Roscid Technologies O2Tracer Specifications

O2Tracer
4 – 20 mA Loop Powered two wire Oxygen Transmitter
Two Wire Process Oxygen analyzer
One User specific measuring range the
unit will offer with different ranges between 0 – 100% of oxygen. Typical range is 0- 25%
Sensor
The O2Tracer uses a special fuel cell to measure the oxygen concentration. The sensor meets the industrial requirements for accuracy, sensitivity, easy to use and operating life.
Calibration
The calibration of the instrument for trace oxygen measurements in gas should be done with a calibration gas. The concentration can be chosen freely within the measuring range. In the percent range the unit can calibrated with air.
Features
compact inexpensive IP65 enclosure reverse voltage protection and temperature
compensation
The flow will be done through the
electronic enclosure
www.roscid.com
O2T-1
0-10%
O2T-2
0-25%
O2T-3
0-100%
Specification
Measuring ranges offered:
Calibration : with calibration gas or
Ambient air
Accuracy : +/- 2% FSD T= const.
+/- 5% FSD 0>T>50°C
Response time : 90 % FSD at 25°C
0 – 100% < 10 s
Operating : 0 - 50°C Temperature
Pressure : 0.1 - 1 bar Signal output : 4 -20 mA / DC Voltage : 10 – 35 VDC
reverse voltage protection up to 40VDC
load : typ. 1000
Ohm, max. at 28Vdc
-max 250 ohms at 10Vdc
Display : no Display
0-5%, 0-25%
0 – 100 % O
Applications are found in
2
Air-, oxygen- and nitrogen generators
Gas manufacturers – filling stations
For Ordering:
Ordering Number Range
oxygen sensor : Micro-Fuel Cell, housing : IP65 Size : 3.15 x 3 x 2.17 ( B x H x T (mm)) Enclosure : aluminum Connection : ¼ tubing Weight : 0.6 lb.
www.roscid.com