This application sheet describes the use of an R&S ZVA vector analyzer and two R&S
ZVA-Z110 frequency converters for 2-port transmission measurements.
Measurements using other converter types are performed in an analogous way.
Basic Operation
Required Equipment
1.1Required Equipment
Measurements with frequency converters can be carried out with the following equipment:
Vector network analyzer (NWA) R&S ZVA or R&S ZVT with an upper frequency limit
●
of 20 GHz or higher (R&S ZVL 20, R&S ZVA 24, R&S ZVA 40 ...).
The required firmware version depends on the frequency converter model:
–R&S ZVA-Z75: at least version V2.47
–R&S ZVA-Z110: at least version V2.20
–R&S ZVA-Z140: at least version V2.71
–R&S ZVA-Z170: at least version V2.71
–R&S ZVA-Z220: at least version V2.72
–R&S ZVA-Z325: at least version V2.45
–R&S ZVA-Z500: at least version V2.80
Two NWA ports per frequency converter port.
●
Alternatively: One NWA port per frequency converter plus a common external generator for the LO signals. The R&S SMF100A signal generator with suitable options
is recommended.
N frequency converters for an N-port measurement
●
Option R&S ZVA<n>-B16, "Direct Generator/Receiver Access" at each port
●
Option R&S ZVA-K8, "Converter Control"
●
a suitable set of calibration standards
●
1.2Measurement Principle
The frequency converters use frequency multipliers to transform the RF source signal
from one of the network analyzer ports into a high-frequency stimulus signal. A second
signal (Local Oscillator, LO) is used for down-conversion of the reference and measurement channels. The LO signal can be provided either by a second analyzer port or by an
external generator.
The measurement involves the following steps:
1. Selection of the converter and test setup, activation of the converter mode
2. Connection of the frequency converters
3Application Sheet 1ZKD-29, M. Jetter, 09/2010 ─ 04
3. Calibration using a suitable waveguide calibration kit
4. Connection of the DUT and measurement
Basic Operation
Activating the Frequency Converter Mode
1.3Activating the Frequency Converter Mode
To activate the converter mode for a setup without external generator,
1. Click "System > System Config ..." and open the "Frequency Converter" tab of the
"System Configuration" dialog.
2. Select your converter "Type": R&S ZVA-Z75, R&S ZVA-Z110, R&S ZVA-Z140, R&S
ZVA-Z170, R&S ZVA-Z220, R&S ZVA-Z325, or R&S ZVA-Z500.
3. Select a test setup with an analyzer port as external source, click "Apply" to activate
the frequency converter mode and "Close".
Analyzer settings with active frequency converter
In frequency converter mode, the frequency and level settings of the network analyzer
are automatically set to be compatible with the selected frequency converters. "Low
Phase Noise" is enabled, Automatic Level Control (ALC) is disabled. The frequency and
levels of all ports are displayed in the "Port Configuration" dialog ("Channel > Mode >
Port Config …").
Fig. 1-1: Frequency Converter dialog (example: R&S ZVA-Z110)
4Application Sheet 1ZKD-29, M. Jetter, 09/2010 ─ 04
Each frequency converter must be connected to the analyzer, the power supply and the
DUT.
Analyzer ports: Connect each of the two frequency converters as shown in the "Fre-
●
quency Converter" tab of the "System Configuration" dialog using the cables for the
REF OUT and MEAS OUT signals supplied with the converters and two additional,
high-quality coaxial cables.
Power supply: Connect the external DC power supply provided with the converter to
●
the "9 V / 0.5 A" DC input.
DUT (usually connected after calibration): Screw on the waveguide flange.
●
Refer to the Quick Start Guide of your converter for details.
1.5Calibration
The output power of the frequency converter can be set manually (at the converter) only,
therefore the standard source power calibration eliminating frequency response errors in
the signal path between the source and the reference plane (external power meter) is not
possible.
A power calibration of the reference receiver (a-wave) using an appropriate external
power meter, however, is possible and recommended for measurements concerning the
wave quantities a and b. Waveguide power meters are configured in the ordinary way
using the "System Configuration External Power Meters" tab. Proceed as follows:
1. Ensure that the output power of the frequency converter is not attenuated (adjust the
adjusting screw (knurled knob) at the top of the converter to 2 mm).
2. Connect the waveguide power meter to the converter's waveguide flange and open
the "Channel > Calibration > Start Power Cal > Source Power Cal" dialog.
This power calibration procedure ensures a reasonable accuracy of the reference power
readings over a wide range of converter output powers (i.e. even if the adjusting screw
is used to reduce the powers).
A receiver power calibration of the b-waves (without external power meter, using the
“Receiver Power Calibration” dialog) is possible after completed power calibration of the
a-wave.
After the power calibration procedure a system error correction is recommended. Due to
the physical properties of the mm-waves and the waveguides, measurements with frequency converters require a special calibration kit for system error correction.
5Application Sheet 1ZKD-29, M. Jetter, 09/2010 ─ 04
Calibration kit R&S ZV-WR15, for converter R&S ZVA-Z75
●
Calibration kit R&S ZV-WR10, for converter R&S ZVA-Z110
●
Calibration kit R&S ZV-WR08, for converter R&S ZVA-Z140
●
Calibration kit R&S ZV-WR06, for converter R&S ZVA-Z170
●
Calibration kit R&S ZV-WR05, for converter R&S ZVA-Z220
●
Calibration kit R&S ZV-WR03, for converter R&S ZVA-Z325
●
Calibration kit R&S ZV-WR02, for converter R&S ZVA-Z500
●
The standards in the calibration kit allow all one-port and two-port calibration types supported by the network analyzer except TNA. Refer to the documentation of the calibration
kit or the help system of your network analyzer for details.
Basic Operation
Measurement
1.6Measurement
After power calibration and system error correction, the mm-wave measurement can be
performed like any other network analyzer measurement. The "Port Configuration" settings (together with the "Stimulus" settings), determine the sweep range of the converted
signals (i.e. the input and output frequencies at the DUT ports). All measured quantities
(S-parameters, wave quantities, ratios etc.) and other trace settings are available.
The following restrictions hold for measurements with external frequency converters:
The measurement is performed at fixed RF source and LO power. No power sweep
●
is possible.
To adjust the actual output power of the converters (e.g. for measuring wave quan-
●
tities or testing compression effects), use the adjusting screw on top of the converters.
Frequency converters R&S ZVA-Z500 do not provide an adjusting screw.
The following example shows the transmission and reflection coefficients of a bandpass
filter in the frequency range between 75 GHz and 110 GHz, which is covered by the
frequency converter R&S ZVA-Z110.
6Application Sheet 1ZKD-29, M. Jetter, 09/2010 ─ 04
Fig. 1-2: Transmission and reflection of a bandpass filter measured with an R&S ZVA-Z110
1.7Additional Information
For a comprehensive description of the frequency converter mode including remote control refer to the R&S ZVA/ZVT online help system or to the printable operating manual,
which is available for download at http://www.rohde-schwarz.com/product/zva.
For an description of the R&S ZVA-Z110 connectors and safety instructions refer to the
R&S ZVA-Z110 Quick Start Guide.
Application notes related to the frequency converter are also available for download, see
http://www.rohde-schwarz.com/product/zva-z.
7Application Sheet 1ZKD-29, M. Jetter, 09/2010 ─ 04
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.