B.5Reset Values of the Status Reporting System....................................................... 792
Contents
List of Commands..............................................................................793
15User Manual 1335.7898.02 ─ 05
R&S®RTA4000
Contents
16User Manual 1335.7898.02 ─ 05
R&S®RTA4000
1Preface
1.1Safety Information
Preface
Safety Information
The R&S RTA4000 digital oscilloscope is designed for measurements on circuits that
are only indirectly connected to the mains or not connected at all. It is not rated for any
measurement category.
The instrument is rated for pollution degree 2 - for indoor, dry location use where only
non-conductive pollution occurs. Temporary conductivity caused by condensation is
possible.
The instrument is intended for use in industrial areas. When used in residential areas,
radio disturbances caused by the instrument can exceed given limits. Additional shielding can be required.
The instrument must be controlled by personnel familiar with the potential risks of measuring electrical quantities. Observe applicable local or national safety regulations and
rules for the prevention of accidents.
Safety information is part of the product documentation. It warns you about the potential dangers and gives instructions how to prevent personal injury or damage caused
by dangerous situations. Safety information is provided as follows:
●
The "Basic Safety Instructions" in different languages are delivered as a printed
brochure with the instrument.
●
Throughout the documentation, safety instructions are provided when you need to
take care during setup or operation.
Risk of injury
Use the instrument in an appropriate manner to prevent electric shock, personal injury,
or fire:
●
Do not open the instrument casing.
●
Do not use the instrument if you detect or suspect any damage of the instrument or
accessories.
●
Do not operate the instrument in wet, damp or explosive atmospheres.
●
Make sure that the instrument is properly grounded.
●
Do not use the instrument to ascertain volt-free state.
●
Do not exceed the voltage limits given in Chapter 2.2.1.1, "Input Connectors",
on page 25.
17User Manual 1335.7898.02 ─ 05
R&S®RTA4000
Preface
Documentation Overview
Risk of instrument damage due to inappropriate operating conditions
An unsuitable operating site or test setup can damage the instrument and connected
devices. Before switching on the instrument, observe the information on appropriate
operating conditions provided in the data sheet. In particular, ensure the following:
●
All fan openings are unobstructed and the airflow perforations are unimpeded. A
minimum distance of 10 cm to other objects is recommended.
●
The instrument is dry and shows no sign of condensation.
●
The instrument is positioned as described in the following sections.
●
The ambient temperature does not exceed the range specified in the data sheet.
●
Signal levels at the input connectors are all within the specified ranges.
●
Signal outputs are connected correctly and are not overloaded.
Electromagnetic interference (EMI) may affect the measurement results.
To suppress generated electromagnetic interference (EMI):
●
Use suitable shielded cables of high quality. For example, use double-shielded RF
and LAN cables.
●
Always terminate open cable ends.
●
Note the EMC classification in the data sheet.
1.2Documentation Overview
This section provides an overview of the R&S RTA4000 user documentation.
1.2.1Manuals and Instrument Help
You find the manuals on the product page at:
www.rohde-schwarz.com/manual/rta4000
Getting started manual
Introduces the R&S RTA4000 and describes how to set up the product. A printed English version is included in the delivery.
User manual
Contains the description of all instrument modes and functions. It also provides an
introduction to remote control, a complete description of the remote control commands
with programming examples, and information on maintenance and instrument interfaces. Includes the contents of the getting started manual.
18User Manual 1335.7898.02 ─ 05
R&S®RTA4000
Preface
Documentation Overview
The online version of the user manual provides the complete contents for immediate
display on the internet.
Instrument help
The help offers quick, context-sensitive access to the functional description directly on
the instrument.
Basic safety instructions
Contains safety instructions, operating conditions and further important information.
The printed document is delivered with the instrument.
Instrument security procedures manual
Deals with security issues when working with the R&S RTA4000 in secure areas.
Service manual
Describes the performance test for checking the rated specifications, module replacement and repair, firmware update, troubleshooting and fault elimination, and contains
mechanical drawings and spare part lists. The service manual is available for registered users on the global Rohde & Schwarz information system (GLORIS, https://
gloris.rohde-schwarz.com).
1.2.2Data Sheet and Brochure
The data sheet contains the technical specifications of the R&S RTA4000. It also lists
the options with their order numbers and optional accessories. The brochure provides
an overview of the instrument and deals with the specific characteristics.
See www.rohde-schwarz.com/brochure-datasheet/rta4000
1.2.3Calibration Certificate
The document is available on https://gloris.rohde-schwarz.com/calcert. You need the
device ID of your instrument, which you can find on a label on the rear panel.
1.2.4Release Notes and Open Source Acknowledgment
The release notes list new features, improvements and known issues of the current
firmware version, and describe the firmware installation. The open source acknowledgment document provides verbatim license texts of the used open source software.
See www.rohde-schwarz.com/firmware/rta4000. The open source acknowledgment
document can also be read directly on the instrument.
19User Manual 1335.7898.02 ─ 05
R&S®RTA4000
1.3Conventions Used in the Documentation
1.3.1Typographical Conventions
Preface
Conventions Used in the Documentation
The following text markers are used throughout this documentation:
ConventionDescription
"Graphical user interface elements"
[Keys]Key and knob names are enclosed by square brackets.
Filenames, commands,
program code
InputInput to be entered by the user is displayed in italics.
LinksLinks that you can click are displayed in blue font.
"References"References to other parts of the documentation are enclosed by quota-
All names of graphical user interface elements on the screen, such as
dialog boxes, menus, options, buttons, and softkeys are enclosed by
quotation marks.
Filenames, commands, coding samples and screen output are distinguished by their font.
tion marks.
1.3.2Conventions for Procedure Descriptions
When operating the instrument, several alternative methods may be available to perform the same task. In this case, the procedure using the touchscreen is described.
Any elements that can be activated by touching can also be clicked using an additionally connected mouse. The alternative procedure using the keys on the instrument or
the on-screen keyboard is only described if it deviates from the standard operating procedures.
The term "select" may refer to any of the described methods, i.e. using a finger on the
touchscreen, a mouse pointer in the display, or a key on the instrument or on a keyboard.
1.3.3Notes on Screenshots
When describing the functions of the product, we use sample screenshots. These
screenshots are meant to illustrate as many as possible of the provided functions and
possible interdependencies between parameters. The shown values may not represent
realistic usage scenarios.
The screenshots usually show a fully equipped product, that is: with all options installed. Thus, some functions shown in the screenshots may not be available in your particular product configuration.
20User Manual 1335.7898.02 ─ 05
R&S®RTA4000
2Getting Started
2.1Preparing for Use
2.1.1Unpacking and Checking the Instrument
Getting Started
Preparing for Use
1. Inspect the package for damage.
If the packaging material shows any signs of stress, notify the carrier who delivered
the instrument.
2. Carefully unpack the instrument and the accessories.
3. Check the equipment for completeness. See section "Delivery contents"
on page 21.
4. Check the equipment for damage.
If there is damage, or anything is missing, immediately contact the carrier as well
as your distributor. Make sure not to discard the box and packing material.
Packing material
Retain the original packing material. If the instrument needs to be transported or shipped later, you can use the material to protect the control elements and connectors.
Delivery contents
The delivery package contains the following items:
●
R&SRTA4000 digital oscilloscope
●
R&S RT-ZP10 probes (4x)
●
Country-specific power cable
●
Printed "Getting Started" manual
●
Printed "Basic Safety Instructions" brochure
2.1.2Positioning the Instrument
The instrument is designed for use under laboratory conditions. It can be used in
standalone operation on a bench top or can be installed in a rack.
For standalone operation, place the instrument on a horizontal bench with even, flat
surface. The instrument can be used in horizontal position, or with the support feet on
the bottom extended.
21User Manual 1335.7898.02 ─ 05
R&S®RTA4000
Getting Started
Preparing for Use
The instrument can be installed in a 19" rack mount using a rack mount kit. The order
number of the rack mount kit is given in the data sheet. The installation instructions are
part of the rack mount kit.
Risk of injury if feet are folded out
The feet can fold in if they are not folded out completely or if the instrument is shifted.
This can cause damage or injury.
●
Fold the feet completely in or out to ensure stability of the instrument. Never shift
the instrument when the feet are folded out.
●
When the feet are folded out, do not work under the instrument or place anything
underneath.
●
The feet can break if they are overloaded. The overall load on the folded-out feet
must not exceed 200 N.
F
max
Risk of instrument damage due to overheating
An insufficient airflow can cause the R&S RTA4000 to overheat, which can impair the
measurement results, disturb the operation, and even cause damage.
●
Ensure that all fan openings are unobstructed and that the airflow perforations are
unimpeded. The minimum distance to a wall is 10 cm.
●
When placing several instruments side by side, keep a minimum distance of 20 cm
between the instruments. Ensure that the instruments do not draw in the preheated
air from their neighbors.
●
When mounting the instrument in a rack, observe the instructions of the rack manufacturer to ensure sufficient airflow and avoid overheating.
2.1.3Starting the Instrument
The R&S RTA4000 can be used with different AC power voltages and adapts itself
automatically to it.
The nominal ranges are:
●
100 V to 240 V AC at 50 Hz to 60 Hz
●
1.6 A to 0.7 A
22User Manual 1335.7898.02 ─ 05
R&S®RTA4000
Getting Started
Preparing for Use
●
max. 160 W
Risk of injury
Connect the instrument only to an outlet that has a ground contact.
Do not use an isolating transformer to connect the instrument to the AC power supply.
To start the instrument
1. Connect the power cable to the AC power connector on the rear panel of the
R&S RTA4000.
2. Connect the power cable to the socket outlet.
3. Switch the main power switch at the rear of the instrument to position I.
The [Standby] key lights up. The key is located in the bottom left corner of the front
panel.
4. Press the [Standby] key.
The instrument performs a system check and starts the firmware.
Warm-up and prepare the instrument
Make sure that the instrument has been running and warming up before you start the
self-alignment and the measurements. The minimum warm-up time is about 30 min.
To power off the instrument
1. Press the [Standby] key.
All current settings are saved, and the software shuts down. All data transfers and
running processes are interrupted.
2. Switch the main power switch at the rear of the instrument to position 0.
3. Disconnect the AC power cable from the AC power supply.
Overview of power switch and [Standby] key actions
ActionConditionResult[Standby]
Set power switch to I.[Standby] key was off
when switching power
switch to 0.
Instrument is in standby mode.Yellow
[Standby] key was on
when switching power
switch to 0.
Switch [Standby] on.Power switch is on.
Instrument performs system check
and boots the firmware. It is ready
for operation.
Green
23User Manual 1335.7898.02 ─ 05
R&S®RTA4000
Getting Started
Instrument Tour
ActionConditionResult[Standby]
Switch [Standby] off.Power switch is on.Software shuts down. All instru-
Set power switch to 0.Instrument is working,
Set power switch to 0.Instrument is in standby
2.1.4Replacing the Fuse
The instrument is protected by a fuse. You can find it on the rear panel between the
main power switch and AC power supply.
Type of fuse: Size 5x20 mm, 250V~, T3.15H (slow-blow), IEC60127-2/5
[Standby] is Green.
mode, [Standby] is Yellow.
Yellow
ment settings are saved, running
data transfers and processes are
interrupted (e.g., self-alignment).
Instrument is in standby mode.
Software shuts down. All instrument settings are saved, running
data transfers and processes are
interrupted (e.g., self-alignment).
No power on the instrument.
No power on the instrument.Off
Off
Risk of electric shock
The fuse is part of the main power supply. Therefore, handling the fuse while power is
on can lead to electric shock. Before opening the fuse holder, make sure that the
instrument is switched off and disconnected from all power supplies.
Always use fuses supplied by Rohde & Schwarz as spare parts, or fuses of the same
type and rating.
1. Pull the fuse holder out of its slot on the rear panel.
2. Exchange the fuse.
3. Insert the fuse holder carefully back in its slot until it latches.
2.2Instrument Tour
2.2.1Front View
Figure 2-1 shows the front panel of the R&S RTA4000. The function keys are grouped
in functional blocks to the right of the display.
24User Manual 1335.7898.02 ─ 05
R&S®RTA4000
Getting Started
Instrument Tour
Figure 2-1: Front panel of R&S RTA4000 with 4 input channels
1 = Display
2 = Horizontal and vertical setup controls
3 = Trigger settings, action and analysis controls
4 = Analog input channels (BNC)
5 = External trigger input
6 = Connectors for demo signal output
7 = Connector for optional function generator output (BNC, R&S RTA-B6)
8 = Connectors for optional pattern generator (R&S RTA-B6)
9 = Connectors for probe compensation
10 = USB connector
11 = [Standby] key
2.2.1.1Input Connectors
BNC inputs (4 and 5)
The R&S RTA4000 has two or four channel inputs (4) to connect the input signals. The
external trigger input (5) is used to control the measurement by an external signal. The
trigger level can be set from -5 V to 5 V.
25User Manual 1335.7898.02 ─ 05
R&S®RTA4000
Getting Started
Instrument Tour
For channel connectors, the input impedance is selectable, the values are 50 Ω and
1 MΩ.
Risk of electrical shock - maximum input voltages
The maximum input voltage on channel inputs must not exceed:
●
400V(peak) and 300V(RMS) at 1MΩ input impedance
●
30 V (peak) and 5 V (RMS) at 50 Ω input impedance
For the external trigger input, the maximum input voltage is 400 V (peak) and
300 V (RMS) at 1 MΩ input impedance.
Transient overvoltages must not exceed 400 V (peak).
For further specifications, refer to the data sheet.
Voltages higher than 30 V (RMS) or 42 V (peak) or 60 V DC are regarded as hazard-
ous contact voltages. When working with hazardous contact voltages, use appropriate
protective measures to preclude direct contact with the measurement setup:
●
Use only insulated voltage probes, test leads and adapters.
●
Do not touch voltages higher than 30V(RMS) or 42V(peak) or 60VDC.
Risk of injury and instrument damage
The instrument is not rated for any measurement category. When measuring in circuits
with transient overvoltages of category II, III or IV circuits, make sure that no such
overvoltages reach the R&S RTA4000 input. Therefore, use only probes that comply
with DIN EN 61010-031. When measuring in category II, III or IV circuits, always insert
a probe that appropriately reduces the voltage so that no transient overvoltages higher
than 400 V (peak) are applied to the instrument. For detailed information, refer to the
documentation and safety information of the probe manufacturer.
Explanation: According to section AA.2.4 of EN 61010-2-030, measuring circuits without any measurement category are intended for measurements on circuits which are
not directly connected to the mains.
2.2.1.2Other Connectors on the Front Panel
[Demo] (6)
The pins are intended for demonstration purposes.
26User Manual 1335.7898.02 ─ 05
R&S®RTA4000
Getting Started
Instrument Tour
[Gen]: Function Generator (7)
BNC output of the function generator (with option R&S RTA-B6).
[Pattern Generator] (8)
Connectors for the pattern generator P0, P1, P2, P3.
[Probe Comp.] (9)
Probe compensation terminal to support adjustment of passive probes to the oscilloscope channel.
Square wave signal for probe compensation.
Ground connector for probes.
[USB] type A (10)
USB 2.0 type A interface to connect a mouse or a keyboard, or a USB flash drive for
storing and reloading instrument settings and measurement data, and to update the
firmware.
2.2.2Side View
Figure 2-2: Side view of R&S RTA4000
1 = Connectors for logic probe (Mixed Signal Option R&S RTA-B1)
Logic probe
The connectors for logic channels can be used if the Mixed Signal Option R&S RTA-B1
is installed. The option provides connectors for two logical probes with 8 digital channels each (D0 to D7 and D8 to D15).
27User Manual 1335.7898.02 ─ 05
R&S®RTA4000
2.2.3Rear View
Getting Started
Instrument Tour
The maximum input voltage is 40 V (peak) at 100 kΩ input impedance. The maximum
input frequency for a signal with the minimum input voltage swing and medium hysteresis of 800 mV (Vpp) is 400 MHz.
Figure 2-3 shows the rear panel of the R&S RTA4000 with its connectors.
Figure 2-3: Rear panel view of R&S RTA4000
1 = Aux Out connector
2 = USB connector, type B
3 = LAN connector
4 = AC power supply connector and main power switch
5 = Kensington lock slot to secure the instrument against theft
6 = Loop for lock to secure the instrument against theft
7 = not used
[Aux Out] (1)
Multi-purpose BNC output that can function as pass/fail and trigger output, and output
of 10 MHz reference frequency.
[USB] type B (2)
USB 2.0 interface of type B (device USB) for remote control of the instrument.
28User Manual 1335.7898.02 ─ 05
R&S®RTA4000
Getting Started
Instrument Tour
Note: Electromagnetic interference (EMI) can affect the measurement results. To avoid
any impact, use only USB connecting cables with a maximum length of 1 m.
[LAN] (3)
8-pin connector RJ-45 used to connect the instrument to a Local Area Network (LAN).
It supports up to 1 Gbit/s.
AC supply: mains connector and main power switch (4)
The instrument supports a wide range power supply. It automatically adjusts to the correct range for the applied voltage. There is no line voltage selector.
The AC main power switch disconnects the instrument from the AC power line.
29User Manual 1335.7898.02 ─ 05
R&S®RTA4000
3Operating Basics
3.1Display Overview
Operating Basics
Display Overview
The touchscreen display of the instrument shows the waveforms and measurement
results, and also information and everything that you need to control the instrument.
Figure 3-1: Display of the R&S RTA4000 with 4 channels
1 = Toolbar
2 = Trigger source, main trigger parameter (here: slope for edge trigger), trigger level
3 = Trigger mode and sample rate
4 = Horizontal scale (time scale) and horizontal position
5 = Acquisition status and acquisition mode
6 = Date, time, education mode if active (here: off), LAN connection status (green = connected, grey = not
connected, yellow = connecting)
7 = Trigger level marker, has the color of the trigger source
8 = Trigger position marker, has the color of the trigger source
9 = Channel markers indicate the ground levels; channel C2S is selected, i.e. it has the focus
10 = Measurement results (here: automatic measurements on the left, cursor measurements on the right)
11 = Vertical settings of active analog channels: vertical scale, bandwidth limitation (no indicator = full band-