The product documentation helps you use the R&S RTH safely and efficiently.
Follow the instructions provided here and in the printed "Safety Instructions".
Keep the product documentation nearby and offer it to other users.
Where do I find safety information?
Safety information is part of the product documentation. It warns you about the
potential dangers and gives instructions how to prevent personal injuries or damage caused by dangerous situations. Safety information is provided as follows:
●
The printed "Safety Instructions" provide safety information in many languages and are delivered with the R&S RTH.
●
Throughout the documentation, safety instructions are provided when you
need to take care during setup or operation.
1.1For Your Safety
To prevent electric shock, personal injury, or fire caused by high voltages, carefully read and observe the following safety information:
●
This chapter, the first part of the safety booklet, provides warnings and safety
information for usage of oscilloscopes, probes, and other measurement
accessories.
●
The second part of the safety booklet provides basic safety instructions
●
Product manuals provide specific safety instructions for procedures and measurement accessories.
Electrical safety
●
Voltages higher than 30 V RMS, or 42 V peak, or 60 V DC are regarded as
hazardous contact voltages. When working with hazardous contact voltages,
use protective measures to preclude direct contact with the measurement
setup:
–Do not touch exposed connections and components when power is
applied.
–Use only insulated voltage probes, test leads and adapters.
5Getting Started 1326.1561.02 ─ 09
R&S®Scope Rider RTH
●
Make sure that only personnel familiar with the potential risks of measuring
electrical quantities controls the instrument. Observe applicable local or
national safety regulations and rules for the prevention of accidents.
●
Use only specified probes and accessories that comply with the measurement
category (CAT) of your measurement task. If you use other than
Rohde & Schwarz accessories, make sure that they are suitable for the instrument and the measurement task.
●
Observe all voltage and current ratings of the instrument, the probes, and the
accessories. The lowest rated component defines the rating of the complete
measurement setup. Limits and ratings are marked on the products and listed
in the data sheets.
For probes, consider that the rated voltage depends on the frequency. The
voltage limitation curves are provided in the data sheet. Do not exceed these
two ratings:
–Maximum measurement voltage from the probe tip to the probe reference
lead.
Safety and Regulatory Information
For Your Safety
–Maximum floating voltage from the probe reference lead to earth ground.
●
Set the correct attenuation factor on the instrument according to the probe
being used. Otherwise, the measurement results do not reflect the actual voltage level, and you might misjudge the actual risk.
●
Set up all connections to the instrument before applying power.
●
Do not open the instrument casing.
●
Do not use the instrument if the instrument casing, the display or any probe or
accessory are damaged. If you detect or suspect any damage, have the
instrument or accessory inspected by qualified service personnel.
●
Do not operate the instrument in wet, damp or explosive atmospheres. Make
sure that all connectors are completely dry before connecting the inputs.
●
Observe the operating conditions specified in the data sheet and in the basic
safety instructions.
Battery replacement
●
Disconnect power supply, probes, test leads and all other cables before opening the battery cover.
●
Use only the specified Li-ion battery pack, which is delivered with the instrument. You can order additional battery packs at Rohde & Schwarz, see data
sheet for order number.
●
Do not operate the instrument with the battery cover open.
6Getting Started 1326.1561.02 ─ 09
R&S®Scope Rider RTH
●
Use only the specified power adapter, which is delivered with the instrument.
Cleaning
Remove all probes, leads, USB and LAN cables and power supply before cleaning the instrument. Use only cleaning materials that are specified in the manual.
Measurement categories
IEC 61010-2-030 defines measurement categories that rate instruments on their
ability to resist short transient overvoltages that occur in addition to the working
voltage. Use the measurement instrument and accessories only in electrical environments for which they are rated.
●
O - Instruments without rated measurement category
For measurements performed on circuits not directly connected to mains, for
example, electronics, circuits powered by batteries, and specially protected
secondary circuits. This measurement category is also known as CAT I.
Safety and Regulatory Information
For Your Safety
●
CAT II:
For measurements performed on circuits directly connected to the low-voltage
installation by a standard socket outlet, for example, household appliances
and portable tools.
●
CAT III:
For measurements performed in the building installation, such as junction
boxes, circuit breakers, distribution boards, and equipment with permanent
connection to the fixed installation.
●
CAT IV:
For measurements performed at the source of the low-voltage installation,
such as electricity meters and primary overcurrent protection devices.
7Getting Started 1326.1561.02 ─ 09
R&S®Scope Rider RTH
Safety and Regulatory Information
Regulatory Information
1.2Regulatory Information
Part 15 of the FCC and RSS-210 of IC Rules
This device complies with Part 15 of the FCC Rules and with Industry Canada
licence-exempt RSS standard(s). Operation is subject to the following two conditions:
●
this device may not cause harmful interference, and
●
this device must accept any interference received, including interference that
may cause undesired operation.
Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux
appareils radio exempts de licence. L'exploitation est autorisée aux deux conditions suivantes:
●
l'appareil ne doit pas produire de brouillage, et
●
l'utilisateur de l'appareil doit accepter tout brouillage radioélectrique subi,
même si le brouillage est susceptible d'en compromettre le fonctionnement.
Changes or modifications made to this equipment not expressly approved by
Rohde & Schwarz may void the FCC authorization to operate this equipment.
This equipment has been tested and found to comply with the limits for a Class A
digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to
provide reasonable protection against harmful interference when the equipment is
operated in a commercial environment. This equipment generates, uses, and can
radiate radio frequency energy and, if not installed and used in accordance with
the instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own
expense.
Administrative Regulations on Low Power Radio Waves Radiated Devices
warning (Republic of China/Taiwan)
Article 12
●
Without permission granted by the NCC, any company, enterprise, or user is
not allowed to change frequency, enhance transmitting power or alter original
characteristic as well as performance to a approved low power radio-frequency devices.
8Getting Started 1326.1561.02 ─ 09
R&S®Scope Rider RTH
Article 14
●
The low power radio-frequency devices shall not influence aircraft security
and interfere legal communications; If found, the user shall cease operating
immediately until no interference is achieved.
●
The said legal communications means radio communications is operated in
compliance with the Telecommunications Act.
●
The low power radio-frequency devices must be susceptible with the interference from legal communications or ISM radio wave radiated devices.
The R&S RTH is the perfect multi-purpose tool for the lab and in the field. Outstanding key features are:
●
Full isolation of all channels and interfaces
●
CAT IV 600 V / CAT III 1000 V safety rating
●
Bandwidth 60 MHz to 500 MHz with 5 GS/s sampling rate
●
Acquisition speed up to 50.000 waveforms per second
●
2 mV/div sensitivity
●
Up to 200 V offset range
●
33 automatic measurement functions
●
Full operation using touch or keypad
●
Wireless LAN and Ethernet for web based remote control and quick data
access (optional)
The R&S RTH combines:
●
Lab performance oscilloscope
●
Logic analyzer with 8 digital inputs (optional)
●
Protocol analyzer with trigger and decode (optional)
●
Data logger
●
Digital multimeter (R&SRTH1002)
2.2Main Specifications
This chapter lists the main characteristics of the R&S RTH. The complete specifications are listed in the Data Sheet, see Chapter 2.4, "Documentation Overview",
on page 15.
10Getting Started 1326.1561.02 ─ 09
R&S®Scope Rider RTH
Main Specifications
Input channels
●
R&SRTH1002: 2 oscilloscope channels, 1 multimeter
●
R&SRTH1004: 4 oscilloscope channels
Maximum input voltage
●
At BNC inputs: CATIV300V
●
With probe R&S RT-ZI10 or R&S RT-ZI11: CATIV600V, CATIII1000V
●
Meter input: CAT IV 600 V; CAT III 1000 V
Voltage ratings: V RMS (50 to 60 Hz) for AC sine wave and V DC for DC applications.
Vertical system
●
Analog bandwidth (–3 dB) of oscilloscope channels:
Preface
–R&S RTH1002 and R&S RTH1004: ≥ 60 MHz
–R&S RTH1002 with -B221 option and R&S RTH1004 with -B241 option:
≥ 100 MHz
–R&S RTH1002 with -B222 option and R&S RTH1004 with -B242 option:
≥ 200 MHz
–R&S RTH1002 with -B223 option and R&S RTH1004 with -B243 option:
≥ 350 MHz
–R&S RTH1002 with -B224 option and R&S RTH1004 with -B244 option:
≥ 500 MHz
●
Rise time (calculated):
R&S RTH1002 and R&S RTH1004 with bandwidth of 60 MHz, 100 MHz,
200 MHz, 350 MHz, 500 MHz: < 5.8 ns, < 3.5 ns, < 1.75 ns, < 1 ns, < 700 ps
respectively
●
DC gain accuracy:
Offset and position set to zero after self-alignment
Lithium-ion rechargeable smart battery
voltage 11.25 V; capacity 72.0 Wh
Environmental conditions
●
Operating temperature
–Battery only: 0 °C to +50 °C
–Power adapter: 0 °C to +40 °C
●
Storage temperature: –20 °C to +50 °C
●
Climatic loading: +25° C/+55 °C at 95 % rel. humidity cyclic, in line with IEC
60068-2-30
●
Altitude
–Operating:
CAT IV 600 V, CAT III 1000 V: up to 2000 m above sea level
CAT III 600 V, CAT II 1000 V: up to 3000 m above sea level
–Nonoperating: up to 4600 m above sea level
●
Pollution degree 2
13Getting Started 1326.1561.02 ─ 09
R&S®Scope Rider RTH
IP rating
●
IP51 in line with IEC 60529
Safety compliance
●
IEC / EN / DIN EN 61010-1
●
IEC / EN / DIN EN 61010-2-030
●
IEC / EN / DIN EN 61010-2-033 (R&SRTH1002)
●
UL / CSA 61010-1
●
UL / CSA 61010-2-030
●
UL / CSA 61010-2-033 (R&SRTH1002)
EMC compliance
●
RF emission
Preface
Input Isolation
–In line with CISPR 11 / EN 55011 group 1 class A (for a shielded test
setup)
–Complies with the emission requirements stipulated by EN 55011, EN
61326-1 and EN 61326-2-1 class A, making the instrument suitable for use
in industrial environments
●
Immunity: in line with IEC / EN 61326-1 table 2, immunity test requirements
for industrial environments
2.3Input Isolation
The instrument has independently floating isolated inputs. Each input channel has
its own signal input and its own reference input. Each input channel is electrically
isolated from the other input channels. Therefore, each reference of the used
inputs must be connected to a reference voltage. Furthermore, input channels are
electrically isolated from the communication ports and the power adapter input.
14Getting Started 1326.1561.02 ─ 09
R&S®Scope Rider RTH
Preface
Documentation Overview
Figure 2-1: Isolation scheme of the R&S RTH
The input isolation has several advantages:
●
You can measure independently floating signals simultaneously.
●
The risk of causing a short circuit while measuring multiple signals is reduced
substantially.
●
When measuring signals with different grounds, the induced ground currents
are kept to a minimum.
2.4Documentation Overview
The user documentation for the R&SRTH consists of the following parts:
●
Instrument Help
The instrument help is part of the instrument's firmware. It offers quick, context-sensitive access to the complete information directly on the instrument.
●
Basic Safety Instructions
This brochure provides safety instructions and operating conditions and further important information. The brochure is delivered with the instrument in
printed form.
15Getting Started 1326.1561.02 ─ 09
R&S®Scope Rider RTH
Documentation Overview
●
Getting Started
The Getting Started manual provides the information needed to set up and
start working with the instrument, and describes basic operations. The English
edition of this manual is delivered with the instrument in printed form. Editions
in other languages and the newest version of the English manual are available
on the product website.
●
User Manual
The user manual describes all instrument modes and functions in detail. It
also provides an introduction to remote control and a complete description of
the remote control commands with programming examples. The newest version of the manual is available in English on the R&S RTH product website at
www.rohde-schwarz.com/manual/rth.
●
Data Sheet
The data sheet contains the complete instrument specification. It also lists the
options and their order numbers, and optional accessories. The data sheet is
available on the R&S RTH product website at www.rohde-schwarz.com/
brochure-datasheet/rth.
Preface
●
Calibration Certificate
The document is available on https://gloris.rohde-schwarz.com/calcert.
●
Open Source Acknowledgment
The Open Source Acknowledgment document provides verbatim license text
of open source software that is used in the instrument's firmware. It is available on the R&S RTH website at www.rohde-schwarz.com/firmware/rth, and it
can be read directly on the instrument.
●
Instrument security procedures manual
Provides information on security issues when working with the R&S RTH in
secure areas.
●
Application cards and application notes
These documents deal with special applications or background information on
particular topics. See www.rohde-schwarz.com/application/rth
16Getting Started 1326.1561.02 ─ 09