Rohde&Schwarz ZVA, ZVT Application Sheet

R&S®ZVA110 Basic Operation
Application Sheet
1ZKD-31, M. Jetter, 01/2011 ─ 01
Application Sheet
Test & Measurement
Muehldorfstr. 15, 81671 Munich, Germany
Phone: +49 89 41 29 - 0
Fax: +49 89 41 29 12 164
E-mail: info@rohde-schwarz.com
Internet: http://www.rohde-schwarz.com
Printed in Germany – Subject to change – Data without tolerance limits is not binding.
R&S® is a registered trademark of Rohde & Schwarz GmbH & Co. KG.
Trade names are trademarks of the owners.
The following abbreviations are used throughout this guide: R&S®ZVAxx as R&S ZVAxx, R&S®ZVTxx as R&S ZVTxx, R&S®ZVA-BU
as R&S ZVA-BU, R&S®ZVA-ZDxx as R&S ZVA-ZDxx, R&S®ZVA-Zxxx as R&S ZVA-Zxxx, R&S®ZV-Zxxx as R&S ZV-Zxxx, R&S®ZV-
WRxx as R&S ZV-WRxx.
R&S®ZVA110
Basic Operation
Required Equipment
1 Basic Operation
This application sheet describes the use of an R&S ZVA110 vector network analyzer with two external test sets for 2-port transmission measurements.
One-port reflection measurements can be performed in a similar way using a single external test set.

1.1 Required Equipment

The R&S ZVA110 system is delivered with all measurement equipment needed. An addi­tional 1 mm calibration kit is required for system error correction (calibration).

1.2 Measurement Principle

The principle of the measurement with external test sets is described at the beginning of the R&S ZVA110 Quick Start Guide, stock no 1314.4502.62.
The measurement involves the following steps:
1. Activation of the "ZVA110-BU" mode for measurements with external test set
2. Entry of power coefficients (when an external test set is used for the first time).
3. Connection of the external test sets
4. Power and frequency settings
5. Power calibration using an appropriate external power meter
6. System error correction (calibration) using a suitable calibration kit
7. Connection of the DUT and measurement

1.3 Activating the ZVA110-BU Measurement Mode

After a factory preset, the R&S ZVA67 is configured for measurements using the internal test sets. You have to activate the "ZVA110-BU" mode explicitly before you connect the external test sets.
To (re-)activate the "ZVA110-BU" mode,
1. Click "System > System Config ..." and open the "Frequency Converter" tab of the "System Configuration" dialog.
2. Select "Type : ZVA110-BU".
3. Click "Apply" to activate the mode.
3Application Sheet 1ZKD-31, M. Jetter, 01/2011 ─ 01
Loading...
+ 5 hidden pages