Roscid Technologies RO120 Specifications

indu
strial applications. Roscid also offers a wall mount and duct mount configuration.
FEATURES
·
RO120 Series
Humidity and Temperature Transmitter
Figure1: RO120-DIS
Dew point: -60°C to +95°C (factory default) for remote units RH 0-95%RH (non-condensing) Calibration performed down to -60C Speed Response: < 12second response time that is T63, 0 to 90% Pressure Rating: 750PSI Dew point accuracy:
+/- 1.0C from +40°C to -20°C
+/- 3.0C <-20C Three 4-20mA outputs or a RS232 (see configuration table) Temperature range: -40°C to +120°C
APPLICATIONS
ENVIRONMENTAL CHAMBERS
· DRYERS
· COMPRESSED AIR
· PHARMACEUTICAL
· HVAC
· FURNACE AND INDUSTRIAL APPLICATIONS
· SEMICONDUCTOR
· DATA STORAGE
Figure 2: RO120
The RO120 Series
The RO120 Series is a precision dew point and temperature transmitter that provides quick
and accurate humidity, barometric pressure, temperature, absolute pressure and dew point measurements in a wide range of applications. It has a wide temperature range of -40°C to 120°C and is resistant to most chemicals. This remote type transmitter comes standard with a 6-foot cable (except for wall mount (W) units) and is ideally suited for many different applications such as environmental chambers, HVAC, compressed air, furnace and
www.roscid.com +1 (781) 933-4007 info@roscid.com
Technical Information
RO120 Series
Humidity and Temperature Transmitter
Measurement Ranges Dew Point: -60 °C to +95 °C
Temperature: -40 °C to +120°C (remote and duct mount only)
Probe Material 316 Stainless Steel (for remote and duct mount only)
Display 2x16 LCD Display, Shows three user selectable parameters.
Fitting 3/4 NPT Included (remote only)
Resolution 0.1 %
Output 4-20mA or RS232
Relative Humidity: 0-100%
Enclosure Dimensions
Cable Length
Supply Voltage
Sensor
Analog Output scaled range Dew Point (4-20mA)
Analog Output scaled range Relative Humidity (4-
20mA)
Analog Output scaled range temperature °C (4-
20mA)
Analog Outputs Pressure range (for units with
pressure transducer) Psia
Analog Outputs Barometric Pressure Range (for
units with pressure transducer) mbar
Operating Temperature
Accuracy
4.88L X 3.70 W X 2.20 H
6 feet is standard. Contact factory for custom lengths. (For remote and duct mount units only).
18 to 30Vdc, unregulated, 200 mA max
Polymer capacitive sensor
-60 °C to +40 °C
0 to 100% RH
-20°C to +120°C
0-200psia
750mb to 1100mb
-25 °C to 200 °C
RH: ±1% H: ±1% over T range -20... 120C, +/- 0.6% over T
range 15...40C
Temperature: ±0.5°C over T range -20... 120C, +/- 0.2C over
T range 15...40C
www.roscid.com +1 (781) 933-4007 info@roscid.com