Rohde&Schwarz RTP-K110 HDMI Compliance Test

R&S®RTP-K110 HDMI Compliance Test User Manual
1179319602 Version 01
This document describes the HDMI compliance test procedures of the following option:
R&S®RTP-K110 (1802.9467.02)
© 2021 Rohde & Schwarz GmbH & Co. KG
Mühldorfstr. 15, 81671 München, Germany
Phone: +49 89 41 29 - 0
Email: info@rohde-schwarz.com
Internet: www.rohde-schwarz.com
Subject to change – data without tolerance limits is not binding.
R&S® is a registered trademark of Rohde & Schwarz GmbH & Co. KG.
Trade names are trademarks of the owners.
1179.3196.02 | Version 01 | R&S®RTP-K110
Throughout this manual, products from Rohde & Schwarz are indicated without the ® symbol , e.g. R&S®ScopeSuite is indicated as
R&S ScopeSuite.
R&S®RTP-K110
1 R&S ScopeSuite Overview....................................................................5
2 Preparing the Measurements................................................................7
2.1 Test Equipment..............................................................................................................7
2.2 Installing Software and License.................................................................................. 7
2.3 Setting Up the Network.................................................................................................8
2.4 Starting the R&S ScopeSuite....................................................................................... 8
2.5 Connecting the R&S RTP............................................................................................. 9
2.6 Report Configuration.................................................................................................. 10
3 Performing Tests..................................................................................12
3.1 Starting a Test Session...............................................................................................12

Contents

Contents
3.2 Configuring the Test................................................................................................... 13
3.2.1 General Test Settings....................................................................................................14
3.3 Initiating the Test.........................................................................................................16
3.4 Getting Test Results....................................................................................................16
3.5 Starting HDMI Tests.................................................................................................... 17
3.6 Test Configuration for HDMI.......................................................................................17
4 HDMI 1.4 Tests......................................................................................20
4.1 Test Equipment............................................................................................................20
4.2 Clock Tests.................................................................................................................. 20
4.2.1 All Clock Tests...............................................................................................................20
4.2.2 Voltage Off Tests........................................................................................................... 22
4.3 Data Tests.................................................................................................................... 23
4.3.1 Single-Ended Tests....................................................................................................... 23
4.3.2 Voltage Off Tests........................................................................................................... 25
4.3.3 Differential Timing Tests................................................................................................ 26
4.3.4 Jitter Mask Tests........................................................................................................... 27
4.3.5 Inter-Pair Skew..............................................................................................................28
5 HDMI 2.1 TMDS Tests...........................................................................30
5.1 Test Equipment............................................................................................................30
5.2 Clock Tests.................................................................................................................. 30
3User Manual 1179.3196.02 ─ 01
R&S®RTP-K110
5.2.1 All Clock Tests............................................................................................................... 30
5.2.2 Measurements.............................................................................................................. 31
5.3 Data Tests.................................................................................................................... 33
5.3.1 Single-Ended Tests....................................................................................................... 33
5.3.2 Differential Timing Tests................................................................................................ 34
5.3.3 Differential Voltage Mask Tests..................................................................................... 35
5.3.4 Jitter Mask Tests........................................................................................................... 36
5.3.5 Inter-Pair Skew..............................................................................................................37
Contents
4User Manual 1179.3196.02 ─ 01
R&S®RTP-K110

1 R&S ScopeSuite Overview

R&S ScopeSuite Overview
The R&S ScopeSuite software is used with R&S RTP oscilloscopes. It can be installed on a test computer or directly on the oscilloscope. For system requirements, refer to the Release Notes.
The R&S ScopeSuite main panel has several areas:
"Settings": connection settings to oscilloscope and other instruments also default report settings
"Compliance Tests": selection of the compliance test
"Demo": accesses demo test cases that can be used for trying out the software without having a connection to an oscilloscope
"Help": opens the help file, containing information about the R&S ScopeSuite con­figuration
"About": gives information about the R&S ScopeSuite software
"Tile View": allows a personalization of the compliance test selection You can configure which tests are visible in the compliance test section and which are hidden, so that only the ones you use are displayed.
► To hide a test from the "Compliance Tests" view, do one of the following:
5User Manual 1179.3196.02 ─ 01
R&S®RTP-K110
R&S ScopeSuite Overview
a) Right-click on the compliance test you want to hide.
The icon of the test changes, see Figure 1-1. Now with a left click you can hide the test.
Figure 1-1: Unpin icon
b) Click on "Title View" to show a list of the available test cases. By clicking a test
case in the show list, you can pin/unpin it from the main panel.
6User Manual 1179.3196.02 ─ 01
R&S®RTP-K110

2 Preparing the Measurements

2.1 Test Equipment

Preparing the Measurements
Installing Software and License
For HDMI compliance tests, the following test equipment is needed:
R&S RTP with 4 channels and minimum 8 GHz bandwidth
R&S RTP-K101 HDMI compliance test option (required option, installed on the R&S RTP)
2 modular probes, R&S ZM160 with 9 GHz bandwidth.
2 modular probe tips R&S RT-ZMA40
HDMI Termination Module and a HDMI TPA adapter
EDID Emulator
The free-of-charge R&S ScopeSuite software, which can be installed on a com­puter or directly on the R&S RTP.

2.2 Installing Software and License

The preparation steps are performed only once for each computer and instrument that are used for testing.
Uninstall older versions of the R&S ScopeSuite
If an older version of the R&S ScopeSuite is installed, make sure to uninstall the old version before you install the new one. You can find the version number of the current installation in "Help" menu > "About". To uninstall the R&S ScopeSuite, use the Win­dows " Control Panel" > "Programs".
For best operation results, we recommend that the installed firmware versions of the R&S ScopeSuite and the oscilloscope are the same.
To install the R&S ScopeSuite
1. Download the latest R&S ScopeSuite software from the "Software" section on the Rohde & Schwarz R&S RTP website:
www.rohde-schwarz.com/product/rtp.html www.rohde-schwarz.com/product/rto.html
2. Install the R&S ScopeSuite software:
On the computer that is used for testing, or
On the R&S RTP.
For system requirements, refer to the Release Notes.
7User Manual 1179.3196.02 ─ 01
R&S®RTP-K110
Preparing the Measurements
To install the license key on the R&S RTP
► When you got the license key of the compliance test option, enable it on the oscil-
loscope using [Setup] > "SW Options". For a detailed description, refer to the R&S RTP user manual, chapter "Installing Options", or to the online help on the instrument.

2.3 Setting Up the Network

If the R&S ScopeSuite software runs on a test computer, the computer and the testing oscilloscope require a LAN connection.
There are two ways of connection:
LAN (local area network): It is recommended that you connect to a LAN with DHCP server. This server uses the Dynamic Host Configuration Protocol (DHCP) to assign all address information automatically. If no DHCP server is available, or if the Tabor WX2182B or WX2182C is used for automatic testing, assign fixed IP addresses to all devices.
Direct connection of the instruments and the computer or connection to a switch using LAN cables: Assign fixed IP addresses to the computer and the instruments and reboot all devices.
Starting the R&S
ScopeSuite
To set up and test the LAN connection
1. Connect the computer and the instruments to the same LAN.
2. Start all devices.
3. If no DHCP server is available, assign fixed IP addresses to all devices.
4. Ping the instruments to make sure that the connection is established.
5. If VISA is installed, check if VISA can access the instruments.
a) Start VISA on the test computer. b) Validate the VISA address string of each device.
See also:
Chapter 2.5, "Connecting the R&S RTP", on page 9

2.4 Starting the R&S ScopeSuite

To start the R&S ScopeSuite on the test computer or on the oscilloscope:
► Double-click the R&S ScopeSuite program icon.
8User Manual 1179.3196.02 ─ 01
R&S®RTP-K110
Preparing the Measurements
To start the R&S ScopeSuite on the instrument, in the R&S RTP firmware:
► In the "Apps" dialog, open the "Compliance" tab.

2.5 Connecting the R&S RTP

If the R&S ScopeSuite is installed directly on the instrument, the software detects the R&S RTP firmware automatically, and the "Oscilloscope" button is not available in the R&S ScopeSuite.
If the R&S ScopeSuite software runs on a test computer, the computer and the testing oscilloscope require a LAN connection, see Chapter 2.3, "Setting Up the Network", on page 8. The R&S ScopeSuite software needs the IP address of the oscilloscope to establish connection.
1. Start the R&S RTP.
2. Start the R&S ScopeSuite software.
3. Click "Settings" > "Oscilloscope".
Connecting the R&S
RTP
4. Enter the IP address of the oscilloscope. To obtain the IP address: press the Rohde & Schwarz logo at the top-right corner of the oscilloscope's display.
5. Click "Get Instrument Information".
The computer connects with the instrument and gets the instrument data.
9User Manual 1179.3196.02 ─ 01
R&S®RTP-K110
Preparing the Measurements
Report Configuration
If the connection fails, an error message is shown.

2.6 Report Configuration

In the "Report Configuration" menu, you can select the format of the report and the details to be included in the report. You can also select an icon that is displayed in the upper left corner of the report.
Also, you can enter common information on the test that is written in the "General Infor­mation" section of the test report.
10User Manual 1179.3196.02 ─ 01
R&S®RTP-K110
Preparing the Measurements
Report Configuration
11User Manual 1179.3196.02 ─ 01
R&S®RTP-K110

3 Performing Tests

3.1 Starting a Test Session

Performing Tests
Starting a Test Session
After you open a compliance test, the "Session Selection" dialog appears. In this dia­log, you can create new sessions, open or view existing report.
The following functions are available for handling test sessions:
Function Description
"Add" Adds a new session
"Open" Opens the selected session
"Remove" Removes the selected session
"Rename" Changes the "Session Name"
"Comment" Adds a comment
"Show report" Generates a report for the selected session
To add a test session
1. In the R&S ScopeSuite window, select the compliance test.
2. In the "Session Selection" dialog press "Add".
3. If necessary change the "Session Name"