The product documentation helps you use the R&S QAR safely and efficiently. Follow
the instructions provided here and in the safety instructions.
Intended use
The product is intended for the development, production and verification of electronic
components and devices in industrial, administrative, and laboratory environments.
Use the product only for its designated purpose. Observe the operating conditions and
performance limits stated in the data sheet.
Where do I find safety information?
Safety information is part of the product documentation. It warns you of potential dangers and gives instructions on how to prevent personal injury or damage caused by
dangerous situations. Safety information is provided as follows:
●
In the safety instructions. The same information is provided in many languages as
printed "Safety Instructions". The printed "Safety Instructions" are delivered with
the R&S QAR.
●
Throughout the documentation, safety instructions are provided when you need to
take care during setup or operation.
1.1Safety instructions
Products from the Rohde & Schwarz group of companies are manufactured according
to the highest technical standards. To use the products safely, follow the instructions
provided here and in the product documentation. Keep the product documentation
nearby and offer it to other users.
Use the product only for its intended use and within its performance limits. Intended
use and limits are described in the product documentation such as the data sheet,
manuals and the printed "Safety Instructions". If you are unsure about the appropriate
use, contact Rohde & Schwarz customer service.
Using the product requires specialists or specially trained personnel. These users also
need sound knowledge of at least one of the languages in which the user interfaces
and the product documentation are available.
Never open the casing of the product. Only service personnel authorized by
Rohde & Schwarz are allowed to repair the product. If any part of the product is damaged or broken, stop using the product. Contact Rohde & Schwarz customer service at
http://www.customersupport.rohde-schwarz.com.
Lifting, carrying and installing the product
The product is heavy and can cause physical injuries when it falls or tips over.
7User Manual 1178.7200.02 ─ 12
Page 8
R&S®QAR
Safety and regulatory information
Safety instructions
The R&S QAR must be installed and set up by trained staff. Make sure to install the
R&S QAR exactly as described in the user manual. During the installation, make sure
that the installation site is only accessible by people working on the installation.
When you carry, lift or move the R&S QAR, use the carrying handles on the back side
of the R&S QAR. Hold the R&S QAR into position until all screws are tight to prevent it
from falling over.
Because of its weight, at least two strong people are required to move or install the
R&S QAR.
If you have to move the pallet the R&S QAR is delivered on, always use a forklift.
During transport, installation or repair, wear appropriate protective clothing that complies with your local rules and regulations. If you are unsure of which equipment to use,
ask your safety inspector.
We recommend to wear:
●
Safety shoes with toe cap and ESD protection.
●
Protective gloves
ESD protected shoes and gloves also protect the hardware from being damaged.
Installing cables
A R&S QAR installation consisting of the R&S QAR, monitor and transmitter module
requires cable connections (power cables, LAN cables etc.). Make sure to secure all
cables on the floor and cover them up.
Loose cables are a tripping hazard. Tripping on cables can cause physical injuries.
Properly covered cables also minimize the risk of cable damage.
Choosing the operating site
Only use the product indoors. The product casing is not waterproof. Water that enters
can electrically connect the casing with live parts, which can lead to electric shock,
serious personal injury or death if you touch the casing. If Rohde & Schwarz provides
accessories designed for your product, e.g. a carrying bag, you can use the product
outdoors.
Unless otherwise specified, you can operate the product up to an altitude of 2000 m
above sea level. The product is suitable for pollution degree 2 environments where
nonconductive contamination can occur. For more information on environmental conditions such as ambient temperature and humidity, see the data sheet.
Supplying the R&S QAR with power
The product is an overvoltage category II product. Connect the product to a fixed
installation used to supply energy-consuming equipment such as household appliances and similar loads. Keep in mind that electrically powered products have risks, such
as electric shock, fire, personal injury or even death.
8User Manual 1178.7200.02 ─ 12
Page 9
R&S®QAR
Safety and regulatory information
Labels on the product
Take the following measures for your safety:
●
Before switching on the product, ensure that the voltage and frequency indicated
on the product match the available power source. Make sure that the mains supply
voltage fluctuations do not exceed ±10 % of the nominal supply voltage. If the
power adapter does not adjust automatically, set the correct value and check the
rating of the fuse.
●
Only use the power cable delivered with the R&S QAR. It complies with countryspecific safety requirements. Only insert the plug into an approved grounded outlet.
●
For components that need an external AC adapter, only use the one that is delivered with the components.
●
Only use intact cables and route them carefully so that they cannot be damaged.
Check the power cables regularly to ensure that they are undamaged.
●
If the voltage or frequency is higher than 230 V or 50 Hz, the leakage current of the
R&S QAR can be above the limits. In that case, a protective ground connection
must be established for each panel before connecting them to the mains supply.
●
Only connect the R&S QAR to supply networks with a fuse protection of maximum
20 A.
●
Ensure that you can disconnect the product from the mains supply at any time. Pull
the power plug to disconnect the product. The power plug must be easily accessible.
Meaning of safety labels
Safety labels on the product warn against potential risks.
Potential hazard
Read the product documentation to avoid personal injury or product damage.
Electrical hazard
Indicates live parts. Risk of electric shock, fire, personal injury or even death.
Protective conductor terminal
Connect this terminal to a grounded external conductor or to protective ground. This protects
you against electric shock should an electric problem occur.
1.2Labels on the product
Labels on the casing inform about:
●
Personal safety, see "Meaning of safety labels"on page 9
●
Product and environment safety, see Table 1-1
●
Identification of the product, see Chapter 3.7, "Device ID", on page 18.
9User Manual 1178.7200.02 ─ 12
Page 10
R&S®QAR
Safety and regulatory information
Regulatory information
Table 1-1: Labels regarding product and environment safety
Labeling in line with EN 50419 for disposal of electrical and electronic equipment after the product has come to the end of its service life. For more information, see "Disposing electrical and
electronic equipment"on page 123.
Take care when handling electrostatic sensitive devices.
1.3Warning messages in the documentation
A warning message points out a risk or danger that you need to be aware of. The signal word indicates the severity of the safety hazard and how likely it will occur if you do
not follow the safety precautions.
WARNING
Potentially hazardous situation. Could result in death or serious injury if not avoided.
CAUTION
Potentially hazardous situation. Could result in minor or moderate injury if not avoided.
NOTICE
Potential risks of damage. Could result in damage to the supported product or to other
property.
1.4Korea certification class B
이 기기는 가정용(B급) 전자파 적합기기로서 주로 가정에서 사용하는 것을 목적으로 하
며, 모든 지역에서 사용할 수 있습니다.
1.5Regulatory information
The R&S QAR complies with the following regulations.
10User Manual 1178.7200.02 ─ 12
Page 11
R&S®QAR
1.5.1Regulations for the USA
FCC compliance
This equipment has been tested and found to comply with the limits for Industrial, Scientific and Medical Equipment (ISM), pursuant to part 18 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in an
industrial installation. This equipment generates, uses, and can radiate radio frequency
energy, and if not installed and used in accordance with the instructions, may cause
harmful interference to radio communications. However, there is no guarantee that
interference will not occur in a particular installation. If this equipment does cause
harmful interference other devices, which can be determined by turning the equipment
off and on, the user is encouraged to try to correct the interference by one or more of
the following measures:
●
Reorient or relocate the receiving antenna.
●
Increase the separation between the equipment and receiver.
●
Consult Rohde&Schwarz customer service for help.
Safety and regulatory information
Regulatory information
1.5.2Regulations for Canada
ICES-001 compliance
This class A ISM device complies with Canadian ICES-001.
Cet appareil ISM de la classe A est conforme à la norme NMB-001 du Canada.
11User Manual 1178.7200.02 ─ 12
Page 12
R&S®QAR
Documentation overview
2Documentation overview
This section provides an overview of the R&S QAR user documentation.
User manual
Contains a description of the R&S QAR system, including the features available in the
R&S QAR application.
A digital version of the user manual is available in the measurement application.
A printed version of the user manual is part of the delivery.
Safety information
Contains safety instructions, operating conditions and further important information.
The printed safety instructions are part of the delivery.
Data sheet and brochure
The data sheet contains the technical specifications of the R&S QAR, its options and
accessories.
The brochure provides an overview of the R&S QAR and its characteristics.
Open-source acknowledgment (OSA)
The open-source acknowledgment documents provide verbatim license texts of the
used open source software.
The OSAs is available in the measurement application ("Info" > "Open Source
Acknowledgement") and on the SSD in the folder C:\OpenSourceAgreement.
12User Manual 1178.7200.02 ─ 12
Page 13
R&S®QAR
R&S QAR tour
3R&S QAR tour
The R&S QAR is a millimeter wave imaging system operating in the frequency band of
automotive radar sensors. With its support of spatially resolved reflection and transmission loss measurements, the R&S QAR provides an intuitive and powerful way to evaluate how radomes and bumpers influence the radar performance.
The R&S QAR consists of a panel that contains the measuring electronics like transmitting and receiving antennas. It also contains a PC with dedicated software to operate and control the measurements.
●
Radome measurements
The basic configuration allows you to measure the reflection of a radome.
The high resolution of the resulting images can identify even the smallest disturbance in the radome. The R&S QAR uses the frequency range between 74 GHz
and 79 GHz to measure the reflection. Therefore, any faults visible in the millimeter
wave image directly correlate to the performance of the radome in combination
with a radar sensor.
An optional transmitter module (R&S QAR-Z10) extends the functionality of the
R&S QAR to measure the transmission loss of a radome. For the transmission, the
R&S QAR uses the frequency range between 72 GHz and 82 GHz.
Additional equipment like an optional platform (R&S QAR-Z20) on which you can
mount the R&S QAR, or a table for the device under test (DUT mounting table,
R&S QAR-Z50) are also available.
The default setup uses all 12 measurement clusters built in the R&S QAR to analyze radomes. With an additional software option (R&S QAR-K50), you can unlock
single cluster measurements that only use one of the built-in clusters.
●
Bumper measurements
An optional reference reflector (R&S QAR-Z60) is available to measure the mean
transmission loss (attenuation) caused by a bumper.
The reflector measures the attenuation in 102 points and therefore enables you to
evaluate the homogeneity of the bumper painting and its overall radar penetrability.
●
Radar positioning measurements
The basic configuration enables you to measure the position of the radar in the car
and compare it with the theoretical position indicated in the manufacturer's CAD
files.
You can set acceptance limits and verify in a few seconds if the radar is correctly
positioned.
Additional equipment like an optional platform (R&S QAR-Z21) on which you can
mount the R&S QAR, or a verification set (R&S QAR-Z42) are also available.
Parts of delivery
●
1 x R&SQAR
●
1 x power cable
Required equipment to run measurements
●
Monitor (with DVI cable)
We recommend a monitor with a 1920 x 1080 resolution.
13User Manual 1178.7200.02 ─ 12
Page 14
R&S®QAR
●
USB Keyboard
●
USB Mouse
Connectors
The R&S QAR has several connectors on its back side.
R&S QAR tour
Main power supply
1
2
3
8
Figure 3-1: Connectors
1 = Power button
2 = USB ports (4x)
3 = Display connector (DVI)
4 = LAN interface
5 = Power supply for a monitor (including 2 fuses)
6 = Transmitter module control
7 = Ground terminal
8 = AC inlet and main power switch
7
4
5
6
3.1Main power supply
The AC inlet and main power switch are located on the back of the R&S QAR.
The power cable to connect the panel to the power supply is part of the delivery.
14User Manual 1178.7200.02 ─ 12
Page 15
R&S®QAR
120V
120V
208V
L1
L2
PE
120V
120V
240V
L1
L2
PE
j
j
j
j
= 120°
Single phase, 2 wire
R&S QAR tour
Main power supply
The main power switch contains an RCD (residual-current device) that protects the
internal power circuit of the R&S QAR. If the RCD detects electrical anomalies, it will
switch off automatically the R&S QAR.
For more information about supplying the panel with power, see Chapter 4.4, "Switch-
ing the R&S QAR on and off", on page 34.
Establishing a ground connection
WARNING! Risk of electric shock. If the voltage or frequency are greater than
1.
230 V or 50 Hz, the leakage current of the R&S QAR can be above the limits and
can cause personal injuries from an electric shock.
In that case, a protective ground connection must be established before connecting
the R&S QAR to the mains supply.
●200 V to 230 V AC and 50 Hz
No additional protective ground needed.
●230 V to 240 V AC or a frequency greater than 50 Hz
Additional protective ground needed.
Exception: The following two 60 Hz AC distribution systems are tested and
work without an additional ground.
For the separate protective ground, you need a green/yellow ground wire with a
cross section of at least 4 mm2 in addition to the main protective earthing terminal.
The ground wire must be connected permanently to the ground terminal on the
side of each R&S QAR panel and earth.
The ground terminal is located on the back of the scanner. Connect the terminal to
the protective grounding system of the building or room where the scanner is
installed.
Figure 3-2: Ground terminal
2. Make sure that the power supply is adequately protected against overcurrent. Otherwise the fuse can blow.
15User Manual 1178.7200.02 ─ 12
Page 16
R&S®QAR
R&S QAR tour
Transmitter module control
We recommend that you use a power cable that is rated for a 16 A (included in the
delivery) and connect each panel to a separately protected power circuit (at least
10 A).
If you cannot connect the panels to separate power circuits, make sure that the
system is protected:
●By a fuse with at least 20 A in 200 V networks.
●By a fuse with at least 16 A in 230 V networks.
3. If the voltage drops to 70 % for 500 ms, a reboot occurs.
3.2Power button
The power button is located on the back of the R&S QAR. Use it to switch the
R&S QAR on and off.
Figure 3-3: Power button
For more information about switching the R&S
"Switching the R&S QAR on and off", on page 34.
3.3Transmitter module control
The connectors for the optional transmitter module for transmission loss measurements are located on the back of the R&S QAR.
QAR on and off, see Chapter 4.4,
16User Manual 1178.7200.02 ─ 12
Page 17
R&S®QAR
R&S QAR tour
LAN interface
Transmitter control
The "External Transmitter Control" connector controls the transmitter module.
RF out
The "RF Out" connector (SMA) outputs the signal for the transmission measurement.
Do not feed a DC or RF signal into the RF output, because it is specified as an output
only. Otherwise, the R&S QAR can get damaged.
3.4Monitor connectors
A DVI interface and power supply (AC outlet) for a monitor are located on the back
side of the R&S QAR. Use them to connect an external monitor to operate the
R&S QAR.
Figure 3-4: DVI interface
Fuses F1 and F2
Two fuses of the same type (1 A, designation: EN 60127-2/5 T1H250V) are located
above the AC outlet. In case of defects, check both fuses.
Replacing a fuse
1. Disconnect the R&S QAR from the power supply.
2. Open the fuse holder by turning it counterclockwise, for example with a screwdriver.
3. Replace the fuse.
4. Close the fuse holder.
3.5LAN interface
The R&S QAR provides a LAN interface on the back of the R&S QAR. Use it to integrate the R&S QAR into a local network.
17User Manual 1178.7200.02 ─ 12
Page 18
R&S®QAR
Figure 3-5: LAN interface
3.6USB ports
The R&S QAR provides 4 USB ports on the back side. The two top ports are USB 3.0
type A. The two bottom ports are USB 2.0 type A.
They offer multiple uses:
●
Connection of a keyboard to operate the R&SQAR.
●
Connection of a mouse to operate the R&SQAR.
●
Connection of a memory stick (for example to update the software).
R&S QAR tour
Device ID
3.7
Figure 3-6: USB port (one of four)
Device ID
The unique device identifier is provided on a barcode sticker on the foot of the panel.
The device ID is made up of the device order number and a serial number (for example
1336.6008K02-123456-aa).
18User Manual 1178.7200.02 ─ 12
Page 19
R&S®QAR
Preparing for use
4Preparing for use
Before you can use the R&S QAR, install it and connect the devices as described in
the next subchapters.
●Unpacking the R&S QAR........................................................................................19
●Installing the R&S QAR...........................................................................................19
●Connecting the R&S QAR.......................................................................................33
●Switching the R&S QAR on and off.........................................................................34
3. Lift the R&S QAR off the pallet using the handles attached to the R&S QAR.
Shipping the R&S QAR
We recommend using the original packaging if the R&S QAR has to be shipped.
4.2Installing the R&S QAR
Using the R&S QAR requires you to install it firmly to some kind of support fixture in
order to prevent it from toppling over. You can install the R&S QAR in different ways,
depending on the equipment you have bought.
If you would like to use a fixture different to those described here, contact
Rohde & Schwarz.
Required tools
Here is a complete list of tools you need for installation on the platform and the installation of the DUT mounting table and the transmitter module.
●
Allen wrench size 5 (for screws ISO 10642 M8x20)
●
Allen wrench size 6 (for screws ISO 4762 M8x40)
●
Allen wrench size 10 (for screws ISO 4762 M12x55)
●
Torx wrench size TX20
●
Open-end wrench size 19
●
Spirit level
●
Torque wrench
●
Power drill (for floor installation only)
19User Manual 1178.7200.02 ─ 12
Page 20
R&S®QAR
●
Anchor bolts size 12 mm, for example a Fischer FAZ II 12 (for floor installation
only)
●Installing the R&S QAR on the floor........................................................................20
●Installing the R&S QAR on the platform (R&S QAR-Z20).......................................22
●Installing the radome measurement elements........................................................ 24
●Installing the bumper measurement elements........................................................ 26
●Installing the R&S QAR on the platform (R&S QAR-Z21).......................................32
4.2.1Installing the R&S QAR on the floor
You can install the R&S QAR on the floor or on the optional platform. When you do not
use the platform, the R&S QAR must be attached to the floor.
Required tools:
●
Allen wrench size 5 (for screws ISO 10642 M8x20)
●
Allen wrench size 6 (for screws ISO 4762 M8x40)
●
Power drill
●
Anchor bolts size 12 mm, for example a Fischer FAZ II 12.
Preparing for use
Installing the R&S QAR
The anchor bolts and power drill are not part of the delivery.
If you have one of the the optional platforms, see:
●
Chapter 4.2.2, "Installing the R&S QAR on the platform (R&S QAR-Z20)",
on page 22
●
Chapter 4.2.5, "Installing the R&S QAR on the platform (R&S QAR-Z21)",
on page 32
1. Drill 4 holes into the ground.
The locations of the drill holes and distances between each other are indicated in
Figure 4-1.
Figure 4-1: Location and distances of drill holes
20User Manual 1178.7200.02 ─ 12
Page 21
R&S®QAR
Preparing for use
Installing the R&S QAR
A = Drill hole location for R&S QAR
B = Drill hole location for DUT mounting table
Whenever you lift, move or carry the R&S QAR, use the handles on its back side.
2. Screw the two supports to the side of the R&S QAR with two socket head screws
(ISO 10642 M8x20, Allen wrench size 5) on each side of the R&S QAR (see Fig-
ure 4-2).
Figure 4-2: R&S QAR installation, including support installation
3. Screw the two supports to the R&S QAR with the socket head screws (ISO 4762
M8x40, Allen wrench size 6) on each side of the R&S QAR (see Figure 4-2).
21User Manual 1178.7200.02 ─ 12
Page 22
R&S®QAR
Preparing for use
Installing the R&S
QAR
Figure 4-3: Support installation (top view)
4. Lift the R&S QAR into a stable position and in a right angle to the floor.
5. Secure the R&S QAR by screwing the two supports to the floor.
The drill holes in the floor define the position of the supports. Two anchor bolts with
a size of 12 mm secure each support.
Make sure that all screws are screwed on tightly.
4.2.2Installing the R&S QAR on the platform (R&S QAR-Z20)
Required tools:
●
Allen wrench size 5 (for screws ISO 10642 M8x20)
●
Allen wrench size 6 (for screws ISO 4762 M8x40)
●
Allen wrench size 10 (for screws ISO 4762 M12x55)
●
Open-end wrench size 19
●
Spirit level.
Installing the platform
1. Place the platform on a stable and horizontal surface that can support the weight of
the R&S QAR.
2. Check with a spirit level if the platform itself is level.
3. If necessary, remove any irregularities by adjusting the height of the feet until the
platform is level.
●A clockwise movement lowers the corresponding platform area.
22User Manual 1178.7200.02 ─ 12
Page 23
R&S®QAR
Preparing for use
Installing the R&S QAR
●A counterclockwise movement raises the corresponding platform area.
4. Tighten the feet with the check nuts using the open-end wrench.
Figure 4-4: Platform alignment
Mounting the R&S QAR
When the platform is in place, you can mount the R&S QAR on the platform.
Whenever you lift, move or carry the R&S QAR, use the handles on its back side.
1. Screw the two supports to the R&S QAR with two socket head screws (size 5) on
each side of the R&S QAR (see Figure 4-5).
Figure 4-5: R&S
QAR installation, including support mounting
23User Manual 1178.7200.02 ─ 12
Page 24
R&S®QAR
Preparing for use
Installing the R&S
QAR
2. Lift the R&S QAR onto the platform into a stable position and in a right angle to the
floor.
3. Secure the R&S QAR by screwing the two supports to the platform.
The mounting holes in the platform define the position of the supports.
Four screws secure each support:
●Two socket head screws with a size 10 on the outside.
●Two socket head screws with a size 6 on the inside.
Note: Insert a washer between screw and support as shown in Figure 4-5.
4. Make sure that all eight screws are screwed on tightly to the platform.
4.2.3
Figure 4-6: Support installation (top view)
Installing the radome measurement elements
Radome measurements require the DUT mounting table and the transmitter module.
Required tools:
●
Allen wrench size 5.
●
Torx wrench size TX20.
●
Power drill (floor installation only).
●
Anchor bolts size 12 mm, for example a Fischer FAZ II 6 (floor installation only).
Installing the DUT mounting table on the platform
You can install the optional mounting table on the platform. The platform has several
mounting holes that you can use to secure the table. The mounting holes you choose
define the distance of the mounting table to the R&S QAR.
1. Place the mounting table on the platform at the distance you need.
2. Screw the mounting table to the platform with the four socket head screws (size 5).
24User Manual 1178.7200.02 ─ 12
Page 25
R&S®QAR
Preparing for use
Installing the R&S
QAR
Note: Insert a washer between screw and mounting table as shown in Figure 4-7.
Figure 4-7: DUT mounting table and transmitter module installation
Installing the DUT mounting table on the floor
Altenatively, you can install the DUT mounting table directly on the floor.
1. Drill holes into the floor.
The locations and distances of the drill holes are indicated in Figure 4-1.
2. Screw the table to the floor with four anchor bolts, size 6 mm, for example a
Fischer FAZ II 6.
The anchor bolts are not included in the delivery.
Attaching the transmitter module to the mounting table
You can install the transmitter module required for transmission loss measurements on
top of the DUT mounting table. The DUT mounting table already comes with the necessary mounting holes.
1. Place the transmitter module on top of the mounting table.
2. Screw the transmitter module to the mounting table using the torx wrench.
25User Manual 1178.7200.02 ─ 12
Page 26
R&S®QAR
Preparing for use
Figure 4-8: DUT mounting table from above
Installing the R&S
QAR
For more information about cable connections, see Chapter 4.3, "Connecting the
R&S QAR", on page 33.
4.2.4Installing the bumper measurement elements
The basic setup for bumper measurements consists of the R&S QAR and a reference
reflector, including a stand for the reference reflector.
The reference reflector has to be placed at a specific height and at specific distances
from the R&S QAR, as shown in Figure 4-10 and Figure 4-9.
There is a tolerance of +/-2mm (lateral distance) and +/-5.8mm (frontal distance).
Figure 4-9: Setup for bumper measurements: height and frontal distance
Required tools:
●
Allen wrench size 5.
26User Manual 1178.7200.02 ─ 12
Page 27
R&S®QAR
Preparing for use
Installing the R&S
●
Allen wrench size 6.
●
Torx wrench size TX25.
●
Spirit level
Installing the reference reflector stand
1. Put the stand on the platform at the specified location.
2. Screw the stand to the platform with the four socket head screws (size 5).
Insert a washer between screw and stand.
QAR
Figure 4-10: Top view - lateral distance
Installing the reference reflector
1. Attach the reference reflector to the stand.
Verify that the top of the reflector pins are parallel to the R&S QAR front by using,
for example, a spirit level. Check the lateral distance by performing the verification
described in Chapter 5.3.5, "Verification", on page 71.
2. Align the reference reflector with a spirit level.
The top of the reflector pins have to be parallel to the front side of the R&S QAR
panel.
3. Screw the reference reflector to the stand with the Torx wrench.
4. Normalize the measurement to verify the position of the reference reflector.
For more information about normalizing the measurement, see Chapter 5.3.3.1,
"Normalization", on page 68.
5. Check the results of the normalization measurement.
If the reference reflector is not aligned correctly, the R&S QAR shows the type of
misalignment.
27User Manual 1178.7200.02 ─ 12
Page 28
R&S®QAR
Preparing for use
Installing the R&S QAR
Figure 4-11: Possible misalignments of the reflector stand
6. If the normalization indicates a misalignment, adjust the position of the reflectorstand as described below.
Do not change the position of the reference reflector itself.
Note that the alignment can take several tries, because of the low tolerances.
Correcting a lateral misalignment (x-axis)
Solves message in user interface: "Shift reference reflector to the right / left".
28User Manual 1178.7200.02 ─ 12
Page 29
R&S®QAR
Preparing for use
Installing the R&S
1. Remove the covers on the base of the stand (2x), for example with a slotted screwdriver.
2. Loosen the screws on the base of the stand with an Allen wrench, size 6 (8x).
3. Loosen the screw on the base of the support bar with an Allen wrench, size 6 (1x).
4. Move the stand to the left or right as necessary.
QAR
5. Repeat normalization measurement.
a) If normalization does no longer indicate a misalignment, reassemble the stand.
b) If normalization still indicates a misalignment, repeat the adjustment until the
error no longer occurs.
Correcting a horizontal misalignment (z-axis)
Solves message in user interface: "Shift reference reflector to toward QAR / from QAR
away".
1. Remove the cover on the top of the stand (1x), for example with a slotted screwdriver.
2. Loosen the screws on the main bar of the stand with an Allen wrench, size 6 (2x).
3. Loosen the screws on the support bar with an Allen wrench, size 6 (2x).
29User Manual 1178.7200.02 ─ 12
Page 30
R&S®QAR
Preparing for use
Installing the R&S QAR
4. Move the reflector to the front or back as necessary.
5. Repeat normalization measurement.
a) If normalization does no longer indicate a misalignment, reassemble the stand.
b) If normalization still indicates a misalignment, repeat the adjustment until the
error no longer occurs.
Correcting a horizontal misalignment (z-axis)
Solves message in user interface: "Shift reference reflector downward / upward".
1. Loosen the screws on the side of the support bar with an Allen wrench, size 6 (2x).
2. Move the reflector up or down as necessary.
3. Repeat normalization measurement.
a) If normalization does no longer indicate a misalignment, reassemble the stand.
b) If normalization still indicates a misalignment, repeat the adjustment until the
error no longer occurs.
Correcting a horizontal tilt
Solves message in user interface: "Rod alignment or rod distance is not correct".
30User Manual 1178.7200.02 ─ 12
Page 31
R&S®QAR
Preparing for use
Installing the R&S
1. Loosen the screws on the side of the support bar with an Allen wrench, size 6 (2x).
2. Turn the reflector to the right or left as necessary.
3. Repeat normalization measurement.
a) If normalization does no longer indicate a misalignment, reassemble the stand.
b) If normalization still indicates a misalignment, repeat the adjustment until the
error no longer occurs.
QAR
Correcting a vertical tilt
Solves message in user interface: "Rod alignment or rod distance is not correct".
1. Remove the cover on the top of the support bar (1x), for example with a slotted
screwdriver.
2. Loosen the screw with an Allen wrench size 6 (1x).
3. Move the support bar up or down to correct the vertical angle of the stand.
4. Repeat normalization measurement.
a) If normalization does no longer indicate a misalignment, reassemble the stand.
31User Manual 1178.7200.02 ─ 12
Page 32
R&S®QAR
Preparing for use
Installing the R&S
b) If normalization still indicates a misalignment, repeat the adjustment until the
error no longer occurs.
4.2.5Installing the R&S QAR on the platform (R&S QAR-Z21)
Required tools:
●
Allen wrench size 5 (for screws ISO 10642 M8x20)
●
Allen wrench size 6 (for screws ISO 4762 M8x40)
●
Allen wrench size 10 (for screws ISO 4762 M12x55)
●
Open-end wrench size 19
●
Spirit level.
Installing the platform
1. Place the platform on a stable and horizontal surface that can support the weight of
the R&S QAR.
2. Check with a spirit level if the platform itself is level.
QAR
3. If necessary, remove any irregularities by adjusting the height of the feet until the
platform is level.
●A clockwise movement lowers the corresponding platform area.
●A counterclockwise movement raises the corresponding platform area.
4. Tighten the feet with the check nuts using the open-end wrench.
Figure 4-12: Platform alignment
Mounting the R&S QAR
When the platform is in place, you can mount the R&S QAR on the platform.
32User Manual 1178.7200.02 ─ 12
Page 33
R&S®QAR
Preparing for use
Connecting the R&S
Whenever you lift, move or carry the R&S QAR, use the handles on its back side.
1. Screw the two supports to the R&S QAR with four socket head screws (size 5) on
each side of the R&S QAR (see Figure 4-13).
2. Lift the R&S QAR onto the platform into a stable position and in a right angle to the
floor.
The position is defined by the two pins on the platform as shown in Figure 4-13.
Those two pins fit into the two mounting holes on the frame of the R&S QAR.
QAR
Figure 4-13: R&S QAR installation, including support mounting
3. Secure the R&S QAR by screwing the two supports to the platform. Two socket
head screws with a size 10 secure each support.
The mounting holes in the platform define the position of the supports.
Note: Insert a washer between screw and support as shown in Figure 4-13.
4. Tighten all four screws to the platform.
4.3Connecting the R&S QAR
You can connect various devices to the R&S QAR. The following overview shows the
possible connections.
33User Manual 1178.7200.02 ─ 12
Page 34
R&S®QAR
Preparing for use
Switching the R&S
Required tools:
●
Torque wrench (order no. 0041.1218.00).
Monitor (DVI + power)
Keyboard (USB)
Mouse (USB)
R&S QAR
Transmitter module
(Radome measurements)
Figure 4-14: Connection overview in a system with optional transmitter module
1. Connect a monitor to the DVI port on the R&S QAR.
As a power source, you can use a household AC outlet or the integrated power
supply to supply the monitor with power (for monitors with a light-current extension
cable).
LAN
AC Power
QAR on and off
Note: The monitor should have at least a resolution of 1650 x 1050 (we recommend 1920 x 1080). Lower resolutions can lead to an incorrect display of the measurement data.
The BIOS supports monitors with a vertical resolution of up to 1080p.
2. Connect a keyboard and a mouse to two of the USB ports.
3. Optional: Establish a connection to your LAN via the LAN interface.
4. Optional (radome measurements): Connect the transmitter module to the
R&S QAR via the SMA and transmitter control connectors on both devices.
Tighten the SMA connector to 0.6 Nm with the torque wrench.
5. Connect the R&S QAR to the mains power supply with the power cord that is part
of the delivery.
Note: Always connect all devices before connecting the R&S QAR to the mains
supply.
4.4Switching the R&S QAR on and off
Supplying the R&S QAR with power
Always complete all connections (LAN, USB etc.) before connecting the R&S QAR to
the mains supply.
Refer to the datasheet for power rating and electrical safety compliance.
1. Make sure that the protective ground connection is established.
34User Manual 1178.7200.02 ─ 12
Page 35
R&S®QAR
Preparing for use
Operating system
For more information, see Chapter 3.1, "Main power supply", on page 14.
2. Connect the power cable to the AC inlet.
3. Connect the power cable to the mains supply.
4. Turn on the main power switch (position "I").
The R&S QAR is supplied with power.
See Chapter 3.1, "Main power supply", on page 14 for more information about the
AC inlet and the main power switch.
Turning on the R&S QAR
1. Press the power button to boot the system.
See Chapter 3.2, "Power button", on page 16 for more information about the power
button.
Note: When you start the R&S QAR for the first time, you have to enter an option
key. The option key and a information about activating the option key is part of the
delivery.
2. Select the application you want to use from the application launcher.
The application launcher automatically starts after you start the system. Alternatively, you can search and select the application from the Windows start menu.
You can now use the R&S QAR.
Warm-up time
Let the R&S QAR warm up for at least 90 minutes before using it.
Turning off the R&S QAR
► Press the power button.
The system shuts down.
See Chapter 3.2, "Power button", on page 16 for more information about the power
button.
Disconnecting the R&S QAR from the mains supply
1. Turn off the main power switch (position "0").
The R&S QAR is disconnected from the mains supply.
2. Remove the power cable from the mains supply.
4.5Operating system
The R&S QAR contains Windows 10 operating system which has been configured
according to the R&S QAR's features and needs. Changes in the system setup are
only required when you install peripherals like a keyboard or if the network configuration does not comply with the default settings.
35User Manual 1178.7200.02 ─ 12
Page 36
R&S®QAR
The drivers and software installed on the R&S QAR are adapted to the R&S QAR.
Only install software updates released by Rohde & Schwarz.
4.5.1Virus protection
Take appropriate steps to protect your instruments from infection. Use strong firewall
settings and scan any removable storage device used with a Rohde & Schwarz instrument regularly. It is also recommended that you install anti-virus software on the instrument. Rohde & Schwarz does NOT recommend running anti-virus software in the
background ("on-access" mode) on Windows-based instruments, due to potentially
degrading instrument performance. However, Rohde & Schwarz does recommend running it during non-critical hours.
For details and recommendations, see the following Rohde & Schwarz white paper:
●
1EF96: Malware Protection Windows 10
4.5.2Service packs and updates
Preparing for use
Operating system
Microsoft regularly creates security updates and other patches to protect Windowsbased operating systems. These are released through the Microsoft Update website
and associated update server. Instruments using Windows, especially those that connect to a network, should be updated regularly.
For details and recommendations, see the following Rohde & Schwarz white paper:
●
1EF96: Malware Protection Windows 10
4.5.3User accounts
The Windows 10 operating system installed on the R&S QAR has two predefined user
accounts:
●
Standard user account with standard windows rights.
The user name for this account is Operator.
The password is operator<serialnumber>, with <serialnumber> being a
placeholder of the serial number of the R&S QAR.
Example: operator123456
●
Administrator account with rights to install software or create user accounts.
The user name for this account is Admin
The password is emodar8Administrator<serialnumber>, with
<serialnumber> being a placeholder of the serial number of the R&S QAR.
Example: emodar8Administrator123456
Note that for older systems with a Windows 7 operating system, or systems that are
updated from Windows 7, the default password does not contain the
<serialnumber>.
36User Manual 1178.7200.02 ─ 12
Page 37
R&S®QAR
Preparing for use
Operating system
Changing the passwords
For security reasons, we recommend that you to change the passwords.
37User Manual 1178.7200.02 ─ 12
Page 38
R&S®QAR
Operating the R&S QAR
5Operating the R&S QAR
The application to operate the R&S QAR is installed on a PC that is integrated into the
R&S QAR. It starts automatically when you turn on the R&S QAR.
Warm-up time
Let the R&S QAR warm up for at least 90 minutes before using it.
Activating licenses
Before you can use the measurement software on your R&S QAR, you have to activate the corresponding license. You can activate and manage licenses with the
R&S License Server Manager. The R&S License Server Manager is already installed
on your R&S QAR.
●
Select the R&S License Server Manager icon ( ) from the system tray in the taskbar ("Open Manager").
The user interface of the R&S License Server Manager opens in a browser.
●
Select the "Licenses" tab to manage licenses.
●
Select "Activate" to add a new license.
The license key is part of the delivery of the option you bought.
For a detailed description of the features of the R&S License Manager, refer to its integrated help (available via the "?" icon) or the documentation available on Gloris.
Accessing the measurement application
You can access the application by connecting a monitor, a keyboard and a mouse to
the R&S QAR. Alternatively, you can access the application via "Remote Desktop" from
another computer, if the R&S QAR is in a LAN.
If the application does not start automatically, you can start it manually with the shortcut on the Windows desktop.
The operation depends on the application you have:
●
Chapter 5.1, "Radome measurements", on page 39
●
Chapter 5.3, "Bumper measurements", on page 64
●
Chapter 5.4, "Radar positioning measurements", on page 73.
Before you can use the measurement software on your R&S QAR, you have to activate the corresponding license. You can activate and manage licenses with the
R&S License Server Manager. The R&S License Server Manager is already installed
on your R&S QAR.
●
Select the R&S License Server Manager icon (
bar ("Open Manager").
The user interface of the R&S License Server Manager opens in a browser.
●
Select the "Licenses" tab to manage licenses.
●
Select "Activate" to add a new license.
The license key is part of the delivery of the option you bought.
) from the system tray in the task-
38User Manual 1178.7200.02 ─ 12
Page 39
R&S®QAR
Operating the R&S QAR
Radome measurements
For a detailed description of the features of the R&S License Manager, refer to its integrated help (available via the "?" icon) or the documentation available on Gloris.
5.1Radome measurements
For radome measurements, the R&S QAR measures the characteristics of a DUT
using its full 12 clusters. The data evaluated from the data of 12 clusters provides an
image of the DUT with a high resolution. You can see the resulting image in the user
interface of the R&S QAR software.
The typical procedure when measuring a radome (DUT) is as follows:
1. Start the R&S QAR to start the system and the measurement application.
For more information about the functionality of the measurement application, see
Chapter 5.1.2, "Graphical user interface", on page 41.
2. Normalize the test setup.
For more information about the normalization, see Chapter 5.1.3.1, "Normaliza-
tion", on page 43.
3. Measure the DUT.
For more information about the measurement and its results, see Chapter 5.1.3,
"Measurements", on page 43.
4. Save and restore measurement data.
For more information about saving and restoring measurement data, see "Manag-
ing measurement results"on page 50.
5.1.1Test setup
Basic test setup
The basic test setup for radome measurements consists of the R&S QAR and a DUT
that is placed at a certain distance from the R&S QAR.
39User Manual 1178.7200.02 ─ 12
Page 40
R&S®QAR
Operating the R&S QAR
Radome measurements
Figure 5-1: Basic test setup (distances in mm)
In Figure 5-1, the box indicates the area in which the R&S QAR takes the image of the
DUT. Note that the DUT must be positioned slightly off-center to the center of the
R&S QAR as indicated in Figure 5-1 (142.5 mm).
Test setup with platform and DUT mounting table
When you use the optional platform and DUT mounting table, the test setup is already
predefined. The R&S QAR is in a given position, and the DUT, when positioned on the
table, is already in the necessary distance to the R&S QAR and the floor. Note that the
position of the mounting table is variable, depending on the mounting holes you have
used to attach the mounting table. For more information, see Chapter 4.2.3, "Installing
the radome measurement elements", on page 24.
Figure 5-2: R&S
QAR test setup with platform, mounting table and transmitter module
40User Manual 1178.7200.02 ─ 12
Page 41
R&S®QAR
Operating the R&S QAR
Radome measurements
Numbers without brackets = dimensions in mm
Numbers in brackets= dimensions in inch
5.1.2Graphical user interface
The graphical user interface (GUI) contains several elements.
1
Figure 5-3: Software graphical user interface
1 = Menu bar
2 = Measurement description
3 = Result display (reflection)
4 = Result display (transmission loss)
5 = Measure button
6 = Normalize transmission button
7 = Numeric results (transmission loss)
8 = System messages
9 = Status bar
10 = Normalize reflection measurement
11 = Reflection mask editor
12 = Numeric results (reflection)
234
56871011129
Result display
The result display shows the data that was collected during the measurement as a millimeter wave image.
Transmission loss result display
Only available with the optional transmitter module.
41User Manual 1178.7200.02 ─ 12
Page 42
R&S®QAR
Operating the R&S QAR
Radome measurements
The transmission loss result display shows the results of the transmission measurement.
Numeric results
Several numeric results of the reflection and transmission measurements are displayed.
●
Mean reflection
●
Standard deviation
●
Optional: One-way mean attenuation within frequency range 76GHz to 77GHz
●
Optional: One-way mean attenuation within frequency range 76GHz to 81GHz
Settings and control buttons
Several settings and buttons allow you to control the measurement.
●
"Measure"
Initiates a measurement.
For more information, see Chapter 5.1.3.2, "Reflection measurement", on page 45
and Chapter 5.1.3.3, "Transmission measurement", on page 45.
●
"Normalize Reflection"
Initiates the normalization of the reflection measurement.
For more information, see Chapter 5.1.3.1, "Normalization", on page 43.
●
Optional: "Normalize Transmission"
Initiates the normalization of the transmission measurement.
For more information, see Chapter 5.1.3.1, "Normalization", on page 43.
●
System messages
Shows various status messages of the system like a missing normalization or the
remaining warm-up time. The full warm-up time is 90 minutes.
Menu bar
The menu bar contains various settings.
●
"File"
Contains functionality to import and export measurement results, and functionality
to close the software.
●
"Options"
Contains functionality to configure the measurement and result displays.
●
"Info"
Contains various information about the application, including information about
open source software licenses and the user manual.
Keyboard shortcuts
The following keyboard shortcuts are available.
42User Manual 1178.7200.02 ─ 12
Page 43
R&S®QAR
Operating the R&S QAR
Radome measurements
Table 5-1: Keyboard shortcuts
Ctrl-SSaves the currently displayed measurement results.
See "Managing measurement results"on page 50
for details.
Ctrl-OOpens a dialog box to load measurement results.
See "Managing measurement results"on page 50
for details.
Ctrl-QCloses the application.
F3Initiates a reflection normalization
F4Initiates a transmission normalization
F5Initiates a measurement.
5.1.3Measurements
The R&S QAR provides several measurement types: reflection measurements and
transmission measurements. Transmission measurements require the optional transmitter module. Both measurements usually consist of two steps, normalization and the
actual measurement.
Measuring a DUT
1. Set up the normalization.
2. Start the normalization with "Normalize Reflection" (or the F3 key) and "Normalize
Transmission" (or the F4 key).
For details on how to normalize the measurement, see Chapter 5.1.3.1, "Normali-
zation", on page 43.
3. Place the DUT in front of the R&S QAR, for example on the optional mounting
table.
4. Start the reflection and transmission measurement with "Measure" (or the F5 key).
The measurement itself lasts less than a second. The results are displayed after a
few seconds.
For details on how to run the measurement, see Chapter 5.1.3.2, "Reflection mea-
surement", on page 45 and Chapter 5.1.3.3, "Transmission measurement",
Normalization is a separate measurement that determines systematic, reproducible
measurement deviations. Determining these measurement deviations allows you to
43User Manual 1178.7200.02 ─ 12
Page 44
R&S®QAR
Operating the R&S QAR
Radome measurements
remove those deviations from the actual measurement results, which in turn improves
the accuracy of the measurement.
Normalizing reflection measurements
Normalizing the reflection measurement requires a clean, smooth and flat metallic
plate for a high reflection, for example aluminum. The recommended size is between
100 mm x 100 mm and 200 mm x 200 mm.
If you have the verification objects, and it fits your setup, you can use the plate labeled
E to normalize the reflection.
1. Place the plate at the same location where you are going to mount the DUT for the
actual measurement.
Make sure that the plate is parallel to the R&S QAR. We recommend using the
same fixture for the metal plate as for the DUT.
2. Select the "Normalize Reflection" button to initiate the measurement.
When the normalization measurement is done, the R&S QAR saves the correction
data. All subsequent measurements are corrected accordingly.
A valid normalization results in an image with high reflection. The mean reflection
should be approximately 0 dB and should not exceed a certain deviation from that
value - refer to the datasheet for exact values.
3. Optional: Scan the metal plate with the "Measure" button to verify the normalization.
If the results do not meet the expectations, repeat the normalization.
If the second normalization still yields unexpected results, verify the test setup. For
details about the verification procedure, see Chapter 5.1.5, "Verification",
on page 52.
Normalizing the transmission measurement
Normalizing the transmission measurement is done by a free-space measurement and
requires no additional devices. However, make sure that the line of sight between
R&S QAR and transmitter module is free of obstacles.
1. Select the "Normalize Transmission" button to initiate the measurement.
When the normalization measurement is done, the R&S QAR saves the correction
data. All subsequent measurements are corrected accordingly.
2. Optional: scan the free space with the "Measure" button to verify the normalization.
A valid normalization results in a straight trace at approximately 0 dB attenuation.
The trace should not exceed a certain deviation from that value - refer to the datasheet for exact values.
If the results do not meet the expectations, repeat the normalization.
If the second normalization still yields unexpected results, verify the test setup. For
details about the verification procedure, see Chapter 5.1.5, "Verification",
on page 52.
44User Manual 1178.7200.02 ─ 12
Page 45
R&S®QAR
5.1.3.2Reflection measurement
Operating the R&S QAR
Radome measurements
The reflection measurement determines the amount of energy that is being reflected by
the DUT. This energy does not pass through the radome and contributes to performance degradation as a result. Reflected signals reduce the performance of the radar
and can even interfere with the received signals, leading to effects as described in
application note 1MA267. Areas with a high reflection can have various causes, for
example material defects, undesired interaction between several layers of materials,
excessive amount of certain materials, or foreign objects.
The R&S QAR achieves spatially resolved reflection measurements of a DUT by linking the information collected by the distributed transmit and receive antennas in a
coherent operation. The resulting millimeter wave image allows for an intuitive evaluation of the DUTs reflection behavior.
Measurement results
The application shows the results of the spatially resolved reflection measurement as a
millimeter wave image with a linear scale from 10 % to 100 %. The x- and y-axis show
the dimensions of the image in mm (300 mm x 300 mm by default). By default, the
image is a grayscale image. Small values, represented by dark areas, correspond to a
low reflection, high values, represented by bright areas, correspond to a high reflection.
A color map next to the image shows the levels that correspond to each shade of color.
In addition to the image, the mean reflection and the standard deviation of the reflection are calculated as numeric results. Both values are shown under the millimeter
wave image in logarithmic units (dB) and as a percentage, with 100 % representing a
maximum reflection of 0 dB.
The application calculates the numeric results over a certain area of the complete millimeter wave image. The evaluation area is indicated by a mask that is superimposed
on the image. You can customize the size and form of this mask to analyze certain
areas in more detail. For more information about working with the mask, see "Working
with evaluation masks"on page 46.
The mean reflection indicates the radar transparency of the DUT. The lower the mean
reflection, the better the radar transparency of the DUT. A higher mean reflection
results in beam refraction, increased transmission loss and standing waves between
the DUT and the radar sensor. For optimal radar compatibility, the DUT's mean reflection has to be minimized.
The reflection's standard deviation indicates the homogeneity of the DUT. Inhomogeneous DUTs alter the incident wave's phase fronts, cause angle errors, blind spots and
increased side lobes. For optimal radar compatibility, the DUT's reflection standard
deviation has to be minimized.
5.1.3.3Transmission measurement
Because the incident signal is split in a reflected and a transmitted part, it is of interest
to measure not only the reflection but also the transmission properties.
Measuring the transmitted signal, requires the optional transmitter module that is located behind the DUT (on the DUT table). The transmitter uses 64 discrete frequency
45User Manual 1178.7200.02 ─ 12
Page 46
R&S®QAR
Operating the R&S QAR
Radome measurements
points to cover a selected frequency span between 72 GHz and 82 GHz. This allows
for an exact evaluation of the DUT's transmission frequency response.
The frequency response yields detailed information regarding the RF transmission loss
of the DUT at the exact frequency band intended for radar operation. This is therefore
independent of the actual signal waveform utilized by the radar unit, which facilitates
the testability and optimization of the radome itself.
The transmission diagram shows the frequency response of the DUT. The minimum
and maximum values indicate the degree of frequency matching achieved by the DUT.
Ideally, the minimum of the frequency response is located at the operating frequency
range of the radar sensor that is to be used in combination with the DUT. Shifted minima indicate issues with the DUT's electrical thickness.
The R&S QAR calculates the one-way mean attenuation for specific frequency ranges
(76 GHz to 77 GHz and 76 GHz to 81 GHz). The results are displayed below the diagram in logarithmic units (dB).
The one-way mean attenuation is calculated as follows:
-20·log10(mean transmission coefficient)
The mean transmission coefficient is calculated over the respective frequency range.
You can also query the minimum attenuation (and its frequency) using SCPI commands. For details ,see Chapter 6.2.4.2, "Measurements (12 cluster)", on page 90.
5.1.4Configuration
The R&S QAR provides several tools to configure the radome measurement.
Working with evaluation masks
The evaluation mask available for the reflection measurement defines the area that is
evaluated for the mean reflection and the standard deviation of the reflection measurement.
1. Select "Edit Mask".
In editing mode, you can change the shape and location of the evaluation mask.
2. Move, delete or design an evaluation mask.
a) Move the mask by dragging it around with the mouse and dropping it in its new
location.
b) Delete a selected mask with the "Del" key on your keyboard. To select a mask,
click it.
c) Create a mask by clicking somewhere in the result display and drawing a line
with the mouse.
One click confirms a line of the mask shape. Add as many lines as you like to
design the mask. A double-click confirms the final shape of the mask.
Using that method, you can design any shape you like around the DUT.
3. Select "Edit Mask" again to exit the editing mode.
46User Manual 1178.7200.02 ─ 12
Page 47
R&S®QAR
Operating the R&S QAR
Radome measurements
The R&S QAR reevaluates the mean reflection and the standard deviation.
4. Optional: Export the evaluation mask.
a) Select "Options" > "Reflection Mask"
b) Select "Save Bitmap" to save the shape and location of the mask as a bitmap
(.bmp file).
c) Select "Export Coordinates" to save the shape and location of the mask as
coordinates in a .json file.
Alternatively, you can design a mask with an external graphics editor or a text editor and import it.
If you design the mask as a bitmap, make sure that the canvas size is 601 x 601
pixels, and that the image has a single channel. Take a mask you have designed in
the application to see what a bitmap looks like.
If you design a mask as text, make sure that you define the coordinates of the xand y-axis correctly. Take a mask you have designed in the software to see what
the .json file looks like.
Importing evaluation masks
If you have designed and saved an evaluation mask previously, you can import it any
time you need it.
Before importing a mask, make sure that the mask editing mode is off.
1. Select "Options" > "Reflection Mask"
2. Select "Load Bitmap" to import the shape and location of the mask from a bitmap
(.bmp file).
3. Select "Import Coordinates" to import the shape and location of the mask from
coordinates in a .json file.
Selecting the image color scheme
The R&S QAR provides several color schemes for the image of the reflection measurement.
1. Select "Options" > "Display".
2. Select one of the color schemes from the "Colormap" dropdown menu.
The R&S QAR adjusts the colors accordingly.
Defining an evaluation grid
Requires R&S QAR-K20.
An evaluation grid helps you to find the areas with highly differing reflection coefficients
on your DUT more easily. The principle is to divide the DUT into several small areas
(cells), and evaluate the characteristics of each cell separately instead of the complete
DUT.
Before you can start to evaluate results based on a grid, you have to design a grid.
47User Manual 1178.7200.02 ─ 12
Page 48
R&S®QAR
Operating the R&S QAR
Radome measurements
You can design two layers in the grid ("Grid 1" and "Grid 2") and thus refine the evaluation areas. Using two grids can be useful to increase the defect localization resolution
without sacrificing the averaging over a larger cell size.
1. If you already have defined a grid previously, select "Options" > "Grid Evaluation
Settings" > "Load Grid Definition" to restore a grid.
If you need a new one or do nor have one yet, you have to create a grid.
2. Select "Options" > "Grid Evaluation Settings" > "Edit Grid Definition".
The preview in the upper part of the dialog shows the current design of the grid(s).
3. Define the absolute size of the grid in mm (height and width).
The grid can have any size, as long as it is not larger than the millimeter wave
image itself.
4. Define the size of the grid in terms of the number of cells.
Grid cells are numbered 1 to x, beginning at the top left cell.
Note that "Grid 2" always has one (horizontal or vertical) cell less than "Grid 1".
5. Define the location of the grid in terms of an offset from the center of the image.
●A positive x offset moves the grid to the right of the center.
●A negative x offset moves the grid to the left of the center.
●A positive y offset moves the grid upwards.
●A negative y offset moves the grid downwards.
6. Define an alarm threshold.
The R&S QAR evaluates the reflection values in each cell against a limit (defined
by a threshold). The threshold is a deviation from the mean reflection of all master
cells.
If the mean reflection in one of the cells exceeds the threshold (in both directions),
the corresponding cell turns red to indicate a failed limit check.
reflection in cell > mean
reflection + threshold = fail
threshold
mean reflection in master cells
threshold
reflection in cell > mean
reflection - threshold = fail
Figure 5-4: Grid evaluation logic
7. Select cells that you want to exclude from the evaluation (for both grids).
Excluded cells are not evaluated at all. These cells are not used to calculate the
mean reflection and not tested against the alarm threshold. In the grid definition
dialog, excluded cells appear in red color.
There are two ways to add cells into the input field.
48User Manual 1178.7200.02 ─ 12
Page 49
R&S®QAR
Operating the R&S QAR
Radome measurements
a) Select the grid whose cells you want to exclude "Grid 1" or "Grid 2". Excluding
cells is not possible while both grids are displayed in the preview.
b) Write the cell numbers into the input field, separated by a ":", for example:
"1:2:4:5".
c) Select the input field (put the cursor in the field) and double-click the cells to
add them to the list.
8. Define master cells.
Master cells are those cells that contribute to the mean reflection result. All other
cells are ignored for the calculation of the mean reflection. If you do not define
master cells, the R&S QAR does not calculate the mean reflection at all. In the grid
definition dialog, master cells appear in green color.
Note that the mean reflection is evaluated separately for "Grid 1" and "Grid 2".
Master cells and undefined cells (= cells that are not excluded and are not master
cells and thus remain white in the grid definition dialog) are both evaluated against
the alarm threshold.
9. Confirm the grid definition with "OK" (closes the dialog) or "Apply".
10. Save your grid with "Save Grid Definition As".
11. Select "Options" > "Grid Evaluation Settings" > "Grid Evaluation" to apply the grid
to the image.
Turning on the grid superimposes an image of the grid over the reflection result display.
Master cells are green. Excluded cells are not displayed. If the reflection in one of
the cells violates the threshold limit you have defined, the corresponding cell turns
red.
49User Manual 1178.7200.02 ─ 12
Page 50
R&S®QAR
Operating the R&S QAR
Radome measurements
21
Figure 5-6: Grid evaluation example
1 = Green cells indicate master cells
2 = Blue grid indicates cells of "Grid 1" (not shown: a yellow grid indicates cells of "Grid 2")
3 = Red cells indicate a violation of the threshold somewhere in the area the cell covers
3
Managing measurement results
You can export the measurement data, for example to store it on an external memory
device or copy it to a different computer for further analysis with different software. You
can also restore data that you have taken at some point and review it in the application.
1. Select "Options" > "Recording".
2. Define the default directory in which to save the results in the "Result Folder" input
field.
For each measurement, the R&S QAR creates a folder that contains the measurement data.
3. Select "Autosave" to use the auto save functionality.
When you are using the autosave feature, the R&S QAR automatically stores the
results after each measurement.
Tip: You can always save the measurement results deliberately with the keyboard
shortcut Ctrl-S or from the "File" menu.
4. Select the data you want to save from the "Results To Save" dropdown menu.
A measurement dataset contains the following files.
<yyyy-mm-dd-hh-mm-ss>_cfg.txt
●
<yyyy-mm-dd-hh-mm-ss>_PanelConf.mat
Contains information about the configuration of the R&S QAR during the measurement.
<yyyy-mm-dd-hh-mm-ss>_image.png
●
<yyyy-mm-dd-hh-mm-ss>_transmissionloss.png
50User Manual 1178.7200.02 ─ 12
Page 51
R&S®QAR
Operating the R&S QAR
Radome measurements
Image of the reflection and transmission loss measurement.
Protocol.csv
●
Contains the numerical measurement results.
<yyyy-mm-dd-hh-mm-ss>_results.mat
●
Contains the raw 2D imaging data.
Only if you have selected the data package "Statistics + Pictures + Data". This
data package is necessary if you want to restore the measurement results later
on.
<yyyy-mm-dd-hh-mm-ss>_volume.mat
●
Contains the raw 3D imaging data.
Only if you have selected the data package "Statistics + Pictures + Data + Volume".
This data package requires approx. 500 MB of disk space for each image. It
also takes a long time to save this data package, approx. 30 seconds for each
image.
●For grid evaluation:
<yyyy-mm-dd-hh-mm-ss>_GridResult.csv
Contains the mean values of every grid cell, the mean value of the master
cells, the threshold value and an evaluation if the limit check in a cell has failed.
●For grid evaluation:
Configuration file (.json) of the specified grid.
Tip: You can restore measurement results at any time with the keyboard shortcut
Ctrl-O or from the "File" menu.
Restoring measurement results requires at least the raw 2D imaging data
(results.mat). The data package "Statistics + Pictures" only contains the
images (.png) and statistics (numerical results etc.) and is therefore not sufficient
to restore results.
Taking pictures with a webcam
You can connect a webcam to the R&S QAR that takes a picture of the DUT during the
measurement. This picture is stored together with the measurement data.
The R&S QAR supports all webcams that are recognized as an "Imaging Device" in
the Windows 10 device manager. Note that the webcam feature of the R&S QAR
always uses the first imaging device in the list to take pictures.
1. Select "Options" > "Recording".
2. Select "Save Webcam Picture".
Using a barcode scanner
You can connect a barcode scanner to the R&S QAR to scan the barcode of a DUT
and use this as a reference that the DUT with a certain barcode has passed or failed
the test. The barcode is displayed in the user interface and is stored together with the
other measurement data.
51User Manual 1178.7200.02 ─ 12
Page 52
R&S®QAR
Operating the R&S QAR
Radome measurements
The R&S QAR supports the barcode scanner Honeywell Xenon 1900 by Honeywell
International Inc.
1. Select "Options" > "Recording".
2. Select "Use Barcode Scanner".
3. Select the (serial) port the barcode scanner is connected to from the dropdown
menu.
5.1.5Verification
The R&S QAR provides a verification procedure to validate the quality of the measurements. Verification consists of a series of measurements with standardized DUTs
which are located at a fix distance from the R&S QAR.
All verification objects are delivered with a quality approved surface finish. Any surface
features are either intentional or are considered to be uncritical for the intended purpose.
If verification fails, validate the verification setup. If verification fails several times in a
row, despite a valid setup, contact Rohde & Schwarz customer support to calibrate the
R&S QAR and check the hardware.
Warm-up time
Let the R&S QAR warm up for at least 90 minutes before using it.
Verification requires one of verification sets available for radome measurements.
●
Verifying 12-cluster measurements requires R&SQAR-Z40 or -Z43.
●
Verifying single cluster measurements requires R&S QAR-Z43.
The setup and installation of the verification sets R&S QAR-Z40 and -Z43 is similar.
The difference is that the frame and objects of the R&S QAR-Z43 are higher.
52User Manual 1178.7200.02 ─ 12
Page 53
R&S®QAR
Operating the R&S QAR
Radome measurements
R&S QAR-Z40R&S QAR-Z43
11
Figure 5-7: Test setup for verification of radome measurements
All required equipment is included in the delivery of the verification set.
●
Verification objects
–1 x metal plate
–1 x metal plate with cut-outs
–3 x plastic plate with different thicknesses
●
Frame for verification objects
●
Allen wrench size 5
Mounting the frame for the verification objects
The verification objects come with a frame that holds the objects. Screw the frame to
the DUT mounting table. The frame base that you screw to the table is the same for
both verification sets.
1. Place the frame on the mounting table. The position of the frame is defined by the
mounting holes in the DUT mounting table.
2. Screw the frame to the mounting table with the Allen wrench. The required socket
head screws (4 x) are included in the delivery of the R&S QAR-Z40.
Verifying the system
The verification process consists of five consecutive measurements with the verification objects (square plates). The verification objects are labeled A to E.
1. Select verification mode in the application ("Options" > "Verification").
In verification mode, the R&S QAR eliminates unwanted responses in the signal by
applying a time gate to evaluate the data. For verification, using this method is
mandatory.
NOTICE! Handling the verification objects.
2.
●Store the verification objects in the bag they were delivered in.
●Keep the verification objects clean and without fingerprints.
●Do not scratch the verification objects.
●Clean the objects with a lint-free duster or a damp rag. Do not use chemicals.
Normalize the reflection measurement.
54User Manual 1178.7200.02 ─ 12
Page 55
R&S®QAR
a) Put the metal plate labeled "E" into the frame.
b) Select "Normalize Reflection" to initiate a measurement.
c) Wait until the results are displayed.
d) Remove the metal plate from the frame.
The resulting image should show a homogeneous image (the colors should not
deviate too much). The mean reflection should be around 0 dB.
The limits for the normalization of the reflection measurement are universal, not
device-specific.
3. Verify the reflection measurement.
a) Put the metal plate with cut-outs labeled D into the frame.
b) Initiate a measurement.
c) Remove the metal plate from the frame.
The resulting image should show a clear image of the verification object. The
patterns should be clearly visible.
Verification OKVerification not OK
Operating the R&S QAR
Radome measurements
4. If you are using the transmitter module:
Normalize the transmission measurement.
a) Remove all objects in front of the R&S QAR (except for the table and the
frame).
b) Select "Normalize Transmission" to initiate a free-space measurement.
The resulting trace should be a straight line at approximately 0 dB attenuation.
5. If you are using the transmitter module:
Verify the transmission measurement.
a) Put the plastic plate labeled A into the frame.
b) Initiate a measurement.
c) Remove the plastic plate from the frame.
55User Manual 1178.7200.02 ─ 12
Page 56
R&S®QAR
Operating the R&S QAR
Single cluster radome measurements
d) Do the same for the other two plastic plates labeled B and C.
After each measurement, check the mean transmission loss against the limits a
and a81 specified on the stickers attached to the verification plates. The sticker
shows the attenuation characteristics of the plate and the deviation the result may
have.
Each sticker contains two values.
The first value "a77" is the one-way mean attenuation in the frequency range from
76 GHz to 77 GHz.
The second value "a81" is the one-way mean attenuation in the frequency range
from 76 GHz to 81 GHz.
The limits for the transmission verification measurement are plate-specific.
If you are using the R&S QAR-Z43 verification set, the plates also have a third
value (r77), which is irrelevant for the verification of the radome measurements that
uses all clusters.
5.2Single cluster radome measurements
The single cluster application is similar to radome measurements. The difference is
that the single cluster application measures the DUT characteristics using only a single
cluster instead of all 12 clusters.
77
The typical procedure when measuring a radome (DUT) with one cluster is as follows:
1. Start the R&S QAR to start the system and the measurement application.
For more information about the functionality of the measurement application, see
Chapter 5.2.2, "Graphical user interface", on page 58.
2. Normalize the test setup.
For more information about the normalization, see Chapter 5.2.3, "Measurements",
on page 59.
3. Measure the DUT.
For more information about the measurement and its results, see Chapter 5.2.3,
"Measurements", on page 59.
4. Save and restore measurement data.
For more information about saving and restoring measurement data, see Chap-
ter 5.2.4, "Configuration", on page 62.
5.2.1Test setup
The basic test setup for radome measurements consists of the R&S QAR and a DUT
that is placed at a certain distance from the R&S QAR.
Note that there has to be at least 30 cm of free space behind the DUT for valid measurement results.
56User Manual 1178.7200.02 ─ 12
Page 57
R&S®QAR
Operating the R&S QAR
Single cluster radome measurements
Figure 5-9: Setup for single cluster measurements
The DUT must be positioned directly opposite to the cluster that measures the DUT
and therefore slightly off-center to the center of the R&S QAR.
In Figure 5-10, the box indicates the area in which the R&S QAR takes the image of
the DUT.
Figure 5-10: Scan volume for single cluster measurements
57User Manual 1178.7200.02 ─ 12
Page 58
R&S®QAR
Operating the R&S QAR
Single cluster radome measurements
5.2.2Graphical user interface
The graphical user interface (GUI) contains several elements.
23514
689107
Figure 5-11: Software graphical user interface for single cluster measurements
1 = Menu bar
2 = Measurement description
3 = Single cluster measurement selection
4 = Result display (reflection)
5 = Overview test setup
6 = Measure button
7 = Status bar
8 = System messages
9 = Normalize measurement
10 = Numeric results
Result display
The result display shows the data that was collected during the measurement as a millimeter wave image.
Numeric results
Several numeric results of the reflection and transmission measurements are displayed.
●
Mean reflection
58User Manual 1178.7200.02 ─ 12
Page 59
R&S®QAR
Operating the R&S QAR
Single cluster radome measurements
Settings and control buttons
Several settings and buttons allow you to control the measurement.
●
"Measure"
Initiates a measurement.
For more information, see Chapter 5.2.3, "Measurements", on page 59.
●
"Normalize Reflection"
Initiates the normalization of the reflection measurement.
For more information, see Chapter 5.2.3, "Measurements", on page 59.
●
System messages
Shows various status messages of the system like a missing normalization or the
remaining warm-up time. The full warm-up time is 90 minutes.
Menu bar
The menu bar contains various settings.
●
"File"
Contains functionality to import and export measurement results, and functionality
to close the software.
●
"Options"
Contains functionality to configure the measurement and result displays.
●
"Info"
Contains various information about the application, including information about
open source software licenses and the user manual.
Keyboard shortcuts
The following keyboard shortcuts are available.
Table 5-2: Keyboard shortcuts
Ctrl-SSaves the currently displayed measurement results.
Ctrl-OOpens a dialog box to load measurement results.
Ctrl-QCloses the application.
F3Initiates a reflection normalization
F5Initiates a measurement.
5.2.3Measurements
For the single cluster application, the R&S QAR provides a reflection measurement.
The reflection measurement usually consists of two steps, normalization and the actual
measurement.
The concept of the single cluster reflection measurement is the same as the reflection
measurement that uses all clusters - see Chapter 5.1.3.2, "Reflection measurement",
See "Managing measurement results"on page 50
for details.
See "Managing measurement results"on page 50
for details.
59User Manual 1178.7200.02 ─ 12
Page 60
R&S®QAR
Operating the R&S QAR
Single cluster radome measurements
on page 45 for details. For single cluster measurements, the R&S QAR uses the cluster indicated in Figure 5-12 and a fix bandwidth of 1 GHz, a typical bandwidth for automotive radar applications.
128
7
1115
16
61014
5
Figure 5-12: Cluster used in single cluster measurements
913
Using only a single cluster with a fix bandwidth of 1 GHz does not yield the same
image resolution compared to measurements with the full set of clusters. The resulting
image therefore gets blurry and details of the DUT are no longer visible.
However, the application does yield results that correlate to the results that you get
using a conventional network analyzer. Since setups with a network analyzer are usually complicated, the single cluster application of the R&S QAR is a good tool to verify
the absolute reflection values of a DUT. If the mean reflection is too high or too low,
you can then use the full cluster application to find the faulty locations. This technique
can be useful in an environment with a high throughput of DUTs, for example production areas.
In the image, each color corresponds to a certain (range of) reflection values as indicated in the color map next to the image.
The displayed mean reflection corresponds to a small area (100 x 100 mm) in the center of the DUT, highlighted by a colored rectangle in the image. In this area, the
R&S QAR calculates the mean reflection over all measurement points that deviate no
more than 3 dB from the maximum reflection value.
60User Manual 1178.7200.02 ─ 12
Page 61
R&S®QAR
Operating the R&S QAR
Single cluster radome measurements
234
1
1
Figure 5-13: Image of a homogenous DUT
1 = Size of the image (in mm)
2 = Evaluation area
3 = Area that contains values for mean reflection
4 = Color map
The measurement points (or areas of measurement points) that the R&S QAR uses to
calculate the mean reflection with are also highlighted. Note that there can be several
such areas in the image, depending on the characteristics of the DUT.
Measuring a homogenous (= flat, highly reflective) DUT usually results in a Gaussian
distribution of reflection values (due to the aperture weights of the antenna array). The
evaluated values are therefore concentrated to a single area in the image.
max. reflection
value range for
mean reflection
- 3 dB threshold
Figure 5-14: Mean reflection on a homogenous DUT
61User Manual 1178.7200.02 ─ 12
Page 62
R&S®QAR
Operating the R&S QAR
Single cluster radome measurements
For inhomogeneous DUTs, the distribution can be different, so there maybe several
highlighted areas in the image, but the evaluation remains the same.
max. reflection
value range for
mean reflection
- 3 dB threshold
Figure 5-15: Mean reflection on an inhomogeneous DUT
Performing the measurement sequence
Similar to measurements using all clusters, you have to normalize the setup before
beginning the actual measurement. For a comprehensive description of the normalization see "Normalizing reflection measurements"on page 44.
1. Normalize the setup with a suitable object.
We recommend using the plate labeled E from the verification kit R&S QAR-Z43.
2. Replace the normalization object with the DUT.
3. Measure the DUT.
5.2.4Configuration
The configuration tools for single cluster measurements are similar to those available
for 12 cluster radome measurements.
For a comprehensive description, see:
●
"Selecting the image color scheme"on page 47
●
"Managing measurement results"on page 50
The folder that contains single cluster measurement data has a suffix _SC. Otherwise, the data is the same.
●
"Taking pictures with a webcam"on page 51
●
"Using a barcode scanner"on page 51
62User Manual 1178.7200.02 ─ 12
Page 63
R&S®QAR
Operating the R&S QAR
Single cluster radome measurements
5.2.5Verification
The R&S QAR provides a verification procedure to validate the quality of the measurements. Verification consists of a series of measurements with standardized DUTs
which are located at a fix distance from the R&S QAR.
All verification objects are delivered with a quality approved surface finish. Any surface
features are either intentional or are considered to be uncritical for the intended purpose.
If verification fails, validate the verification setup. If verification fails several times in a
row, despite a valid setup, contact Rohde & Schwarz customer support to calibrate the
R&S QAR and check the hardware.
Warm-up time
Let the R&S QAR warm up for at least 90 minutes before using it.
The setup is the similar to the verification setup for the standard radome measurements. To verify single cluster measurements, however, you have to use the verification set R&S QAR-Z43. For details, see Chapter 5.1.5, "Verification", on page 52.
Performing verification
The verification process consists of five consecutive measurements with the verification objects (square plates). The verification objects are labeled A to E.
NOTICE! Handling the verification objects.
1.
●Store the verification objects in the bag they were delivered in.
●Keep the verification objects clean and without fingerprints.
●Do not scratch the verification objects.
●Clean the objects with a lint-free duster or a damp rag. Do not use chemicals.
Normalize the reflection measurement.
a) Put the metal plate labeled "E" into the frame.
b) Select "Normalize Reflection" to initiate a measurement.
c) Wait until the results are displayed.
d) Remove the metal plate from the frame.
The resulting image should show a homogeneous image (the colors should not
deviate too much). The mean reflection should be around 0 dB.
The limits for the normalization of the reflection measurement are universal, not
device-specific.
2. Verify the reflection measurement.
a) Put the plastic plate labeled A into the frame.
b) Initiate a measurement.
c) Remove the plastic plate from the frame.
d) Do the same for the other two plastic plates labeled B and C.
63User Manual 1178.7200.02 ─ 12
Page 64
R&S®QAR
Operating the R&S QAR
Bumper measurements
After each measurement, check the mean reflection against the limits r77 specified
on the stickers attached to the verification plates. The sticker shows the reflection
characteristics of the plate and the deviation the result may have.
The other two values a77 and a81 are irrelevant for the verification of the single clus-
ter application.
5.3Bumper measurements
In real cases, the vehicle built-in radar sends a signal that crosses the bumper, is
reflected and crosses the bumper again. Similarly, the R&S QAR antennas send signals that cross the bumper, are reflected from the reference reflector pins and cross
the bumper back. The received signals are measured and compared to the free-space
measurement (normalization) to obtain the transmission loss (or two-way attenuation)
caused by the bumper in multiple points.
The R&S QAR uses 64 discrete frequency points to cover the frequency range 76 GHz
- 81 GHz. Thanks to the multistatic imaging of the reflector pins, the analyzed bumper
is irradiated from different angles (from -38° to +24°), as shown in figure Figure 5-16.
Figure 5-16: Irradiation area: 3D, top view and lateral view
The mean two-way attenuation is calculated as follows:
-20·log10(mean attenuation coefficient).
The mean attenuation coefficient is calculated as an average between the values
obtained for all discrete frequency points and for all the reflector pins used.
64User Manual 1178.7200.02 ─ 12
Page 65
R&S®QAR
Operating the R&S QAR
Bumper measurements
The typical procedure when measuring the transmission loss (attenuation) caused by a
bumper is as follows:
1. Start the R&S QAR.
The software starts automatically. For more information, see Chapter 5.3.2,
"Graphical user interface", on page 65.
2. Normalize the measurement, as described in Chapter 5.3.3.1, "Normalization",
on page 68.
3. Place the bumper to measure in the right position, as described in Chapter 5.3.1,
"Test setup", on page 65.
4. Measure the attenuation, as described in Chapter 5.3.3.2, "Attenuation measure-
ment", on page 68.
5. Save the measurement results, if needed.
See also "Managing measurement results"on page 69.
5.3.1Test setup
► Place the bumper between the R&S QAR and the reference reflector, at a mini-
mum distance of 30 mm from the reflector pins.
5.3.2Graphical user interface
The graphical user interface (GUI) contains several elements.
21
345
6
89710
Figure 5-17: Software user interface
65User Manual 1178.7200.02 ─ 12
Page 66
R&S®QAR
Operating the R&S QAR
Bumper measurements
1 = Menu bar
2 = Bumper type selection
3 = System messages
4 = Normalize button
5 = Result display
6 = Result display hide / unhide
7 = Save button
8 = Status bar
9 = Numeric results (mean attenuation), incl. result indicator
10 = Measure button
10 = Result display
Result display
Shows the millimeter wave image of the data collected from the reference reflector pins
during the scan.
The image indicates the effects of the presence of the object and enables you to evaluate immediately its characteristics, for example, the homogeneity.
You can hide and unhide the image with the arrow icon ( ).
Numeric results and indicator
Two-way mean attenuation within the frequency range 76 GHz - 81 GHz, calculated as
the arithmetical mean of the values registered by the pins used.
If the attenuation is below the threshold, the indicator is green.
Bumper type selection
Selects the configuration for the measurement (type of bumper you are measuring).
The configuration determines which pins of the reference reflector to use for the measurement. The "default" configuration uses all pins.
Settings and control buttons
Several settings and buttons allow you to control the measurement.
●
"Measure"
Starts a measurement.
For more information, see Chapter 5.3.3.2, "Attenuation measurement",
on page 68.
●
"Save"
Saves the measurement results.
For more information, see "Managing measurement results"on page 69.
●
"Normalize"
Starts the normalization.
For more information, see Chapter 5.3.3.1, "Normalization", on page 68.
●
System messages
Shows various status messages of the system like a missing normalization or the
remaining warm-up time. The full warm-up time is 90 minutes.
66User Manual 1178.7200.02 ─ 12
Page 67
R&S®QAR
Operating the R&S QAR
Bumper measurements
Menu bar
The menu bar contains various settings.
●
"File"
Contains functionality to import and export measurement results, and functionality
to close the software.
●
"Options"
Contains functionality to configure the measurement and result displays.
For more information, see "Managing measurement results"on page 69.
●
"Info"
Contains various information about the application, including information about
open source software licenses and the user manual.
Keyboard shortcuts
The following keyboard shortcuts are available.
Table 5-3: Keyboard shortcuts
Ctrl-SSaves the currently displayed measurement results.
See "Managing measurement results"on page 69
for details.
Ctrl-OOpens a dialog box to load measurement results.
Ctrl-QCloses the software.
F5Initiates a measurement.
5.3.3Measurements
For bumper measurements, the R&S QAR provides a reflection measurement. Bumper
measurements require the optional reference reflector. The measurement usually consist of two steps, normalization and the actual measurement.
Measuring a DUT
1. Set up the normalization.
2. Start the normalization with "Normalize".
For details on how to normalize the measurement, see Chapter 5.3.3.1, "Normali-
zation", on page 68.
3. Place the DUT between the R&S QAR and the reference reflector.
4. Start the reflection measurement with "Measure" (or the F5 key).
The measurement itself lasts less than a second. The results are displayed after a
few seconds.
See "Managing measurement results"on page 69
for details.
67User Manual 1178.7200.02 ─ 12
Page 68
R&S®QAR
5.3.3.1Normalization
Operating the R&S QAR
Bumper measurements
For details on how to run the measurement, see Chapter 5.3.3.2, "Attenuation
Normalization is a measurement done in the free space, to obtain a reference value for
the bumper measurements. You do not need any object to measure or additional element.
Normalization enables you also to check the positioning and the status of the reference
reflector. If the position is wrong, or the reflector is damaged, you get a warning.
Perform the following steps:
1. Make sure that nothing is obstructing the view between R&S QAR and the reference reflector.
2. Choose the configuration needed from the "bumper type" drop-down menu.
3. Click the "Normalize" button to start the measurement.
The R&S QAR measures the deviations and saves the correction data. All subsequent measurements are corrected accordingly.
4. Optional: scan the free space by clicking the "Measure" button to verify the normalization.
The expected result is an attenuation around 0 dB. For a range of acceptable
results, see the datasheet.
If the results do not meet the expectations, repeat the normalization.
If the second normalization still yields unexpected results, perform the verification,
as described in Chapter 5.3.5, "Verification", on page 71.
5.3.3.2Attenuation measurement
1. Set a threshold, if needed, by choosing "Options" > "Threshold".
2. Click the "measure" button. Or press F5.
Do not move the bumper during the measurement.
The measurement takes a few seconds, then the results appear on the GUI. See
"Numeric results and indicator"on page 66 and "Result display"on page 66.
If you change the position of the reference reflector, or set a different configuration to
use in the "bumper type" drop-down list, perform a new normalization before measuring.
68User Manual 1178.7200.02 ─ 12
Page 69
R&S®QAR
Operating the R&S QAR
Bumper measurements
5.3.4Configuration
The R&S QAR provides several tools to configure the radome measurements.
Managing bumber types
To get valid results in the bumper measurement, you have to select the bumper model
you want to measure first.
Upon delivery, the R&S QAR does not contain any bumper data. For each bumper type
you want to measure, you have to add the corresponding bumper characteristics first,
based on the CAD data of your bumpers.
If you want to add new bumper types, contact your Rohde & Schwarz representative.
1. For bumper models that are already available on the R&S QAR:
Select the model from the "Bumper Type" menu. This menu contains the data of all
bumper types that are currently stored on your R&S QAR.
2. For bumper models that are not yet available on the R&S QAR:
a) Copy the bumper data that you got from us into the following directory on your
b) Start or restart the software.
c) Select the bumper model from the "Bumper Type" menu.
Managing measurement results
You can export the measurement data, for example to store it on an external memory
device or copy it to a different computer for further analysis with different software. You
can also restore data that you have taken at some point and review them in the software.
1. Select "Options" > "Recording".
2. Define the default directory in which to save the results in the "Result Folder" input
field.
For each measurement, the R&S QAR creates a folder that contains the measurement data.
3. Select "Autosave" to use the auto save functionality..
When you are using the autosave feature, the R&S QAR automatically stores the
results after each measurement.
Tip: You can always save the measurement results deliberately with the keyboard
shortcut Ctrl-S or from the "File" menu.
4. Select the data you want to save from the "Results To Save" dropdown menu.
A measurement dataset contains the following files.
<yyyymmdd-hhmmss>_cfg.txt
●
Contains information about the configuration of the R&S QAR during the measurement.
<yyyymmdd-hhmmss>_check_results.json
●
69User Manual 1178.7200.02 ─ 12
Page 70
R&S®QAR
Operating the R&S QAR
Bumper measurements
Contains information about options.
<yyyymmdd-hhmmss>_image.png
●
Image of the measurement.
<yyyymmdd-hhmmss>_result_QARQT.mat
●
The contents depend on your selection.
–If you have selected the data package "Statistics + Pictures + Data", the file
contains the raw 2D imaging data. This data package is necessary if you
want to restore the measurement results later on.
–If you have selected the data package "Statistics + Pictures + Data + Vol-
ume", the file also contains the 3D imaging data.
This data package requires approx. 500 MB of disk space for each image.
It also takes a long time to save this data package, approx. 30 seconds for
each image.
Tip: You can restore measurement results at any time with the keyboard shortcut
Ctrl-O or from the "File" menu.
Restoring measurement results requires at least the raw 2D imaging data
(result_QARQT.mat). The data package "Statistics + Pictures" only contains the
images (.png) and statistics (numerical results etc.) and is therefore not sufficient
to restore results.
Taking pictures with a webcam
You can connect a webcam to the R&S QAR that takes a picture of the DUT during the
measurement. This picture is stored together with the measurement data.
The R&S QAR supports all webcams that are recognized as an "Imaging Device" in
the Windows 10 device manager. Note that the webcam feature of the R&S QAR
always uses the first imaging device in the list to take pictures.
1. Select "Options" > "Recording".
2. Select "Save Webcam Picture".
Using a barcode scanner
You can connect a barcode scanner to the R&S QAR to scan the barcode of a DUT
and use this as a reference that the DUT with a certain barcode has passed or failed
the test. The barcode is displayed in the user interface and is stored together with the
other measurement data.
The R&S QAR supports the barcode scanner Honeywell Xenon 1900 by Honeywell
International Inc.
1. Select "Options" > "Recording".
2. Select "Use Barcode Scanner".
3. Select the (serial) port the barcode scanner is connected to from the dropdown
menu.
70User Manual 1178.7200.02 ─ 12
Page 71
R&S®QAR
Operating the R&S QAR
Bumper measurements
5.3.5Verification
The R&S QAR provides a verification procedure to validate the quality of the measurements. Verification consists of a series of measurements with standardized DUTs
which are located at a fix distance from the R&S QAR.
All verification objects are delivered with a quality approved surface finish. Any surface
features are either intentional or are considered to be uncritical for the intended purpose.
If verification fails, validate the verification setup. If verification fails several times in a
row, despite a valid setup, contact Rohde & Schwarz customer support to calibrate the
R&S QAR and check the hardware.
Warm-up time
Let the R&S QAR warm up for at least 90 minutes before using it.
Required equipment
All required equipment is included in the delivery of the R&SQAR-Z41.
●
Verification objects
–3 x plastic plate with different thicknesses
●
Frame for verification objects
●
Allen wrench size 5
Mounting the frame
The verification objects come with a frame that can hold the verification objects. The
frame can be screwed to the reference reflector stand.
1. Place the frame onto the reference reflector.
Figure 5-18: Placing the frame
2. Attach the verification plate to the frame. An arrow on the sticker indicates the correct orientation.
71User Manual 1178.7200.02 ─ 12
Page 72
R&S®QAR
Operating the R&S QAR
Bumper measurements
Figure 5-19: Placing the verification plate
Performing verification
The verification process consists of three consecutive measurements using the verification plates.
The plates are labeled "A" to "C". Each plate has a sticker with an arrow indicating the
correct orientation and the expected attenuation value with acceptable error margin.
NOTICE! Handling the verification objects.
1.
●Store the verification objects in the bag they were delivered in.
●Keep the verification objects clean and free from fingerprints.
●Do not scratch the verification objects.
●Clean the objects with a lint-free duster or a damp rag. Do not use chemicals.
Normalize the measurement as described in Chapter 5.3.3.1, "Normalization",
on page 68.
2. Put the plate labeled "A" into the support, as shown in Figure 5-18.
3. Set the "default" bumper type.
4. Initiate a measurement.
5. Check the mean attenuation against the values specified on the sticker.
6. Remove the plate from the support by pushing from the back of the plate.
7. Repeat the steps for the plates labeled "B" and "C".
If the verification fails, calibrate the R&S QAR, as described in the service manual, or
call your local R&S representative.
If after calibration the verification fails again, check the hardware for defects.
72User Manual 1178.7200.02 ─ 12
Page 73
R&S®QAR
Operating the R&S QAR
Radar positioning measurements
5.4Radar positioning measurements
The typical procedure when measuring a radar position is as follows:
1. Start the R&S QAR to start the system and the software.
For more information about the functionality of the software, see Chapter 5.4.2,
"Graphical user interface", on page 74.
2. Park the car in front of R&S QAR.
See Chapter 5.4.1, "Test setup", on page 73.
3. Measure the DUT.
See Chapter 5.4.3, "Measurement", on page 76.
4. Save measurement data.
See "Managing measurement results"on page 80.
5.4.1Test setup
For radar positioning measurements, the test setup consists of the R&S QAR and the
DUT, for example a radar module installed behind the bumper of a car. To measure the
position of the radar module, you simply park the car in front of the R&S QAR. Make
sure to park the car at the correct position as indicated in Figure 5-20 to get valid measurement results.
Figure 5-20: Test setup for radar positioning measurements (side view)
The optimal distance to the front of the car depends on the location of the radar module relative to the radome's location on the car.
●
The distance between R&SQAR and radar module is between 60cm and 80cm.
●
The distance from floor to radar module is between 34cm and 68cm.
●
The offset of the radar module relative to the center of the R&S QAR panel is
14.5 cm to the right of the center of the panel.
73User Manual 1178.7200.02 ─ 12
Page 74
R&S®QAR
Operating the R&S QAR
Radar positioning measurements
Figure 5-21: Test setup for radar positioning measurements (view from above)
5.4.2Graphical user interface
The graphical user interface (GUI) contains several elements.
123
11
10
Figure 5-22: Software graphical user interface
1 = Menu bar
2 = Car type
3 = Description
4 = Graphical results
5 = Display mode
6 = Start measurement
7 = System messages
8 = Status bar
9 = Overall result
10 = Numeric detection results
11 = Numeric positioning results
45
6789
74User Manual 1178.7200.02 ─ 12
Page 75
R&S®QAR
Operating the R&S QAR
Radar positioning measurements
Result display (graphic and numeric)
Graphic and numeric measurement results. For more information, see Chapter 5.4.3,
"Measurement", on page 76.
Settings and control buttons
Several settings and buttons allow you to control the measurement.
●
"Car Type"
Selects the measurement configuration (CAD data of the car). For more information, see "Selecting the working mode"on page 78.
●
"Description"
Optional description of the measurement. The description is part of the folder name
in which the results are stored.
●
"Measure"
Initiates a measurement.
For more information, see Chapter 5.1.3.2, "Reflection measurement", on page 45
and Chapter 5.1.3.3, "Transmission measurement", on page 45.
●
System messages
Shows various status messages of the system like a missing normalization or the
remaining warm-up time. The full warm-up time is 90 minutes.
Menu bar
The menu bar contains various settings.
●
"File"
Contains functionality to import and export measurement results and configurations, and functionality to close the software.
●
"Options"
Contains functionality to configure the measurement and result displays.
●
"Info"
Contains various information about the application, including information about
open source software licenses and the user manual.
Status bar
Shows various information:
●
The folder where measurements are saved.
●
The temperature of the connected R&S QAR.
Double-click on the temperature icon to view the temperature of the last few hours.
●
The connection state ("Online": successful boot up of the connected hardware).
Keyboard shortcuts
The following keyboard shortcuts are supported:
75User Manual 1178.7200.02 ─ 12
Page 76
R&S®QAR
Operating the R&S QAR
Radar positioning measurements
Table 5-4: Keyboard shortcuts
Ctrl-SSaves the currently displayed measurement results.
See "Managing measurement results"on page 80
for details.
Ctrl-OOpens a dialog box to load measurement results.
See "Managing measurement results"on page 80
for details.
5.4.3Measurement
The radar positioning measurement verifies the correct positioning of a radar module,
for example behind the bumper of a car.
To measure the position of the radar module, the R&S QAR generates a three-dimensional point cloud of the car that is in front of it. This image shows the bumper and the
location of the radar sensor mounted behind the bumper. To determine the correct
location of the radar sensor, the R&S QAR compares the recorded image to the CAD
data of the radar sensor and the bumper or radome.
After the position and orientations of both radar and sensor and have been found in the
image, the R&S QAR uses the CAD model of the radar sensors emission cone to calculate the intersection area of the emission cone and the radome.
Finally, the R&S QAR calculates the shortest distance between the intersection area of
the radar emission cone and the boundaries of the allowed area.
Measuring the DUT
1. Select the configuration from the "Car Type" drop-down list.
2. Select "Options" > "Alarm Limits" to define acceptance limits.
For more information, see "Managing measurement results"on page 80.
3. Select "Measure".
The R&S QAR measures the DUT, which takes a few seconds. Do not move the
radar during the measurement.
After a few seconds, the results appear in the user interface.
Graphic results
The image shows the CAD data and, as an overlay, the measurement results.
The colors have the following meaning:
●
The green contour corresponds to the border polygon.
●
The orange contour corresponds to the theoretical radar beam intersection.
●
The red contour corresponds to the measured radar beam intersection.
●
The yellow lines correspond to the measured distances.
You can display the graphical results as a 2D screenshot or as 3D image.
●If you turn on "Save Screenhot", the image is a 2D image that shows the DUT
from the front.
●If you turn off "Save Screenhot", the image is a 3D image. You can turn the
image in any direction and view it from different angles. To do so, select the
image with the mouse pointer and move it in the direction you want.
Numeric results
Only available in advanced mode.
The numeric results show the measured distances and angles.
●
Numeric results that are within the tolerance limits are displayed in a black font.
●
Numeric results that are outside the tolerance limits are displayed in a red font.
The decision if a radar sensor is positioned correctly is based on the evaluation of eight
measurement points. For each measurement point, the R&S QAR calculates the shortest distance between the radar beam intersection on the radome and a specified border polygon (green contour in the graphic results).
The upper part of the table of the numeric result summary shows the deviation
between specified distances and measured distances for all eight measurement points.
By default, the results are shown in ascending order, from smallest to largest distance.
You can also define a different order based on the index number of the measurement
point.
1. Select "Options" > "View Positions".
2. Select "Show following view positions", and define the order of measurement
points, for example "1:2:3:4:5:6:7:8".
The lower part of the table shows the deviation of the DUT to the specified position in
absolute terms.
●
The minimum distance is the smallest distance that has been measured at one of
the measurement points.
●
The gap values indicate the deviation of the position in the corresponding direction.
For example, an x-gap of 1 mm indicates that the DUT has a wrong position on the
longitudinal axis.
●
The angle values indicate that the DUT is tilted by a certain amount in the corresponding direction. For example, an x-angle of 5 ° indicates that the DUT is tilted
slightly by its x-axis.
Overall test result
The overall test passes if all values are within the tolerances and limits.
●
In simple mode, the overall test result is indicated by a green (pass) or red (fail)
traffic light.
●
In advanced mode, all numerical values must have a black font ("Status" = "OK").
A single value in red font results in a failed test ("Status" = "Not OK").
77User Manual 1178.7200.02 ─ 12
Page 78
R&S®QAR
Operating the R&S QAR
Radar positioning measurements
5.4.4Configuration
The R&S QAR provides several tools to configure the measurement.
Selecting the user interface
The radar positioning measurement application provides two user interfaces, simple
and advanced.
► Select "Options" > "Advanced Mode" to switch between user interfaces.
●The simple user interface shows the graphical result and the overall test result
(pass or fail).
●The advanced user interface also shows the numerical results.
A checkmark in front of the "Advanced Mode" menu item indicates that
advanced mode is active.
Selecting the working mode
The system mode selects the radar positioning measurement test setup you are using.
You can either integrate the R&S QAR in a custom automated system ("Inline" mode)
or use it in a stand-alone setup.
Inline mode: The R&S QAR is part of an automated system. An adjustment of the
imaging height for the R&S QAR is not necessary. Instead, the system mechanically
adjusts the location of the R&S QAR in relation to the location of the radar sensor.
Stand-alone mode: The R&S QAR is installed on the platform (R&S QAR-Z21). For a
valid measurement, the R&S QAR has to adjust its imaging height in relation to the
location of the radar sensor. The R&S QAR takes this information from the car type
configuration.
► Select "Options" > "Inline".
A checkmark in front of the "Inline" menu item indicates that inline mode is active.
Selecting the car type
The car type configuration provides information about the car, the location of the radar
sensor and the measurement points (positions). The configuration is based on the
CAD data of the car.
In stand-alone working mode, the car type configuration also defines the imaging
height of the R&S QAR.
Because every car type has different characteristics, you have to load the corresponding configuration.
2. Select the configuration from the "Car Type" dropdown menu.
The R&S QAR loads the corresponding data and adapts its imaging volume and
height. Loading another car type takes about 3 minutes.
78User Manual 1178.7200.02 ─ 12
Page 79
R&S®QAR
Operating the R&S QAR
Radar positioning measurements
3. Optional: If you change any settings that have an effect on the car type configuration (for example limits), select "File" > "Configuration" > "Export Configuration" to
export the updated configuration.
Defining alarm limits
Alarm limits define by how much the position of the radar module may deviate from the
ideal position. If the results show a deviation greater than the limits, the test fails.
1. Select "Options" > "Alarm Limits" to open the corresponding dialog box.
Figure 5-23: Set alarm limits dialog box
2. Define the limits as necessary.
●Angle limits
Maximal difference between theoretical and measured rotation angles. If the
measured difference is higher than the maximum limit, it means that the radar
is too turned around an axis and you get a "Not OK" status (red traffic light).
●Distance limit
Minimum distance from the radar window border. If the measured distance is
lower than the minimum limit, it means that the radar beam is too close to the
border and you get a "Not OK" status (red traffic light).
●Difference limit for the verification set
Only applicable when using the verification sets. Maximal difference between
theoretical and measured position. If the measured difference is higher than the
maximum limit, you get a "Not OK" status (red traffic light).
If verification fails, check the position of all devices (R&S QAR, verification kit
and verification plate). Also check if you have selected the correct verification
plate from the "Car Type" dropdown menu.
If verification still fails, contact Rohde & Schwarz. Stop any further measurements.
79User Manual 1178.7200.02 ─ 12
Page 80
R&S®QAR
Operating the R&S QAR
Radar positioning measurements
Selecting the user interface language
You can display the user interface in various languages.
► Select "Options" > "Languages".
Changing the language requires a restart of the application.
Managing measurement results
You can export the measurement data, for example to store it on an external memory
device or copy it to a different computer for further analysis with different software. You
can also restore data that you have taken at some point and review it in the application.
1. Select "Options" > "Save Settings".
2. Define the default directory in which to save the results in the "Result Folder" input
field.
For each measurement, the R&S QAR creates a folder that contains the measurement data.
3. Select "Autosave After Measurement" to use the auto save functionality.
When you are using the autosave feature, the R&S QAR automatically stores the
results after each measurement.
Tip: You can always save the measurement results deliberately with the keyboard
shortcut Ctrl-S or from the "File" menu.
4. Select "Save Screenshot" to store a screenshot of the graphical result.
5. Select the data you want to save from the "Save Option" dropdown menu.
A measurement dataset contains the following files.
●"Result"
Stores the numeric results in a .txt file.
●"Result and Volume"
Contains the numeric results and the raw measurement data (3D view) in
a .mat file.
Tip: You can restore measurement results at any time with the keyboard shortcut
Ctrl-O or from the "File" menu.
If you restore 3D volume data ("Result and Volume"), the results show:
●A 3D view in the graphic result display.
●All numeric results.
If you restore the numeric results ("Result") and a screenshot ("Save Screenshot"),
the results show:
●A screenshot of the results in the graphic result display.
●All numeric results.
If you restore the results only ("Result") without a screenshot, the results show:
●Nothing in the graphic result display.
●All numeric results.
80User Manual 1178.7200.02 ─ 12
Page 81
R&S®QAR
Operating the R&S QAR
Radar positioning measurements
Remote controlling radar positioning measurements
You can remote control radar positioning measurements via a Modbus protocol.
1. Select "Options" > "Modbus Communication".
Several new input fields and control buttons appear in the user interface.
2. Define a local or remote IP address.
3. Select "Connect" to connect with the remote computer.
Using a barcode scanner
You can connect a barcode scanner to the R&S QAR to scan the barcode of a DUT
and use this as a reference that the DUT with a certain barcode has passed or failed
the test. The barcode is displayed in the user interface and is stored together with the
other measurement data.
The R&S QAR supports the barcode scanner Honeywell Xenon 1900 by Honeywell
International Inc.
1. Select "Options" > "Recording".
2. Select "Use Barcode Scanner".
3. Select the (serial) port the barcode scanner is connected to from the dropdown
menu.
5.4.5Verification
The R&S QAR provides a verification procedure to validate the quality of the measurements. Verification consists of a series of measurements with standardized DUTs
which are located at a fix distance from the R&S QAR.
All verification objects are delivered with a quality approved surface finish. Any surface
features are either intentional or are considered to be uncritical for the intended purpose.
If verification fails, validate the verification setup. If verification fails several times in a
row, despite a valid setup, contact Rohde & Schwarz customer support to calibrate the
R&S QAR and check the hardware.
Required equipment
All required equipment is included in the delivery of the R&SQAR-Z42.
●
3 Verification objects with different characteristics
●
Mounting frame for verification objects
81User Manual 1178.7200.02 ─ 12
Page 82
R&S®QAR
Operating the R&S QAR
Radar positioning measurements
Installing the mounting frame for the verification objects
The verification set includes a mounting frame that holds the verifications objects. For
verification, you have to place the mounting frame in front of the R&S QAR at a defined
location. If the distances deviate from the ones mentioned here, verification is invalid.
●
The distance between the R&S QAR and the front edge of the mounting frame is
65 cm.
Figure 5-24: Verification setup (side view)
●
The verification object has an offset of 14.5 cm to the right of the center of the
R&S QAR.
Figure 5-25: Verification setup (view from above)
Verifying the system
The verification is a series of three consecutive measurements with the verification
objects. The verification objects are labeled A to C.
1. Let the R&S QAR warm up for at least 90 minutes.
NOTICE! Handling the verification objects.
2.
82User Manual 1178.7200.02 ─ 12
Page 83
R&S®QAR
Operating the R&S QAR
Radar positioning measurements
●Store the verification objects in the bag they were delivered in.
●Keep the verification objects clean and free from fingerprints.
●Do not scratch the verification objects.
●Clean the objects with a lint-free duster or a damp rag. Do not use chemicals.
Insert one of the verification objects into the mounting frame.
3. Select the corresponding DUT configuration from the "Car Type" dropdown menu.
4. Start a measurement ("Measure").
5. Repeat the procedure for all three verification objects.
Verification passes if the measurement results for all three objects are within the
limits (measurement result is "OK").
6. Optional: If necessary, change the verification limits ("Options" > "Alarm Limits" >
"Difference Limit for the Verification Set").
However, we recommend to use the default limits (3 mm).
83User Manual 1178.7200.02 ─ 12
Page 84
R&S®QAR
Remote control interface and protocol
6Remote control - SCPI
●Remote control interface and protocol.................................................................... 84
For the R&S QAR, the remote control interface is the LAN interface. The LAN interface
consists of a connector, a network interface card and protocols.
The network card can be operated with the following interfaces:
●
10 Mbit/s Ethernet IEEE 802.3
●
100 Mbit/s Ethernet IEEE 802.3u
●
1Gbit/s Ethernet IEEE 802.3ab
For remote control via a network, the PC and the instrument must be connected via the
LAN interface to a common network with TCP/IP network protocol. They are connected
using a commercial RJ45 cable (shielded or unshielded twisted pair category 5). The
TCP/IP network protocol and the associated network services are preconfigured on the
instrument. Software for instrument control and the VISA program library must be
installed on the PC (controller).
Remote control - SCPI
IP address
Only the IP address or a valid DNS host name is required to set up the connection.
The host address is part of the "VISA resource string" used by the programs to identify
and control the instrument.
The VISA resource string has the form:
TCPIP::host address[::LAN device name][::INSTR]
where:
●
TCPIP designates the network protocol used
●
host address is the IP address or host name of the device
inst0 selects the VXI-11 protocol
●
LAN device name defines the protocol and the instance number of a sub-instrument;
●
INSTR indicates the instrument resource class (optional)
Example:
●
Instrument has the IP address 192.1.2.3; the valid resource string using VXI-11
protocol is:
TCPIP::192.1.2.3::INSTR
84User Manual 1178.7200.02 ─ 12
Page 85
R&S®QAR
Remote control - SCPI
Remote control interface and protocol
Identifying instruments in a network
If several instruments are connected to the network, each instrument has its own IP
address and associated resource string. The controller identifies these instruments by
the resource string.
VISA library
A VISA installation is a prerequisite for remote control using the LAN interface.
VISA is a standardized software interface library providing input and output functions to
communicate with instruments. High level programming platforms use VISA as an
intermediate abstraction layer.
The I/O channel (LAN or TCP/IP) is selected at initialization time by one of the following:
●
The channel–specific address string ("VISA resource string")
●
An appropriately defined VISA alias (short name).
For more information about VISA, refer to the VISA user documentation.
VXI-11 protocol
The VXI-11 standard is based on the open network computing remote procedure call
(ONC RPC) protocol which in turn relies on TCP/IP at the network/transport layer. The
TCP/IP network protocol and the associated network services are preconfigured.
TCP/IP ensures connection-oriented communication, where the order of the
exchanged messages is adhered to and interrupted links are identified. With this protocol, messages cannot be lost.
SCPI compatibility
SCPI commands (standard commands for programmable instruments) are used for
remote control. The SCPI standard is based on standard IEEE 488.2 and aims at the
standardization of instrument-specific commands, error handling and the status registers. The tutorial "Automatic Measurement Control - A tutorial on SCPI and IEEE
488.2" from John M. Pieper (R&S order number 0002.3536.00) offers detailed informa-
tion on concepts and definitions of SCPI. The instrument supports the SCPI version
1999.
Commands that are not taken from the SCPI standard follow the SCPI syntax rules.
Communication ports
You have to open several ports in the firewall to be able to communicate via SCPI. By
default, these ports are already open on the R&S QAR. If you want to close these
ports, you can run a script that closes these ports.
●
To close the ports, run the
script .\firewall\firewall-disableaccess.bat.
Closing the ports disables remote control functionality.
●
To reopen the ports, run the script .\firewall\firewall-setting.bat.