Rohde & Schwarz PR100 User Manual

R&S
®
PR100 Portable Receiver
Manual
© 2008 Rohde & Schwarz GmbH & Co. KG
81671 Munich, Germany
Printed in Germany – Subject to change – Data without tolerance limits is not binding.
R&S
®
is a registered trademark of Rohde & Schwarz GmbH & Co. KG.
Trade names are trademarks of the owners.
The following abbreviations are used throughout this manual:
R&S
®
PR100 is abbreviated as R&S PR100.
Index:
Index:................................................................................................................................................ 3
1 Performance Check .................................................................................................................. 8
1.1 Symbols and safety labels.................................................................................... 9
1.2 Tags and their meaning........................................................................................ 9
1.3 Basic safety instructions .................................................................................... 10
2 Quality certificate................................................................................................................... 12
3 EC certificate.......................................................................................................................... 13
4 Support center address ...........................................................................................................14
5 Operating principle................................................................................................................. 15
6 Set-up ..................................................................................................................................... 20
6.1 Front view.......................................................................................................... 20
6.2 Top view ............................................................................................................ 21
6.3 Unpacking the instrument.................................................................................. 21
6.4 Setting up the instrument ................................................................................... 22
6.5 Inserting the battery ........................................................................................... 23
6.6 Connecting to the power supply ........................................................................ 23
6.7 Charging the Battery.......................................................................................... 24
6.8 Switching on the monitoring receiver................................................................ 25
6.9 Ambient and operating conditions..................................................................... 26
6.10 Preventive maintenance ..................................................................................... 26
6.11 Monitoring receiver connectors......................................................................... 26
6.12 Software update ................................................................................................. 33
6.12.1 Option code activation....................................................................................... 34
7 Configuration Menus .............................................................................................................35
7.1 RX Configuration Menu .................................................................................... 35
7.1.1 General............................................................................................................... 35
7.1.2 Antenna (only with option Field Strength Measurement) ................................. 39
7.1.3 Tone ................................................................................................................... 41
7.1.4 Measure.............................................................................................................. 42
7.1.5 BFO (CW only) .................................................................................................44
7.1.6 Direct Conversion Threshold............................................................................. 45
7.1.7 Inputs / Outputs.................................................................................................. 45
7.2 Scan Configuration Menu.................................................................................. 46
7.2.1 Frequency Scan.................................................................................................. 46
7.3 Memory Scan..................................................................................................... 48
7.3.1 Scan Options...................................................................................................... 50
7.4 Display............................................................................................................... 53
7.4.1 RX Screen.......................................................................................................... 53
7.4.2 IF-PAN Screen................................................................................................... 54
7.4.3 RF-PAN Screen ................................................................................................. 56
7.4.4 Waterfall Screen ................................................................................................ 57
7.5 General Configuration Menu............................................................................. 60
7.5.1 General............................................................................................................... 60
7.5.2 Local Settings .................................................................................................... 61
7.5.3 Display............................................................................................................... 62
7.5.4 Keys ................................................................................................................... 62
7.5.5 Audio ................................................................................................................. 64
7.5.6 LAN ................................................................................................................... 66
7.6 Memory Configuration Menu............................................................................ 69
7.6.1 Direct Save & Auto Save................................................................................... 69
7.7 Antenna Configuration Menu ............................................................................ 71
8 SCPI Interface ........................................................................................................................ 72
8.1 Document Outline.............................................................................................. 72
8.2 Legend ............................................................................................................... 72
Abbreviations Used................................................................................................................ 72
9 SCPI Commands .................................................................................................................... 73
9.1 SCPI introduction .............................................................................................. 73
9.1.1 Common Command Structure ........................................................................... 74
9.1.2 Device-Specific Command Structure ................................................................ 74
9.1.3 Structure of a command line.............................................................................. 75
9.1.4 Responses to queries.......................................................................................... 76
9.1.5 Parameters.......................................................................................................... 77
Syntax Elements..................................................................................................................... 79
9.2 Status Reporting................................................................................................. 79
Structure of an SCPI status register .......................................................................................80
9.2.1 Description of the status registers...................................................................... 83
9.2.2 Use of the Status Reporting System ..................................................................91
9.2.3 Resetting values of the status reporting system................................................. 93
9.3 Error Messages .................................................................................................. 93
9.4 Commands Description ..................................................................................... 98
9.4.1 Notation ............................................................................................................. 98
9.4.2 Common Commands ......................................................................................... 99
10 Instrument Behaviour ...................................................................................... 101
10.1 Error Situations................................................................................................ 102
10.2 Ranging and Rounding .................................................................................... 102
10.3 Value Representation....................................................................................... 103
10.4 Default Values ................................................................................................. 103
10.5 Instrument States ............................................................................................. 103
10.5.1 Introduction...................................................................................................... 103
10.5.2 MR States......................................................................................................... 103
11 Commands Reference...................................................................................... 106
11.1 Common Commands ....................................................................................... 106
11.2 ABORt subsystem ........................................................................................... 107
11.3 CALCulate subsystem ..................................................................................... 107
11.4 DIAGnostic subsystem .................................................................................... 112
11.5 DISPlay subsystem .......................................................................................... 113
11.6 FORMat subsystem ......................................................................................... 132
11.7 INITiate subsystem.......................................................................................... 135
11.8 INPut subsystem .............................................................................................. 137
11.9 MEASure subsystem ....................................................................................... 138
11.10 MEMory subsystem......................................................................................... 140
11.10.1 Memory list subsystem .................................................................................... 147
11.10.2 Memory save subsystem.................................................................................. 149
11.11 MMEMory subsystem ..................................................................................... 152
11.12 OUTPut subsystem .......................................................................................... 160
11.13 Program preset subsystem ...............................................................................173
11.14 SENSe Subsystem ........................................................................................... 176
11.14.1 Sense Memory Scan subsystem MSC .............................................................201
11.14.2 Sense Panorama Scan subsystem PSC ............................................................ 211
Sense Frequency Scan subsystem SWE............................................................................... 217
11.15 STATus subsystem .......................................................................................... 224
11.16 SYSTem subsystem ......................................................................................... 229
11.17 TRACe|DATA subsystem ............................................................................... 246
11.18 TRACe|DATA:UDP subsystem ...................................................................... 259
12 UDP Data Streams........................................................................................... 266
12.1 Stream Packet Structure................................................................................... 266
12.2 Audio ............................................................................................................... 270
12.3 FScan ............................................................................................................... 272
12.4 MScan .............................................................................................................. 274
12.5 CW................................................................................................................... 275
12.6 IFPan................................................................................................................ 276
12.7 IF...................................................................................................................... 277
12.8 PSCAN ............................................................................................................ 279
13 Default Values ................................................................................................. 282
CALCulation subsystem .......................................................................................................... 282
13.1 DISPlay subsystem .......................................................................................... 282
13.2 FORMat subsystem ......................................................................................... 283
13.3 INPut subsystem .............................................................................................. 283
13.4 MEASurement subsystem ............................................................................... 283
13.5 MEMory subsystem......................................................................................... 283
13.6 OUTPut subsystem .......................................................................................... 283
13.7 SENSe subsystem ............................................................................................ 283
13.8 STATus subsystem .......................................................................................... 284
13.9 SYSTem subsystem ......................................................................................... 284
13.10 TRACe subsystem ........................................................................................... 284
R&S PR100 User Manual
1 Performance Check
Before using the product for the first time, please read the following:
Safety Instructions
Rohde & Schwarz makes every effort to maintain the most stringent safety standards as regards its products and to guarantee its customers the highest possible level of safety. Our products and the necessary auxiliary equipment are designed and tested in accordance with the relevant safety standards. Compliance with these standards is continuously monitored by our quality assurance system. This product has been designed and tested in accordance with the EC Certificate of Conformity and has left the manufacturers' plant in a condition that complies fully with safety standards. To maintain this condition and to ensure safe operation, please take note of all the instructions and warnings appearing in this manual. Should you have any questions regarding these safety instructions Rohde&Schwarz will be happy to answer them. Furthermore, you are responsible for using the product in an appropriate manner. This product is designed solely for use in industrial and laboratory environments, or in the field, and must not be used in any way that may cause personal injury or damage to property. The user bears responsibility if the product is used for any purpose other than that for which it was designed or if the manufacturer's instructions are disregarded. The manufacturer accepts no liability for misuse of the product. The product is considered as being used for its designated purpose where it is used in accordance with the relevant operating manual and within its performance limits (see data sheet, documentation, the following safety instructions). Using the products requires technical skills and a knowledge of English. It is therefore essential for the products to be used only by skilled and specialized staff or thoroughly trained personnel with the required skills. Should personal protective equipment be necessary for using R&S products, this will be indicated at the appropriate place in the product documentation.
R&S PR100 User Manual
1.1 Symbols and safety labels
Compliance with safety instructions will help prevent personal injury or damage caused by hazards of any kind. It is therefore essential to carefully read and comply with the following safety instructions before commissioning the product. It is also absolutely vital to comply with additional safety instructions relating to personal safety which appear in other sections of the documentation. In these safety instructions, the word "product" refers to all goods sold and distributed by Rohde&Schwarz, including all instruments, systems and accessories.
1.2 Tags and their meaning
DANGER Indicates a hazard area that carries a high risk of danger for users. The hazard area can
cause death or serious injuries.
WARNING Indicates a hazard area that carries a medium risk of danger for users. The hazazrd
area can cause death or serious injuries.
CAUTION Indicates a hazard area that carries a slight risk of danger for users. The hazard area
can cause minor injuries.
ATTENTION This tag indicates the possibility of incorrect use which may cause damage to the
product.
NOTE This tag indicates a situation where the user should take special care when operating
the product but which will not damage the product.
R&S PR100 User Manual
1.3 Basic safety instructions
1. The product should be operated only under the operating conditions and in the situations specified by the manufacturer; its ventilation must not be obstructed during operation. Unless otherwise specified, the following requirements apply to R&S products: IP protection 2X, pollution level 2, excess voltage category 2, for indoor use only, maximum operating altitude 2000 meters above sea level
2. Applicable local or national safety regulations and accident prevention rules must be observed when performing any operations. The product should only be opened by authorized, specially trained personnel. Prior to carrying out any work on the product or opening the product, it must be disconnected from the mains power supply. Any adjustments, replacement of parts, maintenance or repairs must only be carried out by specialist electricians authorized by R&S. Original parts only should be used to replace safety-related components (e.g. power switches, power transformers or fuses). A safety test must always be carried out after safety-related components have been replaced (visual inspection, ground/earth test, insulation resistance and leakage current measurements, function test).
3. As with all manufactured goods, the use of substances that may cause an allergic reaction (allergens), such as aluminum, cannot be ruled out. Should you develop an allergic reaction (such as a skin rash, frequent sneezing, red eyes or breathing difficulties) when using R&S products, consult a doctor immediately to determine the cause.
4. For certain functions, some products, such as HF radio equipment, can produce a high level of electromagnetic radiation. Given that unborn children require increased protection, appropriate measures should be taken to protect pregnant women. People with pacemakers may also be harmed by electromagnetic radiation. Employers are required to assess workplaces where there is a specific risk of exposure to radiation and, where necessary, take measures to avert the danger.
5. Special training and a high level of concentration is needed to operate the products. Disabled people should only use the products if it is certain that there will be no impairment due to the nature of their disability when operating the products.
6. Before switching on the product, ensure that the rated voltage setting on the product matches the rated voltage of the mains supply. The product's mains fuse should also be changed if it is necessary to alter the voltage setting.
7. In the case of safety class I products with a flexible power cord and connector, operation is only permitted using sockets with an earth/ground contact and a protective earth/ground connection.
8. Intentionally breaking the protective earth/ground connection, either in the feed line or in the product itself, is not permitted; doing so may result in an electric shock hazard from the product. Where extension cables or connector strips are used, they must be checked on a regular basis to ensure that they are safe to use.
9. If the product is not equipped with a power switch for disconnection from the mains supply, the plug on the connecting cable is to be regarded as the disconnecting device. In such cases, you must ensure that the mains plug can be easily reached and is accessible at all times (length of the connecting cable approx. 2m). Function or electronic switches are not suitable for disconnecting the mains supply. If products
R&S PR100 User Manual
without a mains switch are integrated into racks or systems, a disconnecting device must be provided at the system level.
10. Never use the product if the power cable is damaged. Take appropriate safety measures and carefully lay the power cable to ensure that the latter cannot be damaged and that no one can be hurt, for example by tripping over the cable or receiving an electric shock.
11. The product may be operated only from TN/TT mains power networks with a maximum 16A fuse.
12. Do not insert the plug into sockets that are dusty or dirty. Insert the plug firmly and all the way into the socket, otherwise there is a risk of sparks, fire and/or injury.
13. Do not overload any sockets, extension cables or connector strips; doing so may cause fire or electric shocks.
14. For circuit measurements with Vrms voltages above 30V, suitable measures (e.g. appropriate measuring equipment, fuses, current limiting, electrical separation, insulation, etc.) should be taken to avoid any hazards.
15. Ensure that any connections with computer equipment comply with IEC950/EN60950.
16. Never remove the cover or part of the housing while you are operating the product. This will expose circuits and components and may cause injury, fire or damage to the product.
17. If a product is to be permanently installed, the earth/ground connection on site and the product's earth/ground conductor must be connected before any other connection is made. The product must only be installed and connected by a specialist electrician.
18. For permanently installed equipment without built-in fuses, circuit breakers or similar protective devices, the mains circuit must be fuse-protected in such a way that users and products are sufficiently protected.
19. Do not insert any objects which are not designed for this purpose into the openings on the housing. Never pour any liquids onto or into the housing. This may cause a short circuit inside the product and/or electric shocks, fire or injury.
20. Take appropriate measures to protect against excess voltage caused by adverse weather conditions (e.g. thunderstorms) reaching the product, otherwise, operating personnel will be exposed to the risk of electric shocks.
21. R&S products are not protected against water penetration unless otherwise specified (see Point 1). If this is not observed there is a risk of electric shocks or damage to the product, which may also result in personal injury.
22. Never use the product in conditions where condensation has formed or may form in or on the product, for example when the product is moved from a cold to a warm environment.
23. Do not obstruct any slots or openings on the product, since these are necessary for ventilation and prevent the product from overheating. Do not place the product on
R&S PR100 User Manual
surfaces that are not rigid, such as sofas or carpets, or inside a closed housing, unless this is well ventilated.
24. Do not place the product on equipment that generates heat, such as a radiator or fan heater. The ambient temperature must not exceed the maximum temperature specified in the data sheet.
25. Batteries and storage batteries must not be exposed to high temperatures or fire. Store batteries and storage batteries out of the reach of children. If batteries or storage batteries are not replaced appropriately there is a risk of explosion (warning: lithium cells). Batteries and storage batteries must only be replaced with the corresponding R&S type batteries (see spare parts list). Batteries and storage batteries are classed as hazardous waste. Dispose of them only in specially marked containers. Comply with local regulations concerning waste disposal. Do not short­circuit batteries or storage batteries.
26. Please be aware that in the event of a fire, the product may emit toxic gases that can be harmful to your health.
27. Be aware of the weight of the product. Move the product carefully, as its weight may cause back or other physical injuries.
28. Do not place the product on surfaces, vehicles, cabinets or tables whose weight and stability make them unsuitable for this product. Always follow the manufacturer's installation instructions when installing the product and attaching it to objects or structures (e.g. walls and shelves).
29. Should you decide to use the product inside a vehicle, it is the sole responsibility of the driver to drive the vehicle safely. Secure the product properly inside the vehicle to prevent injury or damage in the event of an accident. Never use the product in a moving vehicle if doing so may distract the driver of the vehicle. The driver is always responsible for the safety of the vehicle; the manufacturer assumes no responsibility for accidents or collisions.
30. If a laser product (e.g. a CD/DVD drive) is integrated into an R&S product, do not use any settings or functions other than those described in the documentation. This may otherwise be hazardous to your health, since lasers may cause irreversible optical damage. Never attempt to take such products apart. Never look directly into the laser beam.
2 Quality certificate
Dear Customer,
Thank you for purchasing a Rohde & Schwarz product.
This product is manufactured using state-of-the-art production methods. It is developed, produced and tested in accordance with the rules of our Quality Management System. The Rohde & Schwarz Quality Management System is ISO 9001 certified. Certified Quality System ISO 9001 DQS REG. NO 1954-04
R&S PR100 User Manual
3 EC certificate
R&S PR100 User Manual
4 Support center address
Should you have any questions regarding this Rohde & Schwarz instrument, please call our Support center hotline at Rohde & Schwarz Vertriebs-GmbH. Our team will be happy to answer your questions and work with you to find a solution. The hotline is open Monday to Friday between 8 a.m. and 5 p.m. Should you wish to contact us outside normal business hours, please leave a voice message or send us a fax or email. We will contact you as soon as possible. If you would like to receive information on modifications and updates for a specific instrument, please send us a short email stating which instrument. We will ensure that you regularly receive the latest information.
Support center Tel: +49 180 512 42 42 Fax: +49 89 41 29 137 77 Email: CustomerSupport@rohde-schwarz.com
R&S PR100 User Manual
5 Operating principle
Frontend
Starting from the antenna socket, the frequency in the signal path is limited to 8 GHz. Signal processing then takes place in three paths for three different frequency ranges. Signals from 9 kHz to 30 MHz are routed via a preamplifier directly to the A/D converter. Signals from 20 MHz to 3.5 GHz are taken to the IF section via a preselection and a preamplifier, or via an attenuator in the case of high signal levels. The preselection as well as the attenuator effectively protect the IF section against overloading. This is particularly important in this frequency range, where the maximum signal sum levels occur. Signals from 3.5 GHz to 8 GHz are taken to the IF section via a preamplifier. The three-stage IF section processes the signals from 20 MHz to 8 GHz for the subsequent A/D converter. To provide optimum instrument performance, only signals up to 7.5 GHz are processed in the subsequent stages. The uncontrolled 21.4 MHz IF can also be tapped ahead of the A/D converter via a BNC socket of the R&S®PR100 for further external processing.
Digital signal processing
After A/D conversion of the signal, the signal path is split up: In the first path, the IF spectrum is calculated by means of a digital downconverter (DDC), a digital bandpass filter and an FFT stage. The bandwidth of the bandpass filter can be selected between 10 kHz and 10 MHz. Before the IF spectrum is output on the display or via the LAN interface, results are postprocessed by means of the AVERAGE, MIN HOLD or MAX HOLD function as selected by the user. In the second path, the signal is processed for level measurement or demodulation. Here, too, the signal is taken via a DDC and a bandpass filter. To process the different signals with optimum signal-to-noise ratio, the receiver contains IF filters with demodulation bandwidths from 150 Hz to 500 kHz, which can be selected independently of the IF bandwidth. Prior to the level measurement, the absolute value of the level is determined and weighted by means of the AVERAGE, MAX PEAK, RMS or SAMPLE function, as selected by the user. The measured level is then output on the display or via the LAN interface. For the demodulation of analog signals, the complex baseband data is subjected to automatic gain control (AGC) or manual gain control (MGC) after the bandpass filter. It is then applied to the AM, FM, USB, LSB, ISB, PULSE or CW demodulation stage. The complex baseband data (I/Q data) of digital signals is directly output for further processing after the AGC/MGC stage.
The results obtained are available as digital data and can be output via the LAN interface as required for the particular task. Digital audio data are reconverted to analog signals for output via the loudspeaker.
R&S PR100 User Manual
High receiver sensitivity, high signal resolution
The R&S® PR100 features an IF bandwidth of up to 10 MHz. This allows even very short signal pulses to be captured since the receiver displays the large bandwidth of 10 MHz in a single spectrum about the set center frequency without any scanning being required. The widest IF bandwidth of 10 MHz yields the widest spectral display; the narrowest IF bandwidth of 10 kHz yields maximum sensitivity. The IF spectrum is digitally calculated by means of a Fast Fourier Transform (FFT). The use of FFT computation at the IF offers a major advantage: The receiver sensitivity and signal resolution are clearly superior to those of a conventional analog receiver at the same spectral display width.
IF spectrum
FFT calculation of the IF spectrum is performed in a number of steps. These are described below in simplified form for an IF bandwidth of 10 kHz (BWIF spectrum = 10 kHz), which yields maximum sensitivity. Due to the finite edge steepness of the IF filter, the sampling rate fs must be larger than the selected IF bandwidth BWIF spectrum. The quotient of the sampling rate and the IF bandwidth is thus a value >1 and is a measure of the edge steepness of the IF filter. This relationship is expressed by the following two formulas:
const
B
f
IFspectrum
s
=
or
constBWf
IFspectrums
*=
The value of the constant is dependent on the selected IF bandwidth, i.e. it may vary as a function of the IF bandwidth. For an IF bandwidth of BWIF spectrum = 10 kHz, the constant has a value of 1.28. To display a 10 kHz IF spectrum, therefore, a sampling rate of fs = 12.8 kHz is required.
R&S PR100 User Manual
The R&S®PR100 uses an FFT length N of 2048 points to generate the IF spectrum. To calculate these points, the 12.8 kHz sampling band in the above example is divided into 2048 equidistant frequency slices, which are also referred to as “bins” (see figure "Signal processing for IF spectrum"). The bandwidth BWbin of the frequency slices is obtained as follows:
Hz
kHzf
BW
s
bin
25.6
2048
8.12
2048
===
This means that in the above example only the calculated bandwidth of 6.25 Hz for each bin has to be taken into account as the noise bandwidth in the calculation of the displayed average noise level (DANL) in accordance with the formula below (the effect of the window function (Blackman window) of the FFT is not considered here for simplicity's sake):
)/log(*10174 HzBWNFdBmDANL
bin
++=
The quantity NF represents the overall noise figure of the receiver. The above example shows that, due to the use of the FFT, the actual resolution bandwidth (RBW) to be taken into account in DANL calculation is clearly smaller (i.e. BWbin) than would be expected for the wide display range of 10 kHz. Another advantage of the high spectral resolution used in the FFT calculation is that signals located close together (e.g. f1, f2, f3) can be captured and represented in the IF spectrum as discrete signals (see figure "Signal display in IF spectrum"). If, on an analog receiver, a resolution bandwidth equal to the set IF bandwidth were selected (RBW = BWIF spectrum), a sum signal fsum would be displayed instead of the three discrete signals f1, f2 and f3.
R&S PR100 User Manual
Actual sampling bandwidth compared with selected IF bandwidth
Signal resolution in IF spectrum with digital and analog receiver concept
Panorama scan
The receiver's maximum FFT bandwidth of 10 MHz makes it possible to perform extremely fast scans across a wide frequency range (panorama scan). For this purpose, frequency windows of max. 10 MHz width are linked in succession, and thus the complete, predefined scan range is traversed (see figure "Signal processing in panorama scan mode"). Same as with the IF spectrum, an FFT is used to process the broad window with a finer resolution. The width of the frequency window and the FFT length (number of FFT points) are variable and are selected by the receiver. In the panorama scan mode, the user can select among 12 resolution bandwidths from 125 Hz to 100 kHz. The resolution bandwidth corresponds to the width of the frequency slices (bin width) mentioned under "IF spectrum" above. Based on the selected bin width and start and stop frequency, the R&S®PR100 automatically determines the required FFT length and the width of the frequency window for each scan step. The receiver selects these internal parameters so that the optimum scan speed is achieved for each resolution bandwidth (see figure "Resolution in panorama scan mode").
R&S PR100 User Manual
Basic sequence of steps in fast panorama scan mode
Selection of resolution for panorama scan by varying the bin width.
Selection of 12.5 kHz bin width to capture a radio service using 12.5 kHz channel spacing
R&S PR100 User Manual
6 Set-up
6.1 Front view
14
13
9
7
8
10
11
12
5
4
1
2
3
6
7 On/off switch 1
Aux. / Ext Ref. / IF
interfaces
8 Input keys 9 Unit keys 2
LAN and USB
interface
10 Cursor keys 3 Softkeys 11 Key lock 4 Function keys 12 Rotary knob 5 Function keys 13
Memory access
keys
6 (Alpha-)numeric
keypad
14 SD Card slot
R&S PR100 User Manual
6.2 Top view
The following section describes how to set up the instrument and how to connect external devices including the charger. It then describes typical uses by means of screenshots.
6.3 Unpacking the instrument
The R&S PR100 is supplied in form-fit packaging consisting of an upper and lower shell. The two shells are held together by a sleeve which encloses the packaging.
The packaging contains all accompanying accessories. To unpack the instrument, remove the sleeve. Remove the R&S PR100 and the accessories. Remove the protective film from the screen.
Headphone connector
Antenna connector
Function keys
Key lock
Rotary knob
Function control and display elements
Volume control
AUX connector
R&S PR100 User Manual
6.4 Setting up the instrument
The R&S PR100 portable monitoring receiver is designed for stationary, in-vehicle and in particular for portable use.
Depending on operating conditions, the device can be set up perfectly for both operation and the viewing angle of the display. When used as a desktop instrument, the R&S PR100 can either lie flat or stand up using the folding stand on the back.
For portable use, it is best to attach the receiver to the chest carrying strap. All the control buttons are then easily accessible and the display can be easily read.
Depending on operating conditions, the device can be set up perfectly for both operation and the viewing angle of the display. When used as a desktop instrument, the R&S PR100 can either lie flat or stand up using the folding stand on the back.
R&S PR100 User Manual
6.5 Inserting the battery
The R&S PR100 is fitted with a lithium ion battery. The HA-Z206 battery pack has a charging capacity of 6.75 Ah. The battery is inserted into the bottom right of the instrument. The cover must first be pulled downwards to unlock it and then folded upwards to open it.
The battery is NOT fitted in the R&S PR100 on delivery and must therefore be fitted before the device can be used for the first time.
6.6 Connecting to the power supply
The R&S PR100 can be powered using the mains power adaptor or the internal battery supplied. When fully charged, the built-in lithium ion battery permits approx.
3.5 hours of operation. When the R&S PR100 is delivered, the battery may be completely discharged. Should you wish to use it without a mains power connection you will therefore need to charge it. Charging time is approx. 3.5 hours with the device switched off. During operation using mains power, the R&S PR100 simultaneously charges the internal battery. Insert the power adaptor plug into the POWER ADAPTOR socket on the left-hand side of the device until it clicks into place. Then connect the adaptor to the mains power socket. The adapter voltage range is 100 V to 240 V AC / 50 Hz to 60 Hz.
The PR100's DC supply range is +15 V DC +/-10%, max. 2 A
R&S PR100 User Manual
Caution!
The R&S HA-
Z201 power adaptor supplied should only be
used to operate the device or to charge the battery using
mains power.
Ensure that the mains supply voltage is compatible with
the voltage specified on the adaptor before use. Attach t
he
appropriate connector
before inserting the adaptor into the
AC power outlet.
6.7 Charging the Battery
The R&S PR100 is equipped with a lithium ion battery. The battery permits approximately 3.5 hours' operation at room temperature when it is fully charged.
Caution!
On delivery, the R&S PR100's battery is not fully
charged. The battery therefore needs to be charged
before the device can be used for the first time.
If the unit is stored for an extended period, self-discharge will reduce the battery's charge. The battery should therefore be charged before use if it is intended to be the sole power source for an extended period.
The charge status of the battery pack is shown on the instrument's display.
The battery can either be charged directly in the instrument by using the supplied adaptor or with the optional external R&S HA-Z203 battery charger. Charging takes approx. 7 hours with the device switched on.
For faster charging, switch off the instrument during charging. Charging takes approx. 3.5 hours with the device switched off or by using the external charging unit.
R&S PR100 User Manual
Insert the adaptor plug into the POWER ADAPTER socket on the left side of the instrument until it locks into place. Then connect the adaptor to the mains power supply. The adaptor voltage range is 100 V to 240 V / 50 Hz to 60 Hz.
The device's DC supply range is +15 V DC +/-10%, max. 2 A To charge the battery externally, place it in the external R&S HA-Z203 battery charger and charge it using the plug-in power adaptor.
The plug-in power adaptor is the same R&S HA-Z201 adaptor used for the receiver itself.
6.8 Switching on the monitoring receiver
To switch on the R&S PR100, press the grey button at the bottom left of the front panel.
When the R&S PR100 is switched on, the settings in use when it was last switched off are loaded.
Should you wish to start the R&S PR100 with factory settings, the LOCK key should be pressed and held for approx. 5 seconds when you switch the unit on.
R&S PR100 User Manual
6.9 Ambient and operating conditions
The R&S® PR100 will operate reliably in the following ambient and operating conditions:
Max. humidity 95 % Rated operating altitude max. 4600 m above sea level Transport altitude max. 12000 m above sea level Excess voltage category 2 Pollution level 2
6.10 Preventive maintenance
Any dirt should be removed from the R&S® PR100 with a soft damp cloth and a mild detergent.
In case of a fault the following safety-critical parts should only be replaced with original Rohde & Schwarz spare parts:
Power adaptor 1309.6100.00 Battery charger 1309.6123.00 Six-cell battery pack 1309.6149.00
6.11 Monitoring receiver connectors
The R&S PR100 has the following connectors:
RF input
Connect the RF input to the antenna using a cable with an N connector. Make sure that the input is not overloaded.
Caution!
The maximum permissible continuous power at the HF
input is +20 dBm (100 mW).
The maximum permissible DC voltage at the HF input is 0
VDC.
Headphone connector
A 3.5 mm stereo socket is provided for headphones. The connector's internal resistance is approx. 100 S.
R&S PR100 User Manual
AUX1 IN/OUT (at top)
External control signals can, for example, be fed to the receiver via the AUX1 IN/OUT connector.
AUX2 IN/OUT
Control signals for measurements triggered externally can be fed in via the AUX2 input/output connector (e.g. for coverage measurement applications).
External reference input
A 10 MHz reference signal for frequency synchronization is received via the EXT REF BNC socket. The level for the reference signal must be greater than 0 dBm.
R&S PR100 User Manual
IF output
The unregulated 21.4 MHz IF signal is transmitted via the IF OUT BNC socket.
USB interface
The instrument is equipped with a USB1.1 interface for reading data stored on the SD card.
R&S PR100 User Manual
LAN interface
The instrument is equipped with a 10/100 Base T LAN interface for rapidly reading data stored on the SD card or for operating the receiver remotely.
Mechanical hardware protection
Mechanical hardware protection for the R&S PR100 at a workstation can be provided by installing a Kensington lock in the receiver's housing.
SD memory card
The SD card for storing measurements or user settings is inserted into the upper right side of the R&S PR100.
In order to comply with electro-magnetic compatibility guidelines (R&TTE), only LAN cables shorter than 3 m may be used (see recommended accessories).
R&S PR100 User Manual
Screen settings
The R&S PR100 screen is a 6.5" VGA display. The brightness of the backlight can be adjusted between 0% and 100%. To obtain a balance between battery operating time and screen display quality, set backlighting to the minimum level necessary.
Setting the backlight and color scheme
Press the CONF key.
Press the GENERAL softkey.
Adjust the backlight strength for the screen
Use the rotary knob or cursor keys to select the setting you want and confirm by
pressing ENTER.
Setting the display color
Press the CONF key.
Press the GENERAL softkey.
R&S PR100 User Manual
Adjust the screen colors. Use the rotary knob or the cursor keys to select the setting you want and confirm by pressing ENTER.
Country-specific settings
The R&S PR100 supports multiple languages and can display text in the language of your choice. Softkey lettering is always in English. The default setting (factory setting) is also English.
Operation
Press the CONF key.
Press the GENERAL softkey.
Set the receiver's menu language Use the rotary knob or the cursor keys to select the setting you want and confirm by pressing ENTER.
Setting the date and time
The R&S PR100 has an internal clock which can, for example, provide stored data records with a date and time stamp. The date and time can be reset by the user.
R&S PR100 User Manual
Setting the date
Press the CONF key.
Press the GENERAL softkey.
Set the receiver's date.
Enter the date using the numeric keypad and then confirm by pressing ENTER.
Press the CONF key.
Press the GENERAL softkey.
Set the date format
Use the rotary knob or the cursor keys to select the setting you want and confirm by pressing ENTER.
Setting the time
Press the CONF key.
Press the GENERAL softkey.
Loading...