Roscid Technologies OxyTrans Specifications

www.roscid.com +1 (781) 933-4007 info@roscid.com
Oxytrans II
The Oxytran II Series
4 – 20 mA Loop Powered two wire Oxygen Transmitter
The Model OxyTrans is simple, accurate and economical way to measure traces and % ranges of Oxygen. The unit has a simple display and offers a one point calibration allowing for a quick start up. Optional EX (explosion proof) 1/2 G Ex ia IIC T4 (ATEX).
There are 2 types of units, one for the PPM range and one for the percent range. The PPM range offered is 0-10000ppm (model OXYT-II-T) and the % unit is offered in three different incremental ranges. First range is 0-0.25, 0-2.5 and 0-25% (model OXYT-II-P1), second range is 0-0.5, 0-5, and 0-50% (model OXYT-II-P2) and the third range is 0-1%, 0-10% and 0-100% (model OXYT-II-P3).
FEATURES
Compact
Inexpensive User selectable ranges
Easy to replace sensor
IP65 enclosure Reverse voltage protection and temperature compensation No cross sensitivity to hydrogen Zero calibration for measurement in low ppm range 4 measurement ranges in one unit
APPLICATIONS
SEMICONDUCTOR INDUSTRY
GAS MANUFACTURERS
ENVIRONMENTAL CHAMBERS
METALLURGICAL INDUSTRIES
CHEMICAL INDUSTRY
GLOVEBOX APPLICATIONS
SEMICONDUCTOR
HEAT TREATING AND BRIGHT ANNEALING
Figure 1: Oxytrans-II
www.roscid.com +1 (781) 933-4007 info@roscid.com
Oxytrans II
Measurement Ranges
0-10, 0-100, 0–1000, 0 - 10000 ppm O
2
0 – 0.25, 0- 2.5, 0- 25 %O
2
0 – 0.5, 0- 5, 0- 50 %O
2
0 – 1, 0- 10, 0- 100 %O2
Display
6 Digits, alphanumeric with bar graph
Calibration
Use a known ppm of oxygen in a gas (typical Nitrogen0
For Percent ranges uses air
Resolution
0.1 ppm < 0-100 ppm
1 ppm at 0- 1000, 0-10000ppm
0.1% for 25%
Digital Output
HART Communication
Load
Typical 470 ohm
Pressure
Max 10bar (flow through)
Supply Voltage
10 – 35 VDC
reverse voltage protection
up to 40VDC
Sensor
Micro-Fuel Cell
Alarm Value
3.8 mA (Standard) or 23 mA (if desired)
Signal Output
4-20mA/DC
Operating Temperature
0°C to +50 °C
Weight
1.2kgs
Sample Flow Rate
1-3 scfh
Enclosure Rating
IP65
Process Connection
¼” tube
Accuracy
+/- 2% FSD T= const. +/- 5% FSD 0>T>50°C