This manual describes the Inline Calibration System R&S®ZN-Z3x, comprising
the following components:
●
R&S®ZN-Z30 "Inline Calibration Controller", order no. 1328.7609.02
●
R&S®ZN-Z32 "Inline Calibration Unit 8.5 GHz", order no. 1328.7638.02
●
R&S®ZN-Z33 "Inline Calibration Unit 40 GHz", order no. 1328.7644.02
●
R&S®ZN-Z33 "Inline Calibration Unit 40 GHz, Var. 03 (TVAC)", order no.
1328.7644.03
●
R&S®ZN-CAN025 "CAN-bus control cable (baked out), 0.25 m", order no.
1339.3622.02
●
R&S®ZN-CAN2 "CAN-bus control cable (baked out), 2 m", order no.
1339.3639.02
●
R&S®ZN-CAN10 "CAN-bus control cable (baked out), 10 m", order no.
1339.3654.02
●
R&S®ZN-CANA "CAN-bus adapter cable (baked out), 0.25 m", order no.
1339.3651.02
The software contained in this product uses several valuable open source software packages. For information, see the "Open Source Acknowledgment" on the user documentation CD-ROM (included in delivery).
Rohde & Schwarz would like to thank the open source community for their valuable contribution to embedded computing.
The measurement accuracy of a Vector Network Analyzer (VNA) heavily depends
on a careful user calibration. Any deviation of the test setup after the calibration
process inevitably produces errors. In case of a massive system error correction
(e.g. if long RF cables with high damping or switch matrices are used), those
deviations can lead to strongly altered measurement results. Especially in remote
scenarios such as multiport measurements, a re-calibration using traditional calibration kits or calibration units is a costly and time consuming process.
The Rohde & Schwarz Inline Calibration System R&S ZN-Z3x allows re-calibrating a running measurement setup without mechanically changing the measurement setup. For each DUT port an Inline Calibration Unit (ICU) is inserted
between the corresponding VNA test port and the DUT. The ICU remains within
the signal path throughout the measurement (low-loss through path) and, on
request of the VNA, performs a full one-port calibration (OSM) in-situ.
In order to perform full multiport measurements an initial calibration of the system
has to be done.
Figure 1-1: Inline Calibration Unit
Control connections are established via a central Inline Calibration Controller
(ICC) with
●
2 CAN bus interfaces towards the ICUs
●
LAN and USB interface for remote control
5User Manual 1177.6381.02 ─ 03
R&S®ZN-Z3x
Introduction
Figure 1-2: Inline Calibration Controller
With respect to control, the ICC shall act as the master for the connected ICUs.
However, in order to avoid unnecessary polling of data that might not have
changed (e.g. the temperature of the ICUs) the ICUs themselves are able to send
data autonomously. The CAN protocol supports this kind of multi-master operation.
6User Manual 1177.6381.02 ─ 03
R&S®ZN-Z3x
Introduction
System Overview
1.1System Overview
An Inline Calibration System R&S ZN-Z3x consists of an Inline Calibration Controller R&S ZN-Z30 (order no. 1328.7609.02) and one or more Inline Calibration
Units R&S ZN-Z32/Z33, interconnected via CAN bus.
Table 1-1: Inline Calibration Units
Product NumberNameOrder No.
R&S®ZN-Z32
R&S®ZN-Z33
R&S®ZN-Z33 Var. 03 (TVAC)
RF Inline Calibration Unit,
10 MHz to 8.5 GHz
MW Inline Calibration Unit,
10 MHz to 40 GHz
MW Inline Calibration Unit,
10 MHz to 40 GHz
1328.7638.02
1328.7644.02
1328.7644.03
CAN bus systems are well established in automotive environments and
have proven a high degree of robustness.
The Inline Calibration Controller (ICC) offers two CAN bus interfaces and is able
to control and power supply up to 24 Inline Calibration Units per CAN bus. The
ICC itself can be controlled via LAN or USB.
For technical details see the R&S ZN-Z3x data sheet.
7User Manual 1177.6381.02 ─ 03
R&S®ZN-Z3x
Introduction
About this Manual
1.2About this Manual
The user manual contains the "getting started" chapters as well as descriptions of
all instrument functions and of the remote control of the instrument. In addition, it
contains notes on preventative maintenance for the R&S ZN-Z3x and on troubleshooting on the basis of the warnings and error messages that the instrument
emits.
The user manual is organized into the following chapters:
●
Putting the Instrument into Operation
●
Application
●
Operation (via Remote Control)
●
Maintenance
●
Appendix
The most recent version is always available on the Internet.
8User Manual 1177.6381.02 ─ 03
R&S®ZN-Z3x
Putting the System into Operation
Inline Calibration Controller Tour
2Putting the System into Operation
An Inline Calibration System R&S ZN-Z3x consists of a Inline Calibration Controller R&S ZN-Z30 and one or more Inline Calibration Units R&S ZN-Z32/Z33.
The sections below describe how to put the system into operation and how to
establish the required control connections between the system components. General notes are provided to ensure safety while operating the system.
2.1Inline Calibration Controller Tour
This chapter shows all control elements and connectors on the front and rear
panel of the Inline Calibration Controller R&S ZN-Z30.
2.1.1Front Panel
The front panel of the Inline Calibration Controller is shown on the photo below.
Brief explanations on the controls and connectors can be found on the following
pages.
9User Manual 1177.6381.02 ─ 03
R&S®ZN-Z3x
Mini display
The R&S ZN-Z30 offers an OLED status display which provides the following
information:
●
serial number
●
IP-address
●
CAN bus 1,2: number of connected Inline Calibration Units, bus status (see
ICC:CAN<bus>:GET:STATe? on page 36)
Note that in certain error states the ICC may not be able to detect the number
of connected ICUs.
●
Status of Inline Calibration Controller: "OK" or "Error 6" (configuration corrupted)
Putting the System into Operation
Inline Calibration Controller Tour
If "STAT | Error 6" persists the Inline Calibration Controller has to be
returned to Rohde & Schwarz service.
Display Reset
The Display Reset key deactivates the screensaver.
Status LEDs
●
Power: green = ready, red = standby; see Chapter 2.3.4, "Turning the ICC On
and Starting", on page 17
●
LAN: green = OK, red = not OK or booting
●
USB: green = connected, red = not connected
Standby key
The key serves two purposes:
●
Toggle between standby and ready state; see Chapter 2.3.4, "Turning the ICC
On and Starting", on page 17.
When the instrument is powered up it comes up in ready state.
10User Manual 1177.6381.02 ─ 03
R&S®ZN-Z3x
Putting the System into Operation
Inline Calibration Controller Tour
●
Shut down the instrument; see Chapter 2.3.5, "Switching the ICC Off",
on page 18.
CAN 1,2
Two independent CAN bus interfaces for control connection and
power supply of Inline Calibration Units, see Chapter 2.4, "Connect-
ing ICUs", on page 18.
The CAN bus voltage supplied by the R&S ZN-Z30 is +24 V with a
maximum current of 1.8 A. The Inline Calibration Units will report an error if the
supply voltage is too low (e.g. due to excessive bus length).
Table 2-1: CAN 1,2 pin assignment
ContactDescriptionValue Range
1 (northwest)CAN_HIGH0 ... +5.5 V
2 (southwest)CAN_LOW0 ... +5.5 V
3 (southeast)+24 V supply voltage0 ... +24.5 V
4 (northeast)Ground0 V
The color of the LED above the connectors indicates the state of the respective
CAN bus:
●
red: bus not working (e.g. no Inline Calibration Unit connected) or in error
state
●
blue: bootloader mode (temporary state)
●
pink: calibration data alignment between Inline Calibration Units and Inline
Calibration Controller (temporary state)
●
green: error-free operation
2.1.2Back Panel
The back panel of the Inline Calibration Controller is shown on the photo below.
Brief explanations on the controls and connectors can be found in Table 2-2.
11User Manual 1177.6381.02 ─ 03
R&S®ZN-Z3x
Putting the System into Operation
Inline Calibration Controller Tour
1
2
3
5
6
Table 2-2: Back panel elements
IndexLabelDescription
1O IPower switch
2
3
4LANRJ-45 connector to integrate the instrument to a LAN, primarily for
5USBType B (slave) high speed USB connector for remote control of the
6SD-CARDMicro SD card slot.
Fuse drawer (see Chapter 5.2, "Replacing Fuses", on page 47)
AC power connector
remote control purposes.
R&S ZN-Z30.
The R&S ZN-Z30 offers a Micro SD-Card slot to provide a mass storage device for characterization data gathered from the connected
ICUs. This way the slow process of copying characterization data from
a ICU to the ICC via CAN-bus (data rate 250 kBit/s) is only needed
with the first usage of a new ICU with a dedicated ICC.
The characterization data on the SD-Card can be accessed either via
LAN (using the Inline Calibration Controller's FTP-server) or by utilizing
the USB interface via the ZN-Z3x User tool (ZN-Z3x User Tool).
Note: The SD card has to be formatted to FAT32.
4
See the R&S ZN-Z3x Data Sheet for interface specification details.
12User Manual 1177.6381.02 ─ 03
R&S®ZN-Z3x
Putting the System into Operation
Inline Calibration Unit Tour
2.2Inline Calibration Unit Tour
VNA/DUT
RF connectors, 50 Ω, male/female
●
SMA for R&SZN-Z32
●
2.92 mm (K) for R&S ZN-Z33
It is recommended to use a suitable torque wrench when fastening the RF
cables on the test port connectors.
Maximum input levels
The maximum input levels at all test ports according to the data sheet must
not be exceeded.
In addition, the maximum input voltages of the other input connectors at the
rear panel must not be exceeded.
CAN In/Out
"CAN in" and "CAN out" connect the Inline Calibration Unit to a CAN bus (see
Chapter 2.4, "Connecting ICUs", on page 18). The first is for the direction
towards the Inline Calibration Controller, the latter can be used to connect other
Inline Calibration Units; see Chapter 2.4, "Connecting ICUs", on page 18.
Port Display
In error-free operation this displays the test port number to which the Inline Calibration Unit is connected and which is assigned using ICU:SET:PORT<no>. If an
error occurs, E<error code> will be displayed alternating with the test port number.
13User Manual 1177.6381.02 ─ 03
R&S®ZN-Z3x
Table 2-3: Inline Calibration Unit Errors
Error codeError description
001CAN bus error
002Memory read/write error
003Temperature error
00424 V error
Do not connect any further Inline Calibration Unit R&S ZN-Z32/Z33,
otherwise the calibration precision (effective system data) can no longer be guaranteed
0058 V error
Calibration functionality degraded
0065 V error
CAN bus functionality degraded, risk of system error
Putting the System into Operation
Preparing the System
Start Cal
Button with integrated LED, allowing to start a calibration sweep.
The button is armed using the SCPI command ICU:CAL:START. Its current state
can be read using ICU:SHOW:BUTton:STATE?.
2.3Preparing the System
This section covers the basic steps to set up the Inline Calibration System
R&S ZN-Z3x and to put it into operation.
14User Manual 1177.6381.02 ─ 03
R&S®ZN-Z3x
Risk of injury and instrument damage
The instrument must be used in an appropriate manner to prevent electric
shock, fire, personal injury, or damage.
●
Do not open the instrument casing.
●
Read and observe the "Basic Safety Instructions" delivered as a printed
brochure with the instrument.
In addition, read and observe the safety instructions in the following sections. Notice that the data sheet may specify additional operating conditions.
Putting the System into Operation
Preparing the System
Risk of instrument damage due to inappropriate operating conditions
Specific operating conditions are required to ensure accurate measurements and to avoid damage to the instrument. Observe the information on
appropriate operating conditions provided in the basic safety instructions
and the instrument's data sheet.
Instrument damage caused by electrostatic discharge
Electrostatic discharge (ESD) can damage the electronic components of the
instrument and the device under test (DUT). Electrostatic discharge is most
likely to occur when you connect or disconnect a DUT or test fixture to the
instrument's test ports. To prevent electrostatic discharge, use a wrist strap
and cord and connect yourself to the ground, or use a conductive floor mat
and heel strap combination.
15User Manual 1177.6381.02 ─ 03
R&S®ZN-Z3x
Putting the System into Operation
Preparing the System
Safety precautions
Be absolutely sure to follow the instructions in the sections below to prevent
injury to people or damage to the system components. This is particularly
important the first time that you use the instrument. In addition, be sure to
observe the general safety notes at the beginning of this manual.
2.3.1Unpacking the ICC
The Inline Calibration Controller is delivered in a cardboard box. Proceed as follows to unpack its contents:
1. Remove the instrument from its packaging and check the shipment for completeness by comparing it with the shipping document and the accessory lists
for the various items.
2. Remove the protective covers made of corrugated cardboard from the front
and rear of the instrument.
3. Check the instrument for any damage. If there is any damage, immediately
contact the shipping company that delivered the instrument. If this case, be
absolutely certain that you keep the cardboard box and packing materials.
The delivery comprises two CAN bus control cables R&S ZN-CAN2 (2 m length),
a power cable, a USB cable and the Software and User Documentation CD-ROM.
Packing materials
We recommend that you retain the packaging. It is advisable to keep the
original packing material in order to prevent control elements and connectors from being damaged if the instrument has to be transported or shipped
at a later date.
2.3.2Setting up the System
The Inline Calibration Controller is designed for indoor use and can be set up on
a level surface in a horizontal position.
16User Manual 1177.6381.02 ─ 03
R&S®ZN-Z3x
Putting the System into Operation
Preparing the System
2.3.3Connecting the ICC to the AC Power Supply
Use the supplied power cable to connect the Inline Calibration Controller to the
AC power supply.
Danger of electric shock
The Inline Calibration Controller complies with the specifications for the
EN61010-1 protective class, which means that it can only be connected to a
power outlet that has a protective contact.
2.3.4Turning the ICC On and Starting
Switch the power switch on the back panel to I.
The power switch has two positions:
●
0: The Inline Calibration Controller is disconnected from the AC supply.
●
I: The Inline Calibration Controller is receiving power from the AC supply.
When the instrument receives power from the AC supply it immediately starts
loading the firmware (mini display: "ZN-Z30 <firmware version>") and performs a
self-test.
In standby state press the Standby key to start the instrument.
As soon as the firmware is successfully loaded, the instrument is ready for operation and can be remotely controlled via USB and LAN.
17User Manual 1177.6381.02 ─ 03
R&S®ZN-Z3x
Putting the System into Operation
Connecting ICUs
2.3.5Switching the ICC Off
To switch the R&S ZN-Z30 off, proceed as follows:
1. Press the Standby key on the front of the instrument.
The instrument switches to standby mode; only the POWER (standby) button
remains active.
2. To disconnect the instrument completely from the AC supply, switch the
power switch on the rear of the instrument to 0.
3. Disconnect the power cable at the AC power supply.
2.3.6Function Check
The R&S ZN-Z30 automatically monitors the most important instrument functions
when it is turned on.
2.4Connecting ICUs
Per CAN bus interface up to 24 Inline Calibration Units can be connected in line,
with a maximum total cable length of 40 m. CAN bus termination is provided by
the last Inline Calibration Unit in the respective line.
The connectors are mechanically coded to prevent false connections (e.g. to connect CAN1 with CAN2):
30°0°
CAN 1 and CAN 2 at the Inline Calibration Controller and CAN Out at the Inline
Calibration Unit are coded with 30°, CAN In at the Inline Calibration Unit is coded
with 0°. Each Inline Calibration Unit is shipped with a CAN bus control cable
18User Manual 1177.6381.02 ─ 03
R&S®ZN-Z3x
Putting the System into Operation
Connecting External Devices
R&S ZN-CAN025 (0.25 m length, one end coded 30°, the other coded 0°); other
cable lengths and couplers can be ordered separately.
Order no.Cable lengthConnector types Connector cod-
ing
The R&S ZN-Z30 is equipped with different interfaces for establishing a remote
control connection.
EMI Suppression
To suppress electromagnetic interference (EMI), the instrument may only
be operated when it is closed and all covers are in place. The instrument's
EMC rating is specified on the data sheet.
To prevent influences that cause disturbances, the following conditions
must be adhered to:
●
Use suitable double-shielded cables.
●
Don't use any USB cables that are longer than 1 m.
●
Only use USB devices that maintain the permitted EMC values.
●
For the LAN connection, use CAT6 or CAT7 RJ-45 cables (LAN, Ethernet)
19User Manual 1177.6381.02 ─ 03
Loading...
+ 43 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.