O2Tracer
4 – 20 mA Loop Powered two wire
Oxygen Transmitter
Two Wire Process Oxygen analyzer
The Model O2Tracer is simple to use, accurate and
very economical when measuring Oxygen at %
levels.
One User specific measuring range the
unit will offer with different ranges between
0 – 100% of oxygen. Typical range is 0-
25%
Sensor
The O2Tracer uses a special fuel cell to measure the
oxygen concentration. The sensor meets the
industrial requirements for accuracy, sensitivity, easy
to use and operating life.
Calibration
The calibration of the instrument for trace oxygen
measurements in gas should be done with a
calibration gas. The concentration can be chosen
freely within the measuring range.
In the percent range the unit can calibrated with air.
Features
compact
inexpensive
IP65 enclosure
reverse voltage protection and temperature
compensation
The flow will be done through the
electronic enclosure
www.roscid.com
Specification
Measuring
ranges
offered:
Calibration : with calibration gas or
Ambient air
Accuracy : +/- 2% FSD T= const.
+/- 5% FSD 0>T>50°C
Response time : 90 % FSD at 25°C
0 – 100% < 10 s
Operating : 0 - 50°C
Temperature
Pressure : 0.1 - 1 bar
Signal output : 4 -20 mA / DC
Voltage : 10 – 35 VDC
reverse voltage protection
up to 40VDC
load : typ. 1000
Ohm, max. at
28Vdc
-max 250 ohms
at 10Vdc
Display : no Display
0-5%, 0-25%
0 – 100 % O
Applications are found in
2
•
Air-, oxygen- and nitrogen generators
•
Gas manufacturers – filling stations
For Ordering:
Ordering Number Range
oxygen sensor : Micro-Fuel Cell,
housing : IP65
Size : 3.15 x 3 x 2.17 ( B x H x T (mm))
Enclosure : aluminum
Connection : ¼” tubing
Weight : 0.6 lb.
www.roscid.com