20 mA) for each output. Follow the menu screens in Figure 5-2 to assign values to the outputs.
March 2020
S
1
:
1
.
2
3
4
µ
S
/
c
m
1
2
S
2
:
1
2
.
3
4
p
H
4
µ
tpu
:
ode:
al
S
/
c
m
t
M
Con
S
1
4
-
L
i
i
n
g:
F
i
u
e:
2
1
S
3
:
1
2
3
.
Ou
A
s
s
i
gn
R
an
ge:
S
c
al
e:
D
ampen
F
au
l
t
M
F
au
l
t
V
1: 1.234µS/cm 123.4°C
S
2: 12.34pH 123.4°C
S S3:123.4µS/cm 123.4°C
3
1
2
3
.
0
ed
.0
eas
mA ear 0
fi
s
0
ec
1
gu
mA
4
2
3
r
e M 2
n
x
OutputM Assign S1 Measurement S1 Temperature S2 Measurement S2 Temperature S3 Measurement S3 Temperature
S1: 1.234µS/cm 123.4°C S2: 12.34pH 123.4°C
S3:123.4µS/cm 123.4°C
Output Range OM SN 4mA: 0.000µS/cm OM SN 20mA: 20.00µS/cm OM SN 4mA: 00.00pH OM SN 20mA: 14.00pH
.
4
°
C
°
C
.
4
°
C
Figure 5-2 Configuring and Ranging the Current Outputs
Section 5: Programming the Analyzer - Basics 23
Section 5: Programming the Analyzer - Basics
March 2020
5.5 Setting a Security Code
5.5.1 Purpose
The security codes prevent accidental or unwanted changes to program settings, displays, and calibration. Model 1057 has two levels of security code to control access and use of the instrument to different types of users. The two levels of security are:
All - This is the supervisory security level. It allows access to all menu functions, including
Programming, Calibration, Hold and Display.
Calibration/Hold - This is the operator or technician level menu. It allows access to only calbration
and Hold of the current outputs.
5.5.2 Procedure
1. Press MENU. The main menu screen appears. Choose Program.
2. Scroll down to Security. Select Security.
3. The security entry screen appears. Enter a three digit security code for each of the desired
security levels. The security code takes effect two minutes after the last key stroke. Record the security code(s) for future access and communication to operators or technicians as needed.
4. The display returns to the security menu screen. Press EXIT to return to the previous screen. To
return to the main display, press MENU followed by EXIT. Figure 5-3 displays the
security code screens.
Reference Manual
00809-0100-3157
Figure 5-3 Setting a Security Code
S1: 1.234µS/cm 1.234°C
S2: 12.34pH 1.234°C
S3: 12.34µS/cm 1.234°C
Program
Outputs
Program
MAIN MENU
Measurement Temperature
Security
Diagnostic Setup Rejection Freq: 60Hz Reset Analyzer
S1: 1.234µS/cm 1.234°C
S2: 12.34pH 1.234°C
S3: 12.34µS/cm 1.234°C
Calibration/Hold: 000 All: 000
5.6 Security Access
5.6.1 How the Security Code Works
When entering the correct access code for the Calibration/Hold security level, the Calibration and Hold menus are accessible. This allows operators or technicians to perform routine maintenance.
This security level does not allow access to the Program or Display menus.
When entering the correct access code for all security level, the user has access to all menu functions, including Programming, Calibration, Hold and Display.
Security
Section 5: Programming the Analyzer - Basics24
Reference Manual
00809-0100-3157
Section 5: Programming the Analyzer - Basics
March 2020
5.6.2 Procedure
1: 1.234µS/cm 1.234°C
1. If a security code has been programmed, selecting the Calibrate, Hold,
Program or Display top menu items causes the security access screen
to appear.
2. Enter the three-digit security code for the appropriate security level.
S
2: 12.34pH 1.234°C
S S3: 12.34µS/cm 1.234°C
Security Code
0
00
3. If the entry is correct, the appropriate menu screen appears. If the entry is incorrect, the Invalid
Code screen appears. The Security Code screen reappears after 2 seconds.
5.7 Using Hold
5.7.1 Purpose
The analyzer output is always proportional to measured value. To prevent improper operation of systems or pumps that are controlled directly by the current output, place the analyzer in hold before removing the sensor for calibration and maintenance. Be sure to remove the analyzer from
hold once calibration is complete. During hold, both outputs remain at the last value. Once in hold, all
current outputs remain on Hold indefinitely.
5.7.2 Using the Hold Function
To hold the outputs,
1. Press MENU. The main menu screen appears. Choose Hold.
2. The Hold Outputs and Alarms screen appears. Choose Yes to place the analyzer in hold.
Choose No to take the analyzer out of hold.
Note: There are no alarm relays with this configuration. Current outputs are included with all configurations.
3. The Hold screen appears and remains on indefinitely until Hold is disabled.
Figure 5-4 Using Hold
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
Hold
MAIN MENU
S1 Hold: No S2 Hold: No S3 Hold: No
Hold
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C
S3: 12.34µS/cm 1.234°C
S1 Hold outputs
and alarms? No Yes
5.8 Resetting the Factory Default Settings
5.8.1 Purpose
This section describes how to restore factory calibration and default values. The process also clears all fault messages and returns the display to the first Quick Start screen. The Model 1057 offers three options for resetting factory defaults.
Reset all settings to factory defaults
Reset sensor calibration data only
Reset output calibration only
Section 5: Programming the Analyzer - Basics 25
Section 5: Programming the Analyzer - Basics
March 2020
5.8.2 Procedure
To reset to factory defaults, reset calibration data only or reset analog outputs only, follow the Reset
Analyzer flow diagram (Figure 5-5).
igure 5-5 Resetting Factory Default Settings
F
Reference Manual
00809-0100-3157
5.9 Programming Alarm Relays
5.9.1 Purpose
The Model 1057 24 Vdc (02 order code) and the AC switching power supply (03 order code) provide four alarm relays for process measurement or temperature. Each alarm can be configured as a fault alarm instead of a process alarm. Also, each relay can be programmed independently and each can
be programmed as an interval timer. This section describes how to configure alarm relays, simulate relay
activation, and synchronize timers for the four alarm relays. This section provides details to program the following alarm features:
Table 5-2 Programming alarm relays
Section Alarm Relay Feature
5.9.2 Enter setpoint
5.9.3 Assign measurement
5.9.4 Set relay logic
5.9.5
5.9.6 USP safety
5.9.7
5.9.8 Interval time
5.9.9 On-Time
5.9.10 Recover time
5.9.11 Hold while active
5.9.12 Simulate Manually simulate alarms to confirm relay operation
5.9.13 Synchronize timers
Deadband
Normal state
Default
100.0uS/cm
S1 Measure
High
0.00uS/cm
i
0%
Open
24.0 hr
10 min
60 sec
S1
Yes
Description
Enter alarm trigger value
Select alarm assignment
Program relay to activate at High or Low reading
Program the change in process value after the relay deactivates
Program percentage of the limit to activate the alarm
Program relay default condition as open or closed for failsafe operation
Time in hours between relay activations
Enter the time in seconds that the relay is activated.
Enter time after the relay deactivation for process recovery
Holds current outputs during relay activation
Control the timing of two or more relay timers set as Interval timers
Section 5: Programming the Analyzer - Basics26
Reference Manual
00809-0100-3157
Section 5: Programming the Analyzer - Basics
March 2020
Under the Program/Alarms menu, the adjacent screen appears to allow configuration of the alarm relays. Follow the menu screens in
Figure 5-2 to
configure the outputs.
The adjacent screen appears to allow selection of a specific alarm relay. Select the desired alarm and press ENTER.
The adjacent screen appears next to allow complete programming of each alarm. Factory defaults are displayed as they would appear for an installed contacting conductivity board. USP Safety only appears if alarm logic is set to “USP”. Interval timer, On Time, Recover Time, and Hold While Active only appear if the alarm is configured as an Interval timer.
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
Alarms Configure/Setpoint Simulate Synchronize Timers: Yes
S1: 1.234µS/cm 1.234°C
2: 12.34pH 1.234°C
S
3: 12.34µS/cm 1.234°C
S
Configure/Setpoint Alarm 1 Alarm 2 Alarm 3 Alarm 4
S1: 1.234µS/cm 1.234°C
2: 12.34pH 1.234°C
S S3: 12.34µS/cm 1.234°C
AlarmM Settings Setpoint: 100.0uS/cm Assign: S1 Measure Logic: High Deadband: 0.00uS/cm USP Safety: 0% Interval time: 24.0 hr On Time: 120 sec Recover time: 60 sec Hold while active: Sens1
i
5.9.2 Procedure - Enter Setpoints
Under the Program/Alarms menu, the adjacent screen appears to allow configuration of the alarm relays. Enter the desired value for the process measurement or temperature at which to activate an alarm event.
5.9.3 Procedure - Assign Measurements
Under the Alarms Settings menu, the adjacent screen appears to allow assignment of the alarm relays. select an alarm assignment. Additional assignment choices are shown in Figure 5-2 depending on which measurement board(s) is installed.
5.9.4 Procedure - Set Relay Logic
Under the Alarms Settings menu, the adjacent screen appears to set the alarm logic. Select the desired relay logic to activate alarms at a High reading or a Low reading. USP safety only appears if a contacting conductivity board is installed.
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
Alarm1 S2 Setpoint
+100.0uS/cm
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
AlarmM Assign: S1 Measurement S1 Temperature S2 Measurement S2 Temperature S3 Measurement S3 Temperature Interval Timer Fault Off
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
AlarmM Logic: High Low USP
Section 5: Programming the Analyzer - Basics 27
Section 5: Programming the Analyzer - Basics
March 2020
5.9.5 Procedure - Deadband
Under the Alarms Settings menu, the adjacent screen appears to program the deadband as a measurement value. Enter the change in the process value needed after the relay deactivates to return to normal (and thereby
reventing repeated alarm activation).
p
5.9.6 Procedure - USP Safety
Under the Alarms Settings menu, the adjacent screen appears to program the USP alarm setting. Enter the percentage below the limit at which to activate the alarm.
5.9.7 Procedure - Normal State
The user can define failsafe condition in software by programming the alarm default state to normally open or normally closed upon power up. To display this alarm configuration item, enter the Expert menus by hold ing down the EXIT key for six seconds while in the main display mode. Select Yes upon seeing the screen prompt: “Enable Expert Menu?” Under the Alarms Settings menu, the adjacent screen appears to set the normal state of the alarms. Select the alarm condition that is desired each time the analyzer is powering up.
5.9.8 Procedure - Interval Time
Under the Alarms Settings menu, the adjacent screen appears to set the interval time. Enter the fixed time in hours between relay activations.
Reference Manual
00809-0100-3157
1: 1.234µS/cm 1.234°C
S S2: 12.34pH 1.234°C
3: 12.34µS/cm 1.234°C
S
Alarm1 Deadband
+000.5uS/cm
1: 1.234µS/cm 1.234°C
S S2: 12.34pH 1.234°C
3: 12.34µS/cm 1.234°C
S
Alarm1 USP Safety
i
+0%
1: 1.234µS/cm 1.234°C
S S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
Alarm2 Normal State
Open
Closed
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
Alarm1 Interval Time
024.0 hrs
5.9.9 Procedure - On Time
Under the Alarms Settings menu, the adjacent screen appears to set the relay on time. Enter the time in seconds that the relay is activated.
5.9.10 Procedure - Recovery Time
Under the Alarms Settings menu, the adjacent screen appears to set the relay recovery time. Enter time after the relay deactivation for process recovery.
5.9.11 Procedure - Hold While Active
Under the Alarms Settings menu, the adjacent screen appears to program the feature that holds the current outputs while alarms are active. Select to hold the current outputs for Sensor 1, Sensor 2 or both sensors while the relay is activated.
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
Alarm1 On-Time
00.00sec
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
Alarm1 Recovery
060sec
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
Alarm1 Hold while
active Sensor 1 Sensor 2 Sensor 3 All None
Section 5: Programming the Analyzer - Basics28
Reference Manual
00809-0100-3157
5.9.12 Procedure - Simulate
Alarm relays can be manually set for the purposes of checking devices such as valves or pumps. Under the Alarms Settings menu, the adjacent screen appears to allow manual forced activation of the alarm relays.
elect the desired alarm condition to simulate.
S
5.9.13 Procedure - Synchronize
Under the Alarms Settings menu, the adjacent screen appears to allow synchronization of alarms that are set to interval timers. Select Yes or No to synchronize two or more timers
Section 5: Programming the Analyzer - Basics
March 2020
1: 1.234µS/cm 1.234°C
S
2: 12.34pH 1.234°C
S
3: 12.34µS/cm 1.234°C
S
Simulate Alarm M
on’t simulate
D De-energize Energize
1: 1.234µS/cm 1.234°C
S S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
Synchronize Timers
Yes
No
Section 5: Programming the Analyzer - Basics 29
Section 5: Programming the Analyzer - Basics
March 2020
Reference Manual
00809-0100-3157
Section 5: Programming the Analyzer - Basics30
Reference Manual
00809-0100-3157
Section 6: Programming Measurements
March 2020
Section 6: Programming Measurements
6.1 Programming Measurements - Introduction
The Model 1057 automatically recognizes each installed measurement board upon first power-up
and each time the analyzer is powered. Completion of Quick Start screens upon first power up enable
measurements, but additional steps may be required to program the analyzer for the desired measurement application. This section covers the following programming and configuration functions;
1. Selecting measurement type or sensor type (all sections)
2. Identifying the preamp location (pH - see Section 6.2)
3. Enabling manual temperature correction and entering a reference temperature (all sections)
4. Enabling sample temperature correction and entering temperature correction slope (selected sections)
5. Defining measurement display resolution (pH)
6. Defining measurement display units (all sections)
7. Adjusting the input filter to control display and output reading variability or noise (all sections)
8. Selecting a measurement range (conductivity – see
9. Entering a cell constant for a contacting sensor (see Section 6.4)
10. Entering a temperature element/RTD offset or temperature slope (conductivity - see Section 6.4)
11. Creating an application-specific concentration curve (conductivity - see Section 6.4)
To fully configure the analyzer for each installed measurement board, you may use the following:
1. Reset analyzer function to reset factory defaults and configure the measurement board to the desired measurement. Follow the Reset Analyzer menu (Figure 5-5) to reconfigure the analyzer to display new measurements or measurement units.
2. Program menus to adjust any of the programmable configuration items. Use the following configuration and programming guidelines for the applicable measurement.
Section 6.4)
6.2 pH Measurement Programming
6.2.1 Description
This section describes how to configure the Model 1057 analyzer for pH measurements. The follow ing programming and configuration functions are covered.
Table 6-1 pH measurement programming
Measure Section Menu Function
pH
Section 6: Programming Measurements 31
6.2.2 Measurement type
6.2.3 Preamp location
6.2.4
6.2.5 Temp coefficient
6.2.6 Resolution
6.2.7 Filter
6.2.8
Solution temperature correction
Reference Z
Default Settings
pH
Analyzer
Off
(custom)
0.01pH
4 seconds
Low
Description
Select pH, ORP, redox, ammonia, fluoride, custom ISE
Identify preamp location
Select Off, ultra-pure, high pH, custom
Enter the temp coefficient
Select 0.01pH or 0.1pH for pH display resolution
Override the default input filter, enter 0-999 seconds
Select low or high reference impedance
Section 6: Programming Measurements
March 2020
A detailed flow diagram (Figure 6-1) for pH configuring is provided at the end of Section 6 to guide
you through all basic configuration functions.
To configure the pH measurement board:
1. Press MENU
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
4. Select Sensor 1 or Sensor 2 corresponding to pH. Press ENTER.
The adjacent screen format appears (factory defaults are shown). To program any function, scroll to the desired item and press ENTER.
The following sub-sections provide you with the initial display screen that appears for each configuration function. Use the flow diagram for configuring pH/ORP measurements ( the end of this section and the Model 1057 live screen prompts for each function to complete configuration and programming.
6.2.2 Measurement
The display screen for selecting the Measurement is shown. The default value is displayed in bold type. Refer to configuring pH/ORP measurements flow diagram (Figure 6-1) to complete this function.
Reference Manual
00809-0100-3157
1: 1.234µS/cm 1.234°C
S
2: 12.34pH 1.234°C
S
3: 12.34µS/cm 1.234°C
S
SN Configure
Measure: pH
reamp: Analyzer
P Sol’n Temp Corr: Off T Coeff: -0.029pH/°C
esolution: 0.01pH
R Filter: 4 sec Reference Z: Low
Figure 6-1) at
1: 1.234µS/cm 1.234°C
S
2: 12.34pH 1.234°C
S
3: 12.34µS/cm 1.234°C
S
SN Measurement
pH
ORP Redox Ammonia
6.2.3 Preamp
The display screen for identifying the Preamp location is shown. The default value is displayed in bold type. Refer to configuring pH/ORP measurements flow diagram (Figure 6-1) to complete this function.
6.2.4 Solution Temperature Correction
The display screen for selecting the Solution Temperature Correction algorithm is shown. The default value is displayed in bold type. Refer to configuring pH/ORP measurements flow diagram (Figure 6-1) to complete this function.
6.2.5 Temperature Coefficient
The display screen for entering the custom Solution Temperature Coefficient is shown. The default value is displayed in bold type. Refer to configuring pH/ORP measurements flow diagram (Figure 6-1) to complete this function.
Fluoride Custom ISE
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
SN Preamp
Analyzer
Sensor/JBox
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
SN Sol’n Temp Corr.
Off
Ultra Pure Water High pH Custom
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
SN Sol’n Temp Coeff.
- 0.032pH/ºC
Section 6: Programming Measurements32
Reference Manual
00809-0100-3157
Section 6: Programming Measurements
March 2020
6.2.6 Resolution
The display screen for selecting 0.01pH or 0.1pH for pH display Resolution is shown. The default value is displayed in bold type. Refer to configuring pH/ORP measurements flow diagram (Figure 6-1) to complete this function.
6.2.7 Filter
The display screen for entering the Input filter value in seconds is shown. The default value is displayed in bold type. Refer to configuring pH/ORP measurements flow diagram (Figure 6-1) to complete this function.
6.2.8 Reference Impedence
The display screen for selecting Low or High Reference impedance (Z) is shown. The default value is displayed in bold type. Refer to configuring pH/ORP measurements flow diagram (Figure 6-1) to complete this function.
6.3 ORP Measurement Programming
6.3.1 Description
The section describes how to configure the Model 1057 analyzer for ORP measurements. The following programming and configuration functions are covered:
S1: 1.234µS/cm 1.234°C
2: 12.34pH 1.234°C
S
3: 12.34µS/cm 1.234°C
S
SN Resolution
0.01pH
0.1pH
S1: 1.234µS/cm 1.234°C
2: 12.34pH 1.234°C
S
3: 12.34µS/cm 1.234°C
S
SN Input filter
04 sec
S1: 1.234µS/cm 1.234°C
2: 12.34pH 1.234°C
S S3: 12.34µS/cm 1.234°C
SN Reference Z
Low
High
Table 6-2 ORP measurement programming
Measure Section Menu Function
ORP
6.3.2 Measurement type
6.3.3 Preamp location
6.3.4 Filter
6.3.5 Reference (Z)
A detailed flow diagram (Figure 6-1) for configuring ORP measurements is provided at the end of Section 6 to guide you through all basic configuration functions.
To configure the ORP measurement board:
1. Press MENU.
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
4. Select Sensor 1 or Sensor 2 corresponding to ORP. Press ENTER.
The adjacent screen appears (factory defaults are shown). To program any displayed function, scroll to the desired item and press ENTER.
The following sub-sections provide you with the initial display screen that appears for each configuration function. Use the flow diagram (Figure 6-1) for configuring ORP measurements at the end of this section and the Model 1057 live screen prompts for each function to complete configuration and programming.
Default Settings
pH
Analyzer
4 seconds
Low
Description
Select pH, ORP, redox, ammonia, fluoride, custom ISE
Identify preamp location
Override the default input filter, enter 0-999 seconds
Select low or high reference impedance
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234ºC
SN Configure Measure: pH Preamp: Analyzer Flter: 4 sec Reference Z: Low
Section 6: Programming Measurements 33
Section 6: Programming Measurements
March 2020
Reference Manual
00809-0100-3157
6.3.2 Measurement
The display screen for selecting the Measurement is shown. The default value is displayed in bold type. Refer to configuring pH/ORP measurements
flow diagram (Figure 6-1) to complete this function.
6.3.3 Preamp
The display screen for identifying the Preamp location is shown. The default value is displayed in bold type. Refer to configuring pH/ORP measurements
flow diagram (Figure 6-1) to complete this function.
6.3.4 Filter
The display screen for entering the Input filter value in seconds is shown. The default value is displayed in bold type. Refer to configuring pH/ORP measurements flow diagram (Figure 6-1) to complete this function.
6.3.5 Reference Impedence
The display screen for Selecting Low or high Reference impedance (Z) is shown. The default value is displayed in bold type. Refer to configuring pH/ORP measurements flow diagram (Figure 6-1) to complete this function.
S1: 1.234µS/cm 1.234°C
S2: 12.34pH 1.234°C
3: 12.34µS/cm 1.234°C
S
SN Measurement
pH
ORP
Redox Ammonia
Fluoride
Custom ISE
1: 1.234µS/cm 1.234°C
S S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
SN Preamp
Analyzer
Sensor/JBox
1: 1.234µS/cm 1.234°C
S
2: 12.34pH 1.234°C
S
3: 12.34µS/cm 1.234°C
S
SN Input filter
04 sec
1: 1.234µS/cm 1.234°C
S S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
SN Reference Z
Low
High
6.4 Contacting Conductivity Measurement Programming
6.4.1 Description
The section describes how to configure the Model 1057 analyzer for conductivity measurements using contacting conductivity sensors. The following programming and configuration functions are covered.
Table 6-3 Contacting conductivity measurement programming
Measure Section Menu Function:
Contacting Conductivity
6.4.2 Type
6.4.3 Measure
6.4.4 Range
6.4.5 Cell K
6.4.6 RTD Offset
6.4.7 RTD Slope
6.4.8 Temp Comp
6.4.9 Slope 2.00%/°C Enter the linear temperature coefficient
6.4.10 Ref Temp 25.0 °C Enter the Reference temp
6.4.11 Filter 2 seconds Override the default input filter, enter 0-999 seconds
6.4.12 Custom Setup Enter 2-5 data points in ppm and µS/cm for custom curves
6.4.13 Cal Factor 0.95000/cm Enter the Cal Factor for 4-Electrode sensors from the sensor tag
Default Settings
2- Electrode Conductivity
Auto
1.00000/cm
0.00°C
0
Slope
Description
Select 2-Electrode or 4-Electrode type sensors
Select Conductivity, Resistivity, TDS. Salinity or % conc
Select measurement Auto-range or specific range
Enter the cell Constant for the sensor
Enter the RTD Offset
Enter the RTD Slope
Select Temp Comp: Slope, Neutral Salt, Cation or Raw
Section 6: Programming Measurements34
Reference Manual
00809-0100-3157
Section 6: Programming Measurements
March 2020
A detailed flow diagram (Figure 6-2) for contacting conductivity programming is provided at the end of this section to guide you through all basic programming and configuration functions.
To configure the contacting conductivity measurement board:
1. Press MENU
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
4. Select Sensor 1 or Sensor 2 corresponding to contacting conductivity. Press ENTER.
The adjacent screen appears (factory defaults are shown). To program any displayed function, scroll to the desired item and press ENTER.
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C
3: 12.34µS/cm 1.234°C
S
SN Configure
ype: 2-Electrode
T Measure: Cond Range: Auto Cell K: 1.00000/cm
TD Offset: 0.00°C
R RTD Slope: 0 Temp Comp: Slope Slope: 2.00%/°C Ref Temp: 25.0°C Filter: 2 sec Custom Setup
The following sub-sections provide you with the initial display screen that appears for each configuration function. Use the flow diagram for contacting conductivity programming at the end of this section and the Model 1057 live screen prompts for each function to complete configuration and programming.
6.4.2 Sensor Type
The display screen for selecting 2-Electrode or 4-Electrode Type sensors is shown. The default value is displayed in bold type. Refer to configuring contacting measurements flow diagram (Figure 6-2) to complete this function.
6.4.3 Measure
The display screen for selecting the Measurement is shown. The default value is displayed in bold type. Refer to configuring contacting measurements flow diagram (Figure 6-2) to complete this function.
6.4.4 Range
The display screen for selecting Range (Auto or a specific range) is shown. The default value is displayed in bold type.
Note: Ranges are shown as conductance, not conductivity.
Refer to configuring contacting measurements flow diagram (Figure 6-2) to complete this function.
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
SN Type
2-Electrode
4-Electrode
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
SN Measurement
Conductivity
Resistivity TDS Salinity
NaOH (0-12%) HCl (0-15%) Low H2SO4 High H2SO4 NaCl (0-20%) Custom Curve
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
SN Range
Auto
50 µS 500 µS 2000 µS 20 mS 200 mS 600 mS
Section 6: Programming Measurements 35
Section 6: Programming Measurements
March 2020
Reference Manual
00809-0100-3157
6.4.5 Cell Constant
The display screen for entering a cell Constant for the sensor is shown. The default value is displayed in bold type. Refer to configuring contacting measurements flow diagram (Figure 6-2) to complete this function.
6.4.6 RDT Offset
The display screen for entering the RTD Offset for the sensor is shown. The default value is displayed in bold type. Refer to configuring contacting measurements flow diagram (Figure 6-2) to complete this function.
6.4.7 RDT Slope
The display screen for entering the RTD Slope for the sensor is shown. The default value is displayed in bold type. Refer to configuring contacting measurements flow diagram (Figure 6-2) to complete this function.
6.4.8 Temp Comp
The display screen for selecting Temperature Compensation as Slope, Neutral Salt, Cation or Raw is shown. The default value is displayed in bold type. Refer to configuring contacting measurements flow diagram (Figure 6-2) to complete this function.
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
N Cell Constant
S
.00000 /cm
1
1: 1.234µS/cm 1.234°C
S
2: 12.34pH 1.234°C
S
3: 12.34µS/cm 1.234°C
S
SN RTD Offset
0.00°C
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
SN RTD Slope
2.00/°C
1: 1.234µS/cm 1.234°C
S S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
SN Temp Comp
Slope
Neutral Salt Cation Raw
6.4.9 Slope
The display screen for entering the conductivity/temp Slope is shown. The default value is displayed in bold type. Refer to configuring contacting measurements flow diagram (Figure 6-2) to complete this function.
6.4.10 Reference Temp
The display screen for entering the Reference Temperature is shown. The default value is displayed in bold type. Refer to configuring contacting measurements flow diagram (Figure 6-2) to complete this function.
6.4.11 Filter
The display screen for entering Input filter value is shown in seconds. The default value is displayed in bold type. Refer to configuring contacting measurements flow diagram (Figure 6-2) to complete this function.
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
SN Slope
2.00 /°C
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
SN Ref Temp
(25.0°C normal)
+25.0°C
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
SN Input filter
02 sec
Section 6: Programming Measurements36
Reference Manual
00809-0100-3157
6.4.12 Custom Setup
he display screens for creating a Custom Curve for converting conductivity to
T concentration is shown. Refer to configuring contacting measurements flow diagram (
When the custom curve data entry is complete, press ENTER. The display confirms the determination of a custom curve fit to the entered data by displaying this screen:
If the custom curve fit is not completed or is unsuccessful, the adjacent screen appears and the screen returns to the beginning custom curve screen.
Figure 6-2) to complete this function.
Section 6: Programming Measurements
March 2020
1: 1.234µS/cm 1.234°C
S
2: 12.34pH 1.234°C
S
3: 12.34µS/cm 1.234°C
S
SN Custom Curve Configure Enter Data Points Calculate Curve
S1: 1.234µS/cm 1.234°C
2: 12.34pH 1.234°C
S
3: 12.34µS/cm 1.234°C
S
N Calculate Curve
S Custom curve fit completed. In Process Cal recommended.
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
SN Calculate Curve Failure
6.4.13 Cal Factor
Upon initial installation and power up, if 4-electrode was selected for the sensor type in the Quick Start menus, the user enters a Cell Constant and a Cal Factor using the instrument keypad. The cell constant is needed to convert measured conductance to conductivity as displayed on the analyzer screen. The Cal Factor entry is needed which increases the accuracy of the live conductivity readings, especially at low conductivity readings below 20uS/cm. Both the Cell Constant and the Cal Factor are printed on the tag attached to the 4-electrode sensor/cable.
The display screen for entering Cal Factor is shown. The default value is displayed in bold type. If necessary after initial installation and start-up, enter the Cal Factor as printed on the sensor tag.
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
SN Cal Factor
0
.95000/cm
Section 6: Programming Measurements 37
Section 6: Programming Measurements
March 2020
igure 6-1 Configuring pH/ORP Measurements
F
Reference Manual
00809-0100-3157
Section 6: Programming Measurements38
Reference Manual
00809-0100-3157
Figure 6-2 Configuring Contacting Measurements
Section 6: Programming Measurements
March 2020
Section 6: Programming Measurements 39
Section 6: Programming Measurements
March 2020
Reference Manual
00809-0100-3157
Section 6: Programming Measurements40
Reference Manual
00809-0100-3157
Section 7: Calibration
7.1 Calibration - Introduction
alibration is the process of adjusting or standardizing the analyzer to a lab test or a calibrated
C laboratory instrument, or standardizing to some known reference (such as a commercial buffer). The auto-recognition feature of the analyzer will enable the appropriate calibration screens to allow calibration for any single sensor configuration or dual sensor configuration of the analyzer. Completion of Quick Start upon first power up enables live measurements but does not ensure accurate readings in the lab or in process. Calibration should be performed with each attached sensor to ensure accurate, repeatable readings. This section covers the following programming and configuration functions:
1. Auto buffer cal for pH (pH Cal -
2. Manual buffer cal for pH (pH Cal - Section 7.2)
3. Set calibration stabilization criteria for pH (pH Cal - Section 7.2)
4. Standardization calibration (1-point) for pH, ORP and Redox (pH Cal - Section 7.2 and 7.3)
5. Entering the cell constant of a conductivity sensor (Conductivity Cal - Section 7.4 and 7.5)
6. Calibrating the sensor in a conductivity standard (Conductivity Cal - Section 7.4 and 7.5)
7. Calibrating the analyzer to a laboratory instrument (Conductivity Cal - Section 7.4)
8. Enter a manual reference temperature for temperature compensation of the process measure ment
Section 7.2)
Section 7: Calibration
March 2020
7.2 pH Calibration
7.2.1 Description
New sensors must be calibrated before use. Regular recalibration is also necessary. Use auto calibration instead of manual calibration. Auto calibration avoids common pitfalls and reduces errors. The analyzer recognizes the buffers and uses temperature-corrected pH values in the calibration. Once the Model 1057 successfully completes the calibration, it calculates and displays the calibration slope and offset. The slope is reported as the slope at 25 °C. This section describes how to calibrate
the model 1057 with a ph sensor. the following cali-
Table 7-1 pH calibration routines
Measure Section Menu Function
pH
Section 7: Calibration 41
7.2.2 Auto Calibration
7.2.3 Manual Calibration
7.2.4
7.2.5 Standardization
Entering A Known Slope Value
A detailed flow diagram is provided at the end of this section to guide you through the calibration routines.
To calibrate pH:
1. Press the MENU button
2. Select Calibrate. Press ENTER.
3. Select Sensor 1or Sensor 2 corresponding to pH. Press ENTER.
4. Select pH. Press ENTER.
bration routines are covered.
Default Settings
pH
pH
pH
pH
Description
Two point buffer calibration with auto buffer recognition
Two point buffer calibration with manual buffer value entry
Slope calibration with manual entry of known slope value
One point buffer calibration with manual buffer value entry
Section 7: Calibration
March 2020
The adjacent screen appears. To Calibrate pH or Temperature scroll to the
esired item and press ENTER.
d
The following sub-sections show the initial display screen that appears for each calibration routine. Use the flow diagram (
Figure 7-1) for pH calibration at the end of this section and the live screen
prompts to complete calibration.
7.2.2 Auto Calibration - pH
This screen appears after selecting pH cal.
Note that pH auto calibration criteria can be changed.
The following criteria can be adjusted:
Stabilization time (default 10 seconds)
Stabilization pH value (default 0.02 pH)
Type of Buffer used for AUTO CALIBRATION (default is Standard, non-commercial buffers).
The following commercial buffer tables are recognized by the analyzer:
Reference Manual
00809-0100-3157
S1: 1.234µS/cm 1.234°C
2: 12.34pH 1.234°C
S
3: 12.34µS/cm 1.234°C
S
SN Calibrate?
pH
emperature
T
1: 1.234µS/cm 1.234°C
S S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
SN pH Cal Buffer Cal Standardize
Slope: 59.16mV/pH Offset: 600 mV
Standard (NIST plus pH7)
DIN 19267
Ingold
Merck
The adjacent screen appears to allow adjustment of above criteria:
The adjacent screen appears if the Auto Cal is successful. The screen returns to the pH Buffer Cal menu.
The following screen appears if the Auto Cal is unsuccessful.
1. A High Slope Error generates the adjacent screen display:
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
SN Setup Stable Time: 10 sec Stable Delta: 0.02 pH Buffer: Standard
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
SN pH Auto Cal Slope: 59.16 mV/pH Offset: 60 mV
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
SN pH Auto Cal High Slope Error Calculated: 62.11 mV/pH Max: 62.00 mV/pH Press EXIT
2. A Low Slope Error generates the adjacent screen display:
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234 S3: 12.34µS/cm 1.234
°C
°C
SN pH Auto Cal Low Slope Error Calculated: 39.11mV/pH Min: 40.00 mV/pH Press EXIT
Section 7: Calibration 42
Reference Manual
00809-0100-3157
3. An Offset Error generates the adjacent screen display:
7.2.3 Manual Calibration - pH
New sensors must be calibrated before use. Regular recalibration is also necessary. Use manual calibration if non-standard buffers are being used; otherwise, use auto calibration. Auto calibration avoids common pitfalls and reduces errors.
The adjacent screen appears after selecting pH Manual cal.
7.2.4 Entering A Known Slope Value — pH
If the electrode slope is known from other measurements, it can be entered directly in the Model 1057 analyzer. The slope must be entered as the
Slope at 25 °C.
Section 7: Calibration
March 2020
1: 1.234µS/cm 1.234°C
S S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
SN pH Auto Cal Offset Error Calculated: 61.22mV Max: 60.00mV Press EXIT
1: 1.234µS/cm 1.234°C
S S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
SN pH Manual Cal Buffer 1 Buffer 2
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
SN pH Slope@25°C
59.16 mV/pH
7.2.5 Standardization — pH
The pH measured by the Model 1057 analyzer can be changed to match the reading from a second or referee instrument. The process of making the two readings agree is called standardization. During standardization, the difference between the two pH values is converted to the equivalent volt age. The voltage, called the reference offset, is added to all subsequent measured cell voltages before they are converted to pH. If a standardized sensor is placed in a buffer solution, the measured pH will differ from the buffer pH by an amount equivalent to the standardization offset.
The following screen may appear if ORP Cal is unsuccessful.
An Offset Error generates the adjacent screen display:
If the ORP Cal is successful, the screen returns to the Cal sub-menu.
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.23°C
SN Enter Value
07.00pH
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234°C S3: 12.34µS/cm 1.234°C
SN Standardize Offset Error Calculated: 96mV Max: 60mV Press EXIT
Section 7: Calibration 43
Section 7: Calibration
March 2020
Reference Manual
00809-0100-3157
7.3 ORP Calibration
7.3.1 Description
For process control, it is often important to make the measured ORP agree with the ORP of a standard
olution. During calibration, the measured ORP is made equal to the ORP of a standard solution
s at a single point.
This section describes how to calibrate the model 1057 with ORP sensor. The following calibration routine is covered.
Table 7-2 ORP calibration routine
Measure Section Menu Function
ORP 7.3.2 Standardization ORP One point buffer calibration with manual buffer value entry
Default Settings
Description
A detailed flow diagram (
Figure 7-2) is provided at the end of this section to guide you through the
ORP calibration routines.
To calibrate ORP:
1. Press the MENU button
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to ORP. Press ENTER.
4. Select ORP. Press ENTER.
The adjacent screen appears. To calibrate ORP or Temperature, scroll to the desired item and press ENTER.
The following sub-sections show the initial display screen that appears for each calibration routine. Use the flow diagram (Figure 7-2) for ORP calibration at the end of this section and the live screen prompts to complete calibration.
7.3.1 Standardization — ORP
For process control, it is often important to make the measured ORP agree with the ORP of a standard solution. During calibration, the measured ORP is made equal to the ORP of a standard solution at a single point.
The adjacent screen appears after selecting ORP calibration.
If the ORP Cal is successful, the screen returns to the Cal sub-menu.
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234
S3: 12.34µS/cm 1.234
SN Calibrate?
ORP Temperature
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234
S3: 12.34µS/cm 1.234
SN Enter Value
+0600 mV
°C
°C
°C
°C
The adjacent screen appears if ORP Cal is unsuccessful.
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234
S3: 12.34µS/cm 1.234
SN Standardize Offset Error Calculated: 61.22mV Max: 60.00mV Press EXIT
Section 7: Calibration44
°C
°C
Reference Manual
00809-0100-3157
7.4 Contacting Conductivity Calibration
7.4.1 Description
This section describes how to calibrate a Model 1057 with Contacting Conductivity Sensor attached.
Placing A New Conductivity Sensor In Service
New conductivity sensors rarely need calibration. The cell constant printed on the label is sufficiently accurate for most applications.
Calibrating An In-Service Conductivity Sensor
After a conductivity sensor has been in service for a period of time, recalibration may be necessary. There are three ways to calibrate a sensor.
1. Use a standard instrument and sensor to measure the conductivity of the process stream. It is not necessary to remove the sensor from the process piping. The temperature correction used by the standard instrument may not exactly match the temperature correction used by the Model 1057. To avoid errors, turn off temperature correction in both the analyzer and the standard instrument.
2. Place the sensor in a solution of known conductivity and make the analyzer reading match the conductivity of the standard solution. Use this method if the sensor can be easily removed from the process piping and a standard is available. Be careful using standard solu tions having conductivity less than 100 µS/cm. Low conductivity standards are highly sus ceptible to atmospheric contamination. Avoid calibrating sensors with 0.01/cm cell con stants against conductivity standards having conductivity greater than 100 µS/cm. The resistance of these solutions may be too low for an accurate measurement. Calibrate sen sors with 0.01/cm cell constant using method 3.
3. To calibrate a 0.01/cm sensor, check it against a standard instrument and 0.01/cm sensor while both sensors are measuring water having a conductivity between 5 and 10 µS/cm. To avoid drift caused by absorption of atmospheric carbon dioxide, saturate the sample with air before making the measurements.
To ensure adequate flow past the sensor during calibration, take the sample downstream from the sensor. For best results, use a flow-through standard cell. If the process temperature is much different from ambient, keep connecting lines short and insulate the flow cell.
Section 7: Calibration
March 2020
Table 7-3 Contacting conductivity calibration routines
Measure Section Menu Function
Contacting Conductivity
7.4.2
7.4.3
7.4.4
7.4.5
7.4.6
Cell K 1.00000/cm Enter the cell Constant for the sensor
Zero Cal Zero the analyzer with the sensor attached
In Process Cal Standardize the sensor to a known conductivity
Meter Cal Calibrate the analyzer to a lab conductivity instrument
Cal Factor 0.95000/cm Enter the Cal Factor for 4-Electrode sensors from the sensor tag
Default Settings
Description
A detailed flow diagram (Figure 7-3) is provided at the end of this section to guide you through the calibration routines.
To calibrate contacting conductivity:
1. Press the MENU button
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to contacting conductivity. Press ENTER.
4. Select Conductivity. Press ENTER.
Section 7: Calibration 45
Section 7: Calibration
March 2020
he adjacent screen appears. To calibrate Conductivity or Temperature, scroll
T to the desired item and press ENTER.
The following sub-sections show the initial display screen that appears for each calibration routine. Use the flow diagram (
Figure 7-3) for conductivity calibration at the end of this section and the live
screen prompts for each routine to complete calibration.
The adjacent screen appears after selecting conductivity Calibration.
7.4.2 Entering The Cell Constant
New conductivity sensors rarely need calibration. The cell constant printed on the label is sufficiently accurate for most applications. The cell constant should be entered:
Reference Manual
00809-0100-3157
S1: 1.234µS/cm 1.234°C
2: 12.34pH 1.234
S
3: 12.34µS/cm 1.234
S
SN Calibrate? Conductivity Temperature
S1: 1.234µS/cm 1.234°C
2: 12.34pH 1.234
S
3: 12.34µS/cm 1.234
S
SN Calibration Zero Cal In Process Cal Meter Cal Cell K: 1.00000/cm
°C
°C
°C
°C
When the unit is installed for the first time
When the probe is replaced
The display screen for entering Cell Constant for the sensor is shown. The default value is displayed in bold type.
7.4.3 Zeroing The Instrument
This procedure is used to compensate for small offsets to the conductivity signal that are present even when there is no conductivity to be measured. This procedure is affected by the length of extension cable and should always be repeated if any changes in extension cable or sensor have been made. Electrically connect the conductivity probe as it will actually be used and place the measuring portion of the probe in air. Be sure the probe is dry.
The adjacent screen appears after selecting Zero Cal from the conductivity calibration screen.
The adjacent screen appears if Zero Cal is successful. The screen returns to the conductivity Cal menu.
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234
S3: 12.34µS/cm 1.234
SN Cell Constant
1.00000 /cm
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234
S3: 12.34µS/cm
°CSN Zero Cal
1.234
In Air In Water
S1: 1.234µS/cm 1.234°C
S2: 12.34pH 1.234 S3: 12.34µS/cm 1.234
SN Zero Cal
Sensor Zero Done
°C
°C
°C
°C
°C
The adjacent screen appears if Zero Cal is unsuccessful.
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234
S3: 12.34µS/cm 1.234
°C
SN Zero Cal Sensor Zero Fail Offset too high
Press EXIT
Section 7: Calibration46
°C
Reference Manual
00809-0100-3157
Section 7: Calibration
7.4.4 Calibrating The Sensor In A Conductivity Standard (In Process Cal)
This procedure is used to calibrate the sensor and analyzer against a solution of known conductivity.
his is done by submerging the probe in the sample of known conductivity, then adjusting the dis
T played value, if necessary, to correspond to the conductivity value of the sample. Turn temperature correction off and use the conductivity of the standard. Use a calibrated thermometer to measure temperature. The probe must be cleaned before performing this procedure.
The adjacent screen appears after selecting In Process Cal from the conductivity calibration screen.
The adjacent screen appears if In Process Cal is successful. The screen returns to the conductivity Cal menu.
S1: 1.234µS/cm 1.234°C
2: 12.34pH 1.234
S
3: 12.34µS/cm 1.234
S
SN InProcess Cal Wait for stable reading.
1: 1.234µS/cm 1.234
S S2: 12.34pH 1.234
S3: 12.34µS/cm 1.234
SN InProcess Cal Updated cell constant:
1.00135/cm
March 2020
°C
°C
°C
°C
°C
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234
The adjacent screen may appear if In Process Cal is unsuccessful. The screen returns to the conductivity Cal menu.
S3: 12.34µS/cm 1.234
SN InProcess Cal Calibration Error Press EXIT
7.4.5 Calibrating The Sensor To A Laboratory Instrument (Meter Cal)
This procedure is used to check and correct the conductivity reading of the Model 1057 using a laboratory conductivity instrument. This is done by submerging the conductivity probe in a bath and measuring the conductivity of a grab sample of the same bath water with a separate laboratory instrument. The Model 1057 reading is then adjusted to match the conductivity reading of the lab instrument.
The adjacent screen appears after selecting Meter Cal from the conductivity calibration screen
After pressing ENTER, the display shows the live value measured by the sensor.
If the Meter Cal is successful the screen returns to the conductivity Cal menu.
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234
S3: 12.34µS/cm 1.234
SN Meter Cal Use precision resistors only
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234
S3: 12.34µS/cm 1.234
SN Enter Value
xx.xx k
°C
°C
°C
°C
°C
°C
The adjacent screen appears if Meter Cal is unsuccessful. The screen returns to the conductivity Cal menu.
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234
S3: 12.34µS/cm 1.234
SN Meter Cal Calibration
°C
°C
Error Press EXIT
Section 7: Calibration 47
Section 7: Calibration
March 2020
Reference Manual
00809-0100-3157
7.4.6 Cal Factor
pon initial installation and power up, if 4-electrode was selected for the sensor type in the Quick
U
Start menus, the user enters Cell Constant and Cal Factor using the instrument keypad. The cell
constant is needed to convert measured conductance to conductivity as displayed on the analyzer screen. The Cal Factor entry is needed to increase the accuracy of the live conductivity readings, especially at low conductivity readings below 20 uS/cm. Both the Cell Constant and the Cal Factor are printed on the tag attached to the 4-electrode sensor/cable.
The display screen for entering Cal Factor is shown. The default value is displayed in bold type. If necessary after initial installation and start-up, enter the Cal Factor as printed on the sensor tag.
1: 1.234µS/cm 1.234
S S2: 12.34pH 1.234
S3: 12.34µS/cm 1.234
SN Cal Factor
0.95000 /cm
°C
°C
°C
7.5 Calibrating Temperature
7.5.1 Description
Most liquid analytical measurements require temperature compensation (except ORP). The Model
1057 performs temperature compensation automatically by applying internal temperature correc­tion algorithms. Temperature correction can also be turned off. If temperature correction is off, the Model 1057 uses the manual temperature entered by the user in all temperature correction calculations.
This section describes how to calibrate temperature in the model 1057 analyzer. the following
calibration routine is covered.
Table 7-4 Temperature calibration routine
Default
Measure Section Menu Function
Settings
Description
Temperature 7.5.2 Calibrate Enter a manual reference temperature for temperature
compensation of the process measurement
To calibrate temperature:
1. Press the MENU button
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to the desired measurement. Press ENTER.
4. Select Temperature. Press ENTER.
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234
S3: 12.34µS/cm 1.234
SN Calibrate
+025.0°C
The adjacent screen appears.
The following sub-section provides you with the initial display screen that appears for temperature calibration. Use the flow diagram (Figure 7-4) for temperature calibration at the end of this section to complete calibration.
7.5.2 Calibration
The adjacent screen appears during Temperature Cal.
If the sensor Temperature offset is greater than 5 °C from the default value, the adjacent screen appears. You may continue by selecting Yes or suspend this operation by selecting No. If the Temp Cal is successful, the screen returns to the Cal menu.
Note: To select automatic or manual temp compensation or to program temperature units as °C or °F, refer to Section 5.3 in this manual
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234
S3: 12.34µS/cm 1.234
SN Calibrate Cal in progress. Please wait.
S1: 1.234µS/cm 1.234°C S2: 12.34pH 1.234
S3: 12.34µS/cm 1.234
SN Temp Offset > 5°C
Continue?
No
Yes
°C
°C
°C
°C
°C
°C
Section 7: Calibration48
Reference Manual
00809-0100-3157
Figure 7-1 Calibrate pH
Section 7: Calibration
March 2020
Section 7: Calibration 49
Section 7: Calibration
March 2020
Figure 7-2 Calibrate ORP
Reference Manual
00809-0100-3157
Section 7: Calibration50
Reference Manual
00809-0100-3157
Figure 7-3 Calibrate Contacting Conductivity
Section 7: Calibration
March 2020
Section 7: Calibration 51
Section 7: Calibration
March 2020
Figure 7-4 Calibrate Temperature
Reference Manual
00809-0100-3157
Section 7: Calibration52
Reference Manual
00809-0100-3157
Section 8: Return of Material
8.1 General
o expedite the repair and return of instruments, proper communication between the customer
T and the factory is important. Before returning a product for repair, call 1-949-757-8500 for a Return Materials Authorization (RMA) number.
8.2 Warranty Repair
The following is the procedure for returning instruments still under warranty:
1. Call Emerson for authorization.
2. To verify warranty, supply the factory sales order number or the original purchase order number. In the case of individual parts or sub-assemblies, the serial number on the unit must be supplied.
3. Carefully package the materials and enclose your “Letter of Transmittal” (see Warranty). If possible, pack the materials in the same manner as they were received.
4. Send the package prepaid to:
Emerson
8200 Market Blvd.
Chanhassen, MN 55317
Attn: Factory Repair
RMA No. ____________
Mark the package: Returned for Repair
Model No. ____
Section 8: Return of Material
March 2020
IMPORTANT
Please see second section of “Return of Materials Request” form. Compliance with the OSHA requirements is mandatory for the safety of all personnel. MSDS forms and a certification that the instruments have been disinfected or detoxified are required.
8.3 Non-Warranty Repair
The following is the procedure for returning for repair instruments that are no longer under warran-
ty:
1. Call Emerson for authorization.
2. Supply the purchase order number, and make sure to provide the name and telephone number of the individual to be contacted should additional information be needed.
3. Do Steps 3 and 4 of Section 8.2.
Note: Consult the factory for additional information regarding service or repair.
Section 8: Return of Material 53
Section 8: Return of Material
March 2020
Reference Manual
00809-0100-3157
WARRANTY
eller warrants that the firmware will execute the programming instructions provided by Seller, and that the Goods
S manufactured or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expiration of the applicable warranty period. Goods are warranted for twelve (12) months from the date of initial installation or eighteen (18) months from the date of shipment by Seller, whichever period expires first. Consumables, such as glass electrodes, membranes, liquid junctions, electrolyte, o-rings, catalytic beads, etc.,
and Services are warranted for a period of 90 days from the date of shipment or provision.
Products purchased by Seller from a third party for resale to Buyer ("Resale Products") shall carry only the warranty extended by the original manufacturer. Buyer agrees that Seller has no liability for Resale Products beyond making a reasonable commercial effort to arrange for procurement and shipping of the Resale Products.
If Buyer discovers any warranty defects and notifies Seller thereof in writing during the applicable warranty period, Seller shall, at its option, promptly correct any errors that are found by Seller in the firmware or Services, or repair or replace F.O.B. point of manufacture that portion of the Goods or firmware found by Seller to be defective, or refund the purchase price of the defective portion of the Goods/Services.
All replacements or repairs necessitated by inadequate maintenance, normal wear and usage, unsuitable power sources, unsuitable environmental conditions, accident, misuse, improper installation, modification, repair, storage or handling, or any other cause not the fault of Seller are not covered by this limited warranty, and shall be at Buyer's expense. Seller shall not be obligated to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in writing in advance by an authorized Seller representative. All costs of dismantling, reinstallation and freight and the time and expenses of Seller's personnel for site travel and diagnosis under this warranty clause shall be borne by Buyer unless accepted in writing by Seller.
Goods repaired and parts replaced during the warranty period shall be in warranty for the remainder of the original warranty period or ninety (90) days, whichever is longer. This limited warranty is the only warranty made by Seller and can be amended only in a writing signed by an authorized representative of Seller. Except as otherwise expressly provided in the Agreement, THERE ARE NO REPRESENTATIONS OR WARRANTIES OF ANY KIND, EXPRESS OR IMPLIED, AS TO MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE, OR ANY OTHER MATTER WITH RESPECT TO ANY OF THE GOODS OR SERVICES.
RETURN OF MATERIAL
Material returned for repair, whether in or out of warranty, should be shipped prepaid to:
Emerson
8200 Market Blvd.
Chanhassen, MN 55317
The shipping container should be marked:
Return for Repair
Model _________________________________
The returned material should be accompanied by a letter of transmittal which should include the following information (make a copy of the "Return of Materials Request" found on the last page of the Manual and provide the following thereon):
1. Location type of service, and length of time of service of the device.
2. Description of the faulty operation of the device and the circumstances of the failure.
3. Name and telephone number of the person to contact if there are questions about the returned material.
4. Statement as to whether warranty or non-warranty service is requested.
5. Complete shipping instructions for return of the material.
Adherence to these procedures will expedite handling of the returned material and will prevent unnecessary additional charges for inspection and testing to determine the problem with the device.
If the material is returned for out-of-warranty repairs, a purchase order for repairs should be enclosed.
Section 8: Return of Material 54
Reference Manual
00809-0100-3157
Return of Material
March 2020
Return of Material 55
00809-0100-3157
Rev. AD
March 2020
Emerson.com/Rosemount
Youtube.com/user/Rosemount
Twitter.com/Rosemount_News
Emerson
8200 Market Blvd. Chanhassen, MN 55317, USA Tel +1 800 999 9307 Fax +1 952 949 7001 RMTNA.RCCPO@Emerson.com
Linkedin.com/company/Emerson-Automation-Solutions
Facebook.com/Rosemount
©2020 Emerson. All rights reserved.
The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a mark of one of the Emerson family of companies. All other marks are the property of their respective owners.
The contents of this publication are presented for information purposes only, and while effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, express or implied, regarding the products or services described herein or their use or applicability. All sales are governed by our terms and conditions, which are available on request. We reserve the right to modify or improve the designs or specifications of our products at any time without notice.
Loading...