Rohde&Schwarz NRP-Z81, NRP-Z85, NRP-Z86 Operating Manual

Operating Manual
Wideband Power Sensor
R&S
1137.9009.02
R&S
1411.7501.02
R&S
1417.0109.40/.44
Test and Measurement
1172.9530.12-08 1
Dear Customer,
R&S® is a registered trademark of Rohde & Schwarz GmbH & Co. KG Trade names are trademarks of the owners.
1172.9530.12-08 2
R&S NRP-Z81/Z85/Z86 Safety Information
1172.9530.12 I E-1

Safety Information

Safety information is part of the product documentation. It warns you of potential dangers and gives instructions on how to prevent personal injury or damage caused by dangerous situations. Safety information is provided as follows:
In the Getting Started manual, Chapter 1.1 “Safety instructions”. The same information is provided in many languages as printed "Safety Instructions". Both documents are delivered with the power sensor.
Throughout the documentation, safety instructions are provided when you need to take care during setup or operation.

R&S NRP-Z81/-Z85/-Z86 Supplement

Operation of the
R&S NRP-Z81/-Z85/-Z86 wideband power sensors
on R&S NRP base unit
Dear Customer,
for proper operation of the R&S NRP-Z81/-Z85/-Z86 wideband power sensors on a R&S NRP, the right software is required for the base unit:
Main Program (Application) 5.01 Bootloader 5.01 Keyboard Controller 3.22
The revision numbers for the software components installed in the base unit can be displayed under menu item 'System Info', lines 'Main Program', 'Bootloader' and 'Keybd. Ctrl.'. The 'System Info' can be found in the 'System' menu.
Please note, that the statistics capabilities of the sensors are not yet supported by these revisions.
1172.9530.12 Supplement 1 E-1
R&S NRP-Z81/-Z85/-Z86 Table of Contents Chapter 1
Table of Contents
1 Putting into Operation ......................................................................................................1.1
Unpacking the sensor ......................................................................................................................1.1
Connecting the sensor ....................................................................................................................1.1
Operation with the R&S NRP/NRP2 power meter .........................................................................1.2
Connecting the power sensor .....................................................................................................1.2
PC control .........................................................................................................................................1.2
Hardware and software requirements ........................................................................................1.2
Operation via the R&S NRP-Z4 passive USB adapter ...............................................................1.3
Connecting the sensor to the DUT .............................................................................................1.4
Operation via the Active USB Adapter R&S NRP-Z3 .................................................................1.4
Operation via the R&S NRP-Z5 Sensor Hub ..............................................................................1.5
Equipment Supplied and Alternative Accessories ..........................................................1.5
Ports and LEDs ...............................................................................................................1.6
Test Setup .......................................................................................................................1.7
Troubleshooting ..............................................................................................................1.9
Operation with other Rohde & Schwarz test instruments..........................................................1.10
Hardware and software requirements ......................................................................................1.10
1172.9530.12 I-1.1 E-6
List of Figs. and Tables Chapter 1 R&S NRP-Z81/-Z85/-Z86
Figs.
Fig. 1-1 Displaying the total available power of a USB port ...........................................................1.3
Fig. 1-2 Configuration with Passive USB Adapter R&S NRP-Z4 ...................................................1.3
Fig. 1-3 Configuration with Active USB Adapter R&S NRP-Z3 ......................................................1.4
Fig. 1-4 Changing the primary adapter ..........................................................................................1.5
Fig. 1-5 Ports and LEDs on front panel ..........................................................................................1.6
Fig. 1-6 Ports on rear panel ...........................................................................................................1.6
Fig. 1-7 Unlatching the R&S Instrument connector ........................................................................1.7
Fig. 1-8 Typical test setup with R&S NRP-Z5 and PC ...................................................................1.8
Fig. 1-9 Changing the view in Device Manager .............................................................................1.9
Fig. 1-10 Identification of USB root hub on which R&S NRP-Z5 is operated ..................................1.9
Fig. 1-11 Deactivation of Selective Suspend for a USB root hub ..................................................1.10
Tables
Table 1-1 RF connectors of R&S NRP-Z81/85/86 power sensors ...................................................1.1
1172.9530.12 I-1.2 E-6
R&S NRP-Z81/-Z85/-Z86 Unpacking the sensor
NOTICE
Follow the instructions below precisely to prevent damage to the power sensor – particularly when you are putting it into operation for the first time.
The power sensor contains components which can be destroyed by electrostatic discharges. To prevent this happening, never touch the inner conductor of the RF connector and never open the power sensor.
NOTICE
To prevent EMI, the power sensor must never be operated with its enclosure wholly or partially removed. Only use shielded cables that meet the relevant EMC standards.
Never exceed the maximum RF power limit. Even brief overloads can destroy the power sensor.
Power sensor
Male connector
Matching female connector
Tightening torque
R&S NRP-Z81
N N 1.36 Nm (12" lbs)
R&S NRP-Z85
2.92 mm
3.5 mm/2.92 mm/SMA
0.9 Nm (8" lbs) R&S NRP-Z86
2.4 mm
2.4 mm/1.85 mm
1 Putting into Operation

Unpacking the sensor

Remove the sensor from its packing and check that nothing is missing. Inspect all items for damage. If you discover any damage, inform the carrier responsible immediately and keep the packing to support any claims for compensation.
It is also best to use the original packing if the sensor is to be shipped or transported at a later date.

Connecting the sensor

Caused by their different frequency ranges, the R&S NRP-Z81/85/86 power sensors have different RF connectors. In this context, the following sections will refer to Table 1-1:
Table 1-1 RF connectors of R&S NRP-Z81/85/86 power sensors
1172.9530.12 1.1 E-6
Operation with the R&S NRP/NRP2 power meter R&S NRP-Z81/-Z85/-Z86

Operation with the R&S NRP/NRP2 power meter

Connecting the power sensor

The power sensor can be connected to the R&S NRP/NRP2 base unit when it is in operation. The multiple circular plug-in connector must be inserted, red marking upwards, into one of the R&S NRP/NRP2 base unit’s sensor connectors. When the power sensor is connected, it is detected by the R&S NRP/NRP2 base unit and initialized.
See Table 1-1 for the type of male RF connector a certain power sensor has and which female connectors it mates with. Using light pressure, and keeping the male connector perpendicular, insert it into the female connector and tighten the locking nut (right-hand thread). In many cases, the RF connector only requires manual tightening. However, for maximal measurement accuracy, the RF connector must be tightened using a torque wrench with a nominal torque that corresponds with the tightening torque shown in Table 1-1.

PC control

Hardware and software requirements

The following requirements must be met if the power sensor is to be controlled by a PC via an interface adapter:
The PC must have a USB port.  Either Microsoft Windows XP (32 Bit), Microsoft Windows Vista (32 or 64 Bit), Microsoft Windows 7
(32 or 64 Bit), or x86 Linux (with kernel 2.6.8 oder more recent) must be installed as the operating system of the PC. 64-bit versions of Microsoft Windows XP are not supported. Operation with older versions of Microsoft Windows with USB support (especially Microsoft Windows 2000) is normally possible but cannot be guaranteed.
The USB device drivers in the supplied NRP Toolkit software package must be installed. If these requirements are met, the power sensor can be controlled using a suitable application program
such as the NrpFlashup program contained in the Microsoft Windows version of the NRP Toolkit (includes the modules Power Viewer, USB Terminal, Firmware Update and Update S-Parameters).
When you insert the CD-ROM supplied with the R&S NRP power sensor, the NRP Toolkit is automatically installed on your PC. The rest of the procedure is self-explanatory.
The power sensor can be powered in two ways: Self-powered from a separate power supply via the Active USB Adapter R&S NRP-Z3 or the
Sensor Hub R&S NRP-Z5.
Bus-powered from the PC or a USB hub with its own power supply (self-powered hub) via the
Active USB Adapter R&S NRP-Z3 or via the Passive USB Adapter R&S NRP-Z4.
As the power sensor is a high-power device, there is no guarantee that it can be powered from all types of laptop or notebook in the bus-powered mode. To be sure, you should determine the current at the USB connectors beforehand:
1172.9530.12 1.2 E-6
R&S NRP-Z81/-Z85/-Z86 PC control
In the Windows™ start menu, select Settings – Control Panel Select the System icon Select the Hardware tab By clicking on the button with that name, start the Device Manager Open USB Controller (all USB controllers, hubs and USB devices are listed here) Double-click on USB Root Hub or select Properties in the context menu (use the
right-hand mouse button)
Select the Power tab (Fig. 1-1). If the hub is self-powered and the total power
available is, as indicated by Hub Information, 500 mA per port, high-power devices can be connected.
Fig. 1-1 Displaying the total available power of a USB port
If you have any doubts, ask the manufacturer if the USB port on your laptop or notebook can handle high-power devices.

Operation via the R&S NRP-Z4 passive USB adapter

Fig. 1-2 is a schematic of the measurement setup. The order in which the cables are connected is not critical.
Fig. 1-2 Configuration with Passive USB Adapter R&S NRP-Z4
1172.9530.12 1.3 E-6
PC control R&S NRP-Z81/-Z85/-Z86

Connecting the sensor to the DUT

See Table 1-1 for the type of male RF connector a certain power sensor has and which female connectors it mates with. Using light pressure, and keeping the male connector perpendicular, insert it into the female connector and tighten the locking nut (right-hand thread). In many cases, the RF connector only requires manual tightening. However, for maximal measurement accuracy, the RF connector must be tightened using a torque wrench with a nominal torque that corresponds with the tightening torque shown in Table 1-1.

Operation via the Active USB Adapter R&S NRP-Z3

Fig. 1-3 shows the configuration with the Active USB Adapter R&S NRP-Z3, which also makes it possible to feed in a trigger signal for the Timeslot and Trace modes. The order in which the cables are connected is not critical.
Fig. 1-3 Configuration with Active USB Adapter R&S NRP-Z3 The plug-in power supply for the R&S NRP-Z3 can be powered from a single-phase AC source with a
nominal voltage range of 100 V to 240 V and a nominal frequency between 50 Hz and 60 Hz. The plug­in power supply autosets to the applied AC voltage. No manual voltage selection is required.
The plug-in power supply comes with four primary adapters for Europe, the UK, the USA and Australia. No tools of any kind are required to change the primary adapter. The adapter is pulled out manually and another adapter inserted until it locks (Fig. 1-4).
1172.9530.12 1.4 E-6
R&S NRP-Z81/-Z85/-Z86 PC control
NOTICE
The plug-in power supply is not intended for outdoor use. Keep within the temperature range of 0°C to 50°C. If there is any condensation on the plug-in power supply, dry it off before connecting it to
the AC supply.
Fig. 1-4 Changing the primary adapter The plug-in power supply is short-circuit-proof and has an internal fuse. It is not possible to replace this
fuse or open the plug-in power supply.

Operation via the R&S NRP-Z5 Sensor Hub

The R&S NRP-Z5 sensor hub allows up to four power sensors to be operated on one PC. It combines the following functions:
4-port USB 2.0 hub  Power supply  Through-wired trigger bus  Trigger input and trigger output via BNC sockets
Equipment Supplied and Alternative Accessories
The following equipment is supplied:
Sensor hub  External power supply  Power cable  USB cable
Instead of the supplied, 2 m long USB cable, it is possible to use any other USB-2.0-certified cable (USB connector type A to USB connector type B) with a maximum length of 5 m. If a locking connection is required at the instrument end, it is also possible to use the passive R&S NRP-Z4 interface adapter instead of a standard USB cable.
As an alternative to the supplied external power supply, power can also be supplied using a DC voltage source with an output voltage of 12 V to 24 V and a power output of at least 24 W. Make sure that the polarity is correct (positive pole inside, negative pole outside). The alternative DC voltage source must be in the same building as the R&S NRP-Z5 and must be connected to the latter via a cable with a maximum length of 30 m. Power must not be supplied from a low-voltage supply system.
The supplied external power supply is short-circuit-proof and is also protected by an internal fuse. It is not possible to change the fuse or open the unit.
1172.9530.12 1.5 E-6
PC control R&S NRP-Z81/-Z85/-Z86
NOTICE
The external power supply is not intended for outdoor use. Keep within the temperature range of 0°C to 50°C. If there is any condensation on the external power supply, dry it off before connecting it
to the AC supply.
1–4
Sockets for connecting the power sensors
5–8
Green LEDs for indicating normal operation
9–12
Yellow LEDs for indicating faults
1
Trigger output (BNC socket)
2
Trigger input (BNC socket)
3
Plug allowing connection to Rohde & Schwarz measuring instruments (via R&S NRP-Z2) or to a PC (via R&S NRP-Z4)
4
USB socket (type B) for connection to the USB host (PC or Rohde & Schwarz measuring instrument without sensor socket)
5
Power supply socket (supplied external power supply unit or alternative power supply (12 V to 24 V/24 W))
Ports and LEDs
The power sensors are connected to the front panel of the R&S NRP-Z5. The ports and LEDs on the front panel of the R&S NRP-Z5 are shown in Fig. 1-5.
Fig. 1-5 Ports and LEDs on front panel
On the rear panel are the ports for power supply, USB host, trigger input and trigger output. The ports on the rear panel are shown in Fig. 1-6.
Fig. 1-6 Ports on rear panel
1172.9530.12 1.6 E-6
R&S NRP-Z81/-Z85/-Z86 PC control
Unlatch the R&S Instrument connector on the rear panel by pressing down the unlatching ring of the built-in plug and pulling off the cable jack at the same time (Fig. 1-7).
Fig. 1-7 Unlatching the R&S Instrument connector
1
R&S NRP-Z5 sensor hub
2
External power supply unit (supplied)
3
Power cable (supplied)
4
AC power supply
5
USB cable (supplied)
6
PC with USB host port
7, 8
BNC cable (optional, not supplied)
9
Trigger source (optional)
10
Triggered device (optional)
11-14
R&S NRP-Zxx power sensors (one to four sensors)
15
Signal source
Test Setup
Fig. 1-8 shows a typical test setup:
The sequence in which the cables are connected is not important. As already mentioned in section Ports and LEDs, the R&S NRP-Z4 passive interface adapter can also
be used as an alternative to a standard USB cable for connecting to the PC. This adapter is connected to the R&S Instrument port. Simultaneous operation at two USB hosts is not possible as the USB host connected to the R&S Instrument port always has priority.
It is possible to cascade several R&S NRP-Z5 sensor hubs by connecting the R&S Instrument port of an R&S NRP-Z5 to one of the sensor ports of another R&S NRP-Z5. However, external triggering and the use of the Trigger Master function are then not possible. Instead, it is recommended to connect all R&S NRP-Z5 hubs individually to the USB host or to an interposed USB hub, and to feed the external trigger signal to all R&S NRP-Z5 hubs via their trigger inputs; or to forward the external trigger signal from the trigger output of the R&S NRP-Z5 to which the Trigger Master is connected to the trigger inputs of the other R&S NRP-Z5 hubs.
1172.9530.12 1.7 E-6
PC control R&S NRP-Z81/-Z85/-Z86
Fig. 1-8 Typical test setup with R&S NRP-Z5 and PC
1172.9530.12 1.8 E-6
R&S NRP-Z81/-Z85/-Z86 PC control
Troubleshooting
A frequent problem, especially in the case of operation under Microsoft Windows XP, is that the R&S NRP-Z5 fails to respond after a power sensor connected to the R&S NRP-Z5 is disconnected and reconnected. It is then necessary to restart the PC.
This problem is not specific to the R&S NRP-Z5, but occurs with all USB hubs. You can try to eliminate the problem by switching off the Selective Suspend for the USB port of the PC on which the R&S NRP­Z5 is operated. The following step-by-step description is based on an English version of Windows XP:
1. Connect the R&S NRP-Z5 and put it into operation. Connect at least one power sensor.
2. Start the Device Manager and change the view by selecting View - Devices by connection (Fig. 1-9).
Fig. 1-9 Changing the view in Device Manager
3. Now identify the USB root hub on which the R&S NRP-Z5 is operated. To do so, expand the hierarchical display of the PC components, as shown in Fig. 1-10. Under PCI bus, find the USB host controller to whose root hub the R&S NRP-Z5 is connected. The R&S NRP-Z5 is shown as the Generic USB Hub. It can be distinguished from other USB hubs by the power sensor connected to it.
Fig. 1-10 Identification of USB root hub on which R&S NRP-Z5 is operated
1172.9530.12 1.9 E-6
Operation with other R&S test instruments R&S NRP-Z81/-Z85/-Z86
4. Double-clicking USB Root Hub opens the USB Root Hub Properties dialog box. Select the Power Management tab. Remove the check mark next to Allow the computer to turn off this device to save power and confirm this with OK.
Fig. 1-11 Deactivation of Selective Suspend for a USB root hub

Operation with other R&S test instruments

Hardware and software requirements

Many Rohde & Schwarz test instruments allow power measurements using power sensors of the R&S NRP-Zxx series. The power sensors are generally connected to the USB ports of the test instrument via one of the two interface adapters, R&S NRP-Z3 or R&S NRP-Z4, or the Sensor Hub R&S NRP-Z5. On some instruments, such as the R&S SMU signal generator, there is also an R&S NRP sensor connector available for a direct connection. The R&S NRP-Z5 can be hooked up to this connector via an extension cable R&S NRP-Z2 (Model .03, .05, or .15). For that purpose, the R&S Instrument connector of the R&S NRP-Z5 is used.
1172.9530.12 1.10 E-6
R&S NRP-Z81/-Z85/-Z86 Table of Contents Chapter 2
Table of Contents
2 Virtual Power Meter ...........................................................................................................2.1
Overview............................................................................................................................................2.1
Menus .........................................................................................................................................2.3
1172.9530.12 I-2.1 E-2
List of Figs. and Tables Chapter 2 R&S NRP-Z81/-Z85/-Z86
Figs.
Fig. 2-1 Power Viewer virtual power meter ...................................................................................2.1
Tables
Table 2-1 Virtual power meter keys ..................................................................................................2.2
Table 2-2 Virtual power meter entry fields ........................................................................................2.2
1172.9530.12 I-2.2 E-2
R&S NRP-Z81/-Z85/-Z86 Overview
2 Virtual Power Meter
You will find the NrpFlashup program for controlling sensors with a PC under Windows™ on the CD-ROM that accompanies the sensor. The program comprises several modules which can be started centrally via the Windows™ start-menu entry NRP Toolkit.
This section describes the Power Viewer program module. This is a virtual power meter which only uses a cut-down set of the sensor’s functions. This means that after an extremely brief familiarization period, the user can measure the average power of modulated signals.
The other modules in NrpFlashup are described in Chapter 3 of the operating manual (Terminal and
Update S-Parameters modules) or in the service manual (Firmware Update module).

Overview

Start the virtual power meter using the NRP Toolkit – Power Viewer start-menu entry. The Power Viewer program window is displayed (Fig. 2-1).

Fig. 2-1 Power Viewer – virtual power meter

The result display occupies most of the program window. The result, unit and additional sensor status information are displayed. The serial number of the sensor is displayed in the bottom right. The program window also contains animated buttons and entry fields (see Table 2-1 and Table 2-2).
1172.9530.12 2.1 E-2
Overview R&S NRP-Z81/-Z85/-Z86
Table 2-1 Virtual power meter keys
Button Function Key combination
Exit Terminates the program. The current settings are saved and recalled the
W Selects Watt as the display unit. Alt + W
dBm Selects dBm as the display unit. Alt + M
Zero Zeroes the sensor. Alt + Z
dB Selects dB as the display unit. This is the log of the ratio of the measured
;% Selects % as the display unit. The difference between the measured value
M2Ref Makes the current measured value the reference value for the relative
Offset On/Off
Averaging Man/Auto
Apply
next time the program is started.
value to the reference value.
and the reference value is expressed as a percentage.
display units dB and %.
Turns the offset correction for the sensor on or off. If the offset correction is Off, the Offset/dB entry field has a grey background.
Turns auto-averaging on or off. When auto-averaging is on, the Length entry field has a grey background; the current averaging factor is displayed.
Accepts edited numerical values in the Frequency/Hz, Value/dB and Length entry fields and transfers them to the sensor.
Alt + E
Alt + B
Alt + %
Alt + R
Alt + N
Alt + T
Alt + A or Enter key
Table 2-2 Virtual power meter entry fields
Entry field Function
Frequency/Hz Frequency of the RF carrier in Hertz.
Value/dB Attenuation in dB of the twoport connected to the sensor. The valid range is –100 to 100. The offset
Length Length of the averaging filter (= averaging factor). The valid range is 1 to 65536. Averaging must be set
correction must be activated beforehand with the Offset On/Off button if this entry field is to be edited.
to manual with the Averaging Man/Auto button if this entry field is to be edited.
Scientific notation can also be used for the entry fields. If an invalid entry is made, an error message is output. An edited numerical value will not be transferred to the sensor unless you use the Apply button or the Enter key to terminate the entry.
1172.9530.12 2.2 E-2
R&S NRP-Z81/-Z85/-Z86 Overview

Menus

he menu bar can be used to call less frequently used functions.
T
File Start Log ...
Stop Log
View Display Refresh Rate
Colours Result
Unit Edit Button
Resolution
Options Read Sensor Status …
Opens a file-selection dialog to specify the path and name of the log file. Clicking the Save button starts the recording. All displayed values are written line-by-line to the log file with the date (format: YY/MM/DD) and time (format: hh:mm:ss.ms). Example:
-22.51 dBm (03/02/25 15:37:25.310)
Ends the log-file recording.
Opens a dialog box to adjust the display refresh rate. The time in milliseconds between two refresh operations is entered. The default setting is 200 ms.
Opens a dialog box to select the background colour for
the result,
the unit,
the text in the number fields or
the key labelling.
For setting the result resolution. If auto­averaging has been selected, a higher resolution leads to a greater averaging factor, which means a longer result settling time.
Reads the current sensor status. A parameter list is output.
Read Error Queue …
1172.9530.12 2.3 E-2
Reads the error queue. All the error messages that have been issued since the last call are read line­by-line. A tick before this menu entry indicates that an error has occurred.
Overview R&S NRP-Z81/-Z85/-Z86
Simulation …
Reset Sensor
Help Contents
About
For trying out the functions of the virtual power meter without actually connecting a sensor. The display alternates between
Measurement Value 1 & Measurement Value 2
with a period given by Interval. Simulation can be activated immediately with the Activate check box.
Initializes the sensor. Any previous zeroing remains valid.
Opens the table of contents for the online-help facility.
Displays information about the program version used, etc.
1172.9530.12 2.4 E-2
R&S NRP-Z81/-Z85/-Z86 Table of Contents Chapter 3
Table of Contents
3 Manual Operation ...............................................................................................................3.1
Program module “Terminal” ...........................................................................................................3.1
Main control elements.................................................................................................................3.1
Menus .........................................................................................................................................3.3
Program module “Firmware Update”.............................................................................................3.6
Program module “Update S-Parameters” ......................................................................................3.6
Fundamentals .............................................................................................................................3.6
Procedure ...................................................................................................................................3.9
1172.9530.12 I-3.1 E-3
List of Figs. and Tables Chapter 3 R&S NRP-Z81/-Z85/-Z86
Figs.
Fig. 3-1
Fig. 3-2 Sending commands using command files ........................................................................3.2
Fig. 3-3 Dialog window for loading an s-parameter table...............................................................3.9
Fig. 3-4 Dialog window for loading the backup file of a calibration data set ................................3.10
Fig. 3-5 Subsequently changing the default behaviour of the s-parameter correction ................3.11
Sending commands using the Input field .........................................................................3.1
Tables
Table 3-1
Table 3-2 Buttons assigned to the Command File field ...................................................................3.2
Table 3-3 Buttons assigned to the Output field ................................................................................3.3
Table 3-4 Uncertainties of the s-parameter test system (example) ..................................................3.7
Table 3-5 Interpolated uncertainties of measurement frequencies for s-parameters (example) ......3.7
Buttons assigned to the Input field...................................................................................3.2
1172.9530.12 I-3.2 E-3
R&S NRP-Z81/-Z85/-Z86 Program module “Terminal”
3 Manual Operation
The previous section describes the Power Viewer program module supplied with the instrument. This module simplifies the most frequently used function of a power meter – measuring the average power of an RF signal of almost any modulation. Other program modules are also part of the supplied equipment and can be selected in the Start menu:
Power Viewer: A detailed description of this virtual power meter module is provided in section 2.
Terminal: Program module for sending commands and command sequences to the sensor and for displaying measurement results, status information and other data from the sensor
Firmware Update: Program module for updating the sensor firmware
Update S-Parameters: Program module for loading an s-parameter table into the sensor

Program module “Terminal”

Main control elements

With the USB terminal, commands and command sequences can be sent to the sensor in two different ways:
Commands are entered in the Input field (Fig. 3-1). Consecutive commands can be entered as separate lines, one below the other. The buttons associated with the Input field are described in Table 3-1.
Commands or command sequences are stored in command files. Command files are created with a text editor, for instance, and then stored. They can be called as often as required (Fig. 3-2). The buttons of the Command File field are described in Table 3-2.
Fig. 3-1 Sending commands using the Input field
1172.9530.12 3.1 E-3
Program module “Terminal” R&S NRP-Z81/-Z85/-Z86
Fig. 3-2 Sending commands using command files
Table 3-1 Buttons assigned to the Input field
Button Function Key combination
Send
Loop
Clear
Font key
Colour key
Sends the content of the Input entry field to the sensor.
With Loop the command or command sequence is cyclically sent. Pressing the button again terminates the cyclic transmission. The repetition rate is set in a dialog window that can be opened with View - Loop...
Clears the content of the Input field.
Opens a dialog window where the font for the Input field can be selected.
Opens a dialog window where the background colour of the Input field can be selected.
Alt + S
Alt + L
Alt + R
Table 3-2 Buttons assigned to the Command File field
Button Function Key combination
Send Sends the content of the command file to the sensor. Alt + E
Loop
With Loop the command or command sequence is cyclically sent. Pressing the button again terminates the cyclic transmission. The repetition rate is set in a dialog window that can be opened with View - Loop...
Alt + O
History
Edit Opens the selected command file in the Windows™ text editor. Alt + D
Opens a file opening dialog for selecting the command file.
Opens a window for editing the command file name in the Command File field.
Alt + H
1172.9530.12 3.2 E-3
R&S NRP-Z81/-Z85/-Z86 Program module “Terminal”
where the lines stored in
Only lines starting with the entered character
Only lines not starting with the entered character
Only lines containing the entered character string
Only lines not containing the entered character
Lines not matching the specific filter criterion are blanked but not
A command line starting with a tab, a blank or a special character is considered a comment and not
orwarded to the sensor.
f
Measurement results, parameters and status information returned by the sensor are displayed in the Output field.
Table 3-3 Buttons assigned to the Output field
Button Function Key combination
Clear
Copy
Font button
Colour button
Clears the content of the Output field
Copies the content of the Output field to the clipboard. (Another possiblity: mark the desired information in the output window with the mouse cursor, press the right mouse key or Ctrl+C and then copy the selected text to the clipboard using the menu item Copy in the opened context menu.)
Opens a dialog window where the font for the Output field can be selected.
Opens a dialog window where the background colour of the Output field can be selected.
Close the USB terminal with OK.

Menus

View Post Filter … Opens the Output Postfilter dialog window
the input buffer can be filtered according to different criteria.
Alt + A
Alt + Y
Filter criteria:
Only + Starting with: string are displayed.
Not + Starting with: string are displayed.
Only + Containing: are displayed.
Not + Containing: string are displayed.
cleared.
1172.9530.12 3.3 E-3
Program module “Terminal” R&S NRP-Z81/-Z85/-Z86
The number of lines matching the filter
is
dialog is automatically opened when the
dialog window where the response time of
indicates the time elapsed between dispatch of the last
button is clicked, the response times exceeding the
button is
field are
is active, the time
measurement is terminated as soon as the first response arrives after a
is
active, the time measurement is terminated as soon as the first
dialog
og is
The
is active, the
The dialog
Filtering is started with Apply. criterion is displayed in the Linecounter field. If Open on startup active, the Output Postfilter terminal is started. The dialog window is closed with OK.
.
Response Time …
Opens the Response time the sensor can be set.
Current
command and receipt of an acknowledgement from the sensor. When the Max value in the limit / ms field are recorded. When the Min clicked, the response times within the value in the limit / ms recorded.
If Trigger on first incoming response
command is sent. If Trigger on response matching filter settings
response matching the filter criterion in the Output Postfilter window is received.
If Open on startup is active, the Response Time dial automatically displayed when the Terminal module is started. dialog window is closed with OK.
Loop … Opens the Loop controls dialog
window where the cyclic transfer of commands and command sequences can be controlled.
In the Delay / ms field, the time interval for the cyclic transfer is specified in milliseconds.
The number of completed transfer cycles is displayed in the Counter field. If Open on startup Response time dialog is automatically opened when the Terminal module is started. window is closed with OK.
1172.9530.12 3.4 E-3
Loading...
+ 70 hidden pages