Rosemount 1055-Cl-pH Chlorine/pH Analyzer Abridged Manuals & Guides

Page 1
Dual Input Chlorine/pH Analyzer
Model Options 1055-24-32 and 1055-24
Instruction Sheet
PN 51A-1055CLpH/rev.K August 2006
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
Your purchase from Rosemount Analytical, Inc. has resulted in one of the finest instruments available for your particular application. These instruments have been designed, and tested to meet many national and international standards. Experience indicates that its performance is directly related to the quality of the installation and knowledge of the user in operating and maintaining the instrument. To ensure their con­tinued operation to the design specifications, person­nel should read this manual thoroughly before proceeding with installation, commissioning, opera­tion, and maintenance of this instrument. If this equip­ment is used in a manner not specified by the manufacturer, the protection provided by it against hazards may be impaired.
• Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
• Ensure that you have received the correct model and options from your purchase order. Verify that this manual covers your model and options. If not, call 1-800-854-8257 or 949-757-8500 to request correct manual.
• For clarification of instructions, contact your Rosemount representative.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Use only qualified personnel to install, operate, update, program and maintain the product.
• Educate your personnel in the proper installation, operation, and maintenance of the product.
• Install equipment as specified in the Installation section of this manual. Follow appropriate local and national codes. Only connect the product to electri­cal and pressure sources specified in this manual.
• Use only factory documented components for repair. Tampering or unauthorized substitution of parts and procedures can affect the performance and cause unsafe operation of your process.
• All equipment doors must be closed and protective covers must be in place unless qualified personnel are performing maintenance.
• If this equipment is used in a manner not specified by the manufacturer, the protection provided by it against hazards may be impaired.
WARNINGS
RISK OF ELECTRICAL SHOCK
Equipment protected throughout by double insulation.
• Installation of cable connections and servicing of this product require access to shock hazard voltage levels.
• Main power and relay contacts wired to separate power source must be disconnected before servicing.
• Do not operate or energize instrument with case open!
• Signal wiring connected in this box must be rated at least 240 V.
• Non-metallic cable strain reliefs do not provide grounding between conduit connections! Use grounding type bushings and jumper wires.
• Unused cable conduit entries must be securely sealed by non­flammable closures to provide enclosure integrity in compliance with personal safety and environmental protection requirements. Unused conduit openings must be sealed with NEMA 4X or IP65 conduit plugs to maintain the ingress protection rating (NEMA 4X).
• Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA-70) and/or any other applicable national or local codes.
• Operate only with front and rear panels fastened and in place over terminal area.
• Safety and performance require that this instrument be con­nected and properly grounded through a three-wire power source.
Proper relay use and configuration is the responsibility of the user.
CAUTION
This product generates, uses, and can radiate radio frequency energy and thus can cause radio communication interference. Improper installation, or operation, may increase such interfer­ence. As temporarily permitted by regulation, this unit has not been tested for compliance within the limits of Class A comput­ing devices, pursuant to Subpart J of Part 15, of FCC Rules, which are designed to provide reasonable protection against such interference. Operation of this equipment in a residential area may cause interference, in which case the user at his own expense, will be required to take whatever measures may be required to correct the interference.
WARNING
This product is not intended for use in the light industrial, residential or commercial environments per the instrument’s certification to EN50081-2.
Model 1055 SOLU COMP®II
For additional information, please refer to the Instruction Manuals CD shipped with this product, or visit our website at www.emersonprocess.com/raihome/liquid/.
Page 2
2
MODEL SOLU COMP II SPECIFICATIONS
SPECIFICATIONS - General
Case: ABS (panel mount), polycarbonate (surface/pipe
mount). All versions are NEMA 4X/CSA 4 (IP65).
Dimensions
Panel (code -10): 6.10 x 6.10 x 3.72 in. (155 x
155 x 94.5 mm) Surface/Pipe (code -11): 6.23 x 6.23 x 3.23 in.
(158 x 158 x 82 mm); see page 5 for dimensions of pipe mounting bracket.
Conduit openings: Accepts PG13.5 or 1/2 in. con-
duit fittings
Display: Two line, 16-character, back-lit display.
Character height: 4.8 mm. Display can be cus­tomized to meet individual requirements. Depending on number of sensors, as many as 14 display screens are available.
Ambient temperature and humidity: 0 to 50°C, (32
to 122°F) RH 5 to 95% (non-condensing) Note: The analyzer is operable from -20 to 60°C
(-4 to 140°F) with some degradation in display performance.
Power:
Code -01: 115/230 Vac ±15%, 50/60 Hz ±6%, 8.0W Code -02*: 24 Vdc ±15%, 6.0W
Installation Category II
*
For +24Vdc Power Supply use only devices meeting NEC Class II or UL recognized (UL 1950).
Equipment protected throughout by double insula­tion.
Hazardous Location:
Class I, Division 2, Groups A, B, C, & D
POLLUTION DEGREE 4: Extended Environment
Outdoor use where conductive contamination such as rain, snow, or dust may be present.
(Hazardous Location only)
RFI/EMI: EN-61326 LVD: EN-61010-1
Input: Choice of single or dual sensor input with
measurement choices of pH/ORP, conductivity/ resistivity, toroidal conductivity, flow, chlorine, dis­solved oxygen, and dissolved ozone. Field-com­missioned units allow user to change measure­ments on either or both inputs. See combination guide for valid combinations. For contacting con­ductivity measurements, temperature element must be a Pt 1000 RTD. For other measure­ments, use either a Pt100 RTD, Pt1000 RTD, or 22k NTC (D.O. only).
Outputs: Two 4-20 mA or 0-20 mA isolated outputs.
Continuously adjustable. Linear or logarithmic. Maximum load 500 ohms. Output dampening with time constant of 5 sec is user-selectable.
Alarms: Three alarm relays for process measure-
ment(s) or temperature. Alarm 3 can be configured as a fault alarm, instead of a
process alarm. Each relay can be configured independently. Alarm logic (high or low activation or USP*) and deadband are user-programmable. The USP* alarm can be programmed to activate when the conductivity is within a user-selectable percentage of the limit.
*conductivity/resistivity measurement only
Relays: Form C, single pole double throw, epoxy sealed
Terminal Connections Rating: 26-14 AWG wire size
Weight/Shipping weight (rounded up to nearest lb or
nearest 0.5 kg): 3 lb (1.5 kg)/4 lb (2.0 kg)
Resistive Inductive
28 Vdc 5.0 A 3.0 A 115 Vac 5.0 A 3.0 A 230 Vac 5.0 A 1.5 A
Ordinary Location: (-68 only)
POLLUTION DEGREE 2: Normally only non-con­ductive pollution occurs. Occasionally, however, a temporary conductivity caused by condensation must be expected.
12RN
Page 3
3
QUICK START GUIDE FOR MODEL SOLU COMP II
CHLORINE/pH ANALYZER (Model Options 1055-24-32 and 1055-24)
1. Refer to page 9 for installation instructions.
2. Make alarm, output, and power connections as shown below.
3. Wire sensor(s) to the analyzer. See the drawings below.
Wiring Connections for Solu Comp II Model 1055-01-10
(Panel Mount with 115/230 Vac Power)
Wiring Connections for Solu Comp II Model 1055-02-10
(Panel Mount with 24 Vdc Power)
Wiring Connections for Solu Comp II Model 1055-01-11
(Surface/Pipe Mounting with 115/230 Vac Power)
Wiring Connections for Solu Comp II Model 1055-02-11
(Surface/Pipe Mounting with 24 Vdc Power)
499ACL-01, -02, and -03 sensors only.
The free chlorine (499ACL-01), total chlorine (499ACL-02), and monochloramine (499ACL-03) sensors have identical wiring.
499ACL chlorine sensor with standard cable.
499ACL chlorine sensor with optimum EMI/RFI
or Variopol cable.
Page 4
MODEL SOLU COMP II QUICK-START — WIRING
499ACL-01 (free chlorine sensor) with pH sensor for automatic pH correction.
NOTE
When wiring a pH and free chlorine sensor to the analyzer, connect the anode and pH ref­erence terminals (TB3-1 and TB5-6) with the jumper provided with the analyzer. Depending on the date the instrument was shipped, the jumper may be either a 50 kΩ resistor or a 10 MΩ resistor in parallel with a 0.1 μf capacitor. Either jumper is satisfactory.
Insulate and tape back unused wires.
Wiring diagram: 1055 panel mount; 499ACL with
standard cable; 399-09-62
Wiring diagram: 1055 panel mount; 499ACL with
standard cable; 399VP-09
Wiring diagram: 1055 panel mount; 499ACL with
standard cable; 399-14
Wiring diagram: 1055 panel mount; 499ACL with
EMI/RFI or Variopol cable; 399-09-62
Although the blue wire is connected to a terminal labeled solution ground, the blue wire does not connect to a solution ground in the
sensor. The 399-14 sensor has no solution ground. The blue wire, instead, ties the preamplifier in the sensor to instrument common
4
Page 5
MODEL SOLU COMP II QUICK-START — WIRING
5
Wiring diagram: 1055 panel mount; 499ACL with
EMI/RFI or Variopol cable; 399VP-09
Wiring diagram: 1055 pipe/wall mount; 499ACL with
EMI/RFI or Variopol cable; 399-14
Wiring diagram: 1055 pipe/wall mount; 499ACL with
standard cable; 399-09-62
Wiring diagram: 1055 pipe/wall mount; 499ACL with
standard cable; 399VP-09
Although the blue wire is connected to a terminal labeled solution ground, the blue wire does not connect to a solution ground in the
sensor. The 399-14 sensor has no solution ground. The blue wire, instead, ties the preamplifier in the sensor to instrument common
Page 6
MODEL SOLU COMP II QUICK-START — WIRING
6
Wiring diagram: 1055 pipe/wall mount; 499ACL with
standard cable; 399-14
Wiring diagram: 1055 pipe/wall mount; 499ACL with
EMI/RFI or Variopol cable; 399-09-62
Wiring diagram: 1055 pipe/wall mount; 499ACL with
EMI/RFI or Variopol cable; 399VP-09
Wiring diagram: 1055 pipe/wall mount; 499ACL with
EMI/RFI or Variopol cable; 399-14
Although the blue wire is connected to a terminal labeled solution ground, the blue wire does not connect to a solution ground in the
sensor. The 399-14 sensor has no solution ground. The blue wire, instead, ties the preamplifier in the sensor to instrument common
Although the blue wire is connected to a terminal labeled solution ground, the blue wire does not connect to a solution ground in the sensor. The 399-14 sensor has no solution ground. The blue wire, instead, ties the preamplifier in the sensor to instrument common
Page 7
7
498CL-01 (pH-independent free chlorine sensor)
Integral cable and VP sensors have identical wiring.
MODEL SOLU COMP II QUICK-START — WIRING
Wiring diagram: 1055 pipe/wall mount; 498CL-01
Wiring daigram: 1055 panel mount; 498CL-01
Wiring diagram: 1055 panel mount;
498CL-01; 399-09-62
Wiring diagram: 1055 panel mount;
498CL-01; 399VP-09
498CL-01 sensor used with pH sensor
The 498CL-01 (pH-independent free chlorine sensor) does not require a pH sensor for pH correction. However, the sensor can be used with a pH sensor. Wiring diagrams are shown below.
*
USE RC JUMPER PN 23988-00
Page 8
8
Wiring diagram: 1055 pipe/wall mount;
498CL-01; 399VP-09
Wiring diagram: 1055 pipe/wall mount;
498CL-01; 399-14
MODEL SOLU COMP II QUICK-START — WIRING
Wiring diagram: 1055 panel mount;
498CL-01; 399-14
Wiring diagram: 1055 pipe/wall mount;
498CL-01; 399-09-62
*
USE RC JUMPER PN 23988-00
*
USE RC JUMPER PN 23988-00
Page 9
9. Choose the measurement for sensor 2 (pH). If the analyzer was config­ured to measure chlorine only, this screen does not appear.
10. Choose temperature units.
11. The main display appears. The outputs and alarms are assigned to default values.
12. If free chlorine is to be measured using the 499ACL-01 sensor, the ana­lyzer must be configured for automatic or manual pH correction. Go to the main menu and choose Program followed by Measurement. Choose Sensor 1 then Free Chlorine. Choose Auto or Manual pH correction. For manual pH correction, enter the pH of the process liquid.
If the chlorine is to be measured using the 498CL-01 (pH-independent free chlorine sensor), go to the main menu and choose Program fol- lowed by Measurement. Choose Sensor 1 then Free Chlorine. Choose Manual pH correction and enter 7.00.
If total chlorine (499ACL-02 sensor) or monochloramine (499ACL-03 sen­sor) is to be measured, pH correction is NOT required.
13. To change outputs, alarms, and temperature-related settings, go to the main menu and choose Program. Follow the prompts. For a guide to the Program menu, see the menu tree on the following page.
14. To return the analyzer to the default settings, choose Reset Analyzer in the Program menu.
Temperature in?
°C °
F
S2 Measure?
pH
Redox ORP
4. Once connections are secured and verified, apply power to the analyzer.
5. When the analyzer is powered up for the first time, Quick Start screens appear. Using Quick Start is easy.
a. A blinking field shows the position of the cursor.
b. Use the  or  key to move the cursor left or right. Use the or key to move the cursor up or
down or to increase or decrease the value of a digit. Use the or key to move the decimal point.
c. Press ENTER to store a setting. Press EXIT to leave without storing changes. Pressing EXIT also
returns the display to the previous screen.
9
6. Choose the desired language. Choose >> to show more choices.
7. Choose the number of sensors wired to the analyzer.
NOTE
If one sensor is selected, only S1 (chlorine) will be available. S2 (pH/ORP) cannot be chosen for single measurement.
8. Choose the measurement for sensor 1 (chlorine).
# of sensors?
One
Two
S1 Chlorine Type
free
total >>
English
Franais
Espaol >>
Page 10
QUICK REFERENCE GUIDE
MENU TREE FOR CHLORINE/pH MEASUREMENTS
10
Page 11
11
FIGURE 2-1. Panel Mount Installation
Access to the wiring terminals is through the rear cover. Four screws hold the cover in place.
MILLIMETER
INCH
INSTALLATION
UNPACKING AND INSPECTION.
Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions. Save the box. If there is no apparent damage, unpack the container. Be sure all items shown on the packing list are present. If items are missing, notify Rosemount Analytical immediately.
INSTALLATION
1. Although the analyzer is suitable for outdoor use, do not install it in direct sunlight or in areas of extreme tempera­tures.
2. Install the analyzer in an area where vibrations and elec­tromagnetic and radio frequency interference are mini­mized or absent.
3. Keep the analyzer and sensor wiring at least one foot from high voltage conductors. Be sure there is easy access to the analyzer.
4. AC power and relay wiring should not enter via top conduit openings and should be kept separated from other wiring in the analyzer after installation.
5. To reduce the likelihood of stress on wiring connections,
the hinged front panel (-11 models) shall not be removed from the base during wiring installation, and there shall be sufficient wire leads to avoid stress on conductors.
6. For UL-approved models (-68), the clear wiring shield
must be installed prior to operation.
MODEL SOLU COMP II INSTALLATION
PANEL MOUNTING
Page 12
MODEL SOLU COMP II INSTALLATION
FIGURE 2-2. Pipe Mount Installation
The front panel is hinged at the bottom. The panel swings down for access to the wiring terminals.
PIPE MOUNTING
MILLIMETER
INCH
12
Page 13
MODEL SOLU COMP II INSTALLATION
13
FIGURE 2-3. Surface Mount Installation
The front panel is hinged at the bottom. The panel swings down for access to the wiring terminals.
SURFACE MOUNTING
MILLIMETER
INCH
Page 14
14
NOTES:
Page 15
15
CALIBRATION
Calibration-Zero
1. Place the sensor in the zero solution.
With the exception of the 498CL-01 (pH-independent free chlorine sensor) keep the sensor in the zero solu­tion for AT LEAST TWO HOURS before zeroing.
For the 498CL-01 sensor, place the sensor in FLOWING CHLORINATED water for two hours before zeroing, then transfer the sensor to deionized water. DO NOT EXPOSE THE SENSOR TO NON-FLOWING DEION­IZED WATER FOR LONGER THAN 30 MINUTES.
2. From the main display, press MENU.
3. Choose Calibrate.
4. Choose Sensor1.
5. Choose Measurement.
6. Choose Zero.
7. Once the sensor current is stable, the analyzer will accept the zero value.
8. Press MENU then EXIT.
Calibration-Full scale
1. Free chlorine (499ACL-01): Place the sensor in the flow cell. If automatic pH correction is being used, calibrate the pH sensor and place it in the process liquid. Adjust the sample flow to the value recommended in the sen­sor manual. Adjust the concentration of chlorine to near the upper end of the operating range.
2. Free chlorine (498CL-01): Place the sensor in the flow cell. Adjust the sample flow to the value recommended in the sensor manual. Adjust the concentration of chlo­rine to near the upper end of the operating range.
3. Total chlorine: Keep the sensor in the low flow cell in the
TCL, and start the flow of reagent. Adjust the concentration of chlorine to near the upper end of the operating range.
4. Monochloramine: Place the sensor in the flow cell. Adjust the sample flow to the value recommended in the sensor manual. Adjust the concentration of chlorine to near the upper end of the operating range.
5. From the main display, press MENU.
6. Choose Calibrate.
7. Choose Sensor1.
8. Choose Measurement.
9. Choose InProcess.
10. Take a sample of the process liquid and immediately determine the chlorine concentration. Use the arrow keys to change the concentration reading in the second line of the display to match the test results. Press ENTER.
11. Press MENU then EXIT.
Calibration-pH
1. The analyzer does an automatic two-point calibration.
2. Obtain two buffer solutions. Ideally, the buffer pH should
bracket the expected pH of the sample.
3. From the main display, press MENU.
4. Choose Calibrate.
5. Choose Sensor2.
6. Choose Measurement.
7. Choose BufferCal.
8. Choose Auto.
9. Choose Buffer1.
10. Place the sensor in the first buffer. Once Wait disap-
pears, the analyzer displays the buffer pH. If the dis­played value is not correct use
to scroll through the
list of standard buffer values. Press ENTER.
11. Place the sensor in the second buffer. Choose Buffer2.
Once Wait disappears, the analyzer displays the buffer pH. If the displayed value is not correct use
to scroll
through the list of standard buffer values. Press ENTER.
12. If the calibration was successful, the analyzer will dis-
play the slope and offset.
13. Press MENU then EXIT.
Free chlorine (499ACL-01)
Deionized water containing at least 500 ppm sodium chloride
Free chlorine (498CL-01)
Deionized water
Total chlorine or monochloramine
Deionized water
Page 16
Credit Cards for U.S. Purchases Only.
The right people, the right answers, right now.
ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
http://www.raihome.com
Specifications subject to change without notice.
Emerson Process Management
Liquid Division
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250
http://www.raihome.com
© Rosemount Analytical Inc. 2006
Loading...