Rohde&Schwarz R&S®FSMR Quick Start Guide Getting started

R&S
®
FSMR
1313.9698.02
Measuring Receiver
Getting Started
03
The Getting Started describes the following R&S®FSMR models and options:
& R&S FSMR3 1313.9200.03 & R&S FSMR26 1313.9200.26 & R&S FSMR431313.9200.43 & R&S FSMR50 1313.9200.50
The software contained in this product makes use of several valuable open source software packages. For information, see the "Open Source Acknowledgment" on the user documentation CD-ROM (included in delivery). Rohde & Schwarz would like to thank the open source community for their valuable contribution to embed-
ded computing.
09 Rohde & Schwarz GmbH & Co. KG
© 20 81671 Munich, Germany Printed in Germany – Subject to change – Data without tolerance limits is not binding.
®
R&S
is a registered trademark of Rohde & Schwarz GmbH & Co. KG.
Trade names are trademarks of the owners.
The following abbreviations are used throughout this manual:
®
FSMR is abbreviated as R&S FSMR.
R&S
1
Risk of injury and instrument damage
The instrument must be used in an appropriate manner to prevent personal injury or instrument damage.
Do not open the instrument casing.
Read and observe the "Basic Safety Instructions" delivered as
printed brochure with the instrument.
Read and observe the safety instructions in the following sections.
Note that the data sheet may specify additional operating conditions.
Keep the "Basic Safety Instructions" and the product documentation
in a safe place and pass them on to the subsequent users.
Riesgo de lesiones y daños en el instrumento
El instrumento se debe usar de manera adecuada para prevenir descargas eléctricas, incendios, lesiones o daños materiales.
No abrir la carcasa del instrumento.
Lea y cumpla las "Instrucciones de seguridad elementales"
suministradas con el instrumento como folleto impreso.
Lea y cumpla las instrucciones de seguridad incluidas en las
siguientes secciones. Se debe tener en cuenta que las especificaciones técnicas pueden contener condiciones adicionales para su uso.
Guarde bien las instrucciones de seguridad elementales, así como
la documentación del producto, y entréguelas a usuarios posteriores.
Safety Instructions Instrucciones de seguridad Sicherheitshinweise Consignes de sécurité
1171.1307.42 - 05
2
Gefahr von Verletzungen und Schäden am Gerät
Betreiben Sie das Gerät immer ordnungsgemäß, um elektrischen Schlag, Brand, Verletzungen von Personen oder Geräteschäden zu verhindern.
Öffnen Sie das Gerätegehäuse nicht.
Lesen und beachten Sie die "Grundlegenden Sicherheitshinweise",
die als gedruckte Broschüre dem Gerät beiliegen.
Lesen und beachten Sie die Sicherheitshinweise in den folgenden
Abschnitten; möglicherweise enthält das Datenblatt weitere Hinweise zu speziellen Betriebsbedingungen.
Bewahren Sie die "Grundlegenden Sicherheitshinweise" und die
Produktdokumentation gut auf und geben Sie diese an weitere Benutzer des Produkts weiter.
Risque de blessures et d'endommagement de l'appareil
L'appareil doit être utilisé conformément aux prescriptions afin d'éviter les électrocutions, incendies, dommages corporels et matériels.
N'ouvrez pas le boîtier de l'appareil.
Lisez et respectez les "consignes de sécurité fondamentales"
fournies avec l’appareil sous forme de brochure imprimée.
Lisez et respectez les instructions de sécurité dans les sections
suivantes. Il ne faut pas oublier que la fiche technique peut indiquer
des conditions d’exploitation supplémentaires.
Gardez les consignes de sécurité fondamentales et la
documentation produit dans un lieu sûr et transmettez ces documents aux autres utilisateurs.
1171.1307.42 - 05

Customer Support

Technical support – where and when you need it
For quick, expert help with any Rohde & Schwarz equipment, contact one of our Customer Support Centers. A team of highly qualified engineers provides telephone support and will work with you to find a solution to your query on any aspect of the operation, programming or applications of Rohde & Schwarz equipment.
Up-to-date information and upgrades
To keep your instrument up-to-date and to be informed about new application notes related to your instrument, please send an e-mail to the Customer Support Center stating your instrument and your wish. We will take care that you will get the right information.
Europe, Africa, Middle East
North America
Latin America
Asia/Pacific
China
Phone +49 89 4129 12345
customersupport@rohde-schwarz.com
Phone 1-888-TEST-RSA (1-888-837-8772)
customer.support@rsa.rohde-schwarz.com
Phone +1-410-910-7988
customersupport.la@rohde-schwarz.com
Phone +65 65 13 04 88
customersupport.asia@rohde-schwarz.com
Phone +86-800-810-8228 / +86-400-650-5896
customersupport.china@rohde-schwarz.com
1171.0200.22-06.00
R&S FSMR Tabbed Divider Overview

Tabbed Divider Overview

Contents
Safety Instructions Certificate of Quality EU Certificate of Conformity List of R&S Representatives
Manuals for Measuring Receiver R&S FSMR
Tabbed Divider
Chapter 1: Putting into Operation
Chapter 2: Getting Started
Chapter 3: Menu Overview
Index
1166.6204.62 RE E-1
R&S FSMR Documentation Overview

Documentation Overview

The documentation of the R&S FSMR consists of base unit manuals and option manuals. All manu­als are provided in PDF format on the CD-ROM delivered with the instrument. Each software option available for the instrument is described in a separate software manual.
The base unit documentation comprises the following manuals and documents:
Quick Start Guide
Operating Manual
Service Manual
Release Notes
Apart from the base unit, these manuals describe the models and options of the R&S FSMR Measu­ring Receiver that are listed in the documentation overview of the Operating Manual. All other options are described in separate manuals. These manuals are provided on the CD-ROM. For an overview of all options available for the R&S FSMR visit the R&S FSMR Measuring Receiver Internet site. This manual is delivered with the instrument in printed form. It provides the information needed to set up and start working with the instrument. Basic operations and basic measurements are described. The function descriptions are provided in the Operating Manual. The Quick Start Guide includes gen­eral information (e.g. Safety Instructions) and the following chapters:
Quick Start Guide
Chapter 1 describes the control elements and connectors on the front and rear panel as well
as all procedures required for putting the R&S FSMR into operation and integra­tion into a test system.
Chapter 2 gives an introduction to typical measurement tasks of the R&S FSMR which are
explained step by step.
Chapter 3 describes the operating principles, the structure of the graphical interface and of-
fers a menu overview.
Chapter 10 contains an index for the quick start guide.
1166.6204.62 0.1 E-1
R&S FSMR Contents - Preparing for Operation
Contents - Chapter 1 " Preparing for Operation "
1 Preparing for Operation...................................................................................... 1.1
Description of Front and Rear Panel Views ...................................................................................1.1
Front View ................................................................................................................................1.1
Rear View...............................................................................................................................1.11
Getting Started with the Instrument .............................................................................................1.16
Preparing the Instrument for Operation..................................................................................1.16
Setting Up the Instrument.......................................................................................................1.16
Standalone Operation ..................................................................................................1.16
Safety Instruction for Instruments with Tiltable Feet ....................................................1.17
Rackmounting ..............................................................................................................1.17
EMC Safety Precautions ........................................................................................................1.18
Connecting the Instrument to the AC Supply .........................................................................1.18
Switching the Instrument On/Off ............................................................................................1.18
Switching On the Instrument........................................................................................1.19
Startup Menu and Booting ...........................................................................................1.19
Switching Off the R&S FSMR ......................................................................................1.19
Operating the R&S FSMR with the Removable Hard Disk Option R&S FSMR-B18 ...1.19
Power-Save Mode........................................................................................................1.20
Recalling the Most Recent Instrument Settings .....................................................................1.20
Function Test ..................................................................................................................................1.20
Properties Specific to Windows XP ..............................................................................................1.21
Connecting an External Keyboard................................................................................................1.22
Connecting a Mouse ......................................................................................................................1.23
Connecting an External Monitor ...................................................................................................1.24
Connecting a Printer ......................................................................................................................1.25
Selecting a Printer ..................................................................................................................1.25
Installation of Plug&Play Printers ...........................................................................................1.28
Installation of Non-Plug&Play Printers....................................................................................1.28
Local Printer.................................................................................................................1.30
Configuring a Network Printer ................................................................................................1.34
Connection of USB Devices ..........................................................................................................1.36
Installing Windows XP Software...................................................................................................1.38
Windows XP Software Tested for the Instrument ..................................................................1.38
1313.9723.12 I-1.1 E-1
Contents - Preparing for Operation R&S FSMR
Fig. 1-1 Front View
1313.9723.12 I-1.2 E-1
R&S FSMR Front View

1 Preparing for Operation

Chapter 1 describes the controls and connectors of the Spectrum Analyzer R&S FSMR by means of the front and rear view. Then follows all the information that is necessary to put the instrument into operation and connect it to the AC supply and to external devices.
A more detailed description of the hardware connectors and interfaces can be found in chapter 8. Chapter 2 provides an introduction into the operation of the R&S FSMR by means of typical examples of configuration and measurement; for the description of the concept for manual operation and an overview of menus refer to chapter 3.
For a systematic explanation of all menus, functions and parameters and background information refer to the reference part in chapter 4.
For remote control of the R&S FSMR refer to the general description of the SCPI commands, the instrument model, the status reporting system, and command description in chapter 5 and 6.

Description of Front and Rear Panel Views

Front View

1
Display Screen see Chapter 3
2
Softkeys see Chapter 3
3
Keypad for data input
0...9 input numbers
see Chapter 3
. input decimal point
change sign
ESC – close input field (for uncompleted or CANCEL already closed inputs, the original
entry is kept)
– erase the current entry in input field
(beginning of an input)
– close message window (status, error
and warning messages)
ENTER close the data input.
BACK – erase last character input for
uncompleted input
– restore previous input (undo)
1313.9723.12 1.1 E-2
Front View R&S FSMR
Fig. 1-1 Front View
1313.9723.12 1.2 E-2
R&S FSMR Front View
Keypad for data input
GHz s The units keys close the data
-dBm V input and define the multipli- cation factor for each basic unit. MHz ms For dimension-less or dBm mV alphanumeric inputs, the units keys have weight 1.
kHz µs They behave, in this case, like the dB µV ENTER key.
Hz ns dB.. nV
see Chapter 3
4
FREQ Set frequency axis
SPAN Set span
AMPT Set level indication and configure
RF input.
MKR Select and set standard marker and
delta marker functions.
MKR-> Change instrument settings via
markers
MKR Select further marker and delta FCTN marker functions
see Chapter 4
5
BW – Set resolution bandwidth, video
bandwidth and sweep time,
– Set coupling of these parameters
SWEEP Select sweep
MEAS Select and set power measurements
TRIG Set trigger sources
see Chapter 4
1313.9723.12 1.3 E-2
Front View R&S FSMR
Fig. 1-1 Front View
1313.9723.12 1.4 E-2
R&S FSMR Front View
6
Key group for entering data and for cursor movement
Cursor keys – Move the cursor within the input
fields and tables.
Vary the input value. Define the direction of movement
for the roll-key.
Roll-key – Vary input values.
Move markers and limits. Select letters in the help line editor. Move cursor in the tables Close data input (ENTER)
see Chapter 3
7
Connector for the power sensors NRP-Zx
8
Output connector for an external noise source s. Chapter 3
9
CF Card option R&S FSU-B18
1313.9723.12 1.5 E-2
Front View R&S FSMR
Fig. 1-1 Front View
1313.9723.12 1.6 E-2
R&S FSMR Front View
10
AF OUTPUT
Volume control
Head phone connector
PROBE POWER Power supply and coded socket
(+15 V/ -12 V) for accessories
USB
RF INPUT RF input see Chapter 8
NOTICE!
The maximum DC voltage is 50 V, the maximum power is 1 W (=^ 30 dBm) at 10 dB attenuation.
AUDIO INPUT Audio input with 50 J or 1 MJ input impedance
NOTICE!
The input can handle a maximum power of 1 W or a peak voltage of 20 V.
POWER REF Reference source 50 MHz, 1mW
NOTICE!
USB interfaces
s. Chapter 8
The maximum DC voltage is 0.
11
TRACE Select and activate traces and detectors
LINES Set limit lines
DISP Configure display
FILE – Save and recall instrument data
– Configuration of memory media and data
1313.9723.12 1.7 E-2
see Chapter 4
Front View R&S FSMR
Fig. 1-1 Front View
1313.9723.12 1.8 E-2
R&S FSMR Front View
12
13
14
15
HCOPY
Menu-change keys
NEXT Change to side menu
PREV Call main menu
Hotkeys see Chapter 3
ON/STANDBY switch see Chapter 1
Configure and start a print job see Chapters 1 and 4
see Chapter 3
16
17
18
Define general configuration see Chapter 4
SETUP
Record correction data see Chapter 4
CAL
Call default settings see Chapter 4
1313.9723.12 1.9 E-2
Rear View R&S FSMR
20
100- 240 VAC
3.1- 1.3 A
21
625
SCPI
LPT
22
23
24
MONITOR COM
19
625
IEC 2
SCPI
LAN
IEC 2
OUT
I / Q DATA OUT
VIDEO
IF OUT
REF OUT
640 MHz
404.4 MHz
REF OUT
AUX CONTROL
MOUSE
GATE IN
EXT TRIGGER/
SWEEP
25
26
27
28
29
IF OUT
34
33
REF IN
20.4 MHz
32
1...20 MHz
31
30
Fig. 1-2 Rear View
1313.9723.12 1.10 E-2
R&S FSMR Rear View
Rear View
19
nd
IEC/IEEE bus-connector see Chapter 8
2
20
100 - 2 40 VAC
3. 1 - 1 . 3 A
Power switch and AC power connector see Chapter 1
I 0
21
22
23
24
625
SCPI
COM
MONITOR
LPT
IEC/IEEE bus-connector see Chapter 8
Parallel interface connector
see Chapter 8
(printer connector)
Connector for a serial interface
see Chapter 8
(9-pin socket; COM)
Connector for an external monitor see Chapter 8
1313.9723.12 1.11 E-2
Rear View R&S FSMR
20
100- 240 VAC
3.1- 1.3 A
21
625
SCPI
LPT
22
23
24
MONITOR COM
19
625
IEC 2
SCPI
LAN
IEC 2
OUT
I / Q DATA OUT
VIDEO
IF OUT
REF OUT
640 MHz
404.4 MHz
REF OUT
AUX CONTROL
MOUSE
GATE IN
EXT TRIGGER/
SWEEP
25
26
27
28
29
IF OUT
34
33
REF IN
20.4 MHz
32
1...20 MHz
31
30
Fig. 1-2 Rear View
1313.9723.12 1.12 E-2
R&S FSMR Rear View
25
MOUSE
26
Connector for a PS/2 mouse see Chapter 8
USB
27
AUX CONTROL
28
EXT TRIG /
GATE IN
29
Connector for USB see Chapter 8
Connector to control an external generator ((only with option FSP-B10)
Input connector for an external trigger or an
see Chapter 8
external gate signal
SWEEP
Output connector
see Chapter 8
During a sweep saw toth voltage is output which is proportional to frequency
30
Reserved for options
31
REF IN
1...20 MHz
REF OUT
1313.9723.12 1.13 E-2
REF IN Input connector for an external
reference (1 to 20 MHz)
REF OUT Output connector for an internal
reference (10 MHz)
Rear View R&S FSMR
20
100- 240 VAC
3.1- 1.3 A
21
625
SCPI
LPT
22
23
24
MONITOR COM
19
625
IEC 2
SCPI
LAN
IEC 2
OUT
I / Q DATA OUT
VIDEO
IF OUT
REF OUT
640 MHz
404.4 MHz
REF OUT
AUX CONTROL
MOUSE
GATE IN
EXT TRIGGER/
SWEEP
25
26
27
28
29
IF OUT
34
33
REF IN
20.4 MHz
32
1...20 MHz
31
30
Fig. 1-2 Rear View
1313.9723.12 1.14 E-2
R&S FSMR Rear View
32
20.4 MHz
33
34
IF OUT
LAN
IF OUT
404.4 MHz
IF 20.4 MHz OUT Output connector for
see Chapter 8
20.4 MHz IF
IF 404.4 MHz OUT Output connector for
404.4 MHz IF
LAN-Interface
Reserved for options
1313.9723.12 1.15 E-2
Getting Started with the Instrument R&S FSMR

Getting Started with the Instrument

The following section describes how to activate the instrument and how to connect external devices such as printer and monitor.
Chapter 2 explains the operation of the instrument using simple measurement examples.
Optional notice heading
Prior to switching on the instrument, make sure that the following conditions are fulfilled:
The instrument cover is in place and tightly screwed on
Fan
Signal levels at the inputs are within specified limits
Signal outputs are connected correctly and not overloaded.
Any non-compliance may cause damage to the instrument
openings are not obstructed

Preparing the Instrument for Operation

Take the instrument out of the packaging and check whether the
items listed in the packing list and in the lists of accessories are all included.
Remove the two protective covers from the front and rear of the
R&S FSMR and carefully check the instrument for damage.
remove protective caps
Should the instrument be damaged, immediately notify the carrier
and keep the box and packing material.
For further transport or shipment of the R&S FSMR, the original
packing should be used. It is recommended to keep at least the two protective covers of the front and rear panels in order to prevent damage to the controls and connectors.

Setting Up the Instrument

Standalone Operation
The instrument is designed for use under general laboratory conditions. The ambient conditions required at the site of operation are as follows:
The ambient temperature must be in the range specified in
Wrist strap with cord
Buildin g ground
Ground connection of operational site
Heel strap Floor mat
the data sheet.
All fan openings must be unobstructed and the air flow at
the rear panel and at the side-panel perforations must be unimpeded. The distance to the wall should be at least 10 cm.
he
T
mounting surface should be flat.
To avoid damage of electronic components of the DUT
due to electrostatic discharge on manual touch, protective measures against electrostatic discharge are recommended.
1313.9617.12 1.16 E-2
R&S FSMR Getting Started with the Instrument
instrument be guaranteed and reliable operation be ensured. With the feet out, the total
Safety Instruction for Instruments with Tiltable Feet
The feet must be fully folded in or out. Only in this way can the stability of the
load for the feet must not exceed 500 N (own weight and additional units put onto the instrument). These units must be secured against slipping (e.g. by locking the feet of the unit at the top side of the enclosure).
<500N
When shifting the instrument with the feet out, the feet might collapse and fold in. To avoid injuries, the instrument must therefore not be shifted with the feet out.
The instrument can be operated in any position.
Rackmounting
For rack installation, ensure that the air flow at the side-panel perforations and the air exhaust at the rear panel are not obstructed.
The instrument may be installed in a 19" rack by using a rack adapter kit (Order No. see data sheet). The installation instructions are part of the adapter kit.

EMC Safety Precautions

In order to avoid electromagnetic interference (EMI), the instrument may be operated only with all covers closed. Only adequately shielded signal and control cables may be used (see recommended accessories).
1313.9617.12 1.17 E-2
Getting Started with the Instrument R&S FSMR

Connecting the Instrument to the AC Supply

The R&S FSMR is equipped with an AC voltage selection feature and will automatically adapt itself to the applied AC voltage (range: 100 to 240 V AC, 40 to 400 Hz). External voltage selection or adaptation of the fuses are not necessary. The AC power connector is located on the rear panel (see below).
Connect the instrument to the AC power source using the AC
I
o
Power connector
power cable delivered with the instrument.
As the instrument is designed according to the regulations for safety class EN61010, it must be connected to a power outlet with earthing contact.

Switching the Instrument On/Off

Do not power down during booting.
Such a switch-off may lead to corruption of the hard disk files
AC power switch on the rear panel
P o we r s w i t c h
Power connector
I 0
ON/STANDBY switch on the front panel
ON STANDBY
Power switch
Position I = ON
In the I position, the instrument is in standby mode or in ope
ration, depending on the position of the ON/STANDBY
key at the front of the instrument.
Note: The AC power switch may remain ON continuously.
Switch i n g t o O FF i s o n ly required w h e n t h e instrument must be completely removed from the AC power source.
Position O = OFF
The 0 position implies an all-pole disconnection of the
instrument from the AC power source.
Standby switch
The ON/STANDBY switch activates two different operating modes indicated by coloured LEDs:
Operation ON - ON/STANDBY is depressed
The green LED (ON) is illuminated. The instrument is ready for operation. All modules within the instrument are supplied with power.
Shock hazard
In standby mode, the AC power voltage is present within the instrument
STANDBY - ON/STANDBY switch is not pressed.
The yellow LED (STANDBY) is illuminated. Only the power supply is supplied with power and the quartz oven is maintained at normal operating temperature.
1313.9617.12 1.18 E-2
R&S FSMR Getting Started with the Instrument
Switching On the Instrument
In order to switch on the R&S FSMR, set the power switch on the rear panel to position I.
Set the R&S FSMR to operating mode by pressing the ON/STANDBY key on the front panel. The
green LED must be illuminated.
Startup Menu and Booting
After switching on the instrument, a message indicating the installed BIOS version (e.g. Analyzer BIOS Rev. 1.2) appears on the screen for a few seconds.
Subsequently Windows XP is booted first and after that the instrument firmware will boot. As soon as the boot process is finished the instrument will start measuring. The settings used will be the one that was active when the instrument was previously switched off, provided no other device configuration than
FACTORY had been selected with STARTUP RECALL in the FILE menu.
Switching Off the R&S FSMR
Switch the ON/STANDBY key on the front panel to standby mode by pressing it once.
The R&S FSMR will write the current instrument settings to disk before performing a Windows XP shutdown. At the end of the shutdown procedure the power supply will be switched to STANDBY mode.
The yellow LED must be illuminated.
Only when removing the R&S FSMR completely from the AC power source:
Set the power switch at the rear panel to position 0.
Operating the R&S FSMR with the Removable Hard Disk Option R&S FSMR-B18
The R&S FSMR-B18 option allows the instrument to be operated with differently configured hard disks. Additional hard disks are therefore available as option FSU-B19.
Type and source of hazardous situation
Switch off the instrument before you remove and exchange the hard disk. Exchanging the hard disk during operation can produce severe data errors.
1313.9617.12 1.19 E-2