Rosemount 1054B-DO Dissolved Oxygen Microprocessor Manuals & Guides

Model 1054B DO
Dissolved Oxygen Microprocessor Analyzer
Instruction Manual
PN 51-1054BDO/rev.B April 2003
WARNING
ELECTRICAL SHOCK HAZARD
Be sure to disconnect all hazardous voltage before opening the enclosure.
Relay contacts made to separate power sources must be disconnected before servicing.
Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA-
70) and/or any other applicable national or local codes.
Unused cable conduit entries must be securely sealed by non-flammable closures to provide enclosure integrity in compliance with personal safety and environmental protection require­ments.
The unused conduit openings need to be sealed with NEMA 4X or IP65 conduit plugs to maintain the ingress protection rating (IP65).
For safety and proper performance this instru­ment must be connected to a properly ground­ed three-wire power source.
Proper relay use and configuration is the responsibility of the user.
No external connection to the instrument of more than 69VDC or 43V peak allowed with the exception of power and relay terminals. Any vio­lation will impair the safety protection provided
Do not operate this instrument without front cover secured. Refer installation, operation and servicing to qualified personnel..
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PRO-
CEEDING!
Rosemount Analytical designs, manufactures, and tests its products to meet many national and international stan­dards. Because these instruments are sophisticated tech­nical products, you must properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions must be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the product. If this Instruction Manual is not the correct manual, telephone 1-949-757-8500 and the requested manual will be provided. Save this Instruction Manual for future reference.
• If you do not understand any of the instructions, contact your Rosemount representative for clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper instal­lation, operation, and maintenance of the product.
• Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the proper electrical and pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
• When replacement parts are required, ensure that qual­ified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the product’s performance and place the safe operation of your process at risk. Look alike substitu­tions may result in fire, electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protec­tive covers are in place, except when maintenance is being performed by qualified persons, to prevent electri­cal shock and personal injury.
WARNING
This product is not intended for use in the light industrial, residential or commercial environment, per the instru­ment’s certification to EN50081-2.
Emerson Process Management
Rosemount Analytical Inc.
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250
http://www.RAuniloc.com
© Rosemount Analytical Inc. 2001
MODEL 1054B DO TABLE OF CONTENTS
MODEL 1054B DO
MICROPROCESSOR ANALYZER
TABLE OF CONTENTS
Section Title Page
1.0 DESCRIPTION AND SPECIFICATIONS................................................................. 1
1.1 General Description ................................................................................................. 1
1.2 Physical Specifications ............................................................................................ 1
1.3 Instrument Specifications......................................................................................... 1
1.4 Ordering Information ................................................................................................ 2
2.0 INSTALLATION ....................................................................................................... 3
2.1 General .................................................................................................................... 3
2.2 Unpacking and Inspection........................................................................................ 3
2.3 Mechanical Installation............................................................................................. 3
2.4 Electrical Wiring ....................................................................................................... 4
3.0 DESCRIPTION OF CONTROLS ............................................................................. 12
3.1 Keyboard Functions ................................................................................................. 12
4.0 CONFIGURATION ................................................................................................... 16
4.1 General .................................................................................................................... 16
4.2 Alarm 1 and 2........................................................................................................... 19
4.3 Interval Timer ........................................................................................................... 20
4.4 Temperature Configuration ...................................................................................... 21
4.5 Current Output ......................................................................................................... 21
4.6 Dissolved Oxygen Range Units ............................................................................... 22
4.7 Barometric Pressure Units ....................................................................................... 22
4.8 Solubility Correction Factor...................................................................................... 23
4.9 Sensor Compatibility ................................................................................................ 23
4.10 Defaults....................................................................................................................23
4.11 Software Version Number........................................................................................ 24
4.12 Display Test.............................................................................................................. 24
4.13 Alarm Setpoint ......................................................................................................... 25
4.14 Output Scale Expansion .......................................................................................... 26
4.15 Output Display/Output Simulation............................................................................ 27
4.16 Hold.......................................................................................................................... 27
5.0 START-UP AND CALIBRATION............................................................................. 28
5.1 General.................................................................................................................... 28
5.2 Start-up.................................................................................................................... 28
5.3 Calibration................................................................................................................ 28
6.0 KEYBOARD SECURITY......................................................................................... 31
7.0 THEORY OF OPERATION...................................................................................... 32
7.1 General.................................................................................................................... 32
7.2 Measurement Variables ........................................................................................... 32
8.0 DIAGNOSTICS AND TROUBLESHOOTING ......................................................... 34
8.1 Diagnostics .............................................................................................................. 34
8.2 Troubleshooting....................................................................................................... 35
8.3 Instrument Maintenance .......................................................................................... 35
9.0 RETURN OF MATERIALS ...................................................................................... 37
i
MODEL 1054B DO TABLE OF CONTENTS
TABLE OF CONTENTS CONT’D.
LIST OF FIGURES
Figure No. Title Page
2-1 Panel Mounting Cutout ............................................................................................ 5
2-2 Panel Mounting Tab Installation ............................................................................... 6
2-3 Wall Mounting J-Box Installation.............................................................................. 7
2-4 Wall Mounting J-Box Wiring ..................................................................................... 8
2-5 Pipe Mounting Installation........................................................................................ 9
2-6 Electrical Wiring ....................................................................................................... 10
2-7 Wall Mount Enclosure (Option -20) .......................................................................... 11
3-1 Function Select on Keypad...................................................................................... 12
3-2 Accessing Editing Function...................................................................................... 12
3-3 Accessing Configuration Menus .............................................................................. 12
3-4 LCD Display............................................................................................................. 13
4-1 Set Function Menu................................................................................................... 17
4-2 Alarm Setpoint ......................................................................................................... 25
4-3 Output Scale Expansion .......................................................................................... 26
5-1 DO Standardization Formula ................................................................................... 29
LIST OF TABLES
Table No. Title Page
1-1 Replacement Parts .................................................................................................. 2
1-2 Accessories.............................................................................................................. 2
3-1 Key Description........................................................................................................ 14
3-2 Information Mnemonics............................................................................................ 15
3-3 Set Function Mnemonics ......................................................................................... 15
4-1 Configuration Worksheet.......................................................................................... 18
8-1 Fault Mnemonics...................................................................................................... 34
8-2 RTD Resistance Values ........................................................................................... 34
8-3 Troubleshooting Guide............................................................................................. 36
ii
1
MODEL 1054B DO SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
1.1 GENERAL DESCRIPTION. The Model 1054B DO
Microprocessor Analyzer is designed to continuously measure and control dissolved oxygen in industrial and municipal processes.
Housed in a NEMA 4X (IP65) weatherproof corrosion-resistant, flame retardant enclosure, the Model 1054B is suitable for panel, pipe, or wall mounting. All functions are accessed through the front panel membrane keyboard which features tactile feedback.
The 1054B transmits a user selected isolated current output continu­ously expandable over the measurement range in either direct or reverse action. The output can be displayed in milliamps or percent of full scale. Output dampening is user selectable.
Dual programmable alarms are a standard feature on the Model 1054B and are programmable for either high or low operation. Alarm 2 may be programmed as a fault alarm. Both alarms feature inde­pendent setpoints, adjustable hysteresis, and time delay action. The time delay is convenient when an alarm is used for corrective action. Time delay ignores temporary upsets and prevents relay chatter. An interval timer with relay is also provided.
The 1054B DO analyzer, which is intended for use with a mem­brane-covered amperometric sensor, automatically compensates for changes in membrane permeability with temperature. Temperature can be displayed in either °C or °F.
Calibrating the analyzer is as simple as exposing the sensor to air and keying in the barometric pressure. If removing the sensor from the process is impractical, the analyzer can also be calibrated against a laboratory measurement made on a grab sample.
Solubility correction factors for liquids containing high concentrations of electrolytes can be programmed into the analyzer. The micro­processor automatically calculates ppm dissolved oxygen or % satu­ration.
The 1054B DO Microprocessor Analyzer comes with either LCD or LED display. The display indicates dissolved oxygen in ppm or % saturation as well as temperature, alarm status, hold output, and fault conditions.
1.2 PHYSICAL SPECIFICATIONS
Enclosure: Black, ABS, NEMA 4X, IP65
CSA Enclosure 4 144 X 144 X 192 mm (5.7 X 5.7 X 7.6 in.)
Wall Mount Enclosure: NEMA 4X, Heavy duty fiberglass,
reinforced thermoplastic. 356 X 450 X 180 mm (14 X 17.7 X 7.1 in.) Dimensions include latches and mounting feet
Front Panel: Membrane keypad with tactile feedback and user
selectable security.
Digital Display: LCD, black on grey, Optional red LED
Character Height: 18 mm (0.7 in.)
Electrical Classification:
FM Class I, Div. 2, Group A thru D
28 Vdc relays - 5.0 amps resistive only 150 mA - Groups A & B; 400 mA - Group C ; 540 mA - Group D; Ci = 0; Li = 0
CSA Class I, Div. 2, Group A thru D.
28 Vdc, 110 Vac & 230 Vac relays
5.0 Amps resistive only
Wall Mount Enclosure: General Purpose
Power: 100 - 127 VAC, 50/60 Hz ± 6%, 4.0 W
200 - 253 VAC, 50/60 Hz ± 6%, 4.0 W
EMI/RFI: EN61326 LVD: EN61010-1
Model option -20 (Wall Mount Enclosure) does not meet CE requirements
Current Output: Isolated, 0-20 mA or 4-20 mA into
600 ohms maximum load at 115/230 Vac or 550 ohms maximum load at 100/200 Vac; direct or reverse output; dampening: 0-255 sec
Ambient Temperature: -10 to 65°C (14 to 149°F);
-50 to 65°C (-58 to 149°F) with optional heater in wall mount enclosure.
Ambient Humidity: LED: 0-95% RH
LCD: 0-85% RH @ 50°C
Alarms: Dual, field selectable High/Low, High/High, Low/Low
Alarm 2 configurable as a fault alarm Time Delay 0 to 255 seconds Dual Setpoints, continuously adjustable Hysteresis adjustable up to 2.00 ppm for low side/High Alarm and high side/Low Alarm
Interval Timer: Interval: 10 min. to 2999 days
On Counts: 1 to 60 On Duration: 0.1 to 299.9 seconds Off Duration: 0.1 to 299.9 seconds Wait Duration: 0.1 to 299.9 seconds Controls dedicated relay
Relay Contacts: Epoxy Sealed Form A contacts, SPST,
Normally Open.
Resistive
Inductive
28 VDC 5.0 Amps 3.0 Amps
115 VAC 5.0 Amps 3.0 Amps
230 VAC 5.0 Amps 1.5 Amps
Weight/Shipping Weight: 1.1 kg/1.6 kg (2.5 lb/3.5 lb)
1.3 INSTRUMENT SPECIFICATIONS
Operating Ranges: 0-20 ppm (mg/l); 0-250% saturation; 0-50°C Accuracy: ± 1% full scale Repeatability: ± 0.1% of range Stability: Zero Drift: ± 1% full scale/month
Span Drift: ± 1% full scale/month
Response Time: 0-95% full scale in less than 15 secs Temperature Correction for Membrane Permeability:
Automatic between 0-50°C. Temperature compensation can be disabled if desired.
2
MODEL 1054B DO SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
The Model 1054B Dissolved Oxygen Microprocessor Analyzer: Housed in a NEMA 4X corrosion resistant,
weatherproof housing suitable for panel, pipe, or wall mounting. Standard features include isolated digital display, current outputs, dual programmable alarms and timer relay, default settings, and automatic or manual tempera­ture compensation.
MODEL
1054BDO MICROPROCESSOR ANALYZER (3.5 lbs./1.5 kg)
CODE STANDARD ENCLOSURE OPTIONS
01 LCD Display
02 LED Display
CODE OPTIONS
20 Wall Mount Enclosure (wall mount enclosure does not meet CE requirements)
1054BDO 01 20 EXAMPLE
1.4 ORDERING INFORMATION
P/N DESCRIPTION
22966-00 PCB, LCD Digital Display 23025-01 Panel Mounting Kit 23245-01 PCB, LED Digital Display 32937-00 Gasket, Rear Cover 32938-00 Gasket, Front Cover
9100157 Fuse, .10A, 250V, 3AB,
Slo Blo 23739-00 PCB, Power Supply 23740-00 PCB, Motherboard 23695-12 Keyboard Overlay,
LCD Version 23695-13 Keyboard Overlay,
LED Version 33469-00 Enclosure Body 33470-00 Enclosure, Rear Cover
9100160 Fuse, .250A, 125V,
Axial lead PICO II
9100189 Fuse, .750A, 125V,
Axial lead PICO II 23666-00 PCB, CPU, Dissolved
Oxygen
TABLE 1-1. Replacement Parts TABLE 1-2. Accessories
P/N DESCRIPTION
2001492 Tag, Stainless Steel, Specify
Marking 23053-00 Mounting Bracket, 2-inch Pipe 23054-01 Mounting Bracket, Wall, with
Junction Box 23268-01 Heater, 115 VAC, 50/60 Hz,
1054B (Code 20 Only) 23268-02 Heater, 230 VAC, 50/60 Hz,
1054B (Code 20 Only)
3
MODEL 1054B DO SECTION 2.0
INSTALLATION
SECTION 2.0
INSTALLATION
2.1 GENERAL. The analyzer is suitable for outdoor
use. However, it should be located in an area where temperature extremes and vibrations are minimized or absent. Installation must be performed by a trained technician.
2.2 UNPACKING AND INSPECTION. Inspect the analyzer for shipping damage. If damaged, notify the carrier immediately. Confirm that all items shown on the packing list are present. Notify Rosemount Analytical if items are missing.
2.3 MECHANICAL INSTALLATION. Select an instal­lation site that is at least one foot from any high volt­age conduit, has easy access for operating personnel, and is not in direct sunlight. Mount the analyzer as fol­lows:
1. Remove the four screws that secure the rear cover of the enclosure.
2. Remove the four screws holding the front panel assembly of the enclosure and carefully pull the front panel and connected printed circuit boards straight out.
3. Follow the procedure for the appropriate mounting configuration: Section 2.3.1 for panel mounting, Section 2.3.2 for wall mounting, Section 2.3.3 for pipe mounting.
2.3.1 Panel Mounting (Standard). The Model 1054B
is designed to fit into a DIN standard 137.9 mm X
137.9 mm (5.43 in. X 5.43 in.) panel cutout (Refer to
Figures 2-1 and 2-2).
1. Prepare the analyzer as described in Section 2.3.
2. Install the mounting latches as described in Figure 2-2 (latches are shown oversize for clarity). If the latches are not installed exactly as shown, they will not work correctly. The screws provided are self-tapping. Tap the screw the full depth of the mounting latch (refer to side view) leaving a gap greater than the thickness of the cutout panel.
3. Align the latches as shown and insert the analyz­er enclosure through the front of the panel cutout. Tighten the screws for a firm fit. To avoid damag­ing the mounting latches, do not use excessive force.
4. Replace the front panel assembly. Circuit boards must align with the slots on the inside of the enclo­sure. Assure that the continuity wire is connected to the rear cover and the interface board’s closest mounting screws. Replace the door and four front panel screws.
2.3.2 Wall Mounting Plate with Junction Box
(P/N 23054-01). Refer to Figures 2-3 and 2-4.
1. Prepare the analyzer as described in Section 2.3.
2. Mount the junction box and bracket to the analyz­er with the hardware provided. All wiring can be brought to the terminal strip prior to mounting the analyzer.
3. Place the metal stiffener on the inside of the ana­lyzer and mount the two 1/2-inch conduit fittings using two each weather seals as shown. Mount NEMA 4X conduit plug (included) into center con­duit hole.
4. Mount the analyzer to the junction box using the 1/2-inch conduit fittings.
5. Complete wiring from the analyzer to the junction box (Refer to Figure 2-4).
NOTE
Run sensor wiring out of the left opening (From front view) to J-Box. All others out right opening to J-Box.
2.3.3 Pipe Mounting (P/N 23053-00). The 2-inch pipe
mounting bracket includes a metal plate with a cutout for the analyzer (Refer to Section 2.3 for mounting the analyzer into the plate). Mounting details are shown in Figure 2-5.
2.3.4 Wall Mount Enclosure (option -20). See
Figure 2-7 for installation details.
2.4 ELECTRICAL WIRING.
All electrical connections are made to terminal strips on the rear panel (interface board) of the instrument. To access the interface board, remove the four (4) screws securing the rear cover of the enclosure. Gently pull away the rear cover, which is connected to the back panel by a continuity wire. If the continuity wire is disconnected for any reason, it must be recon­nected to the nearest interface board mounting screw before the rear enclosure cover is replaced.
The three openings in the bottom rear of the Model 1054B analyzer housing accommodate 1/2-inch con­duit fittings. Looking at the analyzer from the front, the conduit opening on the left is for sensor wiring, the center opening is for signal output, and the opening on the right is for timer, alarm, and AC connections. Always run sensor wiring in a separate conduit from power wiring.
NOTE
For best EMI/RFI protection, shield the output cable and enclose it in an earth­grounded, rigid metal conduit. Connect the outer cable shield to the earth ground terminal on TB-A when using the wall mounting junction box (Fig. 2-4) or to terminal 8 on TB3 when wiring direct­ly to the instrument (Fig. 2-6).
The sensor cable should also be shield­ed. When wiring directly to the instru­ment, connect the outer shield of the sensor cable to the earth ground of the instrument on terminal 8 of TB2. If the outer shield of the sensor cable is braid, an appropriate metal cable gland fitting may be used to connect the braid to earth ground by way of the instrument case. When wiring to the wall mounting junction box, connect the outer shield of the sensor cable to the earth ground ter­minal on TB-A.
2.4.1 Power Input Wiring. The Model 1054B can be configured for either 115 VAC or 230 VAC power.
Connect AC power to TB1-8 and -9 (115 VAC) or TB1­7 and -8 (230 VAC) ground to the ground terminal at TB3-8 (refer to Figure 2-6).
4
MODEL 1054B DO SECTION 2.0
INSTALLATION
CAUTION
The sensitivity and stability of the analyzer will be impaired if the input wiring is not grounded. DO NOT apply power to the analyzer until all electrical connections are verified and secure. The following precau­tions are a guide using UL 508 as a safe­guard for personnel and property.
1. AC connections and grounding must be in com­pliance with UL 508 and/or local electrical codes.
2. The metal stiffener is required to provide support and proper electrical continuity between conduit fittings.
3. This type 4/4X enclosure requires a conduit hub or equivalent that provides watertight connect, REF UL 508-26.10.
4. Watertight fittings/hubs that comply with the requirements of UL 514B are to be used.
5. Conduit hubs are to be connected to the conduit before the hub is connected to the enclosure, REF UL 508-26.10.
6. If the metal support plate is not used, plastic fit­tings must be used to prevent structural damage to the enclosure. Also, appropriate grounding lug and awg conductor must be used with the plastic fittings.
2.4.2 Output Wiring. The signal output and alarm
connections are made to terminals 1 through 6 of TB1 and TB3-1 and 2. (Refer to Figure 2-6).
5
FIGURE 2-1. Panel Mounting Cutout
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
DWG. NO. REV.
41054B01 B
MODEL 1054B DO SECTION 2.0
INSTALLATION
6
MODEL 1054B DO SECTION 2.0
INSTALLATION
FIGURE 2-2. Panel Mounting Tab Installation
DWG. NO. REV.
41054A26 A
SIDE VIEW
7
MODEL 1054B DO SECTION 2.0
INSTALLATION
FIGURE 2-3. Wall Mounting J-Box Installation
DWG. NO. REV.
41054A27 A
ITEM PART NUMBER DESCRIPTION QUANTITY
1 23058-01 S Assy, J-Box 1 2 33030-00 Bracket, wall mtg 1 3 9900600 Nut, 6-32 Hex 4 4 9910600 Washer, Flat #6 4 5 9910610 Washer, Lock Int. #6 8 6 9600612 Screw, 6-32 X .75 4 7 9510048 Seal, Weathertight 1
8
MODEL 1054B DO SECTION 2.0
INSTALLATION
FIGURE 2-4. Wall Mounting J-Box Wiring
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
DWG. NO. REV.
41054B17 E
9
MODEL 1054B DO SECTION 2.0
INSTALLATION
FIGURE 2-5. Pipe Mounting Installation
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
DWG. NO. REV.
41054B02 C
10
MODEL 1054B DO SECTION 2.0
INSTALLATION
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
DWG. NO. REV.
41054B07 B
FIGURE 2-6. Electrical Wiring
MODEL 1054B DO SECTION 2.0
INSTALLATION
FIGURE 2-7. Wall Mount Enclosure (Option -20)
DWG. NO. REV.
41054B43 A
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
11
Configuration is all accomplished through a series of menus located within the set mode menu. To access these set mode menus the ACCESS keypad is pressed TWICE in RAPID succession.
Once inside the Set mode menu, use the scroll key­pad to scroll through the menu list. When the menu desired is displayed, release the scroll keypad.
To enter the submenus press the SELECT keypad. If the submenu allows editing, the item will flash that can be edited. If not, use the scroll keypad to scroll through the next list of submenus . SELECT will enter this sub­menu and if it is editable, the field will flash.
To exit the menu and SAVE the new value, press the ENTER keypad.
To exit the menu without saving the edited value, press the PV keypad to jump out of the set menu program with out saving value. To change other parameters will require re-entering the set menu program.
Figure 3-4 explains the various fields surrounding the Primary process on the LC display.
Table 3-1 describes the functions accessible with the 8 keypads, the number of times to press the keypad to access, and its’ function when used with the select keypad and set menu.
Tables 3-2 and 3-3 describe the meaning of the various mnemonics used on the display. They are categorized by their use in either menus, or as process information.
MODEL 1054B DO SECTION 3.0
DESCRIPTION OF CONTROLS
SECTION 3.0
DESCRIPTION OF CONTROLS
3.1 KEYBOARD FUNCTIONS. All operations of the
Model 1054B Microprocessor Analyzer are controlled by the eight (8) keypads on the front of the instrument. These keypads are used to :
1. Display parameters other than the primary param­eter.
2. Edit setpoints for alarms, set up specific output current value for simulation, calibrate temperature, conductivity, etc.
3. Configure display for temperature units, for auto­matic temperature compensation, alarm usage, setting timer functions, security, and output range.
To view, and not change parameters, other than the primary parameter requires only a simple keystroke routine. As shown in Figure 3-1, a single keypress accesses the lower function printed on the keypad. Quick, double keypresses access the top function printed on the keypad.
To edit any of these parameters, requires one more operation. After displaying the value associated with the parameter selected, press the SELECT keypad. As seen in Figure 3-2, this will display the numerical value, and the first digit will be flashing to indicate this value may be edited.
All changes to the operating program that set-up the instrument display are made through the set menu pro­gram. See Figure 4.1.
FIGURE 3-1. Function Select on Keypad.
Single press of the keypad will access the present Dissolved Oxygen reading. Read only.
1. Press Key.
2. AdJ shows briefly.
3. Numbers show with digit flashing.
1 .Press Key twice.
2. Lo shows briefly.
3. Zero point is displayed.
SELECT
ZERO
ALARM
1
OUTPUT
PV
FIGURE 3-2. Accessing Editing Function.
1. Press twice in rapid succession.
2. See SEt on display. Confirms entry into set mode menu.
3. First menu item is displayed. Analyzer now ready to configure.
4. Use the SCROLL keypad to rotate through the available menus.
ACCESS
ENTER
SEt
Cin
é
FIGURE 3-3. Accessing Configuration Menus
Quick double press will access the current output current value in mA or %. Read only.
ç
12
13
MODEL 1054B DO SECTION 3.0
DESCRIPTION OF CONTROLS
FIGURE 3-4. LCD Display
RELAY 1 ACTIVATED
RELAY 2 ACTIVATED
DISSOLVED OXYGEN : PPM­FLAG ON; % SATURATION FLAG BLINKING
% VALUE ­FLAG ON; mA ­FLAG BLINKING
HOLD STATUS FLAG ON; FAULT ­FLAG BLINKING
UPPER FUNC­TION PRESS TWICE QUICKLY
LOWER FUNC­TION PRESS ONCE
3.1.1 Item Selection and Value Adjustment Keys.
The three keys located on the lower right side of the keypad are used for menu navigation, value adjust­ment and entry, and item selection. These keys per­form the following functions:
A. SELECT/Shift (
çç
) Key. This key is used to select the displayed menu, or for shifting to the next digit in the Numeric Display.
B. SCROLL Key (é). This key is used to scroll through menu when selected, or scroll through digits on the active (flash­ing) Numeric Display, or move the deci­mal point and ppm/% SAT display.
Holding key down auto scrolls display.
C. ACCESS/ENTER Key. This key is used to ACCESS the Set Mode (Section
4.1.2) and to ENTER the displayed value into memory (from Numeric Display).
ç
SELECT
é
ACCESS
ENTER
DWG. NO. REV.
41054B27 A
14
MODEL 1054B DO SECTION 3.0
DESCRIPTION OF CONTROLS
TABLE 3-1. Key Description
Displays - current output (mA or % full scale).
Set Function (w/SELECT) - Simulates current output.
Displays – low Lo setpoint; the low end of the dis­solved oxygen range that corresponds to 0 or 4 mA DC output.
Displays – high HI setpoint; the high end of the dissolved oxygen range that corresponds to 20 mA DC output.
Select sub menu (mnemonic display). Shift to next digit (numeric display). Activate decimal point adjustment.
Scroll through menu (mnemonic display). Scroll digits (numeric display). Holding key down autoscrolls digits or set menu items.
Press twice quickly to access set-up menu. Enter displayed value into memory. Enter displayed menu item (flashing) into memory.
Displays - Barometric pressure setting. Used to calibrate the analyzer and the dissolved oxygen sensor loop in air. Set Function (w/select)-One point stan­dardization of % saturation.
Displays - Alarm 1 setpoint.
Set Function (w/SELECT) - Sets Alarm 1 Setpoint.
Displays - DO in ppm or % saturation.
Set Function (w/SELECT) - One point standardization of DO.
(PV = Process Variable)
Initiates or removes analyzer from hold condition.
Displays - process temperature (°C or °F).
Set Function (w/SELECT) - One point standardization of temperature.
Displays - Alarm 2 setpoint.
Set Function (w/SELECT) - Sets Alarm 2 Setpoint.
HOLD
TEMP
OUTPUT
PV
ZERO
ALARM 1
F.S.
ALARM 2
CAL
ç
SELECT
é
ACCESS
ENTER
CAUTION
Air calibrate only when the sensor is fully polarized and stabilized in ambient air.
SECOND FUNCTION (PRESS TWICE QUICKLY)MAIN FUNCTION (PRESS ONCE)
NOTE
When no key is pressed for a period of 60 seconds the analyzer will default to reading DO.
CAUTION
The HOLD function and the CAL function are not read functions. Refer to Sections 4.16 and 5.0 respectively.
15
MODEL 1054B DO SECTION 3.0
DESCRIPTION OF CONTROLS
TABLE 3-2. Information Mnemonics
MNEMONIC DESCRIPTION
AdJ Adjustment to value reading bAd Incorrect entry
HI Displays high range value for current output
itr Interval timer activated
Lo Displays low range value for current output LOC Access locked - enter security code SEt Set mode SiC Simulates current output (mA) SiP Simulates current output (percent) SP1 Displays Alarm 1 setpoint SP2 Displays Alarm 2 setpoint Std Standardize DO
AL1 Alarm 1 setup AL2 Alarm 2 setup Atc Automatic temp. comp. bAr Bars
br Barometric reset bP Barometric pressure
o
C Temperature °C COd Security Code cnt Count on times CUr Config. mA output display Cur Config. fault output cur Default current setpoint dAY Days dFt Fault Configuration
do Dissolved oxygen d-O Display output d-t Display temperature doc Display output in mA doF Delay off time don Delay on time dPn Dampen output dtS Display test dur Waiting period duration
o
F Temperature °F Fct Solubility factor FLt Fault alarm set
Hi Relay action - high
H-L Alarm logic
hr Hours HYS Hysteresis iHg Inches mercury
in Display sensor input int Interval period Int Timer setup
Lo Relay action - low non No action on fault OFF Alarm off oFF Function off ont On time duration
On Alarm on
on Function on OFt Off time duration OUt Current output Pct Display output in percent PPu Parts per million
rng Range units rL1 Relay 1 fault setup rL2 Relay 2 fault setup SAt % saturation SEC Seconds SHO Show fault history SIt Saline solution factor Snr Sensor type SoL Solubility Correction t-C Temperature config. tiL Display time remaining tOn Timer status UEr Display version number uHg Millimeters mercury uin Minutes 420 4mA to 20mA output 020 0mA to 20mA output
-0- Analyzer zero
1 499 or 491 Sensor
Compatible
2 492 or 493 Sensor
Compatible
TABLE 3-3. Set Function Mnemonics
NOTE: See Table 8-1 for Fault Mnemonics.
16
MODEL 1054B DO SECTION 4.0
CONFIGURATION
SECTION 4.0
CONFIGURATION
4.1 GENERAL. This section details all of the items
available in the Set Mode and the setpoint adjustment procedures to configure the Model 1054B Dissolved Oxygen Analyzer. Refer to Table 3-3 and Figure 4-1.
4.1.1 Set Mode (SEt). Most of the analyzer’s configu- ration is done while in the Set Mode. Please refer to Figure 4-1 for the layout of all menu items. All menu variables are written to the analyzer’s EEPROM (memory) when selected and remain there until changed. As these variables remain in memory even after the analyzer’s power is removed, the analyzer configuration may be performed prior to installing it.
1. Make sure the analyzer loop is properly wired. Power up the analyzer. Only power input wiring is required for analyzer configuration. (Refer to Section 5.2 regarding polarization voltage.) The analyzer’s display will begin showing values and/or fault mnemonics. All fault mnemonics will be suppressed while the analyzer is in Set Mode (the fault flag will continue to blink).
2. Enter Set Mode. Pressing the ACCESS key twice will place the analyzer in Set Mode. The display will show SEt to confirm that it is in Set Mode. It will then display the first item in the set menu br. The analyzer is now ready for user configuration.
NOTE:
If LOC displays, the Keyboard Security Code must be entered to access the Set Mode. (Refer to Section 6.0.) To get out of the Set Mode, press the PV key. Refer to the Configuration Worksheet on page 18 for the analyzer ranges and factory set­tings.
4.1.2 Configuration Work Sheet. The Configuration Worksheet provides the range of the various func­tions, the factory settings, and a column for user’s set­tings. As you proceed through the configuration procedures for each function of the analyzer, fill in the appropriate information in the USER column. The con­figuration may be done in any order. However, it is recommended that it be done in the order as shown in the worksheet.
4.1.3 Barometric Pressure. Display Mnemonic br. This function is used to update the barometric pres­sure setting after the analyzer/sensor loop has been air calibrated and the sensor installed in the process. Refer to Section 5.0, Start-up Calibration. It is only used for the % SAT mode.
4.1.4 Analyzer Zero. Display Mnemonic -O-. This function is used to zero the analyzer/sensor loop. Refer to Section 5.3.2, Analyzer Zero.
4.1.5 Sensor Input. Display Mnemonic in. This func- tion is used to display current input from the sensor. Refer to Section 8.2.4, Sensor Troubleshooting, for more information.
17
MODEL 1054B DO SECTION 4.0
CONFIGURATION
H-L
HYS
don
doF
On
OFF
br
-0­in
AL1
AL2
Int
t-C
Out
rng
bP
SoL
Snr dFt
UEr
dtS
COd
Hi
Lo
tOn
int
cnt
ont
OFt
dur
tiL
OFF
on
SEC
uin
hr
dAY
o
F
o
C
doc
Pct
PPu
SAt
Fct
SIt
1
2
on
oFF
non
uHg
iHg
bAr
non
cur
rL1
rL2
Cur
SHO
on
oFF
On
FLt
OFF
420
020
d-t
Atc
dPn
CUr
d-O
FIGURE 4-1. Set Function Menu
SEt
18
RANGE FACTORY SET USER SET
A. Alarm 1 Setup (AL1)
1. Alarm Status (On/OFF) On _________
2. High or Low (H-L) (Hi/Lo) Lo _________
3. Hysteresis (HYS) 0-2.00 0.00 _________
4. Delay Time On (don) 0-255 sec. 000 Seconds _________
5. Delay Time Off (doF) 0-255 sec. 000 Seconds _________
B. Alarm 2 Setup (AL2)
1. Alarm Status (On/FLt/OFF) On _________
2. High or Low (H-L) (Hi/Lo) Hi _________
3. Hysteresis (HYS) 0-2.00 0.00 _________
4. Delay Time On (don) 0-255 sec 000 Seconds _________
5. Delay Time Off (doF) 0-255 sec 000 Seconds _________
C. Interval Timer (int)
1. Active Status (tOn) (OFF/on) OFF _________
2. Interval Time (int) 10 min to 2,999 days 1 Day _________
3. Count (cnt) 1 to 60 5 _________
4. On Time (ont) 0.1 to 299.9 sec 1 Second _________
5. Off Time (OFt) 0.1 to 299.9 sec 1 Second _________
6. Duration (dur) 0.1 to 299.9 sec 2 Seconds _________
D. Temperature Setup (t-C)
1. Display Temperature (d-t) (oC/oF)
o
C _________
2. Automatic TC (Atc) (on/oFF) on _________ a. Manual Temp. Value 0°C to 80°C _________
E. Current Output Setup (OUt)
1. mA Output (CUr) (020/420) 420 _________
2. Display Current Output (d-O) (Pct/doc) doc _________
3. Dampen Current Output (dPn) 0-255 sec. 0.0 Seconds _________
F. Output Range Setup (rng)
1. Parts Per Million (PPu) 0-20 PPu _________
2. Percent Saturation (SAt) 0-250% _________
G. Barometric Pressure Setup (bP)
1. Millimeters Mercury (uHg) 500-1000 uHg _________
2. Inches Mercury (iHg) 19.67-39.37 _________
3. Bars (bAr) 0.666-1.333 _________
H. Solubility Correction Factor Setup (SoL)
1. Solubility Constant (Fct) 0.09 - 9.00 1.00 _________
2. Solubility Factor For Saline Solutions (S1t) 0.00 - 2.00 0.00 _________
I. Sensor compatibility (Snr)
1. Model 499 or 491(1) / Model 492 or 493 (2) 1 _________
J. Default Setup (dFt)
1. Relay 1 Default (rL1) (non/oFF/on) non _________
2. Relay 2 Default (rL2) (non/oFF/on) non _________
3. Current Output Default (Cur) (non/cur) non _________
4. Current Output Held non _________
K. Keyboard Security Setup (COd)
1. Keyboard Security Required 001-999 _ _________
2. Keyboard Security Not Required 000 000 _________
Alarm Setpoints Setup
1. Alarm 1 (SP1) 0-20 ppm 0 ppm _________
2. Alarm 2 (SP2) 0-20 ppm 20 ppm _________
Current Output Setup
1. Zero (0 or 4 mA) 0-20 ppm 0 ppm _________
2. F.S. (20 mA) 0-20 ppm 20 ppm _________
TABLE 4-1. Configuration Worksheet
Use this worksheet to assist in the configuration of the analyzer.
MODEL 1054B DO SECTION 4.0
CONFIGURATION
19
MODEL 1054B DO SECTION 4.0
CONFIGURATION
4.2 ALARM 1 AND 2. Display Mnemonic AL1 or AL2.
Used to set alarm relay logic (See note below). Each alarm is configured separately. Choices are (see note below):
A. On. Display Mnemonic On. Select this item if Alarm 1 or 2 is to be used. If On is selected, AL1 or AL2 may be configured for either high (Hi) or low (Lo) alarm.
B. Off. Display Mnemonic 0FF. Select this item if Alarm 1 or 2 will not be used or to temporarily disable the alarm. Alarm 1 or 2 setpoint will display oFF if this item is select­ed. All other Alarm 1 or 2 settings are ignored.
C. Fault. (In addition to On and 0FF, Alarm 2 may be con- figured to a fault alarm). Display Mnemonic FLt. Select this item to make Alarm 2 a fault alarm. Relay 2 will energize when the analyzer experiences a fault condi­tion. Alarm 2 setpoint will display FLt if this item is select­ed. All other Alarm 2 settings are ignored.
D. Alarm Logic. Display Mnemonic H-L. Select this item for high or low alarm logic. High alarm logic acti­vates the alarm when the reading is greater than the setpoint value. Low alarm logic activates the alarm when the reading is less than the setpoint value.
E. Relay Hysteresis. Display Mnemonic HYS. Sets the relay hysteresis (dead band) for deactivation after reading has passed the alarm setpoint. May be set from 0 to 2.00. Hysteresis is used to delay the alarm relay deactivation on the low side of the high alarm setpoint, or on the high side of the low alarm setpoint. This feature is used to pre­vent or minimize alarm chattering.
Example:
High alarm setpoint is 6.00 ppm, and hysteresis is
1.00 ppm. The alarm will activate when the DO reading exceeds 6.00 ppm and will remain activat­ed until the reading drops to below 5.00 ppm.
F. Delay Time On. Display Mnemonic don. Sets time delay for relay activation after alarm setpoint is reached. May be set from 0 to 255 seconds. Normal (no alarm) state restarts time from zero. Use when a fixed time should pass before relay activation occurs.
G. Delay Time Off. Display Mnemonic doF. Sets time delay for relay deactivation after alarm setpoint is reached. May be set from 0 to 255 seconds. Alarm state restarts time from zero. Use when a fixed time should pass before relay deactivation occurs.
4.2.1 Alarm Setup (AL1/AL2). Refer to Figure 4-1 and Table 4-1.
1. Enter Set Mode by pressing the ACCESS key twice.
2.
SCROLL (é) until AL1 or AL2 appears on the display.
3. SELECT to move to the next menu level. On, OFF or FLt (AL2 only) will display.
4. SCROLL (é ) to display desired item then SELECT.
5. If OFF is selected, display will show oFF to acknowledge. Press the ENTER key to return to AL1 or AL2, concluding routine.
If On is selected, display will show on to acknowl­edge, then display H-L. Proceed to Step 6.
If FLt is selected (AL2 only), display will show FLt to acknowledge. Press the ENTER key to return to AL2.
6. SELECT On. Hi or Lo will display (flashing).
7. SCROLL (é) to the desired item and ENTER it into memory. Display will return to H-L. If changes to relay activation logic are desired, proceed to Step 8, otherwise Step 12.
8. SCROLL (é) to display HYS, don or doF then SELECT. Numerical display will flash to indicate that a value is required.
9. Use the SCROLL (é) and SELECT keys to dis­play the desired value.
10. ENTER the value into memory. Analyzer will acknowledge and return to display of last item selected. Repeat Step 8 if further changes are desired, otherwise Step 12.
11. Repeat Step 3 for the other Alarm’s settings as required.
12. Press the ENTER key to return to the first level of the Set Menu, or press the PV key to get out of the Set Mode.
20
MODEL 1054B DO SECTION 4.0
CONFIGURATION
4.3 INTERVAL TIMER. Display Mnemonic Int. This
item is used to set the interval timer’s relay logic. The timer can be used as a sensor maintenance reminder. Choices are:
A. Interval Timer Enable/Disable. Display Mnemonic tOn. Select this item to begin interval cycle on or dis­able interval cycle OFF.
B. Interval Period. Display Mnemonic int. Select this item to set the time period between end of wait dura­tion and beginning of new on-off cycle. SEC for sec­onds, uin for minutes, hr for hours, and dAY for days. May be set from 1 second to 2999 days. Time of less
than 10 minutes is not recommended.
C. On Periods Per Cycle. Display Mnemonic cnt.
Select this item to enter the number of on periods per cycle. May be set from 1 to 60 on periods.
D. Duration of On Periods. Display Mnemonic ont. Select this item to enter the relay activation time for each on period. May be set from 0.1 to 299.9 seconds.
E. Duration of Off Periods. Display Mnemonic OFt. Select this item to enter the relay
deactivation time between each on period during the control cycle. Valid when cnt is 2 or greater. May be set from 0.1 second to 299.9 seconds.
F. Wait Duration. Display Mnemonic dur. Select this option to enter the wait duration after the last on peri­od in a cycle. May be set from 0.1 to 299.0 seconds. The wait duration can be used for sensor recovery after a cycle to allow the process to stabilize before the next interval time starts again.
NOTE:
The Model 1054B DO is placed on hold during the control cycle (from first on peri­od through the wait duration). The analyz­er will simulate a fault condition and briefly show itr every eight seconds. The dis­play will continue to show the measured value.
G. Interval Time Remaining. Display Mnemonic tiL. Select this item to display the time remaining to the next control cycle. If selected during the control cycle, display will show ---.
4.3.1 Interval Timer Set Up (Int). Refer to Figure 4-1
1. Enter Set Mode by pressing the ACCESS key twice.
2. Press the SCROLL (é) key until Int appears on the display.
3. Press the SELECT key to move to the next menu level. tOn will display. Press the SELECT key again.
4. Press the SCROLL (é) key to display on (if the interval timer is to be used) or OFF (if the interval timer is not to be used) and press the ENTER key. If interval timer configuration is required, pro­ceed to Step 5, otherwise Step 15.
5. Press the SCROLL (é) key to display the next menu item, int. Press the SELECT key.
6. Press the SCROLL (é ) key to display desired duration and SELECT it.
7. Press the SCROLL (é) key and SELECT key to display the desired value and press the ENTER key.
8. Repeat Steps 6 and 7 as needed.
9. Press the ENTER key to return to the interval period int menu.
10. SCROLL (é) down to cnt (on periods per cycle) Press the SELECT key and the current setting will flash. Repeat Step 7.
11. SCROLL (é) down to ont (duration of on times). Press the SELECT key and the current setting will flash. Repeat Step 7.
12. SCROLL (é) down to OFt (duration of off times). Press the SELECT key and the current setting will flash. Repeat Step 7.
13. SCROLL (é) down to dur (waiting time after the last on cycle). Press the SELECT key and the current setting will flash. Repeat Step 7.
14. SCROLL (é) down to tiL (time interval lapse). Press the SELECT key to display the time remaining. Press the ENTER key to return to tiL.
15. Press the ENTER key to return to the first level of the Set Menu, or press the PV key to get out of the Set Mode.
21
MODEL 1054B DO SECTION 4.0
CONFIGURATION
4.4 TEMPERATURE CONFIGURATION. Display
Mnemonic t-C. Select this item for temperature read­ing and compensation choices.
A. Temperature Display. Display Mnemonic d-t. Select this item to toggle between °F and °C tempera­ture display. The analyzer will show all temperatures in the units selected.
B. Automatic Temperature Compensation. Display Mnemonic Atc. The analyzer will use the temperature input from the sensor for temperature compensation when on is selected. When oFF is selected, the analyz­er will use the value entered by the user for manual temperature compensation. This manual temperature option is useful if the temperature sensor is faulty or not on line. Temperature specific faults will be disabled (Refer to Table 8-1).
4.4.1 Temperature Setup (t-C). Refer to Figure 4-1.
1. Enter Set Mode by pressing the ACCESS key twice.
2. Press the SCROLL (é) key until t-C appears on the display.
3. Press the SELECT key to move to the next menu level. d-t will display.
4. Press the SELECT key. oC or oF will be displayed flashing. SCROLL (é) to desired unit.
5. Press the ENTER key when the desired tempera­ture unit is displayed. The analyzer will now dis­play temperature readings in this unit until changed.
6. SCROLL (é) to Atc then press the SELECT key. on or oFF will be displayed flashing. SCROLL (é) to desired condition.
7. Press the ENTER key when the desired condition is displayed.
8. If on was entered, proceed to Step 10. If oFF was entered, the last temperature used for manual compensation will be displayed with the right digit flashing.
9. Use SCROLL (é) and SHIFT (ç) to display the desired value. ENTER value into memory.
10. Press the ENTER key to return to the first level of the Set Menu, or press the PV key to get out of the Set Mode.
4.5 CURRENT OUTPUT. Display Mnemonic is OUt. This item is used to select signal output configuration.
A. Output Dampening. Display Mnemonic dPn. This function is used to filter out and spread out any change in signal output. The number entered is the sampling time (in seconds). Zero to 255 seconds may be entered. If less than 1 second is entered, the signal output change takes place immediately. If 1 to 255 seconds is entered, 63% of the signal output change takes place in the first sampling time, then 63% of the balance of the signal output change takes place in the next sampling time, etc.
B. mA Output Range. Display Mnemonic CUr. Selection of this item will allow choice of either 0 to 20 mA or 4 to 20 mA output range.
C. Display Output. Display Mnemonic d-O. This item is used to select output display logic. Selecting this item will allow the analyzer to display current output in mA (when doc is entered) or in percent of full scale output range (when Pct is entered). When the OUT- PUT key is pressed, a steady flag indicates that the value displayed is percent output. A flashing flag indi­cates mA output.
4.5.1 Output Setup (OUt). Refer to Figure 4-1.
1. Enter Set Mode by pressing the ACCESS key twice.
2. Press the SCROLL (é) key until OUt appears on the display.
3. Press the SELECT key to move to the next menu level. dPn will display.
4. Press the SCROLL ( é ) key then SELECT desired item.
5. If dPn is selected, numerical display will flash indi­cating that a value is required. Proceed to Step 6.
If CUr or d-O is selected, proceed to Step 7.
6. Use SCROLL (é) and SELECT keys to display the desired value. ENTER into memory. Proceed to Step 11.
7. If CUr is selected, 420 or 020 is displayed flashing.
22
MODEL 1054B DO SECTION 4.0
CONFIGURATION
8. SCROLL (é) to the desired mA choice and press ENTER. Proceed to Step 11.
9. If d-O is selected, Pct or doc is displayed flashing.
10. SCROLL (é) to the desired output unit and press ENTER.
11. Press the ENTER key to return to the first level of the Set Menu, or press the PV key to get out of the Set Mode.
4.6 DISSOLVED OXYGEN RANGE UNITS. Display
Mnemonic rng. This function is used to select the desired dissolved oxygen range units. One of the fol­lowing two range units may be chosen:
A. (ppm) Display Mnemonic (PPu). Selecting this item is for parts per million by weight of dissolved oxy­gen concentration.
B. Percent (%) Saturation Display Mnemonic (SAt). Percent (%) saturation is the percent of dissolved oxy­gen in solution compared to the maximum amount of dissolved oxygen the solution can hold at a given tem­perature and partial pressure of oxygen.
NOTE
Dissolved oxygen values will be displayed in the unit selected until changed.
4.6.1 Dissolved Oxygen Analyzer Set Up (rng).
Refer to Figure 4-1.
1. Enter Set Mode by pressing the ACCESS key twice.
2. Press the SCROLL (é) key until rng appears on the display.
3. Press the SELECT key. PPu or SAt will be dis- played.
4. Press the SCROLL (é ) key to display the desired range unit.
5. Press the ENTER key to enter the desired unit into memory.
6. Press the ENTER key to return to the first level of the Set Menu, or press the PV key to get out of the Set Mode.
4.7 BAROMETRIC PRESSURE UNITS. Display Mnemonic bP. This function is used to select the baro­metric pressure units needed by the microprocessor during the air calibration step. One of the following three range units may be chosen:
A. bAr if the barometric pressure is given in bars.
B. iHg if the barometric pressure is given in inches of
mercury.
C. uHg if the barometric pressure is given in millimeters of mercury.
4.7.1 Barometric Pressure Unit Set Up (bP). Refer to Figure 4-1.
1. Enter Set Mode by pressing the ACCESS key twice.
2. Press the SCROLL (é) key until bP appears on the display.
3. Press the SELECT key. uHg, iHg, or bAr will be displayed flashing.
4. Press the SCROLL (é) key to display the desired barometric pressure unit.
5. Press the ENTER key to enter the desired unit into memory.
6. Press the ENTER key to return to the first level of the Set Menu, or press the PV key to get out of the Set Mode.
23
MODEL 1054B DO SECTION 4.0
CONFIGURATION
4.8 SOLUBILITY CORRECTION FACTOR. Display
Mnemonic SOL. This function is used to correct for the solubility of oxygen in the process liquid which is other than fresh water or is unique to the customer. One of the following two factors may be chosen:
A. Fct – Solubility constant of the process liquid. For fresh water, enter a value of 1.00. (The solubility
constant Fct range is 0.09 to 9.00.)
B. SIt – The salinity of the liquid measured in parts per thousand by weight of salt. For fresh water, enter a value of 0.00. (The typical parts per thousand salt by weight range is 0.00 to 2.00.)
NOTE
10 ppt of salt is equivalent to 1.00 percent salt.
4.8.1 Solubility Correction Factor Set Up (SoL). Refer to Figure 4-1.
1. Enter Set Mode by pressing the ACCESS key twice.
2. Press the SCROLL (é) key until SoL appears on the display.
3. Press the SELECT key. Fct or S1t will be dis­played. Press the SCROLL (é) key to display the
desired choice.
4. Press SELECT key. The last choice used will be displayed flashing.
5. Use the SCROLL (é) and SELECT keys to dis- play the desired value. Press the ENTER key to enter the value into memory.
6. Press the ENTER key to return to the first level of the Set Menu, or press the PV key to get out of the Set Mode.
NOTE
Dissolved oxygen values will now be cor­rected according to the current factor value entered until new values are entered.
4.9 SENSOR COMPATIBILITY. Display mnemonic
Snr. This item is used to select the proper sensor com­patibility. There are to menu choices.
A. 1 - For use with the Model 499DO or Model 491.
B. 2 - For use with the Model 492 or Model 493.
The correct selection must be made to match the Model 1054B DO to the output characteristics of the sensor to be used.
4.9.1 Sensor Compatibility Set Up (Snr). Refer to Figure 4-1.
1. Enter Set Mode by pressing the ACCESS key twice.
2. Press the SCROLL (é) key until Snr appears on the display.
3. Press the SELECT key. 1 or 2 will be displayed. Press the SCROLL (é) key to display the appro-
priate choice.
4. Press the ENTER key to enter the value into mem­ory.
5. Press the ENTER key to return to the first level of the set menu, or press the PV key to return to nor­mal operation.
4.10 DEFAULTS. Display Mnemonic dFt. This func-
tion is used to set the configuration of relay and output conditions during a FAULT or HOLD status. A flashing flag beside the HOLD/FAULT label indicates a fault condition.
A. Relays 1 and 2. Display Mnemonic rL1 and rL2. The relays can be set to activate on, deactivate oFF, or hold present status non.
When on is chosen, the relay will be activated at a
FAULT or HOLD condition.
When oFF is chosen, the relay will be deactivated at a
FAULT or HOLD condition.
When non is chosen, the relay will maintain its status
during a FAULT or HOLD condition.
B. Current Output. Display Mnemonic Cur. This item is used to configure the analyzer’s output during a FAULT or a HOLD condition. There are two selections: non and cur. When non is chosen, the present output is frozen at a FAULT or HOLD condition. When cur is chosen, the analyzer uses the milliamp value the user has entered to be the output during a FAULT or HOLD condition.
C. Fault History. Display Mnemonic SHO. Selecting this item will allow the user to view all the faults of the two most recent fault conditions that have occurred since the last reset viewing.
24
MODEL 1054B DO SECTION 4.0
CONFIGURATION
4.10.1 Default Setup (dFt). Refer to Figure 4-1.
1. Enter Set Mode by pressing the ACCESS key twice.
2. Press the SCROLL (é) key until dFt appears on the display.
3. Press the SELECT key to move to the next menu level. rL1 will display.
4. Press the SCROLL (é ) key to display desired item then press the SELECT key.
5. When rL1 or rL2 is selected, non, OFF, or on is dis­played flashing. Press the SCROLL (é) key to display the desired condition and press the ENTER key to enter the selection into memory.
6. When Cur is selected, non or cur is displayed flashing. Press the SCROLL (é) key to display the desired condition and press the ENTER key to enter the selection into memory. When cur is entered, the mA value in memory is displayed with the right digit flashing. Use SCROLL (é) and SELECT to display the desired value and press the ENTER key to enter this value into memory.
7. When SHO is selected, press the SELECT key to view the fault history. Pressing the ENTER key will erase the list from memory. A new list is start­ed as new faults occur.
8. Press the ENTER key to return to the first level of the Set Menu, or press the PV key to get out of the Set Mode.
4.11 SOFTWARE VERSION NUMBER. Display Mnemonic UEr. This function displays the software version number used in the particular analyzer being used. This information may be very important in serv­icing the analyzer.
4.11.1 To Display Software Version Number.
1. Enter Set Mode by pressing the ACCESS key twice.
2. Press the SCROLL (é) key until UEr appears on the display.
3. Press the SELECT key and the software version number will be displayed.
4. Press the ENTER key to return to the first level of the Set Menu, or press the PV key to get out of the Set Mode.
4.12 DISPLAY TEST. Display Mnemonic dtS. This
function allows the user to visually test the LCD display segments. If the display is functioning properly, all the LCD segments are activated.
1. Enter Set Mode by pressing the ACCESS key twice.
2. Press the SCROLL (é) key until dtS appears on the display.
3. Press the SELECT key and all the LCD segments will be displayed for about 5 seconds.
4. Press the ENTER key to return to the first level of the Set Menu, or press the PV key to get out of the Set Mode.
25
MODEL 1054B DO SECTION 4.0
CONFIGURATION
4.13 ALARM SETPOINT. The alarm setpoints should
be adjusted after completing the configuration proce­dure as outlined in Sections 4.2 to 4.10.
1. Press the PV key to ensure that the analyzer is not in Set Mode.
2. Press the ALARM 1 or ALARM 2 key. SP1 or SP2 will show briefly, followed by the Alarm 1 or Alarm 2 setpoint currently in memory.
NOTE
If the alarm is set to OFF or FAULT (AL2 only), the analyzer will display OFF or FLtrespectively. No setpoint value can be entered unless on has been selected in the set menu (Refer to Section 4.2, Alarm Setup).
3. Press the SELECT key to adjust the value. The display will acknowledge briefly with AdJ followed by the numeric display with right digit flashing.
4. SCROLL (é) and SELECT to display the desired setpoint.
5. Press ENTER. SP1 or SP2 will show briefly, then the desired setpoint is displayed.
NOTE
Alarm logic may be changed from normally open (N.O.) to normally closed (N.C.) by cutting circuits (W5, W7, W9) on the power supply PCB and adding jumpers (W4, W6, W8).
ZERO
ALARM
1
F.S.
ALARM
2
ACCESS
ç
SELECT
é
AdJ
SP1/2
ç
SELECT
ENTER
Press
Once
Press
Once
Displays
Briefly
Displays
Briefly
Numeric
Display
Change to
desired
value
Press
Once
Numeric
Display
of
Setpoint
FIGURE 4-2. Alarm Setpoint
26
MODEL 1054B DO SECTION 4.0
CONFIGURATION
4.14 OUTPUT SCALE EXPANSION. This section
should be followed only if the current output needs to be scaled to an operating range other than the factory setting of 0-20 ppm dissolved oxygen. The output ZERO and FULL SCALE value should be adjusted after completing the configuration procedure as out­lined in Sections 4.2 to 4.10.
A. Zero Setpoint. This is the low dissolved oxygen value that the user wants to correspond to 0 or 4 mA DC output. To change the setpoint, perform the follow­ing steps:
1. Press the PV key to ensure that the analyzer is not in Set Mode.
2. Press the ZERO key. LO will show briefly, followed by the ZERO DO value in memory.
3. Press the SELECT key to adjust the value. The display will acknowledge briefly with AdJ followed by the numeric display with right digit flashing.
4. SCROLL (é) and SELECT to display the desired setpoint.
5. Press ENTER. LO will show briefly, then the desired DO valve is entered into memory and dis­played.
B. Full Scale Setpoint. This is the high dissolved oxygen value that the user wants to correspond to 20 mA DC output. To change the setpoint, perform the following steps:
1. Press the PV key to ensure that the analyzer is not in Set Mode.
2. Press the F.S. key. HI will show briefly, followed by the FULL SCALE DO value.
3. Press the SELECT key to adjust the value. The display will acknowledge briefly with AdJ followed by the numeric display with right digit flashing.
4. SCROLL (é) and SELECT to display the desired setpoint.
5. Press ENTER. HI will show briefly, then the desired DO value is entered into memory and dis­played.
NOTE
For reverse output, enter the higher value for zero, and the lower value for the F.S.
ZERO
ALARM
1
F.S.
ALARM
2
ACCESS
ç
SELECT
é
AdJ
LO/HI
ç
SELECT
ENTER
Press Twice
Press
Once
Displays
Briefly
Displays
Briefly
Numeric
Display
Change to
desired
value
Press
Once
Numeric
Display
of
DO value
FIGURE 4-3. Output Scale Expansion
27
MODEL 1054B DO SECTION 4.0
CONFIGURATION
4.15 OUTPUT DISPLAY/OUTPUT SIMULATION.
The output may be displayed by pressing the OUTPUT key. The output is displayed in either milliamp DC or in percent, depending on the output unit selected in the OUt/d-O menu. Refer to Section 4.5.1.
Output simulation allows the user to simulate the cur­rent output to test the operation of the devices con­nected to the output terminals TB3-1 and TB3-2.
Perform the following steps to display output and to simulate output:
1. Press the OUTPUT key to display the output. dOC or Pct will show briefly; then the output is displayed either in milliamp DC or in percent output.
CAUTION
Pressing the SELECT key to enter the sim­ulation mode will make the analyzer imme­diately transmit an output equivalent to the simulation value in memory.
2. Press the SELECT key once. SiC (for current output simulation) or SiP (for percent output sim­ulation) will show briefly, then the simulation value in memory is displayed with right digit flashing. The analyzer is put on HOLD at this time.
3. SCROLL (é) and SELECT to display the desired simulation value.
4. Press ENTER. dOC or Pct will show briefly, then the desired simulation value is displayed flashing and takes effect until the analyzer is taken out of HOLD status. Press HOLD twice to remove the analyzer from HOLD status.
4.16 HOLD. Press HOLD twice to put the analyzer on HOLD, or to get the analyzer out of HOLD. The ana­lyzer is on HOLD when the HOLD/FAULT flag is steady.
Putting the analyzer on HOLD will result in the follow-
ing:
A. If non was chosen in the dFt/Cur menu, the present process current output is frozen until the analyzer is out of HOLD status.
B. If cur was chosen in the dFt/Cur menu, the present current output will switch (default) to the value entered for current output default.
During a HOLD status, the analyzer may be calibrat­ed or configured. The new values will take effect only after the analyzer is taken out of the HOLD condition.
28
MODEL 1054B DO SECTION 5.0
START-UP AND CALIBRATION
SECTION 5.0
START-UP AND CALIBRATION
5.1 GENERAL. This section provides the start-up and
calibration procedures for the Model 1054B Dissolved Oxygen Analyzer with its compatible Rosemount Analytical Models 491, 492, 493 and 499 DO Sensors. The start-up and calibration procedures must be per­formed only after the installation (Section 2.0) and Configuration (Section 4.0) sections have been properly carried out .
5.2 START-UP. The start-up procedure for the1054B DO involves the configuration of the analyzer to your particular process requirements and logging the vari­ous setpoints in the user column of the Configuration Worksheet. Also involved is the complete polarization of the DO sensor.
When the analyzer is powered up, a polarizing voltage is applied between the sensor's anode and cathode. The sensor (electrode) current is initially very high, then quickly falls off and settles to a steady state after a few hours.
It is recommended that the analyzer remains powered up to the sensor while preparing for calibration or while undergoing routine maintenance. Sensor life will not be shortened under these conditions because only a very small current flows through the sensor. If for any rea­son the sensor has to be disconnected (or the analyz­er switched off) the sensor will have to be polarized before it can be ready for further operation.
IMPORTANT NOTE
The first time the analyzer is powered up with a sensor installed, Orn may be flash­ing. Press the CAL key, then the SELECT key, and then press ENTER to remove this fault.
5.3 CALIBRATION.
5.3.1 Temperature Calibration. For accurate temper-
ature compensation and temperature readings, the TEMP function of the analyzer must be calibrated.
1. Place the sensor in a container filled adequately with process sample or any known solution.
2. Place a calibrated temperature reading device in the sample container.
3. Allow the readings to stabilize.
4. When the readings are stable, compare the ana­lyzer's reading to that of the calibrated tempera­ture indicating device.
If the analyzer's reading requires adjusting, follow
these steps:
1. Press the PV key to ensure that the analyzer is not in Set Mode.
2. Press the TEMP key once. oF or oC will show briefly, then the present temperature is displayed in either °F or °C (depending on the unit selected in the t-C/d-t menu).
3. Press the SELECT key to adjust the value. The dis­play will acknowledge briefly with AdJ followed by the numeric display with right digit flashing.
4. SCROLL (é) and SELECT to display the desired correct temperature.
5. Press ENTER.oF or oC will show briefly, then the solution temperature is displayed.
5.3.2 Analyzer Zero. This procedure is required to elec-
tronically zero the analyzer/sensor loop. In order to elim­inate any residual current in the loop the sensor must be placed in a zero Dissolved Oxygen solution and remain there for a minimum of 12 to 24 hours before adjusting the analyzer zero (0 ppm or 0% saturation).
1. Place the sensor in a sodium sulfite solution (1 gram or about 1/10 teaspoonful to a liter of water) or in pure nitrogen gas.
2. Allow the sensor to stabilize in the zero DO envi­ronment.
3. Press the ACCESS key twice to enter the Set Mode.
4. Press the SCROLL (é) key until -0- is displayed.
5. Press the SELECT key. -0- will flash for about five seconds and then freeze. The analyzer loop is now zeroed.
6. Press the ENTER key to return to the first level of the Set Menu, or press the PV key to get out of the Set Mode.
7. Remove the sensor from the solution. Gently clean and dry the sensor.
8. Proceed to Section 5.3.3, ppm Calibration, or Section 5.3.4, % Saturation Calibration.
29
MODEL 1054B DO SECTION 5.0
START-UP AND CALIBRATION
5.3.3 ppm Calibration.
5.3.3.1 Air Calibration The Model 1054B DO Analyzer
and the appropriate DO Sensor loop may be air cali­brated quite easily.
1. Place the sensor in ambient air.
2. Make sure the sensor is clean, dry and in good condition. Make sure the sensor is completely polarized (at least 2 hours). For optimum results, the loop should be powered overnight.
3. Determine the present barometric pressure in your area in one of the following units: bars, inch­es of mercury, or millimeters of mercury.
4. Press the CAL key. The units of pressure select­ed in the bP menu appears briefly and then the value in memory is displayed (Refer to Section
4.7.1).
5. Press the SELECT key. Std appears briefly and then the barometric pressure value in memory is displayed with the right digit flashing.
6. Use SCROLL (é) and SELECT keys to display the correct barometric pressure.
7. Press ENTER. Press the PV key. The oxygen concentration in air is then displayed in ppm as selected on the output range rng menu. This value has been corrected for process solubility. Refer to Section 4.8.
8. The sensor may now be installed in the process. Please refer to the sensor instruction manual for proper installation of the sensor.
NOTE
1 atmosphere = 760 mmHg = 29.92 inches mercury = 1.013 bars
CAUTION
NEVER air calibrate with the sensor in process. This will result in erroneous readings.
5.3.3.2 Grab Sample ppm Standardization.
Air calibration is sufficient for most applications. For greater accuracy, the sensor may be standardized using a grab sample. Single point standardization when ppm (PPu) is chosen is done with the PV key.
Procedure:
1. Place the sensor in process or in a grab sample. Allow it to stabilize.
2. When the analyzer's reading is stable, note the reading. Perform a chemical analysis of the process or grab sample as quickly as possible, making sure that the sample is protected from exposure to air.
3. Note the current DO reading. If it has not changed from the time the sample was taken, proceed to step 5.
4. If the DO reading has changed from the time the sample was taken, use the formula shown in Figure 5-1 to determine the calibration value used in Step 5 C.
5. Standardize the unit loop to the value obtained from the chemical analysis as follows:
A. Press the PV key.
B. Press the SELECT key. Std appears briefly
and then the last dissolved oxygen value is displayed with the right digit flashing.
C. Use SCROLL (é) and SELECT keys to dis-
play the true value from the chemical analysis.
D. Press the ENTER key. The loop is now stan-
dardized.
Analyzer DO Reading at Calibration
DO Calibration Value
=xDO from Chemical Analysis
Analyzer DO Reading at Sampling
FIGURE 5-1. DO Standardization Formula
30
MODEL 1054B DO SECTION 5.0
START UP AND CALIBRATION
5.3.4 % Saturation Calibration.
5.3.4.1 Air Calibration. The Model 1054B DO
Analyzer and the appropriate DO Sensor loop may be air calibrated quite easily.
1. Place the sensor in ambient air.
2. Make sure the sensor is clean, dry and in good condition. Make sure the sensor is completely polarized (at least 2 hours). For optimum results, the loop should be powered overnight.
3. Determine the present barometric pressure in your area in one of the following units: bars, inch­es of mercury, or millimeters of mercury.
4. Press the CAL key. The units of pressure selected in bP menu appears briefly and then the value in memory is displayed (Refer to Section 4.7.1).
5. Press the SELECT key. Std appears briefly and then the barometric pressure value in memory is displayed with the right digit flashing.
6. Use SCROLL (é) and SELECT keys to display the correct barometric pressure.
7. Press ENTER. Press the DO key. The oxygen concentration in air is then displayed in % satura­tion, as selected on the output range rng menu. This value has been corrected for process solubil­ity. Refer to Section 4.8.
8. The sensor may now be installed in the process. Please refer to the sensor instruction manual for proper installation of the sensor.
NOTE
1 atmosphere = 760 mmHg = 29.92 inches mercury = 1.013 bars
CAUTION
NEVER air calibrate with the sensor in process. This will result in erroneous read­ings.
5.3.4.2 Update Barometric Pressure. This function is used to update the barometric pressure after the ana­lyzer/sensor loop has been air calibrated and the sen­sor is installed in the process.
1. Allow the reading to stabilize.
2. Press the ACCESS key twice to enter the Set Mode. SEt will show briefly, then br.
3. Press the SELECT key. The barometric pressure value in memory will be displayed with the right digit flashing.
4. Use the SCROLL and SELECT key to display the current barometric pressure value. Press the
ENTER key.
5.3.4.3 % Saturation Standardization.
Air calibration is sufficient for many applications. For greater accuracy, the sensor may be standardized in process.
1. Install the sensor and perform a barometric update.
a. Allow the reading to stabilize.
b. Press the ACCESS key twice to enter the
Set Mode. SEt will show briefly, then br.
c. Press the SELECT key. The barometric
pressure value in memory entered will be displayed with the right digit flashing.
d. Use the SCROLL and SELECT key to dis-
play the current barometric pressure value. Press the ENTER key.
2. To standardize the sensor, the % saturation of the process must be known.
3. When the analyzer's reading is stable, perform the standardization as follows:
A. Press the PV key.
B. Press the SELECT key. Std appears briefly
and the last % saturation value is displayed with the right digit flashing.
C. Use SCROLL (é ) and SELECT keys as
required to display the processes % satura­tion.
D. Press the ENTER key. The loop is now stan-
dardized.
31
MODEL 1054B DO SECTION 6.0
KEYBOARD SECURITY
SECTION 6.0
KEYBOARD SECURITY
KEYBOARD SECURITY. Display Mnemonic COd.
Select this feature to display the user defined security code. Any three digit number may be used for this code. 000 will disable the security feature. This item is used to prevent accidental changes to the calibration and configuration of the analyzer. When activated, the analyzer will allow all read functions to read normally. If an attempt is made to change a value, LOC will display followed by the Numeric Display ready for the code to be entered. A proper code will unlock the analyzer and the analyzer will return to the last function attempted. Any incorrect value will result in bAd briefly displaying. The analyzer will then return to numeric display and await the entry of the code. Once unlocked, the ana­lyzer will allow access to all functions until the analyzer is either powered down or no keystrokes are made for a period of 2 minutes. If the code should be forgotten, pressing and holding the ACCESS key for 5 seconds will result in display of the code. Releasing the ACCESS key, then pressing ENTER will unlock the analyzer.
Keyboard Security Procedure (COd).
1. Enter Set Mode by pressing the ACCESS key twice.
2. Press the SCROLL (é) key until COd appears on the display.
3. Press the SELECT key.
4. Use the SCROLL (é) and SHIFT (ç) keys to dis­play the desired value, then ENTER it into memory.
NOTE
Entering 000 disables the keyboard securi­ty.
NOTE
Security feature will not activate until 2 minutes without keyboard activity or power is removed from the analyzer, then restored.
32
MODEL 1054B DO SECTION 7.0
THEORY OF OPERATION
SECTION 7.0
THEORY OF OPERATION
7.1 GENERAL. The Model 1054B Dissolved Oxygen
Analyzer automatically and continuously measures concentrations of dissolved oxygen in water or aque­ous solutions. The determination is based on the measurement of the electrical current developed by the DO Sensor in contact with the sample.
7.1.1 Basic DO Measurement.
1. Dissolved Oxygen - The amount of gaseous oxygen,
in mg/l, or ppm by weight, dissolved in a liquid (usual­ly H2O). The presence of dissolved solids affects the solubility of oxygen in water.
2. The amount of oxygen dissolved in fresh water at 100% saturation is inversely proportional to the tempera­ture, and is directly proportional to the pressure.
3. At sea level and a temperature of 20°C, an oxygen saturated solution of water contains 9.2 ppm (parts per million) of oxygen. The figure of 9.2 ppm represents the weight of oxygen with respect to the weight of water.
4. A polarographic oxygen sensor measures oxygen in air as well as in water. In fact, most sensors are air cal­ibrated prior to water measurements.
5. The mineral content of water solution will also alter the amount of dissolved oxygen. For example, salt water in the ocean at 20°C contains only 7.4 ppm of dissolved oxygen compared to fresh water which con­tains 9.2 ppm. This difference may account for the fact that some fish cannot survive when moved from fresh to salt water and vice versa.
7.2 MEASUREMENT VARIABLES. Variables that influence the dissolved oxygen measurement include barometric pressure, relative humidity, sample temper­ature, interfering gases and composition of the liquid medium.
7.2.1 Barometric Pressure and Relative Humidity.
Rate of oxygen diffusion through the sensor mem­brane, and therefore the sensor response, is linear with respect to oxygen partial pressure (assuming con­stant sample temperature).
At the normal sea-level barometric pressure of 760 mm Hg, the oxygen partial pressure of dry air is 160 mm Hg. As atmospheric pressure deviates from the stan­dard value, the oxygen partial pressure varies propor­tionally. Accordingly, the solubility of oxygen in water varies in proportion to the change in the partial pres­sure of oxygen in air. Barometric pressure is therefore a significant factor in analyzer calibration.
In the PPM mode the barometric pressure is only important for air calibration, so the instrument can cal­culate the sensor's output at a known partial pressure. In process, the direct measurement of ppm is not affected by partial pressure.
In the % SAT mode barometric pressure must be known for air calibration and for process measurement at atmospheric pressure.
33
MODEL 1054B DO SECTION 7.0
THEORY OF OPERATION
7.2.2 Relative Humidity. In calibration for dissolved
oxygen measurement, one method is to expose the sensor to a gaseous sample, typically dry air, of accu­rately known oxygen content. The known gaseous oxygen concentration value is then related to a corre­sponding dissolved oxygen value.
Since dry air contains 20.95% oxygen by volume, regardless of the barometric pressure, oxygen can be shown to be directly proportional to the total baromet­ric pressure, according to Dalton’s law of partial pres­sures. Thus for dry air, if the total barometric pressure is known, the partial pressure of oxygen can be com­puted. However, this procedure is valid only for dry air conditions. Humid air has the effect of reducing the partial pressure of oxygen and the other gases in the air without affecting the total barometric pressure.
Thus, for constant barometric pressure, if the humidity in the air is other than zero, the partial pressure of oxy­gen is less than the value for dry air. For most meas­urements taken below 80°F (26.7°C), the effect of water vapor may be ignored.
To determine the partial pressure of oxygen in air at various levels of humidity and barometric pressure, the partial pressure of water is subtracted from the total barometric pressure and the difference is multiplied by
20.95%.
EXAMPLE: If the Barometric pressure = 740 mm Hg
and the Partial Pressure H2O = 20 mm Hg then the Partial pressure O2 = [740 - 20] x 0.2095 mm Hg
= 150 mm Hg (19.95 kPa)
7.2.3 Sample Temperature. The temperature of the sample affects sensor response in two ways:
1. Oxygen Diffusion Rate — The rate of oxygen dif­fusion through the sensor membrane varies with tem­perature at a coefficient of about +3% per degree Celsius, causing a corresponding change in sensor current.
2. Oxygen Solubility — In an oxygen-saturated liquid, partial pressure of dissolved oxygen is equal to the partial pressure of oxygen in the atmosphere above liq­uid. This relationship holds true regardless of the oxy­gen concentration. As sample temperature increases, oxygen partial pressure remains unchanged (except as influenced by vapor pressure of the liquid); however, the dissolved oxygen concentration is reduced.
To compensate for temperature, the analyzer uses the Pt-100 RTD incorporated in the DO Sensor. The Model 1054B DO then corrects for process temperature changes.
7.2.4 Interfering Gases. Gases that are reduced or oxidized at about 0.67 VDC, and thus contribute to sensor current, may cause a readout error. Only a few gases have this characteristic. Common gases that should be avoided include SO2, Cl2 and oxides of nitro­gen. Low-level concentrations of hydrogen-sulfide tend to contaminate the sensor, but do not seriously affect dissolved oxygen measurement. If contaminated, the sensor must be rejuvenated.
7.2.5 Composition of the Liquid Medium. A signifi­cant change in the composition of the solution may change the solubility of oxygen. If the solvent is water, the addition or presence of any water soluble compo­nents, such as sodium chloride, may change the dis­solved oxygen concentration.
In an open equilibrating system, where gas of constant oxygen partial pressure is in direct contact with a salt solution, the solubility of oxygen decreases as salinity increases.
34
MODEL 1054B DO SECTION 8.0
DIAGNOSTICS AND TROUBLESHOOTING
SECTION 8.0
DIAGNOSTICS AND TROUBLESHOOTING
8.1 DIAGNOSTICS. The 1054B DO Analyzer has a
diagnostic feature which automatically searches for fault conditions that would cause an error in the meas­ured dissolved oxygen value. If such a condition occurs, the current output and relays will act as config­ured in default and the fault flag and display will flash. A fault code mnemonic will display at frequent inter­vals. If more than one fault condition exists, the display will sequence the faults at eight second intervals. This will continue until the cause of the fault has been cor­rected. Display of fault mnemonics is suppressed when in Set Mode. Selecting the SHO item will display a history of the two most recent fault conditions unless SHO was cleared. Refer to Section 4.10.
NOTE
If the analyzer is in hold and a fault occurs, the mnemonic HLd will display during the fault sequence.
8.1.1 Fault Mnemonics. Table 8-1 lists the fault mnemonics and describes the meaning of each.
8.1.2 Temperature Compensation. Table 8-2 is a ready reference of RTD resistance values at various temperatures. These are used for test and evaluation of the sensor.
NOTE
Ohmic values are read across the RTD element and are based on the manufactur­er’s stated values (± 1%). Allow enough time for the RTD element to stabilize to the surrounding temperature.
Display Description Display Description
EEP EEPROM write error (bad EEPROM chip). tcH High temperature compensation error. CHS ROM failure (check sum error) (bad ROM chip). tcL Low temperature compensation error.
rci Reverse current input. Orn Overrange error (+20.00 ppm).
SEn Sensor line error or wire length error. Eci Excessive current input. COP Computer not operating properly. FAC Factory calibration required.
Temperature Resistance
0°C 100 ohms
10°C 104 ohms
20°C 108 ohms
30°C 112 ohms
40°C 116 ohms
50°C 119 ohms
60°C 123 ohms
70°C 127 ohms
80°C 131 ohms
90°C 135 ohms
100°C 139 ohms
TABLE 8-1. Fault Mnemonics
TABLE 8-2. RTD Resistance Values
35
MODEL 1054B DO SECTION 8.0
DIAGNOSTICS AND TROUBLESHOOTING
8.2 TROUBLESHOOTING. The Model 1054B DO
Analyzer is designed with state-of-the-art micro­processor circuitry, making troubleshooting simple and direct. Subassembly replacement, i.e. printed circuit board replacement, is all that is usually required.
8.2.1 Installation Failure. If failure does occur, com­plete the following steps:
1. Check for sensor failure first.
2. Check for a fault flag. If a fault condition exists, refer to Table 8-1 for the fault mnemonic explanation.
3. Check wiring connections for proper installation.
4. Refer to Troubleshooting Table 8-3. The table is arranged with the most common problems listed first.
8.2.2 Display Test. Display Mnemonic dtS. Selecting
this option will activate all the display segments. This item is used if a faulty display is suspected. Refer to Section 4.14.
8.2.3 Software Version. Display Mnemonic UEr.
Selection of this item will display the software revision level of the CPU. This number may be requested by factory service personnel if troubleshooting is required. Refer to Section 4.13.
8.2.4 Sensor Troubleshooting. In addition to the fault
mnemonics that directly relate to a possible sensor problem (SLP, tcH, tcL), the analyzer can display the microamp input from the sensor. This information can aid in sensor and application problems.
1. Enter Set Mode by pressing the ACCESS key twice.
2. Press the SCROLL (
é
) key until in appears on
the display.
3. Press the SELECT key. The current input from the sensor will be displayed in microamps. A properly operating sensor will produce approxi­mately 2.5 microamps per ppm at 25°C.
4. Press the ENTER key to return to the first level of the Set Menu, or press the DO key to get out of the Set Mode.
8.2.5 CPU Board Replacement. If there is a problem
with the CPU board resulting in its replacement, spe­cific procedures (included with the order) for calibrat­ing the new board must be followed exactly or the microprocessor will be improperly programmed. Should this occur, it will be necessary to return the analyzer to the factory for reprogramming.
8.2.6 Power Board Replacement. If it becomes nec-
essary to replace the power board, the CPU board will need to be recalibrated following specific procedures that are included with the power board. Failure to fol­low these procedures exactly will cause the micro­processor to be improperly programmed and require the return of the analyzer to the factory for reprogram­ming.
8.3 INSTRUMENT MAINTENANCE. To maintain the
appearance and extend the life of the enclosure, it should be cleaned on a regular basis using a mild soap and water solution followed by a clean water rinse.
36
MODEL 1054B DO SECTION 8.0
DIAGNOSTICS AND TROUBLESHOOTING
SYMPTOM PROBLEM ACTION
Fault code tcH/tcL 1. Miswire. 1. Check wiring between the sensor
and analyzer.
2. Open or shorted RTD. 2. Replace sensor.
Fault code Orn/Eci 1. Overrange error. 1a. Pull the sensor out of the process
clean, and check membrane. 1b. Recalibrate.
Fault code SEn 1. Open wire between sensor 1. Repair wire.
and analyzer.
2. Cable length has been exceeded. 2. Check wiring. Maximum length: consult factory.
Fault code EEP 1. Defective EEPROM. 1. Replace CPU PCB.
Fault code CHS 1. Defective CPU. 1. Replace CPU PCB.
Alarm relay will not close 1. Defective power card. 1. Replace power PCB.
2. Defective CPU. 2. Replace CPU PCB.
No output current 1. Defective output board. 1. Replace power PCB.
2. Miswire. 2. Check for short.
Low output current 1. Circuit loading with excessive. 1. Consult output loading limits.
resistance on output. Analyzer specifications (600 ohms max load).
Reading loses accuracy 1. Incorrect temp. compensation. 1. Verify Snr menu selection (Refer as process temp. changes to Section 4.9.1).
TABLE 8-3. Troubleshooting Guide
37
MODEL 1054B DO SECTION 9.0
RETURN OF MATERIALS
SECTION 9.0
RETURN OF MATERIALS
9.1 GENERAL. To expedite the repair and return of
instruments, proper communication between the cus­tomer and the factory is important. A return material authorization number is required. Call (949) 757-8500. The Return of Materials Request form is provided for you to copy and use in case the situation arises. The accuracy and completeness of this form will affect the processing time of your materials.
9.2 WARRANTY REPAIR. The following is the proce­dure for returning products still under warranty.
1. Contact the factory for authorization.
2. Complete a copy of the Return of Materials Request form as completely and accurately as possible.
3. To verify warranty, supply the factory sales order number or the original purchase order number. In the case of individual parts or sub-assemblies, the serial number on the mother unit must be sup­plied.
4. Carefully package the materials and enclose your Letter of Transmittal and the completed copy of the Return of Materials Request form. If possible, pack the materials in the same manner as it was received.
IMPORTANT
Please see second section of Return of Materials Request Form. Compliance to the OSHA requirements is mandatory for the safety of all personnel. MSDS forms and a certification that the instruments have been disinfected or detoxified are required.
5. Send the package prepaid to:
Rosemount Analytical Inc. 2400 Barranca Parkway Irvine, CA 92606
Attn: Factory Repair
Mark the package:
Returned for Repair RMA No. _______________
Model No. ______________
9.3 NON-WARRANTY REPAIR.
1. Contact the factory for authorization.
2. Fill out a copy of the Return of Materials Request form as completely and accurately as possible.
3. Include a purchase order number and make sure to include the name and telephone number of the right individual to be contacted should additional information be needed.
4. Do Steps 4 and 5 of Section 9.2.
NOTE
Consult the factory for additional infor­mation regarding service or repair.
FROM: RETURN BILL TO:
_____________________________ _____________________________ _____________________________
_____________________________ _____________________________ _____________________________
_____________________________ _____________________________ _____________________________
CUSTOMER/USER MUST SUBMIT MATERIAL SAFETY SHEET (MSDS) OR COMPLETE STREAM COMPOSITION, AND/OR LETTER CERTIFYING THE MATERIALS HAVE BEEN DISINFECTED AND/OR DETOXIFIED WHEN RETURNING ANY PRODUCT, SAMPLE OR MATERIAL THAT HAVE BEEN EXPOSED TO OR USED IN AN ENVIRONMENT OR PROCESS THAT CONTAINS A HAZARDOUS MATERIAL ANY OF THE ABOVE THAT IS SUBMITTED TO ROSEMOUNT ANALYTICAL WITH- OUT THE MSDS WILL BE RETURNED TO SENDER C.O.D. FOR THE SAFETY AND HEALTH OF OUR EMPLOYEES. WE THANK YOU IN ADVANCE FOR COMPLIANCE TO THIS SUBJECT.
SENSOR OR CIRCUIT BOARD ONLY: (Please reference where from in MODEL / SER. NO. Column)
1. PART NO.__________________________1. MODEL_________________________________1. SER. NO. ________________
2. PART NO.__________________________2. MODEL_________________________________2. SER. NO. ________________
3. PART NO.__________________________3. MODEL_________________________________3. SER. NO. ________________
4. PART NO.__________________________4. MODEL_________________________________4. SER. NO. ________________
PLEASE CHECK ONE:
n REPAIR AND CALIBRATE n DEMO EQUIPMENT NO. __________________________
n EVALUATION n OTHER (EXPLAIN) _______________________________
n REPLACEMENT REQUIRED? n YES n NO _________________________________________________
DESCRIPTION OF MALFUNCTION:
______________________________________________________________________________________________________
______________________________________________________________________________________________________
______________________________________________________________________________________________________
WARRANTY REPAIR REQUESTED:
n YES-REFERENCE ORIGINAL ROSEMOUNT ANALYTICAL ORDER NO. ________________________________________
CUSTOMER PURCHASE ORDER NO. _________________________________________________
n NO-PROCEED WITH REPAIRS-INVOICE AGAINST P.O. NO. _________________________________________________
n NO-CONTACT WITH ESTIMATE OF REPAIR CHARGES: LETTER n __________________________________________
PHONE n ___________________________________________
NAME ____________________________________________________ PHONE _________________________________________
ADDRESS ___________________________________________________________________________________________________
______________________________________________________________ ZIP _________________________________________
RETURN AUTHORITY FOR CREDIT ADJUSTMENT [Please check appropriate box(s)]
n WRONG PART RECEIVED n REPLACEMENT RECEIVED
n DUPLICATE SHIPMENT REFERENCE ROSEMOUNT ANALYTICAL SALES ORDER NO. _________
n RETURN FOR CREDIT RETURN AUTHORIZED BY: ______________________________________
WARRANTY DEFECT____________________________________________________________________________________
_____________________________________________________________________________________________________
24-6047
RETURN OF MATERIALS REQUEST
•IMPORTANT!
This form must be completed to ensure expedient factory service.
R E P A
I
R
S T A T U S
R E A S O N
F O R
R E T U R N
C U S T O
M
E R
N O
T
I
C
E
T
O
S E N D E R
Emerson Process Management
Rosemount Analytical Inc.
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250
http://www.RAuniloc.com
© Rosemount Analytical Inc. 2001
WARRANTY
Seller warrants that the firmware will execute the programming instructions provided by Seller, and that the Goods manufactured or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expira­tion of the applicable warranty period. Goods are warranted for twelve (12) months from the date of initial installation or eighteen (18) months from the date of shipment by Seller, whichever period expires first. Consumables, such as glass electrodes,
membranes, liquid junctions, electrolyte, o-rings, catalytic beads, etc., and Services are warranted for a period of 90 days from the date of shipment or provision.
Products purchased by Seller from a third party for resale to Buyer ("Resale Products") shall carry only the warranty extended by the original manufacturer. Buyer agrees that Seller has no liability for Resale Products beyond making a reasonable commercial effort to arrange for procurement and shipping of the Resale Products.
If Buyer discovers any warranty defects and notifies Seller thereof in writing during the applicable warranty period, Seller shall, at its option, promptly correct any errors that are found by Seller in the firmware or Services, or repair or replace F.O.B. point of man­ufacture that portion of the Goods or firmware found by Seller to be defective, or refund the purchase price of the defective por­tion of the Goods/Services.
All replacements or repairs necessitated by inadequate maintenance, normal wear and usage, unsuitable power sources, unsuit­able environmental conditions, accident, misuse, improper installation, modification, repair, storage or handling, or any other cause not the fault of Seller are not covered by this limited warranty, and shall be at Buyer's expense. Seller shall not be obligat­ed to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in writing in advance by an authorized Seller representative. All costs of dismantling, reinstallation and freight and the time and expenses of Seller's person­nel for site travel and diagnosis under this warranty clause shall be borne by Buyer unless accepted in writing by Seller.
Goods repaired and parts replaced during the warranty period shall be in warranty for the remainder of the original warranty peri­od or ninety (90) days, whichever is longer. This limited warranty is the only warranty made by Seller and can be amended only in a writing signed by an authorized representative of Seller. Except as otherwise expressly provided in the Agreement, THERE ARE NO REPRESENTATIONS OR WARRANTIES OF ANY KIND, EXPRESS OR IMPLIED, AS TO MERCHANTABILITY, FIT­NESS FOR PARTICULAR PURPOSE, OR ANY OTHER MATTER WITH RESPECTTO ANY OF THE GOODS OR SERVICES.
RETURN OF MATERIAL
Material returned for repair, whether in or out of warranty, should be shipped prepaid to:
Emerson Process Management
Liquid Division
2400 Barranca Parkway
Irvine, CA 92606
The shipping container should be marked:
Return for Repair Model
_______________________________
The returned material should be accompanied by a letter of transmittal which should include the following information (make a copy of the "Return of Materials Request" found on the last page of the Manual and provide the following thereon):
1. Location type of service, and length of time of service of the device.
2. Description of the faulty operation of the device and the circumstances of the failure.
3. Name and telephone number of the person to contact if there are questions about the returned material.
4. Statement as to whether warranty or non-warranty service is requested.
5. Complete shipping instructions for return of the material. Adherence to these procedures will expedite handling of the returned material and will prevent unnecessary additional charges
for inspection and testing to determine the problem with the device.
If the material is returned for out-of-warranty repairs, a purchase order for repairs should be enclosed.
Credit Cards for U.S. Purchases Only.
The right people, the right answers, right now.
ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
http://www.raihome.com
Emerson Process Management Liquid Division
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250
http://www.raihome.com
© Rosemount Analytical Inc. 2003
Loading...