1 Overview

2 Instrument Models Covered

Overview
It is often imperative that R&S SMBV100B vector signal generators are used in a secured environment. Generally these highly secured environments do not allow any test equipment to leave the area unless it can be proven that no user information leaves with the test equipment. Security concerns can arise when devices need to leave a secured area, e.g. to be calibrated or serviced.
This document describes the types of memory and their usage in the R&S SMBV100B. It provides a statement regarding the volatility of all memory types and specifies the steps required to declassify an instrument through memory clearing or sanitization pro­cedures. These sanitization procedures are designed for customers who need to meet the requirements specified by the US Defense Security Service (DSS).
Table 2-1: vector signal generator models
Product name Order number
R&S SMBV100B 1423.1003K02

3 Security Terms and Definitions

Clearing
The term "clearing" is defined in Section 8-301a of DoD 5220.22-M, "National Industrial Security Program Operating Manual (NISPOM)". Clearing is the process of eradicating the data on media so that the data can no longer be retrieved using the standard inter­faces on the instrument. Therefore, clearing is typically used when the instrument is to remain in an environment with an acceptable level of protection.
Sanitization
The term "sanitization" is defined in Section 8-301b of DoD 5220.22-M, "National Industrial Security Program Operating Manual (NISPOM)". Sanitization is the process of removing or eradicating stored data so that the data cannot be recovered using any known technology. Instrument sanitization is typically required when an instrument is moved from a secure to a non-secure environment, such as when it is returned for ser­vice of calibration.
The memory sanitization procedures described in this document are designed for cus­tomers who need to meet the requirements specified by the US Defense Security Ser­vice (DSS). These requirements are specified in the "Clearing and Sanitization Matrix" in Section 14.1.16 of the ISFO "Manual for the Certification and Accreditation of Classi­fied Systems under the NISPOM".
2Instrument Security Procedures 1178.9032.02 ─ 02
R&S®SMBV100B
4 Types of Memory and Information Storage

Types of Memory and Information Storage in the R&S SMBV100B

Instrument declassification
The term "instrument declassification" refers to procedures that must be undertaken before an instrument can be removed from a secure environment, for example when the instrument is returned for calibration. Declassification procedures include memory sanitization or memory removal, or both. The declassification procedures described in this document are designed to meet the requirements specified in DoD 5220.22-M, "National Industrial Security Program Operating Manual (NISPOM)", Chapter 8.
in the R&S SMBV100B
The R&S SMBV100B contains various memory components.
The following table provides an overview of the memory components that are part of your instrument. For a detailed description regarding type, size, usage and location, refer to the subsequent sections.
Standard configuration (no option R&S SMBVB-B80 installed)
Table 4-1: Types of memory and information storage
Memory type Size Content / Function Volatility User
SDRAM (CPU board) 4 GByte Temporary information storage for
operating system and instrument firmware, or RAM disk for user data
SDRAM (Baseband gen­erator board)
EEPROM/Flash
(one per module)
Flash (CPU board) 8 MByte BIOS Non-volatile No None required (no user
microSD memory card (CPU board)
8 GByte Temporary information storage for
arbitrary waveforms
256 Byte up to 4 MByte
1 GByte
Module-specific data:
Board identification data
Board internal correction data
Operating system
Instrument firmware
Instrument internal correction data
Instrument states, e.g. USB
port, LAN services and set-
ups
Volatile Yes Turn off instrument power.
Volatile Yes Turn off instrument power.
Non-volatile No None required (no user
Non-volatile No None required (no user
Sanitization procedure
data
data)
data)
data)
mSata memory module (IFB.BV interface board)
62 GByte
User data
Instrument setups
Non-volatile Yes Sanitize internal memory
procedure
3Instrument Security Procedures 1178.9032.02 ─ 02