Roscoe Medical InTENSity IF Combo II User Manual User Manual

Page 1
TENSPros
IF Combo II
INSTRUCTION MANUAL
www.roscoemedical.com
Page 2
TENSPros
InTENSity IF Combo II Roscoe Medical
This manual is valid for the
IF Combo II
Roscoe Medical Inc. reserves the right to improve and amend it at any time without prior notice. Amendments may however be published in new editions of this manual.
All Rights Reserved.Rev.V1.0 © 2015, 20150119
United States Federal Law restricts this device to sale by or on the order of a physician or licensed practitioner
Conformity to safety standards
Roscoe Medical Inc. declares that the
device complies with following normative documents:
IEC60601-1, IEC60601-1-2, I EC60601-2-10,
IEC62366, IEC60601-1-11, ISO10993-5, ISO10993-10,
ISO10993-1, ISO7010
- 2 -
Page 3
TENSPros
Roscoe Medical InTENSity IF Combo II
Table of Contents
1. Introduction ........................................4
1.1 General
1.2 Medical Background
1.3 Indication for use
2. Important Safety Precautions and Warnings............6
3. Package Contents .................................13
3.1 Front And Rear Panel
3.2 LCD Display
4. Specification ......................................16
4.1 Technical Information
4.2 Program Parameters
4.3 Waveform Information
5. Instruction For Use................................. 19
5.1 Battery
5.2 Connecting Electrodes
5.3 Connecting Lead Wires
5.4 Turn On The Device
5.5 Setting a New Program
5.6 Turn Off Device
5.7 Other Important Functions Safety Lock Feature Low Battery Indicator Charging The Battery
5.8 Patient Compliance Meter
6. Cleaning And Storage ..............................32
6.1 Cleaning The Unit
6.2 Cleaning The Electrode Pads
6.3 Storing The Electrode Pads And Lead Wires
6.4 Storing The Unit
7. Troubleshooting ...................................34
8. Disposal ..........................................36
9. Glossary Of Symbols ...............................36
10. Electromagnetic Compatibility (EMC) Tables ..........37
11. Warranty .......................................... 43
- 3 -
Page 4
TENSPros
InTENSity IF Combo II Roscoe Medical
1. INTRODUCTION
1.1 General
The InTENsity IF Combo II is a portable electrotherapy device featuring two therapeutic modes: Transcutaneous Electrical Nerve Stimulator (TENS) and Interferential (IF), which are used for pain relief. The stimulator sends gentle electrical current to underlying nerves via electrodes applied on the skin. The parameters of device are controlled by the buttons on the front panel. The intensity level is adjustable according to the needs of patients.
1.2 Medical Background
Explanation of pain
Pain is an unpleasant sensation that can serve a useful purpose by alerting us to a possible injury or disease. When the body is functioning normally, pain serves as a warning system that something is not right. Without pain a person would not know when to avoid danger or seek medical help. Pain becomes a problem when it continues after treatment has started or long after an injury is healed.
There are two types of pain:
Acute — Limited in duration. Examples include but are not limited to:
• Sprains • Incisional pain • Muscle strain
Chronic — Long-lasting, persistent pain that ceases to serve
• Low back pain • Pinched Nerves
• Bursitis • Joint Pain
The IF Combo II was developed to help relieve some types of chronic and acute pain.
as a warning system and becomes a problem. Examples include but are not limited to:
- 4 -
Page 5
TENSPros
Roscoe Medical InTENSity IF Combo II
How does TENS work?
There is nothing “magic” about Transcutaneous Electrical Nerve Stimulation (TENS). TENS is intended to help relieve pain. The TENS unit sends comfortable impulses through the skin to stimulate the nerve (or nerves) in the treatment area. In many cases, this stimulation will greatly reduce or eliminate the pain sensation the patient feels. Pain relief varies by individual patients, mode selected for therapy, and the type of pain. In many patients, the reduction or elimination of pain lasts longer than the actual period of stimulation (sometimes as much as three to four times longer). In others, pain is only modied while stimulation actually occurs. You may discuss this with your physician or therapist.
How IF works
Interferential Stimulation (IF) is an anti-inammatory based treatment modality. Interferential stimulation is characterized by two alternating-current sine waves or square waves of differing frequencies that “work” together to produce an interferential current that is also known as a beat pulse or alternating modulation frequency. One of the two currents is usually held at 5,000Hz,and the other can be held constant or varied over a range of 5,001 – 5,150Hz. Because of the frequency, the interferential wave meets low impedance when crossing the skin to enter deep into soft tissues. The interferential currents reportedly can stimulate sensory, motor, and pain bers. These large impulse bers interfere with the transmission of pain messages at the spinal cord level. This deep tissue penetration stimulates parasympathetic nerve bers for increased blood ow and edema reduction. It utilizes the low electric-current to stimulate muscle nerves to achieve the symptomatic relief of chronic intractable pain, post-traumatic pain, and post-surgical pain.
- 5 -
Page 6
TENSPros
InTENSity IF Combo II Roscoe Medical
1.3 Indication for use
IF Combo II Stimulator may be used for the following conditions:
1) Symptomatic relief of chronic intractable pain.
2) Post traumatic pain.
3) Post surgical pain.
2. IMPORTANT SAFETY PRECAUTIONS AND WARNINGS
It is important that you read all the warning and
DANGER
WARNING
CAUTION
This stimulator must not be used in combination with the following medical devices:
• Internally transplanted electronic medical devices,
• Electronic life support equipment, such as respirators.
• Electronic medical devices attached to the body, such as electrocardiographs. Using this stimulator with other electronic medical devices may cause errone­ous operation of those devices.
precautions included in this manual because they are intended to keep you safe, prevent injury and avoid a situation that could result in damage to the device.
SAFETY SYMBOLS USED IN THIS MANUAL
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, could result in serious injury and equipment damage.
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or damage to the device or other property.
DANGER
such as a pacemaker.
- 6 -
Page 7
TENSPros
Roscoe Medical InTENSity IF Combo II
WARNING
DO NOT USE THIS DEVICE UNDER THESE CONDITIONS
• Consult with your physician before using this device, because the device may cause lethal rhythm disturbances in certain susceptible individuals.
• If you have a cardiac pacemaker, implanted debrillator, or other implanted metallic or electronic device. Such use could cause electric shock, burns, electrical interference, or death.
• Together with a life-supporting medical electronic device such as an articial heart or lung or respirator.
• In the presence of electronic monitoring equipment (e.g., cardiac monitors, ECG alarms), which may not operate properly when the electrical stimulation device is in use.
• On open wounds or rashes, or over swollen, red, infected, or inamed areas or skin eruptions (e.g., phlebitis, thrombophlebitis, varicose veins); or on top of, or in proximity to, cancerous lesions.
• Over areas of skin that lack normal sensation.
• On the opposite sides of your head since the effects of stimulation of the brain are unknown.
DO NOT USE ON THESE INDIVIDUALS
• Pregnant women, because the safety of electrical stimulation during pregnancy has not been established.
• Children or infants, because the device has not been evaluated for pediatric use.
• Persons incapable of expressing their thoughts or intentions.
DO NOT USE THIS DEVICE DURING THESE ACTIVITIES
• When in the bath or shower
• While sleeping
• While driving, operating machinery, or during any activity in which electrical stimulation can put you at risk for injury.
- 7 -
Page 8
TENSPros
InTENSity IF Combo II Roscoe Medical
WARNING (CONTINUED)
PAIN MANAGEMENT WARNINGS
• If you have had medical or physical treatment for your pain, consult with your physician before using this device.
• If your pain does not improve, becomes seriously chronic or severe, or continues for more than ve days, stop using the device and consult with your physician.
• The mere existence of pain functions as a very important warning telling us that something is wrong. Therefore, if you suffer from any serious illness, consult your physician in order to conrm that it is advisable for you to use this TENS Stimulator.
WARNINGS AND PRECAUTIONS REGARDING THE PADS
• Apply pads to normal, healthy, dry, clean skin (of adult patients) because it may otherwise disrupt the healing process.
• If you experience any skin irritation or redness after a session, do not continue stimulation in that area of the skin.
NEVER APPLY THE PADS TO:
• The head or any area of the face.
• Any area of the throat because this could cause severe muscle spasms resulting in closure of the airway, difculty in breathing, or adverse effects on heart rhythm or blood pressure.
• Both sides of the thorax simultaneously (lateral or front and back), or across your chest because the introduction of electrical current may cause rhythm disturbances which could be lethal.
- 8 -
Page 9
TENSPros
Roscoe Medical InTENSity IF Combo II
CAUTION
WARNINGS AND PRECAUTIONS REGARDING THE PADS
• United States Federal Law restricts this device to sale by or on the order of a physician or licensed practitioner
• Do not bend or fold because the pad may not function properly. Place the pads onto the plastic lm and then store into the sealed package when not in use.
• Do not apply ointment or any solvent to the pads or to your skin because it will disrupt the pads from functioning properly.
• The pads are already pre-gelled and will adhere to your skin.
• To avoid damage to the adhesive surface of the pads, put the pads only on the skin or on the plastic lm provided.
• Place the pads at least 1 inch apart on your skin. The pads should never touch each other.
• Make sure the components are connected well and the pads are xed on the part of the body you wish to treat or the therapy may not be effective.
DO NOT USE YOUR PADS THIS WAY
• Pads should not touch each other when placed onto your skin.
• Do not place on your spine or backbone.
• Pad should not touch any metal object, such as a belt buckle or necklace.
• Pads should not be placed simultaneously on the soles of both feet.
• Pads should not be placed simultaneously on the calves of both legs.
• Do not share pads with another person. This may cause a skin irritation or
infection. Pads are intended for use by one person.
• Do not place or relocate the pads while the device is on.
• Always turn the power off before removing or changing the pad location.
• Do not leave pads attached to the skin after treatment.
- 9 -
Page 10
TENSPros
InTENSity IF Combo II Roscoe Medical
CAUTION (CONTINUED)
CAUTION WHILE USING THE STIMULATOR
• If the stimulator is not functioning properly or you feel discomfort, immediately stop using the device.
• Do not use for any other purpose except for what it is intended for.
• Do not insert the electrode plug into any place other than the jack on the main unit.
• Do not mix Alkaline and Manganese batteries as this will shorten the battery life.
• Do not pull on the electrode cord during treatment.
• Do not use the device while wearing electronic devices such as watches as this may damage the device.
• Do not use near a cell phone as this may cause the stimulator to malfunction.
• Do not bend or pull the end of the cord.
• When pulling out the cord from the device, hold the plug and pull.
• Replace the cord when broken or damaged.
• Do not throw the batteries into a re. The batteries may explode.
• Dispose of the device, batteries, and components according to applicable legal regulations. Unlawful disposal may cause environmental pollution.
• The size, shape and type of pads may affect the safety and effectiveness of electrical stimulation.
• The electrical performance characteristics of pads may affect the safety and effectiveness of electrical stimulation.
• Using pads that are too small or incorrectly applied, could result in discomfort or skin burns.
- 10 -
Page 11
TENSPros
Roscoe Medical InTENSity IF Combo II
GENERAL PRECAUTIONS
• The long-term effects of electrical stimulation are unknown.
• Apply stimulation to only normal, intact, clean, dry, and healthy skin.
• TENS is not effective in treating the original source or cause of the pain, including headache.
• TENS is not a substitute for pain medications and other pain management therapies.
• TENS devices do not cure disease or injuries.
• TENS is a symptomatic treatment and, as such, suppresses the sensation of pain that would otherwise serve as a protective mechanism.
• Effectiveness is highly dependent upon patient selection by a practitioner qualied in the management of pain patients.
• You may experience skin irritation or hypersensitivity due to the electrical stimulation or electrical conductive medium (gel).
• If you have suspected or diagnosed heart disease, you should follow precautions recommended by your physician.
• If you have suspected or diagnosed epilepsy, you should follow precautions recommended by your physician.
• Use caution if you have a tendency to bleed internally, such as following an injury or fracture.
• Consult with your physician prior to using the device after a recent surgical procedure, because stimulation may disrupt the healing process.
• Use caution if stimulation is applied over the menstruating or pregnant uterus.
• Use caution if stimulation is applied over areas of skin that lack normal sensation.
Keep unit away from young children. The unit contains small pieces that may be swallowed. The electrode cord can cause strangulation. Immediately contact your physician should any of these things occur.
• Use this device only with the leads, electrodes, and accessories recommended by the manufacturer.
• Keep unit out of the reach of young children.
- 11 -
Page 12
TENSPros
InTENSity IF Combo II Roscoe Medical
POSSIBLE ADVERSE REACTIONS
• Do not use to treat one region for extended periods of time (more than 30 minutes a session, up to 2 times/day) or muscles in that region may become exhausted and sore.
• You may experience skin irritation and burns beneath the stimulation electrodes applied to your skin.
• You should stop using the device and consult with your physician if you experience adverse reactions from the device.
Note: Always use electrodes that are legally marked and sold
in the United States under 510K guidelines.
- 12 -
Page 13
TENSPros
Roscoe Medical InTENSity IF Combo II
TM
3. PACKAGE CONTENTS
IF Combo II Unit 2 × Lead wires 4 × Electrode pads (2" x 2")
IF Combo II
INSTRUCTION MANUAL
1 x Instruction Manual 1 × LI rechargeable 1 × USB Cable Battery Pack and Wall Charger
QUICK REFERENCE GUIDE PAGE 1
The purpose of this guide is to assist you in correctly operating your Intensity™ 10 unit. This device should only be used under the continued supervision of a licensed physician. This guide is not intended to replace or substitute the Intensity™ 10 Instruction Manual. PLEASE READ THE INSTRUCTION MANUAL COMPLETELY BEFORE USING THIS DEVICE. Read, understand and practice the warnings, cautions and operating instructions. Always follow the operating instructions prescribed by your healthcare provider.
INTENSITY™ 10 UNIT
Top
CH1 CH2
Channel 1 and 2 sockets
1
1
2
AC/DC adapter receptacle
Side Front Back
3
T (time) and S (therapeutic part) selection buttons
Channel 1 intensity increase button
4
10
6
Channel 1 intensity decrease button
5
LCD display
6
3
Channel 2 intensity increase button
7
2
4
7
Channel 2 intensity decrease button
8
Power on/o button
9
8
5
9
Belt clip
10
Battery compartment cover
11
11
S
LCD DISPLAY
1
1
TENS therapeutic mode
2
6
Lock function indicator
2
3
ELECTRODES
Low-battery indicator
3
4
(2 SETS)
7
5
Treatment time
4
Timer symbol
5
Channel 1 intensity level
6
8
Channel 2 intensity level
7
LEAD WIRES
Body part selection
8
(2 SETS)
For more information about your Intensity™ 10 unit, please visit us online at roscoemedical.com or call 1-800-871-7858.
1 x Quick Start Guide
- 13 -
Page 14
TENSPros
InTENSity IF Combo II Roscoe Medical
3.1 Front and Rear Panel
9
12
CH1 CH2
8
1
2
7
3
4
1) LCD Display: Operating state of the device
2) Body Part Buttons: Select body part or treatment program
3) Channel 1 Intensity Buttons: Increase or decrease the output intensity of channel 1.
4) Set button: Press this button to enter the setting status
5) Power Button: Press once to turn on. Press and hold for 3 seconds to turn off.
6) Channel 2 Intensity Buttons: Increase or decrease the output intensity of channel 2.
7) Pause/Mode Button: Press this button to change treatment modes (TENS or IF). Press button again to pause active treatment status
8) LED: Charging indicator light.
9) Output Sockets: Lead wire output sockets
10) Battery compartment cover.
11) Belt Clip.
12) USB charging port.
S
5
6
- 14 -
11
10
Page 15
TENSPros
Roscoe Medical InTENSity IF Combo II
3
12 11
3.2 LCD Display
1
2
4
5 6
7 8
9
1) Displays waveform mode.
2) Displays home screen
3) Displays set mode
4) Displays output state: start, pause, lock
5) Hour Indicator for compliance meter
6) Cycle Time
7) Displays output intensity of channel 1, pulse width and pulse rate Beat L
8) Displays therapeutic body part: hand, foot, shoulder, arm, leg and back
9) Displays therapeutic mode: TENS, IF
10) Low-battery indicator
11) Load indicator
12) Indicates compliance meter mode
13) Displays the treatment time
14) Display output intensity of channel 2, Beat H
10
13
14
- 15 -
Page 16
TENSPros
InTENSity IF Combo II Roscoe Medical
4. SPECIFICATION
4.1 Technical Information
Channel Dual, isolated between channels
Power Supply 3.7V LI rechargeable battery pack
Operating Conditions 5°C to 40°C (41°F to 104°F) with a relative
Storage and Transport Conditions
Dimensions 11.7 × 6 × 2.1 cm (L*W*H)
Weight 3.5 oz. (Without battery)
Electrode Detection Function
Technical Specifications
Waveform Symmetrical bi-phase rectangular wave
Pulse Amplitude Adjustable,
Pulse Width
Pulse Rate TENS: 1 ~ 150Hz
Treatment Time Time 5 ~ 90min
Cycle Time 1 ~ 20 sec.
charger output: 5.0V DC, 300mA (optional)
humidity of 30% – 75%,atmospheric pressure from 700 to 1,060 Hpa
-10°C to 55°C (14°F to 131°F) with a relative humidity of 10%-90%,atmospheric pressure from 700 to 1,060 Hpa
The amplitude level will be reset to 0mA when the amplitude level is 10mA or greater and an open circuit at either channel is detected.
TENS: 0 ~ 100mA IF: 3 ~ 35mA (@ 1,000ohm load), 1mA/Step
From 50 to 400µs microseconds
IF: Carrier F 2.5/5/10KHz,
Beat.L 1 ~ 149Hz Beat.H 81 ~ 200Hz
- 16 -
Page 17
TENSPros
Roscoe Medical InTENSity IF Combo II
Neck
Shoulder
Hand
Low back
Knee
Foot
4.2 Program Parameters
TENS
Body
Part
Program
Treatment
Time
Pulse
Rate
Pulse Width
Cycle
Time
Body Part Program
All
Therapeutic
Parts
P1 20 Min. 80 – 100Hz
P2 20 Min. 4Hz
P1 20 Min. 80 – 100Hz
P2 20 Min. 10Hz
P1 20 Min. 100Hz
P2 20 Min. 1 – 10Hz
P1 20 Min. 80 – 100Hz
P2 20 Min. 4Hz
P1 20 Min. 120Hz
P2 20 Min. 1 – 10Hz
P1 20 Min. 80 – 120Hz
P2 20 Min. 1 – 10Hz
Treatment
Time
U1
Adjustable
5 – 90 mins
- 17 -
100 – 120µs
150 – 200µs
100µs
220 – 260µs
100µs
200µs
100µs
200 – 260µs
100 – 120µs
150 – 200µs
100 – 120µs
200µs
Pulse
Rate
Adjustable 1 – 150 Hz
10 Sec
20 Sec
10 Sec
20 Sec
Fixed
20 Sec
10 Sec
20 Sec
10 Sec
20 Sec
10 Sec
20 Sec
Pulse Width
Adjustable 50 – 400µs
Page 18
TENSPros
InTENSity IF Combo II Roscoe Medical
Neck
Shoulder
Hand
Low back
Knee
Foot
4.2 Program (Continued)
IF (Sweep) IF (Sweep)
Body Part Program
P1 20 Min. 10,000Hz
P2 20 Min. 5,000Hz
P1 20 Min. 10,000Hz
P2 20 Min. 5,000Hz
P1 20 Min. 10,000Hz
P2 20 Min. 5,000Hz
P1 20 Min. 10,000Hz
P2 20 Min. 5,000Hz
P1 20 Min. 10,000Hz
P2 20 Min. 5,000Hz
P1 20 Min. 10,000Hz
Treatment
Time
Carrier F
P2 20 Min. 5,000Hz
All Body Parts
U1
- 18 -
Min: 5 - 90,
Step is 5min
Default:
20Min
N/A
Page 19
TENSPros
Roscoe Medical InTENSity IF Combo II
Pulse
Width
100µs
200µs
100µs
200µs
100µs
200µs
100µs
200µs
100µs
200µs
100µs
200µs
100µs, 200µs, 400µs change
in a cycle,
default: 100uS
Beat Low Beat High
1Hz 100Hz 10 Sec
1Hz 100Hz 20 Sec
50Hz 150Hz 10 Sec
50Hz 150Hz 20 Sec
1Hz 20Hz 10 Sec
1Hz 20Hz 20 Sec
1Hz 50Hz 10 Sec
1Hz 50Hz 20 Sec
60Hz 120Hz 10 Sec
60Hz 120Hz 20 Sec
1Hz 30Hz 10 Sec
1Hz 30Hz 20 Sec
1 – 149Hz
step is 1,
default: 1Hz
81 – 200Hz
Step is 1,
Default: 81Hz
Default: 20s
Cycle
Time
1 – 20s
Step is 1,
- 19 -
Page 20
TENSPros
InTENSity IF Combo II Roscoe Medical
Cycle time
Cycle time
Cycle time
4.3 Waveform Information — TENS
Continuous
Pulse Width Modulation
Pulse Rate Modulation
Modulation (Pulse rate and width modulation)
Waveform Information — IF
Interferential
- 20 -
Page 21
TENSPros
Roscoe Medical InTENSity IF Combo II
5. INSTRUCTION FOR USE
5.1 Battery
Installation of Battery
Push down on the belt clip to release it. Remove the battery cover and insert the battery, as shown on the diagram. Replace the battery cover and belt clip. For battery charging instructions, please see page 30.
Disposal of Battery
Depleted batteries do not belong in the household waste. Dispose of the batteries according to the your federal, state and local regulations. As a consumer, you are obligated by law to return depleted batteries.
CAUTION
1) Keep the battery and the product out of the range of children.
2) Battery may not be dismantled, thrown into re or short-circuited.
3) Protect battery from excess heat; Take the battery out of the product if the product is not used for a long period of time.
4) Always replace with the same type of battery.
5.2 Connecting Electrodes
Take the pads out of the sealed package; insert the pin of the lead wire into the electrodes pigtail. Make sure there is no bare metal exposed.
- 21 -
Page 22
TENSPros
InTENSity IF Combo II Roscoe Medical
CAUTION
Always use electrodes with CE mark, or which are legally marketed in the US under 510(K) procedure.
5.3 Connecting Lead Wires
1) Before proceeding to this step, be sure the device is completely turned OFF.
2) Insert the lead wires into the output sockets located on the top of the device.
3) Holding the insulated portion of the connector, push the plug end of the wire into one of the sockets (see drawing); one or two sets of wires may be used.
4) This device has two output receptacles controlled by Channel 1 and Channel 2 at the top of the unit. You may choose to use one channel with one pair of lead wires or both channels with two pairs of lead wires. Using both channels gives the user the advantage of stimulat­ing two different areas at the same time.
Note: If you use only one Channel, only plug in 1 lead wire
at the top of the unit
- 22 -
Page 23
TENSPros
Roscoe Medical InTENSity IF Combo II
5.4 Turn on The Device
Press the [ ] button to turn on the device.
CAUTION
Before using the device for the rst time, you are strongly advised
to take careful note of the contraindications and safety measures
detailed at the beginning of this manual (Safety information),
as this powerful equipment is neither a toy nor a gadget!
5.5 Setting a New Program
There are 6 therapeutic body part buttons available neck, shoulder, hand, low back, knee and foot. Each therapeutic part has 3 programs – P1, P2 and U1 for a total of 36 programs.
1. Press [ M/II ] button to select the treatment mode (TENS or IF).
2. Select a body part on the device and press it.
3. The body part will display on the LCD screen.
4. Press the same button again to select one of the 3 programs (such as P1, P2 or U1).
4
2
1
M /
S
3
- 23 -
Page 24
TENSPros
InTENSity IF Combo II Roscoe Medical
Manual IF Program — U1
A. Select Body Part
Select the body part which you want to treat. Then press the body part button until the LCD displays program "U1" like the following:
B. Set Treatment Time
Press [ S ] button to enter treatment time and the min. will ash. Then press [ + ] or [ ] button to adjust the treatment time.
C. Set Pulse Width
Press [ S ] button to enter the pulse width and the "µs" will ash. Then press [ + ] or [ ] button to adjust the pulse width.
- 24 -
Page 25
TENSPros
Roscoe Medical InTENSity IF Combo II
D. Set Pulse Beat (Low & High)
Press [ S ] button to enter the pulse rate and the Hz will ash. Press [ + ] or [ ] button to adjust pulse rate.
E. Set Cycle Time
Press [ S ] button to set the cycle time and the “sec.” will ash and you will see the below image circled in red on the screen. Then press [ + ] or [ ] button to adjust the treatment time.
After you nished settings, press the [ ] or [ S ] button to conrm, and the device will go back to standby status.
- 25 -
Page 26
TENSPros
InTENSity IF Combo II Roscoe Medical
Set Preset IF Program — P1 or P2
A. Select Body Part
Press the “M/II” button to choose IF mode. Then select the body part which you want to treat. Press the body part button until the LCD displays program "P1" or "P2"
B. Set Treatment Time
Press [ S ] button to enter the treatment time and the "Min" will ash. Press [ + ] or [ ] button to adjust treatment time. After you nished settings, press the [ S ] or [ ]button to conrm, the device will go back to the home screen.
C. Start Treatment
Press [CH1+] or [CH2+] to increase the output intensity of channel 1 and/or channel 2. Press [CH1−] or [CH2−] to decrease the output intensity of channel 1 and/or channel 2.
- 26 -
Page 27
TENSPros
Roscoe Medical InTENSity IF Combo II
Manual TENS Program — U1
A. Select Body Part
Select the body part which you want to treat. Then press the body part button until the LCD displays program "U1" like the following:
B. Set Pulse Rate
Press [ S ] button to enter the pulse rate and the "Hz" will ash. Press [ + ] or [ ] button to adjust pulse rate.
C. Set Pulse Width
Press [ S ] button to enter the pulse width and the "µs" will ash. Then press [ + ] or [ ] button to adjust the pulse width.
- 27 -
Page 28
TENSPros
InTENSity IF Combo II Roscoe Medical
D. Set Treatment Time
Press [ S ] button to enter treatment time and the "min." will ash. Then press [ + ] or [ ] button to adjust the treatment time.
Set Preset Programs P1 and P2
A. Select Body Part
Select the body part which you want to treat. Press the body part button until the LCD displays program "P1" or "P2"
B. Set Treatment Time
Press [ S ] button to enter the treatment time and the "Min" will ash. Press [ + ] or [ ] button to adjust treatment time. After you nished settings, press the [ S ] or [ ]button to conrm, the device will go back to the home screen.
C. Start Treatment
Press [CH1+] or [CH2+] to increase the output intensity of channel 1 and/or channel 2. Press [CH1−] or [CH2−] to decrease the output
intensity of channel 1 and/or channel 2.
- 28 -
Page 29
TENSPros
Roscoe Medical InTENSity IF Combo II
CAUTION
1) If the electrodes are not placed rmly on skin or the device has not
connected with the electrodes or lead wires securely, and the output intensity level is equal to or greater than 10mA, the intensity will stop automatically.
2) If the stimulation levels become uncomfortable, reduce the intensity
to a comfortable level. Contact your medical practitioner if the problems persist.
D. Pause or Stop Treatment
If there is an immediate need to pause treatment, press the [ M/II ] button and the “II” will display on the LCD. Press it again to continue treatment or press [ ] button to stop treatment and the device will return to the home screen.
5.6 Turn OFF Device
Press and hold [ ] button until the device turns off.
CAUTION
If there is no operation in the panel for 3 minutes in the
standby status, the device will shut off automatically and
5.7 Other Important Functions
Safety Lock Feature
The lock function automatically activates after there is no operation in the panel for 20 seconds while in treatment status. The indicator " " will display on the LCD .
you will hear a beep sound.
This is a safety feature to prevent accidental changes to your settings and to prevent accidentally increasing the output intensity level. Press [CH1−] or [CH2−] button to unlock.
- 29 -
Page 30
TENSPros
InTENSity IF Combo II Roscoe Medical
Low Battery Indicator
When the low power indicator " " displays on the LCD, stop the device and charge the battery.
Charging the Battery
NOTE: Battery comes pre-charged.
Proceed as follows to recharge the battery:
• This device cannot be used while charging.
• Make sure that the device has been switched off.
• Make sure that the device is no longer connected to the patient (the output cables and electrodes must be disconnected).
• Connect the USB cable to the charging port on the top of device.
• Connect the USB cable to the wall outlet or USB port.
• When the device is charging, the indicator light will be red.
• It could take up to 6 hours to reach a full charge.
• When charging is completed, the indicator light will be green.
After the battery has been recharged, disconnect the cable from the wall outlet or USB port and the device is ready to be used again.
The life of a rechargeable battery depends on the number of recharging/ rundown cycles it undergoes and how these cycles are performed. The following suggestions will help prolong the life of the battery:
• Whenever the device is not used frequently, charge the battery once a month.
• For longer battery life, discharge the battery as much as possible during battery operation.
CAUTION
1) Please use the standard charger provided by the manufacturer or agent.
2) When you are using the device, never connect it to the charger. If you
do, the device will turn off automatically.
3) When charging is complete, you are strongly advised to disconnect the charger
- 30 -
Page 31
TENSPros
Roscoe Medical InTENSity IF Combo II
5.8 Patient Compliance Meter
You can store 90 sets of treatment records and a total treatment time of up to 100 hours on this device.
A. Check and Delete Individual Records
In standby, press [ M/II ] button and hold for 3 seconds to enter the compliance meter records. The LCD will show the number of records and treatment time (as shown in the below picture). Press [ + ] or [ ] buttons to check each record. To delete a record, press [ S ] button and hold for 3 seconds.
B. Check and Delete Accumulative Records
Individual records menu, press [ II ] button to switch to accumulative records menu. Press [ S ] button holding for 3 seconds and all of records will be deleted followed by a beeping sound. press [ ] button to return to the home screen.
CAUTION
The records will be permanently deleted and can be not restored if you
follow the delete records method mentioned above.
- 31 -
Page 32
TENSPros
InTENSity IF Combo II Roscoe Medical
6. CLEANING AND STORAGE
6.1 Cleaning the Unit
1) Turn unit off and disconnect the lead wires from the unit.
2) Clean the device after use with a soft, slightly moistened cloth and wipe gently.
• Do not use chemicals (like thinner, benzene).
• Do not let water get into the internal area.
Note:
This device and accessories (including the electrodes) do not require sterilization.
6.2 Cleaning the Electrode Pads
1) Turn the power off and remove the lead wires from the electrodes.
2) Wash the electrodes when the adhesive surface becomes dirty and/or the electrodes are difcult to attach.
• To “wash” the pads, place a small drop of water on your clean
ngertip and rub the water across the entire gel part. Place the adhesive part face up and let it air dry until the water is absorbed and has been reconstituted. Do not wipe with a tissue paper or cloth. If the electrode still does not stick properly, replace them with new electrodes.
- 32 -
Page 33
TENSPros
Roscoe Medical InTENSity IF Combo II
CAUTION
1) The life of electrodes may vary by the frequency of washing, skin condition, and storage state.
2) If the electrodes no longer stick to your skin or the electrodes are broken, you should replace new electrodes.
3) Before applying the self-adhesive electrodes, it is recommended to wash and degrease the skin, and then dry it.
4) Do not turn on the device when the electrodes are not positioned on the body.
5) Never remove the self-adhesive electrodes from the skin while the device is still turned on.
6) If replacement electrodes are necessary, use only electrodes that are the same size (2" x 2") as the electrodes provided with the IF Combo II.
7) Use of electrodes that are larger may reduce the effect of the stimulation. Use of electrodes that are much smaller than the electrodes provided with the IF Combo II may increase the chance of skin irritation or electrode burns occurring under the electrodes.
8) Always use electrodes that have been cleared for marketing in the USA by the FDA.
6.3 Storing The Electrode Pads and Lead Wires
1) Turn the device off and remove the lead wires from the unit.
2) Remove the electrodes from your body and disconnect the lead wires from the electrodes.
3) Place the electrodes onto the plastic lm and then store into the sealed package.
4) Wrap the lead wires and store into the sealed package.
6.4 Storing the Unit
1) Place the unit, electrodes, lead wires and manual back into the carrying case. Store the box in a cool, dry place, -10ºC ~ 55ºC; 10% ~ 90% relative humidity.
2) Do not keep in places that can be easily reached by children
3) When not in use for a long period, remove the battery before storage.
- 33 -
Page 34
TENSPros
InTENSity IF Combo II Roscoe Medical
7. TROUBLESHOOTING
Problem Possible Causes Possible Solution
Are the batteries
The unit cannot power on
Stimulation weak or cannot feel any stimulation
Stimulation is uncomfortable
Intermittent output Lead wires
exhausted?
Are the batteries installed correctly?
Electrodes dried out or contaminated
Electrodes are not securely attached to the skin.
Lead wires Old/worn/damaged
Intensity is too high Decrease intensity.
Electrodes are too close together
Electrode active area size is too small.
May not be operating the device according to the manual.
- 34 -
Charge or replace the batteries.
Insert the batteries observing polarity.
Replace with new electrodes
Reconnect the electrodes
Replace with new lead wires
Reposition the electrodes at least 1-1∕2" apart.
Replace electrodes with ones that have an active area no less than 25.0cm2 (5cm*5cm).
Please check the manual before use
Verify connection is secure.
Turn down the intensity.
Rotate lead wires in socket 90°. If still intermittent, replace lead wire.
If still intermittent after replacing lead wire, a component may have failed. Call the repair department.
Page 35
TENSPros
Roscoe Medical InTENSity IF Combo II
7. TROUBLESHOOTING (Continued)
Problem Possible Causes Possible Solution
Stimulation is ineffective.
The skin becomes red and/or you feel a stabbing pain
Output current stops during therapy
Li rechargeable battery pack doesn’t last or life is short
Improper electrode place­ment
Unknown Contact clinician.
Using the electrodes on the same site every time.
The electrodes aren’t stuck onto the skin properly
The electrodes are dirty.
The surface of the electrode was scratched.
The electrodes come off the skin.
The lead wires are disconnected
The power of the batteries has been exhausted.
Brand new or stored batteries
Used Li rechargeable battery has reached end of life
Reposition the electrodes at least 1-1∕2 apart.
Re-position the electrodes. If at any time you feel pain or discomfort stop use immediately.
Ensure the electrodes are stuck securely on the skin.
Clean the electrodes according to description in this manual or replace with new electrodes.
Replace with a new electrode.
Turn off the device and place the electrodes again.
Turn off the device and reconnect the lead wires.
Charge or replace the batteries.
This is normal operation. Please charge and use in device. You must do this 3 – 5 times before full capacity is reached.
Charge the battery. If this does not work, replace the battery.
- 35 -
Page 36
TENSPros
InTENSity IF Combo II Roscoe Medical
8. DISPOSAL
Used fully discharged batteries must be disposed of in a specially labeled collection container, at toxic waste collection points or through an electrical retailer. You are under legal obligation to dispose of batteries correctly.
Please dispose of the device in accordance with the legal obligation.
9. GLOSSARY OF SYMBOLS
Electrical devices are recyclable material and should not
Type BF Applied Part
Please refer to instruction manual because of the higher
be disposed of with household waste after their useful life! Help us to protect the environment and save resources and take this device to the appropriate collection points. Please contact the organization which is responsible for waste disposal in your area if you have any questions.
levels of output.
- 36 -
Page 37
TENSPros
Roscoe Medical InTENSity IF Combo II
10. ELECTROMAGNETIC COMPATIBILITY (EMC) TABLES
Information for accompanying documents in the scope of
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC’s and mobile (cellular) telephones, medical devices in use may be susceptible to electromagnetic interference from other devices. Electromagnetic interference may result in incorrect operation of the medical device and create a potentially unsafe situation. Medical devices should also not interfere with other devices. In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with the aim to prevent unsafe product situations, the EN60601-1-2 standard has been implemented. This standard denes the levels of immunity to electromagnetic interferences as well as maximum levels of electromagnetic emissions for medical devices.
Medical devices manufactured for Roscoe Medical Inc. conform to this EN60601-1-2:2007 standard for both immunity and emissions. Nevertheless, special precautions need to be observed:
• The use of accessories and cables other than those specied by Roscoe Medical, with the exception of cables sold by Roscoe Medical as replacement parts for internal components, may result in increased emission or decreased immunity of the device.
• The medical devices should not be used adjacent to or stacked with other equipment. In case adjacent or stacked use is necessary, the medical device should be observed to verify normal operation in the conguration in which it will be used.
• Refer to further guidance below regarding the EMC environment in which the device should be used.
IEC60601-1-2:2007
- 37 -
Page 38
TENSPros
InTENSity IF Combo II Roscoe Medical
Guidance and manufacturer’s declaration — electromagnetic emissions
The device is intended for use in the electromagnetic environment
specied below. The customer or the user should assure that it is used
Emissions Test Compliance
RF emissions CISPR 11
RF emissions CISPR11
Harmonic emissions lEC 61000-3-2
Voltage uctuations / icker emissions lEC 61000-3-3
in such an environment.
Group 1
Class B
Not applicable
Not applicable
Electromagnetic Environment
— Guidance
The device uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
The device is suitable for use in all establishments including domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
- 38 -
Page 39
TENSPros
Roscoe Medical InTENSity IF Combo II
Guidance And Manufacturer’s Declaration — Electromagnetic Immunity
The device is intended for use in the electromagnetic environment
specied below. The customer or the user should assure that it is used
Immunity Test
Electrostatic discharge (ESD) lEC 61000-4-2
Electrical fast transient/ burst IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short interruptions and voltage varia­tions on power supply IEC 61000-4- 11
Power frequency (50/ 60 Hz) mag­netic eld IEC 61000-4-8
in such an environment.
EC 60601
Test Level
±6 kV contact ±8 kV air
Not applicable Not applicable Not applicable
Not applicable Not applicable Not applicable
Not applicable Not applicable Not applicable
3 A/m 3 A/m
Compliance
Level
±6 kV contact ±8 kV air
Electromagnetic
Environment
— Guidance
Floors should be wood, concrete or ceramic tile. If oors are covered with synthetic material, the relative humidity should be at least 30 %.
Power frequency magnetic elds should be at levels characteristic of a typical location in a typical commercial or hospital environment.
- 39 -
Page 40
TENSPros
InTENSity IF Combo II Roscoe Medical
Guidance And Manufacturer’s Declaration — Electromagnetic Immunity
Immunity
Test
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
EC 60601
Test Level
Not applicable
3 V/m 80 MHz to
2.5 GHz
(Continued)
Compliance
Level
3 V/m
Electromagnetic
Environment
— Guidance
Portable and mobile RF communications equipment should be used no closer to any part of the Electrical Stimulator including cables, than the recommended separation distance calculated from the equation appropriate to the frequency of the transmitter.
Recommend separation distance
d = 1.2 √P d = 1.2 √P 80 MHz to 800 MHz d = 2.3 √P 800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to he transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from xed RF transmitters as determined by an electromagnetic site survey, (a) should be less than the com-
pliance level in each frequency range.
(b) Interference may occur in the
vicinity of equipment marked with he following symbol:
- 40 -
Page 41
TENSPros
Roscoe Medical InTENSity IF Combo II
Guidance And Manufacturer’s Declaration — Electromagnetic Immunity
Note1: At 80 MHz and 800 MHz, the higher frequency range applies. Note2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reection from structures, objects, and people.
a Field strengths from xed transmitters, such as base stations for radio
(cellular/ cordless) telephones and land mobile radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to xed RF transmitters, an electromagnetic site survey should be considered. If the measured eld strength in the location in which the Electrical Stimulator are used exceeds the applicable RF compliance level above, the Electrical Stimulator should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Electrical Stimulator.
b Over the frequency range 150 kHz to 80MHz, eld strengths should be
less than 3 V/m.
(Continued)
- 41 -
Page 42
TENSPros
InTENSity IF Combo II Roscoe Medical
Recommended separation distance between portable and mobile RF
communications equipment and the Electrical Stimulator
The Electrical Stimulator are intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customers or the users of these Electrical Stimulator can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Electrical Stimulator as recommended below, according to the maximum output power of the communications equipment.
Separation distance according to frequency of
Output Power of
Transmitter in
Watt
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.78
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. Note: At 80MHz and 800MHz, the separation distance for the higher pole
frequency range applies
Note: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reection from structures, objects, and people.
150 kHz to
80 MHz
d = 1.2 √P
transmitter in meter
80 MHz to
800 MHz
d = 1.2 √P
800 MHz to
2.5GHz
d = 2.3 √P
Note: EMC tests conducted including attached electrode cord
of 1.5 m length.
- 42 -
Page 43
TENSPros
Roscoe Medical InTENSity IF Combo II
11. WARRANTY
Please contact your dealer or the device center in case of a claim under the warranty. If you have to send in the unit, enclose a copy of your receipt and state what the defect is. The following warranty terms apply:
1) The warranty period for device is one years from date of purchase. In case of a warranty claim, the date of purchase has to be proven by means of the sales receipt or invoice.
2) Repairs under warranty do not extend the warranty period either for the device or for the replacement parts.
3) The following is excluded under the warranty:
• All damage which has arisen due to improper treatment, e.g.
nonobservance of the user instruction.
• All damage which is due to repairs or tampering by the customer or
unauthorized third parties.
• Damage which has arisen during transport from the manufacturer to
the consumer or during transport to the service center.
• Accessories which are subject to normal wear and tear.
4) Liability for direct or indirect consequential losses caused by the unit is excluded even if the damage to the unit is accepted as a warranty claim.
Note:
Shelf life is most inuenced by several factors: exposure to light
and heat, transmission of gases (including humidity), and mechanical stresses, this device and accessories does not require sterilization. The device is supplied to be used under non-sterile conditions, Material is not degraded phenomenon, also won’t produce volatile phenomenon, this device has not restricted shelf-life.
- 43 -
Page 44
TENSPros
Manufactured for:
Roscoe Medical
21973 Commerce Parkway
Strongsville, Ohio 44149
Ph: (800) 3-ROSCOE (376-7263)
www.roscoemedical.com
Loading...