Rohde&Schwarz R&S®FPL1000 Getting Started Getting started

®
R&S
FPL1000 Spectrum Analyzer Getting Started
1179465702 Version 15
This manual applies to the following R&S®FPL1000 models with firmware version
2.00 and later:
R&S®FPL1003 (1304.0004K03) - FPL1000 with maximum frequency 3 GHz
R&S®FPL1007 (1304.0004K07) - FPL1000 with maximum frequency 7.5 GHz
R&S®FPL1014 (1304.0004K14) - FPL1000 with maximum frequency 14 GHz
R&S®FPL1026 (1304.0004K26) - FPL1000 with maximum frequency
26.5 GHz
In addition to the base unit, the following options are described:
R&S FPL1-B4, OCXO (1323.1902.02)
R&S FPL1-B5, Additional Interfaces (1323.1883.02)
R&S FPL1-B9, Internal Generator (1323.1925.03/1323.1925.07)
R&S FPL1-B10, GPIB interface (1323.1890.02)
R&S FPL1-B11, YIG preselector bypass (1323.1619.02)
R&S FPL1-B22, preamplifier (1323.1719.02)
R&S FPL1-B25, electronic attenuator (1323.1990.02)
R&S FPL1-B30, DC power supply (1323.1877.02)
R&S FPL1-B31, Li-Ion battery pack and charger (1323.1725.02)
R&S FPL1-K9, Power Sensor Support (1323.1754.02)
© 2022 Rohde & Schwarz GmbH & Co. KG
Muehldorfstr. 15, 81671 Muenchen, Germany
Phone: +49 89 41 29 - 0
Email: info@rohde-schwarz.com
Internet: www.rohde-schwarz.com
Subject to change – data without tolerance limits is not binding.
R&S® is a registered trademark of Rohde & Schwarz GmbH & Co. KG.
All other trademarks are the properties of their respective owners.
1179.4657.02 | Version 15 | R&S®FPL1000
Throughout this manual, products from Rohde & Schwarz are indicated without the ® symbol , e.g. R&S®FPL1000 is indicated as R&S FPL1000, R&S® FPL1-Bxx/-Kxx is indicated as R&S FPL1-Bxx/-Kxx.
R&S®FPL1000

Contents

1 Safety and regulatory information....................................... 7
1.1 Safety instructions................................................................................7
1.2 Labels on R&S FPL1000..................................................................... 11
1.3 Warning messages in the documentation........................................12
1.4 Korea certification class A.................................................................12
2 Documentation overview.................................................... 13
2.1 Getting started manual.......................................................................13
2.2 User manuals and help.......................................................................13
Contents
2.3 Service manual....................................................................................14
2.4 Instrument security procedures........................................................ 14
2.5 Printed safety instructions.................................................................14
2.6 Data sheets and brochures................................................................14
2.7 Release notes and open source acknowledgment (OSA)...............15
2.8 Application notes, application cards, white papers, etc................. 15
2.9 Calibration certificate......................................................................... 15
3 Key features......................................................................... 16
4 Preparing for use................................................................. 17
4.1 Lifting and carrying............................................................................ 17
4.2 Unpacking and checking....................................................................17
4.3 Choosing the operating site.............................................................. 18
4.4 Setting up the R&S FPL1000..............................................................18
4.5 Connecting to power.......................................................................... 21
4.6 Switching on or off............................................................................. 24
4.7 Connecting to LAN............................................................................. 25
3Getting Started 1179.4657.02 ─ 15
R&S®FPL1000
4.8 Connecting a keyboard...................................................................... 27
4.9 Connecting an external monitor........................................................27
4.10 Windows operating system................................................................29
4.11 Logging on.......................................................................................... 31
4.12 Checking the supplied options..........................................................33
4.13 Performing a self-alignment.............................................................. 34
4.14 Considerations for test setup............................................................ 34
Contents
5 Instrument tour.................................................................... 36
5.1 Front panel view..................................................................................36
5.2 Rear panel view...................................................................................43
6 Trying out the instrument................................................... 51
6.1 Measuring a basic signal................................................................... 51
6.2 Displaying a spectrogram.................................................................. 53
6.3 Activating additional measurement channels..................................55
6.4 Performing sequential measurements..............................................60
6.5 Setting and moving a marker.............................................................61
6.6 Displaying a marker peak list............................................................ 63
6.7 Zooming into the display................................................................... 64
6.8 Saving settings................................................................................... 68
6.9 Printing and saving results................................................................70
7 Operating the instrument....................................................72
7.1 Understanding the display information - spectrum mode.............. 72
7.2 Accessing the functionality............................................................... 81
7.3 Entering data....................................................................................... 87
7.4 Touchscreen gestures........................................................................89
7.5 Getting help......................................................................................... 92
4Getting Started 1179.4657.02 ─ 15
R&S®FPL1000
Contents
8 Customer support................................................................94
8.1 Collecting information for support....................................................94
8.2 Contacting customer support............................................................96
Index..................................................................................... 98
5Getting Started 1179.4657.02 ─ 15
R&S®FPL1000
Contents
6Getting Started 1179.4657.02 ─ 15
R&S®FPL1000
Safety and regulatory information
Safety instructions

1 Safety and regulatory information

The product documentation helps you use the product safely and efficiently. Fol­low the instructions provided here and in the following chapters.
Intended use
The product is intended for the development, production and verification of elec­tronic components and devices in industrial, administrative, and laboratory envi­ronments. Use the product only for its designated purpose. Observe the operating conditions and performance limits stated in the data sheet.
Where do I find safety information?
Safety information is part of the product documentation. It warns you of potential dangers and gives instructions on how to prevent personal injury or damage caused by dangerous situations. Safety information is provided as follows:
In Chapter 1.1, "Safety instructions", on page 7. The same information is provided in many languages as printed "Safety Instructions". The printed "Safety Instructions" are delivered with the product.
Throughout the documentation, safety instructions are provided when you need to take care during setup or operation.

1.1 Safety instructions

Products from the Rohde & Schwarz group of companies are manufactured according to the highest technical standards. To use the products safely, follow the instructions provided here and in the product documentation. Keep the prod­uct documentation nearby and offer it to other users.
Use the product only for its intended use and within its performance limits. Inten­ded use and limits are described in the product documentation such as the data sheet, manuals and the printed "Safety Instructions". If you are unsure about the appropriate use, contact Rohde & Schwarz customer service.
Using the product requires specialists or specially trained personnel. These users also need sound knowledge of at least one of the languages in which the user interfaces and the product documentation are available.
7Getting Started 1179.4657.02 ─ 15
R&S®FPL1000
Never open the casing of the product. Only service personnel authorized by Rohde & Schwarz are allowed to repair the product. If any part of the product is damaged or broken, stop using the product. Contact Rohde & Schwarz customer service at http://www.customersupport.rohde-schwarz.com.
Lifting and carrying the product
The maximum weight of the product is provided in the data sheet. To move the product safely, you can use lifting or transporting equipment such as lift trucks and forklifts. Follow the instructions provided by the equipment manufacturer.
Choosing the operating site
Only use the product indoors. The product casing is not waterproof. Water that enters can electrically connect the casing with live parts, which can lead to elec­tric shock, serious personal injury or death if you touch the casing. If Rohde & Schwarz provides accessories designed for your product, e.g. a carrying bag, you can use the product outdoors.
Safety and regulatory information
Safety instructions
Unless otherwise specified, you can operate the product up to an altitude of 2000 m above sea level. The product is suitable for pollution degree 2 environ­ments where nonconductive contamination can occur. For more information on environmental conditions such as ambient temperature and humidity, see the data sheet.
Setting up the product
Always place the product on a stable, flat and level surface with the bottom of the product facing down. If the product is designed for different positions, secure the product so that it cannot fall over.
If the product has foldable feet, always fold the feet completely in or out to ensure stability. The feet can collapse if they are not folded out completely or if the prod­uct is moved without lifting it. The foldable feet are designed to carry the weight of the product, but not an extra load.
If stacking is possible, keep in mind that a stack of products can fall over and cause injury.
If you mount products in a rack, ensure that the rack has sufficient load capacity and stability. Observe the specifications of the rack manufacturer. Always install the products from the bottom shelf to the top shelf so that the rack stands securely. Secure the product so that it cannot fall off the rack.
8Getting Started 1179.4657.02 ─ 15
R&S®FPL1000
Connecting to power
The product is an overvoltage category II product. Connect the product to a fixed installation used to supply energy-consuming equipment such as household appliances and similar loads. Keep in mind that electrically powered products have risks, such as electric shock, fire, personal injury or even death.
Take the following measures for your safety:
Before switching on the product, ensure that the voltage and frequency indica­ted on the product match the available power source. If the power adapter does not adjust automatically, set the correct value and check the rating of the fuse.
If a product has an exchangeable fuse, its type and characteristics are indica­ted next to the fuse holder. Before changing the fuse, switch off the product and disconnect it from the power source. How to change the fuse is described in the product documentation.
Safety and regulatory information
Safety instructions
Only use the power cable delivered with the product. It complies with country­specific safety requirements. Only insert the plug into an outlet with protective conductor terminal.
Only use intact cables and route them carefully so that they cannot be dam­aged. Check the power cables regularly to ensure that they are undamaged. Also ensure that nobody can trip over loose cables.
If the product needs an external power supply, use the power supply that is delivered with the product or that is recommended in the product documenta­tion or a power supply that conforms to the country-specific regulations.
Only connect the product to a power source with a fuse protection of maxi­mum 20 A.
Ensure that you can disconnect the product from the power source at any time. Pull the power plug to disconnect the product. The power plug must be easily accessible. If the product is integrated into a system that does not meet these requirements, provide an easily accessible circuit breaker at the system level.
Handling batteries safely
The product contains exchangeable or built-in lithium polymer or lithium ion cells or batteries. The use of the word battery in the following always means all types. Only the battery contents are potentially hazardous. As long as a battery is undamaged and the seals remain intact, there is no danger.
9Getting Started 1179.4657.02 ─ 15
R&S®FPL1000
Impact, shock or heat can cause damage such as dents, punctures and other deformations. A damaged battery poses a risk of personal injury. Handle a dam­aged or leaking battery with extreme care. Immediately ventilate the area since the battery releases harmful gases. If you come into contact with the battery fluid, immediately remove all contaminated clothing. Irritation can occur if the battery fluid comes in contact with your skin or eyes. Immediately and thoroughly rinse your skin or eyes with water and seek medical aid.
For safe handling, follow these rules:
Do not short-circuit the battery.
Do not mechanically damage the battery. Do not open or disassemble the bat­tery.
Do not expose the battery to high temperatures such as open flames, hot sur­faces and sunlight.
Only use the battery with the designated Rohde & Schwarz product.
Safety and regulatory information
Safety instructions
Only use the appropriate Rohde & Schwarz charger to charge the batteries. If the batteries are improperly charged, there is a risk of explosion. For charging and discharging temperature ranges, see the product documentation.
Replace exchangeable batteries only with the same battery type.
Store the battery in the product or use the product packaging.
Dispose of exchangeable batteries separately from normal household waste as specified by the local waste disposal agency.
If you disregard these rules, you risk serious personal injury or even death due to explosion, fire or hazardous chemical substances. The product documentation provides further details.
If exchangeable batteries or products with built-in batteries are defective, contact the Rohde & Schwarz customer service. Rohde & Schwarz classifies the severity of the defect. When returning batteries or Rohde & Schwarz products containing batteries, use a carrier qualified to transport dangerous goods and notify the car­rier of this classification. Follow the carrier’s transport stipulations in line with IATA-DGR, IMDG-Code, ADR or RID.
Using headphones
Take the following measures to prevent hearing damage. Before using head­phones, check the volume and reduce it if necessary. If you monitor varying sig­nal levels, take off the headphones and wait until the signal has settled. Then adjust the volume.
10Getting Started 1179.4657.02 ─ 15
R&S®FPL1000
Safety and regulatory information
Labels on R&S FPL1000
Cleaning the product
Use a dry, lint-free cloth to clean the product. When cleaning, keep in mind that the casing is not waterproof. Do not use liquid cleaning agents.
Meaning of safety labels
Safety labels on the product warn against potential hazards.
Potential hazard Read the product documentation to avoid personal injury or product damage.
Electrical hazard Indicates live parts. Risk of electric shock, fire, personal injury or even death.
Hot surface Do not touch. Risk of skin burns. Risk of fire.
Protective conductor terminal Connect this terminal to a grounded external conductor or to protective ground. This
connection protects you against electric shock if an electric problem occurs.

1.2 Labels on R&S FPL1000

Labels on the casing inform about:
Personal safety, see "Meaning of safety labels" on page 11
Product and environment safety, see Table 1-1
Identification of the product, see Chapter 5.2.14, "Device ID", on page 49
Table 1-1: Labels regarding R&S FPL1000 and environment safety
Labeling in line with EN 50419 for disposal of electrical and electronic equipment after the product has come to the end of its service life.
For more information, see the product user manual, chapter "Disposal".
Labeling in line with directive 2006/66/EC for disposal of batteries after they have come to the end of their service life.
For more information, see the product user manual, chapter "Disposal".
11Getting Started 1179.4657.02 ─ 15
R&S®FPL1000
Safety and regulatory information
Korea certification class A

1.3 Warning messages in the documentation

A warning message points out a risk or danger that you need to be aware of. The signal word indicates the severity of the safety hazard and how likely it will occur if you do not follow the safety precautions.
CAUTION
Potentially hazardous situation. Could result in minor or moderate injury if not avoided.
NOTICE
Potential risks of damage. Could result in damage to the supported product or to other property.

1.4 Korea certification class A

이 기기는 업무용(A급) 전자파 적합기기로서 판매자 또는 사용자는 이 점을 주의하 시기 바라며, 가정외의 지역에서 사용하는 것을 목적으로 합니다.
12Getting Started 1179.4657.02 ─ 15
R&S®FPL1000
Documentation overview
User manuals and help

2 Documentation overview

This section provides an overview of the R&S FPL1000 user documentation. Unless specified otherwise, you find the documents at:
www.rohde-schwarz.com/manual/FPL1000

2.1 Getting started manual

Introduces the R&S FPL1000 and describes how to set up and start working with the product. Includes basic operations, typical measurement examples, and gen­eral information, e.g. safety instructions, etc.
A printed version is delivered with the instrument. A PDF version is available for download on the Internet.

2.2 User manuals and help

Separate user manuals are provided for the base unit and the firmware applica­tions:
Base unit manual Contains the description of all instrument modes and functions. It also pro­vides an introduction to remote control, a complete description of the remote control commands with programming examples, and information on mainte­nance, instrument interfaces and error messages. Includes the contents of the getting started manual.
Firmware application manual Contains the description of the specific functions of a firmware application, including remote control commands. Basic information on operating the R&S FPL1000 is not included.
The contents of the user manuals are available as help in the R&S FPL1000. The help offers quick, context-sensitive access to the complete information for the base unit and the firmware applications.
All user manuals are also available for download or for immediate display on the Internet.
13Getting Started 1179.4657.02 ─ 15
R&S®FPL1000
Documentation overview
Data sheets and brochures

2.3 Service manual

Describes the performance test for checking compliance with rated specifications, firmware update, troubleshooting, adjustments, installing options and mainte­nance.
The service manual is available for registered users on the global Rohde & Schwarz information system (GLORIS):
https://gloris.rohde-schwarz.com

2.4 Instrument security procedures

Deals with security issues when working with the R&S FPL1000 in secure areas. It is available for download on the Internet.

2.5 Printed safety instructions

Provides safety information in many languages. The printed document is deliv­ered with the product.

2.6 Data sheets and brochures

The data sheet contains the technical specifications of the R&S FPL1000. It also lists the firmware applications and their order numbers, and optional accessories.
The brochure provides an overview of the instrument and deals with the specific characteristics.
See www.rohde-schwarz.com/brochure-datasheet/FPL1000
14Getting Started 1179.4657.02 ─ 15
R&S®FPL1000
Documentation overview
Calibration certificate
2.7 Release notes and open source acknowledg­ment (OSA)
The release notes list new features, improvements and known issues of the cur­rent firmware version, and describe the firmware installation.
The software makes use of several valuable open source software packages. An open-source acknowledgment document provides verbatim license texts of the used open source software.
See www.rohde-schwarz.com/firmware/FPL1000

2.8 Application notes, application cards, white papers, etc.

These documents deal with special applications or background information on particular topics.
See www.rohde-schwarz.com/application/FPL1000

2.9 Calibration certificate

The document is available on https://gloris.rohde-schwarz.com/calcert. You need the device ID of your instrument, which you can find on a label on the rear panel.
15Getting Started 1179.4657.02 ─ 15
R&S®FPL1000
Key features

3 Key features

The R&S FPL1000 sets standards in RF performance and usability. Outstanding key features are:
One instrument for multiple applications
Spectrum analysis
Signal analysis of analog and digitally modulated signals
Power measurements with power sensors
Noise figure and gain measurements
Phase noise measurements
Solid RF performance
Low spurious response
Low displayed average noise level (DANL)
40 MHz signal analysis bandwidth
Low level measurement uncertainty
Precise spectral measurements due to low phase noise
Intuitive user interface
High-resolution display
Multipoint touchscreen
Flexible arrangement of results and MultiView
Toolbar
Quiet operation
Fully portable
Battery pack and 12 V/24 V power supply
Carrying bag and shoulder harness
Low power consumption
16Getting Started 1179.4657.02 ─ 15
R&S®FPL1000
Unpacking and checking
Preparing for use

4 Preparing for use

Here, you can find basic information about setting up the product for the first time.
Lifting and carrying..........................................................................................17
Unpacking and checking................................................................................. 17
Choosing the operating site............................................................................ 18
Setting up the R&S FPL1000.......................................................................... 18
Connecting to power....................................................................................... 21
Switching on or off...........................................................................................24
Connecting to LAN.......................................................................................... 25
Connecting a keyboard................................................................................... 27
Connecting an external monitor...................................................................... 27
Windows operating system............................................................................. 29
Logging on...................................................................................................... 31
Checking the supplied options........................................................................ 33
Performing a self-alignment............................................................................ 34
Considerations for test setup.......................................................................... 34

4.1 Lifting and carrying

The carrying handles are designed to lift or carry the instrument. Do not apply excessive external force to the handles.
See "Lifting and carrying the product" on page 8.

4.2 Unpacking and checking

1. Unpack the R&S FPL1000 carefully.
2. Retain the original packing material. Use it when transporting or shipping the
R&S FPL1000 later.
3. Using the delivery notes, check the equipment for completeness.
17Getting Started 1179.4657.02 ─ 15
R&S®FPL1000
Setting up the R&S FPL1000
4. Check the equipment for damage.
If the delivery is incomplete or equipment is damaged, contact Rohde & Schwarz.
Preparing for use

4.3 Choosing the operating site

Specific operating conditions ensure proper operation and avoid damage to the product and connected devices. For information on environmental conditions such as ambient temperature and humidity, see the data sheet.
See also "Choosing the operating site" on page 8.
Electromagnetic compatibility classes
The electromagnetic compatibility (EMC) class indicates where you can operate the product. The EMC class of the product is given in the data sheet.
Class B equipment is suitable for use in:
Residential environments
Environments that are directly connected to a low-voltage supply network
that supplies residential buildings
Class A equipment is intended for use in industrial environments. It can cause radio disturbances in residential environments due to possible conducted and radiated disturbances. It is therefore not suitable for class B environments. If class A equipment causes radio disturbances, take appropriate measures to eliminate them.

4.4 Setting up the R&S FPL1000

The R&S FPL1000 is designed for use either on a bench top or in a rack, or as a portable instrument (with optional battery operation) in a transport bag in the field.
See also:
"Setting up the product" on page 8
"Intended use" on page 7
18Getting Started 1179.4657.02 ─ 15
R&S®FPL1000
Setting up the R&S FPL1000
Preparing for use

4.4.1 Placing the R&S FPL1000 on a bench top

To place the product on a bench top
1. Place the product on a stable, flat and level surface.
2. CAUTION! The top surface of the product is too small for stacking. If you
stack another product on top of the product, the stack can fall over and cause injury.
If you want to save space, mount several products in a rack.
3. CAUTION! Foldable feet can collapse. See "Setting up the product"
on page 8.
Always fold the feet completely in or out. With folded-out feet, do not place anything on top or underneath.
4. NOTICE! Overheating can damage the product.
Prevent overheating as follows:
Keep a minimum distance of 10 cm between the fan openings of the prod-
uct and any object in the vicinity.
Do not place the product next to heat-generating equipment such as radia­tors or other products.

4.4.2 Mounting the R&S FPL1000 in a rack

To prepare the rack
1. Observe the requirements and instructions in "Setting up the product" on page 8.
2. NOTICE! Insufficient airflow can cause overheating and damage the product.
Design and implement an efficient ventilation concept for the rack.
19Getting Started 1179.4657.02 ─ 15
R&S®FPL1000
Setting up the R&S FPL1000
To mount the R&S FPL1000 in a rack
1. Use an adapter kit to prepare the R&S FPL1000 for rack mounting.
a) Order the rack adapter kit designed for the R&S FPL1000. For the order
number, see the data sheet.
b) Mount the adapter kit. Follow the assembly instructions provided with the
adapter kit.
2. Lift the R&S FPL1000 to shelf height.
3. Grab the handles and push the R&S FPL1000 onto the shelf until the rack brackets fit closely to the rack.
4. Tighten all screws in the rack brackets with a tightening torque of 1.2 Nm to secure the R&S FPL1000 in the rack.
To unmount the R&S FPL1000 from a rack
Preparing for use
1. Loosen the screws at the rack brackets.
2. Remove the R&S FPL1000 from the rack.
3. If placing the R&S FPL1000 on a bench top again, unmount the adapter kit from the R&S FPL1000. Follow the instructions provided with the adapter kit.

4.4.3 Portable operation

An optional carrying bag designed specifically for the R&S FPL1000 allows you to protect the instrument while working in the field. The bag includes ventilation areas at the position of the ventilation outlets in the casing to ensure air circula­tion. The transparent cover allows you to operate the instrument without removing it from the bag. With the help of the optional vest holster, you can carry the R&S FPL1000 in its bag and keep your hands free. Together with the optional battery pack (see Chapter 4.5.3, "Optional battery pack (R&S FPL1-B31)", on page 23), and packed in the dedicated carrying bag, the R&S FPL1000 is ideally suited for operation directly in the field, even in rough environments.
20Getting Started 1179.4657.02 ─ 15
R&S®FPL1000
Preparing for use
Connecting to power
► Inspect the carrying bag for wear and tear before placing the instrument in it.
For details on optional accessories, see the R&S FPL1000 data sheet.

4.5 Connecting to power

There are various options to supply power to the R&S FPL1000.
The R&S FPL1000 is equipped with an AC power supply connector.
The R&S FPL1000 can also be equipped with an optional (internal) DC power supply connector (R&S FPL1-B30).
The R&S FPL1000 allows for battery operation if option R&S FPL1-B31 is installed.

4.5.1 Connecting the AC power

The R&S FPL1000 can be used with different AC power voltages and adapts itself automatically to it. Refer to the datasheet for the requirements of voltage and frequency.
For safety information, see "Connecting to power" on page 9.
21Getting Started 1179.4657.02 ─ 15
R&S®FPL1000
To connect the AC power
1. Plug the AC power cable into the AC power connector on the rear panel of the instrument. Only use the AC power cable delivered with the R&S FPL1000.
2. Plug the AC power cable into a power outlet with ground contact.
The required ratings are listed next to the AC power connector and in the data sheet.
For details on the connector, refer to Chapter 5.2.1, "AC power supply connection
and main power switch", on page 44.
Preparing for use
Connecting to power

4.5.2 Connecting an optional DC power supply (R&S FPL1-B30)

The R&S FPL1000 can also be equipped with an optional DC power supply con­nector (R&S FPL1-B30). If installed, the R&S FPL1000 can be operated by a DC voltage of +12 V to +24 V. For details on the connector see Chapter 5.2.2, "Li-Ion
battery packs and DC power connector", on page 45.
If you use an external power supply unit to supply safety extra-low DC voltage (SELV) to the instrument, be sure to meet the requirements for reinforced/double insulation in accordance with DIN/EN/IEC 61010 (UL 3111, CSA C22.2 No.
1010.1) or DIN/EN/IEC 60950 (UL 1950, CSA C22.2 No. 950). Provide current
limitation in accordance with DIN EN 61010-1 Appendix F2.1. Use a cable no lon­ger than 3 m.
Also see "Connecting to power" on page 9.
22Getting Started 1179.4657.02 ─ 15
R&S®FPL1000
DC connection
► Connect the DC power connector on the rear panel of the R&S FPL1000 to
the DC power source using a cable as described above.
Preparing for use
Connecting to power

4.5.3 Optional battery pack (R&S FPL1-B31)

As an alternative to the fixed AC or DC power supply, the R&S FPL1000 also allows for battery operation. The "Battery Pack" option R&S FPL1-B31 comprises two Li-ion batteries and an internal charger. The internal charger charges the bat­teries whenever the instrument is connected to AC power. During operation, if neither DC nor AC power is supplied, the R&S FPL1000 automatically switches to battery operation.
The battery pack can be retrofitted by Rohde & Schwarz service.
For safety information, see "Handling batteries safely" on page 9.
Charging batteries
Charge the batteries before using battery operation for the first time. Following a long storage period, it can be necessary to charge and discharge the batteries several times to reach full capacity.
For batteries from third parties, follow the instructions provided by the manufac­turer. For batteries manufactured by Rohde & Schwarz, observe the following:
If inserted in the R&S FPL1000, batteries are charged via the common AC or DC power supply.
You can also use the external battery charger R&S FSV-B34 to charge up to 4 batteries.
Charge in a temperature range from +0 °C to +45 °C. If the temperature is above or below these values, or the temparature varies strongly, charging is interrupted. If the battery temperature rises above +53 °C, charging is stop­ped.
Try not to overcharge the battery too often because overcharging reduces the service life of the battery.
23Getting Started 1179.4657.02 ─ 15
R&S®FPL1000
Preparing for use
Switching on or off
When the battery is being charged in standby mode, the [Power] LED blinks. During operation, the status bar indicates that the battery is
being charged.
Spare battery pack (R&S FPL1-Z4)
In addition to the internal battery pack (option R&S FPL1-B31), spare batteries are available for the R&S FPL1000. The spare battery pack R&S FPL1-Z4 com­prises two additional Li-ion batteries.
Outside the R&S FPL1000, batteries can be charged using the external battery charger R&S FSV-B34. Even during battery operation, you can exchange the internal batteries while the R&S FPL1000 is running, as long as one battery remains in the instrument. However, it is not recommended to operate the R&S FPL1000 with only one battery for a longer period.

4.6 Switching on or off

Table 4-1: Overview of power states
Status LED on Power key Position of main power switch
Off
Standby
Ready
gray
orange
green
To switch on the R&S FPL1000
The R&S FPL1000 is off but connected to power.
1. Set the switch on the power supply to position [I]. See Chapter 5.2.1, "AC power supply connection and main power switch", on page 44.
The LED of the Power key is orange.
See Chapter 5.1.2, "Power key", on page 38.
[0]
[I]
[I]
Note: If the R&S FPL1000 was disconnected from power before it was switched off regularly, it automatically switches back on when the power is reconnected. Using a LAN-controlled power switch, you can take advantage of this behavior to switch the device on and off remotely.
24Getting Started 1179.4657.02 ─ 15
R&S®FPL1000
2. Press the Power key.
The LED changes to green. The instrument operates on battery, DC, or AC power, whichever is supplied. The R&S FPL1000 boots.
After booting, the instrument is ready for operation.
To shut down the product
The product is in the ready state.
► Press the [Power] key.
The operating system shuts down. The LED changes to orange.
To disconnect from power
The R&S FPL1000 is in the standby state.
Preparing for use
Connecting to LAN
1. NOTICE! Risk of data loss. If you disconnect the product from power when it is in the ready state, you can lose settings and data. Shut it down first.
Set the switch on the power supply to position [0]. See Chapter 5.2.1, "AC power supply connection and main power switch", on page 44.
The LED of the standby key is switched off.
2. Disconnect the R&S FPL1000 from the power source.

4.7 Connecting to LAN

You can connect the instrument to a LAN for remote operation via a PC.
For details on the connector, see Chapter 5.2.12, "LAN", on page 49.
Provided the network administrator has assigned you the appropriate rights and adapted the Windows firewall configuration, you can use the interface, for exam­ple:
To transfer data between a controlling device and the test device, e.g. to run a remote control program
To access or control the measurement from a remote computer using the "Remote Desktop" application (or a similar tool)
To connect external network devices (e.g. printers)
25Getting Started 1179.4657.02 ─ 15
R&S®FPL1000
To transfer data from a remote computer and back, e.g. using network folders
Network environment
Before connecting the product to a local area network (LAN), consider the follow­ing:
Install the latest firmware to reduce security risks.
For internet or remote access, use secured connections, if applicable.
Ensure that the network settings comply with the security policies of your com­pany. Contact your local system administrator or IT department before con­necting your product to your company LAN.
When connected to the LAN, the product may potentially be accessed from the internet, which may be a security risk. For example, attackers might mis­use or damage the product. For more information about IT security and how to operate the product in a secure LAN environment, see the Rohde & Schwarz white paper 1EF96: Malware Protection Windows 10.
Preparing for use
Connecting to LAN
NOTICE! Risk of network failure.
Consult your network administrator before performing the following tasks:
Connecting the instrument to the network
Configuring the network
Changing IP addresses
Exchanging hardware
Errors can affect the entire network.
Connect the R&S FPL1000 to the LAN via the LAN interface on the rear panel of the instrument.
Windows automatically detects the network connection and activates the required drivers.
By default, the R&S FPL1000 is configured to use DHCP and no static IP address is configured.
The default instrument name is <Type><variant>-<serial_number>, for example, FPL1003-123456. For information on determining the serial num­ber, see Chapter 5.2.14, "Device ID", on page 49.
For more information on LAN configuration, see the R&S FPL1000 user manual.
26Getting Started 1179.4657.02 ─ 15
R&S®FPL1000
Connecting an external monitor
Preparing for use

4.8 Connecting a keyboard

The keyboard is detected automatically when it is connected. The default input language is English – US.
However, you can also connect foreign language keyboards; currently the follow­ing languages are supported for the R&S FPL1000:
German
Swiss
French
Russian
To configure the keyboard language
1. To access the Windows operating system, press the Windows key on the external keyboard.
2. Select "Start > Settings > Time & language > Region & language > Add a lan­guage" .

4.9 Connecting an external monitor

You can connect an external monitor (or projector) to the "DVI" connector on the rear panel of the R&S FPL1000 (see also Chapter 5.2.13, "DVI", on page 49).
27Getting Started 1179.4657.02 ─ 15
R&S®FPL1000
Connecting an external monitor
Screen resolution and format
The touchscreen of the R&S FPL1000 is calibrated for a 16:10 format. If you connect a monitor or projector using a different format (e.g. 4:3), the calibra­tion is not correct and the screen does not react to your touch actions prop­erly.
The touchscreen has a screen resolution of 1280x800 pixels. Usually, the display of the external monitor is a duplicate of the instrument's monitor.
If you configure the external monitor to be used as the only display in the Windows configuration dialog box ("Show only on 2"), the maximum screen resolution of the monitor is used. In this case, you can maximize the R&S FPL1000 application window and see even more details. You cannot change the monitor's screen resolution via the standard Windows configura­tion dialog box.
The R&S FPL1000 supports a minimum resolution of 1280x768 pixels.
Preparing for use
1. Connect the external monitor to the R&S FPL1000.
2. Press the [Setup] key.
3. Press the "Display" softkey.
4. Select the "Configure Monitor" tab in the "Display" dialog box.
The standard Windows "Screen Resolution" dialog box is displayed.
28Getting Started 1179.4657.02 ─ 15
R&S®FPL1000
Preparing for use
Windows operating system
5. If necessary, change the screen resolution. Consider the information in the note above.
6. Select the instrument for display:
"Display 1" : internal monitor only
"Display 2" : external monitor only
"Duplicate" : both internal and external monitor
7. Tap "Apply" to try out the settings before they are accepted permanently, then you can easily return to the previous settings, if necessary.
8. Select "OK" if the settings are suitable.

4.10 Windows operating system

The instrument contains the Windows operating system which has been config­ured according to the instrument's features and needs. Changes in the system setup are only required when peripherals like a keyboard or a printer are installed or if the network configuration does not comply with the default settings. After the R&S FPL1000 is started, the operating system boots and the instrument firmware is started automatically.
29Getting Started 1179.4657.02 ─ 15
R&S®FPL1000
Windows operating system
Tested software
The drivers and programs used on the instrument under Windows are adapted to the instrument. Only install update software released by Rohde & Schwarz to modify existing instrument software.
You can install additional software on the instrument; however, additional soft­ware can impair instrument function. Thus, run only programs that Rohde & Schwarz has tested for compatibility with the instrument software.
The following program packages have been tested:
Symantec Endpoint Security – virus-protection software
FileShredder - for reliable deletion of files on the hard disk
Service packs and updates
Microsoft regularly creates security updates and other patches to protect Win­dows-based operating systems. They are released through the Microsoft Update website and associated update server. Update instruments using Windows regu­larly, especially instruments that connect to a network.
Preparing for use
Firewall settings
A firewall protects an instrument by preventing unauthorized users from gaining access to it through a network. Rohde & Schwarz highly recommends using the firewall on your instrument. Rohde & Schwarz instruments are shipped with the Windows firewall enabled. All ports and connections for remote control are enabled.
Note that changing firewall settings requires administrator rights.
Virus protection
Take appropriate steps to protect your instruments from infection. Use strong fire­wall settings and scan any removable storage device used with a Rohde & Schwarz instrument regularly. It is also recommended that you install anti-virus software on the instrument. Rohde & Schwarz does NOT recommend running anti-virus software in the background ("on-access" mode) on Windows­based instruments, due to potentially degrading instrument performance. How­ever, Rohde & Schwarz does recommend running it during non-critical hours.
For details and recommendations, see the following Rohde & Schwarz white paper:
1EF96: Malware Protection Windows 10
30Getting Started 1179.4657.02 ─ 15