Rohde&Schwarz R&S®RT-ZVC User Manual

®
R&S
RT-ZVC Multi-Channel Power Probe
User Manual
(=JEW2)
1326213902
Version 03
User Manual
This user manual describes the following R&S®RT-ZVC models:
R&S®RT-ZVC04 2x4 voltage/current channels for R&S®RTO2000/RTE (1326.0259.04 )
R&S®RT-ZVC02 2x2 voltage/current channels for R&S®RTO2000/RTE (1326.0259.02 )
R&S®RT-ZVC04 2x4 voltage/current channels for R&S®CMWrun (1326.0259.24 )
R&S®RT-ZVC02 2x2 voltage/current channels for R&S®CMWrun (1326.0259.22 )
R&S®RT-ZVC04A 2x4 voltage/current channels with autoranging for R&S®CMWrun (1326.0259.34 )
R&S®RT-ZVC02A 2x2 voltage/current channels with autoranging for R&S®CMWrun (1326.0259.32 )
© 2018 Rohde & Schwarz GmbH & Co. KG
Mühldorfstr. 15, 81671 München, Germany
Phone: +49 89 41 29 - 0
Fax: +49 89 41 29 12 164
Email: info@rohde-schwarz.com
Internet: www.rohde-schwarz.com
Subject to change – Data without tolerance limits is not binding.
R&S® is a registered trademark of Rohde & Schwarz GmbH & Co. KG.
Trade names are trademarks of the owners.
1326.2139.02 | Version 03 | R&S®RT-ZVC
The following abbreviations are used in this manual: R&S®RT-ZVC is abbreviated as R&S RT-ZVC. R&S®CMWrun is abbreviated as R&S CMWrun. R&S®RTO2000 and R&S®RTE are abbreviated as R&S RTx. R&S®RTO-B1E and R&S®RTE-B1E are abbreviated as R&S RTx-B1E.
R&S®RT-ZVC

Contents

Contents
1 Product Description.............................................................. 5
1.1 Prerequisites......................................................................................... 6
1.2 Key Characteristics.............................................................................. 6
1.3 Unpacking the Instrument....................................................................8
1.4 Instrument Overview............................................................................ 9
1.4.1 Probe Box............................................................................................. 10
1.4.2 Cables and Accessories....................................................................... 13
1.5 Optional Accessories......................................................................... 14
2 Putting into Operation.........................................................17
2.1 Test Setup with Oscilloscope............................................................ 19
2.2 Test Setup with R&S CMWrun........................................................... 20
2.3 Connecting the Probe to the DUT..................................................... 22
3 Measurement Principles..................................................... 24
3.1 Solder-In Connection..........................................................................24
3.2 Signal Integrity.................................................................................... 26
3.2.1 Ground Connection...............................................................................27
3.2.2 Electromagnetic Interference................................................................ 27
3.2.3 Accuracy............................................................................................... 28
3.2.4 Dynamic Range and Operating Window...............................................28
3.3 Probing Philosophy............................................................................ 29
3.3.1 Signal Flow for Oscilloscope Operation................................................ 29
3.3.2 Signal Flow for R&S CMWrun Operation (USB IF Only).......................30
3.3.3 Internal Shunt Operation.......................................................................31
3.3.4 External Shunt Operation......................................................................32
3.3.5 Autoranging Mode.................................................................................33
3User Manual 1326.2139.02 ─ 03
R&S®RT-ZVC
Contents
4 Maintenance and Service....................................................35
4.1 Service Strategy..................................................................................35
4.2 Returning the Probe for Servicing.................................................... 35
4.3 Cleaning...............................................................................................35
4.4 Calibration Interval............................................................................. 36
4.5 Discarding the Probe..........................................................................36
Index..................................................................................... 37
4User Manual 1326.2139.02 ─ 03
R&S®RT-ZVC
Product Description

1 Product Description

The R&S RT-ZVC is a multi-channel power probe with integrated 2- or 4-channel amperemeter and 2- or 4-channel voltmeter. It provides parallel measurements of analog or digital, voltage/current signals with excellent 18-bit resolution. Integra­ted post-processing logic calculates power and statistical results with high accu­racy.
The R&S RT-ZVC is a useful measurement instrument for such applications as for example:
Embedded electronics
Battery-driven devices (wearables, smartphones)
Power supply of integrated circuits
Battery life-time measurements
Power consumption measurements
Energy-harvesting
The multi-channel power probe is developed for tests with R&S CMWrun and for operation with an oscilloscope.
Table 1-1: Variants of setup with R&S RT-ZVC
Operation with an oscilloscope Standalone device with R&S CMWrun
Embedded electronics/circuit measurements, sophisticated processing
Data transferred via digital interface at full sample rate of 5 MSa/s
Probe powered by scope
System-level power measurements using R&S CMWrun, reduced sample rate via USB
Data transferred via USB to the R&S CMWrun controller PC
Externally powered
5User Manual 1326.2139.02 ─ 03
R&S®RT-ZVC
Product Description
Key Characteristics

1.1 Prerequisites

Operation with oscilloscope
Oscilloscope of type R&S RTO2000 equipped with R&S RTO-B1E or
R&S RTE equipped with R&S RTE-B1E is required.
Note, that old R&S RTx-B1 boards cannot run R&S RT-ZVC.
Operation with R&S CMWrun
R&S CMWrun, version 1.8.8 and later. The option R&S CMW-KT051 of
R&S CMWrun for general-purpose tests is required.
Windows supported operating system includes Windows 7 and Windows
10 (32-bit or 64-bit version).
A Rohde & Schwarz-specific driver is required to be installed at the run
PC, to support USB3.0 SuperSpeed (SS) interface. With the installed driver, no additional connection configuration is required. The probe acts as a plug-and-play device. The driver is delivered with the R&S CMWrun SW.

1.2 Key Characteristics

The R&S RT-ZVC probe houses integrated voltmeter, amperemeter, and post­processing logic. The probe performs built-in A/D conversion and signal post-pro­cessing. Within integrated post-processing, the R&S RT-ZVC calculates power and statistical results (min/max/avg/RMS). It measures the differential and com­mon mode AC and DC signals simultaneously.
The following figure illustrates the scheme of a probe with four voltage and four current channels.
6User Manual 1326.2139.02 ─ 03
R&S®RT-ZVC
Product Description
Key Characteristics
+
-
+
-
+
-
4x Amperemeter
+
-
+
-
+
-
+
4x Voltmeter
-
+
-
A
A
A
A
A
A
A
A
Figure 1-1: Electrical scheme
D
Power Supply
DC/DC
D
D
USB
D
D
Post-
processing
Logic
HDMI Connector
AC/DC Adapter
To CMWrun
(PC)
To Scope
D
D
CLK
D
There are separate A/D converters for each channel that allow for parallel and fully synchronous data acquisition.
The key characteristics of the probe are the following:
Table 1-2: Technical data
Voltmeter (V) Amperemeter (I) Power (internal mul-
tiplier)
No. of channels 2 or 4 2 or 4 2 or 4
Measurement resolu-
18 bit 18 bit 32 bit
tion
Sampling rate with
5 MSa/s per channel 5 MSa/s per channel 5 MSa/s per channel
R&S RTx
Sampling rate with R&S CMWrun
50 kSa/s per channel 50 kSa/s per channel 50 kSa/s per channel
(synchronous multipli­cation of V and I)
For detailed specification and derating information, consider the data sheet.
7User Manual 1326.2139.02 ─ 03
R&S®RT-ZVC

1.3 Unpacking the Instrument

Product Description
Unpacking the Instrument
Figure 1-2: Carrying case (model for R&S CMWrun)
Inspecting the contents
The carrying case contains the following items:
Basic voltage and current leads (1), length 16 cm, one lead per each instru­ment channel
USB 3.0 cable (2) (only for R&S CMWrun)
Ground cable (3) with 4 mm plugs
R&S RT-ZVC probe box (4)
Ground clip (5) for connection to the DUT
AC power adapter (6) (only for R&S CMWrun)
PCB solder-in cables (7):
Blue and red round 1.02 mm, length 5 cm, one pair per each current chan-
nel
Blue and red square 0.64 mm, length 5 cm, one pair per each voltage
channel
Exchangeable clips for AC power plugs (8) for UK, US, Australian and Euro­pean standard
8User Manual 1326.2139.02 ─ 03
R&S®RT-ZVC
Solder-in contacts (9):
8x round headers 1.02 mm
4x shorting links, round 1.02 mm, spacing 5.08 mm
Blue flat oscilloscope interface cable (10) (only for R&S RTx)
User manual
Inspect the package for damage
Keep the package and the cushioning material until the contents have been checked for completeness and the device has been tested.
If the packaging material shows any signs of stress, notify the carrier and your Rohde & Schwarz service center. Keep the damaged package and cushioning material for inspection.
Inspect the product
Product Description
Instrument Overview
If the content is incomplete, damaged, or defect or if the R&S RT-ZVC multi-chan­nel power probe does not operate properly, notify your Rohde & Schwarz service center.
Calibration certificate
The calibration certificate is not included in the standard shipping. The document is available for download at https://gloris.rohde-schwarz.com/calcert.

1.4 Instrument Overview

The probe R&S RT-ZVC consists of the following parts:
The probe box
Leads and accessories for connection to the DUT. See also Chapter 2.3,
"Connecting the Probe to the DUT", on page 22
Variant for operation with R&S CMWrun: AC power adapter and USB 3.0 cable for connection to the PC with R&S CMWrun application, refer to Chapter 2.2, "Test Setup with
R&S CMWrun", on page 20
9User Manual 1326.2139.02 ─ 03
R&S®RT-ZVC
Product Description
Instrument Overview
Variant for operation with an oscilloscope: Blue flat oscilloscope interface cable, refer to Chapter 2.1, "Test Setup with
Oscilloscope", on page 19
The next sections describe the main HW of the probe.

1.4.1 Probe Box

The following section provides an overview of the control elements and sockets of the probe box and explains how to connect external devices.
Figure 1-3: R&S RT-ZVC: upper, front, and rear view
1 = Status LEDs 2 = ID LEDs 3 = Channel LEDs 4 = Voltmeter sockets 5 = Amperemeter sockets 6 = Ground socket 7 = Socket for AC power adapter 8 = HDMI interface 9 = USB 3.0, micro B socket
LEDs
The status LEDs (1) of the probe box are located in the upper left corner of the upper plate. They light to indicate the following detections:
Power: Probe is on (connected to power supply or R&S RTx).
10User Manual 1326.2139.02 ─ 03
R&S®RT-ZVC
Scope: Probe is connected with oscilloscope.
Trigger: Trigger signal is detected on R&S RTx interface.
Overload: Red LED plus blinking affected channel indicate overload at partic­ular voltage or current channel. The overload indication can be reset via a control dialog. During an overload at a current channel, the R&S RT-ZVC switches automati­cally to the next higher range. In case also the 10A range is exceeded, the amperemeter switches to external shunt mode to protect the probe against permanent damage. Thus, the current flow through the DUT is interrupted. In external shunt mode, there is no automated range switching. The user must take care that the external shunt resistor is not damaged. Furthermore, there is no automated range switching at voltmeter channels for overload conditions.
The ID LEDs (2) are located in the middle left part of the upper plate.
Probe Z1 to probe Z4: Indicates on which port of the digital extension board R&S RTx-B1E the probe is attached to.
Product Description
Instrument Overview
Channel LEDs (3) are located in the bottom left corner of the upper plate.
I1 to I4: Blue LEDs for up to four current channels. LEDs indicate which chan­nel is enabled in the GUI. A light-dark flashing indicates active auto-ranging for the corresponding chan­nel (only relevant for variants R&S RT-ZVC04A and R&S RT-ZVC02A).
V1 to V4: Green LEDs for up to four voltage channels. LEDs indicate which channel is enabled in the GUI.
Interfaces
2 or 4 sockets (4) for voltage test leads AWG28. Red color at the socket marks the position of + wire. Number of sockets depends on the instrument model.
2 or 4 sockets (5) for current test leads AWG24. Red color at the socket marks the position of + wire. Number of sockets depends on the instrument model. The voltage and current sockets at the probe box are of different type, to avoid the mixing of voltage and current leads. In addition, current cables have
1.02 mm round pins that allow for currents up to 10 A.
Ground socket (6) - to establish common ground with the DUT. For detailed information on ground connection, refer to Chapter 3.2.1, "Ground Connec-
tion", on page 27.
Socket (7) for AC power adapter is used together with the USB interface.
11User Manual 1326.2139.02 ─ 03
R&S®RT-ZVC
USB 3.0 interface with micro B socket (9) for connection to R&S CMWrun.
HDMI socket (8) for blue oscilloscope interface cable to the digital extension board R&S RTx-B1E of an oscilloscope.
Risk of instrument damage during operation
An unsuitable test setup can damage the instrument and connected devi­ces. When operating the probe together with the oscilloscope, only oscillo­scopes of type R&S RTE and R&S RTO2000 must be used.
Amperemeter
The current measurement is performed with a shunt-amperemeter, i.e. the current measurement is reduced to a measurement of a voltage drop on a resistor where the current is flowing through. It has digitally adjustable gain. There are three built-in shunts that can be switched in series to the circuit under test. For using internal shunts, the circuit under test needs to be interrupted so that the current can flow through the probe.
Product Description
Instrument Overview
In addition to internal shunts, the amperemeter can work as a sensitive voltmeter to measure voltage drops on external shunts mounted directly in the circuit under test (external shunt mode). The maximum input voltage range is 450 mV to GND.
For details, refer to:
Chapter 3.3.3, "Internal Shunt Operation", on page 31
Chapter 3.3.4, "External Shunt Operation", on page 32
Data sheet
Voltmeter
The voltmeter measures both differential and common mode, not floating signals. The input voltage range is ±15 V to GND. The signal can be attenuated or ampli­fied depending on the selected range. Refer to the data sheet.
Power multiplier (only via USB for R&S CMWrun)
Within the integrated post-processing of R&S RT-ZVC, the internal multiplier uses the measured values of the voltmeter and amperemeter and provides high pre­cise statistical results. The minimal, maximal, average and RMS current and volt­age values are internally calculated with the sampling rate of 5 MSa/s per chan-
12User Manual 1326.2139.02 ─ 03