Rohde&Schwarz TS-PSAM User Manual

R&S®TS-PSAM Analog Stimulus Measurement Module User Manual
(;ZðÜ<)
1142987812 Version 12
This manual describes the following R&S®TSVP module:
R&S®TS-PSAM
© 2021 Rohde & Schwarz GmbH & Co. KG Mühldorfstr. 15, 81671 München, Germany Phone: +49 89 41 29 - 0 Email: info@rohde-schwarz.com Internet: www.rohde-schwarz.com Subject to change – data without tolerance limits is not binding. R&S® is a registered trademark of Rohde & Schwarz GmbH & Co. KG. Trade names are trademarks of the owners.
1142.9878.12 | Version 12 | R&S®TS-PSAM
The following abbreviations are used throughout this manual: R&S®PSAM is abbreviated as R&S PSAM.
1
Risk of injury and instrument damage
The instrument must be used in an appropriate manner to prevent personal injury or instrument damage.
Do not open the instrument casing.
Read and observe the "Basic Safety Instructions" delivered as
printed brochure with the instrument.
Read and observe the safety instructions in the following sections.
Note that the data sheet may specify additional operating conditions.
Keep the "Basic Safety Instructions" and the product documentation
in a safe place and pass them on to the subsequent users.
Riesgo de lesiones y daños en el instrumento
El instrumento se debe usar de manera adecuada para prevenir descargas eléctricas, incendios, lesiones o daños materiales.
No abrir la carcasa del instrumento.
Lea y cumpla las "Instrucciones de seguridad elementales"
suministradas con el instrumento como folleto impreso.
Lea y cumpla las instrucciones de seguridad incluidas en las
siguientes secciones. Se debe tener en cuenta que las especificaciones técnicas pueden contener condiciones adicionales para su uso.
Guarde bien las instrucciones de seguridad elementales, así como
la documentación del producto, y entréguelas a usuarios posteriores.
Safety Instructions Instrucciones de seguridad Sicherheitshinweise Consignes de sécurité
1171.1307.42 - 05
2
Gefahr von Verletzungen und Schäden am Gerät
Betreiben Sie das Gerät immer ordnungsgemäß, um elektrischen Schlag, Brand, Verletzungen von Personen oder Geräteschäden zu verhindern.
Öffnen Sie das Gerätegehäuse nicht.
Lesen und beachten Sie die "Grundlegenden Sicherheitshinweise",
die als gedruckte Broschüre dem Gerät beiliegen.
Lesen und beachten Sie die Sicherheitshinweise in den folgenden
Abschnitten; möglicherweise enthält das Datenblatt weitere Hinweise zu speziellen Betriebsbedingungen.
Bewahren Sie die "Grundlegenden Sicherheitshinweise" und die
Produktdokumentation gut auf und geben Sie diese an weitere Benutzer des Produkts weiter.
Risque de blessures et d'endommagement de l'appareil
L'appareil doit être utilisé conformément aux prescriptions afin d'éviter les électrocutions, incendies, dommages corporels et matériels.
N'ouvrez pas le boîtier de l'appareil.
Lisez et respectez les "consignes de sécurité fondamentales"
fournies avec l’appareil sous forme de brochure imprimée.
Lisez et respectez les instructions de sécurité dans les sections
suivantes. Il ne faut pas oublier que la fiche technique peut indiquer
des conditions d’exploitation supplémentaires.
Gardez les consignes de sécurité fondamentales et la
documentation produit dans un lieu sûr et transmettez ces documents aux autres utilisateurs.
1171.1307.42 - 05
R&S®TS-PSAM

1 Usage

1.1 General

Usage
Features of the R&S TS-PSAM
The R&S TS-PSAM Analog Stimulus and Measurement Module is suitable for the R&S CompactTSVP. The module is used to perform analog measuring tasks, in-circuit mea­surements, and the R&S CompactTSVP self-test. The module's floating DC supply voltage is provided via the associated R&S TS-PDC rear I/O module.
The R&S TS-PSAM module is inserted in the front of the R&S CompactTSVP chassis. It is based on the cPCI/PXI standard.
The front connector ends flush with the front panel of the R&S CompactTSVP chassis and is used for contacting the test products or measurement sensors. At the back, the R&S TS-PSAM module is connected to the cPCI control bus and the PXI trigger bus. Instead of using the front connector, measurement signals can be routed via the R&S CompactTSVP's analog measuring bus.
A LabWindows IVI DMM driver is provided for the DMM functions on the card. All other functions are controlled using specific extensions of the driver. As is typical for a Lab­Windows CVI driver, Function Panels and Online Help are available.

1.2 Features of the R&S TS-PSAM

1.2.1 Components

The R&S TS-PSAM modules include a ground-connected discharge circuit, a floating programmable voltage source, and a floating measurement unit. These components can be switched to the R&S CompactTSVP's analog bus via a relay matrix. Measuring tasks can be synchronized using the trigger lines across the PXI bus and the trigger inputs on the front connector. Two relay multiplexers with four channels each are also provided. The major features of these three function blocks is explained in detail in the following section:
floating DC voltage source (DCS)
adjustable voltage and current limiting – ±5 V, 100 mA max.
fast settling time
four-quadrant operation
sense wires
3User Manual 1142.9878.12 ─ 12
R&S®TS-PSAM
Features of the R&S TS-PSAM
floating measurement unit (MU)
Measurement range DC – 10 mV - 125 V – 1 µA - 1 A
Measurement range AC rms
rms
20 mV - 125 V – 100 µA - 1 A
triggered measurements across the PXI Triggerbus
2 triggers derived from the measurement signal with programmable threshold
4 filters
16-bit converter – Sampling rate 200 kHz (max. ) – Single or „multipoint“ measurement with storage depth up to 8 k samples
Discharge circuit (DCH)
Usage
Discharge current 400 mA (max. )
Discharge voltage 125V (max. )

1.2.2 Measuring Functions

The module is used to perform analog measuring tasks, in-circuit measurements, and the R&S CompactTSVP self-test. In these tests, the module functions as a measuring device with adjustable sampling rate for voltage, current and resistance measure­ments. The measurement unit and the adjustable DC voltage source are cross-connec­ted with each other in a suitable manner for measuring resistance. If necessary, the source can also be connected with GND. The measurement unit and the DC voltage source can also be operated independently of one another. In the in-circuit test (ICT), the R&S TS-PSAM module performs the following measuring tasks:
Discharging capacitors
2- and 4-wire resistance measurements (Figure 1-1 to Figure 1-4)
Contact test
Short circuit test
Connection test
1.2.2.1 Resistance Measurements
On this topic, see also Figure 1-1 to Figure 1-4.
Resistance measurements are taken with the aid of the DC voltage source and the measurement unit. 2- and 4-wire measurements are possible. Two different procedures may be followed depending on the resistance value to be measured. If necessary, the source can be connected to GND.
Mode C for small resistances
4User Manual 1142.9878.12 ─ 12
R&S®TS-PSAM
V
LO
R
x
HI
=
MUDCS
V
LO
R
x
HI
SHI
SLO
R
S
R
M
=
MUDCS
Usage
Features of the R&S TS-PSAM
In this method, a constant current is applied and the voltage is measured. (see Fig-
ure 1-1 to Figure 1-2)
Figure 1-1: Cross-connection for 2-wire resistance measuring in Mode C
Figure 1-2: Cross-connection for 4-wire resistance measuring in Mode C
Mode V for large resistances In this method, the voltage is applied and the current is measured. (see Figure 1-3 to Figure 1-4)
5User Manual 1142.9878.12 ─ 12
R&S®TS-PSAM
A
LO
R
x
HI
=
MU
DCS
A
LO
R
x
HI
SHI
SLO
R
M
R
S
=
MU
DCS
Usage
Features of the R&S TS-PSAM
Figure 1-3: Cross-connection for 2-wire resistance measuring in Mode V
Figure 1-4: Cross-connection for 4-wire resistance measuring in Mode V
6User Manual 1142.9878.12 ─ 12
R&S®TS-PSAM
A
B2
R
x
A1
=
R1
R2
A2
C
D
B1
CMU
DCS
1.2.2.2 ICT Measurements with R&S TS-PICT
Usage
Features of the R&S TS-PSAM
On this topic, see also Figure 1-5 and Figure 1-6.
Further in-circuit measurements can be made in conjunction with the R&S TS-PICT module (ICT expansion module). These are:
Diode and transistor test
Guarded resistance measurements (3, 4 and 6 wires)
Impedance measurements (3, 4 and 6 wires)
For this purpose, the R&S TS-PICT module provides a special AC voltage source (AOS) and current measurement unit (CMU).
The UUT is connected for the in-circuit test (ICT) via the R&S TS-PMB module (matrix module B).
Figure 1-5: Cross-connection in a guarded resistance measurement (6-wire)
7User Manual 1142.9878.12 ─ 12
R&S®TS-PSAM
Usage
R&S TS-PDC Features
Figure 1-6: Cross-connection in a guarded impedance measurement (6-wire)

1.3 R&S TS-PDC Features

The R&S TS-PDC module is used as a floating DC voltage source for the R&S TS­PSAM module. It is configured with two identical DC/DC converters. The following floating direct voltages are obtained from an input voltage of 5 VDC:
+15 VDC ±5%, 0,5A (2x)
-15 VDC ±5%, 0,5A (2x)
+5 VDC ±5%, 0,5A (2x)
+3,3 VDC ±5%, 0,25A (2x)
8User Manual 1142.9878.12 ─ 12
R&S®TS-PSAM

2 View

View
Figure 2-1shows the R&S TS-PSAM module without the associated TS-PDC rear I/O
module. The R&S TS-PDC rear I/O module is shown in Figure 2-2.
Figure 2-1: View of the R&S TS-PSAM module
9User Manual 1142.9878.12 ─ 12
R&S®TS-PSAM
View
Figure 2-2: View of the R&S TS-PDC rear I/O module.
The Module R&S TS-PDC exists in 3 different models:
Grouted in a black housing - version up to 1.8 (1157.9804.02 obsolete)
Encapsulated in metal housing with cooling fins - version 1.9 (1157.9804.02 obso­lete)
Without case - version from 2.0 (1157.9804.12 actual version)
10User Manual 1142.9878.12 ─ 12
R&S®TS-PSAM
MU
DCH
DCS_HI
DCS_SLO
DCS_SHI
DCH_HI
a1 c2b1 b2a2 c1 d2d1
Analog Bus (AB)
GND
DCS_LO
Local Analog Bus (LAB)
a1 c2b1 b2a2 c1 d2d1
DCS
PXI_TRIG0.. PXI_TRIG7
XTA1, XTA2
XTI1, XTI2
XTO1, XTO2
Ge ogr aph ic Ad dre ss ing
GA0..GA4
MU_HI
MU_LO
DCS_SLO
DCS_SHI
MU_SHI
MU_SLO
MU
SENSE
SP I Int erfa ce
Analog Bus Connector X 30CPCI Bus Connector X1 Extension Connector X 20
Front Connector X10
CP CI Int e rfa ce
GND
CHA-GND
Isolation
AUX1, AUX2
Trigger
RACH1
RACH2
RACH3
RACH4
RACOM
RBCH1
RBCH2
RBCH3
RBCH4
RBCOM
=
~
=
~

3 Block diagrams

Block diagrams
Figure 3-1shows the block diagram of the R&S TS-PSAM module and Figure 3-2
shows the block diagram of the R&S TS-PDC module. Figure 3-3 is a simplified func­tional block diagram of both modules in the R&S CompactTSVP.
Figure 3-1: Block Diagram of R&S TS-PSAM
11User Manual 1142.9878.12 ─ 12
R&S®TS-PSAM
GND
-15 V
INHIBIT
X 20
COM
-15 V
COM
ON/OFF
Regulator
1
primary
DC-Transducer
Regulator
2
Block diagrams
Figure 3-2: Block Diagram of R&S TS-PDC
Figure 3-3: Functional block diagram of R&S TS-PSAM with R&S TS-PDC in the R&S CompactTSVP
12User Manual 1142.9878.12 ─ 12
R&S®TS-PSAM

4 Layout

4.1 Mechanical Construction of the R&S TS-PSAM

Layout
Mechanical Construction of the R&S TS-PSAM
On this topic, see also Figure 4-1.
The R&S TS-PSAM module is designed as a long cPCI plug-in module for mounting in the front of the R&S CompactTSVP. The board height of the module is 3 HU (134 mm). To ensure that it is inserted correctly into the Compact TSVP, the front panel is fur­nished with a locating pin. The module is secured in place with the two retaining screws on the front panel. Front connector X10 is used for connecting the UUTs. Con­nector X30 connects the R&S TS-PSAM module to the analog bus backplane in the R&S CompactTSVP. Connectors X20/X1 connect the R&S TS-PSAM module to the cPCI backplane/PXI control backplane.
Figure 4-1: Arrangement of connectors and LEDs on the R&S TS-PSAM module
Table 4-1: Connectors on the R&S TS-PSAM module
Symbol Use
X1 cPCI Bus
X10 UUT
X20 Extension (PXI), Rear I/O
X30 Analog Bus
13User Manual 1142.9878.12 ─ 12