List of Commands..............................................................................526
6User Manual 1326.1578.02 ─ 15
R&S®Scope Rider RTH
1For Your Safety
To prevent electric shock, personal injury, or fire caused by high voltages, carefully
read and observe the following safety information:
●
This chapter, the first part of the safety booklet, provides warnings and safety information for usage of oscilloscopes, probes, and other measurement accessories.
●
The second part of the safety booklet provides basic safety instructions
●
Product manuals provide specific safety instructions for procedures and measurement accessories.
Electrical safety
●
Voltages higher than 30 V RMS, or 42 V peak, or 60 V DC are regarded as hazardous contact voltages. When working with hazardous contact voltages, use protective measures to preclude direct contact with the measurement setup:
–Do not touch exposed connections and components when power is applied.
–Use only insulated voltage probes, test leads and adapters.
●
Make sure that only personnel familiar with the potential risks of measuring electrical quantities controls the instrument. Observe applicable local or national safety
regulations and rules for the prevention of accidents.
●
Use only specified probes and accessories that comply with the measurement category (CAT) of your measurement task. If you use other than Rohde & Schwarz
accessories, make sure that they are suitable for the instrument and the measurement task.
●
Observe all voltage and current ratings of the instrument, the probes, and the
accessories. The lowest rated component defines the rating of the complete measurement setup. Limits and ratings are marked on the products and listed in the
data sheets.
For probes, consider that the rated voltage depends on the frequency. The voltage
limitation curves are provided in the data sheet. Do not exceed these two ratings:
–Maximum measurement voltage from the probe tip to the probe reference lead.
–Maximum floating voltage from the probe reference lead to earth ground.
●
Set the correct attenuation factor on the instrument according to the probe being
used. Otherwise, the measurement results do not reflect the actual voltage level,
and you might misjudge the actual risk.
●
Set up all connections to the instrument before applying power.
●
Do not open the instrument casing.
●
Do not use the instrument if the instrument casing, the display or any probe or
accessory are damaged. If you detect or suspect any damage, have the instrument
or accessory inspected by qualified service personnel.
●
Do not operate the instrument in wet, damp or explosive atmospheres. Make sure
that all connectors are completely dry before connecting the inputs.
●
Observe the operating conditions specified in the data sheet and in the basic safety
instructions.
For Your Safety
7User Manual 1326.1578.02 ─ 15
R&S®Scope Rider RTH
Battery replacement
●
Disconnect power supply, probes, test leads and all other cables before opening
the battery cover.
●
Use only the specified Li-ion battery pack, which is delivered with the instrument.
You can order additional battery packs at Rohde & Schwarz, see data sheet for
order number.
●
Do not operate the instrument with the battery cover open.
●
Use only the specified power adapter, which is delivered with the instrument.
Cleaning
Remove all probes, leads, USB and LAN cables and power supply before cleaning the
instrument. Use only cleaning materials that are specified in the manual.
Measurement categories
IEC 61010-2-030 defines measurement categories that rate instruments on their ability
to resist short transient overvoltages that occur in addition to the working voltage. Use
the measurement instrument and accessories only in electrical environments for which
they are rated.
For Your Safety
●
O - Instruments without rated measurement category
For measurements performed on circuits not directly connected to mains, for example, electronics, circuits powered by batteries, and specially protected secondary
circuits. This measurement category is also known as CAT I.
●
CAT II:
For measurements performed on circuits directly connected to the low-voltage
installation by a standard socket outlet, for example, household appliances and
portable tools.
●
CAT III:
For measurements performed in the building installation, such as junction boxes,
circuit breakers, distribution boards, and equipment with permanent connection to
the fixed installation.
●
CAT IV:
For measurements performed at the source of the low-voltage installation, such as
electricity meters and primary overcurrent protection devices.
8User Manual 1326.1578.02 ─ 15
R&S®Scope Rider RTH
2Getting Started
2.1Preface
2.1.1Key Features
The R&S RTH is the perfect multi-purpose tool for the lab and in the field. Outstanding
key features are:
●
Full isolation of all channels and interfaces
●
CAT IV 600 V / CAT III 1000 V safety rating
●
Bandwidth 60 MHz to 500 MHz with 5 GS/s sampling rate
●
Acquisition speed up to 50.000 waveforms per second
●
2 mV/div sensitivity
●
Up to 200 V offset range
●
33 automatic measurement functions
●
Full operation using touch or keypad
●
Wireless LAN and Ethernet for web based remote control and quick data access
(optional)
Getting Started
Preface
The R&S RTH combines:
●
Lab performance oscilloscope
●
Logic analyzer with 8 digital inputs (optional)
●
Protocol analyzer with trigger and decode (optional)
●
Data logger
●
Digital multimeter (R&SRTH1002)
2.1.2Input Isolation
The instrument has independently floating isolated inputs. Each input channel has its
own signal input and its own reference input. Each input channel is electrically isolated
from the other input channels. Therefore, each reference of the used inputs must be
connected to a reference voltage. Furthermore, input channels are electrically isolated
from the communication ports and the power adapter input.
9User Manual 1326.1578.02 ─ 15
R&S®Scope Rider RTH
Getting Started
Preface
Figure 2-1: Isolation scheme of the R&S RTH
The input isolation has several advantages:
●
You can measure independently floating signals simultaneously.
●
The risk of causing a short circuit while measuring multiple signals is reduced substantially.
●
When measuring signals with different grounds, the induced ground currents are
kept to a minimum.
2.1.3Documentation Overview
The user documentation for the R&SRTH consists of the following parts:
●
Instrument Help
The instrument help is part of the instrument's firmware. It offers quick, contextsensitive access to the complete information directly on the instrument.
●
Basic Safety Instructions
This brochure provides safety instructions and operating conditions and further
important information. The brochure is delivered with the instrument in printed form.
●
Getting Started
The Getting Started manual provides the information needed to set up and start
working with the instrument, and describes basic operations. The English edition of
this manual is delivered with the instrument in printed form. Editions in other languages and the newest version of the English manual are available on the product
website.
10User Manual 1326.1578.02 ─ 15
R&S®Scope Rider RTH
●
User Manual
The user manual describes all instrument modes and functions in detail. It also provides an introduction to remote control and a complete description of the remote
control commands with programming examples. The newest version of the manual
is available in English on the R&S RTH product website at www.rohde-
schwarz.com/manual/rth.
●
Data Sheet
The data sheet contains the complete instrument specification. It also lists the
options and their order numbers, and optional accessories. The data sheet is available on the R&S RTH product website at www.rohde-schwarz.com/brochure-data-
sheet/rth.
●
Calibration Certificate
The document is available on https://gloris.rohde-schwarz.com/calcert.
●
Open Source Acknowledgment
The Open Source Acknowledgment document provides verbatim license text of
open source software that is used in the instrument's firmware. It is available on the
R&S RTH website at www.rohde-schwarz.com/firmware/rth, and it can be read
directly on the instrument.
●
Instrument security procedures manual
Provides information on security issues when working with the R&S RTH in secure
areas.
●
Application cards and application notes
These documents deal with special applications or background information on particular topics. See www.rohde-schwarz.com/application/rth
Getting Started
Preparing for Use
2.2Preparing for Use
This section describes the basic steps to be taken when setting up the R&S RTH for
the first time.
11User Manual 1326.1578.02 ─ 15
R&S®Scope Rider RTH
Shock hazard caused by high voltages
The instrument must be used in an appropriate manner to prevent electric shock, fire,
personal injury, or damage.
●
Do not open the instrument casing.
●
Do not use the instrument if the instrument casing, the display or any probe or
accessory are damaged. If you detect or suspect any damage, have the instrument
or accessory inspected by qualified service personnel.
●
Use only specified probes and accessories that comply with the measurement category of your measurement task.
If you use other than Rohde & Schwarz accessories, make sure that they are suitable to the instrument and the measurement task.
●
Do not operate the instrument in wet, damp or explosive atmospheres.
Make sure that all connectors are completely dry before connecting the inputs.
●
Voltages higher than 30 V RMS or 42 V peak or 60 V DC are regarded as hazardous contact voltages. Make sure that only electrically skilled persons use the
R&S RTH for measurements on hazardous contact voltages. These working conditions require special education and experience to perceive risks and to avoid hazards which electricity can create.
●
Observe the operating conditions specified in the data sheet.
●
Read and observe the "Safety Instructions" delivered as a printed brochure with the
instrument. Also read and observe the safety instructions in the following sections.
Getting Started
Preparing for Use
2.2.1Unpacking the Instrument
When you receive your shipping package, unpack and inspect the package and its
contents for damage.
1. Inspect the package for damage.
If the packaging material shows any signs of stress, notify the carrier as well as
your Rohde & Schwarz service center. Keep the package and cushioning material
for inspection. Keep a damaged package and the cushioning material until the contents have been checked for completeness and the instrument has been tested.
2. Unpack the handheld scope and the accessories and check the contents for completeness, see "Package contents"on page 13.
If anything is missing, contact your Rohde & Schwarz service center.
3. Inspect the handheld scope and the accessories.
If there is any damage or defect, or if the R&S RTH does not operate properly,
notify your Rohde & Schwarz service center.
12User Manual 1326.1578.02 ─ 15
R&S®Scope Rider RTH
Packing material
Retain the original packing material. If the instrument needs to be transported or shipped later, you can use the material to protect the control elements and connectors.
Package contents
The delivery package contains the following items:
●
R&SRTH handheld oscilloscope
●
4 Gbyte microSD card, installed in the battery compartment
●
Power adapter with cable and adapter set for various socket types
●
Battery pack
●
R&S RT-ZI10 probes (2x for R&SRTH1002; 4x for R&SRTH1004)
●
DMM test leads (only for R&SRTH1002)
●
Hand strap, attached on the handheld scope
●
Printed "Getting Started" manual and "Basic Safety Instructions" brochure
Getting Started
Preparing for Use
Optional accessories and their order numbers are listed in the data sheet.
2.2.2Inserting and Charging the Battery
Before you can use the handheld oscilloscope for the first time, insert the battery pack
and charge it.
Risk of electrical shock during battery replacement
●
Disconnect power supply, probes, test leads and all other cables before opening
the battery cover.
●
Use only the specified Li-Ion battery pack, which is delivered with the instrument.
You can order additional battery packs at Rohde & Schwarz, see Data Sheet for
order number.
●
Do not operate the instrument with the battery cover open.
●
Use only the specified power adapter, which is delivered with the instrument.
13User Manual 1326.1578.02 ─ 15
R&S®Scope Rider RTH
Getting Started
Preparing for Use
1. Turn off the instrument power. Remove power supply, probes, test leads and all
other cables.
2. Fold out the tilt stand on the back of the instrument.
3. Screw open the battery cover.
4. Insert the battery pack.
5. Screw down the battery cover.
6. Connect the power adapter to the connector on the left side of the scope, and fully
charge the battery. Charging may take a few hours.
14User Manual 1326.1578.02 ─ 15
R&S®Scope Rider RTH
Getting Started
Preparing for Use
If the instrument is on, the battery status is shown on the display.
Replace used batteries periodically by new batteries after 24 months of usage.
Observe the safety regulations in the "Batteries and rechargeable batteries/cells" chapter in the "Basis Safety Instructions" brochure, which is delivered with the instrument.
2.2.3Powering On/Off
►
Press the
The key blinks and turns green after a few seconds.
Table 2-1: Colors of the Power key
GreenPower is on
BlueCharging the battery, power is off
Orange (yellow)Battery is fully charged, power adapter is connected, power is off
If you do not use the instrument for a longer time, the battery gets exhausted. When
you connect the power supply and switch on the instrument with exhausted battery, it
takes a few minutes until the instrument can start.
[Power] key to switch the instrument on or off.
2.2.4Using the Tilt Stand
The R&S RTH has a tilt stand for proper handling while the scope is placed on a table.
15User Manual 1326.1578.02 ─ 15
R&S®Scope Rider RTH
► Pull the tilt stand as shown below.
Getting Started
Preparing for Use
2.2.5EMI Suppression
Electromagnetic Interference (EMI) may affect the measurement results.
To suppress generated Electromagnetic Interference:
●
Use suitable shielded cables of high quality. For example use double-shielded RF
and LAN cables.
●
Always terminate open cable ends.
●
Note the EMC classification in the data sheet.
16User Manual 1326.1578.02 ─ 15
R&S®Scope Rider RTH
2.3Instrument Tour
2.3.1Front View
Getting Started
Instrument Tour
Figure 2-2: Front panel of the R&S RTH1002
1 = Touch display
2 = Waveform setup with [AUTOSET], reset to default with [PRESET]
3 = Analysis functions
4 = Mode selection
5 = Save/Recall
6 = Instrument settings
7 = Power on/off
8 = Navigation controls
9 = Horizontal settings
10 = Run/stop acquisition and trigger settings
11 = Acquisition settings
12 = Screenshot and documentation output
13 = Channels and vertical settings
14 = Multimeter measurements
17User Manual 1326.1578.02 ─ 15
R&S®Scope Rider RTH
Getting Started
Instrument Tour
Figure 2-3: Front panel of the R&S RTH1004
1 = Touch display
2 = Waveform setup with [AUTOSET], reset to default with [PRESET]
3 = Analysis functions
4 = Mode selection
5 = Save/Recall
6 = Instrument settings
7 = Power on/off
8 = Navigation controls
9 = Horizontal settings
10 = Run/stop acquisition and trigger settings
11 = Acquisition settings
12 = Screenshot and documentation output
13 = Channels and vertical settings
For a description of the keys, see Chapter 2.4.1.3, "Using Front Panel Keys",
on page 30.
18User Manual 1326.1578.02 ─ 15
R&S®Scope Rider RTH
2.3.2Top View
The R&S RTH1002 has two BNC input connectors CH1 and CH2, and two 4 mm
banana plug inputs for multimeter measurements. The channel inputs have double
channel-to-channel isolation that allows for independent floating measurements at
each input. The DMM input is fully isolated from scope inputs, interfaces, and ground.
Figure 2-4: Top view of R&S RTH1002
The R&S RTH1004 has four BNC input connectors CH1, CH2, CH3, CH4. The channel
inputs have double channel-to-channel isolation that allows for independent floating
measurements at each input.
Getting Started
Instrument Tour
Figure 2-5: Top view of R&S RTH1004
Shock hazard caused by high voltages
To avoid electrical shock and personal injury, and to prevent damage to the instrument
or any other products connected to it, observe the following:
●
Do not apply input voltages above the rating of the instrument and the accessories.
●
Use only probes, test leads, and adapters that comply with the measurement category (CAT) of your measurement task.
●
Test leads and measurement accessories used for multimeter measurements on a
live mains circuit must be rated for CAT III or CAT IV according to IEC 61010-031.
The voltage of the measured circuit must not exceed the rated voltage value.
Maximum input voltage:
●
At BNC inputs: CATIV300V
●
With probe R&S RT-ZI10 or R&S RT-ZI11: CATIV600V, CATIII1000V
●
Meter input: CAT IV 600 V; CAT III 1000 V
Voltage ratings: V RMS (50 to 60 Hz) for AC sine wave and V DC for DC applications.
19User Manual 1326.1578.02 ─ 15
R&S®Scope Rider RTH
Risk of electrical shock or fire
Voltages higher than 30 V RMS or 42 V peak or 60 V DC are regarded as hazardous
contact voltages. When working with hazardous contact voltages, use appropriate protective avoid electrical shock and injuries:
●
Use only insulated probes, cables, test leads and adapters.
●
Do not touch voltages higher than 30 V RMS or 42 V peak or 60 V DC.
See also: Chapter 2.1.2, "Input Isolation", on page 9.
2.3.3Right View
Getting Started
Instrument Tour
1
= LAN
2 = USB type B for remote control
3 = Probe compensation
4 = USB type A for flash drive
5 = Logic probe connector
Risk of injury or instrument damage
Always close the lids of the communication ports and DC input when they are not in
use.
LAN connector
RJ-45 connector to connect the instrument to a Local Area Network (LAN). It supports
up to 100 Mbit/s.
USB type A connector
USB type A connector to connect a USB flash drive to store and reload instrument settings and measurement data.
20User Manual 1326.1578.02 ─ 15
R&S®Scope Rider RTH
USB type B connector (mini USB)
Mini USB connector to connect a computer for remote control of the instrument.
Probe compensation
Probe compensation terminal to support adjustment of passive probes to the oscilloscope channel.
Logic probe connector
Input for the logic probe R&S RT-ZL04. Logic analysis requires Mixed Signal Option
R&S RTH-B1, which includes the logic probe R&S RT-ZL04.
Risk of electrical shock - no CAT rating for MSO measurements
The logic probe R&S RT-ZL04 is not rated for any measurement category. To avoid
electrical shock or personal injury, and to prevent material damage, make sure that the
ground clips of the R&S RT-ZL04 are connected to protective earth on the DUT.
Getting Started
Instrument Tour
2.3.4Left View
= DC input
1
2 = Kensington lock slot
DC input
Connector for the power adapter to charge the battery.
Kensington lock slot
The Kensington lock is used to secure the instrument against theft.
21User Manual 1326.1578.02 ─ 15
R&S®Scope Rider RTH
2.3.5Rear View
Getting Started
Instrument Tour
1 = Tilt stand to fold out
2 = Thread hole M5
3 = Battery compartment
2.3.6Display Overview
In the most important modes scope, mask and XY, the display shows the following
information.
22User Manual 1326.1578.02 ─ 15
R&S®Scope Rider RTH
Getting Started
Operating Basics
1 = Measurement results, depends on the mode and the selected measurement
2 = Time scale (horizontal scale, in s/division)
3 = Trigger type, trigger source, and trigger mode
4 = Capture status
5 = Battery status and AC connectivity for battery charging; date and time
6 = Trigger level marker, has the color of the trigger source
7 = Trigger position marker, has the color of the trigger source
8 = Channel markers indicate the ground levels. Channel C3 has the focus
9 = Vertical settings for each active channel: vertical scale (vertical sensitivity, in V/division), bandwidth limit
(no indicator = full bandwidth, BW= limited frequency), coupling (AC or DC)
10 = Logic channels (MSO R&S RTH-B1)
11 = Menu button
You can adjust the vertical position of each waveform, the trigger level, and the trigger
position by dragging the corresponding marker on the display. Alternatively, tap a
marker to set the focus, and use the wheel to adjust position.
2.4Operating Basics
2.4.1Accessing the Functionality
The complete functionality is available in the menus and dialogs on the touchscreen.
You can touch the functions directly on the display, or you can use the wheel to navi-
23User Manual 1326.1578.02 ─ 15
R&S®Scope Rider RTH
gate and select. In addition, the most important functions are applied to the keys on the
front panel to set up and perform measurement tasks quickly.
2.4.1.1Using the Touchscreen
Using the touchscreen of the R&S RTH is as easy as using your mobile phone. To
open the menu, tap the "Menu" button - that is the R&S logo in the right bottom corner
of the display.
Getting Started
Operating Basics
Figure 2-6: Open the menu and select a menu item
24User Manual 1326.1578.02 ─ 15
R&S®Scope Rider RTH
Getting Started
Operating Basics
Figure 2-7: Switch on or off (left) and select a parameter value (right)
Figure 2-8: Enter numerical value and unit
2.4.1.2Using the Navigation Wheel
In addition or alternatively to the touchscreen, you can use the wheel to control the
R&S RTH.
25User Manual 1326.1578.02 ─ 15
R&S®Scope Rider RTH
When using the wheel, always observe the position of the focus - the orange frame or
other highlighting that marks the active object on the screen.
●
If the focus is on the menu button or somewhere in the menu or dialogs:
–Turn the wheel to move the focus.
–Press the wheel button to apply the selection.
●
If the focus is on an element in the diagram, for example, on a waveform, cursor
line, or trigger level:
–Turn the wheel to change the position of the active element.
–Press the wheel button to toggle the active element, for example, to toggle the
The [BACK] key closes open dialogs and menus, and resets the focus to the "Menu"
button.
Menu navigation
The following procedure describes how to access and navigate the menu. Navigating
dialogs and selecting parameter values works in the same way. See also Figure 2-9.
Getting Started
Operating Basics
cursor lines, or zoom size and zoom position.
1. Press [BACK] until the focus is on the "Menu" button.
2. Press the wheel button to open the menu.
3. Turn the wheel to move the focus to the required menu item.
4. Press the wheel button to open the dialog, submenu, or keypad for the selected
menu item.
26User Manual 1326.1578.02 ─ 15
R&S®Scope Rider RTH
Getting Started
Operating Basics
Figure 2-9: Open the menu and select a menu item
Set numerical value using the wheel
1. Set the focus to the required setting, and press the wheel button once.
2. Turn the wheel until the required value is shown.
3. Press [BACK].
27User Manual 1326.1578.02 ─ 15
R&S®Scope Rider RTH
Getting Started
Operating Basics
Figure 2-10: Set numerical value using the wheel
Data entry using wheel and keypad
You can enter precise numerical values on the keypad. See also Figure 2-11.
1. Set the focus to the required setting, and press the wheel button twice.
2. Turn the wheel until the focus is on the required number.
3. Press the wheel button.
4. Turn the wheel until the focus is on the required unit.
5. Press the wheel button.
28User Manual 1326.1578.02 ─ 15
R&S®Scope Rider RTH
Getting Started
Operating Basics
Figure 2-11: Enter numerical value and unit in the keypad
The [SHIFT] button toggles the wheel focus in the keypad. If the focus is on the entry
field, turning the wheel changes the value. If the focus is in the lower part, the wheel
selects numbers and unit.
29User Manual 1326.1578.02 ─ 15
R&S®Scope Rider RTH
2.4.1.3Using Front Panel Keys
For an overview of the front panel keys, see Figure 2-3
KeyPress shortlyPress and hold
Getting Started
Operating Basics
[AUTOSET] analyses the active channels,
adjusts the instrument settings, and displays
stable waveforms.
[PRESET] sets the instrument to the default
factory state.
[MEAS] starts or stops the last configured automatic measurements.
[ZOOM] enables or disables the zoom with the
last configuration.
If the zoom is on but not in focus, pressing the
key focuses the zoom.
[CURSOR] starts or stops the last configured
cursor measurement.
If the cursor is on but not in focus, pressing the
key sets the focus to the first cursor line.
[MATH] switches the math waveform on or off.Opens or closes the "Math"
Requires logic analyzer option R&S RTH-B1
(MSO).
The effect depends on the state of digital channels:
If the all digital channels are off, the key
switches them on and sets the focus.
If the digital channels are on but not in focus,
the key sets the focus.
If the focus is on digital channels, the key
switches them off.
Opens or closes the "Meas"
dialog to configure the measurements.
Opens or closes the "Zoom"
dialog to configure the zoom
scale and position.
Opens or closes the "Cursor"
dialog to configure the measurement.
dialog to configure the math
waveform.
Opens or closes the "Logic"
dialog to configure digital channels.
Activates or deactivates the serial bus.
Requires at least one serial triggering and
decoding option. Available options are listed in
the data sheet.
[SHIFT] opens a dialog to save and load instrument settings.
If a dialog or menu is open, [BACK] closes it. If the menu is closed, the key toggles the focus between focused element in the diagram and the Menu button.
Opens or closes the "Bus" dialog to configure serial protocols.
Press and hold for 2 seconds
to disable or enable the touchscreen.
30User Manual 1326.1578.02 ─ 15
Loading...
+ 510 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.