The product documentation helps you use the product safely and efficiently. Follow the
instructions provided here and in the following chapters.
Intended use
The product is intended for the development, production and verification of electronic
components and devices in industrial, administrative, and laboratory environments.
Use the product only for its designated purpose. Observe the operating conditions and
performance limits stated in the data sheet.
Where do I find safety information?
Safety information is part of the product documentation. It warns you of potential dangers and gives instructions on how to prevent personal injury or damage caused by
dangerous situations. Safety information is provided as follows:
●
In Chapter 1.1, "Safety instructions", on page 7. The same information is provided in many languages as printed "Safety Instructions". The printed "Safety
Instructions" are delivered with the product.
●
Throughout the documentation, safety instructions are provided when you need to
take care during setup or operation.
1.1Safety instructions
Products from the Rohde & Schwarz group of companies are manufactured according
to the highest technical standards. To use the products safely, follow the instructions
provided here and in the product documentation. Keep the product documentation
nearby and offer it to other users.
Use the product only for its intended use and within its performance limits. Intended
use and limits are described in the product documentation such as the data sheet,
manuals and the printed "Safety Instructions". If you are unsure about the appropriate
use, contact Rohde & Schwarz customer service.
Using the product requires specialists or specially trained personnel. These users also
need sound knowledge of at least one of the languages in which the user interfaces
and the product documentation are available.
Never open the casing of the product. Only service personnel authorized by
Rohde & Schwarz are allowed to repair the product. If any part of the product is damaged or broken, stop using the product. Contact Rohde & Schwarz customer service at
http://www.customersupport.rohde-schwarz.com.
Lifting and carrying the product
The maximum weight of the product is provided in the data sheet. To move the product
safely, you can use lifting or transporting equipment such as lift trucks and forklifts. Follow the instructions provided by the equipment manufacturer.
7Getting Started 1316.2407.02 ─ 62
R&S®ZND
Safety and regulatory information
Safety instructions
Choosing the operating site
Only use the product indoors. The product casing is not waterproof. Water that enters
can electrically connect the casing with live parts, which can lead to electric shock,
serious personal injury or death if you touch the casing. If Rohde & Schwarz provides
accessories designed for your product, e.g. a carrying bag, you can use the product
outdoors.
Unless otherwise specified, you can operate the product up to an altitude of 2000 m
above sea level. The product is suitable for pollution degree 2 environments where
nonconductive contamination can occur. For more information on environmental conditions such as ambient temperature and humidity, see the data sheet.
Setting up the product
Always place the product on a stable, flat and level surface with the bottom of the product facing down. If the product is designed for different positions, secure the product so
that it cannot fall over.
If the product has foldable feet, always fold the feet completely in or out to ensure stability. The feet can collapse if they are not folded out completely or if the product is
moved without lifting it. The foldable feet are designed to carry the weight of the product, but not an extra load.
If stacking is possible, keep in mind that a stack of products can fall over and cause
injury.
If you mount products in a rack, ensure that the rack has sufficient load capacity and
stability. Observe the specifications of the rack manufacturer. Always install the products from the bottom shelf to the top shelf so that the rack stands securely. Secure the
product so that it cannot fall off the rack.
Connecting to power
The product is an overvoltage category II product. Connect the product to a fixed
installation used to supply energy-consuming equipment such as household appliances and similar loads. Keep in mind that electrically powered products have risks, such
as electric shock, fire, personal injury or even death.
Take the following measures for your safety:
●
Before switching on the product, ensure that the voltage and frequency indicated
on the product match the available power source. If the power adapter does not
adjust automatically, set the correct value and check the rating of the fuse.
●
Only use the power cable delivered with the product. It complies with country-specific safety requirements. Only insert the plug into an outlet with protective conductor terminal.
●
Only use intact cables and route them carefully so that they cannot be damaged.
Check the power cables regularly to ensure that they are undamaged. Also ensure
that nobody can trip over loose cables.
●
If the product needs an external power supply, use the power supply that is delivered with the product or that is recommended in the product documentation or a
power supply that conforms to the country-specific regulations.
8Getting Started 1316.2407.02 ─ 62
R&S®ZND
Safety and regulatory information
Warning messages in the documentation
●
Only connect the product to a power source with a fuse protection of maximum
20 A.
●
Ensure that you can disconnect the product from the power source at any time.
Pull the power plug to disconnect the product. The power plug must be easily
accessible. If the product is integrated into a system that does not meet these
requirements, provide an easily accessible circuit breaker at the system level.
Cleaning the product
Use a dry, lint-free cloth to clean the product. When cleaning, keep in mind that the
casing is not waterproof. Do not use liquid cleaning agents.
Meaning of safety labels
Safety labels on the product warn against potential hazards.
Potential hazard
Read the product documentation to avoid personal injury or product damage.
Electrical hazard
Indicates live parts. Risk of electric shock, fire, personal injury or even death.
Hot surface
Do not touch. Risk of skin burns. Risk of fire.
Protective conductor terminal
Connect this terminal to a grounded external conductor or to protective ground. This connection protects you against electric shock if an electric problem occurs.
1.2Warning messages in the documentation
A warning message points out a risk or danger that you need to be aware of. The signal word indicates the severity of the safety hazard and how likely it will occur if you do
not follow the safety precautions.
WARNING
Potentially hazardous situation. Could result in death or serious injury if not avoided.
NOTICE
Potential risks of damage. Could result in damage to the supported product or to other
property.
9Getting Started 1316.2407.02 ─ 62
R&S®ZND
Safety and regulatory information
Korea certification class A
1.3Korea certification class A
이 기기는 업무용(A급) 전자파 적합기기로서 판매자 또는 사용자는 이 점을 주의하시기
바라며, 가정외의 지역에서 사용하는 것을 목적으로 합니다.
10Getting Started 1316.2407.02 ─ 62
R&S®ZND
Documentation overview
Service manual
2Documentation overview
This section provides an overview of the R&S ZND user documentation. Unless specified otherwise, you find the documents on the R&S ZND product page at:
https://www.rohde-schwarz.com/manual/ZND
2.1Getting started manual
Introduces the R&S ZND and describes how to set up and start working with the product. Includes basic operations, typical measurement examples, and general information, e.g. safety instructions, etc.
A printed version is delivered with the instrument. A PDF version is available for download on the Internet.
2.2User manual and help
The user manual contains the description of all instrument modes and functions. It also
provides an introduction to remote control, a complete description of the remote control
commands with programming examples, and information on maintenance, instrument
interfaces and error messages. Includes the contents of the getting started manual.
The contents of the user manual are available as help in the R&S ZND. The help offers
quick, context-sensitive access to the complete information for the instrument and its
firmware.
The user manual is also available for download or for immediate display on the Internet.
2.3Service manual
Describes the performance test for checking the rated specifications, module replacement and repair, firmware update, troubleshooting and fault elimination, and contains
mechanical drawings and spare part lists.
The service manual is available for registered users on the global Rohde & Schwarz
information system (GLORIS):
https://gloris.rohde-schwarz.com
11Getting Started 1316.2407.02 ─ 62
R&S®ZND
Documentation overview
Application notes, application cards, white papers, etc.
2.4Instrument security procedures
Deals with security issues when working with the R&S ZND in secure areas. It is available for download on the Internet.
2.5Printed safety instructions
Provides safety information in many languages. The printed document is delivered with
the product.
2.6Data sheets and brochures
The data sheet contains the technical specifications of the R&S ZND. It also lists the
firmware applications and their order numbers, and optional accessories.
The brochure provides an overview of the instrument and deals with the specific characteristics.
See https://www.rohde-schwarz.com/brochure-datasheet/ZND
2.7Release notes and open source acknowledgment
(OSA)
The release notes list new features, improvements and known issues of the current
firmware version, and describe the firmware installation.
The open-source acknowledgment document provides verbatim license texts of the
used open source software.
2.8Application notes, application cards, white papers,
etc.
These documents deal with special applications or background information on particular topics.
See https://www.rohde-schwarz.com/brochure-datasheet/ZND
12Getting Started 1316.2407.02 ─ 62
R&S®ZND
Preparing for use
Choosing the operating site
3Preparing for use
Here, you can find basic information about setting up the product for the first time.
3.1Lifting and carrying
The carrying handles are designed to lift or carry the instrument. Do not apply excessive external force to the handles.
See "Lifting and carrying the product"on page 7.
3.2Unpacking and checking
1. Unpack the R&S ZND carefully.
2. Retain the original packing material. Use it when transporting or shipping the
R&S ZND later.
3. Using the delivery notes, check the equipment for completeness.
4. Check the equipment for damage.
If the delivery is incomplete or equipment is damaged, contact Rohde & Schwarz.
3.3Choosing the operating site
Specific operating conditions ensure proper operation and avoid damage to the product and connected devices. For information on environmental conditions such as ambient temperature and humidity, see the data sheet.
See also "Choosing the operating site"on page 8.
Electromagnetic compatibility classes
The electromagnetic compatibility (EMC) class indicates where you can operate the
product. The EMC class of the product is given in the data sheet.
●
Class B equipment is suitable for use in:
–Residential environments
–Environments that are directly connected to a low-voltage supply network that
supplies residential buildings
●
Class A equipment is intended for use in industrial environments. It can cause
radio disturbances in residential environments due to possible conducted and radiated disturbances. It is therefore not suitable for class B environments.
13Getting Started 1316.2407.02 ─ 62
R&S®ZND
If class A equipment causes radio disturbances, take appropriate measures to
eliminate them.
3.4Setting up the product
See also:
●
"Setting up the product"on page 8
●
"Intended use"on page 7
3.4.1Placing the product on a bench top
To place the product on a bench top
1. Place the product on a stable, flat and level surface. Ensure that the surface can
support the weight of the product. For information on the weight, see the data
sheet.
Preparing for use
Setting up the product
CAUTION! Foldable feet can collapse. See "Setting up the product"on page 8.
2.
Always fold the feet completely in or out. With folded-out feet, do not place anything on top or underneath the product.
WARNING! A stack of products can fall over and cause injury. Never stack more
3.
than three products on top of each other. Instead, mount them in a rack.
Stack as follows:
●If the products have foldable feet, fold them in completely.
●It is best if all products have the same dimensions (width and length). If the
products have different dimensions, stack according to size and place the
smallest product on top.
●Do not exceed the permissible total load placed on the product at the bottom of
the stack:
–50 kg when stacking products of identical dimensions (left figure).
–25 kg when stacking smaller products on top (middle figure).
14Getting Started 1316.2407.02 ─ 62
R&S®ZND
Left= Stacked correctly, same dimensions
Middle = Stacked correctly, different dimensions
Right = Stacked incorrectly, too many products
NOTICE! Overheating can damage the product.
4.
Prevent overheating as follows:
●Keep a minimum distance of 10 cm between the fan openings of the product
and any object in the vicinity.
●Do not place the product next to heat-generating equipment such as radiators
or other products.
3.4.2Mounting the product in a rack
To prepare the rack
1. Observe the requirements and instructions in "Setting up the product"on page 8.
Preparing for use
Setting up the product
NOTICE! Insufficient airflow can cause overheating and damage the product.
2.
Design and implement an efficient ventilation concept for the rack.
To mount the R&S ZND in a rack
1. Use an adapter kit to prepare the R&S ZND for rack mounting.
a) Order the rack adapter kit R&S ZZA-KN5 (order number 1175.3040.00).
b) Mount the adapter kit. Follow the assembly instructions provided with the
adapter kit.
2. Lift the R&S ZND to shelf height.
3. Grab the handles and push the R&S ZND onto the shelf until the rack brackets fit
closely to the rack.
4. Tighten all screws in the rack brackets with a tightening torque of 1.2 Nm to secure
the R&S ZND in the rack.
To unmount the R&S ZND from a rack
1. Loosen the screws at the rack brackets.
2. Remove the R&S ZND from the rack.
3. If placing the R&S ZND on a bench top again, unmount the adapter kit from the
R&S ZND. Follow the instructions provided with the adapter kit.
15Getting Started 1316.2407.02 ─ 62
R&S®ZND
3.5Considerations for test setup
Cable selection and electromagnetic interference (EMI)
Electromagnetic interference (EMI) can affect the measurement results.
To suppress electromagnetic radiation during operation:
●
Use high-quality shielded cables, for example, double-shielded RF and LAN
cables.
●
Always terminate open cable ends.
●
Ensure that connected external devices comply with EMC regulations.
Regarding cable length and quality, the following requirements have to be met:
Table 3-1: Cable Requirements
Cable Type (Connector)Requirement
RF cables (PORT 1, ..., PORT N)Double shielded
Preparing for use
Considerations for test setup
BNC cables (various)Double shielded
DB-25 (USER PORT)Double shielded
GPIBStandard cable
Handler I/OStandard cable
DVI-D (Monitor)2 ferrite cores
LANAt least CAT6, S/FTP
USBStandard cables, length ≤ 3m
Signal input and output levels
Information on signal levels is provided in the data sheet. Keep the signal levels within
the specified ranges to avoid damage to the product and connected devices.
Preventing electrostatic discharge (ESD)
Electrostatic discharge is most likely to occur when you connect or disconnect a DUT.
NOTICE! Electrostatic discharge can damage the electronic components of the
►
product and the device under test (DUT).
Ground yourself to prevent electrostatic discharge damage:
a) Use a wrist strap and cord to connect yourself to ground.
b) Use a conductive floor mat and heel strap combination.
During operation, if the firmware observes a serious unexpected disturbance (e.g. due
to ESD), it resets all hardware components to ensure proper instrument functioning. It
then restores the user settings to the state before the disturbance and indicates the
foregone hardware reset by an "Hardware communication problem [...]" information
popup.
16Getting Started 1316.2407.02 ─ 62
R&S®ZND
Preparing for use
Switching the instrument on and off
3.6Connecting the analyzer to the AC supply
For safety information, see "Connecting to power"on page 8.
The network analyzer is automatically adapted to the AC supply voltage, which must
be in the range of 100 V to 240 V at 50 Hz to 60 Hz. The AC power connector is located in the upper part of the rear panel (see Chapter 4.2, "Rear panel", on page 29).
1. Plug the AC power cable into the AC power connector on the rear panel of the
product. Only use the AC power cable delivered with the product.
2. Plug the AC power cable into a power outlet with ground contact.
The required ratings are listed next to the AC power connector and in the data
sheet.
3.
If necessary, also ground the product using the grounding terminal
The maximum power consumption and the typical power consumption of the individual
analyzer models are listed in the data sheet.
3.7Switching the instrument on and off
To switch on the R&S ZND
1. Switch the AC power switch to position I (On).
The AC power switch is located in the upper part of the rear panel, together with
the mains connector; see Chapter 4.2, "Rear panel", on page 29.
After power-on, the analyzer automatically goes to standby or ready state, depending on the state of the standby toggle key at the front panel when the instrument
was switched off last time.
2. If necessary, press the standby toggle key on the front panel to switch the instrument to ready state.
See Chapter 3.8, "Standby and ready state", on page 18
.
The instrument automatically performs a system check, boots the Windows® operating system and then starts the vector network analyzer (VNA) application. If it
was terminated regularly, the VNA application restores all recall sets and instrument settings of the previous analyzer session.
To shut down the instrument
1. Press the standby toggle key.
Pressing the standby toggle key causes the instrument to save all loaded recall
sets, to close the VNA application, to shut down Windows®, and to go to standby
state (see Chapter 3.8, "Standby and ready state", on page 18).
17Getting Started 1316.2407.02 ─ 62
R&S®ZND
Of course, you can also perform these steps manually, like in any Windows session.
2. If desired, set the AC power switch to position O (Off).
To disconnect from power
The R&S ZND is in standby state.
NOTICE! Risk of data loss. If you disconnect the product from power when it is in
1.
the ready state, you can lose settings and data. Shut it down first.
Set the switch on the power supply to position [0].
The LED of the standby toggle key key is switched off.
2. Disconnect the R&S ZND from the power source.
3.8Standby and ready state
Preparing for use
Windows operating system
The standby toggle key is located in the bottom-left corner of the front panel.
In standby state, the right, amber LED is on. The standby power only supplies the
power switch circuits. In this state, it is safe to switch the AC power off and disconnect
the instrument from the power supply.
In ready state, the left, green LED is on and all modules are power-supplied. When
switched to ready state, the analyzer initiates its startup procedure.
Observe the instructions for startup and shutdown in Chapter 3.7, "Switching the instru-
ment on and off", on page 17.
3.9Windows operating system
The analyzer is equipped with a Windows® operating system which has been configured according to the instrument's features and needs. Changes in the system configuration can be necessary to:
●
Establish a LAN connection
●
Customize the properties of the external accessories connected to the analyzer
●
Call up additional software tools
Modifications of the operating system
The operating system is adapted to the network analyzer. To avoid impairment of
instrument functions, only change the settings described in this manual. Existing software must be modified only with update software released by Rohde & Schwarz. Likewise, only programs authorized by Rohde & Schwarz for use on the instrument must
be executed.
18Getting Started 1316.2407.02 ─ 62
R&S®ZND
Preparing for use
Windows operating system
All necessary settings can be accessed from the Windows "Start" menu, in particular
from the "Control Panel". To open the "Start" menu, press the Windows key in the toolbar or on an external keyboard.
User accounts and password protection
The analyzer uses a user name and password as credentials for remote access. Two
user accounts with different levels of access are available on the instrument:
●
"instrument" is the default account with standard rights to change system settings.
Use this account for normal operation of the analyzer.
●
"Admin" is the account for administering the operating system. This account is
required, for instance, if you wish to install programs on the analyzer.
[For instruments running Windows 7, this account is named "Administrator".]
In the factory configuration, "894129" is preset as a password for both users. To protect
the analyzer from unauthorized access, it is recommended to change the preset passwords.
To switch from one user account to another, log off from Windows and then log on
again. The "switch user" functionality is disabled on the R&S ZND.
Service packs and updates
Microsoft regularly creates security updates and other patches to protect Windowsbased operating systems. They are released through the Microsoft Update website and
associated update server. Update instruments using Windows regularly, especially
instruments that connect to a network.
Firewall settings
A firewall protects an instrument by preventing unauthorized users from gaining access
to it through a network. Rohde & Schwarz highly recommends using the firewall on
your instrument. Rohde & Schwarz instruments are shipped with the Windows firewall
enabled. All ports and connections for remote control are enabled.
Note that changing firewall settings requires administrator rights.
Virus protection
Take appropriate steps to protect your instruments from infection. Use strong firewall
settings and scan any removable storage device used with a Rohde & Schwarz instrument regularly. It is also recommended that you install anti-virus software on the instrument. Rohde & Schwarz does NOT recommend running anti-virus software in the
background ("on-access" mode) on Windows-based instruments, due to potentially
degrading instrument performance. However, Rohde & Schwarz does recommend running it during non-critical hours.
For details and recommendations, see the following Rohde & Schwarz white paper:
●
1EF96: Malware Protection Windows 10
●
Windows 7: 1DC01: Malware Protection Windows 7
19Getting Started 1316.2407.02 ─ 62
R&S®ZND
3.10Minimizing the VNA application
Preparing for use
Connecting external accessories
With a minimized VNA application, you can access your analyzer's Windows® desktop
or run other applications.
To exit the default full-screen mode of the VNA application, deselect SYSTEM – [DISPLAY] > "View Bar" > "Title Bar Task Bar". Then use the standard Windows® titlebar
functions to minimize/maximize/close the application window.
To start the VNA application with a minimized window
1. Right-click the Vector Network Analyzer shortcut icon on the desktop and open the
"Properties" dialog.
2. In the "Shortcut" tab, select "Run: Minimized".
A software update restores the original shortcut properties.
3.11Connecting external accessories
The analyzer's standard PC interfaces (Monitor, USB, LAN) can be used to connect
various accessories:
●
An external monitor expands/displays the Windows® desktop, which is, by default,
covered by the vector network analyzer (VNA) (VNA) application window in fullscreen mode.
●
External keyboard and mouse simplify local control, in particular manual (GUI)
operation of the VNA application.
●
A printer can be used to create hard copies of the measurement diagrams and
traces from within the VNA application.
20Getting Started 1316.2407.02 ─ 62
R&S®ZND
3.11.1Connecting a monitor
Preparing for use
Connecting external accessories
●
A LAN connection can be established to access the analyzer's mass storage or
control the analyzer from an external PC.
External monitor, keyboard and mouse are not required for local operation. A
R&S ZND can be fully controlled by tapping the touchscreen and front panel keys.
A standard monitor can be connected to the DVI-D connector on the rear panel of the
R&S ZND. No extra configuration is required.
Safety aspects
The monitor must be connected while the instrument is switched off (or in standby
mode). Otherwise correct operation cannot be guaranteed.
Select SYSTEM > [DISPLAY] > "View Bar" > "Hard Key Panel On" from the menu bar
of the VNA application window to add the (virtual) "Hard Key Panel" to the application
window.
3.11.2Connecting a keyboard
A keyboard can be connected to any of the USB connectors. After being auto-detected
by the operating system, it can safely be disconnected and reconnected even during
measurements.
Keyboard configuration
The default input language is English – US. Select "Settings" > "Time & language" >
"Region & language" from the Windows® Start menu to manage languages and keyboards.
Windows 7: select "Control Panel" > "Clock, Language, and Region" > "Region and
Language" > "Keyboards and Languages" from the Windows® Start menu to configure
the keyboard properties.
To access Windows®, use the button in the toolbar of the application window.
3.11.3Connecting a mouse
A USB mouse can be connected to any of the USB connectors. After being auto-detected by the operating system, it can safely be disconnected and reconnected even during measurements.
21Getting Started 1316.2407.02 ─ 62
R&S®ZND
Preparing for use
Connecting external accessories
Mouse configuration
Select "Settings" > "Devices" > "Mouse & touchpad" from the Windows® "Start" menu
to configure the mouse properties.
Windows 7: select "Control Panel" > "Hardware and Sound" > "Devices and Printers" >
"Mouse" from the Windows® "Start" menu to configure the mouse properties.
To access Windows®, use the
3.11.4Connecting a printer
A printer can be connected to any of the USB connectors. After successful installation,
it can safely be disconnected and reconnected even during measurements.
Before printing (SYSTEM – [PRINT]), the analyzer checks whether a printer is connected and turned on and whether the appropriate printer driver is installed.
Printer driver installation
If necessary, the printer driver installation is initiated using the operating system's "Add
Printer Wizard". The wizard is self-explanatory. A printer driver must be installed only
once.
A great variety of printer drivers is available on the analyzer. To obtain the complete
list, select "Settings" > "Devices" > "Printers & scanners" from the Windows® "Start"
menu.
Windows 7: select "Control Panel" > "Hardware and Sound" > "Devices and Printers"
from the Windows® "Start" menu.
button in the toolbar of the application window.
To access Windows®, use the
You can load updated and improved driver versions or new drivers from an installation
disk, USB memory stick or another external storage medium. Alternatively, if the analyzer is integrated in a network, you can install driver data stored in a network directory.
In either case, use the "Add Printer" wizard to complete the installation.
Printer configuration
Use the "Printer Setup" dialog of the firmware (SYSTEM – [PRINT] > "Print...") or the
Windows® printer management to configure the printer properties and printing preferences.
button in the toolbar of the application window.
22Getting Started 1316.2407.02 ─ 62
R&S®ZND
3.11.5Connecting a LAN cable
Preparing for use
Connecting external accessories
A LAN cable can be connected to the LAN connector on the rear panel of the analyzer.
To establish a LAN connection, proceed as follows:
1. Refer to Chapter 7.2.1, "Assigning an IP address", on page 66.
2. Connect a CAT6 or CAT7 LAN cable to the LAN port.
The LAN port of the analyzer is an auto-crossover Ethernet port. You can connect it to
a network, but you can also set up a direct connection to a computer or another instrument. For both connection types, you can use either crossover or straight through
(patch) cables.
The IP address information is displayed in the SYSTEM – [SETUP] > "Remote Settings" softtool tab.
23Getting Started 1316.2407.02 ─ 62
R&S®ZND
Instrument tour
Front panel
4Instrument tour
This chapter gives an overview of the control elements and connectors of the
R&S ZND and gives all information that is necessary to put the instrument into operation and connect external devices.
The meanings of the labels on the product are described in "Meaning of safety labels"
on page 9.
4.1Front panel
The front panel of the network analyzer consists of the touchscreen with the diagrams
and softtool panels (left side), the hardkey area (right side) and the test port area
below. Brief explanations on the controls and connectors, the hardkey area and the
rear panel can be found on the next pages.
Figure 4-1: Front View of R&S ZND
4.1.1Touchscreen
The analyzer is equipped with a 12.1'' XGA color touchscreen. The touchscreen presents all measurement results, mostly in the form of diagrams. Besides, all instrument
functions can be accessed and operated by tapping the control elements on the touchscreen. For an introduction to touchscreen operation, refer to Chapter 5.1, "Manual
operation", on page 31.
24Getting Started 1316.2407.02 ─ 62
R&S®ZND
Instrument tour
Front panel
The following sections contain further useful information about manual control of the
instrument.
●
Refer to the other sections in chapter Chapter 5, "Operating the instrument",
on page 31 to learn how to handle traces and diagrams, and how to use menus,
keys and softtools.
●
Refer to chapter "Concepts and Features" in the user manual or help system to
obtain information about the results in the diagram.
●
Refer to section "Display Menu" in the user manual or help system to learn how to
customize the screen.
●
Refer to the data sheet for the technical specifications of the display.
Screen saver
The screen saver function of the operating system can be used to switch off the display
if the analyzer receives no command for a selectable period of time. The display is
switched on again if any front panel key is pressed.
To enable the screen saver, access the Windows® operating system (e.g. by pressing
the Windows key in the SYSTEM keypad) and tap "Control Panel > Power Options >
Change Plan Settings > Turn off the display".
(Windows 7: "Control Panel > Hardware and Sound > Power Options > Edit Plan Settings > Turn off the display").
4.1.2Function keys
Most of the keys in the TRACE, CHANNEL, STIMULUS, and SYSTEM keypads call up
a related softtool of the analyzer GUI. Every softtool provides access to a group of related measurement settings.
The TRACE keys give access to all trace settings, to the limit check settings, and to
the marker functions including marker search.
●
The [MEAS] settings select the measured and displayed quantity.
●
The [FORMAT] settings define how measured data (traces) are presented.
●
The [SCALE] settings define how traces are scaled.
●
The [TRACE CONFIG] settings store traces to the memory and perform mathematical operations on traces.
●
The [LINES] settings define limits for the measurement results, visualize them in
the diagrams and activate/deactivate the limit check.
●
The [MARKER] settings position markers on a trace, configure their properties and
select the format of the numerical readout. Markers can also be used to locate specific points on the trace, define the sweep range, and scale the diagram.
The CHANNEL keys give access to the hardware-related (channel) settings.
●
The [POWER BW AVG] settings define the power of the internal signal source, the
IF bandwidth, and the sweep average.
●
The [SWEEP] settings define the scope of measurement, including the sweep type
and the number of measured sweeps.
25Getting Started 1316.2407.02 ─ 62
R&S®ZND
Instrument tour
Front panel
●
[CAL] provides all functions that are necessary to perform a system error correction
(calibration).
●
[CHANNEL CONFIG] provides functions for channel management.
●
The [TRIGGER] settings control the start of the measurement sequence.
●
[OFFSET EMBED] provides a selection of length offset parameters to shift the
measurement plane.
The STIMULUS keys ([START], [STOP], [CENTER], [SPAN]) define the sweep range,
depending on the sweep type.
The SYSTEM keys provide general system settings.
●
[FILE] provides standard Windows® functions used to create, save, recall or print
recall sets, to copy the active screen and to shut down the application.
●
The [PRINT] settings control an external printer that is used to print a hardcopy of
the current recall set.
●
[APPLIC] gives access to external software tools and optional extensions of the
analyzer firmware.
●
[DISPLAY] gives access to all display settings and to the functions which activate,
modify and arrange different diagrams.
●
[SETUP] provides general system settings which are not restricted to a particular
recall set.
●
The Windows® key opens the startup menu of the Windows® operating system
from where you can perform system configurations and call up additional software
utilities.
●
[HELP] opens the context-sensitive help system of the analyzer.
●
[PRESET] performs a preset of the instrument settings.
4.1.3Data entry keys
The keys in the DATA ENTRY keypad are used to enter numbers, units, and characters. The data entry keys are only enabled while the cursor is placed on a data input
field in a dialog or in the Help navigation pane.
●
The keys 0 to 9 enter the corresponding numbers.
●
The function of the "." and "–" keys depends on the data type of the active input
field:
–In numeric input fields, the keys enter the decimal point and a negative sign for
the entered numeric value. Multiple entries have not effect.
–In character input fields, the keys enter a dot and a hyphen, respectively. Both
entries can be repeated as often as desired.
●
The function of the four unit keys depends on the data type of the active input field;
see Chapter 5.5, "Entering data", on page 47.
–In numeric input fields, the G/n, M/μ, k/m or x1 keys multiply the entered value
with factors of 10
is equivalent to ENTER and confirms the previous entry.
(-)9
, 10
(-)6
(-)3
, 10
or 1 and add the appropriate physical unit. x1
26Getting Started 1316.2407.02 ─ 62
R&S®ZND
–In character input fields, the G/n, M/μ, k/m keys enter the letters G, M, K,
respectively. x1 is equivalent to ENTER and confirms the previous entry.
●
ESC is used to:
–Cancel entries / close dialogs without activating the entries made (equivalent to
the "Close" button).
–Close the Help system.
●
ENTER is used to:
–Activate the selected active control element, e.g. a button in a dialog or a link in
the "Contents" page of the Help system.
–Confirm selections and entries made and close dialogs (equivalent to the "OK"
button).
–Compress or expand menus or the Help table of contents
●
BACKSPACE deletes the last character before the cursor position or the selected
character sequence or numeric value.
4.1.4Rotary knob
Instrument tour
Front panel
The rotary knob increases and decreases numerical values, scrolls within lists, activates controls and confirms entries. Turning or pressing the rotary knob is equivalent to
the action of the "Cursor Up" and "Cursor Down" navigation keys or of the ENTER key
in the DATA ENTRY keypad, respectively.
STEP SIZE opens an input box to select the steps (in units of the current physical
parameter) between two consecutive numerical values. The step size is also valid for
value changes using the "Cursor Up" and "Cursor Down" keys. See also Chapter 5.5.2,
"Using the numeric editor", on page 48.
4.1.5Navigation keys
The keys in the NAVIGATION keypad are used to navigate within the touchscreen and
the help system, to access and control active elements.
The "Cursor Up" and "Cursor Down" keys are used to:
●
Scroll up and down in lists, e.g. among menu items, in a list of keywords, in the
Help table of contents, or in the Help index
●
Navigate between table rows in diagrams. Press "OK" to toggle between navigation
mode and data entries.
●
Increase and decrease numerical input values
"Cursor Up" (Down) become inactive as soon as the beginning of the list is reached.
"Cursor Up" (Down) is equivalent to a rotation of the rotary knob to the right (left).
The "Cursor Left" and "Cursor Right" keys are used to:
●
Move the cursor to the left or right within input fields
●
Navigate between table columns in diagrams. Press "OK" to toggle between navigation mode and data entries.
27Getting Started 1316.2407.02 ─ 62
R&S®ZND
●
Compress or expand menus or the Help table of contents
●
Move the highlighted item in the menu bar of the active application
ENT OK is equivalent to the action of the ENTER key in the DATA ENTRY keypad.
UNDO reverses the last action, if possible. REDO reverses the action of the UNDO
button.
4.1.6Standby key
The standby toggle switch is located in the bottom-left corner of the front panel.
The key serves two main purposes:
●
Toggle between standby and ready state; see Chapter 3.8, "Standby and ready
state", on page 18.
●
Shut down the instrument; see Chapter 3.7, "Switching the instrument on and off",
on page 17.
Instrument tour
Front panel
4.1.7Front panel connectors
The test ports and four USB connectors are located on the front panel of the analyzer.
4.1.7.1Test ports
Numbered type N female connectors.
The test ports serve as outputs for the RF stimulus signal and as inputs for the measured RF signals from the DUT (response signals).
●
With a single test port, it is possible to generate a stimulus signal and measure the
response signal in reflection. For a measurement example, refer to Chapter 6.2,
"Reflection S-parameter measurement", on page 63.
●
With 2 test ports, full two-port measurements are possible; see Chapter 6.1,
"Transmission S-parameter measurement", on page 56.
However, to use port 2 of a R&S ZND as driving port, you need one of the "bidirectional measurements" options:
–R&S ZND-K5, for a R&S ZND without "extended frequency range" option
R&S ZND-K1
–R&S ZND-K6, for a R&S ZND with "extended frequency range" option
R&S ZND-K1
●
Use a torque wrench when screwing RF cables on the test port connectors.
●
See also Chapter 3.5, "Considerations for test setup", on page 16.
4.1.7.2USB connectors
Four USB 2.0 connectors of type A (master USB).
28Getting Started 1316.2407.02 ─ 62
R&S®ZND
The USB ports can be used to connect:
●
External PC accessories such as mouse or other pointing devices, a keyboard,
printer or external storage device (USB stick, CD-ROM drive etc.).
●
External measurement equipment such as a calibration unit, power meter or switch
matrix.
4.2Rear panel
This section gives an overview of the rear panel controls and connectors of the network analyzer.
Instrument tour
Rear panel
Figure 4-2: Rear Panel R&S ZND
The following connectors are available on all instruments:
●
LAN is an RJ-45 connector. Use this connector to integrate the instrument to a
Local Area Network, primarily for remote control purposes; see Chapter 7.2.1,
"Assigning an IP address", on page 66.
●
Connector DVI can be used to connect an external monitor; see Chapter 3.11.1,
"Connecting a monitor", on page 21.
●
USER PORT is a 25-pin D-Sub connector used as an input and output for low-voltage (3.3 V) TTL control signals.
●
EXT TRIG IN and EXT TRIG OUT are two BNC connectors for 5 V TTL external
trigger signals.
●
REF OUT is a BNC output for the internal reference frequency of the R&S ZND.
Use this connector to synchronize other instruments to the analyzer.
29Getting Started 1316.2407.02 ─ 62
R&S®ZND
Instrument tour
Rear panel
●
REF IN is a BNC input for an external reference frequency. Use this connector to
synchronize the R&S ZND to another device.
●
SYSTEM DRIVE contains the removable system drive of the R&S ZND, containing
all software (including the operating system and the VNA application) and data. Do
not remove the system drive during operation.
Option R&S ZND-B19 provides an additional removable system drive (including
operating system and firmware).
●
The ground connector in the lower left corner of the rear panel provides the ground
of the analyzer's supply voltage. Use this connector for ESD protection; see "Pre-
venting electrostatic discharge (ESD)"on page 16.
The following optional connectors can be installed in addition:
●
Option R&S ZND-B10 provides a GPIB bus connector according to standard IEEE
488 / IEC 625.
Match signals with 50 Ω to comply with EMC directives. See also Chapter 3.5, "Consid-
erations for test setup", on page 16.
30Getting Started 1316.2407.02 ─ 62
R&S®ZND
Operating the instrument
Manual operation
5Operating the instrument
The following sections describe the basics of manual operation, i.e. how to access
instrument functions and settings via the analyzer GUI. Manual operation is particularly
useful for getting to know the instrument and for trouble shooting.
Manual and remote control of the instrument
Manual control of the R&S ZND is possible either via its touchscreen and frontpanel
keys, via locally connected keyboard and mouse (see Chapter 3.11, "Connecting exter-
nal accessories", on page 20), or via Remote Desktop (see also Chapter 7.2, "Remote
operation in a LAN", on page 66). Alternatively it can be remote-controlled via the
GPIB interface or a LAN connection.
To their full extent, manual operation and remote control are described in the GUI Reference and Command Reference chapters of the user manual, respectively. GUI functions and related remote commands are linked bidirectionally. Background information
is provided in the Concepts and Features chapter of the user manual.
5.1Manual operation
The analyzer functions are accessible via several tabbed softtools, each presenting
related functions and settings. The keys on the front panel or the on-screen "Hard Key
Panel" open the most frequently used softtools.
Manual operation via function keys and softtools provides touch-friendly access to the
instrument functions and settings, avoiding complicated menu structures and long
operating sequences. In general, this approach is recommended. However, sometimes
the toolbar or an object's context menu can offer a shortcut. As a full-fledged alternative for manual operation via mouse and keyboard, also the menu bar provides access
to all instrument functions and settings.
31Getting Started 1316.2407.02 ─ 62
R&S®ZND
Operating the instrument
Manual operation
Figure 5-1: Function Keys
Customizing the screen
The contents of the screen and the size and position of many display and control elements are not fixed. You can display or hide most elements. You can also drag and
drop traces, info fields, and even the softtool panel to your preferred position; see
Chapter 5.4.4, "Using drag and drop", on page 46.
For example, you can show/hide the on-screen hardkey panel by selecting/deselecting
"Display" > "View Bar" > "Hard Key Panel On" from the main menu.
The following table shows possible touchscreen operations with the corresponding
mouse operations.
TouchscreenMouseTypical task
Tap control element for a short timeClick control element (left mouse button)Select button or tab
Tap for an extended time (tap and hold)Click (right button)Open context menu
Tap twice (double-tap)Double-click (left button)Open on-screen keyboard
Using the Front Panel
To access an instrument function:
1. Press a (virtual) key, e.g. the [MEAS] key in the TRACE section.
The corresponding softtool expands at the current docking position.
32Getting Started 1316.2407.02 ─ 62
R&S®ZND
Operating the instrument
Manual operation
left = unidirectional R&S ZND
right = bidirectional R&S ZND
2. Activate the desired softtool tab, e.g. "Z←Sij".
left
= unidirectional R&S ZND
right = bidirectional R&S ZND
3. Select a control element, e.g. "Z←S11".
The diagram immediately reflects your selection. The active trace shows the measurement results for the selected measured quantity.
A control element with three dots (e.g. SYSTEM – [SETUP] > "Setup" > "System Config...") opens a dialog, containing a group of related settings, a wizard or additional
information.
33Getting Started 1316.2407.02 ─ 62
R&S®ZND
Operating the instrument
Manual operation
Using the menu bar
The menu bar at the bottom of the application screen provides alternative access to all
instrument functions. To repeat the measured quantity selection described above,
► Select TRACE – [MEAS] > "Z←Sij" > "Z←S11".
The diagram immediately reflects your selection. The active trace shows the measurement results for the selected measured quantity. At the same time, the related
softtool tab is opened.
Using context menus
Context menus are another alternative for quick access to instrument settings.
1. Touch and hold (right-click) the measured quantity section in the trace info for a
couple of seconds until the context menu appears.
34Getting Started 1316.2407.02 ─ 62
R&S®ZND
Operating the instrument
Manual operation
left = unidirectional R&S ZND
right = bidirectional R&S ZND
2. Select "S-Parameter" to open the "Meas" > "S-Params" softtool tab.
= unidirectional R&S ZND
left
right = bidirectional R&S ZND
3. Select "Z←Sij" > "Z←S11".
35Getting Started 1316.2407.02 ─ 62
R&S®ZND
Operating the instrument
Control elements of the application window
left = unidirectional R&S ZND
right = bidirectional R&S ZND
5.2Control elements of the application window
The application window of the analyzer provides all control elements for the measurements and contains the diagrams for the results. There are several alternative ways for
accessing an instrument function:
●
Using a function key on the (virtual) hardkey panel to open the related softtool (recommended, provides all settings)
●
Using the menus and submenus of the menu bar (alternative to the previous
method)
●
Using the context menus of certain display objects (for important actions in the
context of this object)
●
Using the icons in the toolbar above the diagram area (for frequent global actions)
36Getting Started 1316.2407.02 ─ 62
R&S®ZND
Operating the instrument
Control elements of the application window
These methods are described in more detail in the following sections.
5.2.1Title bar
By default, the analyzer GUI is shown in full screen mode, covering the whole screen
and hiding the Windows taskbar. However, you can toggle the full screen mode using
SYSTEM – [DISPLAY] > "View Bar" > "Title and Task Bar On".
If full screen mode is switched off, the main application window of the vector network
analyzer application provides a standard Windows® title bar.
5.2.2Toolbar
The toolbar above the diagram area contains the most frequently used control elements of the user interface. All controls are also accessible via Softtools.
The toolbar is divided into several icon groups, separated by vertical lines.
37Getting Started 1316.2407.02 ─ 62
R&S®ZND
Operating the instrument
Control elements of the application window
These icons represent the undo and redo actions that are also available
via the menu bar items "System" > "Undo" / "Redo".
Undo reverses the last action, redo reverses the last undo action (if possible).
These icons control the zoom function (TRACE – [SCALE] > " Zoom").
If multiple diagrams are configured, this icon toggles the "Maximize Diagram" action for the active diagram (SYSTEM – [DISPLAY] > "Diagram" >
"Maximize Diagram").
These icons implement the following actions, from left to right:
●
Add a clone of the active trace to the active diagram (single tap; same
as TRACE – [TRACE CONFIG] > "Traces" > "Add Trace") or to an
arbitrary/new one (drag & drop the "Trc+" icon to the diagram area).
●
Add a new marker to the active diagram (single tap; similar to TRACE
– [MARKER] > "Markers" > "Mkr<i>") or to an arbitrary one (drag &
drop the "Mkr+" icon).
●
Delete the active trace (single tap the trash icon), or an arbitrary trace
(drag & drop its trace info field to the trash icon). Or delete all markers
of a trace (drag & drop the marker info field to the trash icon).
You can hide the toolbar using SYSTEM – [DISPLAY] > "View Bar".
5.2.3Softtools
Softtools display groups of related settings as a tabbed panel. They can be opened via
function keys on the front panel or the on-screen "Hard Key" panel, or via menu bar
and context menu items.
These icons provide the following actions, from left to right:
●
Print the current diagrams to a bitmap file (same as SYSTEM – [FILE]
> "Print" > "To File...").
By default, all diagrams are printed, no matter if displayed or not.
However you can also choose to print only the active diagram (see
SYSTEM – [FILE] > "Print" > "Print...").
●
Open the Windows® Start menu.
●
Open the context-sensitive help.
●
Restart the sweep in all channels (same as CHANNEL – [SWEEP] >
"Sweep Control" > "Restart Sweep")
These icons provide the following actions, from left to right:
●
Toggle advanced diagram area editing, which makes rearranging
and/or deleting diagrams a breeze.
●
Toggle other measurements (except the active trace) OFF|ON
38Getting Started 1316.2407.02 ─ 62
R&S®ZND
Operating the instrument
Control elements of the application window
Figure 5-2: Scale softtool
A softtool consists of a title area with a close/re-open icon and a tabbed panel below it.
The title area remains displayed when the softtool is closed, which allows you to
reopen a closed softtool at any time.
Some controls on the softtool tabs allow you to read and modify settings (e.g. "Ref
Value" in the screenshot above), some perform actions (e.g. "Auto Scale Trace"), while
others open additional dialogs (button label ends with "...").
5.2.4Menu bar
You can also access the analyzer functions via textual menus. The menu bar is located
below the diagram area:
As in any Windows® application, menus can be controlled with the touchscreen or a
mouse. A short tap (left mouse click) expands a menu or submenu. If a menu command has no submenu assigned, a short tap (left mouse click) opens a dialog or
directly activates the menu command. When a (sub)menu is selected, the R&S ZND
displays the corresponding softtool.
Overview of menu functions
●
The "File" menu provides functions to handle recall sets and trace files, to print or
save diagram content, and to shut down the VNA application.
●
The "Trace" menu provides all trace settings, the limit check settings, and the
marker functions including marker search.
●
The "Channel" menu provides all channel settings and activates, modifies or stores
different channels.
39Getting Started 1316.2407.02 ─ 62
R&S®ZND
●
The "Display" menu provides all display settings and the functions for activating,
modifying and arranging different diagrams.
●
The "Application" menu gives access to applications and tools that extend the functionality of the analyzer firmware.
●
The "System" menu provides functions that can be used to return to a defined
instrument state, reverse operations, access service functions and define various
system-related settings.
●
The "Help" menu provides assistance with the network analyzer and its operation.
You can toggle the visibility of the menu bar using SYSTEM – [DISPLAY] > "View Bar"
> "Menu Bar".
5.2.5Menu structure
All menus show an analogous structure.
Operating the instrument
Control elements of the application window
●
A menu command with a right arrow expands a submenu with further related settings.
Example: "Marker" expands a submenu with marker-related properties.
●
A menu command with three dots appended calls up a dialog providing several
related settings.
Example: "Search Range" opens a dialog to define the search range for the
marker search.
●
A menu command with no arrow or dots initiates an immediate action.
Example: "Max" sets the active marker to the maximum of the active trace.
40Getting Started 1316.2407.02 ─ 62
R&S®ZND
Operating the instrument
Control elements of the application window
5.2.6Hardkey panel
The (virtual) "Hard Key" panel provides on-screen access to the function keys (plus the
[UNDO] and [REDO] key) of the R&S ZND. Most of the function keys open a related
softtool. For a short description, refer to section Chapter 4.1.2, "Function keys",
on page 25.
The "Hard Key" panel is particularly useful if the analyzer is controlled from an external
monitor or Remote Desktop.
For the R&S ZND, it is hidden by default.
You can display the "Hard Key" panel using one of the following methods:
Select "Display"> "View Bar" > "Hard Key Panel On" from the menu bar.
●
Select "Hard Key" from the context menu of the softtool panel.
5.2.7Status bar
The status bar shows
●
the active channel
●
the current channel's sweep averaging counter (e.g. "Ch<i>: Avg 9/10"), or "Ch<i>:
Avg None" if averaging is disabled
●
the progress of the sweep
The progress bar also shows when the R&S ZND prepares a sweep with new
channel settings
●
the "EXT REF" symbol, if an external reference clock is used for synchronization
41Getting Started 1316.2407.02 ─ 62
R&S®ZND
●
the switch matrix status symbol, if a switch matrix is configured
(See the R&S ZND User Manual)
●
the current date and time
Figure 5-3: R&S ZND with switch matrix
The progress bar shows a moving color gradient if the current sweep is too fast to be
monitored, e.g. because the number of sweep points is low. You can hide/show the status bar using SYSTEM – [DISPLAY] > "View Bar" > "Status Bar".
5.3Working with dialogs
Dialogs provide groups of related settings and allow to make selections and enter data
in an organized way. The settings are visualized, if possible. An example is shown
below.
Operating the instrument
Working with dialogs
All dialogs are operated in a similar way.
●
To open a dialog, select a softtool button with three dots appearing in its label (e.g.
"Start... (Manual)").
●
The title bar of each dialog contains some convenience functions:
42Getting Started 1316.2407.02 ─ 62
R&S®ZND
Operating the instrument
Handling diagrams, traces, and markers
–Use the "Dialog Transparency" function to make the display elements behind
the dialog visible.
Note: The "Dialog Transparency" is a global setting, i.e. it applies to all dialogs.
–
Use the icon to create a screenshot of the dialog.
●
Drag and drop the lower right corner of the dialog to modify its size.
●
Some dialogs are subdivided into tabs, containing groups of related settings. Activate a tab to access those settings.
The Help system provides useful information about each dialog's specific settings.
Select "Help" to open the Help.
5.4Handling diagrams, traces, and markers
The analyzer displays measurement results as traces in rectangular diagrams. Markers
are used to read specific numerical values and to search for points or regions on a
trace. The following section presents some of the graphical tools the R&S ZND provides for trace and marker handling.
For further reference
Refer to chapter "Concepts and Features" in the R&S ZND Help or in the User Manual
to learn more about traces, channels, and screen elements.
5.4.1Adding new traces and diagrams
A new trace is required if you want to measure and display an additional quantity.
Typical scenario: The transmission coefficient S21 is measured as described in Chap-
ter 6.1, "Transmission S-parameter measurement", on page 56. A trace is added to
display the reflection coefficient S11 for comparison.
To create a trace:
1. Drag the "Trc+" icon from the toolbar into a diagram.
The diagram changes its color scheme and contents as shown below.
43Getting Started 1316.2407.02 ─ 62
R&S®ZND
Operating the instrument
Handling diagrams, traces, and markers
2. Select the adequate drop position, depending on whether you want to display the
new trace in the existing diagram, or whether you want to add a new diagram. The
highlighted area indicates the target diagram.
3. In the dialog box that is opened when you release the "New Trace" icon, select the
S-parameter to be measured.
= bidirectional R&S ZND
left
right = unidirectional R&S ZND
The R&S ZND generates a new trace for the selected S-parameter.
44Getting Started 1316.2407.02 ─ 62
R&S®ZND
Alternative control elements
To measure a different quantity, select TRACE – [MEAS]. Drag and drop a softkey representing a measured quantity to create a trace. Or simply select another softkey to
change the measured quantity of the active trace.
Select TRACE – [TRACE CONFIG] to access more trace handling functions. Select
SYSTEM – [DISPLAY] to access more diagram handling functions.
5.4.2Adding new markers
A marker is needed, for instance, to read a particular numerical trace value.
To add a new marker:
●
Drag and drop the "New Marker" icon from the toolbar to the desired position in the
target diagram.
Operating the instrument
Handling diagrams, traces, and markers
The marker appears on the target diagram's active trace. The marker info field displays the stimulus value (x-axis value) and response value (y-axis value) at the
marker position. The response value varies as the analyzer continues performing
sweeps.
Active trace, alternative control elements
The trace line of the active trace in the upper part of the diagram is highlighted. If the
diagram contains several traces, first activate the target trace, then add the marker.
The TRACE – [MARKER] softtool provides more functions for marker handling. In particular, any marker offered in the "Markers" tab can be positioned on the active trace
using drag & drop.
45Getting Started 1316.2407.02 ─ 62
R&S®ZND
5.4.3Deleting display elements
Markers, traces, diagrams, and other display elements are most conveniently deleted
using the "Delete" icon in the toolbar above the diagram area.
●
To delete a single marker, drag it into vertical direction to release it from the trace
and drop it onto the "Delete" icon.
To delete a set of markers, drag and drop their marker info field onto the "Delete"
icon.
Deleting a marker and its info field also disables the associated marker function.
●
To delete a trace, drag and drop its trace line onto the "Delete" icon. The active
trace can also be deleted by tapping/clicking the "Delete" icon.
Note however, that the last remaining trace cannot be deleted.
●
To delete a diagram, drag and drop its diagram number label onto the "Delete"
icon.
Note however, that the last remaining diagram cannot be deleted.
●
To delete a channel, drag and drop all associated traces onto the "Delete" icon.
Note however, that the last remaining channel cannot be deleted.
●
To hide the limit lines and disable the limit check, drag and drop the PASS / FAIL
message onto the "Delete" icon. The limit line itself is not deleted; you can re-use it
any time.
Operating the instrument
Handling diagrams, traces, and markers
The context menu of some display elements also provides the "Delete" function.
Undo function
If you happen to delete a display element unintentionally, you can restore it using the
"Undo" toolbar icon.
5.4.4Using drag and drop
You can drag and drop many of the R&S ZND's control and display elements to
change their size and position. The drag and drop functionality is often more convenient to use than the equivalent buttons of the softtool panels. The following table gives
an overview.
Table 5-1: Drag and drop functionality for various screen elements
Screen elementActionDrag and drop...
DiagramCreateSee Chapter 5.4.1, "Adding new traces and diagrams", on page 43
ResizeSeparator between adjacent diagrams
DeleteSee Chapter 5.4.3, "Deleting display elements", on page 46
TraceCreateSee Chapter 5.4.1, "Adding new traces and diagrams", on page 43
Move verticallyReference line marker (right diagram edge)
Move into other or new
diagram
DeleteSee Chapter 5.4.3, "Deleting display elements", on page 46
Trace line
46Getting Started 1316.2407.02 ─ 62
R&S®ZND
Screen elementActionDrag and drop...
Operating the instrument
Entering data
Reset / suspend
graphic zoom
MarkerCreateSee Chapter 5.4.2, "Adding new markers", on page 45
Move horizontallyMarker symbol
DeleteMarker or marker info field --> "Delete" icon; see Chapter 5.4.3, "Deleting display
Marker info fieldAddSee Chapter 5.4.2, "Adding new markers", on page 45
Move within diagramMarker info field (move to one of several pre-defined positions)
DeleteSee Chapter 5.4.3, "Deleting display elements", on page 46
"Zoom" element in additional trace line --> "Delete" icon; see Chapter 5.6.1,
"Using the graphical zoom", on page 50
elements", on page 46
5.5Entering data
The analyzer provides dialogs with various types of input fields where you can enter
numeric values and character data. Data entry with a mouse and an external keyboard
is a standard procedure known from other Windows® applications. However, there are
various other ways to enter data.
5.5.1Using front panel keys
You can use the keys in the DATA ENTRY keypad to enter numbers, units, and characters.
To enter a numeric value:
1. Select a numeric data input field to activate it.
2. Press the data entry keys.
●Use [0] to [9] to enter the corresponding numbers.
47Getting Started 1316.2407.02 ─ 62
R&S®ZND
Operating the instrument
Entering data
●Use [.] to enter a decimal point.
●Use [-] to change the sign of the value.
●
Use [G/n], [M/μ], [k/m], or [x1] to multiply the entered value with factors of 10
(-)6
10
(-)3
, 10
or 1 and to add the appropriate physical unit.
To enter a character string:
1. Tap a character data input field to activate it.
2. Press the DATA ENTRY keys as if you were writing a short message on your
mobile phone.
●Use [0] to [9] to enter the corresponding numbers.
●Use [.] and [-] to enter a dot or hyphen.
●Use Use [G/n], [M/μ], or [k/m] to enter the letters G, M or K (case-insensitive).
●Use the [←] key to correct wrong entries, deleting the character to the left of
the current cursor position.
●Press [ENTER] to complete an entry.
●Press [ESC] to discard the entries made.
(-)9
,
3. To enter letters other than G, M or K, you can also use one of the methods described in the following sections:
●Chapter 5.5.3, "Using the analyzer's on-screen keyboard", on page 49
●
Chapter 5.5.4, "Using the Windows® on-screen keyboard", on page 49
5.5.2Using the numeric editor
The "Numeric Editor" is a tool for convenient entry and modification of numeric values.
It is available for all numeric input fields in the analyzer GUI.
Operation with touchscreen or mouse:
1. Double-tap (double-click) a numeric input field in a dialog or on a softtool to open
the numeric editor.
2. Use the buttons in the numeric keypad to compose the numeric input value.
48Getting Started 1316.2407.02 ─ 62
R&S®ZND
Operating the instrument
Entering data
3. If desired, select a "Step Size" and use the cursor up/down buttons to increment/
decrement the current value. If a marker is active, you can also set the numeric
value to the current marker value ("Set to Marker").
4. After completing the input string, select "ENTER" to apply your selection and close
the numeric editor.
[STEP SIZE] key
If a numeric input field is active, the [STEP SIZE] front panel key opens a dialog box
containing the "Step Size" panel of the numeric editor. Use this box for efficient operation of the rotary knob (and mouse wheel).
5.5.3Using the analyzer's on-screen keyboard
The on-screen "Keyboard" allows you to enter characters, in particular letters, without
an external keyboard. It is available for all text input fields in the analyzer GUI.
5.5.4
1. Activate a character data input field in a softtool or a dialog.
2. Double-tap/click the input field to open the on-screen keyboard.
3. Select character buttons to compose the input string.
4. Select "Enter" to apply your selection and close the keyboard.
Using the Windows
®
on-screen keyboard
The Windows® on-screen keyboard allows you to enter characters, in particular letters,
even if an input field cannot call up the analyzer's own on-screen keyboard. Examples
are input fields in standard Windows ® dialogs.
49Getting Started 1316.2407.02 ─ 62
R&S®ZND
Operating the instrument
Scaling diagrams
Figure 5-4: Windows 10 on-screen keyboard
To call up the on-screen keyboard:
1. Open the SYSTEM – [APPLIC] softtool.
2. In the "External Tools" tab, select "Screen Keyboard".
5.6Scaling diagrams
The analyzer provides various tools for customizing the diagrams and for setting the
sweep range. Choose the method that is most convenient for you.
5.6.1Using the graphical zoom
The graphical zoom function magnifies a rectangular portion of the diagram (zoom window) to fill the entire diagram area. The sweep points are not affected.
The graphical zoom function is only supported for cartesian trace formats. For (inverted) Smith and polar diagrams, it is not available.
To activate the graphical zoom:
●
Select the "Zoom Config" icon in the toolbar above the diagram area and make
sure that the graphical zoom mode is active:
●
Select the "Zoom Select" toggle button in the toolbar above the diagram area.
50Getting Started 1316.2407.02 ─ 62
R&S®ZND
Operating the instrument
Scaling diagrams
The icon changes its background color from black to blue.
●
In the active diagram area, select an appropriate rectangular area.
The zoomed view shows the selected rectangle, scaled in both horizontal and vertical direction. In general, the zoom window covers only a part of the sweep range;
the horizontal distance between the sweep points increases. The reduced display
range is indicated in an additional zoom line in the channel info area.
Use the "Zoom Reset" icon to restore the original diagram. Alternatively, you can drag
and drop the "Zoom" label from the additional channel info line onto the toolbar button.
51Getting Started 1316.2407.02 ─ 62
R&S®ZND
Operating the instrument
Scaling diagrams
Alternative settings
●
The TRACE – [SCALE] > "Zoom" softtool tab allows you to define the displayed
zoom range numerically.
It can also be opened from the menu of the "Zoom Config" toolbar icon.
●
If the active diagram is graphically zoomed, "Overview On" in the "Zoom" softtool
tab toggles an overview. The upper part of the diagram then shows a small version
of the unzoomed diagram. You can move the zoomed part of the trace by moving
the rectangular area in the overview.
●
To zoom the stimulus range (keeping the number of sweep points constant), use
the "Zoom Config" toolbar icon and select "Mode Stimulus Zoom". Then use the
"Zoom Select" icon to narrow the sweep range and adjust the vertical scaling.
5.6.2Setting the sweep range
The sweep range for all related channels is displayed in the channel info area at the
bottom of each diagram:
To change the sweep range of the active channel, use one of the following methods:
●
Use the [START], [STOP], [CENTER], and [SPAN] function keys from the STIMULUS section.
●
Double-tap (with a mouse: double-click) the "Start" or "Stop" label in the channel
list.
52Getting Started 1316.2407.02 ─ 62
R&S®ZND
●
Tap and hold (with a mouse: right-click) the "Start" or "Stop" label in the channel list
and select "Start Frequency", "Stop Frequency", "Center Frequency", or "Frequency Span" from the context menu.
●
Select "Start Frequency", "Stop Frequency", "Center Frequency", "Span Frequency" from the "Channel" > "Stimulus" menu.
●
Use the "Set by Marker" functions (TRACE – [MARKER] > "Set by Marker"; see
Chapter 5.6.6, "Set by marker", on page 54).
5.6.3Reference value and position
The analyzer provides three parameters for changing the scale of the vertical
(response) axis:
●
Changing the "Ref Value" or "Ref Pos" shifts the trace in vertical direction and
adjusts the labels of the vertical axis. "Ref Value" also works for radial diagrams.
●
Changing the "Scale/Div" modifies the value of the vertical or radial diagram divisions and thus the entire range of response values displayed.
Operating the instrument
Scaling diagrams
The "Scale/Div" and the "Ref Value" are indicated in the scale section of the trace info.
In the example below, a "Scale/Div" of 10 dB and a "Ref Value" of 0 dB is used.
To change such a scale parameter, use one of the following methods:
●
Open the TRACE – [SCALE] > "Scale Values" softtool tab and proceed from there.
●
Tap and hold (with a mouse: right-click) the scale section in the trace info and
select a setting from the context menu.
●
Select a setting from the "Trace" > "Scale" menu.
●
Use "Set by Marker" functions (TRACE – [MARKER] > "Set by Marker"; see Chap-
ter 5.6.6, "Set by marker", on page 54).
5.6.4Auto scale
The "Auto Scale" function adjusts the scale divisions and the reference value so that
the entire trace fits into the diagram. To access "Auto Scale", use one of the following
methods:
●
Open the TRACE – [SCALE] > "Scale Values" softtool tab and select "Auto Scale
Trace" or "Auto Scale Diagram".
●
Tap and hold (with a mouse: right-click) the scale section in the trace info and
select "Auto Scale Trace" from the context menu.
●
Select "Auto Scale Trace" or "Auto Scale Diagram" from the "Trace" > "Scale"
menu.
53Getting Started 1316.2407.02 ─ 62
R&S®ZND
5.6.5Circular diagrams
The radial scale of a circular diagram ("Polar", "Smith" or "Inverted Smith") can be
changed with a single linear parameter, the "Ref Value". The reference value defines
the radius of the outer circumference.
●
Increasing the "Ref Value" scales down the polar diagram.
●
Decreasing the "Ref Value" magnifies the polar diagram.
The "Ref Value" is indicated in the scale section of the trace info.
To change the "Ref Value" setting, use one of the following methods:
●
Locate it on the TRACE – [SCALE] > "Scale Values" softtool tab.
●
Tap and hold (with a mouse: right-click) the scale section in the trace info and
select the parameter from the context menu.
●
Select the parameter from the "Trace" > "Scale" menu.
●
Use the "Set by Marker" functions; see Chapter 5.6.6, "Set by marker",
on page 54.
Operating the instrument
Scaling diagrams
5.6.6Set by marker
The "Set by Marker" functions are a convenient tool for scaling (in particular: magnifying) diagrams without entering explicit numeric values. You simply place a marker to a
trace point and use the marker values to change the sweep range or move the trace
relative to the vertical axis. The touchscreen or a mouse makes it easier to activate
(touch/click) or move (drag and drop) markers.
To set the sweep range using markers, use one of the following methods.
Set "Start" and "Stop" values in the diagram:
1. Create two normal markers, e.g. the markers "Mkr 1" (default label "M1") and "Mkr
2" (default label "M2").
54Getting Started 1316.2407.02 ─ 62
R&S®ZND
Operating the instrument
Scaling diagrams
See Chapter 5.4.2, "Adding new markers", on page 45.
2. Place "M1" to the start value of the desired sweep range and tap TRACE –
[MARKER] > "Set by Marker" > "Start = Marker".
3. Place "M2" to the stop value of the desired sweep range and tap TRACE –
[MARKER] > "Set by Marker" > "Stop = Marker".
Use a definite "Span:"
1. Create a marker.
2. Enable "Delta Mode" for this marker.
The analyzer automatically creates an additional reference marker.
3. Place the reference marker to the start value of the desired sweep range.
4. Set the value of the delta marker to the desired sweep range and tap TRACE –
[MARKER] > "Set by Marker" > "Span = Marker".
To move the trace in vertical direction, proceed as follows:
1. Create a normal marker, e.g. the marker "Mkr 1" (default label "M1").
2. Place "M1" to a particular trace point, e.g. use the "Marker Search" functions to
locate a maximum or minimum on the trace.
3. Select TRACE – [MARKER] > "Set by Marker" > "Max = Marker" to move the trace
towards the upper diagram edge, leaving the values of the vertical divisions
("Scale/Div") and the overall vertical scale unchanged. Analogously, select "Min =
Marker" to move the trace towards the lower diagram edge, or select "Ref Val =
Marker" to move the trace towards the "Ref Value".
You can also use marker values in the "Numeric Editor"; see Chapter 5.5.2, "Using the
numeric editor", on page 48.
5.6.7Enlarging a diagram
The analyzer provides different tools for customizing the contents and size of the diagrams:
●
Select
SYSTEM – [DISPLAY] > "Diagram" > "Maximize Diagram".
If enabled the active diagram is always maximized.
●
The "Menu Bar", the "Status Bar", the "Hard Key Panel", and the "Title Bar" can be
hidden to gain space for the diagrams (SYSTEM – [DISPLAY] > "View Bar").
●
The SYSTEM – [DISPLAY] > "Config" softtool tab defines optional display elements
for the interior of the diagrams.
from the toolbar to maximize the active diagram. Or, equivalently, select
Use the context menu of the diagram, the SYSTEM – [DISPLAY] key or the "Display"
menu to access the display settings.
55Getting Started 1316.2407.02 ─ 62
R&S®ZND
Performing measurements
Transmission S-parameter measurement
6Performing measurements
This chapter takes you through a sample session with a R&S ZND network analyzer
and describes basic operation tasks.
Prerequisite
The instrument is set up, connected to the mains system, and started up as described
in Chapter 3, "Preparing for use", on page 13.
Use the "S-Parameter Wizard" accessible via TRACE – [MEAS] > "S-Params" > "SParam Wizard..." to measure S-parameters in a straightforward way. The wizard provides a series of dialogs where you can select the test setup, screen configuration and
measurement parameters, configure the essential channel settings and perform a guided calibration.
Measurement stages in the wizard
The individual dialogs of the "S-Parameter Wizard" correspond to the typical stages of
any measurement:
1. Select the test setup.
2. Define port impedances.
3. Select the measurement parameters and the diagrams.
4. Define the sweep range.
5. Adjust the receiver and source settings (measurement bandwidth, source power).
6. Perform a calibration.
6.1Transmission S-parameter measurement
In a transmission measurement, the analyzer transmits a stimulus signal to the input
port of the device under test (DUT) and measures the transmitted wave at the DUT's
output port. The trace settings allow you to select the measured quantities and display
formats, depending on what you want to learn from the data. A minimum of two analyzer test ports are required for transmission measurements.
In the following example, the analyzer is set up for a two-port transmission measurement. A frequency sweep range is selected, the instrument is calibrated and the measurement result is analyzed using various display formats.
6.1.1Connecting the instrument for transmission measurements
To prepare a transmission measurement, you have to connect your DUT (which for
simplicity we assume to have appropriate connectors) in-between a pair of analyzer
56Getting Started 1316.2407.02 ─ 62
R&S®ZND
Performing measurements
Transmission S-parameter measurement
test ports. It is recommended that you preset the R&S ZND to start from a well-defined
instrument state.
VNA
12
12
DUT
1. Connect the DUT between test ports 1 and 2 of the network analyzer as shown
above.
2. Use the [PRESET] key to restore a well-defined instrument state.
The analyzer is now set to its default state. The default measured quantity is the
transmission S-parameter S21.
Select TRACE – [TRACE CONFIG] and use the control elements in the "Traces" softtool tab if you wish to create additional traces and diagrams.
6.1.2Selecting the sweep range and other parameters
After a system preset the display shows a diagram with a dB magnitude scale, and the
S-parameter S21 is selected as a measured quantity. This S-parameter is the forward
transmission coefficient of the DUT. It is defined as the ratio of the transmitted wave at
the DUT's output port (port no. 2) to the incident wave at the DUT's input port (port no.
1).
57Getting Started 1316.2407.02 ─ 62
R&S®ZND
Performing measurements
Transmission S-parameter measurement
The R&S ZND automatically adjusts its internal source and receiver to the selected
measured quantities: For an S21 measurement, a stimulus signal (termed a1) is trans-
mitted at the analyzer port no. 1; the transmitted wave (termed b2) is measured at port
2. The stimulus signal from the analyzer port no. 2 is not needed except for some cali-
bration types.
By default the sweep range is set to the frequency range of the analyzer, which can be
unsuitable for your DUT. The following procedure shows you how to configure a
smaller sweep range.
1. Select STIMULUS – [START] and set the "Start Frequency" to the lowest frequency you want to measure (e.g. 1.77 GHz). For convenient numeric entry, open
the "Numeric Editor" (see Chapter 5.5.2, "Using the numeric editor", on page 48).
Tip: If you use the DATA ENTRY keys at the front panel for data entry, type [1][.][7]
[7] and terminate the entry with the [G/n] key.
Refer to Chapter 5.5, "Entering data", on page 47 to learn more about entering
numeric values and characters.
2. In the "Stop Frequency" input field, enter the highest frequency you want to measure (e.g. 2.5 GHz).
3. Select TRACE – [SCALE] > "Scale Values" and activate the "Auto Scale Trace"
function. The analyzer adjusts the scale of the diagram to fit in the entire S21 trace,
leaving an appropriate display margin.
Tip: Refer to Chapter 5.6, "Scaling diagrams", on page 50 to learn more about the
different methods and tools for diagram scaling.
6.1.3Calibrating the instrument
Calibration (system error correction) is the process of eliminating systematic, reproducible errors from the measurement results. E.g., in the current test setup, the connecting cables between the analyzer ports and the DUT introduce an attenuation and a
phase shift of the waves. Both effects impair the accuracy of the S-parameter measurement.
The analyzer provides a wide range of sophisticated calibration methods for all types of
measurements. The calibration method to select depends on the expected system
errors, the accuracy requirements of the measurement, on the test setup and on the
types of calibration standards available.
58Getting Started 1316.2407.02 ─ 62
R&S®ZND
Performing measurements
Transmission S-parameter measurement
The following example requires a calibration kit with a male Through standard with
known transmission characteristics for the related test port connector type and gender.
With a single Through, it is possible to perform a transmission normalization, compensating for a frequency-dependent attenuation and phase shift in the signal paths.
Due to the R&S ZND's calibration wizard, calibration is a straightforward, guided process.
1. Replace the DUT by the Through standard of your calibration kit. Make sure to disconnect all calibration units.
2. Select CHANNEL – [CAL] > "Start... (Manual)" to open the "Calibration Setting"
wizard.
3. Select the port combination Port 1 (P1) and Port 2 (P2) and the calibration type
"Trans Norm". Make sure to define port 1 as the source port.
Tip: For a unidirectional R&S ZND, only port 1 can be the source port. Hence cali-
brations that require both ports as driving port are not available.
4. Select "Next" to proceed to the next page of the "Calibration Setting" wizard.
5. Select the test port connector type and gender (here: N 50 Ω, female, corresponding to a male Through standard), and the calibration kit (here: R&S ZV-Z121).
59Getting Started 1316.2407.02 ─ 62
R&S®ZND
Performing measurements
Transmission S-parameter measurement
6. Tap"Start".
7. The calibration dock widget indicates the standard measurements that make up a
"Trans Norm" calibration.
Select "Through (mm)" to initiate the measurement of the connected Through standard. Measuring the isolation between ports 1 and 2 is optional. Skip it for now.
60Getting Started 1316.2407.02 ─ 62
R&S®ZND
Performing measurements
Transmission S-parameter measurement
The analyzer performs a calibration sweep for the measured quantity S21. The
magnitude and phase of the result is displayed in two diagrams, together with the
expected typical result for a Through standard. The similarity of real and expected
traces indicates that the Through standard has been properly connected. After the
R&S ZND has completed the calibration sweep and calculated the correction data,
the "Apply" button is enabled.
8. Select "Apply" to close the wizard.
The system error correction is calculated and applied to the current channel. A
"Cal" label appears in the trace list.
To proceed with the measurement, remove the Through standard and connect the DUT
again.
6.1.4Evaluating data
The analyzer provides various tools to optimize the display and analyze the measurement data. For instance, you can use markers to determine maxima and minima on the
trace, and change the display format to obtain information about the group delay of the
transmitted wave.
1. Select TRACE – [MARKER] > "Markers" > "Mkr 1".
This places marker "M1" to its default position (center of the sweep range). A
marker symbol (triangle) appears on the trace, a marker info field in the upper right
corner of the diagram. The marker info field displays the stimulus value (frequency)
and response value (magnitude of the transmission coefficient converted to a dB
value) at the marker position.
2. Select TRACE – [MARKER], activate the "Marker Search" softtool tab and activate
"Min" search.
The marker jumps to the absolute minimum of the curve in the entire sweep range.
The marker info field shows the coordinates of the new marker position.
3. Select TRACE – [FORMAT] and choose the "Delay" of the transmission coefficient
as displayed quantity.
61Getting Started 1316.2407.02 ─ 62
R&S®ZND
Performing measurements
Transmission S-parameter measurement
The group delay represents the propagation time of the wave through the DUT; it is
displayed in a Cartesian diagram. The marker info field shows the frequency and
group delay at the marker position.
Refer to the information on trace formats in the help system or in the user manual to
learn more about the diagram properties.
6.1.5Saving and printing data
The analyzer provides standard functions for saving measurement settings and for
saving or printing the results. You can use these functions as if you were working on a
standard PC. Moreover you can export your trace data to an ASCII file and reuse it in a
later session or in an external application.
Data transfer is made easier if external accessories are connected to the analyzer or if
the instrument is integrated into a LAN. Refer to Chapter 3.11, "Connecting external
accessories", on page 20, and Chapter 7.2, "Remote operation in a LAN", on page 66
to obtain information about the necessary steps.
1. Activate the SYSTEM – [FILE] > "Trace Data" softtool tab.
2. In the "Trace Data" softtool tab, select "Export" – "ASCII..." to open the "Export
Data - ASCII Files" dialog.
3. In the "Export Data - ASCII Files" dialog:
a) Select a file location ("Look in:").
b) Enter a file name ("File name:").
c) Select "Save".
The analyzer writes the data of the active trace to an ASCII file and closes the dialog.
5. In the "Print" softtool tab, select "Print" to print the diagram area or "To Clipboard"
to copy it to the Windows clipboard.
62Getting Started 1316.2407.02 ─ 62
R&S®ZND
Performing measurements
Reflection S-parameter measurement
6. Select SYSTEM – [FILE] > "Recall Sets" > "Save..." to open the "Save" dialog for
recall sets.
7. In the "Save" dialog:
a) Select a file location ("Look in:").
b) Enter a name for the recall set file ("File name:").
c) Select "Save".
The analyzer saves the active recall set, containing channel, stimulus and trace
settings, to a znx file. This recall set can be restored in a later session.
6.2Reflection S-parameter measurement
In a reflection measurement, the analyzer transmits a stimulus signal to the input port
of the device under test (DUT) and measures the reflected wave. Different trace formats allow you to express and display the results, depending on what you want to
learn from the data. Only one analyzer test port is required for reflection measurements.
In principle, a reflection measurement involves the same steps as a transmission measurement. Note the following differences:
●
The basic test setup for reflection measurements involves a single DUT and analyzer port. For instance, you can connect the input of your DUT to port 1 of the analyzer as shown below.
VNA
1
1
DUT
You can also use the basic transmission test setup, e.g. if you want to measure
reflection and transmission parameters in parallel.
●
The analyzer provides special calibration types for reflection measurements. Use
the calibration wizard and select an appropriate type.
63Getting Started 1316.2407.02 ─ 62
R&S®ZND
Performing measurements
Reflection S-parameter measurement
A full 2-port calibration (TOSM, UOSM, TNA ...) corrects the system errors for all
transmission and reflection S-parameters.
●
Some of the trace formats are particularly suited for reflection measurements. For
instance, you can display the measured reflection coefficient S11 in a Smith chart to
obtain the complex input impedance at port 1.
Proceed as described in
on page 17 to shut down your analyzer.
Chapter 3.7, "Switching the instrument on and off",
64Getting Started 1316.2407.02 ─ 62
R&S®ZND
Administrative tasks
Firmware installation
7Administrative tasks
This chapter describes some topics that are only needed occasionally, or if a special
instrument configuration is required.
7.1Firmware installation
Upgrade versions of the analyzer firmware are supplied as single executable setup
files (*.exe).
64-bit only
Since version 3.00 of the analyzer firmware, 32-bit Windows is no longer supported.
Only 64-bit firmware is available.
To upgrade your instrument to 64-bit Windows 10, contact Rohde & Schwarz service.
Admin account
You need administrator rights to install a new firmware. See note on "User accounts
and password protection"on page 19 for details.
To perform a firmware update:
1. Copy the setup file to any storage medium accessible from the analyzer. This can
be either the internal mass storage drive, an external storage medium (USB memory stick, external CD-ROM drive) or a network connection (LAN).
The default name of the internal drive is C:. External storage devices are automatically mapped to the next free drive, i.e. D:, E: etc.
2.
Run the setup file from the Windows® Explorer. Follow the instructions of the setup
wizard.
Setup files can be reinstalled.
Factory calibration
A firmware update does not affect the factory calibration.
Downgrade to a firmware version < 3.00
To downgrade the firmware from a version ≥ 3.00 to a version < 3.00, it is required to
uninstall the firmware using Windows 10 "Apps & features" (Windows 7 "Programs and
Features") before proceeding with the installation.
65Getting Started 1316.2407.02 ─ 62
R&S®ZND
7.2Remote operation in a LAN
A LAN connection is used to integrate the analyzer into a home/company network.
LAN connectivity offers several applications, e.g.:
●
Transfer data between a controller and the analyzer, e.g. to run a remote control
program.
●
Control the measurement from a remote computer using Remote Desktop or a similar application.
●
Use external network devices (e.g. printers).
The analyzer uses a user name and password as credentials for remote access; see
note on "User accounts and password protection"on page 19 for details. To protect the
analyzer from unauthorized access, it is recommended to change the factory setting.
7.2.1Assigning an IP address
Depending on the network capacities, the TCP/IP address information for the analyzer
can be obtained in different ways.
●
If the network supports dynamic TCP/IP configuration using the Dynamic Host
Configuration Protocol (DHCP), all address information can be assigned automatically.
●
If the network does not support DHCP, or if the analyzer is set to use alternate
TCP/IP configuration, the addresses must be set manually.
Administrative tasks
Remote operation in a LAN
By default, the analyzer is configured to use dynamic TCP/IP configuration and obtain
all address information automatically. Hence it is safe to establish a physical connection to the LAN without any previous analyzer configuration.
Manual TCP/IP configuration
If your network does not support DHCP, or if you choose to disable dynamic TCP/IP
configuration, you must assign valid address information before you connect the analyzer to the LAN. Contact your network administrator to obtain a valid IP address,
because connection errors can affect the entire network.
Admin account
You need administrator rights to change the TCP/IP configuration. See note on "User
accounts and password protection"on page 19 for details.
To enter the TCP/IP address information manually
1. Obtain the IP address and subnet mask for the analyzer and the IP address for the
local default gateway from your network administrator. If necessary, also obtain the
name of your DNS domain and the IP addresses of the DNS and WINS servers on
your network.
66Getting Started 1316.2407.02 ─ 62
R&S®ZND
Administrative tasks
Remote operation in a LAN
2. Press the Windows key in the SYSTEM keypad or on an external keyboard to
access Windows®.
3. Open the "Control Panel" > "Network and Sharing Center" > "Change adapter settings" dialog and right-click the local connection.
Windows 7: open the "Control Panel" > "Network and Internet" > "Network and
Sharing Center" > "Local Area Connection Status" dialog.
4. Select "Properties" and confirm the user account control message, depending on
your current user account.
●If your current account is an administrator account, select "Yes".
●If your account is an account with standard user rights, enter the password of
the administrator account and select "Yes".
Figure 7-1: User Account Control dialog
5. In the "Connection Properties" dialog opened, select "Internet Protocol Version 4
(TCP/IPv4) > Properties" and enter the IP address information.
67Getting Started 1316.2407.02 ─ 62
R&S®ZND
Administrative tasks
Remote operation in a LAN
7.2.2Using computer names
In a LAN that uses a DNS server (Domain Name System server), each PC or instrument connected in the LAN can be accessed via an unambiguous computer name
instead of the IP address. The DNS server translates the host name to the IP address,
which is especially useful when a DHCP server is used, as a new IP address can be
assigned each time the instrument is restarted.
Each instrument is delivered with an assigned computer name, but this name can be
changed. The default instrument name is a case-insensitive string with the following
syntax:
ZND-<serial_number>
The serial number can be found on the rear panel of the instrument. It is the third part
of the device ID printed on the bar code sticker:
ZND-102030, for example
7.2.3Remote Desktop connection
Remote Desktop is a Windows® application which you can use to access and control
the analyzer from a remote computer through a LAN connection. While the measurement is running, the analyzer screen contents are displayed on the remote computer,
and Remote Desktop provides access to all applications, files, and network resources
of the analyzer.
68Getting Started 1316.2407.02 ─ 62
R&S®ZND
Administrative tasks
Remote operation in a LAN
At the R&S ZND, by default remote connections are enabled using a local group policy
and remote access is granted to users instrument and administrator.
For detailed information about Remote Desktop and the connection, refer to the Windows® Help ("Windows Start Menu > Help and Support").
To establish a remote desktop connection, proceed as follows:
1. Connect remote Windows PC and VNA to the LAN and make sure that an IP connection can be established.
2. At the remote Windows PC, open a remote desktop connection (type RemoteDesktop Connection at the Windows Start/Search menu and hit enter).
3. In the "Remote Desktop Connection" dialog, enter the VNA's computer name or IP
address and select "Connect"
7.2.4
Windows
®
firewall settings
A firewall protects an instrument by preventing unauthorized users from gaining access
to it through a network. In the default configuration of the R&S ZND, the firewall is
enabled. A remote desktop connection does not require any changes in the firewall
settings.
Admin account
You need administrator rights to change the firewall settings. See note on "User
accounts and password protection"on page 19 above for details.
Some actions require a different firewall configuration, e.g.:
●
To transfer data with other hosts in the LAN, you have to allow "File and Printer
Sharing".
To change the firewall settings, proceed as follows:
1.
Access the operating system by pressing the Windows® key on the external keyboard. Open the "Control Panel".
2. Select "Windows Defender Firewall"
("System and Security > Windows Firewall" in Windows 7)
●Select "Allow an app or feature through Windows Defender Firewall" to enable
"File and Printer Sharing".
(Windows 7: select "Change Settings > Allow a program or feature through
Windows Firewall" to enable "File and Printer Sharing").
●Select "Turn Windows Firewall on or off" to enable or disable the firewall.
Confirm the user account control message to allow the desired changes (see Fig-
ure 7-1). For detailed information about the firewall, refer to the Windows® Help.
69Getting Started 1316.2407.02 ─ 62
R&S®ZND
Contacting customer support
8Contacting customer support
Technical support – where and when you need it
For quick, expert help with any Rohde & Schwarz product, contact our customer support center. A team of highly qualified engineers provides support and works with you
to find a solution to your query on any aspect of the operation, programming or applications of Rohde & Schwarz products.
Contact information
Contact our customer support center at www.rohde-schwarz.com/support, or follow this
QR code:
Figure 8-1: QR code to the Rohde & Schwarz support page
70Getting Started 1316.2407.02 ─ 62
R&S®ZND
Index
Index
A
AC supply .......................................................................... 17